Simpson 260 Series 8 Instruction Manual

Page 1
Simpson 260® Series 8 Volt-Ohm-Milliammeters INSTRUCTION MANUAL
Page 2
About this Manual
Notes and Safety Information
This Operator’s Manual contains warning symbols which alert the user to check for hazardous conditions. These appear throughout this manual where appli­cable, and are defined below. To ensure the safety of operating performance of this instrument, these instructions must be adhered to.
Warning, refer to accompanying documents.
Caution, risk of electric shock.
This instrument is designed to prevent accidental shock to the operator when properly used. However, no engineering design can render safe an instrument which is used carelessly. Therefore, this manual must be read carefully and com­pletely before making any measurements. Failure to follow directions can result in serious or fatal accident.
Shock Hazard: As defined in American National Standard, C39.5, Safety Re­quirements for Electrical and Electronic Measuring and Controlling Instrumenta­tion, a shock hazard shall be considered to exist at any part involving a potential in excess of 30 volts RMS (sine wave) or 42.4 volts DC or peak and where a leakage current from that part to ground exceeds 0.5 milliampere, when mea­sured with an appropriate measuring instrument defined in Section 11.6.1 of ANSI C 39.5.
Technical Assistance
SIMPSON ELECTRIC COMPANY offers assistance Monday through Friday 7:30 am to 5:00 pm Central Time by contacting Technical Support or Customer Service at (847) 697-2260. Internet: http://www.simpsonelectric.com
Warranty and Returns
SIMPSON ELECTRIC COMPANY warrants each instrument and other ar ticles manufactured by it to be free from defects in material and workmanship under normal use and service, its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one (1) year after delivery of such instrument or other article of equipment to the original purchaser be returned intact to it, or to one of its authorized service centers, with transportation charges prepaid, and which its examination shall disclose to its satisfaction to have been thus defective; this warranty being expressly in lieu of all other warranties expressed or implied and of all other obligations or liabilities on its part, and SIMPSON ELECTRIC COMPANY neither assumes nor autho­rizes any other persons to assume for it any other liability in connection with the sales of its products.
This warranty shall not apply to any instrument or other article of equipment which shall have been repaired or altered outside the SIMPSON ELECTRIC COMPANY factory or author ized service centers, nor which has been subject to misuse, negligence or accident, incorrect wiring by others, or installation or use not in accord with instructions furnished by the manufacturer.
®
260
is a Registered Trademark of the Simpson Electric Co.
2
Page 3
NOTES
3
Page 4
Contents
1. INTRODUCTION ................................................................................. 7
1.1 General Description ........................................................................... 7
1.2 Overload Protection ............................................................................ 7
1.3 Internal Batteries ................................................................................ 7
1.4 Printed Circuit ..................................................................................... 8
1.5 Phenolic Case .................................................................................... 8
1.6 Adjust-A-Vue Handle .......................................................................... 8
1.7 Test Leads .......................................................................................... 8
1.8 Technical Data .................................................................................... 8
1.9 Definition of Accuracy....................................................................... 1 0
1.10 Safety Considerations ...................................................................... 10
1.11 Accessory AC High Voltage Probe ................................................... 1 1
1.12 Amp-clamp Model 150-2 .................................................................. 1 1
2. INST ALLATION ................................................................................. 11
2.1 Unpacking and Inspection ............................................................... 11
2.2 Warranty ...........................................................................................11
2.3 Power Source Requirements ........................................................... 12
2.4 Operating Position ............................................................................ 1 2
3. CONTR OLS, JA CKS AND INDICATORS........................................... 1 2
3.1 Front Panel Description .................................................................... 12
4. OPERATION ..................................................................................... 14
4.1 Safety Precautions ........................................................................... 14
4.2 Polarity Reversing ............................................................................ 15
4.3 Measurement of Unknown Voltage or Current ................................. 1 5
4.4 Test Leads ........................................................................................ 1 5
4.5 DC Voltage Measurement 0-250mV Range ....................................16
4.6 DC Voltage Measurement 0-1V Range ............................................16
4.7 DC Voltage Measurement 0-2.5 ~ 0-250V Range ...........................17
4.8 DC Voltage Measurement 0-500V Range .......................................17
4.9 DC Voltage Measurement 0-1000V Range ..................................... 1 8
4.10 AC Voltage Measurement 0-2.5 ~ 0-250V Range ...........................18
4.11 AC Voltage Measurement 0-500V Range ........................................ 1 9
4.12 AC Voltage Measurement 0-1000V Range .....................................20
4.13 Output Voltage Measurement ...........................................................21
4.14 Decibel Measurement (–20 to +50 dB) ........................................... 2 2
4.15 Direct Current Measurement ............................................................ 2 3
4.15.1 Direct Current Measurement 0-50µA Range ................................... 23
4.15.2 Direct Current Measurement 0-1mA through 0-500mA Range ...... 24
4.15.3 Direct Current Measurement 0-10A range ...................................... 24
4.16 Resistance Measurements ............................................................... 25
4.16.1 Measuring Resistance ...................................................................... 25
4.16.2 Resistance Measurement of Semiconductors ................................. 26
5. OPERATOR MAINTENANCE .......................................................... 27
5.1 Inspection .........................................................................................27
5.2 Battery Replacement ........................................................................ 2 7
5.3 Fuse Replacement ........................................................................... 28
5.4 Test Lead Inspection ........................................................................28
5.5 Care ..................................................................................................28
4
Page 5
NOTES
5
Page 6
NOTES
6
Page 7
1. INTRODUCTION
1.1 General Description
The Simpson Volt-Ohm-Milliammeter 260 Ser ies 8, (hereafter referred to as the 260 or as the Instrument) is a rugged, accurate, compact, easy-to-use Instrument equipped with mirrored dial to eliminate parallax. The Instrument can be used to make accurate measurements of AC and DC voltage, direct current, resistance, decibels, and Output Voltage. The Output Voltage function is used for measuring the AC component of a mixture of AC and DC voltage. This occurs primarily in amplifier circuits.
New Features in the 260-8 are:
0-25V AC/DC Range Mirrored Dial
1.2 Overload Protection
No single protective device nor even a combination of devices as found in the 260 can fully protect a multipurpose Instrument under all overload conditions. Overloads of such severity as to damage the Instrument despite all the built-in protection provided are possible when the Instrument is misused. The 260-8 has been designed to afford maximum instrument protection under overload condi­tions which might occur within the measuring capabilities of the Instrument, how­ever, the operator must exercise care to avoid severe overloads, particularly when working in high voltage circuits.
Basic overload protection is provided by a small inexpensive and easily obtain­able 3 AG type, 1 Amp, quick-blow fuse. This fuse is rated to interrupt circuit voltages up to 250 V. F or maximum protection it is important to replace the 1A fuse with the Littlefuse type 312001 only.
A supplemental high energy fuse is provided for additional protection from se­vere overloads. This fuse is rated for an interruption capacity of 10,000 Amps up to 600 V. (Littlefuse Type BLS or Bussman Type BBS). If the Instrument fails to indicate a reading, the 1 Amp or the 2 Amp fuse may be burned out. (Refer to paragraph 5.3 for fuse replacement.) A 1 Amp spare fuse is furnished with each Instrument. Fuses are located in the battery and fuse compartment.
NOTE: The 0-10A r ange is not fused. The shunt connects directly to the –10A and +10A jacks.
T wo diodes connected across the meter mo vement affords protection to the mo ve­ment by circumventing excessive current around the movement in the event of an overload.
1.3 Internal Batteries
There are two batteries in the ohmmeter circuits, a NEDA 13F size D cell that furnishes 1.5-volts for the R X 1 and R X 100 ranges and a NEDA 1604 battery that furnishes 9-volts for the R X 10,000 range. The 1.5-volt D cell is held in place with two spring clips which also serve as battery contacts. The polarity symbols for the D cell are marked near the battery contacts. The 9-volt battery is held in place with a spring clip, but contact is made with a separate connector that is polarized. Always observe correct polarity when replacing the 1.5-volt D cell.
7
Page 8
1.4 Printed Circuit
Most of the component parts are mounted on a printed circuit board which simpli­fies assembly and maintenance, thus, extending the useful life of the Instrument.
1.5 Phenolic Case
The phenolic case has heavy reinforced walls for maximum durability and pro­vides excellent protection for the circuit components. Batteries and fuses can be accessed by unscrewing a single captivated screw at the bottom of the case and sliding out the compartment cover. Rubber plug bumpers on the bottom of the case reduce sliding should the 260 be accidentally pulled by the test leads when the Instrument is on a workbench.
1.6 Adjust-A-Vue Handle
A carrying handle is attached to the Instrument case. The handle ma y be used to support the Instrument in a convenient sloping position for easy viewing. The case can also be placed in either a vertical or horizontal position. The horizontal position provides greater accuracy since the Instrument is calibrated in this posi­tion.
1.7 Test Leads
Each Instrument is furnished with one pair (Catalog Number 00043) of probe type test leads (one red and one black, each 48 inches long) for all applications of the Instrument. These test leads have elbow prods on one end to connect the leads to the recessed jacks on the Instrument front panel. The probe tips at the opposite end have threaded shoulders to accept the screw-on insulated alligator clips (furnished with the test leads) or the 5000 volt test probe(s) to extend the Instrument AC or DC voltage range. The test leads and the insulated alligator clips are rated for the voltage and current ranges available on the 260. Maximum ratings are 1000V AC or DC or 10 DC amperes.
NOTE: These ratings for the test leads must not be exceeded in any application.
1.8 Technical Data
TABLE 1-1. Technical Data for the 260-8
1. DC VOLTS: Ranges: 0-1-2.5-10-25-50-250-500-1000V Sensitivity: 20,000 ohms per volt
2. DC MILLIVOLTS: Range: 0-250mV Sensitivity: 20,000 ohms per volt
3. AC V OLTS: Ranges: 0-2.5-10-25-50-250-500-1000V Sensitivity: 5,000 ohms per volt
4. OUTPUT VOLT AGE (AC): Ranges: 0-2.5-10-25-50-250 (limited to 350 VDC)
8
Page 9
5. DC MICROAMPERES: Range: 0-50µA Voltage Drop: 250mV
6. DC MILLIAMPERES: Ranges: 0-1-10-100-500 mA V oltage Drop (Approx.): 250 mV, 255 mV, 300 mV, 500 mV
7. DC AMPERES: Range: 0-10A Voltage Drop (Approx.): 255 mV
NOTE: The 10A range is not internally fused.
8.
ECNATSISER1XR001XR000,01XR
egnaR000,2-0 000,002-0 M02-0
gnidaeRretneC21 0021 000,021
tiucriCnepOlanimoN
egatloV
tiucriCtrohSlanimoN
tnerruC
V5.1V5.1V9
Am521Am52.1Aµ57
9. * ACCURACY: DC Voltage Ranges: 2% of Full Scale DC Current: 0-50 µA Range: 1.5% of Full Scale Other Ranges: 2% of Full Scale **AC Voltage Ranges: 3% of Full Scale Frequency Response: Referenced to 100 Hz (Figures 4-4 and 4-9) Resistance Ranges: R X 1: 2.5° of Arc R X 100: 2° of Arc R X 10,000: 2° of Arc
* Accuracies specified are for the 260 in a horizontal position. ** Responds to the average value of an AC current, and is calibrated to indicate the RMS value of a pure sine wave
10. DECIBELS: Range: –20 to +10 dB, –8 to +22 dB, 0 to +30 dB,
+6 to +36 dB, +20 to +50 dB
Reference Level: With zero dB power level equal to 1 mW across a
600 line.
11. BATTERIES: Voltage: 1.5V, 9V NEDA No.: 13F, 1604
12. FUSE: F1 1A, 250V , type 3A G, quick acting Littlefuse Type
312001, F2 2A, 600V Littlefuse Type BLS or 2A 600V, Bussman Type BBS.
13. TEST LEADS: 1 red, 1 black, 48" long.
14. SIZE: 5-1/2 x 7" x 3-1/8 (13.34 x 17.78 x 7.94 cm)
9
Page 10
15. WEIGHT: 3 lbs. (1.359 kg)
16. ***RATED-CIRCUIT-TO­GROUND VOLTAGE: 1000V AC/DC Max.
***Per ANSI C 39.5 April 1974: The specified voltage with respect to ground, which may be safely and continuously applied to the circuit of an Instrument.
1.9 Definition of Accuracy
The voltage and current accuracy of this Instrument is commonly expressed as a percent of full scale. This should not be confused with accuracy of reading (indi­cation). For example, +2% of full scale on the 10 volt range allows an error of ±0.20V at any point on the dial. This means that at full scale, the accuracy reading would be ±2%, but at half scale it would be ±4%. Therefore, it is advantageous to select a range which gives an indication as near as possible to full scale.
1.10 Safety Considerations
This Operators Manual contains cautions and warnings alerting the user to haz­ardous operating and service conditions. This information is flagged by CAU­TION or WARNING symbols throughout the publication, where applicable, and is defined on the inside front cover of the manual under NOTES AND SAFETY INFORMATION. Adhere to these instructions in order to ensure the safety of operating and servicing personnel and to retain the operating conditions of the Instrument.
TABLE 1-2. Items and Accessories Furnished with this Instrument
Quantity Description Catalog No.
1 Test Lead Set – one red and one black
((48 long) each with combination probe tip
and removable rubber-sleeved alligator clip. 00043 *1 1.5 Volt, D Cell, NEDA 13F *1 9.0 Volt Cell, NEDA 1604 2 1 Amp, 250V Fuse, Littlefuse T ype #312001 1 2A, 600V Fuse, Littlefuse Type BLS or
(Bussman Type BBS) 1 Operators Manual 6-114338
* Batteries are standard items, replaceable from retail stores.
TABLE 1-3. Additional Accessories
Description Catalog Number
Deluxe Case ......................................................................................... 00812
Leatherette Ever-Redy Hard Case, Drop Front ................................... 00805
Model 150-2 Amp-Clamp ..................................................................... 00545
Model 151-2 Line Splitter..................................................................... 00544
10
Page 11
1.11 Accessory AC High Voltage Probe
The 10,000 V AC high voltage probe extends the 260-8 AC voltage range. The probe can be used with the Simpson 260-8 10-volt AC range.
NOTE: Instructions are furnished with each high voltage probe.
1.12 Amp-clamp Model 150-2
The Simpson Amp-Clamp Model 150-2 allows AC current measurements with­out breaking the circuit under test. The Amp-Clamp works as a transformer con­taining a split core for accommodating the conductor which carries the current being measured. The current-carr ying conductor becomes the transformer pri­mary and a coil in the Amp-Clamp serves as the secondary winding. The Amp­Clamp output voltage is proportional to the current measured and can be applied to the 260-8 as an AC voltage.
The Amp-Clamp has a range selector with 6 positions. Any of the following current ranges can be used with the 260-8: 5, 10, 25, 50, 100 or 250 amperes.
NOTE: Instructions are furnished with each Amp-Clamp.
2. INSTALLATION
This section contains information and instructions for the installation and ship­ping of the Simpson 260-8. Included are unpacking and inspection procedures, warranty, shipping, power source requirements and operating position.
2.1 Unpacking and Inspection
Examine the shipping carton for signs of damage. If damaged, notify the carrier and supplier and do not attempt to use the Instrument. If Instrument appears to be in good condition, read Operators Manual in its entirety. Become familiar with the Instrument as instructed in the manual, then check the electrical performance as soon as possible. Also, check that all items listed in Table 1-2 are included with the Instrument.
After unpacking the Instrument, a 1.5V battery and a 9V battery may be found in separate envelopes in the box with the Instrument and the test leads. Two alliga­tor clips for the test leads are in a polyethylene bag. (See Section 5 for instruc­tions on how to install the batteries.)
2.2 Warranty
The Simpson Electric Company warranty policy is printed on the inside front cover of this manual. Read it carefully before requesting any warranty repairs. For all assistance, including help with the Instrument under warranty, contact the nearest Authorized Service Center. If necessary, contact the factory directly, give full details of any installation difficulty and include the Instrument model number, series number, and date of purchase. Service data or shipping instructions will be mailed promptly. If an estimate of charges for non-warranty or other service work is required, a maximum charge estimate will be quoted. This charge will not be exceeded without prior approval.
Pack the Instrument carefully and ship it prepaid and insured.
11
Page 12
2.3 Power Source Requirements
There are two batteries in the ohmmeter circuits, a NEDA 13F D size cell that furnishes 1.5 volts for the R X 1 and R X 100 ranges and a NEDA 1604 battery that furnishes 9 volts for the R X 10,000 range. The 1.5 volt D cell is held in place with two spring clips which also serve as battery contacts. The 9-volt battery is held in place with a spring clip but contact is made with a separate connector that is polarized. (Always observe correct polarity when replacing the 1.5-volt D cell.)
2.4 Operating Position
A handle is attached to the side of the Instrument case. The handle may be used to support the Instrument in a convenient, sloping position for easy viewing. The VOM case can also be placed either in a vertical or horizontal position. The horizontal position promises greater accuracy since the Instrument is calibrated in this position.
3. CONTROLS, JACKS AND INDICATORS
The functions of all the controls, jacks and indicators used to operate the Simpson 260-8 are described in this section. Become familiar with each item before oper­ating the Instrument.
3.1 Front Panel Description
Figure 3-1 depicts the front panel controls, jacks and indicators described below.
1. Front Panel: The 260-8 Volt-Ohm-Milliammeter is a large, easy-
to-read 4-1/4 indicating Instrument. Below the In­strument are four controls and eight circuit jacks.
2. Range Switch: This switch may be rotated in either direction to
select any one of the available voltage, current, or resistance ranges.
3. Function Switch: The function switch has four positions: Off, +DC, –
DC, and AC Volts Only. To measure DC current or voltage, set the function switch at the –DC or +DC position, depending on the polarity of the input sig­nal. To measure A C voltage set the function switch to the AC position. For resistance measurement, the switch may be set in either the +DC or –DC position. The polarity of the test voltage will be as marked at the jacks when the switch is in the +DC position and reversed in the –DC position. Set this switch to off when not using the meter to take mea­surements.
4. Zero Ohms: This control is used when measuring resistance to
calibrate the ohms range selected to read zero with the test leads shorted. Refer to paragraph 4.20.
12
Page 13
O
H
M
OHMS
.C.
D
A.C.
2.5 V.A.C.
1
DB
S
D
.C
.
A
.C
.
6
-10 A.
OUTPUT
350 VDC
A.C. VOLTS
ONLY
- D.C.
+ D.C.
3
COMMON
500 MA.
MAX.
2.5 V. 10 V.
1V.
25 V. 50 A.
OFF
AMPS.
250 V.
500 V. I000 V.
+
-
ALL TERMINALS 1 V MAX
100MA.
10MA. 1MA.
R X I00
R x I0,000
260
5
+10A
+1V.
50 AMPS. +
250 MV.
ZERO OHMS
AMPS.
R x 1
500 V.
1000 V.
A.C. D.C.
A.C. D.C.
4
2
Figure 3-1. Front Panel Controls, Jacks and Indicators
5. Circuit Jacks: There are eight jacks on the front panel marked
with the functions they represent. These jacks pro­vide the electrical connections to the test leads. The COMMON (–) jack is used as the reference point for the measurement of all the functions with the exception of the 10A range. (Refer to the Opera­tion Section 4 for details.)
6. P ointer Adjust For Zero: This control is used to mechanically z ero the Instru-
ment. With the function switch set to an operating position (+DC, –DC, or AC volts only), and no ap­plied input, the pointer should read zero. If it does not, use a screwdriver to turn this adjustment until it does. Once this adjustment is made, back off slightly so that the pointer rests freely over the zero mark.
13
Page 14
4. OPERATION
!
Multifunction instruments (VOMs) such as the 260-8 are intended as general purpose measuring Instruments for use in low power circuitry such as found in consumer appliances, TV and radio receivers, and in general laboratory applica­tions. Their use is not recommended in high voltage, high power circuitry where operator errors and inadequate personal protective measures could result in serious injury from arcing or explosion. Such circuitry is found in industrial or commercial applications such as induction (r-f) heaters and broadcast transmit­ters, power substations and distribution centers, x-ray equipment, large motor controls, etc. Working in the high voltage section of such equipment requires single-function, often specially designed instruments used by specially trained personnel using appropriate protective equipment and procedures.
This section of the manual contains information required to operate the 260-8 in a safe and proper manner.
4.1 Safety Precautions
The following precautions are suggestions of commonly recognized safe prac­tices to adopt and specific hazards to be avoided and are not implied to be sufficient to safeguard untrained personnel in all circumstances. Neither is this manual a substitute for technical manuals covering the equipment in which mea­surements are to be made. Always refer to the equipment manual and its specific warnings and instructions and observe them as well as those contained herein.
a. The 260-8 should only be used by personnel qualified to recognize shock
hazards and who are trained in the safety precautions required to avoid possible injury.
b. Do not connect any terminal of this Instrument to a circuit point at which a
voltage exceeding 1000 volts AC or DC may exist with respect to earth ground. (Refer to Table 1-1, item 16.)
c. Turn off power and discharge any capacitors in the circuit to be measured
before connecting to or disconnecting from it.
d. Before using this Instrument, check accessories (if any) and test leads for
missing, damaged, deteriorated or otherwise faulty insulating parts. Do not use, or permit the use of equipment with faulty insulation until it has been
properly repaired. e. Always wear safety glasses when working in electrical circuitry. f. Do not work alone on high voltage circuits. Make certain that someone
capable of giving aid is alert and located nearby. g. Do not handle the Instrument, its test leads, or the circuitry while high voltage
is being applied. h. Hands, shoes, floor, and workbench must be dry. Avoid making measure-
ments under humid, damp, or other environmental conditions that could
effect the dielectric withstanding voltage of the test leads or Instrument. i. Do not change switch settings or test lead connections while taking mea-
surements. A mistake could result in damage to the Instrument and possible
personal injury.
14
Page 15
j. Locate all voltage sources and accessible current paths before making con-
nections to circuitry. High voltage may appear unexpectedly or in unexpected
locations in faulty equipment. An open bleeder resistor, for example, may
result in a capacitor retaining a dangerous charge. k. Make certain that the equipment being worked on is properly grounded and
fuses are the proper type and rating. l. Double check switch positions and jack connections before applying power
to the Instrument. m. Remain alert for low voltage circuits which may be floating at high voltage
with respect to earth ground and for composite voltages (AC + DC) such as
are found in r-f amplifiers. The floating voltage or composite voltage must not
exceed the Instruments rated maximum circuit-to-ground voltage. n. Do not make electrical measurements where the air may contain explosive
concentrations of gas or dust such as in mines, grain elevators, gasoline
stations or in the presence of charging batteries until determined to be safe
by qualified personnel. Note that even metallic dusts can be explosive. o. No General Purpose VOM is to be used to make electrical measurements on
blasting circuits or blasting caps. Use only VOMs designed for this purpose.
4.2 Polarity Re ver sing
The function switch conveniently reverses polarity of the internal battery to facili­tate testing semiconductor devices. The +DC and –DC positions transpose inter­nal connections to the COMMON and +jacks and therefore can also be used to reverse polarity of external voltages and currents connected to these jacks (only). Because polarity reversing momentarily interrupts the circuit, it should not be done while measuring current or voltage without first turning off the circuit power. When making measurements on the 50mA/250mV, 1 volt, or 10 amp range, po­larity can be corrected only by reversing the test leads.
4.3 Measurement of Unknown Voltage or Current
The approximate voltage or current to be measured is sometimes known and the correct range can easily be selected. However, particularly when working upon faulty equipment, the voltage or current can either be unknown or higher than normal or anticipated. When working upon unfamiliar or unknown equipment, always begin a measurement using the highest range available on the Instru­ment. Once it is determined that the voltage is within the limits of a lower range, change to the lower range.
4.4 Test Leads
Test leads are provided with accessor y screw-on alligator clips which may be attached to the probe tips. Eliminating the need to hand-hold test prods to a circuit for extended periods, the test clips also reduce hand proximity to a high voltage circuit while energized. Circuit power must, of course, be turned off and any charged capacitors discharged before connecting or disconnecting clips to or from the current.
1515
Page 16
4.5 DC Voltage Measurement 0-250mV Range
!
Before making voltage measurements, re vie w the SAFETY PRECAUTIONS listed in paragraph 4.1. Also, when using the 260 as a millivoltmeter, care must be taken to prevent damage to the indicating instrument from excessive voltage. Before using the 250 millivolt range, use the 1.0-volt DC range to determine that the voltage measured is not greater than 250 millivolts (or .25 volt DC).
a. Set the function switch at +DC (Figure 4-1). b. Plug the black test lead into the – COMMON jack and the red test lead into
the +50 µA/250mV jack. c. Set the range switch at 50 µA (common position with 50V). d. Connect the black test lead to the negative side of the circuit being measured
and the red test lead to the positive side of the circuit. e. Read the voltage on the black scale marked DC and use the figures marked
0-250. Read directly in millivolts.
NOTE: For polarity reversing see paragraph 4.2.
4.6 DC Voltage Measurement 0-1V Range
a. Set the function switch at +DC (Fig-
ure 4-1). b. Plug the black test lead into the –
COMMON jack and the red test
lead into the +1V jack.
.
.C
D
D
.C
.
c. Set the range switch at 1V (com-
mon position with 10V). d. Connect the black test lead to the
negative side of the circuit being
measured and the red test lead to
the positive side of the circuit. e. Read the voltage on the black
scale marked DC and use the fig-
- D.C.
+ D.C.
COMMON
ures marked 0-10. Then divide the
reading by 10.
Figure 4-1. Jacks and Switch Positions for Measuring DC Voltages, 0 through 250V Ranges
NOTE: For polarity reversing see paragraph 4.2.
16
+10A
+1V.
50 AMPS. +
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
+
-
260
ALL TERMINALS 1 V MAX
250 MV.
Page 17
4.7 DC Voltage Measurement 0-2.5 ~ 0-250V Range
a. Set the function switch at +DC (Figure 4-1). b. Plug the black test lead into the – COMMON jack and the red test lead into
the + jack. c. Set the range switch at one of the five voltage range positions marked 2.5V,
10V, 25V, 50V or 250V. NOTE: When in doubt as to the voltage present, always use the highest voltage
range as a protection to the Instrument. If the voltage is within a lower range, the switch may be set for the lower range to obtain a more accurate reading. Be sure power is off in the circuit being measured and all capacitors discharged.
d. Connect the black test lead to the negative side of the circuit being measured
and the red test lead to the positive side of the circuit. e. Turn on the power in the circuit being measured. f. Read the voltage on the black scale marked DC. For the 2.5V range, use the
0-250 figures and divide by 100. For the 10V, 50V , and 250V r anges, read the
figures directly. For the 25V range, use the 0-250 figures and divide by 10. NOTE: Turn off power to the circuit and wait until the meter indicates zero before
disconnecting the test leads.
4.8 DC Voltage Measurement 0-500V Range
!
Use extreme care when working with high voltage circuits. Do not touch the Instrument or test leads while power is on in the circuit being measured. Before proceeding with the following steps, review the Safety Precautions in paragraph 4.1.
a. Set the function switch at +DC (Fig-
ure 4-2). b. Set the range switch at the 250V/
500V/1000V position. c. Plug the black test lead into the –
COMMON jack and the red test lead
into the 500V jack. d. Be sure power is off in the circuit be-
ing measured and all capacitors dis-
charged. Connect the black test lead
to the negative side of the circuit be-
ing measured and the red test lead
to the positive side of the circuit. e. Turn on power in circuit being mea-
sured. f. Read the voltage using the 0-50 fig-
ures on the black scale marked DC.
Multiply the reading by 10. NOTE: Tur n off power to the circuit and
wait until the meter indicates zero before disconnecting the test leads.
D.C.
- D.C.
+ D.C.
COMMON
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
500 V.
A.C. D.C.
Figure 4-2. Jacks and Switch Positions for Measuring DC Voltage, 0-500V Range
D.C.
17
Page 18
4.9 DC Voltage Measurement 0-1000V Range
!
Use extreme care when working with high voltage circuits. Do not touch the Instrument or test leads while power is on in the circuit being measured.
Before proceeding with the following steps, review the Safety Precautions in paragraph 4.1.
a. Set the function switch at +DC (Fig-
ure 4-3). b. Set the range switch at the 250V/
500V/1000V position. c. Plug the black test lead into the –
COMMON jack and the red test
lead into the 1000V jack. d. Be sure power is off in the circuit
being measured and all capacitors
discharged. Connect the black test
lead to the negative side of the cir-
cuit being measured and the red
test lead to the positive side of the
circuit. e. Turn on power in circuit being mea-
sured. f. Read the voltage using the 0-10 fig-
ures on the black scale marked DC.
Multiply the reading by 100. NOTE: Turn off power to the circuit and
wait until the meter indicates zero
before disconnecting the test leads.
.
.C
D
- D.C.
+ D.C.
COMMON
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
Figure 4-3. Jacks and Switch Positions for Measuring DC Voltage, 0-1000V Range
D
.C
1000 V.
A.C. D.C.
.
4.10 AC Voltage Measurement 0-2.5 ~ 0-250V Range
!
Branch and distribution circuits (120/240/480V etc.) can deliver dangerous ex­plosive power momentarily into a short circuit before the fuse/breaker opens the circuit. Make certain that the Instrument switches are set properly, jacks are connected properly, and that the circuit power is turned off before making con­nections to such circuits.
The 260 responds to the average value of an AC waveform. It is calibrated in terms of the RMS value of a pure sine wave. If the waveform is nonsinusoidal, and depending upon its waveform, the reading may be either higher or lower than the true RMS value of the measured voltage. Thus an error may be introduced if the 260 is used to measure a nonsinusoidal waveform. Also, accuracy is lessened at higher input frequencies (Figure 4-4).
18
Page 19
Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.1.
a. Set the function switch to AC
Volts Only position (Figure 4-
5).
NOTE: The meter will not indicate if the switch is incorrectly set to a DC position.
b. Set the range switch at one of
the five voltage range positions
marked 2.5V, 10V, 25V, 50V or
250V. (When in doubt as to ac-
tual voltage present, always
use the highest voltage range
as a protection to the Instru-
ment. If the voltage is within a
PERCENT RELATIVE ERROR
+5 +4
+3
+2
+1
0
-1
-2
-3
-4
-5
AC VOLT A GE RANGES
2.5 VAC RANGE 10 VAC RANGE 25 VAC RANGE 50 VAC RANGE
250 VAC RANGE 500 VAC RANGE
1000 VAC RANGE
lower range, the switch may be
10Hz
set for the lower range to ob­tain a more accurate reading.)
c. Plug the black test lead into the
– COMMON jack and the red
100Hz 1KHz 10KHz
FREQUENCY
100KHz
FIGURE 4-4. Frequency Response, AC Voltage Ranges
1MHz
test lead into the + jack.
d. Turn off power to the circuit to
be measured and discharge any capacitors.
e. Connect the test leads across
the circuit voltage to be mea­sured with the black lead to the grounded side.
A.C.
2.5 V.A.C.
A.C.
f. For the 2.5V range read the
value directly on the scale marked 2.5 VAC. For the 10V, 25V , 50V and 250V ranges , read the red scale marked AC and use the black figure immedi­ately above the scale.
A.C . VO LTS
ONLY
COMMON
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
+
-
260
ALL TERMINALS 1 V MAX
4.11 AC Voltage Measurement 0-500V Range
Use extreme care when working in high voltage circuits. Do not handle the Instrument or test leads while the circuit being measured is energized. OB­SERVE ALL SAFETY PRECAUTIONS in paragraph 4.1 and in the instruction manual for the equipment being tested.
Figure 4-5. Jacks and Switch Positions for Measuring AC Voltage 0-2.5 through 0-250V Ranges
19
Page 20
!
Do not attempt any voltage measurement which may exceed 1000 volts or the circuit-to-ground voltage of the Instrument, 1000 volts maximum.
Be sure that the range switch is set to the 250V/500V/1000V range, function switch to AC volts only position, and test leads connected to common and 500V jack.
Do not touch the Instrument or test leads while the power is on the circuit being measured.
Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.1.
a. Set the function switch to AC
Volts Only position (Figure 4-6).
!
.
.C
NOTE: When making AC voltage measurements with the function switch incorrectly set to DC, the Instrument pointer will remain at or near zero, in­dicating a false no voltage present condition that could be hazardous.
b. Set the range switch at 250V/
500V/1000V position.
c. Plug the black test lead into the –
COMMON jack and the red test lead in the 500V jack.
d. Be sure the power is off in the
circuit being measured and that all its capacitors have been dis­charged.
e. Connect the test leads across the
circuit voltage to be measured with the black lead to the
grounded side. f. Turn on the power in the circuit being measured. g. Read the voltage on the red scale marked AC. Use the 0-50 figures and
multiply by 10.
4.12 AC Voltage Measurement 0-1000V Range
Use extreme care when working in high voltage circuits. Do not handle the Instrument or test leads while the circuit being measured is energized. OB­SERVE ALL SAFETY PRECAUTIONS in paragraph 4.1 and in the instruction manual for the equipment being tested. Do not attempt any voltage measure­ment which may exceed 1000 volts or the circuit-to-ground voltage of the Instru­ment, 1000 volts maximum.
Be sure that the range switch is set to the 250V/500V/1000V range, function switch to AC V olts Only position, and test leads connected to common and 1000V
A
A.C . VO LTS
ONLY
COMMON
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
500 V.
A.C. D.C.
FIGURE 4-6. Jacks and Switch Positions for Measuring AC Voltage 0-2.5 Through 0-250V Ranges
5
0
A.C.
ZERO OHMS
20
Page 21
jack. Do not touch the Instrument or test leads while the power is on in the circuit being measured. Before proceeding with the fol­lowing steps, review the Safety Precautions in Paragraph 4.1. a. Set the function switch at
AC (Figure 4-7). b. Set the range switch at
250V/500V/1000V position. c. Plug the black test lead into
.
.C
A
10
A
.C
.
the -COMMON jack and the
red test lead in the 1000V
jack. d. Be sure the power is off in
the circuit being measured
and that all its capacitors
have been discharged. e. Connect the test leads
across the circuit voltage to
be measured with the black
lead to the grounded side. f. T urn on the pow er in the cir-
cuit being measured.
A.C. VOL TS
ONLY
COMMON
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
500 V.
A.C. D.C.
ZERO OHMS
g. Read the voltage on the red
scale marked AC. Use the
0-10 figures and multiply by 100.
FIGURE 4-7. Jacks and Switch Positions for Measuring AC Voltage 0 Through 1000V Range
4.13 Output V oltage Measurement
Its often necessary to measure the AC component of an Output Voltage where both AC and DC voltage levels exist. This occurs primarily in amplifier circuits. The 260-8 has a 0.1 mfd, 400 volt capacitor in series with the OUTPUT jack. The capacitor blocks the DC component of the current in the test circuit, but al­lows the AC or desired compo­nent to pass on to the indicating instrument circuit. The blocking capacitor may alter the AC re­sponse at low frequencies but is usually ignored at audio fre­quencies (Figure 4-8).
Do not connect the OUTPUT jack to a circuit in which the DC voltage component exceeds 350V.
Figure 4-8. Frequency
Response Output Ranges
+10
+8
+6
+4
+2
0
-2
-4
-6
-8
-10
10Hz 100Hz 1KHz 10KHz 100KHz 1MHz
OUTPUT RANGES
2.5 VAC RANGE 10VAC RANGE
25/50 VAC RANGE
250 VAC RANGE
FREQUENCY
21
Page 22
Before proceeding with the fol­lowing steps, review the Safety Precautions in Paragraph 4.1.
a. Set the function switch to
AC volts only position (Fig­ure 4-9).
b. Plug the black test lead into
the -COMMON jack and the
.C.
A
A.C
.
red test lead into the OUT­PUT jack.
c. Set the range switch at one
of the range positions marked 2.5V, 10V, 25V, 50V, or 250V.
d. Connect the test leads
across the circuit being mea­sured with the black test lead to the ground side.
e. T urn on the power in the test
circuit. Read the output volt­age on the appropriate AC
A.C. VOL TS
ONLY
COMMON
OUTPUT
350 VDC
MAX.
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
voltage scale. For the 0-
2.5V range, read the value directly on the scale marked
2.5 VAC. For the 10V, 25V,
Figure 4-9. Jacks and Switch Positions for Output Measurements
50V, or 250V ranges, use the red scale marked AC and read the black figures immediately above the scale.
4.14 Decibel Measurement (–20 to +50 dB)
Some industries measure in terms of voltage or current ratios (decibels) based on a specific reference level. The dB scale on the 260-8 serves this purpose and is calibrated to a reference level (zero dB) of 0.001 watt into 600 ohms. The scale calibration of –20 to +10 dB is for the 0-2.5V range (zero dB = 0.775V). Higher ranges may be used by adding an appropriate dB value to the reading in accordance with the chart on page 23.
a. Review the safety precau-
tions in paragraph 4.1.
b. Refer to Figure 4-10 for
switch settings and jack connections.
Figure 4-10. Jacks and Switch Positions for Measuring Decibels.
22
A.C. VOL TS
ONLY
COMMON
.
.C
A
2.5 V.A.C.
OUTPUT
350 VDC
500 MA.
MAX.
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V.
-
I000 V.
ALL TERMINALS 1 V MAX
A.C
.
260
Page 23
c. Use operating instructions for AC VOLTAGE MEASUREMENT, 0-2.5/10/25/
50/250V RANGES (paragraph 4.10). NOTE: Do not use the 500V or 1000V ranges for decibel readings. d. Read decibels on the bottom scale marked dB. For the 10/25/50/250V ranges,
add the appropriate factor from the chart below:
egnaRegatloVegnaRBdgnidaeR
V5.2-001+ot02-yltceriddaer V01-022+ot8-gnidaerotBd21+dda V52-003ot0gnidaerotBd02+dda V05-063+ot6+gnidaerotBd62dda
V052-005+ot02+gnidaerotBd04dda
NOTE: The maximum voltage ratio that can be measured is +50 dB on the 0­250V range. e. If dB measurements are being made to a 0.006 watt into 500 ohm reference
level, subtract +7 dB from the reading obtained on the 260.
4.15 Direct Current Measurement
!
a. Do not change the range setting of the range or function switches while the
circuit is energized.
b. Never disconnect test leads from the circuit under measurement while it is
energized.
!
c. Always turn the power off and discharge all the capacitors before the setting
of the switches is changed, or the leads disconnected.
d. Never exceed the circuit-to-ground voltage of the instrument (1000 v max.,
table 1-1, item 16). d. Always connect the instrument in series with the grounded side of the circuit. e. In all direct current measurements, make certain the power to the circuit
being tested has been turned off before connecting and disconnecting test
leads or restoring circuit continuity.
4.15.1 Direct Current Measurement 0-50µA Range
a. Set the function switch at +DC. b. Plug the black test lead into the –COMMON jack and the red test lead into the
+50µAMPS/250mV jack. c. Set the range switch at 50µAMPS (dual position with 50V). d. Open the grounded side of the circuit in which the current is being measured.
Connect the VOM in series with the circuit. Connect the red test lead to the
positive side and the black test lead to the negative side. e. Read the current on the black DC scale. Use the 0-50 figures to read directly
in microamperes.
NOTE: For polarity reversing see paragraph 4.2.
2323
Page 24
4.15.2 Direct Current Measurement 0-1mA through 0-500mA Range
a. Set the function switch at
+DC (Figure 4-11).
b. Plug the black test lead into
the – COMMON jack and the red test lead into the + jack.
c. Set the range switch at one
of the four range positions marked 1 mA, 10 mA, 100 mA or 500 mA.
d. Open the grounded side of
the circuit in which the cur­rent is being measured. Connect the VOM in series with the circuit. Connect the red test lead to the positive side and the black test lead to the negative side.
e. Tur n the power on. Read
the current in milliamperes on the black DC scale. For the 1 mA range, use the 0­10 figures and divide by 10. For the 10 mA range, use the 0-10 figures directly. For the 100 mA range, use the 0-10 figures and multi­ply by 10. For the 500 mA range, use the 0-50 figures and multiply by 10.
f. Turn the power off and dis-
connect the test leads.
.
.C
D
-10 A.
OUTPUT
350 VDC
A.C. VOL TS
- D.C.
+ D.C.
ONLY
OFF
COMMON
-
MAX.
+
500 MA.
50 A.
AMPS.
ALL TERMINALS 1 V MAX
100MA.
10MA. 1MA.
260
Figure 4-11. Jacks and Switch Positions for Measuring Direct Current
D
.C.
+10A
+1V.
50 ␮ AMPS.
+
AMPS.
250 MV.
This range is only intended for measurements in low voltage circuits (under 25 V DC) such as the primary power in vehicles or their accessories.
4.15.3 Direct Current Measurement 0-10A range
a. Plug the black test lead into the –10A jack and the red test lead into the +10A
jack.
b. Set the range switch at 10 AMPS (dual position with 10 mA). c. Open the grounded side of the circuit in which the current is being measured.
Connect the V OM in series with the circuit, connecting the red test lead to the positive side and the black test lead to the negative side.
NOTE: The function switch has no effect on polarity for the 10 AMPS range. d. Turn the power on. Read the current directly on the black DC scale. Use the
0-10 figures to read directly in amperes.
!
24
Page 25
e. Turn the power off and disconnect the test leads.
!
NOTE: The 10A range is not internally fused in the 260-8. When using the 10A range, never disconnect a test lead from a jack or from the
circuit while the circuit is energized. Doing so may damage the test jacks or leads and the arcing may be hazardous to the operator. Tu rn off circuit power and wait for the meter reading to drop to zero.
The polarity reversing feature of the function switch does not operate on the 0­10A range. If polarity is found to be incorrect, turn off circuit power, wait for the meter indication to reach zero and then interchange test lead connections to the circuit.
4.16 Resistance Measurements
Voltage applied to a resistance range will cause reading errors if low and dam­age the Instrument if high. When making in-circuit measurements, make certain that the circuit is completely de-energized before making connections to it.
When making in-circuit measurements, circuit paths in parallel with the resis­tance being measured may cause reading errors. Check circuit diagrams for the presence of such components before assuming that the reading obtained is correct.
When resistance is measured, the batteries furnish power for the circuit. Since batteries are subject to variation in voltage and internal resistance, the Instru­ment must be adjusted to zero before measuring a resistance, as follows:
a. Turn range switch to desired ohms range. b. Plug the black test lead into the –COMMON jack and the red test lead into the
+ jack.
c. Connect the ends of test leads together to short the VOM resistance circuit. d. Rotate the ZERO OHMS control until pointer indicates zero ohms. If pointer
cannot be adjusted to zero, one or both of the batteries must be replaced. (For battery replacement, refer to Section 5.)
e. Disconnect shorted test leads.
4.16.1 Measuring Resistance
a. Before measuring resistance in a circuit, make sure the power is off to the
circuit being tested and all capacitors are discharged. Disconnect shunting components from the circuit before measuring its resistance.
b. Set the range switch to one of the resistance range positions as follows
(Figure 4-12):
1. Use R X 1 for resistance readings from 0 to 200 ohms.
2. Use R X 100 for resistance readings from 200 to 20,000 ohms.
3. Use R X 10,000 for resistance readings above 20,000 ohms.
c. Set the function switch at either –DC or +DC position: Operation is the same
in either position unless there are semiconductors in the circuit. (See para­graph 4.16.2.) Adjust ZERO OHMS control for each resistance range.
25
Page 26
d. Observe the reading on the
OHMS scale at the top of the dial.
NOTE: The OHMS scale reads from right to left for increasing values of resistance.
e. To determine the actual re-
sistance value, multiply the reading by the factor at the switch position. (K on the OHMS scale equals one thousand.)
OHMS
O
H
M
S
Figure 4-12. Jacks and Switch Positions for Measuring Resistance
+ D.C.
- D.C.
COMMON
-
+
ALL TERMINALS 1 V MAX
R x 1
R X I00
R x I0,000
260
4.16.2 Resistance Measurement of Semiconductors
Make sure that the OHMS range being used will not damage any of the semicon­ductors (refer to Table 1-1, item 8, Section I).
If there is a forward and backward resistance such as in diodes, the resistance should be relatively low in one direction (for forward polarity) and higher in the opposite direction.
Rotate the function switch between the two DC positions to reverse polarity. This will determine if there is a difference between the resistance in the two directions. To check a semiconductor in or out of a circuit (forward and reverse bias resis­tance measurements) consider the following before making the measurement:
a. The polarity of the battery voltage will be as marked at the jacks when the
switch is in the +DC position and reversed in the –DC position.
b. Ensure that the range selected will not damage the semiconductor. (Refer to
Table 1-1, item 8, and review the specification limits of the semiconductor according to the manufacturers ratings.)
c. If the semiconductor is a silicon diode or conventional silicon transistor, no
precautions are normally required.
d. If the semiconductor material is germanium, check the ratings of the device
and refer to Table 1-1, item 8.
NOTE: The resistance of diodes will measure differently from one resistance range to another on the VOM with the function switch in a given position. For example, a diode which measures 80 ohms on the R X 1 range may measure 300 ohms on the R X 100 range. The difference in values is a result of the diode characteristics and is not indicative of any fault in the VOM.
26
Page 27
5. OPERATOR MAINTENANCE
The following paragraphs in this section describe battery replacement, fuse re­placement, and preventive maintenance procedures for the 260-8.
5.1 Inspection
The user is protected from electrical shock by the insulation of the 260 and its test leads. Frequently examine them for any insulation damage such as cracks, cuts, chips, burns or deterioration that expose internal metal parts or reduce the spac­ing between such metal parts and hand contact by the operator.
Make certain that the battery compartment cover is securely fastened in place before the Instrument is used.
Do not use an Instrument with a broken meter glass. If the battery compartment cover is removed for any reason, check that the proper
fuses are being used.
5.2 Battery Replacement
Figure 5-1. Battery and Fuse Compartment
a. Two batteries supply power for resistance measurements, a 1.5-volt D cell
and a 9-volt battery . When it is no longer possib le to adjust the pointer to zero for the R X 1 and R X 100 ranges (refer to ZERO OHMS ADJUSTMENT paragraph 4.16), replace the 1.5-volt cell. When it is no longer possible to adjust the pointer to zero on the R X 10,000 range, replace the 9-volt battery.
b. To install or replace a battery , de-energize and disconnect test leads from the
Instrument, then remove the cover to the externally accessible battery com­partment by loosening the single captivated screw.
27
Page 28
NOTE: Batteries should be replaced before their useful lif e has e xpired. Failure to
do so may result in corrosion and battery leakage.
c. Observe polarity when replacing the 1.5-volt D cell and connect as indicated.
The D cell is held in place with spring clips which also act as battery contact clips. The 9-v olt battery contacts and connector are polarized. To remove the 9-volt battery, first withdraw battery with mated connector from the compart­ment. Then remove the connector.
5.3 Fuse Replacement
The 1 and 2 ampere fuse (also the 1A spare) is located in the externally acces­sible battery and fuse compartment. Access compartment by disconnecting the test leads and loosening the single captivated screw on the compartment cover. To replace or check a fuse, apply pressure to the 1 amp fuse cup then rotate the plate to expose a selected fuse for removal from the panels cavity.
NOTE: When replacing fuses, it is important to use the same type and value as the ones you have removed.
5.4 T est Lead Inspection
Periodic inspection of the test leads is recommended to detect cuts, burned ar­eas, deterioration or other damage that could reduce the insulation strength of leads.
5.5 Care
a. Immediately clean all spilled materials from the Instrument and wipe dry. If
the spillage is corrosive, use a suitable cleaner to neutralize the corrosive action.
b. When the Instrument is not in use, rotate the function switch to the OFF
position.
c. Avoid prolonged exposure or usage in areas which are subject to tempera-
ture and humidity extremes, vibration or mechanical shock, dust or corrosive fumes, or strong electrical or electromagnetic interferences.
d. Verify Instrument accuracy by performing operational checks using known,
accurate, stable sources. If proper calibration equipment is not available, contact the nearest Authorized Service Center. If the Instr ument has not been used for 30 days, check the batteries for leakage and replace if neces­sary.
e. It is recommended that the Instrument be returned annually to the nearest
Authorized Service Center, or to the factory, for an overall check, adjustment, and calibration.
f. When not in use, store the Instrument in a room free from temperature ex-
tremes, dust, corrosive fumes, and mechanical vibration or shock. If storage time is expected to exceed 30 days, remove batteries.
28
Page 29
NOTES
29
Page 30
NOTES
30
Page 31
NOTES
31
Page 32
SIMPSON ELECTRIC COMPANY 853 Dundee Avenue
Elgin, IL 60120-3090 (847) 697-2260 FAX (847) 697-2272
Printed in U.S.A. Part No. 06-114338 Edition 10, 1/02
Visit us on the web at: www.simpsonelectr ic.com
32
Loading...