Simpson 260 8P User Manual

Simpson 260® Series 8P Volt-Ohm-Milliammeters INSTRUCTION MANUAL
About this Manual
To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument in compliance with its original advertised specifications.
Notes and Safety Information
This Operator’s Manual contains warning headings which alert the user to check for hazardous conditions. These appear throughout this manual where applicable, and are defined below. To ensure the safety of operating performance of this in­strument, these instructions must be adhered to.
Warning, refer to accompanying documents.
!
Caution, risk of electric shock.
!
This instrument is designed to prevent accidental shock to the operator when properly used. However, no engineering design can render safe an instrument whick is used carelessly. Therefore, this manual must be read carefully and com­pletely before making any measurements. Failure to follow directions can result in serious or fatal accident.
Shock Hazard: As defined in American National Standard, C39.5, Safety Re­quirements for Electrical and Electronic Measuring and Controlling Instrumenta­tion, a shock hazard shall be considered to exist at any part involving a potential in excess of 30 volts RMS (sine wave) or 42.4 volts DC or peak and where a leakage current from that part to ground exceeds 0.5 milliampere, when mea­sured with an appropriate measuring instrument defined in Section 11.6.1 of ANSI C 39.5.
Technical Assistance
SIMPSON ELECTRIC COMPANY offers assistance Monday through Friday 7:30 am to 5:00 pm Central Time by contacting Technical Support or Customer Service at (847) 697-2260. Internet: http://www.simpsonelectric.com
Warranty and Returns
SIMPSON ELECTRIC COMPANY warrants each instrument and other articles manufactured by it to be free from defects in material and workmanship under normal use and service, its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one (1) year after delivery of such instrument or other article of equipment to the original pur­chaser be returned intact to it, or to one of its authorized service centers, with transportation charges prepaid, and which its examination shall disclose to its satisfaction to have been thus def ectiv e; this warranty being expressly in lieu of all other warranties expressed or implied and of all other obligations or liabilities on its part, and SIMPSON ELECTRIC COMPANY neither assumes nor authorizes any other persons to assume for it any other liability in connection with the sales of its products.
This warranty shall not apply to any instrument or other article of equipment which shall have been repaired or altered outside the SIMPSON ELECTRIC COMPANY factory or authorized service centers, nor which has been subject to misuse, neg­ligence or accident, incorrect wiring by others, or installation or use not in accord with instructions furnished by the manufacturer.
2
NOTES:
3
Contents
1. INTRODUCTION ............................................................................... 6
1.1 General Description ........................................................................... 6
1.2 Overload Protection ............................................................................ 6
1.3 Internal Batteries ................................................................................ 7
1.4 Overload Circuit Test ..........................................................................7
1.5 Printed Circuit ..................................................................................... 8
1.6 Phenolic Case .................................................................................... 8
1.7 Adjust-A-Vue Handle .......................................................................... 8
1.8 Test Leads ..........................................................................................8
1.9 Technical Data .................................................................................... 8
1.10 Definition of Accuracy....................................................................... 1 0
1.11 Safety Considerations ...................................................................... 10
1.12 Audio Alarm ...................................................................................... 10
1.13 Accessory AC High Voltage Probe ................................................... 1 1
1.14 Amp-Clamp Model 150-2 ................................................................. 11
2. INSTALLATION ................................................................................ 11
2.1 Unpacking and Inspection ............................................................... 11
2.2 Warranty ...........................................................................................11
2.3 Shipping ...........................................................................................12
2.4 Power Source Requirements ........................................................... 12
2.5 Operating Position ............................................................................ 1 2
3. CONTR OLS, JA CKS, AND INDICATORS........................................ 12
3.1 Front Panel Description .................................................................... 1 2
4. OPERATION.................................................................................... 14
4.1 Safety Precautions ........................................................................... 1 4
4.2 Polarity Reversing ............................................................................ 15
4.3 Measurement of Unknown Voltage or Current ................................. 1 5
4.4 Test Leads ........................................................................................1 6
4.5 DC Voltage Measurement ................................................................ 1 6
4.5.1 DC Voltage Measurement 0-250 mV Range .................................... 16
4.5.2 DC Voltage Measurement 0-1V Range .............................................. 16
4.5.3 DC Voltage Measurement 0-2.5 Through 0-250V Range ................. 17
4.5.4 DC Voltage Measurement 0-500V Range ......................................... 17
4.5.5 DC Voltage Measurement 0-1000V Range ....................................... 18
4.6 AC Voltage Measurement .................................................................1 8
4.6.1 AC Voltage Measurement 0-2.5 Thru 0-250V Range ........................ 1 9
4.6.2 AC Voltage Measurement 0-500V Range .......................................... 20
4.6.3 AC Voltage Measurement 0-1000V Range ........................................ 20
4.7 Output Voltage Measurement ...........................................................21
4.8 Decibel Measurement (-20 To +50 dB) ............................................ 2 2
4.9 Direct Current Measurement ............................................................2 3
4.9.1 Direct Current Measurement 0-50µA Range .................................... 23
4.9.2 Direct Current Measurement 0-1mA Through 0-500mA Range ...... 23
4.9.3 Direct Current Measurement 0-10A Range ...................................... 24
4.10 Resistance Measurements ............................................................... 25
4.11 Measuring Resistance ..................................................................... 25
4.12 Resistance Measurement of Semiconductors ................................. 2 5
4.13 Continuity Tests.................................................................................26
5. OPERA T OR MAINTENANCE .......................................................... 27
5.1 Inspection .........................................................................................27
4
5.2 Battery Replacement ........................................................................ 2 7
5.3 Fuse Replacement ........................................................................... 27
5.4 Test Lead Inspection ........................................................................ 28
5.5 Care..................................................................................................28
5
1. INTRODUCTION
1.1 General Description
The Simpson Volt-Ohm-Milliammeter 260 Series 8P, (hereafter referred to as the 260 or as the Instrument) is a rugged, accurate, compact, easy-to-use Instrument equipped with mirrored dial to eliminate parallax. The Instr ument can be used to make accurate measurements of AC and DC voltage, direct current, resistance, decibels, and Output Voltage. The Output Voltage function is used for measuring the AC component of a mixture of AC and DC voltage. This occurs primarily in amplifier circuits. New Features in the 260-8P are:
0 - 25V AC/DC Range Continuity Test Alarm Individual Dual Jacks for the 500V AC/DC and 1000V AC/DC Ranges
1.2 Overload Protection
All the ranges, with exception of 1V, 10A, 50 µA, 250mV, 500/1000V AC and DC are protected by an electronic overload protection system. The protection is from the usual overloads that would damage the VOM. A transistor ized circuit senses the voltage drop across the indicating instrument and actuates a relay when the voltage reaches approximately three to four times rated full scale voltage. Sens­ing of the voltage drop is by means of a bridge network so that overload protec­tion is provided regardless of polarity. The relay operates at a uniform percent of overload since the indicating instrument circuit is common to all ranges. When actuated by an overload, the relay contacts, which are in the -COMMON CIR­CUIT, latch open until the reset button on the front panel is depressed. The reset button is located to the right of the output jack.
!
Due to its sensitivity to overloads, the electronic overload protection relay incor­porated in the 260-8P is susceptible to electrostatic fields from high potentials or transients either in the circuit being measured or in nearby circuitry.
Do not assume the relay to be faulty if it trips out without apparent cause. It may be indicating a dangerous condition such as excessive circuit-to-ground (floating) voltage being applied to the Instrument.
Avoid loading the Instrument or its test leads until certain that excessive circuit-to­ground voltage does not exist.
a. A switch that is opened when the relay latches, opens the relay circuit and
prevents continuous battery drain. This condition occurs when the reset but­ton has been pressed so that the VOM is operational; the protection circuit draws less than the shelf life drain of the battery.
b. Since the 9V battery is used for both the R X 10,000V range and the protec-
tion circuit, the VOM is designed to allow the protection circuit to function normally as long as the R X 10,000V range can be set to zero and an Alka­line Battery is used.
c. The 1 and 2 Amp fuses in the input circuit adds protection to the relay con-
6
tacts. Some overload conditions may cause the fuses to burn out as well as open the relay contacts.
d. When an overload of sufficient magnitude is applied to the VOM, the reset
button will pop-up approximately 3/16" above the surface of the panel. To reset the VOM for normal operation, first remove the overload and depress the button. If the overload remains connected to the Instrument, resetting the overload protection reset button will not connect the VOM input circuit.
In addition to the overload protection circuit, two diodes protect the indicating instrument circuit. These diodes limit the current through the coil in case of over­load.
Fuses: Basic Overload protection is provided by a small inexpensive and easily obtainable 3 AG type, 1 Amp , quic k-b low fuse. This fuse is rated to interrupt circuit voltages up to 250 volts. A supplemental high energy fuse is provided for addi­tional protection from severe over loads. This fuse is the Bussman Type BBS 2A 600V. A Littlefuse Type BLS 2A 600V fuse may be used as a substitute replace­ment.
NOTES: If the 260-8P fails to indicate a reading, the 1 Amp or 2 Amp fuse may be burned out. (Refer to paragraph 5.4 for fuse replacement.) Use only the specified fuse type.
The 0-10A range is not fused. The shunt connects directly to the -10A and +10A jacks.
The relay, fuse and diodes will prevent serious damage to the Instrument in most cases of accidental overload. However, no overload protection system is com­pletely foolproof and misapplication on high voltage circuits can damage the Instrument. Exercise care and caution to protect both the user and the Instrument.
1.3 Internal Batteries
There are two batteries in the ohmmeter circuits, a NEDA 13F size D cell that furnishes 1.5-volts for the R X 1 and R X 100 ranges and a NEDA 1604A alkaline battery that furnishes 9-volts for the R X 10,000 range and overload circuit. The
1.5-volt D cell is held in place with two spring clips which also serve as battery contacts. The polarity symbols for the D cell are mark ed near the battery contacts. The 9-volt battery is held in place with a spring clip, but contact is made with a separate connector that is polarized. Always observe correct polarity when re­placing the 1.5-volt D cell.
!
The 9-volt battery must be installed and tested before the Instrument is ready for use. Perform the overload test described in Paragraph 1.4 to ensure proper over­load circuit operation prior to using the Instrument.
1.4 Overload Circuit Test
a. Set the range switch to R X 10,000 POSITION and the function switch to -DC
POSITION.
b. Plug the black test lead into the -COMMON jack terminal and touch the other
end of the test lead to the +1 volt input terminal. (The relay reset button should rise, indicating that the overload circuit is in working condition.)
NOTE: No damage will occur to the 260-8P as a result of this test.
7
1.5 Printed Circuit
Most of the component parts are mounted on a printed circuit board which simpli­fies assembly and maintenance, thus, extending the useful life of the Instrument.
1.6 Phenolic Case
The phenolic case is designed with heavy reinforced walls for maximum durabil­ity and provides excellent protection for the circuit components. The case has an externally accessible battery and fuse compartment. Access to the batteries and fuses is obtained by unscrewing a single captivated screw at the bottom of the case and sliding out the compartment cover.
Rubber plug bumpers on the bottom of the case reduce sliding should the 260 be accidentally pulled by the test leads when the Instrument is on a workbench.
1.7 Adjust-A-Vue Handle
A carrying handle is attached to the Instrument case. The handle may be used to support the Instrument in a convenient sloping position for easy viewing. The case can also be placed in either a vertical or horizontal position. The horizontal position is preferable for greater accuracy since the Instrument is calibrated in this position.
1.8 Test Leads
Each Instrument is furnished with one pair (Catalog Number 00043) of probe type test leads (one red and one black, each 48 inches long) for all applications of the Instrument. These test leads have elbow prods on one end, to connect the leads to the recessed jacks on the Instrument front panel. The probe tips at the opposite end have threaded shoulders to accept the screw-on insulated alligator clips (furnished with the test leads) or the 5000 volt test probe(s) to extend the Instrument AC or DC voltage range. The test leads and the insulated alligator clips are rated for the voltage and current ranges available on the 260. Maximum ratings are 1000V AC or DC or 10 DC amperes. These ratings for the test leads must not be exceeded in any application.
1.9 Technical Data
1. DC VOLTS: Ranges: 0-1-2.5-10-25-50-250-500-1000V Sensitivity: 20,000 ohms per volt
2. DC MILLIVOLTS: Range: 0-250mV Sensitivity: 20,000 ohms per volt
3. AC VOLTS: Ranges: 0-2.5-10-25-50-250-500-1000V Sensitivity: 5,000 ohms per volt
4. OUTPUT VOL TAGE (AC): Ranges: 0-2.5-10-25-50-250
Table 1-1. Technical Data
(limited to 350VDC)
8
5. DC MICROAMPERES: Range: 0-50µA Voltage Drop: 250mV
6. DC MILLIAMPERES: Ranges: 0-1-10-100-500 mA V oltage Drop (Approx.): 250 mV, 255 mV, 300 mV, 500 mV
7. DC AMPERES: Range: 0-10A Voltage Drop (Approx.): 255mV
NOTE: The 10A range is not internally fused.
8.
egnaR000,2-0 000,002-0 M02-0
:ECNATSISER1XR001XR000,01XR
gnidaeRretneC21 0021 000,021
nepOlanimoN egatloVtiucriC
trohSlanimoN
tnerruCtiucriC
V5.1V5.1V9
Am521Am52.1Aµ57
9. *ACCURACY: DC Voltage Ranges: 2% of Full Scale DC Current:
0-50 µA Range: 1.5% of Full Scale Other Ranges: 2% of Full Scale **AC Voltage
Ranges: 3% of Full Scale Frequency Response: Referenced to 100 Hz (Figures 4-5 and 4-9) Resistance Ranges:
R X 1: 2.5° of Arc
R X 100: 2°of Arc
R X 10,000: 2° of Arc
10. DECIBELS: Range: -20 to +10 dB, -8 to +22 dB,
0 to +30 dB, +6 to +36 dB, +20 to +50 dB
Reference Level: With zero dB power level equal to 1 mW across
a 600 line.
11. BATTERIES: Voltage: 1.5V, 9V NEDA No.: 13F, 1604A
12. FUSE: F1 1A, 250V , type 3A G, quick-acting Littlefuse T ype
312001, F2 2A, 600V Littlefuse Type BLS or 2A 600V, Bussman Type BBS.
13. TEST LEADS: 1 red, 1 black, 48" long.
14. SIZE: 5½” x 7" x 3 (13.34 x 17.78 x 7.94 cm)
15. WEIGHT: 3 lbs. (1.359 kg)
16. ***RATED-CIRCUIT TO-GROUND VOLTAGE: 1000V AC/DC Max.
9
*Accuracies specified are for the 260 in a horizontal position. **Responds to the average value of an AC current, and is calibrated to indicate the RMS value of a pure sine wave. ***Per ANSI C39.5 April 1974: The specified voltage with respect to ground, which may be safely and continuously applied to the circuit of an Instrument.
1.10 Definition of Accuracy
The voltage and current accuracy of this Instrument is commonly expressed as a percent of full scale. This should not be confused with accuracy of reading (indi­cation). For example, +2% of full scale on the 10 volt range allows an error of ±0.20V at any point on the dial. This means that at full scale, the accur acy reading would be ±2%, but at half scale it would be ±4%. Therefore, it is advantageous to select a range which gives an indication as near as possible to full scale.
1.11 Safety Considerations
This Operators Manual contains special caution and warning symbols which alert the user to hazardous operating and service conditions. These symbols appear throughout this publication where applicable, and are defined on the inside front cover of this manual under NOTES AND SAFETY INFORMATION. Please adhere to these instructions to ensure the safety of operating and servic­ing personnel and to retain the operating conditions of the Instrument.
1.12 Audio Alarm
The 260-8P Audio Alarm is a transducer which will emit an audible sound when the operator of the Instrument intends to quickly determine the continuity of a circuit.
NOTE: Not protected by Electronic Overload Protection System.
!
Do not make Circuit Continuity Tests with electrical power “ON”. Turn “OFF” all electrical power.
Table 1-2. Items and Accessories Furnished with This Instrument
Quantity Description Number
1 Test Lead Set - one red and one black (48 long) 00043
each with combination probe tip and removable
rubber-sleeved alligator clip. *1 1.5 Volt, D Cell, NEDA 13F *1 9.0 volt Cell, NEDA 1604A 2 1Amp, 250V Fuse, Littlefuse Type #312001 1 2A, 600V Fuse, Littlefuse Type BLS or
(Bussman Type BBS) 1 Operators Manual 6-114339
*Batteries are standard items replaceable from local retail stores.
10
Table 1-3. Additional Accessories
Description Catalog No. Deluxe Case 00812 Leatherette Ever-Redy Hard Case, Drop Front 00805 Model 150-2 Amp-Clamp 00545 Model 151-2 Line Splitter 00544
1.13 Accessory AC High Voltage Probe
The 10,000 volts AC accessory probe is similar to the high voltage DC probes with the following exceptions: The AC high voltage probe is designed to extend the 260-8P AC voltage range. The probe can be used with the Simpson 260-8P 10-volt AC range.
NOTE: Instructions are furnished with each high voltage probe.
1.14 Amp-Clamp Model 150-2
The Simpson Amp-Clamp Model 150-2 allows AC current measurements with­out breaking the circuit under test. The Amp-Clamp works as a transformer con­taining a split core for accommodating the conductor which carries the current being measured. The current-carr ying conductor becomes the transformer pri­mary and a coil in the Amp-Clamp serves as the secondary winding. The Amp­Clamp output voltage is proportional to the current measured and can be applied to the 260-8P as an AC voltage.
The Amp-Clamp has a range selector with six positions. Any of the following current ranges can be used with the 260-8P: 5, 10, 25, 50, 100 or 250 amperes.
NOTE: Instructions are furnished with each Amp-Clamp.
2. INSTALLATION
This section contains information and instructions for the installation and ship­ping of the Simpson 260-8P. Included are unpacking and inspection procedures, warranty, shipping, power source requirements and operating position.
2.1 Unpacking and Inspection
Examine the shipping carton for signs of damage. If damaged, notify the carrier and supplier and do not attempt further use of the Instrument. If Instrument ap­pears to be in good condition, read Operators Manual in its entirety. Become familiar with the Instrument as instructed in the manual, then proceed to check the electrical performance as soon as possible. Also, check that all items listed in Table 1-2 are included with the Instrument.
After unpacking the Instrument, a 1.5V battery and a 9V battery may be found in separate envelopes in the box with the Instrument and the test leads. Two alliga­tor clips for the test leads are in a polyethylene bag. (See Section 5 for instructions on how to open the battery compartment and install the batteries.
2.2 Warranty
The Simpson Electric Company warranty policy is printed on the inside front cover of this manual. Read it carefully before requesting any warranty repairs. For
11
all assistance, including help with the Instrument under warranty, contact the nearest Authorized Service Center for instructions. If necessary, contact the fac­tory directly, give full details of any installation difficulty and include the Instru­ment model number, series number, and date of purchase. Service data or ship­ping instructions will be mailed promptly. If an estimate of charges for non-war­ranty or other service work is required, a maximum charge estimate will be quoted. This charge will not be exceeded without prior approval.
2.3 Shipping
Pack the Instrument carefully and ship it prepaid and insured.
2.4 Power Source Requirements
There are two batteries in the ohmmeter circuits, a NEDA 13F D size cell that furnishes 1.5 volts for the R X 1 and R X 100 ranges and a NEDA 1604A battery that furnishes 9 volts for the R X 10,000 range. The 1.5 volt D cell is held in place with two spring clips which also serve as battery contacts. The 9-volt battery is held in place with a spring clip but contact is made with a separate connector that is polarized. (Always observe correct polarity when replacing the 1.5-volt D cell.)
2.5 Operating Position
A handle is attached to the side of the Instrument case. The handle may be used to support the Instrument in a convenient, sloping position for easy viewing. The VOM case can also be placed either in a vertical or horizontal position. The horizontal position is preferable for greater accuracy since the Instrument is cali­brated in this position.
3. CONTROLS, JACKS, AND INDICATORS
All operating and adjustment controls, jacks, and indicators for the Simpson 260­8P are illustrated in this section along with a list (Table 3-1) describing their function. Become familiar with each item before operating the Instrument.
3.1 Front Panel Description
The front panel controls, jacks, and indicators as described in Table 3-1, are depicted in Figure 3-1.
Table 3-1. Controls, Jacks, and Indicators
1. Front Panel The 260-8P V olt-Ohm-Milliammeter is a large, easy-to-read 4-1/4 indicating Instrument. Below the Instrument are four controls and eight circuit jacks. Switch positions and circuit jacks are marked in white characters for easy reading.
2. Range Switch: 12 positions: May be rotated 360° in either direction to select any of the five AC and DC voltages, four DC cur­rents and three resistance ranges.
12
O
S
M
H
O
H
M
S
.
.C
D
.
.
.C
A
.C
.A
V .5
2
B
D
7
D
.C
.
A.C
.
6
1
-10 A.
OUTPUT
350 VDC
A.C. VOLTS
ONLY
- D.C.
+ D.C.
3
COMMON
500 MA.
MAX.
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
+
-
ALL TERMINALS 1 V MAX
100MA.
10MA. 1MA.
R X I00
R x I0,000
260
5
+10A
+1V.
50 AMPS. +
250 MV.
ZERO OHMS
AMPS.
R x 1
500 V.
1000 V.
A.C. D.C.
A.C. D.C.
4
2
FIGURE 3-1. Front Panel Controls, Jacks
and Indicators
3. Function Switch: The function switch has four positions: Audio *, +DC, - DC, and AC Volts only. To measure DC current or voltage, set the function switch at the -DC or +DC position, depend­ing on the polarity of the signal applied across the test leads. To measure AC voltage set the function s witch to the AC position. For resistance measurement, the switch may be set in either the +DC or -DC position. The polarity of the internal ohmmeter battery voltage will be as marked at the jacks when the switch is in the +DC position and reversed in the -DC position.
4. Zero Ohms: This control is used to compensate for variation in the volt­age of the internal batteries, when making resistance mea­surements.
5. Circuit Jacks: There are eight jacks on the front panel marked with the functions they represent. These jacks provide the electri­cal connections to the test leads. The COMMON (-) jack is used (in conjunction with the black test lead) as the refer­ence point for the measurement of all the functions with the exception of the 10 A range. (Refer to Section 4 for details.)
6. Pointer Adjust
For Zero: With the Volt-Ohm-Milliammeter in an operating position,
check that the pointer indicates zero at the left end of the scale when there is no input. If pointer is off zero, adjust the screw located in the cover below the center of the dial. Use a small screwdriver to turn the screw slowly clockwise or counterclockwise until the pointer is exactly over the zero mark at the left end of the scale. With the indicating pointer
13
set on the zero mark, reverse the direction of rotation of the zero adjuster slightly to introduce mechanical freedom or play, but insufficient to disturb the position of the indicat­ing pointer. This procedure will avoid disturbances to the zero setting by subsequent changes in temperature, hu­midity, vibration, and other environmental conditions.
7. Reset Button: The white button to the right of the output jack is the reset button of the overload protection circuit.
4. OPERATION
!
Multifunction instruments (VOMs) such as the 260-8P are intended as general purpose measuring Instruments for use in low power circuitry such as found in consumer appliances, TV and radio receivers, and in general laboratory applica­tions. Their use is not recommended in high voltage, high power circuitry where operator errors and inadequate personal protective measures could result in serious injury from arcing or explosion. Such circuitry is found in industrial or commercial applications such as induction (r-f) heaters and broadcast transmit­ters, power substations and distribution centers, X-ray equipment, large motor controls, etc. Working in the high voltage section of such equipment requires single-function, often specially designed instruments used by trained personnel who adopt appropriate protective equipment and procedures.
This section of the manual contains information required to use and operate the 260-8P in a safe and proper manner.
4.1 Safety Precautions
The following precautions are suggestions and reminders of commonly recog­nized safe practices to be adopted and specific hazards to be avoided and are not implied to be sufficient to safeguard untrained personnel in all circumstances. Neither is this manual a substitute for technical manuals covering the equipment in which measurements are to be made. Always refer to the equipment manual and its specific warnings and instructions and observe them as well as those contained herein. a. The 260-8P should only be used by personnel qualified to recognize shock
hazards and trained in the safety precautions required to avoid possible injury.
b. Do not connect any terminal of this Instrument to a circuit point at which a
voltage exceeding 1000V AC or DC may exist with respect to earth ground. (Ref. Table 1-1, Item 16.)
c. Tur n off power and discharge any capacitors in the circuit to be measured
before connecting to or disconnecting from it.
d. Before using this Instrument, check accessories (if any) and test leads for
missing, damaged, deteriorated or otherwise faulty insulating parts. Do not use, or permit the use of equipment with faulty insulation until it has been properly repaired.
e. Always wear safety glasses when working with electrical circuitry.
14
f. Do not work alone on high voltage circuits. Make certain that someone ca-
pable of giving aid is nearby and alert.
g. Do not handle the Instrument, its test leads, or the circuitry while high voltage
is being applied.
h. Hands, shoes, floor, and workbench must be dry. Avoid making measure-
ments under humid, damp, or other environmental conditions that could effect the dielectric withstanding voltage of the test leads or Instruments.
i. Do not change switch settings or test lead connections while the circuit is
energized. A mistake could result in damage to the Instrument and personal injury.
j. Locate all voltage sources and accessible current paths before making con-
nections to circuitry. High voltages may appear unexpectedly or in unex­pected locations in faulty equipment. An open bleeder resistor, for example, may result in a capacitor retaining a dangerous charge.
k. Make certain that the equipment being worked upon is properly grounded
and fuses are of the proper type and rating.
l. Double check switch positions and jack connections before applying power
to the Instrument.
m. Remain alert for low voltage circuits which may be floating at high voltage
with respect to earth ground and for composite voltages (AC + DC) such as are found in r-f amplifiers. The floating voltage or composite voltage must not exceed the Instruments rated maximum circuit-to-ground voltage.
n. Do not make electrical measurements where the air may contain explosive
concentrations of gas or dust such as in mines, grain elevators, gasoline stations or in the presence of charging batteries until determined to be safe by qualified personnel. Note that even metallic dusts can be explosive.
o. No General Purpose VOM is to be used to make electrical measurements on
blasting circuits or blasting caps.
4.2 Polarity Re ver sing
The function switch provides a convenient means to reverse polarity of the inter­nal battery to facilitate testing semiconductor devices. The +DC and -DC posi­tions transpose internal connections to the COMMON and +jacks and therefore can also be used to reverse polarity of external voltages and currents connected to these jacks (only). Because polarity reversing momentarily interrupts the cir­cuit, it should not be done while measuring current or voltage without first turning off the circuit power. When making measurements on the 50µA/250mV, 1 volt, or 10 amp range, polarity can be corrected only by reversing the test leads.
4.3 Measurement of Unknown Voltage or Current
Sometimes the approximate voltage or current to be measured is known and the correct range may easily be selected. It is often the case, however, particularly when working upon faulty equipment, when the voltage or current is either un­known or may be much higher than normal and anticipated. Whenever working upon unfamiliar or unknown equipment, always begin a measurement using the highest range available on the Instrument. Once it is determined that the voltage is within the limits of a lower range, change to the lower range.
15
4.4 Test Leads
Test leads are provided with accessor y screw-on alligator clips which may be attached to the probe tips. Eliminating the need to hand-hold test prods to a circuit for extended periods, the test clips also reduce hand proximity to a high voltage circuit while energized. Circuit power must of course be turned off and any charged capacitors discharged before connecting or disconnecting clips to or from the current.
4.5 DC V oltage Measurement
!
Before making voltage measurements, re vie w the SAFETY PRECAUTIONS listed in paragraph 4.1. Also, when using the 260 as a millivoltmeter, take care to pre­vent damage to the indicating instrument from excessive voltage.
4.5.1 DC Voltage Measurement 0-250 mV Range
Before using the 250 millivolt range, use the 1.0-volt DC range to determine that the voltage measured is not greater than 250 millivolts (or .25 volt DC).
a. Set the function switch at +DC (Figure 4-1). b. Plug the black test lead in the -COMMON jack and the red test lead into the
+50µAmps/250mV/10A jack. c. Set the range switch at 50 µAmps (COMMON) position with 50V. d. Connect the black test lead to the negative side of the circuit being measured
and the red test lead to the positive side of the circuit. e. Read the voltage on the black scale marked DC and use the figures marked
0-250. Read directly in millivolts.
NOTE: For polarity reversing see paragraph 4.2.
4.5.2 DC Voltage Measurement 0-1V Range
a. Set the function switch at +DC (Figure
4-1).
b. Plug the black test lead into the -COM-
MON jack and the red test lead into
the +1V jack.
c. Set the range switch at 1V (COMMON
position with 10V).
d. Connect the black test lead to the nega-
tive side of the circuit being measured
and the red test lead to the positive
side of the circuit.
e. Read the voltage on the black scale
marked DC and use the figures
marked 0-10. Then divide the reading
by 10.
NOTE: For polarity reversing see para-
graph 4.2.
16
- D.C.
- D.C.
+ D.C.
+ D.C.
COMMON
-
Figure 4-1. Jacks and Switch Positions for Measuring DC Voltages, 0 through 1V Ranges
.
.C
D
10 V.
1V.
50 A.
AMPS.
+
260
ALL TERMINALS 1 V MAX
D
.C
.
+10A
+1V.
50 AMPS.
+
250 MV.
4.5.3 DC Voltage Measurement 0-2.5 Through 0-250V Range
a. Set the function switch at +DC (Figure
4-2). b. Plug the black test lead into the -COM-
MON jack and the red test lead into
the + jack. c. Set the range switch at one of the five
voltage range positions marked 2.5V,
.
.C
D
10V, 25V , 50V or 250V. NOTE: When in doubt as to the voltage
present, always use the highest voltage range as a protection to the Instrument. If the voltage is within a lower range, the switch may be set for the lower range to obtain a more accurate reading. Be sure power is off in the circuit being measured and all capacitors discharged.
d. Connect the black test lead to the nega-
- D.C.
+ D.C.
COMMON
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
+
-
260
ALL TERMINALS 1 V MAX
tive side of the circuit being measured
and the red test lead to the positive
side of the circuit. e. Turn on the power in the circuit being
measured.
Figure 4-2. Jacks and Switch Positions for Measuring DC Voltages, 0-250V Range
f. Read voltage on the black scale
marked DC. For the 2.5V range, use the 0-250 figures and divide by 100. For
the 10V, 50V, and 250V ranged, read the figures directly. For the 25V range,
use the 0-250 figures and divide by 10. NOTE: Turn off power to the circuit and wait until the meter indicates zero before
disconnecting the test leads.
D
50 AMPS.
+
.C
.
+10A
250 MV.
4.5.4 DC Voltage Measurement 0-500V Range
!
Use extreme care when working with high voltage circuits. Do not touch the Instrument or test leads while power is on in the circuit being measured.
Before proceeding with the following steps, review the Safety Precautions in Paragraph
4.1. a. Set the function switch at +DC (Figure
4-3). b. Set the range switch at the 250V/550V/
1000V Position. c. Plug the black test lead into the -COM-
MON jack and the red test lead into the
500V jack. d. Be sure power is off in the circuit being
measured and all capacitors dis-
17
.
.C
D
- D.C.
+ D.C.
COMMON
-
Figure 4-3. Jacks and Switch Positions for Measuring DC Voltages, 0-500V Range
250 V.
500 V. I000 V.
ALL TERMINALS 1 V MAX
D
.C
.
1000 V.
260
A.C. D.C.
charged. Connect the black test lead to the negative side of the circuit being
measured and the red test lead to the positive side of the circuit. e. Turn on power in circuit being measured. f. Read the voltage using the 0-50 figures on the black scale marked DC.
Multiple the reading by 10. NOTE: Turn off power to the circuit and wait until the meter indicates zero before
disconnecting the test leads.
4.5.5 DC Voltage Measurement 0-1000V Range
!
Use extreme care when working with high voltage circuits. Do not touch the Instrument or test leads while power is on in the circuit being measured. Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.1. a. Set the function switch at +DC (Figure
4-4). b. Set the range switch at the 250V/500V/
1000V Position. c. Plug the black test lead into the -COM-
MON jack and the red test lead into
the 1000V jack. d. Be sure power is off in the circuit be-
ing measured and all capacitors dis-
charged. Connect the black test lead
to the negative side of the circuit be-
ing measured and the red test lead to
the positive side of the circuit. e. Tur n on power in circuit being mea-
sured. f. Read the voltage using the 0-10 fig-
ures on the black scale marked DC.
Multiple the reading by 100. NOTE: T urn off power to the circuit and wait
until the meter indicates zero before
disconnecting the test leads.
.
.C
D
- D.C.
+ D.C.
COMMON
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
Figure 4-4 Jacks and Switch Positions for Measuring DC Voltages, 0-1000V Range
D
.C
1000 V.
A.C. D.C.
.
4.6 AC Voltage Measurement
!
Branch and distribution circuits (120/240/480V etc.) can deliver dangerous ex­plosive power momentarily into a short circuit before the fuse/breaker opens the circuit. Make certain that the Instrument switches are set properly, jacks are con­nected properly, and that the circuit power is turned off before making connec­tions to such circuits.
18
depending upon its waveform, the reading may be higher or lower than the true RMS value of the measured voltage. Thus an error may be introduced if the 260 is used to measure a nonsinusoidal waveform. Also, accuracy is lessened at higher input frequencies (Figure 4-5).
4.6.1 AC Voltage Measurement 0-2.5 Thru 0-250V Range
Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.1.
a. Set the function switch at AC
Volts Only position (Figure 4-6).
NOTE: The meter will not indicate if
the switch is incorrectly set to a
DC position.
b. Set the range switch at one of the
five voltage range positions
marked 2.5V, 10V, 50V or 250V.
(When in doubt as to actual volt-
age present, always use the
highest voltage range as a pro-
tection to the Instrument. If the
voltage is within a lower range,
the switch may be set for the
lower range to obtain a more ac-
curate reading.)
c. Plug the black test lead into the -
COMMON jack and the red test
lead into the + jack.
d. Turn off power to the circuit to be
measured and discharge any ca-
pacitors.
e. Connect the test leads across the
circuit voltage to be measured
with the black lead to the
grounded side.
f. For the 2.5V range read the value
directly on the scale marked 2.5V
AC. For the 10V, 25V, 50V and
250V ranges, read the red scale
marked AC and use the black fig-
ure immediately above the scale.
+6
+5 +4
+3
+2
+1
-1
-2
PERCENT RELATIVE ERROR
-3
-4
-5
AC VOL TAGE RANGES
2.5 VAC RANGE 10VAC RANGE
25 VAC RANGE
50 VAC RANGE 250 VAC RANGE 500 VAC RANGE
1000 VAC RANGE
0
10Hz 100Hz 1KHz 10KHz 100KHz 1MHz
FREQUENCY
Figure 4-5, Frequency Response AC Voltage Ranges
.
.C
A
.C.
2.5 V.A
A.C . VOLT S
ONLY
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
COMMON
+
-
260
ALL TERMINALS 1 V MAX
Figure 3-1. Fro
A.C.
Figure 4-6. Jacks and Switch Positions for Measuring AC Voltages, 0-2.5 through 0-250V Ranges
19
4.6.2 AC Voltage Measurement 0-500V Range
!
Be sure that the range switch is set to the 250V/500V/1000V range, function switch to AC Volts Only position, and test leads connected to common and 500V jack.
Do not touch the Instrument or test leads while the power is on the circuit being measured.
Use extreme care when working in high voltage circuits. Do not handle the Instru­ment or test leads while the circuit being measured is energized. OBSERVE ALL SAFETY PRECAUTIONS in paragraph 4.1 and in the instruction manual for the equipment being tested. Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.2.
a. Set the function switch at AC Volts
Only position. (Figure 4-7).
NOTE: The meter will not indicate if the
switch is incorrectly set to a DC po­sition.
b. Set the range switch at the 250V/
500V/1000V position.
c. Plug the black test lead into the -
COMMON jack and the red test lead into the 500V jack.
d. Be sure the power is off in the circuit
being measured and that all its ca­pacitors have been discharged.
e. Connect the test leads across the
circuit voltage to be measured with the black lead to the grounded side, if any.
f. Turn on the power in the circuit be-
ing measured.
g. Read the voltage on the red scale
marked AC. Use the 0-50 figures and multiply by 10.
.
.C
A
2.5 V.A.C.
A.C. VOLTS
ONLY
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
COMMON
-
260
ALL TERMINALS 1 V MAX
Figure 4-7. Jacks and Switch Positions for Measuring AC Voltage, 0-500V Range
A.C. D.C.
A.C.
500 V.
4.6.3 AC Voltage Measurement 0-1000V Range
!
Use extreme care when working in high voltage circuits. Do not handle the Instru­ment or test leads while the circuit being measured is energized. OBSERVE ALL SAFETY PRECAUTIONS in paragraph 4.1 and in the instruction manual for the equipment being tested.
20
Be sure that the range switch is set to the 250V/500V/1000V range, function switch to AC position, and test leads connected to common and 1000V jack.
Do not touch the Instrument or test leads while the power is on in the circuit being measured.
Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.2.
a. Set the function switch at AC Volts Only position. (Figure 4-8).
NOTE: The meter will not indicate if the switch is incorrectly set to a DC position. b. Set the range switch at the 250V/500V/
1000V position.
c. Plug the black test lead into the -COM-
.
.C
A
.C.
2.5 V.A
A.C.
MON jack and the red test lead into the 1000V jack.
d. Be sure power is off in the circuit be-
ing measured and that all its capaci­tors have been discharged.
e. Connect the test leads across the cir-
cuit voltage to be measured with the black lead to the grounded side.
f. Turn on the power in circuit being mea-
A.C . VOLT S
ONLY
COMMON
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
1000 V.
A.C. D.C.
sured.
g. Read the voltage on the red scale
marked AC. Use the 0-10 figures and multiply by 100.
Figure 4-8. Jacks and Switch Positions for Measuring AC Voltage, 0-1000V Range
4.7 Output V oltage Measurement
Certain situations necessitatethe measuring of the AC component of an Output Voltage where both AC and DC voltage levels exist. This occurs primarily in amplifier circuits. The 260-8P has a 0.1 mfd, 400 volt capacitor in series with the OUTPUT jack. The capacitor blocks the DC component of the current in the test circuit, but allows the AC or desired component to pass on to the indicating instrument circuit. The blocking capacitor may alter the AC response at low fre­quencies but is usually ignored at audio frequencies (Figure 4-9).
Do not connect the OUTPUT jack to a circuit in which the DC voltage component exceeds 350V. Before proceeding with the following steps, review the Safety Precautions in Paragraph 4.1. a. Set the function switch at AC Volts Only position. (Figure 4-10). b. Plug the black test lead into the -COMMON jack and the red test lead into the
OUTPUT jack.
c. Set the range switch at one of the range positions marked 2.5V, 10V, 25V,
50V, or 250V.
d. Connect the test leads across the circuit being measured with the black test
lead to the grounded side.
21
A.C. VOLTS
ONLY
COMMON
+10
+8
.
.C
A
2.5 V.A.C.
OUTPUT
350 VDC
500 MA.
MAX.
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
-
260
ALL TERMINALS 1 V MAX
A.C.
+6
+4
+2
0
-2
-4
-6
-8
-10
10Hz 100Hz 1KHz 10KHz 100KHz 1MHz
OUTPUT RANGES
2.5 VAC RANGE 10VAC RANGE
25/50 VAC RANGE
250 VAC RANGE
FREQUENCY
Figure 4-10. Jacks and Switch Positions for Output Measurements
Figure 4-9. Frequency Response Output Ranges
e. T urn on the power in the test circuit. Read the output voltage on the appropri-
ate AC voltage scale. For the 0-2.5V range, read the value directly on the scale marked 2.5V AC. For the 10V, 25V, 50V, or 250V ranges, use the red scale marked AC and read the black figures immediately above the scale.
4.8 Decibel Measurement (-20 To +50 dB)
Some industries measure in terms of voltage or current ratios (decibels) based on a specific reference level. The dB scale on the 260-8P serves this purpose and is calibrated to a reference level (zero dB) or 0.001 watt into 600 ohms. The scale calibration of -20 to +10 dB is for the 0-2.5V range (zero dB = 0.775V). Higher ranges may be used by adding an appropriate dB value to the reading in accor­dance with the chart below.
a. Review the safety precautions in paragraph 4.1. b. Refer to Figure 4-11 for switch settings and jack connections. c. Use operating instructions for AC VOLTAGE MEASUREMENT,
0-2.5/10/25/50/250V RANGES (paragraph 4.6.1).
Do not use the 500V or 1000V ranges for decibel readings.
d. Read decibels on the bottom scale marked dB. For the 10/25/50/250V ranges,
add the appropriate factor from the following chart:
EGNAREGATLOVEGNARBdGNIDAER
V5.2-001+ot02-yltceriddaer V01-022+ot8-gnidaerotBd21+dda V52-003ot0gnidaerotBd02+dda V05-063+ot6+gnidaerotBd62dda
V052-005+ot02+gnidaerotBd04dda
22
NOTE: The maximum v oltage ratio that can be measured is +50 dB on the 0-250V
range. e. If dB measurements are being made to a 0.006 watt into 500 ohm reference
level, subtract +7 dB from the reading obtained on the 260.
4.9 Direct Current Measurement
!
a. Do not change the range setting of
the range or function switches while the circuit is energized.
b. Never disconnect test leads from the
circuit under measurement while it is energized.
c. Always turn the power off and dis-
charge all the capacitors before the setting of the switches is changed, or the leads disconnected.
d. Never exceed the circuit-to-ground
voltage of the instrument (1000V Max., Table 1-1, Item 16).
e. Always connect the instrument in se-
ries with the ground side of the cir­cuit.
f. In all direct current measurements,
make certain the power to the circuit being tested has been turned off be­fore connecting and disconnecting test leads or restoring circuit conti­nuity.
A.C . VOLT S
Figure 4-11. Jacks and Switch Positions for Measuring Decibels
ONLY
COMMON
B
D
500 MA.
2.5 V. 10 V.
1V.
25 V. 50 A.
AMPS.
250 V.
500 V. I000 V.
+
-
260
ALL TERMINALS 1 V MAX
4.9.1 Direct Current Measurement 0-50µA Range
a. Set the function switch at +DC. b. Plug the black test lead into the -COMMON jack and the red test lead into the
+50µAMPS/250mV jack. c. Set the range switch at 50µAMPS (dual position with 50V). d. Open the ground side of the circuit in which the current is being measured.
Connect the VOM in series with the circuit. Connect the red test lead to the
positive side and the black test lead to the negative side. e. Read the current on the black DC scale. Use the 0-50 figures to read directly
in microamperes.
NOTE: For polarity reversing see paragraph 4.2.
4.9.2 Direct Current Measurement 0-1mA Through 0-500mA Range
a. Set the function switch at +DC (Figure 4-12). b. Plug the black test lead into the -COMMON jack and the red test lead into the
+ jack.
c. Set the range switch at one of the four range positions marked 1 mA, 10 mA,
100 mA or 500 mA
23
d. Open the grounded side of the circuit in which the current is being measured.
Connect the VOM in series with the circuit. Connect the red test lead to the positive side and the black test lead to the negative side.
e. Turn the power on. Read the current in milliamperes on the black DC scale.
For the 1 mA range, use the 0-10 figures and divide by 10. For the 10 mA range, use the 0-10 figures directly. For the 100 mA range, use the 0-10 figures and multiply by 10. For the 500 mA range, use the 0-50 figures and multiply by 10.
f. Turn the power off and disconnect the test leads.
!
This range is only intended for measurements in low voltage circuits (under 25V DC) such as the primary power in vehicles or their accessories.
4.9.3 Direct Current Measurement 0-10A Range
a. Plug the black test lead into the -10A
jack and the red test lead into the +10A jack.
b. Set the range switch at 10 AMPS (dual
position with 10 mA).
c. Open the grounded side of the circuit
in which the current is being mea­sured. Connect the VOM in series with the circuit, connecting the red test lead to the positive side and the black test lead to the negative side.
NOTE: The function switch has no effect
on polarity for the 10 AMPS range.
d. Turn the power on. Read the current
directly on the black DC scale. Use the 0-10 figures to read directly in am­peres.
e. Turn the power off and disconnect the
test leads.
.
.C
D
-10 A.
COMMON
500 MA.
50 A.
AMPS.
+
-
- D.C.
+ D.C.
260
ALL TERMINALS 1 V MAX
100MA.
10MA.
AMPS.
1MA.
Figure 4-12. Jacks and Switch Positions for Measuring Direct Current
!
The 10A range is not internally fused in the 260-8P. When using the 10A range, never disconnect a test lead from a jack or from the
circuit while the circuit is energized. Doing so may damage the test jacks or leads and the arcing may be hazardous to the operator. Turn off circuit power and wait for the meter reading to drop to zero.
The polarity reversing feature of the function switch does not operate on the 0­10A range. If polarity is found to be incorrect, turn off circuit power, wait for the meter indication to reach zero and then interchange test lead connections to the circuit.
D
.C
.
+10A
50 AMPS.
+
250 MV.
24
4.10 Resistance Measurements
!
Voltage applied to a resistance range will cause reading errors if low and dam­age the Instrument if high. When making in-circuit measurements, make certain that the circuit is completely de-energized before making connections to it.
When making in-circuit measurements, circuit paths in parallel with the resis­tance being measured may cause reading errors. Check circuit diagrams for the presence of such components before assuming that the reading obtained is correct.
When resistance is measured, the batteries furnish power for the circuit. Since batteries are subject to variation in voltage and internal resistance, the Instru­ment must be adjusted to zero before measuring a resistance, as follows: a. Turn range switch to desired ohms range. b. Plug the black test lead into the -COMMON jack and the red test lead into the
+ jack.
c. Connect the ends of test leads together to short the VOM resistance circuit. d. Rotate the ZERO OHMS control until pointer indicates zero ohms. If pointer
cannot be adjusted to zero, one or both of the batteries must be replaced. (For battery replacement, refer to Section 5.)
e. Disconnect shorted test leads.
4.11 Measuring Resistance
a. Before measuring resistance in a circuit, make sure the power is off to the
circuit being tested and all capacitors are discharged. Disconnect shunting components from the circuit before measuring its resistance.
b. Set the range switch to one of the resistance range positions as follows
(Figure 4-13):
1. Use R X 1 for resistance readings from 0 to 200⍀.
2. Use R X 100 for resistance readings from 200 to 20,000⍀.
3. Use R X 10,000 for resistance readings above 20,000⍀.
c. Set the function switch at either -DC or +DC position: Operation is the same
in either position except if there are semiconductors in the circuit. (See para­graph 4.12.) Adjust ZERO OHMS control for each resistance range.
d. Observe the reading on the OHMS scale at the top of the dial.
NOTE: The OHMS scale reads from right to left for increasing values of resis-
tance. e. T o determine the actual resistance value , m ultiply the reading by the f actor at
the switch position. (K on the OHMS scale equals one thousand.)
4.12 Resistance Measurement of Semiconductors
Make sure that the OHMS range being used will not damage any of the semicon­ductors (refer to Table 1-1, item 8, Section 1).
If there is a forward and backward resistance such as in diodes, the resistance should be relatively low in one direction (for forward polarity) and higher in the opposite direction.
25
Rotate the function switch between the two DC positions to reverse polarity. This will determine if there is a difference between the resistance in the two directions.
S
M
H
O
OHMS
To check a semiconductor in or out of a circuit (forward and reverse bias resistance measurements) consider the following
fore
making the measurement:
be-
a. The polarity of the internal ohmmeter
battery voltage will be as marked at the jacks when the switch is in the +DC position and reversed in the -DC posi­tion.
b. Ensure that the range selected will not
damage the semiconductor. (Refer to T able 1-1, item 8, and re view the speci-
- D.C.
+ D.C.
COMMON
+
-
260
ALL TERMINALS 1 V MAX
R X I00
R x I0,000
ZERO OHMS
R x 1
fication limits of the semiconductor ac­cording to the manufacturers ratings.)
c. If the semiconductor is a silicon diode
Figure 4-13. Jacks and Switch Positions for Measuring Resistance
or conventional silicon transistor, pre­cautions are not normally required.
d. If the semiconductor material is germanium, check the ratings of the device
and refer to Table 1-1, item 8.
NOTE: The resistance of diodes will measure differently from one resistance
range to another on the VOM with the function switch in a given position. For example, a diode which measures 80 ohms on the R X 1 range may mea­sure 300 ohms on the R X 100 range. The difference in values is a result of the diode characteristics and is not indicative of any fault in the VOM.
4.13 Continuity Tests
!
Tur n off all power to the circuit under test before proceeding any further.
a. Set the FUNCTION switch to the *”.
Insert the black test lead into the COM­MON (-) and the red test lead into the (+) jacks, (Figure 4-14).
b. Touching the test lead probe tips to-
gether should produce an audible sound
c. To test for continuity, probe the circuit
or component in question. An audible sound indicates continuity.
d. The higher the circuits resistance, the
lower the sound level. The lower the circuits resistance, the higher the sound level.
e. Refer to Paragraph 4.12, Figure 4-13
for measurements of circuit resistance, if any, that exists.
26
R X I00
R x I0,000
OHMS
ZERO OHMS
R x 1
S
M
H
O
- D.C.
+ D.C.
COMMON
+
-
260
ALL TERMINALS 1 V MAX
Figure 4-14. Jacks and Switch
Positions for Continuity Tests
5. OPERATOR MAINTENANCE
The following paragraphs of this section describe battery replacement, fuse re­placement, and preventive maintenance procedures for the 260.
5.1 Inspection
The user is protected from electrical shock by the insulation of the 260 and its test leads. Frequently examine them for any insulation damage such as cracks, cuts, chips, burns or deterioration that expose internal metal parts or reduce the spac­ing between such metal parts and hand contact by the operator.
Make certain that the battery compartment cover is securely fastened in place before the Instrument is used.
Do not permit an instrument with a broken meter glass to be used. Whenever the battery compartment cover is removed for any reason, check that
the proper fuses are being used.
5.2 Battery Replacement
a. Two batteries are used inside the case to supply power for resistance mea-
surements. One is a 1.5-v olt D cell and the other is a 9 v olt battery. When it is no longer possible to adjust the pointer to zero for the R X 1 and R X 100 ranges (refer to ZERO OHMS ADJUSTMENT paragraph 4.21) replace the
1.5 volt cell. When it is no longer possible to adjust the pointer to zero on the R X 10,000 range, replace the 9 volt battery.
b. Recommended replacement batteries are:
1.5V D cell, NEDA 13F (Duracell MN or Eveready No. 95). 9V battery, NEDA 1604A (Duracell MN1604 or Eveready 522).
c. To install or replace a battery , de-energize and disconnect test leads from the
Instrument, then remove the cover to the externally accessible battery com­partment by loosening the single captivated screw.
NOTE: Batteries should be replaced before their useful life has e xpired. Failure to do so may result in corrosion and battery leakage.
d. Observe polarity when replacing the 1.5 volt D cell and connect as indicated.
The D cell is held in place with spring clips which also act as battery contact clips. The 9 volt battery contacts and connector are polarized. To remove the 9 volt battery, first withdraw battery with mated connector from the compart­ment. Then remove the connector.
5.3 Fuse Replacement
The 1 and 2 ampere fuse (also the 1A spare) is located in the externally acces­sible battery and fuse compartment. Access to the compartment is obtained by disconnecting the test leads and loosening the single captivated screw on the compartment cover. To replace or check a fuse: Apply pressure to the 1 amp fuse cup then rotate the plate to expose a selected fuse for removal from the panel’s cavity.
NOTE: When you replace a fuse, make sure it is the same type and value as the fuse you have just removed.
27
5.4 T est Lead Inspection
Periodic inspection of the test leads is recommended to detect cuts, burned areas, rioration or other damage that could reduce the insulation strength of leads.
5.5 Care
a. Immediately clean all spilled materials from the Instrument and wipe dr y. If
the spillage is corrosive, use a suitable cleaner to neutralize the corrosive action.
b. When Instrument is not in use, rotate the function switch to the +DC position. c. Avoid prolonged exposure or usage in areas which are subject to tempera-
ture and humidity extremes, vibration or mechanical shock, dust or corrosive fumes, or strong electrical or electromagnetic interferences.
d. Verify Instrument accuracy by performing operational checks using known,
accurate, stable sources. If proper calibration equipment is not available, contact the nearest Authorized Service Center. If the Instrument has not been used for 30 days, check the batteries for leakage and replace if necessary.
e. It is recommended that the Instrument be returned annually to the nearest
Authorized Service Center, or to the factory, for an overall chec k, adjustment, and calibration.
f. When the Instrument is not in use, store it in a room free from temperature
extremes, dust, corrosive fumes, and mechanical vibration or shock. If stor­age time is expected to exceed 30 days, remove batteries.
dete-
Figure 5-1. Battery and Fuse Compartment
28
29
NOTES
30
NOTES
31
SIMPSON ELECTRIC COMPANY 853 Dundee Avenue
Elgin, IL 60120-3090 (847) 697-2260 FAX (847) 697-2272 Printed in U.S.A. Part No. 06-114339 Edition 8, 12/01
Visit us on the web at: www.simpsonelectric.com
32
Loading...