Any changes to this medical device are fundamentally not allowed.
These Installation Instr uctio ns instructions enable the safe and
efficient assembly and installation of the Sim.LED
450/500/700/5000/7000 SC/MC. The Installation
InstructionsService Manual must be stored so that they are
accessible to the lighting manufacturer's service personnel or
personnel authorised by the manufacturer. The
must be followed in ascending order.
As a rule, medical devices may only be installed, operated,
controlled and maintained by persons who possess the required
training, knowledge and experie nce.
General
assembly sequence
Safety information
Installation, maintenance and repairs may only be conducted by
authorised, qualified personnel as a minimum requirement.
Authorisation is granted upon completion of a training course and
receipt of a certificate from the manufacturer. The qualified
personnel in question must have carefully read these Installation
Instructions and understood their contents before starting any work.
The basic requirement for a safe work process is the adherence to
all safety and handling instructions in these Installation Instructions
in ascending order. Furthermore, local regulations for working with
medical equipment in sterile rooms such as operating theatres,
apply.
The illustrations in this manual are provided for basic
understanding and could deviate from the actual design. Service
work is described in a separate Service Manual. Operation of the
lights and accessories is described in the respective Instructions for
Use.
Safety information in these Installation Instructions is identified by
symbols.
Risk of death or severe bodily injury!
Non-observance could possibly result in death or severe bodily
injury!
Risk of injury!
Non-observance could result in bodily injury!
Risk of material damages: Non-observance of the instructions
may result in damage to or destruction of the product or other
objects, as well as loss of data or loss of working hours.
Indicates a status or an automatic sequence as a result of
a handling step
“Reference title”, page 24
is a cross-reference to a chapter in this document.
Instructions for Use of this lighting
is a cross-reference to a separate document.
Abbreviation Cable colour (en) Cable colour (ger)
BK Black Schwarz
BU Blue Blau
BN Brown Braun
GN Green Grün
GNYE Green-Yellow Grün-gelb
OG Orange Orange
RD Red Rot
YE Yellow Gelb
Copyright protection
TR Transparent Transparent
These Installation Instr uctio ns are copyright-protected.
Transferring these Installation Instructions to a third party,
reproducing them in any type or form – even if only partial – and
using and/or disseminating their contents are prohibited without
written authorisation from the manufacturer.
Infringements will be liable to claims for damages.
We reserve the right to assert further claims.
All specifications and instructions in these Installation Instr uc tions
have been compiled under consideration of applicable norms and
standards, the current state of the art and our many years of
knowledge and experience.
We will accept no liability. The warranty and guarantee will become
inapplicable in the event of damages resulti ng from:
− Non-observance of these Installat ion Instr uct ions
− Non-observance of the Service Manual
− Non-adherence to the designated use / specific function
− Failure to document and observe the defined maintenance
plan and safety-relate d inspections
− Start-up currents and overvoltage
− Use by non-qualified personnel
− Modifications conducted independently
− Technical modifications
− Use of a defective or improperly repaired unit
− Use of unauthorised spare parts or accessories
The actual scope of delivery may deviate from the explanations or
illustrations provided in this manual in the case of special designs,
the use of additional order options, or due to the most recent
technical changes.
The following symbols can be found on the type plate and/or packaging. The symbols must always be
observed.
Adhere to the Instructions for Use
Electric shock: Warns of an electric shock which can cause severe injury or even
death.
Collapse of bearer arm system: Warns of the bearing arm system suddenly
collapsing due to faulty mounting.
Cutting injury:
Warns of cuts. Wear protective gloves.
Spring arm snapping up: Warns of the spring arm snapping up suddenly when the
adaption or end device is disass em bled.
Specifications on minimum and maximum ambient temperature for storage and
transportation
Specifications on minimum and maximum air pressure for storage and
transportation.
Specifications on minimum and maximum air humidity for storage and
transportation
The arrows point towards the top side of the package. They must always point
upwards; otherwise, the contents could be damaged.
This symbol identifies packages that contain breakable or sensitive contents.
Handle the package carefully, do not allow it to drop and do not subject it to any
V. 1.7 10
blows.
Protect the package from wetness and keep dry.
In addition, our personnel is always interested in hearing about new
Date of manufacture and manufacturer's address
Article number
Serial number
Protective grounding (ground)
Medical Equipment – General Medical Equipment
ONLY COMPLIES WITH ANSI/AAMI ES60601-1 (2005), CAN/CSA-C22.2 No. 60601-1 (2008), IEC
60601-2-41:2009, UL 606011:2006, CAN/CSA-C22.2 No. 601.1 AND IEC 60601-2-41:2000
STANDARDS WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS
CE mark of conformity
General
1.3 Spare parts
1.4 Warranty provisions
Procure spare parts from your authorised dealer or directly from the
manufacturer. For the address, see page 2.
Risk of injury due to the use of incorr ect spare parts!
The use of incorrect or defective spare parts may place
personnel and patients at risk as well as cause damage,
malfunctions, or a total breakdown of the unit.
Only use original manufacturer spare parts or manufacturer-
approved spare parts.
If uncertain, always contact the manufacturer.
The warranty provisions are contained in the manufacturer's
General Business Terms and Conditions.
The manufacturer's warranty is voided if unauthorised spare parts
and devices are used. Authorised devices can be found in the valid
Instructions for Use. Authorised spare parts can be found in the
valid Spare Parts Price List.
1.5 Technical Service
V. 1.7 11
Our Technical Service is available to provide technical information.
For contact information, see page 2.
The concept of intended use also includes adherence to all
specifications in these Installation Instructions and the separate
Instructions for Use. Any use that exceeds the intended use, and
any other kinds of uses, are considered to be incorrect uses.
Danger due to incorrect use!
Incorrect use of the unit may lead to dangerous situations. The
following are especially considered to be incorrect uses:
−Use of the unit in facilities that have not been built in
compliance with applicable s tandards and gu ide lines
regulating the construction of medical facilities.
− Use of the unit in explosion-prone areas.
− Use of a damaged unit.
Any type of claims against the manufacturer due to damages
caused by misuse, alteration or modification of the light are
excluded.
1.7 Incoming inspections
Inspect your delivery against the delivery slip for completeness and
integrity, immediately after receipt. The goods must be checked for
transport damages. The manufacturer or dealer must be notified of
any potential transport damages within 5 business days.
− Opening of the unit.
− Use of the unit by unqualified personnel.
− Use of the unit when objects are hanging from its extension
arm, spring arm or light head.
Risk of injury due to falling objects!
Parts may fall off of damaged goods.
Any goods which may be damaged may not be installed and
The operator must inform himself of all applicable accident
prevention and hygiene regulations, and must additionally conduct
a risk assessment at the site where the unit will operate in order to
identify any risks posed by the particular work conditions. The
regulations must be implemented in the form of instructions for
operation of the unit.
During the entire time that the unit is in use, the operator must
check whether the operating instructions that he prepared comply
with current technical regulations and revise them if necessary.
The operator must clearly establish and manage responsibilities in
the areas of installation, operation, troubleshooting, maintenance
and cleaning.
The operator must ensure that all employees who handle the unit
have read and understood the Instructions for Use.
Technical safety inspections
No liability in the event of noncompliance with time limits!
Translation
Furthermore, he must train personnel at regular intervals and
inform them of all dangers. He must also place safeguards so that
unauthorised persons do not use the unit.
The operator must ensure that the maintenance intervals and
technical safety inspections described in these Installation
Instructions are adhered to and documented.
The operator must ensure that only approved accessories and
accessories released by the manufacturer are used together with
the unit.
The operator must allow technical safety inspections to be
conducted annually.
Technical safety checks may only be conducted by the
manufacturer's personnel, or by authorised specialists who have
received written approval from the manufacturer.
The protocol prepared by the authorised specialist, detailing the
measurement procedures, measurement results, and other
evaluations, must be kept until the ligh tin g is disp os ed of
completely.
The manufacturer assumes no liability for personal injury or
material damages if technical safety inspections are not contracted
and conducted within the time limits provided.
In the case of translations into other languages, the German
version shall be binding.Instal lat ion Ins truc ti ons
Malfunctions or unit defects that lead to bodily injury must be
immediately notified to the authorities in charge and to the
manufacturer.
The authorities in charge may request that the operator submit the
incident being notified to a technica l safety evaluation by an
authorised expert, at its own expense, and that the evaluation be
submitted in writing to such authorities. The authorised expert will
be selected in consultation with the author it ies in char g e.
The technical safety evaluation will include determinations on:
− the cause of the incident
− the proper condition of the unit
− the risk situation following rectification of the fault
− the procurement of new knowledge as a result of the incident,
requiring revised or additional precautionary measures to
those already described
As a rule, the medical devices described in these Installation
Instructions may only be installed, operated, controlled and
maintained by persons who have been trained and authorised to do
so by S.I.M.E.O.N. Medical GmbH & Co. KG.
Risk of death due to faulty installation or faulty initial startup!
Errors during the installation or initial start-up may result in lifethreatening situations and cause considerable material damages.
For that reason, please note the following:
The installation and initial start-up may only be conducted by
the manufacturer's personnel or by persons authorised by
the manufacturer.
The manufacturer must also be involved when conducting
unit relocations at a later time.
Unauthorised installation and relocations are prohibited.
1.10 Dismantling and disposal
Disposal of the unit
Disposal of packaging
The unit must be dismantled and undergo environmentally friendly
disposal.
Dismantling may only be conducted by trained, qualified personnel.
Risk of death due to improper dismantli n g!
Errors during dismantling may result in life-threatening situat ions
and cause significant material damages.
Only allow trained, qualified personnel to conduct
The manufacturer must also be involved when conducting
unit relocations at a later time.
Unauthorised dismantling and relocations are prohibited.
The packaging must undergo environmentally friendly disposal.
The packaging may be returned to the manufacturer.
The installation and initial start-up should be exclusively conducted by the manufacturer's personnel or
by persons authorised by the manufacturer.
The case may be, however, that operators or the operator's maintenance personnel are entrusted with
the handling of packaged items during the course of installation and further use. In such a case, the
following instructions must be heeded!
Risk of material damages due to improper transportation!
Transported items may fall or topple over as a result of improper
transportation. This can result in costly material damages.
When unloading items during delivery, and in the case of in-house
transportation, it is important to proceed cautiously and to heed the
instructions on the packaging.
Use only the designated suspension points.
Do not remove packaging until immediately before assembly.
2.1 Transportation inspection
2.2 Packaging
Inspect the delivery immediately upon receipt for completeness and possible
transportation damages.
If any exterior transportation damages are discovered, proceed as follows:
Do not accept delivery or only accept it conditionally.
Write down the extent of damages on the transportation documents or on the
transportation company's delivery note.
Initiate a claim.
Submit a claim for every fault as soon as it is discovered. Claims for damages
can only be asserted within the claim period in effect.
The individual items have been packaged in accordance with the
expected transportation conditions. Only environmentally friendly
materials were used for the packaging.
The packaging should protect the individual components from
transportation damages, corrosion, or other damages, up until the
time of assembly. Therefore, do not destroy the packaging and
only remove it shortly before assembly. Dispose of packaging
materials in accordance with valid legal regulations and local
standards.
Transport all packaged items under the following conditions:
Do not store outdoors.
Store in a dry and dust-free environment.
Transport temperature: -10
Protect from sunlight.
Avoid any mechanical vibrations.
Ambient temperature: - 10°C to max. + 50°C
Relative air humidity: min. 10% max. 90 %
Risk of environmental damage due to improper disposal!
Packaging materials are valuable raw materials and can, in many cases, be
reused or be sensibly processed and recycled. The improper disposal of
packaging materials may result in danger to the environment.
Dispose of packaging materials in an environmentally friendly way.
Observe the applicable local disposal regulations. If needed, contract a
specialist company for such disposal.
°C to 50 °C
Air pressure min. 500 hPa max. 1060 hPa
When transporting parts for longer than three months, check the general condition of all parts
and their packaging regularly.
If required, recondition or replace the preservation measures.
V. 1.7 18
Page 19
2.4 Storage
Store the packaged items under the following conditions:
Do not store outdoors.
Store in a dry and dust-free environment.
Do not expose to any aggressive media.
Protect from sunlight.
Avoid any mechanical vibrations.
Relative air humidity: min. 10% max. 90 %
Air pressure min. 500 hPa max. 1060 hPa
When storing parts for longer than three months, check the general condition of all parts and
their packaging regularly.
If required, recondition or replace the preservation measures.
Storage instructions that exceed the requirements mentioned
herein may be found on the packaged items. These must be
adhered to accordingly!
The different variants of the Surgical LightingSim .LED 450/ 500/ 700 /50 00/7000 SC/MC surgical lights are
exclusively intended for the illumination of treatment and surgical areas used in medicine. The mobility of the
light heads is guaranteed through a system consisting of extension arms, spring arms, a cardanic, and light
heads. The surgical lights and the camera, as an optional accessory, are operated through the control
panels on the light heads and on the wall. The light head is positioned using the unsterile handles and the
sterilisable handle.
The Surgical LightingSim.LED 450/500/700/5000/7000 SC/MC surgical light is available in different variants.
3.1.1 Light head
The various type designations generally denote light heads with different numbers and designs of Sim.PODs
(reflector/LED unit),
“Technical data“, page 119:
Sim.LED 5000 (SC)
24 reflectors
Sim.LED 450 (SC)
12 reflectors
Multi-colour lights (MC) allow you to select the factory-set colour temperatures on the control panel,
Instructions for Use Sim.LED 450/500/700 and Sim.LED5000/7000.
Sim.LED 7000 (SC)
32 reflectors
Sim.LED 500 (SC)
18 reflectors
Sim.LED 700 (MC, with
Sim.CAM)
24 reflectors
3.1.2 Camera
Lights with camera preparation can transmit/process the signals of a Sim.CAM SD or Sim.CAM HD TOUCH
accordingly, Instructions for Use Sim.LED 450/500/700 and Sim.LED 5000/7000.
Risk of injury from failing to follow thes e Installation
Instructions. Installations may only be carried out by trained
and authorised qualified personnel.
Consumption of alcohol or drugs prior to or during
installation are strictly prohibited. The safety, warning and
cautionary instructions must be followed!
Risk of death due to heavy objects!
Parts could fall down while the components are being mount ed.
This could cause severe or even fatal injuries.
Never attempt to lift heavy objects by yourself.
If available, use a hoisting platform or a lift.
Mount with at least 2 trained and authorised technicians!
Wear tight-fitting clothing.
Never wear rings, necklaces and other jewellery.
Cover long hair with hairnets.
Risk of death due to improper faste n ing of stud bolts! Risk
of injury due to reuse of stud bolts!
The collapse of the ceiling structure can result in a lifethreatening situation. Stud bolts are not suited for multiple uses.
Their reuse could thus imperil their bearing capacity and result in
injuries to personnel and patients, as well as material damages to
the unit.
The ceiling must be inspected and approved by a structural
engineer prior to installation.
Only use the stud bolts once
Adhere to the specified torque for the stud bolts.
Use only the stud bolts supplied.
Only mount the ceiling anchor plate on an absolutely horizontal
level surface (< ± 0.1°), there may be no offset areas on the
ceiling.
Have a structural engineer inspect and approve the ceiling,
“Confirmation of static bearing capacity“, page113.
The ceiling must be a flat surface, without any irregularities.
Observe the minimum distance to the wall (see illustration).
Drill and thoroughly blow out 4 holes (3) in the ceiling (4) for
the HSL3G M12 stud bolts (5), in accordance w ith the
attachment bracket manufacturer’s specifications
(www.hilti.de).
Mount stud bolts (5) accordance with the attachment bracket
manufacturer's specifications.
Using stud bolts (5), screw the ceiling anchor plate (1) to the
ceiling (4) and tighten each stud bolt with a torque of 60 Nm.
4.1.2 Mounting the suspended ceiling fitting
Mounting the ceiling plate
Have a structural engineer inspect and approve the ceiling,
“Confirmation of static bearing capacity“, page113.
The ceiling must be a flat surface, without any irregularities.
Observe the minimum distance to the wall (see illustration).
Drill and thoroughly blow out 4 holes (3) in the ceiling (4) for
the HSL3G M12 stud bolts ( 5), in accor danc e with the
attachment bracket manufacturer’s specifications
(www.hilti.de).
Mount stud bolts (5) accordance with the attachment bracket
manufacturer's specifications.
Using stud bolts (5), screw the ceiling plate (1) to the ceiling
(4) and tighten each stud bolt with a torque of 60 Nm.
Risk of death due to improper faste n ing of threaded rods!
If the threaded rods are not screwed in deeply enough into the
ceiling plate, the load-bearing capacity of the structure will not be
guaranteed and the load-bearing system could collapse!
Therefore, always use the entire thickness of the ceiling plate
when screwing in the threaded rods!
The ceiling tube is now mounted onto the threaded rods M12 of the ceiling anchor plate. With an
intermediate ceiling construction, the ceiling tube is mounted onto the threaded rods M12 that were screwed
into the carrier plate.
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above. This is especially the case if
not all of an extension arm's six mounting screws are correctly
tightened in sequence!
Mount the silicone retaining ring before the extension arms!
Before mounting the extension arms, the silicone retaining ring
that accommodates the ceiling tube cover must be correctly
mounted, “Mounting the ceiling tube for Sim.FLEX“, page 31.
This is no longer possible after mounting the extension
arms!
If a monitor mount is included in the lighting system, the central
axis for the corresponding extension arm is delivered together
with a stop piece. When mounting the ceiling tube, it is important
to ensure that the range of rotation of this extension arm is as
desired. To this end, the ceiling tube can be turned in steps of
60° before mounting. During mounting, the central axis allows
additional adjustments of the range of rotation of 120° each,
“Mounting the round central axis“, page 33.
Mounting the threaded rods for the ceiling tube in insulated
form
The threaded rods of the ceiling tube are electrically insulated
from the ceiling tube (10) with the help of insulating washers (9).
To this end, the insulating washers (9) must be mounted as
demonstrated here: Their sleeves are inserted in the holes in the
ceiling tube.
Screw a nut (7) onto each of the six threaded rods M12 (6).
Align the nuts horizontally using a spirit level.
Insert a washer (8) and an insulating washer (9) with the
sleeve facing downwards onto each of the six threaded rods
M12 (6) and then secure together with the ceiling tube
against falling. Here, the sleeves of the insulating washers (9)
must slide from above into the holes in the ceiling tube flange
without diff ic ulty.
Temporarily secure every second threaded rod M12 (6) with
a nut M12 (7).
The ceiling tube (10) is thus secured against falling.
Apply one insulating washer (9) with the sleeve facing
upwards, one washer (8) and one spring was her (11) onto
each of the three free threaded rods, and secure with a nut
M12 (7). Here, the sleeves of the insulating washers must
slide upwards into the holes in the flange without difficulty.
Loosen and remove the temporary nuts without washers.
Apply one insulating washer (9) with the sleeve facing
upwards, one washer (8) and one spring washer (11) onto
each of the three freed threaded rods, and secure with a nut
M12 (7). Here, the sleeves of the insulating washers must
slide upwards into the flange's boreholes without difficulty.
Check the ceiling tube (10) with the spirit level and
realign if necessary. To do so, readjust the nuts M12 (7) as
needed.
Alternately tighten the nuts M12 (7) on the bottom side of the
ceiling tube (10) crosswise with 45 Nm; at the same time,
tighten the respective nuts M12 (7) on the top.
Connect the ceiling tube to the customer-supplied
equipotential bonding using the M6 cylinder screw (1)
that is supplied for this purpose. Here, use a cable with a
cross-section ≥ 6 mm². In addition, adhere to the
effective regulations for installation in the building (e.g.
VDE 0100-710).
Mounting the ceiling tube for Sim.FLEX
Mount the Sim.FLEX system's ceiling tube flange (1) on the six
threaded rods in exactly the same way as described above for the
flange of the round ceiling tube Ø 125 mm.
Align the Sim.Flex extension arm vertically on mounting surface
using a spirit level.
Slide the silicone retaining ring (3) for the ceiling tube cover onto
the flange (4) of the ceiling tube (2) from below, ensuring that the
ceiling tube cover's groove (5) points upwards. This is no longer possible after mounting the extension arms!
The ceiling tube is now ready to mount the Sim.FLEX extension
arms.
Fit Sim.FLEX adapter to round ceiling tube
The adapter enables the Sim.FLEX system to be mounted on a round ceiling tube.
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above. This is especially the case if
not all of an extension arm's six mounting screws are correctly
tightened in sequence!
Insert the adapter (3) for the Sim.FLEX system into the round
ceiling tube from below, so that its threaded holes are visible
through the holes in the ceiling tube for the countersunk screws
(1).
Secure the adapter in the ceiling tube using the six socket head
countersunk screws (1). Tighten countersunk screws (1) with
torque spanner at 7 Nm.
The ceiling tube is now ready to mount the Sim.FLEX extension
arms.
The extension arms for the Sim.FLEX system are mounted on the
system's ceiling tube one after the other.
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
Variant: Power supply unit mounted in the control
cabinet, not on the ceiling tube
Power supply: Connect central axis plug (3) to the
plug socket (2) of the ceiling tube.
If there are multiple extension arms, the longest
cable is for the highest extension arm and the
shortest cable is for the lowest extension arm.
If present, connect the purple CAN bus cable (7).
If present, connect the black video signal cable
(6).
Variant: Power supply unit mounted onto the
ceiling tube
Insert the cable for the central axis upwards
through the ceiling tube, and connect it
professionally with the power supply unit that is
mounted to the ceiling tube and assigned to the
respective lights.
If there are multiple extension arms, the longest
cable is for the highest extension arm and the
shortest cable is for the lowest extension arm.
Both variants
Slide the central shaft (5) into the ceiling tube (1).
Screw in the M6 countersunk screws (4) and,
using a spirit level, align the central shaft so that
the extension arms are horizontal when extended
in every direction.
Insert half (1) of the canopy cover into the groove (2) and screw
it tight.
Swing the extension arm clockwise until it hits the stop.
Loosely wind the cable around the ceiling tube clockwise in 3
windings (3). Fasten the cables to the perforated strip (5) with
cable ties (4).
Note: Choose the cables’ direction of rotation so that they
cannot be damaged when the extension arm is rotated between
the stop position!
Insert the cable for the central axis upwards through the ceiling
tube, and connect it professionally with the power supply unit
that is mounted to the ceiling tube and assigned to the
respective lights.
Remove the adhesive tape from the silicone retaining ring (1)
and apply the silicone retaining ring (1) to the canopy (2).
Mount second half of the canopy with the corresponding
countersunk screws (3).
The extension arms for the Sim.FLEX system are mounted on the system's ceiling tube one after the other.
Each extension arm is mounted in the same way:
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above. This is especially the case if
not all of an extension arm's six mounting screws are correctly
tightened in sequence!
Mount the silicone retaining ring before the extension arms!
Before mounting the extension arms, the silicone retaining ring
that accommodates the ceiling tube cover must be correctly
mounted, “Mounting the ceiling tube for Sim.FLEX“, page 31.
This is no longer possible after mounting the extension
arms!
With the rotating canopy open, position the extension
arm (5) below the central shaft (1).
Connect the grey power supply cable (4).
If present, connect the purple CAN bus cable (3).
Connect the black video signal cable (2), if present
(Caution: Please observe HD Touch connections)
Connecting the Sim.CAM HD Touch on separate
extension arm
Before screwing on the extension arm, ensure that all six M12
screws can be inserted!
Risk of injury due to a collapsing spring arm!
The locking ring may not be overstretched and must
properly snap into the groove when mounted!
Otherwise, injuries may be caused by falling parts!
The locking ring may not be overstretched and must
properly snap into the groove when mounted!
Otherwise, injuries may be caused by falling parts!
The spring arm, adapter (e.g. handle) and the end
device (e.g. flat screen, surgical lights, etc.) are held
in position by a locking ring in the spring arm.
If the locking ring is overexpanded or incorrectly
mounted during disassem bly or mounti ng, it wil l fall
out of the groove in the spring arm. As a result, the
spring arm, adaption and end device will fall, possibly
tearing out internal electrical supply lines. This can
cause severe injuries:
Mounting
• The locking ring may only be installed by trained
and authorised qualified personnel.
• Suitable circlip pliers with expansion control must
be used.
• The installation instructions and installation order
must be complied with.
• Only use the locking ring removed in the initial
installation one time, or use a new, unus ed locking
ring.
• In service or maintenance cases, a new, unused
locking ring will have to be used.
• Check to make sure the locking ring is seated firmly
according to manufacturer specifications!
pol./5 pol./9 pol./TFT) and allo cate the
corresponding spring arm of the central axis or
Sim.Flex to the installation. Incorrect allocation
may damage the central axis and spring ar m!
Precisely align the spring arm with sleeve (5)
horizontally and vertically and insert into the
extension arm as far it will go so that the threaded
holes for the cylinder screws (1) are aligned with
the boreholes in the extension arm.
Fasten the spring arm with the cylinder screws M6
(1) and tighten with torque spanner at 8 Nm.
Insert and tighten the brake screw (2) for the spring
arm “Setting the braking power “, page 112.
Supply variant "Socket pre-mounted on the spring arm"
Remove the brake screw (2) from the end of the
extension arm.
Place the spring arm pivot with sleeve (5) in the
extension arm (3) in such a manner, that the
threaded holes are aligned with the holes (1) in
the extension arm.
4.1.8 Mounting adapter on spring arm / mounting spring arm on Sim.FLEX
extension arm
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
Risk of injury due to a collapsing spring arm!
The locking ring may not be overstretched and must
properly snap into the groove when mounted!
Otherwise, injuries may be caused by falling parts!
The locking ring may not be overstretched and must
properly snap into the groove when mounted!
Otherwise, injuries may be caused by falling parts!
The spring arm, adaption (e.g. handle) and the end
device (e.g. flat screen, surgical lights, etc.) are held
in position by a locking ring in the spring arm.
If the locking ring is overexpanded or incorrectly
mounted during disassem bly or mounti ng, it wil l fall
out of the groove in the spring arm. As a result, the
spring arm, adaption and end device will fall, possibly
tearing out internal electrical supply lines. This can
cause severe injuries:
• The locking ring may only be installed by trained
and authorised qualified per s onnel.
• Suitable circlip pliers with expansion control must
be used.
• The installation instructions and installation order
must be complied with.
• Only use the locking ring removed in the initial
installation one time, or use a new, unused locking
ring.
• In service or maintenance cases, a new, unused
locking ring will have to be used.
• Check to make sure the locking ring is seated firmly
according to manufacturer specifications!
Insert the sleeve (5) on the pivot of the spring arm (12).
Mount the washers (7) - the pivot groove for the locking ring
(6) remains free in the process.
Mount locking ring (6) according to manufacturer’s
specifications!
Identify the spring arm by the product label (3 pol./5
pol./9 pol./TFT) and allocate the corresponding Sim.Flex
spring arm to the installation. Inc o rrect allocation may
damage the slip ring, Sim.Flex extensi o n ar m and spring
arm.
Precisely align the spring arm with sleeve (5) horizontally
and vertically and insert into the extension arm as far it will
go so that the threaded holes for the M8 cylinder screws (4)
are aligned with the boreholes in the extension arm.
Tighten the spring arm using the M8 cylinder screws (4) in
the extension arm, using a torque spanner set to 20 Nm.
Insert and tighten the brake screw (5) for the spring arm
“Setting the braking power“, page 112.
Replace the cap (1).
Insert the retaining screw (2) into the extension arm cap (1)
4.1.9 Mounting spring arm adapter (optional delivery variant)
Mounting
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
The sleeve is installed in the adapter upon delivery.
Identify the adapter by the product label (3 pol./5 pol./9
pol./TFT) and allocate the corresponding spring arm of the
central axis or Sim.Flex to the installation. The adapter can be
installed with normal central axes as well as the Sim.Flex
extension arm. Incorrect allocation may damage the adapter,
central axis/Sim.Flex and spring arm.
Remove the sleeve installed in the bottom side of the adapter.
To do so, remove the mounting screws (marked blue) and
brake screws (marked blue). The 2nd brake screw is offset by
180° on the back side. Do not turn the red marked screws!
Carefully remove the sleeve installed in the adapter bottom
side in the axial direction without using tools. In doing so,
make sure not to touch or damage the inner slip ring!
Align the spring arm with mounted sleeve so that it is
absolutely horizontal and vertical and correctly positioned
based on the holes and carefully insert into the adapter as
far as it will go.
Screw in the M8 screws and tighten at 20 Nm with a torque
spanner.
Screw brake screws in on the front and back side and set
brake force as needed. Turn the brake screw in a clockwise
direction, using the Allen key, to increase the braking power.
Turn the brake screw counterclockwise to decrease it.
V. 1.7 47
Risk of injury due to incorrect mounting!
Incorrect or unsafe mounting of the assembled device may lead
to life-threatening situations and cause significant material
damage. Thus, please note:
−The adapters are not meant to be attached to one another;
instead, only one is to be attached to each central axis
extension arm.
4.1.10 Mounting the light heads onto the spring arm
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
The safety sleeve must be attached and screwed tightly!
Note: When placing the pivot in the spring arm, ensure that the
turning connector (5) is not damaged!
For this purpose, remove the turning connector's protective
cap only shortly before insertion, and carefully align the pin
parallel to the LC spring arm.
A damaged turning connector cannot be used and must be
disposed of.
Bring the spring arm (1) into position.
Loosen and remove the sleeve's retaining screw (6).
Slide the safety sleeve (5) upwards and af f ix .
Remove safety piece (2).
Insert the cardanic's (4) pins carefully and vertically into the
spring arm (1), until the pin's groove can be located in the
safety piece's opening (3).
Insert the safety piece (2) and slide the safety sleeve (5)
downwards.
Screw tightly the sleeve's retaining screw (6).
The light head is safely connected to the spring arm (1).
4.1.11 Mounting and connecting the power supply unit for surgical lights to the
ceiling tube
Danger of electric shock!
A switch is to be provided on-site as a disconnector from the
mains power supply which disconnects the system from the
mains power supply at all poles simultaneously This switch must
meet the creepage and air distance requirements set forth in IEC
61058-1 for a mains voltage peak of 4 kV. The activation device
for this switch must meet the requirements of IEC 60447.
If the switch is not visible to the maintenance personnel, it will
have to be possible to lock it in the OFF position.
The locking mechanism may be located in the mains power
supply switch of the organisation in charge.
The sample image shows a switching power supply mounted
on a ceiling tube. The switching power supply can also be
mounted externally in a control cabinet.
Switch off the building's power supply and safeguard
against restarting!
Screw in the power supply unit (2) using 2 screws M8 (3),
with spring washer and washer on the threaded holes
intended for this purpose, to the ceiling tube's (1) flange.
V. 1.7 49
Applied voltage may damage or destroy the components to
be installed!
Disconnect the on-site power supply from the mains and secure it
to prevent reconnection when
Approved cable lengths for 28V DC direct-current voltage from power supply to central axis (max.
Maximum power: 130 watts
Cable length:
Cable cross-section:
Mounting
Connect a Ø 2.5mm2 cable to Line in (1) of the on-site power supply:
Caution: “Safety Power in” is not controlled. High voltages outside of the toleran ce will lead to
defects or destroy the connected devices!
Line out
voltage drop 5%):
Ambient temperature for approved cable lengths: max. 50° Celsius
≤ 18 metres
> 18 metres and ≤ 30 metres
> 30 metres and ≤ 45 metres
In order to safeguard against the possibility of a lighting system outage,
every switching power supply on site must be powered by a separate feed
line!
Each light must be supplied with a separate switching power supply.
Each camera on a separate arm must be supplied with a separate
switching power supply.
The Line out outputs (2) are only used to supply an optional, medically
approved display (with 100 – 240V AC, if possessing country- / regionspecific approval).
Risk of injury due to voltage fluctuations!
The mains power supply must satisfy overvoltag e category II in
terms of the mains transients. Voltages above 110 % or below 90 %
of the voltage's nominal value may not arise between arbitrary
conductors of the system or between one of these conductors and
earth. The client must also offer Simeon products protection against
excessively high start-up currents.
Next, affix the ferrite core (2) to the Line in (1) mains supply
line. To do so, place the ferrite core around the cable and
allow the ferrite core's snap lock to snap into place. If the cable
is not securely clamped by the ferrite core, secure the ferrite
core in position (3) against sliding using a cable tie.
Make sure that the ferrite core is not mounted at a distance
greater than 40 mm from the terminal (1).
The ferrite core (2) can be dismounted and readjusted using a
If Safety Power in (3) is not supplied with power, Safety Power out (4) will also have no power.
Connect the purple cable from the central axis to the terminal block (7) for CAN communication with
the light:
White:
Black:
-
Connect the RJ45 patch cable from the wall control unit to Wall Control (6) for communication with
the wall control unit.
4.1.12 Installing the Sim.INTERFACE
The Sim.INTERFACE requires 1 installation place on the ceiling tube.
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
The safety sleeve must be attached and screwed tightly!
Danger of electric shock!
The on-site supply line must be disconnected from the mains on
site before further work may be performed! The supply line must
be secured to prevent reconnection.
Power supply and wall controls are connected to the CAN1, CAN2 or CAN3 of the Sim.INTERFACE by one
connection cable each. The interface is connected to an interface card such as OR1 with its RS232 interface
(4) in an integrated OR. This integrated OR is then used to control the surgical lights and cameras.
If further surgical lights are connected, repeat the steps described above. The connection lines must be
plugged into CAN2 and/or CAN3 at the Sim.INTERFACE.
Plug the patch cable (2) on the power supply into
the RJ45 socket Wall Control (1).
Plug the patch cable (2) on Sim.INTERFACE into
the RJ45 socket CAN1 In
correct CAN connection is assigned to CAN1!
CAN2 and CAN3 are intended for additional lights.
(4). Make sure that the
Plug the patch cable for wall control of light 1 on
Sim.INTERFACE into the RJ45 socket
CAN1 Out (3). CAN2 and CAN3 are intended for
additional lights.
If a wall control is not present, close the slot with a
dummy plug and only plug in the cable for the
light.
Danger of electric shock!
The on-site supply line must be disconn ected from the
mains on site before further work may be performed! The
supply line must be secured to prevent re connection.
If the wall controls are configured for multiple lights, the camera
must be positioned on light 2; otherwise, camera control will not
be possible!
One power supply will be needed for each light head. In this
example, 2 of them are connected to the wall controls.
Disconnect the on-site supply line from the mains and
safeguard against reconnection.
Observe the minimum installation depth of 60 mm!
Ensure that the supply line remains disconnected
from the mains until the wall controls have been
mounted.
Pull the patch cable from the Sim.INTERFACE and/or
from power supply of light 1 at the wall controls’
installation point and plug into the RJ45 socket (1) of
the light.
Note: If more than one light is connected, then plug the
patch cable of the other lights into their RJ45 sockets (2).
The CAN connection which is not allocated to any light (3)
remains empty.
Surface mounting
Fasten the housing (5, illustration similar) to the base
with four screws (2) and, if necessary, distance
sleeves (1) so that the connection cables (RJ45) end
in the housing. Borehole spacing: 80 x 210 mm
Connect connection cables to the correct RJ45
socket; see above.
Fasten the wall control unit (4) to the housing with
If the cables are run to the housing externally, they can be
run to the housing through the cable feedthrough (3):
Unscrew the nut (4) and remove the cover (6).
Tighten counter nut (3).
Place nut (4) back on.
First, run the cable with the RJ45 plugs into the
housing (1) without strain relief (5).
Unscrew nut (4) of cable feedthrough (3) and slide it
a bit away from the housing (1) on the cable.
Open pliers (8).
Insert cable (7) between nut (4) and cable feedthrough
(3) into the large opening (9) in the pliers (8).
Close pliers (8) and plug their tip into the feedthrough
intended for the cable (see arrow).
Open pliers (8) wide enough for the cable to be pushed
into the feedthrough of the strain relief (5).
Push cable into the feedthrough of the strain relief (5),
see arrow.
Slide strain relief (5) on the cable into the cable
feedthrough (3).
Guide nut (4) back and tighten it until the strain relief is
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
Wear a protective helmet.
Wear safety boots.
Risk of injury due to pinching!
Multiple parts must be assembled together or mounted onto one
another. As a result, injuries may occur due to pinching.
Proceed cautiously during as s em bl y.
Wear safety boots
The figure shows the canopy on a round ceiling tube.
The Sim.FLEX system's canopy has a larger central aperture of 200
mm and is mounted onto the ceiling tube cover.
Attach retaining ring (3) to ceiling tube (4).
Mount the two half-shells of the canopy (2).
Attach silicone seal (1).
Slide canopy upwards and fasten it with retaining ring (3) (Torx
4.1.16 Mount rotating canopy to Sim.Flex extension arm and normal central axis
Secure on the first half-shell of the rotating canopy with
two Allen screws.
Insert the cable for the central axis upwards through the
ceiling tube, and connect it professionally with the power
supply unit that is mounted to the ceiling tube and
assigned to the respective lights.
Place the cable around the ceiling tube in three
windings to ensure enough reserve cabling for the
extension arm's rotational range. Choose the cables’
direction of rotation so that they cannot be damaged
when the extension arm is rotated between the stop
position!
Secure the rotating canopy's second half-shell (7) to the
bracket and the first half-shell, using two Phillips head
screws (6) in each case.
4.2.2 Mounting the Sim.CAM HD TOUCH onto the camera mount
Disconnect the power suppl y.
Do not touch the sub-D plug (3) in order to protect
from electrostatic discharge!
Plug the D-sub connector (3) found on the rear of the
camera unit into the socket (5) on the camera mount. In
doing so, press in the entire unit firmly and uniform l y.
Tighten the camera unit with the knurled screws (1). To do
so, turn the knurled screws in a clockwise direction. When
the knurled screws have been tightened, the position of
the unit must be checked by uniformly pressing on it once
again.
Slide the sterilisable handle (6) onto the camera unit and
turn until the safety snaps audibly into position.
Mounting
The camera unit is ready to use.
Unsterile positioning may be performed using the
handle (4) on the back side of the bracket, while
sterile positioning can be performed using the
sterilisable handle (6). Changing the sterilisable
handle,Sim.LED Instructions for Use.
Connecting the power supply , S im.CAM HD Touch in light head
Danger of electric shock!
The on-site supply line must be disconnected from the mains on
site before further work may be performed! The supply line must
be secured to prevent reconnection.
Damage or destruction from overvoltage!
The contract unit of the HD Touch may exclusively be
connected and operated with 28 V DC at the specified “Light
Control” of the terminal of the switching power supply. A
100-240 V connection will immediately destroy the device!
Connect the power supply
cable (1) of the camera's
control unit to the power
supply, as shown at left:
Connect the control lines (3) of the light to the right. In this case, the PE terminal must be used for
control lines (3, grounding shield) and camera control unit (1, GNYE) together, as shown in the photo
on the right.
If wall controls are present, they are to be connected to the RJ45 socket,
“Mounting and connecting the power supply unit for surgical lights to the ceiling tube“, page 49.
Connecting the power supply , S im.CAM HD Touch on separate arm
Damage or destruction due to incorrect connection!
The control device of the HD Touch may only be connected
and operated with the switching power supply which is also
connected to the associated ligh t with video camera and
corresponding video line from the central axis!
The Sim.CAM HD Touch on separate arm is always to be connected to a separate switching power
supply!
Connect power supply
for
Sim.CAM HD TOUCH
to power supply 1 (1, on
the left of the photo).
Connect power supply
for HD TOUCH control
unit to power supply 2
(5, on the right of the
photo).
Connect the control lines (3) of the light to the right. In this case, the PE terminal must be used for
control lines (3, grounding shield) and HD TOUCH control unit (5, GNYE) together, as shown in the
photo on the right.
Connect power supply for the light to power supply 2 (6, on the right of the photo).
Connect wall controls (if present) to RJ45 socket (2).
In order to avoid potential communication interference, connect the negative pole of the power
supplies of the Sim.CAM HD TOUCH (on the left of the photo) and the HD TOUCH control unit (on the
right of the photo) with a black cable (4). This cable may be a maximum of 1.5 m long. This will suffice
on the ceiling tube; with other mounting types, ensure that the two power supplies are suitably close to
one another!
Damage or destruction due to incorrect connection!
The control device of the HD Touch may only be connected
and operated with the switching power supply which is also
connected to the associated ligh t with video camera and
corresponding video line from the central axis!
Connect the power cable to the power socket (3).
Connect the camera cable to the camera socket (1).
Connect the video output signal to HD-SDI Out (1 or 2) or DVI
Out.
Mount the ferrite core (1) to the power cable and secure with a
cable tie (2).
Flush mounting the cable
Pull the cable (2) through the cable cover (1).
Screw on the cable cover's (1) lid.
The terminal board for the Sim.CAM HD is normally mounted in the ceiling flange. If the terminal boar d is
found outside of the laminar flow ceiling, a longer earth wire (braided mains cable, minimum cross-section
10 mm²) must be used.
Danger of electric shock!
The on-site supply line must be disconnected fr o m the
mains on site before further work may be performed! The
supply line must be secured to prevent reconnection.
Electrically connect the metal housing of the terminal
board (4) with the metal housing of the ceiling flange, in
an HF-compatible manner. To do so, use the enclosed,
flexible braided mains cable (1). The braided mains
cable must have a minimum cross-section of 10 mm².
Connect the yellow 20 m patch cable (3) (video input
signal), to the control unit for the Sim.CAM HD.
Connect the other end of the patch cable (3) to the
terminal board (as shown in the image).
Plug the cable with the Phoenix connector (5) into the
terminal board and plug the other end (S-video
connector) into the S-video socket in the central axis.
The plug-in connection (2) and S-video socket of the
central axis must be secured against tensile strain!
Remove cap (2).
Connect video cable in the central axis with the blac k
connection cable. Pull cable through ceiling tube and
extension arm.
The plug-in connection and S-video socket of the central axis
must be secured against tensile strain!
Remove cover (5).
Guide the cables through the extension arm (3) without
placing them under mechanical strain, so they are not
damaged during rotational movement.
Push back approx. 5 cm of cable reserve into the extension
arm.
Pull video cable and then power cable through the extension
arm and spring arm,
Mount rotating canopy to Sim.Flex extension arm and
normal central axis
Connect the Sim.CAM HD according to the Instructions for
Use 100-0010175.
4.3.3 Connecting the Sim.STREAM
Use the cold lead to connect Sim.STREAM (2) to the connection (6) with the Schuko socket.
Connect BNC cable to the BNC socket (3, HD-SDI) of the HD control unit (1) and connect it with the
BNC socket (4, SDI-IN) of the Sim.STREAM (2).
Use the cold lead to connect the HD control unit (1) to the connection (7) with the Schuko socket.
The locking ring may not be overstretched and
must properly snap into the groove when mounted!
Otherwise, injuries may be caused by falling parts!
Only mount locking ring once.
Dismantle locking ring (1), washer (2) and
protective sleeve according to manufacturer’s
assembly instructions.
horizontally and place it in the extension arm as far
as it will go so that the threaded holes for the
cylinder screws are aligned with the holes in the
extension arm.
Mount the spring arm with both cylinder screws (1)
M8.
16 Nm torque.
The cable insertion, final assembly and settings
are to be conducted according to the installation
instructions of the company Ondal Medical
Systems OndaSpace.
4.4.6 Sim.SCREEN monitor mount: Positioning the lateral stop pieces
An extension arm for a monitor mount can be provided for at the lowermost or uppermost position of the
central axis. Since the monitor is connected in wired form, this extension arm must be secured against
overtwisting using at least one stop piece at each joint, to prevent the overextension and breaking of the
cables.
Danger of electric shock!
If a stop piece is forgotten during assembly of a monitor mount,
this may result in a short circuit or, in an extreme case, in electric
shock during operation of the monitor.
Danger of injury from falling parts!
Since mounting the arms requires overhead work, injuries can be
caused by parts falling from above.
Wear a protective helmet.
Wear protective gloves.
The stop piece (2) is firmly mounted at the spring arm joint. The
stop pins (1) may be variably used in the monitor mount. A stop pin
is an integral component of the mounting materials that are
included with every monitor mount “mou nti ng materials“, page
84.
Disconnect the power suppl y.
Place the stop pin (1) on the desired position.
Note: The stop pin must be inserted entirely into the opening;
otherwise, the attachment at the spring arm will be problematic.
When using only one stop pin, the swing equals approx.
330°.
With two stop pins, the movement radius for the monitor
mount can be limited; the minimum angle is 45°.
When turning the extension arm, check whether the desired
range of rotation has been achieved. Otherwise, adjust the
stops.
Chafed cables may produce an electric shock! Lay the cabling
without placing it under mechanical strain and use a bending
radius large enough to avoid damage during canopy rotation.
Only use original parts from the manufacturer!
Remove the cap (2).
Pull cable with video output signal through ceiling tube (1)
and extension arm (3, 4): Black video cable for Sim.CAM SD;
cable for HD control unit for Sim.CAM HD.
Remove cover (5).
Pull cable through ceiling tube and extension arm. Guide the
cables through the extension arm without placing them under
mechanical strain, so they are not damaged during rotational
movement.
Push back approx. 5 cm of cable reserve into the extension
arm (3).
Insert the cables carefully into the joint (6) of the spring arm.
Insert the cabling without placing it under mechanical strain
and use a bending radius large enough to avoid damage
during rotational movements.
Note: If the cable plugs do not fit into the opening between the
extension arm and the spring arm adapter, undo the installation
socket and remove cables from the extension arm.
cables (1) around the ceiling tube in 3 windings to
ensure enough reserve cabling for the extension
arm's rotational range.
Choose the direction of rotation so that the cable
cannot be damaged when the extension arm is
rotated form stop to stop. To do so, it must be
exposed without tensile strain in every position of
the extension arm.
Insert the cables into the aperture (3) on the
extension arm (4) and push through until the
cabling is visible at the front end of the arm.
Insert the cables carefully into the joint (7) of the
spring arm.
Insert the cabling without placing it under
mechanical strain and use a bending radius
large enough to avoid damage during rotational
movements.
Note: If the cable plugs do not fit into the opening
between the extension arm and the spring arm
adapter, undo the installation bushing and remove
cables from the extension arm.
Fit the cover (6) using 2 countersunk screws (5).
Carefully loosen the clamps (7) on both sides with a
screwdriver and slide upwards.
Note: The panelling for the spring arm is disassembled for
delivery.
Insert the cabling through the spring arm's (8) cable inlet
without placing it under mechanical strain, and use a
bending radius large enough to avoid damage during
rotational movements.
Note: Lay the cabling without any crossover.
Slide the clamps (9) downwards and allow to snap into
place. In the process, make sure that the cables are not
pinched.
In addition, secure the cables with cable ties (1) so that
they cannot escape from the cable inlet even when the
spring arm moves.
Note: Cables escaping from the cable inlet can be pinched
and become damaged when the spring arm moves.
A CAN cable that is damaged in this manner may result in a
malfunction, for example.
4.5 Mounting the mobile variants (illustration similar)
Risk of injury due to pinching!
Multiple parts must be assembled together or mounted onto one
another. As a result, injuries may occur due to pinching.
Proceed cautiously during as s em bl y.
Mount with at least 2 trained and authorised technicians!
Mounting
Identify the spring arm by the product label and mount
exclusively with the approve lights. I n correct allocation
may damage the spring arm and the li g h t.
Danger of injury from falling parts!
Incorrect allocation may damage the spring arm and the
light and/or errors during assembly can compromise the
stability of the lighting system, which can cause the system
to collapse, potentially resul ting in extremely serious
injuries!
4.5.1 Installation of mobile var iants report
Insert the conical end of the stand tube (6) into the cart (4) so
that it is correctly positioned and tighten w ith the inc lu ded scr e w
(5) at a torque of 30 Nm.
Note: A rivet is integrated into the base. Insert the rivet into the hole
in the tube. This prevents an incorrect assembly!
Insert the pin of the mobile stand connector
into the corresponding groove of the sleeve
and screw together until it stops.
Then remove the cable tie on the stand tube
without letting it fall into the stand tube.
The connector combination is inserted into the
stand tube and the threaded holes in the
adapter sleeve are aligned exactl y vertica lly
with the holes of the stand tube.
After that, the spring arm is inserted into the
stand tube as far as it will go.
Remove the angle plate (red
arrow) and simultaneousl y slide
the locking plate (1) in the correct
position into the corresponding
groove as far as it will go (green
arrow9.
You must be 100% certain that
the locking plate (1) is mounted
as far as it will go!
If the locking plate (1) is not
mounted as far as it will go, the
light can fall down!
Spring arm snapping up
If the locking plate (1) is not
mounted as far as it will go, the
light can fall down and the spring
arm can snap up!
No one may stand beneath the
components of the spring arm
during the mounting process.
It must then be ensured that the
locking plate (1) is mounted as far
as it will go and has not moved
out before the cover (2) for the
locking plate (1) is mounted.
After that, screw on the cover (2)
and make sure that it is fitted
flush with the adjacent areas.
Mount brake screw and adjust
spring force as necessary.
Multiple parts must be assembled together or mounted onto one
another. As a result, injuries may occur due to pinching.
Proceed cautiously during as s em bl y.
Mount with at least 2 trained and authorised technicians!
First, bring the front covers
together in the correct
position and mount them.
After that, mount the
locking rings on both sides
according to
manufacturer’s
specifications.
Bring the rear covers together in the
correct position and mount them.
Pay attention to make sure that the studs
audibly click into place with the
counterpart on the inside and make sure
that the cover is completely closed all the
way around.
The on-site supply line must be disconnected from the mains on
site before further work may be performed! The supply line must
be secured to prevent reconnection.
Risk of death due to heavy objects!
Parts could fall down while the components are being mounted.
This could cause severe or even fatal injuries.
Spirit level
Torque spanner
Circlip pliers with expansion control
Never attempt to lift heavy objects by yourself.
Allow a structural engineer to inspect and approve the wall,
“Confirmation of static bearing capacity“, page113.
Ensure that the wall surface intended for mounting the wall
bearing is even.
There may be no offset areas in the wall.
Disconnect the on-site supply line (1) from the mains and
safeguard against reconnection.
Ensure that the supply line remains de-energised until
mounting of the light head is completed.
Drill holes (2) into the wall for the stud bolts HSL3G M10/20
in accordance with the specifications of the manufacturer of
the attachment bracket (www.hilti.de) and blow them out
thoroughly.
Risk of death due to improper faste n ing of stud bolts!
The collapse of the wall structure can result in a life-threatening
situation.
In order to mount the wall bearing with through-bolts or stud
bolts, an "Individual case approval" by the competent regional
building authority is mandatorily required.
In the case of unstable, lightweight partition walls, the installation
of a metal rail structure will be absolutely required.
Have the static bearing capacity of the structure tested and
approved by a structural engineer.
Adhere to the specified torque for the stud bolts.
Dispose of used stud bolts.
Do not reuse any used stud bolts.
Unscrew the covers (3, 4) from the wall bearing.
Mount the stud bolts in accordance with the specifications of
the manufacturer of the attachment bracket (www.hilti.de).
Hook the wall bearing's slotted hole (2) onto the stud bolt.
Straighten the wall bearing vertically using a spirit level.
Tighten the nuts for the stud bolts with a 35 Nm torque.