Siglent SSA3032X Plus User Manual

SSA3000X Plus Spectrum Analyzer
SSA3000X-R Real-Time spectrum Analyzer
SVA1000X Spectrum & Vector Network Analyzer
User Manual I
User Manual I

Copyright and Declaration

Copyright
SIGLENT TECHNOLOGIES CO., LTD All Rights Reserved.
Trademark Information
SIGLENT is the registered trademark of SIGLENT TECHNOLOGIES CO., LTD
Declaration
SIGLENT products are protected by patent law worldwide
SIGLENT reserves the right to modify or change parts of or all the specifications or pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding material.
Any way of copying, extracting or translating the contents of this manual is not
allowed without the permission of SIGLENT.
SIGLENT will not be responsible for losses caused by either incidental or consequential in connection with the furnishing, use or performance of this manual as well as any information contained.
Product Certification
SIGLENT guarantees this product conforms to the national and industrial standards in
China as well as the ISO9001: 2008 standard and the ISO14001: 2004 standard. Other international standard conformance certification is in progress.
SIGLENT
II User Manual

Safety Information

General Safety Summary

Carefully read the following safety precautions to avoid any personal injury or damage to the instrument and any products connected to it. To avoid potential hazards, please use the instrument as specified.
Use Proper AC Power Line
Only the power cord designed for the instrument and authorized by the local country should be used.
Ground the Instrument
The instrument is grounded through the protective earth conductor of the power line. To avoid electric shock, please make sure the instrument is grounded correctly before connecting its input or output terminals.
Connect the Probe Correctly.
If a probe is used, do not connect the ground lead to high voltage since it has an isobaric electric potential as the ground.
Look Over All Terminals’ Ratings
To avoid fire or electric shock, please look over all ratings and sign instruction of the instrument. Before connecting the instrument, please read the manual carefully to gain more information about the ratings.
Use Proper Overvoltage Protection
Make sure that no over-voltage (such as that caused by a thunderstorm) can reach the product, or else the operator might be exposed to danger of electrical shock.
Electrostatic Prevention
Operate the instrument in an electrostatic discharge protective area environment to avoid damages induced by static discharge. Always ground both the internal and external conductors of the cable to release static before connecting.
Maintain Proper Ventilation
Inadequate ventilation may cause increasing of the instrument’s temperature, which will eventually
damage the instrument. So keep well ventilated and inspect the intake and fan regularly.
Avoid Exposed Circuit or Components
Do not touch exposed contacts or components when the power is on.
Do Not Operate Without Covers
Do not operate the instrument with covers or panels removed.
SIGLENT
Use proper Fuse.
Use Only the Specified Fuse.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in the air, please keep the surface of the device clean and dry.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or electric shock, please do not operate the instrument in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damage to the device or personal injury, it is important to operate the device away from an explosive atmosphere.
Any parts of the device and its accessories are not allowed to be changed or replaced, other than authorized by the manufacturer or his agent. Make sure not to position the equipment so that it is difficult to operate the disconnecting device..

Safety Terms and Symbols

Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in damage to this product or other property.
Terms on the product:
DANGER Indicates direct injuries or hazards that may happen. WARNING Indicates potential injuries or hazards that may happen. CAUTION Indicates potential damages to the instrument or other property that may happen.
Symbols on the product:
Hazardous protective Earth Warning Test Power Switch Voltage Terminal Ground
SIGLENT
IV User Manual

Measurement Category

Measurement Categories
This analyzer can make measurements in other circuits that are not directly connected to mains.
WARNING
This analyzer can only be used for measurements within its specified measurement
categories. Not to use the product for measurements within other measurement categories, such as CAT II, CAT III, CAT IV. Not to use the equipment for measurements on mains circuits
Measurement Category Definitions Measurement category II is for measurements performed on circuits directly connected to the low voltage installation. Examples are measurements on household appliances, portable tools and similar equipment.
Measurement category III is for measurements performed in the building installation. Examples are measurements on distribution boards, circuit-breakers, wiring, including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and equipment for industrial use and some other equipment, for example. Stationary motors with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary over current protection devices and ripple control units.
SIGLENT

Working Environment

Temperature
Operating: 0 to +40 Non-operation: -20 to +70
Humidity
Under +35: 80% relative humidity +35 to +40: ≤60% relative humidity
WARNING
To avoid short circuit inside the instrument or electric shock,
please do not operate in humid environment.
Altitude
Operating: less than 3 km Non-operation: less than 15 km
Degree of protection IP20
Installation (overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.
WARNING
Make sure that no overvoltage (such as that caused by thunderbolt) can reach the
product, or else the operator might expose to danger of electric shock.
Installation (overvoltage) Category Definitions Installation (overvoltage) category I refers to signal level which is applicable to equipment measurement terminals connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to the corresponding low level.
Installation (overvoltage) category II refers to the local power distribution level which is applicable to equipment connected to the AC line (AC power).
SIGLENT
VI User Manual

Ventilation Requirement

This analyzer uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the analyzer in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
WARNING
Inadequate ventilation may cause temperature increase which would damage the instrument. So please keep the instrument well ventilated during operation and inspect the intake and fan regularly.

General Care and Cleaning

Care
Do not store or leave the instrument in direct sunshine for long periods of time.
WARNING
To avoid damages to the instrument or probe, please do not leave them in fog, liquid, or solvent.
Cleaning
Please perform the following steps to clean the instrument and probe regularly according to its operating conditions.
1. Disconnect the instrument from all power sources, and then clean it with a soft wet cloth.
2. Clean the loose dust on the outside of the instrument and probe with a soft cloth. When
cleaning the LCD, take care to avoid scarifying it.
WARNING
To avoid damages to the surface of the instrument and probe, please do not use any corrosive liquid or chemical cleanser.
WARNING
Make sure that the instrument is completely dry before restarting it to avoid short circuits or personal injuries.
SIGLENT

L'information de sûreté

Résumé général de sûreté

Lisez soigneusement les mesures de sécurité suivantes pour éviter n'importe quelles blessures ou les dommages à l'instrument et à tous les produits se sont reliés à eux. Pour éviter des risques, utilisez svp l'instrument comme indiqué.
Employez la ligne à haute tension appropriée
Seulement le cordon de secteur conçu pour l'instrument et autorisé par le pays local a pu être employé.
A rectifié l'instrument.
L'instrument est fondu par le conducteur protecteur de terra de la ligne à haute tension. Pour éviter la décharge électrique, le conducteur moulu doit être relié à la terre. Assurez-vous que l'instrument est fondu correctement avant de relier ses bornes d'entrée ou de rendement.
Reliez le fil de signal correctement.
Le potentiel de l'au sol de fil de signal est égal à la terre, ainsi ne relie pas le fil de signal à une tension.
Regardez estimations au-dessus de toutes les bornes des'
Pour éviter le feu ou la décharge électrique, regardez svp au-dessus de toutes les estimations et instruction de signe de l'instrument. Avant de relier l'instrument, lisez svp le manuel soigneusement pour obtenir plus d'informations sur les estimations.
Employez la protection appropriée de surtension
Assurez-vous qu'aucune surtension (comme cela provoqué par un orage) ne peut atteindre le produit, ou bien l'opérateur pourrait exposer au danger du choc électrique.
Empêchement électrostatique
Fonctionnez dans un environnement protecteur de secteur de décharge électrostatique pour éviter des dommages induits par décharge statique. A toujours rectifié les conducteurs internes et externes du câble pour libérer la charge statique avant de se relier.
La ventilation insatisfaisante de ventilation
bonne de subsistance peut causer l'augmentation de la température, qui endommagera par la suite l'instrument. Gardez ainsi la ventilation bonne et inspectez la prise et éventez régulièrement.
Évitez le circuit ou les composants exposés
ne touchent pas les contacts ou les composants exposés quand le courant passe.
SIGLENT
VIII User Manual
Employez l'utilisation appropriée
de fusible seulement le fusible indiqué.
Ne fonctionnez pas sans couvertures
n'actionnent pas l'instrument des couvertures ou des panneaux étant coupés.
Ne fonctionnez pas avec des échecs suspectés.
Si vous suspectez les dommages se produisent à l'instrument, l'ont inspecté par le personnel de service qualifié avant d'autres opérations. N'importe quel entretien, ajustement ou remplacement particulièrement aux circuits ou aux accessoires doivent être exécutés par le personnel autorisé par SIGLENT.
Ne fonctionnez pas en conditions humides.
Afin d'éviter de court-circuiter à l'intérieur du dispositif ou de la décharge électrique, svp ne fonctionnez pas dans un environnement humide.
Do Not Operate in an Explosive Atmosphere.
Ne fonctionnez pas dans une atmosphère explosive. Afin d'éviter d'endommager le dispositif ou les blessures, il est important d'utiliser le dispositif loin à partir d'une atmosphère explosive.
Maintenez les surfaces de produit propres et sèches.
Pour éviter l'influence de la poussière et/ou de l'humidité en air, maintenez svp la surface du dispositif propre et sèche.
En manipulant la sûreté
manipulez svp avec soin pendant le transport pour éviter d'endommager des boutons, des interfaces de bouton et d'autres parties sur les panneaux.
Le corps ou l'opérateur responsable devrait se référer au manuel d'instruction pour préserver la protection se permettent par l'équipement. Si l'équipement est utilisé en quelque sorte non indiqué par le fabricant, la protection fournie par l'équipement peut être altérée.
On ne permet à aucune pièce du dispositif et de ses accessoires d'être changé ou remplacé, autre qu'autorisé par le fabricant ou son agent. Pas placez l'équipement de sorte qu'il soit difficile d'utiliser le dispositif débranchant (prise détachable).
SIGLENT

Limites et symboles de sûreté

Limites en ce manuel. Ces limites peuvent apparaître en ce manuel :
Les rapports
d'avertissement D'AVERTISSEMENT indiquent les conditions ou les
pratiques qui pourraient avoir comme conséquence les dommages ou la
perte de la ie.
Les rapports
d'attention d'ATTENTION indiquent les conditions ou les pratiques qui pourraient avoir comme consequenceles dommages à ce produit ou à
toute autre propriété.
Limites sur le produit. Ces limites peuvent apparaître sur le produit :
Le DANGER indique les dommages ou les risques directs qui peuvent se produire. Dommages ou risques potentiels de WARNINGIndicates qui peuvent se produire.
L'ATTENTION indique des dommages potentiels à l'instrument ou à toute autre propriété qui peuvent se produire.
Si la trouvaille de tels symboles sur le produit, consultent le manuel pour découvrir la nature du risque et des actions qui doivent être pris.
Symboles sur le produit. Ces symboles peuvent apparaître sur le produit :
Dangereux Protecteur Avertissement Châssis Puissance Tension Au sol de la terre Ground Switch
SIGLENT
X User Manual

Catégorie de mesure

Les analyzers peuvent faire des mesures dans d'autres circuits qui ne sont pas directement reliés aux forces. Pour ne pas employer le produit pour des mesures dans d'autres catégories de mesure, telles que le CAT II, CAT III, CAT IV. Ne pas utiliser l'équipement pour des mesures sur des forces circuite, pour ne pas utiliser l'équipement pour des mesures sur la tension excèdent la gamme de tension décrivent dans le manuel.
Les rapports
cet analyzers peut seulement être employé pour des mesures dans
ses catégories indiquées de mesure.
La catégorie II de mesure
de définitions de catégorie de mesure est pour des mesures effectuées sur des circuits directement reliés à l'installation de basse tension. Les exemples sont des mesures sur des appareils électroménagers, des outils portatifs et l'équipement semblable.
La catégorie III de mesure est pour des mesures effectuées dans l'installation de bâtiment. Les exemples sont des mesures sur des conseils de distribution, des disjoncteurs, le câblage, y compris des câbles, des barres omnibus, des boîtes de jonction, des commutateurs, des douille-sorties dans l'installation fixe, et l'équipement à l'utilisation industrielle et à un autre équipement,
par exemple. Moteurs stationnaires avec le raccordement permanent à l'installation fixe. La catégorie IV de mesure est pour des mesures effectuées à la source d'installation de basse tension.Les exemples sont des mètres et des mesures de l'électricité sur les dispositifs de protection d'excédent primaire et les unités de commande courants d'ondulation.
SIGLENT

Environnement de fonctionnement

Température
En fonctionnement: 0 à + 40 Non-fonctionnement: -20 à + 70
Humidité
Moins de + 35: humidité relative ≤80% + 35 à + 40 : humidité relative ≤60%
EN AVERTISSANT
d'éviter le court-circuit à l'intérieur de l'instrument ou de la décharge électrique, svp ne
fonctionnez pas dans l'environnement humide.
Opération
d'altitude : moins de 3 kilomètres de non-fonctionnement : moins de 15 kilomètres
Degré de protection
IP20
La catégorie d'installation (surtension)
ce produit est actionnée par des forces conformément à la catégorie II. d'installation (surtension).
EN AVERTISSANT
assurez-vous qu'aucune surtension (comme cela provoqué par coup de
foudre) ne peut atteindre le produit, ou bien l'opérateur pourrait
exposer au danger de la décharge électrique.
La catégorie II d'installation de définitions de catégorie d'installation (surtension) (surtension) se rapporte au niveau local de distribution d'énergie qui est applicable à l'équipement relié à la ligne à C.A. (courant alternatif).

Condition de ventilation

This analyzer uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the analyzer in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation.
Cet analyzer utilise le ventilateur pour forcer le refroidissement. Veuillez s'assurer que les secteurs d'entrée et d'échappement d'air sont exempts des obstructions et ont l'air libre. À l'aide
SIGLENT
XII User Manual
de l'analyzer dans un mettre hors jeu-dessus ou un arrangement de support, fournissez au moins le dégagement de 10 centimètres près, au-dessus et derrière de l'instrument pour à ventilation proportionnée.
La ventilation insatisfaisante peut causer l'augmentation de la température qui endommagerait l'instrument. Veuillez ainsi la subsistance l'instrument bien aéré lors du fonctionnement et inspectez la prise et éventez régulièrement.

Soin général et nettoyage

Ne stockez pas ou ne laissez pas l'instrument en soleil direct pendant de longues périodes.
Pour éviter d'endommager l'instrument, svp ne les laissez pas dans le brouillard, le liquide, ou le dissolvant.
nettoyage
Veuillez exécuter les étapes suivantes pour nettoyer l'instrument régulièrement selon ses conditions de fonctionnement.
1. Démontez l'instrument de toutes les sources d'énergie, et puis nettoyez-le avec un tissu humide mou.
2. Nettoyez la poussière lâche sur l'extérieur de l'instrument avec un tissu mou. En nettoyant l'affichage à cristaux liquides, salut pour éviter de le scarifier.
Pour éviter d'endommager la surface de l'instrument, svp n'utilisez aucune épierreuse corrosive de liquide ou de produit chimique.
Assurez-vous que l'instrument est complètement sec avant de le remettre en marche pour éviter des courts-circuits ou des blessures.
SIGLENT

Allgemeine Sicherheitshinweise

Um Verletzungen oder Schäden am Gerät und den damit verbundenen Produkten zu vermeiden, lesen Sie die folgenden Sicherheitshinweise sorgfältig durch. Um mögliche Gefahren zu vermeiden, verwenden Sie bitte das Gerät wie angegeben.
Wartungsarbeiten an diesem Gerät dürfen nur von qualifiziertem Reparaturpersonal durchgeführt werden.
Verwendung eines geeigneten Netzkabels
Verwenden Sie nur das für das Gerät vorgesehene und von den örtlichen Behörden zugelassene Netzkabel.
Erdung des Gerätes
Das Gerät ist über den Schutzleiter des Netzkabels geerdet. Um einen elektrischen Schlag zu vermeiden, vergewissern Sie sich bitte, bevor Sie seine Eingangs- oder Ausgangsklemmen anschließen, dass das Gerät korrekt geerdet ist.
Anschluss der Signalleitung
Die Masse der Signalleitung ist auf dem gleichen Potential wie die Erdung des Gerätes. Schließen Sie daher die Masseleitung nie an eine Hochspannung an. Berühren Sie keine freiliegenden Kontakte oder Komponenten.
Verschaffen Sie sich einen Überblick über alle Anschluss-Ratings
Um Brände oder Stromschläge zu vermeiden, lesen Sie bitte alle Bemerkungen und Anweisungen des Gerätes. Bevor Sie das Gerät anschließen, lesen Sie bitte das Handbuch sorgfältig durch, um weitere Informationen über die Nennleistungen zu erhalten.
Betreiben Sie kein Gerät, bei dem der Verdacht besteht, dass es defekt ist.
Wenn Sie vermuten, dass das Produkt beschädigt ist, lassen Sie es bitte von qualifiziertem Fachpersonal überprüfen. Vermeiden Sie den Kontakt mit freiliegenden Stromkreisen oder Drähten. Berühren Sie keine freiliegenden Kontakte oder Komponenten, wenn das Gerät eingeschaltet ist.
Verwenden Sie einen geeigneten Überspannungsschutz
Stellen Sie sicher, dass keine Überspannung (z.B. durch ein Gewitter) auf das Produkt gelangen kann, da sonst der Bediener der Gefahr eines Stromschlags ausgesetzt sein kann.
Nicht ohne Abdeckungen betreiben
Betreiben Sie das Gerät nicht bei abgenommenen Abdeckungen oder Platten.
Halten Sie die Oberfläche des Gerätes sauber und trocken. Nicht unter nassen/feuchten Bedingungen betreiben. Betreiben Sie das Gerät nicht in einer explosiven Atmosphäre. Vermeiden Sie es, dieses Gerät einem Feuer auszusetzen.
SIGLENT
XIV User Manual

Sicherheitshinweise und Symbole

In diesem Produkt verwendete Begriffe. Diese Begriffe können im Produkt erscheinen:
GEFAHR weist auf eine mögliche, sofortige und direkte Verletzung oder Gefährdung hin. WARNUNG weist auf möglicherweise auftretende Verletzungen oder Gefahren hin VORSICHT weist auf möglicherweise auftretende Schäden am Gerät oder anderen
Gegenständen hin
In diesem Produkt verwendete Symbole. Diese Symbole können auf dem Produkt erscheinen:
Gefährliche Schutzerde Sicherheitswarnung Erdung Netzschalter Spannung
SIGLENT
CONTENTS
COPYRIGHT AND DECLARATION ........................................................................................................................ I
SAFETY INFORMATION ..................................................................................................................................... II
GENERAL SAFETY SUMMARY ...................................................................................................................................... II
SAFETY TERMS AND SYMBOLS .................................................................................................................................... III
MEASUREMENT CATEGORY ....................................................................................................................................... IV
WORKING ENVIRONMENT .......................................................................................................................................... V
VENTILATION REQUIREMENT ..................................................................................................................................... VI
GENERAL CARE AND CLEANING .................................................................................................................................. VI
L'INFORMATION DE SÛ RETÉ ........................................................................................................................... VII
RÉSUMÉ GÉNÉRAL DE SÛ RETÉ ................................................................................................................................... VII
LIMITES ET SYMBOLES DE SÛRETÉ ................................................................................................................................ IX
CATÉGORIE DE MESURE.............................................................................................................................................. X
ENVIRONNEMENT DE FONCTIONNEMENT ..................................................................................................................... XI
CONDITION DE VENTILATION ...................................................................................................................................... XI
SOIN GÉNÉRAL ET NETTOYAGE ................................................................................................................................... XII
ALLGEMEINE SICHERHEITSHINWEISE ............................................................................................................. XIII
SICHERHEITSHINWEISE UND SYMBOLE ....................................................................................................................... XIV
CHAPTER 1 QUICK START .......................................................................................................................... 1
1.1 GENERAL INSPECTION ................................................................................................................................... 1
1.2 APPEARANCE AND DIMENSION ....................................................................................................................... 1
1.3 PREPARING FOR USE ..................................................................................................................................... 2
1.3.1 Adjust the Supporting Legs .............................................................................................................. 2
1.3.2 Connect to AC Power Supply ............................................................................................................ 2
1.4 FRONT PANEL .............................................................................................................................................. 3
1.4.1 Front Panel Function Keys................................................................................................................ 3
1.4.2 Front Panel Key Backlight ................................................................................................................ 5
1.4.3 Using the Numeric Keyboard ........................................................................................................... 5
1.4.4 Front Panel Connectors ................................................................................................................... 6
1.5 REAR PANEL ................................................................................................................................................ 7
1.6 USER INTERFACE .......................................................................................................................................... 9
1.7 MODE...................................................................................................................................................... 12
1.8 FIRMWARE OPERATION ............................................................................................................................... 13
1.8.1 Check System Information ............................................................................................................. 13
1.8.2 Load Option ................................................................................................................................... 13
1.8.3 Firmware Upgrade ........................................................................................................................ 13
1.9 REMOTE CONTROL ..................................................................................................................................... 13
1.10 TOUCH OPERATION..................................................................................................................................... 14
1.11 USING BUILT-IN HELP .................................................................................................................................. 14
CHAPTER 2 SPECTRUM ANALYZER MODE .............................................................................................. 15
SIGLENT
XVI User Manual
2.1 BASIC SETTINGS ......................................................................................................................................... 15
2.1.1 Frequency ...................................................................................................................................... 15
2.1.2 Span ............................................................................................................................................... 20
2.1.3 Amplitude ...................................................................................................................................... 22
2.1.4 Auto Tune ...................................................................................................................................... 25
2.2 SWEEP AND FUNCTIONS .............................................................................................................................. 27
2.2.1 BW ................................................................................................................................................. 27
2.2.2 Trace .............................................................................................................................................. 29
2.2.3 Detect ............................................................................................................................................ 31
2.2.4 Sweep ............................................................................................................................................ 32
2.2.5 Trigger ........................................................................................................................................... 34
2.2.6 Limit ............................................................................................................................................... 35
2.2.7 TG (Tracking Generator) ................................................................................................................ 37
2.2.8 Demod ........................................................................................................................................... 39
2.3 MARKER ................................................................................................................................................... 41
2.3.1 Marker ........................................................................................................................................... 41
2.3.2 Marker -> ....................................................................................................................................... 44
2.3.3 Marker Fn ...................................................................................................................................... 45
2.3.4 Peak ............................................................................................................................................... 47
2.4 MEASUREMENT ......................................................................................................................................... 50
2.4.1 Meas .............................................................................................................................................. 50
2.4.2 Meas setup .................................................................................................................................... 51
CHAPTER 3 VECTOR NETWORK ANALYZER ............................................................................................. 62
3.1 USER INTERFACE ........................................................................................................................................ 62
3.2 BASIC SETTINGS ......................................................................................................................................... 63
3.2.1 Frequency ...................................................................................................................................... 63
3.2.2 Span ............................................................................................................................................... 64
3.2.3 Amplitude ...................................................................................................................................... 65
3.3 SWEEP AND FUNCTIONS .............................................................................................................................. 66
3.3.1 BW ................................................................................................................................................. 66
3.3.2 Trace .............................................................................................................................................. 67
3.3.3 Sweep ............................................................................................................................................ 69
3.3.4 TG .................................................................................................................................................. 70
3.4 MARKER ................................................................................................................................................... 70
3.4.1 Marker ........................................................................................................................................... 70
3.4.2 .Peak .............................................................................................................................................. 73
3.4.3 Marker Fn ...................................................................................................................................... 74
3.5 MEAS SETUP ............................................................................................................................................. 74
3.5.1 Stimulus ......................................................................................................................................... 74
3.5.2 Meas .............................................................................................................................................. 75
3.5.3 Format ........................................................................................................................................... 75
3.5.4 Scale .............................................................................................................................................. 76
3.5.5 Trace .............................................................................................................................................. 76
3.5.6 Calibration ..................................................................................................................................... 76
SIGLENT
CHAPTER 4 DISTANCE-TO-FAULT MODE .................................................................................................. 79
4.1 USER INTERFACE ........................................................................................................................................ 79
4.2 MEASUREMENT ......................................................................................................................................... 80
4.2.1 Disp Mode ..................................................................................................................................... 80
4.2.2 Start Distance ................................................................................................................................ 80
4.2.3 Stop Distance ................................................................................................................................. 81
4.2.4 Unit ................................................................................................................................................ 81
4.2.5 Velocity Factor ............................................................................................................................... 81
4.2.6 Cable Atten .................................................................................................................................... 82
4.2.7 Window ......................................................................................................................................... 82
4.2.8 Calibration ..................................................................................................................................... 83
CHAPTER 5 MODULATION ANALYZER .................................................................................................... 84
5.1 USER INTERFACE ........................................................................................................................................ 84
5.2 BASIC SETTINGS ......................................................................................................................................... 85
5.2.1 Frequency ...................................................................................................................................... 85
5.3 MEASUREMENT ......................................................................................................................................... 86
5.3.1 Digital Modulation Analysis .......................................................................................................... 86
5.3.2 Analog Modulation Analysis .......................................................................................................... 90
5.4 SWEEP AND FUNCTIONS .............................................................................................................................. 91
5.4.1 Trigger ........................................................................................................................................... 91
5.4.2 Sweep ............................................................................................................................................ 91
CHAPTER 6 REAL-TIME SPECTRUM ANALYZER ........................................................................................ 93
6.1 BASIC SETTINGS ......................................................................................................................................... 93
6.1.1 Frequency ...................................................................................................................................... 93
6.1.2 Span ............................................................................................................................................... 95
6.1.3 Amplitude ...................................................................................................................................... 96
6.2 SWEEP AND FUNCTIONS .............................................................................................................................. 98
6.2.1 BW ................................................................................................................................................. 98
6.2.2 Trace .............................................................................................................................................. 98
6.2.3 Detect .......................................................................................................................................... 100
6.2.4 Sweep .......................................................................................................................................... 100
6.2.5 Trigger ......................................................................................................................................... 102
6.2.6 FMT ............................................................................................................................................. 103
6.3 MARKER ................................................................................................................................................. 104
6.3.1 Marker ......................................................................................................................................... 104
6.3.2 Peak ............................................................................................................................................. 106
6.3.3 Marker-> ...................................................................................................................................... 106
6.4 MEASUREMENT ....................................................................................................................................... 107
6.4.1 Meas ............................................................................................................................................ 107
6.4.2 Meas setup .................................................................................................................................. 111
CHAPTER 7 EMI MEASUREMENT .......................................................................................................... 112
7.1 INTRODUCTION ........................................................................................................................................ 112
SIGLENT
XVIII User Manual
7.2 BASIC SETTINGS ....................................................................................................................................... 114
7.2.1 Frequency .................................................................................................................................... 114
7.2.2 Span ............................................................................................................................................. 115
7.2.3 Amplitude .................................................................................................................................... 116
7.3 SWEEP AND FUNCTIONS ............................................................................................................................ 119
7.3.1 BW ............................................................................................................................................... 119
7.3.2 Trace ............................................................................................................................................ 120
7.3.3 Detect .......................................................................................................................................... 121
7.3.4 Sweep .......................................................................................................................................... 121
7.3.5 Limit ............................................................................................................................................. 123
7.4 MARKER ................................................................................................................................................. 124
7.4.1 Marker ......................................................................................................................................... 124
7.4.2 Marker-> ...................................................................................................................................... 126
7.4.3 Peak ............................................................................................................................................. 126
7.5 MEASUREMENT ....................................................................................................................................... 127
7.5.1 Sequence ..................................................................................................................................... 127
7.5.2 Start / Stop Sequence .................................................................................................................. 128
7.5.3 Scan Config .................................................................................................................................. 128
7.5.4 Search .......................................................................................................................................... 129
7.5.5 Meas ............................................................................................................................................ 129
7.5.6 Signal List..................................................................................................................................... 129
7.5.1 Meter ........................................................................................................................................... 131
CHAPTER 8 SYSTEM SETTINGS .............................................................................................................. 131
8.1 SYSTEM .................................................................................................................................................. 131
8.1.1 Language ..................................................................................................................................... 131
8.1.2 Power On/Preset ......................................................................................................................... 132
8.1.3 Interface ...................................................................................................................................... 132
8.1.4 Calibration ................................................................................................................................... 134
8.1.5 System Info .................................................................................................................................. 134
8.1.6 Date and Time ............................................................................................................................. 135
8.1.7 Self Test........................................................................................................................................ 135
8.2 DISPLAY .................................................................................................................................................. 136
8.2.1 Grid Brightness ............................................................................................................................ 136
8.2.2 Screenshot ................................................................................................................................... 136
8.2.3 Touch Settings ............................................................................................................................. 137
8.2.4 Power Saving ............................................................................................................................... 138
8.2.5 Annotation................................................................................................................................... 138
8.2.6 Display Line ................................................................................................................................. 138
8.3 FILE ....................................................................................................................................................... 139
8.3.1 Browser ....................................................................................................................................... 139
8.3.2 Open/Load ................................................................................................................................... 139
8.3.3 Back ............................................................................................................................................. 139
8.3.4 View Type .................................................................................................................................... 139
8.3.5 Save Type ..................................................................................................................................... 139
SIGLENT
8.3.6 Save ............................................................................................................................................. 140
8.3.7 Create Folder ............................................................................................................................... 140
8.3.8 Operate ....................................................................................................................................... 140
8.4 SHORTCUT KEY ........................................................................................................................................ 141
8.4.1 Preset........................................................................................................................................... 141
8.4.2 Couple .......................................................................................................................................... 151
8.4.3 Help ............................................................................................................................................. 152
8.4.4 Save ............................................................................................................................................. 152
CHAPTER 9 PROGRAMMING OVERVIEW .............................................................................................. 153
9.1 REMOTELY OPERATING THE ANALYZER .......................................................................................................... 153
9.1.1 USB: Connecting the Analyzer via the USB Device port ............................................................... 153
9.1.2 LAN: Connecting the Analyzer via the LAN port .......................................................................... 153
9.1.3 GPIB: Connecting the Analyzer via the USB-Host port ................................................................. 154
9.2 BUILD COMMUNICATION ........................................................................................................................... 155
9.2.1 Build Communication Using VISA ................................................................................................ 155
9.2.2 Build Communication Using Sockets/Telnet ................................................................................ 157
9.3 REMOTE CONTROL CAPABILITIES ................................................................................................................. 157
9.3.1 User-defined Programming ......................................................................................................... 157
9.3.2 Send SCPI Commands via NI MAX ............................................................................................... 157
9.3.3 Easy Spectrum Software .............................................................................................................. 160
9.3.4 Web Control ................................................................................................................................. 161
CHAPTER 10 SERVICE AND SUPPORT...................................................................................................... 162
10.1 SERVICE SUMMARY ................................................................................................................................... 162
10.2 TROUBLESHOOTING .................................................................................................................................. 162
SIGLENT

Chapter 1 Quick Start

This chapter guides users to quickly get familiar with the appearance, dimensions, front/rear panel and the user interface, as well as announcements during the first use of the analyzer.

1.1 General Inspection

1. Inspect the shipping container
Keep the damaged shipping container or cushioning material until the contents of the shipment have been completely checked and the instrument has passed both electrical and mechanical tests. The consigner or carrier will be responsible for damages to the instrument resulting from shipment.
SIGLENT will not provide free maintenance or replacement.
2. Inspect the instrument
If the instrument is found to be damaged, defective or fails in electrical or mechanical tests, please contact SIGLENT.
3. Check the accessories
Please check the accessories according to the packing list in the box. If the accessories are incomplete or damaged, please contact your SIGLENT sales representative.

1.2 Appearance and Dimension

Figure 1-1 Front View
SIGLENT
2 User Manual
Figure 1-2 Top View

1.3 Preparing for Use

1.3.1 Adjust the Supporting Legs

Adjust the supporting legs properly to use them as stands to tilt the analyzer upwards for stable placement as well as easier operation and observation of the instrument display.
Figure 1-3 before adjusting Figure 1-4 after adjusting

1.3.2 Connect to AC Power Supply

The spectrum analyzer accepts 100-240V, 50/60Hz or 100-120V 400Hz AC power supply. Please use the provided power cord to connect the instrument to the power source as shown in the figure below. Before powering on, make sure the analyzer is protected by a fuse.
Figure 1-5 Power Cord Connection
The Plug
Fuse holder
SIGLENT

1.4 Front Panel

Figure 1-6 the Front Panel
Table 1-1 Front Panel Description
NO.
Description
NO.
Description
1
User Graphical Interface, touch support
7
RF Input, VNA port 2
2
Menu Control Keys
8
TG Output, VNA port 1
3
Function Keys
9
3.5 mm Earphone interface
4
Knob
10
USB Host
5
Numeric / Letter Keyboard
11
Power Switch
6
Arrow Keys

1.4.1 Front Panel Function Keys

Figure 1-7 Function Keys area
SIGLENT
4 User Manual
Table 1-2 Function keys description
Control Keys
Description
Frequency
Set the parameters of frequency, and Peak→CF, CF→Step.
Span
Set the parameters of span, and X-scale (Log-Linear) setup.
Amplitude
Set the parameters of amplitude, including Ref Level, Attenuator, Preamp, etc.; and Correction setup.
Auto Tune
Scan the full span rapidly and move the biggest signal to center freq, and automatically sets the optimal parameters according to the signal.
Setting Keys
Description
BW
Set the parameters of RBW and VBW, Average Type (Log power, Power, Voltage), and Filter Type (-3 dB Gauss\ -6 dB EMI).
Trace
Select Trace, Trace setup and Trace math.
Sweep
Set the parameters of sweep, and EMI QPD Dwell Time.
Detect
Select the detector type for each trace independently.
Trigger
Select triggers in Free Trigger, Video Trigger and External Trigger.
Limit
Set the Pass\Fail Limit.
TG
Set the parameters of tracking generator. Including TG Level, TG Level offset Normalization setup. The backlight LED is on when TG source is working.
Demod
Set the demodulation parameters of the AM and FM for audio listening.
Marker Keys
Description
Marker
Set the Markers and Marker Table.
Marker->
Set other system parameters based on the current marker’s value.
Marker Fn
Special functions of the marker such as noise marker, N dB bandwidth measurement and frequency counter.
Peak
Search for the peak signal, peak search configuration and peak table.
Meas Keys
Description
Meas
In spectrum analyzer mode, selects the Advanced Measurement function. In non-spectrum analyzer mode, select corresponding settings.
Meas Setup
Set the measurement parameters.
System Keys
Description
System
Set the system parameters.
Mode
Select the working modes.
Display
Set the display parameters.
File
Use the file system and files.
Shortcut Keys
Description
Preset
Sets the system to certain status.
Couple
Set the parameters of some functions between auto and manual.
Help
Turn on the built-in help.
Save
Save Shortcut Key.
SIGLENT

1.4.2 Front Panel Key Backlight

The on/off state and the color of the backlights of some keys at the front panel indicate the working state of the analyzer. The states are as listed below.
1. Power Switch
Constant on: indicates the instrument is in normal operating state.
2. Mode
When the function is Spectrum Analyzer, the backlight turns off. When in other modes, the backlight turns on.
3. TG
When the TG source is on, the backlight of TG turns on and turns off when the function is off.

1.4.3 Using the Numeric Keyboard

The analyzer provides a numeric keyboard at the front panel. The numeric keyboard supports English uppercase/lowercase characters, numbers and common symbols (including decimal point, #, space and +/-) and are mainly used to edit file or folder names and set parameters.
Figure 1-8 Numeric Keyboards
1. +/- In number input, set the sign of number; in file input, switch in number and letter.
2. 1 A/a In number input, enter number 1; in file input, switch between uppercase and lowercase letter.
3. . # In number input, enter a decimal point. In English input, enter special characters.
4. Back In parameter editing, press this key to delete the character on the left of the cursor.
5. Esc
SIGLENT
6 User Manual
During parameter editing process, press this key to clear the inputs in the active function area and
exit parameter input.
When the instrument is in remote mode, use this key to return to local mode.
6. Enter
In parameter editing, the system will complete the input and insert a default unit for the parameter.

1.4.4 Front Panel Connectors

1 2 3
Figure 1-9 Front Panel Connectors (1)
1. Power Switch
Power on / Power down the instrument
2. USB Host
The analyzer can serve as a “host” device to connect external USB devices. This interface is
available for USB storage devices, the SIGLENT GPIB-USB adapter, wireless or wired mouse and keyboard, or the SIGLENT Ecal electronic calibration module.
Read and write functions for an external USB storage device or store the filesfor instance
contents currently displayed on the screen in the USB storage device in .png or .jpg or .bmp format.
3. Earphone Jack
The analyzer can demodulate AM and FM signals. Insert a 3.5 mm earphone into to the jack to acquire the audio output of the demodulated signal. You can turn on or off the earphone output and adjust the volume via Demod ->Volume.
CAUTION
Protect your hearing. Please turn the volume down to zero before using the earphone. Gradually turn the volume up to a comfortable level after putting in the earphone.
SIGLENT
Figure 1-10 Front Panel Connectors (2)
4. TG SOURCE, VNA PORT 1
The TG SOURCE can be connected to a device-under-test (DUT) through a cable with a male
N-type male connector.
In the VNA mode, this port is used as the single port of S11 and the output port of S21.
CAUTION
To avoid damaging to the tracking generator, the reverse DC voltage cannot exceed 50 V
5. RF INPUT, VNA PORT 2
The RF INPUT can be connected to the DUT through a cable with a male N-type connector
In the VNA mode, this port is used as the input port for S21 measurements.
CAUTION
To avoid damaging to the instrument, he DC voltage component and the maximum continuous power of the AC (RF) signal component cannot exceed 50 V and +30 dBm respectively.

1.5 Rear Panel

Figure 1-11 Rear Panel
SIGLENT
8 User Manual
1. Handle
Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down.
2. USB Device Interface
3. Through this interface, the analyzer can be connected to PC for remote control.
4. LAN Interface
Through this interface, the analyzer can be connected to your local network (LAN) for remote control.
5. REF IN 10 MHz
The analyzer can use the internal or an external reference source.
When a 10 MHz external clock signal is received through the [10 MHz IN] connector, this signal is
used as the external reference source and “Ext Ref” is displayed in the status bar of the user interface. When the external reference is lost or not connected, the instrument switches to its
internal reference source automatically and “Ext Ref” on the screen disappears.
The [10 MHz IN] and [10 MHz OUT] connectors are usually used to build synchronization among
multiple instruments.
6. REF OUT 10 MHz
The analyzer can use the internal or an external reference source.
When an internal reference source is used, the [10 MHz OUT] connector can output a 10 MHz
clock signal generated by the analyzer. This signal can be used to synchronize other instruments.
The [10 MHz OUT] and [10 MHz IN] connectors are usually used to build synchronization among
multiple instruments.
7. Trigger in
In external trigger mode, the analyzer will update the trace scan after the Trigger In connector receives an external trigger signal that meets the trigger input specifications.
8. Security Lock Hole If needed, you can use a security lock(purchased separately) to lock the analyzer to a desired location.
9. AC Power Supply The analyzer accepts 100-240V, 50/60 or 100-120VA 400Hz power supply. Please use the power cord provided as accessories to connect the instrument.
10. Fuse
Before power on, make sure the analyzer is protected by the proper input fuse.
SIGLENT

1.6 User Interface

21
1 2 3 4 5 976 10
17
41 40
38 37
39
36 35 34 33
32 31
30
29
28 27 25 24 23 22 20 19 18
1614
13
1211
8
15
26
42
45
43
44
Figure 1-12 User Interface of spectrum analyzer mode
Table 1-3 User Interface labels of spectrum analyzer mode
NO.
Name
Description
1
SIGLENT
SIGLENT logo
2
Ref
Reference level
3
UNCAL
The sweep time is less than the auto couple time, the measure result may have decreased accuracy
4
EXT REF
Valid Ext 10 MHz reference clock detected indicator
5
Att
Attenuator Value
6
Day and time
System time
7
Pass/Fail status
Limit Pass/Fail status
8
Marker
Current active marker
9
Trace
Active trace
10
Marker instruction
Current marker, touch to open a new marker
11
Marker x value
Unit: frequency, frequency delta or time
12
State indication
Auto Tune: Automatically sets the optimal parameters according to the characteristics of the signal. Waiting for Trigger
13
Marker y value
Amplitude value or amplitude delta value
SIGLENT
10 User Manual
14
USB storage device identification
The identification is displayed when a USB flash drive is inserted 15
Main menu touch logo
Clicking this button will bring up the main menu
16
Menu title
Function of the current menu.
17
Menu items
Menu items of the current function
18
Operation status
LOCAL or REMOTE mode. When REMOTE, keys would be locked.
19
Sweep indication
Indicates the currently scanned frequency position
20
Stop frequency
Stop frequency value
21
Sweep time
Time duration of a single sweep
22
Center frequency
Center frequency value
23
Span
Span value
24
VBW
Video bandwidth
25
Start frequency
The first frequency of a sweep
26
RBW
Resolution bandwidth
27
Manually instructions
When it appears, this parameter is not automatically coupled but manually configured
28
Touch assistant
Click to open the common functions for measurement. Touch Assist can be moved to any position on the screen and turned off in the DISPLAY menu
29 30 31 32
Trace A\B\C\D status
Trace type: C&W: Clear Write MaxH: Max Hold MinH: Min Hold View: View AVG: average.
Detect type: P-PK: Positive peak N-PK: Positive peak Samp: Sample Norm: Normal AVG: average Q-PK: Quasi-peak
33
Correction
Indicates that there is a user-configured amplitude correction table being mathematically applied to the displayed trace data
34
AM or FM
AM or FM demodulation activated
35
PA
Enable or disable the Preamplifier
36
FFT
Sweep mode is FFT
37
Single or Continue
Sweep mode single or continuous
38
Average type
Log power\Power\Voltage power
39
Trigger type
Free\Video\External trigger
40
Ref offset
34:Ref offset identification;35:Ref offset value
41
Scale/Div
Scale value
42
Scale type
Logarithm or linearity
43
Limit line
Limit Pass/Fail level
44
Trigger level
Video trigger level
45
Display line
Reference display line
SIGLENT
1 2
3 4 5 976
10
15
16
14
13
12 11
8
Figure 1-13 User Interface of Vector Network Analyzer Mode
Table 1-4 Vector Network Analyzer Mode User Interface labels
NO.
Name
Description
1
SIGLENT
SIGLENT logo
2
Calibration Status
Cor: Calibrated; Off : Correction Off; C?: Need to re-calibrate;
3
Port Extensions
P: Port Extensions is On
4
Marker Table
Active Marker indication,every trace 6 markers
5
Marker
Current trace active marker
6
Marker x value
Marker frequency
7
Marker y value
Unit depend on trace format
8
Menu title
Function of the current menu.
9
Menu items
Menu items of the current function
10
Stop frequency
Stop frequency value
11
Points
Measurement points number,101~751
12
Start frequency
The first frequency of a sweep
13
Mode
Mode indication
14
Trace active
Highlight active trace indication depend on trace format
15
Trace deactive
Gray indicationd
16
Trace
Active trace
SIGLENT
12 User Manual

1.7 Mode

The analyzer offers a variety of operating modes that can be purchased separately. They can be selected via the Mode key:
Spectrum Analyzer Mode The default mode of the analyzer. It offers a general-purpose spectrum analyzer with a tracking generator, and some advanced measurement.
Vector Network Analyzer Mode In this mode, the analyzer offers a vector network analysis function and operation UI. It is able to measure full 1 port S11 and single direction S21 parameters. This mode is only available on SVA1000X series.
Distance-To-Fault Mode It is a performance verification and failure analysis tool, used for antenna and transmission line service and maintenance. It returns VSWR versus distance, to show the effects of poor connections, damaged cables, or faulty antennas. This mode is only available on SVA1000X series.
Modulation Analyzer Mode In this mode, the analyzer offers analog signal analysis (AM/FM) and digital vector signal analysis (ASK/FSK/PSK/MSK/OAM) functions. The original RF data can also be recorded to PC, from EasyVSA software.
EMI Measurement Mode In this mode, the analyzer provides the EMI pre-compatibility measurement function, including a sequence test, a CISPR EMC filter and QP detector, and a set of EMC limit standard lines.
Real-Time Spectrum Analyzer Mode In this mode, the analyzer offers real-time spectrum analysis with density, 3D, spectrogram, PvT, etc.
from multi-views and dimensions to monitor complex signals.
This mode is only available on SSA3000X-R series.
Front panel key menus may be different in different modes. Different modes have their own Preset function. Refer to User Manual for more details about Mode.
SIGLENT

1.8 Firmware Operation

1.8.1 Check System Information

Users can get the system information by press System-> “System Info”, including
Product Model, Serial and Host ID
Software Version and hardware Version
Option Information

1.8.2 Load Option

Refer to the procedures below to activate the options you have purchased.
1. Press System-> “System Info-> “Load Option
2. Enter the license key in the onscreen window. Press Enter to confirm your input and terminate the
license key input. Or
3. Load the .lic file provided by pressing File -> “Load” from internal memory or USB stick. The option will be enabled after rebooting.

1.8.3 Firmware Upgrade

Follow this procedure to update the instrument firmware:
1. Download the firmware package from an official SIGLENT website.
2. Extract and copy the .ADS file into the root directory of an USB stick.
3. Plug the USB stick into the USB Host connector. Press System-> System Info-> Firmware
Update”; find the .ADS file in USB stick.
4. Press the Load, the analyzer will perform the update process automatically.
The upgrade process will take several minutes. When the upgrade is completed, the machine will
reboot.
Any interruption during the update process will result in update failure and system data loss. This
is not covered under the warranty and the user will bear repair costs and shipping.
Do not remove the USB storage device until the update is finished.

1.9 Remote Control

The analyzer supports communication with computers via USB, LAN, and GPIB-USB interfaces. By using these interfaces, in combination with programming languages and/or NI-VISA software, users can remotely control the analyzer based on a SCPI (Standard Commands for Programmable Instruments) compliant command set, LabView and IVI (Interchangeable Virtual Instrument), to interoperate with other programmable instruments.
SIGLENT
14 User Manual
You can also remote monitor and control the analyzer in Web Browser or Easy Spectrum. For more details, refer to the ‘Programming Guide’ or contact your nearest SIGLENT office.

1.10 Touch Operation

The analyzer has a 10.1-inch multi-touch screen and supports various gesture operations. Including:
Press or click on the upper-right-corner of the screen to enter the main menu;
Swipe up and down or left and right in the waveform area to change the X-axis center coordinate
or Y-axis reference level;
Perform two-points scaling in the waveform area to change the X-axis span;
Click on a screen parameter or menu for parameter selection or editing;
Open and drag the marker;
Use auxiliary shortcuts to perform common operations.
You can turn the touch screen function on and off via Display->’Touch Settings’.

1.11 Using Built-in Help

The built-in help system provides information about every function key at the front panel and every menu soft key.
Press Help and the embedded help would show up.
Click on the items in the contents tree on the left to navigate to any topics interested.
Click the green back or forward arrow to go back or forward to the contents just read.
Click the close sign button in the top right corner or press the Esc front-panel key to quit the help
system.
Figure 1-14 help information
SIGLENT

Chapter 2 Spectrum Analyzer Mode

Press Mode, select ‘Spectrum Analyzer’ to enter spectrum analyzer mode. The ‘Spectrum Analyzer’ mode is the default mode of the analyzer. In this mode, the Mode backlight does not light up; in other modes, the Mode backlight will light up.
This chapter introduces in detail the function keys and menu functions of the front panel in Spectrum Analyzer Mode.

2.1 Basic Settings

2.1.1 Frequency

Set the frequency-related parameters and functions of the analyzer. The sweep will restart every time the frequency parameters are modified.
The frequency range of a channel can be expressed by three parameters: Start Frequency, Center Frequency and Stop Frequency. If any of the parameters change, the others will be adjusted automatically in order to ensure the coupling relationship among them:
startstopspan
stopstartcenter
fff
)/2f(ff
=
+=
, Where
span
f
is the span.
2.1.1.1 Center Frequency
Set the center frequency of the current sweep. The center frequency and span values are displayed at the bottom of the grid respectively.
Modifying the center frequency will modify both the start frequency and stop frequency when the
span is constant (except when the start frequency or stop frequency reaches the boundary).
In Zero Span, the start frequency, stop frequency and center frequency are always the same
value.
Table 2-1 Center Frequency
Parameter
Explanation
Default
Full Span/2
Range
Zero Span, 0 Hz ~ Full Span Nonzero Span, 50 Hz ~ (Full Span -50Hz)
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200, min 1 Hz Span = 0, step = RBW/100, min 1 Hz
SIGLENT
16 User Manual
Direction Key Step
Freq Step
Related to
Start Freq, Stop Freq
2.1.1.2 Start Frequency
Set the start frequency of the current sweep. The start and stop frequencies are displayed at the bottom of the grid respectively.
The span and center frequency vary with the start frequency when the Span does not reach the
minimum. For more details, please refer to “Span”.
In Zero Span, the start frequency, stop frequency and center frequency are the same value.
Table 2-2 Start Frequency
Parameter
Explanation
Default
0 Hz
Range
Zero Span, 0 Hz ~ Full Span Nonzero Span, 0 Hz ~ (Full Span-100Hz)
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200, min 1 Hz Span = 0, step = RBW/100, min 1 Hz
Direction Key Step
Freq Step
Related to
Center Freq, Span
2.1.1.3 Stop Frequency
Set the stop frequency of the current sweep. The start and stop frequencies are displayed at the lower right sides of the grid respectively.
The span and center frequency vary with the stop frequency. The change of the span will affect
other system parameters. For more details, please refer to “Span”.
In Zero Span, the start frequency, stop frequency and center frequency are always the same
value.
Table 2-3 Stop Frequency
Parameter
Explanation
Default
Full Span
Range
Zero Span: 0 Hz ~ Full Span Nonzero Span: 100 Hz ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200, min 1 Hz Span = 0, step = RBW/100, min 1 Hz
Direction Key Step
Freq Step
Related to
Center Freq, Span
SIGLENT
2.1.1.4 Freq Offset
Set the frequency offset value to illustrate the frequency conversion between the measured device and the input of the spectrum analyzer.
This parameter does not affect any hardware settings of the spectrum analyzer, but only changes
the display values of center frequency, start frequency and stop frequency.
To eliminate the frequency offset value, the frequency offset value can be set to 0 Hz.
Table 2-4 Freq Offset
Parameter
Explanation
Default
0 Hz
Range
-100GHz ~ 100GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, Step = Span/200, min 1 Hz Span = 0, Step = RBW/100, min 1 Hz
Direction Key Step
Freq Step
Related to
Center Freq, Start Freq, Stop Freq
2.1.1.5 Freq Step
Setting the value of Freq Step will change the direction key step of center frequency, start frequency, stop frequency and frequency offset.
At a fixed step change the value of the center frequency can reach the purpose of switching
measurement channels rapidly and continuously.
There are two kinds of frequency step modesAuto and Manual. In Auto mode, the Freq step is
1/10 of the span in Non-zero span or equals the RBW while in Zero Span. In Manual mode, you can set the step using the numeric keys.
Table 2-5 Freq Step
Parameter
Explanation
Default
Full Span/10
Range
1Hz ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, Step = Span/200, min 1 Hz Span = 0, Step = RBW/100, min 1 Hz
Direction Key Step
In 1-2-5 sequence
Related to
RBW, Span and related parameters
2.1.1.6 Signal Track
Turn on or off the signal tracking function. It is used to track the signal whose frequency is unstable and the instantaneous change of the amplitude is less than 3 dB. By marking cursor 1 on the measured
SIGLENT
18 User Manual
signal, the change of the measured signal can be tracked and measured all the time. The signal tracking process is shown in the following figure:
Figure 2-1 Signal Tracking Flow
When Marker1 is on, turn on signal tracking, a point whose amplitude does not change more than
3 dB near Marker1 will be searched and marked, and the frequency at that point will be set to center frequency.
When Maker1 is off, turn on signal tracking, Marker1 will be activated, and a peak search will be
performed, and then the peak frequency will be set to center frequency.
Signal tracking function is only available in sweep analysis. Signal tracking function is turned off in
following cases:
Zero Span mode
Tracking Generator (TG) open
Traces are not updated, including single sweep mode or View mode.
Cont Peak function turned on
Other non-SA measurement modes
2.1.1.7 Peak -> CF
Executes a peak search and sets the center frequency (CF) of the display to the frequency of the current peak. The function is invalid in Zero Span.
SIGLENT
Figure 2-2 before Peak -> CF
Figure 2-3 after Peak -> CF
2.1.1.8 CF -> Step
Set the current center frequency as the Freq Step. At this point, the Freq Step will switch to “Manual
SIGLENT
20 User Manual
mode automatically. This function is usually used in channel switching. For example, in harmonic measurement, firstly locate the signal at the center frequency (CF) of the display, and then execute CF->Step. Next you can press the upward direction key continuously to measure each order of harmonic in sequence.

2.1.2 Span

Set the span of the analyzer. Any change of this parameter will affect the frequency parameters and restart the sweep.
2.1.2.1 Span
Set the frequency range of the current sweep. The center frequency and span are displayed at the bottom of the grid respectively.
The start and stop frequency vary with the span when the center frequency is constant.
In non-zero span mode, the span can be set down to 100 Hz and up to the full span described in
Specifications. When the span is set to the maximum, the analyzer enters full span mode.
Modifying the span in non-zero span mode may cause an automatic change in both Freq Step and
RBW if they are in Auto mode. Besides, the change of RBW may influence VBW (in Auto VBW mode).
Variation in the span, RBW or VBW would cause a change in the sweep time.
Table 2-6 Span
Parameter
Explanation
Default
Full Span
Range
0 Hz ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
In 1-2-5 sequence
Related to
Start Freq, Stop Freq, Freq Step, RBW, Sweep time
Note: 0 Hz is available only in zero span.
2.1.2.2 Full Span
Set the span of the analyzer to the maximum available frequency span.
2.1.2.3 Zero Span
Set the span of the analyzer to 0Hz. Both the start and stop frequencies will equal to center frequency and the horizontal axis will denote time. The analyzer measures the time domain characteristics of the amplitude of the corresponding frequency point of the input signal.
The following functions are invalid in Zero span:
Frequency: Peak->CF and Signal Track;
SIGLENT
SPAN: Zoom In and Zoom Out;
Sweep: Sweep Mode;
Marker->: M->CF, M->CF step, M->Start Freq, M->Stop Freq, 󰅿M->Span and 󰅿M->CF;
Marker Fn : Read Out (default option: 󰅿Time);
2.1.2.4 Zoom In
Set the span to half of its current value. At this point, the signal on the screen will be amplified to observe signal details.
2.1.2.5 Zoom Out
Set the span to twice the current value. At this point, the signal on the screen will be reduced to gain more information about the nearby spectrum.
2.1.2.6 Last Span
Set the span to the previous span setting.
2.1.2.7 X-Scale
Set the scale type of X-axis to Linear (Lin) or Logarithmic (Log) scale. In Log scale type, the frequency scale of X-axis is displayed in the logarithmic form, and Meas function is invalid.
Figure 2-4 Logarithmic X Scale
SIGLENT
22 User Manual

2.1.3 Amplitude

Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error. Any change of Ref Level, Attenuator Value, Preamp mode and Ref Offset will restart sweep.
2.1.3.1 Ref Level
Set the maximum power or voltage that can be currently displayed in the trace window. The value is displayed at the upper left corner of the screen grid. The maximum reference (Ref) level available is affected by the maximum mixing level; input attenuation is adjusted under a constant maximum mixing level in order to fulfill the following condition:
Ref <= ATT - PA - 20dBm, where ATT = Attenuation value, PA = Preamplifier value
Table 2-7 Ref Level
Parameter
Explanation
Default
0 dBm
Range
-200 dBm ~ 20 dBm
Unit
dBm, dBmV, dBuV, dBuA, V, W
Knob Step
In Log scale mode, step = Scale/10 In Lin scale mode, step = 0.1 dBm
Direction Key Step
In Log scale mode, step = Scale In Lin scale mode, step = 1 dBm
Related to
Attenuator, Preamp, Ref Offset
Note: the maximum reference level of different machine models may be different, please refer to the data manual specifically.
2.1.3.2 Attenuator
Set the value for the internal attenuator of the RF input. So that the large signal can be low distortion and the small signal can pass through the mixer with low noise.
Ref <= ATT - PA - 20dBm, where ATT = Attenuation value, PA = Preamplifier value
Input attenuation can be set up to auto or manual mode.
Auto mode: the attenuation value is automatically adjusted according to the state of preamplifier
and the current reference level.
The maximum input attenuation can be set to 31 dB. When the set parameters do not meet the
above formula, you can adjust the reference level.
SIGLENT
Table 2-8 Attenuator
Parameter
Explanation
Default
20 dB
Range
0 ~ 31 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
5 dB
Related to
Preamp, Ref level
Note: the maximum attenuator value of different machine models may be different, please refer to the data manual specifically.
2.1.3.3 RF Preamp
Control the state of the internal preamplifier (PA) located in the RF input signal path. When the signal-under-measurement is small, turning on the preamplifier can reduce the displayed noise level and aid distinguishing small signals from the noise. The corresponding icon “PA” will appear at the left side of the screen when the preamplifier is turned on.
2.1.3.4 Units
Set the unit of the Y-axis to dBm, dBmV, dBuV, dBuA, Volts (RMS) or Watts. Default is dBm. The conversion relationships between units are as follows.
Where, R denotes the reference impedance. The default value is 50Ω and can be adjusted by pressing
Correction -> RF input”. The “75 Ω” impedance is just a numeric value, not a real impedance. Setting the RF input to 75 Ω will not change the actual input impedance. A 75 Ω feed-through adapter is required to match 75 Ω circuits to the 50 Ω input of the analyzer.
SIGLENT
24 User Manual
2.1.3.5 Scale
Set the logarithmic units per vertical grid division on the display. This function is only available when the
scale type is set to “log”.
By changing the scale, the displayed amplitude range is adjusted.
The Minimum range: Reference level –10 × current scale value.
The Maximum range: The reference level.
Table 2-9 Scale/Div
Parameter
Explanation
Default
10 dB
Range
0.1 dB ~ 20 dB
Unit
dB
Knob Step
Scale>=1, 1 dB, Scale<1, 0.1dB
Direction Key Step
1-2-5 sequence
Related to
Scale Type
2.1.3.6 Scale Type
Set the scale type of the Y-axis to Lin or Log. The default is Log.
In Lin mode, the vertical Scale value cannot be changed. The Display area is set for reference
level of 0%.
In Log scale type, the Y-axis denotes the logarithmic coordinate. The value shown at the top of the
grid is the reference level and each grid represent the scale value. The unit of Y-axis will automatically switch to the default unit (dBm) in Log scale type when the scale type is changed from Lin to Log.
In Lin scale type, the Y-axis denotes the liner coordinate; the values shown at the top of the grid
and the bottom of the grid are the reference level and 0 V. The scale setting function is invalid. The unit of Y-axis will automatically switch to the default unit (Volts) in Lin scale type when the scale type is charged from Log to Lin.
2.1.3.7 Ref Offset
Assign an offset to the reference level to compensate for gains or losses generated between the device under measurement and the analyzer. The change of this value changes both the reference level readout and the amplitude readout of the marker; but does not impact the position of traces on the screen.
Table 2-10 Ref Offset
Parameter
Explanation
Default
0 dB
SIGLENT
Range
-100 dB ~ 100 dB
Unit
dB
Knob Step
Not support
Direction Key Step
Not support
Related to
Ref Level
2.1.3.8 Correction
Correct the displayed amplitude to compensate for gains or losses from external devices such as antennas and cables. When using this function, you can view the correction data table and save or load the current correction data. When amplitude correction is enabled, both the trace and related measurement results will be mathematically corrected. Positive correction values are added to the measured values. Negative (-) correction values are subtracted from the measured values.
1. RF Input
Set the input impedance for numeric voltage-to-power conversions. To measure a 75 Ω device, you should use a 75 Ω to 50 Ω adapters to connect the analyzer with the system-under-test and then set the input impedance to 75 Ω.
2. Apply Correction
Enable or disable amplitude corrections. Default is Off. The analyzer provides four correction factors that can be created and edited separately, but they can be applied independently in any combination.
3. Edit Correction factors
Table 2-11 Edit Correction table
Function
Explanation
Correction
Select the correction factor on or off.
Add Point
Add a point into correction table.
Point Num
Select a point to edit by point num.
Frequency
Edit the frequency value for the current selected point.
Amplitude
Edit the amplitude value for the current selected point.
Del Point
Delete the selected correction point.
Del All
Clear all data of the correction table.
Save/Load
Save or load correction data. You can save the current correction data into or load correction data from a specified file.

2.1.4 Auto Tune

The analyzer will search for signals automatically throughout the full frequency range and adjust the frequency and amplitude settings for optimum display of the strongest signal.
In the process of auto search, The “Auto Tune” is shown in the status bar on the screen until the
SIGLENT
26 User Manual
search is finished.
Some parameters such as the reference level, scale, input attenuation and maximum mixing level
may be changed during the auto search.
Figure 2-5 before Auto Tune
Figure 2-6 after Auto Tune
SIGLENT

2.2 Sweep and Functions

2.2.1 BW

The bandwidth menu contains the RBW (Resolution Bandwidth), VBW (Video Bandwidth), average type and filter type. Filter type includes the EMI filter type that enables EMI measurement controls.
2.2.1.1 Resolution Bandwidth
Set the resolution bandwidth in order to distinguish between signals which have frequency components that are near one another.
Reducing the RBW will increase the frequency resolution, but will also increase the sweep time
dramatically (Sweep Time is affected by a combination of RBW and VBW when the analyzer is in Auto mode).
Generally, the frequency resolution ability is affected by RBW, RBW Filter shape factor, LO Phase
noise, and LO Residual FM.
RBW varies with the span (non-zero span) in Auto RBW mode.
Under EMI filter, RBW can only be set to 200 Hz, 9 kHz, 120 kHz and 1 MHz with a 6dB shape
factor.
Table 2-12 RBW
Parameter
Explanation
Default
1 MHz
Range
1 Hz ~ 1 MHz
Unit
MHz, kHz, Hz
Knob Step
in 1, 3, 10 sequence
Direction Key Step
in 1, 3, 10 sequence
Relation
Span, VBW, V/R Ratio, Sweep Time
Notes: RBW of above table is in Gauss filter type.
2.2.1.2 Video Bandwidth
Set the video bandwidth in order to filter out the noise outside the video band.
Reducing the VBW will smooth the trace and helps to highlight small signals from noise, but it will
also increase the sweep time (Sweep Time is affected by a combination of RBW and VBW when it is in Auto mode).
VBW varies with RBW when it is in Auto mode. While in Manual mode, VBW is not affected by
RBW.
SIGLENT
28 User Manual
Table 2-13 VBW
Parameter
Explanation
Default
1 MHz
Range
1 Hz ~ 1 MHz
Unit
MHz, kHz, Hz
Knob Step
in 1, 3 sequence
Direction Key Step
in 1, 3 sequence
Relation
RBW, V/R Ratio, Sweep Time
2.2.1.3 V/R Ratio
Set the ratio of VBW to RBW. This value is different while measuring different kinds of signals:
Sine/Continuous Wave (CW) signals: Use 1 to 3 (for faster sweeps)
Pulsed/transient signals: Use 10 (to reduce the influence on the amplitude of transient signals)
Noise signals: Generally use 0.1 (to obtain the average of noises)
Table 2-14 V/R Ratio
Parameter
Explanation
Default
1
Range
0.001 ~ 1000
Unit
N/A
Knob Step
in 1, 3 sequence
Direction Key Step
in 1, 3 sequence
Relation
RBW, VBW
2.2.1.4 Average Type
Choose one of the following averaging types: log power (video), power (RMS), or voltage averaging. When trace average is on, the average type is shown on the left side of the display.
1. Log Power
Select the logarithmic (decibel) scale for all filtering and averaging processes. This scale is "Video" because it is the most common display and analysis scale for the video signal within analyzer. This scale is excellent for finding Sine/CW signals near noise.
2. Power Average
In this average type, all filtering and averaging processes work on the power (the square of the magnitude) of the signal, instead of its log or envelope voltage. This scale is best for real-time power measurement of complex signals.
3. Voltage Average
In this Average type, all filtering and averaging processes work on the voltage of the envelope of the
SIGLENT
signal. This scale is suitable for observing rise and fall behavior of AM or pulse-modulated signals such as radar and TDMA transmitters.
2.2.1.5 Filter
Set the RBW filter type. The analyzer supports two kinds of RBW filters: “Gauss” (-3 dB bandwidth) and
EMI” (-6 dB bandwidth). When “EMI” is selected, resolution bandwidth can only be 200 Hz, 9 kHz, 120 kHz and 1 MHz Quasi-Peak” detector is available only when EMI” filter is turned on.

2.2.2 Trace

The sweep signal is displayed as a trace on the screen.
2.2.2.1 Select Trace
The analyzer allows for up to four traces to be displayed at the same time. Each trace has its own color (Trace A - Yellow, Trace B - Purple, Trace C - Light blue and Trace D - Green). All traces can be set parameter independently. As a default, analyzer will choose Trace A and set the type of the trace as Clear Write.
2.2.2.2 Trace Type
Set the type of the current trace or disable it. The system calculates the sampled data using a specific operation method according to the trace type selected and displays the result. Trace types include Clear Write, Max Hold, Min Hold, View, Average and Blank. The corresponding icon of the trace type will be displayed in the status bar at the left of the screen, as shown in the figure below.
Figure 2-7 Trace Type
SIGLENT
30 User Manual
1. Clear Write
Erases any data previously stored in the selected trace, and display the data sampled in real-time of each point on the trace.
2. Max Hold
Retain the maximum level for each point of the selected trace. Update the data if a new maximum level is detected in successive sweeps. Max Hold is very effective when measuring events that may take successive scans to measure accurately. Some common applications include FM Deviation, AM NRSC, and frequency hopping or drift.
3. Min Hold
Display the minimum value from multiple sweeps for each point of the trace and update the data if a new minimum is generated in successive sweeps.
4. View
Freezes and holds the amplitude data of the selected trace. The trace data is not updated as the analyzer sweeps.
5. Blank
Disable the trace display and all measurements of this trace.
6. Average
Set the averages times of the selected trace. More averages can reduce the noise and the influence of other random signals; thus highlighting the stable signal characteristics. The larger the averages times is, the smoother the trace will be. Enabling averaging will take more time to collect the full spectral information because the analyzer needs to sweep the set average count. The displayed data is averaged in a first-in-first-out fashion.
Table 2-15 Average Times
Parameter
Explanation
Default
100
Range
1 ~ 999
Unit
N/A
Knob Step
1
Direction Key Step
5
2.2.2.3 Math
Set the computational method of the math trace.
1. Output Z
The Math result is denoted by the Z variable and can be displayed by trace A, B, C, or D.
SIGLENT
2. Input X, Y
Input X, Y can be applied to trace A, B, C, or D.
3. Calculation Type
The analyzer provides the calculation types as shown below:
Power Diff: X-Y+Offset→Z
Power Sum: X+Y+Offset→Z
Log Offset: X+ Offset→Z
Log Diff: X-Y-Ref→Z
4. Offset value
Table 2-16 Offset value
Parameter
Explanation
Default
0 dB
Range
-100 dB ~ 100 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
1 dB

2.2.3 Detect

The analyzer displays the sweep signal on the screen in the form of a trace. For each trace point, the analyzer always captures all the data within a specific time interval and processes (Peak, Average, etc.) the data using the detector currently selected, then it displays the processed data (a single data point) on the screen. Select an appropriate detector type according to the actual application in order to ensure the accuracy of the measurement. The available types are Pos Peak, Neg Peak, Sample, Normal, Average and Quasi Peak. The default is Pos peak.
1. Positive Peak
For each trace point, Positive Peak detector displays the maximum value of data sampled within the corresponding time interval.
2. Negative Peak
For each trace point, Negative Peak detector displays the minimum value of data sampled within the corresponding time interval.
3. Sample
For each trace point, Sample detector displays the transient level corresponding to the central time
SIGLENT
32 User Manual
point of the corresponding time interval. This detector type is applicable to noise or noise-like signal.
4. Normal
Normal detector (also called ROSENFELL Detector) displays the maximum value and the minimum value of the sample data segment in turn: Odd-numbered data points display the maximum value and even-numbered data points display the minimum value. In this way, the amplitude variation range of the signal is clearly shown.
5. Average
For each trace point, Average detector displays the average value of data sampled within the corresponding time interval.
6. Quasi-Peak
Quasi-Peak (QP) detector, which is a weighted form of peak detector, is used for EMC pulse testing. The QP detector is designed to follow CISPR-16 response specifications. For a single frequency point, the detector detects the peaks within the QP dwell time. The peaks detected are weighted using a digital model that follows a defined response curve as well as the time constant specified in the CISPR 16 standards. The measurement time for QP is far longer than Peak Detector.

2.2.4 Sweep

Sets parameters about the Sweep functions, including sweep time, sweep times, sweep mode, etc.
2.2.4.1 Sweep Time
Sets the time needed for the analyzer to finish a sweep within the span range. The sweep time can be set in “Auto” or “Manual” mode and the default is “Auto”.
In non-zero span, the analyzer selects the shortest sweep time on the basis of the current RBW
and VBW settings if Auto is selected.
Decreasing the sweep time will decrease measurement time. However, an error may be caused if
the specified sweep time is less than the minimum sweep time in Auto coupling; at this point, UNCAL” is shown in the status bar on the screen. Measurements taken with “UNCAL” showing may not meet the specifications of the instrument and can have significant error.
Table 2-17 Sweep Time
Parameter
Explanation
Default
N/A
Range
900 us ~ 1.5 ks (Quasi Peak: 900us ~ 15ks)
Unit
ks, s, ms, us
Knob Step
Sweep time/100, min =1 ms
Direction Key Step
in 1, 3 sequence
SIGLENT
2.2.4.2 Sweep Rule
The analyzer provides two sweep time rules to meet the different sweep time requirements:
Speed: Activates the default fast sweep time rule.
Accuracy: Activates the normal sweep time rule to ensure increased measurement accuracy.
The Speed sweep time rule provides a fast measurement function that decreases the sweep time. While the Accuracy Sweep rule will increase the measurement accuracy.
2.2.4.3 Sweep
Set the sweep mode in single or continuous, the default is continuous. The corresponding icon of the sweep will be displayed in the status bar at the left of the screen.
1. Single Set the sweep mode to “Single”. The number on the parameter icon denotes the current sweep times.
2. Numbers
Set the sweeps times for a single sweep. In single sweep mode, the system executes the specified sweeps times and the number shown on the icon in the status bar at the left of the screen varies with the process of the sweep.
3. Continue
Set the sweep mode to “Continue”. The character Cont on the parameter icon denotes the analyzer is sweeping continuously.
If the instrument is in single sweep mode and no measurement function is enabled, press this key
and the system will enter continuous sweep mode and sweep continuously if the trigger conditions are satisfied.
If the instrument is in single sweep mode and a measurement function is on, press this key and the
system will enter continuous sweep mode and measure continuously if the trigger conditions are satisfied.
In continuous sweep mode, the system will send a trigger initialization signal automatically and
enter the trigger condition judgment directly after each sweep.
Table 2-18 Sweep Times
Parameter
Explanation
Default
1
Range
1 ~ 99999
Unit
N/A
Knob Step
1
Direction Key Step
1
SIGLENT
34 User Manual
2.2.4.4 Sweep Mode
Sweep mode includes auto mode, sweep mode and FFT mode.
1. Auto When the sweep mode is auto, the analyzer selects the sweep mode automatically between Sweep and FFT Mode in the shortest time.
2. Sweep True swept operation including point-by-point scanning. The Sweep mode is available when the RBW is in 30 Hz – 1 MHz. For a spectrum analyzer that scans at a single point, each time it tunes to the corresponding frequency, it only completes the energy detection of the current frequency point.
3. FFT The FFT mode is only available when RBW is in 1 Hz - 10 kHz. For a spectrum analyzer that performs scanning in FFT steps, each time it tunes to a certain frequency, it will stay longer, and it also analyzes all the energy in a frequency band. Therefore. The selected range of FFT step scanning become wider, but the basic process has not changed fundamentally compared to scaning at a single point. Simply put, FFT can be regarded as a series of parallel IF filters working at the same time, thus speeding up the frequency sweep. When the tracking generator (TG) is on, the sweep mode is forced to Sweep.
2.2.4.5 QPD Dwell Time
Dwell time is the measurement time at a single frequency. QP detector gets its weighted envelope response during this dwell time. The longer dwell time is, the more sufficiently the QP detector responses to a single frequency, and the more accurate the QP detector envelope is.
Table 2-19 QPD Dwell Times
Parameter
Explanation
Default
50 ms
Range
0 s ~ 10 s
Unit
ks, s, ms, us
Knob Step
1 ms
Direction Key Step
1 ms

2.2.5 Trigger

2.2.5.1 Free Run
The trigger conditions are satisfied at any time and the analyzer generates trigger signals continuously.
SIGLENT
2.2.5.2 Video Trigger
A trigger signal will be generated when the system detects a video signal of which the voltage exceeds the specified video trigger level. Set the trigger level with the video trigger menu entry. At this point, the trigger level line (Trig Line) and value are displayed on the screen.
Table 2-20 Trigger Setup
Parameter
Explanation
Default
0 dBm
Range
-300 dBm ~ 50 dBm
Unit
dBm
Knob Step
1 dBm
Direction Key Step
10 dBm
2.2.5.3 External
In this mode, an external signal (TTL signal) is input from the [TRIGGER IN] connector at the rear panel and trigger signals are generated when this signal fulfills the specified trigger edge condition. Set the trigger edge in external trigger to the rising (Pos) or falling (Neg) edge of the pulse.

2.2.6 Limit

The analyzer supports Pass/Fail test function. In this function, the measured curve will be compared with the pre-edited curve. If the related rules are met, the result is “Pass, else the result is “Fail”.
2.2.6.1 Limit1
Enable or disable limit1.
2.2.6.2 Limit1 Edit
Edit the properties of limit1.
Table 2-21 Limit1 Edit Menu
Function
Explanation
Type
Select upper or lower limit type. The default value is Upper.
Mode
Select limit line or limit point. The default value is Line. Set the number of the point to be edited if you selected the point type, and the range is 1 ~ 100.
Add point
Add a new point for editing.
X-axis
Edit the X-axis value (frequency or time) of the current point.
Amplitude
Edit the amplitude of the current point or line.
SIGLENT
36 User Manual
Del Point
Delete the point whose number is selected in Mode.
Del All
Delete all the points.
Save/Load
Save or load the limit file.
X Offset
Set offsets of X axis.
Y Offset
Set offsets of Y axis
2.2.6.3 Limit2
Enable or disable limit2.
2.2.6.4 Limit2 Edit
Edit the properties of limit2.
Table 2-22 Limit2 Edit Menu
Function
Explanation
Type
Select upper or lower limit type. The default value is Lower.
Mode
Select limit line or limit point. The default value is Line. Set the number of the point to be edited if you selected the point type, and the range is 1 ~ 100.
Add point
Add a new point for editing.
X-axis
Edit the X-axis value (frequency or time) of the current point.
Amplitude
Edit the amplitude of the current point or line.
Del Point
Delete the point whose number is selected in Mode.
Del All
Delete all points.
Save/Load
Save or load the limit file.
X Offset
Set offsets of X axis.
Y Offset
Set offsets of Y axis
2.2.6.5 Test
Enable or disable the limit test function.
2.2.6.6 Setup
1. Fail to stop
Turn on or off the Fail to stop function. If the function is on, the analyzer will stop sweep and retain the test result when the test result is “Fail”.
2. Buzzer
Turn on or off the buzzer. When the buzzer is on, it beeps when the test result is “Fail”.
3. X Axis
Set the X-axis to frequency or time domain. The default value is Freq.
SIGLENT
Note that all the points of the current limit line will be deleted if the X-axis unit changes.

2.2.7 TG (Tracking Generator)

Set the parameters related to the tracking generator (TG) and normalize.
2.2.7.1 TG
The tracking generator is a signal source with an adjustable frequency and amplitude. When the TG is enabled, a signal with the same frequency of the current sweep signal will be output from the [TG SOURCE] connector at the front panel. The power of the signal could be set through the menu. The TG output frequency follows the analyzer sweep frequency. For example, if the sweep is set to scan from 1 MHz to 10 MHz, the TG output frequency will change from 1 MHz to 10 MHz in coordinated steps with the sweep. In Zero Span mode, the TG frequency will match the center frequency of the analyzer.
2.2.7.2 TG Level
Set the output power of the signal of the tracking generator.
Table 2-23 TG Level
Parameter
Explanation
Default
-20 dBm
Range
-20 dBm ~ 0 dBm
Unit
dBm
Knob Step
1 dBm
Direction Key Step
10 dBm
2.2.7.3 TG Level Offset
Assigns a certain offset to the output power of the TG when gains or losses occur between the TG output and external device in order to display the actual power value.
This parameter only changes the readout of the TG output power, rather than the actual value.
The offset could be either a positive (gain in the external output) or a negative (loss in the external
output).
Table 2-24 TG Level Offset
Parameter
Explanation
Default
0 dB
Range
-200 dB ~ 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB
SIGLENT
38 User Manual
2.2.7.4 Normalize
Normalization can eliminate errors in the TG Level. Before using this function, connect the [TG SOURCE] output terminal of the TG with the [RF INPUT] input terminal of the analyzer. The reference trace can be stored by pressing the “Store Ref” button after the current sweep finished. Then the normalize function can be enabled. When normalization is enabled, the corresponding value of the reference trace will be subtracted from the trace data after every sweep.
The reference trace must be stored before the normalize function can be used.
Default
reference plane
Normalized
reference plane
Figure 2-8 Normalize
2.2.7.5 Store Ref
Before enable the normalize function, you should press the “Store Ref” button to store the data of Trace
A to Trace D. Then Trace D is the reference trace.
The tracking generator needs to be turned on first, and the reference trace can be saved only
when the first screen is scanned.
The normalization can be enabled only after the reference trace is stored.
After you stored the reference trace and enabled the normalization, if then you change some
parameters which will change the sweep frequency and sweep time, the "UNCAL" sign will be displayed in the upper left corner of the screen, indicating that the reference trace is no longer applicable to the new trace, and prompting you to save the reference trace again.
2.2.7.6 Ref Trace
Set whether to display the reference trace or not. If “View” is selected, the reference trace saved will be shown in “View” type.
Note: When normalization is enabled, the unit of Y-axis is “dB” and will not be influenced by the definition in AMPT->Units. At this point, “(dB)” is displayed under the Y-axis scale in the user interface.
2.2.7.7 Norm Ref Level
Adjust the vertical position of the trace on the screen by adjusting the reference level when normalization is enabled.
SIGLENT
This operation differs from the Ref Level function in the AMPT menu. This parameter has no influence on the reference level of the analyzer.
Table 2-25 Reference level under normalization
Parameter
Explanation
Default
0 dB
Range
-200 dB ~ 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB
2.2.7.8 Norm Ref Pos
Adjust the vertical position of the normalization reference level on the screen by adjusting the reference position when normalization is enabled.
The function of this menu is like that of Norm Ref Level. When it is set to 0%, the normalization
reference level is displayed at the bottom of the screen grid and at the top when it is set to 100%.
Table 2-26 TG reference position under normalization
Parameter
Explanation
Default
100%
Range
0 ~ 100%
Unit
100%
Knob Step
1%
Direction Key Step
10%

2.2.8 Demod

Press Demod at the front panel to enter the demodulation setting menu. Both AM and FM demodulations are available on the analyzer.
2.2.8.1 Demod (AM/FM)
Set the demodulation type to AM or FM or off. The default is off.
The analyzer features an earphone jack and the demodulated signal can be output in audio
frequency (AF) mode through the earphone. The frequency and intensity of AF denotes the frequency and amplitude of the signal respectively.
A marker would be enabled automatically, place it at the desired frequency and set a demod time.
The analyzer would demod at the marker frequency for the demod time, and then performs a normal sweep.
Zero span and linear scale type amplitude can be used to see the time domain waveform of the
SIGLENT
40 User Manual
modulated signal.
More demodulation analysis, please refer to AM/FM modulation analyzer in MA option.
2.2.8.2 Earphone
Set the status of the earphone. When it is on, the demodulated signal can be heard through the earphone during the demodulation. By default, it is off.
2.2.8.3 Volume
Set the volume of the earphone.
Table 2-27 Volume
Parameter
Explanation
Default
6
Range
0 ~ 10
Unit
N/A
Knob Step
1
Direction Key Step
1
2.2.8.4 Demod Time
Set the time for the analyzer to complete a signal demodulation after each sweep. If the Earphone is set to “On”, you will hear the demodulated signal through the earphone during the demodulation. A longer demod dwell time is recommended for demodulating audio signals.
Table 2-28 Demod time
Parameter
Explanation
Default
5 s
Range
5 ms ~ 1000 s
Unit
ks, s, ms, us
Knob Step
0 ms ~ 100 ms, step = 1 ms 100 ms ~ 1 s, step = 10 ms 1 s ~ 10 s, step = 100 ms 10 s ~ 100 s, step = 1 s 100 s ~ 1000 s, step = 10 s
Direction Key Step
In 1-2-5 step sequence
SIGLENT

2.3 Marker

2.3.1 Marker

The marker appears as a rhombic sign (as shown below) for identifying points on a trace. You can easily read the amplitude, frequency and sweep time of the marked point on the trace.
Figure 2-9 Marker
The analyzer allows for up to eight/four pairs of markers to be displayed at one time, but only one
pair or a single marker is active every time.
You can use the numeric keys, knob or direction keys to modify the desired frequency or time as
well as view the readouts of different points on the trace.
2.3.1.1 Select Marker
Select one of the eight markers. The default is Marker1. When a marker is selected, you can set its type, trace to be marked, readout type and other related parameters. The enabled marker will appear on the trace selected through the Select Trace option and the readouts of this marker are also displayed in the active function area and at the upper right corner of the screen.
SIGLENT
42 User Manual
Table 2-29 Marker parameters
Parameter
Explanation
Default
Center Frequency
Range
0 ~ Full Span
Unit
Readout = Frequency: GHz, MHz, kHz, Hz Readout = Time or Period: s, ms, us, ns, ps
Knob Step
Readout = Frequency, Step = Span/(Sweep Points - 1) Readout = Time or Period, Step = Sweep Time/(Sweep Points - 1)
Direction Key Step
Readout = Frequency, Step = Span/10 Readout = Time or Period, Step = Sweep Time/10
2.3.1.2 Select Trace
Select the trace to be marked by the current marker. Valid selections include A, B, C, or D.
2.3.1.3 Normal
One of the marker types. It is used to measure the X (Frequency or Time) and Y (Amplitude) values of a
certain point on the trace. When selected, a marker with the number of the current marker (such as “1”)
appears on the trace.
If no active marker exists currently, a marker will be enabled automatically at the center frequency
of the current trace.
You can use the numeric keys, knob or direction keys to move the marker. The readouts of the
marker will be displayed at the upper right corner of the screen.
The readout resolution of the X-axis (frequency or time) is related to the span. For higher readout
resolution, reduce the span.
2.3.1.4 Delta
One of the marker types. It is used to measure the delta values of X (Frequency or Time) and Y (Amplitude) between the reference point and a certain point on the trace. When selected, a pair of markers appears on the trace: Fixed Related Marker (marked by a combination of the marker number
and letter “+”, such as “2+”) and the Delta Marker (marked by the “∆”, such as “1∆2”).
After the marker selects “Delta”, the original marker will become the delta measurement marker,
and the related marker of the incrementing sequence number will become the reference “fixed” marker
The delta marker is in the "relative to" state, and its X-axis position can be changed; the related
marker is in the "fixed" state by default (the X-axis and Y-axis positions are fixed), but the X-axis can be adjusted by changing to the "normal" state.
The first row in the upper right corner of the trace area shows the frequency (or time) difference
and amplitude difference between the two markers; the second row in the upper right corner of the trace area shows the X axis and amplitude value of the related marker.
SIGLENT
2.3.1.5 Fixed
One of the marker types. When “Fixed” is selected, the X-axis and Y-axis of the marker will not change by the trace and can only be changed through the menu. The fixed marker is marked with "+".
After the marker selects “Delta”, the original marker will become the delta measurement marker, and the related marker of the incrementing sequence number will become the reference “fixed” marker
2.3.1.6 Off
Turn off the marker currently selected. The marker information displayed on the screen and functions based on the marker will also be turned off.
2.3.1.7 Relative To
Relative to is used to measure the delta values of X (Frequency or Time) and Y (Amplitude) between two markers which can mark on different traces.
After the marker selects “Delta”, the original marker will become the delta measurement marker, and the related marker of the incrementing sequence number will become the reference “fixed” marker
2.3.1.8 Marker Table
Enable or disable the Marker Table. Display all the markers enabled on the lower portion of the screen, including marker number, trace number, marker readout type, X-axis readout and amplitude. Through this table you can view the measurement values of multiple points. The table allows for up to eight markers to be displayed at one time.
Figure 2-10 Marker table
SIGLENT
44 User Manual

2.3.2 Marker ->

1. M->CF
Set the center frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the center frequency will be set to the frequency of the current
marker.
If the Delta or Delta Pair marker is selected, the center frequency will be set to the frequency of
the Delta Marker.
The function is invalid in Zero span.
2. M -> CF Step
Set the center frequency step of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the center frequency step will be set to the frequency of the
current marker.
If the Delta or Delta Pair marker is selected, the center frequency step will be set to the frequency
of the Delta Marker.
The function is invalid in Zero span.
3. M -> Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the start frequency will be set to the frequency of the current
marker.
If the Delta or Delta Pair marker is selected, the start frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span.
4. M -> Stop Freq
Set the stop frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the stop frequency will be set to the frequency of the current
marker.
If the Delta or Delta Pair marker is selected, the stop frequency will be set to the frequency of the
Delta Marker.
The function is invalid in Zero span.
5. M ->Ref Level
Set the reference level of the analyzer to the amplitude of the current marker.
If the Normal marker is selected, the reference level will be set to the amplitude of the current
marker.
If the Delta or Delta Pair marker is selected, the reference level will be set to the amplitude of the
Delta Marker.
6. ΔM->Span Set the span of the analyzer to the frequency difference between the two markers in Delta marker type.
If the Normal marker is selected, this function is invalid.
SIGLENT
The function is invalid in Zero span.
7. ΔM->CF
Set the center frequency of the analyzer to the frequency difference between the two markers in Delta marker type.
If the Normal marker is selected, this function is invalid.
The function is invalid in Zero span.

2.3.3 Marker Fn

Special marker functions including Noise Marker, N dB BW and Freq Counter.
2.3.3.1 Select Marker
Select one of the eight markers (1, 2, 3, 4, 5, 6, 7 and 8) and the default is Marker1.
2.3.3.2 Noise Marker
Execute the Noise marker function for the selected marker and read the normalized noise power spectral density.
If the current marker is “Off” in the Marker menu, pressing Noise Marker will first set it to Normal
type automatically; then measure the average noise level at the marked point and normalize this value to 1 Hz bandwidth. During this process, certain compensation is always made based on the detection and trace types. The measurement will be more precise if RMS Avg or Sample detection type is used.
This function can be used for measuring the C/N ratio.
2.3.3.3 N dB BW
Enable the N dB BW measurement or set the value of N dB. The N dB BW denotes the frequency difference between two points that are located on both sides of the current marker and with N dB fall (N<0) or rise (N>0) in amplitude as shown in the figure on the next page.
SIGLENT
46 User Manual
Figure 2-11 N dB BW
When the measurement starts, the analyzer will search for the two points which are located at both sides of the current point with N dB fall or rise in amplitude and display the frequency difference between the two points in the active function area. "----" would be displayed if the search fails. The parameters in the figure are shown as
Figure 2-12 N dB parameter
Table 2-30 N dB Noise
Parameter
Explanation
Default
-3 dB
Range
-100 dB ~ 100 dB
Unit
dB
SIGLENT
Knob Step
0.1 dB
Direction Key Step
1 dB
2.3.3.4 Freq Counter
Turn on or off the frequency counter. The frequency readout is accuracy is up to 0.01 Hz.
The function is valid only when selecting marker 1.
If marker 1 is selected but not active, turning on the frequency counter will open marker 1 Normal
marker automatically.
The frequency counter measures the frequency near the center frequency in Zero span.
2.3.3.5 Off
Turn off the noise marker, N dB BW measurement or Frequency Counter, but not the marker itself.
2.3.3.6 Read Out
Select a desired readout type for the X-axis for the marker. Different markers can use different readout types. This setting will change the readout type and affect the marker readings in the active function area and at the upper right corner of the screen, but will not change the actual value.
1. Frequency
In this type, Normal marker shows the absolute frequency. Delta markers and Delta Pair markers show the frequency difference between the delta marker and reference marker. The default readout mode in non-zero span is “Frequency”.
Note: This type is invalid in Zero span.
2. Period
In this type, the Normal marker shows the reciprocal of frequency; while Delta marker and Delta Pair marker show the reciprocal of frequency difference. When the frequency difference is zero, the reciprocal is infinite and 100 Ts is displayed.
Note: This type is invalid in Zero span.
3. Δ Time
In this type, the Normal marker shows the time difference between the marker and the start of the sweep; while Delta marker and Delta Pair marker show the sweep time difference between the delta marker and reference marker. The default readout mode in Zero span is Δ Time.

2.3.4 Peak

Open the peak search setting menu and execute peak search.
SIGLENT
48 User Manual
2.3.4.1 Peak -> CF
Execute peak search and set the center frequency of the analyzer to the frequency of the peak.
2.3.4.2 Next Peak
Search for and mark the peak whose amplitude is closest to that of the current peak and which meets the peak search condition.
2.3.4.3 Next Left Peak
Search for and mark the nearest peak which is located at the left side of the current peak and meets the peak search condition.
2.3.4.4 Next Right Peak
Search for and mark the nearest peak which is located at the right side of the current peak and meets the peak search condition.
2.3.4.5 Peak Peak
Execute peak search and minimum search at the same time and mark the results with delta pair markers. Wherein, the result of peak search is marked with the delta marker and the result of minimum search is marked with the reference marker.
2.3.4.6 Count Peak
Enable or disable continuous peak search. The default is Off. When enabled, the system will always execute a peak search automatically after each sweep in order to track the signal under measurement.
2.3.4.7 Peak Table
Open the peak table (in the lower window) which lists the peaks (with frequency and amplitude) that meet the peak search condition. Up to 16 peaks can be displayed in the table.
2.3.4.8 Search Config
Define the conditions of peak search for various peak searches. A real peak should meet the
requirements of both the “Peak Excursion” and “Peak Threshold”.
1. Peak Threshold
Assign a minimum for the peak amplitude. Peaks whose amplitudes are greater than the specified peak threshold are treated as real peaks.
SIGLENT
Table 2-31 Peak Threshold
Parameter
Explanation
Default
-140 dBm
Range
-200 dBm ~ 200 dBm
Unit
dBm
Knob Step
1 dB
Direction Key Step
5 dB
2. Peak Excursion
Set the excursion between the peak and the minimum amplitude on both sides of it. Peaks whose excursions are beyond the specified excursion are treated as real peaks.
Table 2-32 Peak Excursion
Parameter
Explanation
Default
15 dB
Range
0 dB ~ 200 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
5 dB
3. Peak Type
Set the peak search condition. The available options are Maximum and Minimum.
SIGLENT
50 User Manual

2.4 Measurement

2.4.1 Meas

Provide optional measurement functions. When activated, the screen will be divided into two parts. The above part is the measure screen which displays traces, and the other part is used to display measurement results.
2.4.1.1 Reflection
Measure the DUT scalar s parameter. In this time, the spectrum analyzer automatically turns on TG. After selecting the measurement type as "reflection measurement", press Meas Setup to set the relevant parameters.
2.4.1.2 Channel Power
Measure the power and power density within the specified channel bandwidth. When this function is enabled, the span and resolution bandwidth are automatically adjusted to smaller values. Select Channel Power and press Meas Setup to set the corresponding parameters.
2.4.1.3 ACPR
Measure the power of the main channel and adjacent channels as well as the power difference between the main channel and each of the adjacent channels. When this function is enabled, the span and resolution bandwidth of the analyzer are adjusted to smaller values automatically. Select ACPR and press Meas Setup to set the corresponding parameters.
2.4.1.4 Occupied BW
Integrates the power within the whole span and calculates the bandwidth occupied by this power according to the specified power ratio. The OBW function also indicates the difference (namely “Transmit Freq Error”) between the center frequency of the channel under measurement and the center frequency of the analyzer. Select Occupied BW and press Meas Setup to set the corresponding parameters.
2.4.1.5 T-Power
The system enters Zero span and calculates the power within the time domain. The types of powers available include Peak, Average and RMS. Select T-Power and press Meas Setup to set the corresponding parameters.
2.4.1.6 TOI
Automatic measurement of IP3 (Third order Intercept Point), including the power of fundamental wave
SIGLENT
and the Third order in the power, and calculate the adjustable Intercept Point.
2.4.1.7 Spectrum Monitor
Display the power of the swept spectrum as an intensity color map commonly referred to as a waterfall chart. Select Spectrum Monitor and press Meas Setup to set the corresponding parameters.
2.4.1.8 CNR
Measure the power of the carrier and noise of the specified bandwidth and their ratio. Select CNR and press Meas Setup to set the corresponding parameters.
2.4.1.9 Harmonics
The harmonic power and total harmonic distortion of carrier signal are measured. The maximum measurable harmonic is 10th harmonic. The fundamental wave amplitude of carrier signal must be greater than - 50 dBm, otherwise the measurement result is invalid. Select Harmonics and press Meas Setup to set the corresponding parameters.
2.4.1.10 Meas Off
Turn off all the Meas functions.

2.4.2 Meas setup

2.4.2.1 Reflection
Figure 2-13 Reflection measurement
Return LossRefi CoefficientVSWR:
Return LossRatio of incident power to reflected power
Refi CoefficientRatio of reflected to incident voltage.
VSWRRatio of standing wave belly voltage to valley voltage amplitude.
SIGLENT
52 User Manual
1. Reset
Clear calibration data
2. Calibration
In the ssa3000x plus series spectrometer, an external bridge is used for reflection measurement. The connection between the bridge and the spectrometer is shown in the figure below.
Figure 2-14 external bridge Dotted arrows indicate the direction of reflected signal Solid line arrow indicates the direction of incident signal Table 2-33 Calibration Type
Calibration
instruction
Cal Open Load
Connect open calibration to DUT port
Open+Short/2
Connect open and short calibration to DUT port respectively
Open+Load
Connect open and load calibration to DUT port respectively
3. Ref Pos
Set the position of the reference level in Y axis.
4. Average
Turn average measurement on or off
5. Pause/Resume
Pause or resume the current measurement
SIGLENT
6. Restart Restart the measurement
2.4.2.2 Channel Power
Figure 2-15 Channel Power
Measurement Results: Channel power and power spectral density.
Channel Power: Power within the integration bandwidth.
Power Spectral Density: Power (in dBm/Hz) normalized to 1Hz within the integration bandwidth.
Measurement Parameters: Center Freq, Integration BW, Span, Span power.
1. Center Freq
Set the center frequency, this center frequency which is the same with the center frequency of the analyzer. Modifying this parameter will change the center frequency of the analyzer.
2. Integration BW
Set the frequency width of the channel to be tested and the power of the channel is the power integral within this bandwidth. You can use the numeric keys, knob or direction keys to modify this parameter.
Table 2-34 Integration BW
Parameter
Explanation
Default
2 MHz
Range
100 Hz ~ Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Integration BW/100, the minimum is 1 Hz
SIGLENT
54 User Manual
Direction Key Step
In 1-1.5-2-3-5-7.5 sequence
3. Span
Set the frequency range of the channel. This span which is the same with the span of the analyzer is the frequency range of the sweep. Modifying this parameter will change the span of the analyzer. The channel power span is related to the integration bandwidth.
Table 2-35 Channel Power Span for Chan Power Measurement
Parameter
Explanation
Default
current span
Range
100 Hz ~ Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Channel Power Span/100, the minimum is 1Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 sequence
4. Span Power
Set the integrated bandwidth to the sweep span of display. The channel power and power spectral density display on the screen simultaneously.
2.4.2.3 ACPR
Figure 2-16 ACPR
Measurement Results: Main CH Power, Left channel power and Right channel power.
Main CH Power: Displays the power within the bandwidth of the main power
Left channel power : Displays the power of left channel and the power difference between the left
channel and the main channel (in dBc)
Right channel power: Display the power of the right channel and the power difference between the
SIGLENT
right channel and the main channel(in dBc) Measurement parameter: Center frequency, main channel bandwidth, adjacent channel bandwidth and channel spacing
1. Center Freq
Set the center frequency. The center frequency is the same with the center frequency of the analyzer display. Modifying this parameter will change the center frequency of the analyzer.
2. Main Channel
Set the bandwidth of the main channel and the power of the main channel is the power integral within this bandwidth.
Table 2-36 Main Channel
Parameter
Explanation
Default
1 MHz
Range
100 Hz ~ full span
Unit
GHz, MHz, kHz, Hz
Knob Step
Main Channel /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
3. Adjacent Chn
Set the frequency width of the adjacent channels. The adjacent channel bandwidth is related to the main channel bandwidth.
Table 2-37 Adjacent channel bandwidth
Parameter
Explanation
Default
1 MHz
Range
100 Hz ~ (full span - 2*100 Hz)
Unit
GHz, MHz, kHz, Hz
Knob Step
Adjacent Chn/10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
4. Adj Chn space
Set the difference between the center frequency of the main channel and the center frequency of the adjacent channels. Adjusting this parameter will also adjust the distance between the upper/lower channel and the main channel.
Table 2-38 adjacent channel space
Parameter
Explanation
Default
3 MHz
Range
100 Hz ~ (full span – 100 Hz) /2
SIGLENT
56 User Manual
Unit
GHz, MHz, kHz, Hz
Knob Step
Adj Chn space /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
2.4.2.4 OBW
Figure 2-17 OBW
Measurement Results: occupied bandwidth and transmit frequency error.
Occupied Bandwidth: Integrates the power within the whole span and then calculates the
bandwidth occupied by the power according to the specified power ratio.
Transmit Frequency Error: The difference between the center frequency of the channel and the
center frequency of the analyzer.
2.4.2.5 T-Power
Figure 2-18 T-Power
SIGLENT
Measurement Results: T-Power
T-Power: The power of the signal from the start line to the stop line.
Measurement Parameter: Center frequency, start line, stop line.
1. Center Frequency
Set the center frequency, this center frequency which is the same with the center frequency of the analyzer. Modifying this parameter will change the center frequency of the analyzer.
2. Start line
Set the left margin (in time unit) of T-Power measurement. The data calculated under this measurement is between the start line and stop line.
Table 2-39 start line
Parameter
Explanation
Default
0 s
Range
0 s ~ stop line
Unit
ks, s, ms, us, ns
Knob Step
Sweep time/751
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
3. Stop line
Set the right margin (in time unit) of T-Power measurement. The data calculated under this measurement is between the start line and stop line.
Table 2-40 stop line
Parameter
Explanation
Default
900 us
Range
Start line ~ sweep time
Unit
ks, s, ms, us, ns
Knob Step
Sweep time /751
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
SIGLENT
58 User Manual
2.4.2.6 TOI
Figure 2-19 TOI
TOI is an automatic measurement. There are no user controlled parameters.
2.4.2.7 Spectrum Monitor
Figure 2-20 Spectrum Monitor
Display the power of spectrum of successive scans as a color map. Also call a waterfall chart.
Measurement Parameter: Spectrogram, Restart.
1. Spectrogram: Sets the meas state of spectrum monitor.
SIGLENT
2. Restart: clear the measurement and then restart it.
2.4.2.8 CNR
Figure 2-21 CNR
Measurement Results: C/N, Carrier Power, Noise Power.
C/N: the ratio of Carrier Power to Noise Power.
Carrier Power: the total power of the carrier bandwidth.
Noise Power: the total power of the selected noise bandwidth.
Measurement Parameter: Carrier BW, Noise BW, Freq Offset.
1. Carrier BW
Set the bandwidth of the carrier to be measured.
Table 2-41 Carrier BW
Parameter
Explanation
Default
3 MHz
Range
100 Hz ~ (2*span - 2*|Freq Offset| - Noise BW)
Unit
GHz, MHz, kHz, Hz
Knob Step
Carrier BW /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
2. Noise BW
Set the bandwidth of the noise to be measured.
SIGLENT
60 User Manual
Table 2-42 Noise BW
Parameter
Explanation
Default
3 MHz
Range
100 Hz ~ (2*span - 2*|Freq Offset| - Carrier BW)
Unit
GHz, MHz, kHz, Hz
Knob Step
Noise BW /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
3. Freq Offset
Set the difference between carrier center frequency and noise center frequency.
Table 2-43 Freq Offset
Parameter
Explanation
Default
3 MHz
Range
-(span - (Carrier BW + Noise BW) /2) ~ (span - (Carrier BW + Noise BW) /2)
Unit
GHz, MHz, kHz, Hz
Knob Step
Freq Offset /10, the minimum is 1 Hz
Direction Key Step
In 1-1.5-2-3-5-7.5 Sequence
2.4.2.9 Harmonics
Figure 2-22 Harmonics
Measurement Results: each harmonic amplitude and total harmonic distortion of carrier signal. It can measure up to 10th harmonic.
Measurement Parameter:
SIGLENT
1. Fundamental
Set the frequency of the fundamental wave. If the automatic mode is turned on, the fundamental wave will be automatically found from the first scan. If the automatic mode is turned off, the user can input the fundamental frequency manually.
2. Freq Step
Set the harmonic step. In auto mode, the frequency of a harmonic is a multiple of the fundamental frequency.
3. Harmonic Num
Set the total number of the harmonics to be measured.
Table 2-44 Harmonic Num
Parameter
Explanation
Default
10
Range
2 ~ 10
Unit
None
Knob Step
1
Direction Key Step
1
4. Select Harmonic
When "All" is selected, the trace shows the fundamental wave and all harmonics in the sweep bandwidth. When 1-10 is selected, the trace shows a zero span trace corresponding to the fundamental wave or the measured harmonic.
2.4.2.10 Pause/Resume
1. Pause/resume
Pause or resume the current measurement
2. Restart
Restart measurement
SIGLENT
62 User Manual

Chapter 3 Vector Network Analyzer

This chapter introduces in detail the function keys and menu functions of the front panel under the Vector Network Analyzer Mode.

3.1 User Interface

1 2
3 4 5 976
10
17
18
19
16
14
13
12 11
8
15
Figure 3–1 Vector Network Analyzer User Interface
Table 3–1 Vector Network Analyzer User Interface
NO.
Name
Description
1
SIGLENT
SIGLENT logo
2
C/C!/C?/---
Calibration status
3 P Port Extensions indication
4
Marker active
Active Marker indication
5
Marker
Current trace active marker
6
Marker x value
Unit: frequency, frequency delta or time
7
Marker y value
Unit depend on trace format
8
Menu title
Function of the current menu.
9
Menu items
Menu items of the current function
10
Stop frequency
Stop frequency value
SIGLENT
11
Points
Measurement points number
12
Start frequency
The first frequency of a sweep
13
Mode
Mode indication
14
Trace active
Highlight active trace indication
15
Trace format
Unit depend on trace format
16
Trace scale
Unit depend on trace format
17
Trace division
Unit depend on trace format
18
Gray indication
Trace deactive
19
Trace
Active trace

3.2 Basic Settings

3.2.1 Frequency

3.2.1.1 Center Freq
Set the center frequency of the current sweep. Modifying the center frequency will modify both the start frequency and stop frequency when the span is constant (except when the start frequency or stop frequency reaches the boundary).
Table 3-2 Center Frequency
Parameter
Explanation
Default
755 MHz
[1]
Range
10.00005 MHz ~ 1.49999995 GHz
[2]
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
Span/10, min 1 Hz
Related to
Start Freq, Stop Freq
3.2.1.2 Start Freq
Set the start frequency of the current sweep. The start and stop frequencies are displayed at the bottom of the grid respectively.
Table 3-3 Start Freq
Parameter
Explanation
Default
10 MHz
[3]
Range
10 MHz ~ 1.4999999 GHz
[4]
Unit
GHz, MHz, kHz, Hz
SIGLENT
64 User Manual
Knob Step
Span/200, min 1 Hz
Direction Key Step
Span/10, min 1 Hz
Related to
Center Freq, Stop Freq
3.2.1.3 Stop Freq
Set the start frequency of the current sweep. The start and stop frequencies are displayed at the bottom of the grid respectively.
Table 3-4 Stop Freq
Parameter
Explanation
Default
1.5 GHz
[5]
Range
10.0001 MHz ~ 1.5 GHz
[6]
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
Span/10, min 1 Hz
Related to
Start Freq, Center Freq
Note:
[1] - [6] Different models have different values. Please refer to Data Sheet.

3.2.2 Span

Switch to the Span Settings menu by pressing Span on the front panel. When entering the span menu, the “Span” is selected by default. Any change of this parameter will affect the frequency parameters and restart the sweep.
3.2.2.1 Span
Set the frequency range of the current sweep.
The start and stop frequency vary with the span when the center frequency is constant.
In vector network analysis mode, the minimum sweep span is 100 Hz, and zero span cannot be
set.
When the span is set to the maximum, the analyzer enters full sweep mode.
Table 3-5 Span
Parameter
Explanation
Default
Full Span
Range
100 Hz ~ Full Span
Unit
GHz, MHz, kHz, Hz
Knob Step
Span/200, min 1 Hz
Direction Key Step
In 1-2-5 sequence
SIGLENT
Related to
Start Freq, Center Freq, Stop Freq
3.2.2.2 Full Span
Set the span of the analyzer to the maximum available frequency span.
3.2.2.3 Last Span
Set the span to the previous span setting.

3.2.3 Amplitude

Set the amplitude parameters of the analyzer. Through modifying these parameters, signals under measurement can be displayed in a proper mode for easier observation and minimum error.
3.2.3.1 Auto Scale
Automatically adjust the grid scale and reference level of the currently selected trace to optimize the trace display.
3.2.3.2 Auto Scale All
Automatically adjust the grid scale and reference level of all display traces to optimize the display of traces.
After setting “Auto Scale All”, the grid scale and reference level of different traces may be different.
In the left status bar of the screen, the grid scale and reference level of each trace are displayed
below the trace mark.
3.2.3.3 Scale
Set the vertical scale of each grid to adjust the range of amplitude that can currently be displayed. This value is also displayed in the left status bar of the screen.
Table 3-6 Scale/Div
Parameter
Explanation
Default
10 dB
Range
0.1 dB ~ 1000 dB
Unit
dB
Knob Step
0.1 dB ~ 1 dB, step = 0.01 dB
1 dB ~ 10 dB, step = 0.1 dB 10 dB ~ 100 dB, step = 1 dB 100 dB ~ 1000 dB, step = 10 dB
Direction Key Step
In 1-2-5 sequence
SIGLENT
66 User Manual
3.2.3.4 Ref Level
Set the reference level to indicate the minimum value that the current grid can display. This value is also displayed in the left status bar of the screen.
Table 3-7 Ref Level
Parameter
Explanation
Default
0 dB
Range
-1000 dB ~ 1000 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB
3.2.3.5 Ref Position
The vertical position of the currently selected trace in the screen can be adjusted by adjusting the reference position. When set to 5, the reference level of the trace is in the middle of the screen, 0 is at the bottom of the screen grid, and 10 is at the top of the screen grid.
Table 3-8 Ref Position
Parameter
Explanation
Default
5 Div
Range
0 Div ~ 10 Div
Unit
Div
Knob Step
1 Div
Direction Key Step
1 Div

3.3 Sweep and Functions

3.3.1 BW

3.3.1.1 IFBW
The VNA converts the received signal from its source to a lower intermediate frequency (IF).Reducing the IF receiver bandwidth reduces the effect of random noise on a measurement.However, narrower IF bandwidths cause longer sweep times. The default value 10 kHz.
SIGLENT

3.3.2 Trace

3.3.2.1 Select Trace
Select the trace in order to set the corresponding trace parameters. You can also select the trace by clicking on the trace mark displayed in the left status bar of the screen. By default, track 1 is selected and opened. When the trace is selected, the background color will appear on the trace mark of the left status bar on the screen.
The number of traces that can be selected is affected by the "Num of Traces" parameter. For example, if you set the "Num of Traces" to 3, traces 1 through 3 can be selected.
3.3.2.2 Num of Traces
Set the upper limit of displayed trace numbers. Up to four traces can be displayed simultaneously in the screen window.
3.3.2.3 Display
Set the display content of the currently selected trace:
Data
Only display the measured data.
Memory
Only display the memory trace. A [M] icon is displayed in the left status bar of the screen.
Data & Mem
Display both the measured data and the memory trace. A [D&M] icon is displayed in the left
status bar of the screen.
Trace Off
Neither the measured data nor the memory trace is displayed.
“Data -> Mem” operation is needed before trace can display memory data. If “Data -> Mem” is not performed first, the two options of “Memory” and “Data & Mem” are grey and not optional.
3.3.2.4 Data -> Mem
Store the measured data in memory. After "Data -> Mem" is executed, the selected trace for which measured data is displayed is provided with an additional trace, called a memory trace, that temporarily stores measured data. When selecting the trace to display "Memory" or "Data & Mem", the memory trace is displayed in the screen. The memory trace is slightly thinner than the data trace. Memory trace can be used as a reference for the data trace.
SIGLENT
68 User Manual
Figure 3-2 data trace & memory trace
3.3.2.5 Trace Hold
1. Max
Retain the maximum level for each point of the selected trace. Update the data if a new maximum level is detected in successive sweeps.
2. Min
Display the minimum value from multiple sweeps for each point of the trace and update the data if a new minimum is generated in successive sweeps.
3. Off
Erases any data previously stored in the selected trace, and display the data sampled in real-time of each point on the trace.
4. Restart
When select max or min trace hold, the operation “Restart” can empty the trace data and restart the max or min trace hold.
3.3.2.6 Math
After "Data -> Mem" is executed, you can perform complex data math between the memory trace and measured data. The following data math operations are available:
1. Data/Mem
Divide the measured data by the data in the memory trace. This function can be used to evaluate the ratio of two traces (e.g., evaluating gain or attenuation).
SIGLENT
2. Data*Mem
Multiply the measured data by a memory trace.
3. Data-Mem
Subtract a memory trace from the measured data. This function can be used, for example, to subtract a vector error that has been measured and stored (e.g., directivity) from data subsequently measured on a device.
4. Data+Mem
Add the measured data and the data in the memory trace.
5. Off
Turn off the math operations.
"Data -> Mem" must be performed first to select math operations.
Trace math operations are mutually exclusive, that is, when a math function is applied to a trace,
the last selected math function will be turned off.
3.3.2.7 Average
Turn on or off the average function of traces, and set the average times. More averages can reduce the noise and the influence of other random signals; thus highlighting the stable signal characteristics. The larger the averages times is, the smoother the trace will be.
Table 3-9 Average Times
Parameter
Explanation
Default
100
Range
1 ~ 999
Unit
N/A
Knob Step
1
Direction Key Step
10

3.3.3 Sweep

3.3.3.1 Points
The number of points is the number of data items collected in one sweep.
To obtain a higher trace resolution against the stimulus value, choose a larger value for number of points.
To obtain higher throughput, keep the number of points to a smaller value within an allowable trace resolution.
SIGLENT
70 User Manual
To obtain higher measurement accuracy after calibration, perform calibration using the same number of points as in actual measurements.
Table 3-10 Points
Parameter
Explanation
Default
201
Range
101 ~ 751
Unit
N/A
Knob Step
1
Direction Key Step
50
3.3.3.2 Sweep
Set the sweep mode in single or continuous, the default is continuous.
1. Single
Set the sweep mode to “Single”. Every time the “Single” key is pressed, a sweep is performed.
2. Continue
Set the sweep mode to “Continue”. In continuous sweep mode, the system will automatically enter into next sweep directly after each sweep.

3.3.4 TG

Set the signal amplitude of [PORT 1] output.

3.4 Marker

3.4.1 Marker

The marker can be used in the following ways:
Reading a measured value as numerical data (as an absolute value or a relative value from the reference point).
Moving the marker to a specific point on the trace (marker search).
Analyzing trace data to determine a specific parameter.
The analyzer is capable of displaying up to 4 markers including the reference marker on each trace. Each marker has a stimulus value (the value on the X-axis in rectangular display format) and one or two response value (the value on the Y-axis in rectangular display format). The Smith chart and polar formats each have two marker response values (log amplitude and phase). And the Phase format has two marker response values (phase and expanded phase) as well.
SIGLENT
3.4.1.1 Select Trace
The same function as in Trace -> “Select Trace”, please refer to section 3.3.2.1.
3.4.1.2 Select Marker
Select one of the four markers. The default is Marker1. When a marker is selected, you can set its type, trace to be marked and other related parameters. The enabled marker will appear on the trace selected through the Select Trace option and the readouts of this marker are also displayed in the active function area and at the upper right corner of the screen.
The current active marker is a filling diamond mark, while the non-active marker is an unfilled diamond mark. On the left side of the current active marker in the upper right corner of the screen, the ">" sign appears.
When marker R is opened, whether it is currently active or not, the upper right corner of the screen will display its reading.
Figure 3-3 Marker
3.4.1.3 Normal
The same function as in Marker -> “Normal” in Spectrum Analyzer Mode, please refer to section
2.3.1.3.
3.4.1.4 Delta
The same function as in Marker -> “Delta” in Spectrum Analyzer Mode, please refer to section 2.3.1.4.
SIGLENT
72 User Manual
3.4.1.5 Off
The same function as in Marker -> “Off” in Spectrum Analyzer Mode, please refer to section 2.3.1.6.
3.4.1.6 Discrete
Turning on discrete mode, a marker moves only between actual measurement points. When a specific marker stimulus value is specified as a numerical value, the marker is placed at the measurement point closest to the specified value. A marker placed between the interpolated points with the discrete mode off automatically moves to the nearest measurement point when the discrete mode is turned on. Turning off discrete mode, the marker can move from one actual measurement point to another. Because it is interpolated, it can also move in the space between measurement points.
3.4.1.7 Couple
When Marker Couple is on, Markers are set up and moved in coupled operation on all traces. When Marker Couple is off, Markers are set up and moved independently for each trace.
Figure 3-4 Marker Couple is ON
SIGLENT
Figure 3-5 Marker Couple is OFF
3.4.1.8 All Off
Turn off all the markers of all traces.

3.4.2 .Peak

3.4.2.1 Select Trace
The same function as in Trace -> “Select Trace”, please refer to section 3.3.2.1.
3.4.2.2 Select Marker
The same function as in Marker -> “Select Marker”, please refer to section 3.4.1.2.
3.4.2.3 Peak
Search the greatest measured value of the trace.
3.4.2.4 Valley
Search the smallest measured value of the trace.
3.4.2.5 Peak->CF
Execute peak search and set the center frequency of the analyzer to the frequency of the peak.
SIGLENT
74 User Manual
3.4.2.6 Valley->CF
Execute valley search and set the center frequency of the analyzer to the frequency of the valley.
3.4.2.7 Cont Peak
Turn on or off continuous peak search, which is off by default. When continuous valley search is turned on, the spectrum analyzer performs peak search once after each scan.
3.4.2.8 Cont Valley
Turn on or off continuous valley search, which is off by default. When continuous valley search is turned on, the spectrum analyzer performs valley search once after each scan.Measurement

3.4.3 Marker Fn

3.4.3.1 Select Trace
The same function as in Trace -> “Select Trace”, please refer to section 3.3.2.1.
3.4.3.2 Select Marker
The same function as in Marker -> “Select Marker”, please refer to section 3.3.2.1.
3.4.3.3 N dB BW
The same function as in Marker Fn -> “N dB BW”, please refer to section 2.3.3.3
3.4.3.4 Close
Close Marker Function

3.5 Meas Setup

3.5.1 Stimulus

Set the stimulus conditions, including the sweep range and the number of points. In Frequency, Span and Sweep menu, there are corresponding settings parameters.
SIGLENT

3.5.2 Meas

Select S11 or S21 as the current measurement item. This value is also displayed in the status bar on the left side of the screen.

3.5.3 Format

Set the display type of measurement result, enter “Format” submenu, and select the corresponding
display type. This value is also displayed in the status bar on the left side of the screen.
1. Log Mag
The trace represents the logarithmic magnitude of the measurement result, unit: dB.
2. Phase
The trace represents the phase of the measurement result, displayed in range from -180 ° to +180 °, as well as the expanded phase, can be displayed above +180 ° and below -180 °. Unit: degrees (°).
3. Group Delay
The trace represents the transmission delay of the signal through the DUT. Unit: seconds (s).
4. Smith
The Smith chart format is used to display impedances based on the reflection measurement data of the DUT. In this format, traces are plotted at the same spots as in the polar format. The Smith chart format allows users to select one of the following five data groups for displaying the marker response values.
Linear magnitude and phase (°)
Log magnitude and phase (°)
Real and imaginary parts
Resistance (ohm), Reactance (ohm), and inductance (H) or capacitance (F)
Conductance (S), susceptance (S), and capacitance (F) or inductance (H)
5. Polar
In the polar format, traces are drawn by expressing the magnitude as a displacement from the origin (linear) and phase in an angle counterclockwise from the positive X-axis. This data format does not have a stimulus axis, so frequencies must be read by using the marker. The polar format allows users to select one of the following three data groups for displaying the marker response values.
Linear magnitude and phase (°)
Log magnitude and phase (°)
Real and imaginary parts
6. Lin Mag
The trace represents the linear magnitude of the measurement result, units: 1.
7. SWR
SIGLENT
76 User Manual
The trace represents , where ρ is the reflection coefficient, units: 1.

3.5.4 Scale

Press the “Scale” key to open the Amplitude menu.

3.5.5 Trace

Press the “Trace” key to open the Trace menu.

3.5.6 Calibration

Set calibration related items. The calibration status is displayed in the upper left corner of the screen. The calibration status and display are as follows: No calibration data --- (displayed in gray) Calibrated Cor (displayed in blue) Correction Off Off (displayed in gray) Need to re-calibrate C? (displayed in blue)
Note: The calibration status is shown as “C?” which indicates that the scan frequency range is different from the time of calibration. Users need to re-calibrate in the current scan frequency range. When performing calibration, the calibration wizard pops up. Please follow the calibration wizard to perform calibration. After completing the calibration step, the calibration data is automatically saved as user calibration data. Turn on the calibration switch and the error correction function will be enabled.
Figure 3-6 Calibration Wizard
3.5.6.1 Correction
Turn on or off the calibration.
3.5.6.2 Calibrate
1. 1-Port Cal
Calibrate with the specified mechanical calibration kit, and need to connect an OPEN standard, a SHORT standard, and a LOAD standard to the [Port 1]. This calibration effectively eliminates the frequency response reflection tracking error, directivity error, and source match error from the test
SIGLENT
setup in a reflection test using that port. This function is only available when the measurement item is S11. Calibration data is saved as user calibration data.
2. Response Through
When operating, connect the port 1 and port 2 of the analyzer with an optional Through Adapter. The normalization operation moves the measurement reference plane to both ends of the Through Adapter. This function is only available when the measurement item is S21. Enter the normalized submenu to make the appropriate selection.
3.5.6.3 Ecal
Use the optional SIGLENT electronic calibration kit for calibration. Calibration data is saved as user calibration data.
3.5.6.4 Cal Kit
Specify the calibration kit used for mechanical calibration.
F503ME:Type-N 50Ω 3 GHz Cal Kit (Siglent)
85032F:Type-N 50Ω 9 GHz Cal Kit (KeySight)
85032B/E
User
Before executing calibration, you need to select a calibration kit. If you use a calibration kit other than a predefined one, you need to define it. If the connector type of the standard of the calibration kit you use has polarity (the distinction between male and female), you need to change the standard class definition of the calibration kit depending on the standard you actually use.
3.5.6.5 Port Extensions
When extending the calibrated plane to other planes (i.e. port extension) instead of the standard calibration process, the port extension function can be used to compensate for the delay (phase shift) caused by fixtures and other possible losses.
1. Extensions
Please set whether to open the port expansion function according to the actual instrument device. Close by default. After opening the port extension function, the "P" icon will be displayed on the right side of the calibration status in the upper left corner of the screen.
2. Delay Port1
Set the delay of Port1 extension. The relationship between Port Delay and Port Length is as follows:
SIGLENT
78 User Manual
Port Length = Port Delay * c0 * Velocity Factor, where c0=299792458m/s
3. Length Port1
Set the length of Port1 extension.
4. Delay Port2
Set the delay of Port2 extension.
5. Length Port2
Set the length of Port2 extension.
6. Auto Open Port1
Automatically calculate the delay and length of Port1.
3.5.6.6 System Z0
Set the System characteristic impedance (Z0).
3.5.6.7 Velocity Factor
Set the velocity factor of the cable to be measured relative to the speed of light in vacuum. Make sure that the velocity factor of the cable to be measured conforms to the reality.
Table 3-11 Velocity Factor
Parameter
Explanation
Default
0.66
Range
0.1 ~ 1
Unit
N/A
Knob Step
0.01
Direction Key Step
In 1-2-5 sequence