SIGLENT SDG5000 series Service Manual

Page 1
Service Manual
SDG5000 Series Function/Arbitrary Waveform Generator
2017 SIGLENT TECHNOLOGIES CO., LTD
Page 2
II SDG5000 Service Manual
Page 3
SDG5000 Service Manual I
Guaranty and Declaration
Copyright
SIGLENT TECHNOLOGIES CO., LTD. All Rights Reserved.
Trademark Information
SIGLENT is the registered trademark of SIGLENT TECHNOLOGIES CO., LTD
Declaration
SIGLENT products are protected by patent law in and outside of P.R.C.  SIGLENT reserves the right to modify or change parts of or all the
specifications or pricing policies at company‟s sole decision.
Information in this publication replaces all previously corresponding
material.
Any way of copying, extracting or translating the contents of this manual is
not allowed without the permission of SIGLENT.
SIGLENT will not be responsible for losses caused by either incidental or
consequential in connection with the furnishing, use or performance of this manual as well as any information contained.
Product Certification
SIGLENT guarantees this product conforms to the national and industrial
standards in china as well as the ISO9001: 2008 standard and the ISO14001: 2004 standard. Other international standard conformance certification is in progress.
Page 4
II SDG5000 Service Manual
General Safety Summary
Carefully read the following safety precautions to avoid person injury and prevent damage to the instrument or any products connected to it. To avoid potential hazards, please use the instrument as specified.
Only qualified technician should perform service procedures
Use Proper Power Line
Use only the special power line of the instrument that approved by local state.
Ground the Instrument
The instrument grounds through the protective terra conductor of the power line. To avoid electric shock, the ground conductor must be connected to the earth. Make sure the instrument is grounded correctly before connect its input or output terminals.
Connect the Signal Wire correctly
The potential of the signal wire is equal to the earth, so do not connect the signal wire to a high voltage. Do not touch the exposed contacts or components.
Look Over All Terminals’ Ratings
To avoid fire or electric shock, please look over all ratings and sign instruction of the instrument. Before connecting the instrument, please read the manual carefully to gain more information about the ratings.
Do not Operate with Suspected Failures
If you suspect that there is a damage of the instrument, please let a qualified service personnel check it.
Avoid Circuit or Components Exposed
Do not touch exposed contacts or components when the power is on.
Do not Operate in Wet/Damp conditions
Do not Operate in an Explosive Atmosphere
Keep the Surface of the Instrument Clean and Dry
Page 5
SDG5000 Service Manual III
Safety Terms and Symbols
Terms used on the instrument. Terms may appear on the instrument:
DANGER: Indicates an injury or hazard that may immediately happen. WARNING: Indicates an injury or hazard that may not immediately happen. CAUTION: Indicates a potential damage to the instrument or other property
that might occur.
Symbols used on the instrument. Symbols may appear on the instrument:
Hazardous Protective Warning Earth Ground Power
Voltage Earth Ground Switch
Page 6
IV SDG5000 Service Manual
Overview for the Document
The document is for SDG5000 series arbitrary waveform generator, which will be mostly written as generator for short in the following text. The main contents described in this manual are:
SDG5000 Series Generator at a glance
This part introduces the main technology characteristics for SDG5000 generator.
The Front Panel at a glance
This part introduces briefly all the buttons and the knob on the front panel.
The Rear Panel at a glance
This part introduces all the ports for easy communication on the rear panel.
Specification Chapter 1 lists the generator‟s specifications.
Quick Start
Chapter 2 prepares the generator for use and helps you get familiar with a few of its front-panel features.
Calibration
Chapter 3 provides calibration, verification and adjustment procedures for the generator.
Assembly Procedures
Chapter 4 provides disassembly procedures for you to get an understanding of the structure of the generator, thus to install or replace some needed modules, or troubleshoot faults you encounter while operating it.
Troubleshooting Chapter 5 provides troubleshooting procedures for the internal main board and power supply board, as well as the quick guide for general troubles. Before any operation, you should read the ESD Precautions to avoid personal injuries or damages to the generator
Maintenance
Chapter 6 provides information on maintenance, daily care and unpacking inspection of the instrument. The contact information is attached in the end in case of some unsolvable troubles you encounter.
Page 7
SDG5000 Service Manual V
Convention for the whole Contents
All the description for function and performance in this document are according to SDG5162 series generator, and apply to generator of the other types. The SDS5000 series contains the following types:
Type
Analog Bandwidth
Channel
SDG5082
80 MHz
2
SDG5122
120 MHz
2
SDG5162
160 MHz
2
Page 8
SDG5000_ServiceManual_SM0205
0-E02A
VI
SDG5000 Series at a Glance
SDG5000 Series adopt the direct digital synthesis (DDS) technology, which can provide stable, high-precision, pure and low distortion signals. Its combination of excellent system features, easiness in usage and versatile functions makes this generator a perfect solution for your testing now and in the future.
Characteristics
Reading the characteristics and specifications given below, you will understand how SDG5000 can satisfy your requirements.
DDS technology provides precise, stable and low distortional output signal. 4.3inch‟TFT color LCD display. 500MSa/s sampling rate, 14-bit resolution. Frequency characteristics:
Sine: 1μHz to 160 MHz
Square: 1μHz to 50 MHz
Ramp: 1μHz to 4MHz
Pulse: 1μHz to 40MHz White Noise: 100MHz bandwidth (-3dB) Arbitrary: 1μHz to 40MHz
5 standard waveforms: Sine, Square, Ramp, Pulse, Noise. Self-defined arbitrary waveform. Multiple modulation function, various modulated waveform: AM, DSB-AM,
FM, PM, ASK, FSK, PWM, Sweep and Burst.
Multiple I/O: external modulation source, external 10 MHz reference input,
external trigger source, waveform output, synchronous signal output.
Support USB storage device. Software updating could also be performed
using USB devices.
Up to 512k sample points of internal waveform depth, which can rebuild or
Page 9
SDG5000_ServiceManual_SM02050-E02A
VII
simulate any complex waveform.
Remote control is realized using the USB cable. Multiple interfaces: USB host & device. USB-GPIB (IEEE-488) and LAN
(option).
Support the seamless connection with SIGLENT Series Digital
Oscilloscopes; Being able to directly read and rebuild the stored waveform in the oscilloscopes.
2 languages(English and Chinese)user interface and built-in help system.
Note: All the specifications described in this manual are according to
SDG5162.
Page 10
SDG5000_ServiceManual_SM0205
0-E02A
VIII
The Front Panel at a Glance
No.
Description
No.
Description
1
Power Switch
7
Universal Knob
2
LCD Display
8
Numeric Keypad
3
Menu Operation
9
Waveform Key
4
Function Keys
10
Type
5
Output Control
11
Logo
6
Direction Keys
12
USB Host
2 3 4 5
11 10 9 8 7
6
1
12
Page 11
SDG5000_ServiceManual_SM02050-E02A
IX
The Front Panel Display at a Glance
Figure1 Display Interface (Sine Wave is the default waveform)
Here are the character definitions in this Service Manual:
The signs for buttons in this manual are the same as the panel buttons. Please note that, the signs for the functional buttons on the operation panel are represented by squared words, such as Waveforms, which represents the transparent functional key with Sine on it on the front panel, while the menu buttons are represented by brighten words such as Frequency, which means the frequency option in the Sine menu.
Function Status
Channel 1 information
Channel 2 information
Waveform Parameters
Waveform Display
Waveform type operational key
Page 12
SDG5000_ServiceManual_SM0205
0-E02A
X
The Rear Panel at a Glance
1. Ext Trig/Gate/FSK/Burst connector
2. 10MHz Output connector
3. USB Device connector
4. Power Socket
5. 10MHz Input connector
6. Sync Output connector
7. Modulation Input connector
WARNING: For protection from electric shock, the grounding power cord must not be defeated. If only a two-contact electrical outlet is available, connect the
instrument’s chassis ground screw (see above) to a good earth ground.
6 5
1 2 3 4
7
Page 13
SDG5000_ServiceManual_SM02050-E02A
XI
Contents
Guaranty and Declaration .......................................................................................................... I
General Safety Summary .......................................................................................................... II
Safety Terms and Symbols ...................................................................................................... III
Overview for the Document ..................................................................................................... IV
SDG5000 Series at a Glance .................................................................................................... VI
The Front Panel at a Glance ................................................................................................. VIII
The Rear Panel at a Glance ...................................................................................................... X
Specification .................................................................................................................................. 1
Specifications ........................................................................................................................ 1
General Specifications ........................................................................................................ 7
Quick Start ..................................................................................................................................... 8
Prepare the Generator for Use .......................................................................................... 9
Adjust the carrying Handle .............................................................................................. 10
Set the Output Frequency ................................................................................................ 11
Set the Output Amplitude ................................................................................................. 13
Set the DC offset ................................................................................................................ 14
Set the Duty Cycle of a Square Wave ........................................................................... 15
Set the Symmetry of a Ramp Wave ............................................................................... 16
Generate a Pulse Waveform ............................................................................................ 17
Generate a Noise Waveform ............................................................................................ 19
Set the DC Voltage ............................................................................................................. 21
Output a Built-In Arbitrary Waveform ........................................................................... 22
Use the Built-In Help System .......................................................................................... 23
Calibration .................................................................................................................................... 24
Calibration Interval ............................................................................................................ 24
Adjustment Recommendations ...................................................................................... 24
Automating Calibration Procedures.............................................................................. 24
Recommended Test Equipment ..................................................................................... 25
Test Considerations ........................................................................................................... 25
Performance Verification Test ......................................................................................... 26
DC Output Verification .............................................................................................. 27
AC Amplitude Verification ........................................................................................ 28
Frequency Response Verification .......................................................................... 29
General Adjustment Procedure ...................................................................................... 31
Scripts for Calibration ............................................................................................... 32
Software Environment .............................................................................................. 32
Warming up .................................................................................................................. 32
Channel Adjustment .................................................................................................. 33
Frequency Response Adjustments ....................................................................... 37
Page 14
SDG5000_ServiceManual_SM0205
0-E02A
XII
Assembly Procedures ............................................................................................................... 38
Security Consideration ..................................................................................................... 38
List of Modules ................................................................................................................... 39
Required Tools .................................................................................................................... 39
Disassembly Procedures ................................................................................................. 39
A view of the whole Instrument ...................................................................................... 40
Removing the Handle ........................................................................................................ 41
Removing the Metal Shell and Rear Cabinet ............................................................... 42
Removing the Front Cabinet ........................................................................................... 43
Removing the Display Module ........................................................................................ 44
Removing the Main Body ................................................................................................. 45
Troubleshooting ......................................................................................................................... 47
ESD Precautions ................................................................................................................ 47
Required Equipment .......................................................................................................... 47
Channel Board Drawing ................................................................................................... 48
Main Board Drawing .......................................................................................................... 49
Check the Power Supply .................................................................................................. 50
Check the Channel Board ................................................................................................ 52
Voltage Checking ........................................................................................................ 52
FPGA Checking ........................................................................................................... 53
Check the Main Board ....................................................................................................... 54
Voltage Checking ........................................................................................................ 54
Mainboard Clock Checking...................................................................................... 54
DSP/CPLD Checking ................................................................................................. 55
Quick Guide for General Failures .................................................................................. 56
Maintenance ................................................................................................................................ 57
Maintain Summary ............................................................................................................. 57
Repackaging for Shipment .............................................................................................. 58
Contact SIGLENT ............................................................................................................... 59
Page 15
SDG5000 Service Manual
1
Specification
These specifications apply to SDG5000 series Arbitrary Waveform Generator. To verify that an oscilloscope meets specifications, it must first meet the following conditions:
The generator must have been operating continuously for fifteen minutes
within the specified operating temperature.
You must perform the Self Adjust operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
The oscilloscope must be within the factory calibration interval of one year.
Specifications
Model
SDG5082
SDG5122
SDG5162
Maximum output frequency
80MHz
120MHz
160MHz
Output channels
2
Sample rate
500MSa/s
Arbitrary waveform length
CH1: 16kpts , CH2: 512kpts
Frequency resolution
1μHz
Vertical resolution
14 bits
Modulation
AM, DSB- AM, FM, PM, FSK, ASK, PWM, Sweep, Burst
Frequency counter
Frequency range100mHz ~ 200MHz
Standard interface
USB Host & Device
Dimension
W x H x D=261mm x 105mm x 344mm
Page 16
SDG5000 Service Manual
2
Frequency Specification
Model
SDG5082
SDG5122
SDG5162
Waveform
Sine, Square, Ramp, Pulse, Noise, Arbitrary waveform
Sine
1μHz ~ 80MHz
1μHz ~120MHz
1μHz ~ 160MHz
Square
1μHz ~ 30MHz
1μHz ~ 40MHz
1μHz ~ 50MHz
Pulse
1μHz ~ 20MHz
1μHz ~ 30MHz
1μHz ~ 40MHz
Ramp
1μHz ~ 2MHz
1μHz ~ 3MHz
1μHz ~ 4MHz
Gaussian Noise
100MHz (-3dB)
Arbitrary waveform
1μHz ~ 20MHz
1μHz ~ 30MHz
1μHz ~ 40MHz
Resolution
1μHz
Accuracy
within 1 year ±2ppm 0°C ~ 55°C
Square Wave
Rise/Fall time10% ~ 90%
6ns
Overshoot
< 3%
Duty Cycle
1μHz ~ 10 MHz
20% ~ 80%
10 MHz(exclude)~ 40 MHz
40% ~ 60%
40 MHz(exclude)~ 50 MHz
50%
Asymmetric50% Duty Cycle
1% of Cycle + 5 ns(typical value
Jitter (cycle to cycle)
< 100pstypical value, rms
Pulse Wave
Period
Maximum: 1000000s Minimum: 25 ns
Pulse width
12ns, resolution: 100ps
Rise/Fall time10% ~ 90%
6ns ~ 6s, resolution: 100ps
Duty Cycle
0.0001% ~ 99.9999%
Overshoot
< 3%
Jitter (cycle to cycle)
< 100pstypical value, rms
Sine Wave Spectrum Purity
Harmonic distortion DC-1 MHz
<-54dBc
1 MHz - 10 MHz
<-46dBc
10 MHz - 100 MHz
<-36dBc
100 MHz - 160 MHz
<-30dBc
Total harmonic waveform distortion
DC ~ 20 kHz
1Vpp <0.2%
Spurious signalnon-harmonic DC ~ 1 MHz
< -70dBc
1 MHz ~ 10 MHz
<-70dBc+6dB/spectrum phase
Phase noise
100kHz Offset, –116dBc / Hztypical value
Page 17
SDG5000 Service Manual
3
Ramp Wave
Linearity
< 0.1% of output Peak value typical value, 1 kHz, 1Vpp, symmetric 100%
Symmetry
0% ~ 100%
Arbitrary Waveform
Channel
CH1
CH2
Waveform length
16kpts
512kpts
Vertical resolution
14 bits
14 bits
Sample rate
500MSa/s
500MSa/s
Minimum Rise/Fall time
5nstypical value
5nstypical value
JitterRMS
2nsmaximum
2nsmaximum
Output Characteristics
Channel
CH1
CH2
Amplitude (50Ω )
1mVpp~ 10Vpp≤40MHz
1mVpp ~ 10Vpp≤40MHz
1mVpp~5Vpp 40MHz~100MHz
1mVpp~5Vpp 40MHz~100MHz
1mVpp~2.5Vpp 100MHz~130MHz
1mVpp~2.5Vpp 100MHz~130MHz
1mVpp ~ 1.5Vpp 130MHz~160MHz
1mVpp ~ 1.5Vpp 130MHz~160MHz
Vertical resolution 100 kHz sine waveform
±(0.3dBm+1mVpp)
±(0.3dBm+1mVpp)
Amplitude flatnessbased on 100kHz Sine Waveform, 1Vpp
≤10MHz, ±0.1 dB
≤10MHz, ±0.1 dB
60MHz, ±0.2 dB
60MHz, ±0.2 dB
100MHz, ±0.4 dB
100MHz, ±0.4 dB
160MHz, ±0.8 dB
160MHz, ±0.8 dB
Isolate channel depth
< -80dB
Channel delay
< 1ns
DC
Output characteristic
CH1
CH2
RangeDC
±5 V50ohm ±10 Vhigh resistance
±5 V50ohm ±10 Vhigh resistance
Offset accuracy
±(1%*|setting offset |+1 mV)
±(1%*|setting offset |+1 mV)
Page 18
SDG5000 Service Manual
4
Waveform Output
Channel
CH1/CH2
Impedance
50Ω (typical value
Protection
short-circuit protection
Isolation
The BNC connectors for channel output, synchronous and modulation input are isolated to the Re Rack Earth, and the allowable voltage range is ± 42Vpk
AM/DSB-AM ModulationCH1/CH2
Carrier
Sine, Square, Ramp, Arbitraryexcept DC
Source
Internal/External
Modulated wave
Sine, Square, Ramp, Noise, Arbitrary
Modulation depth
0% ~ 120%
Modulation frequency
1mHz~50kHz
FM ModulationCH1/CH2
Carrier
Sine, Square, Ramp, Arbitraryexcept DC
Source
Internal/External
Modulation waveform
Sine, Square, Ramp, Noise, Arbitrary
Modulation frequency
1mHz ~ 50kHz
PM ModulationCH1/CH2
Carrier
Sine, Square, Ramp, Arbitraryexcept DC
Source
Internal/External
Modulation waveform
Sine, Square, Ramp, Noise, Arbitrary
Modulation frequency
1mHz ~ 50kHz
Phase deviation
0 ~ 360°
FSK ModulationCH1/CH2
Carrier
Sine, Square, Ramp, Arbitraryexcept DC
Source
Internal/External
Modulation waveform
50% duty square waveform
Key frequency
1mHz ~ 1MHz
ASK ModulationCH1/CH2
Carrier
Sine, Square, Ramp, Arbitraryexcept DC
Source
Internal/External
Modulation waveform
50% duty square waveform
Key frequency
1mHz ~ 1MHz
PWM Modulation (CH1/CH2)
Carrier
Pulse
Source
Internal/External
Modulation waveform
Sine, Square, Ramp, Noise, Arbitrary
Modulation frequency
1mHz ~ 50kHz
SweepCH1/CH2
Carrier
Sine, Square, Ramp, Triangle, Arbitrary (except DC)
Page 19
SDG5000 Service Manual
5
Type
linear/logarithmic
Direction
Up/ down
Sweep time
1ms ~ 500s ± 0.1%
Trigger source
Manual, external, internal
BurstCH1/CH2
Waveform
Sine, Square, Ramp, Pulse, Noise, Arbitrary (except DC)
Carrier frequency
2mHz ~ 100MHz
Type
Count1 ~ 1000000 cycles, infinite, Gated
Start/Stop phase
0° ~ 360°
Internal cycle
1μs ~ 1000s ± 1%
Trigger delay
296ns ~ 34s
Gated source
External trigger
Trigger source
Manuel, External or Internal
External Modulation Input
Connector
The rear panel BNC connector, isolated to the Re Rack Earth
Input amplitude
± 4.5Vpk=100% modulation, input impedance >5kΩ
The external input voltage can‟t be over ± 5Vpk, or the instrument will get damaged.
Trigger Input
Connector
The rear panel BNC connector, rack (reference)
Voltage level input
TTL compatible
Slope
Up or down (optional)
Pulse width
> 50 ns
Input impedance
> 5kΩ, DC coupling
Response time
380nstypical value
Trigger Output
Connector
The rear panel BNC connector, rack (reference)
Voltage level
TTL compatible
Pulse width
> 60ns(typical value)
Output impedance
50Ω(typical value
Maximum frequency
1 MHz
SYNC Output
Connector
The rear panel BNC connector, isolated to the Re Rack Earth
Voltage level
TTL compatible
Pulse width
> 50 nstypical value
Output impedance
50Ω(typical value
Maximum frequency
10MHz
Page 20
SDG5000 Service Manual
6
REFCLK Input(10MHz ~ In)
Connector
The rear panel BNC connector, isolated to the Re Rack Earth and all the other connectors.
Lock range
10MHz ± 50Hz
Signal level
2.3Vpp ~ 3.3Vpp
Lock Time
< 2 s
Input impedance
1KΩ, AC coupling
REFCLK Input(10MHz ~ Out)
Connector
The rear panel BNC connector, rack (reference)
Output frequency
10MHz
Signal level
>1Vpp
Output impedance
50Ω, AC coupling
Frequency Counter
Measurement
Frequency, Period, positive/negative Pulse Width, Duty Cycle
Frequency range
Single Channel: 100mHz ~ 200 MHz
Frequency resolution
6 bits/s
Voltage range and sensitivity(non-modulated signal
Manual
DC coupling
DC deviation range
±1.5 VDC
100mHz ~ 100 MHz
50mVrms ~ ±2.5V
100 MHz ~ 200 MHz
100mVrms ~ ±2.5V
AC coupling
1 Hz ~ 200 MHz
100mVrms ~ 5Vpp
Pulse Width and Duty measurement
6Hz ~ 10MHz80mVrms ~ 5Vpp
Input adjustment Input impedance
1 MΩ
Coupling mode
AC/DC
HFR
On/Off
Trigger method
Trigger level range: -3 v ~ 1.8v
Page 21
SDG5000 Service Manual
7
General Specifications
Display
Display type
4.3‟TFT-LCD
Resolution
480×RGB×272
Color
24bit
Contrasttypical value
500:1
Backlight intensity (typical )
300cd/m2
Power
Voltage 100~240 VACRMS, 45~66 Hz, CATII
100~127 VACRMS, 45~440 Hz, CATII
Power
< 30W
Fuse
1.25A, 250V
Environment
Environmental Temperature Operation: 0°C ~ 40°C
Non-operation: -20°C ~ 60°C
Cooling method
natural cooling down
Humidity range Below +35: ≤90 relative humidity
+35 ~ +40: ≤60 relative humidity
Altitude Operation: < 3,000m
Non-operation: <15,000m
Mechanical
Dimension Width: 261mm
Height: 105mm
Depth: 344mm
Weight
N.W: 2.8 Kg
IP Protection
IP2X
Calibration Cycle
1 year
Page 22
SDG5000 Service Manual
8
Quick Start
One of the first things you will want to do with your generator is to become acquainted with the front panel. We have written the exercises in this chapter to prepare the instrument for use and help you get familiar with some of its front-panel operations. This chapter is divided into the following sections:
To Prepare the Generator for Use To Adjust the Carrying Handle To Set the Output Frequency To Set the Output Amplitude To Set a DC Offset Voltage To Set the Duty Cycle of a Square Wave To Set the Symmetry of a Ramp Wave To Configure a Pulse Waveform To Configure a Noise Waveform To Select “DC Volts” To Output a Built-In Arbitrary Waveform To Use the Built-In Help System
Page 23
SDG5000 Service Manual
9
Prepare the Generator for Use
1. Check the list of accessories
Accessories supplied with the instrument are listed below. If anything is missing or damaged, please contact your nearest SIGLENT Sales Office.
A Quick Start A Calibration Certificate A CD(EasyWave software, Datasheet and User Manual) A power cord that fits the standard of destination country A USB cable
2. Connect the power cord and turn on the generator
The instrument runs a short power-on self test, which takes a few seconds. The generator powers up in the sine wave function at 1 kHz with an amplitude
of 4Vpeak-to-peak (Channel 1). At power-on, the Output connector is disabled. To enable the Output connector, press the Output key.
If the generator does not turn on, verify that the power cord is firmly connected to the power socket on the rear panel (the power-line voltage is automatically sensed at power-on). You should also make sure that the generator is connected to a power source that is energized.
Then, verify that the generator is turned on.
Note: If the power-on self test fails, the generator may stop with black screen. For solutions, please contact the nearest SIGLENT sales office or return the generator to SIGLENT for service.
Page 24
SDG5000 Service Manual
10
Adjust the carrying Handle
To adjust the position, grasp the handle by the sides and pull outward. Then, rotate the handle to the desired position.
Pull the handle ahead for easy
carrying
Pull the handle down
Page 25
SDG5000 Service Manual
11
Set the Output Frequency
The following steps show you how to change the frequency to 20 KHz.
1. Press the Store/Recall softkey
Press the Store/Recall softkey to enter the parameter set interface. The waveform parameters including Frequency/Period, Amplitude/HighLevel, Offset/LowLevel and Phase are for you to modify using the numeric keypad.
2. Press the Frequency/Period softkey The displayed frequency is either the power-on value or the frequency
previously selected. When you change the function, the same frequency is used if the current value is valid for the new waveform. If you want to set the period, press the softkey again to switch to the period parameter
2. Input the desired frequency Use the digital keypad to input the value directly, enter the value “20”.
Page 26
SDG5000 Service Manual
12
Then press the corresponding softkey to select the desired unit. For example, press KHz.
Note: You can also enter the desired value using the knob and direction keys.
Page 27
SDG5000 Service Manual
13
Set the Output Amplitude
The following steps show you how to change the amplitude to 8Vpp.
1. Press the Ampl/HLevel softkey
The displayed amplitude is either the power-on value or the amplitude previously selected. When you change the function, the same amplitude is used if the current value is valid for the new waveform. If you want to set the HLevel for the waveform, press the softkey again to switch to the HLevel parameter.
2. Input the desired amplitude Use the digital keypad to input the value directly, enter the value “8”.
Then press the corresponding softkey to select the desired unit. For example, press Vpp.
Note: You can also enter the desired value using the knob and direction keys.
Page 28
SDG5000 Service Manual
14
Set the DC offset
The following steps show you how to change offset to 1Vdc.
1. Press the Offset/LLevel softkey
The displayed offset is either the power-on value or the offset previously selected. When you change the function, the same offset is used if the current value is valid for the new waveform. If you want to set the LLevel for the waveform, press the softkey again to switch to the LLevel parameter.
2. Input the desired offset Use the digital keypad to input the value directly, enter the value “1”.
Then press the corresponding softkey to select the desired unit. For example, press Vdc.
Note: You can also enter the desired value using the knob and direction keys.
Page 29
SDG5000 Service Manual
15
Set the Duty Cycle of a Square Wave
At power-on, the default duty cycle for square wave is 50%. You can adjust the duty cycle from 20% to 80% for output frequencies up to 50 MHz. The following steps show you how to change the duty cycle to 80%.
1. Select the square wave function
Press Square button and then select the desired output frequency to any value up to 50 MHz.
2. Press the Duty softkey
Firstly you should press Store/Recall button to enter the parameter setting interface. The displayed duty is either the power-on value or the percentage previously selected.
.
3. Input the desired duty
Use digital keypad to input the value directly, enter the value 80”, then press the corresponding softkey to select the desired unit “%”.
Page 30
SDG5000 Service Manual
16
Set the Symmetry of a Ramp Wave
At power-on, the Symmetry for Ramp wave is 50%. You can adjust the symmetry from 0% to 100%. The following steps show you how to change the symmetry to 60%.
1. Select the Ramp wave function
Press Ramp button and then select the desired output frequency to 4MHz.
2. Press the Symmetry softkey
Firstly you should press Store/Recall button to enter the parameter setting interface. The displayed symmetry is either the power-on value or the percentage previously selected.
.
3. Input the desired symmetry
Use digital keypad to input the value directly, enter the value “60”, then press the corresponding softkey to select the desired unit “%”.
Page 31
SDG5000 Service Manual
17
Generate a Pulse Waveform
You can set the generator to output a pulse waveform with variable width, edge time and delay time. The following steps show you how to generate a 500 µs pulse waveform with a pulse width of 100 µs, rise time of 8ns and delay time of 50 ns.
1. Select the pulse function
Press the Pulse button to select the pulse function and output a pulse waveform with the default parameters.
2. Set the Pulse period
Press the Frequency/Period softkey twice and then set the period to 500 µs.
3. Set the Pulse Width
Press the Width/Duty softkey and then set the pulse width to 100 µs. The pulse width represents the time from the 50% threshold of the rising edge to the 50% threshold of the next falling edge.
Page 32
SDG5000 Service Manual
18
4. Set the Edge Time
Press the Rise/Fall softkey and then set the rise edge to 8 ns.
5. Set the Pulse Delay
Press the Delay softkey and then set the delay time to 50 ns.
Page 33
SDG5000 Service Manual
19
Generate a Noise Waveform
You can set the generator to output a noise waveform with Stdev and Mean. The following steps show you how to generate a noise waveform with 500mV Stdev and 1mV Mean.
1. Select the Noise function
Press the Noise button to select the noise function and output a noise waveform with the default parameters.
2. Set the Stdev
Press the Stdev softkey and then set the Stdev to 500mV.
Page 34
SDG5000 Service Manual
20
3. Set the Mean
Press the Mean softkey and then set the mean to 1mV.
Page 35
SDG5000 Service Manual
21
Set the DC Voltage
You can set the DC Voltage feature from the Utility Menu, and then set a dc voltage as an “Offset” value. The following steps show you how to set a dc voltage with +1Vdc.
1. Select the Store/Recall function Press the Store/Recall button to enter the parameter setting interface.
2. Set the DC Voltage Press the DC button to select the DC function and output a DC waveform
with the default parameter.
3. Set the DC Offset
Press the Offset softkey and then set the offset to 1V.
Page 36
SDG5000 Service Manual
22
Output a Built-In Arbitrary Waveform
There are 36 built-in arbitrary waveforms stored in non-volatile memory. The following steps show you how to output the built-in exponential fall waveform from the front panel.
1. Set the arbitrary waveform function Press the Arb button to select the arbitrary waveform function. Then
press the Built-In softkey to enter the arbitrary waveform setting interface.
2. Set the Math waveform
Press the Math softkey to display the mathematic waveform as below.
3. Output the exponential fall waveform
Rotate the knob to select the ExpFall waveform and press the Done softkey.
The waveform is output with the present settings unless you change them.
Page 37
SDG5000 Service Manual
23
Use the Built-In Help System
The built-in help system is designed to provide context-sensitive assistance of some functions. A list of help topics is also available to assist you with several operations.
1. Read the help information for SDG5000
Press Help button in utility system, you will see a list of help topics as below. With easy operations on the front panel, you could read any help item you desired.
2. Press Cancel or any function button to exit the help system
Page 38
SDG5000 Service Manual
24
Calibration
This calibration procedure contains performance verification procedure and adjustment procedure. After receiving a generator, it is recommended to first confirm that the performance meets the published specifications by following the performance verification process. If the instrument performance is outside of the specifications, then perform the appropriate calibration adjustments to correct the issues.
Calibration Interval
The instrument should be calibrated on a regular interval determined by the measurement accuracy requirements of your application. A 1-year interval is adequate for most applications. Accuracy specifications will remain valid only if performance verification is made at regular calibration intervals. Accuracy specifications are not valid beyond the1-year calibration interval. SIGLENT does not recommend extending calibration intervals beyond 2 years for any application.
Adjustment Recommendations
Specifications are valid only within the period from the last performance verification. Whatever calibration interval you select, SIGLENT recommends that re-adjustment should always be performed within the calibration interval. This is necessary to ensure the accuracy of the performance data measured during the calibration interval.
Automating Calibration Procedures
The complete verification and adjustment procedures can be automated with the use of appropriate automated test instrumentation. The complete instrument configurations specified for each test may be programmed via the remote interface. Verification data can then be read-back from the instrument into a test program and compared to the appropriate test limit values.
The instrument can also be adjusted from the remote interface. Remote adjustment is similar to the local front-panel procedure. A PC can be used to perform the adjustment by first selecting the required function and range. The adjustment command is sent to the instrument and then the adjustment is initiated over the remote interface.
Page 39
SDG5000 Service Manual
25
Recommended Test Equipment
The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, substitute with an instrument of equivalent accuracy.
Instrument
Requirements
Recommended Model
Digital Multimeter(DMM)
DC volts accuracy: 100ppm resolution: 100 μV
Agilent 34401A/34461A Power Meter
Frequency: 10KHz to 50MHz
Agilent U2004A GPIB cable
GPIB(IEEE488)
Frequency Counter
Accuracy: 0.5ppm
Test Considerations
For optimum performance, all procedures should comply with the following recommendations:
Updating SDG‟s firmware to the newest version is strongly recommended.  Assure that the ambient temperature is stable and between 18 and 28.
Ideally, the calibration should be performed at 23℃±1.
Assure ambient relative humidity is less than 80%.
Allow a 30 minutes warm-up period before verification or adjustment.
Keep the measurement cables as short as possible, consistent with the
impedance requirements.
Use a 50Ω coaxial cable.
Page 40
SDG5000 Service Manual
26
Performance Verification Test
The performance verification tests are recommended as acceptance tests when you first receive the instrument. The acceptance test results should be compared against the specifications given in chapter 1. After acceptance, you should repeat the performance verification tests at every calibration interval.
If the instrument fails performance verification, adjustment or repair is required.
Page 41
SDG5000 Service Manual
27
DC Output Verification
This test checks if the DC offset listed in the table below are within the specified range using a DMM.
1. Set the DMM to measure DC voltage. Connect the DMM to the channel 1 output of the generator as shown below.
2. Turn on CH1 and select the DC waveform.
3. Set the instrument to each output value described in the table below, and measure the output voltage with the DMM. Be sure the generator output impedance is set to High–Z and the output is enabled.
DC Offset
CH1
CH2
Spec Range
± (1% + 2 mV)
0 mV
-2 mV ~ 2 mV
10 mV
7.9 mV ~ 12.1 mV
100 mV
97 mV ~ 103 mV
1 V
0.988 V ~ 1.012 V
3 V
2.968 V ~ 3.032 V
10 V
9.898 V ~ 10.102 V
-10m V
-12.1 mV ~ -7.9 mV
-100 mV
-103 mV ~ -97 mV
-1 V
-1.012 V ~ -0.988 V
-3 V
-3.032 V ~ -2.968 V
-10 V
-10.102 V ~ -9.898 V
4. Move the BNC cable to CH2 output and perform the same verification as
channel 1.
5. Compare the measured voltage to the spec range shown in the table above.
Page 42
SDG5000 Service Manual
28
AC Amplitude Verification
This test checks the ac amplitude output accuracy at the frequency of 10 kHz using a DMM.
1. Turn on the generator and choose CH1 as the operating channel. Set the
Load to HiZ under the CH1/CH2 menu.
2. Connect the DMM and generator as shown below.
3. Select Sine waveform of the generator and set the amplitude to the values listed below in sequence.
Amplitude (V)
CH1 (V)
CH2 (V)
Spec Range (V) ± (1%+1 mV)
11
10.889~11.111
5.6
5.543~5.657
2
1.979~2.021
0.9
0.89~0.91
0.4
0.395~0.405
0.142
0.13958~0.14442
0.064
0.06236~0.06564
0.022
0.02078~0.02322
0.01
0.0089~0.0111
0.004
0.00296~0.00504
4. Move the BNC cable to CH2 output and perform the same verification as
channel 1.
5. Compare the value measured from the DMM to the spec range shown in the
table above.
Page 43
SDG5000 Service Manual
29
Frequency Response Verification
This test checks if the amplitude flatness is within the spec range using a PC controlled Power Meter. If you do not have a PC controlled Power Meter, simply ignore the USB connections shown.
1. Turn on the generator and choose CH1 as the operating channel. Set the
Load to 50Ω under the CH1/CH2 menu.
2. Connect the Power Meter, PC and generator as shown below.
3. Select Sine waveform of the generator and set the amplitude to 1 V, and
then step through the frequencies listed below in sequence. Then, set the amplitude to 2.5V and repeat the frequency steps.
Voltage
1V
Frequency
CH1 (dBm)
CH2 (dBm)
Value
Spec Range
Value
Spec Range
10KHz
V1
3.8833 ~ 4.0744
V2
3.8833 ~ 4.0744
100KHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
1MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
5MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
10MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
20MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
30MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
50MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
80MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
100MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
110 MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
120 MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
160MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
Page 44
SDG5000 Service Manual
30
Voltage
2.5V
Frequency
CH1 (dBm)
CH2 (dBm)
Value
Spec Range
Value
Spec Range
10KHz
V1
11.8474 ~ 12.0281
V2
11.8474 ~ 12.0281
100KHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
1MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
5MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
10MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
20MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
30MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
50MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
80MHz
V1-0.2 ~ V1+0.2
V2-0.2 ~ V2+0.2
100MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
110 MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
120 MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
160MHz
V1-0.8 ~ V1+0.8
V2-0.8 ~ V2+0.8
4. Move the BNC cable to CH2 output and perform the same verification as channel 1.
5. Compare the value measured from Power Meter to the spec range shown in the table above.
Page 45
SDG5000 Service Manual
31
General Adjustment Procedure
This chapter explains how to adjust the SDG5000 series generator for optimum operating performance.
Channel Adjustment which includes three steps: feedback channel
adjustment, channel self adjustment and vertical accuracy calibration.
Feedback channel Adjustment which act as a standard calibration
module, providing the accuracy assurance for Self Adjust.
Channel Self Adjustment which includes internal control channel
adjustment and signal channel adjustment.
Vertical Accuracy Calibration which calibrates the amplitude at low
frequency.
Frequency Response Adjustment which could properly compensate the
amplitude as the frequency increases.
Page 46
SDG5000 Service Manual
32
Scripts for Calibration
SIGLENT provides a free suite for calibrating SDG series products, in which all
scripts are written in Python. Customers can download the package from the website and just run the scripts to complete the whole calibration process after some preparation works. For more information, please refer to the instructions accessible in the provided package.
Download Link: http://www.siglentamerica.com/prodcut-gjjrj.aspx?id=4705&tid=16&T=2
Software Environment
1. Python
Make sure you have installed Python 2.7 in your computer. The following modules of Python are required as well: PyVISA, PyQt4 and PyWin32.
2. Microsoft Office
As test reports are saved as „.xls‟, it is necessary that you are working with MS
Office 2003 or higher version.
3. NI VISA
The Python module, PyVISA, is based on the National Instruments VISA I/O library which can be downloaded from online by searching for PyVISA.
Warming up
Before performing the adjustment procedures, let the generator and other test equipment warm up for at least 30 minutes in an ambient temperature between 18 °C and 28 °C. Adjustments performed prior to warm-up or outside this temperature range may result in poor performance.
Page 47
SDG5000 Service Manual
33
Channel Adjustment
The channel adjustment includes three parts: feedback channel adjustment, channel self adjustment and vertical accuracy calibration.
1. Turn on the generator.
2. Connect the DMM, PC and SDG5000 generator as shown below:
3. Double click and run the script ChannelAdjust.py prepared previously.
Then change the channel connection according to the prompt message.
Note:
If there is a problem in the process of the channel adjustment, you can separately perform the three parts mentioned above. After the feedback channel adjustment has been completed, perform a self adjustment. Then perform the vertical accuracy calibration.
Page 48
SDG5000 Service Manual
34
Feedback Channel Adjustment
The internal feedback channel receives the feedback signal from all the other channels that are performing self adjustment. This adjustment acts as a standard test module and must be performed prior to Self Adjust to make sure it is performed based on a best adjusting accuracy.
1. Connect the DMM, PC and SDG5000 generator as shown below:
2. Double click and run the script FeedbackAdjust.py prepared previously.
The CH1 will be automatically turned on upon running of the script.
Page 49
SDG5000 Service Manual
35
Channel Self Adjustment
1. This adjustment is for internal control channel and signal channel. Through
feedback circuits of the feedback channel and signal channel, this adjustment helps to obtain the actual working performance for components from signal channels. After the feedback channel adjustment completed, disconnect the two BNC cables on channel 1 and channel 2.
2. Select “Self Adjust” to perform the channel Self Adjustment in Test/Cal
system under the Utility menu.
3. Once the “Self Adjust” begins, the generator turns to adjustment interface
and the progress bar displays on the screen. In about 15 seconds it will reach to 100%, which indicates completing of the self-adjust. You could press any functional button to exit.
Page 50
SDG5000 Service Manual
36
Vertical Accuracy Calibration
This adjustment aims to calibrate the amplitude at low frequency, using a DMM with high precision. The SDG5000 begins this portion of the calibration after the completion of the self adjustment process. It calibrates the Sine wave with a 10 kHz frequency and selects several amplitudes to calibrate. Compare the measured amplitude and the standard to obtain a set of compensation values.
1. Connect the DMM, PC and SDG5000 generator as shown below:
2. Double click and run the script VertAccCalib.py prepared previously.
3. Change the channel connection according to the prompt message.
Page 51
SDG5000 Service Manual
37
Frequency Response Adjustments
This adjustment aims to obtain the actual frequency response characteristics of the signal channel, thus to perform appropriate compensation. Since the waveform amplitude will decrease as the input frequency increases, it is essential to compensate for the amplitude to match it to the frequency. This adjustment should be performed after the channel adjustment.
1. Turn on the generator and connect it to PC using a USB cable.
2. Connecting the USB end of the Power Meter to the USB port of the PC,
upon which the Power Meter indicator light will be brightened. Only after the indicator light goes out from red, could it be connected to the generator output.
3. Run the script FrqRespCalib.py prepared previously. In approximately 10
seconds the prompt message will display on the screen to indicate the operator to connect the Power Meter. After the generator completes the adjustment for CH1, the prompt message displays on the screen will prompt the operator to remove the BNC cable to CH2 to start adjustment.
4. After the two channel adjustment completed, please disconnect all
connections.
Page 52
SDG5000 Service Manual
38
Assembly Procedures
This chapter describes how to remove the major modules from the SDG5000 series generator. To install the removed modules or replace new modules, please follow corresponding operating steps in reverse order.
The following contents are what mainly included in this chapter:
Security Consideration which describes security information needed to
considerate while operating.
List of Module in which the modules to remove are listed.  Required Tool which describes the tools needed to perform the
procedures
Disassembly Procedures which describes in detail how to remove and
install the modules
Security Consideration
Only qualified personnel should perform the disassembly procedures. Whenever possible, disconnect the power before removing or replacing the modules. Otherwise, any personal injuries or damages to the components may occur.
Avoid Electric Shock Hazardous voltages exist on the LCD module and power supply module. To avoid electrical shock, disconnect the power cord from the oscilloscope, and then wait at least three minutes for the capacitors in the oscilloscope to discharge before you begin disassembly.
Preventing ESD Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damages can occur at electrostatic discharge voltages as low as 50 volts. The following guidelines will help preventing ESD damage when servicing the instrument or any electronic device.
Disassemble instruments only in a static-free work area. Use a conductive work area to reduce static charges. Use a conductive wrist strap to reduce static charge accumulation.
Page 53
SDG5000 Service Manual
39
Minimize handling. Keep replacement parts in original static-free packaging. Remove all plastic, foam, vinyl, paper and other static-generating materials
from the immediate work area.
Use only anti-static solder suckers.
List of Modules
The following removable modules are listed in the order of performing disassembly procedures.
Number of Module
Module
1
Handle
2
Metal Shell and Rear Cabinet
3
Front Cabinet
4
Display Module
5
Main Body
Required Tools
Use these tools to remove or replace the modules in the oscilloscope: PH2 phillips screwdriver
Disassembly Procedures
This section describes how to remove and install the modules listed above in the generator in detail. Complete disassembly will be best achieved through the following operating steps.
Page 54
SDG5000 Service Manual
40
A view of the whole Instrument
The disassembly drawing are shown as below for you to disassemble the generator in right steps. Before disassembling, please cut the power to avoid any personal injuries or damages to inside components. Since some modules of the instrument are sharp, you should also take care while operating to prevent being scratched.
Table 4-1 A view of the whole instrument
Page 55
SDG5000 Service Manual
41
Removing the Handle
Table 4-2 Removing the handle
Removing steps: Pull down the handle hard from the two sides of the generator.
To install the handle, please operate as the reverse steps.
Page 56
SDG5000 Service Manual
42
Removing the Metal Shell and Rear Cabinet
Table 4-3 Removing the Metal Shell and Rear Cabinet
Removing steps:
1. Remove the two screws from the rear cabinet of the generator using a PH2
phillips screwdriver.
2. Remove the rear cabinet.
3. Remove the metal shell carefully from the main body to avoid being
scratched.
To install the metal shell and rear cabinet, please operate as the reverse steps.
Rear Cabinet screw
Page 57
SDG5000 Service Manual
43
Removing the Front Cabinet
Table 4-4 Removing the Front Cabinet
Removing steps:
1. Remove the four front cabinet screws using the PH2 screwdriver.
2. Firmly pull up the two screws connected to the main board from the front
panel BNC connectors.
3. Disconnect the display module cable and the USB module cable from the
main board.
4. Separate the front cabinet from the main body of the generator.
To install the front cabinet, please operate as the reverse steps.
Front Cabinet screw
BNC Connector
Page 58
SDG5000 Service Manual
44
Removing the Display Module
Table 4-5 Removing the display module screws
Removing steps:
1. Remove the six screws fixed the keypad circuit board using the PH2 screw
driver.
2. Remove the two screws fixed the USB module using the PH2 screwdriver.
3. Remove the four screws fixed the metallic support using the PH2
screwdriver.
4. Remove the keypad circuit board, silica gel soft keypad, metallic support
and LCD screen from the front cabinet in sequence.
5. Remove the front cabinet universal knob by firmly pull it out.
To install the display module, please operate as the reverse steps.
Keypad Board screw
USB screw
Metallic Support screw
Page 59
SDG5000 Service Manual
45
Removing the Main Body
Table 4-6 Disconnecting all the cables
Table 4-7 Removing all the screws
Page 60
SDG5000 Service Manual
46
Removing steps:
1. Disconnect all cables from the main board and power supply board.
2. Remove all the 20 screws fixed the main board and power supply board
using a the PH2 screwdriver.
3. Separate the main board and power supply board carefully from the main
body of the generator to avoid being scratched by the sharp metal edge.
To install the main body, please operate as the reverse steps.
Page 61
SDG5000 Service Manual
47
Troubleshooting
The internal structure of the generator consists of main board, channel board and power supply board, and they are linked through cables or connectors. This chapter explains the main checking procedures for these three boards by measuring the rated test points on them, thus to help you decide the reason for the failure you encounter while operating SDG5000 series arbitrary waveform generator.
ESD Precautions
While performing any internal test of the generator, please refer to the following precautions to avoid damages to its internal modules or components result from ESD.
Touch circuit boards by the edges as possible as you can. Reduce handling of static-sensitive modules when necessary. . Wear a grounded antistatic wrist strap to insulate the static voltage from
your body while touching these modules.
Operate static-sensitive modules only at static-free areas. Avoid handling
modules in areas that allow anything capable of generating or holding a static charge.
Required Equipment
The equipment listed in the table are required to troubleshoot the generator.
Table 5-1 Required equipment
Equipment
Critical Specifications
Example
Digital Multimeter
DC Accuracy
±0.015%
SIGLENT SDM3055
Oscilloscope
300MHz Bandwidth
SIGLENT SDS2304X
Page 62
SDG5000 Service Manual
48
Channel Board Drawing
Channel board is a kind of signal conditioning board for output analog signal. It mainly works on the adjusting of signal parameters such as frequency, amplitude. Please refer to the following drawing to quickly locate the test points on the channel board for easy resolution of the failures you encounter.
Page 63
SDG5000 Service Manual
49
Main Board Drawing
Main board is used to control and manage the whole internal system of the generator. It completes the GUI function, controlling and configuration function for channel board as well as man-machine interaction. Please refer to the following drawing to quickly locate the test points on the main board for easy resolution of the failures you encounter.
Page 64
SDG5000 Service Manual
50
Check the Power Supply
There are two power connectors through which the channel board and main board can be supplied electricity. For the channel board, there are four voltage test points on its power connector. For the main board, there is one test point.
Before performing the power supply testing procedure, please make sure that the generator is grounded correctly through the protective lead of the power cord. Take care not to touch or even disassemble the power supply module without any safety precautions, or you may probably suffer from electric shock or burn. Here are procedures for testing the power supply:
1. Disconnect the power cord of the generator and then check whether the fuse has been burnt out.
2. Remove metal shell of the generator using a driver, and then disconnect the power connector connected to the main board.
3. Focus on the Power Connector for channel board, which contains 6 pins from Pin1 to Pin6. Since two of the six pins are ground wire, you can test the other four pins that are marked with blue, yellow, red and white to check whether the voltage value is within the corresponding specified range using a digital multimeter. The voltage parameters to be tested are listed in table below:
Table 5-2 Test voltages for the channel board power connector
Voltage value
Pins
Error limit
15.5V
VH+(red)
±5%
-16.5V
VH-(white)
±5%
5.5V
VL+(yellow)
±5%
-6.5V
VL-(blue)
±5%
Table 5-3 Test voltage for the main board power connector
Voltage value
Pins
Error limit
6.5V
VH+(red)
±5%
If each tested voltage value is within the corresponding spec range referring to the table above, then the power supply works normally. Otherwise, it proves to be faulted, please return it to the factory to have it repaired or contact SIGLENT.
Page 65
SDG5000 Service Manual
51
Note: The main power supply provides an input fuse to protect against the danger of fire in the event of a failure of the power supply circuitry. However, this fuse will not fail ("open" or "blow") in normal power supply operation except that a significant overload occurs. Replace the entire main power supply assembly if the input fuse fails.
Page 66
SDG5000 Service Manual
52
Check the Channel Board
If you want to remove the main board from the metal shelf inside the generator, youd better place it on a clean, insulated mat. In addition, to avoid some chips or components on the main board being damaged for overheating, it is essential to cool the main board whenever possible using a fan. Here are procedures for testing the main board:
1. Several kinds of connectors including Fan Connector, Power Connector and
Keypad Connector are located on the main board. Check if all these connectors are connected properly.
2. Make sure that the connectors on the channel board are properly connected,
then connect the generator to AC power and turn it on. Check if the voltage values at all test points are within the spec range using a digital multimeter. The voltage parameters to be tested are listed in table below:
Voltage Checking
Test the voltage points on the channel board in the table below. If each tested voltage value is within the corresponding spec range referring to the table above, then the main board works normally. Otherwise, it proves to be faulted, please return it to the factory to have it repaired or contact SIGLENT.
Table 5-3 Test voltages of the channel board
Test point
Voltage value
Error limit
+15V
+15V
±5%
-15V
-15V
±5%
-5V
-5V
±5%
+5V
+5V
±5%
VCC3.3V
+3.3V
±5%
VCC5V
+5V
±5%
VCC1.8V
+1.8V
±5%
VCC1.2V
+1.2V
±5%
FAN_12V
+12V
±5%
PLL_3.3V
+3.3V
±5%
Page 67
SDG5000 Service Manual
53
FPGA Checking
To check if the FPGA works normally, please look at the test point marked with FPGA_LED on the channel board drawing. The LED light twinkles at the frequency of 1Hz in normal case, if it cannot be lighted or twinkles at incorrect frequency, then the FPGA may be faulted.
Page 68
SDG5000 Service Manual
54
Check the Main Board
If you want to remove the main board from the metal shelf inside the generator, youd better place it on a clean, insulated mat. In addition, to avoid some chips or components on the main board being damaged for overheating, it is essential to cool the main board whenever possible using a fan. Here are procedures for testing the main board:
1. Several kinds of connectors including Fan Connector, Power Connector and
Keypad Connector are located on the main board. Check if all these connectors are connected properly.
2. Make sure that the connectors on the main board are properly connected,
then connect the generator to AC power and turn it on. Check if the voltage values at all test points are within the spec range using a digital multimeter. The voltage parameters to be tested are listed in table below:
Voltage Checking
Test the voltage points on the main board in the table below. If each tested voltage value is within the corresponding spec range referring to the table above, then the main board works normally. Otherwise, it proves to be faulted, please return it to the factory to have it repaired or contact SIGLENT.
Table 5-4 Test voltages of the main board
Test point
Voltage value
Error limit
+16V
+16V
±5%
3.3V
+3.3V
±5%
VCC5V
+5V
±5%
VCC1.2V
+1.2V
±5%
Mainboard Clock Checking
Mainboard clock is the internal system clock of the generator. To verify if the clock on the main board works normally, please test the clock frequency listed below using an oscilloscope.
Test point
Frequency
Stability
Clock_25MHz
25MHz
±50ppm
Page 69
SDG5000 Service Manual
55
DSP/CPLD Checking
DSP and CPLD respectively represent the main control chip and programmable logic device of the main board. Only when the corresponding codes loaded successfully, can the two chips work normally.
Look at the DSP_LED/CPLD_LED light on the main board, which indicates the working state of DSP/CPLD chip. If the light turns on, then the corresponding codes have been loaded successfully and the chip is in good working state. Otherwise, there may be failure within it.
Page 70
SDG5000 Service Manual
56
Quick Guide for General Failures
The general hardware failures are described in the following. Reading the following information can help you quickly handle some easy hardware failures with more convenience.
1. No start-up after pressing the Power button:
(1) Check if the power cord is correctly connected. (2) Check if the power button is usable. (3) Check whether the fuse has been burnt out. If the fuse needs to be
changed, please contact SIGLENT as soon as possible and return the
instrument to the factory to have it repaired by qualified personnel. (4) Check if the power connector is properly connected to the main board. (5) If the instrument still does not work normally, please contact SIGLENT.
2. The instrument starts up with a dark screen:
(1) Check if the power connector is properly connected to the main board. (2) Test if the voltages on the test points marked “-5V” and “LED_ANODE”
are within the spec range. (3) If the instrument still does not work normally, please contact SIGLENT.
3. No response after pressing any button or abnormal display of the screen:
(1) Check if the two end of the connector between the keypad circuit board
and the main board is properly connected.
(2) If the instrument still does not work normally, please contact SIGLENT.
Page 71
SDG5000 Service Manual
57
Maintenance
Maintain Summary
SIGLENT warrants that the products it manufactures and sells are free from
defects in materials and workmanship for a period of three years from the date of shipment from an authorized SIGLENT distributor. If a product or CRT proves defective within the respective period, SIGLENT will provide repair or replacement as described in the complete warranty statement.
To arrange for service or obtain a copy of the complete warranty statement, please contact your nearest SIGLENT sales and service office.
Except that as provided in this summary or the applicable warranty Statement, SIGLENT makes no warranty of any kind, express or implied, including without limitation the implied warranties of merchantability and fitness for a particular purpose. In no case shall SIGLENT be liable for indirect, special or consequential damages.
Page 72
SDG5000 Service Manual
58
Repackaging for Shipment
If the unit needs to be shipped to SIGLENT for service or repair, be sure to:
1. Attach a tag to the unit identifying the owner and indicating the required
service or repair.
2. Place the unit in its original container with appropriate packaging material
for shipping.
3. Secure the container with strong tape or metal bands.
If the original shipping container is not available, place your unit in a container which will ensure at least 4 inches of compressible packaging material around all sides for the instrument. Use static-free packaging materials to avoid additional damage to your unit.
Page 73
SDG5000 Service Manual
59
Contact SIGLENT
North America
SIGLENT Technologies America, Inc
6557 Cochran Rd Solon, Ohio 44139 Tel: 440-398-5800 Toll Free:877-515-5551 Fax: 440-399-1211
info@siglent.com www.siglentamerica.com
Headquarters
SIGLENT TECHNOLOGIES CO., LTD.
Blog No.4 & No.5, Antongda Industrial Zone, 3rd Liuxian Road, Bao‟an District, Shenzhen, 518101, China. Tel:+ 86 755 3661 5186 Fax:+ 86 755 3359 1582
sales@siglent.com www.siglent.com/ens
Europe
SIGLENT TECHNOLOGIES EUROPE GmbH
Liebigstrasse 2-20, Gebaeude 14, 22113 Hamburg Germany Tel: +49(0)40-819-95946 Fax: +49(0)40-819-95947
info-eu@siglent.com www.siglenteu.com
Loading...