Carefully read the following safety precautions to avoid any personal injuries
or damages to the instrument and any product connected to it. To avoid
potential hazards, please use the instrument as specified.
Only qualified technical personnel should service this instrument.
Avoid fire or open flame.
Use properly rated power line connections.
Use only the specified power line which has been approved by your local
regulatory agency.
Ground the Instrument.
The instrument is grounded through the protective ground conductor of the
power line. To avoid electric shock, the ground conductor must be connected
to the earth ground. Make sure the instrument is grounded correctly before
connecting its input or output terminals.
Connect the signal wire correctly.
The potential of the signal wire ground is equal to the earth, therefore do not
connect the signal wire to a high voltage. Do not touch the exposed contacts
or components.
Observe all terminal ratings.
To avoid fire or electric shock, please observe all ratings and sign instructions
on the instrument. Before connecting the instrument, please read the manual
carefully to gain more information about the ratings.
Do not operate with suspected failures.
If you suspect that the product is damaged, please let only qualified service
personnel check it.
Avoid circuit or wire exposure.
Do not touch exposed contacts or components when the power is on.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep the surface of the instrument clean and dry.
SDG2000X User Manual III
SIGLENT
Safety Terms and Symbols
Terms used on the instrument. Terms may appear on the instrument:
DANGER: Indicates an injury or hazard that may immediately happen.
WARNING: Indicates an injury or hazard that may not immediately happen.
CAUTION: Indicates that a potential damage to the instrument or other
property might occur.
Symbols used on the instrument. Symbols may appear on the instrument:
Hazardous Protective Warning Chassis Power
Voltage Earth Ground Ground Switch
IV SDG2000X User Manual
SIGLENT
Introduction of SDG2000X
The manual covers the following 3 models of SDG2000X Series
Function/Arbitrary Waveform Generators: SDG2042X, SDG2082X and
SDG2122X.
SIGLENT‘s SDG2000X is a series of dual-channel function/arbitrary waveform
generators with specifications of up to 120MHz maximum bandwidth,
1.2GSa/s sampling rate and 16-bit vertical resolution. The proprietary TrueArb
& EasyPulse techniques help to solve the weaknesses inherent in traditional
DDS generators when generating arbitrary, square and pulse waveforms.
Using these techniques the SDG2000X provides users with a variety of high
fidelity, low jitter signals in order to meet the growing requirements for a
multitude of complex applications.
Key Features
Dual-channel, 120MHz maximum bandwidth, 20Vpp maximum output
amplitude, output with 80dB dynamic range
High-performance sampling system with 1.2GSa/s sampling rate and
16-bit vertical resolution. No detail in your waveforms will be lost
Innovative TrueArb technology, based on a point-by-point architecture,
supports any 8pts~8Mpts Arb waveform with a sampling rate in range of
1μSa/s~75MSa/s
Innovative Easy Pulse technology, capable of generating lower jitter
Square or Pulse waveforms, brings a wide range and extremely high
precision in pulse width and rise/fall times adjustment
A variety of analog and digital modulation types: AM、DSB-AM、FM、PM、
FSK、ASK 、PSK and PWM
Sweep and Burst functions
SDG2000X User Manual V
SIGLENT
Harmonic waveforms generating function
Waveforms combining function
High precision Frequency Counter
196 kinds of built-in arbitrary waveforms
Standard interfaces: USB Host, USB Device(USBTMC), LAN(VXI-11)
Optional interface: GPIB
4.3‖ touch screen display for easier operation
VI SDG2000X User Manual
SIGLENT
Catalog
General Safety Summary ............................................................................... III
Introduction of SDG2000X ............................................................................. V
Appendix B: Daily Maintenance and Cleaning ..................................... 162
SDG2000X User Manual IX
1 Quick Start
This chapter covers the following topics:
Handle Adjustment
The Front/Rear Panel
To Select a Waveform
To Set Modulation/Sweep/Burst
To Turn On/Off Output
To Use Numeric Input
SIGLENT
To Use Common Function Keys
SDG2000X User Manual 1
SIGLENT
1.1 Handle Adjustment
To adjust the handle position of SDG2000X, please grip the handle by the
sides and pull it outward. Then, make the handle rotate to the desired position.
Figure 1-1 Viewing Position and Carrying Position
2 SDG2000X User Manual
SIGLENT
USB
Host
Numeric Keyboard
Power
Key
User Interface
Menu
Softkeys
Function
Keys
Channel
Control
Arrow
Keys
1.2 The Front/Rear Panel
This chapter will provide a brief introduction and description for the operation
and functions of the front/rear panel.
Front Panel
SDG2000X has a clear and simple front panel which includes 4.3 inch touch
screen, menu softkeys, numeric keyboard, knob, function keys, arrow keys,
and channel control area, etc.
Figure 1-2 Front Panel of SDG2000X
SDG2000X User Manual 3
SIGLENT
Aux
In/Out
10 MHz Clock Input/Output
Earth Terminal
LAN Interface
USB Device
AC Power Supply
Input
Counter
Rear Panel
The rear panel provides multiple interfaces, including Counter, 10MHz In/Out,
Aux In/Out, LAN, USB Device, Earth Terminal and AC Power Supply Input.
Figure 1-3 Rear Panel of SDG2000X
4 SDG2000X User Manual
SIGLENT
1 2
4 3 8 7 6 5
Touch Screen Display
SDG2000X can only display parameters and waveform of one channel at a
time. The picture below shows the interface when CH1 chooses AM
modulation of sine waveform. The information displayed may vary depending
on the function selected.
The entire screen of the SDG2000X is a touch screen. You can use your figure
or touch pen to control the instrument. Most functions and selections can be
chosen using the touch screen in a similar manner to the front panel keys and
knob.
Figure 1-4 Touch Screen Display
1. Waveform Display Area
Displays the currently selected waveform of each channel.
2. Channel Status Bar
Indicates the selected status and output configuration of the channels.
SDG2000X User Manual 5
SIGLENT
3. Basic Waveform Parameters Area
Shows the current waveform‘s parameters of each channel. Press Parameter
and select the corresponding softkey to highlight the parameter to configure.
Then use number keys or knob to change the parameter value.
4. Channel Parameters Area
Displays the load and output settings of the currently selected channel.
Load ----Value of the output load, as selected by the user.
Press Utility → Output → Load, then use the softkeys, number keys or knob to
change the parameter value; or continue pressing the corresponding output
key for two second to switch between High Impedance and 50Ω.
High Impedance: display HiZ.
Load: display impedance value (the default is 50Ω and the range is 50Ω to
100kΩ).
Note: This setting does not actually change the instrument‘s output impedance
of 50Ω but rather is used to maintain amplitude accuracy into different load
values.
Output ----Channel output state.
After pressing corresponding channel output control port, the current channel
can be turned on/off.
5. LAN Status Icon
SDG2000X will show different prompt messages based on the current network
status.
This mark indicates LAN connection is successful.
This mark indicates there is no LAN connection or LAN connection is
unsuccessful.
6. Mode Icon
SDG2000X will show different prompt messages based on the current mode.
This mark indicates current mode is Phase-locked.
6 SDG2000X User Manual
SIGLENT
This mark indicates current mode is Independent.
7. Menu
Shows the menu corresponding to the displayed function. For example, Figure
1-4 Touch Screen Display shows the parameters of ―AM modulation‖.
8. Modulation Parameters Area
Shows the parameters of the current modulation function. After selecting the
corresponding menu, use number keys or knob to change the parameter
value.
SDG2000X User Manual 7
SIGLENT
1.3 To Select a Waveform
Press Waveforms to enter the menu as Figure 1-5 shows. The example below
will help familiarize with the waveform selection settings.
Figure 1-5 Waveform Selections
1. Press Waveforms key and then press Sine softkey. The SDG2000X can
generate sine waveforms with frequencies from 1μHz to 120MHz. By
setting Frequency/Period, Amplitude/High level, Offset/Low level and
Phase, a sine signal with different parameters can be generated.
Figure 1-6 Sine Display Interface
2. Press Waveforms key and then press Square softkey. The generator can
generate square waveforms with frequencies from 1μHz to 25MHz and
variable duty cycle. By setting Frequency/Period, Amplitude/High level,
Offset/Low level, Phase and DutyCycle, a square waveform with different
parameters can be generated.
8 SDG2000X User Manual
SIGLENT
Figure 1-7 Square Display Interface
3. Press Waveforms key and then press Ramp softkey. The generator can
generate ramp waveforms with frequencies from 1μHz to 1MHz and
variable symmetry. By setting Frequency/Period, Amplitude/High level,
Offset/Low level, Phase and Symmetry, a ramp waveform with different
parameters can be generated.
Figure 1-8 Ramp Display Interface
4. Press Waveforms key and then press Pulse softkey. The generator can
generate pulse waveforms with frequencies from 1μHz to 25 MHz and
variable pulse width and rise/fall times. By setting Frequency/Period,
Amplitude/High level, Offset/Low level, PulWidth/Duty, Rise/Fall and Delay,
SDG2000X User Manual 9
a pulse waveform with different parameters can be generated.
Figure 1-9 Pulse Display Interface
SIGLENT
5. Press Waveforms key and then press Noise softkey. The generator can
generate noise with bandwidth from 20MHz to 120MHz. By setting Stdev
and Mean, noise with different parameters can be generated.
Figure 1-10 Noise Display Interface
6. Press Waveforms key and then press Page 1/2, last press the DC softkey.
The generator can generate a DC signal with a level up to ±10V into a
HighZ load or ±5V into a 50Ω load.
10 SDG2000X User Manual
SIGLENT
Figure 1-11 DC Display Interface
7. Press Waveforms key and then press Page 1/2, last press the Arb softkey.
The generator can generate repeatable arbitrary waveforms with length
from 8 to 8M points and frequencies up to 20MHz. By setting
Frequency/Period, Amplitude/High level, Offset/Low level, Phase and Arb
Mode, an arbitrary signal with different parameters can be generated.
Figure 1-12 Arbitrary Waveform Display Interface
SDG2000X User Manual 11
SIGLENT
1.4 To Set Modulation/Sweep/Burst
As shown in Figure 1-13, there are three keys on the front panel which are
used for modulation, sweep and burst settings. The instructions below will help
to explain these functions.
Figure 1-13 Modulate/Sweep/Burst Key
1. Press Mod, the Modulation function will be enabled.
The modulated waveform can be changed by modifying the parameters such
as Type, Source, AM Depth, AM Freq, Shape, etc. The SDG2000X can
modulate waveforms using AM, FM, PM, ASK, FSK, PSK, PWM and DSB-AM,
etc. Pulse waveforms can only be modulated using PWM. Noise and DC
waveforms cannot be modulated.
Figure 1-14 Modulation Display Interface
2. Press Sweep, the Sweep function will be enabled.
Sine, square, ramp and arbitrary waveforms support the sweep function. In
sweep mode, the SDG2000X can generate signals with variable frequency.
The available range of sweep time is from 1ms to 500s. The trigger source can
be ―Internal‖, ―External‖ or ―Manual‖.
12 SDG2000X User Manual
Figure 1-15 Sweep Waveform Display Interface
3. Press Burst, the Burst function will be enabled.
SIGLENT
Burst signals for sine, square, ramp, pulse or arbitrary waveforms may be
generated. Start Phase ranges from 0° to 360° and Burst Period ranges from
1μs to 1000s.
Figure 1-16 Burst Waveform Display Interface
SDG2000X User Manual 13
SIGLENT
1.5 To Turn On/Off Output
As shown in Figure 1-17 there are two keys on the right side of the operation
panel which are used to enable / disable the output of the two channels.
Choose a channel and press the corresponding Output key, the key backlight
will be lighted and the output will be enabled. Press the Output key again, the
key backlight will be extinguished and the output will be disabled.
Keep pressing the corresponding output key for two seconds to switch
between High Impedance and 50Ω load.
Figure 1-17 Output Keys
14 SDG2000X User Manual
SIGLENT
1.6 To Use Numeric Input
As shown in Figure 1-18, there are three sets of keys on the front panel, which
are arrow keys, knob and numeric keyboard. The instructions below will help
to familiarize you with the digital input selection.
Figure 1-18 Front Panel Digital Input
1. The numeric keyboard is used to enter the parameter‘s value.
2. The knob is used to increase (clockwise) or decrease (counterclockwise)
the current digit when setting parameters
3. When using knob to set parameters, the arrow keys are used to select the
digit to be modified; When using numeric keyboard to set parameters, the
left arrow key is used as a Backspace function.
SDG2000X User Manual 15
SIGLENT
1.7 To Use Common Function Keys
As shown in Figure 1-19, there are five keys on the operation panel, which are
labeled Parameter, Utility, Store/Recall, Waveforms, and Ch1/Ch2. The
instructions below will help to familiarize you with these functions.
Figure 1-19 Waveforms Utility and Parameter Key
1. The Parameter key makes it convenient for the operator to set the
parameters of basic waveforms directly.
2. The Utility key is used to set the auxiliary system function, such as output
configurations, interface setting, system setting information, performing
the instrument self-test and reading the calibration information, etc.
3. The Store/Recall key is used to store and recall waveform data and
configuration information.
4. The Waveforms key is used to select basic waveforms.
5. The Ch1/Ch2 key is used to switch the currently selected channel
between CH1 and CH2. After start-up, CH1 is selected as default. At this
point, press the key to select CH2.
16 SDG2000X User Manual
SIGLENT
2 Front Panel Operations
Up to now, you have got a brief understanding about SDG2000X with the
front/rear panel, every function control area and keys. You should also know
how to set your Function/Arbitrary Waveform Generator for your usage. If you
are not familiar with these operations, you are suggested to read chapter one
‗Quick Start‘ again.
This chapter covers the following topics:
To Set Sine
To Set Square
To Set Ramp
To Set Pulse
To Set Noise
To Set DC
To Set Arbitrary
To Set Harmonic Function
To Set Modulation Function
To Set Sweep Function
To Set Burst Function
To Store and Recall
To Set Utility Function
It is recommended that you read this chapter carefully so as to understand the
SDG2000X‘s versatile waveform setting functions and additional operation
methods.
SDG2000X User Manual 17
SIGLENT
FuSe
Explanations
Frequency/
Period
Set the signal frequency or period;
The current parameter will be switched at a second press.
Amplitude/
HighLevel
Set the signal amplitude or high level;
The current parameter will be switched at a second press.
Offset/
LowLevel
Set the signal offset or low level;
The current parameter will be switched at a second press.
Phase
Set the phase of the signal.
2.1 To Set Sine Waveform
Press Waveforms key to select the waveform function and then press the Sine
softkey. The sine waveform parameters are set by using the sine operation
menu.
The parameters available for sine waveforms include frequency/period,
amplitude/high level, offset/low level and phase. Different sine signals can be
generated by setting these parameters. As shown in Figure 2-1, in the soft key
menu, select Frequency. The frequency parameter area is highlighted in the
parameter display window, and users can set the frequency value here.
Figure 2-1 Sine Parameters Display Interface
Table 2-1 Menu Explanations of Sine Waveform
18 SDG2000X User Manual
SIGLENT
To Set the Frequency/Period
Frequency is one of the most important parameters of basic waveforms. For
different instrument models and waveforms, the available ranges of frequency
are different. For detailed information, please refer to ―SDG2000X Datasheet‖.
The default frequency is 1kHz.
1. Press Waveforms → Sine → Frequency, to set the frequency parameter.
The frequency shown on the screen when the instrument is powered on is the
default value or the set value of last power down. If Period (rather than
Frequency) is the desired parameter, press Frequency/Period again to enter
the Period mode. The current value for the waveform‘s period is now
displayed in inverse color. Press the Frequency/Period key once again to
return to the Frequency entry mode.
2. Input the desired frequency.
Use the numeric keyboard to input the parameter value directly, and press the
corresponding key to select the parameter unit. Or use the arrow keys to
select the digit to edit, and then use the knob to change its value.
Figure 2-2 Setting the Frequency
Note:
When using the numeric keyboard to enter the value, the left arrow key can be
used to move the cursor backward and delete the value of the previous digit.
SDG2000X User Manual 19
SIGLENT
To Set the Amplitude
The amplitude setting range is limited by the ―Load‖ and ―Frequency/Period‖
settings. For detailed information, please refer to ―SDG2000X Datasheet‖.
1. Press Waveforms → Sine → Amplitude, to set the amplitude.
The amplitude shown on the screen when the instrument is powered on is the
default value or the set value of last power down. If setting the waveform‘s
high level is desired, press the Amplitude/HighLevel key again to switch into
the high level parameter (the current operation is displayed in inverse color).
2. Input the desired amplitude.
Use the numeric keyboard to input the parameter value directly, and press the
corresponding key to select the parameter unit. Or use the arrow keys to
select the digit to edit, and then use the knob to change its value.
Figure 2-3 Setting the Amplitude
To Set the Offset
The offset setting range is limited by the ―Load‖ and ―Amplitude/HighLevel‖
settings. For detailed information, please refer to ―SDG2000X Datasheet‖. The
default value is 0Vdc.
20 SDG2000X User Manual
Loading...
+ 142 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.