Sierra Monitor Corporation 5100-03-IT, 5100-05-IT, 5100-04-IT, 5100-06-IT, 5100-10-IT User Manual

...
Page 1
Manual # T12020
Document Revision: D4
Model 5100-XX-IT
IT Series
Toxic Gas Sensor Module
5100-03-IT 5100-04-IT 5100-05-IT (FM Approved) 5100-06-IT 5100-08-IT 5100-10-IT 5100-12-IT 5100-21-IT 5100-25-IT 5100-26-IT 5100-88-IT
APPLICABILITY & EFFECTIVITY
Effective for all systems manufactured after May 2017.
Page 2
Model 5100-XX-IT Toxic Gas Sensor Module
Page 2 of 62
©2017 Sierra Monitor Corporation. All rights reserved. Manual Number: T12020 This manual contains intellectual property that was developed by Sierra Monitor Corporation and is
protected by the copyright laws of the United States, international copyright treaties, and all other applicable national laws. This manual contains information regarding technology that is protected under one or more issued or pending United States and foreign patents.
Sierra Monitor and Sentry IT are trademarks of Sierra Monitor Corporation. All other trademarks are the property of their respective owners.
This manual is licensed to users of Sierra Monitor products and may be copied, in whole or in part. The offices of Sierra Monitor are located at:
Sierra Monitor Corporation 1991 Tarob Court Milpitas, CA 95035 USA
Customer Service: Telephone: (408) 262-6611
Fax: (408) 262-9042 Toll-Free: (800) 727-4377
Email: support@sierramonitor.com Website: www.sierramonitor.com
Page 3
Model 5100-XX-IT Toxic Gas Sensor Module
Page 3 of 62
FM PERFORMANCE APPROVAL
ONLY THE FOLLOWING ITEMS, FUNCTIONS AND OPTIONS ARE FM* APPROVED
Model 5100-05-IT Hydrogen Sulfide Gas Sensor Module
Sensor Module
Model 5100-05-IT
Sensor Module - Hydrogen Sulfide
Calibration Equipment
Model 1250-01
Gas Sensor Calibration Kit, Type A
Model 1260-05
Hydrogen Sulfide in N2, 25 ppm Gas Cylinder
Model 1260-45
Hydrogen Sulfide in N2, 10 ppm Gas Cylinder
Model 5358-01
Calibration Adapter
Notes:
1) Apparatus must be installed in accordance with National Electrical Code.
2) FM Comments *FM Approvals, a subsidiary of FM Global
Project# 3034435
Page 4
Model 5100-XX-IT Toxic Gas Sensor Module
Page 4 of 62
TABLE OF CONTENTS
Product Description ...................................................................................................................... 7
1.1 General ......................................................................................................................................................... 7
1.2 Product Configuration .................................................................................................................................. 8
1.3 Theory of Operation ..................................................................................................................................... 8
1.4 Modes of Operation ...................................................................................................................................... 8
1.4.1 Sentry Interface ..................................................................................................................................... 8
1.4.2 Modbus Operation ................................................................................................................................ 8
1.4.3 Analog Operation .................................................................................................................................. 8
1.4.4 Remote Sensor ..................................................................................................................................... 8
1.4.5 HART Connection ................................................................................................................................. 9
1.4.6 Remote Alarm Reset ............................................................................................................................. 9
1.4.7 Optional Integral Relays ........................................................................................................................ 9
1.5 Mechanical ................................................................................................................................................... 9
1.5.1 Enclosure .............................................................................................................................................. 9
1.5.2 Transmitter Electronics ....................................................................................................................... 11
1.5.3 Sensor Assembly ................................................................................................................................ 11
1.6 Interconnect Wiring .................................................................................................................................... 11
1.7 Power Requirements .................................................................................................................................. 11
Cautions & Warnings .................................................................................................................. 12
2.1 Introduction ................................................................................................................................................. 12
2.2 Gas Sensor Modules – General ................................................................................................................. 12
2.3 Wiring Warnings ......................................................................................................................................... 12
2.4 Calibration Frequency ................................................................................................................................ 12
Quick Start ................................................................................................................................... 13
3.1 Overview .................................................................................................................................................... 13
3.2 Wiring ......................................................................................................................................................... 13
3.3 Module Installation ..................................................................................................................................... 13
3.4 Transmitter Installation ............................................................................................................................... 13
3.5 Startup and Operation ................................................................................................................................ 14
3.6 Zero Stabilization ....................................................................................................................................... 14
Installation ................................................................................................................................... 15
4.1 Sensor Module Locations ........................................................................................................................... 15
4.2 Wiring ......................................................................................................................................................... 16
4.2.1 Analog 4-20 mA Operation ................................................................................................................. 16
4.2.2 Modbus Operation Using RS-485 Connection.................................................................................... 16
4.2.2.1 Termination Resistor Jumpers .................................................................................................... 16
4.2.2.2 Bias Jumpers (BIAS A, BIAS B) ................................................................................................. 16
4.2.3 Sentry Operation Using Sentry Connection ........................................................................................ 16
4.2.4 General ............................................................................................................................................... 17
4.3 Enclosure Installation ................................................................................................................................. 17
4.4 Transmitter and Sensor Installation ........................................................................................................... 18
4.5 Module Address Switch .............................................................................................................................. 19
Operation ..................................................................................................................................... 25
5.1 Introduction ................................................................................................................................................. 25
5.2 Data Entry Keypad ..................................................................................................................................... 25
5.3 Main Menu .................................................................................................................................................. 26
5.4 Set-Up ........................................................................................................................................................ 27
5.5 Maintenance Sub-Menu ............................................................................................................................. 29
Calibration ................................................................................................................................... 30
6.1 Calibration Frequency ................................................................................................................................ 30
6.2 Calibration Preparation .............................................................................................................................. 30
6.3 Calibration Gas Delivery Methods.............................................................................................................. 30
Page 5
Model 5100-XX-IT Toxic Gas Sensor Module
Page 5 of 62
6.4 Sensor Exposure to Gas ............................................................................................................................ 31
6.5 Calibration Sub-Menu ................................................................................................................................ 32
6.5.1 5100-03-IT Oxygen Deficiency ........................................................................................................... 33
6.5.2 5100-88-IT Carbon Dioxide................................................................................................................. 33
Appendix A. Service .......................................................................................................................... 34
Appendix A.1. Sensor Module Configuration ........................................................................................................ 34
Appendix A.2. Enclosure Replacement ................................................................................................................ 34
Appendix A.3. Transmitter Replacement .............................................................................................................. 35
Appendix A.4. Sensor Replacement ..................................................................................................................... 35
Appendix A.5. Installation Inspection .................................................................................................................... 36
Appendix A.5.1. Controller Installation .............................................................................................................. 36
Appendix A.5.2. Cabling Installation ................................................................................................................. 36
Appendix A.5.3. Sensor Module Installation ..................................................................................................... 36
Appendix A.5.4. Moisture Traps and Rainshields ............................................................................................. 36
Appendix A.5.5. Standard Voltages .................................................................................................................. 36
Appendix A.6. Inspection and Troubleshooting Guide.......................................................................................... 36
Appendix A.6.1. Ir Module Does Not Respond to Gas ...................................................................................... 36
Appendix A.6.2. Module Displays “Starting” for more than 1 Hour ................................................................... 36
Appendix A.6.3. Module Does Not Display the Correct %PPM ........................................................................ 37
Appendix A.6.4. Display Shows ‘F’ – Sensor Missing or Bad Sensor .............................................................. 37
Appendix A.6.5. Module Shows “***Calibration Required***” ........................................................................... 37
Appendix A.6.6. Display Shows “H” or “L” ......................................................................................................... 37
Appendix A.6.7. Display Shows “M” – Not Calibrated ....................................................................................... 37
Appendix A.6.8. Display Shows “C” – Calibration Mode ................................................................................... 37
Appendix A.6.9. Display Shows “S” – Sentry Connection ................................................................................. 37
Appendix A.6.10. Diagnostic LEDs ................................................................................................................... 37
Appendix B. Specifications ............................................................................................................... 38
Appendix C. Model Numbers and Parts List .................................................................................... 41
Appendix D. Limited 2 Year Warranty .............................................................................................. 42
Appendix E. Remote Sensor Drawing .............................................................................................. 43
Appendix F. Modbus Memory Map ................................................................................................... 44
Appendix G. HART ............................................................................................................................. 45
Appendix H. Gas Sensor Module Calibration .................................................................................. 52
Appendix H.1. Calibration for 5100-25-IT ............................................................................................................. 52
Appendix H.2. Calibration for 5100-26-IT ............................................................................................................. 56
Appendix I. 5100-88-IT Dimensions .................................................................................................. 57
Appendix J. 5100-05-IT FM Performance Approval ......................................................................... 58
Appendix K. 5100-XX-IT SIL-2 Certificates ....................................................................................... 60
Appendix L. 5100-05-IT ATEX Certificate ......................................................................................... 60
Page 6
Model 5100-XX-IT Toxic Gas Sensor Module
Page 6 of 62
LIST OF FIGURES
Figure 1: Model 5100-XX-IT Toxic Gas Sensor Module – Mounting Options ............................................................9
Figure 2: Model 5100-05-IT-S1/S2 Toxic Gas Sensor – Stainless Steel Enclosure, Dimensions .......................... 10
Figure 3: Model 5100-05-IT-A1/A2 Toxic Gas Sensor – Cast Aluminum Enclosure, Dimensions ......................... 10
Figure 4: Specific Gravities ...................................................................................................................................... 15
Figure 5: Typical Mounting Options ......................................................................................................................... 15
Figure 6: Minimum Wire Gauges ............................................................................................................................. 17
Figure 7: Sensor Module Address Switch Positions ................................................................................................ 19
Figure 8: Transmitter Face Plate ............................................................................................................................. 19
Figure 9: Interface Board Connectors ..................................................................................................................... 20
Figure 10: RS-485 – Termination, BIAS Jumper ..................................................................................................... 20
Figure 11: mA Circuit Types .................................................................................................................................... 21
Figure 12: 4-20 mA Circuit Type Connections for 5100-XX-IT ................................................................................ 22
Figure 13: Wiring Connections for Modbus and Sentry Interface ........................................................................... 23
Figure 14: Wiring Connections for Remote Alarm Reset ........................................................................................ 24
Figure 15: Face Plate with Operator Keypad .......................................................................................................... 25
Figure 16: Master Menu .......................................................................................................................................... 26
Figure 17: Operation Display Values ....................................................................................................................... 26
Figure 18: Set-Up Configuration Part 1 ................................................................................................................... 28
Figure 19: Set-Up Configuration Part 2 ................................................................................................................... 29
Figure 20: Maintenance Menu ................................................................................................................................. 29
Figure 21: Model 5358-01 Calibration Adapter ....................................................................................................... 30
Figure 22: Calibration Sub-Menu ............................................................................................................................. 32
Figure 23: Module Components .............................................................................................................................. 34
Figure 24: Remote Sensor Drawing ........................................................................................................................ 43
Figure 25: Wiring Connections Table ...................................................................................................................... 45
Figure 26: 4-20 mA Circuits Types 5100-XX-IT – Connections - HART ................................................................. 46
Figure 27: Electrolyte Replacement ........................................................................................................................ 52
Figure 28: Sensor Assembly – Exploded View ....................................................................................................... 54
Figure 29: Model 5100-25-IT-A1/A2 and Model 5100-26-IT-A1/A2 ........................................................................ 55
Figure 30: Model 5100-25-IT-S1 and Model 5100-26-IT-S1 ................................................................................... 55
Figure 31: Model 5100-88-IT-A1 Dimensions ......................................................................................................... 57
Page 7
Model 5100-XX-IT Toxic Gas Sensor Module
Page 7 of 62
PRODUCT DESCRIPTION
1.1 General
The Model 5100-XX-IT Toxic Gas Sensor Module is a smart transmitter and member of the IT Series family and it offers a broad array of features including:
Integral alphanumeric LED display
Up to 180 day calibration frequency
FM Approval for performance and safety
SIL-2 Certified
Optional integral alarm relays (3)
4-20 mA output
Modbus® RTU interface 1
SMC Sentry interface
Optional HART interface
316 Stainless steel enclosure option
Remote sensor option
Low maintenance and operation costs
ATEX Approval (5100-05-IT)
The 5100-XX-IT is designed and approved for installation and operation in hazardous locations. Members of the 5100-XX-IT Series Toxic Gas family include:
5100-03-IT Oxygen Gas Sensor Module
5100-04-IT Carbon Monoxide Gas Sensor Module
5100-05-IT Hydrogen Sulfide Gas Sensor Module
5100-06-IT Chlorine Gas Sensor Module
5100-08-IT Chlorine Dioxide Gas Sensor Module
5100-10-IT Sulfur Dioxide Gas Sensor Module
5100-12-IT Nitrogen Dioxide Gas Sensor Module
5100-21-IT Hydrogen Chloride Gas Sensor Module
5100-25-IT Ammonia Gas Sensor Module
5100-26-IT Hydrogen Fluoride Gas Sensor Module
5100-88-IT Carbon Dioxide Gas Sensor Module
1 ®
Registered trademark of Schneider Electric
Page 8
Model 5100-XX-IT Toxic Gas Sensor Module
Page 8 of 62
1.2 Product Configuration
Various module mounting configurations can be implemented without special fixtures. Where applicable, these options are factory configured prior to shipment. Mounting configuration can be selected by the installer or field technician and are fully described in this manual. Sensor must always be oriented downward.
1.3 Theory of Operation
Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode, and electrolyte. The components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical reaction and generate a current. The cells are diffusion limited so the rate the gas enters the cell is solely dependent on the gas concentration. The current generated is proportional to the rate of consumption of the subject gas in the cell.
Sierra Monitor electrochemical sensors provide improved reliability by allowing the gas to diffuse into the sensor through a capillary port, rather than diffusing through membranes. The result is an extremely stable sensor with very low temperature and pressure coefficients and the capability to monitor gas as ppm.
1.4 Modes of Operation
1.4.1 Sentry Interface IT gas sensor modules can be installed on Sierra Monitor Sentry Model 5000 controllers Ver. 6.XX MFD after
9/1/95. When it is installed in a Sentry system the IT module must have a unique address which can be established by setting an address between 1 and 8 on the Module Address Switch accessible from the cover plate as illustrated in Figure 8. Figure 13 in this manual provides the wiring terminations for connections to the Sentry controller.
NOTE: The Model # 5100-88-IT CO2 gas sensor module range is 0-5000 ppm. Therefore, the Sentry
Controller must be configured for a 0-100 % full scale input. The Sentry cannot display engineering units up to 5000. The alarm set points need to be set as a percentage of full scale. For example, a 2500 ppm alarm set point is equal to 50% scale.
When the module is operated in conjunction with a Sentry controller, the alarm relay set-up (Section 5.3) should
be set to “Sentry”, allowing the Sentry controller to manage alarm relay action rather than the 5100-XX-IT Gas Sensor Module.
1.4.2 Modbus Operation An RS-485 Modbus RTU serial interface allows direct connection to standard PLCs and DCSs. The Module
Address Switch (Section 4.5) allows the user to select up to 15 different Modbus addresses. Also, an additional 254 Modbus addresses are available via menu selection. Figure 13 in this manual provides the wiring terminations for Modbus connections. The 5100-XX-IT provides the wiring terminations for Modbus connections.
1.4.3 Analog Operation The analog 4-20 mA interface allows direct connection to standard controller and distributed system. The module
is an active current source. The standard configuration is set up for a 3-wire non-isolated connection. An optional 4-wire isolated connection
is also available and can be enabled by changing JP1 and JP2. (Figure 12)
1.4.4 Remote Sensor The Remote Sensor option can be used to remotely mount some sensors up to 15’ from the transmitter. Refer to
Appendix E for allowable distances.
Page 9
Model 5100-XX-IT Toxic Gas Sensor Module
Page 9 of 62
1.4.5 HART Connection A HART interface option is available. Refer to Appendix G for information.
1.4.6 Remote Alarm Reset An input is available for connection of remote alarm reset. Figure 14 provides the wiring termination for
connecting the remote alarm reset. This only resets local alarms, not Sentry alarms.
1.4.7 Optional Integral Relays The optional relays are integral to the gas sensor module and are rated as SPDT, 250V AC, 5 Amps* for the High
Alarm and Low Alarm relays and SPDT, 250V AC, 0.25 Amp for the Trouble relay. * HART option alarm relays are all SPDT, 250V AC, 2 Amp If the gas sensor module is provided with the optional relays, it will include Terminal P4 on the interface board
(Figure 9 and Figure 10). Relay output connections are on P4.
1.5 Mechanical
The sensor module is comprised of the following three primary components:
1.5.1 Enclosure Standard on the 5100-XX-IT is an explosion-proof, rain-tight cast aluminum electrical housing (Figure 3) with
three ¾” FNPT conduit hubs. The 5100-XX-IT-S has a 316 Stainless Steel enclosure (Figure 2). Both enclosure covers have a viewing window. The design of the enclosure allows 3-way mounting choices as shown in
Figure 1.
NOTE: Module installation must be such that it is mounted vertically with sensor pointing downward.
Figure 1: Model 5100-XX-IT Toxic Gas Sensor Module – Mounting Options
Page 10
Model 5100-XX-IT Toxic Gas Sensor Module
Page 10 of 62
NOTE: Figure 2 and Figure 3 are valid for all except 5100-25/26-IT. See Appendix H.
Figure 2: Model 5100-05-IT-S1/S2 Toxic Gas Sensor – Stainless Steel Enclosure, Dimensions
Figure 3: Model 5100-05-IT-A1/A2 Toxic Gas Sensor – Cast Aluminum Enclosure, Dimensions
Page 11
Model 5100-XX-IT Toxic Gas Sensor Module
Page 11 of 62
1.5.2 Transmitter Electronics Electronic Assembly consisting of one printed circuit board assembly mounted under a cover plate, plugged into
one field termination board. Connectors for wiring for power, signal interface and alarm relays are located on the bottom of the termination board.
1.5.3 Sensor Assembly The sensor assembly includes an explosion proof housing containing the gas sensor and a wiring harness for
connection to the transmitter. The sensor assembly threads into one hub of the enclosure. The exposed end of the sensor assembly is threaded to allow connection of a rain-shield or calibration gas.
1.6 Interconnect Wiring
Not supplied with the sensor module, but necessary to the installation and operation is the shield cable which connects the module to its power source and controller. Before this wiring is installed it is important to read and understand the control system installation instructions to determine wiring requirements and alternatives.
1.7 Power Requirements
The modules operate on 24V DC power between 10V DC and 30V DC. Filtered, regulated DC power must be supplied from a separate source or from an approved Sentry or IT controller.
Page 12
Model 5100-XX-IT Toxic Gas Sensor Module
Page 12 of 62
CAUTIONS & WARNINGS
2.1 Introduction
Although the IT Transmitter Modules are designed and constructed for installation and operation in industrial
applications including “hostile” environments, caution should be taken to insure that the installation is made in
compliance with this instruction manual and that certain procedures and conditions are avoided. This chapter discusses the necessary cautions. Read the entire chapter prior to installation of the equipment.
2.2 Gas Sensor Modules – General
Avoid installing sensor modules where they will be unnecessarily exposed to wind, dust, water (i.e. direct hose down), shock, or vibration. Observe temperature range limitations.
Sensors may be adversely affected by prolonged exposure to certain materials. Loss of sensitivity, or corrosion, may be gradual if such materials are present in low concentrations. These materials include: Halides (compounds containing chlorine, fluorine, bromine, iodine), acid vapors, caustic liquids or mists.
Care has been taken by the manufacturer to ship modules in protective packaging to avoid contamination prior to installation. It is recommended that the modules remain protected during installation and that the covering be removed immediately prior to system start-up.
During normal use the O2, CO, H2S, and SO2 sensors are protected from dirt and oil contamination by a sintered metal cover. If this cover becomes clogged, the response of the sensor will be reduced. Protect the sensor from contamination by careful placement, or by use of rain shields and dust shields.
Sensor modules must not be painted. Paint may contain compounds which will contaminate the sensor. Paint will cause clogging of the sintered metal cover and will cause difficulties during attachment of the calibration head or other maintenance activity. It is recommended that the module be tagged "DO NOT PAINT".
2.3 Wiring Warnings
The manufacturer recommends that extra caution be taken where the installation is near any sources of electromagnetic or radio frequency interference. Precautions include:
Avoid running sensor module cable close to high power cables, radio transmission lines, or cables subject to pulses of high current.
Avoid running cables near large electric motors or generators.
When the sensor module is to be operated in analog (4-20mA output) mode shielded cable is required.
When shielding is used, it is recommended that shields be grounded at the controller and nowhere else.
All splices must be via either a termination hardware system or soldered. Improperly spliced cable can
result in corrosion, resistance changes and system errors. The use of wire nuts and crimp-on connectors is unacceptable.
NOTE: Installation and wiring must be in accordance with the National Electrical Code. Voltage AC
conductors are not to be run in the same conduit as voltage DC conductors.
2.4 Calibration Frequency
The H2S calibration frequency is six months or 180 days. However, prudent gas detector maintenance practices normally suggest a simple recalibration following incidences of exposure to high levels of gas that would lead to alarm activation within any system utilizing the output of the gas sensor module.
Page 13
Model 5100-XX-IT Toxic Gas Sensor Module
Page 13 of 62
QUICK START
3.1 Overview
The gas sensor module has been supplied factory calibrated and ready for immediate installation and operation. An installer familiar with installation and operation of gas detection products can use this section to begin immediate use of the module.
3.2 Wiring
See Section 4.2 to determine if 3-wire or 4-wire operation is necessary. Provide two conductor twisted shielded wiring from the power supply/control device to the sensor module location.
Use wire that is 18 AWG or larger.
3.3 Module Installation
The sensor can be mounted in a variety of configurations supported by the conduit. See Figure 3 to determine which configuration is best for your specific application. The default configuration enables the modules to be put in line with other modules with the sensor element below the transmitter. To change the configuration simply remove the transmitter and rotate to the appropriate configuration and remount the standoffs and transmitter.
The module is designed to be installed on a ¾” NPT or M20 conduit. During installation heed the following:
The installation must meet any hazardous environment codes for electrical equipment.
The sensor module enclosure mounting must be spaced far enough from any vertical surface to allow
removal and replacement of the sensor assembly which is threaded into one ¾” conduit entry.
Sensor housing must be oriented vertically pointing downward.
If module is installed outdoors it is recommended that it be sheltered from direct sunlight.
3.4 Transmitter Installation
To install the transmitter printed circuit assembly into the housing, carefully turn the faceplate so that the printing is in the correct horizontal position for the mounting configuration and slide the assembly over the two stand-offs in the enclosure.
Hand tighten the two captive panel thumb screws into the stand-offs. Replace the enclosure cover prior to providing power to the transmitter.
NOTE: If the transmitter is installed in a classified hazardous area, replace the threaded cover prior to
providing power.
Page 14
Model 5100-XX-IT Toxic Gas Sensor Module
Page 14 of 62
3.5 Startup and Operation
To begin operation of the sensor module, activate the instrument loop with 10-30V DC. Each time the sensor module is powered up it will perform a warm-up for approximately 1.0 minutes (60 seconds). During this time the
display will read “Starting”. The loop output will be held at 4 mA.
NOTE: If the sensor is uncalibrated, the startup display will state “UNCALIB” instead of “STARTING”. NOTE: The 5100-06-IT, 5100-08-IT, 5100-25-IT, 5100-26-IT and 5100-88-IT require 300 seconds warm-up
period. After the warm-up period has expired, the display will indicate the gas concentration. Also, the instrument loop will be released to output current in the range of 4 to 20 mA. The actual current is linear with the gas concentration and depends on the selected range. For instance, when the range of 0 to 50 PPM has been selected, a current of 4 mA corresponds to 0 PPM and a current of 20 mA corresponds to 50 PPM.
3.6 Zero Stabilization
All electrochemical sensors require at least 30 minutes on power prior to calibration. This allows the electrode potentials to equilibrate, resulting in a stable zero signal level.
Page 15
Model 5100-XX-IT Toxic Gas Sensor Module
Page 15 of 62
INSTALLATION
4.1 Sensor Module Locations
NOTE: All IT modules are factory pre-configured and calibrated. All modules are tagged to indicate the
configuration including the sensor module number Identify all components during unpacking and install using the factory configuration.
The gas sensor module utilizes a diffusion type sensor which should be located close to either the expected source or destination of the gas hazard. If the gas is heavier than air, the sensor module should be installed within 24 inches of the ground or floor. If it is lighter than air, move it above 6’.
After optimum locations are determined based on the above recommendations, consideration should be given to placing the sensor modules in locations which are accessible for calibration service. Slight adjustments to the location of the sensor module may have little impact on effectiveness but major effect on accessibility.
Modules should be placed close to the potential source of gas.
Modules should be placed in areas accessible for calibration.
Sensors should be pointed down and the conduit should include an
inverse trap to reduce moisture (condensation) from accumulating in the electronics enclosure.
Remote calibration fitting (5360-00) should be used to facilitate calibration gas delivery. Run polyurethane tubing (1/4” O.D. x 1/8” I.D.) from fitting to an accessible location.
Model
Gas
Gas Density
N/A
Air
1.00
5100-04-IT
CO
0.97
5100-05-IT
H2S
1.19
5100-06-IT
Cl2
2.49
5100-08-IT
ClO2
3.09
5100-10-IT
SO2
2.26
5100-12-IT
NO2
2.12
5100-21-IT
HCl
1.27
5100-25-IT
NH3
0.60
5100-26-IT
HF
1.86
5100-88-IT
CO2
1.53
Figure 4: Specific Gravities
Figure 5: Typical Mounting Options
Page 16
Model 5100-XX-IT Toxic Gas Sensor Module
Page 16 of 62
4.2 Wiring
4.2.1 Analog 4-20 mA Operation
For a 3-wire non-isolated connection, set jumpers, located on the bottom of the transmitter board, to the lower position as illustrated in Figure 12. Verify that both jumpers are in the position marked by 3-wire. When using a 3-wire connection, a minimum of an 18 AWG, 3-conductor shielded cable must be used. A cable shield must never be used as a conductor. Larger gauge wire is recommended with distances over 1000’. Connect wires as shown in Figure 10.
For a 4-wire isolated connection, set jumpers, located on the bottom of the transmitter board, to the upper position as illustrated in Figure 10. Verify that both jumpers are in the position marked by 4-wire. When using a 4-wire connection, a minimum of 2 each of an 18 AWG, 2 conductor twisted/shielded pair cable must be used. A cable shield must never be used as a conductor. Larger gauge wire is recommended with distances over 1000’. Connect wires as shown in Figure 10.
4.2.2 Modbus Operation Using RS-485 Connection
Use a minimum of 18 AWG, 2-conductor for DC power connection. No shield required. In addition, use a minimum of 24 AWG, low capacitance, shielded data cable for RS485 half-duplex communication. The installation may be planned in a manner which provides up to 32 sensor modules on a single home run.
Standard default RS485 Settings are: 38,400 baud, 8 bits, 1 stop bit, no parity
4.2.2.1 Termination Resistor Jumpers
Termination resistors are used in RS-485 wire runs to provide impedance matching. The IT series modules use a 120 Ohm resistor for this function. The cable being used for this RS-485 connection must have a minimum of 100 Ohm impedance with a maximum of 120 Ohms.
Installations where the cable length is under 100’, termination resistors may not be required. In installations where
the cable length is greater than 100’, it is recommended to place the termination jumpers on the first device and
last device on the RS-485 wire run. Termination jumpers must be removed from all other modules connected between the first and last device. The first device in the RS-485 multiplexed bus is usually a gas controller or PLC. Factory term resistor setting is “not enabled.”
4.2.2.2 Bias Jumpers (BIAS A, BIAS B)
Bias resistors are used to force RS-485 receiver outputs to a known (fail-safe) state, when the bus is idle. Bias jumpers are always installed in pairs as the bias must be placed on both the TX A and TX B lines. Sierra Monitor’s IT series of toxic gas sensors automatically apply the bias jumpers, and are factory installed so that the bias is always enabled.
4.2.3 Sentry Operation Using Sentry Connection
See Figure 13. Use a minimum of 18 AWG, 3-conductor cable up to 2000’. The cable may or may not be shielded. We recommend shielded cable in circumstances that there could be RF or EM interference present. Shield to be terminated and grounded only at the Sentry controller. Shield must be cut and dressed at the module end so that no part of it comes in contact with the conduit or ground.
NOTE: Be sure to follow all local electric code and safety requirements when installing the 5100-XX-IT
Gas Sensor Module.
Page 17
Model 5100-XX-IT Toxic Gas Sensor Module
Page 17 of 62
4.2.4 General
Install conduit as required by local code or construction specifications. Provide for splice boxes where multiple modules will be wired to a single run. Pull conductors of the correct gauge wire from the controller to each splice box and from the respective splice box to each planned module location. See for proper wire termination in the splice box. Twisted wire secured with wire nuts is not an acceptable splice.
NOTE: The drain wire of shielded cable must NOT be used as one of the conductors. Installation and
wiring must be in accordance with the National Electrical Code. Temperature rating of cable wire must be at least 75°C. If cable runs through higher temperature environments, it must be specified for that environment.
4.3 Enclosure Installation
To protect the transmitter and sensor assembly they should be removed from the enclosure and preserved until final installation and wiring termination.
Number
of
Modules
Maximum Length of Wire Run (feet)
500
1,000
2,000
3,000
5,000
1
18
18
16
16
14 2 18
18
14
12
xx 3 18
16
12
xx
xx
4
16
14
12
xx
xx
Figure 6: Minimum Wire Gauges
Prior to installation and wiring complete the following steps.
1. Remove the transmitter from the module housing by:
a) Unscrew the two captive panel screws on the faceplate. b) Lift the transmitter out of the enclosure. c) Unplug the sensor cable from transmitter connector P2. d) Remove the sensor assembly from the enclosure hub.
2. Install the module enclosure onto the end of the supply conduit and/or bolt into position as required.
NOTE: When enclosure earth grounding is required for the installation a grounding lug is located in the
base of the enclosure. Install the earth ground under the green ground screw.
Page 18
Model 5100-XX-IT Toxic Gas Sensor Module
Page 18 of 62
4.4 Transmitter and Sensor Installation
When all pre-wire is complete:
1. Install sensor assembly in the open hub on the module enclosure. The sensor assembly thread must be fully seated into the hub and tightened to maintain explosion proof assembly.
2. Connect the wires which return to the controller to interface board connectors P1, P2, P3 and P4 according to
Figure 9 and as detailed in the connector tables below.
3. Connect the sensor assembly cable to top transmitter board connector P2.
4. Align the headers between the top transmitter board and the lower interface board then push together.
5. Turn the rotary switch to correct sensor address if required.
6. Carefully return the transmitter to the enclosure, installing it over the two stand-off’s. Tighten the retaining screws into the stand-offs.
7. Cycle power to accept module address change.
8. Establish the module address according to Section 4.5.
NOTE: The starting delay period normally takes approximately 3 minutes but under some circumstances
can take longer. For optimum performance, it is recommended that a calibration be performed after 24 hours of operation.
P1
PCB Label
Function
1
Switch
IN +
Digital Input SW +
2
IN -
Digital Input SW -
3
4-20
IN +
4-20 mA Input +
4
IN -
4-20 mA Input -
5
GND
Ground
6
4-20
OUT +
4-20 mA Output +
7
OUT -
4-20 mA Output -
P2
1
RS 485
+
RS 485 (+) (A)
2 - RS 485 (-) (B)
3 S RS 485 shield (Isolated GND)
4
+
RS 485 (+) (A)
5 - RS 485 (-) (B)
6 S RS 485 shield (Isolated GND)
P3B
1 P V DC Power
2
S
Sentry Signal or Communication
3 G V DC Ground
P3A
4 P V DC Power
5
S
Sentry Signal or Communication
6 G V DC Ground
P4*
PCB Label
Function
1
WARNING
N/C
Low Alarm Relay NC
2
COM
Low Alarm Relay COM
3
N/O
Low Alarm Relay NO
4
ALARM
N/C
High Alarm Relay NC
5
COM
High Alarm Relay COM
6
N/O
High Alarm Relay NO
7
TRBL**
N/C
Trouble Alarm Relay NC
8
COM
Trouble Alarm Relay COM
9
N/O
Trouble Alarm Relay NO
* P4 Connections are installed only when the optional
Relays are included
** Trouble relay is a fail-safe so it is energized for normal operation, functions are labeled for normal operation.
Page 19
Model 5100-XX-IT Toxic Gas Sensor Module
Page 19 of 62
4.5 Module Address Switch
For digital interface applications the module address switch (or Modbus node) Figure 8 must be set per the table below.
Position
Address
Position
Address
1
Sensor 1
9
Sensor 09
2
Sensor 2
A
Sensor 10
3
Sensor 3
B
Sensor 11
4
Sensor 4
C
Sensor 12
5
Sensor 5
D
Sensor 13
6
Sensor 6
E
Sensor 14
7
Sensor 7
F
Sensor 15
8
Sensor 8
0
Software Menu
Figure 7: Sensor Module Address Switch Positions
NOTE: For Sentry (PSG) applications only sensor addresses 1-8 are allowed. If using Modbus output
sensor addresses 1-15 are available. Position 0 allows the Modbus Address to be set by software menu, in the range 16-254.
Figure 8: Transmitter Face Plate
Page 20
Model 5100-XX-IT Toxic Gas Sensor Module
Page 20 of 62
Figure 9: Interface Board Connectors
Figure 10: RS-485 – Termination, BIAS Jumper
Page 21
Model 5100-XX-IT Toxic Gas Sensor Module
Page 21 of 62
Figure 11: mA Circuit Types
Page 22
Model 5100-XX-IT Toxic Gas Sensor Module
Page 22 of 62
Figure 12: 4-20 mA Circuit Type Connections for 5100-XX-IT
Page 23
Model 5100-XX-IT Toxic Gas Sensor Module
Page 23 of 62
5100-XX-IT TO MODBUS DEVICE
5100-XX-IT TO SENTRY
Figure 13: Wiring Connections for Modbus and Sentry Interface
(dry contact only unpowered)
Page 24
Model 5100-XX-IT Toxic Gas Sensor Module
Page 24 of 62
Figure 14: Wiring Connections for Remote Alarm Reset
5100-XX-IT Remote Alarm Reset (Unsupervised)
(dry contact only unpowered)
SWITCH IN
SWITCH IN
+
-
+
-
+
GND
+
- -
IN IN
OUT
SWITCH
4-20 4-20
P1
4.7K OHM (CUSTOMER SUPPLIED)
5100-XX-IT Remote Alarm Reset (Supervised)
Page 25
Model 5100-XX-IT Toxic Gas Sensor Module
Page 25 of 62
OPERATION
5.1 Introduction
The Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A magnetic tool (5358-
50) is supplied for this purpose. The menu system is used to configure alarm set-points, calibrate the sensor
module, and for maintenance procedures and alarms acknowledge.
5.2 Data Entry Keypad
The module menu system is operated by means of directing the magnet stick toward each of four independent hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are located under the faceplate above and below the LED display and are labeled M , E , and as shown in
Figure 15.
Figure 15: Face Plate with Operator Keypad
Page 26
Model 5100-XX-IT Toxic Gas Sensor Module
Page 26 of 62
5.3 Main Menu
Key
Function
Display
Description
Reference
M
E
Mode
Switch [M]
M
E
Enter
Switch [E]
M
E
Up
Switch [▲] Previous Menu
M
E
Down
Switch [▼] Next Menu
5100-XX
First screen at power up: Model No.
VXX-XX--
Second screen at power up: Version No.
STARTING
Third screen at start up: Starting Delay
XXX PPM
Normal condition - default display
M
E
Mode
ALMRSET:
Mode Function - Alarm Reset
Banner: "Press [E] to reset alarm"
M
E
RESET
Alarm Reset
XXX PPM
Default Display
M
E
Mode
ALMRSET:
Mode Function - Alarm Reset
M
E
Mode
CALIB:--
Mode Function - Calibrate
Figure 21
M
E
Mode
SETUP:--
Mode Function - Set up alarms, relays, etc
Figure 18 & Figure 19
M
E
Mode
MAINT:--
Mode Function - Maintenance
Figure 20
M
E
Mode
EXIT-?--
Exit Menu
M
E
Enter
XXX PPM
Apply Selected Mode (Exit)
XXX PPM
Default Display (Once a minute the sensor displays module address)
Figure 16: Master Menu
Figure 17 defines the key operational displays on the operator interface.
Display
Description
STARTING
Delay from loss of power at start-up
XXX PPM
Concentration
LXXX PPM
Low Alarm (Warning)
HXXX PPM
High Alarm (Alarm)
HIGH
Measures gas, concentration exceeds 100% of Full Scale
CXXX PPM
Calibration Mode
ACK
Acknowledged Function
Figure 17: Operation Display Values
NOTE: If display shows “Start” instead of “Starting”, the module must be calibrated before using it.
Page 27
Model 5100-XX-IT Toxic Gas Sensor Module
Page 27 of 62
5.4 Set-Up
The sensor module set-points menu is used to initially set-up the alarm set points, relay actions, gas type and range, 4-20 mA action or RS-485/Sentry address and baud rates, Digital Input and Warm-up time. Use the [▲] or [▼] keys to select following menus and press [E] to select.
Alarms – U se the [▲] or [▼] keys to select Low Alarm (Warning) or High Alarm (Alarm) menu. Key [▲]
will adjust the setpoint upwards and Key [▼] will adjust the value downwards. Once it reaches the desired
setpoint, Key [E] will accept it and ACK will appear.
Set-points can be configured using this menu up to the following values:
Factory Alarm Set Points
Model
Warning
Alarm
5100-03-IT
19.5%
16.5%
5100-04-IT
50 PPM
100 PPM
5100-05-IT
10 PPM
20 PPM
5100-06-IT
2.0 PPM
5.0 PPM
5100-08-IT
0.3 PPM
1.0 PPM
5100-10-IT
10 PPM
20 PPM
5100-12-IT
2 PPM
4 PPM
5100-21-IT
5 PPM
10 PPM
5100-25-IT
25 PPM
50 PPM
5100-26-IT
2.0 PPM
5.0 PPM
5100-88-IT
1000 PPM
2500 PPM
NOTE: To enable rising O2 Alarm set warning setpoint higher than 20.9%
Relays Use the [▲] or [▼] keys to select High Alarm (Alarm) or Low Alarm (Warning) relay menu and
press [E]. Use the [▲] or [▼] keys to select the correct alarm relay action for the application, Latch, Sentry or Non-Latch. Selecting “Sentry” enables the Sentry controller to make all alarm action decisions. * indicates the current selection.
Range – Use the [▲] or [▼] keys to select Range menu and press [E]. When “Range” is selected menu
provides any choices of ranges available for the gas type selected. Use the [▲] or [▼] keys to select the
desired range. If the “User” range is selected, use the [▲] or [▼] keys to adjust the high end of the range
desired.
Factory Range Set-up
Model
Range
Other Available
5100-03-IT
5-25%
0-25%
5100-04-IT
0-500 PPM
Variable
5100-05-IT
0-100 PPM
Variable
5100-06-IT
0-10 PPM
None
5100-08-IT
0-3 PPM
None
5100-10-IT
0-100 PPM
Variable
5100-12-IT
0-20 PPM
None
5100-21-IT
0-20 PPM
None
5100-25-IT
0-100 PPM
Variable
5100-26-IT
0-10 PPM
None
5100-88-IT
0-5000 PPM
Variable
Page 28
Model 5100-XX-IT Toxic Gas Sensor Module
Page 28 of 62
4-20mA – Use the [▲] or [▼] keys to select Calib or CalibOut menu and press [E]. The “Calib” section of
the menu allows the user to calibrate the 4 mA and 20 mA outputs. To calibrate the 4 mA and 20 mA outputs it is necessary to have an amp meter connected to the 5100-XX-IT and upon selecting the 4 mA
output calibration then the [▲] or [▼] keys can be used to adjust the 4 mA reading on the amp meter until
it reads 4 mA. Similar steps can then be performed for the 20 mA output. The CalibOut section allows the user to select the 4-20 mA output action desired during calibration. * indicates the current selected value. Available selections include:
o Track – the 4-20mA value tracks the calibration gas exposed to the gas sensor module o C2.50mA – the 4-20mA value is held at 2.50mA during calibration o C4.00mA – the 4-20mA value is held at 4.0mA during calibration
RS-485Use the [▲] or [▼] keys to select Address or Baud rate menu and press [E]. Note that the
5100-XX-IT has a rotary switch on the faceplate and it is used to select addresses 1-15. When connected to Sentry the user can select 1-8 and when using Modbus RS-485 the user can select addresses 1-15.
For Modbus addresses above 15, set the rotary switch to 0 and then use the “Address” menu to select
any address between 16 and 254. The Baud rate menu allows the user to select a baud rate of 38400, 19200, 9600, 4800 or 2400. * indicates current selection. The Sentry default is 9600.
DigInput – Use the [▲] or [▼] keys to select supervised or Non-supervised digital input and then press
[E] to change.
Key
Function
Display
Description
Reference
--0%LEL-
Default Display
M
E
Mode
ALMRSET:
Mode Function - Alarm Reset
M
E
Mode
CALIB:--
Mode Function - Calibrate
M
E
Mode
SETUP:---
Mode Function - Set Point Adjustments
M
E
Enter
Alarms
S.P. Function - Alarm Adjust
* A Below
M
E
Down
Relays
S.P. Function - Relays Adjust
* B Below
M
E
Down
Relays
S.P. Function - Range Adjust
* C Below
M
E
Down
4-20mA
S.P. Function - 4-20 mA Adjust
* D Below
M
E
Down
Dig Input
S.P. Function - RS-485/Sentry Output Adjust
* E Below
High Alarm Set Point Adjustment Example
M
E
Enter
H.Alarm
S.P. Function - High Alarm Adjust
*A
M
E
Enter
HASP:60-
High Alarm Set Point: current = 60
Use or keys to adjust to new set point
M
E
Down (x5)
HASP:55-
High Alarm Set Point: new = 55
M
E
Enter
ACK
Momentary Acknowledge of new Set Point
H.Alarm
S.P. Function - High Alarm Adjust
Relays Set Point Adjustment Example
M
E
Enter
H. Relay
S.P. Function - High Alarm Relay Adjust
*B
M
E
Down
L.Relay
S.P. Function - Low Alarm Relay Adjust
M
E
Enter
Latch
Use or keys to adjust to new relay action (Latch, Sentry, NonLatc) * indicates current
M
E
Down
Sentry
NOTE: Sentry indicates that Sentry controls relay action and not the IT Sensor Module
M
E
Down
*Sentry
Low Alarm Relay set to Sentry
Figure 18: Set-Up Configuration Part 1
Page 29
Model 5100-XX-IT Toxic Gas Sensor Module
Page 29 of 62
Gas Range Adjustment Example
M
E
Enter
Range
S.P. Function - Range Adjust
*C
M
E
Enter
*100 PPM
Select [E] to select or or to select another and press [E]
M
E
Down
10 PPM
Press [E] if selecting 0-10 PPM range
M
E
Down
USER
Press [E] if selecting User adjustable range
M
E
Enter
100 PPM
Select [E] to select or or to select another and press [E]
4-20 mA Adjustment Example
M
E
Enter
Calib
S.P. Function - Calib Adjust
*D
M
E
Enter
Out: 4mA
Use or keys to select 4 mA or 20mA
M
E
Enter
4mA
Selects 4 mA
M
E
Enter
ACK
Momentary Acknowledge of new Set Point
M
E
Mode
Calib
S.P. Function - Calib Adjust
M
E
Down
CalibOut
S.P. Function - Output during Calibration Adjust
M
E
Enter
Track
Use or keys to select Track, C2.50mA, C4.00mA
Track = Output during calibration tracks the calibration gas, C2.50mA = Output during calibration is 2.50 mA, C4.00mA = Output during calibration is 4.0 mA
M
E
Enter
*Track
* = Current selection
RS-485 Adjustment Example
M
E
Enter
Address
S.P. Function - RS-485 Address Adjust
*E
M
E
Enter
Addr:016
Use or keys to enter new address
M
E
Enter
ACK
New address selected
M
E
Enter
Address
S.P. Function - RS-485 Address Adjust
M
E
Down
Baud
S.P. Function - RS-485 Baud Rate Adjust
M
E
Enter
*38400
Press [E] to select or or to select another
Figure 19: Set-Up Configuration Part 2
5.5 Maintenance Sub-Menu
The maintenance menu enables the operator to view sensor and software versions. Sierra Monitor technical support has access to other values as needed. The maintenance menu operation is described in Figure 20.
Key
Function
Display
Description
Reference
XXXPPM
Default Display
M
E
Mode
ALMRSET:
Mode Function - Alarm Reset
M
E
Mode
CALIB:--
Mode Function - Calibrate
M
E
Mode
SETUP:-
Mode Function - Set Point Adjustments
M
E
Mode
MAINT:--
Mode Function - Maintenance
M
E
Enter
3.XX
Module Software Version Number
M
E
Enter
CCC001
Module Custom Configuration Control number
Figure 20: Maintenance Menu
Page 30
Model 5100-XX-IT Toxic Gas Sensor Module
Page 30 of 62
CALIBRATION
6.1 Calibration Frequency
The Gas Sensor Module has been calibrated in the factory prior to shipment. It is recommended that the user calibrate before placing into service. The sensor module must be calibrated every 180 days at a minimum. Periodic functional tests are advisable for critical applications and hostile environments. Oxygen sensors will require more frequent adjustment.
The sensor module microprocessor software includes high-level self-checking algorithms which provide continuous sensor diagnostic and self-adjustment. Users may select to increase calibration frequencies based on low-drift experience during first two calibration periods.
6.2 Calibration Preparation
Calibration of the sensor is accomplished by simple menu based steps and application of span gas.
NOTE: If an error is made during any stage of the calibration process, hold the magnet stick at the [M] for
10 seconds. A scrolling display will indicate “calibration aborted” and the sensor module will exit
the calibration activity and return to normal operating mode. The calibration procedure can then be restarted.
Calibration must be performed only when the area is known to be clear of the gas of interest. When in doubt, use a portable instrument to confirm that there is no background.
For compliance with Factory Mutual (FM) Approvals, the Sierra Monitor Model 1250-01, 1260 -05 and 1260-45 are the FM Approved calibration gas delivery devices. Use the Model 5358-01 Calibration Adapter delivery fitting.
NOTE: Refer to Appendix H.1 for information on use of permeation tube for 5100-25-IT and Appendix H.2
for 5100-26-IT calibration.
6.3 Calibration Gas Delivery Methods
Calibration gas can be delivered to the sensors via the Model 5358-01: Calibration Adapter (Figure 21) - used with portable calibrators.
Calibration Procedure
The Calibration Menu is described on Figure 22. The procedure requires that the menu “keys” be activated using the magnet stick.
Each key press steps through the process of setting the zero value for clean air and then setting the span value.
At each of these steps, apply calibration gas of the value corresponding to the setting accepted on the sensor module display.
Figure 21: Model 5358-01
Calibration Adapter
Page 31
Model 5100-XX-IT Toxic Gas Sensor Module
Page 31 of 62
6.4 Sensor Exposure to Gas
Calibration gas must be delivered to the sensor using the flow rate and duration listed below:
Model
Calibration Gas
Flow
Period
5100-03-IT
Air
300 cc of Zero Air or exposure to Ambient Air (3 minutes)
5100-04-IT
Carbon Monoxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-05-IT
Hydrogen Sulfide
300 cc/min
Until Stable (minimum 3 minutes)
5100-06-IT
Chlorine
300 cc/min
Until Stable (minimum 3 minutes)
5100-08-IT*
Chlorine Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-10-IT
Sulfur Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-12-IT
Nitrogen Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-21-IT
Hydrogen Chloride
300 cc/min
5100-25-IT
Ammonia
See Appendix H.1
5100-26-IT
Hydrogen Fluoride
See Appendix H.2
5100-88-IT
Carbon Dioxide
300 cc/min
Until Stable (minimum 3 minutes)
5100-04-IT
Carbon Monoxide
300 cc/min
Until Stable (minimum 3 minutes)
* Use 5PPM Cl2 to calibrate (set module to 1.5 PPM ClO2)
Page 32
Model 5100-XX-IT Toxic Gas Sensor Module
Page 32 of 62
6.5 Calibration Sub-Menu
Key
Function
Display
Description
Reference
XXXPPM
Default Display
M
E
Mode
ALMRSET:
Mode Function - Alarm Reset


M
E
Mode
CALIB:--
Mode Function - Calibrate


M
E
Enter
CAL-0PPM--
Banner: Apply zero gas, enter <E> when done (not applicable for 5100-03-IT)
*


Operation: Confirm area clear of gas, or apply zero air to sensor
M
E
Enter
ACK
Zero gas setting acknowledged


10PPM-SP
Banner: Select span, enter <E> to calibrate sensor Banner: (5100-03-IT) Specify Cal gas <E> when done
Sub A
M
E
Enter
C 4 PPM
Banner: Apply span gas, then enter <E> to calibrate gas sensor
Sub B


C 4 PPM
Operation: Apply calibration gas
CXXXPPM
Operation: As gas is applied the reading will increase - wait 3 minutes
M
E
Enter
CAL-OK
Calibration Passed - now remove gas


WAIT-300
Operation: Five minute time out before sensor is returned to service.
Sub-Routines
10PPM-SP
Banner: Select span, enter <E> when done
Sub A
M
E
Down
25PPM-SP
Operation: Change Span Gas Value to 25PPM

M
E
Down
Adj-SPAN
Operation: User adjustable value


M
E
Enter
25PPM-SP
Operation: Ready for user adjustment


M
E
Enter
C 25PPM
Banner: Apply 25PPM gas, then enter <E> to calibrate gas sensor
Sub B


C 25PPM
Operation: Apply calibration gas
M
E
Enter
CAL-FAIL
Operation: No calibration gas applied, or sensor did not respond correctly.


WAIT-300
Operation: Five minute time out before sensor is returned to service.
M
E
Mode
(Any)
Operation: Hold magnet over Mode Switch for ten seconds to abort calibration
Sub C


Banner: Calibration Aborted
XXXPPM
Default Display
Figure 22: Calibration Sub-Menu
Page 33
Model 5100-XX-IT Toxic Gas Sensor Module
Page 33 of 62
6.5.1 5100-03-IT Oxygen Deficiency
It is recommended to use fresh Air as the calibration source for the Oxygen detector. If the area is known to be safe with adequate air flow, you may use the surrounding Air to set the instrument span at 20.9 %. If the condition of the environment is unknown, then the customer may use a cylinder of Zero Air as the span gas. Please note that the Oxygen concentration contained in Zero Air cylinders may have anywhere from 20.0 to 21.0 % Oxygen as it is not a precise standard.
To calibrate the Oxygen sensor module, place the module into the calibration mode using the magnetic wand. Press the Mode key until CALIB is displayed. Press Enter to begin. Select Calibration value to be 20.9%. Press Enter to confirm. Apply span gas or use the surrounding Air then press Enter to complete procedure.
6.5.2 5100-88-IT Carbon Dioxide
The CO2 transmitter also has a span only calibration adjustment. There is no zero adjustment or requirement to apply Zero gas. To calibrate the CO2 detector, place the module into the calibration mode using the magnetic wand. Press the Mode key until CALIB is displayed. Press Enter to begin. Select calibration value, typical 2500 ppm, and press Enter to confirm. Apply span gas for at least 3 minutes or until stable. Press Enter to complete procedure.
NOTE: The Analog output values to be transmitted during the calibration procedure are selected in the
maintenance configuration. Normal 20.9% value provides 17.3 mA.
Page 34
Model 5100-XX-IT Toxic Gas Sensor Module
Page 34 of 62
Appendix A. Service
Appendix A.1. Sensor Module Configuration
The gas sensor module is comprised of the following sub-assemblies (Figure 23):
5100-XX-IT Gas Sensor Module (See Appendix C)
SPL21810 Aluminum Enclosure SPL21823 316SS Enclosure XXXXXXX Transmitter Assembly (See Appendix C) XXXXXXX Sensor Assembly (See Appendix C) 5200-XX-IT Sensor (See Appendix C)
Appendix A.2. Enclosure Replacement
The enclosure should be replaced if the cover threads or conduit threads have been damaged, or if the enclosure has been damaged sufficiently that it no longer meets the required NEMA classification.
To replace the enclosure follow the transmitter and sensor assembly removal instructions, remove the damaged enclosure from its conduit, install a new enclosure and continue the transmitter and sensor assembly replacement instructions.
Figure 23: Module Components
Page 35
Model 5100-XX-IT Toxic Gas Sensor Module
Page 35 of 62
Appendix A.3. Transmitter Replacement
The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be adjusted for calibration. This may occur due to age, corrosion or failed components.
To replace the transmitter assembly:
1. Remove the cover of the main enclosure.
2. Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90o to relieve the
wiring service loop.
3. Remove all wires from bond, marking them carefully so they can be correctly replaced.
4. Unplug the sensor connector from the transmitter.
5. Reverse the preceding steps to install the new transmitter.
6. Restore power and allow a minimum of 30 minutes for stabilization before re-calibration.
Appendix A.4. Sensor Replacement
NOTE: Refer to Appendix H for information on sensor electrolyte replacement for 5100-25-IT and
5100-26-IT.
The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. The gas sensor needs replacement when:
The “CAL-FAIL” message appears after calibration.
The sensor output signal is noisy, causing erroneous gas level readings.
The “Sensor Failure” message displays.
To replace the sensor:
1. Confirm that system power has been removed.
2. Remove the gas sensor module enclosure cover.
3. Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90o to
relieve the wiring service loop.
4. Unplug the sensor connector from the transmitter.
5. Unscrew the old sensor assembly from the enclosure conduit hub.Remove the sensor assembly with its
harness.
6. Unscrew sensor housing cover from the sensor.
7. Carefully pull the old sensor straight up from the socket.
8. Press the new sensor into the socket.
9. Reverse the preceding steps to install the sensor assembly.
10. Allow the new sensor to stabilize for a minimum of 60 minutes and then calibrate using the procedure in
Section 6.
Page 36
Model 5100-XX-IT Toxic Gas Sensor Module
Page 36 of 62
Appendix A.5. Installation Inspection
Prior to system start-up or trouble shooting, the entire system should be visually inspected. The following are guidelines for that inspection.
Appendix A.5.1. Controller Installation
Controller installed in conformance to instruction manual recommendations.
AC power is correctly grounded.
Hot AC and relay connections have safety covers installed.
Appendix A.5.2. Cabling Installation
All splices are soldered or via terminal block.
Cabling is away from sources of electrical noise or RFI where possible.
Appendix A.5.3. Sensor Module Installation
Module installation in conformance with this manual.
Modules accessible for calibration.
Wiring terminations clean and correct.
Appendix A.5.4. Moisture Traps and Rainshields
Conduit seals and drains installed to avoid moisture build up in electronics enclosure. Water accumulation
in sensor module enclosures is a major cause of damage and system failures - take precautions to seal electrical conduits and provide moisture traps and drains to avoid water damage.
Rain-shields installed where applicable.
Appendix A.5.5. Standard Voltages
Regulated DC voltage to be applied to the sensor module must be between 10 VDC and 30 VDC.
Appendix A.6. Inspection and Troubleshooting Guide
The inspection and troubleshooting guide can be used to step through the system startup and to determine the appropriate corrective action if a fault occurs.
Appendix A.6.1. Ir Module Does Not Respond to Gas
1. Repeat calibration procedure.
2. Remove the gas and wait for the timer to completely count down.
3. Apply calibration gas and verify that the sensor sees calibration gas after calibration.
4. If the sensor still does not see gas, power cycle the unit and repeat calibration.
Appendix A.6.2. Module Displays “Starting” for more than 1 Hour
1. Make sure the sensor is placed in an ambient room temperature environment.
2. Power cycle the sensor.
3. Ensure that the sensor is not exposed to gas of interest during warm-up.
Page 37
Model 5100-XX-IT Toxic Gas Sensor Module
Page 37 of 62
Appendix A.6.3. Module Does Not Display the Correct %PPM
1. Power cycle the unit.
2. Recalibrate the sensor.
Appendix A.6.4. Display Shows ‘F’ – Sensor Missing or Bad Sensor
1. Power down the unit.
2. Open the enclosure and unplug the sensor assembly from the transmitter board.
3. Plug the sensor back into the transmitter board carefully and ensure a secure fit.
4. Check all other connections.
5. Power up the unit.
Appendix A.6.5. Module Shows “***Calibration Required***”
Calibrate the module.
Appendix A.6.6. Display Shows “H” or L
This means the local high or low alarm is active.
Appendix A.6.7. Display Shows “M” – Not Calibrated
Calibrate the module.
Appendix A.6.8. Display Shows “C” – Calibration Mode
Complete calibration and exit to operating mode.
Appendix A.6.9. Display Shows “S” – Sentry Connection
Check connections with Sentry Controller
Appendix A.6.10. Diagnostic LEDs
LEDs 1 = 1 Hz Heartbeat normal, 2 Hz Panic Error 2 = Computation process 3 = 1 Hz Heartbeat normal 4 = HART or Sentry comms activity
Page 38
Model 5100-XX-IT Toxic Gas Sensor Module
Page 38 of 62
Appendix B. Specifications
Sensor
Type Electrochemical
Standard
Range
(1)
Optional
Max Range
Zero Drift
Repeatability
Linearity
Resolution
Response
Time
(2)
Accuracy
Sensor
Life
(3)
5100-03-IT O2
5-25% Vol
0-25% Vol
+/- 0.1%
+/- 0.2%
+/- 0.2%
+/- 0.1%
<10 sec.
+/- 0.2%
2 years
5100-04-IT CO
0-500 PPM
1200 PPM
+/- 1 PPM
+/- 1 PPM
+/- 1 PPM
0.5 PPM
<35 sec.
+/- 1 PPM
2 years
5100-05-IT H2S
(5)
0-100 PPM
100 PPM
+/- 0.5 PPM
+/- 1 PPM
+/- 1 PPM
0.1 PPM
<44 sec.
+/- 3 PPM
(6)
2 years
5100-06-IT Cl2
0-10.0 PPM
10.0 PPM
+/- 0.5 PPM
+/- 0.5 PPM
+/- 0.5 PPM
0.1 PPM
<60 sec.
+/- 0.5 PPM
2 years
5100-08-IT ClO2
0-3.0 PPM
3.0 PPM
+/- 0.5 PPM
+/- 0.5 PPM
+/- 0.5 PPM
0.1 PPM
<60 sec.
+/- 0.5 PPM
2 years
5100-10-IT SO2
0-100 PPM
100 PPM
+/- 0.5 PPM
+/- 1 PPM
+/- 1 PPM
0.5 PPM
<20 sec.
+/- 0.5 PPM
2 years
5100-12-IT NO2
0-20 PPM
20.0 PPM
+/- 0.5 PPM
+/- 0.5 PPM
+/- 0.5 PPM
0.2 PPM
<35 sec.
+/- 0.5 PPM
2 years
5100-21-IT HCl
0-20 PPM
20.0 PPM
+/- 0.5 PPM
+/- 0.5 PPM
+/- 0.5 PPM
0.2 PPM
<35 sec.
+/- 0.5 PPM
2 years
5100-25-IT NH3
(4)
0-100 PPM
100 PPM
+/- 1.0 PPM
+/- 1.0 PPM
+/- 1.0 PPM
+/- 1 PPM
<30 sec.
+/- 1 PPM
1 year
(7)
5100-26-IT HF
(4)
0-10 PPM
10 PPM
--
--
--
0.5 PPM
<30 sec.
+/- 1 PPM
1 year
(7)
5100-88-IT CO2
(8)
0-5000 PPM
5000 PPM
+/- 1.0 PPM
+/- 1.0 PPM
+/- 1.0 PPM
+/- 1 PPM
<20 sec.
+/- 2%
2 years
(1) Optional ranges available (2) Response time to 90% full signal value for applied concentration (3) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest (4) Diffusion via membrane. Requires electrolyte (5) Max range, zero drift, linearity, resolution, and sensor life, have not been verified by FM Approvals (6) Or +/- 10% of reading, whichever is greater (7) Electrolyte recharge only (8) Infrared
Output
Display Fixed and Scrolling LED Relays Option 5 Amp, High Alarm, Low Alarm, .25 Amp Trouble all SPDT Analog Output Analog 4-20 mA (Trouble 1.5mA, Calibration 2.5mA) Loop Resistance 800 ohm Signal Output Sentry digital bus Analog 4-20 mA 3-wire Non-Isolated 4-wire Isolated Serial RS-485 Modbus RTU HART
Input
Remote Alarm Reset Normally open digital input
Power
Power consumption 2 watts (4 watts for 5100-88-IT) Input voltage 24 VDC nominal: 10-30 VDC
Page 39
Model 5100-XX-IT Toxic Gas Sensor Module
Page 39 of 62
Operating Range
Ambient Temp (oF)
Ambient Temp (oC)
Relative Humidity
5100-03-IT O2
5 to 122 oF
-15 to 50 oC
15 – 90%
5100-04-IT CO
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-05-IT H2S
(1)
- 40 to 122 oF
-40 to 50 oC
15 – 90%
5100-06-IT Cl2
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-08-IT ClO2
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-10-IT SO2
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-12-IT NO2
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-21-IT HCl
- 4 to 122 oF
-20 to 50 oC
15 – 90%
5100-25-IT NH3
+14 to 113 oF
-10 to 45 oC
20 – 95%
5100-26-IT HF
+14 to 113 oF
-10 to 45 oC
20 – 95%
5100-88-IT CO2
- 4 to 140 oF
-20 to 60 oC
15 – 90%
(1) Performance tested by FM Approvals to -40oC/oF
Enclosure
Material - Aluminum Polyester powder-coated, sand-cast, copper-free aluminum Material - Stainless Steel 316 Stainless Steel Dimensions (HxWxD) (differs for 5100-25-IT & 5100-26-IT)
(A1 & A2) 7.9 x 5.7 x 3.9 in. (20.1 x 14.5 x 9.9 cm)
(S1 & S2) 7.6 x 4.4 x 4.9 in. (19.3 x 11.2 x 12.4 cm)
Weight (A1 & A2) 3.4 lb. (1.5 Kg) (S1 & S2) 7.2 lb. (3.3 Kg) Housing NEMA 4X, NEMA 7 Hazardous Area Approval Explosion proof, Class 1, Div. I, Groups B, C, D Group IIB + H2 IP66, IP65, NEMA 4X
Electrical Approvals
Factory Mutual (FM) Class 1, Div. I, Groups B, C, D FM Standards 3600, 3615, 3810 UL UL Standard 508A ATEX (5100-XX-IT) II 2 G Ex d IIB+H2 T6 Gb Ta = -40oC to +50oC
Performance Approvals
FM Performance Approval FM Performance Approval (5100-05-IT) (Appendix J) SIL-2 Certified (5100-04-IT, 5100-05-IT, 5100-06-IT)
CE Mark (5100-05-IT) ABS (5100-05-IT) CQST
Remote Sensor Option (Appendix E)
Distance between sensor and transmitter
5100-03-IT
10 feet
5100-06-IT
10 feet
5100-04-IT
15 feet
5100-08-IT
10 feet
5100-05-IT
15 feet
5100-10-IT
15 feet
Warranty
Limited Warranty 2 years
Specifications subject to change without notice
Page 40
Model 5100-XX-IT Toxic Gas Sensor Module
Page 40 of 62
Cross Sensitivity Data
5100-04-IT Carbon Monoxide
Gas
Conc.
Response
HS
15 PPM
~38 PPM
SO2
5 PPM
~3 PPM
NO
35 PPM
~10 PPM
NO2
5 PPM
~-3 PPM
H2
100 PPM
<60 PPM
C2H4
100 PPM
<100 PPM
HCN
10 PPM
~5 PPM
HCl
5 PPM
0 PPM
5100-05-IT Hydrogen Sulfide
Gas
Conc.
Response
CO
400 PPM
<.1 PPM
SO2
20 PPM
<10 PPM
NO
50 PPM
<3 PPM
NO2
10 PPM
<-30 PPM
Cl2
10 PPM
<-25 PPM
H2
400 PPM
1 PPM
C2H4
400 PPM
<0.1 PPM
NH3
20 PPM
<0.1 PPM
5100-06-IT Chlorine & 5100-08-IT Chlorine
Dioxide
Gas
Conc.
Response
CO
300 PPM
0 PPM
H2S
15 PPM
0 to -3.38 PPM
SO2
5 PPM
-0.05 PPM
NO
35 PPM
0 PPM
NO2
5 PPM
~5 PPM
H2
100 PPM
0 PPM
C2H4
100 PPM
0 PPM
HCN
10 PPM
0 PPM
HCl
5 PPM
0 PPM
ClO2
3 PPM
10 PPM
5100-10-IT Sulfur Dioxide
Gas
Conc.
Response
H2S
15 PPM
0 PPM
NO
35 PPM
-7 to 0 PPM
NO2
5 PPM
-5 PPM
Cl2
5 PPM
-1.5 to 0 PPM
NO2
5 PPM
~5 PPM
H2
100 PPM
0 PPM
HCN
10 PPM
<5 PPM
HCl
5 PPM
0 PPM
C2H4
100 PPM
0 PPM
5100-12-IT Nitrogen Dioxide
Gas
Conc.
Response
CO
300 PPM
0 PPM
H2S
15 PPM
-1.5 to 0 PPM
SO2
5 PPM
-0.05 to 0 PPM
NO
35 PPM
0 PPM
Cl2
1 PPM
~1 PPM
H2
100 PPM
0 PPM
HCN
10 PPM
0 PPM
HCl
5 PPM
0 PPM
C2H4
100 PPM
0 PPM
5100-21-IT Hydrogen Chloride
Gas
Conc.
Response
CO
300 PPM
<3 PPM
H2S
15 PPM
27 to 45 PPM
SO2
5 PPM
1.5 to 3.5 PPM
NO
35 PPM
0 PPM
NO2
5 PPM
0.5 to 1 PPM
Cl2
1 PPM
-0.05 to 0.1 PPM
H2
100 PPM
<0.5 PPM
HCN
10 PPM
<0.3 PPM
5100-25-IT Ammonia
Gas
Conc.
Response
CO
1000 PPM
0 PPM
SO2
15 PPM
>40 PPM
NO
100 PPM
0 PPM
NO2
60 PPM
0 PPM
H2
4%
0 PPM
CH4
4%
0 PPM
HCN
10 PPM
75 PPM
H2S
3 PPM
75 PPM
5100-26-IT Hydrogen Fluoride
Gas
Conc.
Response
Cl2
5 PPM
8 PPM
HCl
5 PPM
9 PPM
NH3
75 PPM
0 PPM
SiH4
15 PPM
0 PPM
CO
1000 PPM
0 PPM
SO2
5 PPM
9 PPM
NO
100 PPM
0 PPM
NO2
15 PPM
2 PPM
H2
4%
0 PPM
H2S
30 PPM
0 PPM
Page 41
Model 5100-XX-IT Toxic Gas Sensor Module
Page 41 of 62
Appendix C. Model Numbers and Parts List
Sensor Module
Ordering Information: The following part number is essential in selection of the correct product. Please build up the desired part by inserting the code for each element
into the appropriate space (ie. 5100-12-IT-S1-02-00-0-C).
Model Gas
Series Enclosure
Relay/Connection
Output
XX Sensor Protection
Remote
Option
ATEX
5100
  
   
- 03 = O2 - 10 = SO2
-IT
- A1 = AL ¾” NPT
- 01 = No Relays (std)
- 0 = Standard
- 0 = None
- C = ATEX/IECEX
- 04 = CO - 12 = NO2
- A2 = AL M20x1.5
- 02 = Relays Option
- 1 = Conformal
- 1 = Remote
(5100-05-IT only)
- 05 = H2S - 21 = HCl
- S1 = SS ¾” NPT
- 05 = HART Option
Sensor
Sensor
- 06 = Cl2 - 25 = NH3
- S2 = SS M20x1.5
- 08 = ClO2 - 26 = HF
- 27 = ETO
Spare Parts
5200-03-IT
Sensor, for 5100-03-IT
5200-04-IT
Sensor, for 5100-04-IT
5200-05-IT
Sensor, for 5100-05-IT
5200-06-IT
Sensor, for 5100-06-IT / 5100-08-IT
5200-10-IT
Sensor, for 5100-10-IT
5200-12-IT
Sensor, for 5100-12-IT
5200-21-IT
Sensor, for 5100-21-IT
SPL21834
Sensor Assembly, Aluminum, for 5100-03-IT ­Oxygen
SPL21832
Sensor Assembly, Aluminum, for 5100-04-IT - CO
SPL21830
Sensor Assembly, Aluminum, for 5100-05-IT ­H2S
SPL21833
Sensor Assembly, Aluminum, for 5100-06-IT / 5100-08-IT - Cl2 & ClO2
SPL21844
Sensor Assembly, Aluminum, for 5100-10-IT ­SO2
SPL21845
Sensor Assembly, Aluminum, for 5100-12-IT – NO2
SPM21878
Sensor Assembly, Plastic, for 5100-26-IT
SPM27082
Sensor Assembly, 316SS, for 5100-03-IT ­Oxygen
SPM27083
Sensor Assembly, 316SS, for 5100-04-IT - CO
SPM27074
Sensor Assembly, 316SS, for 5100-05-IT - H2S
SPM27084
Sensor Assembly, 316SS, for 5100-06-IT / 5100­08-IT - Cl2 & ClO2
SPM27087
Sensor Assembly, 316SS, for 5100-10-IT - SO2
SPM27088
Sensor Assembly, 316SS, for 5100-12-IT - NO2
SPL21824
Transmitter
SPL21810
Enclosure, Transmitter, Aluminum
SPL21823
Enclosure Transmitter, 316SS
SPL21825
Interface Board without Relay
SPL21829
Interface Board with Relay
Options
5311-00
Rainshield
5311-02
Rainshield with calibration port
Calibration Items
1250-01
Gas Sensor Calibration Kit, Type A (CO, H2S, Air)
1250-02
Gas Sensor Calibration Kit, Type B (SO2, NO2)
1250-03
Gas Sensor Calibration Kit, Type C (Cl2)
1260-00
Gas Cylinder, Air, (Type A), 105 liters
1260-04
Gas Cylinder, CO 100 PPM, (Type A) 57 liters
1260-05
Gas Cylinder, H2S in N2, 25 PPM, (Type A) 57 liters
1260-06
Gas Cylinder, Cl2, 5 PPM, (Type C) 104 liters
1260-10
Gas Cylinder, SO2, 50 PPM, (Type B) 74 liters
1260-12
Gas Cylinder, NO2, 10 PPM, (Type B) 74 liters
1260-13
Gas Cylinder, CO, 1000 PPM, (Type A), 105 liters
5360-00
Calibrator Head Standard
1256-01
Regulator Type A Calibrator
1256-02
Regulator Type B Calibrator
1265-03
Regulator Type C Calibrator
5358-01
Calibration Adapter - Direct, Standard
5358-51
Calibration/Configuration Magnetic Tool, 4501-XX
9210-00
Calibrator, Permeation Tube
9211-09
Permeation Tube, Ammonia, NH3
SPX27057
Kit, Sensor Recharge, for 5100-25-IT
SPX27061
Kit, Sensor Recharge, for 5100-26-IT
See Appendix E for 5100-25-IT and 5100-26-IT
Page 42
Model 5100-XX-IT Toxic Gas Sensor Module
Page 42 of 62
Appendix D. Limited 2 Year Warranty
Sierra Monitor Corporation warrants its products to be free from defects in workmanship or material under normal use and service for two years after date of shipment. SMC will repair or replace without charge any equipment found to be defective during the warranty period. Final determination of the nature and responsibility for defective or damaged equipment will be made by SMC personnel.
All warranties hereunder are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without SMC approval or which have been subjected to accident, improper maintenance, installation or application, or on which original identification marks have been removed or altered. This Limited Warranty also will not apply to interconnecting cables or wires, consumables (i.e. calibration gases, batteries, sensors), nor to any damage resulting from battery leakage.
In all cases SMC’s responsibility and liability under this warranty shall be limited to the cost of the equipment. The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty provision and compliance with such instruction shall be a condition of this warranty.
Except for the express warranty stated above, SMC disclaims all warranties with regard to the products sold hereunder including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of SMC for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product.
Page 43
Model 5100-XX-IT Toxic Gas Sensor Module
Page 43 of 62
Appendix E. Remote Sensor Drawing
NOTE: Drawing using aluminum enclosure available from Sierra Monitor.
Unless otherwise specified, NOTE the following:
Provide access for sensor replacement during installation.
For field upgrade, relocate 5100-XX-IT-A1/A2 Sensor to 5394-50 as shown in Detail A (Figure 24).
Use belden cable #9925 or SMC P/N 53056 (see in Figure 24).
Terminal blocks float in each enclosure.
The remote sensor assembly is meant to be supported by rigid conduit.
Calibrate post installation.
For conduit with M20 thread, install adapter 39218 (2X) where shown (see in Figure 24).
NOTE: The distance between the sensor and transmitter is 15 feet for models 5100-04/05/10-IT. However,
the distance is 10 feet for Model 5100-03-IT.
Typical Wire Connections at all Terminations:
Color
Position
Red
1
Black
2
White
3
Bare Wire
4
Figure 24: Remote Sensor Drawing
Page 44
Model 5100-XX-IT Toxic Gas Sensor Module
Page 44 of 62
Appendix F. Modbus Memory Map
Register Description Read/ Comments
Write
40001 Concentration R Gas concentration multipl ied by Gas Scale (e .g. 209 = 20.9%) 40002 Temperature R Temperature in degrees Cel cius scaled by a factor of 10
40003 Alarm Relay R/W
Boolean indi cating the Al arm relay status (0 = No Alarm, 1 - (High) Alarm). Clearing bit wi ll reset alarm.
40004 Warning Relay R/W Boolean indi cating the Warning relay status (0 = No Alarm, 1 - Warning Al arm). Clearing bit wil l reset alarm. 40005 Warning Setpoint R/W Warning Alarm set point, used for activating Low Alarm multiplie d by Gas Scale 40006 Alarm Setpoint R/W Al arm set point, used for activating High Alarm multipl ied by Gas Scale 40007 CCC detail R e .g. 0001, 0007, etc 40008 Software Revision U R e .g. 201 denote s ve rsi on 2.01 40009 Software Revision L R e .g. version aA is denoted by 0x6141 40010 Modbus Map Revi sion R e .g. version 2 is denoted as v2 = 0x7602 40011 Trouble Bits R 0 = no trouble 40012 Trouble R 1 for any trouble, 0 = no trouble 40013 Alarm Immediate R Like 40003, but neve r latched 40014 Warning Immediate R Like 40004, but never latched 40015 Serial Numbe r U R Se rial Number - 24 bits (presently set to Random ID) 40016 Serial Numbe r L R 40017 Run time U R In seconds 40018 Run time L R In se conds 40019 Gas (MAX) value R Gas concentration multipl ied by Gas Scale 40020 Model R Read as 51028 for IR, 51002 for Cat Bead, 51005 for TOXIC H2S etc. 40021 Range RW 40022 Units RW 1 = %LEL, 2 = %VOL
40023 Module Status R
A bit is defi ned for each of the following states: (no bit set also means some sort of trouble) Module State - In Calibration (0x40) = cal failed Module State - In Calibration (0x20) = cal passed Module State - In Calibration (0x10) Module State - Trouble (0x08) Module State - Warming (0x04) Module State - RUN (ALARM) (0x02) Module State - RUN (0x01)
40024 Gas Scale R V alue li ke 1, 10 or 100 40025 Cal Concentration leve l R/W Gas concentration value used during calibration, multiplied by Gas Scale 40026 Calibration due, in days R Gets rese t to 365 / 180 if cal successful , Less then 0 = calibration due 40027 Calibration Count R Counter increments if cal successful (0 = not cali brated) 40028 Display Gas Output R Gas value seen on display (i.e. use ful during calibration) 40029 Gas (MIN) value R Minimum gas value multi plied by Gas Scale 40030 Alarm Time U R In seconds 40031 Alarm Time L R 40032 Warning Time U R In seconds 40033 Warning Time L R 40034 Trouble Time U R In se conds 40035 Trouble Time L R 40036 Max Gas Time U R I n seconds 40037 Max Gas Time L R 40038 Min Gas Time U R In seconds 40039 Min Gas Time L R 40040 Rotary Switch R Module address (If val ue < 16, the value refle cts positi on of rotary switch) 40041 Restart Count R Restart count 40042 Alarm Relay Action Setup RW Alarm Relay Action Setup - Latch/Sentry/Non-l atch 40043 Warning Relay Action Setup RW Warning Rel ay Action Setup - Latch/Sentry/Non-l atch 40044 4-20mA Cal ibOut Value R 40045 4-20mA TroubleOut Value R 40046 Gas Factor RW Applies only to -02 40047 Force Value RW 40048 Module Number RW
40049 Command Register RW
0 indicates command failed, 1 = passed, supported commands: Lock keys = 101 Unlock keys = 102 Abort calibration = 103 Apply zero gas = 104 Apply span gas = 105 Ack alarms = 106 Reset sensor = 107 Clear mem = 108 Speed up startup = 109 Clear eeprom = 110 Force address disp= 111 Clear address disp= 112
40050 Last command executed R Displays last exe cuted command from register 40049
Page 45
Model 5100-XX-IT Toxic Gas Sensor Module
Page 45 of 62
Appendix G. HART
Figure 25: Wiring Connections Table
Page 46
Model 5100-XX-IT Toxic Gas Sensor Module
Page 46 of 62
Figure 26: 4-20 mA Circuits Types 5100-XX-IT – Connections - HART
Page 47
Model 5100-XX-IT Toxic Gas Sensor Module
Page 47 of 62
Hart Protocol Menu
HART (Highway Addressable Remote Transducer) Protocol is the global standard for sending and receiving digital information across analog wires between smart devices and control or monitoring system. HART is a bi­directional communication protocol that provides data access between intelligent field instruments and host systems. A host can be any software application from technician's hand-held device or laptop to a plant's process control, asset management, safety or other system using any control platform. The HART protocol used on this gas sensor module has been developed to comply fully with the HART specifications outlined by the HART Communication Foundation (HCF).
If the HART interface board has been provided with the gas sensor module, then HART will operate via the 4-20 mA interface automatically. While no specific handheld device is necessary for HART operation, the gas sensor module does comply with the HART standard so any handheld or PC-based HCF certified tool will work.
The information below provides the operator instructions on connecting HART and the HART Protocol Menu.
Common Practice Commands
The following Common Practice commands are implemented.
Command
Number
Byte Number
Description
38
N/A
Reset Configuration Changed Flag
48
0
Returns Module Status: Module State - In Calibration (0x10) (0x30 = cal passed) (0x50 = cal failed) Module State - Trouble (0x08) Module State – Warming (0x04) Module State - RUN (ALARM) (0x02) Module State – RUN (0x01)
48
1
Returns Alarm Relay Status: 0x01 = Alarm Relay ON 0x02 = Warning Relay ON 0x03 = Both Relays ON
48
2
Returns Trouble Status (non-zero value indicates Trouble)
48
3
Returns maximum gas value
48
4
Returns gas value used during calibration
48
5
Returns IT-Series Model Number: 0x28 = 5100-IT-28
Device Specific Commands
The Device Specific commands are used for accessing the unique features of the 5100-IT Series of gas sensor modules.
A response code of 0 indicates SUCCESS. This applies to all commands. As per HART specification, all command responses will have a response code and status byte, plus any command specific data bytes as indicated in the following tables.
Page 48
Model 5100-XX-IT Toxic Gas Sensor Module
Page 48 of 62
Device Specific Commands Summary
Command
Number
Description
130
Key Press
131
Set Alarm Level
132
Set Warning Level
133
Set Alarm Relay Action
134
Set Warning Relay Action
135
Reset Alarms
136
Abort Calibration
137
Set Calibration Gas Level
138
Apply ZERO Gas
139
Apply SPAN Gas
140
RESERVED
141
Force Gas Value
142
Reset Force Gas Value
143
Read MODBUS RTU Register
144
Set Transducer Serial Number
145
Set Calibration Mode output current
146
Set Trouble Mode output current
147
RESERVED
148
RESERVED
149
Set Gas Factor
150
Write MODBUS RTU Register
Command 130: Key Press
This command mimics the front plate key inputs.
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
1=Enter 2=Down 4=Up 8=Menu
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns same value as passed
Command 131: Set Alarm Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Alarm level, range 0 to 60. Default is 60
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Alarm Level
Command 132: Set Warning Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Warning level, range 0 to 60. Default is 20.
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Warning Level.
Page 49
Model 5100-XX-IT Toxic Gas Sensor Module
Page 49 of 62
Command 133: Set Alarm Relay Action
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
0=Latching (default) 2=Non-Latching
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Alarm Relay Action value
Command 134: Set Warning Relay Action
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
0=Latching 2=Non-Latching (default)
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Warning Relay Action value
Command 135: Reset Alarms
This command will reset any alarm relays that have been latched, providing the alarm condition is no longer present on the sensor.
Request Data Bytes: None Command Specific Response Data Bytes: None
Command 136: Abort Calibration
This command aborts the calibration procedure. Request Data Bytes: None Command Specific Response Data Bytes: None
Command 137: Set Calibration Gas Level
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Calibration Gas Level, range 0 to 100
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Calibration Gas Level
Command 138: Apply ZERO Gas
This command will signal the module to accept the present sensor output to as the ZERO gas condition.
Request Data Bytes: None Command Specific Response Data Bytes: None
Command 139: Apply SPAN Gas
This command will signal the module to accept the present sensor output to as the SPAN gas condition.
Request Data Bytes: None Command Specific Response Data Bytes: None
Page 50
Model 5100-XX-IT Toxic Gas Sensor Module
Page 50 of 62
Command 141: Force Gas Value
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Sets the Force Gas Value, range 0 to 100
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Force Gas Value
Command 142: Reset Force Gas Value
This command will signal the module to set the Force Gas Value to 0.
Request Data Bytes: None Command Specific Response Data Bytes: None
Command 143: Read MODBUS RTU Register
This command allows the reading of sensor module data as defined in the MODBUS RTU register map.
Request Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Address of Requested Modbus Register 40001 = lowest address 65535 = highest address
Command Specific Response Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Returns contents of selected Modbus Register
Command 144: Set Transducer Serial Number
This command allows user to add a device specific serial number, if required. This is different from the serial number assigned by the Manufacturer.
Request Data Bytes:
Byte
Format
Description
0-2
Unsigned-24
Serial Number
Command Specific Response Data Bytes:
Byte
Format
Description
0-2
Unsigned-16
Returns new Serial Number
Command 145: Set Calibration Mode output current
Request Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Current output required during calibration, in micro­amperes
Command Specific Response Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Returns new Calibration current value
Page 51
Model 5100-XX-IT Toxic Gas Sensor Module
Page 51 of 62
Command 146: Set Trouble Mode Output Current
Request Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Current output required to indicate Trouble, in micro­amperes
Command Specific Response Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Returns new Trouble current value
Command 149: Set Gas Factor
Request Data Bytes:
Byte
Format
Description
0
Unsigned-8
Sets the Gas Factor value, range 30-200
Command Specific Response Data Bytes:
Byte
Format
Description
0
Unsigned-8
Returns new Gas Factor value
Command 150: Write MODBUS RTU Register
This command allows the writing of sensor module data as defined in the MODBUS RTU register map.
Request Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Address of Designated Modbus Register 40001 = lowest address 65535 = highest address
2-3
Unsigned-16
Data for Designated Modbus Register
Command Specific Response Data Bytes:
Byte
Format
Description
0-1
Unsigned-16
Returns MODBUS register value
Page 52
Model 5100-XX-IT Toxic Gas Sensor Module
Page 52 of 62
Appendix H. Gas Sensor Module Calibration
Appendix H.1. Calibration for 5100-25-IT
NOTE: Supplement to Instructions in Section 6.
1. Equipment Required
The following tools and equipment will be required for calibration:
Permeation Device Calibrator (Model 9210-00), with Ammonia
Permeation Tube (Model 9211-09).
Calibration Adapter (Model 5358-01)
2. Permeation Device Calibrator Operation
The calibrator is a portable device with a pump that operates from a 1.5 V alkaline size D battery. A fresh battery should allow 8 hours of continuous operation. The permeation device will probably be shipped separately, usually in a pipe with screw cap ends, one of which will be marked for opening. The permeation device should come fitted with a scrubber, but, as a precaution, open the pipe and remove the permeation device in a well ventilated area.
a) Remove the scrubber from the permeation device and attach the device
to the stainless steel “Tee” assembly inside the calibrator box. Tighten the connection with a wrench, with the device resting inside one of the
openings in the rack. b) Close the door of the calibrator and turn on the pump switch. c) Set the flow rate for the desired ppm of ammonia. The desired flow rate
for calibration of the Model 5100-25-IT is 300-500 cc/min. and the
concentration range should be set between the high alarm point and
100 ppm full scale.
The carrier gas flow rate can be calculated from the formula:
P = (CxF)/K
Where P = the permeation rate in nanograms/min., and C = ppm of span gas, F = carrier gas flow rate in
ml./min. and K = 1.437 for ammonia. The permeation rate for the temperature read on the front of the calibrator can be obtained from permeation rate vs. temperature chart.
d) Run the gas calibrator for 30 min. before attaching the output hose to the ammonia sensor. e) When the use of the calibrator is completed, do not turn off the pump unit until the permeation device has
been removed from the “Tee” and the gas scrubber reattached. Failure to remove the permeation device
when the pump has been turned off will contaminate the system with high amounts of the permeating gas
and result in erroneous concentrations. The device, with scrubber, can be stored in the storage rack
overnight, but should not stay in the gas calibrator for longer storage periods.
3. Sensor Assy. Rebuild Procedure
Follow procedures outlined in Electrolyte/Membrane Replacement (Figure 28). The electrolyte needs to be replaced at least every six months, or if there is evidence of low sensor
output. The membrane and O-Ring should be replaced at the same time that the electrolyte is replaced.
Figure 27: Electrolyte
Replacement
Page 53
Model 5100-XX-IT Toxic Gas Sensor Module
Page 53 of 62
4. Equipment Required
Electrolyte Recharge Kit (SPX27057 for Model 5100-25-IT and SPX27061 for Model 5100-26-IT)
consisting of the electrolyte, package of 5 membranes, 5 0-Rings and alcohol wipes to clean the
electrode.
Tweezers to aid in removing and replacing the membrane.
5. Preparation
a) Transfer the necessary electrolyte into a plastic wash bottle. b) Confirm that system power has been removed. c) Remove the transmitter electronics board from the main housing and unplug the sensor harness from the
transmitter electronics. d) Unscrew the sensor assembly from the bottom of the enclosure. e) Remove the assembly cover from the sensor assembly.
CAUTION: Only remove the sensor cover from the assembly, DO NOT attempt to unscrew the sensor
from the mounting assembly or you could damage the wiring.
6. Electrolyte Replacement (Figure 27)
a) Invert the sensor assembly so that the membrane is upward. b) Insert the sensor assembly in a clamp, if available, and unscrew the box nut counterclockwise, and take
out the hold-down ring, membrane and O-Ring. Then pour out the old electrolyte into a beaker for
disposal.
CAUTION: Avoid contact of the electrolyte with skin, eyes and clothing. In case of contact wash
immediately - rubber gloves and safety glasses are recommended.
c) Clean the electrode with the alcohol wipes included in the electrolyte replacement kit. All oxidation
residue must be removed before re-assembly. When dirt is found inside the sensor assembly, wash the
dirt off with electrolyte. d) Hold the sensor assembly with the sensing electrode up. Close the pressure compensation port with a
tissue. Pour approximately 10cc of fresh electrolyte from the wash bottle into the sensor, and rinse the
inside of the sensor. Empty the electrolyte into the beaker. e) Place the O-ring on top of the washer. Then transfer one drop of electrolyte from the plastic wash bottle to
the exposed electrode surface. f) Place a new membrane into the hold-down ring and carefully place the plastic hold-down ring on top of
the sensor body with the ring ears in the slots of the sensor body. Then fasten the sensor cover over the
assembly.
CAUTION: Never touch the membrane with your hands. Use rubber gloves or tweezers.
g) Invert the sensor so that the membrane is down. If this is the first time that electrolyte has been replace,
there should be a round Mili-Seal covering the pressure compensation screw. This Mili-Seal prevented
loss of electrolyte during shipment. Remove and discard the round Mili-Seal and remove the screw to
expose the electrolyte supply port. h) Transfer electrolyte from the plastic wash bottle into the electrolyte supply port (Figure 27) until the level
is at the maximum level line on the side of the sensor body. i) Replace the pressure compensation screw into the electrolyte supply port. j) Replace the sensor housing cover. k) Install the sensor assembly into the enclosure hub and tighten firmly. l) Reconnect the sensor harness to the transmitter, carefully install the transmitter into the housing and
restore system power.
Page 54
Model 5100-XX-IT Toxic Gas Sensor Module
Page 54 of 62
m) Recalibrate the transmitter following the instructions in Section 6.
Spare Parts
SPX27057 Recharge Kit NH3 SPX27061 Recharge Kit HF SPX57009 Membrane Kit (both) SPX99017 Electrolyte NH3 SPX99018 Electrolyte HF 5200-25-IT Sensor Assy NH3 5200-26-IT Sensor Assy HF
Figure 28: Sensor Assembly – Exploded View
(WHITE)
Page 55
Model 5100-XX-IT Toxic Gas Sensor Module
Page 55 of 62
Figure 29: Model 5100-25-IT-A1/A2 and Model 5100-26-IT-A1/A2
Figure 30: Model 5100-25-IT-S1 and Model 5100-26-IT-S1
Page 56
Model 5100-XX-IT Toxic Gas Sensor Module
Page 56 of 62
Appendix H.2. Calibration for 5100-26-IT
NOTE: Supplement to Instructions in Section 6.
1. The calibration of the Model # 5100-26-IT HF gas sensor module is accomplished by using a 5 ppm Chlorine standard (Type C kit with # 1260-06 cylinder). Scale range is 0-10 ppm with 5 ppm being mid-scale. 5 ppm Cl2 is equivalent to 5 ppm HF.
2. Reference Figure 22 for the calibration procedure sub menu. This procedure requires the use of the magnetic wand to activate the keys and confirm each entry step. Follow prompts on the module display as navigation is detailed in Figure 22.
3. Enter actual span gas concentration that is printed on the cylinder being used when prompted.
4. Refer to the sensor rebuild and electrolyte replacement procedure as detailed in Appendix H.1.
Page 57
Model 5100-XX-IT Toxic Gas Sensor Module
Page 57 of 62
Appendix I. 5100-88-IT Dimensions
Figure 31: Model 5100-88-IT-A1 Dimensions
Page 58
Model 5100-XX-IT Toxic Gas Sensor Module
Page 58 of 62
Appendix J. 5100-05-IT FM Performance Approval
Page 59
Model 5100-XX-IT Toxic Gas Sensor Module
Page 59 of 62
Page 60
Model 5100-XX-IT Toxic Gas Sensor Module
Page 60 of 62
Appendix K. 5100-XX-IT SIL-2 Certificates
Our SIL-2 Certificates for the 5100-04-IT, the 5100-05-IT and the 5100-06-IT can be found on our website at:
http://www.sierramonitor.com/gas/support/downloads.php
Appendix L. 5100-05-IT ATEX Certificate
Page 61
Model 5100-XX-IT Toxic Gas Sensor Module
Page 61 of 62
Page 62
Model 5100-XX-IT Toxic Gas Sensor Module
Page 62 of 62
Loading...