Sierra Monitor Corporation 5000 User Manual

0 (0)

MODEL 5000

S E N T R Y

GAS MONITORING SYSTEM

Version 6

Sensors Service

APPLICABILITY & EFFECTIVITY

Effective for all Sentry systems manufactured after September 1, 1995.

Instruction Manual Part Number T12001-A1

Sierra Monitor Corporation

1991 Tarob Court, Milpitas, CA 95035

(408) 262-6611

Applicability & Effectivity

Sierra Monitor Corporation 5000 User Manual

SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

 

 

 

7.5 COMBUSTIBLE GAS SENSOR MODULE

 

 

(5100-02)

 

 

 

7.5.1 DESCRIPTION

 

 

The Combustible Gas Module includes the sensor

 

 

and electronic assembly installed in an explosion

 

 

proof housing. The sensor screws into one hub of

 

 

the enclosure and plugs into the bottom

 

 

electronics card via a six pin connector. Cabling

 

 

to the controller connects to a three pin spring

 

 

loaded terminal strip.

 

 

7.5.2 TROUBLE ANALYSIS

 

 

Electrical adjustment, or replacement of the

 

 

sensor will be necessary under the following

 

 

conditions:

 

 

 

Controller displays the following error

 

 

 

messages:

 

 

 

CHK BRIDGE VOLT

 

 

 

SENSOR FAILURE

 

 

 

LOW SENSITIVITY

 

 

 

False readings or alarms are received

 

 

 

due to sensor inaccuracy.

 

 

Warning: : During sensor adjustments the

 

 

 

Figure 7-6

concentration reading on the controller will be

 

Cover Plate - Combustible Gas Module

inaccurate and alarm level concentrations may be

 

 

displayed. If false activation of the alarm relays

 

 

will cause a problem disconnect the relay wiring

Comparing to the "GND" (ground) test point measure

prior to adjustment or turn the module off using the

"BRIDGE VOL" (bridge voltage) and adjust to 2.00 VDC

CHANGE MODULE mode.

using "BRIDGE ADJ" potentiometer. Turn counter-

 

 

 

 

 

NOTE

 

clockwise to increase.

 

Although all the necessary data can be collected with a

 

 

 

voltmeter at the sensor module, some helpful information

 

7.5.3.2 ZERO ADJUSTMENT

 

can be displayed or printed. See diagnostic codes 0004

 

 

and 0008 in Appendix C.

 

Comparing to ground measure SIGNAL OUT and adjust to

 

 

 

 

 

 

 

 

7.5.3 ADJUSTMENT PROCEDURE

0.18 VDC by turning ZERO ADJ potentiometer. Turn

counter clock-wise to increase.

Prior to reading voltages and making adjustments perform

 

a visual inspection to confirm that there are no physical

7.5.3.3 SPAN ADJUSTMENT

problems such as water in the electronics enclosure,

1. Determine the concentration of the calibration

wiring damage or corrosion.

Use Figure 7-6 to locate test points during the following

gas (CG) in % LEL.

2. Use the following formula to determine the

procedures:

Caution: The area must be free of flammable vapors or

response to the calibration gas at SIGNAL OUT.

Voltage (Signal Out) = (CG+12)/60

gases during any adjustments or maintenance procedures.

 

 

 

 

For Example:

7.5.3.1 BRIDGE VOLTAGE ADJUSTMENT

Confirm that no combustible gas is present and remove sensor module cover.

If CG = 50% L.E.L Signal Out = (50+12)/60

= 1.03 volts

SERVICE

Page: 69

SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

3.Using a Sierra Monitor Model 1200-26 Gas Calibrator with a Sierra Monitor Model 535800 calibration adapter, or Model 5360 Gas Delivery Fitting. Apply the calibration gas at a flow rate of 100 cc/min. until the signal out voltage stops changing (approx. 2 minutes). Then adjust span adj. potentiometer to the calculated value.

4.Remove the calibration gas and wait for the voltage to drop. If the voltage is below 0.30 VDC the procedure is complete. If the voltage is above 0.30 VDC repeat the zero and span adjustments.

5.Calibrate the sensor using the instruction manual procedure.

A LOW SENSITIVITY message during calibration indicates one of the two conditions for a combustible sensor:

The zero gas voltage must be less than

.3 volts.

The new span change (span voltage less zero voltage) must exceed 30% of the calculated span change.

For Example:

Zero gas voltage= .20 VDC Span voltage = .43 VDC

Span change is .43-.2 = .23 VDC

As the calculated span value (from step 5 above) is 1.03, the calculated span change is: 1.03-.2 = .87

By calculation: 30% of .87 = .26

The new span change (.23) is less than 30% of the calculated span change so a

LOW SENSITIVITY message will be

displayed.

The LOW SENSITIVITY calculation is made to insure that the calibration does not adjust the electronics so that the "gain" is large enough to cause false alarms due to minor drift or due to minimal electronic noise.

7.5.4 SENSOR REPLACEMENT

If any of the above adjustments cannot be completed successfully the sensor and/or electronics will require replacement. To check the sensor unplug it from the electronics and remove from the housing so that continuity and resistance tests can be made. Table 7-4 provides the pin out for the sensor assembly.

PIN

FUNCTION

WIRE COLOR

 

 

 

1

DETECTOR

WHITE

 

 

 

2

COMPENSATOR

BROWN

 

 

 

3

CALIB. SW.

RED

 

 

 

4

DETECTOR

BLACK

 

 

 

5

DRAIN

BARE

 

 

 

6

CALIB. SW.

GREEN

 

 

 

Table 7-5

Combustible Gas Sensor

Wiring Pin Out

1.Detector element resistance should be between 1.5 to 2 ohms.

2.Compensator element resistance should be between 1.5 to 2 ohms.

3.Calibration switch should be open in normal operation and should close when the calibration magnet is applied.

If the sensor fails any of these tests it should be replaced. If the sensor passes these tests the electronics are suspect and BOTH the sensor and electronics should be replaced. (The electronics and sensor which are removed should be returned to the factory for evaluation).

When a new sensor is installed the following actions must be taken:

Make the bridge voltage adjustment immediately to avoid over-voltage damage to the sensor.

Use diagnostic code 0020 to establish nominal value calibration.

Allow 24 hours for full stabilization of the sensor, recheck the zero and span adjustments and calibrate the sensor module.

SERVICE

Page: 70

SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

7.5.5 COMBUSTIBLE GAS SCALING FACTORS

For combustible gas monitoring, a calibration standard of Methane or Propane my be used in conjunction with scaling factors to cause Sentry concentration display and alarm function in %LEL scale of another gas as follows:

 

GAS

 

METHANE

 

PROPANE

 

 

 

 

FACTOR

 

 

FACTOR

 

 

Acetaldehyde

 

 

60

 

 

109

 

 

Acetic Acid

54

 

98

 

 

Acetic Anhydride

46

 

83

 

 

Acetone

52

 

94

 

 

Acetylene

57

 

103

 

 

Alkyl Alcohol

51

 

92

 

 

Ammonia

126

 

229

 

 

n-Amyl Alcohol

33

 

59

 

 

Aniline

39

 

71

 

 

Benzene

41

 

74

 

 

Biphenyl

25

 

45

 

 

1,3-Butadiene

56

 

101

 

 

n-Butane

58

 

106

 

 

iso-Butane

52

 

94

 

 

Butene-1

45

 

82

 

 

cis-Butene-2

48

 

88

 

 

trans-Butene-2

51

 

92

 

 

n-Butyl Alcohol

34

 

62

 

 

iso-Butyl Alcohol

53

 

96

 

 

tert-Butyl-Alcohol

74

 

134

 

 

n-Butyl Benzene

31

 

57

 

 

iso-Butyl Benzene

32

 

58

 

 

n-Butyric Acid

38

 

69

 

 

Carbon Disulfide

18

 

32

 

 

Carbon Monoxide

75

 

137

 

 

Carbon Oxysulphide

93

 

169

 

 

Chlorobenzene

34

 

62

 

 

Cyanogen

89

 

162

 

 

Cyclohexane

41

 

74

 

 

Cyclopropane

62

 

113

 

 

n-Decane

33

 

59

 

 

Diethylamine

49

 

88

 

 

Dimethylamine

58

 

105

 

 

2,3-Dimethylpentane

40

 

72

 

 

2,3-Dimethylpropane

40

 

72

 

 

Dimethylsulphide

43

 

79

 

 

1,4-Dioxane

45

 

81

 

 

Epichlorohydrin

45

 

82

 

 

Ethane

68

 

123

 

 

Ethyl Acetate

51

 

93

 

 

Ethyl Alcohol

73

 

132

 

 

Ethylamine

53

 

95

 

 

Ethyl Benzene

36

 

65

 

 

Ethyl Bromide

91

 

165

 

 

Ethyl Chloride

57

 

103

 

 

Ethylcyclopentane

40

 

72

 

 

Ethylene

71

 

128

 

 

Ethylenedichloride

66

 

120

 

 

Ethyleneoxide

52

 

94

 

NOTES:

 

GAS

 

METHANE

 

PROPANE

 

 

 

 

FACTOR

 

 

FACTOR

 

 

Diethyl Ether

 

 

46

 

 

84

 

 

Dimethoxyethane

42

 

75

 

 

Dimethyl Ether

63

 

113

 

 

Dimethylformamide

46

 

83

 

 

Ethyl Formate

44

 

80

 

 

Ethylmercaptan

56

 

102

 

 

n-Heptane

39

 

70

 

 

n-Hexane

37

 

67

 

 

Hydrazine

45

 

82

 

 

Hydrogencyanide

48

 

86

 

 

Hydrogen

77

 

139

 

 

Hydrogen Sulfide

41

 

74

 

 

Methane

100

 

181

 

 

Methyl Actetate

50

 

90

 

 

Methyl Alcohol

86

 

156

 

 

Methylamine

77

 

140

 

 

Methyl Bromide

90

 

162

 

 

Methyl Chloride

102

 

186

 

 

Methylcyclohexane

44

 

80

 

 

Methylenedichloride

93

 

168

 

 

Methylethylether

44

 

80

 

 

Methylethylketone

41

 

75

 

 

Methyl Formate

67

 

121

 

 

Methylmercaptan

61

 

110

 

 

Methylpropionate

51

 

93

 

 

Methyl n-propylketone

40

 

73

 

 

Napthalene

34

 

62

 

 

Nitromethane

34

 

62

 

 

n-Nonane

31

 

57

 

 

n-Octane

37

 

68

 

 

n-Pentane

46

 

83

 

 

i-Pentane

46

 

84

 

 

Propane

55

 

100

 

 

n-Propyl Alcohol

47

 

85

 

 

n-Propylamine

48

 

88

 

 

n-Propylchloride

50

 

90

 

 

Propylene

52

 

93

 

 

Propyleneoxide

46

 

83

 

 

iso-Propylether

44

 

79

 

 

Propyne

42

 

75

 

 

Toluene

40

 

73

 

 

Triethylamine

40

 

72

 

 

Trimethylamine

48

 

88

 

 

Vinylethylether

42

 

76

 

 

o-Xylene

36

 

65

 

 

m-Xylene

39

 

71

 

 

p-Xylene

39

 

71

 

 

JP-4 (Jet Fuel)

41

 

73

 

1.Scaling factors are not FMRC approved.

2.Base data source: EEV sensor specification catalog. (EEV claims some data is the result of specific tests, other data is empirically derived.)

SERVICE

Page: 71

SMC sierra monitor corporation

Sentry Instruction Manual - Version 6

7.6 OXYGEN MODULE (5100-03)

7.6.1 DESCRIPTION

The Oxygen Module includes the electronic assembly installed in an explosion proof housing and the electrochemical sensor connected to one hub of the enclosure. Cabling from the controller connects to a three pin spring loaded terminal strip.

7.6.2 TROUBLE ANALYSIS

Electrical adjustment, or replacement of the sensor will be necessary under the following conditions:

Controller displays the following error messages:

CHANGE SENSOR

SENSOR FAILURE

REPLACE SENSOR

False readings or alarms are received due to sensor inaccuracy.

Warning: During

sensor

adjustments

the

 

concentration reading on the controller will be

 

inaccurate and alarm level concentrations may be

Figure 7-7

displayed. If false activation of the alarm relays

Cover Plate - Oxygen Module

will cause a problem disconnect the relay wiring

 

prior to adjustment or turn the module off using

 

the "Change Module" mode.

 

 

 

7.6.4 SENSOR REPLACEMENT

 

 

 

 

 

NOTE

 

The oxygen sensor should be replaced when it can no

 

Although all the necessary data can be collected with a

 

 

voltmeter at the sensor module, some helpful information

 

longer be calibrated correctly or when the signal output

 

can be displayed on the controller or printed. See

 

drops to zero. Generally this is every twelve to eighteen

 

diagnostic codes 0004 and 0008 in Appendix C.

 

months. The Sentry clock keeps track of the age of the

 

 

 

 

 

 

7.6.3 ADJUSTMENT PROCEDURE

sensor if it is correctly initialized in the "change module"

mode.

 

 

 

 

 

Prior to reading voltages and making adjustments perform

 

 

 

 

 

When a new sensor is installed use the CHANGE MODULE

a visual inspection to confirm that there are no physical

menu selection to

answer yes

to

the "new

sensor?"

problems such as water in the electronics enclosure,

question. After

nine months

the

CHANGE

SENSOR

wiring damage or corrosion.

message will be displayed to warn that a new sensor

Use Figure 7-7 to locate test points during the following

should be installed.

 

 

 

 

 

procedures.

The SENSOR FAILURE message for Oxygen sensors does

 

 

 

7.6.3.1 SIGNAL ADJUSTMENT

not indicate a specific failure of the sensor but indicates

that the sensor is not correctly connected to the

 

 

 

It is unlikely that any electrical adjustment will be

electronics. If this message appears check that the sensor

required except when a new sensor is installed.

harness is correctly installed to the connector on the

To make the electrical adjustment connect a voltmeter to

bottom electronics board.

 

 

 

 

When sensor replacement is required, open the cover of

SIGNAL OUT (pos) and GND (ground) and use

SENSITIVITY ADJ potentiometer to set the voltage equal to

the sensor module, remove the transmitter and disconnect

1/10 of the actual oxygen concentration. In clean air the

the sensor wiring harness from the back of the transmitter.

concentration is 20.9% so the voltage should be set at

 

 

 

 

 

 

2.09 VDC.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SERVICE

 

 

 

 

 

 

 

Page:

72

 

 

 

 

 

Loading...
+ 10 hidden pages