Sierra Monitor Corporation 5000 User Manual

MODEL 5000
S E N T R Y
GAS MONITORING SYSTEM
Sensors Service
APPLICABILITY & EFFECTIVITY
Effective for all Sentry systems manufactured after September 1, 1995.
Instruction Manual Part Number T12001-A1
Sierra Monitor Corporation
1991 Tarob Court, Milpitas, CA 95035
(408) 262-6611
Applicability & Effectivity
SMC sierra monitor corporation Sentry Instruction Manual - Version 6
7.5 COMBUSTIBLE GAS SENSOR MODULE
(5100-02)
7.5.1 DESCRIPTION
The Combustible Gas Module includes the sensor and electronic assembly installed in an explosion proof housing. The sensor screws into one hub of the enclosure and plugs into the bottom electronics card via a six pin connector. Cabling to the controller connects to a three pin spring loaded terminal strip.
7.5.2 TROUBLE ANALYSIS
Electrical adjustment, or replacement of the sensor will be necessary under the following conditions:
Controller displays the following error
messages:
CHK BRIDGE VOLT
SENSOR FAILURE
LOW SENSITIVITY
False readings or alarms are received
due to sensor inaccuracy.
Warning: : During sensor adjustments the concentration reading on the controller will be inaccurate and alarm level concentrations may be displayed. If false activation of the alarm relays will cause a problem disconnect the relay wiring prior to adjustment or turn the module off using the
CHANGE MODULE mode.
NOTE
Although all the necessary data can be collected with a
voltmeter at the sensor module, some helpful information
can be displayed or printed. See diagnostic codes 0004
and 0008 in Appendix C.
7.5.3 ADJUSTMENT PROCEDURE
Prior to reading voltages and making adjustments perform a visual inspection to confirm that there are no physical problems such as water in the electronics enclosure, wiring damage or corrosion.
Use Figure 7-6 to locate test points during the following procedures:
Caution: The area must be free of flammable vapors or gases during any adjustments or maintenance procedures.
7.5.3.1 BRIDGE VOLTAGE ADJUSTMENT
Confirm that no combustible gas is present and remove sensor module cover.
Figure 7-6
Cover Plate - Combustible Gas Module
Comparing to the "GND" (ground) test point measure "BRIDGE VOL" (bridge voltage) and adjust to 2.00 VDC using "BRIDGE ADJ" potentiometer. Turn counter­clockwise to increase.
7.5.3.2 ZERO ADJUSTMENT
Comparing to ground measure SIGNAL OUT and adjust to
0.18 VDC by turning
ZERO ADJ potentiometer. Turn
counter clock-wise to increase.
7.5.3.3 SPAN ADJUSTMENT
1. Determine the concentration of the calibration
gas (CG) in % LEL.
2. Use the following formula to determine the
response to the calibration gas at
SIGNAL OUT.
Voltage (Signal Out) = (CG+12)/60 For Example: If CG = 50% L.E.L Signal Out = (50+12)/60
= 1.03 volts
SERVICE
Page: 69
SMC sierra monitor corporation Sentry Instruction Manual - Version 6
3. Using a Sierra Monitor Model 1200-26 Gas
Calibrator with a Sierra Monitor Model 5358­00 calibration adapter, or Model 5360 Gas Delivery Fitting. Apply the calibration gas at a flow rate of 100 cc/min. until the signal out voltage stops changing (approx. 2 minutes). Then adjust span adj. potentiometer to the calculated value.
4. Remove the calibration gas and wait for the
voltage to drop. If the voltage is below 0.30 VDC the procedure is complete. If the voltage is above 0.30 VDC repeat the zero and span adjustments.
5. Calibrate the sensor using the instruction
manual procedure.
LOW SENSITIVITY message during
A
calibration indicates one of the two conditions for a combustible sensor:
The zero gas voltage must be less than
.3 volts.
The new span change (span voltage less
zero voltage) must exceed 30% of the calculated span change.
For Example:
Zero gas voltage = .20 VDC Span voltage = .43 VDC Span change is .43-.2 = .23 VDC As the calculated span value (from
step 5 above) is 1.03, the calculated span change is: 1.03-.2 = .87
By calculation: 30% of .87 = .26 The new span change (.23) is less than
30% of the calculated span change so a
LOW SENSITIVITY message will be
displayed.
LOW SENSITIVITY calculation is made to insure that
The the calibration does not adjust the electronics so that the "gain" is large enough to cause false alarms due to minor drift or due to minimal electronic noise.
PIN FUNCTION WIRE COLOR
1 DETECTOR WHITE 2 COMPENSATOR BROWN 3 CALIB. SW. RED 4 DETECTOR BLACK 5 DRAIN BARE 6 CALIB. SW. GREEN
Table 7-5
Combustible Gas Sensor
Wiring Pin Out
1. Detector element resistance should be between 1.5 to
2 ohms.
2. Compensator element resistance should be between
1.5 to 2 ohms.
3. Calibration switch should be open in normal
operation and should close when the calibration magnet is applied.
If the sensor fails any of these tests it should be replaced. If the sensor passes these tests the electronics are suspect and BOTH the sensor and electronics should be replaced. (The electronics and sensor which are removed should be returned to the factory for evaluation).
When a new sensor is installed th e follo wing action s must be taken:
Make the bridge voltage adjustment
immediately to avoid over-voltage damage to the sensor.
Use diagnostic code 0020 to establish nominal
value calibration.
Allow 24 hours for full stabilization of the
sensor, recheck the zero and span adjustments and calibrate the sensor module.
7.5.4 SENSOR REPLACEMENT
If any of the above adjustments cannot be completed successfully the sensor and/or electronics will require replacement. To check the sensor unplug it from the electronics and remove from the housing so that continuity and resistance tests can be made. Table 7-4 provides the pin out for the sensor assembly.
SERVICE
Page: 70
SMC sierra monitor corporation Sentry Instruction Manual - Version 6
7.5.5 COMBUSTIBLE GAS SCALING FACTORS
For combustible gas monitoring, a calibration standard of Methane or Propane my be used in conjunction with scaling factors to cause Sentry concentration display and alarm function in %LEL scale of another gas as follows:
GAS METHANE PROPANE
FACTOR FACTOR
Acetaldehyde 60 109 Acetic Acid 54 98 Acetic Anhydride 46 83 Acetone 52 94 Acetylene 57 103 Alkyl Alcohol 51 92 Ammonia 126 229 n-Amyl Alcohol 33 59 Aniline 39 71 Benzene 41 74 Biphenyl 25 45 1,3-Butadiene 56 101 n-Butane 58 106 iso-Butane 52 94 Butene-1 45 82 cis-Butene-2 48 88 trans-Butene-2 51 92 n-Butyl Alcohol 34 62 iso-Butyl Alcohol 53 96 tert-Butyl-Alcohol 74 134 n-Butyl Benzene 31 57 iso-Butyl Benzene 32 58 n-Butyric Acid 38 69 Carbon Disulfide 18 32 Carbon Monoxide 75 137 Carbon Oxysulphide 93 169 Chlorobenzene 34 62 Cyanogen 89 162 Cyclohexane 41 74 Cyclopropane 62 113 n-Decane 33 59 Diethylamine 49 88 Dimethylamine 58 105 2,3-Dimethylpentane 40 72 2,3-Dimethylpropane 40 72 Dimethylsulphide 43 79 1,4-Dioxane 45 81 Epichlorohydrin 45 82 Ethane 68 123 Ethyl Acetate 51 93 Ethyl Alcohol 73 132 Ethylamine 53 95 Ethyl Benzene 36 65 Ethyl Bromide 91 165 Ethyl Chloride 57 103 Ethylcyclopentane 40 72 Ethylene 71 128 Ethylenedichloride 66 120 Ethyleneoxide 52 94
NOTES:
Diethyl Ether 46 84 Dimethoxyethane 42 75 Dimethyl Ether 63 113 Dimethylformamide 46 83 Ethyl Formate 44 80 Ethylmercaptan 56 102 n-Heptane 39 70 n-Hexane 37 67 Hydrazine 45 82 Hydrogencyanide 48 86 Hydrogen 77 139 Hydrogen Sulfide 41 74 Methane 100 181 Methyl Actetate 50 90 Methyl Alcohol 86 156 Methylamine 77 140 Methyl Bromide 90 162 Methyl Chloride 102 186 Methylcyclohexane 44 80 Methylenedichloride 93 168 Methylethylether 44 80 Methylethylketone 41 75 Methyl Formate 67 121 Methylmercaptan 61 110 Methylpropionate 51 93 Methyl n-propylketone 40 73 Napthalene 34 62 Nitromethane 34 62 n-Nonane 31 57 n-Octane 37 68 n-Pentane 46 83 i-Pentane 46 84 Propane 55 100 n-Propyl Alcohol 47 85 n-Propylamine 48 88 n-Propylchloride 50 90 Propylene 52 93 Propyleneoxide 46 83 iso-Propylether 44 79 Propyne 42 75 Toluene 40 73 Triethylamine 40 72 Trimethylamine 48 88 Vinylethylether 42 76 o-Xylene 36 65 m-Xylene 39 71 p-Xylene 39 71 JP-4 (Jet Fuel) 41 73
GAS METHANE PROPANE
FACTOR FACTOR
1. Scaling factors are not FMRC approved.
2. Base data source: EEV sensor specification catalog. (EEV claims some data is the result of specific tests, other
data is empirically derived.)
SERVICE
Page: 71
SMC sierra monitor corporation Sentry Instruction Manual - Version 6
7.6 OXYGEN MODULE (5100-03)
7.6.1 DESCRIPTION
The Oxygen Module includes the electronic assembly installed in an explosion proof housing and the electrochemical sensor connected to one hub of the enclosure. Cabling from the controller connects to a three pin spring loaded terminal strip.
7.6.2 TROUBLE ANALYSIS
Electrical adjustment, or replacement of the sensor will be necessary under the following conditions:
Controller displays the following error
messages:
CHANGE SENSOR
SENSOR FAILURE
REPLACE SENSOR
False readings or alarms are received due to
sensor inaccuracy.
Warning: During sensor adjustments the concentration reading on the controller will be inaccurate and alarm level concentrations may be displayed. If false activation of the alarm relays will cause a problem disconnect the relay wiring prior to adjustment or turn the module off using the "Change Module" mode.
NOTE
Although all the necessary data can be collected with a
voltmeter at the sensor module, some helpful information
can be displayed on the controller or printed. See
diagnostic codes 0004 and 0008 in Appendix C.
7.6.3 ADJUSTMENT PROCEDURE
Prior to reading voltages and making adjustments perform a visual inspection to confirm that there are no physical problems such as water in the electronics enclosure, wiring damage or corrosion.
Use Figure 7-7 to locate test points during the following procedures.
7.6.3.1 SIGNAL ADJUSTMENT
It is unlikely that any electrical adjustment will be required except when a new sensor is installed.
To make the electrical adjustment connect a voltmeter to
SIGNAL OUT (pos) and GND (ground) and use SENSITIVITY ADJ potentiometer to set the voltage equal to
1/10 of the actual oxygen concentration. In clean air the concentration is 20.9% so the voltage should be set at
2.09 VDC.
Figure 7-7
Cover Plate - Oxygen Module
7.6.4 SENSOR REPLACEMENT
The oxygen sensor should be replaced when it can no longer be calibrated correctly or when the signal output drops to zero. Generally this is every twelve to eighteen months. The Sentry clock keeps track of the age of the sensor if it is correctly initialized in the "change module" mode.
When a new sensor is installed use the
CHANGE MODULE
menu selection to answer yes to the "new sensor?" question. After nine months the
CHANGE SENSOR
message will be displayed to warn that a new sensor should be installed.
SENSOR FAILURE message for Oxygen sensors does
The not indicate a specific failure of the sensor but indicates that the sensor is not correctly connected to the electronics. If this message appears check that the sensor harness is correctly installed to the connector on the bottom electronics board.
When sensor replacement is required, open the cover of the sensor module, remove the transmitter and disconnect the sensor wiring harness from the back of the transmitter.
SERVICE
Page: 72
Loading...
+ 10 hidden pages