Effective for all Sentry systems manufactured after September 1, 1995.
Instruction Manual Part Number T12001-A1
Sierra Monitor Corporation
1991 Tarob Court, Milpitas, CA 95035
(408) 262-6611
Applicability & Effectivity
SMC sierra monitor corporationSentry Instruction Manual - Version 6
7.5 COMBUSTIBLE GAS SENSOR MODULE
(5100-02)
7.5.1 DESCRIPTION
The Combustible Gas Module includes the sensor
and electronic assembly installed in an explosion
proof housing. The sensor screws into one hub of
the enclosure and plugs into the bottom
electronics card via a six pin connector. Cabling
to the controller connects to a three pin spring
loaded terminal strip.
7.5.2 TROUBLE ANALYSIS
Electrical adjustment, or replacement of the
sensor will be necessary under the following
conditions:
•Controller displays the following error
messages:
CHK BRIDGE VOLT
SENSOR FAILURE
LOW SENSITIVITY
•False readings or alarms are received
due to sensor inaccuracy.
Warning: : During sensor adjustments the
concentration reading on the controller will be
inaccurate and alarm level concentrations may be
displayed. If false activation of the alarm relays
will cause a problem disconnect the relay wiring
prior to adjustment or turn the module off using the
CHANGE MODULE mode.
NOTE
Although all the necessary data can be collected with a
voltmeter at the sensor module, some helpful information
can be displayed or printed. See diagnostic codes 0004
and 0008 in Appendix C.
7.5.3 ADJUSTMENT PROCEDURE
Prior to reading voltages and making adjustments perform
a visual inspection to confirm that there are no physical
problems such as water in the electronics enclosure,
wiring damage or corrosion.
Use Figure 7-6 to locate test points during the following
procedures:
Caution: The area must be free of flammable vapors or
gases during any adjustments or maintenance procedures.
7.5.3.1 BRIDGE VOLTAGE ADJUSTMENT
Confirm that no combustible gas is present and remove
sensor module cover.
Figure 7-6
Cover Plate - Combustible Gas Module
Comparing to the "GND" (ground) test point measure
"BRIDGE VOL" (bridge voltage) and adjust to 2.00 VDC
using "BRIDGE ADJ" potentiometer. Turn counterclockwise to increase.
7.5.3.2 ZERO ADJUSTMENT
Comparing to ground measure SIGNAL OUT and adjust to
0.18 VDC by turning
ZERO ADJ potentiometer. Turn
counter clock-wise to increase.
7.5.3.3 SPAN ADJUSTMENT
1. Determine the concentration of the calibration
gas (CG) in % LEL.
2. Use the following formula to determine the
response to the calibration gas at
SIGNAL OUT.
Voltage (Signal Out) = (CG+12)/60
For Example:
If CG = 50% L.E.L
Signal Out = (50+12)/60
= 1.03 volts
SERVICE
Page: 69
SMC sierra monitor corporationSentry Instruction Manual - Version 6
3. Using a Sierra Monitor Model 1200-26 Gas
Calibrator with a Sierra Monitor Model 535800 calibration adapter, or Model 5360 Gas
Delivery Fitting. Apply the calibration gas at a
flow rate of 100 cc/min. until the signal out
voltage stops changing (approx. 2 minutes).
Then adjust span adj. potentiometer to the
calculated value.
4. Remove the calibration gas and wait for the
voltage to drop. If the voltage is below 0.30
VDC the procedure is complete. If the voltage
is above 0.30 VDC repeat the zero and span
adjustments.
5. Calibrate the sensor using the instruction
manual procedure.
LOW SENSITIVITY message during
• A
calibration indicates one of the two
conditions for a combustible sensor:
• The zero gas voltage must be less than
.3 volts.
• The new span change (span voltage less
zero voltage) must exceed 30% of the
calculated span change.
For Example:
Zero gas voltage = .20 VDC
Span voltage = .43 VDC
Span change is .43-.2 = .23 VDC
As the calculated span value (from
step 5 above) is 1.03, the calculated
span change is: 1.03-.2 = .87
By calculation: 30% of .87 = .26
The new span change (.23) is less than
30% of the calculated span change so a
LOW SENSITIVITY message will be
displayed.
LOW SENSITIVITY calculation is made to insure that
The
the calibration does not adjust the electronics so that the
"gain" is large enough to cause false alarms due to minor
drift or due to minimal electronic noise.
PIN FUNCTION WIRE COLOR
1 DETECTOR WHITE
2 COMPENSATOR BROWN
3 CALIB. SW. RED
4 DETECTOR BLACK
5 DRAIN BARE
6 CALIB. SW. GREEN
Table 7-5
Combustible Gas Sensor
Wiring Pin Out
1. Detector element resistance should be between 1.5 to
2 ohms.
2. Compensator element resistance should be between
1.5 to 2 ohms.
3. Calibration switch should be open in normal
operation and should close when the calibration
magnet is applied.
If the sensor fails any of these tests it should be replaced.
If the sensor passes these tests the electronics are suspect
and BOTH the sensor and electronics should be replaced.
(The electronics and sensor which are removed should be
returned to the factory for evaluation).
When a new sensor is installed th e follo wing action s must
be taken:
•Make the bridge voltage adjustment
immediately to avoid over-voltage damage to
the sensor.
•Use diagnostic code 0020 to establish nominal
value calibration.
•Allow 24 hours for full stabilization of the
sensor, recheck the zero and span adjustments
and calibrate the sensor module.
7.5.4 SENSOR REPLACEMENT
If any of the above adjustments cannot be completed
successfully the sensor and/or electronics will require
replacement. To check the sensor unplug it from the
electronics and remove from the housing so that
continuity and resistance tests can be made. Table 7-4
provides the pin out for the sensor assembly.
SERVICE
Page: 70
SMC sierra monitor corporationSentry Instruction Manual - Version 6
7.5.5 COMBUSTIBLE GAS SCALING FACTORS
For combustible gas monitoring, a calibration standard of Methane or Propane my be used in conjunction with scaling
factors to cause Sentry concentration display and alarm function in %LEL scale of another gas as follows:
2. Base data source: EEV sensor specification catalog. (EEV claims some data is the result of specific tests, other
data is empirically derived.)
SERVICE
Page: 71
SMC sierra monitor corporationSentry Instruction Manual - Version 6
7.6 OXYGEN MODULE (5100-03)
7.6.1 DESCRIPTION
The Oxygen Module includes the electronic
assembly installed in an explosion proof housing
and the electrochemical sensor connected to one
hub of the enclosure. Cabling from the controller
connects to a three pin spring loaded terminal
strip.
7.6.2 TROUBLE ANALYSIS
Electrical adjustment, or replacement of the
sensor will be necessary under the following
conditions:
•Controller displays the following error
messages:
CHANGE SENSOR
SENSOR FAILURE
REPLACE SENSOR
•False readings or alarms are received due to
sensor inaccuracy.
Warning: During sensor adjustments the
concentration reading on the controller will be
inaccurate and alarm level concentrations may be
displayed. If false activation of the alarm relays
will cause a problem disconnect the relay wiring
prior to adjustment or turn the module off using
the "Change Module" mode.
NOTE
Although all the necessary data can be collected with a
voltmeter at the sensor module, some helpful information
can be displayed on the controller or printed. See
diagnostic codes 0004 and 0008 in Appendix C.
7.6.3 ADJUSTMENT PROCEDURE
Prior to reading voltages and making adjustments perform
a visual inspection to confirm that there are no physical
problems such as water in the electronics enclosure,
wiring damage or corrosion.
Use Figure 7-7 to locate test points during the following
procedures.
7.6.3.1 SIGNAL ADJUSTMENT
It is unlikely that any electrical adjustment will be
required except when a new sensor is installed.
To make the electrical adjustment connect a voltmeter to
SIGNAL OUT (pos) and GND (ground) and use
SENSITIVITY ADJ potentiometer to set the voltage equal to
1/10 of the actual oxygen concentration. In clean air the
concentration is 20.9% so the voltage should be set at
2.09 VDC.
Figure 7-7
Cover Plate - Oxygen Module
7.6.4 SENSOR REPLACEMENT
The oxygen sensor should be replaced when it can no
longer be calibrated correctly or when the signal output
drops to zero. Generally this is every twelve to eighteen
months. The Sentry clock keeps track of the age of the
sensor if it is correctly initialized in the "change module"
mode.
When a new sensor is installed use the
CHANGE MODULE
menu selection to answer yes to the "new sensor?"
question. After nine months the
CHANGE SENSOR
message will be displayed to warn that a new sensor
should be installed.
SENSOR FAILURE message for Oxygen sensors does
The
not indicate a specific failure of the sensor but indicates
that the sensor is not correctly connected to the
electronics. If this message appears check that the sensor
harness is correctly installed to the connector on the
bottom electronics board.
When sensor replacement is required, open the cover of
the sensor module, remove the transmitter and disconnect
the sensor wiring harness from the back of the transmitter.
SERVICE
Page: 72
Loading...
+ 10 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.