![](/html/89/89b4/89b429f6ebe9018938ba454e1c89ed22ad56785a25b54272f0c672953be310ff/bg1.png)
The Model 4501-XX Series T oxic Gas 2-wire gas
sensor modules allow easy direct connection to
many DCS, PLC and other control devices. The
sensor module, in a stainless steel enclosure, is
best in class for ease of installation, reliable
operation and ease of use.
The 4501-XX Gas Sensor Module utilizes the electrochemical cell to monitor for the specified gas over various gas specific ranges. This compact Module requires
minimal maintenance. Internal continuous self-diagnostics will automatically indicate any fault problems. The
4501-XX is best in class providing all the built-in features desired by instrumentation and safety professionals, including:
• Industry’s longest calibration interval of 180 days
• 2-wire loop powered for easy installation and
direct interface to common PLC and DCS
• Integral scrolling LCD 2-line, 16 character display
• Non-intrusive one-person calibration
• 316 Stainless S teel or aluminum enclosure
• 4-20 mA output
• Long sensor life
• Robust RFI immunity
• Ability to rotate face plate to enable the enclo-
sure to be mounted in any orientation.
Toxic Gas Sensor Module
2-Wire
Model 4501-XX
How It Works
Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode, and electrolyte. The components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical reaction and generate a current. The cells are diffusion
limited so the rate the gas enters the cell is solely dependent on the gas concentration. The current generated is proportional to the rate of consumption of the
subject gas in the cell.
Sierra Monitor electrochemical sensors provide improved
reliability by allowing the gas to diffuse into the sensor
through a capillary port, rather than diffusing through
membranes. The result is an extremely stable sensor
with very low temperature and pressure coefficients and
the capability to monitor gas as ppm (toxics).
4501-04 Carbon Monoxide, CO
4501-05 Hydrogen Sulfide, H2S
4501-06 Chlorine, Cl
4501-08 Chlorine Dioxide, ClO
4501-25 Ammonia, NH
Model 4501 Series Sensors
Sierra Monitor Corp. 1991 Tarob Court Milpitas, CA 95035 USA (408) 262-6611 (800) 727-4377 Fax: (408) 262-9042
Visit our website: www.sierramonitor .com E-mail: sales@sierramonitor .com
2
2
3
Explosion proof all models
FM Performance Approval for 4501-05
![](/html/89/89b4/89b429f6ebe9018938ba454e1c89ed22ad56785a25b54272f0c672953be310ff/bg2.png)
Specifications:
2-Wire Toxic Gas Sensor Module Model 4501-XX
Performance 4501-04
Max. Range (1) 0-1200 PPM
Resolution 0.5 PPM
Response Time (2) <25 sec.
Sensor Life 2 years
(1) Other ranges available
(2) To 90% range
(3) Ammonia sensor requires electrolyte recharge
up to 1 year
4501-05
0-100 PPM
0.1 PPM
<30 sec.
2 years
Output:
Display: Scrolling LCD, 2-line, 16-characters
Signal Output: 2-wire sensor powered analog 4-20 mA
Loop Resistance 800 ohm
(Trouble: 3.6 mA)
Electrical Data:
Power consumption: <160 milliwatts
Connection type: 2-wire, loop-powered
Input voltage: 24 VDC nominal: 14-30VDC
RFI/EMI Protection: EN50081-2, EN50082-2
Operating Range:
Ambient T emp Range:
Standard: See Table below
Extended T emp Sensor and 4-20 mA will continue to
Storage Temp: -4o to 122 oF(-20o to 50oC)
operate to -40oF (-40oC) but the LCD
display will decay over time if exposed
to temperatures below -4o (-20o C) for
significant time (not for 4501-25)
4501-06
0-10 PPM
0.1 PPM
<60 sec
2 years
4501-08
0-3 PPM
0.1 PPM
<60 sec
2 years
4501-25
0-100 PPM
1.0 PPM
<30 sec
(3)
Construction:
Dimensions: 7.0 x 3.6 x 4.0 inches (H x W x D)
4501-25 12,6 x 3,6 x 4.0 inches
Weight: 4.4 lb. (2.0 Kg)
4501-25 5.0 lb. (2.3 Kg)
Enclosure: Stainless S teel, NEMA 4X
Remote Sensor
Option: 4501-04 Max 25 feet
4501-05 Max 25 feet
4501-06 Max 15 feet
(17.8 x 9.1 x 11.2 cm)
Approvals and Warranty:
FM Approval: Explosion proof, Enclosure
4501-05 Approval FM Performance Approval Standards:
Limited warranty: 2 years
Class 1, Div. I, Group s B, C, D
Class 2, Groups B, C, D
Class 1, Zone 1, Group IIB+H2 IP66
4501-05 Approval FM ISA-92.0.01, Part I1998 and FM 3600
Specifications subject to change without notice
4501-04 CO
Ambient T emp (oF) -4 to 122 oF
Ambient T emp (oC) -20 to 50 oC
Relative Humidity 5 – 99%
4501-05 Hl2S
-40 to 122 oF
-40 to 50 oC
5 – 99%
4501-06 Cl
-4 to 122 oF
-20 to 50 oC
5 – 99%
2
Ordering Information
4501-04-AL Sensor Module, 2-Wire, Carbon Monoxide, CO, Aluminum Enclosure
4501-04-SS Sensor Module, 2-Wire, Carbon Monoxide, CO, 316 SS Enclosure
4501-05-AL Sensor Module, 2-Wire, Hydrogen Sulfide, H2S, Aluminum Enclosure
4501-05-SS Sensor Module, 2-Wire, Hydrogen Sulfide, H2S, 316 SS Enclosure
4501-06-AL Sensor Module, 2-Wire, Chlorine, Cl2, Aluminum Enclosure
4501-06-SS Sensor Module, 2-Wire, Chlorine, Cl2, 316 SS Enclosure
4501-25-AL Sensor Module, 2-wire, Ammonia, NH3, Aluminum Enclosure
4501-25-SS Sensor Module, 2-wire, Ammonia, NH3, SS Enclosure
4501-51 Remote sensor assembly, 4501 Series
Calibration Accessories
5358-01 Calibration Adapter, Standard
5360-00 Calibration Gas Delivery Fitting
5358-51 Calibration/Configuration Magnetic Tool,
4501-XX
Remote Sensor Distances
4501-01 25 feet
4501-05 25 feet
4501-06 15 feet
4501-08 15 feet
High Capacity Calibration Accessories
1250-01 Gas Sensor Calibration Kit, Type A (CO and H2S)
1250-03 Gas Sensor Calibration Kit, Type C (Chlorine)
1260-04 Gas Cylinder, Carbon Monoxide, 100 ppm, Type A
1260-05 Gas Cylinder, H2S in N2, 25 ppm Type A
1260-06 Gas Cylinder, Chlorine, 5 ppm, Type C
1260-13 Gas Cylinder, Carbon Monoxide, 1000 ppm, Type A
1260-45 Gas Cylinder, H2S in N2, 10 ppm, Type A
9210-00 Calibrator, Permeation tube
9211-09 Permeation Tube, Ammonia, NH3
4501-25 15 feet
4501-08 ClO
-4 to 122 oF
-20 to -50 oC
5-95%
2
4501-25 NH
-14 to 113 oF
-10 to 45 oC
20-95%
3
Sierra Monitor Corp. 1991 Tarob Court Milpitas, CA 95035 USA (408) 262-6611 (800) 727-4377 Fax: (408) 262-9042
Visit our website: www.sierramonitor .com E-mail: sales@sierramonitor .com
Rev. C 10/02/08