3.6 ZERO STABILIZATION ...................................................................................................................................5
5.2 DATA ENTRY KEYPAD ..................................................................................................................................8
5.3 MAIN MENU ....................................................................................................................................................9
6.2.2 SENSOR EXPOSURE TO GAS.............................................................................................................13
7. SERVICE .......................................................................................................................................................15
The 4501-XX is designed and approved for installation and
operation in hazardous locations.
Members of the 4501-XX Two-Wire Toxic Gas family include:
▪ 4501-04 Carbon Monoxide Gas Sensor Module
▪ 4501-05 Hydrogen Sulfide Gas Sensor Module
▪ 4501-06 Chlorine Gas Sensor Module
1.2 MOUNTING CONFIGURATION
Various module mounting configurations can be implemented without special fixtures. Where applicable,
these options are factory configured prior to shipment. Mounting configuration can be selected by the installer
or field technician and are fully described in this manual.
1.3 THEORY OF OPERATION
Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode, and electrolyte. The
components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical
reaction and generate a current. The cells are diffusion limited so the rate the gas enters the cell is solely
dependent on the gas concentration. The current generated is proportional to the rate of consumption of the
subject gas in the cell.
Sierra Monitor electrochemical sensors provide improved reliability by allowing the gas to diffuse into the
sensor through a capillary port, rather than diffusing through membranes. The result is an extremely stable
sensor with very low temperature and pressure coefficients and the capability to monitor gas as ppm.
Page: 1
Model 4501-XX Toxic Gas Sensor Module
1.4 MECHANICAL
The sensor module is comprised of the following three primary components:
1.4.1 ENCLOSURE
Explosion proof, rain-tight stainless steel electrical housing with three ¾” FNPT conduit hubs. The
enclosure cover has a viewing window. The 3 conduit hubs combined with the rotatable faceplate enable
the user to mount the module in a variety of configurations. See Figure 1-2 for details.
1.4.2 TRANSMITTER ELECTRONICS
Electronic Assembly consisting of one printed circuit board assembly mounted under a cover plate. Wiring
connections for power and signal interface are located on the back side of the display.
1.4.3 SENSOR ASSEMBLY
The sensor assembly includes a stainless steel, explosion proof, housing containing the gas sensor and a
wiring harness for connection to the transmitter. The sensor assembly threads into one hub of the
enclosure. The exposed end of the sensor assembly is threaded to allow connection of a rain-shield or
calibration gas delivery fitting.
1.5 INTERCONNECT WIRING
Not supplied with the sensor module, but necessary to the installation and operation is the two conductor
wiring which connects the module to its power source and controller. Before this wiring is installed it is
important to read and understand the control system installation instructions to determine wiring requirements
and alternatives.
1.6 POWER REQUIREMENTS
The 4501-XX module operates on DC power between 14 VDC and 30 VDC. The DC power must be supplied
by the loop controller.
Page: 2
Model 4501-XX Toxic Gas Sensor Module
Model 4501-05 Hydrogen Sulfide Gas Sensor Module– Dimensions
Figure 1-1
Figure 1-2
Model 4501-05 Hydrogen Sulfide Gas Sensor Module– Orientation Options
Page: 3
Model 4501-XX Toxic Gas Sensor Module
2. CAUTIONS & WARNINGS
2.1 INTRODUCTION
Although the 4501-XX Gas Sensor Module is designed and constructed for installation and operation in
industrial applications including hostile environments, caution should be taken to insure that the installation is
made in compliance with this instruction manual and that certain procedures and conditions are avoided. This
chapter discusses the necessary cautions. Read the entire chapter prior to installation of the equipment.
2.2 GAS SENSOR MODULES - GENERAL
Avoid installing sensor modules where they will be unnecessarily exposed to wind, dust, water (i.e. direct hose
down), shock, or vibration. Observe temperature range limitations.
Sensors may be adversely affected by prolonged exposure to certain materials. Loss of sensitivity, or
corrosion, may be gradual if such materials are present in low concentrations. These materials include:
Halides (compounds containing chlorine, fluorine, bromine, iodine), acid vapors, caustic liquids or mists.
Care has been taken by the manufacturer to ship modules in protective packaging to avoid contamination prior
to installation. It is recommended that the modules remain protected during installation and that the covering
be removed immediately prior to system start-up.
During normal use the sensor is protected from dirt and oil contamination by a sintered metal cover. If this
cover becomes clogged, the response of the sensor will be reduced. Protect the sensor from contamination
by careful placement, or by use of rain shields and dust shields.
Sensor modules must not be painted. Paint may contain compounds which will contaminate the sensor. Paint
will cause clogging of the sintered metal cover and will cause difficulties during attachment of the calibration
head or other maintenance activity. It is recommended that the module be tagged
2.3 WIRING
"DO NOT PAINT".
The manufacturer recommends that extra caution be taken where the installation is near any sources of
electromagnetic or radio frequency interference. Precautions include:
• Avoid running sensor module cable close to high power cables, radio transmission lines, or cables
subject to pulses of high current.
• Avoid running cables near large electric motors or generators.
• Analog devices require shielded cable.
• In conduit installations and cable applications the shield should be connected to instrument ground.
• All splices must be via either a termination hardware system or soldered. Improperly spliced cable can
result in corrosion, resistance changes and system errors.
Installation and wiring must be in accordance with the National Electrical Code.
Voltage AC conductors are not to be run in the same conduit as voltage DC
2.4 CALIBRATION FREQUENCY
NOTE
conductors.
The 4501-XX calibration frequency is six months (180 days). However, prudent gas detector maintenance
practices normally suggest a simple recalibration following incidences of exposure to high levels of H
would lead to alarm activation within any system utilizing the output of the 4501-05.
S that
2
Page: 4
Model 4501-XX Toxic Gas Sensor Module
3. QUICK START
3.1 OVERVIEW
The gas sensor module has been supplied factory calibrated and ready for immediate installation and
operation. An installer familiar with installation and operation of gas detection products can use this section to
begin immediate use of the module.
3.2 WIRING
Provide two conductor shielded wiring from the power supply/control device to the sensor module location.
Use wire that is 16 AWG or larger.
3.3 MODULE INSTALLATION
The 4501-XX can be mounted in a variety of configurations supported by the conduit. See figure 1-2 to
determine which configuration is best for your specific application. The default configuration enables the
modules to be put in line with other modules with the sensor element below the transmitter. To change the
configuration simply remove the transmitter and rotate to the appropriate configuration and remount the
standoffs and transmitter.
The module is designed to be installed on a ¾” conduit. Two important warnings:
• The installation must meet any hazardous environment codes for electrical equipment
• The sensor module enclosure mounting must be spaced far enough from any vertical surface
to allow removal and replacement of the sensor assembly which is threaded into one ¾”
conduit entry.
3.4 TRANSMITTER INSTALLATION
To install the transmitter printed circuit assembly into the housing, carefully turn the faceplate so that the
printing is in the correct horizontal position for the mounting configuration and slide the assembly over the two
stand-offs in the enclosure.
Hand tighten the two captive panel thumb screws into the stand-offs. Replace the enclosure cover prior to
providing power to the transmitter
•If the transmitter is installed in a classified hazardous area, replace the threaded cover prior to
providing power.
3.5 START-UP & OPERATION
To begin operation of the sensor module activate the instrument loop with 14-30 VDC. Each time the sensor
module is powered up it will perform a warm-up for approximately 2.5 minutes (148 seconds). During this time
the display will read “Warm. 148” and count down. The loop output will be held at 4 mA.
After the warm-up period has expired, the display will indicate the gas concentration. Also, the instrument loop
will be released to output current in the range of 4 to 20 mA. The actual current is linear with the gas
concentration and depends on the selected range. For instance, when the range of 0 to 50 PPM has been
selected, a current of 4 mA corresponds to 0 PPM and a current of 20 mA corresponds to 50 PPM.
3.6 ZERO STABILIZATION
All electrochemical sensors require at least 30 minutes on power prior to calibration. This allows the electrode
potentials to equilibrate, resulting in a stable zero signal level.
Page: 5
Model 4501-XX Toxic Gas Sensor Module
4. INSTALLATION
4.1 SENSOR MODULE LOCATIONS
The gas sensor module utilizes a diffusion type sensor which should be located close to either the expected
source or destination of the gas hazard. If the gas is heavier than air, the sensor module should be installed
within 24 inches of the ground or floor. If it is lighter than air, move it above 6’.
After optimum locations are determined based on the above recommendations, consideration should be given
to placing the sensor modules in locations which are accessible for calibration service. Slight adjustments to
the location of the sensor module may have little impact on effectiveness but major effect on accessibility.
• Modules should be placed in areas accessible for calibration.
• The cover should face out from the wall for easy access.
• Sensors should be pointed down and the conduit should include an inverse trap to reduce moisture
(condensation) from accumulating in the electronics enclosure.
4.2 WIRING
•Wire should be 16 AWG minimum at a maximum distance is 5,000 feet. Install conduit as required by
local code or construction specifications.
•See figure 4-1 for typical 2 terminal and 4 terminal connections.
Figure 4-1
4-20 mA Connection
4.3 ENCLOSURE INSTALLATION
To protect the transmitter and sensor assembly they should be removed from the enclosure and preserved
until final installation and wiring termination.
Prior to installation and wiring:
Page: 6
Model 4501-XX Toxic Gas Sensor Module
1. Remove the transmitter from the module housing by:
• Unscrew the two captive panel screws on the faceplate.
• Lift the transmitter out of the enclosure.
• Unplug the sensor cable from transmitter connector J1.
• Remove the sensor assembly from the enclosure hub.
2. Install the module enclosure onto the end of the supply conduit and/or bolt into position as required.
When enclosure earth grounding is required for the installation a grounding lug is located in
the base of the enclosure. Install the earth ground under the green ground screw.
4.4 TRANSMITTER AND SENSOR INSTALLATION
When all pre-wiring is complete:
1. Install sensor assembly in the open hub on the module enclosure. The sensor assembly thread
must be fully seated into the hub and tightened to maintain explosion proof assembly.
2. Connect the loop wires (Loop + and Loop -) to transmitter connector P1, pins 1 and 2 (Figure 4-2)
3. Connect the sensor assembly cable to transmitter connector J1. (Figure 4.2)
4. Carefully return the transmitter to the enclosure installing it over the two standoffs. Finger tighten
the retaining screws into the standoffs.
Sensor
Connector
NOTE
Loop Power
Figure 4-2
Connector Locations
Page: 7
Model 4501-XX Toxic Gas Sensor Module
5. OPERATION
5.1 INTRODUCTION
The 4501-XX Toxic Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A
magnet stick is supplied for this purpose. The menu system is used to calibrate the sensor module, and for
maintenance and diagnostic procedures.
The following adjustments can be made to the sensor module:
Range adjustments (see section 5.4)
4mA and 20mA calibration adjustments (see section 5.4)
Span adjustment (see section 6.3.1)
5.2 DATA ENTRY KEYPAD
The module menu system is operated by means of directing the magnet stick toward each of four independent
hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are
located under the faceplate above and below the LCD and are labeled
Figure 5-1.
M , E , S and T as shown in
• Key
: MODE
• Key ENTER
• Key : UP (+)
• Key : DOWN (-)
Figure 5-1
Face Plate with Operator Keypad
Page: 8
Model 4501-XX Toxic Gas Sensor Module
5.3 MAIN MENU
Key
FunctionDisplayDescriptionReference
M E
ST
M E
ST
ME
S T
ME
S T
M E
ST
M E
ST
ME
S T
ME
S T
Table 5-2 defines the key operational displays on the operator interface.
Banner: Use [S] or [T] to change 4mA output, [E] when done
A multi-meter should be inserted into the current loop
to measure the loop current between the power and
the sensor module No gas is required for this step
Acknowledges 4mA calibrationor Sub-Routine A
Enter Calibration 20 mA Menu
Banner: Use [S] or [T] to change 20mA output, [E] when
done
A multi-meter should be inserted into the current loop
to measure the loop current between the power and
the sensor module No gas is required for this step
Acknowledges 20mA calibrationor Sub-Routine A
Enter Range Menu
Enter Calibration Range Menu
Banner: Use [S] or [T] to change range, [E] when done
M E
ST
Enter
Range
DONE
CAL: XmA
ERROR!
Acknowledge range selection
Sub-Routine A
The expected range of the 4mA and 20mA is limited to factory
defined values. If the actual signal is outside of this
predefined range, an error is displayed indicating a hardware
or installation problem. Contact technical support.
Table 5-3
Maintenance Sub Menu
Page: 10
5.5 DIAGNOSTICS SUB-MENU
Key
FunctionDisplay
M E
ST
M E
ST
ME
S T
ME
S T
M E
ST
M E
ST
M E
ST
ME
S T
M E
ST
M E
ST
ME
S T
M E
ST
M E
ST
ME
S T
M E
ST
M E
ST
ME
S T
M E
ST
M E
ST
ME
S T
M E
ST
Mode
Enter
Down
Down
Enter
Enter
Mode
Down
Enter
Mode
Down
Enter
Mode
Down
Enter
Mode
Down
Enter
Mode
Down
Enter
H2S
XXX PPM
H2S
CALIB
H2S
MAINT:
H2S
DIAG:
DIAG:
RawSig
D-RawSig
XXXXmV
DIAG:
RawSig
DIAG:
Ref
D-Ref
XXXXmV
DIAG:
Ref
DIAG:
Temp
TEMP
XXXC
DIAG:
Temp
DIAG:
4-20 mA
D-4-20 mA
Normal
DIAG:
4-20mA
DIAG:
Version
D-Ver
VX.XX
DIAG:
Version
DIAG:
FactInit
Accept?
N NO Y
Default Display(example 4501-05)
Go to previous maintenance menu item
Enter diagnostics sub-menu
Voltage reading of the output of the sensor element
Return to diagnostic menu
Voltage reading of the output of the reference
Temperature of sensor use [S] or [T] to switch between Celsius and Fahrenheit
Enter 4-20 mA output sub-menuSub-Routine A
Version number of firmware
Banner: Use [S] or [T] to change, [E] to acceptSub-Routine B
D-4-20mA
Normal
ME
S T
ME
S T
ME
S T
ME
S T
ME
S T
M E
ST
M E
ST
Down
Down
Down
Down
Down
Enter
Enter
D-4-20mA
3.6 mA
D-4-20mA
4.0 mA
D-4-20mA
10-.0 mA
D-4-20mA
20.0 mA
Accept?
N YES Y
Accept?
N YES Y
FactInit
Done
Accept?
N NO Y
FactInit
Aborted
Press [E] to send 3.6 mA signal
Press [E] to send 4.0 mA signal
Press [E] to send 10.0 mA signal
Press [E] to send 20.0 mA signal
If accepting [T] to change to yes and [E] to accept
Factory initialization completed
If not accepting, [E] to accept NOSub-Routine C
Factory initialization not done
Description
Sub-Routine A
Sub-Routine B
Sub-Routine C
Table 5-4
Diagnostics Sub Menu
Model 4501-XX Toxic Gas Sensor Module
Reference
Sub-Routine C
Page: 11
Model 4501-XX Toxic Gas Sensor Module
5.5.1 NOTES
When selecting the output current requested, an acknowledge message will appear on the screen. Select
[E] to accept this message.
Whenever a message appears on the screen, select [E] to accept the message. No other action can be
taken with messages.
Except for the FactInit Menu, the sub-menus in the diagnostics section only provide information and do not
cause action.
Page: 12
Model 4501-XX Toxic Gas Sensor Module
6. CALIBRATION
6.1 CALIBRATION FREQUENCY
The sensor module must be calibrated every 180 days. Periodic functional tests are advisable for critical
applications and hostile environments.
The sensor module microprocessor software includes high-level self checking algorithms which provide
continuous sensor diagnostic and self adjustment.
6.2 CALIBRATION PREPARATION
Calibration of the sensor is accomplished by simple menu based steps and application of span gas.
If an error is made during any stage of the calibration process, hold the magnet stick at the
[M] for 10 seconds. A scrolling display will indicate “Operate Mode” and the sensor module
will exit the calibration activity and return to normal operating mode. The calibration
procedure can then be restarted.
NOTE
Calibration must be performed only when the area is known to be clear of Hydrogen Sulfide gas. When in
doubt, use a portable instrument to confirm that there is no background Hydrogen Sulfide gas.
For compliance with Factory Mutual (FM) Approvals, the Sierra Monitor Model 1250-01, 1260 -05 and 1260-45
are the FM Approved calibration gas delivery devices. Use the Model 5358-01 Calibration Adapter delivery
fitting.
6.2.1 CALIBRATION GAS DELIVERY METHODS
Calibration gas can be delivered to the sensors via the Model 5358-01:
Calibration Adapter (Figure 6-1) - used with portable calibrators.
CALIBRATION PROCEDURE
The Calibration Menu is described on Table 6-1:
The procedure requires that the menu “keys” be activated using the magnet stick.
Each key press steps through the process of setting the zero value for clean air
and then setting the span value.
Sample Gas In
At each of these steps, apply calibration gas of the value corresponding to
the setting accepted on the sensor module display.
6.2.2 SENSOR EXPOSURE TO GAS
Calibration gas must be delivered to the sensor using the flow rate and
duration listed in below:
Model Gas Flow Period
4501-04 Carbon Monoxide 300 cc/min Until Stable (minimum 3 minutes)
4501-05 Hydrogen Sulfide 300 cc/min Until Stable (minimum 3 minutes)
4501-06 Chlorine 300 cc/min Until Stable (minimum 3 minutes)
Page: 13
Figure 6-1
Model 5358-01
Calibration Adapter
Model 4501-XX Toxic Gas Sensor Module
6.3 DIAGNOSTICS SUB-MENU
FunctionDisplayDescriptionReference
Key
M E
ST
M E
ST
M E
ST
M E
ST
M E
ST
Mode
Enter
Enter
Enter
Mode
H2S
CALIB:
CAL 0 PPM
Cal0PPM
Wait.XXX
ZERO
DONE
XXPPM
H2S
WAIT-148
H2S
CAL-OK
CALIB
ABORTED
H2S
XXX PPM
Default Display (example 4501-05)
Banner: Apply zero gas, [E] when done
Operation: Confirm area clear of gas, or apply zero air to sensor.
There is a minimum time to calibrate at 0PPM. If [E] is hit prior to this time
has elapsed, a countdown screen will appear.
Zero gas setting completed (Message will appear on the screen and after 3
seconds will go to the next menu - 25PPM Span)
Banner: [S
Operation: 148 second time out before sensor is returned to service.
Banner: Remove gas from sensor
] or [T] to change span value, [E] when done
Sub-Routine A - Abort Calibration
Operation: Hold magnet over [M] for 10 seconds to abort calibration
Default Display
or Sub-Routine A
Sub-Routine B
Sub-Routine C
ME
S T
M E
ST
M E
ST
Down
Enter
Enter
25PPM SPAN
C##PPM
H2S
REMOVE GAS
H2S
CAL-OK
READ
MANUAL
H2S
CAL-FAIL
Sub-Routine B - Select Span
Banner: [S
Banner: Apply xx-PPM gas,
Operation: Remove gas from sensor
Calibration successful
] or [T] to change span value, [E] when done
[E] when stable
Sub-Routine C - Calibration Fail
Operation: Remove gas from sensor read manual, particularly section 6.3 and
7.6, to check calibration procedures.
Calibration Failed
Table 6-1
Calibration Sub-Menu
Should be stable after 3
minutes of gas exposure
If calibration failed, then
Sub-Routine C
Page: 14
Model 4501-XX Toxic Gas Sensor Module
7. SERVICE
7.1 SENSOR MODULE CONFIGURATION
The gas sensor module is comprised of the following sub-assemblies (Figure 7-1):
4501-XX Gas Sensor Module
SPL27077 Enclosure
SPM27071 Transmitter Assembly
XXXXXXX Sensor Assembly (see Appendix B)
XXXXXXX Sensor (see Appendix B)
There are no field serviceable components below the sub assembly level.
ENCLOSURE REPLACEMENT
The enclosure should be replaced if the cover threads or conduit threads have been damaged, or if the
enclosure has been damaged sufficiently that it no longer meets the required NEMA classification.
To replace the enclosure follow the transmitter and sensor assembly removal instructions, remove the
damaged enclosure from its conduit, install a new enclosure and continue the transmitter and sensor
assembly replacement instructions.
Figure 7-1
Module Components
Page: 15
Model 4501-XX Toxic Gas Sensor Module
TRANSMITTER REPLACEMENT
The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be
adjusted for calibration. This may occur due to age, corrosion or failed components.
To replace the transmitter assembly:
a. Remove the cover of the main enclosure
b. Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90
wiring service loop
c. Unplug the sensor connector from the transmitter
d. Remove the two wires from J1 terminals (see figure 4-2)
e. Reverse the preceding steps to install the new transmitter
f. Restore power and allow a minimum of 30 minutes for stabilization before re-calibration
SENSOR REPLACEMENT
The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of
the housing. The gas sensor needs replacement when:
It is no longer possible to obtain correct Zero and Span values
The “LOW SENS” message appears after calibration
The sensor output signal is noisy, causing erroneous gas level readings
To replace the sensor:
a. Confirm that system power has been removed
b. Remove the gas sensor module enclosure cover,
c. Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90
relieve the wiring service loop.
d. Unplug the sensor connector from the transmitter
e. Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its
harness
f. Unscrew sensor housing cover from the sensor
g. Carefully pull the old sensor straight up from the socket
h. Press the new sensor into the socket.
i. Reverse the preceding steps to install the sensor assembly.
j. Allow the new sensor to stabilize for a minimum of 30 minutes and then calibrate using the procedure in
Section 6.
o
to relieve the
o
to
INSTALLATION INSPECTION
Prior to system start-up or trouble shooting, the entire system should be visually inspected. The following are
guidelines for that inspection:
7.1.1 CONTROLLER INSTALLATION
• Controller installed in conformance to manufacturer’s instruction manual recommendations.
7.1.2 CABLING INSTALLATION
• All splices are soldered or via terminal block.
• Cabling is away from sources of electrical noise or RFI where possible.
7.1.3 SENSOR MODULE INSTALLATION
• Module installation in conformance with this manual.
• Modules accessible for calibration.
Page: 16
Model 4501-XX Toxic Gas Sensor Module
• Wiring terminations clean and correct.
7.1.4 MOISTURE TRAPS AND RAINSHIELDS
• Conduit seals and drains installed to avoid moisture build up in electronics enclosure. Water
accumulation in sensor module enclosures is a major cause of damage and system failures - take
precautions to seal electrical conduits and provide moisture traps and drains to avoid water damage
•Rain-shields installed where applicable.
7.1.5 STANDARD VOLTAGES
• Loop voltage to be applied to the sensor module must be between 14 VDC and 30 VDC.
7.2 INSPECTION AND TROUBLESHOOTING GUIDE
The inspection and troubleshooting guide can be used to step through the system start-up and to determine
the corrective action if a fault occurs.
7.2.1 IF MODULE DOES NOT DISPLAY THE CORRECT PPM GAS
1. Repeat calibration procedure.
2. Remove the gas and wait for the timer to completely count down.
3. Apply Span Gas and verify that the sensor sees the correct value of the span gas after calibration.
4. If the sensor still does not respond to gas, power cycle the unit and repeat calibration.
7.2.2 IF THE MODULE DOES NOT DISPLAY THE CORRECT PPM
1. Power cycle the module.
2. Recalibrate the sensor.
7.2.3 IF THE DISPLAY SHOWS ‘NO SENSOR’ – SENSOR FAILING
1. Power down the unit
2. Open the enclosure and unplug the sensor from the transmitter board.
3. Plug the sensor back into the transmitter board carefully and ensure a secure fit.
4. Power up the unit.
7.2.4 IF THE DISPLAY SHOWS “C” – CALIBRATION MODE
1. Complete calibration or exit to operating mode.
7.2.5 IF THE DISPLAY SHOWS “ T ” – DOWNSCALE INDICATOR
1. The reaction of the sensor to gas may, for a short period of time, result in a downscale indicator being
shown on the display.
2. If this indicator remains for a longer period of time, the sensor may need re-calibration.
7.2.6 IF THE DISPLAY SHOWS “HIGH” – UPSCALE INDICATOR
1. The concentration of the gas is outside the calibrated range.
Page: 17
Model 4501-XX Toxic Gas Sensor Module
8. APPENDICES
APPENDIX A: SPECIFICATIONS
Specifications:
Sensor:
(1) Optional ranges available
(2) Response time to 90% full signal value for applied concentration
(3) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest
Output:
Display: LCD, 2-line, 20-characters
Signal Output: 2-wire sensor powered analog 4-20 mA (Trouble: 3.6 mA)
Loop Resistance 800 ohm
Extended Range Sensor and 4-20 mA will continue to operate to -40oF (-40oC) but the
Operation LCD display will decay over time if exposed to temperatures below -4o (-
Storage Temperature: -4o to 122 oF(-20o to 50oC)
Enclosure:
Dimensions: 7.0 x 3.6 x 4.0 inches (H x W x D)
(17.8 x 9.1 x 11.2 cm)
Weight: 4.4 lb. (2.0 Kg)
Material of Construction: 316 Stainless Steel
Enclosure: NEMA 4X
Electrical Classification:
FM Approved: Explosion proof, Class 1, Div. I, Groups B, C, D
Class 1, Zone 1, Group IIB+H2 IP66
Approvals:
FM Performance Approval: Approval Standards: ISA-92.0.01, Part I-1998 and FM 3600
(4501-05 only)
Remote Sensor Option: Distance Sensor to Transmitter: 4501-04 25 feet
4501-05 25 feet
4501-06 15 feet
Max. Range (1) 0-1200 PPM 0-100 PPM 0-10 PPM
Zero Drift +/- 1 PPM +/- 0.5 PPM +/- 0.5 PPM
Repeatability +/- 1 PPM +/- 1 PPM +/- 0.5 PPM
Linearity +/- 1 PPM +/- 1 PPM +/- 0.5 PPM
Resolution 0.5 PPM 0.1 PPM 0.1 PPM
Response Time (2) <25 sec. <30 sec. <60 sec
Sensor Life (3) 2 years 2 years 2 years
Ambient Temp (oF) -4 to 122 oF -40 to 122 oF -4 to 122 oF
Ambient Temp (oC) -20 to 50 oC -40 to 50 oC -20 to 50 oC
Relative Humidity 5 – 99% 5 – 99% 5 – 99%
4501-04 4501-05 4501-06
4501-04 CO 4501-05 H2S 4501-06 Cl2
20o C) for significant time
Page: 18
Model 4501-XX Toxic Gas Sensor Module
Warranty:
Limited warranty: 2 year specifications subject to change without notice
Model Number Gas Type CO H2S CI2 H2 SO2 NO2 NO HCI HCN C2H4
SPM33041 Sensor for 4501-05
SPM33050 Sensor for 4501-06
Page: 20
Model 4501-XX Toxic Gas Sensor Module
APPENDIX C: LIMITED WARRANTY
SIERRA MONITOR CORPORATION warrants its products to be free from defects in workmanship or material
under normal use and service for two years after date of shipment. SMC will repair or replace without charge
any equipment found to be defective during the warranty period. Final determination of the nature and
responsibility for defective or damaged equipment will be made by SMC personnel.
All warranties hereunder are contingent upon proper use in the application for which the product was intended
and do not cover products which have been modified or repaired without SMC approval or which have been
subjected to accident, improper maintenance, installation or application, or on which original identification
marks have been removed or altered. This Limited Warranty also will not apply to interconnecting cables or
wires, consumables (i.e. calibration gases, batteries, sensors), nor to any damage resulting from battery
leakage.
In all cases SMC’s responsibility and liability under this warranty shall be limited to the cost of the equipment.
The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty
provision and compliance with such instruction shall be a condition of this warranty.
Except for the express warranty stated above, SMC disclaims all warranties with regard to the products sold
hereunder including all implied warranties of merchantability and fitness and the express warranties stated
herein are in lieu of all obligations or liabilities on the part of SMC for damages including, but not limited to,
consequential damages arising out of/or in connection with the use or performance of the product.
Page: 21
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.