Sierra Monitor Corporation 4400 User Manual

In-Stream Hydrogen Sulfide Analyzer
The Model 4400 In-Stream Hydrogen Sulfide Analyzer is used to continuously verify the H2S content of a natu­ral gas sample for the presence of 0-100 ppm levels of Hydrogen Sulfide. This system is used to insure that H2S levels in natural gas collection lines remains at ac­ceptable levels to help guard against the corrosive ef­fects of H2S on compressors, pipelines and process equipment. It is intended for continuous monitoring prior to the addition of odorants to the gas.
The 4400-10 receives a low pressure natural gas slip­stream sample. A flow controller in the system delivers a constant flow to the gas sensor and allows for easy calibration and maintenance. An in-line coalescing fil­ter protects the system from sample contaminants.
A 4-20 mA linear signal proportional to sample concen­tration is provided from the sensor assembly. This sig­nal can interface with a Sierra Monitor Model 4107-99 Transmitter/Display, a single channel controller, the Sentry multi-channel controller or most industry stan­dard controllers or PLCs.
Model 4400
The Sierra Monitor Model 4400 provides the user with a low-cost, easy-to-use solution to effective monitoring of natural gas streams. This system accurately mea­sures H2S concentration in the 0-100 ppm range. Fast response of 90% of step change within 30 seconds enables the user to respond quickly to changes in natural gas supply conditions to prevent product contamina­tion.
The Model 4107-99 Transmitter/Display option can ac­cept the 4-20 mA signal from the Model 4400-10 and provide digital display of the H2S concentration, dual concentration alarm contacts, 4-20 mA output and se­rial MODBUS communication. The Model 4107-99 can be mounted directly to the Model 4400 H2S Analyzer of at a remote location. The Model 4107-99 is in an ex­plosion proof enclosure.
There is minimal maintenance with the Model 4400 with no tapes to replace, no conditioning solutions, no me­chanical stream blenders to calibrate. Calibration is easy with external application of span gas and simple one person adjustments. Calibration and swapping of the gas sensor is a quick and simple operation per­formed every two weeks.
Features Benefits
n No tapes to replace or conditioning solutions Minimal maintenance, no hazardous material to
dispose
n Fast response Prevention of product contamination n Simple one-person adjustments Easy calibration, minimal service time
Sierra Monitor Corporation 1991 Tarob Court Milpitas, CA 95035 USA (408) 262-6611 (800) 727-4377 Fax: (408) 262-9042
Specifications:
Model 4400
In-Stream Hydrogen Sulfide Analyzer
4400 In-Stream Analyzer
Power: 24VDC Output: 4-20 mA = 0-100 ppm H
Environmental:
Temperature Range
Operating:: -20 to 50 Storage: -40 to 50oC (-40 to 122oF)
Humidity: 0-95% RH (non-condensing)
Sensor:
Range: 0-100 ppm Hydrogen Sulfide Type: Electrochemical Life: Approximately 1 year in cycled
service
Response: <30 seconds to 90% FS
Mechanical:
Area Classification: Div 1, Class 1, Groups B,C,D Enclosure: NEMA-1 with removable cover panel Mounting: installs on 9” x 7” “H” Frame Connections: 3 position terminal strip with
separate conduit
Dimensions (WxHxD): 9.0 x 11.0 x 4.0 inches
(22.9 x 27.9 x 10.2 cm)
Weight: 5 lbs (2.2 Kg)
Warranty: 2 years as non non-consumable items
o
C (-4 to 122oF)
S
2
4107-99 Transmitter/Display
Power: 24VDC
Operating Temperature Range: 14 to 140
Display: 2-line, 32 character alphanumeric
backlit LCD
Alarms: Dual level alarms plus fault alarm Alarm Contacts: Dry contacts from each alarm rated to
switch at least 1 amp @ 30 volts or
0.5 amp @ 125 volts for resistive loads
Serial Output: Bi-directional RS-232 and RS-485
MODBUS
Mechanical:
Housing: Explosion proof (NEMA 7/NEMA 4X)
(Div. 1, Class 1, Groups B,C,D
Class 2, Groups E,F,G) Mounting: 3/4" NPT Dimensions( WxHxD): 5.3 x 5.7 x 5.7 inches
(13.5 x 14.5 x 14.5 cm) Weight: 6.0 lbs (3.2 Kg)
o
C (-10 to 60oF)
protocol
Ordering Information:
4400-10 In-Stream Hydrogen Sulfide Monitor 4107-99 Single Point Explosion Proof Transmitter/Display SPM27064 Maintenance Kit (consists of 1 spare sensor assembly & 2 flow harnesses)
Model 4400-10 with 4107-99 Transmitter/Display
Sierra Monitor Corporation 1991 Tarob Ct. Milpitas, CA 95035 (408) 262-6611 800-727-4377 FAX: (408) 262-9042
Visit our Web site at: www.sierramonitor.com E-Mail: sierra@sierramonitor.com
© 1997 by Sierra Monitor Corporation
Loading...