Models: 820, 820S, 822, 822S, 824, 824S, 826 and 827
Instruction Manual
Part Number IM-82
Revision H, November 2017
Table of Contents Series 820 Instruction Manual
GLOBAL SUPPORT LOCATIONS: WE ARE HERE TO HELP!
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940
Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
www.sierrainstruments.com
EUROPE HEADQUARTERS
Bijlmansweid 2 1934RE Egmond aan den Hoef The Netherlands
Phone +31 72 5071400 Fax +31 72 5071401
ASIA HEADQUARTERS
Second Floor Building 5, Senpu Industrial Park
25 Hangdu Road Hangtou Town
Pu Dong New District, Shanghai, P.R. China
Postal Code 201316
Phone: + 8621 5879 8521 Fax: +8621 5879 8586
IMPORTANT CUSTOMER NOTICE: OXYGEN SERVICE
Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from
the use of Sierra Instruments standard ma ss flow meters or controllers for oxygen gas. You are responsible for determining if this mass flow meter or controller is appropriate for your oxygen application. You are responsible for cleaning the mass flow meter or controller to the degree required
for your oxygen flow application
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any me ans,
electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change
without notice.
TRADEMARKS
Top-TrakTM and Cal-Bench TM is a Registered Trademark of Sierra Instruments, Inc. Other product
and company names listed in this manual are tradem arks or trade names of their respective manufacturers.
Returning Equipment to the Factory ....................................... 4-13
3-1
IM-82 0-3
Table of Contents Series 820 Instruction Manual
Appendix A Conversion Formulas and Gas Tables
Appendix B 822/824 Product Specifications
Appendix C 822S/824S Product Specifications
Appendix D 826/827 Product Specifications
List of Figures
1-1. Top-Trak Features (Typical) ........................................ 1-3
1-2. Flow Paths through the Transducer ............................. 1-4
Only qualified personnel should install the transducer.
Do not supply +DC power at the D-connector while using a
power supply at the power jack. Both supplies may be damaged.
Caution!
Operating a 12 VDC transducer at 24 VDC will cause equip-
ment damage.
Caution!
Only qualified personnel should perform transducer service, ca-
libration or troubleshooting procedures.
Caution!
When using toxic or corrosive gases, purge the unit thoroughly
with inert dry gas before disconnecting from the gas line.
Caution!
Printed circuit boards are sensitive to electrostatic discharge. To
avoid damaging the board, follow these precautions to minimize the risk
of damage:
before handling the assembly, discharge your body by touching
a grounded, metal object
handle all cards by their edges unless otherwise required
when possible, use grounded electrostatic discharge wrist
straps when handling sensitive components
IM-82 0-5
Series 820 Instruction Manual Chapter 1 Introduction
Chapter 1 Introduction
This instruction manual covers the installation, operation and
maintenance of Sierra’s 820 Series product line including the following Top-Trakª Models:
822 Mass Flow Meter with display (nylon flow body)
824 Mass Flow Meter without display (nylon flow body)
826 Hi-Flow Meter with display (aluminum flow body)
827 Hi-Flow without display (aluminum flow body)
822-S Mass Flow Meter with display (stainless steel flow body)
824-S Mass Flow Meter without display (stainless steel flow body)
Sierra’s Top-Trakª Mass Flow Meters are designed for precise
measurement of gas mass flow. The 820 Series offers a broad
range of sizes and process connections for flexibility and versatility. The primary standard calibration ensures starting point accuracy and NIST traceability. The meter’s 0-5 VDC or 4-20 mA output
signal is provided for recording, data-logging or control. The optional display reads the mass flow rate directly in engineering units
or percentage of full scale.
Using This Manual
This manual is organized into four chapters:
Chapter 1 includes the introduction and theory of operation
Chapter 2 provides installation and wiring instructions
Chapter 3 describes transducer operation and features
Chapter 4 covers maintenance, calibration and troubleshooting
IM-82 1-1
Gas tables and conversion formulas are found in Appendix A.
The product specifications and dimensional drawings are found
in Appendix B.
Throughout this manual, we use the word
transducer
as a generic term to represent all models of Sierra’s 820 Series TopTrak Mass Flow Meters.
Chapter 1 Introduction Series 820 Instruction Manual
Safety Information
Caution and warning statements are used throughout this book to
draw your attention to important information.
Warning! Caution!
This statement appears with information that
is important to protect people and equipment
from damage. Pay very close attention to all
warnings that apply to your application.
Receipt of System Components
When receiving a Sierra transducer, carefully check the outside
packing carton for damage incurred in shipment. If the carton is
damaged, notify the local carrier and submit a report to the factory
or distributor. Remove the packing slip and check that all ordered
components are present and match your specifications (as ordered).
Make sure any spare parts or accessories are not discarded with the
packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service.
This statement appears with information that is
important for protecting your equipme nt and
performance. Read and follow all cautions that
apply to your application.
Technical Assistance
If you encounter a problem with your transducer, review the configuration information for each step of the installation, operation and
set up procedures. Verify that your settings and adjustments are
consistent with factory recommendations. Refer to Chapter 4,
Troubleshooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 4, contact Sierra Instruments by fax or by
E-mail (see inside front cover). For urgent phone support you may
call (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00
p.m. PST. In Europe contact Sierra Instruments bv at +31 20
6145810. When contacting Technical Support, make sure to include
this information:
the flow range, serial number, Sierra order number and
model number (all marked on the transducer nameplate)
the problem you are encountering and any corrective action
taken
application information (gas, pressure, temperature, pipe
and fitting configuration)
1-2 IM-82
Series 820 Instruction Manual Chapter 1 Introduction
Top-Trak Features
Standard Top-Trak Mass Flow Meters require a 12 to 15 VDC external power source (24 VDC input power optional). The transducer’s 0 to 5 VDC output signal allows for flow recording, data-logging
or control. A 4 to 20 mA output signal is optionally available. Input
power and output signal connections are made via the 9-pin subtype D-connector located on the side of the transducer. An additional input power jack is located just below the D-connector. (It is
important to connect input power at only one location.)
The transducer shown below is a typical example of a 822 Series
Top-Trak Mass Flow Meter. Other models may vary slightly in
their appearance but are operationally equivalent.
IM-82 1-3
Figure 1-1. Top-Trak Features (Typical)
Chapter 1 Introduction Series 820 Instruction Manual
The 820 Series Flow Sensing Principle
The operating principle of Top-Trak transducers is based on heat
transfer and the first law of thermodynamics. During operation
process gas enters the instrument’s flow body and divides into two
flow paths, one through the sensor tube, the other through the laminar flow element bypass. The laminar flow bypass generates a
pressure drop, P1–P2, forcing a small fraction of the total flow to
pass through the sensor tube (m1).
Figure 1-2. Flow Paths through the Transducer
Two resistance temperature detector (RTD) coils around the sensor
tube direct a constant amount of heat (H) into the gas stream. In
actual operation, the gas mass flow carries heat from the upstream
coil to the downstream coil. The resulting temperature difference
(?T) is detected by the RTD coils and gives the output signal.
Since the molecules of the gas carry away the heat, the output signal is linearly proportional to gas mass flow.
Figure 1-3. Flow Measuring Principle
1-4 IM-82
Series 820 Instruction Manual Chapter 2 Installation
Chapter 2 Installation
Installation Overview
To ensure a successful installation, inlet and outlet tubing should be clean
and free from burrs or rims caused by cutting prior to plumbing the transducer into the system. The protective caps covering the inlet/outlet fittings
should not be removed until immediately prior to installation.
Before installing the transducer, verify the following:
1. Make sure the installation site meets the specific operating parameters
recorded on the transducer’s nameplate. Each transducer is factoryconfigured for a specific gas and flow range. If the operating pressure is
more than 50 psi (3.4 bar) away from the calibration pressure, it is advisable to return the unit to the factory for re-calibration. (Adjusting zero
may be sufficient to remain within specification.)
2. Do not locate the transducer in areas subject to sudden temperature
changes, moisture, drafts or near equipment radiating significant amounts
of heat. Make sure to allow adequate space for cable connectors and wiring.
3. For 1/2-inch size inlet/outlet process connections on models 826 and 827
make sure the location meets the minimum number of recommended pipe
diameters upstream and downstream of the transducer. A minimum of 5
inches (127 mm) upstream and 2-1/2 inches (64 mm) downstream is always recommended. (not necessary for other models)
4. Horizontal mounting is preferable. Vertical mounting is possible with
best results achieved when the factory calibration is specifically performed for vertical mounting. In vertical positions zero shift will occur
depending on the gas pressure at zero flow.
5. If the gas contains any particulate matter, install an in-line filter prior to
the transducer. Recommended filter size: 15 micron for flows of 10
sccm to 30 slpm, 30 micron for above 30 slpm.
6. If a potential over-flow condition exists, insert a valve or critical orifice
in the line to limit flow to approximately 25 percent above the full scale
range of the meter.
7. Confirm that the transducer o-ring material is compatible with the gas to
be measured.
8. For remote displays, confirm the supplied cable is of sufficient length to
connect the components.
IM-82 2-1
Chapter 2 Installation Series 820 Instruction Manual
Installing the Transducer
Follow the installation instructions that apply to your transducer’s
Caution!
Only qualified personnel
should install the
transducer.
process connection. For all 1/2-inch size process connections, observe
the piping recommendations given on page 2-3. Before operation, all
plumbing should be checked carefully for leaks and the transducer
purged with dry nitrogen.
Compression Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
2. Verify the position of the front
and back ferrule. Insert the
tubing into the fitting. Make
sure that the tubing rests firmly
on the shoulder of the fitting
and that the nut is finger tight.
(Do not mix or interchange
parts of tube fittings made by different manufactur- ers.)
3. Hold the body steady with a backup wrench.
tighten the nut 1-1/4 turns from finger tight. For 1/8-inch, 1/4- inch
and 3∕8-inch sizes, tighten only 3/4 turn from finger tight.
over-tighten!
4. Check the system’s entire flow path thoroughly for leaks. (Do not
use liquid leak detectors, instead monitor pressure decay. Overexposing the transducer to leak detector fluid may damage the unit.)
For 1/2-inch size,
Do not
VCO Fittings
1. Position the transducer with the flow direction arrow pointing
2. Install new o-rings compatible with the gas to be used. (Do not mix
3. Hold the body steady with a backup wrench.
4. Check the system’s entire flow path thoroughly for leaks. (Do not use
2-2 IM-82
downstream in the direction of flow.
or interchange parts of tube fittings made by different manufacturers.)
Tighten the nut
finger tight and then 1/4 turn tighter with a wrench.
tighten!
Do not over-
liquid leak detectors, instead monitor pressure decay. Over- exposing
the transducer to leak detector fluid may damage the unit.)
Series 820 Instruction Manual Chapter 2 Installation
1/4 inch Female NPT Connections
1. Position the transducer with the flow direction arrow pointing in the direction of flow.
Use a good quality paste pipe thread sealant. Apply to pipe threads.
2.
3. Tighten the pipe no more than 1 turn past hand-tight. Caution! Do not
over-tighten. Damage to the instrument may result.
1/2-Inch Size NPT Connections
1. Install a section of straight pipe at least ten pipe diameters in length up-
stream of the transducer. Also, allow at least five pipe diameters downstream for accurate operation. DO NOT use reducers. If the preceeding
components in the flow path create disturbances extend the upstream pipe
length.
2. Position the transducer with the flow direction arrow pointing down-
stream in the direction of flow.
3. Tighten fittings until leak tight (refer to published standards for specific
recommendations).
(standard on nylon flow bodies
)
4. Check the system’s entire flow path thoroughly for leaks. (Do not use liq-
uid leak detectors, instead monitor pressure decay. Over-exposing the
transducer to leak detector fluid may damage the unit.)
Figure 2-1 Piping Requirements for all 1/2-Inch Size Process Connections
IM-82 2-3
Chapter 2 Installation Series 820 Instruction Manual
r
e
t
Caution!
Do not supply +DC powe
at the D-connector whil
using a power supply a
the power jack. Both sup-
plies may be damaged.
Caution!
Operating a 12 VDC trans-
ducer at 24 VDC will cause
equipment damage.
Wiring the Transducer
Standard Top-Trakrm transducers require a 12 to 15 VDC power supply
(15 VDC nominal, 100 mA maximum). 24 VDC input power is
optional. Transducers are connected to the power supply through
either
the dedicated DC power jack or through the 9-pin D-connector
located on the side of the enclosure. Before powering the unit, check
the transducer’s nameplate to confirm input power:
PV1 = 12 to 15 VDC
PV2 = 24 VDC
Note: operating a 24 VDC transducer at 12 to 15 VDC will result in unreliable
operation.
The transducer’s standard 0 to 5 VDC (4-20 mA optional) output sig- nal is
available through the D-connector. The mating connector is in- cluded with the
transducer. Connection details are given on the fol- lowing pages.
When the transducer is configured for a remote display, signal connections are made via the 9-pin connector. Input power connections
are not included in the standard interface cable. Power connections
for the display and the transducer are shared when the connector is
wired for a remote display. Optionally the input power can be applied
at the power jack (either 12 volts (PV 1) or 24 volts (PV2). Remote
display mounting dimensions are given at the end of this chapter.
2-4 IM-82
Series 820 Instruction Manual Chapter 2 Installation
Note: The power jack requires a plug that has a 2.5mm inner diameter and a 5mm outer
diameter. The center is positive (+).
IM-82 2-5
Chapter 2 Installation Series 820 Instruction Manual
Standard 0-5 VDC Output Signal Wiring
The standard 0-5 VDC output signal flows from Pin 3 (0-5 VDC Out)
through the load (1K Ohm minimum) to Pin 7 (Power Common). The
figure below is a typical example of input power and output signal
connections.
Figure 2-3. Standard 0-5 VDC Output Signal Wiring
Optional 4-20 mA Output Signal Wiring
The optional 4-20 mA output signal flows from Pin 9 (4-20 mA Out)
through the load (50 to 500 Ohms maximum) to Pin 7 (Power Common). The figure below is a typical example of input power and output
signal connections. (Multiple transducer current loop output connections are given on the next page.)
Figure 2-4. Single Transducer Current Loop Connection
2-6 IM-82-E
s
Series 820 Instruction Manual Chapter 2 Installation
Figure 2-5. Multiple Transducer Current Loop Connection
Remote Display Installation
Mount the remote display at a convenient location within reach of the
supplied interface cable. The maximum cable length is 100 feet (30 m).
IM-82 2-7
Figure 2-6. Mounting the Remote Display
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.