Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instruments standard mass flow meters or
controllers for oxygen gas. You are responsible for determining if this mass flow
meter or controller is appropriate for your oxygen application. You are responsible
for cleaning the mass flow meter or controller to the degree required for your oxygen flow application.
TRADEMARKS
Side-Trak™, Auto-Trak™ and Cal-Bench™ are trademarks of Sierra Instruments, Inc.
Other product and company names listed in this manual are trademarks or trade names of
their respective manufacturers.
Series 830/840/860 Instruction Manual Chapter 1 Introduction
IM-83/84/86-H 1-1
Chapter 1 Introduction
Sierra’s Side-Trak™ and Auto-Trak™ mass flow meters and controllers are designed to accurately measure and control flows of
process gases. This instruction manual covers the installation, operation and maintenance of the entire Side-Trak and Auto-Trak
product lines, which include the following Sierra models:
• 830 Side-Trak Mass Flow Meter
• 840 Side-Trak Mass Flow Controller
• 860M Auto-Trak Mass Flow Meter with RS-485 interface
• 860C Auto-Trak Mass Flow Controller with RS-485 interface
Sierra’s mass flow meters and controllers offer a broad range of sizes,
control electronics, process connections, cables and connectors for
flexibility and versatility. The meter’s patented straight sensor tube
with access ports permits easy cleaning and reduces maintenance
down-time. The platinum sensor eliminates zero-drift and ensures
long-term repeatability. The primary standard calibration ensures
starting point accuracy and NIST traceability. All wetted surfaces are
constructed of 316 stainless steel with Viton® seals standard. Controllers incorporate either an electromagnetic valve (up to 500 slpm) or a
motor-operated rotary valve (100 slpm to 200 scfm). The Auto-Trak
860 digital microprocessor features automatic bridge-balancing circuitry and provides simultaneous analog and digital RS-485 communication.
Using This Manual
This manual is organized into five chapters:
• Chapter 1 includes the introduction and theory of operation
• Chapter 2 provides installation and wiring instructions
• Chapter 3 describes system operation and controller features
• Chapter 4 covers maintenance and calibration procedures
• Chapter 5 provides troubleshooting advice
Gas tables and conversion formulas are found in Appendix A. The
product specifications and dimensional drawings are found in Appendix B.
The Model 860 Auto-Trak RS-485 communications program includes an “Auto-Net” Software Addendum as a supplement to this
manual. Please refer to this addendum for all RS-485 operating instructions.
Series 830/840/860 Instruction Manual Chapter 1 Introduction
IM-83/84/86-H 1-3
Safety Information
Caution and warning statements are used throughout this book to
draw your attention to important information.
Warning!
Caution!
This statement appears with information that
is important to protect people and equipment
from damage. Pay very close attention to all
warnings that apply to your application.
This statement appears with information that is
important for protecting your equipment and
performance. Read and follow all cautions that
apply to your application.
Receipt of System Components
When receiving a Sierra transducer, carefully check the outside
packing carton for damage incurred in shipment. If the carton is
damaged, notify the local carrier and submit a report to the factory
or distributor. Remove the packing slip and check that all ordered
components are present and match your specifications (as ordered).
Make sure any spare parts or accessories are not discarded with the
packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service.
Technical Assistance
If you encounter a problem with your transducer, review the
configuration information for each step of the installation, operation
and set up procedures. Verify that your settings and adjustments are
consistent with factory recommendations. Refer to Chapter 5, Troubleshooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 5, contact Sierra Instruments by fax or by
E-mail (see inside front cover). For urgent phone support you may
call (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00
p.m. PST. In Europe contact Sierra Instruments bv at +31 20
6145810. When contacting Technical Support, make sure to include
this information:
• the flow range, serial number and Sierra order
number and model number (all marked on the
transducer nameplate)
• the problem you are encountering and any correc-
tive action taken
• application information (gas, pressure, temperature
Chapter 1 Introduction Series 830/840/860 Instruction Manual
1-4 IM-83/84/86-H
The Series 800 Flow Sensing Principle
The operating principle of Sierra’s Side-Trak and Auto-Trak flow
meters and controllers is based on heat transfer and the first law of
thermodynamics. During operation process gas enters the instrument’s flow body and divides into two flow paths, one through the
sensor tube, the other through the laminar flow element bypass.
The laminar flow element bypass generates a pressure drop, P1–P2,
forcing a small fraction of the total flow to pass through the sensor
tube (m1) which is then monitored.
Figure 1-1. Flow Paths through the Transducer
Two resistance temperature detector (RTD) coils around the sensor
tube direct a constant amount of heat (H) into the gas stream. In actual operation, the gas mass flow carries heat from the upstream
coil to the downstream coil. The resulting temperature difference
(∆T) is detected by the RTD coils.
Series 830/840/860 Instruction Manual Chapter 1 Introduction
IM-83/84/86-H 1-5
Figure 1-3. Sensor Temperature Distribution
Figures 1-2 and 1-3 show the mass flow through the sensor tube as
inversely proportional to the temperature difference of the coils. The
coils are legs of a bridge circuit with an output voltage in direct proportion to the difference in the coils’ resistance; the result is the temperature difference (∆T). Two other parameters, heat input (H) and
coefficient of specific heat (Cp) are both constant. The resulting output is nearly linear over the transducer’s normal operating range.
Figure 1-4. Linear Range of the Transducer’s Output Signal
For mass flow controllers, once the gas flows through the monitoring section, it is then controlled by the built-in servo-control valve.
For flows up to 500 slpm, Sierra’s proprietary high-efficiency FastTrak Electromagnetic Valve is used. The normally closed FastTrak valve is similar to an on/off solenoid valve, except that the
current to the valve coil, and hence the magnetic field, is modulated so that the ferromagnetic valve armature, or valve plug, assumes the exact height above the valve’s orifice required to maintain the valve’s command flow. The result is nearly infinite resolution. All controllers with the Fast-Trak valves are available with a
“Soft-Start” option for those processes that require it.
For higher flows of 50 to 200 scfm, controllers are equipped with Sierra’s microprocessor-controlled, motor-operated rotary valve. While
the rotary valve does not provide a completely positive shut-off, it
does offer accurate control for higher flow installations.
Series 830/840/860 Instruction Manual Chapter 2 Installation
IM-83/84/86-H 2-1
Chapter 2 Installation
Quick-Installation Instructions For
830/840 L, M and H Meters And Controllers
1. NEVER TEST FOR LEAKS WITH LIQUID LEAK DETECTOR. If
liquid seeps into the electronics or the sensor compartment below,
the instrument may be damaged. Use a pressure-decay test instead.
2. Install an appropriate in-line filter upstream if the gas contains
any particulate matter or condensed moisture. (A 15-micron particulate filter for full-scale flows up to 30 liters/minute is recommended, 30 micron for flows above 30).
3. Mount with a horizontal gas-flow. This orientation is preferable
unless the factory calibration was specifically performed for a vertical flow. Consult your calibration certificate. (Horizontal flow is
assumed unless vertical flow upward or downward is specified).
4. DO NOT APPLY POWER TO THE OUTPUT LOOP on units
equipped and calibrated for a 4-20 mA output signal. This is NOT
a loop-powered device. Damage will occur.
5. Wire the instrument per the diagrams on the back of this card or
the Operator’s Manual, Chapter 2. Controllers require a separate
ground wire for the Valve Return signal as well as a Setpoint signal (no wire connected to the setpoint may cause the valve to float
open).
6. Apply the gas listed on the label to the inlet at the recommended
inlet pressure as listed on the calibration certificate. On a meter,
obtain this pressure by blocking off the flow downstream. On a
controller, confirm that there is no gas flowing through it with a
zero setpoint. (If there is a flow, consult the Valve Adjustment
Procedure in the manual).
7. Apply power and verify or adjust the zerosetting after allowing
a 15-minute warm-up period. The zero adjustment potentiometer
is accessed through the upper hole in the side of the electronics
cover. (See the Operator’s Manual, Chapter 3).
8. DO NOT LEAVE A SETPOINT APPLIED TO A CONTROLLER WHEN NO GAS IS AVAILABLE TO THE INLET FITTING. The
control circuit will apply the maximum voltage to the valve coil resulting in eventual overheating. Damage may occur. (Instead,
consult the Operator’s Manual for use of the “Valve Off” feature).
9.An ANNUAL factory evaluation and calibration is recom-
Chapter 2 Installation Series 830/840/860 Instruction Manual
2-2 IM-83/84/86-H
Installation Overview
Side-Trak™ and Auto-Trak™ transducers are supplied with either
compression, VCO, VCR or NPT process connections. To ensure a
successful installation, inlet and outlet tubing should be in a clean
state prior to plumbing the transducer into the system. The shipping caps covering the inlet/outlet fittings should not be removed
until immediately prior to installation.
Before installing the transducer, verify the following:
1. Make sure the installation site meets the specific operating pa-
rameters recorded on the transducer’s nameplate. Each transducer is factory-configured for a specific gas and flow range,
pressure differential, temperature range and mounting position.
2. Do not locate the transducer in areas subject to sudden tem-
perature changes, moisture, drafts or near equipment radiating
significant amounts of heat. Make sure to allow adequate space
for cable connectors and wiring.
3. Make sure the location meets the minimum number of recom-
mended pipe diameters upstream and downstream of any NPT
or 1/2-inch process connection. A minimum of two inches is
always recommended.
4. If the gas contains any particulate matter, install an in-line fil-
ter prior to the transducer. Recommended filter size: 15 micron for flows of 10 to 30 slpm, 30 micron for above 30 slpm.
5. If the transducer is mounted in any position other than horizontal
and it was not calibrated specifically for that position, contact Sierra’s Customer Service Department for instructions.
6. For meters, if a potential over-flow condition exists, insert a
valve or critical orifice in the line to limit flow to approximately 25 percent above the full scale range of the meter.
7. For controllers, use a properly-sized pressure controller and
verify the controller orifice size is the smallest in the system.
There can be no restrictions (such as valves, tubing or piping
internal diameters, reducers, etc.) upstream or downstream of
the controller less than the valve orifice diameter.
8. Confirm that the transducer o-ring material is compatible with
Series 830/840/860 Instruction Manual Chapter 2 Installation
IM-83/84/86-H 2-3
Installing the Transducer
Follow the installation instructions that are applicable to your transducer’s process connection. For all NPT and 1/2-inch size fittings,
see the piping recommendations given on page 2-3. Before use, all
plumbing should be checked carefully for leaks and the transducer
purged with dry nitrogen. Make sure that the tubing is free from
burrs, or rims caused by cutting.
Compression Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction
of flow.
2. Verify the position of the
front and back ferrule as
shown at right. Insert the
tubing into the fitting. Make
sure that the tubing rests
firmly on the shoulder of the fitting and that the nut is finger
tight. Scribe the nut at the six o’clock position.
3. While holding the fitting body steady with a backup wrench,
tighten the nut 1-1/4 turns, watching the scribe mark make one
complete revolution and continue to the nine o’clock position.
For 1/16-inch, 1/8-inch and 3/16-inch (2, 3 and 4 mm) sizes,
tighten only 3/4 turn from finger tight. Do not over-tighten!
4. Check the system’s entire flow path thoroughly for leaks. (Do
not use liquid leak detectors, instead monitor pressure decay.
Over-exposing the transducer to leak detector fluid may damage the unit.)
VCO Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
2. Tighten the nut finger tight, and then 1/4 turn tighter with a
wrench. Do not over-tighten!
3. Check the system’s entire flow path thoroughly for leaks. (Do
not use liquid leak detectors, instead monitor pressure decay.
Over-exposing the transducer to leak detector fluid may damage the unit.)
Chapter 2 Installation Series 830/840/860 Instruction Manual
2-4 IM-83/84/86-H
VCR Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
2. Install new washers compatible with the gas to be used. (Do
not mix or interchange parts of tube fittings made by different
manufacturers.)
3. Tighten the nut finger tight, and then 1/4 turn tighter with a
wrench. Do not over-tighten!
4. Check the system’s entire flow path thoroughly for leaks. (Do
not use liquid leak detectors, instead monitor pressure decay.
Over-exposing the transducer to leak detector fluid may damage the unit.)
NPT Fittings and 1/2-Inch Connections - All Types
1. Install a section of straight pipe at least five pipe diameters in
length upstream of the transducer. For flow meters only, also
allow at least two pipe diameters downstream for accurate operation. DO NOT use reducers.
2. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
3. Tighten fittings until leak tight (refer to published standards for
specific recommendations).
4. Check the system’s entire flow path thoroughly for leaks. (Do
not use liquid leak detectors, instead monitor pressure decay.
Over-exposing the transducer to leak detector fluid may damage the unit.)
Figure 2-1.
Piping Requirements for NPT and 1/2-Inch Process Connections
Series 830/840/860 Instruction Manual Chapter 2 Installation
Follow National Electric
Warning!
Code or your local code
safety practices when wir-
ing or connecting this uni
to a power source. Only
qualified personnel should
wire the transducer.
Wiring the Transducer
Side-Trak and Auto-Trak transducers require a +15, 0, –15 VDC power
supply and a readout device. Additionally, controllers require a set
point input(0-5 VDC standard, 4-20 mA optional). Transducers are
connected to the power supply, output signal and set point control
through either a 20-pin card-edge connector or 15-pin D-connector.
Motor-oper ated rotary valves use a 10-pin circular connector. Wiring
instructions are given for each type of connector on the following pages.
If you are replacing a not he r m anuf ac tu rer’ s t ran sdu cer wi th a Si err a pr odu ct ,
see the factory-supplied wiring supplement for instructions specific to your
installation.
Recommended Wire Gauge
Using the correct wire gauge for cabling runs to transducers enhances system reliability. For cable runs over 150 feet (46 m), consider using a local power supply.
Series 830/840/860 Instruction Manual Chapter 2 Installation
Wiring Transducers with 20-Pin Card-Edge Connectors Input
Power Connections
Figure 2-2. Input Power Connections (20-Pin Connector)
Output Signal Connections
Standard output for all transducers is a 0-5 V DC signal, which directly corresponds to the0 to 100% mass flow full scale range.
Output signals are linear and require a minimum load resistance of 1000
Ohms (4-20 mA output600 Ohms maximum loop resistance).
The Auto-Trak 860 allows transmission of all available input/output
signals to your computer’s R S-485 communications port (connection
details are given below). For additional information, refer to the “Auto Net” Software Addendum included with this manual.
Figure 2-3. Output Signal Connections (20-Pin Connector)
Notes: 1. Controllers require a separate, dedicated ground wire between pin 2 and the power
supply common to carry the valve coil current.
2. 4-20 mA output signal is ground-referenced (“non-isolated”).
Warning: Do NOT apply any external voltage to this loop.
Chapter 2 Installation Series 830/840/860 Instruction Manual
Input Signal Connections (Controllers Only)
The0-5 VDC (4-20 mA optional) set point command input signal
allows you to set the flow to any desired value within the range of
the controller. The set point input signal is a direct linear representation of 0 to 100 percent of the controller’s mass flow full scale
value.
For 0–5 V DC set point control, connect a manual-adjust potentiometer as shown below. Any potentiometer value between 5K and 100K
may be used but is limited to a V DC input signal only. For 4-20
set point control, use a 4-20 mA generator.
mA
If the set point input is not connected to some type of control
device,
no set
up and
the valve on/off switch must be activated in the off position. If
point command is present on the controller when powered-
the valve is not switched off, the valve will drift wide open.
Figure 2-4.
Local Command Set Point Potentiometer (20-Pin Connector)
Figure 2-5.
External Command Set Point Source (20-Pin Connector)
Note: Prior to2001, sierrasupplied separate cablesforcontrollers and meters.
These arenowcombined into onestyle for 830/840devices andonestyle for
860M/C.
Series 830/840/860 Instruction Manual Chapter 2 Installation
Wiring Transducers with 15-Pin D-Connectors
Input Power Connections
1
9
+
Case
10
11
12
1
14
1
2
3
COM
4
5
6
-15
7
8
Pin 3 connection fo
+15/0/-15
Vdc
Power
Supply
controllers only
Figured 2-9. Input Power (15-Pin D-Connector)
Note: Use separate, dedicated wires for pins 9 and 10 to the power supply common on meters and pins 9,10 and 3 on controllers.
Output Signal Connections
Standard output for all transducers is a 0-5 V DC signal, which directly corresponds to the0 to 100% mass flow full scale range.
Output signals are linear and require a minimum load resistance of
1000 Ohms (4-20 mA output600 Ohms maximum loop resistance).
Figure 2-10. Output Signal (15-Pin D-Connector)
Notes:
1. Panel meter or read-out device should be wired to beat the
same ground potential as the power supply.
2. 4-20 mA output signal is ground referenced (“non-isola ted”).
Warning: Do not apply any external voltage to this loop.
Chapter 2 Installation Series 830 /840/860 Instruction Manual
Input Signal Connections (Controllers Only)
A 0-5 VDC (4-20 mA optional) set point command inputs ignal allows
you to set the f low to any desired value within the range of
the
controller. The set point input signal is a direct linear representation of
0 to 100 percent of the controller’s mass flow full scale
value.
For 0–5 V DC set point control, connect a manual-adjust potentiometer
as shown below. Any potentiometer value between 5K and 100K
be used but is limited to a V DC input signal only. For 4-20 mA
V DC set point control, use a PLC or other digital device.
may
or 0–5
If the set point input is not connected to some type of control device,
the valve on/off switch must be activated in the off position. If no set
point command is present on the controller when powered-up and the
valve is not switched off, the valve w ill drif t wid e open .
Figure 2-11.
Local Command Set Point Potentiometer (15-Pin D-Connector)
Figure 2-12.
External Command Set Point Source (15-Pin D-Connector)
Series 830/840/860 Instruction Manual Chapter 2 Installation
IM-83/84/86-H 2-13
Valve Purge and Valve Off Connections (Controllers Only)
Depending on the valve voltage, connect as shown below.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Purge
On
Off
Open
collector
output
Valve Off
On
Off
15 Volt Valve
COM
Figure 2-13. Valve Purge and Valve Off (15-Pin D-Connector)
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Purge
On
Off
Open
collector
output
Valve Off
On
Off
30 Volt Valve
COM
+15V
Figure 2-14. Valve Purge and Valve Off (15-Pin D-Connector)
Note: Typically controllers for 10 slpm or less will be equipped with a 15volt valve circuit, while those above 10 slpm with a 30-volt circuit. Consult Technical Support if necessary.
Note: Prior to2001, Sierra suppliedseparate cables forcontrollers and meters.
These arenowcombined into onestyle for 830/840devices andonestyle for
860M/C
Series 830/840/860 Instruction Manual Chapter 3 Operation
IM-83/84/86-H 3-1
Chapter 3 Operation
This chapter covers transducer operation and controller features
available on Sierra’s Side-Trak™ and Auto-Trak™ models. For
more information on the advanced digital RS-485 commands
available on Auto-Trak transducers, refer to the Series 860 “AutoNet” Software Addendum included with this manual.
Standard output for all transducers is a linear 0-5 VDC output signal
directly corresponding to 0 to 100% of the mass flow full scale
range. For mass flow controllers an input signal of 0-5 VDC (4-20
mA optional) allows set point flow control to any desired value
within the range of the model. The input signal is a direct linear representation of 0 to 100% of the mass flow full scale value. A 0 VDC
(or 4 mA) set point will cause a condition of 0% flow to occur and a
5.00 VDC (20 mA) set point will cause a flow condition equivalent
to 100% of flow to occur.
Mass Flow Meter Operation
When the transducer is installed and the system has undergone a
complete leak check:
1. Apply power. Allow thirty minutes of warm-up time. (When
power is first applied, the output signal from the transducer
remains fixed at a much higher than normal level until the sensor warms up to its normal operating temperature range.)
2. Turn on the gas supply. Perform an initial zero output check
(only required for first-time start ups). Set flow to zero. Connect
a digital multimeter to V+out (4-20 mA out) and COM terminals.
3. Check the reading. If it is not within ±10 mV (±0.10 mA) of
zero, adjust the zero potentiometer to zero the transducer. (The
zero pot is located behind the upper metal button on the side of
the transducer or behind the swing-out plastic door.)
Chapter 3 Operation Series 830/840/860 Instruction Manual
3-2 IM-83/84/86-H
Mass Flow Controller Operation
Electromagnetic Valve Controllers
When the transducer is installed and the system has undergone a
complete leak check:
1. Make sure the set point is zero before applying power. If the
set point input is not connected to some type of control device,
set the valve on/off switch in the off position. (No connection
may cause the setpoint to float). Note: The valve opens momentarily when power is first applied.
2. Apply power. Allow thirty minutes of warm-up time. (When
power is first applied, the output signal from the transducer
remains fixed at a much higher than normal level until the sensor warms up to its normal operating temperature range.)
3. Turn on the gas supply. Perform an initial zero output check
(only required for first-time start ups). Set flow to zero. Connect
a digital multimeter to V+out (4-20 mA out) and COM terminals.
4. Check the reading. If it is not within ±50 mV (±0.16 mA) of
zero, adjust the zero potentiometer to zero the transducer. (The
zero pot is located behind the upper metal button on the side of
the transducer or behind the swing-out plastic door.)
5. Adjust the controller set point to the desired flow rate.
Caution: Never leave the controller with no gas supply while supplying a
setpoint. The valve will open to its maximum position and eventually
overheat causing possible damage to the control circuit. If it is desired
to interrupt the gas flow for an extended period, consider utilizing the
“Valve-Off” circuit.
Series 830/840/860 Instruction Manual Chapter 3 Operation
IM-83/84/86-H 3-3
Motor-Operated Rotary Valve Controllers
When the transducer is installed and the system has undergone a
complete leak check:
1. Make sure the valve set point is zero before applying power.
2. Apply power. Allow thirty minutes of warm-up time. (When
power is first applied, the output signal from the transducer
remains fixed at a much higher than normal level until the sensor warms up to its normal operating temperature range.)
3. Turn on the gas supply. Verify that no flow exists. Perform an
initial zero output check (only required for first-time start ups).
Set flow to zero. Connect a digital multimeter to V+out
(4-20 mA out) and COM terminals.
4. Check the reading. If it is not within ±10 mV (±0.10 mA) of
zero, adjust the zero potentiometer to zero the transducer. (The
zero pot is located behind the upper metal button on the side of
the transducer or behind the swing-out plastic door.)
5. Adjust the controller set point to the desired flow rate.
All motor-operated rotary valves seek a “deadband” where the
flow is identical to the set point within specified tolerances. Keep
this in mind when checking the instrument’s performance. For instance, a 100 scfm flow controller may have a set point of 2.500
volts (50 scfm) and a flow indication of 2.550 volts (51 scfm). This
controller is working correctly, since the difference between the
flow and the set point is within 1 scfm, which is the specified accuracy of ±1% of full scale (0.01 x 100 scfm = 1 scfm). The width of
the deadband varies with the valve size and range of each controller.
During operation, the motor runs only when it needs to. If the pressure differential across the valve is insufficient to obtain set point,
the valve will open fully and the motor will cease to run. Conversely, if the set point is zero, the valve will close, and the motor
will cease to run. In either case, there will be no control action.
Check for adequate pressure differential and that set point command
voltage is present at the motor operator connector.
Chapter 3 Operation Series 830/840/860 Instruction Manual
3-4 IM-83/84/86-H
Mass Flow Controller Features
Over-Range and Cold Sensor Lockout Circuit
Side-Trak Models incorporate an over-range circuit which disables
the output signal when the sensor is heating to its normal operating
range, or if an over flow condition exists. The operation of this circuit can be verified by observing the 5 VDC (20 mA) output signal
during power up.
The cold sensor lockout safety circuit automatically closes the
valve if a fault condition is detected that could result in uncontrolled flow (valve wide open). The circuit operates by monitoring
the temperature of the sensor elements. If the temperature falls below a preset limit the circuit forces the output signal to a fixed high
level which disables the output signal. The cold sensor lockout circuit is enabled during initial start-up. This feature is not available if
liquid is present in the flow path.
There are several conditions which could trigger the cold sensor
lockout circuit:
1. Operation at a temperature below the instrument’s rating.
2. Power failure while running at or near full scale. Upon resump-
tion of power, the valve will remain closed until the minimum
operating temperature is again reached.
3. Sensor failure.
If needed, the cold sensor lockout circuit and/or over-range indicator
can be disabled by removing CR1 from the main 830 circuit board.
Controller Auto Shut-Off Feature
All flow controllers are normally provided with an Auto Shut-Off
feature that closes the valve at a command signal level of approximately 2% of full scale, or less. (All electromagnetic valves fail
normally-closed). On 840 controllers, this value can be reduced.
Contact the factory for details on this option.
Series 830/840/860 Instruction Manual Chapter 3 Operation
IM-83/84/86-H 3-5
On-Off Control and Soft Start Option
On-off control is provided for all models equipped with an electromagnetic valve. This feature is available by connecting a manual
on-off switch to the transducer connector as shown in Chapter 2.
Certain processes cannot tolerate the normally fast response of Sierra mass flow controllers. In these cases a “soft start” circuit
slows the mass flow controller’s valve response characteristics to
avoid any bursts of flow that could disrupt a delicate process or
perhaps stir up particulate.
The standard soft start time constant is 1.5 seconds, providing a 7.5
second response time to within 2 percent of final value. The time
constant is factory-set to fit your exact requirements.
Figure 3-1. Soft Start Response Times
The Model 860 provides valve on-off and soft start control with either the manual hardware option, or with digital (software) control
via RS-485. For more information, refer to the Series 860 “AutoNet” Software Addendum.
Chapter 3 Operation Series 830/840/860 Instruction Manual
3-6 IM-83/84/86-H
Valve Purge Function
Electromagnetic valves are driven with either 15 or 30 volts depending on the range and pressure. In general, units below 10 slpm
have the 15 volt valve. When the purge function occurs, electromagnetic valves are driven fully open regardless of the set point
input. Furthermore, the purge function will override any valve-off
function. This option is not available on motor-operated rotary
valves. Motor-operated valves are purged by setting the command
signal to full scale. There is no equivalent monitor function.
The Model 860 allows valve monitoring and the purge function to
be utilized in either analog or digital (RS-485) mode. If both hardware and software control options are used simultaneously on an
860, conflicts may occur that can produce undesired results.
For +15 Volt Valve Controllers: The valve purge function is activated when the Valve Test Point/Purge connection is connected to
ground. The maximum valve voltage is indicated in the “comments”
field of the calibration certificate. (Units with ranges below 10
slpm)
For +30 Volt Valve Controllers: The valve purge function is activated when the Valve Test Point/Purge connection is connected to
+15 VDC. The maximum valve voltage is indicated in the “comments” field of the calibration certificate.
A Sierra controller intended as a replacement for another manufacturer’s controller may have Pin D and other pins disconnected. Some
options may not be available.
Purging Non-Reactive Gases:
Purge the transducer with clean, dry nitrogen for a minimum of two hours.
Purging Reactive Gases:
One of the following methods may be used:
• Cycle purge. This is done by alternately evacuating and purg-
ing the transducer for two to four hours with clean, dry nitrogen.
• Purge the transducer with clean, dry nitrogen for eight to
twenty-four hours.
• Evacuate the transducer for eight to twenty-four hours.
When toxic or corrosive gases are used, purge unit thoroughly with
inert dry gas before disconnecting from the gas line. If a transducer
used with a toxic or corrosive gas is returned to the factory, a Ma-
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-1
Chapter 4 Maintenance
Side-Trak™ and Auto-Trak™ transducers essentially require no
scheduled maintenance other than periodic flow path cleaning if
the gas is dirty. If an in-line filter is used, the filtering element
should be periodically be replaced or ultrasonically cleaned. Included in this chapter are general instructions for:
• Flow Path Maintenance ............... page 4-1
• Sensor Maintenance.....................page 4-6
• Valve Maintenance......................page 4-8
• Valve Adjustment......................page 4-12
• Transducer Calibration .............. page 4-15
Flow Path Maintenance
The transducer flow path is 316 stainless steel (wetted magnetic
parts of the solenoid valve are 430F stainless steel) with Viton®
Neoprene® or Kal-Rez® (or equivalent) seals, depending on the
gas used. The flow path should be periodically inspected and
cleaned as required.
When toxic or corrosive gases are used, purge unit thoroughly with
inert dry gas before disconnecting from the gas line. A transducer
used with toxic or corrosive gas must NOT be returned to the factory. Contact Technical Support for details.
Cleaning the Laminar Flow Element
The laminar flow element (LFE) is a precision flow divider which
diverts a preset amount of flow through the sensor tube. The LFE
is made of precision machined 316 stainless steel. The particular
LFE used depends on the gas and flow range of the instrument and
is identified by the number on the transducer nameplate. Should
the LFE require cleaning or inspection due to deposition, use the
appropriate cleaning procedure given on the following pages.
CAUTION! : REMOVING THE LFE WILL SHIFT THE
CALIBRATION OF THE INSTRUMENT. DO NOT REMOVE THE LFE
UNLESS YOU PLAN ON RE-CALIBRATING THE TRANSDUCER.
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-2 IM-83/84/86-H
Figure 4-1. Low Flow Transducer LFE Cleaning
Low Flow Transducers:
The LFE is accessed by unscrewing the main inlet fitting and removing it from the flow body. The LFE is screwed into the inlet
fitting, which has been specially machined for this purpose. To access the components:
1. Remove the unit from the system.
2. The inlet filter screen is held in place in the inlet fitting by the
LFE. Disassemble by holding the fitting steady with a wrench
and unscrewing the LFE with a medium flat-tipped screwdriver.
3. Remove the LFE assembly taking care not to bend the inlet
screen. Inspect the sealing O-ring and replace if necessary. Inspect the inlet screen and replace if corroded or damaged. Light
to medium particulate contamination can be cleaned by back
washing with a suitable solvent. Air dry thoroughly.
4. Inspect the LFE for damage and replace if necessary. Replace-
ment of the LFE or inlet screen requires transducer recalibration.
5. Re-assemble components. When the transducer is re-installed
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-3
Figure 4-2. Medium Flow Transducer LFE Cleaning
Medium Flow Transducers:
In the medium flow body, the LFE assembly consists of the honeycomb laminar flow element, inlet screen, 0.63 inch long standoff,
two ranging washers, 2-1/4 inch long 4-40 screw and 4-40 nut.
Range changes in the honeycomb element are made with various
diameter ranging washers. To access the components:
7. Remove the unit from the system.
8. Access the LFE by unscrewing the four 10-32 socket head cap
screws from the inlet side of the flow body and remove the inlet end cap.
9. Remove the LFE assembly taking care not to bend the inlet
screen. Inspect the sealing O-ring and replace if necessary. Inspect the inlet screen and replace if corroded or damaged. Light
to medium particulate contamination can be cleaned by back
washing with a suitable solvent. Air dry thoroughly.
10. Inspect the honeycomb element for damage and replace if nec-
essary. Replacement of the LFE or inlet screen requires transducer re-calibration.
11. Re-assemble components. When the transducer is re-installed
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-4 IM-83/84/86-H
Figure 4-3. High Flow Transducer LFE Cleaning
High Flow Transducers:
The high flow LFE is similar to the honeycomb element used in the
medium flow body but larger in diameter. The high flow body consists of four parts: inlet tube, inlet cap, main flow body and end cap.
The inlet tube is only removed to inspect and replace the sealing Oring between the inlet tube and inlet cap. To access the components:
1. Remove the unit from the system.
2. To remove the inlet screen, remove the four 1/4-28 socket head cap
screws on the inlet side of the flow body and separate the inlet cap
from the main flow body.
3. Inspect the inlet screen for damage and corrosion and replace if neces-
sary. Light to medium particulate contamination can be cleaned by back
washing with a suitable solvent. Air dry thoroughly.
4. Inspect the sealing O-ring for damage and replace if necessary.
The inlet screen is mounted with the fine mesh side facing the inlet.
5. To remove the LFE loosen and remove the four threaded rods hold-
ing the end cap to the main flow body. Separate the end cap from the
main flow body and remove the LFE assembly. The LFE assembly
consists of: 6-32 x 31/8 inch long screw, a #6 washer, two ranging
washers, honeycomb LFE, spacer, inlet filter and 6-32 nut.
6. Inspect the honeycomb element for damage and replace if necessary.
Replacement of the LFE or inlet screen requires transducer recalibration.
`
7. Re-assemble components. When the transducer is re-installed in the system,
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-5
Figure 4-4. NPT Transducer LFE Cleaning
NPT-Type Transducers:
The NPT flow body is comprised of a stainless steel pipe terminated on either end by an end cap sealed with an O-ring. The end
caps are held in place by four tie rods which run the length of the
flow meter and are fastened with eight 3/8-16 nuts and washers.
The inlet screen is mounted to the inside of the inlet end cap on
three standoffs. The LFE is a cylindrical honeycomb element located in the flow body between the two sensing ports. In most
cases, the LFE can be cleaned without removing it from the flow
body. If it cannot be cleaned by back washing, return for factory
service. To access the components:
1. Remove the unit from the system.
2. To remove the inlet filter screen, stand the flow body on end
with the inlet facing up and remove the four 3/8-16 nuts on the
ends of the tie rods. Carefully lift the end cap and filter screen
assembly up and off the meter.
3. Inspect the inlet screen for damage and corrosion and replace if
necessary. Light to medium particulate contamination can be
cleaned by back washing with a suitable solvent. Air dry thoroughly before re-assembly. Mount the inlet screen with the fine
mesh side facing the inlet.
4. Inspect the sealing O-ring for damage and replace if necessary.
5. Re-assemble components. When the transducer is re-installed
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-6 IM-83/84/86-H
Sensor Maintenance
The sensor tube is straight and has a relatively large, 0.031 inch
ID, which makes inspection and cleaning easy. Sensor maintenance consists of:
• inspecting and cleaning the sensor flow path
• checking the sensor for proper electrical function
Sensor Cleaning and Inspection
Cleaning is accomplished by simply rodding out the sensor with
the Sensor Cleaning Stylette, part number “CK”, available from
Sierra for this purpose. (A 0.020 inch to 0.028 inch diameter piano
wire may also be used.) To access the sensor for inspection or
cleaning:
1. Remove the unit from the system.
2. Remove the two socket head ac-
cess port plugs with a 1/4 inch
Allen wrench.
3. Visually inspect the sensing ports
and sensor.
4. Use a hemostat or tweezers to push the cleaning wire into the
downstream opening of the sensor tube. Do not force the cleaning wire; move it back and forth–DO NOT TWIST OR
ROTATE.
5. Flush the sensor tube with a non-residuous solvent. In cases
where solids are deposited in sensor, units should be returned
to factory for complete cleaning and re-calibration.
6. Blow dry all parts with dry nitrogen and re-assemble.
7. When the transducer is re-installed in the system, leak test the
To check the electrical integrity of the sensor windings:
1. Remove the unit from the system.
2. Remove the two 4-40 Phillips head screws located on top of
the electronics enclosure and slide the enclosure up and off.
Use a nut driver to remove the two standoffs supporting the
printed circuit boards. Tilt the boards to access the three sensor
wires as shown below.
3. Locate the black, red and white wires connecting the sensor to
the main circuit board.
4. Connect one lead of an Ohm meter to the white wire and measure the resistance between the red and white wires. Next,
measure the resistance between the black and white wires.
These readings should each be approximately 50 Ohms. Low
or zero Ohm readings on either indicates a short circuit. High
or infinite Ohms readings indicate an open circuit.
5. Next, measure the resistance between the case (metal part of
the flow body) and any one of the sensor wires. This reading
should be 2 Megohms or greater. Incorrect readings require
sensor replacement and re-calibration.
6. When the transducer is re-installed in the system, leak test the
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-8 IM-83/84/86-H
Valve Maintenance
Electromagnetic valves may require spring-tension adjustments for
various reasons (See Valve Adjust Procedure). No maintenance is
required under normal operating conditions other than an occasional cleaning. Use of certain corrosive gases may require frequent replacement of the valve plug and O-rings. This indicates a
need for a different elastomer. Viton is standard, with Neoprene,
Kal-Rez (or equivalent), and PFA Teflon offered as options.
Do not attempt any valve adjustments while the meter is “on-line”
with any dangerous gas. All controllers should be thoroughly leaktested following any valve adjustment.
For the Model 860 Auto-Trak, refer to “Piston Tube Calibration
Procedure” for instructions on how to re-adjust the electromagnetic
valve for proper operation.
Electromagnetic Valves - Low and Medium Flow Controllers
Controllers with a low or medium flow body use the same valve;
high flow models use a larger version. Cleaning can often be accomplished by opening the valve, using the purge function and
flushing in both directions. Alternatively, the valve may be manually opened by loosening the 6-32 lock nut on top of the valve and
turning the adjustment screw fully counterclockwise.
To disassemble the low and medium flow valve for inspection:
1. Remove the unit from the system.
2. Remove the two 4-40 Phillips head screws on top of the enclo-
sure. Remove the enclosure by sliding it up and off. Remove
the metal cap on top of the valve by inserting a flat tip screwdriver into the slots provided and prying upward.
3. Use a 5/8 inch nut driver to loosen and remove the 5/8 inch
nut at the top of the valve. Remove the coil, coil enclosure and
warp washer.
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-9
4. The small circuit board may be separated from the main board
to ease removal of the coil. To separate, first remove the plastic
#4 mounting screw located in the center of the main circuit
board and carefully pull the two boards apart. Remove the four
4-40 socket head cap screws at the base of the valve. Separate
the valve from the flow body.
5. There are three O-rings sealing the valve assembly: one between the base and the flow body, one under the valve seat
(orifice), and one on the top adjusting screw inside the valve.
Inspect the O-rings for damage and replace as necessary. (You
should replace all O-rings whenever the valve is disassembled.)
6. Inspect the valve seat and plug for corrosion or roughness and
replace as necessary.
7. Re-assemble components. Note: the 5/8 inch nut must not be
tightened with more than 10 in-lb of torque.
8. When the controller is re-installed in the system, leak test the
connection.
9. Adjust the valve for proper operation as given in the controller
valve adjustment procedure.
Electromagnetic Valves - High Flow Controllers
The high flow electromagnetic valve is similar in design to the low
flow valve. As with the low flow valve, cleaning can often be accomplished by opening the valve using the purge function and
flushing in both directions. Alternatively, the valve may be manually opened by loosening the 6-32 lock nut on top of the valve and
turning the adjustment screw fully counterclockwise.
To disassemble the high flow valve:
1. Remove the unit from the system.
2. Remove the two 4-40 Phillips head screws from the top of the
enclosure. Remove the enclosure by sliding it up and off. Remove the metal cap on top of the valve by inserting a flat tip
screwdriver between the two parts and lifting upward.
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-10 IM-83/84/86-H
3. Unscrew the coil cover with a 11/16 inch wrench. Remove the
coil and coil enclosure. Note the location of any washers or
spacers for re-assembly.
4. The small circuit board may be separated from the main board
to ease removal of the coil. To separate, first remove the
mounting screw located in the center of the main circuit board
and carefully pull the two boards apart.
5. Remove the four 1/4-28 socket head cap screws at the base of
the valve. Separate the valve from the flow body, taking care
not to lose any small parts.
6. There are three O-rings sealing the valve assembly: one be-
tween the base and the flow body, one under the valve seat
(orifice), and one on the top adjusting screw inside the valve.
Inspect the O-rings for damage and replace as necessary. (You
should replace all O-rings whenever the valve is disassembled.)
7. Inspect the valve seat and plug for corrosion or roughness and
replace as necessary.
8. Re-assemble the components. When the controller is installed
in the system, leak test the connection.
9. Adjust the valve for proper operation as given in the controller
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-11
Motor-Operated Rotary Valves
The electronic components in the Model 840 motorized valve flow
controllers require no maintenance with the exception of the motor,
which requires replacement when its service life has ended. This
will be evidenced by “hunting,” or the inability to maintain set
point without a constant motion back and forth, or outright failure.
The motor normally has a long service life, typically more than
five years, and all other possibilities should be explored before
considering replacement.
To disassemble the motor-operated rotary valve:
1. To allow for component removal, set the valve to the fully
open position before removing power.
2. Remove the unit from the system.
3. If desired, remove the valve from the flow body to make disassembly easier. Unplug the connector and remove the four bolts
holding the valve assembly to the flow body. Inspect the sealing
O-ring and replace as necessary.
4. Unscrew the six 4-40 Phillips head screws from the valve motor enclosure and remove the lid. Loosen the lower set screws
in the motor shaft coupling (there are two set screws in each
location to prevent loosening during operation). Inspect the
valve rotor shaft and coupling for scoring and replace if necessary.
5. Remove the three 8-32 screws holding the valve to the enclosure and separate the valve body from the enclosure.
6. Unscrew the four 6-32 x 7/8 inch flat head screws from the top
of the valve and remove the valve base plate and spacer. Remove the valve bonnet, rotor and Teflon washers. Inspect the
valve rotor and body for wear and replace the O-ring and seal
in the valve bonnet.
7. Inspect and replace the Teflon valve leaves and upper and
lower Teflon washers if they are worn or damaged.
8. Re-assemble the components. When the controller is reinstalled in the system, leak test the connection.
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-12 IM-83/84/86-H
840 L, M and H series
Valve Adjustment Procedure
The Sierra 840 L, M and H (low, medium or high-flow units) control the flow with a proportional electromagnetic valve that is set
up for certain process conditions. Variables, which affect their operation, include orifice size, spring selection and adjustment, input
and output pressures and gas-flow orientation (horizontal or vertical). If operating conditions change it may be necessary to make a
valve spring adjustment. An adjustment may also become necessary due to a valve-seat depression or spring sag, both of which
can cause a change in the internal tension of the valve spring and
electrical current. An incorrectly adjusted valve can be detected by
OSCILLATION (UNSTABLE OUTPUT), NOT OBTAINING
MAXIMUM FLOW, OR LEAKAGE AT A ZERO SET
POINT (LEAK-BY).
The following valve adjustment procedure may correct the above
conditions.
Equipment needed:
L and M:
1/16” hex wrench
5/16” nut driver
H:
Standard screwdriver
3/8” nut driver
(Quick fix for leak-by). If your unit is experiencing leak-by, first
confirm that this condition is mechanical in nature rather than electrical by shutting down the power. If leak-by stops (check for
downstream flow), the problem is electrical in nature and the instrument may require factory repair. If leak-by continues, the
problem is mechanical and may be fixed by an adjustment. Remove the small cover that is on top of the silver solenoid to the
right of the rectangular electronic enclosure. (On some 840L-units
this may require removing the plastic electronic enclosure to expose the valve coil.) With the power off, apply the appropriate inlet pressure to the controller per the inlet/outlet operating pressures listed on your label and apply a zero setpoint. Loosen the
lock nut. Adjust the center adjustment screw slowly clockwise
(increasing the spring tension) 1/4 turn at a time until this leak-by
subsides. (Putting your finger over the outlet fitting can confirm
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-13
the no-leak condition). Now, adjust this screw clockwise one quarter turn past this point, tighten the locknut and replace the cap.
Apply power and test the controller for proper operation.
(Quick fix for oscillation). If your unit is experiencing oscillations
at lower flows, apply power, apply the desired setpoint, and adjust
the valve spring adjustment screw ¼ turn at a time clockwise (increasing the spring tension – see procedure above) during flow until the symptoms disappear. Adjust ¼ turn further and tighten the
locknut.
If your unit is experiencing other symptoms listed above, you
may require this more involved adjustment method. Obtain a
voltmeter and prepare to monitor the voltage found on pins 4 to 6
on a D-connector, or D to F on a card-edge connector. Or, if these
pins are inaccessible, turn off the power and remove the cable.
1. Remove the screws and/or nuts on the top of the controller
electronics cover. Cut the yellow calibration sticker. Remove
the silver zero and/or span plugs on the right side of the enclosure or turn the plastic access-hole doors to the vertical position. Pull the cover up and off. Re-connect the cable, apply
gas pressure, and apply power to the controller. Connect a
voltmeter to the two wires on the back PCA that are routed to
the solenoid and set the scale for 30 volts or higher.
2. Determine if this controller has a 15-volt control circuit or a
30-volt control circuit. (Don’t confuse this with the setpoint or
output signals – this is an internal voltage that energizes the
valve coil to raise the valve seat and open the internal orifice).
M and H instruments are ALWAYS 30-volt circuits, while L
instruments can be either one, based mainly on their flow-rates.
Units scaled for below 10 slpm are typically set up for 15 volts
while those 10 slpm or over are set up for 30 vdc. (Note: this
voltage is typically listed on the calibration certificate in the
“Comment” section). If there is any question, contact the factory for clarification.
3. Apply a 100% set point and monitor this valve voltage, allowing at least three minutes for stabilization. The voltage “window” for a 100% setpoint is 13-18 volts for the 30-volt circuit
and 9-11 volts for the 15-volt circuit. If the voltage is out of
this specification, loosen the lock nut, and adjust the center adjustment screw. Clockwise (downward) will increase the valve
voltage, while counterclockwise will decrease it. Adjust
Chapter 4 Maintenance Series 830/840/860 Instruction Manual
4-14 IM-83/84/86-H
slowly, 1/8 turn at a time and observe until the desired voltage
is obtained.
4. Now apply a 5% setpoint signal and confirm that the voltage
does not drop below 11 volts on the 30-volt circuit or 6 volts
on the 15-volt circuit. Go back and forth between these settings and make sure that the voltages stay within these boundaries. When this voltage “window” is obtained, the symptoms
should disappear and the lock nut can be re-tightened.
5. Re-install the electronics cover by powering down the control-
ler, removing the cable, putting the cover back on and replacing the access door plugs. Reconnect the cable and apply the
power. A final check for leak-by with a zero set point and
proper flow with various control signals should be conducted.
Notes: Adjusting the valve voltage does not change the
calibration of the controller. If these adjustments do not
obtain the desired operation, contact Sierra directly.
Series 830/840/860 Instruction Manual Chapter 4 Maintenance
IM-83/84/86-H 4-15
Transducer Calibration
Calibration of Sierra’s flow meters and controllers requires a calibration standard of at least equal accuracy and preferably an order
of magnitude of at least four times better than the transducer. Sierra’s Cal-Bench Automated Primary Calibration System is the
preferred method of calibration and is used at the factory for all
calibrations from 10 sccm up to 5600 slpm.
Sierra Instruments maintains a fully-equipped calibration laboratory. All measuring and test equipment used in the calibration of
Sierra transducers are traceable to NIST standards. Sierra is ISO9001 registered and conforms to the requirements of ANSI/NCSLZ540 and ISO/IEC Guide 25. If the flow body or electronics have
been damaged or you simply want to have the transducer recalibrated, contact the factory for return shipping instructions
Series 830/840/860 Instruction Manual Chapter 5 Troubleshooting
IM-83/84/86-H 5-1
Chapter 5 Troubleshooting
This chapter is provided to help locate the cause of a transducer
failure. It is not intended to be an all inclusive repair manual. In
the case of major repairs, the unit should be returned to the factory for service. Contact Sierra Customer Service for return
shipping instructions (page 5-4).
When you suspect that the transducer is not operating correctly,
there are a few simple checks that can be made before dismantling
the unit:
1. Make certain that there are no leaks in the line.
2. Check that all cables are connected and are in good condition.
3. Verify that the power supply is of the correct polarity and
voltage and properly connected to the transducer.
4. For controllers, check for adequate pressure differential across
the controller. Check the valve voltage.
5. Double check connector pin outs when replacing another
manufacturer’s transducer.
After verifying the factors above, follow the troubleshooting procedures outlined on the next page.
Chapter 5 Troubleshooting Series 830/840/860 Instruction Manual
5-4 IM-83/84/86-H
Returning Equipment to the Factory
Factory Calibration—All Models
Sierra Instruments maintains a fully-equipped calibration laboratory. All
measuring and test equipment used in the calibration of Sierra transducers
are traceable to NIST Standards. Sierra is ISO-9001 registered and conforms to the requirements of ANSI/NCSL-Z540 and ISO/IEC Guide 25.
Instructions for Returning Your Instrument for Service
The following information will help you return your instrument to Sierra
Instruments' Factory Service Center and will ensure that your order is
processed promptly. We will assign an RMA (Return Materials Authorization) number prior to the unit’s return and send you an estimate of charges
within 24 hours of receiving your unit. Prices may vary depending on the
flow range, type of gas and operating pressure of your unit. To request detailed pricing contact your local Sierra Instruments distributor or contact
one of our offices directly. Our expedite fees are: three-day turnaround
25%, two-day turnaround 40%.
Please follow these easy steps to return your instrument for factory service:
1. Obtain a Return Materials Authorization (RMA) number from Si-
erra Instruments. You may obtain this from the factory by calling
(800) 866 0200 between 8:00 a.m. and 5:00 p.m. PST Monday
through Friday. You may also obtain this number via e-mail by
contacting service@sierrainstruments.com.
2. Once you have obtained an RMA number, complete the form. If
you require service beyond calibration, but do not know which
service(s) will be required, describe the symptoms as accurately as
possible on the RMA form. Submit electronically or by fax to
(831) 373-2414.
3. Pack your instrument carefully (bubble wrap or molded foam sug-
gested-NOT PEANUTS) and include a copy of the RMA form
(complete with Sierra supplied RMA number) with the unit(s).
Conversion of Flow Rate to Other T and P Conditions
The flow rate of your transducer is referenced to certain “standard”
conditions of temperature and pressure. Unless otherwise specified
in your order, these standard conditions are 21°C (70°F) and 760 mm
of mercury (1 atmosphere). If you wish to convert to other “standard” conditions or to find the “actual” conditions in the pipe where
your instrument is installed, use the following relationship:
P1 T2
Q2 = Q1 (1)
P2 T1
( )1 = The standard conditions under which your instrument
was calibrated,
( ) 2 = The new standard conditions or the actual temperature
and pressure conditions in the pipe,
Q1 = The gas mass flow rate referenced to the calibrated standard
conditions (sccm or slm),
Q2 = The gas mass flow rate referenced to the new standard or
actual conditions (sccm or slm—“S” means “standard,”
accm or alm—“A” means “actual”),
P = Absolute pressure (kg/cm2 or psia), and
T = Absolute temperature (°K or °R) (°K = °C + 273, °R = °F + 460).
Example 1: Changing “Standard” Conditions
If your transducer has a flow rate reading of 10.00 slm and was calibrated at standard conditions of 70°F (21°C) and 1 atmosphere (14.7
psia), and if you wish to convert this reading to standard conditions of
32°F (0°C) and 1 atmosphere, then you would use Equation (1) as follows:
14.7 460 + 32
Q2 = (10.0) = 9.28 slm
14.7 460 + 70
The flow rate referenced to 0°C will be approximately 7% lower than
when referenced to standard conditions of 21°C.
Example 2: Finding the “Actual” Flow Rate
If the flow rate and calibrated standard conditions are as given in
Example 1 and you wish to find the actual flow rate at 100°F and
30 psig, then you would use Equation (1) as follows:
The following tables provide K-factors and thermodynamic properties of gases commonly used with mass flow meters and controllers. The purpose of these tables is two-fold:
1. Calibrating an “actual” gas with a reference gas. This is particularly useful if the actual gas is not a common gas or if it is
toxic, flammable, corrosive, etc.
2. Interpreting the reading of a flow meter or flow controller
which has been calibrated with a gas other than the actual gas.
In applying the tables, the following fundamental relationship is used:
Q1/Q2 = K1/K2 (1)
Where:
Q = The volumetric flow rate of the gas referenced to standard
conditions of 0°C and 760 mm Hg (sccm or slm),
K = The K-factor defined in equation (6),
( ) 1 = Refers to the “actual” gas, and
( ) 2 = Refers to the “reference” gas.
The K-factor is derived from the first law of thermodynamics applied to the sensor tube, as described in Chapter 1:
For our purposes, we want the ratio of the flow rate, Q1, for an actual gas to the flow rate of a reference gas, Q2, which will produce
the same output voltage in a particular mass flow meter or controller. We get this by combining equations (1) and (5):
Q1/Q2 = K1/K2 = (N1/ ρ1Cp1)/(N2/ρ2CP2) (6)
Please note that the constant b cancels out. Equation (6) is the fundamental relationship used in the accompanying tables. For convenience, the tables give “relative” K-factors, which are the ratios K1/K2,
instead of the K-factors themselves. In the tables, the relative K-factor
is
K
actual
/KN2 where the reference gas is the commonly used gas, nitrogen
(N2). The remaining columns give Cp and ρ, enabling you to calculate
K1/K2 directly using Equation (6). In some instances, K1/K2 from the
tables may be different from that which you calculate directly. The
value from the tables is preferred because in many cases it was obtained by experiment. Sierra calibrates every transducer with primary
standards using the actual gas or a molecular equivalent reference gas.
The calibration certificate accompanying the transducer cites the reference gas used.
Example 1:
A transducer is calibrated for nitrogen (N2), and the flow rate is
1000 sccm for a 5.000 VDC output signal. The flow rate for carbon dioxide at a 5.000 VDC output is:
Q
CO2/QN2
= K
CO2/K N2
, or
Q
CO2
= (0.74/1.000)1000 = 740 sccm
Example 2:
A transducer is calibrated for hydrogen (H2), and the flow rate is
100 sccm for a 5.000 VDC output signal. The flow rate for nitrous
oxide (N2O) is found as follows:
Q
N2O/QH2
= K
N2O/K H2
, or
Q
N2O
= (0.71/1.01) 100 = 70.3 sccm
Note that the K-factors relative to nitrogen must be used in each case.
Example 3:
We want a transducer to be calibrated for use with dichlorosilane
(SiH2Cl2) at a 100 sccm full scale flow. We wish to use the preferred reference gas Freon-14 (CF4). What flow of CF4 must we
generate to do the calibration?
Equation (6) is used for gas mixtures, but we must calculate N/ρCp
for the mixture. The equivalent values of ρ, Cp, and N for a dual
gas mixture are given as follows:
The equivalent gas density is:
m
•
!
= ( / ) + ( / )
m
•
1T
!
1
m
•
2
m
•
T!2
Where:
•
T
m = m
•
1
m
•
2
+= Total mass flow rate (gm/min),
( )1 = Refers to gas #1, and
( )2 = Refers to gas #2
The equivalent specific heat is:
Cp = F1Cp1 + F2Cp2
Where:
1
m
•
!
m
•
1!1
m
•
2
m
•
T
!
2
F = ( )/( ) and
T
!
F = ( )/( )
2
The equivalent value of N is:
m
•
! = ( / ) ! + ( / ) !m
•
1T1
m
•
2
m
•
T2
The equivalency relationships for ρ, Cp, and N for mixtures of
more than two gases have a form similar to the dual-gas relationship given above.
IMPORTANT NOTE ABOUT K-FACTORS:
Please note that if you have a transducer calibrated for a gas such
as methane and wish to use the K-factors to measure a gas such as
air, that the inaccuracy of the measurement can range from ±5 to
10%. The use of K-factors is, at best, only a roughapproximation
and should not be used in applications that require a better than ±5
to 10% accuracy.
It should also be noted that certain gases, in similar “families,” will
work exceptionally well with K-factors; however, those instances
are only true when similar thermal properties of the gas are present.
Appendix B Specifications Series 830/840/860 Instruction Manual
Control Range 840/860 (all sizes).. 2 to 100% of full scale; valve shuts between 1 to 2.5% of full
840 (NPT) .............. 10 to 100% of full scale
Command Signal 0-5 VDC, 20 megaohms minimum input impedance
4-20 mA, 250 Ohms maximum input impedance, optional RS-485 digi-
Output Signal Linear 0-5 VDC, 1000 Ohms minimum load resistance
Linear 4-20 mA, 500 Ohms maximum loop resistance
scale
tal signal standard on Model 860C
Performance Specifications
Accuracy ±1% of full scale including linearity over 15 to 25°C and 10 to 60 psia
For the models 840 (NPT), the accuracies above are 1.5%, ±3%, and
Repeatability ±0.15% of full scale
±0.5% of full scale for 840 (NPT)
Temperature Coefficient 0.08% of full scale per °F (0.15% of full scale per °C), or better
Pressure Coefficient 0.01% of full scale per psi (0.15% of full scale per bar), or better
Response Time 830 (all) .................. 300 ms time constant; 2 seconds (typical) to within ±2%
of set point over 20 to 100% of full scale
840 (low, med) ....... 300 ms time constant; 2 seconds (typical) to within ±2%
of set point over 20 to 100% of full scale
840 (high)............... 600 ms time constant; 4 seconds (typical) to within ±2%
of set point over 20 to 100% of full scale
840 (NPT) .............. 800 ms time constant; 4 seconds (typical) to within ±5%
of set point over 20 to 100% of full scale
860 (all).................. 0.6 to 1 second to within ±2% of set point over 20 to 100%
of full scale
(0.7 to 4 bara); ±2% of full scale including linearity over 0 to 50°C and 5
to 50 psia (0.3 to 10 bara); ±1% of full scale accuracy at a specific temperature and pressure is available with special calibration;
Series 830/840/860 Instruction ManualAppendix B Specifications
Transducer Mounting Dimensions
All dimensions are inches, millimeters are in parentheses. Certied drawings are available on request. Shaded portion of dimensional drawings indicates controller models.