Sierra 820 User Manual

Series 820 Instruction Manual Table of Contents
Sierra 820 Series
Top-Trakª Mass Flow Meters
Instruction Manual
Part Number IM-82
5 Harris Court, Building L Monterey, CA 93940
(831) 373-0200 (800) 866-0200 Fax (831) 373-4402
http://www.sierrainstruments.com
Sierra Instruments b.v. Bolstoen 30A 1046 AV Amsterdam The Netherlands
+31(0)20-6145810 Fax +31(0)20-6145815
IM-82-C 0-1
Table of Contents Series 820 Instruction Manual
Customer Notice
Sierra Instruments, Inc. is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instruments standard mass flow meters or control­lers for oxygen gas. You are responsible for determining if this mass flow meter or controller is appropriate for your oxygen application. You are responsible for cleaning the mass flow meter or controller to the degree required for your oxygen flow application.
© COPYRIGHT SIERRA INSTRUMENTS 1994 No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change without notice.
TRADEMARKS Top-Trakª and Cal-Benchª are trademarks of Sierra Instruments, Inc. Other product and company names listed in this manual are trademarks or trade names of their respective manufacturers.
All Sierra products are Year 2000 compliant.
0-2 IM-82-C
Series 820 Instruction Manual Table of Contents

Table of Contents

Chapter 1 Introduction
Introduction.............................................................1-1
Using this Manual .................................................1-1
Safety Information.................................................1-2
Receipt of System Components..................................1-2
Technical Assistance ..............................................1-2
Top Trak Features......................................................1-3
The 820 Series Flow Sensing Principle .............................1-4
Chapter 2 Installation
Installation Overview ..................................................2-1
Installing the Transducer ..............................................2-2
Compression Fittings .............................................2-2
VCO Fittings .......................................................2-2
VCR Fittings .......................................................2-3
All 1/2-inch Size Connections....................................2-3
Wiring the Transducer.................................................2-4
Standard 0-5 VDC Output Signal Wiring.......................2-5
Optional 4-20 mA Output Signal Wiring........................2-5
Remote Display Installation...........................................2-6
Chapter 3 Operation
Transducer Operation..................................................3-1
Transducer Accuracy..............................................3-1
Referencing the Transducer to Non-Standard Conditions ....3-2
Transducer Over-Ranging ........................................3-2
Zero and Span Adjustments ......................................3-3
Chapter 4 Maintenance and Repair
Transducer Cleaning...................................................4-1
Flow Path Cleaning Model 822/824 .................................4-2
Inlet and Outlet Screens...........................................4-2
Laminar Flow Element............................................4-3
Flow Path Cleaning Model 826/827 .................................4-4
Laminar Flow Element............................................4-4
Flow Path Cleaning Model 822-S/824-S............................4-5
Laminar Flow Element............................................4-5
Sensor Cleaning and Inspection .................................4-8
Transducer Calibration ................................................4-9
Transducer Troubleshooting............ . . . . . . ...................... 4-11
Returning Equipment to the Factory ............................... 4-12
IM-82-C 0-3
Table of Contents Series 820 Instruction Manual
Appendix A Conversion Formulas and Gas Tables
Appendix B Production Specifications
List of Figures
1-1. Top-Trak Features (Typical)..................................1-3
1-2. Flow Paths through the Transducer..........................1-4
1-3. Flow Measuring Principle.....................................1-4
2-1. Piping Requirements for 1/2-inch Size Connections.......2-3
2-2. Transducer D-Connector Pin Assignments .. . . .............2-4
2-3. Standard 0-5 VDC Output Signal Wiring ...................2-5
2-4. Single Transducer Current Loop Connection...............2-5
2-5. Multiple Transducer Current Loop Connections ...........2-6
2-6. Mounting the Remote Display ................................2-6
4-1. Model 822/824 Flow Components.................. . . . . . . . . . 4-2
4-2. Correct LFE Position ..........................................4-3
4-3. Model 826/827 Flow Components.................. . . . . . . . . . 4-4
4-4. Low Flow Transducer LFE Cleaning........................4-5
4-5. Medium Flow Transducer LFE Cleaning ...................4-6
4-6. High Flow Transducer LFE Cleaning .......................4-7
4-7. Printed Circuit Board Component Locations ............. 4-10
0-4 IM-82-C
Series 820 Instruction Manual Table of Contents
Cautions
Caution! Only qualified personnel should install the transducer.
Caution! Do not supply +DC power at the D-connector while using a power supply at
the power jack. Both supplies may be damaged.
Caution! Operating a 12 VDC transducer at 24 VDC will cause equipment damage.
Caution! Only qualified personnel should perform transducer service, calibration or
troubleshooting procedures.
Caution! When using toxic or corrosive gases, purge the unit thoroughly with inert dry gas before disconnecting from the gas line.
Caution! Printed circuit boards are sensitive to electrostatic discharge. To avoid dam­aging the board, follow these precautions to minimize the risk of damage:
· before handling the assembly, discharge your body by touching a grounded,
metal object
· handle all cards by their edges unless otherwise required
· when possible, use grounded electrostatic discharge wrist straps when handling
sensitive components
IM-82-C 0-5
Table of Contents Series 820 Instruction Manual
0-6 IM-82-C
Series 820 Instruction Manual Chapter 1 Introduction

Chapter 1 Introduction

This instruction manual covers the installation, operation and main­tenance of SierraÕs 820 Series product line including the following Top-Trakª Models:
· 822 Mass Flow Meter with display (nylon flow body)
· 824 Mass Flow Meter without display (nylon flow body)
· 826 Hi-Flow Meter with display (aluminum flow body)
· 827 Hi-Flow without display (aluminum flow body)
· 822-S Mass Flow Meter with display (stainless steel flow body)
· 824-S Mass Flow Meter without display (stainless steel flow body)
SierraÕs Top-Trak measurement of gas mass flow. The 820 Series offers a broad range of sizes and process connections for flexibility and versatility. The primary standard calibration ensures starting point accuracy and NIST traceability. The meterÕs 0-5 VDC or 4-20 mA output signal is provided for recording, data-logging or control. The optional dis­play reads the mass flow rate directly in engineering units or per­centage of full scale.
Using This Manual
This manual is organized into four chapters:
· Chapter 1 includes the introduction and theory of operation
· Chapter 2 provides installation and wiring instructions
· Chapter 3 describes transducer operation and features
· Chapter 4 covers maintenance, calibration and troubleshooting
Gas tables and conversion formulas are found in Appendix A. The product specifications and dimensional drawings are found in Appendix B.
Throughout this manual, we use the word transducer as a generic term to represent all models of SierraÕs 820 Series Top-Trak Mass Flow Meters.
ª
Mass Flow Meters are designed for precise
IM-82-C 1-1
Chapter 1 Introduction Series 820 Instruction Manual
Safety Information
Caution and warning statements are used throughout this book to draw your attention to important information.
Warning! Caution!
This statement appears with information that is important to protect people and equipment from damage. Pay very close attention to all warnings that apply to your application.
This statement appears with information that is important for protecting your equipment and per­formance. Read and follow all cautions that ap­ply to your application.
Receipt of System Components
When receiving a Sierra transducer, carefully check the outside packing carton for damage incurred in shipment. If the carton is damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered com­ponents are present and match your specifications (as ordered). Make sure any spare parts or accessories are not discarded with the packing material. Do not return any equipment to the factory without first contacting Sierra Customer Service.
Technical Assistance
If you encounter a problem with your transducer, review the configu­ration information for each step of the installation, operation and set up procedures. Verify that your settings and adjustments are consis­tent with factory recommendations. Refer to Chapter 4, Trouble­shooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 4, contact Sierra Instruments by fax or by E-mail (see inside front cover). For urgent phone support you may call (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00 p.m. PST. In Europe contact Sierra Instruments bv at +31 20 6145810. When contacting Technical Support, make sure to include this information:
· the flow range, serial number, Sierra order number and
model number (all marked on the transducer nameplate)
· the problem you are encountering and any corrective action
taken
· application information (gas, pressure, temperature, pipe and
fitting configuration)
1-2 IM-82-C
Series 820 Instruction Manual Chapter 1 Introduction
Top-Trak Features
Standard Top-Trak Mass Flow Meters require a 12 to 15 VDC exter­nal power source (24 VDC input power optional). The transducerÕs 0 to 5 VDC output signal allows for flow recording, data-logging or control. A 4 to 20 mA output signal is optionally available. Input power and output signal connections are made via the 9-pin sub-type D-connector located on the side of the transducer. An additional input power jack is located just below the D-connector. (It is important to connect input power at only one location.)
The transducer shown below is a typical example of a 822 Series Top-Trak Mass Flow Meter. Other models may vary slightly in their appearance but are operationally equivalent.
Figure 1-1. Top-Trak Features (Typical)
IM-82-C 1-3
Chapter 1 Introduction Series 820 Instruction Manual
The 820 Series Flow Sensing Principle
The operating principle of Top-Trak transducers is based on heat transfer and the first law of thermodynamics. During operation process gas enters the instrumentÕs flow body and divides into two flow paths, one through the sensor tube, the other through the lami­nar flow element bypass. The laminar flow bypass generates a pres­sure drop, P1ÐP2, forcing a small fraction of the total flow to pass through the sensor tube (m1).
Figure 1-2. Flow Paths through the Transducer
Two resistance temperature detector (RTD) coils around the sensor tube direct a constant amount of heat (H) into the gas stream. In ac­tual operation, the gas mass flow carries heat from the upstream coil to the downstream coil. The resulting temperature difference (DT) is detected by the RTD coils and gives the output signal. Since the molecules of the gas carry away the heat, the output signal is linearly proportional to gas mass flow.
Figure 1-3. Flow Measuring Principle
1-4 IM-82-C
Series 820 Instruction Manual Chapter 2 Installation

Chapter 2 Installation

Installation Overview
To ensure a successful installation, inlet and outlet tubing should be clean and free from burrs or rims caused by cutting prior to plumb­ing the transducer into the system. The protective caps covering the inlet/outlet fittings should not be removed until immediately prior to installation.
Before installing the transducer, verify the following:
1. Make sure the installation site meets the specific operating pa-
rameters recorded on the transducer’s nameplate. Each trans-
ducer is factory-configured for a specific gas and flow range. If
the operating pressure is more than 50 psi (3.4 bar) away from
the calibration pressure, it is advisable to return the unit to the
factory for re-calibration. (Adjusting zero may be sufficient to
remain within specification.)
2. Do not locate the transducer in areas subject to sudden tempera-
ture changes, moisture, drafts or near equipment radiating sig-
nificant amounts of heat. Make sure to allow adequate space for
cable connectors and wiring.
3. For 1/2-inch size inlet/outlet process connections on models 826
and 827 make sure the location meets the minimum number of
recommended pipe diameters upstream and downstream of the
transducer. A minimum of 5 inches (127 mm) upstream and 2-
1/2 inches (64 mm) downstream is always recommended. (not
necessary for other models)
4. Horizontal mounting is preferable. Vertical mounting is possible
with best results achieved when the factory calibration is specifi-
cally performed for vertical mounting. In vertical positions zero
shift will occur depending on the gas pressure at zero flow.
5. If the gas contains any particulate matter, install an in-line filter
prior to the transducer. Recommended filter size: 15 micron for
flows of 10 sccm to 30 slpm, 30 micron for above 30 slpm.
6. If a potential over-flow condition exists, insert a valve or critical
orifice in the line to limit flow to approximately 25 percent above
the full scale range of the meter.
7. Confirm that the transducer o-ring material is compatible with
the gas to be measured.
8. For remote displays, confirm the supplied cable is of sufficient
length to connect the components.
IM-82-C 2-1
Chapter 2 Installation Series 820 Instruction Manual
Caution!
Only qualified
personnel should install
the transducer.
Installing the Transducer
Follow the installation instructions that apply to your transducer’s process connection. For all 1/2-inch size process connections, observe the piping recommendations given on page 2-3. Before operation, all plumbing should be checked carefully for leaks and the transducer purged with dry nitrogen.
Compression Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
1. Verify the position of the front
and back ferrule. Insert the tubing into the fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger tight. (Do not mix or inter­change parts of tube fittings made by different manufacturers.)
2. Hold the body steady with a backup wrench. For 1/2-
inch size, tighten the nut 1-1/4 turns from finger tight. For 1/8­inch, 1/4-inch and 3⁄8-inch sizes, tighten only 3/4 turn from fin­ger tight. Do not over-tighten!
3. Check the system’s entire flow path thoroughly for leaks. (Do
not use liquid leak detectors, instead monitor pressure decay. Over-exposing the transducer to leak detector fluid may damage the unit.)
VCO and VCR Fittings
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
2. Install new o-rings compatible with the gas to be used. (Do not
mix or interchange parts of tube fittings made by different manu­facturers.)
3. Hold the body steady with a backup wrench. Tighten
the nut finger tight and then 1/4 turn tighter with a wrench. Do
not over-tighten!
4. Check the system’s entire flow path thoroughly for leaks. (Do not
use liquid leak detectors, instead monitor pressure decay. Over­exposing the transducer to leak detector fluid may damage the unit.)
2-2 IM-82-C
Series 820 Instruction Manual Chapter 2 Installation
1/4 Inch Female NPT (standard on nylon flow bodies)
1. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
2. Use a good quality paste pipe thread sealant.
Apply to the pipe threads.
3. Tighten the pipe no more than 1 turn past hand-tight.
Caution! Do not over-tighten,
damage to the instrument may result.
4. Check the system’s entire flow path thoroughly for leaks. (Do not
use liquid leak detectors, instead monitor pressure decay. Over-
exposing the transducer to leak detector fluid may damage the unit.)
1/2-Inch Size NPT Connections (Models 826,827 only)
1. Install a section of straight pipe at least ten pipe diameters in length
upstream of the transducer. Also, allow at least five pipe diameters
downstream for accurate operation. DO NOT use reducers. If the
preceeding components in the flow path create disturbances extend
the upstream pipe length.
2. Position the transducer with the flow direction arrow pointing
downstream in the direction of flow.
3. Tighten fittings until leak tight (refer to published standards for
specific recommendations).
4. Check the system’s entire flow path thoroughly for leaks. (Do not
use liquid leak detectors, instead monitor pressure decay. Over-
exposing the transducer to leak detector fluid may damage the unit.)
Figure 2-1.
Piping Requirements for all 1/2-Inch Size Process Connections
IM-82-C 2-3
Chapter 2 Installation Series 820 Instruction Manual
Caution!
Do not supply +DC power
at the D-connector while
using a power supply at the
power jack. Both supplies
may be damaged.
Wiring the Transducer
Standard Top-Trak™ transducers require a 12 to 18 VDC power supply (15 VDC nominal, 100 mA maximum). 24 VDC input power is optional. Transducers are connected to the power supply through either the dedicated DC power jack or through the 9-pin D-connector located on the side of the enclosure. Before powering the unit, check the transducer’s nameplate to confirm input power:
PV1 = 12 to 18 VDC
PV2 = 24 VDC
Note: operating a 24 VDC transducer at 12 to 18 VDC will result in unreliable operation.
Caution!
Operating a 12 VDC trans-
ducer at 24 VDC will cause
equipment damage.
The transducer’s standard 0 to 5 VDC (4-20 mA optional) output signal is available through the D-connector. The mating connector is included with the transducer. Connection details are given on the following pages.
When the transducer is configured for a remote display, signal con­nections are made via the 9-pin connector. Input power connections are not included in the standard interface cable. Remote display mounting dimensions are given at the end of this chapter.
Figure 2-2. Transducer D-Connector Pin Assignments
2-4 IM-82-C
Loading...
+ 31 hidden pages