Sierra 221, 220 User Manual

Sierra Series 220 and 221
Innova-Flo™ Vortex Flow Meters
Instruction Manual
Part Number IM-22
05/03 Revision C
IMPORTANT: This manual is for use with 220/221 Vortex Products
CORPORATE HEADQUARTERS
5 Harris Court, Building L Monterey, CA 93940
Phone (831) 373-0200 (800) 866-0200 Fax (831) 373-4402
Bijlmansweid 2 1934RE Egmond a/d Hoef The Netherlands
Phone +31 72 5071400 Fax +31 72 5071401
Tomson Centre, Rm. A618, 188 Zhang Yang Road
Phone: + 8621 5879 8521 Fax: +8621 5879 8586
www.sierrainstruments.com
EUROPE HEADQUARTERS
ASIA HEADQUARTERS
Pu Dong New District, Shanghai, P.R.China
Table of Contents Series 220/221 Instruction Manual
Customer Notice
Unless you have specifically ordered Sierra’s optional O2 cleaning, this flow meter may not be fit for oxygen service. Sierra Instruments, Inc., is not liable for any damage or personal injury, whatsoever, resulting from the use of Sierra Instru­ments standard mass flow meters for oxygen gas. You are responsible for clean­ing the mass flow meter to the degree required for your oxygen flow application.
© COPYRIGHT SIERRA INSTRUMENTS 1997 No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission of Sierra Instruments. The information contained in this manual is subject to change without notice.
TRADEMARKS Innova-Flo™ is a trademark of Sierra Instruments, Inc. Other product and company names listed in this manual are trademarks or trade names of their respective manufacturers.
Revision B. Added CE required wiring information. Improved insertion graphics. Added in­formation on: wafer-style installation, adjusting meter orientation, remote sensor wiring and component field replacement procedures.
Revision C. Changed to new style 220 body and adapter. Removed 4-20 A/D counts from calibration menu. Various minor updates.
0-2 IM-22-C
Series 220/221 Instruction Manual Table of Contents
Table of Contents

Chapter 1 Introduction

Innova-Flo™ Vortex Flow Meters........................................................ 1-1
Using this Manual .........................................................................1-1
Note and Safety Information ......................................................... 1-2
Receipt of System Components ....................................................1-2
Technical Assistance ..................................................................... 1-2
How the Innova-Flo Vortex Flow Meter Operates .............................. 1-3
Velocity Measurement/Pressure Drop........................................... 1-3
Flow Meter Configurations.................................................................. 1-7

Chapter 2 Installation

Installation Overview........................................................................... 2-1
Flow Meter Installation Requirements.......................................... 2-1
Unobstructed Flow Requirements................................................. 2-2
Series 220 In-Line Flow Meter Installation......................................... 2-3
Wafer-Style Flow Meter Installation............................................. 2-4
Flange-Style Flow Meter Installation............................................ 2-5
Series 221 Insertion Flow Meter Installation....................................... 2-6
Cold Tap Guidelines...................................................................... 2-7
Hot Tap Guidelines .......................................................................2-8
Flow Meter Insertion ...........................................................................2-9
Installing Meters with a Compression Connection .....................2-10
Installing Meters with a Packing Gland Connection................... 2-12
Installing Meters (Packing Gland), No Insertion Tool................ 2-15
Adjusting Meter Orientation.............................................................. 2-17
Display/Keypad Adjustment ....................................................... 2-17
Enclosure Adjustment ................................................................. 2-18
Wiring Connections ........................................................................... 2-19
Input Power Connections ............................................................ 2-19
Pulse Output Connections ...........................................................2-21
Remote Electronics Wiring .........................................................2-22

Chapter 3 Operating Instructions

Flow Meter Display/Keypad................................................................ 3-1
Start Up................................................................................................ 3-2
Using the Setup Menus........................................................................ 3-3
Programming the Flow Meter ....................................................... 3-3
Output Menu ................................................................................. 3-4
Display Menu ................................................................................3-5
Totalizer Menu ..............................................................................3-6
Units Menu.................................................................................... 3-7
Diagnostics Menu.......................................................................... 3-8
Calibration Menu .......................................................................... 3-9
Password Menu ........................................................................... 3-10
IM-22-C 0-3
Table of Contents Series 220/221 Instruction Manual

Chapter 4 Troubleshooting and Repair

Hidden Diagnostics Menus.................................................................. 4-1
Column One Hidden Diagnostics Values...................................... 4-2
Column Two Hidden Diagnostics Values ..................................... 4-2
Analog Output Calibration .................................................................. 4-3
Troubleshooting the Flow Meter .........................................................4-3
Symptom: Output at no Flow ........................................................ 4-3
Symptom: Erratic Output .............................................................. 4-3
Symptom: No Output ....................................................................4-4
Electronics Assembly Replacement..................................................... 4-5
Returning Equipment to the Factory.................................................... 4-6

Appendix A Product Specifications

Appendix B Glossary

List of Figures

1-1. Series 220 In-Line Vortex Flow Meter ................................... 1-3
1-2. Measurement Principle of Vortex Flow Meters...................... 1-4
1-3. Reynolds Number Range for the Innova-Flo.......................... 1-6
2-1. Recommended Pipe Length Requirements ............................. 2-2
2-2. Flange Bolt Torquing Sequence ............................................. 2-3
2-3. Wafer-Style Flow Meter Installation ......................................2-4
2-4. Flange-Style Flow Meter Installation ..................................... 2-5
2-5. Hot Tap Sequence ................................................................... 2-8
2-6. Insertion Calculation (Compression Type) ........................... 2-10
2-7. Flow Meter with Compression Type Fitting......................... 2-11
2-8. Insertion Calculation (Meters with Insertion Tool) ..............2-12
2-9. Flow Meter with Permanent Insertion Tool.......................... 2-13
2-10. Flow Meter with Removable Insertion Tool......................... 2-14
2-11. Insertion Calculation (Meters without Insertion Tool) .........2-15
2-12. Display/Keypad Viewing Adjustment ..................................2-17
2-13. Enclosure Viewing Adjustment ............................................ 2-18
2-14. DC Power Connections......................................................... 2-19
2-15. Load Resistance Versus Input Voltage................................. 2-20
2-16. Pulse Output with External Power Supply ...........................2-21
2-17. Junction Box Sensor Connections ........................................ 2-22
3-1. Flow Meter Display/Keypad................................................... 3-1

List of Tables

2-1. Minimum Stud Bolt Lengths for Wafer-Style Meters ............2-3
0-4 IM-22-C
Series 220/221 Instruction Manual Table of Contents

Warnings and Cautions

Warning!
Agency approval for hazardous location installations varies between flow meter models. Consult the factory for specific flow meter approvals before any hazardous location instal­lation. For explosion proof installations, you must use solid metal conduit and follow NFPA (or your local code) approved wiring methods.
Hot tapping must be performed by a trained professional. U.S. regulations often require a hot tap permit. The manufacturer of the hot tap equipment and/or the contractor perform­ing the hot tap is responsible for providing proof of such a permit.
All flow meter connections, isolation valves and fittings for cold/hot tapping must have the same or higher pressure rating as the main pipeline.
For Series 221 insertion flow meter installations, an insertion tool must be used for any installation where a flow meter is inserted under pressure greater than 50 psig.
To avoid serious injury, DO NOT loosen a compression fitting under pressure.
All wiring procedures must be performed with the power Off.
Before attempting any flow meter repair, verify that the line is de-pressurized.
Always remove main power before disassembling any part of the mass flow meter.
Caution!
Calibration must be performed by qualified personnel. Sierra Instruments, Inc., strongly rec­ommends that you return your flow meter to the factory for calibration.
In order to achieve accurate and repeatable performance, the flow meter must be in­stalled with the specified minimum length of straight pipe upstream and downstream of the flow meter’s sensor head.
When using toxic or corrosive gases, purge the line with inert gas for a minimum of four hours at full gas flow before installing the flow meter.
Wafer-style flow meter gaskets must be carefully aligned to ensure accurate flow meas­urement.
For Series 221 insertion flow meter installations, the sensor alignment pointer must point downstream in the direction of flow.
IM-22-C 0-5
Table of Contents Series 220/221 Instruction Manual
0-6 IM-22-C
Series 220/221 Instruction Manual Chapter 1 Introduction
Chapter 1 Introduction

Innova-Flo™ Vortex Flow Meters

The Sierra Instruments’ Series 220 In-Line and the Series 221 Insertion Innova-Flo™ Vortex Flow Meters provide a reliable solution for process flow measurement. From a single entry point in the pipeline, Innova-Flo meters offer precise measurements of mass or volumetric flow. The ve­locity sensor reduces the effects of pipeline vibration by incorporating a unique piezoelectric element that senses the vortex frequency. To extend rangeability at the low end of flow, the meter’s smart electronics calcu­lates the Reynolds number (Re) based on constant values of fluid density and viscosity stored in memory and automatically corrects for any non­linearity down to Re = 5,000.
Innova-Flo digital electronics allows reconfiguration for most gases, liq­uids and steam. The instrument is loop powered (12 to 36 VDC) with two output signals. The pulse output signal is proportional to volumetric flow rate; the analog linear 4-20 mA signal offers your choice of volu­metric flow rate or mass flow rate. The mass flow rate is based on a con­stant value for fluid density stored in the instrument’s memory. The local keypad/display provides instantaneous flow rate in engineering units or totalized flow.
The Sierra Series 220 and 221 Innova-Flo Meters simple installation combines with an easy-to-use interface that provides quick set up, long term reliability and accurate flow measurement over a wide range of flows and conditions.

Using This Manual

This manual provides the information needed to install and operate the Series 220 In-Line and Series 221 Insertion Innova-Flo Vortex Flow Me­ters. The four chapters of this manual cover these areas:
Chapter 1 includes the introduction and product description
Chapter 2 provides information needed for installation
Chapter 3 describes system operation and programming
Chapter 4 covers troubleshooting and repair
The product specifications are found in Appendix A. Appendix B con­tains a glossary of terms.
IM-22-C 1-1
Chapter 1 Introduction Series 220/221 Instruction Manual

Note and Safety Information

We use note, caution and warning statements throughout this book to draw your attention to important information.
Warning!
This statement appears with information that is important to protect people and equipment from damage. Pay very close attention to all warnings that apply to your application.
Caution!
This statement appears with information that is important for protecting your equipment and performance. Read and follow all cautions that apply to your application.

Receipt of System Components

When receiving a Sierra flow meter, carefully check the outside packing carton for damage incurred in shipment. If the carton is damaged, notify the local carrier and submit a report to the factory or distributor. Remove the packing slip and check that all ordered components are present. Make sure any spare parts or accessories are not discarded with the packing material. Do not return any equipment to the factory without first con­tacting Sierra Customer Service.

Technical Assistance

If you encounter a problem with your flow meter, review the configura­tion information for each step of the installation, operation and set up pro­cedures. Verify that your settings and adjustments are consistent with fac­tory recommendations. Refer to Chapter 4, Troubleshooting, for specific information and recommendations.
If the problem persists after following the troubleshooting procedures outlined in Chapter 4, contact Sierra Instruments, Technical Support at (800) 866-0200 or (831) 373-0200 between 8:00 a.m. and 5:00 p.m. PST. When calling Technical Support, have the following information on hand:
the flow range, serial number and Sierra order number (all marked on the meter nameplate)
the problem you are encountering and any corrective action taken
application information (gas, pressure, temperature and pip­ing configuration)
Note
This statement appears with a short message to alert you to an important detail.
1-2 IM-22-C
Series 220/221 Instruction Manual Chapter 1 Introduction

How the Innova-Flo Vortex Flow Meter Operates

Figure 1-1. Series 220 In-Line Vortex Flow Meter
Sierra Series 220 and 221 Innova-Flo™ Vortex Flow Meters use a unique velocity sensor head to monitor volumetric flow rate. The built-in flow computer calculates mass flow rate based on a constant value of fluid density stored in the instrument’s memory. To measure fluid velocity, the flow meter incorporates a bluff body (shedder bar) in the flow stream, and the velocity sensor measures the frequency of vortices cre­ated by the shedder bar. The velocity sensor head is located downstream of the shedder bar within the flow body.

Velocity Measurement

The Innova-Flo vortex velocity sensor is a patented mechanical design that minimizes the effects of pipeline vibration and pump noise, both of which are common error sources in flow measurement with vortex flow meters. The velocity measurement is based on the well-known Von Kar­man vortex shedding phenomenon. Vortices are shed from a shedder bar, and the vortex velocity sensor located downstream of the shedder bar senses the passage of these vortices. This method of velocity measurement has many advantages including inherent linearity, high turndown, reliabil­ity and simplicity.
IM-22-C 1-3
Chapter 1 Introduction Series 220/221 Instruction Manual
Vortex Shedding Frequency
Von Karman vortices form downstream of a shedder bar into two distinct wakes. The vortices of one wake rotate clockwise while those of the other wake rotate counterclockwise. Vortices generate one at a time, al­ternating from the left side to the right side of the shedder bar. Vortices interact with their surrounding space by over-powering every other nearby swirl on the verge of development. Close to the shedder bar, the distance (or wave length) between vortices is always constant and meas­urable. Therefore, the volume encompassed by each vortex remains con­stant, as shown below. By sensing the number of vortices passing by the velocity sensor, the Innova-Flo
Vortex Flow Meter computes the total
fluid volume.
Velocity sensor
Vortex shedder bar
Flow
Constant wave length
Vortices
Figure 1-2. Measurement Principle of Vortex Flow Meters
Vortex Frequency Sensing
The velocity sensor incorporates a piezoelectric element that senses the vortex frequency. This element detects the alternating lift forces pro­duced by the Von Karman vortices flowing downstream of the vortex shedder bar. The alternating electric charge generated by the piezoelec­tric element is processed by the transmitter’s electronic circuit to obtain the vortex shedding frequency. The piezoelectric element is highly sensi­tive and operates over a wide range of flows, pressures and temperatures.
1-4 IM-22-C
Series 220/221 Instruction Manual Chapter 1 Introduction
Flow Velocity Range
To ensure trouble-free operation, vortex flow meters must be correctly sized so that the flow velocity range through the meter lies within the measurable velocity range (with acceptable pressure drop) and the linear range.
The measurable range is defined by the minimum and maximum velocity using the following table.
Vmin Vmax 300 ft/s 30 ft/s
Vmin Vmax 91 m/s 9.1 m/s
Gas Liquid
37
25
ρ
ft/s
ρ
m/s
1 ft/s
0.3 m/s
English
Metric
ρ
(lb/ft3)
ρ
(kg/m3)
The pressure drop for series 221 insertion meters is negligible. The pres­sure drop for series 220 in-line meters is defined as:
2
P = .00024 ρ V P = .000011 ρ V
English units (P in psi, ρ in lb/ft3, V in ft/sec)
2
Metric units (P in bar, ρ in kg/m3, V in m/sec)
The linear range is defined by the Reynolds number. The Reynolds num­ber is the ratio of the inertial forces to the viscous forces in a flowing fluid and is defined as:
Where
Re =
ρ V D
µ
Re = Reynolds Number ρ = mass density of the fluid being measured V = velocity of the fluid being measured D = internal diameter of the flow channel µ = viscosity of the fluid being measured
The Strouhal number is the other dimensionless number that quantifies the vortex phenomenon. The Strouhal number is defined as:
Where
St = Strouhal Number f = frequency of vortex shedding d = shedder bar width V = fluid velocity
IM-22-C 1-5
St =
f d
V
Chapter 1 Introduction Series 220/221 Instruction Manual
As shown below, Innova-Flo™ Vortex Flow Meters exhibit a constant Strouhal number across a large range of Reynolds numbers, indicating a consistent linear output over a wide range of flows and fluid types. Be­low this linear range, the intelligent electronics in Innova-Flo automati­cally corrects for the variation in the Strouhal number. Innova-Flo’s smart electronics correct for this non-linearity by calculating the Rey­nolds number based on constant values of the fluid’s density and viscos­ity stored in the instrument’s memory. Innova-Flo Vortex Flow Meters automatically correct down to a Reynolds number of 5,000.
0.3
Corrected range
Linear range
0.2
0.1
Strouhal Number, St
0.0
34
10
10
5000
5
10 10
Reynolds Number, Re
10
6
10
7
8
Figure 1-3. Reynolds Number Range for the Innova-Flo Meter
1-6 IM-22-C
Series 220/221 Instruction Manual Chapter 1 Introduction

Flow Meter Configurations

Innova-Flo™ Vortex Flow Meters are available in two configurations:
Series 220 in-line flow meter (replaces a section of the pipeline)
Series 221 insertion flow meter (requires a “cold” tap or a “hot” tap
into an existing pipeline)
Both the in-line and insertion configurations are similar in that they both use identical electronics and have similar sensor heads. Besides installa­tion differences, the main difference between an in-line flow meter and an insertion flow meter is their method of measurement.
For an in-line vortex flow meter, the shedder bar is located across the en­tire diameter of the flow body. Thus, the entire pipeline flow is included in the vortex formation and measurement. The sensing head, which di­rectly measures velocity is located just downstream of the shedder bar.
An insertion vortex flow meter has its sensing head at the end of a 0.750 inch diameter tubular stem. The stem is inserted into the pipe until the sensing head is properly located in the pipe’s cross section. The sensing head fits through any entry port with an 1.875 inch minimum internal di­ameter.
The sensing head of an insertion vortex flow meter directly monitors the velocity at a point in the cross-sectional area of a pipe, duct, or stack (re­ferred to as “channels”). The velocity at a point in the pipe varies as a func­tion of the Reynolds number. The insertion vortex flow meter computes the Reynolds number based on constant values of the fluid’s density and vis­cosity stored in its memory and then computes the total flow rate in the channel. The output signal of insertion meters is the total flow rate in the channel. The accuracy of the total flow rate computation depends on ad­herence to the piping installation requirements given in Chapter 2. If ad­herence to those guidelines cannot be met, contact the factory for specific installation advice.
Flow Meter Electronics
Innova-Flo electronics are available mounted directly to the flow body, or remotely mounted. The electronics housing may be used indoors or outdoors, including wet environments. The instrument requires 4-20 mA loop power (12 to 36 VDC). One analog output signal is available for your choice of volumetric flow rate or mass flow rate. A pulse output is available for totalization.
IM-22-C 1-7
Chapter 1 Introduction Series 220/221 Instruction Manual
The meter includes a local 2 x 16 character LCD display housed within the enclosure. Local operation and reconfiguration is accomplished using six push buttons. For hazardous locations, the six push buttons can be operated through the sealed enclosure using a hand-held magnet, thereby not compromising the integrity of the hazardous location certification.
The electronics include nonvolatile memory that stores all configuration information. The memory allows the flow meter to function immediately upon power up, or after an interruption in power.
1-8 IM-22-C
Series 220/221 Instruction Manual Chapter 2 Installation
Warning!
Consult the flow meter name-
plate for specific flow meter
approvals before any hazard-
ous location installation.
Chapter 2 Installation

Installation Overview

Innova-Flo™ meter installations are simple and straightforward. Both the Se­ries 220 In-Line and Series 221 Insertion type flow meter installations are covered in this chapter. After reviewing the installation requirements given below, see page 2-3 for Series 220 installation instructions. See page 2-6 for Series 221 installation instructions. Wiring instructions begin on page 2-19.

Flow Meter Installation Requirements

Before installing the flow meter, verify the installation site allows for these considerations:
1. Line pressure and temperature will not exceed the flow meter rating.
2. The location meets the required minimum number of pipe di­ameters upstream and downstream of the sensor head as illus­trated Figure 2-1.
3. Safe and convenient access with adequate overhead clearance for maintenance purposes.
4. Verify that the cable entry into the instrument meets the specific standard required for hazardous area installations.
5. For remote installations, verify the supplied cable length is suf-
ficient to connect the flow meter sensor to the remote electron­ics.
Also, before installation check your flow system for anomalies such as:
leaks
valves or restrictions in the flow path that could create disturbances in
the flow profile that might cause unexpected flow rate indications
IM-22-C 2-1
Chapter 2 Installation Series 220/221 Instruction Manual

Unobstructed Flow Requirements

Select an installation site that will minimize possible distortion in the flow profile. Valves, elbows, control valves and other piping components may cause flow disturbances. Check your specific piping condition against the ex­amples shown below. In order to achieve accurate and repeatable perform­ance install the flow meter using the recommended number of straight run pipe diameters upstream and downstream of the sensor.
Note: For liquid applications in vertical pipes, avoid installing with flow in the downward direction because the pipe may not be full at all points. Choose to install the meter with flow in the upward direction if possible.
AB
AB
C' C
Flow meter
Flow conditioner
(if used)
Example 1. One 90° elbow before meter
Flow meter
Example 4. Reduction before meter
AB
AB
C
C'
Example 2. Two 90° elbows before meter in one plane
Flow meter
Flow conditioner
(if used)
Example 5. Expansion before meter
C' C
Flow meter
Flow conditioner
(if used)
AB
AB
C
C'
Example 3. Two 90° elbows before meter out of plane (if three 90° bends present, double recommended length)
Flow meter
Flow conditioner
(if used)
C' C
Example 6. Regulator or valve partially closed before meter (If valve is always wide open, base length requirements on fitting directly preceding it)
Flow meter
Flow conditioner (if used)
Minimum Required
Upstream Diameters
No Flow
Conditioner
With Flow Conditioner
Example A A C C´ B B
1 10 D N/A N/A N/A 5 D 5 D 2 15 D 10 D 5 D 5 D 5 D 5 D 3 25 D 10 D 5 D 5 D 10 D 5 D 4 10 D 10 D 5 D 5 D 5 D 5 D 5 20 D 10 D 5 D 5 D 5 D 5 D 6 25 D 10 D 5 D 5 D 10 D 5 D
D = Internal diameter of channel. N/A = Not applicable
Minimum Required
Downstream Diameters
No Flow
Conditioner
With Flow
Conditioner
Figure 2-1. Recommended Pipe Length Requirements for Installation, Series 220 and 221
2-2 IM-22-C
Series 220/221 Instruction Manual Chapter 2 Installation

Series 220 In-Line Flow Meter Installation

Install the Series 220 In-Line Flow Meter between two conventional pipe flanges as shown in Figures 2-3 and 2-4. Table 2-1 provides the recom­mended minimum stud bolt lengths for wafer-style meter body size and dif­ferent flange ratings.
The meter inside diameter is equal to the same size nominal pipe ID in schedule 80. For example, a 2” meter has an ID of 1.939” (2” schedule 80).
Do not install the meter in a pipe with an inside diameter smaller than the inside diameter of the meter. For schedule 160 and higher pipe, a spe-
cial meter is required. Consult the factory before purchasing the meter.
Series 220 Meters require customer-supplied gaskets. When selecting gasket material make sure that it is compatible with the process fluid and pressure ratings of the specific installation. Verify that the inside diameter of the gas­ket is larger than the inside diameter of the flow meter and adjacent piping. If the gasket material extends into the flow stream, it will disturb the flow and cause inaccurate measurements.
Flange Bolt Specifications
Stud Bolt Lengths for Each Flange Rating (inches)
Line Size Class 150 Class 300 Class 600
1 inch 6.00 7.00 7.50
1.5 inch 6.25 8.50 9.00 2 inch 8.50 8.75 9.50 3 inch 9.00 10.00 10.50 4 inch 9.50 10.75 12.25
Table 2-1. Minimum Recommended Stud Bolt Lengths for Wafer Meters
The required bolt load for sealing the gasket joint is affected by several ap­plication-dependent factors, therefore the required torque for each application may be different. Refer to the ASME Pressure Vessel Code guidelines for bolt tightening standards.
1
34
2
1
8
6
5
34
7
2
4
10
4-bolt 8-bolt 12-bolt
1
12
8
6
5
9
3
7
11
2
Figure 2-2. Flange Bolt Torquing Sequence
IM-22-C 2-3
Chapter 2 Installation Series 220/221 Instruction Manual

Wafer-Style Flow Meter Installation

Install the wafer-style meter between two conventional pipe flanges of the same nominal size as the flow meter. If the process fluid is a liquid, make sure the meter is located where the pipe is always full. This may require lo­cating the meter at a low point in the piping system. Note: Vortex flow me­ters are not suitable for two-phase flows (i.e., liquid and gas mixtures). For horizontal pipelines having a process temperature above 300° F, mount the meter at a 45 or 90-degree angle to avoid overheating the electronics enclo­sure. To adjust the viewing angle of the enclosure or display/keypad, see page 2-17.
Caution!
When using toxic or cor-
rosive gases, purge the
line with inert gas for a minimum of four hours
at full gas flow before
installing the flow meter.
Figure 2-3. Wafer-Style Flow Meter Installation
When installing the meter make sure the section marked “inlet” is positioned up­stream of the outlet, facing the flow. This ensures that the sensor head is posi­tioned downstream of the vortex shedder bar and is correctly aligned to the flow. Installing the meter opposite this direction will result in completely inaccurate flow measurement. To install the meter:
1. Turn off the flow of process gas, liquid or steam. Verify that the line is not pressurized. Confirm that the installation site meets the required minimum upstream and downstream pipe diameters.
2. Insert the studs for the bottom side of the meter body between the pipe flanges. Place the wafer-style meter body between the flanges with the end stamped “inlet” facing flow. Center the meter body inside the diameter with respect to the inside diameter of the adjoining piping.
3. Position the gasket material between the mating surfaces. Make sure both gaskets are smooth and even with no gasket material extending into the flow profile. Obstructions in the pipeline will disturb the flow and cause inaccurate measurements.
4. Place the remaining studs between the pipe flanges. Tighten the nuts in the sequence shown in Figure 2-2. Check for leaks after tightening the flange bolts.
2-4 IM-22-C
Loading...
+ 42 hidden pages