The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPR8-125.061.03.02.02 Release Date:08.03
Replaces: SPR8-125.061.03.01.02
Page 2
0 - 2Revision
ChapterPageRevision
AllAll02
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Letzte Seite des Kapitel
POLYMOBIL PlusSPR8-125.061.03Page 2 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
8 ______ Changes to previous version _____________________________________8 - 1
Letzte Seite des Kapitel 0
POLYMOBIL PlusSPR8-125.061.03Page 4 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 5
Prerequisites1
Text emphasis1
1 - 1
DANGER
WAR NING
CAUTION
NOTICE
NOTE
DANGER indicates an immediate danger that is likely to result in
death or serious physical injury.
WARNING indicates a risk of danger that may lead to death or to
serious physical injury.
CAUTION used in conjunction with the safety alert symbol indicates a risk of danger that is likely to result in slight or moderate
physical injury and/or damage to property.
NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads to or may lead to an undesirable
result or state other than death, physical injury or damage to
property.
NOTE contains information that emphasizes proper use of the
equipment or proper execution of a procedure, i.e. hints, tips.
Symbols1
Checks and adjustments that must be performed with radiation ON are identified by the
radiation warning symbol.
This symbol means "Dangerous voltage".
Checks marked with this symbol are to be entered into the test certificate at the end of
these instructions.
This symbol means "Attention, consult the documentation".
This symbol indicates components sensitive to electrostatic discharge (ESD).
Siemens AGSPR8-125.061.03Page 1 of 6POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 6
1 - 2Prerequisites
Required documents1
• Block circuit diagramX 069I
• Function descriptionSPR8-125.041...
• Instructions for useSPR8-125.201.02...
NOTE
A detailed A 3 circuit diagram can be ordered under item
number 18 20 377 X069I.
Tools, measurement and auxiliary devices1
NOTE
.
• Standard service kit*
• Digital multimeter
• 2-channel memory oscilloscope
• mAs meter
• Dose meter
• Protective conductor meter
• Equivalent leakage current meter
• Torque wrench 20 Nm - 100 Nm
• Loctite 242
• Current transformer (50 A : 50 mA)*
In addition, a metal film resistor with
10 Ohms; 0.5 W; 1 % is required*
Refer also to Speed Info 06/91
• Calibration tool for DAP measuring system (Part No. 65 84 978)
All tools, measurement and auxiliary devices with the exception
of those marked " * " are listed along with their specifications in
the STC (Service Tools Catalog).
(Part No. 31 51 289)
WAR NING
Dangerous electrical voltage present during service work!
• Risk of death or serious physical injury.
• The oscilloscope has to be connected to ground to perform
measurements. A TEK video isolator and the trigger set have to
be used when ground loops may distort the measurement
results (see ARTD-002.731.22..).
POLYMOBIL PlusSPR8-125.061.03Page 2 of 6Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 7
Prerequisites1 - 3
Safety information and protective measures1
NOTICE
WAR NING
• The following items must be complied with when performing
service work and tests:
- The product-specific safety information in the document
- The safety information in TD00-000.860.01.. in the Register of
the POLYMOBIL Plus binder
- The safety information contained in ARTD Part 2.
Dangerous X-radiation during checks and adjustment work steps.
Risk of death or serious physical injury.
For checks and adjustments that must be performed with radiation switched on, the prescribed radiation safety measures must
be observed; if necessary, wear radiation protective clothing (see
also ARTD-002.731.02.. and ARTD-002.731.38.. General Guidelines for Technical Service). These checks and adjustments are
explicitly identified on the following pages with the radiation
warning symbol .
Siemens AGSPR8-125.061.03Page 3 of 6POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 8
1 - 4Prerequisites
4
R
3
R
5
R
1
C
C
C
V
-
+
Fig. 1
DANGER
When working on the open system, there is a danger of
Electric shock
!
Contact with current-carrying parts will lead to death or to serious
physical injury.
• The capacitor bank can still be charged.
Do not attempt to work on the system while this condition exists.
• After switching off the system, approximately 450 V may still be
present in the system, even after disconnecting the line voltage
plug. Within 10 minutes this voltage will drop to approximately
10 V.
• Always measure the actual voltage present with the DVM at test
points -VCC and + VCC on the D 960 inverter board or (more
accessible) on the D 970 capacitor board at the + connection
point of capacitor C3 and on the right side of fuse F3 (see Fig. 1).
• LED’s V1 ... V10 on D 970 go out at a significantly higher voltage level
and therefore are not reliable safety indicators.
• If a fuse on the D 970 has responded, high voltage may still be present
at the affected capacitor even after a prolonged period of time.
• The capacitor discharging circuit utilizes the D925 board, CS and LS
relays. If connectors X3 or X9 on D 925 or D 950 are not inserted or if
there is a defect in the circuit, the C-bank will not discharge.
This can cause life-threatening voltage to be present in the system
even after a prolonged period of time.
• Refer also to chapter "Replacing the capacitor bank".
POLYMOBIL PlusSPR8-125.061.03Page 4 of 6Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 9
Prerequisites1 - 5
• Connect the POLYMOBIL Plus only to a line voltage supply (line voltage receptacle) that
complies with the requirements of VDE 0107 or corresponds to the local national
regulations.
• Disconnect the POLYMOBIL Plus at the line voltage OFF switch on the operating
console and disconnect the line voltage plug prior to any service work.
• Remove or insert boards with the generator switched OFF only; observe
ESD guidelines when handling boards.
Replacing damaged or missing screws1
• Damaged or missing screws may only be replaced by steel screws as specified in the
installation drawings conforming to DIN267 and must have the specified tensile strength.
NOTICE
Replacement of screws.
All Allen screws must have a tensile strength rating of 8.8 !
Cleaning1
• The unit must always be switched off or disconnected before cleaning.
• Use only water to clean the unit or a lukewarm, diluted solution consisting of water and a
household cleaning agent.
• Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot
remover. Do not spray with water.
• For additional information, refer to the Instructions for use, chapter "Cleaning/
Disinfection".
Siemens AGSPR8-125.061.03Page 5 of 6POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 10
1 - 6Prerequisites
This page intentionally left blank.
POLYMOBIL PlusSPR8-125.061.03Page 6 of 6Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 11
Checking the assemblies2
Preparation2
4
R
3
R
5
R
1
C
C
C
V
-
+
Fig. 1
2 - 1
DANGER
When working on the open system, there is a danger of
Electric shock
!
Contact with current-carrying parts will lead to death or to serious
physical injury.
• The capacitor bank can still be charged.
Do not attempt to work on the system while this condition exists.
• For details see chapter "Prerequisites".
Siemens AGSPR8-125.061.03Page 1 of 8POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 12
2 - 2Checking the assemblies
1
Fig. 2Fig. 3
To check the assemblies and remove the upper covers, the handles and the lower front
cover of the switchbox.
Proceed as follows:
• Switch POLYMOBIL Plus OFF and disconnect the line voltage plug.
• Remove the four Allen screws (1/Fig. 2) in the handle and remove the handle
(see Fig. 2 and Fig. 3).
POLYMOBIL PlusSPR8-125.061.03Page 2 of 8Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 13
Checking the assemblies2 - 3
2
3
2
3
Fig. 4Fig. 5
• Remove the four Allen screws at the bottom of the switchbox cover (2/Fig. 4) and remove
the cover.
• After removing the four panel screws (3/Fig. 4), remove the bottom cover at the front with
the cassette tray.
WAR NING
When working on the open system, there is a danger of
Electric shock
Risk of death or serious physical injury.
Always measure the actual voltage present with the DVM at test
points -VCC and + VCC on the D 960 inverter board or (more
accessible) on the D 970 capacitor board at the + connection
point of capacitor C3 and on the right side of fuse F3.
!
4
• Remove the four screws of the top housing (4/Fig. 5)
• To stay clear of the exposure guide holder, tilt the housing forward and lift it up to
remove it.
Siemens AGSPR8-125.061.03Page 3 of 8POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 14
2 - 4Checking the assemblies
Position and function of the LED’s, fuses, test points and
potentiometers
NOTE
D915 CPU, control and measurement value acquisition2
LED:
V22Preparation ON
V23Exposure ON
V24Error occurred
V25not used
V34Radiation ON
Test points:
X15PinColor
1blackGround (+ 5 V)
3blackGround (+ 5 V)
4red+ 5 V
5brown+15 V
6blueGround (+/- 15 V)
7yellow-15 V
Potentiometers:
See block diagram for the location of components and boards.
± 5 %
± 10 %
± 10 %
2
P1Setting of the mAs counter
P2OFFSET mAS - converter
Set at the factory - Do not change!
P3Max. filament inverter frequency
P4Max. main inverter frequency
P5Adjustment/setting of filament heating circuit
P6Nominal filament current (standby), nominal tube current (exposure)
Set at the factory - Do not change!
P7Nominal kV value
POLYMOBIL PlusSPR8-125.061.03Page 4 of 8Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
VP+15 V for inverter frequency ok
+12V+12.5 V auxiliary voltage ok
I/O+12.5 V auxiliary voltage for I/O ok
ARLine voltage for anode starter ok
+DC, -DC 300 V voltage for filament ok
Fuses:
F110 AT+ 12.5 V for collimator lamp
F30.6 ATAuxiliary voltage +12.5 V for I/O
F40.6 ATAuxiliary voltage +12.5 V
F53.2 AFLine voltage for anode starter
F61.0 ATLine voltage for filament circuit
F710 ATPower
F810 ATPower
Siemens AGSPR8-125.061.03Page 5 of 8POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 16
2 - 6Checking the assemblies
Measurement points:
VP+15 V
P12+ 12.5 V
GNDLGround (+ 12.5 V)
X4PinColor
1red+12.5 V
2red+12.5 V
3red+12.5 V
4blackGround (+12.5 V)
5blackGround (+12.5 V)
6blackGround (+12.5 V)
± 10 % for inverter frequency
± 10 % auxiliary voltage
± 10 %
± 10 %
± 10 %
U1 power supply unit 5 V ± 5% / 15 V ± 10 %2
Fuse: 2 A T
U2 power supply unit 12.5 V ± 10 %2
Fuse: 5 A T
POLYMOBIL PlusSPR8-125.061.03Page 6 of 8Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 17
Checking the assemblies2 - 7
D950 capacitor charging circuit2
LED:
LINE (V25)Line voltage ok
450 VCC (V30) Capacitor voltage present (the voltage can also be considerably lower
than 450 V)
Fuse:
F110 ATCharging circuit
Test points:
ERRError output charging error
VCCapacitor voltage 1 V 100 V
ICCapacitor charging current
FR1Charging circuit frequency (dependent on the line voltage and VC)
GNDGround
CARSignal to enable the charger
FR2Frequency of the auxiliary power supply
OKSignal to indicate the full charge
D970 capacitor bank2
LED:
V1 - V10Capacitor voltage present at C1 - C10
Fuse:
F1 - F1020AFCapacitor charging voltage for C1 - C10
F1180AFCapacitor voltage for inverter
F1220AFCharging voltage for capacitor bank
Test points:
-VCC, +VCCCapacitor voltage
Siemens AGSPR8-125.061.03Page 7 of 8POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 18
2 - 8Checking the assemblies
Status of the LED’s (optimum condition)2
a) POLYMOBIL Plus OFF, line voltage plug connected:
D925:LED+12 VON
LED I/OON
D950:LED450 VCC (V30) glows for 10 minutes following shutdown
D970:all LED’s glows for 10 minutes following shutdown
b) POLYMOBIL Plus ON
D915:The exposure values are displayed
D925:LED VPON
LED+12 VON
LED I/OON
LED ARON
LED+DCON
LED-DCON
D950:LEDLINE (V25)ON
LED450 VCC (V30)ON
D970:all LED’sON
c) Preparation
All LED’s as described under b) and also V22 on D915.
d) Exposure
, standby
All LED’s as described under b) and also V22 and V23 as well as radiation indicator V34
on D915.
In the case of an error, V24 on D915 glows (red).
POLYMOBIL PlusSPR8-125.061.03Page 8 of 8Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 19
Troubleshooting3
Error messages3
3 - 1
NOTE
• Error messages Err 1 and initialization errors Err 90, Err 96 and
Err 97 block the generator so that the system must be switched
off and switched back on again.
• "StandBy" error messages disappear when the error is cleared.
• Error messages which occur during an exposure will continue to
Err 1Capacity charge errorCharging circuit error
Err 2+15 VError
Err 3Vfil < Vstby
Err 4Vfil > Vstby
Err 5KVact ≠ 0
Err 6mAact ≠ 0
Err 7Rot ≠ 0
Err 8Braking failureError in the rotating anode circuit
Err 9Pressure switchOil pressure switch in the single tank
is active
Err 30Fast capacitor chargeCapacitor bank is charging too fast
Err 31Slow capacitor chargeCapacitor bank is charging too slow
Err 33Inverter Error
Err 34Filament circuit Inverter Short Cir-
cuit
c) During exposure
Err 10Rot
Err 11Short circuit in main inverterShort circuit in the main inverter
Err 12KVmax
Err 13Imax
Err 14KVact < KVnom
Err 15mAact < mAnom
Err 17Maximum exposure time Max. exposure reached
Err 18Exposure not completedExposure time not completed (expo-
Err 20Capacitor not readyCapacitor bank not completely
Err 21Vfil > Vmax during prepMaximum filament heating exceeded
Err 22Maximum prep timeMax. preparation time exceeded
Siemens AGSPR8-125.061.03Page 1 of 4POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
= 0No rotating anode acceleration
sure release switch released)
charged
during preparation
Page 20
3 - 2Troubleshooting
Troubleshooting after error messages3
Err 1Capacitor charge error
- Has the system not been used for a long period of time
(higher capacitor leakage current)?
wIf the error occurs after switching the system off and back
on again, configure the capacitor bank
(see chapter "Service program", page 4-2).
- Are the discharge resistors permanently connected?
- Has the temperature sensor on D950 responded
(D950 TP ’TP’)?
- Has overvoltage been detected in the C-bank? (D950 TP ’VC’)
- Is there a short on D950? The error appears only when the short circuit is
continuous.
Replace D950.
Err 2+ 15 V - Error
- Check the fuse on power supply U1.
- Is power supply U1 defective?
Err 3 Vfil < Vstby (no or insufficient filament current)
- Do ’V1’ (-DC) and ’V2’ (+DC) flash on D925? Otherwise check fuse F6.
- No filament frequency? Check D915.
- Connection D925 =>X-ray tube unit (X5).
- Connection D925 => D915 (X11).
- Check X-ray tube assembly.
- D925 defective.
Err 4Vfil > Vstby (filament current is too high)
- Do ’V1’ (-DC) and ’V2’ (+DC) flash on D925? Otherwise check fuse F6.
- Filament frequency too high?
- A/D converter on D915 defective? => Replace D915.
- Short circuit at the output of the filament inverter =>check connection to/in
the X-ray tube assembly.
Err 5KVact 0
Err 6mAact 0
≠
- No connection to the single tank generator (D915 X8)?
- Plug-in connection on D900 in the single tank generator ok?
≠
- Error in the measurement acquisition on D915? => Replace D915.
- No connection to the single tank generator (D915 X8)?
- Plug-in connection on D900 in the single tank generator ok?
Err 7Rot 0 (rotating anode speed 0)
POLYMOBIL PlusSPR8-125.061.03Page 2 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
≠≠
- Error on D915 or D925.
Page 21
Troubleshooting3 - 3
Err 8Braking failure (error when braking the rotating anode - a current was
measured in the auxiliary capacitor C2 during braking.)
- D915TP ’I1’. No current
- D915 TP ’I2’. Braking current must be measured here.
- TRIAC ’V7’ or ’V8’ defective? => Replace D925.
Err 9Pressure switch (oil pressure switch in the single tank generator is
active)
- Tube is overheated?
- Measure D915 plug X8 at pin 8. In case of malfunction, 0 V are measured.
Err 10Rot = 0 (rotating anode speed is 0 on exposure)
The error is only displayed as long as preparation is depressed.
- LED ’AR’ on D925 flashes; is fuse F5 ok?
- Measure at TP ’I1’ and ’I2’. Current has to be present during acceleration.
- Check D925.
- Is the power factor correction capacitor C2 defective?
- Connection to/in the single tank generator ok?
Err 11Short circuit in main inverter
- Check D960.
- Check D915.
- Main inverter frequency too high?
must be measured here when braking.
Err 12KVmax
- Interruption of KVact measurement. Measure TP ’KVN’ and ’KVP’ on D915.
Both signals must have the same amplitude.
- ’KVN’ or
’KVP’ > 70 kV.
- Error in kV measurement circuit?
- Error in kV control circuit?
Err 13Imax
The tube or filament current is recorded.
- Check the filament circuit.
- Exposure filament basic value incorrectly set (P5 on D915)?
- Tube arcing?
If kVact < kVnom, Err 14 or Err 13 is detected and displayed. See also troubleshooting information under Err 14.
Err 14KVact < KVnom
- Voltage in the capacitor bank too low (D950 TP ’VC’)?
- Defective capacitors /fuses on the capacitor bank (D970)?
-No KVact?
- Error in the measurement circuit on D915?
- Main inverter frequency too low?
- Defect in the main inverter D960?
Err 15mA
ACT
< mA
NOM
- Check the filament circuit.
Siemens AGSPR8-125.061.03Page 3 of 4POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 22
3 - 4Troubleshooting
Err 17Maximum exposure time (max. exposure time reached)
- Filament current too low?
- mAs counter is set incorrectly (P1 on D915)?
Err 18Exposure not completed (exposure not completed by the system - expo-
sure switch was released)
- Operator error?
- Intermittent contact in the release switch?
- Error in the exposure release circuit?
Err 20Capacitor bank not ready
Device was not ready yet / had not released yet.
- F1 on D950 defect?
- Error on D950?
Err 21Vfil > Vmax (max. filament heating exceeded during preparation)
- Error during the acquisition of the filament heating?
- Basic value of the exposure filament incorrectly set (P5 on D915)?
- Error on D915?
Err 22Maximum preparation (preparation time too long)
- Prep level pressed too long.
Err 30Fast capacitor charge (capacitor bank charges too fast)
- Are all fuses on the capacitor bank (D970) ok?
- Are there defective capacitors?
- Is the line voltage too high (> 250 V)?
- Is the charging frequency too high?
- Configure the capacitor bank.
- D950 defect?
Err 31Slow capacitor charge (capacitor bank charges too slow
- ’V25’ (line) on D950 flashing? Check fuse F1
- No or insufficient charging frequency?
- Check D950 (’V22 - V24’ defective?)
- Poor connections to/in the capacitor bank?
Err 33Inverter Error
- Check ribbon cable X20.
- FMAX not adjusted?
- IGBT short-circuited?
Err 34Filament Inverter short-circuit
- FMAX not adjusted?
- IGBT short-circuited?
)
Err 90EPROM checksum
- EPROM defective.
Err 96kVnom failure (hardware error, see also Err 5)
Err 97mAnom failure (hardware error, see also Err 6)
POLYMOBIL PlusSPR8-125.061.03Page 4 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 23
Service programs4
SW2
4 - 1
2
Fig. 1
P4
P3
P2
P1
GNDJR
ST
X12
X15
1
"OPEN"
V22
V23 V24 V25
Service
D 915
P5
P6
P7
X11
11
X10X20
1
REG
CAL
KVP
KVN IH JRS KVS
KV
To bypass the service mode, the ST slot (1/Fig. 1, below the display on D915) has to be
jumpered before switching on the POLYMOBIL Plus.
As soon as "Pr. 1" appears in the display, bring the jumper back to the "OPEN" position.
F1 I
I1 I2
TRG
JIM
SCHGND
VC BAT KVM SCM VR
KV +/-Keys toggle between the programs. When you are
already in a program, the next or previous program step
is selected.
Setting of the kV values in "Pr. 5" and "Pr. 6".
mAs +/-Setting of the mAs values in "Pr. 5" and "Pr. 6".
Lamp switchActivating or quitting the current program.
To quit the service mode, switch the POLYMOBIL Plus off/on or press the Reset button
SW 2 (2/Fig. 1).
Siemens AGSPR8-125.061.03Page 1 of 2POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 24
4 - 2Service programs
Service programs available4
Pr. 1Configures the capacitor bank.
After starting the service program, "CAP 150" appears on the display.
The voltage of the capacitor increases in increments (approx.) from 150
V to 250 V, 330 V, 400 V, 430 V up to 440 V.
The program remains at each step for about 6 minutes at the 150 V to
440 V level.
The system remains at the 440 V level for one hour after which
the program automatically ends; "CAP END" appears on the display.
Pr. 2Reads out the exposure counter.
Pr. 3Error memory with up to 20 errors.The left display indicates how many
errors occured.The error message is displayed in the right display. The
latest error is displayed first.
Pr. 4Deletes error memory.
Press mAs + key for approximately 4 seconds.
Pr. 5Sets the default exposure parameters (according to customer request).
Pr. 6Sets the maximum possible KV and mAs values (according to customer
request).
Pr. 7Adjusts FMAX on the main inverter.
"ADJ F" appears on the display after starting the service program.
Press S27.
"F on" appears on the display.
FMAX can be measured at TP "REG" on D 915 and can be set with
potentiometer P4.
Pr. 8Tests the anode starter unit.
Press S 27.
Pr. 9 Tests the filament unit.
"FIL 3" appears on the display after starting the service program.
Press S27.
Filament boosts; "FIL 5" appears on the display.
NOTE
There is no radiation present during these service programs.
POLYMOBIL PlusSPR8-125.061.03Page 2 of 2Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 25
Measurements5
Line voltage5
The line voltage must be set to 110 V (+ 10%) or 230 V (+ 10%).
The line voltage is adjusted automatically.
Measuring KVact and Itube5
• Connect the oscilloscope to TP "JR", "KV" and "GND" on D915.
CH1 TP "JR"1.0 V/Div
CH2TP "KV"1.0 V/Div
Ground at "GND"
10 ms/Div
Trigger external at TP "TRG"
• Set 90 kV and 10 mAs at the POLYMOBIL Plus and release an exposure.
The oscillogram should correspond to that shown in Fig. 1.
5 - 1
Fig. 1
Siemens AGSPR8-125.061.03Page 1 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 26
5 - 2Measurements
Capacitor charging circuit5
NOTE
Signals are only present during charging of the capacitors.
Charging frequency5
• Connect the oscilloscope to D 950 at TP "FR1" and "GND".
CH1 TP "FR1"5.0 V/Div
Ground at "GND"
20 µs/Div
• After switch on, the following oscillogram can be measured:
Fig. 2
The capacitor charging frequency is 25 kHz and cannot be adjusted.
POLYMOBIL PlusSPR8-125.061.03Page 2 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 27
Measurements5 - 3
Capacitor charging current5
• Connect the oscilloscope to D 950 at TP "IC" and "GND".
CH1 TP "IC"500 mV/Div
Ground at "GND"
20 µs/Div
• After switch on, the following oscillogram can be measured:
Fig. 3
The charging current depends on the line voltage present and on the charge status
(capacitor voltage U
Fig. 3 shows the charging current in standby
) of the capacitor bank.
c
, i.e. when the C-bank is charged.
Siemens AGSPR8-125.061.03Page 3 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 28
5 - 4Measurements
Capacitor voltage (Uc) and charging time5
In exposure-ready status, the capacitor voltage is approximately 440 V and can be measured on D 950 TP "VC".
If the device is switched off because of overvoltage, the signal "ERR" is output on D950
WAR NING
The charging time depends upon the residual charge of the capacitor bank and on the
available line voltage, e.g. with 230 V the max. charging time is about 15 s.
In the worst case, (line voltage <100 V and residual charge = 0 V), the charging time
can be 40 seconds.
For the following measurement, the line voltage was 230 V and the residual charge 30 V.
When working on the open system, there is a danger of
Electric shock
Risk of death or serious physical injury.
You can measure the capacitor voltage at TP "VC" only if
POLYMOBIL Plus is switched on.
!
D915 test point "VC" (1 V 100 V)
Fig. 4
POLYMOBIL PlusSPR8-125.061.03Page 4 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 29
Measurements5 - 5
Rotating anode starter5
Acceleration5
Fig. 5 shows the stator currents during rotating anode acceleration (230 V).
Fig. 5
CH2 D915 test point "I1" (1 V 0.25 A)
CH1 D915 test point "I2" (1 V 0.25 A)
10 ms /Div
NOTE
The currents are correspondingly lower at 100 V line voltage.
Siemens AGSPR8-125.061.03Page 5 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 30
5 - 6Measurements
Brakes5
Fig. 6 shows the stator currents during braking (230 V)
Fig. 6
CH2 D915 test point "I1" (1 V 0.25 A)
CH1 D915 test point "I2" (1 V 0.25 A)
10 ms /Div
NOTE
The currents are correspondingly lower at 100 V line voltage.
POLYMOBIL PlusSPR8-125.061.03Page 6 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 31
Measurements5 - 7
kV Control5
Checking the maximum main inverter frequency5
NOTE
It is generally not necessary to set the main inverter frequency in
the field. When ordered as a replacement, D915 is preset at the
factory.
• Connect the oscilloscope to D915 at TP "REG" and "GND".
• Measure F
(See "Service program", section on "Service programs available")
5 V/Div.
in the service program Pr. 7
MAX
Fig. 7
20
µs/Div.
0 →
F
= 18 kHz
MAX
T
MIN
= 55.5 µs
Tolerance: -1.5 µs /+1.5 µs
• Setting the maximum main inverter frequency with potentiometer P4
on D 915.
Siemens AGSPR8-125.061.03Page 7 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 32
5 - 8Measurements
Measuring the oscillating current5
WAR NING
When working on the open system, there is a danger of
Electric shock
Risk of death or serious physical injury.
Prior to beginning the following instructions, you must discharge
the capacitor bank because life-threatening DC voltage is
present. Check the actual charge with the DVM.
!
• Switch OFF POLYMOBIL Plus and wait until the capacitor voltage has dropped to a
value < 10 V.
• Disconnect the cable from the inverter to capacitor C1. Push the cable through the
current transformer 50 A : 50 mA (with a parallel resistor of 10 ohms) and reconnect the
cable.
• Connect the oscilloscope to the current transformer.
• Switch ON POLYMOBIL Plus.
• Set 60 KV, 10 mAs on POLYMOBIL Plus and release exposure.
The diagram must correspond to Fig. 8.
NOTICE
Maximum main inverter frequency
The maximum main inverter frequency can only be measured in
the ascending part of the high voltage, i.e. in the first microseconds, because afterwards the frequency is lower.
POLYMOBIL PlusSPR8-125.061.03Page 8 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 33
Measurements5 - 9
Fig. 8
CH2: 1 V/Div, Trigger: Test point "TRG" external on D 915
20 µs /Div
Siemens AGSPR8-125.061.03Page 9 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 34
5 - 10Measurements
Measuring the kVact and kVnom5
• Connect the oscilloscope to D915 at TP "KV", "KVS" and "GND".
• Set 60 KV, 10 mAs on POLYMOBIL Plus and release exposure.
• The oscillogram must appear as shown in Fig. 9.
• The KVnom value can be set with P7 on D915.
Fig. 9
kVact:CH1 D915 test point "KV"1 V 20 KV
kVnom:CH2 D915 test point "KVS"1 V 30 KV
20 ms /Div
POLYMOBIL PlusSPR8-125.061.03Page 10 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 35
Measurements5 - 11
Error in the kVact acquisition5
Fig. 10 shows that one kVact cable from the single tank generator to the D915 is damaged
or disconnected.
The monitor (KVM) responds (curve 3) because KVN and KVP have different amplitudes;
the exposure is terminated with ERR 12.
Refer to function description "KV monitoring".
Fig. 10
Exposure data: 102 KV 10 mAs
KVact:CH1: D915 test point "KV"
Itube:CH2: D915 test point "JR"
CH3: D915 test point "KVM"
1 ms /Div
Siemens AGSPR8-125.061.03Page 11 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 36
5 - 12Measurements
Testing the filament circuit5
Filament circuit in standby5
• Connect the oscilloscope to D915 at TP "CAL", "I" and "GND".
Fig. 11
Filament inverter frequencyCH1: D915 test point "CAL"
Filament currentCH2: D915 test point "I"
POLYMOBIL PlusSPR8-125.061.03Page 12 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 37
Measurements5 - 13
Setting the maximum filament inverter frequency5
The maximum filament inverter frequency is 25 kHz and can be set with P3 on D915.
Proceed as follows:
• Switch POLYMOBIL Plus OFF.
• Remove fuse F6 on D925.
• Connect the oscilloscope to D915 at TP "CAL" and "GND".
• Switch POLYMOBIL Plus ON (Err 03 is blinking).
The oscillogram must appear as shown in Fig. 12.
CH1: D915 test point "CAL"
20 µs /Div
Fig. 12
• Setting the maximum filament inverter frequency with potentiometer P3 on D915.
Siemens AGSPR8-125.061.03Page 13 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 38
5 - 14Measurements
,
Measuring the filament nominal value5
The filament nominal value is set at the factory and should not be changed.
The filament nominal value can be measured on D915 at TP "JRS" and "GND" with the
DVM and is 1.5 V
± 4% in standby.
NOTE
If this value is changed, the nominal values for preparation and
exposure are also affected.
Setting the tube current for the exposure5
The tube current must be reset for each tube.
Proceed as follows:
• Connect the oscilloscope to D915 at TP "JR", "IH" and "GND".
• Set 60 kV, 10 mAs on POLYMOBIL Plus and release an exposure.
The oscillogram must appear as shown in Fig. 13.
Fig. 13
Actual tube current valueCH1: D915 test point "JR"
Actual fil. current valueCH3: D915 test point "IH"
If this is not the case, adjust P5 on D915 until there is minimal current overshoot at the
beginning of the exposure.
The standby heating of the exposure is set via P5 ().
Fig. 13 shows the correct setting.
In Fig. 14 the standby heating is set too high.
In Fig. 15 the standby heating is set too low.
POLYMOBIL PlusSPR8-125.061.03Page 14 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
IH
IH
Page 39
Measurements5 - 15
Tek Stop: Single Seq 2.50kS/s
T
T
3
Ch1
1.00 VM20.0ms
Ch3
500mV
Standby heating too high
1.56VCh1
Fig. 14
Standby heating too low
Fig. 15
Siemens AGSPR8-125.061.03Page 15 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 40
5 - 16Measurements
Testing and setting the mAs counter5
• Unscrew the lid of the single tank cover.
• Pull out the mAs jumper on D900 and connect the mAs meter.
• Set 70 kV, 80 mAs on POLYMOBIL Plus and release exposure.
If the value displayed on the mAs meter does not coincide with the value set on
POLYMOBIL Plus, adjust the mAs counter with P1 on D915.
Release another exposure and compare the display.
CAUTION
Overheating of the tube during radiation.
If not observed the tube can be damaged.
Observe the cooling intervals for the tube.
POLYMOBIL PlusSPR8-125.061.03Page 16 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 41
Measurements5 - 17
Length
Marker
Side marker
Fig. 16
Coincidence of light and radiation fields5
NOTE
Procedure:
If the POLYMOBIL is equipped with a DAP measuring system,
remove the DAP ionization chamber first.
• Insert film into the 24 cm x 30 cm or 10” x 12” cassette and place it on a table or similar
surface.
• Adjust the vertical SID of 100 cm or 40” to the top edge of the cassette using a measuring
tape.
• Use the knobs to set a format of 18 cm x 24 cm or 8” x 10”.
• Switch on the light localizer and align the cassette.
• Apply radiopaque markers (e.g. washers, coins) as shown in Fig. 16.
Use one washer as a side marker.
• Release an exposure (60 kV, 10 mAs) and develop the film.
• Log the following data on the developed film using a water-resistant felt-tip pen.
- SID setting
- Film size
- Radiation field size
Siemens AGSPR8-125.061.03Page 17 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 42
5 - 18Measurements
Evaluation:
• Measure the deviations (X1, X2, Y1, Y2) between the edges of the light field and the
radiation field on all four sides, corresponding to Fig. 17.
• Calculate the total deviations in the X and Y directions (disregarding the algebraic sign).
Y
1
Film
Light field
Fig. 17
X
2
X
1
Radiation field
Y
2
• The deviations in length (total Y) as well as the deviations in width (total X) must be
smaller than 1.6 cm respectively.
POLYMOBIL PlusSPR8-125.061.03Page 18 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 43
Measurements5 - 19
MOB00672
Fig. 18Fig. 19
• If the deviation is larger, loosen the four Allen screws a little (arrow/Fig. 18/Fig. 19) and
shift the collimator accordingly.
After adjusting the collimator, tighten the screws again (arrow/Fig. 18/Fig. 19).
• Repeat the test and readjust the collimator until the light and radiation fields are within
the acceptable tolerance (< 1.6 cm).
Siemens AGSPR8-125.061.03Page 19 of 20POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 44
5 - 20Measurements
3
4
2
Fig. 20Fig. 21
1
Testing and readjusting the counterbalance5
• Loosen the rotating knob (2/Fig. 20) and unlock the C-arm (1/Fig. 20).
It should be possible to move the C-arm without any attached accessories easily across
the whole range and the arm should come to a stop in any desired position.
Readjusting the counterbalance with accessories attached5
If the C-arm with accessories attached does not come to a stop in any desired position,
readjust the spring tension for the counterbalance.
• Remove the cover (3/Fig. 20) at the stand.
• Release the supplemental brake at the C-arm (2/Fig. 20) completely.
• Move the C-arm into a vertical position.
• Tighten the screw using a 17 mm Allen key (4/Fig. 21).
Adjust the spring tension in such a way that there is a balance when the C-arm is in a
vertical position, i.e. that equal force is required to lift and lower the tube assembly.
POLYMOBIL PlusSPR8-125.061.03Page 20 of 20Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 45
Replacing important components6
6 - 1
NOTICE
To repair paint chips, the POLYMOBIL Plus colors
can be ordered in spray cans:
To replace the single tank generator, remove it from the stand as described below:
• Remove the collimator (see “Replacing the collimator”).
• Move the arm into the lowest position and check whether the safety mechanism (bolt 1 /
Fig. 1) is locked into position.
Now the stand cannot be moved upwards.
• Remove the four Allen screws of the single tank generator cover (2/Fig. 2), remove the
cover, disconnect the connections and the strain relief of the corrugated tubing (3/Fig. 2)
and remove it.
Siemens AGSPR8-125.061.03Page 1 of 10POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 46
6 - 2Replacing important components
4
Fig. 3
• Take out one of the four mounting screws and insert it into the threaded hole and tighten
it (arrow/Fig. 3).
WAR NING
Accurate usage of safety device for the arm of the unit.
If not observed, death or serious physical injury can occur.
This is a safety device (see arrow/Fig.03) to prevent the arm of the
unit from flying upward if the lever for the transport lock is inadvertently activated.
• Place a pad under the single tank generator.
• Disconnect the single tank generator together with the bracket from the stand by
removing the three remaining screws (4/Fig. 3).
• Lower the single tank generator to the pad.
• The single tank generator is installed in reverse sequence.
- Secure the three mounting screws (4/Fig. 3) with Loctite 242 and tighten them.
- Remove the fourth mounting screw from the threaded hole (arrow/Fig. 3) and reinstall it
in its former position. Secure the screw with Loctite 242.
- Test the torque of the mounting screws;
Nom: 25 Nm, Tolerance
± 10%.
• Reinstall the collimator in reverse sequence.
• Set the tube current (see "Setting the tube current for the exposure", page 5-13).
• Check the coincidence of the light and radiation fields and adjust, if necessary
(see "Coincidence of light and radiation fields", in chapter 5)
POLYMOBIL PlusSPR8-125.061.03Page 2 of 10Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 47
Replacing important components6 - 3
5
Fig. 4Fig. 5
6
Fig. 6
Replacing the collimator6
In case of damage, the collimator has to be completely replaced.
Proceed as follows:
• Remove the two screws (6/Fig. 6) and remove the back panel.
• Mark the five connecting cables of the collimator and disconnect them.
• Remove two screws on each side (5/Fig. 5) of the upper cover.
• Turn the two knobs to the end position and remove them.
• Remove the front panel after you have removed the two screws (7/Fig. 7).
Caution: The front panel is connected to the collimator via a cable (9/Fig. 8).
Siemens AGSPR8-125.061.03Page 3 of 10POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 48
6 - 4Replacing important components
• Remove the collimator after removing the four Allen screws (arrow /Fig. 4-Fig. 5).
CAUTION
Accurate replacement of the collimator
If not observed, the collimator can be damaged.
Hold the collimator securely.
Lift the upper cover and pull the cable through the housing.
• Attach the new collimator with the four screws (arrow /Fig. 4-Fig. 5) and center it.
• Reconnect the cables and the back panel.
• Check the coincidence of the light and radiation fields and adjust, if required.
(See “Coincidence of light and radiation field”, page 5-16.)
7
9
10
8
Fig. 7Fig. 8
Replacing the lamp for the localizer 6
• The replacement light localizer lamp is located below the cassette module.
• Turn both knobs (8/Fig. 7) to the end position and remove them.
• Remove the front panel after removing both screws (7/Fig. 7).
Caution: The front panel is connected to the collimator via a cable (9/Fig. 8).
• Remove both screws (10/Fig. 8) and take off the metal bracket for the temperature
switch).
• Take out the defective lamp (11/Fig. 8) and replace it with a new lamp
11
POLYMOBIL PlusSPR8-125.061.03Page 4 of 10Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 49
Replacing important components6 - 5
NOTICE
Danger of burn-in by fingerprint.
Do not touch the glass envelope with your fingers.
Use a fluffless cloth during the replacement of the lamp.
• Attach the bracket and the front panel.
• Adjust both knobs and attach them.
• Check the function of the light localizer lamp.
• Check the coincidence of light and radiation fields (see "Coincidence of light and
radiation fields" in chapter 5).
Siemens AGSPR8-125.061.03Page 5 of 10POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 50
6 - 6Replacing important components
Fig. 9
Replacing the steering caster 6
Step:
1Loosen only Allen screws on the left and right (2.5 mm wrench.
2Pull back hexagonal shaft
3Remove pedal
4Remove Allen screws (wrench size 6 mm).
5Pull out the caster.
6Align the threaded holes on the new caster.
• Install in the reverse sequence.
• Test the pedal positions (see Operating Instructions).
POLYMOBIL PlusSPR8-125.061.03Page 6 of 10Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 51
Replacing important components6 - 7
4
R
3
R
5
R
1
C
C
C
V
-
+
Fig. 10
Replacing the capacitor bank6
DANGER
When working on the open system, there is the danger of
Electric shock
!
Contact with current-carrying parts will lead to death or to serious
physical injury.
• The capacitor bank can still be charged.
Do not attempt to work on the system while this condition exists.
• For details see chapter "Prerequisites".
Siemens AGSPR8-125.061.03Page 7 of 10POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 52
6 - 8Replacing important components
2
1
1
2
Fig. 11Fig. 12
• Switch the unit off; disconnect the line voltage plug;
• Remove the two cover screws on each side (1/Fig. 11) and remove the cover (with
cassette drawer).
• Measure the actual voltage present in the capacitor bank with the DVM at test point -
VCC and +VCC on the inverter board D960 or (more accessible) on the capacitor board
D970 at the + connecting point of the C3 capacitor and on the right-hand side of fuse F3
(see Fig. 12).
4
R
3
R
5
R
1
C
C
C
V
-
+
•
DANGER
Danger of Electric shock
Contact with current-carrying parts will lead to death or to serious
physical injury.
!
• Continue work only if the capacitor unit is discharged (U < 10 V)
• Remove one cover screw on each side (2/Fig. 11).
• Unscrew the three Allen screws (2/Fig. 12).
• Pull out the connectors X3 and X9 (3/Fig. 12) on the charging board D950.
• Remove flat ribbon cable X20 (4/Fig. 12) from the inverter board D960 and from the
charging board D950.
• Disconnect and remove the cables of oscillating circuit capacitor C1 and oscillating
circuit coil L1 to inverter board D960.
• Pull the entire capacitor unit out towards the front.
POLYMOBIL PlusSPR8-125.061.03Page 8 of 10Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 53
Replacing important components6 - 9
DANGER
Danger of Electric shock
Contact with current-carrying parts will lead to death or to serious
physical injury.
!
• If a fuse on D970 is defective, high voltage may still be present at
the affected capacitor.
• In this case, discharge the affected capacitor with the
discharging resistor R5 (Fig. 12).
• Assembly is performed in the reverse sequence.
• Check the capacitor voltage (Uc) and the charging time (see "Capacitor voltage (Uc) and
charging time ", page 5-4)
Siemens AGSPR8-125.061.03Page 9 of 10POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 54
6 - 10Replacing important components
This page intentionally left blank.
POLYMOBIL PlusSPR8-125.061.03Page 10 of 10Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 55
DAP measuring system (option)7
Recalibration of DAP measuring system7
This chapter shows how to recalibrate the DAP measuring system by using an external
ionization chamber (dose meter).
• Connect the DAP calibration tool between both of the connectors that just were
disconnected.
NOTE
It is irrelevant which connector is the input or the output of the
DAP calibration tool.
• System ON.
• Set up the POLYMOBIL and the dose meter as shown in Fig. 2.
NOTICE
Measurement errors.
If a semiconductor detector is used, the temperature/pressure
factor of the dose meter must be considered.
Otherwise a deviation against the factory calibration may occur.
Siemens AGSPR8-125.061.03Page 1 of 4POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 56
7 - 2DAP measuring system (option)
• Switch the collimator light on. The dose meter should be positioned in the center of the
light field and freely in the air to avoid back scattering radiation.
MOB00676
A
B
length
width
Fig. 2 Setup
• Measure the Focus - dose meter distance (A / Fig. 2) and then the width and the length
of the light field on the surface. For optimal accuracy, place an unexposed film cassette
on the surface, and measure the exposed area width and length on the film after an
exposure has been released.
• Measure the Focus-light field/surface distance (B / Fig. 2).
• Calculate the reference area according to the following formula:
2
Reference area
A
------
width length⋅() unit m
2
B
()⋅=
2
• Set kV to 102 and mAs to 20 and release an exposure.
• Record the dose value measured with the dose meter.
POLYMOBIL PlusSPR8-125.061.03Page 2 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 57
DAP measuring system (option)7 - 3
• Calculate the reference DAP value using the measured dose value according to the
following formula:
Reference DAP valuemeasured dose value reference area unit µGy m
()⋅=
2
• Record the DAP value measured with the DAP measuring system (DAP value from
display).
• Calculate a comparison with the following formula:
⋅
100 DAP value from display
Comparison value (%)
The comparison value shows the deviation between the actual DAP value (measured
with the dose meter) and the DAP value measured with the DAP measuring system. If
the comparison value is negative, the DAP display shows a value that is lower than the
actual value, and vice versa.
If the comparison value is less than ±5%, the calibration is completed.
------------------------- --------------------------------------- -------------100–=
Calculated DAP value
• To correct the deviation, press the “UP” button on the calibration tool (for negative
comparison values) or the “Down” button (for positive comparison values). Pressing a
button once corresponds to approximately 0.5 % change of the comparison value.
NOTE
The middle push-button on the DAP calibration tool is for test
purposes. It has the same function as the test button on the DAP
display.
• Repeat the calibration sequence until the deviation is less than ±5 %.
• System OFF.
• Remove the DAP calibration tool and reconnect the DAP ionization chamber as shown in
(pos. 1 / Fig. 1).
• System ON.
• Press the reset and the test button to finish the calibration procedure.
• Release a test exposure.
Miscellaneous service issues7
No further service or maintenance shall be performed on the DAP ionization chamber and
the DAP display. Note that the warranty will become void if the DAP ionization chamber or
the DAP display are opened or altered in any fashion.
Siemens AGSPR8-125.061.03Page 3 of 4POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 58
7 - 4DAP measuring system (option)
This page intentionally left blank.
POLYMOBIL PlusSPR8-125.061.03Page 4 of 4Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Page 59
Changes to previous version8
Chapter 1
Page 1 -2: Section "Tools, measurement and auxiliary devices required"
completely revised.
Revision of the warning symbols.
Chapters 2 and 5
Revision of the warning symbols.
Chapter 6
Page 6 -1: Part No. for spray paint "medical blue" added.
Revision of the warning symbols.
Chapter 7: completely revised.
8 - 1
Siemens AGSPR8-125.061.03Page 1 of 2POLYMOBIL Plus
Medical SolutionsRev. 0208.03CS PS 24
Page 60
8 - 2Changes to previous version
This page intentionally left blank.
POLYMOBIL PlusSPR8-125.061.03Page 2 of 2Siemens AG
Rev. 0208.03CS PS 24Medical Solutions
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.