The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: RXB8-115.041.02.03.02 Doc. Gen. Date:10.97
Replaces: RXB8-115.041.02.02.02
0 - 2Revision
ChapterPageRevision
0all03
1all03
2all03
POLYMOBIL IIIRXB8-115.041.02Page 2 of 4Siemens AG
Rev. 0310.97TDSP 1Medical Engineering
Contents0 - 3
Page
1 _______POLYMOBIL III _________________________________________________ 1 - 1
2 _______Changes to previous version _____________________________________ 2 - 1
Siemens AGRXB8-115.041.02Page 3 of 4POLYMOBIL III
Medical EngineeringRev. 0310.97TDSP 1
0 - 4Contents
This page intentionally left blank.
POLYMOBIL IIIRXB8-115.041.02Page 4 of 4Siemens AG
Rev. 0310.97TDSP 1Medical Engineering
POLYMOBIL III1
Generator principle1
The POLYMOBIL III makes use of high-frequency inverter technology.
An 8-bit microcontroller on board D 910 controls the entire system. The control circuits,
operating switches and operating panel indicators are connected to this controller.
Line connection1
The POLYMOBIL III should be connected to a grounded outlet.
The line voltage must be either 110 V or 230 V.
When switching the unit on at board D 910 ( 3 / 1 ), the line voltage is automatically
switched over via the LU contactor ( 3 / 3 ) at transformer T1. The switch-over voltage is
160 V.
Switching power supply M6 ( 3 / 6 ) generates the internal supply voltages for the electronic components.
By actuating the line voltage switch on the operating console on D 910 ( 3 / 1 ), the HR
relay on board D 920 is switched on. The DC intermediate circuit ( 3 / 7 ) is charged via
resistor R 5 ( 3 / 1 ).
1 - 1
As soon as initialization has ended, the US contactor ( 3 / 5 ) energizes and shorts out
charging resistor R5.
Following successful completion of the self-test, the default values:
60 kV and 10 mAs are displayed.
The Polymobil III is now ready to operate.
Due to high-frequency oscillation generated in the inverter, an integrated EMC line filter
Z1 (3 / 1 ) has been connected to suppress interference signals.
DC intermediate circuit1
The line voltage is rectified by rectifier V1 ( 3 / 7 ) and supplied to the charging capacitors
C1 and C2 in the intermediate circuit on board D 111. Parallel resistors R1 and R2
effect a symmetrical voltage distribution across the charging capacitors. They also discharge the intermediate circuit after switching the system off.
The intermediate circuit supplies voltage to the main inverter.
At 230 V line voltage, V1 ( 3 / 7 ) acts as a standard rectifier and the LU contactor is open.
At 100 V line voltage, the LU contactor ( 3 / 6 ) closes and rectifier V1 in combination with
intermediate circuit capacitors C1 and C2 acts as a voltage doubling circuit.
For safety reasons, intermediate circuit capacitors C1 and C2 automatically discharge via
the auxiliary contact 21-22 ( 3 / 7 ) of the US contactor and resistor R3 after the Polymobil
has been switched off.
The intermediate circuit voltage is monitored by microcontroller J15 on board D 910.
Siemens AGRXB8-115.041.02Page 1 of 6POLYMOBIL III
Medical EngineeringRev. 0310.97TDSP 1
1 - 2POLYMOBIL III
Main inverter1
The main inverter generates high frequency energy packets for the high-voltage transformer. The microcontroller regulates high voltage during exposure via the control frequency.
By triggering two "Isolated Gate Bipolar Transistors", called IGBT’s, across a bridge diagonal, a charge voltage flows through the primary serial circuit L2, C4 and the high voltage
transformer in the X-ray tube single tank generator ( 4 / 2 ).
Subsequently, the charge voltage across serial capacitor C4 is higher than the intermediate circuit voltage. This higher capacitor voltage forces a reverse voltage impulse across
the antiparallel transistor-diodes and the intermediate circuit. The amplitude of the reverse
voltage impulse is always smaller than that of the charge voltage impulse.
One cycle of oscillation is now complete and another one can be transmitted in the opposite direction.
Two bridge diagonals are always required in an inverter, since otherwise the magnetic
flow in the iron core of the high-voltage transformer would reach a saturation state.
Additional energy packets can be transmitted by alternating control of the bridge diagonals. The maximum inverter frequency is 10 kHz.
During the exposure, the current IS (4 / 3 ) is measured on board D 920 and used to synchronize the control pulses ( 4 / 5 ).
Monitoring the inverter1
A possible bridge short circuit is detected by monitoring the collector-emitter voltage Uce
of every IGBT and, if necessary, by blocking the inverter. The microcontroller then indicates an error message:
ERR 11Inverter short circuit
Refer to chapter 3 in the service instructions.
Remarks:If the corrugated hose between the switch module and the X-ray tube sin-
gle tank is not connected, the Polymobil III cannot be switched on.
Testing the inverter1
If jumper S1 on board D 920 ( 92 B / 7 ) is removed, the inverter can be operated without
high voltage.
The exposure is interrupted after 15 ms. An error message is displayed:
ERR 14kV-ACT < kV-NOM
POLYMOBIL IIIRXB8-115.041.02Page 2 of 6Siemens AG
Rev. 0310.97TDSP 1Medical Engineering
POLYMOBIL III1 - 3
X-ray tube single tank generator1
The main inverter feeds the primary winding of the high-voltage transformer (5 / 1 ). The
secondary winding is connected to two high voltage cascades. These supply the stationary anode with high voltage.
The oil pressure switch S1 ( 5 / 1 ) in the tube housing responds if the single tank generator becomes overheated during high use. The following error message is displayed:
ERR 9Oil pressure in single tank too high
Actual value acquisition1
The momentary tube voltage is acquired via a frequency-compensated voltage divider
( 10 000 : 1 ). This kV ACT value can be measured on board D 910 ( 5 / 2 ) via an oscilloscope.
The tube current is acquired via resistor R2 on board D 1. It can be measured via an oscilloscope at board D 910. In addition, the mAs jack is located on board D 1 ( 5 / 4 ).
Overvoltage arrester U1 protects the filament circuit during no-load operation. The V1
suppressor diode protects the filament circuit from over voltage. Suppressor diode V2 protects the mA ACT circuit during no-load operation.
The kV measurement circuit is protected by overvoltage arrester U2.
kV regulation1
The analog kV NOM value ( 6 / 3 ) results from pulse width modulation and can be measured on the oscilloscope on board D 910.
KV NOM and kV ACT are compared and forwarded to PI regulator J33. The output signal
of this regulator is forwarded to a voltage frequency converter. The output frequency can
be measured on the oscilloscope at measurement point REG ( 6 / 1 ). The pulse patterns
for the inverter diagonals are subsequently generated.
The current is measured in the inverter and synchronized with the pulse patterns. The current can be measured on the oscilloscope on board D 920, measurement point IS ( 4 / 3 ).
The microcontroller controls the U/F converter J21 ( 6 / 1). Refer to measurement point
TRIG on board D 910 ( 6 / 4 ). Exposure is also switched on and off via this component.
Siemens AGRXB8-115.041.02Page 3 of 6POLYMOBIL III
Medical EngineeringRev. 0310.97TDSP 1
1 - 4POLYMOBIL III
kV monitoring1
The kV ACT value is continuously being compared with the maximum acceptable kV limit
values in the threshold components J29 ( 6 / 2). The inverter is immediately blocked by
switching off the SS relay ( 6 / 5 ) when the kV limit value is exceeded.
The inverter is disabled in response to the following errors:
Bridge short circuitERR 11
kV-MAX ERR 12100 kV + 10 %
I-MAXERR 1350 mA + 10 %
I-filament-MAXERR 135 A + 20 %
kV-ACT < kV-NOMERR 14kV-NOM - 25 %
Filament regulation1
The analog nominal value of the filament circuit or tube current is generated by the microcontroller J 15 via pulse width modulation and can be measured on the oscilloscope on
board D 910 at measurement points :
IHS ( 7 / 3 )Filament current
IRS ( 7 / 3 )Tube current
The actual value of the filament current can be measured across resistor R 58 and monitored by the microcontroller.
I-FIL-NOM and I-FIL-ACT are linked to each other and forwarded to the Pl regulator J 32.
The output signal of this regulator is forwarded to a voltage frequency converter. This output frequency can be measured via an oscilloscope at measurement point CAL ( 7 / 2 ) on
board D 910. This impulse sequence controls switching transistor V 32 and also the current through the filament.
The actual tube current IR is used for regulation during the exposure.
If the voltage in the intermediate circuit is lower than 260 V=, the tube current is automatically reduced by amplifier J 18 on board D 910. The intermediate circuit voltage can be
measured at measurement point VL ( 7 / 5 ).
Filament monitoring1
If the filament current IH or the tube current IR exceed the maximum acceptable value, the
threshold components J 29 on board D 910 will respond.
The generator is disabled and the following error codes will be displayed on the control
panel:
ERR 13I-TU-MAX + 10 %
ERR 13I-FIL-MAX + 20 %
POLYMOBIL IIIRXB8-115.041.02Page 4 of 6Siemens AG
Rev. 0310.97TDSP 1Medical Engineering
POLYMOBIL III1 - 5
mAs counter1
The tube current acquired on board D 1 ( 7 / 3 ) is forwarded to the voltage frequency converter J 26 ( 7 / 4 ) on board D 910. This converter generates a proportional frequency at
a ratio of:
10 mA => 1 V corresponds to 3.2 kHz.
The output pulses can be measured on the oscilloscope at measurement point F1 on
board D 910 ( 7 / 4 ). The pulses are digitally integrated in the microcontroller and the
exposure is ended when the mAs NOMINAL value is attained.
Microcontroller unit1
Central monitoring and control is processed via an 8-bit microcontroller ( 8 / 3 ).The controller communicates via a bus system with:
J8EPROM, contains the firmware,
J11RAM, not used,
J19Display-Controller.
High voltage, filament and tube current are controlled via parallel port inputs and outputs.
In addition, several interrupt inputs and an internal watchdog are available via the microcontroller. The watchdog monitors the software functions in POLYMOBIL III. There is no
external signal or display for the watchdog on board D 915.
The component J 3 on board D 910 monitors the voltage supply Vref = +5 V ( 6 / 3 ). If the
voltage drops below 4 V , a RESET is automatically triggered.
Exposure circuit1
Preparation:
When selecting preparation -ZB- on exposure release switch S 27 ( 8 / 4 ), an interrupt is
generated in the microcontroller.
If the microcontroller is not blocked by any error message, the tube filament will be heated
up in preparation for exposure.
The display controller J 19 ( 91 A / 3 ) triggers LED V 13 - ZB (91 A /7) on board D 910.
The exposure is released after a preparation time of 1.9 seconds has elapsed.
Exposure release:
When the main contact - HK- is pressed on the exposure release switch S 27 ( 8 / 4 ), the
main inverter, measurement point: TRIG ( 6 / 4 ), is released and the exposure begins.
Exposure shutoff:
When the mAs counter reaches the mAs NOMINAL value, the exposure is ended by the
microcontroller.
Siemens AGRXB8-115.041.02Page 5 of 6POLYMOBIL III
Medical EngineeringRev. 0310.97TDSP 1
1 - 6POLYMOBIL III
Interrupting exposure1
An exposure can be interrupted by one of the following events:
• The time limit (T-Limit) is reached before the mAs counter reaches the NOMINAL value.
A safety timer ends the exposure and the following error message appears:
ERR 17maximum exposure time.
>> T-Limit = 2 x theoretical exposure time + 100 ms <<
• kV-ACT > kV-MAX. As soon as this error occurs, the inverter is immediately blocked and
the error message:
ERR 12kV-MAX, is displayed.
• I TU > I TU-MAX. As soon as this error occurs, the inverter is immediately blocked and
the error message:
ERR 13I-MAX, is displayed.
• If the exposure is prematurely interrupted manually, the error:
ERR 18exposure not ended, is displayed.
Refer also to the service instructions, chapter 3.