Weitergabe sowie VervielfBltigung dieser Unterlage, Verwertung und
Mitt&lung ihrer Inhalts nicht gestattet. soweit nicht aurdriitklich
zugestanden Zuwiderhandlungen verpflichten zu Schadenerratz. Alle
Rechteftirden Fall der Patenterteilung oder GM-Eintragung vorbehalten
Copyright(C) Siemens AG 1994
Proprietary data. company-confidential.All rights reserved
Confie B titre de secretd’ entreprise Tour droitr r&w&
Confiadocomo secrete industrial. NOS reservamartodor lorderechos
English
06.94
Page 2
2
Startup instructions
Cover page
Table of contents, revision status
1
2
3
Requirements
General information
Required documentation
Required measuring instruments and devices
Scope of delivery
Safety notes
Notes on start-up
Preparations for dose rate setting
Notes on the VIDOMATIC P
Suppression of light distributor (220 V,not for 1.1.40)
Notes on dose rate setting
Startup
Programming the options
Fluoroscopy programming
Programming the lontomat P
Programming for SIRCAM (if connected)
Programming for digital image processing
(if connected)
Setting the max. skin dose rate
Final check of the skin dose rate
Setting the dose rate
Checking the dose rate for fluoroscopy
Checking the dose/image for SIRCAM
Calibration of the transfer
Setting the TV iris diaphragm
Mirror correction
Programming the fluoroscopic characteristic curves
High-contrast fluoroscopy
Therapy simulation
Generating separate fluoroscopic characteristic curves
(short description)
Page
1
2
1-I
l-1
l-l
l-2
l-2
1-2
1-2
2-1
2-1
2-l
2-2
3-1
3-1
3-1
3-4
3-5
3-6
3-7
3-8
3-8
3-11
3-12
3-13
3-14
3-16
3-17
3-l 8
3-19
3-19
Rev.
01
01
01
01
4
5
6
Siemens AG
Medical Engineering Group
Transfer curves (O-point technique)
General
Generating separate curves
Assignment of curves to organ button
Protocol
Programming
Options
Error notes
4-1
4-1
4-3
4-5
4-6
5-1
5-l
6-1
RX63-050.034.07.01.02
01
01
01
Rev. 01
Page 3
Requirements
General information
l
These instructions apply for
POLYDOROS SX 50180 generators
The SIRECON 2 connection is the extension of the basic generator (see
generator installation instructions, chap. 3-21) and contains the functions of the
POLYMATIC system:
. VIDOMATIC P
. SIRCAMATIC P
l
DIGIMATIC P (startup is described in the product documentation)
l
TV iris diaphragm
l
Radiographic system selection control
1-I
l
Light distributor control
For systems with RAS (x-ray system interface) components, see RAS
instructions.
Required documentation
l
TV-1.1. A connection wiring diagram Xl768
l
POLYDOROS SX wiring diagram X2075
l
PL SX system configuration with service PC RX63-050.034.06
Oscilloscope >50 MHz with delayed time base and post-triggering of the 2nd
time base (e.g. Tektronix 2232)
l
Service PC with connecting cable, e.g. 99 00 400 RE999
l
Adapter (fur N93)
l
Dosimeter Dali or NOMEX
1
cm 3 measuring chamber 97 02 341 Y0388 for skin dose
99 00 218 RE990
measurement
l
Digital multimeter (DMM)
l
Set of copper filters (10 PCS.), part no. 44 06 120 RV090
Scope of delivery
l
SIRECON 2 connection
l
M99 TV-1.1. connection = KK version
l
Standard cable: KK23 = KK version
l
N93 TV-1.1. connection = RAS version
. Yl
l
N 21 SIRCAM 103 or 106 control
l
2 standard cables:
SIRCAM 103 or 106 camera
KK 42, KK 45
Safety notes
Observe the “Safety notes” contained in the installation instructions for the
generator.
l
All checks and adjustments to be carried out with x-radiation switched on, are
marked with the radiation warning symbol ‘;;’ . Provide for suitable radiation
protection measures during these procedures (see TI 216).
Notes on start-up
l
For systems already prestaged and tested according to BZ at the factory (can
be taken from the “test certificate, jumper protocol” contained in the
“Certificates” register) only slight checks and adjustments are necessary.
l
Record the measured values resulting from the measurements marked with
the ” > W symbol in the “Measured value/start-up” column contained in the test
certificate
Siemens
Medical Engineering Group
AG
RX63-050.034.07.01.02
Rev. 02
Page 5
Preparations for dose rate setting
Notes on the VIDOMATIC P
l
For systems in KK version, 2 fluoroscopic characteristic curves can be
programmed
(Fluoro 1 and Fluoro 2).
l
For systems in RAS version, up to 4 fluoroscopic characteristic curves can
be programmed if they can be selected from the unit (Fluoro 1. . . Fluoro 4)
Normal fluoroscopy (max. 450 W)
l
Can be selected from fluoroscopic characteristic curves in the firmware
(COO, CO1 , C02, C03, Cl 0, Cl 1) or
High-contrast fluoroscopy (max. 1200 W with changed high-voltage generator;
as from Serial No. 01056, max. 2000 W)
l
Can be selected from fluoroscopic characteristic curves in the firmware
For both modes of operation, separate curves can be generated:
- within the range C80 . . .
- within the range C90 . . .
l
Dose-saving zoom can be programmed with existing iris diaphragm.
C89 for normal fluoroscopy
C99 for high-contrast fluoroscopy
Therapy simulation
l
It is possible to program 2 fluoroscopic curves which can be selected from:
- fluoroscopic curves in the firmware (C40, C41, C42, C50, C60, C61, C62,
C63, C64, C65).
- fluoroscopic curves within the range of C90 . . . C99 (separate curves)
l
Dose-saving zoom can be programmed with existing iris diaphragm.
Suppression of light distributor (220 V, not for 1.1. 40)
Voltage peaks caused by switching off the motor voltage of the mirror motor in
the light distributor result in generator interferences.
l
For this reason, generally incorporate two RC elements, part no. 31 33 147
B2003 parallel to the motor (Fig. 3)
For 220V mirror motor voltage, reconnect
jumper x7 on board D271
Siemens
Medical Engineering Group
AG
0.22pF
2
1
V-7
24V mirror (40~cm 1.1.)
3
220V mirror
l
RX63-050.034.07.01.02
Rev. 02
Page 6
2-2
Preparations for dose rate setting
Notes on dose rate setting
Requirement:
For determination of the dose at the 1.1. input ( (KB) or the dose rate at the 1.1.
input (KrS), position the dose measuring chamber as close as possible to the
image intensifier; in the case of spotfilm devices in the cassette shaft:
- With OV units, on the tabletop,
- with UT units, on the spotfilm device rear panel,
- with C-arm units, on the grid.
By means of the correction factor m (unit attenuation factor), the value resulting at
the 1.1. input can thus be determined.
The unit attenuation factor m includes all attenuating layers lying between the
patient and the image receptor, and the distance factor as well. The unit
attenuation factor is indicated in the unit test certificates; it can also be calculated
by multiplying the attenuation factors of all individual layers.
The basic generator and the IONTOMAT as well as the l.l.-TV
system must function properly.
m
Unit
attenuation
factor
KB/KB
KB#B~
K~gf KT~
f-B
rT
rl
rB
f
T
= dose/dose rate directly at the 1.1. input
= measured dose/dose rate in the 1.1. plane
= measured dose/dose rate in front of the attenuating layers.
= source-image receptor distance
= distance between source and start of the attenuating layers
Cu filter
Meas. chamber in front of the attenuating layers
Tabletop or spotfilm device rear panel
Grid
lontomat chamber
Measuring chamber in 1.1. plane (cassette shaft)
1.1. input
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 02
Page 7
Preparations for dose rate setting
Determining the correction factor
Correction factor m (total) =
x m-factor (tabletop etc., if between dose measuring chamber and 1.1.) etc.
- Distance correction factor (mA)
2-3
source-1.1. distance (rB)
source-meas. chamber
distance (rT)
- Grid correction factor (mB) = 1.7
- IONTOMAT chamber correction factor (ml) = 1 .l
- Tabletop correction factor (m-r) = 1 .l
When setting the dose/dose rate, multiply the dose/dose rate required at the 1.1.
input by the correction factors.
e.g. 174
nGy/s x (mA x mn x ml x mr)
174nGy/sx(1.17x1.7xl.l x1.1)=
174 nGy/s x 2.4
(1 pR = 8.7
nGy)
2
Example:
=
= 418 nGyls
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 02
Page 8
Preparations for dose rate setting
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev.02
Page 9
Startup
Programming the options
l
Establish connection between generator and service PC.
0
“System”/“Connect”:ON “Connect,BACKUP and make...”
l
Call up “Configuration” on the service PC and program correspondingly.
I Definition of the dose rate units
0
L-~_.--_-. -~~--_- _____
&stem Pata Lontiqulalion ‘. Dlepno~lcs Info tilndow
0 with ml-PI tsy and Orysnproyfamcbsnyc
3-l
1
Guide&Reference...
,,.:.. : :..
0
E~posurc Rclcssc
0 al Generator
Measuring field selection
0 at Generator
DiagnostiA !nfo Mr
Unit 2
0
Unit 3
SIREGFIWH D3
0 at Unil
0
Unit 4
@I Wilb O-R key and Ovjanplogranlchange
1
l
Perform the system programming.
(With system delivery, the system has already been programmed at the factoVI
l
With “Data”/“Restore”, load the “system configuration” from PC into the
generator after programming.
Siemens AG RX63-050.034.07.01.02
Medical Engineering Group
Rev. 01
Page 10
3-2
Startup
Programming of fluoroscopy
Icm Qata Z;onfiguration .:. :::.: .:
__. .,__ ..” - _...
Unit selection button:1
Notes on:
1 Programming the kind of FL
0
- Therapy simulation :
- High-contrast FC
: Programmable for units with MEGALIX tube assem-
-‘. -I--...
Diagnostics !nto Win
4
0
5
0
6
0
Unit for therapy simulation
bly and FL footswitch with 2nd pressure point
2 Programming the max. FL output for all operating modes:
0
- 450 W for normal FL
- With therapy simulation, an FL output of max. 1000 W can be programmed for tube assembly OPTILIX 150/12/50 C.
- Max. 2000 W for high-contrast FL
(with high-voltage generation from ser. no. 01056)
3 Programmable for units with feedback of the current source-1.1. distance to
0
the generator via RAS, currently only POLYSTAR
MPID = (Measuring Point Image Distance) always a distance of 0300 mm
lonization chamber to the 1.1. input
5 Enter SID and set identically to dose adjustment:
0
OT-units = 115 cm
UT-units = max. SID (no distator)
e.g. SIRESKOP 5/45 = 115 cm
C-arm units
6 Programming takes place after “Setting the dose rate”
0
and “Calibration of the takeover”(3-13).
Siemens AG
Medical Engineering Group
SIRESKOP 5 = 100 cm
= 100 cm
RX63-050.034.07.01.02
Rev.
01
Page 11
Startup
3-3
i
i
Doscratc: 07
Normal Curve: 11
Ls] TS/HC-Curve: 42
tic-Doserate: 139
Zoom doserate:
: 4 : 5
1:2
FC-lonlomatchanncl
OD OE @F
1
C
Doserate: 87
Normal Curve: 10
Ix] TS/HC-Curve: 41
HC-Dosers\c: 139
Automatic 1
Doserate: 87
Normal Curve: 3
[x1 TS/HC-CUIVC: 41
HC-Ooseratc: 139
Jmagc-Intensifier (II-Type]:
40/28/20/l 4
I
nGds
nGyls
nGyls
$I
Notes on :
u With input of the 1.1. type, dose rate factors are preset automatically accor-
ding tothe 1.1. input areas.
The dose rate factors resulting from the 1.1. formats are displayed in the
“Zoom dose rate”mask.
It is possible to modify the factors independently of one another for dose-
saving zoom.
Prerequisite for this is a TV iris diaphragm.
@ The dose rate selected for fluoroscopy always refers to the largest 1.1. format!
(see
3 -4)
l
For automatic 1 and 2 (3/4), set the same dose rate values.
Urological systems: If the TV iris diaphragm is used for automatic 2,
DSA systems:
l
FL characteristics Normal curve :
@ FL blocking always
(Blocking takes place after max. 10 min. according to the German law, within
the area of validity of the DHHS regulations after 5 min.)
5 Select channel in which the multiplier or SDM cable has been inserted.
0
Siemens AG
Medical Engineering Group
set the double dose rate.
For the control stage with high-contrast FL (stage 2)
program the dose rate according to the unit-specific
documentation
see 3 - 17
TS/HC curve : see3-18
q
RX63-050.034.07.01.02
Rev. 01
Page 12
Startup
33cm HDR
27cm
SIRECON
Kompakt (alt/old) 17
23cm HDR
Sirecon Kompakt 17
neu/new 13
UMultiplier =
700
Tolerance:
33
22
17
27
17
22
23
- 950 v *1
113
235
435 13
680 (430)
174
445 14
655 (435)
261
435
261
440
730 (470)
1 nGyls
= 0.115 pR/s / 1 pR/s = 8.7 nGy/s
Full 1.1. format + 10%; with zoom formats only standard values.
Values in brackets ( ) apply for TV iris diaphragm.
*l applies from 7.93 (possible only with dark multiplier)
1:
2.3:
6.43i:.8)
1:
,.,‘i;:,,
1:
1.7
1:
3.1’$:8)
Programming of lontomat P
l
Select “Configuration” / “lontomat’”
l
Under “CHANNEL-Prog”, program the
selected detector multiplier or PDA (SDM)
Siemens
Medical Engineering Group
AG
RX63050.034.07.01.02
Rev. 01
Page 13
Startup
3-5
Programming for SIRCAM (if connected)
.:: :‘.
1
0 Unit 2 0 Unit 3
EL(posurc Rclcasc
0 al Generator
c
-Measuring field scleciion
0
Diagnostics Info &in
SIREGRAPH D3
at Generator
l
l
Perform programming for SIRCAM connected.
Depending on
BV-Type /
1.1. Type
0
0 at Unit
Unit 4
SIRCAM-Type
0
Slrcam 103
0 Slrcam 106
the 1.1. used. we recomm ?d the following setti
@l, @2
Formate / Dosisleistung/
formats
Dose rate
VW4
Framespeeds -
1:
2:
1
3:
4:
r----
SIIICAU dose:
~-___
POLYDOROS SEX
Zoom Dose
Factors
(Polymatic)
Igs :
40 cm HD
40
218 1
28 400 2
20
790
14 1600 (1090) 8.24(5)
33 cm HD
33
23
261 1
560 2.1
17 1050 3.8
33 cm HDR
33
261 1
22 545
1010
1565 (990) 6.43(:.8)
1110
27 cm
17
13
27 435
17
14 1635 (1090) 3.7 (2.5)
23 cm HDR
23
574 1:
17 965
13
l
With use of a diaphragm, a reduced dose/frame (
1610 (1030) 3.1’(::8)
ding on the zoom position: “Zoom dose factors”
2.3
1
2.5
) can be preset depen-
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 01
Page 14
Programming for digital image processing (if connected)
Diagnostics !nfo Wil
0
Unit 2 0 Unit 3
0
Unit 4
Startup
I
Exposure Release
-Measuring field selection
SIREGRAPH 03
0 al Generalor
0 al Generator
With selection of the imaging
systems:
- POLYTRON or
- FLUOROSPOT
all details are preallocated.
0 at Unit
I
stem Data Contiouration :,:j.. .: :, .: Diannoslics lnlo tiindw
w Pcrlphcral gngingrnphic
BP reduction per step:
step 1:
Slep 2:
step 3:
step 4: 4 EPIEJPI
II
_.. . _.... . _r .
1 EP[BP)
1
EP(Bp)
3 rr~llrl
System-specific indications are made in the instructions for the imaging system
connected.
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 01
Page 15
Startup 3-7
Setting the maximum skin dose rate
The maximally admissible values are to be set according to the country-specific
regulations.
Within the area of validity of the DHHS regulations, 9 R/min. (1320 pGy/s) must be
set.
If there are no regulations, the module must be performed with “SS” OFF. For
110 kV and 70 kV, 18 mA must be entered in each case.
Note: Before measuring the dose, the dosimeter must be switched on for 20
min. at least.
l
Set minimum source-image distance on the unit (SID) : with POLYSTAR
0866
mm)
l
Attach ionization chamber at a distance of 30 cm centrally to the 1.1.
This distance must be maintained independently of the SID distance.
l
Open collimator until the ionization chamber is fully exposed.
l
Remove fuses from the 1.1. power supply.
l
Cover 1.1. input additionally, e.g. with lead rubber (due to risk of burn-inr).
l
On the service PC, select “Adjustment “/” Skindoserate”.
Feedback of the current SID by the unit with program. SID tracking
1
0
in “Unit selection“ / “Primary FC-Data”
input.
Within the area of validity of the DHHS regulations, the following applies in
2
0
addition: With high-contrast FL:
Reduced skin dose rate 5 RImin. (733 pGy/s) for normal fluoroscopy
(1 st pressure point)
With manual mode:
3
0
Reduced skin dose rate 5 R/min. (733 gGy/.s)
b.4
.
0
Fluoroscopy “ON/OFF” on board D211.
0
By means of key +/+ on the PC, modify the mA until a maximum skin dose
rate of 9 R/min. (1320 pGy/s) is measured.
Adaptation for @ @ takes place automatically if programming has
beenperformed with (XI
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev.
01
Page 16
3-8
4.4
.
Startup
Final check of the skin dose rate
l
Modify the source-image distance and check the dose rate. Maintain distance
between ionization chamber and 1.1.
A dose rate of 9 Wmin must not be exceeded with none of the distances.
l
Finally remove the measuring devices and reinsert the fuses in the 1.1. power
supply.
Dose rate setting
Preparation:
l
Select 1.1. full format
l
Source-1.1. distance: OT units = 115 cm
UT units = max. SID (no distator)
C-arm units = 100 cm
l
Attach Dali measurement chamber centrally in relation to the 1.1. input and
connect to the Dali measuring device (collimate to chamber measuring field
with collimator)
l
Attach 1.2 mm Cu to the collimator (p 4 plates, 0.3 mm)
l
Collimate to measurement chamber field
a) for multiplier
l
Select “Adjustment” / “Doserate”.
SlFlEGFtAPH 03
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev.
01
Page 17
Startup
4.4
*
l
Perform adjustment via PC by means of the on-line instructions
l
By means of the tube current ( f mA), set 174 nGy/s or 20 @s,
(observe unit attenuation factors)
3-9
Note:
Fluoroscopy “ON/OFF” always from PC board 0211.
Attention: Perform dose adjustment of 770 nGls or 20 $3/s as accurately as
possible, since all dose values are derived from this.
With automatic adjustment, adjustment must
result
for multiplier
ranges 7 and 2, that means the adjusted multiplier voltages are
displayed.
If no adjustment takes place, check among others the multiplier and
the cabling (hum, drift measurement) (see generator setting
instructions: lontomat startup).
Notes on software version VCOOB
During dose rate adjustment with multiplier, it is possible that 20 $3/s are not increased up to200 $3/s (174 nGy/s up to 1740 nGy/s) when fluoroscopy is swit-
ched on. The tube current is not increased and the LED in the lower right area on
PC board D901 does not flash.
Remedy:
l
Leave adjustment module
l
Reinitiaize generator with “System”/“Disconnect”
l
Select “System”/“Connect” with “Connect only”
l
Reselect “Adjustment”/“Doserate” and perform adjustment
If this is not successful, leave the service mode and switch the generator off
and on again.
For adjustment of multiplier area 2 (AEREA Z), proceed additionally as follows:
l
Establish connection between D271.X6.2 (MPL ref) and D271.X5 (GND).
l
Within 7 . ..lO s, fluoroscopy ON
l
After fluoroscopy ON, separate connection 0...2 s later.
l
Fluoroscopy OFF
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev.
01
Page 18
3-10
Startup
b) SDM
Information:
SDM (Selective Dominant Measurement) is an acquisition system for the initial
brightness of the image intensifier. A PDA (Photo Diode Array) is used as recei-
ver which assumes basically the function of the multiplier.
The PDA consists of individual segments which can be selected also parallel as
measuring fields.
The size of the measuring fields is switched over depending on the zoom.
Prerequisite:
l
PDA must be programmable for fluoroscopy in the mask “lontomat” / “CHAN-
NEL Prog. ‘I.
0
“Invert MF”must be selected with
Preparations:
l
Do measuring arrangement identical to multiplier
l
Connect DVM to D901 X2.1 (OV) and X2.7 (signal).
In stand-by (FL OFF), OV, + 1OmV must be displayed.
If more than 10mV is displayed, there is possibly external light on PDA.
q
,if POLYSTAR existing.
b.4
.
rFC4JNIT----- I
tkasurcflelds:
l
Perform adjustment via PC by means of the on-line instructions.
l
By means of the tube current ( f. mA), set 870nG/s or lOO~R/s.
I’OL YSl AH
J
(observe unit attenuation factors)
Note: Fluoroscopy “ON/OFF” always from PC board 0217
l
Check after adjustment, with ionization chamber removed :
- DVM to D901 X2.1 and X2.7
- FL ON on D211, fully open collimator
- with Rl (on DlOO in the 1.1. light distributor), set 700mV+ 35mV
. FLOFF
l
Deselect ” PDA Preadjust”
l
Attach hood of the 1.1. ligtht distributor and check setting of the 700 mV.
l
Perform sensitivity adjustment for all 1.1. formats by means of the on-line in-
structions.
Siemens
Medical Engineering Group
AG
RX63-050.034.07.01.02
Rev. 01
Page 19
Startup
Checking the dose rate for fluoroscopy
l
FL curve: Always program the anti-isowatt curve (00) for dose rate
measurements.
l
Record the programmed dose rate value and the programmed zoom-dose
factors (image quality test certificate 4.1.2)
l
ADC measurement chamber; with 40 cm I.I., select central measurement
field.
l
Move in grid.
3-I 1
A.4
.
l
Source-1.1. distance:
OT units = 115 cm,
UT units = min. SID (no Distator).
C-arm units = 100 cm
l
Attach 2.1 mm Cu to the collimator (with UT units on the tabletop) .
l
Attach dose measuring chamber centrally to the 1.1. and collimate to ionization
chamber size.
l
If measurement cannot be performed directly at the 1.1. input, perform
measurement on the tabletop or spotfilm device rear panel and convert into
the 1.1. input by means of the unit attenuation factor (m).
l
Measure and record the dose rate values for all 1.1. formats (KQ or Krg).
l
Calculate and record the dose rate at the 1.1. input (KB).
Note: With 12 cm 1.1. format and with 14 cm 1.1. format (if measurement is not
performed directly at the 1.1. input), the dose measuring chamber is not
exposed completely. As a result, the measured dose rate values are
lower than the true ones.
indicate this fact under “Remarks” in the test certificate.
l
Record the dose rate values and curves programmed at the customer for all
steps (see 3-17 and 3-l 8) (image quality test certificate, 4.2).
l
Remove the dose measuring chamber from the beam path.
l
Open the collimator fully and record the entire kV and mA values resulting
during fluoroscopy for all 1.1. formats in the image quality test certificate, 4.2.
This table contains the values for indirect dose checking for future servicing.
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev.
01
Page 20
3-12 Startup
Checking the dose/image for SIRCAM
l
For measuring arrangement, see “Checking the dose rate”.
l
Select SIRCAM
l
Call up (or generate) zero-point technique organ program
- small focus
- 1 f/s or single frame
- 70 kV plateau (Cl4 with preliminary version display in HEX : OE ,
selectable with ms buttons)
44
.
l
Measure and record dose/image for all 1.1. formats (Kag or KT~).
If the pointer of the dose measuring device does not deflect enough, make
several exposures and divide the measured dose value by the number of
exposures.
l
Calculate and record the dose at the 1.1. input:
KB= KB~
x 0.94;
KB = Kra x 0.94
m
Factor 0.94= 6% radiation attenuation caused by dose measuring chamber.
l
Remove the dose measuring chamber from the beam path and record the
indicated kV values and the resulting mAs values for one exposure each for
all 1.1. formats in the image quality test certificate (7.1 .I .2). If the mAs values
are not indicated, measure the mAs values with the mAs meter.
Note: With 12 cm 1. I. format and with 14 cm 1.1. format (if measurement is not
performed direct/y at the 1.1. input), the dose measuring chamber is not
exposed completely. As a result, the measured dose rate values are lower
than the true ones.
Indicate this fact under “remarks” in the test certificate.
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 01
Page 21
Startup
Cali bration of transfer
0
Place 20 cm (must be maintained exactly!) water phantom in the beam path.
l
Set the source-1.1. distance entered in the mask “Primary FC-Data”/” SID”.
0
Allocate arbitrary organ field with transfer curve COO and select.
For
DR
systems, transfer curve C29 can be selected.
Grid must be moved in
Select zero-point technique
Select full format
Fluoroscopy ON using the footswitch or from the spotfilm device
Collimate to water phantom; 70 kV must be obtained as radiographic voltage.
+ deviation of 0 EP.
3-13
0
In the case of deviations, read off the corresponding exposure points (EP)
from the graph, kV graduation in integral EP, and enter the values in the
Measure B-signal values with expanded time base (higher measurement ac-
of one line in the image center (noise center) at the TP8 of the W-PC
bias light portion, see table).
l
FL ON (on PC board D211 in the generator)
l
On the service PC, vary dia
uired B-si
ii 9
-Signal va ues without bias-light or dark current
nal is measure
hragm opening by means of key +/ + until re-
8
B-Signals
Hivicon 1 180 mV + 15/ -15 mV 1
I
110 mV +5/-10 mV
max. dark current 60 mV
20 +
Bias light
1
5mV
110 mV +5/-10 mV* 40 + 5 mV with DFR
20 + 5 mV without DFR
(DG-Saticon [ 130mV +5/-lOmV*I 30 f 5mV
Plumbicon
150 mV . . . 180 mV
I
I I
I
* With high-contrast FL: + 15/ -10 mV
l
After performing the module, leave the system programming (“System’ /“Disconnect”) and check the values set.
Checking the B-signal values
l
increase Cu up to 2.1 mm
l
Select FL characteristic COO.
l
FL on the unit ON, measure B-signal with all 1.1. formats and record it in the IQ
test certificate (53.2.).
Mirror correction
With DR exposures, the iris position to be expected is inquired. From a certain opening (> 60 mm, with SDM > 50 mm), the mirror is moved out automati-
cally and the iris opening becomes smaller.
l
Perform mirror correction also if there is no mirror.
Siemens AG
Medical Engineering Group
km Uata Qmfiguration ~djtrstment Diagnostics !nlo Wandow
: :
~~~..~~~~~....~ li _._.__. :...::..:..::.:.r:;:_ ,__.._.......___~..~.i_...............................’ .__________..................~:
Normally, two FL curves are used.
Depending on the unit, selection of more characteristic curves is possible with
systems in RAS version, e.g. POLYSTAR with three automatic buttons.
If a footswitch with two preps (pressure points) is connected, high-contrast
fluoroscopy is possible with a separate characteristic curve at the 2nd prep.
l
For determining the characteristic curves, please contact the customer and
program and record the characteristic curves (in the image quality test
certificate 4.2. and in the operating instructions as well).
l
Die Programmierung erfolgt unter “Configuration” / “Unit selection” und “FCDetails” (3-l bis 3-3)
l
Select module JOO.
l
Program the FL curves, as far as provided, for automatic 1 - 4.
Normal fluoroscopy
Recommendation:
Preferably program the following characteristic curves:
- Automatic 1: Cl 0
- Automatic 2: CO2
The same dose rate for automatic 1 and automatic 2 (autom. 3 - 4)
1lOkV
90
77
66
57
50
44
40
10 -1
Fig. 1
Application of the characteristic curves:
COl, ClO:
C02, C03, Cl 1: Curve with especially high contrast
coo:
Siemens AG
Medical Engineering
Group
1
10 mA
Curve with low exposure to radiation, especially appropriate for
pediatrics.
Anti-isowatt curve, as known from the OPTIMATIC-SYSTEM.
RX63-050.034.07.01.02
Rev. 01
Page 26
3-18
Startup
High-contrast fluoroscopy
A MEGALIX tube assembly with circulation cooling system is required for this
operating mode.lt is selected at the 2nd prep of a 2-step footswitch.
1 Selection of this mode of operation is displayed acoustically, with IZI HC gond
Fi
!a 2
0.1
1 I II 1111
/I I II bill
l
I
1 2 3
I II III1
I
4 10 14 18
I ‘1
mA
The characteristic curves shown in Fig. 2 are stored under numbers C4O....C79.
Application of the characteristic curves
C40 : Anti-isowatt curve, as known from the OPTIMATIC SYSTEM
c41 :
Curve with low exposure to radiation
C42 : Curve with especially high contrast
Additional curves are stored as follows (for the sake of clarity, they are not drawn
- For definition of the characteristic, 2 points at least,
3
0
4 - Confirm programming of the respective entry
0
max. 10 points must be entered in ascending order.
The characteristics must be located within the respective range
y
I:.; ::....-.: : .:.
:.
i.;...:. : .-’
1
.
1..
1. ,. j
10 mA11
llOk\
5 -
0
90
85
77
66
40 I
Leave mask with “Cancel”
2 3 4
I
I
I
18
mA
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 01
Page 29
Startup
l
After programming, the mask displaying the curve and the list of the program-
med points is shown.
3-21
Inter Name:
Test
Points of Curve:
Confirm with “OK” and store (“Data”/“Restore editable...” )
Under “Configuration” / “Unit selection” / “FC-Details”, allocate the characteristic to the automatic stage requested and program. (3-1 to 3-3)
Record the modified or new characteristics also in the operating instructions.
The FL characeristics can be printed out via “Data”/“Print”.
Note: Characteristics can also be produced graphically with the mouse.
Select “Insert” and select the points of the characteristics.
Siemens
Medical Engineering Group
AG
RX63-050.034.07.01.02
Rev.
01
Page 30
3-22
Startup
Siemens AG
Medical Engineering Group
RX63-050.034.07.01.02
Rev. 01
Page 31
Transfer curves (O-point technique)
General
POLYMATIC curves (EXP-Curves)
150 kV
133
102
90
81
73
66
60
55
50
46
42
40
0 10
,-
C61
20 30 40
4-1
/
VVW in cm Hz0
General
In the Polymatic system, the object thickness is determined in fluoroscopy. For
direct and indirect technique, the exposure parameters are determined on the
basis of this value:
Direct technique:
l
Radiographic kV
l
Density corrrection
Memory locations for 99 curves are available.
These memory locations are divided as follows:
l
Curves 01...79 are stored in the firmware.
* Curve 00 is a check curve for checking the transfer.
For FLUOROSPOT operation, the curve can be used as C 29..
These characteristics must not be programmed for operation.
* Under the numbers 01...27, constant kV curves are stored in ascending
order (see table) at kV point intervals of 40 kV to 150 kV. They include:
small focus, auto focus, no density correction. As delta kV value, the
difference to the next higher kV value + 1 kV is entered.
These curves are used for all indirect exposure techniques.
Indirect technique
l
Radiographic kV
. Focus
l
Radiographic current
l
Density correction
l
Exposure time
Siemens AG
Medizinische Technik
RX63-050.034.07.01.02
Rev.
02
Page 32
4-2
Transfer curves (O-point technique)
General
* The numbers 31 to 57 include constant kV curves at kV-point intervals of
40.5 kV to 145 kV (see table). The other conditions correspond to curves
01 . ..27.
* For various medical applications, the following curves are programmed
additionally (see page 6-l):
C60: gastrointestinal tract I 85...125 kV
C61: thorax 66...125 kV
C62: gastrointestinal tract II 70...109 kV
C63: gallbladder/kidney 60...70 kV
l
The numbers 80 to 99 are reserved to curves which can be programmed
freely as requested by the customer.
It is possible to use existing curves unchanged or to program separate
curves.
In the following, generation of the customer-specific curves is described.
The following constant kV curves are filed in memory locations 01 to 27 and 31 to
56:
Characteristic curves
CO1 ............ 40 kV
CO2 ............ 41
CO3 ............ 42
CO4 ............ 44
CO5
CO6 ............ 48
CO7 ............ 50
CO8 ............ 52
............ 46 kV
kV
kV
kV
kV
kV
kV
CO9 ............ 55 kV
Cl0 ............ 57 kV
Cl1 ............ 60kV
Cl2 ............ 63 kV
Cl3 ............ 66 kV
Cl4 ............ 70 kV
Cl5 ............ 73kV
Cl6 ............ 77 kV
Cl7 ............ 81 kV
Cl8 ............ 85 kV
Cl9 ............ 90 kV
C20
C21
C22
............ 96 kV
........... 102 kV
........... 109 kV
C23 ........... 117kV
C24
C25
C26
C27
........... 125 kV
........... 133 kV
........... 141 kV
........... 150 kV
C31 ............ 40.5 kV
C32 ............ 41.5
C33 ............ 43
C34 ............ 45
C35 ............ 47
C36
C37 ............ 51
C38 ............ 53.5
C39 ............ 56
C40 ............ 58.5
............ 49 kV
kV
kV
kV
kV
kV
kV
kV
kV
C41 ............ 61.5 kV
C42 ............ 64.5
C43 ............ 68
C44
C45 ............ 75
C46
C47 ............ 83
C48
C49
C50
C51 ............ 105
C52 ............ 113
C53 ............ 121
C54
C55 ............ 137
C56 ............ 145
............ 71.5 kV
............ 79 kV
............
............
87.5
93 kV
............ 99 kV
............ 129 kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
Siemens AG
Medizinische Technik
RX63-050.034.02.02.02
Rev. 02
Page 33
Transfer curves (O-point technique)
Generating separate characteristics
l
Separate characteristics can be generated within the memory range 80 to 99.
l
Selection takes place as displayed in the following:
1 EXP-Curves
0
anwahlen
4-3
Edilable EXP - Curves (EPPROM):
I
CR1 - could not be read1
CB2 - could not be read1
Cfl3 - could not be read!
CfI4 - could not be read1
CE5 - could not be read1
CEl6 - could nol be read1
CR7 - could not be read!
CBII - could nol be read1
ClJ9 - could not be read!
C90 - could not be read!
C91 - could not be read1
Siemens AG
Medizinische Technik
RX63-050.034.07.01.02
Rev. 02
Page 34
4-4
Transfer curves (O-point technique)
Generating separate characteristics
0
After selection of the characteristic to be programmed, the following mask is
displayed:
~
Points of Curve: (PO-> p91
Voltaqe
1 Watervalue I Dens Ad.:
1 -
0
2 -
0
3
0
4,
0
5 -
0
Enter Name: not programmed
Nominal Value: [NPXJ
kV: WW: Dens Ad.:
not programmed
Enter designation of the characteristic
Program nominal value (for exposure without previous fluoroscopy)
e.g. 60kV for mean water value, e.g. 20 EP (20 cm H20)
and confirm with “ok”
Select “New Curve” (the display of the NPO values disappears)
Enter data of the characteristic with ascending water value, in each case
confirm with “ok”.
6 -
0
0
Siemens
Medizinische Technik
AG
Terminate programming
, Voltage:
_ Watetvalue:
6
r] kV
[zo ) BP
Points 01 Curve: (PO-> PS]
Voltage f Watervalue I Dens Ad.:
RX63-050.034.02.02.02
Rev. 02
Page 35
Transfer curves (O-point technique)
l
After programming, the mask displaying the characteristic and the list of the
programmed points is shown.
Enter Name: test
Points of Curve: [PO-> PS)
Ei
FOCUS
@I r&j,;
I..
Focus Switching 1
@ Eix Focus
0 Auto Focus
Delta k’j: (i!]
0
Small
Voltage 1 Watervalue 1 Dens Ad.:
“PO: 42.OkV! 1 BP I D.A. -1 .O
‘Pl: 60.2kV 18 BP i D.A. -0.5
NP2: 60.2kV 1 20 BP / D.A. to.0
“P3: 60.2kV I30 BP / D.A. to.5
*~4: i 24.8kv I 40 BP I D.A. ti
4-5
.o
l
Program focus: with ” Autofocus”,
enter switchover from small to large focus
(Delta kV, default 0).
l
Confirm with “OK” and store.
Assignment of curves to organ buttons
With the following sequence, the curves are assigned (see also operating