s
INFINITY Medside Data Station
Windows 2000 Version
Service Manual
EM Guidelines, 1997-04-02
ASK-T962-03-7600
ADVISORY
This document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Medside Data Station |
Field Service Manual |
|
|
Table of Contents
1 |
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . .1 |
||
2 |
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .1 |
||
|
2.1 |
Power Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .1 |
|
|
|
2.1.1 No Response when power On/Off switch toggled ON . . . . . . . . . . . . . . |
. . . . .1 |
|
|
|
2.1.2 Power On/Off Piezo Tone Fails to Sound . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .2 |
|
|
|
2.1.3 Power-Up Sequence Fails to Complete Properly . . . . . . . . . . . . . . . . . . |
. . . . .2 |
|
|
|
2.1.4 |
No Video display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . 3 |
|
|
2.1.5 MDS Fails to boot properly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .3 |
|
|
2.2 |
BIOS Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .3 |
|
3 |
MDS Installation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .7 |
||
4 |
MDS Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 7 |
||
|
4.1 |
MDS Mounting Arm Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 7 |
|
|
4.2 |
MDS Table Top Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .8 |
|
5 |
Monitor Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .8 |
||
|
5.1 |
CRT (Art. No. 57 35 894 E5310) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . .9 |
|
|
5.2 |
Flat Screen Display (Art. No. 59 55 567 E531U) . . . . . . . . . . . . . . . . . . . . . . . . |
. . . .10 |
|
|
5.3 |
Locally Supplied Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 12 |
|
6 |
Keyboard, Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 12 |
||
|
6.1 |
Wall Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 12 |
|
|
6.2 |
Table Top . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 12 |
|
7 |
Connecting Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 12 |
||
8 |
Reinstallation Windows 2000 Operating Systems and Installing Optional Drivers |
. . . . .14 |
||
|
8.1 |
Laptop Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 14 |
|
|
|
8.1.1 |
TCP/IP Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . .15 |
|
|
8.1.2 |
CDROM Share Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 16 |
|
|
8.1.3 |
Install MDS Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . 18 |
|
|
|
||
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
i |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
||
|
|
|
|
8.2 |
MDS to Service Laptop Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . 19 |
|
8.3 |
Launch MDS Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . . 20 |
|
8.4 |
MDS Network Boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .21 |
|
8.5 |
Mapping MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 21 |
|
8.6 |
Install Windows 2000 Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 22 |
|
9 Phoenix BIOS Phlash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .32 |
||
9.1 |
Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 32 |
|
9.2 |
Software Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 32 |
|
|
9.2.1 |
CDROM Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 32 |
|
9.2.2 |
Download Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 33 |
9.3 |
MDS Hard Drive Phlash Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .34 |
|
10 Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .36 |
||
10.1 Opening MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .36 |
||
10.2 Replacing Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .38 |
||
10.3 Replacing Hard Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .39 |
||
10.4 Replacing Memory / Daughterboard . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 40 |
||
10.5 Closing MDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . . 41 |
||
11 Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .42 |
||
12 Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .43 |
||
Appendix A: Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . 47 |
||
Appendix B: BIOS Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . .51 |
||
Appendix C: POST Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . |
. . . . . . . . . 55 |
ii |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
In keeping with the Service Strategy for the Medside Data Station (MDS), this Service Manual provides the necessary information required to troubleshoot and service a Windows 2000 based MDS. The MDS is powered by an AC/DC power adapter, and can be placed on a desktop or attached to a wall bracket. Control of all Medside Data Station functions is done by means of a keyboard and a mouse. (It is recommended that the keyboard and mouse be purchased through SIEMENS, to avoid any possible incompatibility problems). The display screen has a 1280 x 1024 resolution capabilities which enables the user to display and run applications such as INFINITYTM Explorer and Webview.
For the purpose of clarification, special text in this Service Manual is described below:
Bold Characters |
text that is to be typed in by the User. |
^ Character |
a space required between typed characters. |
Italic Characters |
a selection that is required by the User. |
If the Medside Data Station should fail to respond properly, use the procedures below to aid in identifying and remedying the problem.
2.1.1 No Response when power |
Refer to Table 2-1 to troubleshoot Power-On problems. |
|
|||
|
On/Off switch toggled ON |
|
|
|
|
Table 2-1 Power-On Problem |
|
|
|
||
|
|
|
|
|
|
|
Conditions |
Possible Cause(s) |
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS connected |
Power source. |
1) Assure Power Adapter is connected to an active |
|
|
|
directly to Power |
Power Adapter |
hospital power source. |
|
|
|
Adapter; Power |
|
|
||
|
malfunction. |
2) If problem persists, disconnect power adapter from |
|
||
|
Adapter LED not |
|
|||
|
MDS Malfunction. |
MDS and measure voltage between Power Adapter |
|
||
|
illuminated |
|
|||
|
|
|
output pins. |
|
|
|
|
|
|
|
|
|
|
|
|
• If voltage q 11.6 VDC or Q 13.8 VDC, replace |
|
|
|
|
|
Power Adapter. |
|
|
|
|
|
• If voltage = 11.6 to 13.8 VDC, contact TSS in |
|
|
|
|
|
Solna or Danvers. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS directly |
Power Adapter |
1) Disconnect power adapter from MDS and measure |
|
|
|
connected to Power |
malfunction. |
voltage between Power Adapter output pins. |
|
|
|
Adapter; Power |
MDS Power Switch |
• If voltage q 11.6 VDC or Q 13.8 VDC, replace |
|
|
|
Adapter LED On, MDS |
|
|||
|
MDS malfunction. |
power adapter. |
|
||
|
charger LED not |
|
|||
|
|
|
• If voltage = 11.6 to 13.8 VDC, contact TSS in Solna |
|
|
|
illuminated. |
|
|
|
|
|
|
|
|
or Danvers. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS directly |
MDS Power Switch |
1) Switch MDS Power to On. |
|
|
|
connected to Power |
malfunction. |
2) If MDS fails to Power-up, contact TSS in Solna or |
|
|
|
Adapter; MDS Charger |
MDS malfunction. |
|
||
|
Danvers. |
|
|||
|
LED illuminated. No |
|
|||
|
|
|
|
|
|
|
Power On LED. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
1 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
|
|
|
Medside Data Station |
|
|
|
|
|
|
|
Table 2-1 Power-On Problem (Continued) |
|
|
|
||
|
|
|
|
|
|
|
Conditions |
Possible Cause(s) |
|
|
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS not connected to |
Internal UPS Battery |
1) |
Connect MDS to A/C Power Adapter. |
|
|
A/C Power Adapter; |
discharged. |
2) |
Switch MDS Power switch to On and verify that |
|
|
Power switch On, |
Replace Internal battery. |
|||
|
|
battery charger LED illuminates. |
|||
|
Power LED not |
|
|||
|
MDS malfunction |
|
Note: If Power On LED fails to illuminate, contact |
||
|
illuminated. |
|
|||
|
|
|
|
TSS in Solna or Danvers. |
|
|
|
|
3) |
If battery charger LED fails to illuminate, leave |
|
|
|
|
|
power adapter connected to MDS for i 1hr. |
|
|
|
|
4) |
After 1 hr. disconnect MDS from Power Adapter |
|
|
|
|
|
and switch MDS Power On/Off switch Off, and then |
|
|
|
|
|
On. |
|
|
|
|
|
• |
If Power LED is green, reconnect MDS to Power |
|
|
|
|
|
Adapter and leave MDS connected an additional |
|
|
|
|
|
8 hours to charge internal battery. |
|
|
|
|
• |
If Power LED is not green, replace Internal |
|
|
|
|
|
batteries. |
|
|
|
5) |
If problem still persists, contact TSS in Solna or |
|
|
|
|
|
Danvers. |
|
|
|
|
|
|
|
|
|
|
|
|
|
Table 2-2 Power-off Alarm Malfunction
|
|
|
|
|
|
Symptom(s) |
Possible Cause(s) |
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
Piezo tone fails to |
Speaker. |
Contact TSS in Solna or Danvers. |
|
|
sound when MDS |
MDS malfunction. |
|
|
|
powered On, if MDS |
|
|
|
|
|
|
|
|
|
loses power, or when |
|
|
|
|
MDS is powered-Off. |
|
|
|
|
|
|
|
|
|
|
|
|
|
2.1.3Power-Up Sequence Fails to Complete Properly
Table 2-3 Power-up Process Malfunction
|
|
|
|
|
|
|
|
Symptom(s) |
Possible Cause(s) |
|
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Power inputs OK, but |
BIOS failure. |
1) |
If MDS displays error message proceed to |
|
|
|
MDS fails to complete |
Software program |
|
“Appendix B: BIOS Messages” on page 51, and |
|
|
|
boot up. |
|
“Appendix C: POST Error Codes” on page 55 to |
|
||
|
corrupted. |
|
|
|||
|
|
|
identify failure. |
|
|
|
|
|
Hard Drive failure. |
|
|
|
|
|
|
2) |
Check BIOS configuration according to Section 2.2. |
|
||
|
|
MDS malfunction. |
|
|||
|
|
3) |
If BIOS configuration OK, reinstall WindowsTM 2000 |
|
||
|
|
|
|
|||
|
|
|
|
according to Section 8 starting on page 14. |
|
|
|
|
|
4) |
If problem still persists, contact TSS in Solna or |
|
|
|
|
|
|
Danvers. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
2 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
a |
|
|
|
|
|
|
|
|
|
|
|
|
|
s |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
SW 2 |
|
|
|
|
|
|
|
|
|
|
SW 1 |
|
|
|
|
|
|
|||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Figure 2-1 MDS switch settings 2.1.4 No Video display
Table 2-4 Video malfunction
|
|
|
|
|
|
|
Symptom(s) |
Possible Cause(s) |
|
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS power LED On. |
Cable problem. |
1) |
Check both ends of video cable and ensure cable is |
|
|
No video on LCD/CRT |
No power to Display. |
|
connected. |
|
|
Display. |
|
|
|
|
|
Bad display. |
2) |
If problem persists, verify power source to display. |
|
|
|
|
|
|||
|
|
Video switch set |
3) |
If problem persists, switch out display with a known |
|
|
|
incorrectly. |
|
good display. |
|
|
|
MDS malfunction. |
4) |
If problem persists, verify MDS video switch position |
|
|
|
|
|
(see a in Figure 2-1) at rear of MDS is set to right. |
|
|
|
|
5) |
If problem still persists, contact TSS Danvers/Solna. |
|
|
|
|
|
|
|
|
|
|
|
|
|
2.1.5 MDS Fails to boot properly |
|
|
|
||
Table 2-5 Power On/MDS Malfunction |
|
|
|
||
|
|
|
|
|
|
|
Symptom(s) |
Possible Cause(s) |
|
Troubleshooting and Remedial Action |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
MDS power LED On. |
Watchdog timer switch |
1) |
Verify Watchdog timer switch position (see s in |
|
|
MDS Resets after |
set incorrectly. |
|
Figure 2-1) at rear of MDS is set to left. |
|
|
successful boot. |
BIOS problem. |
2) |
If problem persists, check BIOS configuration as |
|
|
|
|
|||
|
|
MDS malfunction |
|
described in Section 2.2. |
|
|
|
|
3) |
If problem still persists, contact TSS Danvers/Solna. |
|
|
|
|
|
|
|
|
|
|
|
|
|
The MDS is configured at the factory for default settings that provide proper operation. Use the following procedure to check MDS BIOS setup, if MDS does not boot to Windows 2000 logon screen. Changing BIOS settings is not necessary on a new MDS, unless system failure occurs.
1) Switch MDS Power On/Off switch to On.
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
3 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
a s
Figure 2-2 Main screen
2)Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
3)Press F9 key, then press <Enter> to load default BIOS configuration settings.
4)Press F10 key, then press <Enter> to save configuration settings. Note: After <Enter> key is pressed, MDS will reboot.
5)Verify that MDS boots to Windows 2000 logon screen. If MDS does not boot to Window 2000 logon screen, proceed to step 6.
6)Toggle MDS On/Off power switch Off, and then On to reboot MDS.
7)Press and hold F2 key to gain access to PhoenixBIOS Setup Utility.
8)Enter correct date/time (a in Figure 2-2) for clinical site, using arrow/ number keys.
9)Verify correct “Primary Master” settings as shown in (s in Figure 2-2).
Note: If Primary Master is incorrect, use up/down arrow keys to select Primary Master, and then press <Enter> key.
4 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
a
Figure 2-3 Hard drive settings
10)Use +/- keys to select type to Auto (a in Figure 2-3).
11)Press ESC key to get back to “Main” tab.
a
s
Figure 2-4 Advanced settings
12)Use left/right arrow keys to select “Advanced” tab (a in Figure 2-4).
13)Verify correct “Advanced” settings as shown in Figure 2-4. Note: If changes need to be made use up/down arrow keys.
14)Use up/down arrow keys to select I/O Configuration (s in Figure 2-4), and then press <Enter> key.
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
5 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
Serial port A:
Figure 2-5 I/O Device configuration
15)Verify correct “I/O Device Configuration” settings as shown in Figure 2-5. Note: Use left/right arrow and +/- keys to make changes.
16)Press ESC key to return to main menu.
Boot
a
Figure 2-6 Boot settings
17)Use up/down arrow keys to select “Boot” tab (a in Figure 2-6).
18)Verify correct “Boot” order as shown in Figure 2-6. Note: Use left/right arrow and +/- keys to make changes.
19)Press F10 key, and then press <Enter> to save configuration settings.
|
Note: After <Enter> key is pressed, MDS reboots. |
|
|
|
|
6 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
4.1 MDS Mounting Arm
Installation
20)Verify that Windows 2000 logon screen appears.
21)If MDS fails to boot to Windows 2000 logon screen, proceed to Section 9 to Phlash BIOS.
22)If Phlash BIOS setup procedure does not boot to Windows 2000 logon screen, proceed to Section 8 and re-install Windows 2000. If problem still exists, contact TSS in Danvers/Solna.
Install the Medside Data Station in a location that has good air circulation and is reasonably free from dust, extreme temperatures, and humidity. The MDS and the devices for the MDS are not intended for use in the same room with magnetic resonance equipment. Make sure a hospital grade power outlet and ethernet terminal (if connecting to a LAN) are located near MDS.
Caution: !
Do not place anything on top or bottom of Medside Data Station that can obstruct air flow to the ventilation holes on each side.
Do not place any liquid containers on MDS, to avoid possibility of a liquid spill damaging MDS.
Do either a or b as appropriate:
aIf installing an MDS on mounting arm, go on to Section 4.1.
bIf installing an MDS on table top, go on to Section 4.2.
Refer to front cover for an illustration of a complete wall mount setup of a Medside Data Station (Flat screen display shown).
1) Secure mount to wall (see instructions included with mounting arm).
a s
Figure 4-1 Mounting plate
2)Align mounting plate (a in Figure 4-1) on bottom of MDS to slots (a in Figure 4-2) on left side of mounting bracket.
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
7 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
4.2 MDS Table Top
Installation
a
s
Figure 4-2 Mounting bracket
3)Pull down and hold spring loaded locking pin (s in Figure 4-2), and slide in MDS to align hole on mounting plate (s in Figure 4-1) with locking pin.
4)Release locking pin to secure mounting plate to mounting bracket. Note: The locking pin snaps into place when properly installed.
5)Proceed to Section 5.
Monitor |
Medside |
|
Data |
|
Station |
Keyboard
Mouse
Figure 4-3 Medside Data Station (table top configuration)
1)Set MDS on flat clean surface, and within close proximity of monitor, keyboard, and mouse. See Figure 4-3.
2)Proceed to Section 5.
aIf installing optional CRT to wall mount, go to Section 5.1.
bIf installing optional Flat Screen to Medside Data Station, go to Section 5.2.
cIf installing locally supplied monitor, go to Section 5.3.
8 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
5.1CRT (Art. No. 57 35 894 E5310)
The same type of mounting bracket used to mount the Medside Data Station is used to mount the CRT monitor.
a
Figure 5-1 Monitor (bottom view)
1) Set Monitor upside down on clean surface.
a
s
Figure 5-2 Mounting plate
2)Align clearance holes on mounting plate (a in Figure 5-2) to threaded holes on bottom of monitor (a in Figure 5-1), orientated so that raised area around holes on plate are against bottom of monitor.
Note: This creates a gap so that the plate can easily slide into mounting bracket. Make sure the drill hole (s in Figure 5-2) is towards back of monitor. This hole is used for a locking pin from the mounting bracket.
3)Insert and tighten 4 Phillips-head screws (supplied).
4)Set monitor upright with mounting plate aligned to slots on left side of mounting bracket (see Figure 4-2 on page 8), and slide monitor into bracket so that hole on mounting plate aligns with locking pin.
5)Secure with locking pin. See step 3 in Section 4.1.
6)Proceed to Section 6.
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
9 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
|
|
|
|
Medside Data Station |
|
|
|
|
|
|
5.2 Flat Screen Display |
|
|
|
|
|
(Art. No. 59 55 567 |
d |
|
|
|
|
|
|||||
|
|
|
|
||
E531U) |
|
|
|
|
|
|
|
|
|
|
a
s
Figure 5-3 MDS with mounting bracket
1)Set MDS upright on flat surface.
2)Remove and discard three plastic plugs from mounting holes on top of MDS.
3)Align mounting bracket screw holes (a in Figure 5-3) to screw holes on top of MDS (s in Figure 5-3).
Note: Mounting bracket and screws ship with flat screen display.
4)Insert and tighten 3 Phillips-head screws (supplied, d in Figure 5-3).
aa
Figure 5-4 Flat Screen Display (rear view)
10 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
a a
s
Figure 5-5 Flat Screen Support Mount
5)Place flat screen display face down on clean surface. See Figure 5-4 on page 10.
6)Align 8 support mount clearance holes (a in Figure 5-5) to threaded holes on back of flat screen display (a in Figure 5-4).
7)Insert and tighten 8 Phillips-head screws (supplied).
8)Insert 1 cable tie (supplied) through back of each slot (s in Figure 5-5) to enable power supply to be secured to support mount.
a
d s
Figure 5-6 Flat Screen Display mounted to MDS
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
11 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
5.3 Locally Supplied
Monitor
9)Set power supply (a in Figure 5-6) between both slots on support mount, and tighten cable tie to secure supply to mount.
10)Set flat screen display upright with bottom of support mount aligned to slots on mounting bracket (installed in step 2 and 3 above), and slide mount into bracket (s in Figure 5-6 on page 11), so that mount is positioned in center.
11)Insert and tighten 2 Phillips-head screws (supplied, d in Figure 5-6) on rear of mounting bracket to secure flat screen display in bracket.
12)Proceed to Section 6.
Refer to Installation instructions that were provided with monitor.
1)Set monitor on secure flat surface in close proximity to MDS.
2)Proceed to Section 6.
Do either a or b as appropriate:
aIf installing keyboard and mouse to wall mount, go to Section 6.1.
bIf installing keyboard and mouse on table top, go to Section 6.2.
1)Set keyboard on shelf (refer to illustration on front cover), and slide in side clamps to secure keyboard to shelf.
2)Set mouse on shelf (refer to illustration on front cover).
3)Proceed to Section 7.
1)Set keyboard and mouse in close proximity to MDS. See Figure 4-3 on page 8.
2)Proceed to Section 7.
1)Insert and tighten 15 pin video cable from monitor into video out connector (s in Figure 7-1 on page 13) on rear of MDS.
2)Plug in Keyboard cable into keyboard connector (J in Figure 7-1) on rear of MDS.
3)Plug in mouse cable into mouse connector (H in Figure 7-1) on rear of MDS.
4)Plug power cord from monitor into hospital grade outlet. (If using flat screen monitor connect power cable from 12V DC power supply into back of monitor, and then plug in power cord from power adapter into hospital grade outlet).
5)Plug in power connector from MDS AC power adapter into Power In connector (j in Figure 7-1) on back of Medside Data Station, and then plug in power cord from AC adapter into hospital grade outlet.
6)Proceed to Section 11, Functional Check - Power Circuits and Startup.
12 |
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
Field Service Manual |
|
|
a s |
|
d |
fgh j k |
||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
l ; A S D F G H JK
Figure 7-1 MDS (rear view)
Table 3: MDS Part Description
Item |
|
Item |
|
|
Description |
Numbe |
Description |
||
Number |
||||
|
r |
|
||
|
|
|
||
|
|
|
|
|
1 |
VIDEO IN / OPTIONAL COM 3/4 |
10 |
COM 1 |
|
|
|
|
|
|
2 |
VIDEO OUT |
11 |
BOOT ROM SELECT (default, switch to right) |
|
|
|
|
|
|
3 |
*ETHERNET HUB |
12 |
VIDEO OVERRIDE (default, switch to right) |
|
|
|
|
|
|
4 |
AUDIO OUT |
13 |
PARALLEL PORT |
|
|
|
|
|
|
5 |
AUDIO IN |
14 |
PIEZO OVERRIDE (default, switch to right) |
|
|
|
|
|
|
6 |
MICROPHONE |
15 |
MAIN ETHERNET (auto-negotiating) |
|
|
|
|
|
|
7 |
POWER IN |
16 |
MOUSE |
|
|
|
|
|
|
8 |
POWER (future use) |
17 |
KEYBOARD |
|
|
|
|
|
|
9 |
COM 2 |
18 |
USB |
|
|
|
|
|
* Ethernet Hub will not auto-negotiate to a 10Mbs device. If a 10Mbs device is connected to the Ethernet Hub all devices must be manually set to 10Mbs operation. If All ports are occupied by 100Mbs devices, no manual settings are required. Consult your local IT department regarding manual settings of 10Mbs devices.
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
13 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
Windows 2000
Reinstallation
Optional Hardware
14
This section explains how to reinstall Windows 2000 and also references how to install Optional Hardware Drivers onto a Medside Data Station hard drive.
Note: Windows 2000 reinstallation should not be required on a new MDS. The factory installs Windows 2000 on the MDS prior to shipment.
In addition to these instructions, the following Hardware and Software is required:
•Siemens Service Laptop meeting minimum Hardware requirements as specified by Med QM document ARTD001.719.06.04.02 (V01) M4 or greater.
•RJ45 crossover cable.
•MDS Software Recovery CDROM (shipped with MDS).
Windows 2000 reinstallation requires making a network boot connection between an MDS RJ-45 network port, and a Windows 95/98 Laptop/PC RJ-45 network port, and then transferring the Windows 2000 image from the Laptop/PC CDROM drive to the MDS Hard Drive.
MDS Optional Hardware includes a TeacTM CDROM Drive, BackpackTM CDRW Drive, and MicrosoftTM Optical Mouse. Refer to “Loading Software” section in INFINITYTM Medside Data Station Reference Manual for Teac CDROM drive and Backpack CDRW drive operation.
An Optical Mouse driver CDROM is shipped with each Optical Mouse. This driver must be loaded through a network connection to the MDS. Contact local IT department for network connection configuration.
Setup the Service laptop as follows to allow network connectivity between a Laptop CDROM drive and an MDS Hard drive.
Figure 8-1 Windows Main Menu
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak
Medside Data Station |
|
|
Field Service Manual |
|
|
|
|
|
|
8.1.1 TCP/IP Setup |
1) |
Boot Service Laptop to WindowsTM 95/98 screen. |
|
|
|
2) |
Right-click on Network Neighborhood icon (see Figure 4-1 on page 7) |
|
|
|
|
|
and click on Properties. |
|
|
|
|
|
|
|
|
|
|
|
Figure 8-2 Network Window
3)At “Network” window (see Figure 8-2) click on Configuration Tab, scroll down to Laptop TCP/IP Ethernet Adapter, and select Properties.
Note: TCP/IP Ethernet Adapter name is unique according to specific adapter used on laptop. Refer to Service Laptop Ethernet Adapter vendor document for specific name.
Laptop Settings
Figure 8-3 TCP/IP Window
4)Copy “IP Address” to the following line:_____________________________
ASK-T962-03-7600 |
Siemens Medical Solutions, EM-PCS Danvers |
15 |
MDS2K_sm.fm/04-02/Sulak
Field Service Manual |
Medside Data Station |
|
|
Note: IP Address is needed to reconfigure the service laptop back to its original configuration, after completing Windows 2000 reinstallation.
16
192. 168. 0. 1
255. 255. 255. 0
Figure 8-4 TCP/IP Window
5)Type new IP address 192.168.0.1 as shown in Figure 8-4, then click on OK button.
Laptop Settings
Figure 8-5 Network window
6)At “Network” window (see Figure 8-5) click on Identification Tab.
7)Copy Computer Name to the following line:_______________________
The service laptop CDROM must be set up for file sharing. Complete the following section to configure service laptop CDROM for file sharing.
Siemens Medical Solutions, EM-PCS, Danvers |
ASK-T962-03-7600 |
MDS2K_sm.fm/04-02/Sulak