LITHOSTAR Multiline
CS
Update Instructions
Title:
Reason for Update: Performance
Urgency: Immediate Within 6 months
Update material required? Yes No
Materials free of charge? Yes No
Return of parts? Yes No
Estimated completion time: 1 hr Number of CSE’s: 1
Customer application training? Yes No
Systems/Products affected/System identifying IVK
Name Material No. Serial No.
Software Update Rev. 6.3
x
x
SP012/02/P
X
x
x
x
LITHOSTAR Multiline 31 27 011 J1042 06000 - 07164
Remark: n.a.
Components affected/to be modified
Name Material No. Serial No. Component status
affected
n.a. n.a. n.a. n.a.
Remark: n.a.
© Siemens AG 2002
Chg. Ref. No.: 112 570
Name: Erlwein
Dept.: CS PS 24
Print No.: RXL2-120.896.03.01.02 Doc. Gen. Date: 09.02
Replaces: n.a. Version 2.0; A ugust 1, 2002 Page 1 of 8
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model
reserved.
_or_ design,_are_
2 Update Instructions SP012/02/P
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products Affected 0
LITHOSTAR Multiline - Systems with serial numbers according to page 1.
Reason for the Update 0
• Uniform software level for all LITHOSTAR Multiline systems installed worldwide.
• Elimination of errors 429X.
• Increase of limit to 1.2 Million.
Prerequisites 0
UI SP013/01/P Software Update, Rev. 06.
NOTICE
If revision 06 is not yet installed, install it first according to
UI SP013/01/P, and then perform a combined function test.
Special Tools / Documents 0
• Service PC (max. 133 MHz) with WIN 3.11/ 95 operating system
• PC connection cable 99 00 440 RE999
• Service tool kit
LITHOSTAR Multiline RXL2-120.896.03 Page 2 of 8 Siemens AG
Rev. 01 09.02 CS PS 24 Medical Solutions
Update Instructions SP012/02/P 3
Ordering Information 0
Place an order as follows for each system to be updated:
11 57 853 J1042 update kit VD00C
The update kit is free of charge.
Contents of the update kit 0
Update kit 11 57 853 J1042 contains the following parts:
Pos. Quan. Part No. Name
1 1 11 57 614 J1042 HOST Software (VD00C)
2 1 Technical documents
Work Steps 0
NOTICE
There may appear problems while do wnloading the host software
in case a service pc with a frequency > 133 MHz is used.
If a more powerful PC is used the download will be terminated
and the system cannot be operated anymore.
Recording Multiline-specific system parameters 0
• Switch the system on.
• Connect the service PC to the system (N11 board D1.X10).
(Refer to Operating the service software RXL2-120.113.01).
• If service software VA00C has not yet been installed on the PC, install it.
(Refer to Operating the service software RXL2-120.113.01).
• Call up the service software (To start, put the HOST software diskette VD00A from the
logbook into the drive "a").
• Carry out a backup of all system parameters. (Data → Backup).
• Read out the following parameters and record them in the table.
(Adjustment
(Refer to setting parameters RXL2-120.032.01.).
→ Parameters → Get from Unit).
Siemens AG RXL2-120.896.03 Page 3 of 8 LITHOSTAR Multiline
Medical Solutions Rev. 01 09.02 CS PS 24