Urgency: Immediate Within6months
Update material required? Yes No
Materials free of charge? Yes No
Return of parts? Yes No
Estimated completion time: 1 hr Number of CSE’s: 1
Customer application training? Yes No
Print No.: RXL2-120.896.03.01.02Doc. Gen. Date:09.02
Replaces: n.a.Version 2.0; A ugust 1, 2002Page 1 of 8
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model
reserved.
_or_ design,_are_
2Update Instructions SP012/02/P
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products Affected0
LITHOSTAR Multiline - Systems with serial numbers according to page 1.
Reason for the Update0
• Uniform software level for all LITHOSTAR Multiline systems installed worldwide.
• Elimination of errors 429X.
• Increase of limit to 1.2 Million.
Prerequisites0
UI SP013/01/P Software Update, Rev. 06.
NOTICE
If revision 06 is not yet installed, install it first according to
UI SP013/01/P, and then perform a combined function test.
Special Tools / Documents0
• Service PC (max. 133 MHz) with WIN 3.11/ 95 operating system
• PC connection cable 99 00 440 RE999
• Service tool kit
LITHOSTAR MultilineRXL2-120.896.03Page 2 of 8Siemens AG
Rev. 0109.02CS PS 24Medical Solutions
Update Instructions SP012/02/P3
Ordering Information0
Place an order as follows for each system to be updated:
11 57 853 J1042 update kit VD00C
The update kit is free of charge.
Contents of the update kit0
Update kit 11 57 853 J1042 contains the following parts:
There may appear problems while do wnloading the host software
in case a service pc with a frequency > 133 MHz is used.
If a more powerful PC is used the download will be terminated
and the system cannot be operated anymore.
Recording Multiline-specific system parameters0
• Switch the system on.
• Connect the service PC to the system (N11 board D1.X10).
(Refer to Operating the service software RXL2-120.113.01).
• If service software VA00C has not yet been installed on the PC, install it.
(Refer to Operating the service software RXL2-120.113.01).
• Call up the service software (To start, put the HOST software diskette VD00A from the
logbook into the drive "a").
• Carry out a backup of all system parameters. (Data → Backup).
• Read out the following parameters and record them in the table.
(Adjustment
(Refer to setting parameters RXL2-120.032.01.).
→ Parameters → Get from Unit).
Siemens AGRXL2-120.896.03Page 3 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0109.02CS PS 24
4Update Instructions SP012/02/P
Parameter GroupParameterExisting
value
rev 06
BIOS Host parameterMultiline Ve rsion
Serial No*
Clinic dataClinic name**
Configuration: Motor
devices
Configuration: othersTable tilting yes/no
Configuration: User termi-
nal
Interface: CollimatorCassette format horizon-
MC1 am9 cone rotation
control at unit (slave 2)
control at unit (slave 7)
control at unit (slave 8)
Language
User Terminal Ultrasound
tal
Entered in new
parameter file
rev 6.2
Cassette format vertical
Pointformat
Slotformat horizontal
Slotformat vertical
Interface: IntensifierSystem type
Zoom 0
Zoom 1
Zoom 2
Zoom 3
Interface:
Image storage
Interface:
Shock wave system
Interface: UltrasoundLine frequency 50/60 Hz
Line frequency 50/60 Hz
System type
ECG Trigger available
Resp-Trigger available
Shock frequency
System available
Target
LITHOSTAR MultilineRXL2-120.896.03Page 4 of 8Siemens AG
Rev. 0109.02CS PS 24Medical Solutions
Update Instructions SP012/02/P5
Parameter GroupParameterExisting
value
rev 06
Interface:
XCS Controller Generator
Parameter: am7 TV-IrisDR Zoom 0
Parameter:
am8 Cone length
FLUORO Automatic 2
Generator type
DR Zoom 1
DR Zoom 2
DR Zoom 3
Fluoro Zoom 0
Fluoro Zoom 1
Fluoro Zoom 2
Fluoro Zoom 3
Incremental encoder
Position factor, mantisse
Entered in new
parameter file
rev 6.2
Safety position (1/10mm)
Parameter:
am9 Cone rotation
* if no serial number is entered, enter the serial number of the system into the new
parameter
** 32 blanks are entered as clinic name. These have to be deleted in order to enter a
name.
Incremental encoder
Position factor, mantisse
• Read out the shock wave count, and record the values.
(Refer to service software RXL2-120.032.02., monitoring chapter).
• Switch the system off.
Siemens AGRXL2-120.896.03Page 5 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0109.02CS PS 24
6Update Instructions SP012/02/P
Updating the host software on the LITHOSTAR Multiline0
Required parts from the upgrade kit:
Pos.Quan.Part No.Name
1111 57 614 J1042HOST software VD00C
Tab. 1
• Switch the system on.
• Place the disk with the new host software version VD00C into the drive of the service PC.
• Call up the service software.
• Download the new host software version VD00C.
In the Data menu: select Download Select "ML_Host".
(Refer to Operating the service software RXL2-120.113.01.)
• Despite the message, no automatic reset is performed. Perform the reset with S1 on
board D1 (in N11).
NOTICE
If no manual reset is carried out, no restore of the parameters is
possible. Disregard the status display on the operating monitor,
service PC or board D1 and D6.
• Use Menu, Data Restore to transfer the new parameter set LITHO_M.P00 from the
Host SW version VD00C to the unit.
(Refer to Operating the service software RXL2-120.113.01.).
• Record the parameters as well as the default values from the table; transfer them back to
the unit as unit and system-specific values (Refer to Operating the Service Software
RXL2-120.113.01. Menu Adjustment, Parameter.).
NOTICE
The changed values have to be transmitted separately for each
parameter group to the host.
• If error EE00 occurs in the MEMOSKOP 3
- set incorrect mains frequency and wait until reset was carried out.
- Subsequently set correct mains frequency again.
• Create backup copies for the following data (Menu, Data, Backup)
- Parameter, Customer Data, System Data, Technical Info, Service Data.
• Quit the service program and remove the disk from the drive.
• Switch the system off.
LITHOSTAR MultilineRXL2-120.896.03Page 6 of 8Siemens AG
Rev. 0109.02CS PS 24Medical Solutions
Update Instructions SP012/02/P7
What is new in software Rev 6.3?0
- Improved error detection.
- Constancy limit increased to 1.2 million.
NOTICE
Due to changes in the monitoring of the shock wave system, the
shock wave cabinet must be switched on when switching on the
system. The shock wave cabinet can be switched off, when the
system is ready for use.
Final Work Steps0
• Reattach all covers.
• Perform a function test for all the drives.
• Using the service PC, enter the Text "Software Rev. 6.2" into the Menu.
→ Technical Info.
Info
• Make backup copies of all parameters (menu, data, backup).
• Place the disks in the logbook. Remove the old software and destroy it.
• Update the system documentation.
Update the revision level, the operating instructions and the technical documentation.
Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update
Completion Form" and file it in the corresponding System Binder/User Handbook.
• Updates that have already been completed prior to publication of this SI must also be
reported.
• The update is reported as follows:
- The modification reply cards (Type 606) previously distributed with the publication of
updates no longer apply.
- The modification reply report has to be prepared by authorized personnel using an
application on the Intranet.
Siemens AGRXL2-120.896.03Page 7 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0109.02CS PS 24
8Update Instructions SP012/02/P
Completion Protocol0
The update with the number SP012/02/P has been completed.