Siemens Lithostar Multiline Update Instructions

LITHOSTAR Multiline
CS
Update Instructions
Title:
Reason for Update: Performance
Urgency: Immediate Within 6 months Update material required? Yes No Materials free of charge? Yes No Return of parts? Yes No Estimated completion time: 1 hr Number of CSE’s: 1 Customer application training? Yes No
Systems/Products affected/System identifying IVK
Name Material No. Serial No.
Software Update Rev. 6.3
x x
SP012/02/P
X
x
x
x
LITHOSTAR Multiline 31 27 011 J1042 06000 - 07164
Remark: n.a.
Components affected/to be modified
Name Material No. Serial No. Component status
affected
n.a. n.a. n.a. n.a.
Remark: n.a.
© Siemens AG 2002
Chg. Ref. No.: 112 570 Name: Erlwein Dept.: CS PS 24
Print No.: RXL2-120.896.03.01.02 Doc. Gen. Date: 09.02
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model reserved.
_or_ design,_are_
2 Update Instructions SP012/02/P
Document Revision Level
This document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.

Systems/Products Affected 0

LITHOSTAR Multiline - Systems with serial numbers according to page 1.

Reason for the Update 0

Uniform software level for all LITHOSTAR Multiline systems installed worldwide.
Elimination of errors 429X.
Increase of limit to 1.2 Million.

Prerequisites 0

UI SP013/01/P Software Update, Rev. 06.
NOTICE
If revision 06 is not yet installed, install it first according to UI SP013/01/P, and then perform a combined function test.

Special Tools / Documents 0

Service PC (max. 133 MHz) with WIN 3.11/ 95 operating system
PC connection cable 99 00 440 RE999
Service tool kit
LITHOSTAR Multiline RXL2-120.896.03 Page 2 of 8 Siemens AG Rev. 01 09.02 CS PS 24 Medical Solutions
Update Instructions SP012/02/P 3

Ordering Information 0

Place an order as follows for each system to be updated:
11 57 853 J1042 update kit VD00C
The update kit is free of charge.

Contents of the update kit 0

Update kit 11 57 853 J1042 contains the following parts:
Pos. Quan. Part No. Name
1 1 11 57 614 J1042 HOST Software (VD00C) 2 1 Technical documents

Work Steps 0

NOTICE
There may appear problems while do wnloading the host software in case a service pc with a frequency > 133 MHz is used. If a more powerful PC is used the download will be terminated and the system cannot be operated anymore.

Recording Multiline-specific system parameters 0

Switch the system on.
Connect the service PC to the system (N11 board D1.X10).
(Refer to Operating the service software RXL2-120.113.01).
If service software VA00C has not yet been installed on the PC, install it.
(Refer to Operating the service software RXL2-120.113.01).
Call up the service software (To start, put the HOST software diskette VD00A from the
logbook into the drive "a").
Carry out a backup of all system parameters. (Data Backup).
Read out the following parameters and record them in the table.
(Adjustment (Refer to setting parameters RXL2-120.032.01.).
Parameters Get from Unit).
Siemens AG RXL2-120.896.03 Page 3 of 8 LITHOSTAR Multiline Medical Solutions Rev. 01 09.02 CS PS 24
4 Update Instructions SP012/02/P
Parameter Group Parameter Existing
value rev 06
BIOS Host parameter Multiline Ve rsion
Serial No*
Clinic data Clinic name** Configuration: Motor
devices Configuration: others Table tilting yes/no Configuration: User termi-
nal
Interface: Collimator Cassette format horizon-
MC1 am9 cone rotation
control at unit (slave 2) control at unit (slave 7) control at unit (slave 8) Language User Terminal Ultrasound
tal
Entered in new parameter file rev 6.2
Cassette format vertical Pointformat Slotformat horizontal Slotformat vertical
Interface: Intensifier System type
Zoom 0 Zoom 1 Zoom 2 Zoom 3
Interface: Image storage
Interface: Shock wave system
Interface: Ultrasound Line frequency 50/60 Hz
Line frequency 50/60 Hz System type ECG Trigger available Resp-Trigger available Shock frequency
System available Target
LITHOSTAR Multiline RXL2-120.896.03 Page 4 of 8 Siemens AG Rev. 01 09.02 CS PS 24 Medical Solutions
Update Instructions SP012/02/P 5
Parameter Group Parameter Existing
value rev 06
Interface: XCS Controller Generator
Parameter: am7 TV-Iris DR Zoom 0
Parameter: am8 Cone length
FLUORO Automatic 2 Generator type
DR Zoom 1 DR Zoom 2 DR Zoom 3 Fluoro Zoom 0 Fluoro Zoom 1 Fluoro Zoom 2 Fluoro Zoom 3 Incremental encoder
Position factor, mantisse
Entered in new parameter file rev 6.2
Safety position (1/10mm)
Parameter: am9 Cone rotation
* if no serial number is entered, enter the serial number of the system into the new
parameter
** 32 blanks are entered as clinic name. These have to be deleted in order to enter a
name.
Incremental encoder
Position factor, mantisse
Read out the shock wave count, and record the values.
(Refer to service software RXL2-120.032.02., monitoring chapter).
Switch the system off.
Siemens AG RXL2-120.896.03 Page 5 of 8 LITHOSTAR Multiline Medical Solutions Rev. 01 09.02 CS PS 24
6 Update Instructions SP012/02/P

Updating the host software on the LITHOSTAR Multiline 0

Required parts from the upgrade kit:
Pos. Quan. Part No. Name
1 1 11 57 614 J1042 HOST software VD00C
Tab. 1
Switch the system on.
Place the disk with the new host software version VD00C into the drive of the service PC.
Call up the service software.
Download the new host software version VD00C.
In the Data menu: select Download Select "ML_Host". (Refer to Operating the service software RXL2-120.113.01.)
Despite the message, no automatic reset is performed. Perform the reset with S1 on
board D1 (in N11).
NOTICE
If no manual reset is carried out, no restore of the parameters is possible. Disregard the status display on the operating monitor, service PC or board D1 and D6.
Use Menu, Data Restore to transfer the new parameter set LITHO_M.P00 from the
Host SW version VD00C to the unit. (Refer to Operating the service software RXL2-120.113.01.).
Record the parameters as well as the default values from the table; transfer them back to
the unit as unit and system-specific values (Refer to Operating the Service Software RXL2-120.113.01. Menu Adjustment, Parameter.).
NOTICE
The changed values have to be transmitted separately for each parameter group to the host.
If error EE00 occurs in the MEMOSKOP 3
- set incorrect mains frequency and wait until reset was carried out.
- Subsequently set correct mains frequency again.
Create backup copies for the following data (Menu, Data, Backup)
- Parameter, Customer Data, System Data, Technical Info, Service Data.
Quit the service program and remove the disk from the drive.
Switch the system off.
LITHOSTAR Multiline RXL2-120.896.03 Page 6 of 8 Siemens AG Rev. 01 09.02 CS PS 24 Medical Solutions
Update Instructions SP012/02/P 7

What is new in software Rev 6.3? 0

- Improved error detection.
- Constancy limit increased to 1.2 million.
NOTICE
Due to changes in the monitoring of the shock wave system, the shock wave cabinet must be switched on when switching on the system. The shock wave cabinet can be switched off, when the system is ready for use.

Final Work Steps 0

Reattach all covers.
Perform a function test for all the drives.
Using the service PC, enter the Text "Software Rev. 6.2" into the Menu.
Technical Info.
Info
Make backup copies of all parameters (menu, data, backup).
Place the disks in the logbook. Remove the old software and destroy it.
Update the system documentation.
Update the revision level, the operating instructions and the technical documentation. Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update Completion Form" and file it in the corresponding System Binder/User Handbook.
Updates that have already been completed prior to publication of this SI must also be
reported.
The update is reported as follows:
- The modification reply cards (Type 606) previously distributed with the publication of updates no longer apply.
- The modification reply report has to be prepared by authorized personnel using an application on the Intranet.
Siemens AG RXL2-120.896.03 Page 7 of 8 LITHOSTAR Multiline Medical Solutions Rev. 01 09.02 CS PS 24
8 Update Instructions SP012/02/P

Completion Protocol 0

The update with the number SP012/02/P has been completed.
Material Number: 31 27 011 LITHOSTAR Multiline
Serial Number: ................................
Customer: ................................ Site: ................................
Customer No.: ................................
Name ( CSE ): ................................ Telephone: ................................
Country: ................................ Location: ................................
Date: ................................ Signature: ................................
Remark: ....................................................................................................
....................................................................................................
NOTE
LITHOSTAR Multiline RXL2-120.896.03 Page 8 of 8 Siemens AG Rev. 01 09.02 CS PS 24 Medical Solutions
After completing the update, make a copy of this page, fill it out and file it in the corresponding System Binder/User Handbook.