English
Print No.: RXL2-120.072.03.04.02 Doc. Gen. Date:06.02
Replaces: RXL2-120.072.03.03.02
0 - 2Revision level
ChapterPageRevision
allall04
Document revision level0
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcop y documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer0
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Yellow LED V20 (charging time exceeded Board D1 display: 5391 . . . . . . . . .8 -3
9 ______ Errors without error numbers_____________________________________ 9 -1
10 _____ Changes as compared to previous version ________________________ 10 -1
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
10 LM_628_error
11 Control_error
12 Position accuracy not reached in stated stabilizing time
13 Motor not initialized, drive command received
14 Motor moves without actuation
15 Motor does not move when actuated
16 Speed not reached or exceeded
17 Motor moves in wrong direction
18 Motor unsuccessfully actuated three times
19 Faulty initialization of the ASIC
1 - 1
ZL am2 (Lifting column left)1
Error number
↓ Meaning:
20 LM_628_error
21 Control_error
22 Position accuracy not reached in stated stabilizing time
23 Motor not initialized, drive command received
24 Motor moves without actuation
25 Motor does not move when actuated
26 Speed not reached or exceeded
27 Motor moves in wrong direction
28 Motor actuated unsuccessfully three times
29 Faulty initialization of the ASIC
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
1 - 2Motor controller D4a
W drive am31
Error number
↓ Meaning:
30 LM_628_error
31 Control_error
32 Position accuracy not reached in stated stabilizing time
33 Motor not initialized, drive command received
34 Motor moves without actuation
35 Motor does not move when actuated
36 Speed not reached or exceeded
37 Motor moves in wrong direction
38 Motor unsuccessfully actuated three times
39 Faulty initialization of the ASIC
O drive am4 (C-Arm)1
Error number:
↓ Meaning:
40 LM_628_error
41 Control_error
42 Position accuracy not reached in stated stabilizing time
43 Motor not initialized, drive command received
44 Motor moves without actuation
45 Motor does not move when actuated
46 Speed not reached or exceeded
47 Motor moves in wrong direction
48 Motor unsuccessfully actuated three times
49 Faulty initialization of the ASIC
Board errors, general1
Error number:
↓ Meaning:
60 No first contact with HOST
61 TIMEOUT, data exchange disrupted
62 CHECKSUM_error
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 2Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
10 LM_628_error
11 Control_error
12 Position accuracy not reached in stated stabilizing time
13 Motor not initialized, drive command received
14 Motor moves without actuation
15 Motor does not move when actuated
16 Speed not reached or exceeded
17 Motor moves in wrong direction
18 Motor actuated unsuccessfully three times
19 Faulty initialization of the ASIC
2 - 1
Y drive am62
Error number
↓Meaning:
20 LM_628_error
21 Control_error
22 Position accuracy not reached in stated stabilizing time
23 Motor not initialized, drive command received
24 Motor moves without actuation
25 Motor does not move when actuated
26 Speed not reached or exceeded
27 Motor moves in wrong direction
28 Motor actuated unsuccessfully three times
29 Faulty initialization of the ASIC
Siemens AGRXL2-120.072.03Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
2 - 2Motor controller D4b
TVI Television iris am72
Error number
↓Meaning:
30 LM_628_error
31 Control_error
32 Position accuracy not reached in stated stabilizing time
33 Motor not initialized, drive command received
34 Motor moves without actuation
35 Motor does not move when actuated
36 Speed not reached or exceeded
37 Motor moves in wrong direction
38 Motor unsuccessfully actuated three times
39 Faulty initialization of the ASIC
TL Cone drive, longitudinal, am82
Error number
↓Meaning:
40 LM_628_error
41 Control_error
42 Position accuracy not reached in stated stabilizing time
43 Motor not initialized, drive command received
44 Motor moves without actuation
45 Motor does not move when actuated
46 Speed not reached or exceeded
47 Motor moves in wrong direction
48 Motor unsuccessfully actuated three times
49 Faulty initialization of the ASIC
TR Cone drive, rotation, am92
Error number
↓Meaning:
50 LM_628_error
51 Control_error
52 Position accuracy not reached in stated stabilizing time
53 Motor not initialized, drive command received
54 Motor moves without actuation
55 Motor does not move when actuated
56 Speed not reached or exceeded
57 Motor moves in wrong direction
58 Motor unsuccessfully actuated three times
59 Faulty initialization of the ASIC
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Motor controller D4b2 - 3
Board errors, general2
Error number:
↓ Meaning:
60 No first contact with HOST
61 TIMEOUT, data exchange disrupted
62 CHECKSUM_error
Siemens AGRXL2-120.072.03Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
2 - 4Motor controller D4b
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Communication controller D63
3 - 1
NOTICE
Code "55" is displayed when the system is operating without
errors.
Code "60" also appears if the system is not equipped with the
ultrasound option.
Error displays of the collimator interface:3
Error No. Meaning
10 No first contact to the HOST (status message)
11 Timeout error from the HOST
12 Fatal error in the data reception block from the HOST
13 Illegal data in the reception block
14 Illegal value in the data reception flag from the HOST
15 No first contact to collimator N
16 Collimator timeout error
17 Data reception error from collimator
18 Continuous communication error to collimator
19 Reception queue full
1A DPR reception area error
1B CAN module error
1C Invalid values in reception block
Error displays of the operating master interface:3
Error No.Meaning
20 No first contact to the HOST (status message)
21 Timeout error from the HOST
22 Fatal error in the data reception block from the HOST
23 Wrong block type in the data reception block from the HOST
24 Illegal value in the data reception flag from the HOST
25 No first contact to slaves (status message)
26 Slave timeout error
27 Data reception error from slave
28 Serial transmitter not free
2A DPR reception area error
7x Timeout slave no. x
8x Data reception error slave no. x
Siemens AGRXL2-120.072.03Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
3 - 2Communication controller D6
Error displays of the image memory interface:3
Error No.Meaning
30 No first contact to the HOST (status message)
31 Timeout error from the HOST
32 Fatal error in the data reception block from the HOST
33 Illegal data in the data reception block from the HOST
34 Illegal value in the data reception flag from the HOST
35 No first contact to the image memory (status message)
36 Image memory timeout error
37 Data reception error from image memory
38 Serial transmitter not free
3A Error in the DPR reception area
Error displays of the shock wave system interface:3
Error No.Meaning
40 No first contact to the HOST (status message), System not registered
41 Timeout error from the HOST
42 Fatal error in the data reception block from the HOST
43 Illegal data in the data reception block from the HOST
44 Illegal value in the data reception flag from the HOST
45 No first contact to the shock wave system (status message)
46 Shock wave system timeout error
47 Data reception error from shock wave system
48 Serial transmitter not free
4A Error in the DPR reception area
4E Respiratory trigger signal timeout
Error displays of the brightness actual value interface:3
Error No. Meaning
50 No first contact to the HOST (status message)
51 Timeout error from the HOST
52 Fatal error in the data reception block from the HOST
53 Illegal data in the data reception block from the HOST
54 Illegal value in the data reception flag from the HOST
55 No first contact to the brightness actual value (status mes-
sage)
56 Timeout error from the brightness value actual value signal
device
57 Data reception error of brightness actual value
58 Serial transmitter not free
5A Error in the DPR reception area
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Communication controller D63 - 3
Error displays of the ultrasound interface:3
Error No. Meaning
60 No first contact to the HOST (status message)
61 Timeout error from the HOST
62 Fatal error in the data reception block from the HOST
63 Illegal data in the reception block from the HOST
64 Illegal value in the data reception flag from the HOST
65 No first contact to ultrasound system
66 Timeout error from ultrasound system
67 Error in block reception from ultrasound system
68 Error in block transmission to ultrasound system
69Transmission buffer not empty (component)
6A Error in the DPR reception area
Siemens AGRXL2-120.072.03Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
3 - 4Communication controller D6
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14
Overview and explanations of the unit drives4
O = am4
4 - 1
W = am3
TVI=Television iris
on the television camera
TL=Cone longitudinal
TR=Cone rotation
SWO=Shock wave head, orbital
I.I.=Image intensifier lift
ZR = am1
ZL = am2
Y = am6
X = am5
OC arm
YUnit, longitudinal
XUnit, lateral
ZR, ZLTable, up
WShockwave head
Siemens AGRXL2-120.072.03Page 1 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 2CPU board D1
Definition of the slave numbers4
Assigment: Slave no./Slave name
Slave no.Slave nameAbbreviation
0PC-BoardHBF
2Tableside control panel on PCNBF
3Memoskop image memoryBS
4Memoskop image memory BSHDBSHD
7Tableside control panel 1 (basic unit)NBF
8FootswitchNBFFT
9Tableside control panel (Ultrasound unit)NBFUS
10Shock Wave SystemSWS
NOTICE
Errors in the error log are identified with a letter in addition to the error number.
E = Error
These errors can be cleared by the software during the next check.
F = Fatal Error
System operation is stopped. The error may be cleared by resetting the system
(switching the system ON/OFF).
W = Warn
I = Information
ing
System operation may continue without restrictions.
These error numbers are for information only. They are not relevant to service.
Error messages in ERRO.LOG starting with FD... (Error text not
available) are internal error messages without any
field service.
meaning for
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 3
General errors4
Error No. Error display on the service PCMeaning of the error
0001 Switch-on routineSwitch-on routine: point in time at
which the unit was switched on.
Identifies associated errors and
required switch off/on proce-
dures.
0002 Service reset
0010 Contactors K1/K2 closed, no 27V present
0011 Contactor K3 closed, no 30V present no safety contactor closed?
0012 Contactors K1/K2 open, no 0V present
0013 Contactor K3 open, no 0V present
0020 Error: reference switch check / ZRam 1Check the reference
0021 Error: reference switch check / ZLam 2
0022 Error: reference switch check /Wam 3
0023 Error: reference switch check /O am 4
0025 Error: reference switch check / X am 5
0026 Error: reference switch check / Y am 6
0028 Error: reference switch check / TLam 8
0029 Error: reference switch check / TRam 9
0050 Emergency stop is operated Emergency stop active
0070 ADC time is exceeded / converter 1Error on board D5
0071 ADC time is exceeded / converter 2Error on board D5
0080 ADC value out of tolerance / channel 0ZR Drive
0081 ADC value out of tolerance / channel 1ZL Drive
0082 ADC value out of tolerance / channel 2W Drive
0083 ADC value out of tolerance / channel 3O Drive
0084 ADC value out of tolerance / channel 4X Drive
0085 ADC value out of tolerance / channel 5Y Drive
0086 ADC value out of tolerance / channel 6
0087ADC value out of tolerance / channel 7
0088 ADC value out of tolerance / channel 8TVI Drive
0089ADC value out of tolerance / channel 9
008AADC value out of tolerance / channel 10TL Drive
008BADC value out of tolerance / channel 11TR Drive
008CADC value out of tolerance / channel 12
008DADC value out of tolerance / channel 13
008EADC value out of tolerance / channel 14
008FADC value out of tolerance / channel 15
switch
for errors 0020 to 002B
Siemens AGRXL2-120.072.03Page 3 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 4CPU board D1
Error No. Error display on the service PCMeaning of the error
0090ADC reference voltage out of tolerance5.12V
009124V ADC supply out of tolerance24V 3V
009227V ADC supply out of tolerance27V - 9V / + 13V
009330V ADC supply out of tolerance3DV - 10V / + 9V
SS relay dropped out
0094No signal of stopping potential
0F00System is not initialized
The system is not initialized or the ROM version is wrong. The version number of the HOST software and the parameters in the EEPROM do not match
(the first two places).
If, however, the software is present, then parameters are missing or the version numbers of the HOST software or the parameters in the EEPROM do
not match.
The error will be displayed after the do wnload routine . Repeat the do wnload
routine.
± 60mV
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 5
Boot software4
Error No. Error display on the
service PC
1000Flash - EEPROM is emptyDownload HOST software again
1003Flash EEPROM checksum
error
1004No access to clockSet the clock again and perform Reset. If the
1005Clock is not initializedSet the clock again and perform Reset. If the
1009Incorrect checksum
with block receipt
100AIncorrect command
from Service PC
100BIncorrect block length
from Service PC
100C Incorrect segment address
from Service PC
100DParity errorError in the communication between the ser-
100EOverrun errorReceive buffer of the serial interface at the
100FFraming errorError in the communication between the ser-
1010Break errorService PC was switched off before the serial
1011Timeout error
1012Error: Switch S2.8
(D1) = On
Meaning of the error
Incorrect checksum in the Flash EEPROMs.
Download HOST software again.
error still exists, replace PC board D1.
error still exists, replace PC board D1.
Incorrect checksum when a block is received
from the service PC/interface driver. The
problem is in the service PC or HOST. Interference during transmission.
Error in the data transmission between the
service PC and the HOST. The service software version and the boot software version
do not match. Compare with the compatibility
list in the logbook.
Error in the data transmission between the
service PC and the HOST. The service software version and the boot software version
do not agree. Compare with the compatibility
list in the logbook.
Incorrect segment address from the service
PC for writing on EPROM. Error in IDX file;
install the backup copy.
vice PC and the HOST. Hardware problem or
electromagnetic compatiblity (EMC) interference.
If necessary, perform the transfer again.
service PC has been exceeded.
vice PC and the HOST.
Hardware problem or electromagnetic compatibility (EMC) interference.
If necessary, perform the transfer again.
connection closed. This message is also displayed if hardware errors are present on
board D1.
After three unsuccessful attempts to transfer,
transmission between the HOST and the service PC is terminated.
Siemens AGRXL2-120.072.03Page 5 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
9 = Ultrasound operating panel (BFUS)
3001HBF timeout
3002Receive error of D6 from HBF(D6 = Communication Controller)
3003NAK error of HBF from D6(NAK = no acknowledge)
3004HBF Sent invalid block length
3005HBF didn't respond ACK/NAK
3009Serial Interface of D6 is faulty
3030HBF: Data error from D1
3032HBF FIFO overrun during transmission
3033HBF FIFO overrun during reception
3040PC doesn't react to transmission
3041Reception timeout, PC does not transmit
3042Timeout of First PC contact after 60s
3043Invalid data from PC
3044Checksum error from PC
3045New init request from PC
3050Data overrun
3051HBF: Timeout to D6
3081HBF: Watchdog hit
3083No contact to PC
3201NBF Timeout
3202Reception error of D6 from NBF
3203NAK error of NBF from D6(NAK = no acknowledge)
3204NBF Sent invalid block length
3205NBF didn´t respond ACK/NAK
3209Serial interface of D6 is faulty
3231NBF: Data error from D1
3250Data overrun
3251NBF: Timeout to D6
3252NBF: Keyboard buffer overflow
3260NBF: Actuated or defective keyNBF:Actuated or defective key
3281NBF: Watchdog hit
3701NBF Timeout
3702Reception error of D6 from NBF
➝ HBF data update
LITHOSTAR MultilineRXL2-120.072.03Page 6 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 7
Error No. Error text displayed on the service PC
3703NAK error of NBF from D6(NAK = no acknowledge)
3704NBF Sent invalid block length
3705NBF didn't respond ACK/NAK
3709Serial Interface of D6 is faulty
3731NBF: Data error from D1
3750Data overrun
3751NBF:Timeout to board D6
3752NBF: Keyboard buffer overflow
3760NBF: Actuated or defective keyNBF: Actuated or defective key
3781NBF: Watchdog hit
3801NBFFT timeout
3802Reception error of D6 from NBFFT
3803NAK error of NBFFT from D6(NAK = no acknowledge)
3804NBFFT Sent invalid block length
3805NBFFT didn't respond ACK/NAK
3809Serial Interface of D6 is faulty
3831NBFFT: Data error from D1
3850Data overrun
3851NBFFT: Timeout to board D6
3852NBFFT: Keyboard buffer overflow
3860NBFFT: actuated or defective keyNBFFT: Actuated or defective
3881NBFFT: Watchdog hit
3901NBFUS:Timeout
3902Reception error of D6 from NBFUS
3903NAK error of NBFUS from D6(NAK = no acknowledge)
3904NBFUS: Sent invalid block length
3905NBFUS: didn‘t respond ACK/NAK
3909Serial Interface of D6 is faulty
3931NBFUS: Data error from D1
3950Data overrun
3951
3952NBFUS: Keyboard buffer overflow
3960NBFUS: actuated or defective keyNBFUS: Actuated or defective
3981NBFUS: Watchdog hit
NBFUS: Timeout to board D6
➝ NBF data update
➝ NBFFT data update
key
➝ NBFUS data update
key
Siemens AGRXL2-120.072.03Page 7 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 8CPU board D1
Drive4
Error No. Error text displayed on the service PC
4000Protocol error of MC1 (D4a)
4001Protocol error of MC2 (D4b)
4002Protocol error of MC3 (D4c)
4120Target position from D1 too large / Z
4121Target position from D1 too large / Z
4122Target position from D1 too large / Wam 3
4123Target position from D1 too large / Oam 4
4125Target position from D1 too large / Xam 5
4126Target position from D1 too large / Yam 6
4127Target position from D1 too large / TVIam 7
4128Target position from D1 too large / TLam 8
4129Target position from D1 too large / TRam 9
412ATarget position from D1 too large / SWOam 10
412BTarget position from D1 too large / I.I.am 11
4130Invalid command from D1 / ZRam 1
4131Invalid command from D1 / ZLam 2
4132Invalid command from D1 / Wam 3
4133Invalid command from D1 / Oam 4
4135Invalid command from D1 / Xam 5
4136Invalid command from D1 / Yam 6
4137Invalid command from D1 / TVIam 7
4138Invalid command from D1 / TLam 8
4139Invalid command from D1 / TRam 9
413AInvalid command from D1 / SWOam 10
413BInvalid command from D1 / I.I.am 11
R
L
am 1
am 2
4140Control error / ZRam 1
4141Control error / ZLam 2
4142Control error / Wam 3
4143Control error / Oam 4
4145Control error / Xam 5
LITHOSTAR MultilineRXL2-120.072.03Page 8 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
4159Invalid parameter from D1 / TRam 9
415AInvalid parameter from D1 / SWOam 10
415BInvalid parameter from D1 / I.I.am 11
4160Invalid position from D1 / ZRam 1
4161Invalid position from D1 / ZLam 2
4162Invalid position from D1 / Wam 3
4163Invalid position from D1 / Oam 4
4165Invalid position from D1 / Xam 5
4166Invalid position from D1 / Yam 6
4167Invalid position from D1 / TVIam 7
4168Invalid position from D1 / TLam 8
4169Invalid position from D1 / TRam 9
416AInvalid position from D1 / SWOam 10
416BInvalid position from D1 / I.I.am 11
4170Invalid data for motor initialization / ZRam 1
4171 Invalid data for motor initialization / ZLam 2
4172Invalid data for motor initialization / Wam 3
4173 Invalid data for motor initialization / Oam 4
4175 Invalid data for motor initialization / Xam 5
4176 Invalid data for motor initialization / Yam 6
4177 Invalid data for motor initialization / TVIam 7
4178 Invalid data for motor initialization / TLam 8
4179Invalid data for motor initialization / TRam 9
417AInvalid data for motor initialization / SWOam 10
417BInvalid data for motor initialization / I.I.am 11
Siemens AGRXL2-120.072.03Page 9 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 10CPU board D1
Error No. Error text displayed on the service PC
4180Motor activated without drive signal / ZRam 1
4181Motor activated without drive signal / ZLam 2
4182Motor activated without drive signal / Wam 3
4183Motor activated without drive signal / Oam 4
4185Motor activated without drive signal / Xam 5
4186Motor activated without drive signal / Yam 6
4187Motor activated without drive signal / TVIam 7
4188Motor activated without drive signal / TLam 8
4189Motor activated without drive signal / T
418AMotor activated without drive signal / SWOam 10
418BMotor activated without drive signal / I.I.am 11
4190Motor not activated by drive signal / ZRam 1
4191Motor not activated by drive signal / ZLam 2
4192Motor not activated by drive signal / Wam 3
4193Motor not activated by drive signal / Oam 4
4195Motor not activated by drive signal / Xam 5
4196Motor not activated by drive signal / Yam 6
4197Motor not activated by drive signal / TVIam 7
4198Motor not activated by drive signal / TLam 8
4199Motor not activated by drive signal / TRam 9
419AMotor not activated by drive signal / SWOam 10
419BMotor not activated by drive signal / I.I.am 11
R
am 9
Check the fuses,
check the
output stages
(D7...),
check the LED on
D4
41A0Drive speed not reached / ZRam 1
41A1Drive speed not reached / ZLam 2
41A2Drive speed not reached / Wam 3
41A3Drive speed not reached / Oam 4
41A5Drive speed not reached / Xam 5
41A6Drive speed not reached / Yam 6
41A7Drive speed not reached / TVIam 7
41A8Drive speed not reached / TLam 8
41A9Drive speed not reached / TRam 9
41AADrive speed not reached / SWOam 10
41ABDrive speed not reached / I.I.am 11
41B0Motor rotation in reverse direction / ZRam 1
41B1Motor rotation in reverse direction / ZLam 2
41B2Motor rotation in reverse direction / Wam 3
41B3Motor in reverse direction / Oam 4
41B5Motor in reverse direction / Xam 5
LITHOSTAR MultilineRXL2-120.072.03Page 10 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 11
Error No. Error text displayed on the service PC
41B6Motor in reverse direction / Yam 6
41B7Motor in reverse direction / TVIam 7
41B8Motor in reverse direction / TLam 8
41B9Motor in reverse direction/ TRam 9
41BAMotor in reverse direction / SWOam 10
41BBMotor in reverse direction / I.I.am 11
41C0Repeated fatal error of motor / ZRam 1
41C1Repeated fatal error of motor / ZLam 2
41C2Repeated fatal error of motor / Wam 3
41C3Repeated fatal error of motor / Oam 4
41C5Repeated fatal error of motor / Xam 5
41C6Repeated fatal error of motor / Yam 6
41C7Repeated fatal error of motor / TVIam 7
41C8Repeated fatal error of motor / TLam 8
41C9Repeated fatal error of motor / TRam 9
41CARepeated fatal error of motor / SWOam 10
41CBRepeated fatal error of motor / I.I.am 11
41D0Drive timeout for motor indicated / ZRam 1
41D1Drive timeout for motor indicated / ZLam 2
41D2Drive timeout for motor indicated / Wam 3
41D3Drive timeout for motor indicated / Oam 4
41D5Drive timeout for motor indicated / Xam 5
41D6Drive timeout for motor indicated / Yam 6
41D7Drive timeout for motor indicated / TVIam 7
41D8Drive timeout for motor indicated / TLam 8
41D9Drive timeout for motor indicated / TRam 9
41DADrive timeout for motor indicated / SWOam 10
41DBDrive timeout for motor indicated / I.I.am 11
Siemens AGRXL2-120.072.03Page 13 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 14CPU board D1
Shock Wave System4
Error
No.
51XXRemovable error restricts interface function marginally. The D1 can reset
5101 SWS timeout(Shock Wave System)
5102 Reception error of D6
5103 NAK error of SWS from
5104 SWS Sent invalid block
5105 SWS didn' t respond ACK/
5109 Serial Interface of D6 is
5111 Respiratory trigger time-
5150 Receive FIFO overrun
5151 Polling timeout
5152 Watchdog reset
5153 Nominal energy is not
5154 Shock counter doesn’t
Error text displayed
on the service PC
the error and restart operation.
from SWS
D6.
length
NAK
faulty
out
actual
work D19
Meaning of the error
(NAK = no acknowledge)
Check the setting D19.R47
52XX
5201 Valve Y1 overheatingW11/X1,z22 = high (no associated LED)
5202 Valve Y1 open loadW11/X1,d4 = high (no associated LED)
5203 Valve Y1 short circuitW11/X1,d12 = high (no associated LED)
5204 Cooling system faultyW11/X1,d16 = high and yellow LED W11/V11 lights
5205 Water container emptyW11/X1,z10 = high and yellow LED W11/V9 lights up
5206 Supply reserveW11/X1,b12 = high and yellow LED W11/V10 lights up
5207 Temperature too highW11/X1,z14 = high (no associated LED)
5241 Motor M20 overloadW12/X1,b32 = high and red LED W12/V21 lights up
5261 Energy generation errorW15/X1,z10 = high and yellow LED W15/V2 lights up
5262 Shockwave head errorW15/X1,z14 = high and yellow LED W15/V3 lights up
5281 D19 self ignitionD19/X1,b14 = high and yellow LED D19/V33 lights up
5282 D19 kA
5283 D19 U
Error. The interface controller malfunctioned.
One or several active jobs will be aborted.
The D1 can no longer reset the error.
up. Display at cooling water overtemperature is okay
despite flow
actual
residual
< kA
min
is too highD19/X1,b12 = high and red LED D19/V9 lights up
D19/X1,b16 = high and yellow LED D19/V32 lights up
LITHOSTAR MultilineRXL2-120.072.03Page 14 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 15
Error
No.
53XX
Error text displayed on the
service PC
Fatal Error:
Meaning of the error
The occurrence of fatal errors blocks the controller interface. Resetting
these errors by software is no longer possible.
5301Pressure too high switch S4 Pressure > 800mbar
± 30 mbar switch S4
W11/X1,b14 = high
and red LED W11/V14 lights up
5302Pressure too low sensor SH Pressure < 210mbar + 30 mbar
W11/X1,z20 = high and red LED W11/V5 lights
up; the cooling circuit pressure sensor signals
that the system has not reached the control
range.
5303Temperature too high sen-
sor SH
Temperature > 38 C sensor W11/X1,b18 = high
and red LED W11/V15 lights up; the cooling circuit pressure temperature sensor signals that the
system has exceeded the control range.
5304Flowmeter faultyW11/X1,b14 = high and red LED W11/V13 lights
up. During the switch-on reset, flow okay has
been signalled.
5305Pressure too high sensorPressure > 500mbar + 30 mbar sensor
W11/X1,d14 = high and red LED W11/V13 lights
up; the cooling circuit pressure sensor signals
that the control range has been exceeded.
5306Flow (l/H) too lowW11/X1,z12 = high
and red LED W11/V12 lights up
5307Cooling system faultyW11/X1,d18 = low and green LED W11/V3 is dark
(= shot triggering inhibited)
5341End of lift min/maxW12/X1,z18 = high
and red LED W12/V17 lights up. (either the min.
or the max. limit of lift is exceeded).
5361kV-limit errorW15/X1,z26 = high
and red LED W15/V5 lights up.
5362Energy generation errorW15/X1,z8 = high
and red LED W15/V1 lights up
5363Shockwave generator error W15/X1,z16 = high
and red LED W15/V4 lights up
5364Invalid kV value acknowl-
edge message
W15/X1,z16 = high
and red LED W15/V4 lights up.
Acknowledged kV value is wrong: Message software W16 (no associated LED) from W15. The
kV value received deviates by >14% from the
Siemens AGRXL2-120.072.03Page 15 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 16CPU board D1
General errors4
Error No. Error text displayed on the service PC
6100Checksum error / MC1 (D4a)
6101Checksum error / MC2 (D4b)
6102Checksum error / MC3 (D4c)
6103Checksum error / Collimator
6104Checksum error / Operator-Master
6105Checksum error / Image Storage
6106Checksum error / Shock Wave System
6107Checksum error / Actual Brightness Value
6108Checksum error / Ultrasound
6110Wrong block length / MC1 (D4a)
6111Wrong block length / MC2 (D4b)
6112Wrong block length / MC3 (D4c)
6113Wrong block length / Collimator
6114Wrong block length / Operator-Master
6115Wrong block length / Image Storage
6116Wrong block length / Shock Wave System
6117Wrong block length / Actual Brightness Value
6118Wrong block length / Ultrasound
6120Subsystem timeout / MC1 (D4a)
6121Subsystem timeout / MC2 (D4b)
6122Subsystem timeout / MC3 (D4c)
6123Subsystem timeout / Collimator
6124Subsystem timeout / Operator-Master
6125Subsystem timeout / Image Storage
6126Subsystem timeout / Shock Wave System
6127Subsystem timeout / Actual Brightness Value
6128Subsystem timeout / Ultrasound
6130Start up tim out MC1 (D4a)
6131Start up timeout MC2 (D4b)
6132Start up timeout MC3 (D4c)
6133Start up timeout / Collimator
6134Start up timeout / Operator-Master
6135Start up timeout / Image Store
6136Start up timeout / Shock Wave System
6137Start up timeout / Actual Brightness Value
6138Start up timeout / Ultrasound
LITHOSTAR MultilineRXL2-120.072.03Page 16 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 17
Ultrasound4
Error
No.
7001Timeout from US deviceTimeout from ultrasound system
7002Error with block reception from
7003Error during transmission to US
7004Module emitter buffer not freeModule emitter buffer not empty
7010Incorrect ID in reception blockIncorrect ID in reception block
7011Incorrect length of reception
7100No Litho US applicator connected No ultrasound probe for Litho connected
7101Error with autom. selection of
7102Error with autom. realtime selec-
7103Error with autom. cursor display Error in automatic cursor display
7104Reported target mark position
Error text displayed
on the service PC
US unit
unit
block
probe 2
tion
incorrect
Meaning of the error
Error in block reception from ultrasound
system
Error during transmission to ultrasound
system
Incorrect length of reception block
Error in automatic selection of probe 2
Error in automatic realtime selection
Reported target position incorrect
Collimator4
Error
No.
8001Collimator timeout
8002Collimator data reception error
8003CAN-communication disturbance Possible defect on board D1 in the colli-
8004Collimator signalling invalid data
8005CAN chip is faulty
8010RAM test D1 failure
8011EEPROM test D1 failure
8012CAN Error D6
8015Collimator height: calculation
Siemens AGRXL2-120.072.03Page 17 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 18CPU board D1
Generator4
Error No. Error text displayed on the service PC
9000Subsystem not configured
9010Received block too long
9011Timeout Board D2
9012Init data transmitted after runup
9013Faulty writing attempt
9020XCU not readyCheck fuse F8 in the
Videomed DIM
9021Contact to XCU interrupted
9022Turn-off message during operation
9023XCU not transmitting data
9100Invalid telegram mode
9101Invalid AP_ID in telegram
9102Invalid object ID from Board D2
9103Invalid object length
9110Invalid object ID for control unit
9111Invalid value for automatic fluoro data
9112Invalid value for fluoro alarm data
9113Invalid value for mode fluoro data
9114Invalid value for I.I. zoom
9115Invalid value for TV image reversion
9116Invalid value for x-ray data
9117Invalid value for system data
9118Invalid value for focus data
9119Invalid operating mode
911AInvalid value for tube loading
911BInvalid value for acknowledgement byte
9120Invalid obj.ID for I.I. functional unit
9121Invalid value for I.I. func. unit: ZOOM
LITHOSTAR MultilineRXL2-120.072.03Page 18 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 19
Image memory errors received from MEMOSKOP4
NOTICE
Error text displayed on the service PC: "Error text not available"
Error No.Meaning of the error
E101Hardware error on board D1 in the Memoskop.
E102External RAM test error of the D1 processor in the Memoskop.
E103Checksum test for the configuration byte is wrong (board D1).
E106CPU 2 bus on board D2 defective. The internal tests in the Memoskop
E108Memoskop: Prozessor D1 Stack overflow
E109Image text incorrectly loaded the checksum test.
E402No communication with CPU 2 after a Reset.
Error codes E101 to EDFF are messages from MEMOSKOP.
Internal test routine in the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
During the initialization routine, the RAMs are erased- if this is not possible, this error is displayed.
External test not possible. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Check the configuration (refer to the service instructions below).
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Check the software status in the Download menu.
Ensure that the Memoskop status is "up to date".
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Check whether board D2 is correctly inserted in the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
show errors. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
No test
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Was a text allocated to the image?
Check the 50/60 Hz configuration.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Correct the configuration.
Replace board D6.
Check whether boards D1 and D2 are correctly inserted in the Memoskop.
Replace board D2 in the Memoskop. Replace the backplane board D5.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Siemens AGRXL2-120.072.03Page 19 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 20CPU board D1
Error No. Meaning of the error
E403Timeout during communication with CPU 2
Check the software status in the Download menu.
BILD.HEX Memoskop HARD.HEX hard disk
Ensure that the Memoskop status is "up to date".
If the status is "different" or "not available", perform the software download
routine.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images)
E404Error during polling of CPU 2 version.
Check the software status in the Download menu.
Ensure that the Memoskop status is "up to date".
If the statis is "different" or "not available", perform the software download
routine.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E480
bis
E4E2
E601Text controller test incorrect.
E602Text controller RAM test incorrect.
E610Keyboard interface test Baud rate too low.
E611Keyboard interface test Baud rate too high.
E612Keyboard interface test Baud clock missing.
E620Keyboard interface test, feedback error
MEMOSKOP Hard disk error
Board D2 defective
Cable defective
Hard disk defective
Internal test in the Memoskop.
No external test possible.
Replace text controller D6. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Internal test in the Memoskop.
No external test possible.
Replace text controller D6.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard of the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard of the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard or the keyboard cable of the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard or the keyboard cable of the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
LITHOSTAR MultilineRXL2-120.072.03Page 20 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 21
Error No. Meaning of the errors
E621Keyboard Interface-Test Feedback Timeout.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E710Memoskop video sync test: H sync signal period too long.
Check connections transformer T1 - plug D1/X13 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E711Memoskop video sync test: H sync signal period too short.
Check connections transformer T1 -plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E712Memoskop BAS (CVS) syc test: H sync signal missing.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E720Memoskop video sync test: 50 Hz VR sync period too long.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E721Memoskop video sync test: 50 Hz VR sync period too short.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E722Memoskop video sync test: 60 Hz VR sync period too long.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E723Memoskop video sync test: 60 Hz VR sync period too short.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Siemens AGRXL2-120.072.03Page 21 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 22CPU board D1
Error No. Meaning of the errors
E724Memoskop BAS (CVS) sync test: VR sync missing.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E730Memoskop BAS (CVS) sync test: H sync level error
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E731Memoskop BAS (CVS) sync test: Error PLL lock.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E740Memoskop BAS (CVS) sync test: VR interrupt missing.
Check connections transformer T1 - plug D1/X3 in Memoskop.
Check the fuses F2; F3; F4; F5. Replace the defective fuses.
Correct connections transformer T1 - plug D1/X13.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E750Memoskop BAS (CVS) sync test: Incorrect callup parameters.
Multiline: check pixel clock and composite video signal of TV system.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the camera head of the Videomed DI.
E800Memoskop voltage test: Reference voltage ADC too high.
No test. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E801Memoskop voltage test: Reference voltage ADC too low.
No test. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E810Memoskop voltage test: Voltage VPPmin. too high.
(Programming voltage for Flash-EPROMS).
No test.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E811Memoskop voltage test: Voltage VPPmin. too low.
No test. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E812Memoskop voltage test: Voltage VPPmax. too high.
No test. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
LITHOSTAR MultilineRXL2-120.072.03Page 22 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 23
Error No. Meaning of the errors
E813Memoskop voltage test: Voltage VPPmax. too low.
No test. Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E814Memoskop voltage test. Voltage VPP not present.
Check the fuses in the Memoskop.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E820Memoskop voltage test: Voltage 5V analog too high (>6V).
The LED +5V in the Memoskop must light with normal brightness.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E821Memoskop voltage test: Voltage 5V analog too low (<4V).
Check the LED display in the Memoskop, the LED +5V (green) must light up.
Check fuse F3 in Memoskop and replace, if necessary.
Check connections X13 in Memoskop and correct, if necessary.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E830Memoskop voltage test: Voltage -5.2V too high.
The LED 5V2 in the Memoskop must light with normal brightness.
Check connections X13 in Memoskop and correct, if necessary.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E831Memoskop voltage test: Voltage -5.2V too low.
Check fuse F4 in Memoskop and replace, if necessary.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E840Memoskop voltage test: Voltage VEE analog -5V too high.
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
E841Memoskop voltage test: Voltage VEE analog -5V too low.
Check the LED "VEE" (green) on the Memoskop - the LED m ust light up.
Check fuse F5 in Memoskop and replace, if necessary.
(Monitor synchronization no longer present).
Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images).
EA1HDPR err: write access invalid subsys.no.
EA2HDPR err: write access too high subsys.no.
EA3HDPR err: read access invalid subsys.no.
EA4HDPR err: read access too high subsys.no.
EC21Memoskop not ready for download.
Check the LED displays on the Memoskop.
Check the connection: Service PC HOST/D1.
Try download routine again. Adjust the service PC, if necessary
Image memory (errors from Multiline)
Siemens AGRXL2-120.072.03Page 23 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 24CPU board D1
Error No.Meaning of the error
ED14Memoskop Board D2 data checksum error during download
Error detected in checksum during data transfer.
during download of Memoskop Board D2
help: Malfunction of data line; Check plug connections; Board D1 defective
ED21Memoskop D2 not ready for download
Request sent for download, however, Memoskop not yet ready.
during download of Memoskop Board D2
ED22Memoskop Board D2 DBI too long during download
Excessively long data word detected during download.
during download of Memoskop Board D2
help: Fault in Memoskop; Board D2
ED23Memoskop D2 address outside of area during download
Selected address lies outside of permissible area
during download of Memoskop Board D2
ED24Memoskop Board D2 incorrect data checksum during download
Data transferred with incorrect checksum.
during download of Memoskop Board D2
Help: Malfunction of data line. Check plug connections.
ED31Memoskop Board D2 FEPROM erase error during download
Memoskop cannot be cleared
during download of Memoskop Board D2
Help: Board D2, D1 defective. +15 V supply voltage, X13
ED32Memoskop Board D2 FEPROM programming error during download
After each write operation to the FEPROM, the input value is read back. If
the value read does not agree with the value written, an error is displayed.
during download of Memoskop Board D2
Help: FEPROM defective; +15 V supply voltage,
X13 Board D2, D1 defective
ED33Memoskop Board D2 FEPROM1 programming error during download
No data can be written to FEPROM1
during download of Memoskop Board D2
Help: FEPROM defective; +15 V supply voltage
X13 Board D2, D1 defective
ED34Memoskop Board D2 FEPROM2 programming error during download
No data can be written to FEPROM2.
during download of Memoskop Board D2
Help: FEPROM defective +15 V supply voltage,
X13 Board D2, D1 defective
ED35Memoskop Board D2 FEPROM erase error during download
The FEPROM is completely erased prior to programming.
If this proves impossible, an error message appears
during download of Memoskop Board D2
Help: FEPROM defective + 15 V supply voltage,
X13 Board D2, D1 defective
ED36Memoskop Board D2 FEPROMS all bytes already 0
LITHOSTAR MultilineRXL2-120.072.03Page 24 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 25
Error No.Meaning of the errors
ED41Memoskop Board D2 unknown test number during download
An unknown test number was called up
during download of Memoskop Board D2
ED42Memoskop Board D2 unknown DBI in HX_REQUEST during download
Unknown data bit detected during polling of configuration
during download of Memoskop Board D2
ED43Memoskop Board D2 wrong segment during download
Software error
during download of Memoskop Board D2
ED44Memoskop Board D2 wrong test DBI during download
Software error
during download of Memoskop Board D2
ED45Memoskop Board D2 wrong segment during download
Software error
during download of Memoskop Board D2
ED51Memoskop Board D2 unknown telegram from HOST during download
ED52Memoskop Board D2 receive error during download
Memoskop-HOST connection
ED61Memoskop Board D2 +12 V Vpp out of tolerance during download
+ 12 V programming voltage not o.k.
during download of Memoskop Board D1
Help: Connector X13 on board D1; Board D1 defective;
Power unit voltage supply
ED62Memoskop D2 + 12,4V Vpp out of tolerance dur. download
+ 12 V programming voltage not o.k.
during download of Memoskop D1
Help: Connector X13 on board D1; Board D1 defective;
Power unit voltage supply
ED63Memoskop Board internal test failed during download
Error during internal test
during download of Memoskop Board D2
ED71Memoskop Board D2 FEPROM checksum error during download
Checksum of FEPROMs incorrect
End of download of Memoskop Board D2
Help: FEPROM defective
Siemens AGRXL2-120.072.03Page 25 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 26CPU board D1
Image memory errors received from Multiline4
NOTICE
Error No. Error text displayed on the service PC
EE00Start of reset not signalled
EE01End of reset not signalled
EE02Unexpected reset timeout
EE03Timeout during job acknowledge
EE04Job terminated by error
EE05Job acknowledged due to invalid DBI-NR
EE10Invalid message of operator master
EE17Invalid DBI received
EE18Data block of wrong length received
EE19Loss of polling signal from BS detectedBS = MEMOSKOP
EF00Invalid operating block (INFTYP)
EF01Invalid master command byte
EF02Invalid M_to_H-Message (Master to D1)
EF03Unknown error signalled
EF04Slave number not configured
EF07Master transmission error
EF08BS subsystem configured incorrectly
EF09BS subsystem configured incorrectly
EF13BS does not respond
EF14BSHD does not respond
EF23NAK reception error for BSNAK = no acknowledge
EF24NAK reception error for BSHDBSHD =
EF33NAK reception error from BS
EF34NAK reception error from BSHD
EF43Invalid block length from BS
EF44Invalid block length from BSHD
EF53Image Storage does not respond ACK/NAK
EF54Hard disk does not respond ACK/NAK
EF60Acquisition active without fluoro
EF70Acquisition timeout during fluoro
EF80Timeout acknowledgement "SH_Ready"(SH = from Slave to D1)
EF90Timeout acknowledgement "SH_Stop"
Error codes EE00 to EF90 are messages from Multiline.
MEMOSKOP Harddisk
LITHOSTAR MultilineRXL2-120.072.03Page 26 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 27
General errors4
Error No. Error text displayed on the service PCMeaning of the error
F000Incompatible AST version
F001Z_No. of execution control too large
F010Software timer array overflow
F100Invalid Watchdog configuration
F200Watchdog has responded
F500C-MOS Data (D1) are deleted
F600EEPROM error (checksum)
F700RAM error (checksum)
Boot Software4
Error No.Error text displayed on the ser-
vice PC
F800Error in watchdog testPerform Reset;
F801Error in RAM test
F802Battery dischargeReplace the battery
F803Error in Test1 COM1 D1.X10Perform Reset;
F804Error in Test1 COM2 D1.X11
F807Timer 0 error (D1)
F808Timer 1 error (D1)
Siemens AGRXL2-120.072.03Page 27 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 28CPU board D1
Internal software errors4
Error No. Error text displayed on the service PCMeaning of the error
FE00Invalid motor number
FE01Invalid ADC channel number
FE02Queue overrun for LED control
FE03Software internal error
FE04Temporary buffer overflow messages
FE05Alternating buffer overflow messages
FE06Acknowledgeable buff. overflow messages
FE07Buffer overflow of keyboard storage
FE08Msg No <> Err No with msg delection
FE40Invalid BS command (PUSH_BS_CMD)
FE41Invalid BS status (S_BS_Stat)
FE42Invalid parameter (G_BS_Slave)
FE43Invalid parameter (G_BS_CmdAct)
FE44Invalid parameter (S_BS_CmdAct)
FE45Invalid parameter (G_BS_Stat)
FE46Invalid parameter (HS-Modus)
FE47Transfer data buffer overflow
FE48Data block buffer overflow
FE70Parameter error "PUSH_US_CMD"
FE80Target position outside
admission position / X
FE81Target position outside
admission position / Y
FE82Target position outside
admission position / Z
FE90Parameter error "PUSH_XCS_CMD"
FE91Parameter error "Fluoro mode"
FE92Parameter error "Direct_mode"
FE93Inadmissible value in "FE_BV_ZOOM"Unacceptable value
FE94Inadmissible value in "XCS_STAT()"Unacceptable value
Target position outside permissible position (X drive)
Target position outside permissible position (Y drive)
Target position outside permissible position (Z drive)
in "FE_BV_Zoom"
in "XCS_STAT()"
LITHOSTAR MultilineRXL2-120.072.03Page 28 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Boot and error messages of the operating PC5
Operating system errors5
"Start-Up-Message: No HOST-Communication"
Meaning:This message is displayed during booting of the operating PC as long as no
Dual Port RAM (DPR) communication is running;
"Interface-Error: DUEP-POLLING-TIME-OUT"
Meaning:The plug-in card reports via the DPR interface that polling is no longer rec-
ognized at the DUEP interface
(Communication fault);
"Interface-Error: HOST-DATA-OVERFLOW"
Meaning:If the HOST transmits data faster than can be forw arded to the operating PC ,
then this message is displayed;
"Interface-Error: NEW-INIT-REQUEST"
Meaning:The plug-in card requests resynchronization of the DPR communication
because of internal errors (e.g. watchdog reset);
5 - 1
"Interface-Error: TIME-OUT-INIT"
Meaning:A timeout was detected during the booting synchronization of the DPR inter-
face;
"Interface-Error: TIME-OUT-SEND"
Meaning:A timeout has occurred while transmitting data via DPR interface to the plug-
in card (DPR communication error);
"Interface-Error: INVALID-DPR-FLAG"
Meaning:The DPR communication protocol was not observed
(DPR communication fault);
"Interface-Error: CHECKSUM-ERROR"
Meaning:A checksum error has been detected while receiving data via DPR interface
(DPR communication error);
"Interface-Error: RECEIVE-BUFFER FULL"
Meaning:The operating PC has received data too fast and an internal data overflow
has occurred;
"Interface-Error: SEND-BUFFER FULL"
Meaning:The operating PC has generated more data blocks than can be sent and
thus triggered an internal data overflow;
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
5 - 2Boot and error messages of the operating PC
Board D51 of the operating PC5
7-segment display
The following messages are shown on the 7-segment display:
0 Normal operation (DUEP polling and DPR interface OK)
1 DUEP communication error, but DPR interface OK
2 DPR communication error, but DUEP interface OK
3 DPR and DUEP communication error
4 Error in the analog part of the board or on the processor
5 Error on the internal timer of the processor
6 External RAM error
7 EPROM or program RAM error
8 Status after reset
9nc
Decimal point
80C535 works=>DP flashes (ADC/D<DUEP card sign of life)
LED displays:
The following messages are output via LEDs:
- Watchdog has responded
- Card is being booted by PC
- AT bus working (PC sign of life)
AP - IDxx displays5
The following AP-ID codes are displayed on the control monitor
MeaningMeaning
AP-ID: 16VIDEOMED DIAP-ID: 241Service PC
AP-ID:80POLYDOROSAP-ID: 242MULTILINE
AP-ID: 240XCUAP-ID : 243I.I. system
These error codes indicate the error group. In order to receive the specific error information, the service PC must be connected to the XCU (generator D200) and then the error
memory can be read out.
and theVIDEOMED DI/DIMService instructions RX52-022.061...
contain information on the causes of errors.
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 2Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
XCS controller D26
Error No.:Meaning:Remedial actions:
00Normal operation ok
One of the HW tests
not successful
10-IFSwitch the component off
39-IFACS software errorsand then on again or re51-53place PC board or (PROMs)
20-2F
37-38Operating error by HOST-SWCorrect on HOST system
55-5F
36Temporary ACS
50high load problemRecovers by itself
Replace PC board
6 - 1
34
35Communication with the service PC
63disrupted or terminated
30Power failure with subsequent
reconfiguration
31No network partner found
or FOC not connected
32Node-ID in network is present
several times
33Transmission repeatnormal or unstable
54ACS firmware in network
not compatible
60No network partner foundnormal or FOC PCBs must
normal or [power line] unstable
PCB. FOC must be checked
normal or FOC. PCBs must
be checked
Check node-IDs in the
network
PCB, FOC must be checked
Check the ACS SW version
be checked
61, 62No connection to the
HOST system
V1Decimal point changes frequently: ACS net is not stable; along with the
message number 30
V2Decimal point flashes: correct program run
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
Check the HOST system
6 - 2XCS controller D2
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 2Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D197
Overview of operating status displays7
D19
7 - 1
Shot release, V
act
Test ON, I-monitoring
(flashes when S1 in position 2)
UC
Rest
kV
kV
actual
actual
≠ kV
> kV
General fault
Short circuit
Open in discharge circuit
Spark gap error, after 2 unsuccessful attempts without triggering
= kV
> 60V
nominal
nominal
nom.
V6
V36
V9
V7
V10
V26
V31
V32
V33
R7A
S1
S2
R60
R47
MP2:
kV
nominal
(3.0 V ± 0.03 V with 15.1 kV)
Position 1: Normal operating mode
Position 2: Error simulation
Reset for error memory
MP4: X3 x 1.33
MP3: X3 x 2.5
Fig. 1
Check the LEDs on circuit board D19 (Fig.1) to determine which monitoring control has
responded.
Red LED V7:kV nominal value (kV
kV actual value (kV
nominal
nominal
Red LED V10:The kV actual value (kV
Yellow LED V31:The current through the shock wave head l
Yellow LED V32:The current through the shock wave head l
) from board D15 and the
) at the capacitor differ.
) at the capacitor is greater than 23 kV.
actual
is too high.
dis
is too low.
dis
Yellow LED V33:Spark gap ignition occurred twice without triggering.
NOTICE
For settings on board D19, refer to instructions "Repair of shock
wave system" RXL2-120.091...
Siemens AGRXL2-120.072.03Page 1 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
7 - 2Error displays on D19
V7 (kV
actual
≠ kV
nominal
Produce oscilloscope
display of the voltages
on board D19
actual
nominal
(MP1)
(MP2)
kV
act
not o.k.
kV
and kV
Check kV indication
on charging unit to determine
whether the charging unit
has charged
up to the required value
yes
no
Measuring cable U,
shock wave generator X4
or
board D19.X2 defective
) / Board D1 display: 53827
kV
o.k.
act
kV
act
no
Does kV
correspond
to the value set ?
Compare with
table 1
no
Replace circuit
board D19
=kV
?
nom.
nominal
yes
Check and adjust
as necessary
± 15V ± 0.1 V
supply voltage
Replace board
D19
yes
Charging unit
not charging up
Charging unit
defective
Unplug high-voltage
cable K26
Charging unit
charging up
High-voltage cable K26 or
shock wave generator
defective
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 8Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D197 - 3
Red LED V10 (kV
Produce oscilloscope display of the voltages on board D19 kV
kV
< 5
nom.
V
> kV
actual
nominal
(measuring point MP1) and kV
kV
< 23 kV
act
Check the kV code at cable K108 on the
charging unit side
(table 2)
Code not o.k.
Check the kV code on board D15
connector X2
) / Board D1 display: 53817
actual
nominal
(measuring point MP2)
kV
< 23 kV
act
1 V 10 kV
Code o.k.
Replace board D19
Replace charging unit
(table 2)
Code not o.k.
Code o.k.
Replace cable K108
Replace board D15
Siemens AGRXL2-120.072.03Page 3 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
7 - 4Error displays on D19
Yellow LED V32 (discharge current too low) / Board D1 display:52827
Check lead X3 (I
meas
)
(if necessary, change over two leads at the
capacitor and at D19)
Error removed
Error not removed
Check adjustment of the circuit board D19
Error not removed
Replace board D19
Error not removed
Check trigger cable X2
Interchange pick-
up leads X3
Error not removed
Replace shock head
Error not removed
Replace spark gap module
Yellow LED V33 (spontaneous ignition of the spark gap)
Board D1 Display: 5281 7
Reset on keyboard with ESC
Error occurs repeatedly
Replace spark gap module
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 8Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D197 - 5
Yellow LED V31 (discharge current too high) Board D1 display: 53837
Replace shock
wave head
Error not
removed
Replace high-voltage cable
between shock wave generator and shock wave head
Error not
removed
Check adjustment of the circuit board D19
Error not
removed
Replace
board D19
Siemens AGRXL2-120.072.03Page 5 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
7 - 6Error displays on D19
kV reference values
Energy levels
on
keyboard
0.19.51.9 V ± 0.03 V
0.29.9
0.310.4
0.410.8
0.511.2
0.611.7
0.712.1
0.812.6
0.912.9
1.013.4
1.513.7
2.013.9
2.514.2
3.014.5
kV Indication
charging unit
± 0.2 kV
D19 MP 2
(1 kV = 0.2 V)
3.514.7
4.015.13.0 V ± 0.03 V
4.515.3
5.015.5
5.515.9
6.016.2
6.516.5
7.016.8
7.517.1
8.017.4
8.517.8
9.018.13.64 V ± 0.03 V
9.518,3
10.018.4
10.518.7
11.018.9
11.519.3
12.019.6
Tab. 1
LITHOSTAR MultilineRXL2-120.072.03Page 6 of 8Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D197 - 7
MSBLSB0 V
W15d32d30d28d26d24d22d20d18z24
K6345671011122
K108. N123456710111218
Tab. 2
Energy
levels
K108.
charging unit
0.101010001
0.201010010
0.301011000
0.401011100
0.501011111
0.601100011
0.701100111
0.801101011
0.901101110
1.001110010
1.501110100
2.001110110
2.501111001
3.001111011
3.501111101
4.010000000
4.510000010
5.010000100
5.510000111
6.010001010
6.510001100
7.010001111
7.510010001
8.010010101
8.510010111
9.010011010
9.510011011
10.010011100
10.510011111
11.010100001
11.510100100
12.010100111
123456789
K108 pin assignment
18
on the charging unit
9
Siemens AGRXL2-120.072.03Page 7 of 8LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
15
7 - 8Error displays on D19
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 8 of 8Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D918
Overview of operating status displays8
D91
Nom. charging value
R38
8 - 1
Fig. 1
V20
V4
V7
t
charge actual
U/
actual
Start pulse, charging unit
≠ U/2
> t
charge nominal
nominal
Reset
Energy monitoring
S1
Check the LEDs on circuit board D91 (Fig.1) to determine which monitoring controls have
responded.
Red LED V4:Half the capacitor voltage U/2 and full capacitor voltage U differ.
Yellow LED V20:The acceptable charging time has been exceeded.
(The LED lights only briefly in the event of an error)
When the shock wave generator has produced more than one million shots, the
high-volt age stability may have decreased. This makes it likely that board D91 has
responded.
NOTICE
For settings on board D91, refer to the instructions "Maintenance
of shockwave system" RXL2-120.091...
Siemens AGRXL2-120.072.03Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
8 - 2Error displays on D91
Red LED V4 (kV
actual
Produce oscilloscope display
of the voltages on D91.b20
kV
actual
and
(cathode diode V1).
1
kV
act/2
/2 kV
kV
act
kV
> kV
actual/2
not equal to kV
1
/2 kV
≠ 0
act/2
= kV
act/2
kV
act/2
act
< kV
act/2
nominal
act
= 0
) / Board D1 display: 53918
Check signal path
to board D19
Check lead X5 U/
Check supply voltage (±15V ±0,1 V)
on board D1;
replace board D91
2
Replace
leads X4
and X5
kV
kV
act
act
< kV
> kV
act/2
Check leads
act/2
Replace shock wave generator
kV
kV
act
act
> kV
< kV
Replace
leads X4
and X5
act/2
act/2
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Error displays on D918 - 3
Yellow LED V20 (charging time exceeded Board D1 display: 53918
Check automatic circuit breaker
on charging unit
Check switch position on charging unit
Charging unitSwitchPosition
HPE
LS21 - 1,5
VoltageNom value
ModeRemote
Outputon
Display the voltages kV
(D91.b20) on
board D91 on oscilloscope
Display the voltages kV
kV
actual
not o.k.
act
actual
kV
act
1 V = 10 kV
(D91.b20 on board D91 on oscilloscope)
noyes
Measuring cable
to theboard D91
Disconnect high-volt-
age cable K26
defective
o.k.
Display charge end pulse
on board D91.z24
on oscilloscope
Charge end not o.k.
Check signal path
from D91.z24
up to the charging
unit
Charge end o.k.
Check setting of
charge time
Setting
correct
Replace circuit
Charging unit not charging up
board D91
Replace shock
Charging unit
charging up
Charging unit
defective
Siemens AGRXL2-120.072.03Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
High voltage cable K26 or
Shock wave generator defective
wave generator
8 - 4Error displays on D91
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Errors without error numbers9
Letters appear scattered across the monitor.
Cause: defective text module on board D1 in the Memoskop.
9 - 1
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
9 - 2Errors without error numbers
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 2Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Changes as compared to previous version10
Chapter 0Revision level increased, table of contents adjusted
Chapter 4 - 27 Error F812 - new inserted
Error F813 - new inserted
Error F814 - new inserted
10 - 1
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
10 - 2Changes as compared to previous version
This page intentionally left blank.