Siemens Lithostar Multiline Troubleshooting Instructions

LITHOSTAR Multiline
System
Troubleshooting Instructions
SP
Errors by error codes
© Siemens AG 1996
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: RXL2-120.072.03.04.02 Doc. Gen. Date: 06.02
Replaces: RXL2-120.072.03.03.02
0 - 2 Revision level
Chapter Page Revision
all all 04
Document revision level 0
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcop y documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer 0
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Contents 0 - 3
Page
Document revision level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1 _______Motor controller D4a ____________________________________________ 1 -1
General errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -1
ZR am1 (Lifting column right) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -1
ZL am2 (Lifting column left) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -1
W drive am3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -2
O drive am4 (C-Arm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -2
Board errors, general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -2
2 _______Motor controller D4b____________________________________________ 2 -1
General errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -1
X drive am5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -1
Y drive am6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -1
TVI Television iris am7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -2
TL Cone drive, longitudinal, am8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -2
TR Cone drive, rotation, am9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -2
Board errors, general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -3
3 _______Communication controller D6 ____________________________________ 3 -1
Error displays of the collimator interface: . . . . . . . . . . . . . . . . . . . . . . . . 3 -1
Error displays of the operating master interface: . . . . . . . . . . . . . . . . . . . . 3 -1
Error displays of the image memory interface: . . . . . . . . . . . . . . . . . . . . . 3 -2
Error displays of the shock wave system interface: . . . . . . . . . . . . . . . . . . . 3 -2
Error displays of the brightness actual value interface: . . . . . . . . . . . . . . . . . 3 -2
Error displays of the ultrasound interface:. . . . . . . . . . . . . . . . . . . . . . . . 3 -3
4 _______CPU board D1 _________________________________________________ 4 -1
Overview and explanations of the unit drives . . . . . . . . . . . . . . . . . . . . . . 4 -1
Definition of the slave numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -2
General errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -3
Boot software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -5
Operating system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -6
Drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -8
Shock Wave System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -14
General errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -16
Ultrasound. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -17
Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -17
Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -18
Image memory errors received from MEMOSKOP . . . . . . . . . . . . . . . . . . 4 -19
Image memory errors received from Multiline. . . . . . . . . . . . . . . . . . . . . 4 -26
General errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -27
Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
0 - 4 Contents
Page
Boot Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -27
Internal software errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -28
5 ______ Boot and error messages of the operating PC _______________________ 5 -1
Operating system errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 -1
Board D51 of the operating PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 -2
AP - IDxx displays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 -2
6 ______ XCS controller D2 ______________________________________________ 6 -1 7 ______ Error displays on D19 ___________________________________________ 7 -1
Overview of operating status displays . . . . . . . . . . . . . . . . . . . . . . . . . .7 -1
V7 (kV Red LED V10 (kV
Yellow LED V32 (discharge current too low) / Board D1 display:5282 . . . . . . . .7 -4
Yellow LED V33 (spontaneous ignition of the spark gap)
Board D1 Display: 5281 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 -4
Yellow LED V31 (discharge current too high) Board D1 display: 5383. . . . . . . .7 -5
actual
¼ kV
nominal
actual
) / Board D1 display: 5382 . . . . . . . . . . . . . . . . .7 -2
> kV
nominal
) / Board D1 display: 5381 . . . . . . . . . . . .7 -3
8 ______ Error displays on D91 ___________________________________________ 8 -1
Overview of operating status displays . . . . . . . . . . . . . . . . . . . . . . . . . .8 -1
Red LED V4 (kV
not equal to kV
actual
nominal
) / Board D1 display: 5391 . . . . . . .8 -2
Yellow LED V20 (charging time exceeded Board D1 display: 5391 . . . . . . . . .8 -3
9 ______ Errors without error numbers_____________________________________ 9 -1 10 _____ Changes as compared to previous version ________________________ 10 -1
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Motor controller D4a 1

General errors 1

Error number
Meaning:
01 PROM_error 02 INT_RAM_error 03 EXT_RAM_error 04 TIMER_0_error 05 TIMER_1_error 06 KERN_TEST_error 07 SFR_error 09 WATCHDOG_error

ZR am1 (Lifting column right) 1

Error number
Meaning:
10 LM_628_error 11 Control_error 12 Position accuracy not reached in stated stabilizing time 13 Motor not initialized, drive command received 14 Motor moves without actuation 15 Motor does not move when actuated 16 Speed not reached or exceeded 17 Motor moves in wrong direction 18 Motor unsuccessfully actuated three times 19 Faulty initialization of the ASIC
1 - 1

ZL am2 (Lifting column left) 1

Error number
Meaning:
20 LM_628_error 21 Control_error 22 Position accuracy not reached in stated stabilizing time 23 Motor not initialized, drive command received 24 Motor moves without actuation 25 Motor does not move when actuated 26 Speed not reached or exceeded 27 Motor moves in wrong direction 28 Motor actuated unsuccessfully three times 29 Faulty initialization of the ASIC
Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
1 - 2 Motor controller D4a

W drive am3 1

Error number
Meaning:
30 LM_628_error 31 Control_error 32 Position accuracy not reached in stated stabilizing time 33 Motor not initialized, drive command received 34 Motor moves without actuation 35 Motor does not move when actuated 36 Speed not reached or exceeded 37 Motor moves in wrong direction 38 Motor unsuccessfully actuated three times 39 Faulty initialization of the ASIC

O drive am4 (C-Arm) 1

Error number:
Meaning:
40 LM_628_error 41 Control_error 42 Position accuracy not reached in stated stabilizing time 43 Motor not initialized, drive command received 44 Motor moves without actuation 45 Motor does not move when actuated 46 Speed not reached or exceeded 47 Motor moves in wrong direction 48 Motor unsuccessfully actuated three times 49 Faulty initialization of the ASIC

Board errors, general 1

Error number:
Meaning:
60 No first contact with HOST 61 TIMEOUT, data exchange disrupted 62 CHECKSUM_error
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Motor controller D4b 2

General errors 2

Error number
Meaning:
01 PROM_error 02 INT_RAM_error 03 EXT_RAM_error 04 TIMER_0_error 05 TIMER_1_error 06 KERN_TEST_error 07 SFR_error 09 WATCHDOG_error

X drive am5 2

Error number
Meaning:
10 LM_628_error 11 Control_error 12 Position accuracy not reached in stated stabilizing time 13 Motor not initialized, drive command received 14 Motor moves without actuation 15 Motor does not move when actuated 16 Speed not reached or exceeded 17 Motor moves in wrong direction 18 Motor actuated unsuccessfully three times 19 Faulty initialization of the ASIC
2 - 1

Y drive am6 2

Error number
Meaning:
20 LM_628_error 21 Control_error 22 Position accuracy not reached in stated stabilizing time 23 Motor not initialized, drive command received 24 Motor moves without actuation 25 Motor does not move when actuated 26 Speed not reached or exceeded 27 Motor moves in wrong direction 28 Motor actuated unsuccessfully three times 29 Faulty initialization of the ASIC
Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
2 - 2 Motor controller D4b

TVI Television iris am7 2

Error number
Meaning:
30 LM_628_error 31 Control_error 32 Position accuracy not reached in stated stabilizing time 33 Motor not initialized, drive command received 34 Motor moves without actuation 35 Motor does not move when actuated 36 Speed not reached or exceeded 37 Motor moves in wrong direction 38 Motor unsuccessfully actuated three times 39 Faulty initialization of the ASIC

TL Cone drive, longitudinal, am8 2

Error number
Meaning:
40 LM_628_error 41 Control_error 42 Position accuracy not reached in stated stabilizing time 43 Motor not initialized, drive command received 44 Motor moves without actuation 45 Motor does not move when actuated 46 Speed not reached or exceeded 47 Motor moves in wrong direction 48 Motor unsuccessfully actuated three times 49 Faulty initialization of the ASIC

TR Cone drive, rotation, am9 2

Error number
Meaning:
50 LM_628_error 51 Control_error 52 Position accuracy not reached in stated stabilizing time 53 Motor not initialized, drive command received 54 Motor moves without actuation 55 Motor does not move when actuated 56 Speed not reached or exceeded 57 Motor moves in wrong direction 58 Motor unsuccessfully actuated three times 59 Faulty initialization of the ASIC
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Motor controller D4b 2 - 3

Board errors, general 2

Error number:
Meaning:
60 No first contact with HOST 61 TIMEOUT, data exchange disrupted 62 CHECKSUM_error
Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
2 - 4 Motor controller D4b
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Communication controller D6 3

3 - 1
NOTICE
Code "55" is displayed when the system is operating without errors.
Code "60" also appears if the system is not equipped with the ultrasound option.

Error displays of the collimator interface: 3

Error No. Meaning
10 No first contact to the HOST (status message) 11 Timeout error from the HOST 12 Fatal error in the data reception block from the HOST 13 Illegal data in the reception block 14 Illegal value in the data reception flag from the HOST 15 No first contact to collimator N 16 Collimator timeout error 17 Data reception error from collimator 18 Continuous communication error to collimator 19 Reception queue full 1A DPR reception area error 1B CAN module error
1C Invalid values in reception block

Error displays of the operating master interface: 3

Error No.Meaning
20 No first contact to the HOST (status message) 21 Timeout error from the HOST 22 Fatal error in the data reception block from the HOST 23 Wrong block type in the data reception block from the HOST 24 Illegal value in the data reception flag from the HOST 25 No first contact to slaves (status message) 26 Slave timeout error 27 Data reception error from slave 28 Serial transmitter not free
2A DPR reception area error
7x Timeout slave no. x 8x Data reception error slave no. x
Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
3 - 2 Communication controller D6

Error displays of the image memory interface: 3

Error No.Meaning
30 No first contact to the HOST (status message) 31 Timeout error from the HOST 32 Fatal error in the data reception block from the HOST 33 Illegal data in the data reception block from the HOST 34 Illegal value in the data reception flag from the HOST 35 No first contact to the image memory (status message) 36 Image memory timeout error 37 Data reception error from image memory 38 Serial transmitter not free 3A Error in the DPR reception area

Error displays of the shock wave system interface: 3

Error No.Meaning
40 No first contact to the HOST (status message), System not registered 41 Timeout error from the HOST 42 Fatal error in the data reception block from the HOST 43 Illegal data in the data reception block from the HOST 44 Illegal value in the data reception flag from the HOST 45 No first contact to the shock wave system (status message) 46 Shock wave system timeout error 47 Data reception error from shock wave system
48 Serial transmitter not free 4A Error in the DPR reception area 4E Respiratory trigger signal timeout

Error displays of the brightness actual value interface: 3

Error No. Meaning
50 No first contact to the HOST (status message)
51 Timeout error from the HOST
52 Fatal error in the data reception block from the HOST
53 Illegal data in the data reception block from the HOST
54 Illegal value in the data reception flag from the HOST
55 No first contact to the brightness actual value (status mes-
sage)
56 Timeout error from the brightness value actual value signal
device 57 Data reception error of brightness actual value 58 Serial transmitter not free
5A Error in the DPR reception area
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Communication controller D6 3 - 3

Error displays of the ultrasound interface: 3

Error No. Meaning
60 No first contact to the HOST (status message) 61 Timeout error from the HOST 62 Fatal error in the data reception block from the HOST 63 Illegal data in the reception block from the HOST 64 Illegal value in the data reception flag from the HOST 65 No first contact to ultrasound system 66 Timeout error from ultrasound system 67 Error in block reception from ultrasound system 68 Error in block transmission to ultrasound system 69 Transmission buffer not empty (component)
6A Error in the DPR reception area
Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
3 - 4 Communication controller D6
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

CPU board D1 4

Overview and explanations of the unit drives 4

O = am4
4 - 1
W = am3
TVI = Television iris
on the television camera
TL = Cone longitudinal
TR = Cone rotation
SWO = Shock wave head, orbital
I.I. = Image intensifier lift
ZR = am1
ZL = am2
Y = am6
X = am5
OC arm Y Unit, longitudinal X Unit, lateral
ZR, ZL Table, up
W Shockwave head
Siemens AG RXL2-120.072.03 Page 1 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 2 CPU board D1

Definition of the slave numbers 4

Assigment: Slave no./Slave name
Slave no. Slave name Abbreviation
0 PC-Board HBF 2 Tableside control panel on PC NBF 3 Memoskop image memory BS 4 Memoskop image memory BSHD BSHD 7 Tableside control panel 1 (basic unit) NBF 8 Footswitch NBFFT 9 Tableside control panel (Ultrasound unit) NBFUS
10 Shock Wave System SWS
NOTICE
Errors in the error log are identified with a letter in addition to the error number.
E = Error
These errors can be cleared by the software during the next check.
F = Fatal Error
System operation is stopped. The error may be cleared by resetting the system (switching the system ON/OFF).
W = Warn
I = Information
ing
System operation may continue without restrictions.
These error numbers are for information only. They are not relevant to service.
Error messages in ERRO.LOG starting with FD... (Error text not available) are internal error messages without any field service.
meaning for
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 3

General errors 4

Error No. Error display on the service PC Meaning of the error
0001 Switch-on routine Switch-on routine: point in time at
which the unit was switched on. Identifies associated errors and required switch off/on proce-
dures. 0002 Service reset 0010 Contactors K1/K2 closed, no 27V present 0011 Contactor K3 closed, no 30V present no safety contactor closed? 0012 Contactors K1/K2 open, no 0V present 0013 Contactor K3 open, no 0V present 0020 Error: reference switch check / ZR am 1 Check the reference 0021 Error: reference switch check / ZL am 2 0022 Error: reference switch check /W am 3 0023 Error: reference switch check /O am 4 0025 Error: reference switch check / X am 5 0026 Error: reference switch check / Y am 6 0028 Error: reference switch check / TL am 8 0029 Error: reference switch check / TR am 9
002A Error: reference switch check / SWO am 10 002B Error: reference switch check / I.I. am 11
0050 Emergency stop is operated Emergency stop active 0070 ADC time is exceeded / converter 1 Error on board D5 0071 ADC time is exceeded / converter 2 Error on board D5 0080 ADC value out of tolerance / channel 0 ZR Drive 0081 ADC value out of tolerance / channel 1 ZL Drive 0082 ADC value out of tolerance / channel 2 W Drive 0083 ADC value out of tolerance / channel 3 O Drive 0084 ADC value out of tolerance / channel 4 X Drive 0085 ADC value out of tolerance / channel 5 Y Drive 0086 ADC value out of tolerance / channel 6 0087 ADC value out of tolerance / channel 7 0088 ADC value out of tolerance / channel 8 TVI Drive 0089 ADC value out of tolerance / channel 9
008A ADC value out of tolerance / channel 10 TL Drive 008B ADC value out of tolerance / channel 11 TR Drive 008C ADC value out of tolerance / channel 12 008D ADC value out of tolerance / channel 13 008E ADC value out of tolerance / channel 14
008F ADC value out of tolerance / channel 15
switch for errors 0020 to 002B
Siemens AG RXL2-120.072.03 Page 3 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 4 CPU board D1
Error No. Error display on the service PC Meaning of the error
0090 ADC reference voltage out of tolerance 5.12V 0091 24V ADC supply out of tolerance 24V 3V 0092 27V ADC supply out of tolerance 27V - 9V / + 13V 0093 30V ADC supply out of tolerance 3DV - 10V / + 9V
SS relay dropped out
0094 No signal of stopping potential
0F00 System is not initialized
The system is not initialized or the ROM version is wrong. The version num­ber of the HOST software and the parameters in the EEPROM do not match (the first two places). If, however, the software is present, then parameters are missing or the ver­sion numbers of the HOST software or the parameters in the EEPROM do not match. The error will be displayed after the do wnload routine . Repeat the do wnload routine.
± 60mV
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 5

Boot software 4

Error No. Error display on the
service PC
1000 Flash - EEPROM is empty Download HOST software again 1003 Flash EEPROM checksum
error
1004 No access to clock Set the clock again and perform Reset. If the
1005 Clock is not initialized Set the clock again and perform Reset. If the
1009 Incorrect checksum
with block receipt
100A Incorrect command
from Service PC
100B Incorrect block length
from Service PC
100C Incorrect segment address
from Service PC
100D Parity error Error in the communication between the ser-
100E Overrun error Receive buffer of the serial interface at the
100F Framing error Error in the communication between the ser-
1010 Break error Service PC was switched off before the serial
1011 Timeout error 1012 Error: Switch S2.8
(D1) = On
Meaning of the error
Incorrect checksum in the Flash EEPROMs. Download HOST software again.
error still exists, replace PC board D1.
error still exists, replace PC board D1. Incorrect checksum when a block is received
from the service PC/interface driver. The problem is in the service PC or HOST. Inter­ference during transmission.
Error in the data transmission between the service PC and the HOST. The service soft­ware version and the boot software version do not match. Compare with the compatibility list in the logbook.
Error in the data transmission between the service PC and the HOST. The service soft­ware version and the boot software version do not agree. Compare with the compatibility list in the logbook.
Incorrect segment address from the service PC for writing on EPROM. Error in IDX file; install the backup copy.
vice PC and the HOST. Hardware problem or electromagnetic compatiblity (EMC) interfer­ence. If necessary, perform the transfer again.
service PC has been exceeded.
vice PC and the HOST. Hardware problem or electromagnetic com­patibility (EMC) interference. If necessary, perform the transfer again.
connection closed. This message is also dis­played if hardware errors are present on board D1.
After three unsuccessful attempts to transfer, transmission between the HOST and the ser­vice PC is terminated.
Siemens AG RXL2-120.072.03 Page 5 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 6 CPU board D1

Operating system 4

Error No. Error text displayed on the service PC
Meaning of the error
3YXX 3 = Op. system Y = Slave XX = error type
0 = Operating PC (HBF) keyboard + mouse 2 = Tableside control/control desk(NBF) 7 = Tableside control/unit (NBF) 8 = Footswitch (NBFFT)
9 = Ultrasound operating panel (BFUS) 3001 HBF timeout 3002 Receive error of D6 from HBF (D6 = Communication Controller) 3003 NAK error of HBF from D6 (NAK = no acknowledge) 3004 HBF Sent invalid block length 3005 HBF didn't respond ACK/NAK 3009 Serial Interface of D6 is faulty 3030 HBF: Data error from D1 3032 HBF FIFO overrun during transmission 3033 HBF FIFO overrun during reception 3040 PC doesn't react to transmission 3041 Reception timeout, PC does not transmit 3042 Timeout of First PC contact after 60s 3043 Invalid data from PC 3044 Checksum error from PC 3045 New init request from PC 3050 Data overrun 3051 HBF: Timeout to D6 3081 HBF: Watchdog hit 3083 No contact to PC 3201 NBF Timeout 3202 Reception error of D6 from NBF 3203 NAK error of NBF from D6 (NAK = no acknowledge) 3204 NBF Sent invalid block length 3205 NBF didn´t respond ACK/NAK 3209 Serial interface of D6 is faulty 3231 NBF: Data error from D1 3250 Data overrun 3251 NBF: Timeout to D6 3252 NBF: Keyboard buffer overflow 3260 NBF: Actuated or defective key NBF:Actuated or defective key 3281 NBF: Watchdog hit 3701 NBF Timeout 3702 Reception error of D6 from NBF
HBF data update
LITHOSTAR Multiline RXL2-120.072.03 Page 6 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 7
Error No. Error text displayed on the service PC
3703 NAK error of NBF from D6 (NAK = no acknowledge) 3704 NBF Sent invalid block length 3705 NBF didn't respond ACK/NAK 3709 Serial Interface of D6 is faulty 3731 NBF: Data error from D1 3750 Data overrun 3751 NBF:Timeout to board D6 3752 NBF: Keyboard buffer overflow 3760 NBF: Actuated or defective key NBF: Actuated or defective key 3781 NBF: Watchdog hit 3801 NBFFT timeout 3802 Reception error of D6 from NBFFT 3803 NAK error of NBFFT from D6 (NAK = no acknowledge) 3804 NBFFT Sent invalid block length 3805 NBFFT didn't respond ACK/NAK 3809 Serial Interface of D6 is faulty 3831 NBFFT: Data error from D1 3850 Data overrun 3851 NBFFT: Timeout to board D6 3852 NBFFT: Keyboard buffer overflow 3860 NBFFT: actuated or defective key NBFFT: Actuated or defective
3881 NBFFT: Watchdog hit 3901 NBFUS:Timeout 3902 Reception error of D6 from NBFUS 3903 NAK error of NBFUS from D6 (NAK = no acknowledge) 3904 NBFUS: Sent invalid block length 3905 NBFUS: didn‘t respond ACK/NAK 3909 Serial Interface of D6 is faulty 3931 NBFUS: Data error from D1 3950 Data overrun 3951 3952 NBFUS: Keyboard buffer overflow 3960 NBFUS: actuated or defective key NBFUS: Actuated or defective
3981 NBFUS: Watchdog hit
NBFUS: Timeout to board D6
NBF data update
NBFFT data update
key
NBFUS data update
key
Siemens AG RXL2-120.072.03 Page 7 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 8 CPU board D1

Drive 4

Error No. Error text displayed on the service PC
4000 Protocol error of MC1 (D4a) 4001 Protocol error of MC2 (D4b) 4002 Protocol error of MC3 (D4c)
4100 Hardware error. MC1 (D4a) 4101 Hardware error. MC2 (D4b) 4102 Hardware error. MC3 (D4c)
4110 MC1 (D4a) disabled 4111 MC2 (D4b) disabled 4112 MC3 (D4c) disabled
4120 Target position from D1 too large / Z 4121 Target position from D1 too large / Z 4122 Target position from D1 too large / W am 3 4123 Target position from D1 too large / O am 4 4125 Target position from D1 too large / X am 5 4126 Target position from D1 too large / Y am 6 4127 Target position from D1 too large / TVI am 7 4128 Target position from D1 too large / TL am 8
4129 Target position from D1 too large / TR am 9 412A Target position from D1 too large / SWO am 10 412B Target position from D1 too large / I.I. am 11
4130 Invalid command from D1 / ZR am 1
4131 Invalid command from D1 / ZL am 2
4132 Invalid command from D1 / W am 3
4133 Invalid command from D1 / O am 4
4135 Invalid command from D1 / X am 5
4136 Invalid command from D1 / Y am 6
4137 Invalid command from D1 / TVI am 7
4138 Invalid command from D1 / TL am 8
4139 Invalid command from D1 / TR am 9 413A Invalid command from D1 / SWO am 10 413B Invalid command from D1 / I.I. am 11
R L
am 1 am 2
4140 Control error / ZR am 1
4141 Control error / ZL am 2
4142 Control error / W am 3
4143 Control error / O am 4
4145 Control error / X am 5
LITHOSTAR Multiline RXL2-120.072.03 Page 8 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 9
Error No. Error text displayed on the service PC
4146 Control error / Y am 6 4147 Control error / TVI am 7 4148 Control error / TL am 8
4149 Control error / TR am 9 414A Control error / SWO am 10 414B Control error / I.I. am 11
4150 Invalid parameter from D1 / ZR am 1
4151 Invalid parameter from D1 / ZL am 2
4152 Invalid parameter from D1 / W am 3
4153 Invalid parameter from D1 / O am 4
4155 Invalid parameter from D1 / X am 5
4156 Invalid parameter from D1 / Y am 6
4157 Invalid parameter from D1 / TVI am 7
4158 Invalid parameter from D1 / TL am 8
4159 Invalid parameter from D1 / TR am 9 415A Invalid parameter from D1 / SWO am 10 415B Invalid parameter from D1 / I.I. am 11
4160 Invalid position from D1 / ZR am 1
4161 Invalid position from D1 / ZL am 2
4162 Invalid position from D1 / W am 3
4163 Invalid position from D1 / O am 4
4165 Invalid position from D1 / X am 5
4166 Invalid position from D1 / Y am 6
4167 Invalid position from D1 / TVI am 7
4168 Invalid position from D1 / TL am 8
4169 Invalid position from D1 / TR am 9 416A Invalid position from D1 / SWO am 10 416B Invalid position from D1 / I.I. am 11
4170 Invalid data for motor initialization / ZR am 1
4171 Invalid data for motor initialization / ZL am 2
4172 Invalid data for motor initialization / W am 3
4173 Invalid data for motor initialization / O am 4
4175 Invalid data for motor initialization / X am 5
4176 Invalid data for motor initialization / Y am 6
4177 Invalid data for motor initialization / TVI am 7
4178 Invalid data for motor initialization / TL am 8
4179 Invalid data for motor initialization / TR am 9 417A Invalid data for motor initialization / SWO am 10 417B Invalid data for motor initialization / I.I. am 11
Siemens AG RXL2-120.072.03 Page 9 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 10 CPU board D1
Error No. Error text displayed on the service PC
4180 Motor activated without drive signal / ZR am 1 4181 Motor activated without drive signal / ZL am 2 4182 Motor activated without drive signal / W am 3 4183 Motor activated without drive signal / O am 4 4185 Motor activated without drive signal / X am 5 4186 Motor activated without drive signal / Y am 6 4187 Motor activated without drive signal / TVI am 7 4188 Motor activated without drive signal / TL am 8
4189 Motor activated without drive signal / T 418A Motor activated without drive signal / SWO am 10 418B Motor activated without drive signal / I.I. am 11
4190 Motor not activated by drive signal / ZR am 1
4191 Motor not activated by drive signal / ZL am 2
4192 Motor not activated by drive signal / W am 3
4193 Motor not activated by drive signal / O am 4
4195 Motor not activated by drive signal / X am 5
4196 Motor not activated by drive signal / Y am 6
4197 Motor not activated by drive signal / TVI am 7
4198 Motor not activated by drive signal / TL am 8
4199 Motor not activated by drive signal / TR am 9 419A Motor not activated by drive signal / SWO am 10 419B Motor not activated by drive signal / I.I. am 11
R
am 9
Check the fuses,
check the
output stages
(D7...),
check the LED on
D4
41A0 Drive speed not reached / ZR am 1 41A1 Drive speed not reached / ZL am 2 41A2 Drive speed not reached / W am 3 41A3 Drive speed not reached / O am 4 41A5 Drive speed not reached / X am 5 41A6 Drive speed not reached / Y am 6 41A7 Drive speed not reached / TVI am 7 41A8 Drive speed not reached / TL am 8 41A9 Drive speed not reached / TR am 9
41AA Drive speed not reached / SWO am 10 41AB Drive speed not reached / I.I. am 11
41B0 Motor rotation in reverse direction / ZR am 1 41B1 Motor rotation in reverse direction / ZL am 2 41B2 Motor rotation in reverse direction / W am 3 41B3 Motor in reverse direction / O am 4 41B5 Motor in reverse direction / X am 5
LITHOSTAR Multiline RXL2-120.072.03 Page 10 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 11
Error No. Error text displayed on the service PC
41B6 Motor in reverse direction / Y am 6 41B7 Motor in reverse direction / TVI am 7 41B8 Motor in reverse direction / TL am 8 41B9 Motor in reverse direction/ TR am 9 41BA Motor in reverse direction / SWO am 10 41BB Motor in reverse direction / I.I. am 11
41C0 Repeated fatal error of motor / ZR am 1 41C1 Repeated fatal error of motor / ZL am 2 41C2 Repeated fatal error of motor / W am 3 41C3 Repeated fatal error of motor / O am 4 41C5 Repeated fatal error of motor / X am 5 41C6 Repeated fatal error of motor / Y am 6 41C7 Repeated fatal error of motor / TVI am 7 41C8 Repeated fatal error of motor / TL am 8
41C9 Repeated fatal error of motor / TR am 9 41CA Repeated fatal error of motor / SWO am 10 41CB Repeated fatal error of motor / I.I. am 11
41D0 Drive timeout for motor indicated / ZR am 1
41D1 Drive timeout for motor indicated / ZL am 2
41D2 Drive timeout for motor indicated / W am 3
41D3 Drive timeout for motor indicated / O am 4
41D5 Drive timeout for motor indicated / X am 5
41D6 Drive timeout for motor indicated / Y am 6
41D7 Drive timeout for motor indicated / TVI am 7
41D8 Drive timeout for motor indicated / TL am 8
41D9 Drive timeout for motor indicated / TR am 9 41DA Drive timeout for motor indicated / SWO am 10 41DB Drive timeout for motor indicated / I.I. am 11
4280 ADC/MC pos. comparison failure / ZR am 1 4281 ADC/MC pos. comparison failure / ZL am 2 4282 ADC/MC pos. comparison failure / W am 3 4283 ADC/MC pos. comparison failure / O am 4 4285 ADC/MC pos. comparison failure / X am 5 4286 ADC/MC pos. comparison failure / Y am 6 4287 ADC/MC pos. comparison failure / TVI am 7 4288 ADC/MC pos. comparison failure / TL am 8
4289 ADC/MC pos. comparison failure / TR am 9 428A ADC/MC pos. comparison failure / SWO am 10 428B ADC/MC pos. comparison failure / I.I. am 11
Siemens AG RXL2-120.072.03 Page 11 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 12 CPU board D1
Error No. Error text displayed on the service PC Meaning of the
error
4290 Drive moves without actuation / ZR am 1 Drive moves without
signal / ZR
4291 Drive moves without actuation / ZL am 2 Drive moves without
signal / ZL
4292 Drive moves without actuation / W am 3 Drive moves without
signal / W
4293 Drive moves without actuation / O am 4 Drive moves without
signal / O
4295 Drive moves without actuation / X am 5 Drive moves without
signal / X
4296 Drive moves without actuation / Y am 6 Drive moves without
signal / Y
4297 Drive moves without actuation / TVI am 7 Drive moves without
signal / TVI
4298 Drive moves without actuation / TL am 8 Drive moves without
signal / TL
4299 Drive moves without actuation / TR am 9 Drive moves without
signal / TR
429A Drive moves without actuation / SWO am 10 Drive moves without
signal / SWO
429B Drive moves without actuation / I.I. am 11 Drive moves without
signal / I.I.
42A0 Error in averaging / ZR Error in averaging /
ZR
42A1 Error in averaging / ZL Error in averaging /
ZL
42A2 Error in averaging / W Error in averaging/
W
42A3 Error in averaging / O Error in averaging/
O
42A5 Error in averaging / X Error in averaging /
X
42A6 Error in averaging / Y Error in averaging /
Y
42A7 Error in averaging / TVI Error in averaging/
TVI
42A8 Error in averaging / TL Error in averaging /
TL
42A9 Error in averaging / TR Error in averaging /
TR
42AA Error in averaging / SWO Error in averaging /
SWO
42AB Error in averaging / I.I. Error in averaging /
I.I.
LITHOSTAR Multiline RXL2-120.072.03 Page 12 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 13
Error No. Error text displayed on the service PC Meaning of the
error
42B0 Outside max. operating limits / ZR Outside max. oper-
ating limits /ZR
42B1 Outside max. operating limits / ZL Outside max. oper-
ating limits /ZL
42B2 Outside max. operating limits / W Outside max. oper-
ating limits /W
42B3 Outside max. operating limits / O Outside max. oper-
ating limits /O
42B5 Outside max. operating limits / X Outside max. oper-
ating limits / X
42B6 Outside max. operating limits / Y Outside max. oper-
ating limits /Y
42B7 Outside max. operating limits / TVI Outside max. oper-
ating limits /TVI
42B8 Outside max. operating limits / TL Outside max. oper-
ating limits /TL
42B9 Outside max. operating limits / TR Outside max. opera-
ing limits /TR
42BA Outside max. operating limits / SWO Outside max. oper-
ating limits /SWO
42BB Outside max. operating limits / I.I. Outside max. oper-
ating limits /I.I.
Siemens AG RXL2-120.072.03 Page 13 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 14 CPU board D1

Shock Wave System 4

Error
No.
51XX Removable error restricts interface function marginally. The D1 can reset
5101 SWS timeout (Shock Wave System) 5102 Reception error of D6
5103 NAK error of SWS from
5104 SWS Sent invalid block
5105 SWS didn' t respond ACK/
5109 Serial Interface of D6 is
5111 Respiratory trigger time-
5150 Receive FIFO overrun 5151 Polling timeout 5152 Watchdog reset 5153 Nominal energy is not
5154 Shock counter doesn’t
Error text displayed on the service PC
the error and restart operation.
from SWS
D6.
length
NAK
faulty
out
actual
work D19
Meaning of the error
(NAK = no acknowledge)
Check the setting D19.R47
52XX
5201 Valve Y1 overheating W11/X1,z22 = high (no associated LED) 5202 Valve Y1 open load W11/X1,d4 = high (no associated LED) 5203 Valve Y1 short circuit W11/X1,d12 = high (no associated LED) 5204 Cooling system faulty W11/X1,d16 = high and yellow LED W11/V11 lights
5205 Water container empty W11/X1,z10 = high and yellow LED W11/V9 lights up 5206 Supply reserve W11/X1,b12 = high and yellow LED W11/V10 lights up 5207 Temperature too high W11/X1,z14 = high (no associated LED) 5241 Motor M20 overload W12/X1,b32 = high and red LED W12/V21 lights up 5261 Energy generation error W15/X1,z10 = high and yellow LED W15/V2 lights up 5262 Shockwave head error W15/X1,z14 = high and yellow LED W15/V3 lights up 5281 D19 self ignition D19/X1,b14 = high and yellow LED D19/V33 lights up 5282 D19 kA 5283 D19 U
Error. The interface controller malfunctioned. One or several active jobs will be aborted. The D1 can no longer reset the error.
up. Display at cooling water overtemperature is okay despite flow
actual
residual
< kA
min
is too high D19/X1,b12 = high and red LED D19/V9 lights up
D19/X1,b16 = high and yellow LED D19/V32 lights up
LITHOSTAR Multiline RXL2-120.072.03 Page 14 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 15
Error
No.
53XX
Error text displayed on the service PC
Fatal Error:
Meaning of the error
The occurrence of fatal errors blocks the controller interface. Resetting these errors by software is no longer possible.
5301 Pressure too high switch S4 Pressure > 800mbar
± 30 mbar switch S4
W11/X1,b14 = high and red LED W11/V14 lights up
5302 Pressure too low sensor SH Pressure < 210mbar + 30 mbar
W11/X1,z20 = high and red LED W11/V5 lights up; the cooling circuit pressure sensor signals that the system has not reached the control range.
5303 Temperature too high sen-
sor SH
Temperature > 38 C sensor W11/X1,b18 = high and red LED W11/V15 lights up; the cooling cir­cuit pressure temperature sensor signals that the system has exceeded the control range.
5304 Flowmeter faulty W11/X1,b14 = high and red LED W11/V13 lights
up. During the switch-on reset, flow okay has been signalled.
5305 Pressure too high sensor Pressure > 500mbar + 30 mbar sensor
W11/X1,d14 = high and red LED W11/V13 lights up; the cooling circuit pressure sensor signals that the control range has been exceeded.
5306 Flow (l/H) too low W11/X1,z12 = high
and red LED W11/V12 lights up
5307 Cooling system faulty W11/X1,d18 = low and green LED W11/V3 is dark
(= shot triggering inhibited)
5341 End of lift min/max W12/X1,z18 = high
and red LED W12/V17 lights up. (either the min. or the max. limit of lift is exceeded).
5361 kV-limit error W15/X1,z26 = high
and red LED W15/V5 lights up.
5362 Energy generation error W15/X1,z8 = high
and red LED W15/V1 lights up
5363 Shockwave generator error W15/X1,z16 = high
and red LED W15/V4 lights up
5364 Invalid kV value acknowl-
edge message
W15/X1,z16 = high and red LED W15/V4 lights up. Acknowledged kV value is wrong: Message soft­ware W16 (no associated LED) from W15. The kV value received deviates by >14% from the
expected value. 5381 D19 kV 5382 D19 kV 5383 D19 kA
actual actual actual
> kV
max
< > kV
> kA
max
selected
see chapter 7
5391 D91 Error memory see chapter 8 5392 D19Error memory
Siemens AG RXL2-120.072.03 Page 15 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 16 CPU board D1

General errors 4

Error No. Error text displayed on the service PC
6100 Checksum error / MC1 (D4a) 6101 Checksum error / MC2 (D4b) 6102 Checksum error / MC3 (D4c) 6103 Checksum error / Collimator 6104 Checksum error / Operator-Master 6105 Checksum error / Image Storage 6106 Checksum error / Shock Wave System 6107 Checksum error / Actual Brightness Value 6108 Checksum error / Ultrasound 6110 Wrong block length / MC1 (D4a) 6111 Wrong block length / MC2 (D4b) 6112 Wrong block length / MC3 (D4c) 6113 Wrong block length / Collimator 6114 Wrong block length / Operator-Master 6115 Wrong block length / Image Storage 6116 Wrong block length / Shock Wave System 6117 Wrong block length / Actual Brightness Value 6118 Wrong block length / Ultrasound 6120 Subsystem timeout / MC1 (D4a) 6121 Subsystem timeout / MC2 (D4b) 6122 Subsystem timeout / MC3 (D4c) 6123 Subsystem timeout / Collimator 6124 Subsystem timeout / Operator-Master 6125 Subsystem timeout / Image Storage 6126 Subsystem timeout / Shock Wave System 6127 Subsystem timeout / Actual Brightness Value 6128 Subsystem timeout / Ultrasound 6130 Start up tim out MC1 (D4a) 6131 Start up timeout MC2 (D4b) 6132 Start up timeout MC3 (D4c) 6133 Start up timeout / Collimator 6134 Start up timeout / Operator-Master 6135 Start up timeout / Image Store 6136 Start up timeout / Shock Wave System 6137 Start up timeout / Actual Brightness Value 6138 Start up timeout / Ultrasound
LITHOSTAR Multiline RXL2-120.072.03 Page 16 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 17

Ultrasound 4

Error
No.
7001 Timeout from US device Timeout from ultrasound system 7002 Error with block reception from
7003 Error during transmission to US
7004 Module emitter buffer not free Module emitter buffer not empty 7010 Incorrect ID in reception block Incorrect ID in reception block 7011 Incorrect length of reception
7100 No Litho US applicator connected No ultrasound probe for Litho connected 7101 Error with autom. selection of
7102 Error with autom. realtime selec-
7103 Error with autom. cursor display Error in automatic cursor display 7104 Reported target mark position
Error text displayed on the service PC
US unit
unit
block
probe 2
tion
incorrect
Meaning of the error
Error in block reception from ultrasound system
Error during transmission to ultrasound system
Incorrect length of reception block
Error in automatic selection of probe 2
Error in automatic realtime selection
Reported target position incorrect

Collimator 4

Error
No.
8001 Collimator timeout 8002 Collimator data reception error 8003 CAN-communication disturbance Possible defect on board D1 in the colli-
8004 Collimator signalling invalid data 8005 CAN chip is faulty 8010 RAM test D1 failure 8011 EEPROM test D1 failure 8012 CAN Error D6 8015 Collimator height: calculation
8016 Collimator width: calculation error 8020 Collimator height: stepping error 8021 Collimator width: stepping error 8024 Collimator height: limit switch
8025 Collimator width: limit switch
Error text displayed on the service PC
error
actuated
actuated
Meaning of the error
mator
Siemens AG RXL2-120.072.03 Page 17 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 18 CPU board D1

Generator 4

Error No. Error text displayed on the service PC
9000 Subsystem not configured 9010 Received block too long 9011 Timeout Board D2 9012 Init data transmitted after runup 9013 Faulty writing attempt 9020 XCU not ready Check fuse F8 in the
Videomed DIM 9021 Contact to XCU interrupted 9022 Turn-off message during operation 9023 XCU not transmitting data 9100 Invalid telegram mode 9101 Invalid AP_ID in telegram 9102 Invalid object ID from Board D2 9103 Invalid object length 9110 Invalid object ID for control unit 9111 Invalid value for automatic fluoro data 9112 Invalid value for fluoro alarm data 9113 Invalid value for mode fluoro data 9114 Invalid value for I.I. zoom 9115 Invalid value for TV image reversion 9116 Invalid value for x-ray data 9117 Invalid value for system data 9118 Invalid value for focus data 9119 Invalid operating mode
911A Invalid value for tube loading 911B Invalid value for acknowledgement byte
9120 Invalid obj.ID for I.I. functional unit 9121 Invalid value for I.I. func. unit: ZOOM
LITHOSTAR Multiline RXL2-120.072.03 Page 18 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 19

Image memory errors received from MEMOSKOP 4

NOTICE
Error text displayed on the service PC: "Error text not available"
Error No. Meaning of the error
E101 Hardware error on board D1 in the Memoskop.
E102 External RAM test error of the D1 processor in the Memoskop.
E103 Checksum test for the configuration byte is wrong (board D1).
E104 Checksum error Flash EPROM (board D1, Memoskop).
E105 Board D2 in the Memoskop does not respond.
E106 CPU 2 bus on board D2 defective. The internal tests in the Memoskop
E108 Memoskop: Prozessor D1 Stack overflow
E109 Image text incorrectly loaded the checksum test.
E402 No communication with CPU 2 after a Reset.
Error codes E101 to EDFF are messages from MEMOSKOP.
Internal test routine in the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
During the initialization routine, the RAMs are erased- if this is not possi­ble, this error is displayed. External test not possible. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Check the configuration (refer to the service instructions below). Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Check the software status in the Download menu. Ensure that the Memoskop status is "up to date". Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Check whether board D2 is correctly inserted in the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
show errors. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
No test Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Was a text allocated to the image? Check the 50/60 Hz configuration. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). Correct the configuration. Replace board D6.
Check whether boards D1 and D2 are correctly inserted in the Memoskop. Replace board D2 in the Memoskop. Replace the backplane board D5. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Siemens AG RXL2-120.072.03 Page 19 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 20 CPU board D1
Error No. Meaning of the error
E403 Timeout during communication with CPU 2
Check the software status in the Download menu. BILD.HEX Memoskop HARD.HEX hard disk Ensure that the Memoskop status is "up to date". If the status is "different" or "not available", perform the software download routine. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images)
E404 Error during polling of CPU 2 version.
Check the software status in the Download menu. Ensure that the Memoskop status is "up to date". If the statis is "different" or "not available", perform the software download routine. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E480
bis
E4E2
E601 Text controller test incorrect.
E602 Text controller RAM test incorrect.
E610 Keyboard interface test Baud rate too low.
E611 Keyboard interface test Baud rate too high.
E612 Keyboard interface test Baud clock missing.
E620 Keyboard interface test, feedback error
MEMOSKOP Hard disk error Board D2 defective Cable defective Hard disk defective
Internal test in the Memoskop. No external test possible. Replace text controller D6. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Internal test in the Memoskop. No external test possible. Replace text controller D6. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard of the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard of the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard or the keyboard cable of the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Replace the keyboard or the keyboard cable of the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
LITHOSTAR Multiline RXL2-120.072.03 Page 20 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 21
Error No. Meaning of the errors
E621 Keyboard Interface-Test Feedback Timeout.
Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E710 Memoskop video sync test: H sync signal period too long.
Check connections transformer T1 - plug D1/X13 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E711 Memoskop video sync test: H sync signal period too short.
Check connections transformer T1 -plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E712 Memoskop BAS (CVS) syc test: H sync signal missing.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E720 Memoskop video sync test: 50 Hz VR sync period too long.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E721 Memoskop video sync test: 50 Hz VR sync period too short.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E722 Memoskop video sync test: 60 Hz VR sync period too long.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E723 Memoskop video sync test: 60 Hz VR sync period too short.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
Siemens AG RXL2-120.072.03 Page 21 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 22 CPU board D1
Error No. Meaning of the errors
E724 Memoskop BAS (CVS) sync test: VR sync missing.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E730 Memoskop BAS (CVS) sync test: H sync level error
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E731 Memoskop BAS (CVS) sync test: Error PLL lock.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E740 Memoskop BAS (CVS) sync test: VR interrupt missing.
Check connections transformer T1 - plug D1/X3 in Memoskop. Check the fuses F2; F3; F4; F5. Replace the defective fuses. Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E750 Memoskop BAS (CVS) sync test: Incorrect callup parameters.
Multiline: check pixel clock and composite video signal of TV system. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). Replace the camera head of the Videomed DI.
E800 Memoskop voltage test: Reference voltage ADC too high.
No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E801 Memoskop voltage test: Reference voltage ADC too low.
No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E810 Memoskop voltage test: Voltage VPPmin. too high.
(Programming voltage for Flash-EPROMS). No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E811 Memoskop voltage test: Voltage VPPmin. too low.
No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E812 Memoskop voltage test: Voltage VPPmax. too high.
No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
LITHOSTAR Multiline RXL2-120.072.03 Page 22 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 23
Error No. Meaning of the errors
E813 Memoskop voltage test: Voltage VPPmax. too low.
No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E814 Memoskop voltage test. Voltage VPP not present.
Check the fuses in the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E820 Memoskop voltage test: Voltage 5V analog too high (>6V).
The LED +5V in the Memoskop must light with normal brightness. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E821 Memoskop voltage test: Voltage 5V analog too low (<4V).
Check the LED display in the Memoskop, the LED +5V (green) must light up. Check fuse F3 in Memoskop and replace, if necessary. Check connections X13 in Memoskop and correct, if necessary. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E830 Memoskop voltage test: Voltage -5.2V too high.
The LED 5V2 in the Memoskop must light with normal brightness. Check connections X13 in Memoskop and correct, if necessary. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E831 Memoskop voltage test: Voltage -5.2V too low.
Check fuse F4 in Memoskop and replace, if necessary. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E840 Memoskop voltage test: Voltage VEE analog -5V too high.
Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images).
E841 Memoskop voltage test: Voltage VEE analog -5V too low.
Check the LED "VEE" (green) on the Memoskop - the LED m ust light up. Check fuse F5 in Memoskop and replace, if necessary. (Monitor synchronization no longer present). Replace board D1 in the Memoskop
(Observe the configuration 50/60 Hz, 3 or 100 images). EA1H DPR err: write access invalid subsys.no. EA2H DPR err: write access too high subsys.no. EA3H DPR err: read access invalid subsys.no. EA4H DPR err: read access too high subsys.no. EC21 Memoskop not ready for download.
Check the LED displays on the Memoskop.
Check the connection: Service PC HOST/D1.
Try download routine again. Adjust the service PC, if necessary
Image memory (errors from Multiline)
Siemens AG RXL2-120.072.03 Page 23 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 24 CPU board D1
Error No. Meaning of the error
ED14 Memoskop Board D2 data checksum error during download
Error detected in checksum during data transfer.
during download of Memoskop Board D2
help: Malfunction of data line; Check plug connections; Board D1 defective
ED21 Memoskop D2 not ready for download
Request sent for download, however, Memoskop not yet ready.
during download of Memoskop Board D2
ED22 Memoskop Board D2 DBI too long during download
Excessively long data word detected during download.
during download of Memoskop Board D2
help: Fault in Memoskop; Board D2
ED23 Memoskop D2 address outside of area during download
Selected address lies outside of permissible area
during download of Memoskop Board D2
ED24 Memoskop Board D2 incorrect data checksum during download
Data transferred with incorrect checksum.
during download of Memoskop Board D2
Help: Malfunction of data line. Check plug connections.
ED31 Memoskop Board D2 FEPROM erase error during download
Memoskop cannot be cleared
during download of Memoskop Board D2
Help: Board D2, D1 defective. +15 V supply voltage, X13
ED32 Memoskop Board D2 FEPROM programming error during download
After each write operation to the FEPROM, the input value is read back. If
the value read does not agree with the value written, an error is displayed.
during download of Memoskop Board D2
Help: FEPROM defective; +15 V supply voltage,
X13 Board D2, D1 defective
ED33 Memoskop Board D2 FEPROM1 programming error during download
No data can be written to FEPROM1
during download of Memoskop Board D2
Help: FEPROM defective; +15 V supply voltage
X13 Board D2, D1 defective
ED34 Memoskop Board D2 FEPROM2 programming error during download
No data can be written to FEPROM2.
during download of Memoskop Board D2
Help: FEPROM defective +15 V supply voltage,
X13 Board D2, D1 defective
ED35 Memoskop Board D2 FEPROM erase error during download
The FEPROM is completely erased prior to programming.
If this proves impossible, an error message appears
during download of Memoskop Board D2
Help: FEPROM defective + 15 V supply voltage,
X13 Board D2, D1 defective
ED36 Memoskop Board D2 FEPROMS all bytes already 0
LITHOSTAR Multiline RXL2-120.072.03 Page 24 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 25
Error No. Meaning of the errors
ED41 Memoskop Board D2 unknown test number during download
An unknown test number was called up during download of Memoskop Board D2
ED42 Memoskop Board D2 unknown DBI in HX_REQUEST during download
Unknown data bit detected during polling of configuration during download of Memoskop Board D2
ED43 Memoskop Board D2 wrong segment during download
Software error during download of Memoskop Board D2
ED44 Memoskop Board D2 wrong test DBI during download
Software error during download of Memoskop Board D2
ED45 Memoskop Board D2 wrong segment during download
Software error
during download of Memoskop Board D2 ED51 Memoskop Board D2 unknown telegram from HOST during download ED52 Memoskop Board D2 receive error during download
Error in data reception by HOST
during download of Memoskop Board D2
Memoskop-HOST connection; Software error ED53 Memoskop Board D2 transmit error during download
Error in data transmission by HOST
during download of Memoskop Board D2
Memoskop-HOST connection ED61 Memoskop Board D2 +12 V Vpp out of tolerance during download
+ 12 V programming voltage not o.k.
during download of Memoskop Board D1
Help: Connector X13 on board D1; Board D1 defective;
Power unit voltage supply ED62 Memoskop D2 + 12,4V Vpp out of tolerance dur. download
+ 12 V programming voltage not o.k.
during download of Memoskop D1
Help: Connector X13 on board D1; Board D1 defective;
Power unit voltage supply ED63 Memoskop Board internal test failed during download
Error during internal test
during download of Memoskop Board D2 ED71 Memoskop Board D2 FEPROM checksum error during download
Checksum of FEPROMs incorrect
End of download of Memoskop Board D2
Help: FEPROM defective
Siemens AG RXL2-120.072.03 Page 25 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 26 CPU board D1

Image memory errors received from Multiline 4

NOTICE
Error No. Error text displayed on the service PC
EE00 Start of reset not signalled EE01 End of reset not signalled EE02 Unexpected reset timeout EE03 Timeout during job acknowledge EE04 Job terminated by error EE05 Job acknowledged due to invalid DBI-NR EE10 Invalid message of operator master EE17 Invalid DBI received EE18 Data block of wrong length received EE19 Loss of polling signal from BS detected BS = MEMOSKOP EF00 Invalid operating block (INFTYP) EF01 Invalid master command byte EF02 Invalid M_to_H-Message (Master to D1) EF03 Unknown error signalled EF04 Slave number not configured EF07 Master transmission error EF08 BS subsystem configured incorrectly EF09 BS subsystem configured incorrectly EF13 BS does not respond EF14 BSHD does not respond EF23 NAK reception error for BS NAK = no acknowledge EF24 NAK reception error for BSHD BSHD =
EF33 NAK reception error from BS EF34 NAK reception error from BSHD EF43 Invalid block length from BS EF44 Invalid block length from BSHD EF53 Image Storage does not respond ACK/NAK EF54 Hard disk does not respond ACK/NAK EF60 Acquisition active without fluoro EF70 Acquisition timeout during fluoro EF80 Timeout acknowledgement "SH_Ready" (SH = from Slave to D1) EF90 Timeout acknowledgement "SH_Stop"
Error codes EE00 to EF90 are messages from Multiline.
MEMOSKOP Harddisk
LITHOSTAR Multiline RXL2-120.072.03 Page 26 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
CPU board D1 4 - 27

General errors 4

Error No. Error text displayed on the service PC Meaning of the error
F000 Incompatible AST version F001 Z_No. of execution control too large F010 Software timer array overflow F100 Invalid Watchdog configuration F200 Watchdog has responded F500 C-MOS Data (D1) are deleted F600 EEPROM error (checksum) F700 RAM error (checksum)

Boot Software 4

Error No. Error text displayed on the ser-
vice PC
F800 Error in watchdog test Perform Reset; F801 Error in RAM test
F802 Battery discharge Replace the battery F803 Error in Test1 COM1 D1.X10 Perform Reset; F804 Error in Test1 COM2 D1.X11 F807 Timer 0 error (D1) F808 Timer 1 error (D1)
F809 Timer 2 error (D1) F80A Timer 3 error (D1) F80B Timer 4 error (D1) F80C Timer 5 error (D1) F80D Flash - EEPROM on D1 defective F80E EEPROM - checksum is incorrect
F80F Watchdog counter init error
F810 EPROM counter too high (D1)
F811 The system is not initialized
F812 Error in writing process
F813 Error in comparison of desired and
actual values
F814 Error while writing checksum
Remedial actions
if no change: Perform DOWN­LOADroutine
If no change: Replace board D1.
if no change: Perform DOWN­LOADroutine
If no change: Replace board D1.
Siemens AG RXL2-120.072.03 Page 27 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
4 - 28 CPU board D1

Internal software errors 4

Error No. Error text displayed on the service PC Meaning of the error
FE00 Invalid motor number FE01 Invalid ADC channel number FE02 Queue overrun for LED control FE03 Software internal error FE04 Temporary buffer overflow messages FE05 Alternating buffer overflow messages FE06 Acknowledgeable buff. overflow messages FE07 Buffer overflow of keyboard storage FE08 Msg No <> Err No with msg delection FE40 Invalid BS command (PUSH_BS_CMD) FE41 Invalid BS status (S_BS_Stat) FE42 Invalid parameter (G_BS_Slave) FE43 Invalid parameter (G_BS_CmdAct) FE44 Invalid parameter (S_BS_CmdAct) FE45 Invalid parameter (G_BS_Stat) FE46 Invalid parameter (HS-Modus) FE47 Transfer data buffer overflow FE48 Data block buffer overflow FE70 Parameter error "PUSH_US_CMD" FE80 Target position outside
admission position / X
FE81 Target position outside
admission position / Y
FE82 Target position outside
admission position / Z FE90 Parameter error "PUSH_XCS_CMD" FE91 Parameter error "Fluoro mode" FE92 Parameter error "Direct_mode" FE93 Inadmissible value in "FE_BV_ZOOM" Unacceptable value
FE94 Inadmissible value in "XCS_STAT()" Unacceptable value
Target position outside permis­sible position (X drive)
Target position outside permis­sible position (Y drive)
Target position outside permis­sible position (Z drive)
in "FE_BV_Zoom"
in "XCS_STAT()"
LITHOSTAR Multiline RXL2-120.072.03 Page 28 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Boot and error messages of the operating PC 5

Operating system errors 5

"Start-Up-Message: No HOST-Communication"
Meaning: This message is displayed during booting of the operating PC as long as no
Dual Port RAM (DPR) communication is running;
"Interface-Error: DUEP-POLLING-TIME-OUT"
Meaning: The plug-in card reports via the DPR interface that polling is no longer rec-
ognized at the DUEP interface (Communication fault);
"Interface-Error: HOST-DATA-OVERFLOW"
Meaning: If the HOST transmits data faster than can be forw arded to the operating PC ,
then this message is displayed;
"Interface-Error: NEW-INIT-REQUEST"
Meaning: The plug-in card requests resynchronization of the DPR communication
because of internal errors (e.g. watchdog reset);
5 - 1
"Interface-Error: TIME-OUT-INIT"
Meaning: A timeout was detected during the booting synchronization of the DPR inter-
face;
"Interface-Error: TIME-OUT-SEND"
Meaning: A timeout has occurred while transmitting data via DPR interface to the plug-
in card (DPR communication error);
"Interface-Error: INVALID-DPR-FLAG"
Meaning: The DPR communication protocol was not observed
(DPR communication fault);
"Interface-Error: CHECKSUM-ERROR"
Meaning: A checksum error has been detected while receiving data via DPR interface
(DPR communication error);
"Interface-Error: RECEIVE-BUFFER FULL"
Meaning: The operating PC has received data too fast and an internal data overflow
has occurred;
"Interface-Error: SEND-BUFFER FULL"
Meaning: The operating PC has generated more data blocks than can be sent and
thus triggered an internal data overflow;
Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
5 - 2 Boot and error messages of the operating PC

Board D51 of the operating PC 5

7-segment display
The following messages are shown on the 7-segment display:
0 Normal operation (DUEP polling and DPR interface OK) 1 DUEP communication error, but DPR interface OK 2 DPR communication error, but DUEP interface OK 3 DPR and DUEP communication error 4 Error in the analog part of the board or on the processor 5 Error on the internal timer of the processor 6 External RAM error 7 EPROM or program RAM error 8 Status after reset 9nc
Decimal point
80C535 works=>DP flashes (ADC/D<DUEP card sign of life)
LED displays:
The following messages are output via LEDs:
- Watchdog has responded
- Card is being booted by PC
- AT bus working (PC sign of life)

AP - IDxx displays 5

The following AP-ID codes are displayed on the control monitor
Meaning Meaning AP-ID: 16 VIDEOMED DI AP-ID: 241 Service PC AP-ID: 80 POLYDOROS AP-ID: 242 MULTILINE AP-ID: 240 XCU AP-ID : 243 I.I. system
These error codes indicate the error group. In order to receive the specific error informa­tion, the service PC must be connected to the XCU (generator D200) and then the error memory can be read out.
The POLYDOROS LX 30/50 Troubleshooting instructions R63-020.072...
POLYDOROS SX 65/80 Troubleshooting instructions R63-055.072...
and the VIDEOMED DI/DIM Service instructions RX52-022.061...
contain information on the causes of errors.
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

XCS controller D2 6

Error No.: Meaning: Remedial actions:
00 Normal operation ok
One of the HW tests
not successful
10-IF Switch the component off 39-IF ACS software errors and then on again or re­51-53 place PC board or (PROMs)
20-2F 37-38 Operating error by HOST-SW Correct on HOST system 55-5F
36 Temporary ACS 50 high load problem Recovers by itself
Replace PC board
6 - 1
34 35 Communication with the service PC 63 disrupted or terminated
30 Power failure with subsequent
reconfiguration
31 No network partner found
or FOC not connected
32 Node-ID in network is present
several times
33 Transmission repeat normal or unstable
54 ACS firmware in network
not compatible
60 No network partner found normal or FOC PCBs must
normal or [power line] unstable PCB. FOC must be checked
normal or FOC. PCBs must be checked
Check node-IDs in the network
PCB, FOC must be checked
Check the ACS SW version
be checked
61, 62 No connection to the
HOST system
V1 Decimal point changes frequently: ACS net is not stable; along with the
message number 30
V2 Decimal point flashes: correct program run
Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
Check the HOST system
6 - 2 XCS controller D2
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Error displays on D19 7

Overview of operating status displays 7

D19
7 - 1
Shot release, V
act
Test ON, I-monitoring (flashes when S1 in position 2)
UC
Rest
kV
kV
actual
actual
kV
> kV
General fault
Short circuit
Open in discharge circuit
Spark gap error, after 2 unsuccess­ful attempts without triggering
= kV
> 60V
nominal
nominal
nom.
V6
V36
V9
V7
V10
V26
V31
V32
V33
R7A
S1
S2
R60
R47
MP2:
kV
nominal
(3.0 V ± 0.03 V with 15.1 kV)
Position 1: Normal operating mode Position 2: Error simulation
Reset for error memory
MP4: X3 x 1.33
MP3: X3 x 2.5
Fig. 1
Check the LEDs on circuit board D19 (Fig.1) to determine which monitoring control has responded.
Red LED V7: kV nominal value (kV
kV actual value (kV
nominal
nominal
Red LED V10: The kV actual value (kV
Yellow LED V31: The current through the shock wave head l
Yellow LED V32: The current through the shock wave head l
) from board D15 and the
) at the capacitor differ.
) at the capacitor is greater than 23 kV.
actual
is too high.
dis
is too low.
dis
Yellow LED V33: Spark gap ignition occurred twice without triggering.
NOTICE
For settings on board D19, refer to instructions "Repair of shock wave system" RXL2-120.091...
Siemens AG RXL2-120.072.03 Page 1 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
7 - 2 Error displays on D19
V7 (kV
actual
kV
nominal
Produce oscilloscope
display of the voltages
on board D19
actual
nominal
(MP1)
(MP2)
kV
act
not o.k.
kV
and kV
Check kV indication
on charging unit to determine
whether the charging unit
has charged
up to the required value
yes
no
Measuring cable U,
shock wave generator X4
or
board D19.X2 defective
) / Board D1 display: 5382 7
kV o.k.
act
kV
act
no
Does kV
correspond
to the value set ?
Compare with
table 1
no
Replace circuit
board D19
=kV
?
nom.
nominal
yes
Check and adjust
as necessary
± 15V ± 0.1 V
supply voltage Replace board
D19
yes
Charging unit not charging up
Charging unit
defective
Unplug high-voltage
cable K26
Charging unit charging up
High-voltage cable K26 or
shock wave generator
defective
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Error displays on D19 7 - 3
Red LED V10 (kV
Produce oscilloscope display of the voltages on board D19 kV
kV
< 5
nom.
V
> kV
actual
nominal
(measuring point MP1) and kV
kV
< 23 kV
act
Check the kV code at cable K108 on the
charging unit side
(table 2)
Code not o.k.
Check the kV code on board D15
connector X2
) / Board D1 display: 5381 7
actual
nominal
(measuring point MP2)
kV
< 23 kV
act
1 V 10 kV
Code o.k.
Replace board D19
Replace charging unit
(table 2)
Code not o.k.
Code o.k.
Replace cable K108
Replace board D15
Siemens AG RXL2-120.072.03 Page 3 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
7 - 4 Error displays on D19

Yellow LED V32 (discharge current too low) / Board D1 display:5282 7

Check lead X3 (I
meas
)
(if necessary, change over two leads at the
capacitor and at D19)
Error removed
Error not removed
Check adjustment of the circuit board D19
Error not removed
Replace board D19
Error not removed
Check trigger cable X2
Interchange pick-
up leads X3
Error not removed
Replace shock head
Error not removed
Replace spark gap module

Yellow LED V33 (spontaneous ignition of the spark gap) Board D1 Display: 5281 7

Reset on keyboard with ESC
Error occurs repeatedly
Replace spark gap module
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Error displays on D19 7 - 5

Yellow LED V31 (discharge current too high) Board D1 display: 5383 7

Replace shock
wave head
Error not removed
Replace high-voltage cable
between shock wave generator and shock wave head
Error not removed
Check adjustment of the circuit board D19
Error not removed
Replace
board D19
Siemens AG RXL2-120.072.03 Page 5 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
7 - 6 Error displays on D19
kV reference values
Energy levels
on
keyboard
0.1 9.5 1.9 V ± 0.03 V
0.2 9.9
0.3 10.4
0.4 10.8
0.5 11.2
0.6 11.7
0.7 12.1
0.8 12.6
0.9 12.9
1.0 13.4
1.5 13.7
2.0 13.9
2.5 14.2
3.0 14.5
kV Indication charging unit
± 0.2 kV
D19 MP 2
(1 kV = 0.2 V)
3.5 14.7
4.0 15.1 3.0 V ± 0.03 V
4.5 15.3
5.0 15.5
5.5 15.9
6.0 16.2
6.5 16.5
7.0 16.8
7.5 17.1
8.0 17.4
8.5 17.8
9.0 18.1 3.64 V ± 0.03 V
9.5 18,3
10.0 18.4
10.5 18.7
11.0 18.9
11.5 19.3
12.0 19.6
Tab. 1
LITHOSTAR Multiline RXL2-120.072.03 Page 6 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Error displays on D19 7 - 7
MSB LSB 0 V
W15 d32 d30 d28 d26 d24 d22 d20 d18 z24
K6 3 4 5 6 7 10 11 12 2
K108. N12 3 4 5 6 7 10 11 12 18
Tab. 2
Energy
levels
K108.
charging unit
0.1 01010001
0.2 01010010
0.3 01011000
0.4 01011100
0.5 01011111
0.6 01100011
0.7 01100111
0.8 01101011
0.9 01101110
1.0 01110010
1.5 01110100
2.0 01110110
2.5 01111001
3.0 01111011
3.5 01111101
4.0 10000000
4.5 10000010
5.0 10000100
5.5 10000111
6.0 10001010
6.5 10001100
7.0 10001111
7.5 10010001
8.0 10010101
8.5 10010111
9.0 10011010
9.5 10011011
10.0 10011100
10.5 10011111
11.0 10100001
11.5 10100100
12.0 10100111
123456789
K108 pin assignment
1 8
on the charging unit
9
Siemens AG RXL2-120.072.03 Page 7 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
15
7 - 8 Error displays on D19
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 8 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Error displays on D91 8

Overview of operating status displays 8

D91
Nom. charging value
R38
8 - 1
Fig. 1
V20
V4
V7
t
charge actual
U/
actual
Start pulse, charging unit
U/2
> t
charge nominal
nominal
Reset Energy monitoring
S1
Check the LEDs on circuit board D91 (Fig.1) to determine which monitoring controls have responded.
Red LED V4: Half the capacitor voltage U/2 and full capacitor voltage U differ.
Yellow LED V20: The acceptable charging time has been exceeded.
(The LED lights only briefly in the event of an error)
When the shock wave generator has produced more than one million shots, the high-volt age stability may have decreased. This makes it likely that board D91 has responded.
NOTICE
For settings on board D91, refer to the instructions "Maintenance of shockwave system" RXL2-120.091...
Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
8 - 2 Error displays on D91
Red LED V4 (kV
actual
Produce oscilloscope display
of the voltages on D91.b20
kV
actual
and
(cathode diode V1).
1
kV
act/2
/2 kV
kV
act
kV
> kV
actual/2
not equal to kV
1
/2 kV
0
act/2
= kV
act/2
kV
act/2
act
< kV
act/2
nominal
act
= 0
) / Board D1 display: 5391 8
Check signal path
to board D19
Check lead X5 U/
Check supply voltage (±15V ±0,1 V)
on board D1;
replace board D91
2
Replace
leads X4
and X5
kV
kV
act
act
< kV
> kV
act/2
Check leads
act/2
Replace shock wave generator
kV
kV
act
act
> kV
< kV
Replace
leads X4
and X5
act/2
act/2
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions
Error displays on D91 8 - 3

Yellow LED V20 (charging time exceeded Board D1 display: 5391 8

Check automatic circuit breaker
on charging unit
Check switch position on charging unit
Charging unit Switch Position
HPE
LS21 - 1,5
Voltage Nom value
Mode Remote
Output on
Display the voltages kV
(D91.b20) on
board D91 on oscilloscope
Display the voltages kV
kV
actual
not o.k.
act
actual
kV
act
1 V = 10 kV
(D91.b20 on board D91 on oscilloscope)
noyes
Measuring cable
to theboard D91
Disconnect high-volt-
age cable K26
defective
o.k.
Display charge end pulse
on board D91.z24
on oscilloscope
Charge end not o.k.
Check signal path
from D91.z24
up to the charging
unit
Charge end o.k.
Check setting of
charge time
Setting correct
Replace circuit
Charging unit not charging up
board D91
Replace shock
Charging unit charging up
Charging unit
defective
Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
High voltage cable K26 or
Shock wave generator defective
wave generator
8 - 4 Error displays on D91
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Errors without error numbers 9

Letters appear scattered across the monitor. Cause: defective text module on board D1 in the Memoskop.
9 - 1
Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
9 - 2 Errors without error numbers
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions

Changes as compared to previous version 10

Chapter 0 Revision level increased, table of contents adjusted
Chapter 4 - 27 Error F812 - new inserted
Error F813 - new inserted
Error F814 - new inserted
10 - 1
Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System
10 - 2 Changes as compared to previous version
This page intentionally left blank.
LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions