English
Print No.: RXL2-120.072.03.04.02 Doc. Gen. Date:06.02
Replaces: RXL2-120.072.03.03.02
0 - 2Revision level
ChapterPageRevision
allall04
Document revision level0
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcop y documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer0
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Yellow LED V20 (charging time exceeded Board D1 display: 5391 . . . . . . . . .8 -3
9 ______ Errors without error numbers_____________________________________ 9 -1
10 _____ Changes as compared to previous version ________________________ 10 -1
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
10 LM_628_error
11 Control_error
12 Position accuracy not reached in stated stabilizing time
13 Motor not initialized, drive command received
14 Motor moves without actuation
15 Motor does not move when actuated
16 Speed not reached or exceeded
17 Motor moves in wrong direction
18 Motor unsuccessfully actuated three times
19 Faulty initialization of the ASIC
1 - 1
ZL am2 (Lifting column left)1
Error number
↓ Meaning:
20 LM_628_error
21 Control_error
22 Position accuracy not reached in stated stabilizing time
23 Motor not initialized, drive command received
24 Motor moves without actuation
25 Motor does not move when actuated
26 Speed not reached or exceeded
27 Motor moves in wrong direction
28 Motor actuated unsuccessfully three times
29 Faulty initialization of the ASIC
Siemens AGRXL2-120.072.03Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
1 - 2Motor controller D4a
W drive am31
Error number
↓ Meaning:
30 LM_628_error
31 Control_error
32 Position accuracy not reached in stated stabilizing time
33 Motor not initialized, drive command received
34 Motor moves without actuation
35 Motor does not move when actuated
36 Speed not reached or exceeded
37 Motor moves in wrong direction
38 Motor unsuccessfully actuated three times
39 Faulty initialization of the ASIC
O drive am4 (C-Arm)1
Error number:
↓ Meaning:
40 LM_628_error
41 Control_error
42 Position accuracy not reached in stated stabilizing time
43 Motor not initialized, drive command received
44 Motor moves without actuation
45 Motor does not move when actuated
46 Speed not reached or exceeded
47 Motor moves in wrong direction
48 Motor unsuccessfully actuated three times
49 Faulty initialization of the ASIC
Board errors, general1
Error number:
↓ Meaning:
60 No first contact with HOST
61 TIMEOUT, data exchange disrupted
62 CHECKSUM_error
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 2Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
10 LM_628_error
11 Control_error
12 Position accuracy not reached in stated stabilizing time
13 Motor not initialized, drive command received
14 Motor moves without actuation
15 Motor does not move when actuated
16 Speed not reached or exceeded
17 Motor moves in wrong direction
18 Motor actuated unsuccessfully three times
19 Faulty initialization of the ASIC
2 - 1
Y drive am62
Error number
↓Meaning:
20 LM_628_error
21 Control_error
22 Position accuracy not reached in stated stabilizing time
23 Motor not initialized, drive command received
24 Motor moves without actuation
25 Motor does not move when actuated
26 Speed not reached or exceeded
27 Motor moves in wrong direction
28 Motor actuated unsuccessfully three times
29 Faulty initialization of the ASIC
Siemens AGRXL2-120.072.03Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
2 - 2Motor controller D4b
TVI Television iris am72
Error number
↓Meaning:
30 LM_628_error
31 Control_error
32 Position accuracy not reached in stated stabilizing time
33 Motor not initialized, drive command received
34 Motor moves without actuation
35 Motor does not move when actuated
36 Speed not reached or exceeded
37 Motor moves in wrong direction
38 Motor unsuccessfully actuated three times
39 Faulty initialization of the ASIC
TL Cone drive, longitudinal, am82
Error number
↓Meaning:
40 LM_628_error
41 Control_error
42 Position accuracy not reached in stated stabilizing time
43 Motor not initialized, drive command received
44 Motor moves without actuation
45 Motor does not move when actuated
46 Speed not reached or exceeded
47 Motor moves in wrong direction
48 Motor unsuccessfully actuated three times
49 Faulty initialization of the ASIC
TR Cone drive, rotation, am92
Error number
↓Meaning:
50 LM_628_error
51 Control_error
52 Position accuracy not reached in stated stabilizing time
53 Motor not initialized, drive command received
54 Motor moves without actuation
55 Motor does not move when actuated
56 Speed not reached or exceeded
57 Motor moves in wrong direction
58 Motor unsuccessfully actuated three times
59 Faulty initialization of the ASIC
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Motor controller D4b2 - 3
Board errors, general2
Error number:
↓ Meaning:
60 No first contact with HOST
61 TIMEOUT, data exchange disrupted
62 CHECKSUM_error
Siemens AGRXL2-120.072.03Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
2 - 4Motor controller D4b
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Communication controller D63
3 - 1
NOTICE
Code "55" is displayed when the system is operating without
errors.
Code "60" also appears if the system is not equipped with the
ultrasound option.
Error displays of the collimator interface:3
Error No. Meaning
10 No first contact to the HOST (status message)
11 Timeout error from the HOST
12 Fatal error in the data reception block from the HOST
13 Illegal data in the reception block
14 Illegal value in the data reception flag from the HOST
15 No first contact to collimator N
16 Collimator timeout error
17 Data reception error from collimator
18 Continuous communication error to collimator
19 Reception queue full
1A DPR reception area error
1B CAN module error
1C Invalid values in reception block
Error displays of the operating master interface:3
Error No.Meaning
20 No first contact to the HOST (status message)
21 Timeout error from the HOST
22 Fatal error in the data reception block from the HOST
23 Wrong block type in the data reception block from the HOST
24 Illegal value in the data reception flag from the HOST
25 No first contact to slaves (status message)
26 Slave timeout error
27 Data reception error from slave
28 Serial transmitter not free
2A DPR reception area error
7x Timeout slave no. x
8x Data reception error slave no. x
Siemens AGRXL2-120.072.03Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
3 - 2Communication controller D6
Error displays of the image memory interface:3
Error No.Meaning
30 No first contact to the HOST (status message)
31 Timeout error from the HOST
32 Fatal error in the data reception block from the HOST
33 Illegal data in the data reception block from the HOST
34 Illegal value in the data reception flag from the HOST
35 No first contact to the image memory (status message)
36 Image memory timeout error
37 Data reception error from image memory
38 Serial transmitter not free
3A Error in the DPR reception area
Error displays of the shock wave system interface:3
Error No.Meaning
40 No first contact to the HOST (status message), System not registered
41 Timeout error from the HOST
42 Fatal error in the data reception block from the HOST
43 Illegal data in the data reception block from the HOST
44 Illegal value in the data reception flag from the HOST
45 No first contact to the shock wave system (status message)
46 Shock wave system timeout error
47 Data reception error from shock wave system
48 Serial transmitter not free
4A Error in the DPR reception area
4E Respiratory trigger signal timeout
Error displays of the brightness actual value interface:3
Error No. Meaning
50 No first contact to the HOST (status message)
51 Timeout error from the HOST
52 Fatal error in the data reception block from the HOST
53 Illegal data in the data reception block from the HOST
54 Illegal value in the data reception flag from the HOST
55 No first contact to the brightness actual value (status mes-
sage)
56 Timeout error from the brightness value actual value signal
device
57 Data reception error of brightness actual value
58 Serial transmitter not free
5A Error in the DPR reception area
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
Communication controller D63 - 3
Error displays of the ultrasound interface:3
Error No. Meaning
60 No first contact to the HOST (status message)
61 Timeout error from the HOST
62 Fatal error in the data reception block from the HOST
63 Illegal data in the reception block from the HOST
64 Illegal value in the data reception flag from the HOST
65 No first contact to ultrasound system
66 Timeout error from ultrasound system
67 Error in block reception from ultrasound system
68 Error in block transmission to ultrasound system
69Transmission buffer not empty (component)
6A Error in the DPR reception area
Siemens AGRXL2-120.072.03Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
3 - 4Communication controller D6
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 4Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14
Overview and explanations of the unit drives4
O = am4
4 - 1
W = am3
TVI=Television iris
on the television camera
TL=Cone longitudinal
TR=Cone rotation
SWO=Shock wave head, orbital
I.I.=Image intensifier lift
ZR = am1
ZL = am2
Y = am6
X = am5
OC arm
YUnit, longitudinal
XUnit, lateral
ZR, ZLTable, up
WShockwave head
Siemens AGRXL2-120.072.03Page 1 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 2CPU board D1
Definition of the slave numbers4
Assigment: Slave no./Slave name
Slave no.Slave nameAbbreviation
0PC-BoardHBF
2Tableside control panel on PCNBF
3Memoskop image memoryBS
4Memoskop image memory BSHDBSHD
7Tableside control panel 1 (basic unit)NBF
8FootswitchNBFFT
9Tableside control panel (Ultrasound unit)NBFUS
10Shock Wave SystemSWS
NOTICE
Errors in the error log are identified with a letter in addition to the error number.
E = Error
These errors can be cleared by the software during the next check.
F = Fatal Error
System operation is stopped. The error may be cleared by resetting the system
(switching the system ON/OFF).
W = Warn
I = Information
ing
System operation may continue without restrictions.
These error numbers are for information only. They are not relevant to service.
Error messages in ERRO.LOG starting with FD... (Error text not
available) are internal error messages without any
field service.
meaning for
LITHOSTAR MultilineRXL2-120.072.03Page 2 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 3
General errors4
Error No. Error display on the service PCMeaning of the error
0001 Switch-on routineSwitch-on routine: point in time at
which the unit was switched on.
Identifies associated errors and
required switch off/on proce-
dures.
0002 Service reset
0010 Contactors K1/K2 closed, no 27V present
0011 Contactor K3 closed, no 30V present no safety contactor closed?
0012 Contactors K1/K2 open, no 0V present
0013 Contactor K3 open, no 0V present
0020 Error: reference switch check / ZRam 1Check the reference
0021 Error: reference switch check / ZLam 2
0022 Error: reference switch check /Wam 3
0023 Error: reference switch check /O am 4
0025 Error: reference switch check / X am 5
0026 Error: reference switch check / Y am 6
0028 Error: reference switch check / TLam 8
0029 Error: reference switch check / TRam 9
0050 Emergency stop is operated Emergency stop active
0070 ADC time is exceeded / converter 1Error on board D5
0071 ADC time is exceeded / converter 2Error on board D5
0080 ADC value out of tolerance / channel 0ZR Drive
0081 ADC value out of tolerance / channel 1ZL Drive
0082 ADC value out of tolerance / channel 2W Drive
0083 ADC value out of tolerance / channel 3O Drive
0084 ADC value out of tolerance / channel 4X Drive
0085 ADC value out of tolerance / channel 5Y Drive
0086 ADC value out of tolerance / channel 6
0087ADC value out of tolerance / channel 7
0088 ADC value out of tolerance / channel 8TVI Drive
0089ADC value out of tolerance / channel 9
008AADC value out of tolerance / channel 10TL Drive
008BADC value out of tolerance / channel 11TR Drive
008CADC value out of tolerance / channel 12
008DADC value out of tolerance / channel 13
008EADC value out of tolerance / channel 14
008FADC value out of tolerance / channel 15
switch
for errors 0020 to 002B
Siemens AGRXL2-120.072.03Page 3 of 28LITHOSTAR Multiline
Medical SolutionsRev. 0406.02CS PS 24System
4 - 4CPU board D1
Error No. Error display on the service PCMeaning of the error
0090ADC reference voltage out of tolerance5.12V
009124V ADC supply out of tolerance24V 3V
009227V ADC supply out of tolerance27V - 9V / + 13V
009330V ADC supply out of tolerance3DV - 10V / + 9V
SS relay dropped out
0094No signal of stopping potential
0F00System is not initialized
The system is not initialized or the ROM version is wrong. The version number of the HOST software and the parameters in the EEPROM do not match
(the first two places).
If, however, the software is present, then parameters are missing or the version numbers of the HOST software or the parameters in the EEPROM do
not match.
The error will be displayed after the do wnload routine . Repeat the do wnload
routine.
± 60mV
LITHOSTAR MultilineRXL2-120.072.03Page 4 of 28Siemens AG
SystemRev. 0406.02CS PS 24Medical Solutions
CPU board D14 - 5
Boot software4
Error No. Error display on the
service PC
1000Flash - EEPROM is emptyDownload HOST software again
1003Flash EEPROM checksum
error
1004No access to clockSet the clock again and perform Reset. If the
1005Clock is not initializedSet the clock again and perform Reset. If the
1009Incorrect checksum
with block receipt
100AIncorrect command
from Service PC
100BIncorrect block length
from Service PC
100C Incorrect segment address
from Service PC
100DParity errorError in the communication between the ser-
100EOverrun errorReceive buffer of the serial interface at the
100FFraming errorError in the communication between the ser-
1010Break errorService PC was switched off before the serial
1011Timeout error
1012Error: Switch S2.8
(D1) = On
Meaning of the error
Incorrect checksum in the Flash EEPROMs.
Download HOST software again.
error still exists, replace PC board D1.
error still exists, replace PC board D1.
Incorrect checksum when a block is received
from the service PC/interface driver. The
problem is in the service PC or HOST. Interference during transmission.
Error in the data transmission between the
service PC and the HOST. The service software version and the boot software version
do not match. Compare with the compatibility
list in the logbook.
Error in the data transmission between the
service PC and the HOST. The service software version and the boot software version
do not agree. Compare with the compatibility
list in the logbook.
Incorrect segment address from the service
PC for writing on EPROM. Error in IDX file;
install the backup copy.
vice PC and the HOST. Hardware problem or
electromagnetic compatiblity (EMC) interference.
If necessary, perform the transfer again.
service PC has been exceeded.
vice PC and the HOST.
Hardware problem or electromagnetic compatibility (EMC) interference.
If necessary, perform the transfer again.
connection closed. This message is also displayed if hardware errors are present on
board D1.
After three unsuccessful attempts to transfer,
transmission between the HOST and the service PC is terminated.
Siemens AGRXL2-120.072.03Page 5 of 28LITHOSTAR Multiline