2Service Instructions SP005/03/I
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products Affected0
LITHOSTAR Multiline
Reason for the Update0
When using the IWAKI pump, error message 5305 could occur at the LITHOSTAR
Multiline.
Prerequisites0
n.a.
Special Tools / Documents0
Wrap tool
LITHOSTAR MultilineRXL2-120.898.03.Page 2 of 6Siemens AG
Rev. 0111.03CS PS 24Medical Solutions
Service Instructions SP005/03/I3
Ordering Information0
The following update kit has to be ordered from CSML (SAP Distribution Channel, factory
2050) or BU Logistics (factory BU) (only with an order via BU Logistics):
Update Kit 11 57 929 (Material Number)
Contents of the Update Kit0
Update kit 11 57 929 contains the following parts:
Pos.Quan.Material No.Name
1130 96 062Special resistor
21Mechanical part
Tab. 1
Return of Parts0
n.a.
Siemens AGRXL2-120.898.03.Page 3 of 6LITHOSTAR Multiline
Medical SolutionsRev. 0111.03CS PS 24
4Service Instructions SP005/03/I
14
15
Fig. 1Fig. 2
Work Steps0
• Loosen the control frame of the water system and swivel it out.
• Attach the resistor to the left side of the N12 control frame, acc. to Fig. 1.
• Run the connecting cables towards the rear through the control frame to the K2 plug.
• Wrap one end of the cable at point K 2.14.
• Wrap the other end of the cable at point K 2.15.
• Swivel the N12 control frame back in and fasten it with screws.
Final Check0
• Monitor the system. Error 5305 must no longer obstruct the water system.
Customer Information0
n.a.
Final Work Steps0
• Update the system documentation.
Update the revision level, the operating instructions and the technical documentation.
Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update
Completion Form" and file it in the corresponding System Binder/User Handbook.
• Updates that have already been completed prior to publication of this SI must also be
reported.
• The update is reported as follows:
- The modification reply cards (Type 606) previously distributed with the publication of
updates no longer apply.
- The modification reply report has to be prepared by authorized personnel using an
application on the Intranet.)
LITHOSTAR MultilineRXL2-120.898.03.Page 4 of 6Siemens AG
Rev. 0111.03CS PS 24Medical Solutions
Service Instructions SP005/03/I5
Completion Protocol / Update Completion Form0
The update with the number SP005/03/I has been completed.