Siemens Lithostar Multiline Disposal Instructions

LITHOSTAR Multiline
Disposal Instructions
SP
Environmental protection measures
© Siemens AG 1995
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 3 English Print No.: RXL2-120.163.01.03.02 Doc . Gen. Date: 04.97
Replaces: RXL2-120.163.01.02.02
0 - 2 Revision
Chapter Page Revision
01 - 403 11 - 202 21 - 203 3 1 - 10 03 41 - 202
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 2 of 4 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
Contents 0 - 3
Page
1 _______General_______________________________________________________1 - 1
General inf o rmatio n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
2 _______Outline _______________________________________________________2 - 1
Disposal list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Materials/components subject to mandatory disposal . . . . . . . . . . . . . . . . .2 - 2
3 _______Disassembly and Disposal_______________________________________3 - 1
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Insulating oil in HV battery charger
(when scrapping only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Coolant in the cooling unit
(when scrapping only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Protective measures, storage and handling . . . . . . . . . . . . . . . . . . . . . 3 - 1
Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 2
Lithium battery in the MEMOSKOP . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Lithium battery in control frame N11. . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Buffer battery in POLYDOROS LX . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Battery in POLYDOROS SX. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Lithium battery in the SONOLINE . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Battery in the SIRECUST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 4
Battery in the laser light localizer . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Insulating oil (
High-voltage generator (POLYDOROS LX and SX). . . . . . . . . . . . . . . . . 3 - 5
Shock wave generator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
X-ray tube assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
Lead and components containing lead . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
X-ray tube assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 6
Grid and centering cross on the I.I. . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Grid on the cassette holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Implosive vacuum components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
X-ray tube assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Image intensifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
System PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 9
MULTISPOT 2000, optional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 9
when scrapping only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
4 _______Changes as compared to previous version _________________________4 - 1
Siemens AG Register 3 RXL2-120.163.01 Page 3 of 4 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
0 - 4 Contents
Page
This page intentionally left blank.
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 4 of 4 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering

General 1

General information 1

The regulation regarding electronic scrap presently being formulated requires that manu­facturers of systems and equipment dispose of the hazardous materials and components contained in them. Proper disposal procedures for hazardous waste apply not only when scrapping the system but also when replacing hazardous materials while the system is operational.
The list of hazardous materials b y p roduc t locat ed on pa ge 2-1 of t hese in structions gi v es an outline of the assemblies and components subject to mandatory disposal procedures. To assist you in locating t he assemb l ies/co mponents containi ng hazardous s ubst ances in the system, we have included disposal information and illustrations on page 2-2.
The disposal information contained in these instructions applies when scrapping the sys­tem or when replacing components and parts containing substances hazardous to the environment.
1 - 1
Siemens AG Register 3 RXL2-120.163.01 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 04.97 TD SP 1
1 - 2 General
This page intentionally left blank.
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 2 of 2 Siemens AG Rev. 02 04.97 TD SP 1 Medical Engineering

Outline 2

Disposal list 2

2 - 1
Component
Assembly
Location
Battery charger Shock wave cabinet A13 Oil ~ 2 ltr. Page 3-1 Cooling unit Shock wave cabinet A13
Battery
Control cabinet/ MEMOSKOP/Board D1
Harmful
substances/
Hazard
1
Coolant:
2
R134A
3
Lithium 1 Page 3-2
Quan-
tity
0.88 lb
0.4 kg
Page 3-1
Mea-
sure
Control cabinet/
Battery
Control frame
4
Lithium 1 Page 3-2
N11Board D1
Buffer battery in component J32
High-voltage genera­tor
Shock wave generat.
POLYDOROS LX Board D200
POLYDOROS LX / SX Basic unit
HV capacitor
5
Lithium 1 Page 3-3 Transformer
6
oil PCB-free Oil
7
inflammable
~15 ltr. Page 3-5
~2 ltr . Page 3-6
Lead Page 3-6
X-ray tube asse mbly C-arm, lower secti o n
Oil Page 3-6
8
Implosion of X-ray tube
1 Page 3-7
Collimator X-ray tube assembly Lead 1 Page 3-6 Grid
Centering cross
Image intensifier
Image intensifier Grid Cassette holder Lead 1 Page 3-7 Image intensifier with
Videomed
C-arm, upper section
PC monitor Control console FL double monitor Control console Impl. of tube 1 Page 3-8
Physiolog. monitor optional
MULTISPOT, optional
FL monitor System PC, keyboard
and mouse
Control console
Control console
Monitor trolley or ceiling
mounted support
Control console Battery 1 Page 3-9
SONOLINE monitor SONOLINE Battery SONOLINE Lithium 1 Page 3-4
9
10
Lead 1 Page 3-7
11
Implosion of
12
X-ray tube Implosion of
13
picture tube
14
Implosion of
15
picture tube Implosion of
16
1 Page 3-7
1 Page 3-8
1 Page 3-8
1 Page 3-8
picture tube Implosion of
17
picture tube
18
Implosion of
19
picture tube
20
1 Page 3-8
1 Page 3-8
Siemens AG Register 3 RXL2-120.163.01 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
2 - 2 Outline
Component
Location
Assembly
SIRECUST monitor SIRECUST Battery SIRECUST Lithium 1 page 3-4 Buffer batter y
POLYDOROS SX in CPU unit
Battery Laser light localizer dep. on model 4 page 3-4
21
22
23
24
Harmful
substances/
Hazard
Implosion of picture tube
Quan-
tity
1 page 3-8
Lithium 1 page 3-3

Materials/components subject to mandatory disposal 2

234
17
24
12
11
10
1
23
13
5
6
14
15
Mea-
sure
20
7
8
19
9
21
18
16
22
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 2 of 2 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
3 - 1

Preparation 3

Make sure that the unit and the on-si te power supply are switched off.
Ensure that the system and the on-site power distr ibutor cannot be switched on again.
Battery charger
Cooling unit
6,(0(16
Insulating oil in HV battery charger
(when scrapping only) 3
Remove the front cover of the shock wave cabinet.
Take out the battery charger.
Pour out or draw off the oil of th e charger into a suitable
container.
The oil removed must be disposed of by a qualified
recycling company.
Coolant in the cooling unit
(when scrapping only)

Procedure 3

Disassemble the cooling unit
Pour the coolant, fluorocarbon R134a, into a suitable
container.
The liquid removed must be disposed of by a qualified
recycling company.

Protective measures, storage and handling 3

Technical protective measures:
- If vapors or mist can accumulate, ensure s ufficient ventilation or exhaust to comply wi th specified thresholds.
- Avoid inhaling high vapor concentrat ions.
- Since mist and vapor are heavier than air and di splace the oxygen, avoid narrow or low worki ng positions. Exhaust must be provided at the lowest wor king positions.
- Keep the liquid away from heat and sources of ignition, since it can decompose into toxic by-products i f held near open flames or placed on red-hot surfaces.
3
Individual protective measur es:
- Use respiratory equipment in closed or ins ufficiently ventilated rooms.
- Use safety glasses for eye protection .
- Use gloves made of chlorbutyl rubber to protec t hands.
- Use lightweight protective clothes and skin protection.
Work hygiene
- Do not inhale the vapors.
- Do not allow the liquid to contact ey es or hands.
Fig. 1 Shockwave cabinet
Siemens AG Register 3 RXL2-120.163.01 Page 1 of 10 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
3 - 2 Disassembly and Disposal
'
MEMOSKOP
/LWKLXPEDWWHULH
0HPRVNRS
9
*
-
9 9 9
Lithium battery
;
Control frame N11
7
1HW]EDX JUXSSH 0
D1
/LWKLXPEDWWHULH
Lithium battery
;
'

Batteries 3

Remove the front panel of the cont rol cabinet (Fig. 2).
Remove the front panel of the genera tor cabinet (Fig. 3)
Lithium battery in the MEMOSKOP 3
Remove the front cover of the MEMOSKOP in the
control cabinet (Fig. 2) by removing the screws.
Remove board D1 from the MEMOSKOP and place it on
a flat surface.
Remove the lithium battery from position J 144 on board
D1 and have a qualified recycling company dispose of it.
Lithium battery in control frame N11 3
Remove board D1 from control frame N11 in the control
cabinet and place it on a flat surface (Fig. 2).
Remove the lithium battery from position G4 on boar d
D1 and have a qualified recycling company dispose of it.
Fig. 2
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 2 of 10 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
Disassembly and Disposal 3 - 3
X14
X1
XCS
X3
X4
Buffer battery
V1 V2
X5
X13
Fig. 3 POLYDOROS LX
3XIIHUEDWWHULH
HQWKDOWHQ
SIEMENS
11 72 720 X2169 D200 E2
DATA
MODEM
'
D200
D100
D110
D115
D200
J32 J11 J28
X5
X2
D291
D160
X9
X5
T2
H1
K20
F1F2F3
D220
K30
M16
Buffer battery

Buffer battery in POLYDOROS LX 3

Remove board D200 from the generator
cabinet and place it on a flat surf ace (Fig.
3).
Remove component J32 containing the
buffer battery from board D200. (Fig. 3).
Have the component containing the buffer
battery disposed of by a qualif ied recycling company.

Battery in POLYDOROS SX 3

Remove the battery from the CPU HD unit
and have a qualified recycling company dispose of it (Fig.4).
Fig. 4 POLYDOROS LX
Siemens AG Register 3 RXL2-120.163.01 Page 3 of 10 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
3 - 4 Disassembly and Disposal
Lithium battery in the SONOLINE 3
Remove the side panel from the
SONOLINE:
Remove the lithium battery from t he
SONOLINE and have it disposed of by a qualified recycling company.
Battery
Fig. 5 SONOLINE

Battery in the SIRECUST 3

Remove the battery from the back of
the SIRECUST and have it disposed of by a qualified recycling company.
Fig. 6 SIRECUST
Battery

Battery in the laser light localizer 3

Remove the battery from the laser light l ocalizer and have it disposed of by a qualif ied
recycling company .
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 4 of 10 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
Disassembly and Disposal 3 - 5
Cover with installed parts
24 pan head screws distributed on the periphery
Trough filled with transformer oil
Fig. 7 High-voltage generator
Insulating oil (when scrapping only) 3

High-voltage generator (POLYDOROS LX and SX) 3

Remove the high-voltage generator H1fr om the generator cabinet (Fig. 3)
Loosen the 24 pan head screws on the high-voltage generator (Fig.7).
Lift the cover with the install ed parts from the trough of the high-voltage generator and
allow the oil to run into the trough .
Pour the oil from the trough into a su itable container and have it disposed of by a
qualified recycling co mpany.
Siemens AG Register 3 RXL2-120.163.01 Page 5 of 10 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
3 - 6 Disassembly and Disposal

Collimator

X-ray tube assem.

Shock wave generator

Fig. 8
Shock wave generator 3
Remove the panels of the basic unit.
Remove the shock wave generator fro m the basic unit (Fig. 8, left).
Open the shock wave generator by removing the screws and take out the high-voltage
capacitor filled with oil.
Have the entire high-voltage capaci tor disposed of by a qualified recycli ng company or
send it back as returned goods to SIEMENS AG, Medical Enginee ring Group, for reprocessing.

X-ray tube assembly 3

Remove the covers of the tube assembly by remo ving the screws.
Remove the collimator fr om the tube assembly (Fig.8, right) by removin g the screws.
Remove the tube assembly filled with oi l from the C-arm and send it back as returned
goods to SIEMENS AG, Medical Engineering Group for reprocessing.

Lead and components containing lead 3

X-ray tube assembly 3

Due to the lead content of the screens, the tube asse mbly must be sent back as returned
goods to SIEMENS AG, Medical Engineering Group for reprocessing.
Collimator 3
Due to the lead content of the blades, the col limator previously removed from the t ube
assembly must be sent back as returned goods to SIEMENS AG, Medical Engineeri ng Group for reprocessing.
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 6 of 10 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
Disassembly and Disposal 3 - 7
Cover

Image Intensifier

Grid Centering cross
Fig. 9

Grid and centering cross on the I.I. 3

Remove the grid and centering cross atta ched to the I.I. input (Fig. 9).
Have the grid and the centering cross di sposed of by a qualified recyc ling company
(Fig. 9).

Grid on the cassette holder 3

Remove the grid from the cassette hold er and have it disposed of by a qual ified recycling
company.

Implosive vacuum components 3

X-ray tube assembly 3

Due to the vacuum X-ray tube, the tube assembly must be sent back as previously
described to SIEMENS AG, Medical Engineering Group for reprocessing.
Image intensifier 3
Remove the side panels of the I.I. by r emoving the screws (Fig. 9, left)
Remove the I.I. from the C-arm (Fig. 9, ri ght)
and send it back as returned goods to SIEMENS AG, Medical Engineering Group for reprocessing.
Siemens AG Register 3 RXL2-120.163.01 Page 7 of 10 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
3 - 8 Disassembly and Disposal
FL Monitor
SIRECUST
Monitor
PC Monitor
Keyboard
FL double monitor
Mouse
Physiolog. monitor
System PC
Multispot 2000
(if present)
Fig. 10

Monitors 3

Send the PC monitor to SIEMENS AG, Medical
Engineering or to a q ualified recycli ng company.
Send the FL double monitor as returned goods to
SIEMENS AG, Medical Engineering or to a qualified recycling company.
Send the FL monitor to SIEMENS AG, Medical
Engineering, or to a qualified recycling company.
Send the physiological monitor, i f present, to the
Lucius & Baer company or to a qualified recyc ling company.
Have the entire SIRECUST system, including the built-
in monitor, disposed of by a quali fied recycling company or send it as returned goods to SIEMENS AG, Medical Engineering Group.
Send the SONOLINE monitor as returned goods to
SIEMENS AG, Medical Engineering Group, or have it disposed of by a qualified recyc ling company.
Monitor
Fig. 11
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 8 of 10 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering
Disassembly and Disposal 3 - 9

System PC 3

Turn over the sy stem PC (Fig. 10) together with the keyboard and the mouse to
SIEMENS AG, Medical Engineering Group or to a qualified recycling company.

MULTISPOT 2000, optional 3

Send the MULTISPOT 2000 (Fig. 10), if present, to the Hertel company, Electronics
Recycling, Nuremberg, or to SIEMENS AG, Medical Engineering Group.
Siemens AG Register 3 RXL2-120.163.01 Page 9 of 10 LITHOSTAR Multiline Medical Engineering Rev. 03 04.97 TD SP 1
3 - 10 Disassembly and Disposal
This page intentionally left blank.
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 10 of 10 Siemens AG Rev. 03 04.97 TD SP 1 Medical Engineering

Changes as compared to previous version 4

Chapter 0 Cover page ad apted; page containi ng re vi sion le vel and table of contents
expanded Chapter 1 Layout changed Chapter 2 Table and figure changed Chapter 3 Fig. 4 new; text changes and additions Chapter 4 Changed in accordance with the new revision level
4 - 1
TD SP 1 / Schlee TD SP 2 / Wendt SMS Iselin / Weiss Costa
Siemens AG Register 3 RXL2-120.163.01 Page 1 of 2 LITHOSTAR Multiline Medical Engineering Rev. 02 04.97 TD SP 1
4 - 2 Changes as compared to previous version
This page intentionally left blank.
LITHOSTAR Multiline Register 3 RXL2-120.163.01 Page 2 of 2 Siemens AG Rev. 02 04.97 TD SP 1 Medical Engineering
Loading...