The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
Register 3English
Print No.: RXL2-120.163.01.03.02 Doc . Gen. Date:04.97
Replaces: RXL2-120.163.01.02.02
0 - 2Revision
ChapterPageRevision
01 - 403
11 - 202
21 - 203
31 - 1003
41 - 202
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 2 of 4Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
4 _______Changes as compared to previous version _________________________4 - 1
Siemens AGRegister 3RXL2-120.163.01Page 3 of 4LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
0 - 4Contents
Page
This page intentionally left blank.
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 4 of 4Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
General1
General information1
The regulation regarding electronic scrap presently being formulated requires that manufacturers of systems and equipment dispose of the hazardous materials and components
contained in them. Proper disposal procedures for hazardous waste apply not only when
scrapping the system but also when replacing hazardous materials while the system is
operational.
The list of hazardous materials b y p roduc t locat ed on pa ge 2-1 of t hese in structions gi v es
an outline of the assemblies and components subject to mandatory disposal procedures.
To assist you in locating t he assemb l ies/co mponents containi ng hazardous s ubst ances in
the system, we have included disposal information and illustrations on page 2-2.
The disposal information contained in these instructions applies when scrapping the system or when replacing components and parts containing substances hazardous to the
environment.
1 - 1
Siemens AGRegister 3RXL2-120.163.01Page 1 of 2LITHOSTAR Multiline
Medical EngineeringRev. 0204.97TD SP 1
1 - 2General
This page intentionally left blank.
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 2 of 2Siemens AG
Rev. 0204.97TD SP 1Medical Engineering
Siemens AGRegister 3RXL2-120.163.01Page 1 of 2LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
2 - 2Outline
Component
Location
Assembly
SIRECUST monitorSIRECUST
Battery SIRECUSTLithium1page 3-4
Buffer batter y
POLYDOROS SX in
CPU unit
BatteryLaser light localizerdep. on model4page 3-4
21
22
23
24
Harmful
substances/
Hazard
Implosion of
picture tube
Quan-
tity
1page 3-8
Lithium1page 3-3
Materials/components subject to mandatory disposal2
234
17
24
12
11
10
1
23
13
5
6
14
15
Mea-
sure
20
7
8
19
9
21
18
16
22
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 2 of 2Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
3 - 1
Preparation3
• Make sure that the unit and the on-si te power supply are switched off.
• Ensure that the system and the on-site power distr ibutor cannot be switched on again.
Battery
charger
Cooling
unit
6,(0(16
Insulating oil in HV battery charger
(when scrapping only)3
• Remove the front cover of the shock wave cabinet.
• Take out the battery charger.
• Pour out or draw off the oil of th e charger into a suitable
container.
• The oil removed must be disposed of by a qualified
recycling company.
Coolant in the cooling unit
(when scrapping only)
Procedure3
• Disassemble the cooling unit
• Pour the coolant, fluorocarbon R134a, into a suitable
container.
• The liquid removed must be disposed of by a qualified
recycling company.
Protective measures, storage and handling3
• Technical protective measures:
- If vapors or mist can accumulate, ensure s ufficient
ventilation or exhaust to comply wi th specified
thresholds.
- Avoid inhaling high vapor concentrat ions.
- Since mist and vapor are heavier than air and di splace
the oxygen, avoid narrow or low worki ng positions.
Exhaust must be provided at the lowest wor king
positions.
- Keep the liquid away from heat and sources of ignition,
since it can decompose into toxic by-products i f held
near open flames or placed on red-hot surfaces.
3
• Individual protective measur es:
- Use respiratory equipment in closed or ins ufficiently
ventilated rooms.
- Use safety glasses for eye protection .
- Use gloves made of chlorbutyl rubber to protec t hands.
- Use lightweight protective clothes and skin protection.
• Work hygiene
- Do not inhale the vapors.
- Do not allow the liquid to contact ey es or hands.
Fig. 1 Shockwave cabinet
Siemens AGRegister 3RXL2-120.163.01Page 1 of 10LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
3 - 2Disassembly and Disposal
'
MEMOSKOP
/LWKLXPEDWWHULH
0HPRVNRS
9
*
-
999
Lithium battery
;
Control frame N11
7
1HW]EDXJUXSSH0
D1
/LWKLXPEDWWHULH
Lithium battery
;
'
Batteries3
• Remove the front panel of the cont rol cabinet (Fig. 2).
• Remove the front panel of the genera tor cabinet (Fig. 3)
Lithium battery in the MEMOSKOP3
• Remove the front cover of the MEMOSKOP in the
control cabinet (Fig. 2) by removing the screws.
• Remove board D1 from the MEMOSKOP and place it on
a flat surface.
• Remove the lithium battery from position J 144 on board
D1 and have a qualified recycling company dispose of it.
Lithium battery in control frame N113
• Remove board D1 from control frame N11 in the control
cabinet and place it on a flat surface (Fig. 2).
• Remove the lithium battery from position G4 on boar d
D1 and have a qualified recycling company dispose of it.
Fig. 2
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 2 of 10Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
Disassembly and Disposal3 - 3
X14
X1
XCS
X3
X4
Buffer battery
V1 V2
X5
X13
Fig. 3 POLYDOROS LX
3XIIHUEDWWHULH
HQWKDOWHQ
SIEMENS
11 72 720 X2169 D200 E2
DATA
MODEM
'
D200
D100
D110
D115
D200
J32 J11 J28
X5
X2
D291
D160
X9
X5
T2
H1
K20
F1F2F3
D220
K30
M16
Buffer battery
Buffer battery in POLYDOROS LX3
• Remove board D200 from the generator
cabinet and place it on a flat surf ace (Fig.
3).
• Remove component J32 containing the
buffer battery from board D200. (Fig. 3).
• Have the component containing the buffer
battery disposed of by a qualif ied recycling
company.
Battery in POLYDOROS SX3
• Remove the battery from the CPU HD unit
and have a qualified recycling company
dispose of it (Fig.4).
Fig. 4 POLYDOROS LX
Siemens AGRegister 3RXL2-120.163.01Page 3 of 10LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
3 - 4Disassembly and Disposal
Lithium battery in the SONOLINE3
• Remove the side panel from the
SONOLINE:
• Remove the lithium battery from t he
SONOLINE and have it disposed of
by a qualified recycling company.
Battery
Fig. 5 SONOLINE
Battery in the SIRECUST3
• Remove the battery from the back of
the SIRECUST and have it disposed
of by a qualified recycling company.
Fig. 6 SIRECUST
Battery
Battery in the laser light localizer3
• Remove the battery from the laser light l ocalizer and have it disposed of by a qualif ied
recycling company .
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 4 of 10Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
Disassembly and Disposal3 - 5
Cover with installed parts
24 pan head screws distributed
on the periphery
Trough filled with transformer oil
Fig. 7 High-voltage generator
Insulating oil (when scrapping only)3
High-voltage generator (POLYDOROS LX and SX)3
• Remove the high-voltage generator H1fr om the generator cabinet (Fig. 3)
• Loosen the 24 pan head screws on the high-voltage generator (Fig.7).
• Lift the cover with the install ed parts from the trough of the high-voltage generator and
allow the oil to run into the trough .
• Pour the oil from the trough into a su itable container and have it disposed of by a
qualified recycling co mpany.
Siemens AGRegister 3RXL2-120.163.01Page 5 of 10LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
3 - 6Disassembly and Disposal
Collimator
X-ray tube assem.
Shock wave generator
Fig. 8
Shock wave generator3
• Remove the panels of the basic unit.
• Remove the shock wave generator fro m the basic unit (Fig. 8, left).
• Open the shock wave generator by removing the screws and take out the high-voltage
capacitor filled with oil.
• Have the entire high-voltage capaci tor disposed of by a qualified recycli ng company or
send it back as returned goods to SIEMENS AG, Medical Enginee ring Group, for
reprocessing.
X-ray tube assembly3
• Remove the covers of the tube assembly by remo ving the screws.
• Remove the collimator fr om the tube assembly (Fig.8, right) by removin g the screws.
• Remove the tube assembly filled with oi l from the C-arm and send it back as returned
goods to SIEMENS AG, Medical Engineering Group for reprocessing.
Lead and components containing lead3
X-ray tube assembly3
• Due to the lead content of the screens, the tube asse mbly must be sent back as returned
goods to SIEMENS AG, Medical Engineering Group for reprocessing.
Collimator3
• Due to the lead content of the blades, the col limator previously removed from the t ube
assembly must be sent back as returned goods to SIEMENS AG, Medical Engineeri ng
Group for reprocessing.
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 6 of 10Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
Disassembly and Disposal3 - 7
Cover
Image Intensifier
GridCentering cross
Fig. 9
Grid and centering cross on the I.I.3
• Remove the grid and centering cross atta ched to the I.I. input (Fig. 9).
• Have the grid and the centering cross di sposed of by a qualified recyc ling company
(Fig. 9).
Grid on the cassette holder3
• Remove the grid from the cassette hold er and have it disposed of by a qual ified recycling
company.
Implosive vacuum components3
X-ray tube assembly3
• Due to the vacuum X-ray tube, the tube assembly must be sent back as previously
described to SIEMENS AG, Medical Engineering Group for reprocessing.
Image intensifier3
• Remove the side panels of the I.I. by r emoving the screws (Fig. 9, left)
• Remove the I.I. from the C-arm (Fig. 9, ri ght)
and send it back as returned goods to SIEMENS AG, Medical Engineering Group for
reprocessing.
Siemens AGRegister 3RXL2-120.163.01Page 7 of 10LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
3 - 8Disassembly and Disposal
FL Monitor
SIRECUST
Monitor
PC Monitor
Keyboard
FL double monitor
Mouse
Physiolog. monitor
System PC
Multispot 2000
(if
present)
Fig. 10
Monitors3
• Send the PC monitor to SIEMENS AG, Medical
Engineering or to a q ualified recycli ng company.
• Send the FL double monitor as returned goods to
SIEMENS AG, Medical Engineering or to a qualified
recycling company.
• Send the FL monitor to SIEMENS AG, Medical
Engineering, or to a qualified recycling company.
• Send the physiological monitor, i f present, to the
Lucius & Baer company or to a qualified recyc ling
company.
• Have the entire SIRECUST system, including the built-
in monitor, disposed of by a quali fied recycling
company or send it as returned goods to SIEMENS
AG, Medical Engineering Group.
• Send the SONOLINE monitor as returned goods to
SIEMENS AG, Medical Engineering Group, or have it
disposed of by a qualified recyc ling company.
Monitor
Fig. 11
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 8 of 10Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
Disassembly and Disposal3 - 9
System PC3
• Turn over the sy stem PC (Fig. 10) together with the keyboard and the mouse to
SIEMENS AG, Medical Engineering Group or to a qualified recycling company.
MULTISPOT 2000, optional3
• Send the MULTISPOT 2000 (Fig. 10), if present, to the Hertel company, Electronics
Recycling, Nuremberg, or to SIEMENS AG, Medical Engineering Group.
Siemens AGRegister 3RXL2-120.163.01Page 9 of 10LITHOSTAR Multiline
Medical EngineeringRev. 0304.97TD SP 1
3 - 10Disassembly and Disposal
This page intentionally left blank.
LITHOSTAR MultilineRegister 3RXL2-120.163.01Page 10 of 10Siemens AG
Rev. 0304.97TD SP 1Medical Engineering
Changes as compared to previous version4
Chapter 0Cover page ad apted; page containi ng re vi sion le vel and table of contents
expanded
Chapter 1Layout changed
Chapter 2Table and figure changed
Chapter 3Fig. 4 new; text changes and additions
Chapter 4Changed in accordance with the new revision level