The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: RXL2-120.021.01.12.02 Doc. Gen. Date:10.04
Replaces: RXL2-120.021.01.11.02
0 - 2Revision
ChapterPageRevision
allall01
allall02
allall03
allall04
allall05
allall06
allall07
allall08
allall09
allall10
allall11
allall12
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Transport Dimensions, Basic System with Transport Frame . . . . . . . . . . . . . . 6 - 1
Siemens AGRXL2-120.021.01Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
0 - 4Contents
Page
7 ______ Changes to Previous Version _____________________________________ 7 - 1
LITHOSTAR MultilineRXL2-120.021.01Page 4 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
General Notes1
1 - 1
General notes 1
- With distribution of these revision level, all preceding planning guides, Speed - Infos (PG’s) and
drafts lose their validity.
- All layouts issued by the Planning Departments must bear a note referring to the installation and
delivery conditions of Siemens Medical Engineering. The installation and delivery conditions must
be submitted with the layouts.
- Unless otherwise specified, all dimensions are indicated in "mm".
- The symbol indicates a change (see revision status).
- Orientation points
Points specific to system components to which reference is made when positioning system
components to each other or in the room.
The isocenter of a radiographic system is always illustrated as the orientation point.
-Fixpoints
Clearly marked points on system components, installation ceiling, walls or floor on which cable
outlets are located.
Illustration in the drawings: octagon with letter/number-combination.
The cable lengths specify the maximum fixpoint distances and thus the maximum distances
between the individual system components.
- Room height
The room height is the distance measured from the top surface of the floor to the bottom surface of
the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
- Room lighting
According to DIN 68 68-57 (international standard in preparation), the lighting in rooms in which
image playback devices (monitors) are used for diagnosis, the following requirements must be
met:
adjustable, no anti-glare screen, reproducible adjustment of the lighting (e. g. dimmer with
scala),
no glare or reflection from windows, lights and light boxes in the standard working position of
the monitors.
Hotline+ 49 (9191) 18 - 8080
Siemens AGRXL2-120.021.01Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
1 - 2General Notes
Safety1
- The provisions of the relevant fire safety regulations must be observed for the premises.
- The system has been developed according to EN 60601 - 1.
- Minimum dimensions (e. g. room heights, safety distances) indicated in the planning guides are
marked "min."
- Basic strength against electromagnetic sources of interference.
Result of lightning discharges.
The protection targets of the different lightning protection areas up to the unit connection are also
specified in the IEC 1024, DIN 48810, VDE 0675 and in the DEMVT recommendations.
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
General Notes1 - 3
System Configuration1
(Basic layout), LITHOSTAR Multiline with POLYDOROS LX1
The following options are not shown in the basic layout:Respiratory ECG triggering
Ultrasound unit
Monitor cart
Diamentor
DICOM
Siemens AGRXL2-120.021.01Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
1 - 4General Notes
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.021.01Page 4 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Room Planning2
2 - 1
Wall Distances and Room Height, Version I2
LITHOSTAR Multiline with POLYDOROS LX
*1
*5
*2
*1
*4
Orientation point
*3
*6
*5
*1
*7
Radiation protection
shield, on-site
1 : 50
*1Safety distance according to the equipment safety law
*2System excursion
*3Only in countries with UL regulations apply
*4Option: Ultrasound unit
*5Option: 3rd monitor with monitor trolley
Cable feed max. 5 cm above floor.
Cable outlet, item no. 14 76 241 R 4189 also required
*6Option: Diamentor
*7Option: DICOM
Room size
AreaLengthWidthHeight
Technically possible minimum
examination room size
Siemens AGRXL2-120.021.01Page 1 of 10LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
22.70 m
2
5.80 m3.90 mmin. 2.40 m
2 - 2Room Planning
Wall Distances and Room Height, Version II2
LITHOSTAR Multiline with POLYDOROS SX
*1
*3
*2
*3
*5
*1
Orientation point
*4
*6
*7
Radiation protection
shield, on-site
1 : 50
*1Safety distance according to the equipment safety law
*2System excursion
*33rd monitor with monitor trolley
Cable feed max. 5 cm above floor.
Cable outlet, item no. 14 76 241 R 4189 also required
*4Only in countries with UL regulations apply
*5Option: Ultrasound unit
*6Option: Diamentor
*7Option: DICOM
Room size
AreaLengthWidthHeight
Technically possible minimum
examination room size
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 10Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
22.70 m
2
5.80 m3.90 mmin. 2.40 m
Room Planning2 - 3
Dimensions: LITHOSTAR Multiline 2
Side view
Drawing shows the installation
plate embedded in the floor
Top vi e w
*1System travel range
*2Tabletop extension
right or left
Option:
preferred position:
water connection kit
*1
Orientation point
*1
*1
*2
Siemens AGRXL2-120.021.01Page 3 of 10LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
Orientation
point
*2
1 : 25
2 - 4Room Planning
Dimensions: UMR Cabinets with POLYDOROS LX 2
POLYDOROS LX
Shock wave
cabinet
Control cabinet
1 : 20
NOTICE
LITHOSTAR MultilineRXL2-120.021.01Page 4 of 10Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
For improved visual effect, the front panels of the cabinets can be aligned
with one another.
Room Planning2 - 5
Dimensions: UMR Cabinets with POLYDOROS SX 2
POLYDOROS SX
Shock wave
cabinet
Control cabinet
1 : 20
Siemens AGRXL2-120.021.01Page 5 of 10LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
2 - 6Room Planning
Dimensions: System Operating Console 2
*2
System
ON/OFF
E. STOP
Connection for footswitch (optional) 60 mm above the
floor (note adjustment range) and 500 mm from the front
Alternate
cable outlet
Operating keyboard for
option: DICOM
Door, system control console
Cable outlet ∅ 100 mm
Push-button pad
permanently installed
Operating
keyboard
1 : 20
Option: MULTISPOT
*1
LITHOSTAR MultilineRXL2-120.021.01Page 6 of 10Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
*1Face plate is removed when installing the
MULTISPOT
Option: DICOM
*2Display only at option DICOM available
Room Planning2 - 7
Dimensions: Monitor Trolley, Option2
Dimensions of monitor trolley with 44 cm monitor (for Uro Plus equipment, included in delivery)
Swivel range
+ 15
o
... - 10
o
Radiation
indicator
Radiation indicator
Cable holder
both front
castors
with brakes
both rear castors
without brakes
Dimensions: Respiratory ECG Gating, Option 2
Patients monitor
A location for installation of the patients monitor must
be provided on-site.
Refer also to the accessory catalog for patient
monitoring
Siemens AGRXL2-120.021.01Page 7 of 10LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
2 - 8Room Planning
Dimensions: Ultrasound System with Monitor, Option2
1 : 20
The ultrasound option can be disconnected from device B1. A support bracket must be provided onsite for the excess corrugated tubing (6m).
Laser Camera Connection, Option2
The optical video separation device between the laser camera and the Memoskop is installed in the
control cabinet. The cable from the optical video separation device to the laser camera must not
exceed a maximum length of 24 m.
LITHOSTAR MultilineRXL2-120.021.01Page 8 of 10Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Room Planning2 - 9
Static Load Prerequisites 2
Orientation point
Cable outlet
Free area for levelling concrete
1 : 20
The pull forces that occur at the floor mounting points A are max. 1688 N per mounting point.
Total weight:
System
Floor plate
Patient
approx. 660 kg
approx. 130 kg
approx. 150 kg
∑approx. 940 kg
• = Alternate bore
Siemens AGRXL2-120.021.01Page 9 of 10LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
2 - 10Room Planning
Suggested Construction for Radiation Protection Shield2
NOTICE
Variable side beams
depending on ceiling
height
Type Kyowa H 35 = 1.5 Pb at 110 kV can be used for the radiation protection
shield.
1 : 25
LITHOSTAR MultilineRXL2-120.021.01Page 10 of 10Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Preparation for Installation3
3 - 1
Installation Information3
The installation plate must be ordered in advance under item no. 31 27 052 J 1042.
Liebig expansion bolts/anchors, type LSP I 14/80 item no. 70 54 828 F 1107 and corresponding installation material are included in the shipment for the installation. Only the Liebig expansion bolts packed
with the unit may be used.
The system must be installed on a solid sub-floor with sufficient load-bearing capacity, such as concrete. If the existing self-leveling grout has sufficient load bearing capacity, the system can be installed
directly on the self-leveling grout.
Other types of floors and floor coverings without sufficient load bearing capacity should be removed
prior to the installation.
If the system is not being delivered by a third party (contractor, outside of Europe/overseas),
the transport roller set (identical to POLYSTAR) item no. 11 53 654 G 5338 must be ordered
in advance.
The UMR cabinets must be installed at the same level as the system.
After the installation, the space between the system and the floor must be sealed.
(Refer to installation instructions RXL2-120.031.01...)
NOTICE
NOTICE
NOTICE
The UMR cabinets should be planned left to right.
Generator - Shock wave cabinet - Control cabinet or
Control cabinet - Shock wave cabinet - Generator
The installation plate must be levelled.
Maximum deviation over the longest dimension =1 mm/m
Recommendations from Liebig and the Installation instructions enclosed
for the dowels RXL2-120.038.01 must be strictly observed.
With 440/480 V, connect the patients monitor monitor before the building
system transformer.
NOTICE
Option:
1 Water connection kitItem no. 16 12 972 G 5354
Siemens AGRXL2-120.021.01Page 1 of 6LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
For information on installing LITHOSTAR Multiline in mobile environments,
refer to PG RXL2-120.021.02.01...
3 - 2Preparation for Installation
Dimensions: Installation Plate 3
*1
*3
*1
*3
*2
*3
*3
*3
*1
Orientation point
*1
*2
*1
*2
t = 12
*1
*1
*1
*2
1 : 10
*1Alternate bore*2System base mounting M10*3System mounting M8
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 6Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Preparation for Installation3 - 3
Installation Plate3
Solid concrete
Insulation plates 2.0 mm
Installation
plate
System
Concrete
Direct installation on self-leveling grout
Screed
Liebig expansion bolt*1
System
Levelling concrete
Insulating plates 2.0 mm
Concrete
Installation
plate
Self-leveling grout
*1The instructions from the Liebig company must be observed.
Anchoring in the levelling concrete is not permitted.
Comply with the Installation Instructions RXL2-120.038.01.
Siemens AGRXL2-120.021.01Page 3 of 6LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
3 - 4Preparation for Installation
Water Installation 3
Cover plate
Sealing with
sanitary silicone
Screed
Use caution due to
static forces when
cutting out self-leveling grout and
concrete
Shutoff valve
Flush with finished
floor
*1
*1On-site water connection 1/2" external thread
System-front
Floor recess
(flush with top edge of finished floor)
Water installation
materials:
16 12 972 G 5354
(complete material
from on-site pipe)
1 : 2.5
on-site drain 1" external thread
NOTICE
LITHOSTAR MultilineRXL2-120.021.01Page 4 of 6Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Prevent introduction of foreign substances into the drinking water supply.
Observe all national and local laws and regulations.
Preparation for Installation3 - 5
On-site Electrical Installation3
Proposal for on-site power distribution per DIN VDE 0100-710 or national regulations
Power supply line
must be provided on-site
*1
*1The fuse protection must
be selected to match the
generator used
Generator cabinet
SIEMENS
must be provided on-site
On/Off button with pilot lamp
Emergency shutdown button with lock
Generator cabinet
Power connection box
Patients monitor (optional)
Ultraschallgerät (optional)
1 To extraneous conductive parts
2 Per DIN EN 50178 Classification VDE 0160, "Equipping of high voltage systems with electronic
operating elements", the following FI switches must be used exclusively:
For U(N) = 3 ~ 400/415 V: I(N) = 63A, IdN = 30 mA for AC and pulsed as well as smooth DC currents.
Order from wholesale electrical supply company, Order No.FI 5SZ3 466 0KG05 all-current sensitive
Order from SPH2, Part No. 49 54 470 Y7933
(Width of the FI switch is 144 mm = 8TE, install on standard rail)
For countries in which this standard does not apply, the following ground fault interrupters can be
used for voltages of
> 415 V: With U(N) > 3 ~ 415 V; I(N) = 125 A, IdN = 30 mA
Doepke-Norden (order from SPH2, Part No. 51 41 168 Y7933)
2a Ground fault current interrupter (GF I) IDN 30 mA
3 Unit breaker
4 Option: Radiation display (24 V/5 W)
Siemens AGRXL2-120.021.01Page 5 of 6LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
3 - 6Preparation for Installation
5 Applicable only in countries with line voltage of 115 V, ±10%, 50/60 Hz, ±1 Hz:
If the ultrasound option is configured, at least one power outlet with the specifications 115 V, ± 10%,
50/60 Hz, ±1 Hz must be installed on site, in the vicinity of the ultrasound unit. Depending on how
the room is set up, additional power outlets must be provided for the various working positions.
[ ] Numbers are free lead ends in m
LITHOSTAR MultilineRXL2-120.021.01Page 6 of 6Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
System Connections4
4 - 1
Overview of Connections 4
Option:
3rd Monitor
*3
System cabinet (Control
cabinet) I.I./TV
System
Option:
Ultrasound
unit
*2
*2
*1
*1
*1
*1
Generator and
System control panel
Option: Diamentor
Option: DICOM
Patients monitor
Shock wave cabinet
POLYDOROS SX
POLYDOROS LX
*1
Cable duct between the system and cabinets min. W 200 x H 50 mm.
Cable duct between the system control panel and cabinets min. W 50 x H 50 mm.
Tap holes of
*1Zipper hose
If a conduit is used in place of a cable channel, a conduit with a dia. of at least 100 mm must be
provided for each zipper hose.
*2The fiber optic cable and video cable are included in the Ultrasound option.
*3Cable lengths = 18 m
NOTICE
∅ 80 mm must be provided for zipper hoses for ceiling feed-through.
∅ 50 mm
System cabling
The cable length between the system and each cabinet is 9 m.
The cable length from the control console to the control cabinet is 9 m.
The cable lengths indicated are fixed lengths that cannot be changed. If the
cable length required is shorter, storage must be provided for the excess
cable length. (Excess cable lengths may not be stored in the cabinets due
to EMC).
Siemens AGRXL2-120.021.01Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
4 - 2System Connections
Options:
The cable length for patients monitor is 18 m.
The ultrasound unit can be disconnected from the system at B1. A support bracket must be
provided on-site for the remaining corrugated tubing (6m).
The measuring cable for the Diamentor is pulled in from the unit to the system cabinet. If the Diamentor
option is configured, lay the measuring cable further to the control console.
Die Option DICOM wird nur innerhalb der Bodenplatte installiert. Es sind keine zusätzlichen Kabel zu
verlegen.
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 2Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Technical Data5
5 - 1
Electrical Data 5
Power consumption
Power connection
POLYDOROS LX 50
3/N/PE
50/60 Hz
Power connection
POLYDOROS SX 65/80
3/N/PE
50/60 Hz
*1If this power supply is not available, an on-site system transformer is required.
∼ 400 V ± 10%
± 1 Hz *1
∼ 400 V ± 10%
± 1 Hz *1
NOTICE
Max. Ri at the output terminals
(including the line Ri).
(observe UL regulations)
Rating = line voltage x internal fuse x
Ri max. = 170 m Ω POLYDOROS LX, Ri max. = 170 m Ω POLYDOROS SX
Fluoroscopy
and lithotripsy
2.5 kVA94 kVA
2.1/3.3 kVA120/145 kVA
≤ 170 m Ω POLYDOROS LX, ≤ 170 m Ω POLYDOROS SX
Transient power con-
sumption during
radiography
√3 [kVA]
Internal fuse
35 A
slow-blow
50 A
slow-blow
Weights and Heat Dissipation 5
Weight [kg]Heat dissipation [W]
LITHOSTAR Multiline
Systemapprox. 660approx. 250
Control cabinetapprox. 160approx. 800
Shock wave cabinetapprox. 240approx. 600
System control consoleapprox. 110approx. 150
Generator cabinet LXapprox. 230approx. 300
Generator cabinet SXapprox. 290approx. 900
Monitor and monitor trolleyapprox. 100approx. 90
Ultrasound systemapprox. 85approx. 450
Patients monitorapprox. 5approx. 55
Physiological monitorapprox. 11approx. 50
Siemens AGRXL2-120.021.01Page 1 of 4LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
5 - 2Technical Data
Environmental Conditions 5
LITHOSTAR Multiline
without Ultrasound option
Permissible ambient temperature+ 10° C ... + 35° C- 20° C ... + 35° C
970 kg with transport frame and palette
840 kg with transport frame, without packaging
minimum door width and door height
for transport
min.
approx.
min.
1100 mm door width
1800 mm corridor width
2000 mm door height
Surface Colors 5
Main colorMedical White C610
Combination colorAnthracite C614
Additional Data 5
POLYDOROS LX 50refer to PG RX63-020.021.01...
POLYDOROS SX 65/80refer to PG RX63-065.021.01...
Monitor trolleys (optional)refer to PG RX53-020.021.01...
Installation in mobile environmentsrefer to PG RXL2-120.021.02...
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Technical Data5 - 3
Noise Level5
Noise generation during shock wave release at 1 m distance from the patient’s head
Noise level [db (A)]
Energy level 4
Energy level 7
≤ 70.4
≤ 73.3
Siemens AGRXL2-120.021.01Page 3 of 4LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
5 - 4Technical Data
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.021.01Page 4 of 4Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Transportation Specifications6
6 - 1
Transport Dimensions, Basic System with Transport Frame 6
*1
*1
*1Required for removal of the transport pallet
NOTICE
Siemens AGRXL2-120.021.01Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
Minimum door width 1100 mm, minimum corridor width approx. 1800 mm,
minimum door height 2000 mm.
Transport height without transport rollers: 23 cm I.I.1860 mm
Transport height without transport rollers: 33 cm I.I. 1989 mm
6 - 2Transportation Specifications
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 2Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Changes to Previous Version7
ChapterPageChange
0 - 7Rev. level of document changed from 11 to 12.
22-7SIRECUST by patients monitor replaces.
33-1SIRECUST by patients monitor replaces.
33-5Text updated.
44-1 and 4-2SIRECUST by patients monitor replaces.
55-1SIRECUST by patients monitor replaces.
55-2Surface colors updated.
77-1Changes to previous version updated.
7 - 1
Siemens AGRXL2-120.021.01Page 1 of 2LITHOSTAR Multiline
Medical SolutionsRev. 1210.04CS SD 21
7 - 2Changes to Previous Version
This page intentionally left blank.
LITHOSTAR MultilineRXL2-120.021.01Page 2 of 2Siemens AG
Rev. 1210.04CS SD 21 Medical Solutions
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.