The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.05
1995
Page 2
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 2 of 52
Medical Solutions
Page 3
Table of Contents3
0Table of Contents
1 _______ General information______________________________________________ 5
13 ______ Changes to Previous Version _____________________________________ 51
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 4 of 52
Medical Solutions
Page 5
General information5
1-
1General information
1.1Required Documents
•Isocenter AdjustmentRXL2-120.071.01
•Spare Parts ListRXL2-120.081..
•Start-up InstructionsRXL2-120.815.01
•PlanningRXL2-120.021..
•Maintenance ProtocolRXL2-120.105.01.05.
•Image quality binderRXD0-000.064.01..
•Ultrasound Localizer Start-upRXL2-120.091.10...
•Shock Wave Pressure and Position CheckSPL2-120.074.01...
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 5 of 52
Page 6
6General information
1.2 Test and Auxiliary Equipment
ToolFor exampleMaterial no.:
•Protective conductor meter44 15 899 RV 090
•Digital voltmeterFluke 18799 94 831
•Service PC with connecting cable according to ARTD
•Storage oscilloscopeFluke Scope Meter73 92 074
•Dose meterDIADOS97 17 612
PTW Nomex97 08 637 Y0388
•Copper filter - set (10 x 0.3 mm)44 06 120 RV 090
•17-mm Cu strip (minimum contrast check)11 67 662 G 5247
•2.1-mm precision copper filter99 00 598 XE 999
•ESD kits, ARTD 3- Wrist band34 63 874
- Cable34 64 062
•TV dynamic test case37 90 156 X 1963
•PTU pressure tester
- PTU adapter
30 95 408
70 41 028
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 6 of 52
Medical Solutions
Page 7
General information7
1.3 Tools
NOTE
The indicated items are listed in the STC (Service Tools Catalog)
unless otherwise stated (the STC is a component of the Spare
Parts Catalog) with the exception of items marked with "*".
•Start-up tool kit
•High-pressure grease gun28 84 609 RE 012
•1 kg ”Longtime PD2” lubricant or
20 g ”Longtime PD2” lubricant
•1 l turbo oil bottle HN 32 SRN 30 021
Squirt oiler 0.5 l
73 95 445 RH 999
34 91 271
37 83 359 J 1042
38 02 415
•Torque wrench (20 to 100 Nm)44 30 906 RH 090
•for POLYDOROS: Installation material "washers"88 67 665 X 1122
•1 liter transformer oil as needed 99 00 945
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 7 of 52
Page 8
8General information
1.4 Tolerance data
General tolerances for linear dimensions in compliance with ISO 2768
Limit value
for the nominal range
Admissible
tolerance
These tolerances apply to all dimensions indicated in these instructions, unless other tolerances are specifically indicated after the value.
The tolerance for torque values is ± 10%.
over 3
mm to 6
mm
± 0.5
mm
over 6
mm to
30 mm
± 1 mm ± 1.5
over 30
mm to
120 mm
mm
over 120
mm to
400 mm
± 1.5
mm
over 400
mm to
1,000
mm
± 4 mm± 6 mm± 8mm
over
1,000
mm to
2,000
mm
over
2,000
mm to
4,000
mm
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 8 of 52
Medical Solutions
Page 9
General information9
1.5 Abbreviations
Abbrev.Description
SISafety Inspection
SIEElectrical Safety
SIMMechanical Safety
PMPreventive Maintenance
PMPPeriodic Preventive Maintenance
PMAPreventive Maintenance Adjustments
PMFPreventive Operating Value Check and Functional Check
QQuality Check
QIQImage Quality
QSQSystem Quality Check
SWSoftware Maintenance
The steps identified by these abbreviations are part of the maintenance report and should
be checked off upon completion.
NOTE
Unless otherwise indicated, the maintenance interval is 12
months.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 9 of 52
Page 10
10Inspection of exterior and inspection of surroundings
2-
2Inspection of exterior and inspection of surroundings
2.1Inspection of exterior
PMPUnit
PMPCover
PMPTable pad
PMPShock wave head/coupling bellows
PMPImage intensifier
PMPX-ray tube assembly cover
PMPControl PC
PMPKeyboard
PMPMouse
•Check the entire system for mechanical damage.
•Replace or repair any defective parts.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 10 of 52
Medical Solutions
Page 11
Inspection of exterior and inspection of surroundings11
2.2 Inspection of surroundings
2.2.1 Unit movements
PMPPatient monitor
PMPMonitor trolley
PMPAdditional medical devices
•Ensure that there is free space around the unit to avoid collisions during unit movement.
•Mark the locations of the additional devices or secure the units.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 11 of 52
Page 12
12Safety inspections
3-
3Safety inspections
SIRemove the unit covers
•Remove the side covers of the unit.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 12 of 52
Medical Solutions
Page 13
Safety inspections13
3.1 Electrical Safety
SIEmergency-off switch function
•Operate the emergency-off switch in the room.
- The system must switch off.
- The system cannot be turned on until the emergency-off switch is released and the
voltage supply is switched back on.
SIFunction of the fault current circuit breaker
•Check the mechanical functioning of the fault current circuit breaker by operating the
built-in test button.
•Check electrical functioning according to ARTD-002.731.17.
SISafety circuit switch function
•Check the safety circuit for the O-W-X-Y motor movements in table G1 using the
J1042-7 circuit diagram.
•After connecting and logging on to the service PC, the Adjustment/Calibration service
software program can be used to move the motor drives to the safety switches without
software monitoring.
(See Operating the Service Software)
•Use the tableside control panel to move the drives into the safety circuit while monitor-
ing the appropriate limit switch.
NOTE
- When checking the X-drive using the "urology insert", move the O-drive to the +10°
position to avoid risk of collision with the tube assembly housing. During "litho operation", move the shock wave head to the coupling position and move the X-drive to
S78 in a limited range.
- In the case of a Y-drive check using the urology insert, move the O-drive to the 0° position. Switches S64/65 for the large range must be activated. During "litho operation", switches S76/77 for the limited range must be activated.
NOTE
SILines and cables
Do not move to the mechanical end positions!
If the safety switches are actuated, you can move the drives out of
the safety zone by activating service button S59 on N11 or service
button S58 on D32 in the unit.
•Check the line and cable insulation for damage or kinks to the extent possible.
•Check charging line K26 for a secure connection and a good ground connection at the
charger and the shock wave generator.
SIHV cable shielding
DIN VDE 0750 Part 21 (IEC 601-2-7) Paragraph 16
The high voltage cables that can be touched and that conduct the tube current must be
provided with a flexible, conductible shielding that has a maximum resistance of 1 ohm/m.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 13 of 52
Page 14
14Safety inspections
The shielding must be connected to both the metal housing of the X-ray generator and the
X-ray tube assembly.
Conduct a check using measurements as follows:
•Switch off the system contactor.
NOTE
Wait until the red LEDs on D110 are no longer lit (approx. 1 min.).
•Disconnect the high voltage cable from the socket at the HV generator and measure the
shielding mesh against protective ground. Move the cable, particularly at bends and
loops.
- High voltage cables with a shielding mesh resistance
exceeding the maximum of 1 ohm/m are to be replaced.
- Reinsert the cables
Only use transformer oil for inserting the high voltage plugs at the HV generator. Do not
use silicone washers!
Set of spare parts:
Installation material "washers"Part no. 88 67 665 X 1122
1 l transformer oil as neededPart no. 99 00 945
SITime - radiation indicator
Check functioning of the radiation indicator and functioning of the different FL release buttons while the FL time display and/or time-out for fluoroscopy are running.
•Reset the fluoroscopy timer to zero.
X
X
X
•Move the filter (approx. 0.3 mm Cu) into the fluoro unit beam projection.
•Switch fluoroscopy ON at the fluoro unit.
The timer counts up.
Observe radiation protection regulations!
•Check the radiation indicator at the control monitor.
During fluoroscopy, a symbol for radiation release appears on the left of the control
monitor in the status field.
•Check functioning of the FL release buttons on the connection panel of the control con-
sole and the footswitch at the console and/or unit.
•During fluoroscopy and exposure, check functioning of the radiation indicator lamp on
monitor trolley N72 (optional) or the local radiation indicators.
After a total of 4.5 minutes, a signal generator must sound.
•Switch off the signal generator. Do not reset the FL timer.
After 9.5 minutes, the signal generator in the control console must sound.
•Switch off the signal generator. Do not reset the FL timer.
Fluoro must switch off within 10 minutes.
•Switch fluoroscopy OFF at the fluoro unit.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 14 of 52
Medical Solutions
Page 15
Safety inspections15
•Reset the fluoroscopy timer to zero.
(See Start-up Instructions, paragraph 4-6)
SIFunction switch of the generator
•Check functioning of the service switch on board D100 in the generator. When the ser-
vice switch is off, radiation release must be blocked (no activation of the inverter!)
•Check functioning of switch S2 on board D160. When S2 is in the OFF position, the line
contactor of the generator may not switch on. (Multiline, not operating)
NOTE
SIProtective conductor check
When the generator is switched off, there is still line voltage
present at the T1 transformer and the D160 switch-on circuit.
(See LX circuit diagram X2169-11 or SX circuit diagram X2206-11)
After switching off the generator, approx. 600 V DC is still available
for the inverter! This is indicated by LEDs V135 and V136 on board
D110.
(See LX circuit diagram X2169-16 or SX circuit diagram X2206-16)
The voltage dissipates within approx. 1.5 minutes to 0 V; the LEDs
go off at approx. 30 V.
•Test the protective conductor with the protective conductor tester according to
ARTD-002.731.17.
The protective conductor connection in the generator is the central protective
conductor.
(The protective conductor check must be performed again as a final step when the
system is closed!)
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 15 of 52
Page 16
16Safety inspections
3.2 Mechanical Safety
SIToothed belt of the W-drive
SIToothed belt of the O-drive
•Visually inspect the toothed belts of the drive for wear.
•Cracks on the back or at the tooth base; replace toothed belts if necessary.
•Ensure that the clamping pieces of the toothed belts are secure.
SIX-ray tube assembly
SICollimator
SIImage intensifier with camera head
•Ensure that the screw connections of the X-ray tube assembly on the C-arm are secure.
•Ensure that the collimator is securely attached to the X-ray tube assembly.
•To check the mounting of the collimator, remove the large rectangular cover of the radi-
ation protection housing and then reinstall it. The collimator may not rotate at the housing flange.
•To check the image intensifier mounting, remove its rubber cover and the protective
cover of the camera.
•Check the mounting of the image intensifier.
•Check the mounting of the camera at the image intensifier.
•Reinstall covers.
SIUnit mounting
SISafety distance of the unit
•Use a torque wrench to check the attachment of the table to the base plate. Ensure that
8 x Allen screws (at the lifting columns) M8 x 40 with 24 Nm ± 10% 4 x external hexagons M10x25 are secure.
•Check the maneuverability ranges of the unit, the O-W -X-Y-Z-drive during lithotripsy
and urology operation. Maintain the safety distances in the end positions of the unit
movements.
SIUrology insert
•Insert the urology insert and ensure that it is secure.
•Ensure that S22 is properly switched.
•Adjust switch S22 if necessary.
•If needed, adjust both guide pieces of the urology insert. For this purpose, insert the
urology insert and press both guide pieces toward the outside.
•Tighten the screws.
SIShock wave heads
•Check the maneuverability of the shock wave system support (W-drive) from the
parked position to the working position and back.
•Check the limited range by moving unit G1 in the table cutout.
SIUnit grip strip
•Ensure that the unit grip strip is secure.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 16 of 52
Medical Solutions
Page 17
Safety inspections17
•Check the accessory parts for proper functioning.
SITable extensions
•Ensure that the table extensions are securely mounted.
•Check the patient position detection (head/foot) for proper functioning and check the
patient position display on the control monitor.
SIUS applicator
•Insert the US applicator in the shock wave head and secure it.
•Ensure that the US applicator is secure.
•Ensure that the US applicator module is undamaged. If there is damage, replace the
US applicator.
SICassette holder mounting
SIExposure display on the control monitor
•Install the cassette holder and insert a cassette with film.
• Select 77kV, 100 mAs, 80% output on the control PC.
X
•Releasing an exposure:
- The radiation symbol must be displayed on the control PC during the exposure (X-ray
radiation).
- The buzzer (whistle) must be audible in the control PC for the duration of the exposure.
•Change the cassette and develop the film after every exposure.
•Ensure that the cassette holder is secure.
- When the cassette holder is lifted, the collision protection (yellow safety button
-) must light up yellow.
SICollision protection
•Check the collision protection of the X-ray tube assembly housing. Given blockage of
the X-ray tube assembly housing when moving the drives in the O-W-X-Y-Z direction,
the collision protection must stop the movement.
The yellow safety button on the unit lights up.
Given unit blockage (collision protection) during movement of a drive, this button can be
used to carry out the unit movement in the opposite direction. In the event of blockage, the
corresponding unit movement button blinks.
In the event of unit blockage (collision protection) during unit drive movement:
•The yellow safety button lights up.
•Move in the opposite direction of the unit blockage: Operate the safety button and the
button on the tableside control panel.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 17 of 52
Page 18
18Safety inspections
Fast beeping
¹ Wrong direction - the unit is still moving into the blockage.
Slow beeping
¹ Correct direction - the unit is moving away from the blockage.
•Check the collision protection of the image intensifier. Given blockage of the X-ray tube
assembly housing when moving the drives in the O-W-X-Y-Z direction, the collision protection must stop the movement.
On the connection panel of the control console and the unit the yellow safety button
lights up.
•Interrupt the light barrier, if present.
Unit movements should not be possible.
Yellow danger buttons light up.
Remove any interruptions in the light barrier.
•Activate the safety cover on the shock wave head (L+R).
Unit movements should not be possible.
Yellow danger buttons light up.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 18 of 52
Medical Solutions
Page 19
Inspection for wear and replacement of parts19
4-
4Inspection for wear and replacement of parts
4.1Periodic preventive replacement of parts
PMPBattery
•Check the battery on board D1 (≥ 2.7 V) and replace it at least once every 36 months.
•Document this in the operating data report/log book.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 19 of 52
Page 20
20Inspection for wear and replacement of parts
4.2 Inspection for wear
4.2.1 Unit/mechanical components
PMPO-W-X-Y-Z drives
PMPO-drive toothed belt
PMPW-drive toothed belt
PMPX-drive toothed belt
PMPY-drive toothed belt
PMPCL-drive toothed belt
PMPZ-drive plastic clutch
•Check the drives of the motors (in the O-W-X-Y-Z direction) for wear.
•Check the drive toothed belts (in the O-W-X-Y-CL directions) for wear and replace them
as needed. If the toothed belt of the cone lenght (CL) drive is worn, the cone drive must
be replaced.
(see the spare parts list).
•Ensure that the teeth of the toothed belt engage securely with the recesses of the drive
wheel.
•Check for wear on the plastic clutch between the motor axis and the drive axis of the
Z-drives.
PMPActual value potentiometer
•Check the drive belts for the actual value potentiometer (O-W-X-Y drive) for wear and
replace them as needed
(see the spare parts list).
PMPPressure tester
PMPPTU adapter-M
PMPShock wave head
•Perform a pressure test at the shock wave head.
•Check and document the pressure value.
See Shock Wave Pressure and Position Check Service Instructions SPL2-120.074.01.
•Check functioning of switch S28 "adjustment aid attached".
(Blocking of the W-drive)
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 20 of 52
Medical Solutions
Page 21
Inspection for wear and replacement of parts21
4.3 Respiratory belt
PMFWear and tear
•Check the respiratory belt for damage/wear and tightness and replace it in the event of
damage.
(See the spare parts list).
PMPContacts at the PS contactor (generator)
Due to the low impedance of the intermediate circuit, the power-on sequence occurs in
two steps. In the first step, the intermediate circuit capacitors are charged via the LS contactor. After the charging time, the switch-on damping circuit is bypassed with the PS contactor.
Given errors in the charging process or when disconnecting the PS contactor contacts
under load (fuses, FI switch, etc.), these can be overloaded.
•Switch off the system contactor.
NOTE
Wait until the red LEDs on D110 are no longer lit (approx. 1 min.).
•Disconnect the RC element from the PS contactor.
•Remove the spark quenching chamber from the PS contactor (remove 2 screws).
- In the case of an older POLYDOROS LX (< series no. 1020), the contactor must be
taken out in order to remove the spark quenching chamber.
The following is required for conversion:
- Cable deflector 31 57 141
- Holder for cable deflector 31 57 364
•Check the contacts for excessive melting loss.
•Reattach the spark quenching chamber.
•Reattach the RC element.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 21 of 52
Page 22
22Maintenance
5-
5Maintenance
5.1Changing the water in the coupling circuit
Removing air pockets in the coupling bellows
PMAShock wave head
The water in the coupling circuit is to be changed:
•Move switch S2 on board W12 to the "Empty coupling bellows" position (Y3 opens), see
the start-up instructions.
•Connect the service hose with a funnel to the water tank and insert the hose end into an
available container. The water tank is emptied.
•If the system parts are very dirty, clean them prior to refilling. (If necessary, remove the
tank, fill it halfway with water, and shake it to remove the dirt).
•After emptying the water tank, fill the coupling circuit in accordance with the start-up in-
structions.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 22 of 52
Medical Solutions
Page 23
Maintenance23
5.2 Changing the water in the cooling circuit,
Checking flow and the water tank
Fig. 1:Quick vent valveFig. 2:Flowmeter
5.2.1 Cleaning the quick vent valve
PMAWater preparation
PMAWater system: Cleaning the system parts
Contamination in the cooling circuit can settle in the quick vent valve (Fig.1/p.23) so
that the valve can no longer close properly and the water leaks.
•Disconnect the hose connections from the front and back of the filter housing and re-
move the hoses.
•Loosen the four fastening screws of the filter housing and remove the entire housing in-
cluding the installed quick vent valve.
NOTE
•Empty the water from the filter housing.
•Unscrew the cover of the vent valve. Use open-end wrench SW19 to provide
counter-support to the housing base. (If the cover cannot be loosened manually, carefully use water pump pliers to loosen it).
•Remove the cover and lift the float out of the housing.
Do no exert any pressure on the sides of the quick vent valve.
NOTE
The float (Fig.1/p.23) is accommodated in the cover and can fall
out of its holder during removal.
•Remove the float from the holder in the cover.
•Clean the sealing surfaces of the float holder and the vent hole (Fig.1/p.23) with flow-
ing water.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 23 of 52
Page 24
24Maintenance
•Remove build-up in the float housing and rinse it out with water.
•Check the sealing ring (O ring) on the cover for damage and replace the complete quick
vent valve if necessary.
•Put the float back in the correct position in the holder and insert it in the housing.
•Manually tighten the cover. Use open-end wrench SW19 to provide counter-support to
the housing base.
•Install the filter housing from below and secure it again.
•Turn the top of the vent valve housing counter-clockwise to OPEN (approx. 1/8 rota-
tion).
•Connect the hoses to the front and back of the filter housing.
PMACooling circuit: Water replacement
PMACooling circuit: Flow/bubble-free
PMAWater tank: Check the level
The water in the cooling circuit is to be changed:
•Remove the water hoses at interfaces 5 and 8 at the water interface of the shock wave
cabinet.
•Open the seals with the matching ends of the hoses from the service cabinet and emp-
ty the cooling water into the prepared container.
•Fill the cooling system and set the flow according to the start-up instructions.
•Replace the water filter, part no. 70 52 012 F 0428, if necessary.
•Refill the water supply tank after the water has been changed.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 24 of 52
Medical Solutions
Page 25
Maintenance25
5.3 Lubricating and oiling the linear guides of the drives
PMALinear guide, longitudinal (lubricate)
PMALinear guide, transverse (lubricate)
Use the high-pressure grease gun to lubricate the bearing blocks of the linear guides.
Y
X
1
Fig. 3:Longitudinal drive and transverse drive
•To access the front points to be lubricated (on the patient table), move the system to the
"gel position".
•Loosen and remove the bottom white cover plate on the front side under the orbital
drive by unscrewing the 5 screws (Allen 3 mm) (Fig.3/p.25).
The lubricating nipples of the bearing blocks are then accessible from the front, on the
bottom.
To access the back points to be lubricated, move the system to the "lift position".
The lubricating nipples of the bearing blocks are accessible from the back, on the bottom.
NOTE
After the linear guide has been lubricated, move the system to the "uro position" and then
move it several times to the end positions of the X and Y-drives to lubricate the linear
guide evenly.
Apply the grease gun axially in order to prevent damage to the lubricating nipples!
•Remove excess lubricant from the bearing blocks and linear guides.
Reattach the removed cover plate.
PMAGuide rail, W-drive (oil)
PMAGuide rails, orbital (oil)
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 25 of 52
Page 26
26Maintenance
PMAGuide rails, orbital (oil)
PMAGear wheel, O-drive (lubricate)
•Oil the guide rails of the O and W-drives via the dirt removal device and oil container.
O
W
1
Fig. 4:Orbital drive and W-drive
•Oil the guide rails integrated in the C-arm at the vertical drive. Mechanically clean the
dirt removal areas at the points indicated by arrows and apply a few drops of oil to the
oiling means (felt).
•Oil each drive on the right and left side.
•Oil the guide rails on the W drive, see arrows in ((Fig.4/p.26)) and fill the containers
between the running wheels of the shock generator cart with oil.
•Move the W-drive to the middle position.
•Use the oilcan to fill the containers at the top and bottom of the guide rail with oil.
•Lubricate the crown gears of the O-drive motor on the orbital drive (Fig.4/p.26).
•Remove old or excess lubricant.
•After oiling the W and O-drives, move them to the end positions several times to oil the
guide rails evenly.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 26 of 52
Medical Solutions
Page 27
System heat dissipation27
6-
6System heat dissipation
6.1Cooling unit in the service cabinet
PMPSwitching area
PMPSeals/leaks
•Check whether the thermostat of the cooling unit is responding.
- Mark the controller position for this purpose.
- Turn the thermostat to the left to set a lower temperature. This causes the cooling
unit to switch on.
- Turn the temperature controller back to the marked position.
•Check the complete cooling unit for leaks.
•Ensure that the intake and exhaust air slots at the cooling unit are dust-free. Clean them
if necessary.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 27 of 52
Page 28
28System heat dissipation
6.2 Cabinet ventilation
PMPIntake air and ventilation shafts unobstructed
•Check whether the intake air and ventilation shafts are unobstructed when the cabinet
is closed in order to ensure a fresh-air supply.
•Remove any objects placed on the top of the cabinet!
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 28 of 52
Medical Solutions
Page 29
System heat dissipation29
6.3 MEMOSKOP
PMPCleaning the ventilation slots
•Remove dust from the bottom of the ventilation slots.
•Perform a functional check of the fan installed in MEMOSKOP.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 29 of 52
Page 30
30System heat dissipation
6.4 X-ray tube assembly fan
PMPFunction
PMPCleaning the ventilation slots
•Check the X-ray tube assembly fan for proper functioning.
•Clean the ventilation slots at the X-ray tube assembly housing.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 30 of 52
Medical Solutions
Page 31
Operating value inspection31
7-
7Operating value inspection
7.1Respiratory gate control
PMALearning time, window
•Ensure proper functioning of the respiratory gate control at the physiological monitor.
NOTE
The boards cannot be adjusted due to the complicated measuring
setup!
•Perform a functional check in accordance with Start-up Instructions RXL2-120.815...
•Ensure proper functioning of the respiratory belt at both connection points.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 31 of 52
Page 32
32Operating value inspection
7.2 ECG control
PMAFunction
•Ensure proper functioning of the ECG control. For this purpose, attach electrodes to a
test person. The signal is displayed on the physiological monitor. A trigger is only released when a QRS complex is detected!
•Perform a functional check in accordance with Start-up Instructions RXL2-120.815...
(See the Triggering Shock Waves section).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 32 of 52
Medical Solutions
Page 33
Operating value inspection33
7.3 Dose control
QIQAdjustment values
•Check the dose during fluoroscopy.
X
•(See Start-up Instructions RXL2-120.815... for the assigned value
see the test report).
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 33 of 52
Page 34
34Operating value inspection
7.4 Generator
PMAChecking/setting the tube filament heating
Not necessary if the check according to the start-up instructions is within the tolerance!
XObserve radiation protection regulations!
In the case of POLYDOROS, the filament heating functions as a constant current regulator.
Filament vaporization increases the resistance. Given a constant filament current, the filament power and the filament temperature also increase. If the filament current is not corrected, filament vaporization accelerates.
An increase in filament temperature of approx. 2-3% reduces the service life of the tube
by half.
Keep prep. time (preparation) short!
Always push the exposure release button down fully.
Let go of the release button immediately following the exposure.
X
•Connect the oscilloscope to
POLYDOROS LX:
•Connect the oscilloscope to
POLYDOROS SX:
- CH 1 = D100.X18 (kV actual)
- CH 2 = D100.X61 (mA actual)
- Trig.ext. ( - ) = D100.X48 (SWR )
- CH 1 = D100.X61 (kV_ACT)
- CH 2 = D100.X64 (MA_ACT)
- Trig.ext. ( - ) = D100.X64 (SWR )
•Set the following for each focus:- 73 kV
-100 %
- Change the mAs value until an exposure time of approx. 100 ms is
displayed.
•Release an exposure.
•Evaluate the oscillogram.
- The high voltage should be ± 5 % of the set value within 10 ms.
- The tube current should be ± 5 % of the set value within 50 ms.
- It is favorable for the service life of the tube for the tube current to be below the set
value (max. 20%) during power-on.
- An initial brief spiked overshoot (e-function < 3 ms, depending on the length of the HV
cable) is the charging current of the HV capacitors and can be ignored (does not flow
over the tube and therefore does not affect dose).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 34 of 52
Medical Solutions
Page 35
Functional inspections35
8-
8Functional inspections
8.1PC keyboard
PMFKey functions
•Ensure proper functioning of all control keys on the keyboard in accordance with the op-
erating instructions.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 35 of 52
Page 36
36Functional inspections
8.2 Triggering shock waves
PMFTriggering
•Check shock wave triggering with and without respiratory and ECG triggering.
(See the start-up instructions).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 36 of 52
Medical Solutions
Page 37
Functional inspections37
8.3 Footswitches
PMFFunction
•Ensure proper functioning of the footswitch for fluoroscopy (Electrical Safety / p. 13)
section: Time - radiation indicator.
X
•Check the footswitch at the unit and control console.
(See the radiation indicator function).
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 37 of 52
Page 38
38Functional inspections
8.4 Unit footswitches - urology
PMFFunction
•Check all footswitch functions in accordance with the operating instructions.
•Check the cables and plug connections for damage.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 38 of 52
Medical Solutions
Page 39
Functional inspections39
8.5 Surveillance monitor
PMFBasic settings
•Check the basic settings and functioning of the physiological monitor. Take the appro-
priate amplification into account when displaying the monitoring signals.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 39 of 52
Page 40
40Functional inspections
8.6 Tableside control panel
PMFFunction
•Connect the tableside control panel for the unit via connection points A/B on the unit.
•Ensure that all button functions are operating properly.
•Check the cables and plug connections for damage.
•Ensure proper functioning of the tableside control panel of the control console.
•Ensure proper functioning of the ultrasound tableside control panel.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 40 of 52
Medical Solutions
Page 41
Functional inspections41
8.7 Control PC mouse
PMFFunction
•Ensure proper movement of the mouse pointer on the control monitor.
•If necessary, remove the ball from the mouse and clean the sensor.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 41 of 52
Page 42
42Quality inspections
9-
9Quality inspections
9.1Component inspection
9.1.1 Television chain
QSQImage quality
•Use the image quality test to check the image quality of the television chain.
(See Start-up Instructions RXL2-120.815...).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 42 of 52
Medical Solutions
Page 43
Quality inspections43
9.2 Final result quality inspection
9.2.1 Log book
QSQCompare/document check results
•Save the current system parameters and delete fault storage.
(See Operating the Service Software "Backup").
•Document the various counter values in the "Operational certificate" log book.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 43 of 52
Page 44
44System adjustment
10-
10System adjustment
10.1Collimator
PMARadiation field
PMAFixed formats
•Check the predefined fixed collimator formats and adjust them if necessary in accor-
dance with the setting instructions.
•Check the settings of the collimator iris and adjust them if necessary in accordance with
the setting instructions.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 44 of 52
Medical Solutions
Page 45
System adjustment45
10.2 Cassette holder
PMACassette center
•Check the centering of the cassette in the cassette holder with respect to the radiation
field.
•If necessary, adjust it in accordance with the setting instructions.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 45 of 52
Page 46
46System adjustment
10.3 Isocenter
PMAAdjustment
•Check the isocenter at the table with the adjustment phantom and readjust it in the case
of deviations outside of the tolerance.
•Check the adjustment of the shock wave head.
(See "Adjusting the isocenter" in the adjustment instructions).
•If the ultrasound option is available, also check the isocenter with ultrasound. (See in-
structions RXL-120.091....).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 46 of 52
Medical Solutions
Page 47
Maintenance47
11-
11Maintenance
11.1Unit and system cabinets
PMPCleaning the exterior
PMPCleaning the interior
PMPPaint
•Clean the unit and the system cabinets
•Remove contamination from inside the cabinets
•Clean the fan and ventilation slots
•Repair paint damage as necessary, for the paint number
see Planning Guide RXL2-120.021.01.
•Clean the tube cone from the inside
(Only perform if the ultrasound option is available).
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 47 of 52
Page 48
48Maintenance
11.2 Log book
PMPCompleteness
•Check and document completed updates.
•Fill out the system papers, check them for completeness, and add any missing informa-
tion.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 48 of 52
Medical Solutions
Page 49
Final Work Steps49
12-
12Final Work Steps
12.1Covers
PMFInstalling unit covers
•Reattach all covers and check them for damage.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 49 of 52
Page 50
50Final Work Steps
12.2 Unit function
PMFFunctions when the system is closed
•Check all unit functions.
•Perform a functional check of the accessory parts (see the operating instructions).
- Replace them in the case of damage (see the spare parts list).
SIProtective conductor test
•Check the protective conductor when the system is closed in accordance with
ARTD-002.731.17.
(Also see the start-up instructions).
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 50 of 52
Medical Solutions
Page 51
Changes to Previous Version51
13-
13Changes to Previous Version
Toothed belt, cone lenght (CL) drive added to "Inspection for Wear."
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 51 of 52
Page 52
52Changes to Previous Version
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 52 of 52
Medical Solutions
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.