The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.05
1995
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 2 of 52
Medical Solutions
Table of Contents3
0Table of Contents
1 _______ General information______________________________________________ 5
13 ______ Changes to Previous Version _____________________________________ 51
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 4 of 52
Medical Solutions
General information5
1-
1General information
1.1Required Documents
•Isocenter AdjustmentRXL2-120.071.01
•Spare Parts ListRXL2-120.081..
•Start-up InstructionsRXL2-120.815.01
•PlanningRXL2-120.021..
•Maintenance ProtocolRXL2-120.105.01.05.
•Image quality binderRXD0-000.064.01..
•Ultrasound Localizer Start-upRXL2-120.091.10...
•Shock Wave Pressure and Position CheckSPL2-120.074.01...
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 5 of 52
6General information
1.2 Test and Auxiliary Equipment
ToolFor exampleMaterial no.:
•Protective conductor meter44 15 899 RV 090
•Digital voltmeterFluke 18799 94 831
•Service PC with connecting cable according to ARTD
•Storage oscilloscopeFluke Scope Meter73 92 074
•Dose meterDIADOS97 17 612
PTW Nomex97 08 637 Y0388
•Copper filter - set (10 x 0.3 mm)44 06 120 RV 090
•17-mm Cu strip (minimum contrast check)11 67 662 G 5247
•2.1-mm precision copper filter99 00 598 XE 999
•ESD kits, ARTD 3- Wrist band34 63 874
- Cable34 64 062
•TV dynamic test case37 90 156 X 1963
•PTU pressure tester
- PTU adapter
30 95 408
70 41 028
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 6 of 52
Medical Solutions
General information7
1.3 Tools
NOTE
The indicated items are listed in the STC (Service Tools Catalog)
unless otherwise stated (the STC is a component of the Spare
Parts Catalog) with the exception of items marked with "*".
•Start-up tool kit
•High-pressure grease gun28 84 609 RE 012
•1 kg ”Longtime PD2” lubricant or
20 g ”Longtime PD2” lubricant
•1 l turbo oil bottle HN 32 SRN 30 021
Squirt oiler 0.5 l
73 95 445 RH 999
34 91 271
37 83 359 J 1042
38 02 415
•Torque wrench (20 to 100 Nm)44 30 906 RH 090
•for POLYDOROS: Installation material "washers"88 67 665 X 1122
•1 liter transformer oil as needed 99 00 945
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 7 of 52
8General information
1.4 Tolerance data
General tolerances for linear dimensions in compliance with ISO 2768
Limit value
for the nominal range
Admissible
tolerance
These tolerances apply to all dimensions indicated in these instructions, unless other tolerances are specifically indicated after the value.
The tolerance for torque values is ± 10%.
over 3
mm to 6
mm
± 0.5
mm
over 6
mm to
30 mm
± 1 mm ± 1.5
over 30
mm to
120 mm
mm
over 120
mm to
400 mm
± 1.5
mm
over 400
mm to
1,000
mm
± 4 mm± 6 mm± 8mm
over
1,000
mm to
2,000
mm
over
2,000
mm to
4,000
mm
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 8 of 52
Medical Solutions
General information9
1.5 Abbreviations
Abbrev.Description
SISafety Inspection
SIEElectrical Safety
SIMMechanical Safety
PMPreventive Maintenance
PMPPeriodic Preventive Maintenance
PMAPreventive Maintenance Adjustments
PMFPreventive Operating Value Check and Functional Check
QQuality Check
QIQImage Quality
QSQSystem Quality Check
SWSoftware Maintenance
The steps identified by these abbreviations are part of the maintenance report and should
be checked off upon completion.
NOTE
Unless otherwise indicated, the maintenance interval is 12
months.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 9 of 52
10Inspection of exterior and inspection of surroundings
2-
2Inspection of exterior and inspection of surroundings
2.1Inspection of exterior
PMPUnit
PMPCover
PMPTable pad
PMPShock wave head/coupling bellows
PMPImage intensifier
PMPX-ray tube assembly cover
PMPControl PC
PMPKeyboard
PMPMouse
•Check the entire system for mechanical damage.
•Replace or repair any defective parts.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 10 of 52
Medical Solutions
Inspection of exterior and inspection of surroundings11
2.2 Inspection of surroundings
2.2.1 Unit movements
PMPPatient monitor
PMPMonitor trolley
PMPAdditional medical devices
•Ensure that there is free space around the unit to avoid collisions during unit movement.
•Mark the locations of the additional devices or secure the units.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 11 of 52
12Safety inspections
3-
3Safety inspections
SIRemove the unit covers
•Remove the side covers of the unit.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 12 of 52
Medical Solutions
Safety inspections13
3.1 Electrical Safety
SIEmergency-off switch function
•Operate the emergency-off switch in the room.
- The system must switch off.
- The system cannot be turned on until the emergency-off switch is released and the
voltage supply is switched back on.
SIFunction of the fault current circuit breaker
•Check the mechanical functioning of the fault current circuit breaker by operating the
built-in test button.
•Check electrical functioning according to ARTD-002.731.17.
SISafety circuit switch function
•Check the safety circuit for the O-W-X-Y motor movements in table G1 using the
J1042-7 circuit diagram.
•After connecting and logging on to the service PC, the Adjustment/Calibration service
software program can be used to move the motor drives to the safety switches without
software monitoring.
(See Operating the Service Software)
•Use the tableside control panel to move the drives into the safety circuit while monitor-
ing the appropriate limit switch.
NOTE
- When checking the X-drive using the "urology insert", move the O-drive to the +10°
position to avoid risk of collision with the tube assembly housing. During "litho operation", move the shock wave head to the coupling position and move the X-drive to
S78 in a limited range.
- In the case of a Y-drive check using the urology insert, move the O-drive to the 0° position. Switches S64/65 for the large range must be activated. During "litho operation", switches S76/77 for the limited range must be activated.
NOTE
SILines and cables
Do not move to the mechanical end positions!
If the safety switches are actuated, you can move the drives out of
the safety zone by activating service button S59 on N11 or service
button S58 on D32 in the unit.
•Check the line and cable insulation for damage or kinks to the extent possible.
•Check charging line K26 for a secure connection and a good ground connection at the
charger and the shock wave generator.
SIHV cable shielding
DIN VDE 0750 Part 21 (IEC 601-2-7) Paragraph 16
The high voltage cables that can be touched and that conduct the tube current must be
provided with a flexible, conductible shielding that has a maximum resistance of 1 ohm/m.
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 13 of 52
14Safety inspections
The shielding must be connected to both the metal housing of the X-ray generator and the
X-ray tube assembly.
Conduct a check using measurements as follows:
•Switch off the system contactor.
NOTE
Wait until the red LEDs on D110 are no longer lit (approx. 1 min.).
•Disconnect the high voltage cable from the socket at the HV generator and measure the
shielding mesh against protective ground. Move the cable, particularly at bends and
loops.
- High voltage cables with a shielding mesh resistance
exceeding the maximum of 1 ohm/m are to be replaced.
- Reinsert the cables
Only use transformer oil for inserting the high voltage plugs at the HV generator. Do not
use silicone washers!
Set of spare parts:
Installation material "washers"Part no. 88 67 665 X 1122
1 l transformer oil as neededPart no. 99 00 945
SITime - radiation indicator
Check functioning of the radiation indicator and functioning of the different FL release buttons while the FL time display and/or time-out for fluoroscopy are running.
•Reset the fluoroscopy timer to zero.
X
X
X
•Move the filter (approx. 0.3 mm Cu) into the fluoro unit beam projection.
•Switch fluoroscopy ON at the fluoro unit.
The timer counts up.
Observe radiation protection regulations!
•Check the radiation indicator at the control monitor.
During fluoroscopy, a symbol for radiation release appears on the left of the control
monitor in the status field.
•Check functioning of the FL release buttons on the connection panel of the control con-
sole and the footswitch at the console and/or unit.
•During fluoroscopy and exposure, check functioning of the radiation indicator lamp on
monitor trolley N72 (optional) or the local radiation indicators.
After a total of 4.5 minutes, a signal generator must sound.
•Switch off the signal generator. Do not reset the FL timer.
After 9.5 minutes, the signal generator in the control console must sound.
•Switch off the signal generator. Do not reset the FL timer.
Fluoro must switch off within 10 minutes.
•Switch fluoroscopy OFF at the fluoro unit.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 14 of 52
Medical Solutions
Safety inspections15
•Reset the fluoroscopy timer to zero.
(See Start-up Instructions, paragraph 4-6)
SIFunction switch of the generator
•Check functioning of the service switch on board D100 in the generator. When the ser-
vice switch is off, radiation release must be blocked (no activation of the inverter!)
•Check functioning of switch S2 on board D160. When S2 is in the OFF position, the line
contactor of the generator may not switch on. (Multiline, not operating)
NOTE
SIProtective conductor check
When the generator is switched off, there is still line voltage
present at the T1 transformer and the D160 switch-on circuit.
(See LX circuit diagram X2169-11 or SX circuit diagram X2206-11)
After switching off the generator, approx. 600 V DC is still available
for the inverter! This is indicated by LEDs V135 and V136 on board
D110.
(See LX circuit diagram X2169-16 or SX circuit diagram X2206-16)
The voltage dissipates within approx. 1.5 minutes to 0 V; the LEDs
go off at approx. 30 V.
•Test the protective conductor with the protective conductor tester according to
ARTD-002.731.17.
The protective conductor connection in the generator is the central protective
conductor.
(The protective conductor check must be performed again as a final step when the
system is closed!)
Siemens AGRXL2-120.101.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 15 of 52
16Safety inspections
3.2 Mechanical Safety
SIToothed belt of the W-drive
SIToothed belt of the O-drive
•Visually inspect the toothed belts of the drive for wear.
•Cracks on the back or at the tooth base; replace toothed belts if necessary.
•Ensure that the clamping pieces of the toothed belts are secure.
SIX-ray tube assembly
SICollimator
SIImage intensifier with camera head
•Ensure that the screw connections of the X-ray tube assembly on the C-arm are secure.
•Ensure that the collimator is securely attached to the X-ray tube assembly.
•To check the mounting of the collimator, remove the large rectangular cover of the radi-
ation protection housing and then reinstall it. The collimator may not rotate at the housing flange.
•To check the image intensifier mounting, remove its rubber cover and the protective
cover of the camera.
•Check the mounting of the image intensifier.
•Check the mounting of the camera at the image intensifier.
•Reinstall covers.
SIUnit mounting
SISafety distance of the unit
•Use a torque wrench to check the attachment of the table to the base plate. Ensure that
8 x Allen screws (at the lifting columns) M8 x 40 with 24 Nm ± 10% 4 x external hexagons M10x25 are secure.
•Check the maneuverability ranges of the unit, the O-W -X-Y-Z-drive during lithotripsy
and urology operation. Maintain the safety distances in the end positions of the unit
movements.
SIUrology insert
•Insert the urology insert and ensure that it is secure.
•Ensure that S22 is properly switched.
•Adjust switch S22 if necessary.
•If needed, adjust both guide pieces of the urology insert. For this purpose, insert the
urology insert and press both guide pieces toward the outside.
•Tighten the screws.
SIShock wave heads
•Check the maneuverability of the shock wave system support (W-drive) from the
parked position to the working position and back.
•Check the limited range by moving unit G1 in the table cutout.
SIUnit grip strip
•Ensure that the unit grip strip is secure.
LITHOSTAR MultilineRXL2-120.101.01.06.02Siemens AG
08.05CS PS 24
Page 16 of 52
Medical Solutions
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.