The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.05
1995
Cust.-No.: Date: Protocol
1Protocol
SIEMENS Office:
Address:
Region:
Country:
Contact person:
Tel.:
CSE in charge:
Tel.:
Remarks Regarding the Protocol:
The protocol is valid as proof of quality for one check that must be performed on the system / component in one year.
The check must be performed in the specified intervals.
The results of the check are entered in this protocol.
The chapter numbers in front of the checkpoints indicate the corresponding chapters in
the particular instructions (see cover page).
The protocol must be completely filled out by the Customer Service Engineer, i.e.:
•All boxes must be filled out. If a box does not apply to the system or if no entry needs to
be made, check the “n.a.“ box.
•Enter the customer number (Cust. No.:) and the date of the check in the header of each
page so that each page can be allocated to a customer and to a check date.
•If there are complaints, the IVKs for the component about which a complaint has been
made as well as the type of complaint must be entered in the “Open Points“ table provided for this. Correction of these open points also must be documented in this table
with the date and a signature. If there are no open points, check “No“ and document this
with the date and a signature.
•If movable components (also test phantoms that are part of the system) that can be
used in different systems are used for the check, they must be entered in the “Movable
Components“ table provided for this.
•The measurement values for the measurements that must be performed during the
check must also be entered in the open spaces / tables provided for them.
•After completing the check, Page 3 of this protocol must be filled out and signed.
LITHOSTAR MultilineRXL2-120.105.01.06.02Siemens AG
08.05CS PS 24
Page 2 of 12
Medical Solutions
ProtocolDate: Cust.-No.:
Further Processing and Archiving of the Protocol
The protocol is a document and thus must be archived. After completing the test, it must
be filed in the corresponding register in the “System Owner Manual“ binder. If needed, a
copy can be handed to the customer.
System:
Serial No.:
Software Version:
Number of the Service Contract:
Type of Maintenance:
Evaluating the Condition of the System / Component
The system has no deficiencies. The image quality test
resulted in no differences from required reference values.
The system / component has slight deficiencies that have
no affect on continued operation of the system. However
they should be corrected preventively.
The image quality test resulted in no differences from
required reference values.
The system / component has serious deficiencies. For
safety reasons, continued operation of the system is permitted only after successfully correcting the deficiencies.
After completing all work steps, an evaluation was performed.
Signature:
Date:Name:
The operator or a person assigned for this has taken note of this evaluation.
(if national regulations require this)
Signature:
Date:Name:
Siemens AGRXL2-120.105.01.06.02LITHOSTAR Multiline
Medical Solutions
Only activities that are not described in the instructions for the system / component need
to be listed.
Date:
Additional activities performed:
OKnot OKn.a.
Open Points:
Yes:No:Signature:
Date:Name:
If “Yes“, enter the component with the IVK and the open point (only the number) in the
table. After completing maintenance, record this in the table.
IVKComponentOpen PointsCompleted
DateSignature
LITHOSTAR MultilineRXL2-120.105.01.06.02Siemens AG
08.05CS PS 24
Page 4 of 12
Medical Solutions
ProtocolDate: Cust.-No.:
Measuring Devices queried electronically:
Yes:No:Signature:
Date:Name:
If the measurement devices are queried electronically, for example with a Scout Mobile
Device, entry of the measuring devices in the table can be skipped.
Measuring DevicesTypeSerial No.Date Used Next Cali-
bration
Due
Movable Components:
Yes:No:Signature:
Date:Name:
If “Yes“, enter the movable component with which the check was performed along with the
with the Serial No. in the table.
Movable components (also test phantoms that are part of the system) are parts that can
be used on different systems).
ComponentSerial No.
Siemens AGRXL2-120.105.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 5 of 12
Cust.-No.: Date: Protocol
OK not n.a.
OK
1General information
2Inspection of exterior and inspection of surroundings
2.1Inspection of exterior
PMPUnit
PMPCover
PMPTable pad
PMPShock wave head/coupling bellows
PMPImage intensifier
PMPX-ray tube assembly cover
PMPControl PC
PMPKeyboard
PMPMouse
2.2Inspection of surroundings
2.2.1Unit movements
PMPPatient monitor
PMPMonitor trolley
PMPAdditional medical devices
3Safety inspections
SIRemove the unit covers
3.1Electrical Safety
SIEmergency-off switch function
SIFunction of the fault current circuit breaker
SISafety circuit switch function
SILines and cables
SIHV cable shielding
Measured value:
SITime - radiation indicator
SIFunction switch of the generator
SIProtective conductor check
Measured value:
3.2Mechanical Safety
SIToothed belt of the W-drive
SIToothed belt of the O-drive
SIX-ray tube assembly
SICollimator
LITHOSTAR MultilineRXL2-120.105.01.06.02Siemens AG
08.05CS PS 24
Page 6 of 12
Medical Solutions
ProtocolDate: Cust.-No.:
OK not n.a.
OK
SIImage intensifier with camera head
SIUnit mounting
SISafety distance of the unit
SIUrology insert
SIShock wave heads
SIUnit grip strip
SITable extensions
SIUS applicator
Option present:Yes:No:
Signature:
Date:Name:
SICassette holder mounting
SIExposure display on the control monitor
SICollision protection
4Inspection for wear and replacement of parts
4.1Periodic preventive replacement of parts
PMPBattery
4.2Inspection for wear
4.2.1Unit/mechanical components
PMPO-W-X-Y-Z drives
PMPO-drive toothed belt
PMPW-drive toothed belt
PMPX-drive toothed belt
PMPY-drive toothed belt
PMPCL-drive toothed belt
PMPZ-drive plastic clutch
PMPActual value potentiometer
PMPPressure tester
PMPPTU adapter-M
PMPShock wave head
4.3Respiratory belt
Option present:Yes:No:
Signature:
Date:Name:
Siemens AGRXL2-120.105.01.06.02LITHOSTAR Multiline
Medical Solutions
08.05CS PS 24
Page 7 of 12
Cust.-No.: Date: Protocol
OK not n.a.
OK
PMFWear and tear
PMPContacts at the PS contactor (generator)
5Maintenance
5.1Changing the water in the coupling circuit
PMAShock wave head
5.2Changing the water in the cooling circuit,
5.2.1Cleaning the quick vent valve
PMAWater preparation
PMAWater system: Cleaning the system parts
PMACooling circuit: Water replacement
PMACooling circuit: Flow/bubble-free
PMAWater tank: Check the level
5.3Lubricating and oiling the linear guides of the drives