The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 09.05
2005
Page 2
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
Analog boardRemove the resistor between X902 and X903,
refer to the service note "Adapting a
MEMOSKOP"
Digital
control
board
"Load Default Value" is possible the first time only
with the switch, not with the service software.
Select the LITHOSTAR measurement field;
perform a dose rate check;
perform an IQ test
For 33 cm I.I. only:
select horizontal and vertical image reversal
(Board
D1, analog
board / p.
30)
n.a.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 13 of 72
Page 14
14Overview of Measures
Memoskop component0
Use Service Instructions RX57-029.061.01... when working on the Memoskop.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 14 of 72
Medical Solutions
Page 15
Overview of Measures15
Multispot component0
Use the Setting Instructions RX57-091.032.01... when working on the Multispot 2000.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 15 of 72
Page 16
16Overview of Measures
Laser light localizer component0
The complete laser light localizer unit is replaced. It can be converted from battery operation to rechargeable battery pack operation (refer to Instructions RXL2-120.033.03...).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
Replace completely, no special attention
necessary
Replace completely;n.a.
switch
Detail
info
(Checking /
Replacing the
Lifting
Columns
R /
L/p.56)
n.a.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 18 of 72
Medical Solutions
Page 19
Overview of Measures19
Generator component0
The following instructions should be used when working on the generator.
LITHOSTAR Multiline Serial no. ≤ 6073RX63-020.034.12...
Serial no. > 7000POLYDOROS SXRXL2-120.034.05...
Serial no. > 7000POLYDOROS LXRXL2-120.034.02...
Troubleshooting
Instructions
Troubleshooting
Instructions
When replacing the image intensifier, X-ray tube or camera, use the Start-up instructions
for the specific component. In addition, adjust the system in accordance with the adjust-
ment instructions, "Isocenter Adjustment" RXL2-120.071.01....
POLYDOROS LXRX63-020.072.01...
POLYDOROS SXRX63-055.072.01...
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 19 of 72
Page 20
20Overview of Measures
Ultrasound option0
Use Start-up Instructions RXL2-120.091.10.01... when working on the ultrasound system.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 20 of 72
Medical Solutions
Page 21
PC Board Replacement21
3PC Board Replacement
4-
Component N 110
NOTE
Refer to wiring diagram J1042-3A for the rated current values of
the different fuses listed in this chapter.
Board D1, (new type) host computer or when replacing the battery0
78 28 841 J 1042, refer to wiring diagram J1042-21-4
30 82 138battery 3.0 V
Follow items A + B + D when replacing board D1.
Follow items A + C + D when replacing the battery.
•If possible, perform a backup of all current parameters and data.
A
(Refer to Operation of the Service Software RXL2-120.113.01. Chap. 3-3)
•Read out the counter values for
shock counter generatorshock counter shockwave
shock counter spark-gapshock counter total
shock counter US probe (starting with Rev. 05)
and document them in the operating log (Logbook, Reg. 8).
•Remove board D1 from electronics rack N11.
B
•Change EMC shielding plate, if present, from old to new board.
•Set or check the switches on the new board.
•Switch settings, see wiring diagram excerpt in the Appendix to this manual.
C
•Measure the battery voltage; U
> 2.7 V is normal.
batt
•If the battery voltage is < 2.7 V, replace the battery.
•Desolder the old battery and replace it with a new one.
D
•Measure the battery voltage; U
> 2.7 V is normal.
batt
•If the battery voltage is < 2.7 V, replace the battery.
•Insert board D1.
•After switching on the system:
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 21 of 72
Page 22
22PC Board Replacement
- Establish a connection between the service PC and D1 X10
- Start the service software (Refer to the service software operating instructions
RXL2-120.113...)
- Enter the current time and date "Realtime Clock"
- Execute a download of "Host Software, Parameter file" in the Data menu
- Execute a restore of the "Customer Data", "System Data" parameters in the
Data menu
- Call up "Technical Info" and "Service Data" in the Info menu with "Load from
File" and save by selecting "Put to Unit".
- Call up "Parameters" in the Adjustment menu
- Select "Interface Shockwavesystem" from the Parameter group
- The value 00 must appear in the "Value" column for:
Reset shock counter generator / Reset shock counter shockwave
Reset shock counter spark-gap / Reset shock counter total
Reset shock counter US probe (starting with Rev. 05)
•If necessary, reset the values in the Value column
(Reset with "FF"/normal operation with "00", refer to "Setting the software parameters" RXL2-120.032.01..)
NOTE
The Appendix of this manual contains a layout of the new D1
board. When board D1 is replaced, this layout is to be attached to
the LITHOSTAR Multiline wiring diagram.
Board D1, (old type) host computer or when replacing the battery0
31 25,825 J 1042, refer to wiring diagram J1042-21-4
31 46 073 B 2905battery 3.6 V
Follow items A + B + D when replacing board D1.
Follow items A + C + D when replacing the battery.
•If possible, perform a backup of all current parameters and data.
A
(Refer to Operation of the Service Software RXL2-120.113.01. Chap. 3-3)
•Read out the counter values for
shock counter generatorshock counter shockwave
shock counter spark-gapshock counter total
shock counter US probe (starting with Rev. 05)
and document them in the operating log (Logbook, Reg. 8).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 22 of 72
Medical Solutions
Page 23
PC Board Replacement23
•Remove board D1 from electronics rack N11.
B
•Change EMC shielding plate, if present, from old to new board.
•Set or check the switches on the new board.
•Replace Proms J66/J67.
•Solder in jumper X26 - X27 to activate the battery.
C
•Measure the battery voltage; U
•If the battery voltage is < 3.0 V, replace the battery.
•Desolder the old battery and replace it with a new one.
D
•Measure the battery voltage; U
> 3.0 V is normal.
batt
> 3.0 V is normal.
batt
•If the battery voltage is < 3.0 V, replace the battery.
•Insert board D1.
•After switching on the system:
- Establish a connection between the service PC and D1 X10
- Start the service software (Refer to the Service Software Operating Instructions RXL2-120.113...)
- Enter the current time and date "Realtime Clock"
- Execute a download of "Host Software, Parameter file" in the Data menu
- Execute a restore of the "Customer Data", "System Data" parameters in the
Data menu
- Call up "Technical Info" and "Service Data" in the Info menu with "Load from
File" and save by selecting "Put to Unit".
- Call up "Parameters" in the Adjustment menu
- Select "Interface Shockwavesystem" from the Parameter group
- The value 00 must appear in the "Value" column for:
Reset shock counter generator / Reset shock counter shockwave
Reset shock counter spark-gap / Reset shock counter total
Reset shock counter US probe (starting with Rev. 05)
•If necessary, reset the values in the Value column
(Reset with "FF"/normal operation with "00", refer to "Setting the software parameters" RXL2-120.032.01..)
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 23 of 72
Page 24
24PC Board Replacement
Board D4xx, Motor controller0
89 49 588 G 5334Refer to jumper log wiring diagram J 1042-24
•Set or check jumpers
•Check switch S2 (S2 defines the board as D4A, B or C).
NOTICE
PAL modules
¹ The PAL modules J5, J31 and J32 on the replacement
board must be removed and destroyed. Only the original
PAL modules of the old board D4 may be used.
•Replace Prom J38.
•Mark the board according to its location.
•Insert the board and switch the system on.
•Check the operating voltage at test point X10 +5 V +/- 250 mV (0V, X11).
Risk of short circuit! Do not use the 96-pin extension board 81 33
944 X1454 to extend the front plug D5.X2.
¹ Use only the short 96-pin extension board, part no. 85 73
628 G5206.
•Remove the front plug X2; remove board D5.
•Set or check jumpers.
•Check the Proms J1/J2 and replace them, if necessary.
•Insert the board and set front plug X2 to the short extension; switch on the system.
•Check the reference voltage U
Mp X31 - X30 (0V). If necessary, set it with potentiometer R220 on board D8.
Checking the A/D converter
= 5.120 V ± 5 mV by connecting a DVM to board D5 at
ref
•Measure and record the voltage value for every actual value potentiometer (r1 ... r6)
(Mp X10 ... X15).
•Read out and record the value of the ADC value for the corresponding potentiometer
with the service PC in the Diagnostic, Monitoring menu.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 24 of 72
Medical Solutions
Page 25
PC Board Replacement25
•Ensure that the relevant voltage value corresponds to the ADC value.
Conversion:
Voltage value (mV) = ADC value (bit) x 5mV/bit-> +/ - 20 mV tolerance
Voltage value (mV))
ADC value (bit) =
Example:
Measured value at D5 M
mV
Calculated valueADC value (bit) =
- Compare with recorded ADC value
-> the ADC value in bits for this channel is correct as long as the values do
not deviate from one another by more than +/- 4 bits.
Check the calculation: 587 bit x 5 mV/bit = 2935 mV +/- 20 mV
___________________
5mV/bit
X = 2935
p
2935 mV
___________ =
5 bit/mV
-> +/ - 4 bit tolerance
= 587 bit
•Switch off the system, remove the extension board and reconnect the front plug X2.
•Switch the system on.
•Check the operation of the various drives.
SMP bus board D100
11 14 839 G5347,wiring diagram J 1042-30
•Remove the cable connections from module N11.
•Remove the boards according to ESD guidelines and remove module N11 from the
cabinet.
•Replace board D10.
•Reinstall module N11 in reverse order.
•When installing the EMC shielding plate, ensure that the cables fit in the recesses pro-
vided and are not damaged.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 25 of 72
Page 26
26PC Board Replacement
•Attach the exposed cable shielding with clamps.
Starting with serial no. 7001:
Transfer the jumper from the old board to the new board. X112.4 - X110.4
The plug must connect 1-5. (Refer to the jumper log).
Board D11, backplane0
31 25 577 J1041,wiring diagram J 1042-31
•Remove the cable connections from module N11.
•Remove the boards according to ESD guidelines and remove module N11 from the
cabinet.
•Replace board D11.
•Reinstall module N11 in reverse order.
•When installing the EMC shielding plate, ensure that the cables fit in the recesses pro-
vided and are not damaged.
•Attach the exposed cable shielding with clamps.
•Check or set the jumper positions according to the jumper log or wiring diagram.
•Check the fuses F10...F16 according to the wiring diagram.
Board D 21, power supply M160
31 25 577 J1042,wiring diagram J 1042-41
•Check fuses F1 and F2.
•Replace the board.
•Switch the system on.
•Check LEDs V10... V16 and operating voltages.
Supply voltage of the magnetic brakesLED V10
MPX3.3/4+25 V to +29 V regulated
GndX3.1/2
Supply voltage of function switches and safety circuitsLED V11
MPX2.10+23 V to +26 V regulated
GndX2.6
Supply voltage of the DUEP interface to the image memoryLED V12
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 26 of 72
Medical Solutions
Page 27
PC Board Replacement27
MPX2.1+8.5 to +9.5 V regulated
GndX2.2
Supply voltage of the collimatorLED V13
MPX4.4+27 V to +35 V
GndX4.2
Supply voltage of the large drives (K1/K2 are energized)LED V14
MPX15+28 V to +35 V
GndX14
Supply voltage of the small drives (K3 is energized)LED V15
MPX17+28 V to +35 V
GndX16
Supply voltage of the Com controller (DUEP)LED V16
MPX2.9+14 V to +15.5 V regulated
GndX2.6
Board D4, power supply, respiratory and ECG triggering0
84 37 337 K2013,wiring diagram J 1042-44
•Check fuses F1 and F2.
•Measure the operating voltage (place the board on the extension)
+ 12 V ± 10%(X1.d/b/z4)LED V4The - 5V are not used in the
- 12 V ± 10%(X1.d/b/z4)LED V11
GND(X1.d/b/z4)
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 27 of 72
LITHOSTAR Multiline
.
Page 28
28PC Board Replacement
Collimator N0
Fig. 1:Strain relief
Replacing the collimator0
•Unpack the collimator.
•Remove the two colored screws from the transport bracket on the side of the blades
close to the focus.
•Check the adjustment of the leaves close to the focus with the template.
•Attach the collimator to the tube unit.
•If plug X1 has a connector shell, unscrew it, cut it open, remove it.
•Plug the X1 plug into the collimator and bolt it
•Attach the cable at the panel using strain relief clamps ((1/Fig.1/p.28)).
•Perform the adjustment test when you replace the entire collimator.
Replacing electric components0
•Switch off the system
•Replace the electric components according to the spare parts list and perform the ad-
justment test.
D1/S2.5
OFFCollimator without prefiltration
ONCollimator with prefiltration
(On D1X7, check that the cable for the drive is connected)
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 28 of 72
Medical Solutions
Page 29
PC Board Replacement29
Adjustment test0
•Set DIP switch S2.4 to "OFF".
•Close the "height" and "width" collimator blades completely by hand.
•Remove the prefilter (if present).
•Move the iris blades to an opening of 70 x 70 mm (measure at the iris blades).
•Switch the system back on or reinsert fuse F1 on D1.
•Switch DIP switch S2.4 to "ON" for initialization.
The system will initialize. DIP switch S2.4 should remain "ON" during operation.
•The height and width blades will open or close to the 44mm format (± 0.75%) x 44 mm
(± 0.75%) at a SID of 125 cm (35 mm x 35 mm at 100 cm SID).
Control position of the collimator blades0
•Install the cassette holder. Do not insert a film cassette.
•Move the system into the mid-range position.
•Release fluoroscopy.
•A radiation field of comparable size
44 mm (± 0.75%) x 44 mm (± 0.75%) - is visible on the monitor at a SID of 125 cm.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 29 of 72
Page 30
30PC Board Replacement
VIDEOMED DI-M component0
Board D1, analog board0
30 97 727 X2170wiring diagram X 2170-10
•Starting with revision level 01, the 620 ohm resistor between test points X902 and X903
should be removed.
Board D3, analog board0
30 97 651 X2170wiring diagram X 2170-13
•Select the "Multiline" measurement field from the specific parameters (refer to the Ser-
vice Instructions RX52-022.061.01...).
•Check the dose setting according to Start-up Instructions RXL2-120.034.01...).
•Perform an IQ test
•For 33 cm I.I. only:
Select "H-image reverse" and "V-image reverse" in the specific parameters and
transfer them individually to the system (Refer to the Service Instructions
RX52-022.061.01...).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 30 of 72
Medical Solutions
Page 31
PC Board Replacement31
Unit/control console component0
Board D32, distributor board, supporting frame0
31 25 676 J1042,wiring diagram J1042-52/52A
•Plug in and check the connection cables according to the wiring diagram.
•Check connection (X56) and the operation of the tube assembly fan.
•Check the operation of service switch S 58.
•Check the operation LED’s V14 to V18 for the magnetic brake by moving the associat-
ed drive (The LED lights up when the brake is released).
I.I. supply Z96 "Röderstein"0
31 36 194 X2134,wiring diagram X2134
•Replace the board and the power transistor at the same time.
•Check fuses F2 and F3.
•Check the operating voltage 27 V ± 10% for V11 "ON"; Tp 27 V → Tp 0.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 31 of 72
Page 32
32PC Board Replacement
Tableside control panel (NBF), board D410
31 23 879 J1042,wiring diagram J1042-61
Board D41 is coded in the factory to "Patient table basic unit".
Recoding for use on the control console, board D410
•Place the tableside control panel with the membrane keyboard side on a level surface.
•Remove four screws (M 2.5x16) from the lower part of the housing.
•Carefully lift off the lower housing.
Four metal spacers are lying loosely on the PC board.
•Set or check the DUEP slave address of switch S3 according to the application and the
wiring diagram:
Address 7: Tableside control panel
(NBF) basic unit (patient table) factory
setting
Address 2: Tableside control panel
(NBF) basic unit control console
Fig. 2:S3 Switch
Assembly0
•Center the four metal spacers on the contact surfaces and insert the four screws
(M2.5x16) in the holes in the lower part of the housing.
•Join the housing parts together so that one of the screws can be inserted through the
hole of the grounding tape.
The metal spacers must not move during assembly.
•Tighten the screws.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 32 of 72
Medical Solutions
Page 33
PC Board Replacement33
System unit0
30 10 175 JE002
•Replace the system unit, if necessary, according to the Repair Instructions
RXL2-120.091.02... .
Board D51, ADC/DUEP plug-in board0
31 25 627 J1042,wiring diagram J1042-71
Refer also to Repair Instructions "Replacement of the system unit" RXL2-120.091...
•Set and check the jumpers and Proms according to the jumper or wiring diagram.
•Check the 7-segment display
(Refer to the error list RXL2-120-072.03...).
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 33 of 72
Page 34
34Potentiometers and Reference Switches
4Potentiometers and Reference Switches
5-
Overview of the potentiometers and reference switches0
Fig. 3:Overview of potentiometers and reference switches
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 34 of 72
Medical Solutions
Page 35
Potentiometers and Reference Switches35
Basic information on the replacement and adjustment of
potentiometers, reference switches and motor encoders
Fig. 4:Replacing the potentiometer
General replacement and adjustment0
0
a) Potentiometers
•Move the unit so that the set screw becomes accessible ((M/Fig.4/p.35)).
•Switch the system off.
•Remove the Allen 1.5 mm set screw on the connecting piece of the potentiometer shaft
((M/Fig.4/p.35)).
•Remove the potentiometer retaining plate ((ISK 3 mm/Fig.4/p.35)).
•Remove the adjustment wheel ((E/Fig.4/p.35)) and the retaining parts from the po-
tentiometer shaft.
•Desolder the connection wires from the potentiometer and solder them to the new po-
tentiometer in the same order.
•Install the potentiometer in the reverse order. Do not tighten the set screw yet (Allen 1.5
mm).
•Switch the system on.
•Move the drive in "unregulated motion" with the service PC (Adjustment, Calibration) in
mm steps so that the corresponding reference switch is actuated (refer to
(Fig.5/p.36)). Check with DVM at D5.X2.A1...A11, 0V=X30.
NOTE
The potentiometer values are set in accordance with ADC values.
•Exit "unregulated motion" and select the "ADC value" for the corresponding drive in Di-
agnostic, Monitoring.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 35 of 72
Page 36
36Potentiometers and Reference Switches
•Set the ADC value to the corresponding value using the adjustment wheel
((E/Fig.4/p.35)) (Refer to the detailed description for the corresponding potentiome-
ter).
•Retighten the set screw on the potentiometer shaft.
•Reset board D1 with S1.
•Check the drive function. No errors may be displayed.
•If reference switch errors continue to be displayed after adjusting the potentiometer (Er-
rors 0020-002B), check the A/D converter on board D5 according to (Board D 5, A/D
converter / p. 24).
b) Reference switches
NOTE
A 24 V lamp may be used to check switch functions instead of an
oscilloscope.
•Switch the system off.
•Remove the reference switch in question and remove the switch rocker.
•Replace the switch and wire.
•Reinstall the switch with the switch rocker and check the operation of the switch rocker.
•Switch the system on.
Fig. 5:Replacing the reference switch
Connect an oscilloscope (or lamp ≥ 24 V) to the input of board D5 X2 a... to check the
reference switch function (GND: X30/X40) (refer to the Overview circuit diagram for the
respective reference switch J1042-8...).
•Move the corresponding drive far enough so that the reference position is reached on
the service PC (Diagnostic; Monitoring) (Refer to the chapter on adjusting the drives).
•Adjust the reference switch by moving it in the X direction. The switch should activate
within the mechanical limits of the switching hysteresis ((Fig.5/p.36)).
Check it using the service PC (Diagnostic; Monitoring; Refswitch; Y; X; 0; W; ZL; ZR;
TL; TR) or on the oscilloscope (lamp ≥ 24 V).
Move the drive over the reference switch position several times to check.
•Check the drive function.
•Move the activator of the drive up to approx. 1 cm in front of the reference switch.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 36 of 72
Medical Solutions
Page 37
Potentiometers and Reference Switches37
•Continue to move toward the switch in inching mode ((Fig.5/p.36)).
In addition, check the status of the reference switch on the service PC (Diagnostic,
Monitoring).
•If the switch is activated from both sides, check the switching hysteresis for both sides.
No errors may be displayed.
Overview of the switching hysteresis
•The test points of the actual value potentiometers and reference switches on board D5
are shown in the wiring diagram J1042-8x.
Fig. 6:Overview of the switching hysteresis
The switch can be ON or OFF within the range "X".
For the actual value potentiometers and reference switch test points on board D5, refer to
wiring diagram J1042-8x.
c) Checking synchronization of the actual value potentiometers to the motor
encoder
NOTE
Adjustment of the MC position with the ADC position
- When the system is switched ON or board D1 is reset with button S1, the value of
"MC position" is adjusted to the value of the "ADC position.“ (initialized).
- If the ADC position deviates from the MC position, the error "428x AD / MC position
comparison failure / x“ is displayed (x corresponds to the number of the drive).
Reason for the error message
- Potentiometer Actual value Rx not synchronous to the motor encoder Bx.
Maximum acceptable differences in the ADC/MC positions
ZL, ZR = 25 in 1/10 mm W = 30 in 1/10 mmX = 30 in 1/10 mmY = 30 in 1/10 mm
Starting with software VA00D (Software Upgrade Rev. 3), the physical limits of the drive are monitored.
O = 90 in 1/100°TL = 18 in 1/10 mmTR = 175 in 1/100°
Checking the display for the corresponding drive in the service PC
"MC position" and "ADC position" are displayed on the service PC (Diagnostic, Monitoring).
•Move the respective drive into an end position.
•Reset board D1 with button S1.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 37 of 72
Page 38
38Potentiometers and Reference Switches
•Check the respective adjustment for the drive ("MC position" to "ADC position") on the
service PC (Diagnostic, Monitoring).
•Move the drive to the opposite end stop in 5 steps.
•Perform a check at the service PC.
NOTE
Procedure for tensioning the toothed belt:
If the deviations of the ADC/MC positions are greater than the maximum acceptable deviations, the drives must be checked. If there
are errors in the lifting columns and/or the cone drive, they must
be replaced, since the actual value potentiometer cannot be adjusted.
•Loosen the screws at the idler pulley of the corresponding potentiometer toothed belt.
•Tighten the toothed belt by shifting the idler pulley and retighten the retaining plate.
•Reset board D1 with button S1.
•Perform a functional check as described above.
•Loosen the idler pulley again if shifting the idler pulley does not adjust the belt ade-
quately.
•Remove the retaining plate for the toothed belt end stop.
•Open the clamping piece of the toothed belt, shorten the belt by removing 4 teeth (2
teeth each parallel), insert the belt and reassemble completely.
•Perform a functional check as described above (check hysteresis of the reference
switch).
Error display:
„Antriebe stehen außerhalb der physikalischen Betriebsgrenzen"
"Drives are located outside of physical limits"
"Moteurs endehors des limites physiques"
"Los Motores se eneuentran fuera de los limites fisicos"
Error Cause:
Remedy:
•One of the drives (O, W, X, Y, Z, TL, TR) is in the ADC position in
the area of the hardware proximity switch outside the operating
range (Monitoring) (ZL, ZR = 50 in 1/10 mm; W, X, Y = 40 in 1/10
mm; O = 100 in 1/100°; TL = 20 in 1/10 mm; TR = 400 in 1/100 °)
•Check the respective drive using the service PC (Diagnostic, Moni-
toring).
•Move the drive into the mid-range position (inside the operating
limits) in "unregulated mode" (Avoid collisions).
•Reset board D1 with button S1.
•Perform a functional check. If necessary, repeat the check of the
display ("MC/ADC position") for the corresponding drive in the service PC. If there is no improvement, adjust the toothed belt.
Error display:
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 38 of 72
Medical Solutions
Page 39
Potentiometers and Reference Switches39
„Gerät befindet sich außerhalb der Betriebsgrenzen“
"Unit is outside the operating limits"
"L appereil est en dehors des limites d utilisation"
"Equipo se halla fuera de los limites de servicio"
Error Cause:- After exiting service mode "Unregulated motion" or
- when changing the Uro table, (switching over from Lithotripsy mode<-> Urology mode) O, W, X, Y, Z, TL, TR
•At least one drive is outside the respective operating limits. (Refer
above for the drive values)
Remedy:
•Move drives to the basic position or ascending position (inside the
operating limits) using the →0← button.
•Perform a functional check. If necessary, repeat the check of the
display ("MC/ADC position") for the corresponding drive in the service PC.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 39 of 72
Page 40
40Potentiometers and Reference Switches
Checking, replacement and/or adjustment0
Reference switches S3..6, S8, S9 or actual value potentiometers r3..6, r8, r9
NOTICE
Make the following preparations according to the respective work step:
Danger of collision!
¹ Move the drives to a mid-range position to avoid collision
when moving the system in "unregulated motion" mode.
¹ The shock wave head (W drive) is in the park position!
•Connect the service PC to board D1 X10.
•The ADC values, position values and reference switch settings can be read out in the
"Diagnostic, Monitoring" menu.
•As an alternative, you can proceed as follows:
- Remove the EMC shielding plate at the front of N11.
The test points on board D5 are now accessible from the front.
The test points of the ACTUAL value potentiometers are shown in wiring diagram J
1042-25.
- Remove the front plug X2 of board D5 and reconnect it using the short 96-pin extension board 85 73 628 G5206.
(service accessory for the engineer).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 40 of 72
Medical Solutions
Page 41
Potentiometers and Reference Switches41
Cone drive TL, reference switch S8 and potentiometer r80
•Connect an oscilloscope to D5.X2.a21 (for S8) (0 V to X30); 5 V/div., 0.1 s/div.
•Switch the system on.
•Move the shock wave head into the therapy position.
•Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30) ADC Value = 102
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
•Connect the DVM to D5.X2.a22 (0 V at X30).
•Check the function of potentiometer r8.
•Perform the check in normal mode.
•Check the potentiometer voltage at reference switch S8 = "On". U
at D5.X22 and 0 V at D5.X30; ADC value = 102
= 512 mV ± 5 mV
ref
•Check the switching hysteresis of reference switch S8 according to (Fig.7/p.41) in
"Monitoring".
Fig. 7:Switching hysteresis S8
Checking the toothed belt0
•Check the toothed belt according to Service Instructions Shock Wave System
RXL2-120.841.01.
Checking the travel range0
•The shock wave head is in the therapy position.
•The C-arm is in the + 30° fluoroscopy position.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 41 of 72
Page 42
42Potentiometers and Reference Switches
X•Release fluoroscopy FL.
¹ The cone drive moves to the front end position. Safety limit switch S40 must not
be activated.
¹ The service PC (Diagnostic, Monitoring) display indicates: ADC value = MC
position 600 +/- 1 bit.
•Release a shock wave.
¹ The cone drive travels back to MCESWL position 470 ± 1 bit.
•Uncouple the W drive.
•The cone longitudinal drive travels
- Cone drive serial no. < 2000 to the MC position 300 ± 1 bit
- Cone drive serial no. > 2000 to the MC position 360 ± 1 bit
Checking the travel range with US probe (optional)0
Prerequisite:- US probe inserted- Shock wave head in therapy position
•Use the cone longitudinal forward button on the US tableside control panel to move the
cone drive to the front end position.
¹ The LED on the US tableside control panel lights up.
¹ The service PC (Diagnostic, Monitoring) display indicates: ADC value = MC
position 600 ± 1 bit.
•Release shock waves.
¹ The cone drive travels back to MCESWL position 470 ± 1 bit.
•Use the cone longitudinal back button to move the cone drive to the back end position.
¹ The LED on the US tableside control panel lights up.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 42 of 72
Medical Solutions
Page 43
Potentiometers and Reference Switches43
Cone drive TR, reference switch S 9 and potentiometer r90
NOTE
Only perform if the ultrasound option is available.
•Connect the oscilloscope to D5.X2.a23 (for S9) (0 V to X30); 5 V/div, 0.1 s/div.
•Switch the system on.
•Move the shock wave head into the therapy position.
•Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30)
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
•Connect a DVM to D5.X2.a24 (0 V at X30).
•Check the function of potentiometer r9.
Prerequisite: US probe not inserted.
•Move to the reference position in Normal mode using the tableside control panel.
Use the "Cone rotation right or left" key for this.
¹ The cone rotation drive travels to the 0° position and stops.
¹ The 0° marks on the toothed ring and the lifting cylinder line up.
(If in "unregulated motion" mode, readjust the cone rotation )
¹ The position is displayed on the US tableside control panel.
•When reference switch S9 is actuated, the DVM should display a reading of 2.560 V ± 2
mV. ADC value = 512
•Check the switching hysteresis of the reference switch S9 according to (Fig.8/p.43).
Fig. 8:Switching hysteresis S9
Setting the parameters0
Refer to the "Parameter Groups" section of the "Service software, Software parameter /
Monitoring" instructions RXL2-120.032.02..
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 43 of 72
Page 44
44Potentiometers and Reference Switches
Checking the travel range0
Prerequisite: US probe is inserted.
•Move the US probe +90° and -90° by pressing the "Cone rotation left or right" key.
•Check the angular position on the US tableside control panel display.
•Safety limit switch S42 must not be actuated in the ± 90° setting.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 44 of 72
Medical Solutions
Page 45
Potentiometers and Reference Switches45
Longitudinal system Y drive0
Adjusting belt tension
Fig. 9:Measurement_1
Fig. 10: Measurement_2
•Move the unit left to the end stop.
•The distance from the center of roller (1/Fig.9/p.45) to the end stop (1/Fig. 10 / p. 45)
is 420 ± 5mm.
•Attach the spring balance (3/Fig. 10 / p. 45) and lift it with a force of 20 N + 5 N.
•It must be possible to raise the belt 3 mm + 1 mm (2/Fig. 10 / p. 45).
•If the belt tension varies from this, readjust it.
•Move the orbital drive in inching mode to the mechanical indicator "+ 30°" on the ser-
vice PC (Diagnostic, Monitoring); and the C-arm to the ADC position in 1/10 mm display to "+ 3000".
Fig. 18: Adjusting S4
•Adjust reference switch S4 so that switching is assured within the stated switching hys-
teresis tolerance range (refer to (Fig.5/p.36) and (Fig. 18 / p. 51)).
•Reset the system on board D1 with S1.
•Move the drive over the reference switch position several times to check.
Replacing reference switch S4 or potentiometer r4
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 51 of 72
Page 52
52Potentiometers and Reference Switches
•Connect a DVM to D5.X13 (0 V an X30).
•Connect an oscilloscope to D5.X2.a7 (0 V toX40) ; 5 V/div, 0,1 s/div.
•Switch the system on.
•Bring the X drive and the Y drive into mid-range position.
•Bring the shock wave head (W drive) into the park position.
•Select "unregulated motion" mode for the O drive.
•Bring the arm into +30° position.
•Align the reference edge of the unit cover and the reference mark (arrow)
((Fig. 17 / p. 51)).
•Replace reference switch S4 (refer to (b) Reference switches / p. 36)).
•Check the potentiometer voltage for the reference switch S4 = On U
mV (D5.X13 0V = D5 X30) ADC value = 966
NOTICE
After replacement or for potentiometer deviations:
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 4.830 V ± 2
REF
= 5.120 V ±
ref
•If necessary, replace the potentiometer according to (a) Potentiometers / p. 35).
•Reference switch is adjusted and closed.
•Loosen the set screw on potentiometer R4.
•Set the potentiometer value to ADC value = 966 by turning the adjustment wheel
((E/Fig.4/p.35)).
•Tighten the set screw.
•Reset board D1 with button S1.
•Check the switching hysteresis of the reference switch S4 according to drawing
((Fig. 18 / p. 51)).
•Move to the +30° or -10° end position with the tableside control.
- The safety limit switches S70/S71 must not be activated.
- When the 0° position keys on the tableside control are activated, the orbital drive
moves into the vertical position (check the position indicated on the control monitor
and at the reference mark).
- Fluoroscopy is enabled in the + 30° / 0° / -10° positions.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 52 of 72
Medical Solutions
Page 53
Potentiometers and Reference Switches53
Shock head mount W system0
Adjusting belt tension
Fig. 19: Adjusting belt tension
•Move the W drive to the uppermost position.
•Switch the system off.
•Loosen the 4 fastening screws ((1/Fig. 19 / p. 53)) (two screws are not visible) and set a
value of 28 mm ((2/Fig. 19 / p. 53)) with the two adjustment screws ((3/Fig. 19 / p. 53)).
•Retighten the 4 mounting screws.
Adjusting reference switch S3 and potentiometer r3
•Switch the system on.
•Move the shock wave system into the therapy position.
•Attach the adjustment phantom (switch S28 has switched).
•Check the isocenter in the + 30 degree or - 10 degree position and, if necessary, adjust
according to the "Isocenter adjustment" instructions " RXL2-120.071.01....
(Perform only the work necessary for shock head replacement).
NOTICE
Isocenter
¹ The isocenter is the starting point for the adjustment of
the reference switch or of the actual value potentiometer.
•Adjust reference switch S3 so that switching is assured within the stated switching hys-
teresis tolerance range (refer to (Fig.5/p.36) and (Fig. 20 / p. 54)).
The shock wave head is located in the isocenter.
Replacing reference switch S3 and potentiometer r3
•Switch the system on.
•Bring the X drive and the Y drive into mid-range position (basic position)..
•Bring the orbital drive into +30°.
•Select "unregulated motion" mode for the W drive.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 53 of 72
Page 54
54Potentiometers and Reference Switches
•Bring the shock head into the therapy position.
•Attach the adjustment phantom and adjust the isocenter (refer to Isocenter Adjustment
RXL2-120.071.01...)
•Replace reference switch S3 (refer to (b) Reference switches / p. 36)).
•Check the potentiometer voltage for reference switch S3 = ON; Uref = 4.770 V ± 2 mV
(D5.X12 0V = D5.X30) ADC value = 954.
NOTICE
Fig. 20: Replacing S3
After replacement or for potentiometer deviations:
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ±
ref
•If necessary, replace the potentiometer according to (a) Potentiometers / p. 35).
•Reference switch is adjusted and closed.
•The shock wave system is in the adjusted isocenter.
•Loosen the set screw on potentiometer R3.
•Set the potentiometer value to ADC value = 954 by turning the adjustment wheel
((E/Fig.4/p.35)).
•Tighten the set screw.
•Reset board D1 with button S1.
•Remove the adjustment device (switch S28 is open).
•Check the switching hysteresis of reference switch S3 according to (Fig. 20 / p. 54).
•Move to the "Coupling or park position" end positions.
- The safety limit switches S69/S68 must not be activated.
•Check the isocenter according to "Isocenter adjustment" RXL2-120.071.01....
•Check the Litho park position S79 (release of the X/Y travel range for Uro mode).
•Move to the park position of the W drive; contact S79 must be closed. If necessary,
check "Switch ON" with the DVM on board D32 X41 PIN 3 and at ground GND.
•Install the URO insert.
•Move to the basic →0← position.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 54 of 72
Medical Solutions
Page 55
Potentiometers and Reference Switches55
•The system may now be moved outside the "Litho Position". (X = -130 / + 70; Y = ± 320)
•Safety limit switches S 76/77 as well as S 78 must not be activated.
(Refer to circuit diagram J1042-7).
•Remove the Uro insert and move to the →0← basic position.
The system will only allow travel in the "Litho Position".
•Move the W drive out of the park position.
Switch S79 must open in the MC position - 430 ± 2 mm and switches S 76; S77; S78
must be activated. If necessary, check "Switch ON" with the DVM on board D32 X41
PIN 3 and at ground GND.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 55 of 72
Page 56
56Checking / Replacing the Lifting Columns R / L
5Checking / Replacing the Lifting Columns R / L
6-
Checking potentiometers r1/r2 and reference switches S1 and
S25
•Connect an oscilloscope to D5.X2.a1/a3 (for S1/2) (0 V at X30); 5 V/div, 0.1s/div.
•Switch the system on.
•Install the Uro insert.
•Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30)
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
•Connect a DVM to D5.X2.a2/a4 (0 V at X30).
•Check the functions of potentiometers R1/2.
•Perform the check in normal mode.
•Adjust the table height with the tableside control (NBF).
•Check the potentiometer voltage at reference switches S1/2 = "On" U
mV at D5.X10/X11 and 0 V at D5.X30. ADC value = 742
= 3.710 V ± 2
ref
0
•Check the switching hysteresis of the reference switch S1/2 according to
(Fig. 21 / p. 56).
Fig. 21: Switching hysteresis S1_S2
NOTE
It is not possible to adjust potentiometers r1/r2 or reference
switches S1/S2. The corresponding lifting column must be replaced if there are defects or malfunctions.
•Move the system into the basic position→0← with the tableside control.
•The position indicator on the control monitor will display the value "1033". Check this
with the service PC (MC position) in 1/10 mm (display "10330" / control monitor Z =
1033).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 56 of 72
Medical Solutions
Page 57
Checking / Replacing the Lifting Columns R / L57
Fig. 22: Lifting column
Preliminary
remarks:
NOTICE
Fig. 23: Removing cover
To avoid replacing the lifting columns unnecessarily, troubleshoot
thoroughly and localize errors accurately.
Removing the lifting column
¹ When removing the lifting column, use a support
(wooden plank) approx. 80 cm to 95 cm high and with a
load capacity of approx. 40 kg to support the tabletop.
The engineer must obtain this item locally.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 57 of 72
Page 58
58Checking / Replacing the Lifting Columns R / L
Preparation0
•Switch on the system and install the Uro insert.
•Loosen the four cover screws of the front, gray base frame cover and pull the plate out
to the side (refer to (Fig. 23 / p. 57)).
•Move the unit into the left or right end position. Disconnect the L/R cables of the lifting
columns from the cable harness in the base frame.
•Move the tabletop to the max. height.
•Remove the table extensions.
•Remove all covers of the lifting column affected ((Fig. 22 / p. 57))
•Remove the holding brackets of the covers by removing the screws.
•Unscrew the holder of the energy chain on the left lifting column.
•To remove the tabletop, push the support (wood plank) between the base frame and ta-
ble.
•Slowly move the table down onto this support until the weight is taken off the column.
NOTICE
Lifting column
¹ Do not retract the lifting column any further after lower-
ing the table onto the support.
•Switch the system off.
•Remove the 4 M8 Allen screws at the top of the column being replaced.
•Disconnect the plug for the power supply to the motor on the column that is not being
replaced.
•Switch the system on.
•Select "unregulated motion" with System-Logon-Adjustment-Calibration " as per the
"Operating the Service Software" instructions RXL2-120.113.01....
•Select the lifting column drive ZR or ZL.
NOTICE
Danger of collision!
¹ The system is now in "unregulated operation". The
movement sequences and the unit movements are no
longer monitored. Use caution to prevent collision hazards.
•Retract the column using the "Table down" button on the tableside control panel.
•Switch the system off.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 58 of 72
Medical Solutions
Page 59
Checking / Replacing the Lifting Columns R / L59
Removing the lifting column0
•Disconnect the plug from the actual value acquisition of the lifting columns (board D31
X10/ X11).
•Disconnect the motor power supply plug.
•Unlock and remove the motor encoder plug.
Fig. 24: Allen_screws_A
•Turn the lifting column drive shaft by hand until the Allen screws of the drive shaft are
accessible from above ((Fig. 24 / p. 59)).
•Remove the 2.5 mm Allen set screws ((1/Fig. 24 / p. 59) and (1/Fig. 25 / p. 59)).
If necessary, remove "Loctite" using a heat gun.
•Remove the mounting screws on the base of the lifting column (Fig. 4) and remove the
column.
NOTE
•Remove the drive shaft bearing to replace the cables ((Fig. 26 / p. 60)).
•Remove the strain relief on the two cables for the lifting column.
•Remove the mounting screws on the base of the lifting column ((Fig. 24 / p. 59)) and re-
move the column.
Fig. 25: Allen_screws_B
Ensure that the encoder on the motor shaft does not become damaged.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 59 of 72
Page 60
60Checking / Replacing the Lifting Columns R / L
Installing the new lifting column0
•Insert the connection cable of the new lifting column. Ensure that the leads are not twist-
ed.
•Position the column on the base plate.
•Apply Loctite 221 to the screws (Allen M8 x 40). Insert them and tighten them with a
torque of 24 Nm ± 10%.
•Insert the cable in the bearing guide track of the drive shaft and tighten it gently.
•Secure the strain relief with cable ties.
•Reinstall the drive shaft of the lifting column.
Ensure that the screw engages on the flattened side of the drive shaft.
Fig. 26: Drive axle_A bearing
Fig. 27: Drive axle_B_bearing
•Reattach the drive shaft bearing with M10x20 screws. Torque: 50 Nm ± 10%
•Connect the drive shaft with the PVC coupling (Allen screws 2.5 mm) and bolt the mo-
tor down ((Fig. 25 / p. 59)).
•Reconnect the actual value cable to board D31 X 10 or X11.
•Reconnect the power supply and the motor encoder cable.
•Switch the system on.
•Select "unregulated motion" with System-Logon-Adjustment-Calibration " as per the
"Operating the Service Software" instructions RXL2-120.113.01....
•Select the lifting column drive ZR or ZL.
•Raise the column in unregulated mode until it just makes contact with the tabletop.
•Apply Loctite 221 to the four Allen screws (M8) and bolt the column to the tabletop from
below.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 60 of 72
Medical Solutions
Page 61
Checking / Replacing the Lifting Columns R / L61
•Switch the system off.
•Install the holding brackets for the covers and the chain holder.
•Reconnect the power supply for the second column.
•Switch the system on.
•Raise the table and remove the tabletop support from the travel range.
•Check the function of the lifting columns.
If necessary, check the "Table tilt" option.
•Reinstall all the covers of the lifting column that was just replaced. Reconnect the pro-
tective conductor.
•Reattach the two-part plastic cover at the base of the lifting column. Seal the edges of
the plastic cover against the unit or floor with silicon.
•Reconnect the cables to the cable harness in the base frame.
•Close the covers of the unit base frame.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 61 of 72
Page 62
62Working on the Motor Drives
6Working on the Motor Drives
7-
The following
general issues
apply:
NOTICE
No special safety measures are required when working on the X/Y
drives.
Working on the orbital / W drive
¹ When working on the orbital drive, secure the C-arm
against uncontrolled movement. When working on the W
drive, secure the bearing block from backing up into the
mechanical stop.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 62 of 72
Medical Solutions
Page 63
Working on the Orbital Drive63
7Working on the Orbital Drive
8-
Removing the am4 drive motor0
Fig. 28: Covers_A
Fig. 30: Retaining rails
•Move the shock wave head to the therapy position and the C-arm to the 30° position.
•Remove the covers ((Fig. 28 / p. 63)) from the basic system and the X-ray tube.
Fig. 29: Covers_B
Tools required:
- A tensioning belt or similar device
approximately 4.5 m long and 25mm
wide must be purchased locally.
- 5mm Allen wrench with guide pin
•Remove the side and lower C-arm covers ((Fig. 29 / p. 63)).
•Remove the retaining rails ((1/Fig. 30 / p. 63)) for the cover plate.
•Switch the system on.
•Select "unregulated mode" on the service PC.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 63 of 72
Page 64
64Working on the Orbital Drive
Fig. 31: D7 board
Fig. 33: Potentiometer r4
Fig. 32: Board D7_tilted
Fig. 34: Rubber end stop
•Move the O drive to the mechanical end stop, passing the 30° position, in unregulated
mode so that the rubber end stops are against the mechanical end stop
((Fig. 34 / p. 64)).
•Switch the system off.
•Unplug the motor connector ((A/Fig. 32 / p. 64)).
•Remove board D7 ((2/Fig. 31 / p. 64)).
•Loosen the screws for the D7 board holder ((3/Fig. 31 / p. 64)) and tilt it forward
((5/Fig. 32 / p. 64)).
•Loosen the screw ((4/Fig. 31 / p. 64)) at the energy chain guide.
•Lift up the energy chain guide and secure it above ((6/Fig. 32 / p. 64)).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 64 of 72
Medical Solutions
Page 65
Working on the Orbital Drive65
•Position the tools as indicated in the figure ((7/Fig. 32 / p. 64)) and loosen the screws,
but do not remove them.
•Loosen the set screw M (1.5 Allen) at potentiometer r4 ((9/Fig. 33 / p. 64)) on the left
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 65 of 72
Page 66
66Working on the Orbital Drive
Replacing the toothed belt0
(The am4 drive motor must have been removed).
NOTICE
Risk of accidents!
¹ Two persons are always required when working on the
C-arm. One person must press against the X-ray tube
while the other tightens or loosens the tensioning belt.
•Remove the I.I. collision protection.
•Attach the tensioning belt ((10/Fig. 35 / p. 65) and (10/Fig. 37 / p. 65)) on one of the
sides; route the tensioning belt over the upper concealed roller(11/Fig. 35 / p. 65).
•Lift up the C-arm using the tensioning belt until the cable guides ((12/Fig. 37 / p. 65))
are accessible.
•For both the left and right cable guides:
- Loosen the three screws ((13/Fig. 37 / p. 65)) and remove the cover plate.
- Remove the rubber end stop ((14/Fig. 37 / p. 65)); the attachment screw (Allen 6
mm) is accessible from above, through the cable harness.
- Remove the two screws ((15/Fig. 37 / p. 65)).
- After removing the 4 attachment screws from above ((16/Fig. 37 / p. 65)) and below
(in (Fig. 37 / p. 65) concealed), push the holder downward as far as possible.
Fig. 38: Attachment plates
Fig. 39: X-ray tube assembly holder
•Move the C-arm outward as far as possible, passing the -10° position, using the ten-
sioning belt until the attachment plates on the back ((17/Fig. 38 / p. 66)) can be re-
moved.
•Loosen and remove the attachment plates using a 5 mm Allen key with guide pins.
•Carefully lower the C-arm to the 30° position; the end stop has been removed.
•Remove both upper screws from the X-ray tube holder ((18/Fig. 39 / p. 66)).
•Remove both tensioning screws ((19/Fig. 39 / p. 66) and (19/Fig. 40 / p. 67)).
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 66 of 72
Medical Solutions
Page 67
Working on the Orbital Drive67
•Remove the 5 screws ((20/Fig. 39 / p. 66)).
•Pull the clamp with the belts downward and out. Note the original position to facilitate in-
stalling the new belt later.
•Open the clamp and remove both old toothed belts.
•Prepare both new belts by:
- Cutting them to the length of the old belts.
- Drilling one side of the toothed belt,
Using the old belt as a guide.
- Inserting them into the clamp;
ensuring that the holes of the toothed belt are not inside the clamp.
•Position the clamp for securing the toothed belt with the threaded holes facing down-
ward between the X-ray tube flange and insert the holder on the C-arm ((Fig. 39 / p. 66)
and (Fig. 40 / p. 67)).
•Route both toothed belts into the C-arm from below.
•Insert the 5 screws ((20/Fig. 39 / p. 66)) into the clamp and tighten only slightly.
•Insert the tensioning screws ((19/Fig. 39 / p. 66) and (19/Fig. 40 / p. 67)) into the clamp
from below and tighten only slightly.
Fig. 40: Toothed belt placement
•Re-tighten both upper screws of the X-ray tube holder ((18/Fig. 39 / p. 66)).
•Raise the C-arm, passing the -10° position, to make the attachment holes visible.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 67 of 72
Page 68
68Working on the Orbital Drive
•Route the toothed belt according to (Fig. 40 / p. 67).
•Pull the toothed belt upward and out of the C-arm.
•Insert the screws first through the attachment plate ((17/Fig. 38 / p. 66)), then through
the toothed belt and finally tighten them in the C-arm, securing them with Loctite. Use a
5 mm Allen wrench with guide pin to tighten them.
NOTICE
Ensure that no steel cables are lying in the path of the screws.
¹
•Lower the C-arm far enough and hold it with the tensioning belt until both cable guides
and rubber end stops ((Fig. 37 / p. 65)) can be reassembled in reverse order.
•Lower the C-arm to the 30° end stop and remove the tensioning belt.
•Move the C-arm through its entire range at least four times to align the belt.
•Tighten the toothed belt with both tensioning screws ((19/Fig. 39 / p. 66) and
(19/Fig. 40 / p. 67)).
•Tighten the 5 screws ((20/Fig. 39 / p. 66)).
•Move the C-arm through its entire travel range again and check whether the alignment
is still correct.
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 68 of 72
Medical Solutions
Page 69
Working on the Orbital Drive69
Installing the am4 drive motor0
•Install the drive motor by reversing the order in which it was removed, as described in
the section "Removing the am4 drive motor".
•Install the holder for board D7.
•Insert board D7.
•Plug in the motor connector.
•Install the holder for the energy chain guide.
•Install the I.I. collision protection.
•Switch the system on.
•Move the C-arm out of the safety circuit in unregulated mode.
•Adjust potentiometer r4 as described in Chapter 4.
•Move the C-arm within the range from -10° to +30° and check whether the toothed
gears of the drive motor make any noise.
•Re-install all covers.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 69 of 72
Page 70
70Appendix
8Appendix
9-
Fig. 41: D1
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 70 of 72
Medical Solutions
Page 71
Changes to Previous Version71
9Changes to Previous Version
10-
No changes; initial version.
Siemens AGRXL2-120.841.02.01.02LITHOSTAR Multiline
Medical Solutions
09.05CS PS 24
Page 71 of 72
Page 72
72Changes to Previous Version
LITHOSTAR MultilineRXL2-120.841.02.01.02Siemens AG
09.05CS PS 24
Page 72 of 72
Medical Solutions
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.