Siemens Lithostar Multiline Replacements of Parts

Page 1
Erlwein
LITHOSTAR Multiline
Replacements of Parts
System
SP
Unit not including shock wave system
03127011
Print No.: Replaces: n.a.
RXL2-120.841.02.01.02
© Siemens AG
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
English Doc. Gen. Date: 09.05
2005
Page 2

2 Revision / Disclaimer

1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 2 of 72
Medical Solutions
Page 3
Table of Contents 3
1- 0Table of Contents
1 _______ Prerequisites ___________________________________________________ 5
Documents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Required measurement and adjusting devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Disposal information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Service information on the Intranet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2 _______ Overview of Measures___________________________________________ 10
Component N11 (in control cabinet A12) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Collimator component N. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
VIDEOMED DI-M component. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Memoskop component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Multispot component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Laser light localizer component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Unit/control console component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Generator component. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ultrasound option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3 _______ PC Board Replacement __________________________________________ 21
Component N 11 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Board D1, (new type) host computer or when replacing the battery. . . . . . . . . . . . . 21
Board D1, (old type) host computer or when replacing the battery. . . . . . . . . . . . . . 22
Board D4xx, Motor controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Board D 5, A/D converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
SMP bus board D10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Board D11, backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Board D 21, power supply M16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Board D4, power supply, respiratory and ECG triggering. . . . . . . . . . . . . . . . . . . . . 27
Collimator N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Replacing the collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Replacing electric components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Adjustment test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Control position of the collimator blades. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
VIDEOMED DI-M component. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Board D1, analog board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Board D3, analog board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Unit/control console component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Board D32, distributor board, supporting frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
I.I. supply Z96 "Röderstein" . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Tableside control panel (NBF), board D41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Recoding for use on the control console, board D41 . . . . . . . . . . . . . . . . . . . . . . . . 32
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
System unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Board D51, ADC/DUEP plug-in board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 3 of 72
Page 4
4 Table of Contents
4 _______ Potentiometers and Reference Switches ___________________________ 34
Overview of the potentiometers and reference switches . . . . . . . . . . . . . . . . . . . . . . . . 34
Basic information on the replacement and adjustment of potentiometers, reference switch-
es and motor encoders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
General replacement and adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Checking, replacement and/or adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Cone drive TL, reference switch S8 and potentiometer r8 . . . . . . . . . . . . . . . . . . . . . . . 41
Checking the toothed belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Checking the travel range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Checking the travel range with US probe (optional) . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cone drive TR, reference switch S 9 and potentiometer r9 . . . . . . . . . . . . . . . . . . . . . . 43
Setting the parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Checking the travel range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Longitudinal system Y drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Transverse system, X drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
C-arm orbital drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Shock head mount W system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5 _______ Checking / Replacing the Lifting Columns R / L ______________________ 56
Checking potentiometers r1/r2 and reference switches S1 and S25 . . . . . . . . . . . . . . . 56
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Removing the lifting column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Installing the new lifting column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6 _______ Working on the Motor Drives _____________________________________ 62
7 _______ Working on the Orbital Drive _____________________________________ 63
Removing the am4 drive motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Replacing the toothed belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Installing the am4 drive motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8 _______ Appendix______________________________________________________ 70
9 _______ Changes to Previous Version _____________________________________ 71
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 4 of 72
Medical Solutions
Page 5

Prerequisites 5

1Prerequisites
2-
Observe the EMC meaures for systems with serial number starting with 06022 (EMC =
electromagnetic compatibility).
Observe ESD protective measures (ESD = electrostatically sensitive device).
Ensure that the EMC shielding is reinstalled correctly after removing it to replace or
check boards.
Check the connection pins when inserting boards and connectors.
Remove the system covers for the area of the system being serviced and reinstall them
after the service work is completed.
NOTICE
Protective conductor test and backup:
¹ When service work is completed, perform a protective
conductor test and a backup of the new parameters.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 5 of 72
Page 6
6 Prerequisites

Documents required 0

SAFETY NOTES RA0-000.012.29...
LITHOSTAR Multiline Startup Instructions RXL2-120.034.01...
LITHOSTAR Multiline wiring diagram set J1042
Operation of the service software RXL2-120.113.01...
Setting the software parameters RXL2-120.032.01...
Jumper log 31 27 011 EBP 01S
Start-up Instructions RXL2-120.034.01...
MEMOSKOP Service Instructions RX57-029.061.01...
VIDEOMED DI/DI-M Service Instructions RX52-022.061.01...
MULTISPOT 2000 Setting Instructions RX57-091.032.01...
POLYDOROS LX 30/50 Troubleshooting Instructions RX63-020.072.01...
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 6 of 72
Medical Solutions
Page 7
Prerequisites 7

Required measurement and adjusting devices 0

Special tool, PLCC extractor 99 00 549 LE999
96-pin extension board, short 85 73 628 G5206
2x96-pin extension board 73 94 943 EE999
1x64-pin extension board
(This board can also be used on the 48-pin connectors)
21 67 179 N2030
Service PC (for details see Intranet service laptop for CSEs)
Serial PC connection cable, 5m 99 00 440 RE999
Multimeter (Class 0.1%; Ri > 10 M), e. g., Fluke 8060 A 97 02 101 Y4290
Oscilloscope, e.g., Fluke Scope Meter 73 92 074
ESD tools according to ARTD, Part 3
Standard service equipment kit
A stretching belt or similar device (approximately 4.5 m long and
25 mm wide), must be purchased locally.
5 mm Allen wrench with guide pin
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 7 of 72
Page 8
8 Prerequisites

Disposal information 0

When disposing of boards and components no longer required, adhere to the Disposal
Instructions RXL2-120.163.01... .
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 8 of 72
Medical Solutions
Page 9
Prerequisites 9

Service information on the Intranet 0

The most recent information is available on the Intranet.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 9 of 72
Page 10

10 Overview of Measures

2Overview of Measures
3-

Component N11 (in control cabinet A12) 0

BoardPart no. Name Remarks regarding replacement
- Tool tip
Fuses Replace according to wiring diagram J1042-3A n.a.
D1 71 28 841
J1042
Host com­puter
Check switches according to wiring diagram excerpt in Appendix. (Appendix / p. 70)
Wiring diagram 1042-21-4
Set time, restore the parameters, download software
30 82 138 Battery Set time, restore parameters
D1 31 25 825
J1042
Host com­puter
Activate battery: Solder in jumper X26-X27; check jumpers/switches; jumper log
Wiring diagram J1042-21
Replace Proms, set time, restore the parameters, download software
31 46 073
Battery Set time, restore parameters
B2905
Detail info
(Board D1,
(new type) host com­puter or when replac­ing the battery/p.2
1).
(Board D1, (old type) host com­puter or when replac­ing the battery/p.2
2)
D2 31 25 643
J1042
D4 89 49 588
G5334
D5 31 25 528
J1042
D5 37 82 714
J1042
XCS
controller
Motor
controller
A/D con­verter
A/D
converter
Check jumpers/switches; replace Proms; jumper log, wiring diagram J1042-22
Change EMC shielding plate, if present, from old to new board
Check jumpers/switches;
replace Proms
Check jumpers; replace Proms
Check U
, setting on D8 with R220 (Board D 5,
ref
- Board backward-compatible
(starting with HOST Software VA00D)
- May only be oper­ated with Prom:
- J1 Part no. 47 81 857 J1042
- J2 Part no. 47 81 865 J1042
n.a.
(Board D4xx, Motor controller / p.
24)
A/D converter / p.
24)
Check jumpers; replace Prom; check U
setting on
ref
D8 with R220
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 10 of 72
Medical Solutions
Page 11
Overview of Measures 11
BoardPart no. Name Remarks regarding replacement
- Tool tip
D6 31 25 569
J1042
COM
controller
Check jumpers/switches; replace Proms; jumper log, wiring diagram J1042-26
Change EMC shielding plate, if present, from old to new board
D7 89 49 505
G5334
Motor
output
No special attention necessary n.a.
stages
D8 31 73 911
B3305
+5V power supply
Set the reference voltage U R220; test point U
: D5.X31 (0V; X30)
ref
= 5.120V +/- 5 mV with
ref
Prior to installing the board, set potentiometer R220 to the lowest value
D10 11 14 839
G5347
Bus board starting with serial no. 7000: Transfer the jumper from
the old board to the new board. Jumper log
X112.4 - X110.4. The plug must connect 1-5
D11 31 25 692
J1042
Backplane Check the fuses/jumpers
Jumper log, wiring diagram J1042-31/
Detail info
n.a.
n.a.
(SMP bus board D10 / p. 25)
(Board D11, backplane / p. 26)
D21 31 25 577
J1042
D25 88 21 068
G1005
D26 88 21 068
G1005
D4 84 37 337
K2013
Power supply M16
ECG/QRS
separation
Dig. mem­ory
Respira­tion Ana­log Gate
Power supply D25/D26
Check the fuses,
wiring diagram J1042-6A/6B
Check the fuse,
wiring diagram J1042-45 jumper
Note: Do not change the potentiometer setting!
Check the jumper,
jumper log, wiring diagram J1042-46
Note: Do not change the potentiometer setting!
Check fuses F1/F2,
wiring diagram J1042-44
(Board D 21, power supply M16 / p. 26)
n.a.
n.a.
(Board D4, power sup­ply, respira­tory and ECG triggering / p.
27)
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 11 of 72
Page 12
12 Overview of Measures

Collimator component N 0

Collimator N is maintenance-free and should be replaced in the event of a mechanical
malfunction. An adjustment test must be performed following replacement ((Replacing
electric components / p. 28)).
In case of electronics errors, parts will be replaced according to spare parts list
RX74-020.081.10....
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 12 of 72
Medical Solutions
Page 13
Overview of Measures 13

VIDEOMED DI-M component 0

When servicing or replacing the VIDEOMED DI-M, adhere to the Service Instructions RX52-022.061.01...
Board Part no. Name Remarks regarding replacement / tool tips Info
D1 30 97 727
X2170
D3 30 97 651
X2170
Analog board Remove the resistor between X902 and X903,
refer to the service note "Adapting a
MEMOSKOP"
Digital
control
board
"Load Default Value" is possible the first time only with the switch, not with the service software.
Select the LITHOSTAR measurement field;
perform a dose rate check;
perform an IQ test
For 33 cm I.I. only:
select horizontal and vertical image reversal
(Board D1, ana­log board / p.
30)
n.a.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 13 of 72
Page 14
14 Overview of Measures

Memoskop component 0

Use Service Instructions RX57-029.061.01... when working on the Memoskop.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 14 of 72
Medical Solutions
Page 15
Overview of Measures 15

Multispot component 0

Use the Setting Instructions RX57-091.032.01... when working on the Multispot 2000.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 15 of 72
Page 16
16 Overview of Measures

Laser light localizer component 0

The complete laser light localizer unit is replaced. It can be converted from battery opera­tion to rechargeable battery pack operation (refer to Instructions RXL2-120.033.03...).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 16 of 72
Medical Solutions
Page 17
Overview of Measures 17

Unit/control console component 0

Board Part no. Name Remarks regarding replacement - tool
tips
D7 ... 89 49 505 G5334 Motor output
stage
D31 31 25 585 J1042 Base frame,
distributor board
D32 31 25 676 J1042 Support frame,
distributor board
Z96 31 36 194 X2134 I.I. power sup-
ply,
operator panel
No special attention necessary n.a.
Check jumper log, wiring diagram J1042-51
No special attention
necessary
Check fuses F2 and F3; wiring diagram J1042-6
Detail info
n.a.
(Board D32, dis­tributor board, support­ing frame / p.
31)
(I.I. sup­ply Z96 "Röderst ein" / p. 3
1)
n.a. 31 23 879 J1042 Tableside con-
trol
D51 31 25 627 J1042 ADC/DUEP
plugin board in the PC
r3-6, 8 30 28 009 B0611 Potentiometer
Completely replaced!
check the switch,
jumper log, wiring diagram J1042-61
Check jumpers, switches;
replace Prom;
jumper log, wiring diagram J1042-71
n.a.
(Table­side con­trol panel (NBF), board D41 / p. 3
2)
(Board D51, ADC/DU EP plug-in board / p.
33)
(a) Potentio meters / p. 35)
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 17 of 72
Page 18
18 Overview of Measures
Board Part no. Name Remarks regarding replacement - tool
tips
am 1/2 70 54 422 F0721 Lifting column
n.a.
n.a. 31 23 994 J1042 Shock wave
release
n.a. 47 82 707 J1042 System floor
Replace completely, no special attention necessary
Replace completely; n.a.
switch
Detail info
(Check­ing / Replac­ing the Lifting Columns R / L/p.56)
n.a.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 18 of 72
Medical Solutions
Page 19
Overview of Measures 19

Generator component 0

The following instructions should be used when working on the generator.
LITHOSTAR Multiline Serial no. 6073 RX63-020.034.12...
Serial no. > 7000 POLYDOROS SX RXL2-120.034.05...
Serial no. > 7000 POLYDOROS LX RXL2-120.034.02...
Troubleshooting Instructions
Troubleshooting Instructions
When replacing the image intensifier, X-ray tube or camera, use the Start-up instructions for the specific component. In addition, adjust the system in accordance with the adjust-
ment instructions, "Isocenter Adjustment" RXL2-120.071.01....
POLYDOROS LX RX63-020.072.01...
POLYDOROS SX RX63-055.072.01...
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 19 of 72
Page 20
20 Overview of Measures

Ultrasound option 0

Use Start-up Instructions RXL2-120.091.10.01... when working on the ultrasound system.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 20 of 72
Medical Solutions
Page 21

PC Board Replacement 21

3PC Board Replacement
4-

Component N 11 0

NOTE
Refer to wiring diagram J1042-3A for the rated current values of the different fuses listed in this chapter.

Board D1, (new type) host computer or when replacing the battery 0

78 28 841 J 1042, refer to wiring diagram J1042-21-4
30 82 138 battery 3.0 V
Follow items A + B + D when replacing board D1.
Follow items A + C + D when replacing the battery.
If possible, perform a backup of all current parameters and data.
A
(Refer to Operation of the Service Software RXL2-120.113.01. Chap. 3-3)
Read out the counter values for
shock counter generator shock counter shockwave
shock counter spark-gap shock counter total
shock counter US probe (starting with Rev. 05)
and document them in the operating log (Logbook, Reg. 8).
Remove board D1 from electronics rack N11.
B
Change EMC shielding plate, if present, from old to new board.
Set or check the switches on the new board.
Switch settings, see wiring diagram excerpt in the Appendix to this manual.
C
Measure the battery voltage; U
> 2.7 V is normal.
batt
If the battery voltage is < 2.7 V, replace the battery.
Desolder the old battery and replace it with a new one.
D
Measure the battery voltage; U
> 2.7 V is normal.
batt
If the battery voltage is < 2.7 V, replace the battery.
Insert board D1.
After switching on the system:
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 21 of 72
Page 22
22 PC Board Replacement
- Establish a connection between the service PC and D1 X10
- Start the service software (Refer to the service software operating instructions RXL2-120.113...)
- Enter the current time and date "Realtime Clock"
- Execute a download of "Host Software, Parameter file" in the Data menu
- Execute a restore of the "Customer Data", "System Data" parameters in the Data menu
- Call up "Technical Info" and "Service Data" in the Info menu with "Load from File" and save by selecting "Put to Unit".
- Call up "Parameters" in the Adjustment menu
- Select "Interface Shockwavesystem" from the Parameter group
- The value 00 must appear in the "Value" column for: Reset shock counter generator / Reset shock counter shockwave Reset shock counter spark-gap / Reset shock counter total Reset shock counter US probe (starting with Rev. 05)
If necessary, reset the values in the Value column
(Reset with "FF"/normal operation with "00", refer to "Setting the software parame­ters" RXL2-120.032.01..)
NOTE
The Appendix of this manual contains a layout of the new D1 board. When board D1 is replaced, this layout is to be attached to the LITHOSTAR Multiline wiring diagram.

Board D1, (old type) host computer or when replacing the battery 0

31 25,825 J 1042, refer to wiring diagram J1042-21-4
31 46 073 B 2905 battery 3.6 V
Follow items A + B + D when replacing board D1.
Follow items A + C + D when replacing the battery.
If possible, perform a backup of all current parameters and data.
A
(Refer to Operation of the Service Software RXL2-120.113.01. Chap. 3-3)
Read out the counter values for
shock counter generator shock counter shockwave
shock counter spark-gap shock counter total
shock counter US probe (starting with Rev. 05)
and document them in the operating log (Logbook, Reg. 8).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 22 of 72
Medical Solutions
Page 23
PC Board Replacement 23
Remove board D1 from electronics rack N11.
B
Change EMC shielding plate, if present, from old to new board.
Set or check the switches on the new board.
Replace Proms J66/J67.
Solder in jumper X26 - X27 to activate the battery.
C
Measure the battery voltage; U
If the battery voltage is < 3.0 V, replace the battery.
Desolder the old battery and replace it with a new one.
D
Measure the battery voltage; U
> 3.0 V is normal.
batt
> 3.0 V is normal.
batt
If the battery voltage is < 3.0 V, replace the battery.
Insert board D1.
After switching on the system:
- Establish a connection between the service PC and D1 X10
- Start the service software (Refer to the Service Software Operating Instruc­tions RXL2-120.113...)
- Enter the current time and date "Realtime Clock"
- Execute a download of "Host Software, Parameter file" in the Data menu
- Execute a restore of the "Customer Data", "System Data" parameters in the Data menu
- Call up "Technical Info" and "Service Data" in the Info menu with "Load from File" and save by selecting "Put to Unit".
- Call up "Parameters" in the Adjustment menu
- Select "Interface Shockwavesystem" from the Parameter group
- The value 00 must appear in the "Value" column for: Reset shock counter generator / Reset shock counter shockwave Reset shock counter spark-gap / Reset shock counter total Reset shock counter US probe (starting with Rev. 05)
If necessary, reset the values in the Value column
(Reset with "FF"/normal operation with "00", refer to "Setting the software parame­ters" RXL2-120.032.01..)
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 23 of 72
Page 24
24 PC Board Replacement

Board D4xx, Motor controller 0

89 49 588 G 5334 Refer to jumper log wiring diagram J 1042-24
Set or check jumpers
Check switch S2 (S2 defines the board as D4A, B or C).
NOTICE
PAL modules
¹ The PAL modules J5, J31 and J32 on the replacement
board must be removed and destroyed. Only the original PAL modules of the old board D4 may be used.
Replace Prom J38.
Mark the board according to its location.
Insert the board and switch the system on.
Check the operating voltage at test point X10 +5 V +/- 250 mV (0V, X11).

Board D 5, A/D converter 0

31 25 528 J1042, Refer to jumper log, wiring diagram J 1042-25
37 83 714 J1042
NOTICE
Risk of short circuit! Do not use the 96-pin extension board 81 33 944 X1454 to extend the front plug D5.X2.
¹ Use only the short 96-pin extension board, part no. 85 73
628 G5206.
Remove the front plug X2; remove board D5.
Set or check jumpers.
Check the Proms J1/J2 and replace them, if necessary.
Insert the board and set front plug X2 to the short extension; switch on the system.
Check the reference voltage U
Mp X31 - X30 (0V). If necessary, set it with potentiometer R220 on board D8.
Checking the A/D converter
= 5.120 V ± 5 mV by connecting a DVM to board D5 at
ref
Measure and record the voltage value for every actual value potentiometer (r1 ... r6)
(Mp X10 ... X15).
Read out and record the value of the ADC value for the corresponding potentiometer
with the service PC in the Diagnostic, Monitoring menu.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 24 of 72
Medical Solutions
Page 25
PC Board Replacement 25
Ensure that the relevant voltage value corresponds to the ADC value.
Conversion:
Voltage value (mV) = ADC value (bit) x 5mV/bit -> +/ - 20 mV tolerance
Voltage value (mV))
ADC value (bit) =
Example:
Measured value at D5 M mV
Calculated value ADC value (bit) =
- Compare with recorded ADC value
-> the ADC value in bits for this channel is correct as long as the values do not deviate from one another by more than +/- 4 bits.
Check the calculation: 587 bit x 5 mV/bit = 2935 mV +/- 20 mV
___________________
5mV/bit
X = 2935
p
2935 mV
___________ =
5 bit/mV
-> +/ - 4 bit tolerance
= 587 bit
Switch off the system, remove the extension board and reconnect the front plug X2.
Switch the system on.
Check the operation of the various drives.

SMP bus board D10 0

11 14 839 G5347, wiring diagram J 1042-30
Remove the cable connections from module N11.
Remove the boards according to ESD guidelines and remove module N11 from the
cabinet.
Replace board D10.
Reinstall module N11 in reverse order.
When installing the EMC shielding plate, ensure that the cables fit in the recesses pro-
vided and are not damaged.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 25 of 72
Page 26
26 PC Board Replacement
Attach the exposed cable shielding with clamps.
Starting with serial no. 7001:
Transfer the jumper from the old board to the new board. X112.4 - X110.4
The plug must connect 1-5. (Refer to the jumper log).

Board D11, backplane 0

31 25 577 J1041, wiring diagram J 1042-31
Remove the cable connections from module N11.
Remove the boards according to ESD guidelines and remove module N11 from the
cabinet.
Replace board D11.
Reinstall module N11 in reverse order.
When installing the EMC shielding plate, ensure that the cables fit in the recesses pro-
vided and are not damaged.
Attach the exposed cable shielding with clamps.
Check or set the jumper positions according to the jumper log or wiring diagram.
Check the fuses F10...F16 according to the wiring diagram.

Board D 21, power supply M16 0

31 25 577 J1042, wiring diagram J 1042-41
Check fuses F1 and F2.
Replace the board.
Switch the system on.
Check LEDs V10... V16 and operating voltages.
Supply voltage of the magnetic brakes LED V10
MP X3.3/4 +25 V to +29 V regulated
Gnd X3.1/2
Supply voltage of function switches and safety circuits LED V11
MP X2.10 +23 V to +26 V regulated
Gnd X2.6
Supply voltage of the DUEP interface to the image memory LED V12
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 26 of 72
Medical Solutions
Page 27
PC Board Replacement 27
MP X2.1 +8.5 to +9.5 V regulated
Gnd X2.2
Supply voltage of the collimator LED V13
MP X4.4 +27 V to +35 V
Gnd X4.2
Supply voltage of the large drives (K1/K2 are energized) LED V14
MP X15 +28 V to +35 V
Gnd X14
Supply voltage of the small drives (K3 is energized) LED V15
MP X17 +28 V to +35 V
Gnd X16
Supply voltage of the Com controller (DUEP) LED V16
MP X2.9 +14 V to +15.5 V regulated
Gnd X2.6

Board D4, power supply, respiratory and ECG triggering 0

84 37 337 K2013, wiring diagram J 1042-44
Check fuses F1 and F2.
Measure the operating voltage (place the board on the extension)
+ 12 V ± 10% (X1.d/b/z4) LED V4 The - 5V are not used in the
- 12 V ± 10% (X1.d/b/z4) LED V11
GND (X1.d/b/z4)
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 27 of 72
LITHOSTAR Multiline
.
Page 28
28 PC Board Replacement

Collimator N 0

Fig. 1: Strain relief

Replacing the collimator 0

Unpack the collimator.
Remove the two colored screws from the transport bracket on the side of the blades
close to the focus.
Check the adjustment of the leaves close to the focus with the template.
Attach the collimator to the tube unit.
If plug X1 has a connector shell, unscrew it, cut it open, remove it.
Plug the X1 plug into the collimator and bolt it
Attach the cable at the panel using strain relief clamps ((1/Fig.1/p.28)).
Perform the adjustment test when you replace the entire collimator.

Replacing electric components 0

Switch off the system
Replace the electric components according to the spare parts list and perform the ad-
justment test.
D1/S2.5
OFF Collimator without prefiltration
ON Collimator with prefiltration
(On D1X7, check that the cable for the drive is connected)
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 28 of 72
Medical Solutions
Page 29
PC Board Replacement 29

Adjustment test 0

Set DIP switch S2.4 to "OFF".
Close the "height" and "width" collimator blades completely by hand.
Remove the prefilter (if present).
Move the iris blades to an opening of 70 x 70 mm (measure at the iris blades).
Switch the system back on or reinsert fuse F1 on D1.
Switch DIP switch S2.4 to "ON" for initialization.
The system will initialize. DIP switch S2.4 should remain "ON" during operation.
The height and width blades will open or close to the 44mm format (± 0.75%) x 44 mm
(± 0.75%) at a SID of 125 cm (35 mm x 35 mm at 100 cm SID).

Control position of the collimator blades 0

Install the cassette holder. Do not insert a film cassette.
Move the system into the mid-range position.
Release fluoroscopy.
A radiation field of comparable size
44 mm (± 0.75%) x 44 mm (± 0.75%) - is visible on the monitor at a SID of 125 cm.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 29 of 72
Page 30
30 PC Board Replacement

VIDEOMED DI-M component 0

Board D1, analog board 0

30 97 727 X2170 wiring diagram X 2170-10
Starting with revision level 01, the 620 ohm resistor between test points X902 and X903
should be removed.

Board D3, analog board 0

30 97 651 X2170 wiring diagram X 2170-13
Select the "Multiline" measurement field from the specific parameters (refer to the Ser-
vice Instructions RX52-022.061.01...).
Check the dose setting according to Start-up Instructions RXL2-120.034.01...).
Perform an IQ test
For 33 cm I.I. only:
Select "H-image reverse" and "V-image reverse" in the specific parameters and transfer them individually to the system (Refer to the Service Instructions RX52-022.061.01...).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 30 of 72
Medical Solutions
Page 31
PC Board Replacement 31

Unit/control console component 0

Board D32, distributor board, supporting frame 0

31 25 676 J1042, wiring diagram J1042-52/52A
Plug in and check the connection cables according to the wiring diagram.
Check connection (X56) and the operation of the tube assembly fan.
Check the operation of service switch S 58.
Check the operation LED’s V14 to V18 for the magnetic brake by moving the associat-
ed drive (The LED lights up when the brake is released).

I.I. supply Z96 "Röderstein" 0

31 36 194 X2134, wiring diagram X2134
Replace the board and the power transistor at the same time.
Check fuses F2 and F3.
Check the operating voltage 27 V ± 10% for V11 "ON"; Tp 27 V Tp 0.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 31 of 72
Page 32
32 PC Board Replacement

Tableside control panel (NBF), board D41 0

31 23 879 J1042, wiring diagram J1042-61
Board D41 is coded in the factory to "Patient table basic unit".

Recoding for use on the control console, board D41 0

Place the tableside control panel with the membrane keyboard side on a level surface.
Remove four screws (M 2.5x16) from the lower part of the housing.
Carefully lift off the lower housing.
Four metal spacers are lying loosely on the PC board.
Set or check the DUEP slave address of switch S3 according to the application and the
wiring diagram:
Address 7: Tableside control panel (NBF) basic unit (patient table) factory setting
Address 2: Tableside control panel (NBF) basic unit control console
Fig. 2: S3 Switch

Assembly 0

Center the four metal spacers on the contact surfaces and insert the four screws
(M2.5x16) in the holes in the lower part of the housing.
Join the housing parts together so that one of the screws can be inserted through the
hole of the grounding tape.
The metal spacers must not move during assembly.
Tighten the screws.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 32 of 72
Medical Solutions
Page 33
PC Board Replacement 33

System unit 0

30 10 175 JE002
Replace the system unit, if necessary, according to the Repair Instructions
RXL2-120.091.02... .

Board D51, ADC/DUEP plug-in board 0

31 25 627 J1042, wiring diagram J1042-71
Refer also to Repair Instructions "Replacement of the system unit" RXL2-120.091...
Set and check the jumpers and Proms according to the jumper or wiring diagram.
Check the 7-segment display
(Refer to the error list RXL2-120-072.03...).
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 33 of 72
Page 34

34 Potentiometers and Reference Switches

4Potentiometers and Reference Switches
5-

Overview of the potentiometers and reference switches 0

Fig. 3: Overview of potentiometers and reference switches
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 34 of 72
Medical Solutions
Page 35
Potentiometers and Reference Switches 35

Basic information on the replacement and adjustment of potentiometers, reference switches and motor encoders

Fig. 4: Replacing the potentiometer

General replacement and adjustment 0

0
a) Potentiometers
Move the unit so that the set screw becomes accessible ((M/Fig.4/p.35)).
Switch the system off.
Remove the Allen 1.5 mm set screw on the connecting piece of the potentiometer shaft
((M/Fig.4/p.35)).
Remove the potentiometer retaining plate ((ISK 3 mm/Fig.4/p.35)).
Remove the adjustment wheel ((E/Fig.4/p.35)) and the retaining parts from the po-
tentiometer shaft.
Desolder the connection wires from the potentiometer and solder them to the new po-
tentiometer in the same order.
Install the potentiometer in the reverse order. Do not tighten the set screw yet (Allen 1.5
mm).
Switch the system on.
Move the drive in "unregulated motion" with the service PC (Adjustment, Calibration) in
mm steps so that the corresponding reference switch is actuated (refer to
(Fig.5/p.36)). Check with DVM at D5.X2.A1...A11, 0V=X30.
NOTE
The potentiometer values are set in accordance with ADC values.
Exit "unregulated motion" and select the "ADC value" for the corresponding drive in Di-
agnostic, Monitoring.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 35 of 72
Page 36
36 Potentiometers and Reference Switches
Set the ADC value to the corresponding value using the adjustment wheel
((E/Fig.4/p.35)) (Refer to the detailed description for the corresponding potentiome- ter).
Retighten the set screw on the potentiometer shaft.
Reset board D1 with S1.
Check the drive function. No errors may be displayed.
If reference switch errors continue to be displayed after adjusting the potentiometer (Er-
rors 0020-002B), check the A/D converter on board D5 according to (Board D 5, A/D
converter / p. 24).
b) Reference switches
NOTE
A 24 V lamp may be used to check switch functions instead of an oscilloscope.
Switch the system off.
Remove the reference switch in question and remove the switch rocker.
Replace the switch and wire.
Reinstall the switch with the switch rocker and check the operation of the switch rocker.
Switch the system on.
Fig. 5: Replacing the reference switch
Connect an oscilloscope (or lamp 24 V) to the input of board D5 X2 a... to check the reference switch function (GND: X30/X40) (refer to the Overview circuit diagram for the respective reference switch J1042-8...).
Move the corresponding drive far enough so that the reference position is reached on
the service PC (Diagnostic; Monitoring) (Refer to the chapter on adjusting the drives).
Adjust the reference switch by moving it in the X direction. The switch should activate
within the mechanical limits of the switching hysteresis ((Fig.5/p.36)).
Check it using the service PC (Diagnostic; Monitoring; Refswitch; Y; X; 0; W; ZL; ZR; TL; TR) or on the oscilloscope (lamp 24 V).
Move the drive over the reference switch position several times to check.
Check the drive function.
Move the activator of the drive up to approx. 1 cm in front of the reference switch.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 36 of 72
Medical Solutions
Page 37
Potentiometers and Reference Switches 37
Continue to move toward the switch in inching mode ((Fig.5/p.36)).
In addition, check the status of the reference switch on the service PC (Diagnostic, Monitoring).
If the switch is activated from both sides, check the switching hysteresis for both sides.
No errors may be displayed.
Overview of the switching hysteresis
The test points of the actual value potentiometers and reference switches on board D5
are shown in the wiring diagram J1042-8x.
Fig. 6: Overview of the switching hysteresis
The switch can be ON or OFF within the range "X".
For the actual value potentiometers and reference switch test points on board D5, refer to wiring diagram J1042-8x.
c) Checking synchronization of the actual value potentiometers to the motor encoder
NOTE
Adjustment of the MC position with the ADC position
- When the system is switched ON or board D1 is reset with button S1, the value of "MC position" is adjusted to the value of the "ADC position.“ (initialized).
- If the ADC position deviates from the MC position, the error "428x AD / MC position comparison failure / x“ is displayed (x corresponds to the number of the drive).
Reason for the error message
- Potentiometer Actual value Rx not synchronous to the motor encoder Bx.
Maximum acceptable differences in the ADC/MC positions
ZL, ZR = 25 in 1/10 mm W = 30 in 1/10 mm X = 30 in 1/10 mm Y = 30 in 1/10 mm
Starting with software VA00D (Software Upgrade Rev. 3), the phys­ical limits of the drive are monitored.
O = 90 in 1/100° TL = 18 in 1/10 mm TR = 175 in 1/100°
Checking the display for the corresponding drive in the service PC
"MC position" and "ADC position" are displayed on the service PC (Diagnostic, Monitor­ing).
Move the respective drive into an end position.
Reset board D1 with button S1.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 37 of 72
Page 38
38 Potentiometers and Reference Switches
Check the respective adjustment for the drive ("MC position" to "ADC position") on the
service PC (Diagnostic, Monitoring).
Move the drive to the opposite end stop in 5 steps.
Perform a check at the service PC.
NOTE
Procedure for tensioning the toothed belt:
If the deviations of the ADC/MC positions are greater than the max­imum acceptable deviations, the drives must be checked. If there are errors in the lifting columns and/or the cone drive, they must be replaced, since the actual value potentiometer cannot be ad­justed.
Loosen the screws at the idler pulley of the corresponding potentiometer toothed belt.
Tighten the toothed belt by shifting the idler pulley and retighten the retaining plate.
Reset board D1 with button S1.
Perform a functional check as described above.
Loosen the idler pulley again if shifting the idler pulley does not adjust the belt ade-
quately.
Remove the retaining plate for the toothed belt end stop.
Open the clamping piece of the toothed belt, shorten the belt by removing 4 teeth (2
teeth each parallel), insert the belt and reassemble completely.
Perform a functional check as described above (check hysteresis of the reference
switch).
Error display:
„Antriebe stehen außerhalb der physikalischen Betriebsgrenzen"
"Drives are located outside of physical limits"
"Moteurs endehors des limites physiques"
"Los Motores se eneuentran fuera de los limites fisicos"
Error Cause:
Remedy:
One of the drives (O, W, X, Y, Z, TL, TR) is in the ADC position in
the area of the hardware proximity switch outside the operating range (Monitoring) (ZL, ZR = 50 in 1/10 mm; W, X, Y = 40 in 1/10 mm; O = 100 in 1/100°; TL = 20 in 1/10 mm; TR = 400 in 1/100 °)
Check the respective drive using the service PC (Diagnostic, Moni-
toring).
Move the drive into the mid-range position (inside the operating
limits) in "unregulated mode" (Avoid collisions).
Reset board D1 with button S1.
Perform a functional check. If necessary, repeat the check of the
display ("MC/ADC position") for the corresponding drive in the ser­vice PC. If there is no improvement, adjust the toothed belt.
Error display:
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 38 of 72
Medical Solutions
Page 39
Potentiometers and Reference Switches 39
„Gerät befindet sich außerhalb der Betriebsgrenzen“
"Unit is outside the operating limits"
"L appereil est en dehors des limites d utilisation"
"Equipo se halla fuera de los limites de servicio"
Error Cause: - After exiting service mode "Unregulated motion" or
- when changing the Uro table, (switching over from Lithotripsy mode<-> Urology mode) O, W, X, Y, Z, TL, TR
At least one drive is outside the respective operating limits. (Refer
above for the drive values)
Remedy:
Move drives to the basic position or ascending position (inside the
operating limits) using the 0 button.
Perform a functional check. If necessary, repeat the check of the
display ("MC/ADC position") for the corresponding drive in the ser­vice PC.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 39 of 72
Page 40
40 Potentiometers and Reference Switches

Checking, replacement and/or adjustment 0

Reference switches S3..6, S8, S9 or actual value potentiometers r3..6, r8, r9
NOTICE
Make the following preparations according to the respective work step:
Danger of collision!
¹ Move the drives to a mid-range position to avoid collision
when moving the system in "unregulated motion" mode.
¹ The shock wave head (W drive) is in the park position!
Connect the service PC to board D1 X10.
The ADC values, position values and reference switch settings can be read out in the
"Diagnostic, Monitoring" menu.
As an alternative, you can proceed as follows:
- Remove the EMC shielding plate at the front of N11. The test points on board D5 are now accessible from the front. The test points of the ACTUAL value potentiometers are shown in wiring diagram J
1042-25.
- Remove the front plug X2 of board D5 and reconnect it using the short 96-pin exten­sion board 85 73 628 G5206.
(service accessory for the engineer).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 40 of 72
Medical Solutions
Page 41
Potentiometers and Reference Switches 41

Cone drive TL, reference switch S8 and potentiometer r8 0

Connect an oscilloscope to D5.X2.a21 (for S8) (0 V to X30); 5 V/div., 0.1 s/div.
Switch the system on.
Move the shock wave head into the therapy position.
Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30) ADC Value = 102
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
Connect the DVM to D5.X2.a22 (0 V at X30).
Check the function of potentiometer r8.
Perform the check in normal mode.
Check the potentiometer voltage at reference switch S8 = "On". U
at D5.X22 and 0 V at D5.X30; ADC value = 102
= 512 mV ± 5 mV
ref
Check the switching hysteresis of reference switch S8 according to (Fig.7/p.41) in
"Monitoring".
Fig. 7: Switching hysteresis S8

Checking the toothed belt 0

Check the toothed belt according to Service Instructions Shock Wave System
RXL2-120.841.01.

Checking the travel range 0

The shock wave head is in the therapy position.
The C-arm is in the + 30° fluoroscopy position.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 41 of 72
Page 42
42 Potentiometers and Reference Switches
X Release fluoroscopy FL.
¹ The cone drive moves to the front end position. Safety limit switch S40 must not
be activated.
¹ The service PC (Diagnostic, Monitoring) display indicates: ADC value = MC
position 600 +/- 1 bit.
Release a shock wave.
¹ The cone drive travels back to MCESWL position 470 ± 1 bit.
Uncouple the W drive.
The cone longitudinal drive travels
- Cone drive serial no. < 2000 to the MC position 300 ± 1 bit
- Cone drive serial no. > 2000 to the MC position 360 ± 1 bit

Checking the travel range with US probe (optional) 0

Prerequisite: - US probe inserted - Shock wave head in therapy position
Use the cone longitudinal forward button on the US tableside control panel to move the
cone drive to the front end position.
¹ The LED on the US tableside control panel lights up.
¹ The service PC (Diagnostic, Monitoring) display indicates: ADC value = MC
position 600 ± 1 bit.
Release shock waves.
¹ The cone drive travels back to MCESWL position 470 ± 1 bit.
Use the cone longitudinal back button to move the cone drive to the back end position.
¹ The LED on the US tableside control panel lights up.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 42 of 72
Medical Solutions
Page 43
Potentiometers and Reference Switches 43

Cone drive TR, reference switch S 9 and potentiometer r9 0

NOTE
Only perform if the ultrasound option is available.
Connect the oscilloscope to D5.X2.a23 (for S9) (0 V to X30); 5 V/div, 0.1 s/div.
Switch the system on.
Move the shock wave head into the therapy position.
Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30)
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
Connect a DVM to D5.X2.a24 (0 V at X30).
Check the function of potentiometer r9.
Prerequisite: US probe not inserted.
Move to the reference position in Normal mode using the tableside control panel.
Use the "Cone rotation right or left" key for this.
¹ The cone rotation drive travels to the 0° position and stops.
¹ The 0° marks on the toothed ring and the lifting cylinder line up.
(If in "unregulated motion" mode, readjust the cone rotation )
¹ The position is displayed on the US tableside control panel.
When reference switch S9 is actuated, the DVM should display a reading of 2.560 V ± 2
mV. ADC value = 512
Check the switching hysteresis of the reference switch S9 according to (Fig.8/p.43).
Fig. 8: Switching hysteresis S9

Setting the parameters 0

Refer to the "Parameter Groups" section of the "Service software, Software parameter / Monitoring" instructions RXL2-120.032.02..
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 43 of 72
Page 44
44 Potentiometers and Reference Switches

Checking the travel range 0

Prerequisite: US probe is inserted.
Move the US probe +90° and -90° by pressing the "Cone rotation left or right" key.
Check the angular position on the US tableside control panel display.
Safety limit switch S42 must not be actuated in the ± 90° setting.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 44 of 72
Medical Solutions
Page 45
Potentiometers and Reference Switches 45

Longitudinal system Y drive 0

Adjusting belt tension
Fig. 9: Measurement_1
Fig. 10: Measurement_2
Move the unit left to the end stop.
The distance from the center of roller (1/Fig.9/p.45) to the end stop (1/Fig. 10 / p. 45)
is 420 ± 5mm.
Attach the spring balance (3/Fig. 10 / p. 45) and lift it with a force of 20 N + 5 N.
It must be possible to raise the belt 3 mm + 1 mm (2/Fig. 10 / p. 45).
If the belt tension varies from this, readjust it.
Adjusting reference switch S6 (Prerequisite: potentiometer r6 functioning cor­rectly)
Remove the cover of the cable inlet.
Remove the metal shutter on the left and right at the back of the unit base frame.
Switch the system on.
Move the Y drive in inching mode to display "00" on the control monitor or on the ser-
vice PC (Diagnostic, Monitoring) Y Length ADC position in 1/10mm display to "000".
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 45 of 72
Page 46
46 Potentiometers and Reference Switches
Adjust the reference switch S6 so that switching is assured within the stated tolerance
range indicated (refer to (Replacing reference switches) and (Switching hystereses S6)).
Reset the system on board D1 with S1.
Move the drive over the reference switch position several times to check.
Replacing reference switch S6 and potentiometer r6
Switch adjustment
Switch the system on.
Bring the orbital drive into approx. 0° position.
Bring the X drive into the mid-range position.
Bring the shock wave head (W drive) into the park position.
Select "unregulated motion" mode for the Y drive.
Approach number 3 of the 5 reference points on the back of the unit base frame from
the left.
(Reference edge is the right outer side of the cable inlet, refer to (Fig. 11 / p. 46)).
Fig. 11: Reference edge
Align the reference edge and the reference mark.
Replace reference switch S6 (refer to (b) Reference switches / p. 36)).
Check the potentiometer voltage for reference switch S6 = On U
(D5.X15 0V = D5 X30) ADC value = 512.
NOTICE
After replacement or for potentiometer deviations:
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 2.560 V + 2 mV
REF
= 5.120 V ±
ref
If needed, replace the potentiometer according to (a) Potentiometers / p. 35).
Reference switch is adjusted and closed.
Loosen the set screw on potentiometer R6.
Set the potentiometer value to ADC value = 512 by turning the adjustment wheel
((E/Fig.4/p.35)).
Tighten the set screw.
Reset board D1 with button S1.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 46 of 72
Medical Solutions
Page 47
Potentiometers and Reference Switches 47
Check the switching hysteresis of the reference switch S6 according to (Fig. 12 / p. 47).
Fig. 12: Switching hysteresis S6
Move to the end positions on the left or right in the Uro mode.
- The safety limit switches S64/S65 must not be activated.
- The unit movement must stop on the left at - 320 mm ± 2 mm and on the right at + 320 ± 2 mm (control monitor display).
Move to the left or right end positions in Litho mode (shock wave head in therapy posi-
tion). The safety limit switches S76/S77 must not be activated.
The unit movement must stop on the left at - 25 mm ± 2 mm and on the right at +25 ± 2 mm (control monitor display).
If necessary, reattach the metal shutter on the left and right onto the back of the base
frame.
When doing this, be sure to correctly position the mounting holes and to use the upper threads of the metal shutter.
Reattach the cover of the cable inlet.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 47 of 72
Page 48
48 Potentiometers and Reference Switches

Transverse system, X drive 0

Adjusting belt tension
Fig. 13: Screws
Fig. 14: Reference value
(Checking the belt tension with a spring balance is not possible when the system is assembled)
The unit should not have any play transversally to the patient.
To readjust, loosen the 4 fastening screws (1/Fig. 13 / p. 48) (two screws are not visi-
ble) and adjust with the adjustment screw ((2/Fig. 13 / p. 48)); reference value for ad- justment: 30 mm ((1/Fig. 14 / p. 48)).
Retighten the 4 mounting screws (1/Fig. 13 / p. 48).
Adjusting reference switch S5
(Prerequisite: potentiometer r5 functioning correctly)
Switch on the system.
Move the X drive in inching mode to display "20" mm on the control monitor or on the
service PC (Diagnostic, Monitoring) X Length ADC position in 1/10mm display to "200".
Adjust the reference switch S5 so that switching is assured within the stated tolerance
range (refer to (Fig.5/p.36)).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 48 of 72
Medical Solutions
Page 49
Potentiometers and Reference Switches 49
Reset the system on board D1 with S1.
Move the drive over the reference switch position several times to check.
Replacing reference switch S5 and potentiometer r5
Fig. 15: Transverse system_X_drive
Switch the system on.
Bring the orbital drive into approx. 0° position.
Bring the Y drive into the mid-range position (basic system).
Bring the shock wave head (W drive) into the park position.
Select "unregulated motion" mode for the X drive.
Move the unit carrier to the front far enough so that the front edge of the front plate is 23
mm ± 0.5 mm in front of the rail edge. ((Fig. 15 / p. 49))
Replace reference switch S5 (refer to (b) Reference switches / p. 36)).
Check the potentiometer voltage for reference switch S5 = On; U D5.X14 and 0V at D5.X30. ADC value = 780
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 3.900 V + 2 mV at
ref
= 5.120 V ±
ref
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 49 of 72
Page 50
50 Potentiometers and Reference Switches
Fig. 16: Switching hysteresis S5
After replacement or for potentiometer deviations:
Replace the potentiometer according to (a) Potentiometers / p. 35).
Reference switch is adjusted and closed.
Loosen the set screw on potentiometer R5.
Set the potentiometer value to ADC value = 780 by turning the adjustment wheel
((E/Fig.4/p.35)).
Tighten the set screw.
Reset board D1 with button S1.
Check the switching hysteresis of reference switch S5 according to (Fig. 16 / p. 50) Fig.
9.
Move to the front or back end positions in Uro mode.
- The safety limit switches S67/S66 must not be activated.
- Unit movement must stop at the front at +120 ± 2 mm and at the back at - 350 ± 2 mm.
(control monitor display).
Move to the back end position in Litho mode (shock wave head in therapy position).
- The safety limit switch S78 must not respond
- Restricted unit movement must stop at - 30 ± 2 mm to the back. (control monitor display).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 50 of 72
Medical Solutions
Page 51
Potentiometers and Reference Switches 51

C-arm orbital drive 0

Adjusting belt tension
The adjustment of the belt tension is described in "Working on the orbital drive" (Work-
ing on the Orbital Drive / p. 63) in this manual.
Fig. 17: C_arm_orbital
Adjusting the S4 reference switch
(Prerequisite: potentiometer r4 functioning correctly)
Switch the system on.
Move the orbital drive in inching mode to the mechanical indicator "+ 30°" on the ser-
vice PC (Diagnostic, Monitoring); and the C-arm to the ADC position in 1/10 mm dis­play to "+ 3000".
Fig. 18: Adjusting S4
Adjust reference switch S4 so that switching is assured within the stated switching hys-
teresis tolerance range (refer to (Fig.5/p.36) and (Fig. 18 / p. 51)).
Reset the system on board D1 with S1.
Move the drive over the reference switch position several times to check.
Replacing reference switch S4 or potentiometer r4
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 51 of 72
Page 52
52 Potentiometers and Reference Switches
Connect a DVM to D5.X13 (0 V an X30).
Connect an oscilloscope to D5.X2.a7 (0 V toX40) ; 5 V/div, 0,1 s/div.
Switch the system on.
Bring the X drive and the Y drive into mid-range position.
Bring the shock wave head (W drive) into the park position.
Select "unregulated motion" mode for the O drive.
Bring the arm into +30° position.
Align the reference edge of the unit cover and the reference mark (arrow)
((Fig. 17 / p. 51)).
Replace reference switch S4 (refer to (b) Reference switches / p. 36)).
Check the potentiometer voltage for the reference switch S4 = On U
mV (D5.X13 0V = D5 X30) ADC value = 966
NOTICE
After replacement or for potentiometer deviations:
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 4.830 V ± 2
REF
= 5.120 V ±
ref
If necessary, replace the potentiometer according to (a) Potentiometers / p. 35).
Reference switch is adjusted and closed.
Loosen the set screw on potentiometer R4.
Set the potentiometer value to ADC value = 966 by turning the adjustment wheel
((E/Fig.4/p.35)).
Tighten the set screw.
Reset board D1 with button S1.
Check the switching hysteresis of the reference switch S4 according to drawing
((Fig. 18 / p. 51)).
Move to the +30° or -10° end position with the tableside control.
- The safety limit switches S70/S71 must not be activated.
- When the 0° position keys on the tableside control are activated, the orbital drive moves into the vertical position (check the position indicated on the control monitor and at the reference mark).
- Fluoroscopy is enabled in the + 30° / 0° / -10° positions.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 52 of 72
Medical Solutions
Page 53
Potentiometers and Reference Switches 53

Shock head mount W system 0

Adjusting belt tension
Fig. 19: Adjusting belt tension
Move the W drive to the uppermost position.
Switch the system off.
Loosen the 4 fastening screws ((1/Fig. 19 / p. 53)) (two screws are not visible) and set a
value of 28 mm ((2/Fig. 19 / p. 53)) with the two adjustment screws ((3/Fig. 19 / p. 53)).
Retighten the 4 mounting screws.
Adjusting reference switch S3 and potentiometer r3
Switch the system on.
Move the shock wave system into the therapy position.
Attach the adjustment phantom (switch S28 has switched).
Check the isocenter in the + 30 degree or - 10 degree position and, if necessary, adjust
according to the "Isocenter adjustment" instructions " RXL2-120.071.01....
(Perform only the work necessary for shock head replacement).
NOTICE
Isocenter
¹ The isocenter is the starting point for the adjustment of
the reference switch or of the actual value potentiometer.
Adjust reference switch S3 so that switching is assured within the stated switching hys-
teresis tolerance range (refer to (Fig.5/p.36) and (Fig. 20 / p. 54)).
The shock wave head is located in the isocenter.
Replacing reference switch S3 and potentiometer r3
Switch the system on.
Bring the X drive and the Y drive into mid-range position (basic position)..
Bring the orbital drive into +30°.
Select "unregulated motion" mode for the W drive.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 53 of 72
Page 54
54 Potentiometers and Reference Switches
Bring the shock head into the therapy position.
Attach the adjustment phantom and adjust the isocenter (refer to Isocenter Adjustment
RXL2-120.071.01...)
Replace reference switch S3 (refer to (b) Reference switches / p. 36)).
Check the potentiometer voltage for reference switch S3 = ON; Uref = 4.770 V ± 2 mV
(D5.X12 0V = D5.X30) ADC value = 954.
NOTICE
Fig. 20: Replacing S3
After replacement or for potentiometer deviations:
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ±
ref
If necessary, replace the potentiometer according to (a) Potentiometers / p. 35).
Reference switch is adjusted and closed.
The shock wave system is in the adjusted isocenter.
Loosen the set screw on potentiometer R3.
Set the potentiometer value to ADC value = 954 by turning the adjustment wheel
((E/Fig.4/p.35)).
Tighten the set screw.
Reset board D1 with button S1.
Remove the adjustment device (switch S28 is open).
Check the switching hysteresis of reference switch S3 according to (Fig. 20 / p. 54).
Move to the "Coupling or park position" end positions.
- The safety limit switches S69/S68 must not be activated.
Check the isocenter according to "Isocenter adjustment" RXL2-120.071.01....
Check the Litho park position S79 (release of the X/Y travel range for Uro mode).
Move to the park position of the W drive; contact S79 must be closed. If necessary,
check "Switch ON" with the DVM on board D32 X41 PIN 3 and at ground GND.
Install the URO insert.
Move to the basic 0 position.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 54 of 72
Medical Solutions
Page 55
Potentiometers and Reference Switches 55
The system may now be moved outside the "Litho Position". (X = -130 / + 70; Y = ± 320)
Safety limit switches S 76/77 as well as S 78 must not be activated.
(Refer to circuit diagram J1042-7).
Remove the Uro insert and move to the →0 basic position.
The system will only allow travel in the "Litho Position".
Move the W drive out of the park position.
Switch S79 must open in the MC position - 430 ± 2 mm and switches S 76; S77; S78 must be activated. If necessary, check "Switch ON" with the DVM on board D32 X41 PIN 3 and at ground GND.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 55 of 72
Page 56

56 Checking / Replacing the Lifting Columns R / L

5Checking / Replacing the Lifting Columns R / L
6-

Checking potentiometers r1/r2 and reference switches S1 and S25

Connect an oscilloscope to D5.X2.a1/a3 (for S1/2) (0 V at X30); 5 V/div, 0.1s/div.
Switch the system on.
Install the Uro insert.
Check the reference voltage, if necessary set it with R220 on D8. (U
mV; check on D5 MP X31, 0V at X30)
NOTICE
Changing the reference voltage
¹ When changing the reference voltage D5 U
5 mV, check the switching hysteresis for all drives.
= 5.120 V ± 5
ref
= 5.120 V ±
ref
Connect a DVM to D5.X2.a2/a4 (0 V at X30).
Check the functions of potentiometers R1/2.
Perform the check in normal mode.
Adjust the table height with the tableside control (NBF).
Check the potentiometer voltage at reference switches S1/2 = "On" U
mV at D5.X10/X11 and 0 V at D5.X30. ADC value = 742
= 3.710 V ± 2
ref
0
Check the switching hysteresis of the reference switch S1/2 according to
(Fig. 21 / p. 56).
Fig. 21: Switching hysteresis S1_S2
NOTE
It is not possible to adjust potentiometers r1/r2 or reference switches S1/S2. The corresponding lifting column must be re­placed if there are defects or malfunctions.
Move the system into the basic position0 with the tableside control.
The position indicator on the control monitor will display the value "1033". Check this
with the service PC (MC position) in 1/10 mm (display "10330" / control monitor Z =
1033).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 56 of 72
Medical Solutions
Page 57
Checking / Replacing the Lifting Columns R / L 57
Fig. 22: Lifting column
Preliminary remarks:
NOTICE
Fig. 23: Removing cover
To avoid replacing the lifting columns unnecessarily, troubleshoot thoroughly and localize errors accurately.
Removing the lifting column
¹ When removing the lifting column, use a support
(wooden plank) approx. 80 cm to 95 cm high and with a load capacity of approx. 40 kg to support the tabletop. The engineer must obtain this item locally.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 57 of 72
Page 58
58 Checking / Replacing the Lifting Columns R / L

Preparation 0

Switch on the system and install the Uro insert.
Loosen the four cover screws of the front, gray base frame cover and pull the plate out
to the side (refer to (Fig. 23 / p. 57)).
Move the unit into the left or right end position. Disconnect the L/R cables of the lifting
columns from the cable harness in the base frame.
Move the tabletop to the max. height.
Remove the table extensions.
Remove all covers of the lifting column affected ((Fig. 22 / p. 57))
Remove the holding brackets of the covers by removing the screws.
Unscrew the holder of the energy chain on the left lifting column.
To remove the tabletop, push the support (wood plank) between the base frame and ta-
ble.
Slowly move the table down onto this support until the weight is taken off the column.
NOTICE
Lifting column
¹ Do not retract the lifting column any further after lower-
ing the table onto the support.
Switch the system off.
Remove the 4 M8 Allen screws at the top of the column being replaced.
Disconnect the plug for the power supply to the motor on the column that is not being
replaced.
Switch the system on.
Select "unregulated motion" with System-Logon-Adjustment-Calibration " as per the
"Operating the Service Software" instructions RXL2-120.113.01....
Select the lifting column drive ZR or ZL.
NOTICE
Danger of collision!
¹ The system is now in "unregulated operation". The
movement sequences and the unit movements are no longer monitored. Use caution to prevent collision haz­ards.
Retract the column using the "Table down" button on the tableside control panel.
Switch the system off.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 58 of 72
Medical Solutions
Page 59
Checking / Replacing the Lifting Columns R / L 59

Removing the lifting column 0

Disconnect the plug from the actual value acquisition of the lifting columns (board D31
X10/ X11).
Disconnect the motor power supply plug.
Unlock and remove the motor encoder plug.
Fig. 24: Allen_screws_A
Turn the lifting column drive shaft by hand until the Allen screws of the drive shaft are
accessible from above ((Fig. 24 / p. 59)).
Remove the 2.5 mm Allen set screws ((1/Fig. 24 / p. 59) and (1/Fig. 25 / p. 59)).
If necessary, remove "Loctite" using a heat gun.
Remove the mounting screws on the base of the lifting column (Fig. 4) and remove the
column.
NOTE
Remove the drive shaft bearing to replace the cables ((Fig. 26 / p. 60)).
Remove the strain relief on the two cables for the lifting column.
Remove the mounting screws on the base of the lifting column ((Fig. 24 / p. 59)) and re-
move the column.
Fig. 25: Allen_screws_B
Ensure that the encoder on the motor shaft does not become dam­aged.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 59 of 72
Page 60
60 Checking / Replacing the Lifting Columns R / L

Installing the new lifting column 0

Insert the connection cable of the new lifting column. Ensure that the leads are not twist-
ed.
Position the column on the base plate.
Apply Loctite 221 to the screws (Allen M8 x 40). Insert them and tighten them with a
torque of 24 Nm ± 10%.
Insert the cable in the bearing guide track of the drive shaft and tighten it gently.
Secure the strain relief with cable ties.
Reinstall the drive shaft of the lifting column.
Ensure that the screw engages on the flattened side of the drive shaft.
Fig. 26: Drive axle_A bearing
Fig. 27: Drive axle_B_bearing
Reattach the drive shaft bearing with M10x20 screws. Torque: 50 Nm ± 10%
Connect the drive shaft with the PVC coupling (Allen screws 2.5 mm) and bolt the mo-
tor down ((Fig. 25 / p. 59)).
Reconnect the actual value cable to board D31 X 10 or X11.
Reconnect the power supply and the motor encoder cable.
Switch the system on.
Select "unregulated motion" with System-Logon-Adjustment-Calibration " as per the
"Operating the Service Software" instructions RXL2-120.113.01....
Select the lifting column drive ZR or ZL.
Raise the column in unregulated mode until it just makes contact with the tabletop.
Apply Loctite 221 to the four Allen screws (M8) and bolt the column to the tabletop from
below.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 60 of 72
Medical Solutions
Page 61
Checking / Replacing the Lifting Columns R / L 61
Switch the system off.
Install the holding brackets for the covers and the chain holder.
Reconnect the power supply for the second column.
Switch the system on.
Raise the table and remove the tabletop support from the travel range.
Check the function of the lifting columns.
If necessary, check the "Table tilt" option.
Reinstall all the covers of the lifting column that was just replaced. Reconnect the pro-
tective conductor.
Reattach the two-part plastic cover at the base of the lifting column. Seal the edges of
the plastic cover against the unit or floor with silicon.
Reconnect the cables to the cable harness in the base frame.
Close the covers of the unit base frame.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 61 of 72
Page 62

62 Working on the Motor Drives

6Working on the Motor Drives
7-
The following general issues apply:
NOTICE
No special safety measures are required when working on the X/Y drives.
Working on the orbital / W drive
¹ When working on the orbital drive, secure the C-arm
against uncontrolled movement. When working on the W drive, secure the bearing block from backing up into the mechanical stop.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 62 of 72
Medical Solutions
Page 63

Working on the Orbital Drive 63

7Working on the Orbital Drive
8-

Removing the am4 drive motor 0

Fig. 28: Covers_A
Fig. 30: Retaining rails
Move the shock wave head to the therapy position and the C-arm to the 30° position.
Remove the covers ((Fig. 28 / p. 63)) from the basic system and the X-ray tube.
Fig. 29: Covers_B
Tools required:
- A tensioning belt or similar device approximately 4.5 m long and 25mm wide must be purchased locally.
- 5mm Allen wrench with guide pin
Remove the side and lower C-arm covers ((Fig. 29 / p. 63)).
Remove the retaining rails ((1/Fig. 30 / p. 63)) for the cover plate.
Switch the system on.
Select "unregulated mode" on the service PC.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 63 of 72
Page 64
64 Working on the Orbital Drive
Fig. 31: D7 board
Fig. 33: Potentiometer r4
Fig. 32: Board D7_tilted
Fig. 34: Rubber end stop
Move the O drive to the mechanical end stop, passing the 30° position, in unregulated
mode so that the rubber end stops are against the mechanical end stop ((Fig. 34 / p. 64)).
Switch the system off.
Unplug the motor connector ((A/Fig. 32 / p. 64)).
Remove board D7 ((2/Fig. 31 / p. 64)).
Loosen the screws for the D7 board holder ((3/Fig. 31 / p. 64)) and tilt it forward
((5/Fig. 32 / p. 64)).
Loosen the screw ((4/Fig. 31 / p. 64)) at the energy chain guide.
Lift up the energy chain guide and secure it above ((6/Fig. 32 / p. 64)).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 64 of 72
Medical Solutions
Page 65
Working on the Orbital Drive 65
Position the tools as indicated in the figure ((7/Fig. 32 / p. 64)) and loosen the screws,
but do not remove them.
Loosen the set screw M (1.5 Allen) at potentiometer r4 ((9/Fig. 33 / p. 64)) on the left
side of the system.
Fig. 35: Tensioning belt_A Fig. 36: Tensioning belt_B
Fig. 37: Cable guides
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 65 of 72
Page 66
66 Working on the Orbital Drive

Replacing the toothed belt 0

(The am4 drive motor must have been removed).
NOTICE
Risk of accidents!
¹ Two persons are always required when working on the
C-arm. One person must press against the X-ray tube while the other tightens or loosens the tensioning belt.
Remove the I.I. collision protection.
Attach the tensioning belt ((10/Fig. 35 / p. 65) and (10/Fig. 37 / p. 65)) on one of the
sides; route the tensioning belt over the upper concealed roller (11/Fig. 35 / p. 65).
Lift up the C-arm using the tensioning belt until the cable guides ((12/Fig. 37 / p. 65))
are accessible.
For both the left and right cable guides:
- Loosen the three screws ((13/Fig. 37 / p. 65)) and remove the cover plate.
- Remove the rubber end stop ((14/Fig. 37 / p. 65)); the attachment screw (Allen 6 mm) is accessible from above, through the cable harness.
- Remove the two screws ((15/Fig. 37 / p. 65)).
- After removing the 4 attachment screws from above ((16/Fig. 37 / p. 65)) and below (in (Fig. 37 / p. 65) concealed), push the holder downward as far as possible.
Fig. 38: Attachment plates
Fig. 39: X-ray tube assembly holder
Move the C-arm outward as far as possible, passing the -10° position, using the ten-
sioning belt until the attachment plates on the back ((17/Fig. 38 / p. 66)) can be re- moved.
Loosen and remove the attachment plates using a 5 mm Allen key with guide pins.
Carefully lower the C-arm to the 30° position; the end stop has been removed.
Remove both upper screws from the X-ray tube holder ((18/Fig. 39 / p. 66)).
Remove both tensioning screws ((19/Fig. 39 / p. 66) and (19/Fig. 40 / p. 67)).
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 66 of 72
Medical Solutions
Page 67
Working on the Orbital Drive 67
Remove the 5 screws ((20/Fig. 39 / p. 66)).
Pull the clamp with the belts downward and out. Note the original position to facilitate in-
stalling the new belt later.
Open the clamp and remove both old toothed belts.
Prepare both new belts by:
- Cutting them to the length of the old belts.
- Drilling one side of the toothed belt, Using the old belt as a guide.
- Inserting them into the clamp; ensuring that the holes of the toothed belt are not inside the clamp.
Position the clamp for securing the toothed belt with the threaded holes facing down-
ward between the X-ray tube flange and insert the holder on the C-arm ((Fig. 39 / p. 66) and (Fig. 40 / p. 67)).
Route both toothed belts into the C-arm from below.
Insert the 5 screws ((20/Fig. 39 / p. 66)) into the clamp and tighten only slightly.
Insert the tensioning screws ((19/Fig. 39 / p. 66) and (19/Fig. 40 / p. 67)) into the clamp
from below and tighten only slightly.
Fig. 40: Toothed belt placement
Re-tighten both upper screws of the X-ray tube holder ((18/Fig. 39 / p. 66)).
Raise the C-arm, passing the -10° position, to make the attachment holes visible.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 67 of 72
Page 68
68 Working on the Orbital Drive
Route the toothed belt according to (Fig. 40 / p. 67).
Pull the toothed belt upward and out of the C-arm.
Insert the screws first through the attachment plate ((17/Fig. 38 / p. 66)), then through
the toothed belt and finally tighten them in the C-arm, securing them with Loctite. Use a 5 mm Allen wrench with guide pin to tighten them.
NOTICE
Ensure that no steel cables are lying in the path of the screws.
¹
Lower the C-arm far enough and hold it with the tensioning belt until both cable guides
and rubber end stops ((Fig. 37 / p. 65)) can be reassembled in reverse order.
Lower the C-arm to the 30° end stop and remove the tensioning belt.
Move the C-arm through its entire range at least four times to align the belt.
Tighten the toothed belt with both tensioning screws ((19/Fig. 39 / p. 66) and
(19/Fig. 40 / p. 67)).
Tighten the 5 screws ((20/Fig. 39 / p. 66)).
Move the C-arm through its entire travel range again and check whether the alignment
is still correct.
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 68 of 72
Medical Solutions
Page 69
Working on the Orbital Drive 69

Installing the am4 drive motor 0

Install the drive motor by reversing the order in which it was removed, as described in
the section "Removing the am4 drive motor".
Install the holder for board D7.
Insert board D7.
Plug in the motor connector.
Install the holder for the energy chain guide.
Install the I.I. collision protection.
Switch the system on.
Move the C-arm out of the safety circuit in unregulated mode.
Adjust potentiometer r4 as described in Chapter 4.
Move the C-arm within the range from -10° to +30° and check whether the toothed
gears of the drive motor make any noise.
Re-install all covers.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 69 of 72
Page 70

70 Appendix

8Appendix
9-
Fig. 41: D1
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 70 of 72
Medical Solutions
Page 71

Changes to Previous Version 71

9Changes to Previous Version
10-
No changes; initial version.
Siemens AG RXL2-120.841.02.01.02 LITHOSTAR Multiline Medical Solutions
09.05 CS PS 24
Page 71 of 72
Page 72
72 Changes to Previous Version
LITHOSTAR Multiline RXL2-120.841.02.01.02 Siemens AG
09.05 CS PS 24
Page 72 of 72
Medical Solutions
Loading...