Erlwein
LITHOSTAR Multiline
Software
System
SP
Operation of the Service Software
03127011
Print No.:
Replaces: RXL2-120.113.01.03.02
RXL2-120.816.01.01.02
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 12.05
1995
2 Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
Page 2 of 38
Medical Solutions
Table of Contents 3
1- 0Table of Contents
1 _______ General information and overview__________________________________ 5
Important notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Documentation required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Aids required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
General overview of a connection setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 _______ Installation of the service software _________________________________ 9
Service software Version VB01C and higher (Win 98, 2000, XP) . . . . . . . . . . . . . . . . . . 9
Service software up to version VA00C (Win 3.11 und 95) . . . . . . . . . . . . . . . . . . . . . . . 10
3 _______ System _______________________________________________________ 11
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Logon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Logoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Real-time clock. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Id codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Printer setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Quit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4 _______ Data __________________________________________________________ 19
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Backup, version VB01C and higher . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Backup, versions up to VA00C. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5 _______ Adjustment ____________________________________________________ 24
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6 _______ Diagnostic ____________________________________________________ 27
Error Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7 _______ Info __________________________________________________________ 30
Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Customer Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
System Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Technical Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Siemens AG RXL2-120.816.01.01.02 LITHOSTAR Multiline
Medical Solutions
12.05 CS PS SP
Page 3 of 38
4 Table of Contents
Service Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8 _______ Help __________________________________________________________ 36
9 _______ Changes to previous version _____________________________________ 37
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
Page 4 of 38
Medical Solutions
General information and overview 5
1General information and overview
2-
Important notes 0
• Due to the technologies used in the LITHOSTAR Multiline, tasks may be performed
only by technicians who have been trained in the use of this material.
• Since the service program works under MS WINDOWS, basic knowledge of MS WIN-
DOWS is indispensable.
• If the system performs an automatic reset, the connection to the PC is maintained.
• Some messages depend upon the language of the Windows version installed.
Documentation required 0
• Setting the software parameters / Monitoring RXL2-120.032.02...
Aids required 0
• Service PC with Windows 98, 2000, XP (see Intranet/Product Information/Common
Services/Service Laptop for CSE’s).
• Serial PC connection cable, 5 m 99 00 440 RE999
• Service software 11 57 630 J1042
• Lithostar Multiline software Memoskop
• Lithostar Multiline host software
NOTE
Install service software on
the service PC (Installa-
tion of the service
software / p. 9)
This flowchart is intended only as a guide.
No ⇐
⇓⇓
Is the LITHOSTAR Multiline Service Soft-
ware installed on the service PC?
⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐ ⇐
⇒ Yes
Connect service PC in the
Multiline
N11.D1.X10
⇓
Start the service software
(General information and
overview / p. 5)
⇓
(System/p.11) ⇒ Connect.../ Logon ... / Logoff... / Realtime-Clock ...
/Id-Codes ... / Printer Setup... / Quit ...
Siemens AG RXL2-120.816.01.01.02 LITHOSTAR Multiline
Medical Solutions
12.05 CS PS SP
Page 5 of 38
6 General information and overview
(Data/p.19) ⇒ Download... / Backup ... / Restore...
Configuration ⇒ without function
(Adjustment / p. 24) ⇒ Parameters... / Calibration...
(Diagnostic / p. 27) ⇒ Error log ... / Tests... / Monitoring...
(Info / p. 30) ⇒ Versions ... / Customer Data... / System Data... / Technical
Info ... / Service Data ... / Operating Time ...
(Help/p.36) ⇒ General Help... / Index... / About... /
General overview of a connection setup 0
Establish connection between HOST PC board D1.X10 and service PC
⇓
Insert floppy with host software in drive A
⇓
Start the Lithostar Multiline service software
⇓
Perform User Identification on the service PC
...Username: enter name
.....Group: if the word is not known
see Speedinfo Knowledge Base PASSWORD LIST
⇓
In the System menu, open Connect
(See also System(Connect / p. 12))
⇓
⇒
⇓ ⇐ ⇐ ⇐
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
The time can be set in the Real-
time... menu
Page 6 of 38
Medical Solutions
General information and overview 7
Perform desired tasks with the service PC
⇓
Make a backup.
⇓
Delete error log in unit, if necessary
(see also Diagnostic (Error Log / p. 28) )
⇓
In the System menu, open Logoff...
(See also System (Logoff / p. 14))
⇓
In the System menu, open Quit...
⇓
Remove floppy with host software and put it back in logbook
⇓
Interrupt connection between HOST PC board D1.X10 and service PC
It is possible to view the following files without connecting the service PC to the unit.
However, this is recommended only to evaluate or print out data (e.g. in the office).
In the
Parameters
(Adjustment / p. 24)
menu:
In the
Error Log
(Diagnostic / p. 27)
menu:
In the (Info / p. 30)
menu:
Customer Data...
System Data...
Technical Data...
Service Data...
In the (Help / p. 36)
menu:
General Help
Index
About
Insert floppy disk with the Host-Software into drive A
⇓
Start the service software
Siemens AG RXL2-120.816.01.01.02 LITHOSTAR Multiline
Medical Solutions
12.05 CS PS SP
Page 7 of 38
8 General information and overview
⇓
Perform User Identification on the service PC
...Username: enter name
...Group: if the word is not known
see Speedinfo Knowledge Base RX PASSWORD LIST
⇓
Open corresponding menu
⇓
Click "Load from File"
⇓
See corresponding chapter for further information
⇓
End: Click Quit in the System menu
⇓
Remove floppy disk with HOST software from drive A
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
Page 8 of 38
Medical Solutions
Installation of the service software 9
2Installation of the service softw are
3-
Service software Version VB01C and higher (Win 98, 2000, XP) 0
Perform only if the service software is not yet installed on the service PC.
Open Windows Explorer
⇓
Insert floppy disk with service software in drive A
⇓
Click drive A
⇓
Double-click Setup.exe
⇓
The Install Shield starts
⇓
Next
⇓
Choose destination folder
(C:\SSW\Multiline\VB01C)
⇓
Next
⇓
Choose program folder
(C:\SSW\Multiline\VB01C)
⇓
Next
⇓
Service software is installed
The following message is dis-
played: “Setup complete”
⇓
End
⇓
End
Siemens AG RXL2-120.816.01.01.02 LITHOSTAR Multiline
Medical Solutions
12.05 CS PS SP
Page 9 of 38
10 Installation of the service software
Service software up to version VA00C (Win 3.11 und 95) 0
Perform only if the service software is not yet installed on the service PC.
Open file manager
⇓
Insert floppy disk with service software in drive A
⇓
Click drive A
⇓
Double-click Install.exe
⇓
⇐ The Install Shield starts ⇒
⇓⇓
⇓ Select Custom Install .
Default path ⇓
Select Full install .
Multiline service software
⇓⇓
⇓ Click Continue .
⇓⇓
⇒ Service software is installed ⇐
⇓
The following message is displayed:
A set of icons . . . . . . or existing group below
⇓
Click Create .
⇓
Click ok .
Select
⇓
When finished, answer the question in File Manager
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
Page 10 of 38
Medical Solutions
System 11
3System
4-
Overview 0
In the System ........................ menu:
Connect... ⇒ Initiate data transfer to unit
Logon... ⇒ Start service
Logoff... ⇒ Terminate service (another service start is possible with
LOGON)
Realtime clock... ⇒ Set time and date on the host PC board
Id-Codes... ⇒ Request of system data for the generation of a password (not
used with Multiline)
Printer Setup... ⇒ Set up printer
Quit... ⇒ Terminate service software
Siemens AG RXL2-120.816.01.01.02 LITHOSTAR Multiline
Medical Solutions
12.05 CS PS SP
Page 11 of 38
12 System
Connect 0
In the System menu, open Connect...
⇓
Select Com 1 or Com 2, depending on the service PC
⇓
⇐ Select a baud rate of 19200 Bd ⇒ Only necessary when
setting the time
⇓⇓
Click Logon ... .C l i c k ok.
⇓⇓
Click Realtime-clock.. in the System menu
⇓⇓
Click Logon.. in the System menu
⇓⇓
⇒ At “Please enter Password” ⇐
enter Siemens if it does not already appear
⇓
Click ok
⇓
The service software-host software connection is established
LITHOSTAR Multiline RXL2-120.816.01.01.02 Siemens AG
12.05 CS PS SP
Page 12 of 38