LITHOSTAR Modularis
CS
Service Instructions
Title:
Reason for update:
Urgency: Immediate Within 12 months
Update material required? Yes No
Materials free of charge? Yes No
Return of parts? Yes No
Estimated completion time: 0.5 h Number of CSE’s: 1 person
Customer application training? Yes No
Systems/Products affected/System identifying IVK
Name Material No. Serial No.
Replacing the hand-held control
x
x
SP008/02/I
X
Infrequent Service
x
x
x
LITHOSTAR Modularis 55 31 103 J1048 01001 - 01148
Remark: n.a.
Components affected/to be modified
Name Material No. Serial No. Component status
Affected
Control Unit 55 31 475 n.a. n.a.
Remark: n.a.
© Siemens AG 2002
Chg. Ref. No.: TDSP-CR-01/02 0022
Name: Erlwein
Dept.: CS PS 24
Print No.: SPL1-130.898.01.02.02 Doc. Gen. Date: 09.02
Replaces: SPL1-130.898.01.01.02 Version 2.0; A ugust 1, 2002 Page 1 of 6
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
2 Service Instructions SP008/02/I
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products affected 0
All LITHOSTAR Modularis systems.
Reason for the Update 0
With the Modularis control unit it is possible that switches do not respond or remain
enabled on actuation. This results in error messages and interferences in the patient treatment.
Prerequisites 0
Control units with the following serial numbers have already been updated:
1032, 1034, 1088, 1113, 1131, 1134, 1136, 1140, 1150, 1155, 1160, 1163, 1166, 1167,
1168, 1174, 1176, 1177, 1178, 1179, 1183, 1186, 1189, 1190, 1191, 1192, 1193, 1195.
Control units starting from serial number 1197 have already been delivered as changed
versions and do not have to be replaced.
Special Tools / Documents 0
n.a.
LITHOSTAR Modularis SPL1-130.898.01 Page 2 of 6 Siemens AG
Rev. 02 09.02 CS PS 24 Medical Solutions