Urgency: Immediate Within12 months
Update material required? Yes No
Materials free of charge? Yes No
Return of parts? Yes No
Estimated completion time: 0.5 h Number of CSE’s: 1 person
Customer application training? Yes No
Print No.: SPL1-130.898.01.02.02Doc. Gen. Date:09.02
Replaces: SPL1-130.898.01.01.02Version 2.0; A ugust 1, 2002Page 1 of 6
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
2Service Instructions SP008/02/I
Document Revision Level
This document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Systems/Products affected0
All LITHOSTAR Modularis systems.
Reason for the Update0
With the Modularis control unit it is possible that switches do not respond or remain
enabled on actuation. This results in error messages and interferences in the patient treatment.
Prerequisites0
Control units with the following serial numbers have already been updated:
1032, 1034, 1088, 1113, 1131, 1134, 1136, 1140, 1150, 1155, 1160, 1163, 1166, 1167,
1168, 1174, 1176, 1177, 1178, 1179, 1183, 1186, 1189, 1190, 1191, 1192, 1193, 1195.
Control units starting from serial number 1197 have already been delivered as changed
versions and do not have to be replaced.
Special Tools / Documents0
n.a.
LITHOSTAR ModularisSPL1-130.898.01Page 2 of 6Siemens AG
Rev. 0209.02CS PS 24Medical Solutions
Service Instructions SP008/02/I3
Ordering Information0
Place an order as follows for each system to be updated:
70 41 325 update kit
Contents of the Update Kit0
Update Kit 70 41 325 contains the following parts:
Pos.Quan.Part No.Name
1155 31 475Control Unit
21Technical Document
Tab. 1
Returns0
The following part must be returned to Erlangen:
55 31 475 Control Unit
The part must be correctly packed!
No credit will be issued for returns that have been damaged as a result of incorrect
handling.
A correctly completed returns form must accompany the parts to be returned.
Spare Parts Handling0
n.a.
Siemens AGSPL1-130.898.01Page 3 of 6LITHOSTAR Modularis
Medical SolutionsRev. 0209.02CS PS 24
4Service Instructions SP008/02/I
Work Steps0
• If problems arise during the operation of Modularis and if movements are not performed
properly a malfunction of the control unit can be assumed.
• The defective control unit should then be replaced by a modified control unit during the
next service call.
Final Check0
• Performance of a final check. Check of the overall function of the system.
Customer Information0
n.a.
Final Work Steps0
• Update the system documentation.
Update the Revision level, the User Instructions and the technical documentation.
Fill out, and if needed, copy the attached "Update Completion Form" and insert it into
Register 10 of the System Binder or System Logbook.
• Updates that were completed prior to publication of this SI must also be reported.
• The update is reported as follows:
- The modification reply report has to be made exclusively through an application in the
Intranet by authorized persons. Do not send the update completion from to CSPS.
Changes to Previous Version0
Page 1Number changed from SP008/02/R to SP008/02/I
Cross at "Materials free of charge ?" changed from "Yes" to "No".
LITHOSTAR ModularisSPL1-130.898.01Page 4 of 6Siemens AG
Rev. 0209.02CS PS 24Medical Solutions
Service Instructions SP008/02/I5
Completion Protocol / Update Completion Form0
The update with the number SP008/02/I has been completed.