Laser Camera Connection
Installation Instructions
SP
Connection for laser camera Part No. 16 61 524
LITHOSTAR Multiline
UROSKOP D
© Siemens AG 1996
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: RXL2-120.031.02.02.02 Doc. Gen. Date: 05.03
Replaces: RXL2-120.031.02.01.02

2
Connection for Laser Camera1
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Laser Camera Connection RXL2-120.031.02 Page 2 of 4 Siemens AG
Rev. 02 05.03 CS PS 24 Medical Solutions

Connection for Laser Camera 3
Tools required 1
• Service tools
• Protective ground wire tester 44 15 899 RV090
Installation locations of the optical video isolation amplifier
OPTV 105B
Videomed DIM
MEMOSKOP
I
N
7
5
S
O
U
T
OPTV
105B
N11
M16
M20
If T-piece for 3rd
monitor installed, alt.
location here
S
I
N
O
U
T
OPTV
105B
IN
OUT 75S
Z96
OPTV
105B
7
5
14
1
M20
I
N
7
5
S
O
9
U
T
OPTV
105B
10
5
Control cabinet
LITHOSTAR Multiline
Control cabinet
UROSKOP D1
Control cabinet
UROSKOP D2
Mechanical installation of the video isolation amplifier 1
Installation locations in the units, see Fig.
• Fasten the video isolation amplifier OPTV 105B to the support plate.
• Replace the supplied plastic screws for the installation of the support plate by steel
screws.
• Install the support plate with the video isolation amplifier according to system in
accordance with the Fig. in the control cabinet.
Siemens AG RXL2-120.031.02 Page 3 of 4 Laser Camera Connection
Medical Solutions Rev. 02 05.03 CS PS 24

4 Connection for Laser Camera
Connection of the cables 1
• Remove the small EMC plate on the MEMOSKOP.
• Connection of the Remote lead in the MEMOSKOP
- If a cable is connected to the connection X4, remove this.
- Connect the supplied cable to the connection X4 and run it to the laser camera.
• Connection of the video cable to the MEMOSKOP/video isolation amplifier
- If a cable is connected to the MEMOSKOP connection X10, remove this.
- Connect the short supplied video cable to the MEMOSKOP connection X10 and to the
video isolation amplifier "Video in".
- Run the supplied long video cable from the video isolation amplifier "Video 75
the laser camera.
- Connect the shielding of the video cable with the cabinet.
• Connection of the power cable
- Plug the power cable into the video isolation amplifier.
- LITHOSTAR Multiline: - Connect power cable to M16.X2.L2, N, PE
- UROSKOP D1 / D2: - Connect power cable to M10.X2. 3.8.PE in the image in-
tensifier TV cabinet below.
S out" to
• Attach the small EMC plate to the MEMOSKOP again.
• Perform a functional test.
• Measure the protective ground wire.
Changes to previous version 1
Page 1 - 4 Document complete new revised
Laser Camera Connection RXL2-120.031.02 Page 4 of 4 Siemens AG
Rev. 02 05.03 CS PS 24 Medical Solutions