Siemens Hematek Operating Manual

Page 1
Page 2
© 2008 Siemens Healthcare Diagnostics Inc. All rights reserved.
No part of this manual or the products it describes may be reproduced by any means in any form without prior consent in writing from Siemens.
Hematek is a trademark of Siemens Healthcare Diagnostics.
Origin: United States
The information in this manual was correct at the time of printing. However, Siemens continues to improve products and reserves the right to change specifications, equipment, and maintenance procedures at any time without notice.
If the Hematek system is used in a manner not specified by Siemens, the protection provided by the equipment may be impaired. See warning and hazard statements.
Page 3

Contents

Section 1: Introduction............................................................................................ 1-3
General Description and Intended Use...........................................................................................1-3
Theory of Operation .................................................................................................................1-4
Physical Characteristics....................................................................................................................1-5
Circular Bubble Level ...............................................................................................................1-5
Levelers ..................................................................................................................................1-5
Slide Transport System.....................................................................................................................1-6
Conveyor Spirals .....................................................................................................................1-6
Platen .....................................................................................................................................1-6
Slide Drying System .................................................................................................................1-7
Slide Drawer ...........................................................................................................................1-7
Waste Tank .............................................................................................................................1-7
Staining System.................................................................................................................................1-8
Volume Control Panel ..............................................................................................................1-8
Solution Pumps .......................................................................................................................1-8
Pump Tubing and Cannulas ......................................................................................................1-9
Sensing Switches ....................................................................................................................1-10
Electrical System ...............................................................................................................................1-11
Operating Lever .......................................................................................................................1-11
POWER Light ..........................................................................................................................1-12
LOW STAIN Light ....................................................................................................................1-12
Power Module .........................................................................................................................1-12
Line Cord Receptacle ...............................................................................................................1-12
Fuse Holder ............................................................................................................................1-13
ON / OFF Switch .....................................................................................................................1-13
Specifications .....................................................................................................................................1-13
Disposal of System Waste and Supplies ..................................................................................1-14
Section 2: Installation............................................................................................... 2-3
Overview .............................................................................................................................................2-3
Environmental Factors ......................................................................................................................2-3
Unpacking...........................................................................................................................................2-4
Instrument Setup ...............................................................................................................................2-4
Level the Instrument ................................................................................................................2-4
Plug the Line Cord into an Outlet ...............................................................................................2-5
Performance Check (Prior to Installing Tubing) ...........................................................................2-5
Install the Pump Tubing ............................................................................................................2-6
Install the Stain Pak .................................................................................................................2-9
Inspect Waste Tank and Slide Drawer .......................................................................................2-10
Familiarize Yourself with the Instrument .....................................................................................2-10
Check Pump Volumes ..............................................................................................................2-10
Section 3: Operating Instructions ........................................................................... 3-3
Contents
Page 4
General Guidelines for Optimal Staining ........................................................................................3-3
Operating Procedures.......................................................................................................................3-3
Start the Instrument .................................................................................................................3-3
Prime the Tubing .....................................................................................................................3-4
Load the Blood Smear Slides ....................................................................................................3-5
Stain the Smears .....................................................................................................................3-6
Clean the Tubing after Use .......................................................................................................3-7
Turn the Instrument Off at the End of the Day .............................................................................3-7
Stain Pak Replacement ....................................................................................................................3-7
Section 4: Specimens............................................................................................... 4-3
Overview .............................................................................................................................................4-3
Peripheral Blood Smears..................................................................................................................4-3
Bone Marrow Smears .......................................................................................................................4-5
Section 5: Maintenance............................................................................................. 5-3
General Cleaning...............................................................................................................................5-3
Daily Cleaning ....................................................................................................................................5-3
Cleaning the Platen ..................................................................................................................5-3
Cleaning the Stain Tubing and Cannula .....................................................................................5-4
Emptying the Waste Tank .........................................................................................................5-6
Weekly Cleaning................................................................................................................................5-6
Cleaning Drain Troughs and Rear Guide Rail .............................................................................5-6
Decontamination and Removal from Operation ............................................................................5-9
Section 6: Minor Replacements and Adjustments ................................................. 6-3
Fuse Replacement ............................................................................................................................6-3
Replacement of Light Assemblies ...................................................................................................6-4
Replacement of the Pump Tubing...................................................................................................6-7
Replacement of the Underplaten Tubing........................................................................................6-11
Pump Volume Adjustment................................................................................................................6-14
Check and Correct Instrument Operation Prior to Pump Volume Adjustment .................................6-15
Adjust the Stain, Buffer, and Rinse Volumes ..............................................................................6-15
Volume and Ratio Determination.....................................................................................................6-17
Sensing Switch Fingers Adjustment................................................................................................6-18
Section 7: Troubleshooting ..................................................................................... 7-3
Checklist for Quality of Stained Blood Smears..............................................................................7-3
Troubleshooting Chart.......................................................................................................................7-4
Section 8: Service, Supplies, and Replacement Parts .......................................... 8-3
Contents
Page 5
When to Call for Service...................................................................................................................8-3
For Service .........................................................................................................................................8-3
Siemens Authorized Representative ..........................................................................................8-3
Siemens Offices Worldwide .....................................................................................................8-4
Returning the Instrument for Repair, Exchange, Replacement, or Loaner................................8-7
Supplies and Replacement Parts ....................................................................................................8-8
Hematek Stain Pak – Modified Wright’s Stain .............................................................................8-8
Hematek Stain Pak – Modified Wright-Giemsa Stain ...................................................................8-8
Hematek Cannula Set ..............................................................................................................8-8
Hematek Pump Tube Set .........................................................................................................8-9
Hematek Underplaten Tubing ...................................................................................................8-9
Ordering Information for Supplies and Replacement Parts ...........................................................8-10
Appendix A: List of Symbols................................................................................... A-1
Symbols Used With This System ....................................................................................................A-1
Appendix B: Hematek Preservice Checklist........................................................... B-1
Contents
Page 6
Contents
Page 7
Section 1: Introduction............................................................................................ 1-3
General Description and Intended Use...........................................................................................1-3
Theory of Operation .................................................................................................................1-4
Physical Characteristics....................................................................................................................1-5
Circular Bubble Level ...............................................................................................................1-5
Levelers ..................................................................................................................................1-5
Slide Transport System.....................................................................................................................1-6
Conveyor Spirals .....................................................................................................................1-6
Platen .....................................................................................................................................1-6
Slide Drying System .................................................................................................................1-7
Slide Drawer ...........................................................................................................................1-7
Waste Tank .............................................................................................................................1-7
Staining System.................................................................................................................................1-8
Volume Control Panel ..............................................................................................................1-8
Solution Pumps .......................................................................................................................1-8
Pump Tubing and Cannulas ......................................................................................................1-9
Sensing Switches ....................................................................................................................1-10
Electrical System ...............................................................................................................................1-11
Operating Lever .......................................................................................................................1-11
POWER Light ..........................................................................................................................1-12
LOW STAIN Light ....................................................................................................................1-12
Power Module .........................................................................................................................1-12
Line Cord Receptacle ...............................................................................................................1-12
Fuse Holder ............................................................................................................................1-13
ON / OFF Switch .....................................................................................................................1-13
Specifications .....................................................................................................................................1-13
Disposal of System Waste and Supplies ..................................................................................1-14
Section 1: Introduction, Page 1—1
Page 8
Page 1—2, Section 1: Introduction
Page 9

Section 1: Introduction

General Description and Intended Use

The Hematek automated, bench-top instrument designed specifically for the staining of hematology slides for in vitro diagnostic use. This self-contained precision instrument accepts, conveys, fixes, stains, and delivers dry blood smear preparations that are spread on standard thickness 25-mm x 75-mm or 1” x 3” glass slides. The slides are stained at the rate of one slide per minute.
Two conveyor spirals move the slides along the platen. Three sensing switches are triggered sequentially as the slide moves along the platen. Each switch activates its respective solution pump, which meters and delivers the stain, buffer, or rinse into the capillary space between the slide and platen. After staining and rinsing have been accomplished, the slide is dried by a flow of air from a low velocity blower and it is then deposited in a slide drawer.
Optimal results with the Hematek Slide Stainer are obtained by using one of the Hematek Stain Paks. The Stain Pak consists of one bottle each of stain, buffer, and rinse solutions, and is designed for easy installation and removal from the instrument.
NOTE:
Only Hematek Stain Paks should be used with the Hematek Slide Stainer. Use of other stain packs or solutions may void the warranty.
®
Slide Stainer, shown in Figure 1-1. Hematek Slide Stainer is a fully
Section 1: Introduction, Page 1—3
Page 10

Theory of Operation

The Hematek Slide Stainer is designed to produce stained slides of consistent quality in a continuous process. This is accomplished by having a fixed length of time in each of the three phases of stain, buffer, and rinse, as well as a predetermined ratio of stain-to-buffer volumes in the buffer phase.
A properly stained slide is the result of an interactive process involving the pump volumes, the mixing process, and the stain-to-buffer ratio. A minimum ratio of 1:2 is recommended; however, if the platen is filling properly, good mixing is occurring, and the stained slides are acceptable under the microscope, the stain-to-buffer ratio should not be adjusted.
Figure 1-1. Hematek Slide Stainer
Page 1—4, Section 1: Introduction
Page 11

Physical Characteristics

Figure 1-2. Circular Bubble Level and Leveling Feet, shows the physical characteristics of the Hematek Slide Stainer, including the circular bubble level and the leveling feet.
Figure 1-2. Circular Bubble Level and Leveling Feet
1 Circular Bubble Level 2 Leveling Feet

Circular Bubble Level

CAUTION
Do not move the circular bubble level from its location directly behind the operating lever. Proper leveling is essential for optimal mixing of the stain and buffer.
The circular bubble level is a water bubble gauge with an inscribed circle that can be observed through the clear plastic lid of the instrument. The level is located on the instrument panel to the left of the reagent pumps, directly behind the operating lever. It is used to ensure the Slide Stainer is in a level position.

Levelers

The feet under the front corners of the instrument act as levelers and may be adjusted to raise or lower the instrument to a level position. Observe the circular bubble level to verify that the feet have been properly adjusted and the instrument is level.
Section 1: Introduction, Page 1—5
Page 12

Slide Transport System

Figure 1-3. Slide Transport System and Figure 1-4. Slide Transport and Staining System show the slide transport system, including the conveyor spirals, the platen, the platen guide rails, and the slide drawer.
Figure 1-3. Slide Transport System
1 Conveyor Spirals 2 Platen 3 Platen Guide Rails 4 Slide Drawer

Conveyor Spirals

The conveyor spirals are two parallel spirals with opposing grooves. Slides to be stained are fitted separately into the opposing grooves, which move the slides across the staining surface. When loading slides onto the instrument, the side covered by the blood smear is faced to the left of the operator, with the feathered edge to the back of the instrument. Slides are moved by the conveyor spirals side by side, from right to left, first at the vertical plane, then at the horizontal plane with the blood smear facing downward toward the platen.
Page 1—6, Section 1: Introduction
Page 13

Platen

The platen spans the entire front of the instrument, between the conveyor spirals. The outside ridges of the platen are elevated guide rails, which give support to the slides as they are moved along the platen. It is a precision-machined component made from a high­performance plastic polymer material and is designed specifically for two main functions:
1. It maintains the exact volumes of the required solutions within a capillary space between the platen and the slide.
2. It provides a mixing system for the stain and buffer.
The platen also provides the necessary time interval after the rinsing step for proper drying of the slides before they are deposited in the slide drawer. A trough around the perimeter of the platen allows for drainage of used solutions into the waste tank below.

Slide Drying System

The dryer is a blower type fan that runs continuously when the instrument is in operation. The airflow serves to cool the mechanical components inside the Slide Stainer, as well as to dry the slides.

Slide Drawer

The slide drawer is located below the left end of the platen. It receives the slides as they drop from the platen after being stained, rinsed, and dried. The drawer will hold 100 slides. See Figure 1-4. Slide Transport and Staining System .

Waste Tank

The waste tank is located underneath the platen and collects used and overflow staining solutions. The waste tank should be emptied and rinsed once each day and whenever a new Stain Pak is installed. Refer to Section 5: Maintenance for more information.
Section 1: Introduction, Page 1—7
Page 14

Staining System

Figure 1-4. Slide Transport and Staining System , Figure 1-5. Staining Systems, and Figure 1-6. Sensing Switches, illustrate the staining system, including the volume control panel, pump assemblies, pump cap, pump arm, pump tubing, and cannula.
Figure 1-4. Slide Transport and Staining System
1 Slide Drawer 2 Waste Tank 3 Volume Control Panel

Volume Control Panel

The volume control panel tips out from the right front corner of the instrument and contains three graduated adjustment knobs. The volume of reagent being delivered can be adjusted by rotating the respective control dial clockwise to increase the volume or counterclockwise to decrease the volume.

Solution Pumps

The instrument has three pump assemblies, one for each solution. Each assembly consists of a pump motor, four rollers that are attached to the underside of a pump cap, and a pump arm. All work together to maintain a constant metering speed to provide consistent volumes of stain, buffer, and rinse, even though the line voltage may fluctuate. The amount of solution pumped is electronically adjusted through the use of the volume control knobs.
Page 1—8, Section 1: Introduction
Page 15
Figure 1-5. Staining Systems
1 Pump Assemblies 2 Pump Cap 3 Pump Arm 4 Pump Tubing 5 Cannula

Pump Tubing and Cannulas

CAUTION
Use only the Hematek Pump Tube Set with the Hematek Slide Stainer. Use of other tubing may result in incorrect measurement and improper staining, buffering, and rinsing.
The pump tubing is provided as a set of three separate pieces of tubing. Notice that each pump tube is identified with a number and symbol, which corresponds to the reagent carried by the tubing:
1 for stain
2 for buffer
3 for rinse
Each tube has a clear plastic cuff that fits snugly up to the pump arm and holds the tube in position without slippage. The tubing is a special type of rubber that is resistant to leaching and brittleness. The special diameters and lengths of the tubing assure precise measurement of solutions. Each pump tube is connected to a cannula that is inserted into the appropriate bottle in the Stain Pak. Three cannulas are provided with the Hematek Slide Stainer. With proper routine cleaning, the cannulas can be used for an extended period of time.
Section 1: Introduction, Page 1—9
Page 16

Sensing Switches

The sensing switches are three finger-like devices located just above the back edge of the platen. When contacted, the respective sensing switch is activated, which tells the instrument that a slide is in position for the pump to run. After a specific time delay, the pumping motor for the specific reagent is activated. The solution pumps are set so that precise volumes of stain, buffer, and rinse are delivered to their respective areas on the platen. Each reagent is delivered through its respective cannula and tubing network to the platen orifice. The capillary space between the slide and the platen is then filled with the measured volume of reagent.
Figure 1-6. Sensing Switches
1 Sensing Switches
Page 1—10, Section 1: Introduction
Page 17

Electrical System

Figure 1-7. Electrical System and Figure 1-8. Electrical System–Rear of Instrument illustrate the components of the electrical system.
Figure 1-7. Electrical System
1 Operating Lever 2 Power Light 3 Low Stain Light

Operating Lever

The operating lever is a multiple function, bar-shaped lever located at the front left side of the instrument, near the top. The three positions of the lever are labeled and their related functions are as follows:
UNLOCK releases the pressure of the reagent pump arms against the pump tubing.
LOCK locks the reagent pump arms into their proper position against the pump tubing.
PRIME provides an override control to allow the pumps to run continuously so the tubes
can be primed with solutions and cleared of air bubbles. The lever must be held continuously in the
LOCK position.
PRIME position. When released, it returns automatically to the
Section 1: Introduction, Page 1—11
Page 18

POWER Light

The green POWER light is located at the right side of the instrument near the top. When the instrument is turned on, the light illuminates.

LOW STAIN Light

The yellow LOW STAIN light is to the left of the POWER light. It is off under usual operating conditions; however, when the Hematek Stain Pak contains sufficient reagents to stain only about 20 slides, a weight-sensing device activates the circuit and illuminates the light. The Stain Pak should be replaced at this time.

Power Module

The power module is located on the rear of the instrument; on one side of the module is the panel that contains the line cord receptacle, fuse, and power switch. The power module converts the line current coming into the instrument to 12 volts DC, which is the voltage within the instrument.
Figure 1-8. Electrical System–Rear of Instrument
1 Power Module 2 ON / OFF Switch 3 Fuse Holder 4 Line Cord Receptacle

Line Cord Receptacle

The line cord connects into the line cord receptacle, which is located at the top of the panel on the power module.
Page 1—12, Section 1: Introduction
Page 19

Fuse Holder

The fuse holder, in the center of the panel, holds the fuse that protects against serious electrical overload. A spare fuse is also located in the fuse holder.

ON / OFF Switch

The ON / OFF Switch is located at the bottom of the panel and controls all power to the instrument.

Specifications

Power Requirements 100 - 230 VAC + 10%, 50/60 Hz, 0.75
Amps
Fuse Rating 250 Volt, 1.0 Amps, 5mm x 20mm, Type T
Line Leakage Current <5 milliamperes
Testing protocol and allowable limits as specified by the safety standards for laboratory equipment outlined in UL 1262 and CSA 22.2 No. 151.
Dimensions Depth - 43.4 cm (17.1 in)
Width - 47.0 cm (18.5 in) Height - 19.0 cm (7.5 in)
Weight 17.7 kg (39 lbs)
Throughput Stains 1 slide per minute
Section 1: Introduction, Page 1—13
Page 20

Disposal of System Waste and Supplies

Laws and regulations enacted to protect the environment and to encourage resource conservation require the disposal of hazardous and biohazardous wastes in a specified manner. Some of the wastes from the Hematek Slide Stainer can be classified as hazardous or biohazardous wastes. It is essential that the laboratory take appropriate steps to determine the laws and regulations applicable to their disposal and to effect compliance. If it is necessary to sample instrument wastes and effluent in order to evaluate compliance with applicable regulations, the laboratory should contact a local licensed biohazardous waste disposal firm for assistance.
The principal wastes associated with the use of the Hematek Slide Stainer are pump and underplaten tubing, effluents from the staining operation, and the container for stain, buffer, and rinse.
Slides with human specimens, control materials, and all reagents, should also be handled and disposed of in accordance with the prevailing regulations and guidelines of agencies with jurisdiction over the laboratory. Refer to the product label and to Material Safety Data Sheets for details concerning any special precautions related to the handling of
Hematek Stain Pak containers. Material Safety Data Sheets are available from Siemens.
BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.
Page 1—14, Section 1: Introduction
Page 21
Section 2: Installation............................................................................................... 2-3
Overview .............................................................................................................................................2-3
Environmental Factors ......................................................................................................................2-3
Unpacking...........................................................................................................................................2-4
Instrument Setup ...............................................................................................................................2-4
Level the Instrument ................................................................................................................2-4
Plug the Line Cord into an Outlet ...............................................................................................2-5
Performance Check (Prior to Installing Tubing) ...........................................................................2-5
Install the Pump Tubing ............................................................................................................2-6
Install the Stain Pak .................................................................................................................2-9
Inspect Waste Tank and Slide Drawer .......................................................................................2-10
Familiarize Yourself with the Instrument .....................................................................................2-10
Check Pump Volumes ..............................................................................................................2-10
Section 2: Installation, Page 2—1
Page 22
Page 2—2, Section 2: Installation
Page 23

Section 2: Installation

Overview

This section provides detailed installation and setup instructions for the Hematek Slide Stainer. The installation steps must be followed correctly to ensure proper installation, operation, and service. Read this Operator’s Guide carefully before attempting to operate the instrument. Follow all instructions carefully.
The Hematek Slide Stainer is a precision instrument and must be handled accordingly. Rough handling or dropping of the instrument will disturb or damage internal components. Always handle the instrument with care.

Environmental Factors

As with all sensitive electronic instruments, prolonged exposure to excessive humidity and temperature should be avoided. Temperature should be held relatively constant to obtain the highest degree of operating stability. The ambient temperature range for operating the instrument is 18ºC to 30ºC (64ºF to 86ºF). The ambient operating humidity range is 20% to 85% relative humidity.
Place the instrument in a well-ventilated area, avoiding exposure to corrosive vapors or temperature extremes. Be sure it is near a power source that meets the electrical requirements (voltage) specified on the rating label located on the rear of the instrument. Avoid proximity to open windows, sinks, ovens, hot plates, open burners, radiators, and dry ice baths. The instrument should not be used in an explosive atmosphere.
Section 2: Installation, Page 2—3
Page 24

Unpacking

Before opening the shipping carton, inspect it for visible signs of damage. Use the following steps to unpack the instrument.

1. Carefully remove the Hematek instrument and supplies from the shipping carton.

The following items are provided with the instrument:
Hematek pump tube set
Hematek cannula set
operating manual
line cord
warranty card - for use by US customers only
FedEx PRP label - for use by US customers only

2. If the instrument shows any visible signs of damage, immediately file a complaint with the carrier.

3. Retain the shipping carton for further use. If the instrument ever needs to be shipped, the shipping carton will afford the best protection.

4. After the instrument has been unpacked, place it on a firm, level work surface in the designated work are by lifting the instrument by its frame.

Instrument Setup

Complete the following procedures to ensure proper installation and performance of your Slide Stainer.

Level the Instrument

CAUTION
Do not move the circular bubble level from its location directly behind the operating lever. Proper leveling is essential for optimal mixing of the stain and buffer.
1. Raise the hinged lid of the instrument and locate the circular bubble level.
The circular bubble level is located on the instrument panel to the left of the reagent pumps, directly behind the operating lever.
Page 2—4, Section 2: Installation
Page 25
2. Adjust the two feet located under the front corners of the instrument to raise or lower the instrument to a level position.
Figure 2-1. Leveling the Instrument
3. Observe the circular bubble level to indicate when the feet have been properly adjusted and the instrument is level. The bubble in the level should be centered within the inscribed circle.
NOTE:
If necessary, make a final leveling check and adjustment by watching the flow of the waste fluids during staining. Fluids should flow evenly to the drain hole at the left-front corner of the waste trough.

Plug the Line Cord into an Outlet

CAUTION
Be sure the outlet supplies the proper voltage for your instrument. Refer to the rating label located on the rear of the instrument to determine the proper voltage rating
Plug the appropriate end of the line cord into the instrument and the other end into an appropriately grounded AC electrical outlet.

Performance Check (Prior to Installing Tubing)

1. Turn the instrument on by pressing the ON / OFF switch, located on the left side of the power module at the rear of the instrument, to the
The green revolves.
2. Inspect the slide dryer area for noticeable airflow.
3. Place five blank slides into the grooves on the right side of the conveyor spirals.
Section 2: Installation, Page 2—5
POWER light illuminates, the fan starts, and the conveyor spirals slowly
ON position.
Page 26
Be sure the slides are positioned in opposing slots, parallel to the inscribed lines on the platen.
4. Allow the slides to automatically feed onto the platen.
5. As the slides move down the platen, make sure the leading edge of the slides contact and activate each of the three sensing switch fingers, located along the back wall of the platen above the platen trough.
As each switch is activated, the appropriate pump activates and you are able to observe the pump cap rotating.
6. If the instrument functions properly in these steps, continue with the instrument setup procedures.
7. If a problem occurred, contact the your local technical support provider or distributor. Refer to Section 8: Service, Supplies, and Replacement Parts for more information.

Install the Pump Tubing

1. Remove the three cannulas and pump tube sets from their packaging.
Notice that each pump tube is identified with a number, which corresponds to the reagent carried by the tubing:
• 1 for stain
• 2 for buffer
• 3 for rinse
2. Attach the labeled end of each pump tube to a cannula.
Figure 2-2. Attaching Pump Tubing to Cannula
Page 2—6, Section 2: Installation
Page 27
3. Push the operating lever down to the UNLOCK position.
4. Extend each new tube to its respective pump (in numerical order from right to left) and thread the end of the tubing into the hole in the pump arm.
5. Push the thumb tab on the pump arm to the extreme left and push the tubing through until the plastic cuff is flush against the pump arm.
Figure 2-3. Threading Tubing through Pump Assembly
NOTE:
If you encounter difficulty in threading the tubing through the pump housing, lift the operating lever to
PRIME for just a few seconds. This will cause the rollers inside the
pump housing to rotate slightly and relieve the interference. Then return the operating lever to the
UNLOCK position.
6. Release the pump arm.
7. Repeat steps 4 and 5 for the other two pumps.
99E44889 Rev. D Section 2: Installation, Page 2—7
Page 28
8. Connect each tube to its proper recessed nipple, located on the backside of the circuit board housing.
9. Be sure that at least 7 mm (0.25 in.) of tubing is connected to the nipple for a secure connection.
Figure 2-4. Attaching Pump Tubing to Nipple
NOTE:
If it is difficult to connect the tube to its nipple, use forceps or hemostats to grasp and attach the tubing.
CAUTION
Take care to not damage the tubing when using forceps or hemostats.
Page 2—8, Section 2: Installation
Page 29

Install the Stain Pak

1. Remove the perforated tabs from the Hematek Stain Pak. Insert the carton, with the
STAIN bottle to the right, into the well at the rear of the instrument.
2. Make sure the carton is all the way down and resting on the tray at the bottom of the well. The carton should be level when properly installed.
Figure 2-5. Installing the Stain Pak
3. Insert the appropriate cannula into its respective bottle by puncturing the center of the indentation on the bottle.
Figure 2-6. Installing Cannula into Stain Pak Bottle
Section 2: Installation, Page 2—9
Page 30
4. Remove the cannula, turn it 1/4 turn, and insert it again into the same puncture.
The double puncture creates a slightly larger hole for venting.
NOTE:
If additional venting is desired, a second hole can be made in the top of the bottle, near the indentation. A 20-gauge needle can be placed into the hole, if needed.
5. Push the cannula down until the guard at the top touches the plastic container.
6. Repeat steps 3 through 5 for each bottle.

Inspect Waste Tank and Slide Drawer

Inspect the waste tank and slide drawer for proper positioning below the platen. Each should be pushed completely into its respective cavity in the front of the instrument so it is flush with the control panel on the right front corner.

Familiarize Yourself with the Instrument

Before beginning normal instrument use, carefully review Section 3: Operating Instructions, and Section 5: Maintenance, to become familiar with operating techniques
and instrument cleaning requirements.

Check Pump Volumes

Check the pump volumes and adjust them if necessary by following the procedures in Pump Volume Adjustment on page 6-14 and Volume and Ratio Determination on page 6-
17. Do this before staining any patient slides for clinical evaluations.
Page 2—10, Section 2: Installation
Page 31
Section 3: Operating Instructions ........................................................................... 3-3
General Guidelines for Optimal Staining ........................................................................................3-3
Operating Procedures.......................................................................................................................3-3
Start the Instrument .................................................................................................................3-3
Prime the Tubing .....................................................................................................................3-4
Load the Blood Smear Slides ....................................................................................................3-5
Stain the Smears .....................................................................................................................3-6
Clean the Tubing after Use .......................................................................................................3-7
Turn the Instrument Off at the End of the Day .............................................................................3-7
Stain Pak Replacement ....................................................................................................................3-7
Section 3: Operating Instructions, Page 3—1
Page 32
Page 3—2, Section 3: Operating Instructions
Page 33

Section 3: Operating Instructions

Following initial installation (see Section 2: Installation), the Hematek Slide Stainer is ready for routine operation. Carefully read this section before beginning any slide staining.

General Guidelines for Optimal Staining

• Use high quality slides. Do not use slides of variable thickness. Bevel-edged slides
should not be used, as they may break in the instrument.
• Make sure the blood smears are thoroughly dry before placing on the instrument for
staining.
• Clean the stain tubing with methanol after each run, especially if the Slide Stainer is not
to be used for one hour or more.
• Keep the staining grooves and guide rails clean, as instructed in Cleaning the Platen on
page 5-3. Use only methanol to clean.
• Change the pump tubing and underplaten tubing on a routine basis. Refer to
Replacement of the Pump Tubing on page 6-7 and Replacement of the Underplaten Tubing on page 6-11 for recommended frequency.
• Check the alignment of the sensing switch fingers, as instructed in step 5 of Check and
Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15.

Operating Procedures

CAUTION
Use Hematek Stain Paks only. Other solutions may contain non-dissolved or particulate materials.

Start the Instrument

1. Turn the instrument on by pressing the ON / OFF switch, located on the left side of the power module at the rear of the instrument, to the
The green will begin to slowly revolve.
NOTE:
When the Hematek Stain Pak contains sufficient reagents to stain only about 20 slides, the yellow Pak. Refer to Stain Pak Replacement on page 3-7.
LOW STAIN light will illuminate to indicate the need for replacement of the Stain
ON position.
POWER light will illuminate, the fan will start, and the conveyor spirals
Section 3: Operating Instructions, Page 3—3
Page 34

Prime the Tubing

1. Lift the stainer-operating lever to the PRIME position and hold in this position until the stain, buffer, and rinse reagents all flow evenly through their tubes to the platen without any air bubbles.
NOTE:
If the tubing is new and does not prime easily, it may be necessary to assist the priming initially. Continue holding the operating lever in the arm inwards (toward the pump) until the reagent fills the tubing, then release the pump arm. It may also be helpful to pinch the tubing several times with the fingers, pinching in the area between the cannula and the pump.
PRIME position and push the pump
Figure 3-1. Priming the Tubing
2. Release the lever, which will return automatically to the LOCK position.
CAUTION
Always wipe the platen from right to left. Damage to the position or shape of the sensing switch fingers may result from not following recommended cleaning practices. Check the position of the fingers after cleaning as described in step 5 of Check and Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15.
3. After priming, wipe the platen with a soft disposable, lint-free absorbent cloth or tissue.
Page 3—4, Section 3: Operating Instructions
Page 35

Load the Blood Smear Slides

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.
1. Prime the platen.
The platen should be primed and wetted with reagents in order to ensure optimal results on the specimen slides.
2. Mark blank or old blood smear slides as priming slides (for example, label them with the name
You can reuse these priming slides
CAUTION
Slides must be inserted into the spiral grooves so they are parallel to the slide loading lines inscribed on the platen. If slides are not placed correctly in the spiral grooves, breakage can occur.
3. Load two to four priming slides and allow them to be transported across the platen ahead of the patient slides.
PRIME).
Figure 3-2. Priming the Platen
Section 3: Operating Instructions, Page 3—5
Page 36
4. Load the patient slides.
Place the properly prepared slides into the grooves of the conveyor spirals with the blood smear side facing to the left of the operator and the feathered edge of the blood smear toward the back of the instrument.
Figure 3-3. Loading the Blood Smear Slides

Stain the Smears

As the slides move along the platen, first vertically, then horizontally with the blood smear side down, the instrument performs the following steps.
1. Three sensing switches are triggered sequentially.
Each switch senses the slide and passes the information to another switch, which activates its respective solution pump as long as a slide is detected. The solution pumps meter and deliver the stain, buffer, and then rinse into the capillary space between the slide and platen.
NOTE:
If the reagents overflow, as opposed to filling the capillary space between the platen and slide, or if you obtain improper staining results, the reagent pumps may need adjustment. Refer to Pump Volume Adjustment on page 6-14 for more information.
2. After staining and rinsing have been accomplished, the slide is dried by a flow of air from a low velocity blower.
3. The slide is delivered into the slide drawer, ready for examination.
NOTE:
If the instrument is accidentally unplugged or there is a power failure during operation, it may be necessary to remove the slides on the platen and reprocess them.
Page 3—6, Section 3: Operating Instructions
Page 37

Clean the Tubing after Use

If no slides are to be processed for an extended period (one hour or more), it is recommended that the stain tubing be cleaned with methanol as described in Cleaning the Stain Tubing and Cannula on page 5-4. Push the operating lever down to the position to relieve the pressure against the pump tubing. You must re-prime the tubing and the platen before processing any more slides.

Turn the Instrument Off at the End of the Day

CAUTION
Always clean the instrument after daily use. See Daily Cleaning on page 5-3. If using frosted-end slides, it is especially important to clean the front guide rail on a regular and frequent basis. This is because the stain may spread across the frosted portion of the slide to the front rail. If allowed to accumulate, slide breakage may occur.
1. At the end of the day, clean the platen and stain tubing, and empty the waste tank.
UNLOCK
2. Push the operating lever to the
3. Turn the instrument off.

Stain Pak Replacement

If the LOW STAIN light illuminates when the instrument is first turned on or while slides are being processed, a new Hematek Stain Pak is needed. Follow these steps to replace the Stain Pak.
CAUTION
After replacement of the Stain Pak, the instrument must always be primed to remove any air bubbles that may be present.

1. Remove the three cannulas from the used Stain Pak and lift the empty carton out of the well at the rear of the instrument.

2. Remove the perforated tabs from the new Stain Pak carton.

3. Insert the carton with the instrument.
Make sure the carton is all the way down and resting on the tray at the bottom of the well. The carton should be level when properly installed.
UNLOCK position.
STAIN bottle to the right into the well at the rear of the

4. Vent each bottle and insert the cannulas as described in Install the Stain Pak on page 2-9.

NOTE:
Check the cannulas with each new Stain Pak and replace them if they appear bent or damaged.

5. Empty the waste tank into an appropriate receptacle and rinse it with water after each Stain Pak replacement. See Emptying the Waste Tank on page 5-6.

Section 3: Operating Instructions, Page 3—7
Page 38
Page 3—8, Section 3: Operating Instructions
Page 39
Section 4: Specimens............................................................................................... 4-3
Overview .............................................................................................................................................4-3
Peripheral Blood Smears..................................................................................................................4-3
Bone Marrow Smears .......................................................................................................................4-5
Section 4: Specimens, Page 4—1
Page 40
Page 4—2, Section 4: Specimens
Page 41

Section 4: Specimens

Overview

The Hematek Slide Stainer is designed specifically for the automatic staining of peripheral blood smears that have been prepared on standard 25 mm x 75 mm or 1” x 3” glass slides. Blood smears that are stained according to instructions will provide the examiner with high quality differential staining characteristics for all cytologic blood components. See Section 3: Operating Instructions for staining instructions.

Peripheral Blood Smears

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.
It is crucial to start with a properly prepared blood smear in order to obtain the best results on the stained slide. The following suggestions are recommended:
• Use high quality slides that are new and thoroughly cleaned.
• Slides must be free of oil and grease.
• Do not touch the slide surfaces with the fingers or against the skin of the patient.
• Protect blank slides from moisture and high humidity, as well as contamination by dust, flies, and other insects.
• Store the slides covered in a cool, dry place.
• Never use oxalated or heparinized blood for making blood smears.
• EDTA is the anticoagulant of choice.
• Protect blood smears from excessive heat (such as radiators and ovens), water splatters, and high humidity.
Use the following procedure to prepare a blood smear.

1. Remove the cap from a tube of well-mixed anticoagulated whole blood. EDTA is the anticoagulant of choice.

NOTE:
Blood should be kept at room temperature prior to preparing the smear. Adaquate mixing requires approximately 20 inversions prior to blood film preparation.
2. Place a drop of well-mixed blood near one end of a high quality slide. Hold a second spreader slide at about a 45 degree angle and approach the drop of blood. Allow the blood to spread almost to the width of of the slide edge. Then rapidly and smoothly push the spreader slide to the opposite end of the slide, pulling the blood behind it.
NOTE:
The angle and speed at which the blood drop is spread determines the thickness or thiness, and length of the blood film.
Section 4: Specimens, Page 4—3
Page 42

3. The ‘feather’ end of the blood film should be at least 1.57 cm (5/8 inch) from the opposite end of the slide to avoid extending beyond the platen surface and contact with the staining reagents.

Do not allow the smear to touch the edges of the slide as large cells tend to accumulate there. This may mechanically effect the distribution of the cell types.
Since frosted-end glass slides limit the spreading area, extra care is needed when preparing a blood film.
Figure 4-1. Making a Blood Smear

4. Allow the smear to dry thoroughly before staining. See Section 3: Operating Instructions for staining instructions.

After staining the slide, take the following precautions.
• Examine stained slides as soon as possible.
• Protect stained slides from direct sunlight and store them in a cool, dry place.
Page 4—4, Section 4: Specimens
Page 43

Bone Marrow Smears

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.
Many laboratories stain bone marrow smears with the same stain as is used for blood smears. The procedure involves using the “squash” technique for smear preparation and staining the slide once or twice on the Hematek Slide Stainer. The following procedure has been verified as giving clinically useful staining results on bone marrow slides; however, the stain quality is dependent upon the thickness and evenness of the bone marrow smear. A thick smear is more likely to require a second pass through the instrument than a thinner smear; an uneven smear will not stain uniformly.

1. Use bone marrow smears that have been prepared using the “squash” technique.

Be sure the smear is thoroughly dry before staining.

2. Place the slide onto the Hematek Slide Stainer and stain according to instructions in Section 3: Operating Instructions.

3. Remove the slide after it has been stained and examine it under high power (dry) for the
staining quality.
Do not use any oil on the slide for this examination.

4. If the slide is under-stained, place it on the slide stainer again and stain it a second time. In rare instances, a third staining may be necessary.

When using the Hematek Modified Wright Stain Pak (#4481), the bone marrow smears must be stained twice; one pass through the instrument is generally insufficient to produce distinct nuclear intensity. A third pass generally does not increase or decrease the quality of the staining.
When using the Hematek Modified Wright-Giemsa Stain Pak (#4405), a single staining is generally sufficient for most bone marrow smears. The smear may become over stained, with the appearance of precipitate and nuclear artifacts, on slides stained two or more times.

5. Thoroughly clean the platen after staining bone marrow slides to remove the greasy residue that can be left on the platen by the fat droplets in the marrow

Section 4: Specimens, Page 4—5
Page 44
Page 4—6, Section 4: Specimens
Page 45
Section 5: Maintenance............................................................................................. 5-3
General Cleaning...............................................................................................................................5-3
Daily Cleaning ....................................................................................................................................5-3
Cleaning the Platen ..................................................................................................................5-3
Cleaning the Stain Tubing and Cannula .....................................................................................5-4
Emptying the Waste Tank .........................................................................................................5-6
Weekly Cleaning................................................................................................................................5-6
Cleaning Drain Troughs and Rear Guide Rail .............................................................................5-6
Decontamination and Removal from Operation ............................................................................5-9
Section 5: Maintenance, Page 5—1
Page 46
Page 5—2, Section 5: Maintenance
Page 47

Section 5: Maintenance

The Hematek Slide Stainer is a precision instrument, designed to provide trouble-free operation with a minimum of maintenance.
BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.

General Cleaning

Keep the exterior surfaces of the instrument free of dust at all times. If needed, the exterior may be cleaned using a damp cloth and mild detergent. A small amount of methanol may be used to clean stain from the instrument.

Daily Cleaning

It is vitally important to clean the platen and tubing at least once each day in order to maintain consistently high quality staining results. It is also important to empty the waste tank at the end of each day.

Cleaning the Platen

After staining a large number of blood smears, precipitated stain solution tends to accumulate in the mixing grooves of the platen. This precipitate must be removed at regular intervals. Daily cleaning of the platen is imperative; after each run is optimal. Cleaning is especially important if bone marrow slides have been stained because the fat droplets can leave a greasy residue on the platen. Use the following procedure to clean the platen:
1. Turn the instrument off.
2. Carefully flood the working area of the platen with methanol.
Avoid splashing the methanol.
CAUTION
Always wipe the platen from right to left. Damage to the position or shape of the sensing switch fingers may result from not following recommended cleaning practices. Check the position of the fingers after cleaning as described in step 5 of Check and Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15.
Section 5: Maintenance, Page 5—3
Page 48
3. After priming, wipe the platen with a soft, disposable, lint-free absorbent cloth or tissue.
Wipe from right to left only.
Figure 5-1. Cleaning the Platen
CAUTION
If using frosted-end slides, it is especially important that the front guide rail be cleaned on a regular and frequent basis. This is because the stain may spread across the frosted portion of the slide to the front rail. If allowed to accumulate, slide breakage may occur.
4. Thoroughly clean the mixing grooves and the front guide rail.

Cleaning the Stain Tubing and Cannula

The tubing carrying the stain should be cleaned at least once daily (or after each run), as described in the following procedure:
1. Remove the stain, buffer, and rinse cannulas from the Stain Pak bottles.
NOTE:
The buffer and rinse cannulas and tubing do not require cleaning. These cannulas are removed so the buffer and rinse solutions are not wasted while the stain tubing is being cleaned.
Page 5—4, Section 5: Maintenance
Page 49
2. Using the prime function, purge the stain from the tubing.
a. Place the stain cannula, with tubing attached, into a small container of methanol
and lift the operating lever to
Figure 5-2. Rinsing the Stain Tubing
PRIME.
b. Holding the lever in this position, continue to pump methanol through the stain
tubing until it is thoroughly rinsed and clear solution appears on the platen.
3. After the stain tubing has been cleaned, remove the cannula from the methanol and continue to prime until the tubing is emptied of all methanol
CAUTION
Always wipe the platen from right to left. Damage to the position or shape of the sensing switch fingers may result from not following recommended cleaning practices. Check the position of the fingers after cleaning as described in step 5 of Check and Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15.
4. Carefully wipe the platen with a soft, disposable, lint-free absorbent cloth or tissue.
Wipe from right to left only.
Section 5: Maintenance, Page 5—5
Page 50

Emptying the Waste Tank

Empty the waste tank once each day, as well as after installing a new Stain Pak. To avoid spilling the contents while emptying the waste tank, carefully pull the tank away from the instrument, supporting the bottom of the tank to hold it level during removal. After emptying into an appropriate receptacle, rinse the waste tank with water and reinstall it into the instrument.
Figure 5-3. Removing the Waste Tank

Weekly Cleaning

Clean the drain troughs and the rear guide rail on a weekly basis.

Cleaning Drain Troughs and Rear Guide Rail

The back and front drain troughs of the stainer function as drains for excess stain, and residues from the reagents that may accumulate and interfere with proper drainage. The recommended procedure for proper cleaning is as follows:
1. Turn the instrument off and remove the line cord from the AC electrical outlet.
2. Raise the lid of the instrument.
Page 5—6, Section 5: Maintenance
Page 51
3. Loosen the two thumbscrews that are inserted through the notches in the back of the circuit board cover.
Figure 5-4. Location of Thumbscrews
4. Carefully raise the circuit board cover so the back trough is completely exposed and easily accessible.
Figure 5-5. Raising the Circuit Board Cover
5. Move the panel up and back, out of the way.
The circuit board cover is connected to the instrument by the two connectors of the
LOW STAIN and POWER lights. Rotate the panel carefully so the connectors are not
pulled loose.
Section 5: Maintenance, Page 5—7
Page 52
6. Flood both the front and back troughs with methanol to loosen any precipitated stain that may be present. Take care not to splash methanol onto the circuit board.
7. Using an applicator stick with a cotton swab attached, wipe from right to left along the length of the back and front troughs to remove the entire excess residue.
Figure 5-6. Cleaning the Back Trough
8. Clean other exposed areas that might be accidentally stained, including the rear guide rail, in the same careful manner.
NOTE:
Ensure that you do not scratch the platen.
9. Return the circuit board cover to its normal position, between the circuit board panel and the back edge of the screws, and tighten the two thumbscrews.
Be sure the panel is not resting on top of the screw heads before tightening.
Page 5—8, Section 5: Maintenance
Page 53

Decontamination and Removal from Operation

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to Appendix A, page A-1 for recommended precautions when working with biohazardous materials.
Use this procedure to remove the Hematek system from operation for extended periods and prior to packing and shipping the system to another location or to an off-site service facility.

1. Remove and dispose of all slides in the appropriate receptacle.

2. Follow the first five steps of the Replacement of the Pump Tubing on page 6-7, procedure to empty and remove the pump tubing and discard it in an appropriate waste container.

3. Lift the used Hematek Stain Pak carton out of the well at the rear of the instrument.

4. Follow steps 3 through 8 of the Replacement of the Underplaten Tubing on page 6-11 to turn off the instrument and remove the underplaten tubing. Discard it in an appropriate waste container

5. Clean the platen with methanol, always wiping from right to left.

6. Empty the waste tank, fill with 10% solution of household bleach and water, empty the solution, and then rinse the tank with regular water.

7. Clean the exterior surfaces of the instrument with a damp cloth and mild detergent. A small amount of methanol may be used to clean stain from the instrument.

8. Screw the front feet all the way up.

The instrument is now ready for storage or for packing and shipping.
Section 5: Maintenance, Page 5—9
Page 54
Page 5—10, Section 5: Maintenance
Page 55
Section 6: Minor Replacements and Adjustments ................................................. 6-3
Fuse Replacement ............................................................................................................................6-3
Replacement of Light Assemblies ...................................................................................................6-4
Replacement of the Pump Tubing...................................................................................................6-7
Replacement of the Underplaten Tubing........................................................................................6-11
Pump Volume Adjustment................................................................................................................6-14
Check and Correct Instrument Operation Prior to Pump Volume Adjustment .................................6-15
Adjust the Stain, Buffer, and Rinse Volumes ..............................................................................6-15
Volume and Ratio Determination.....................................................................................................6-17
Sensing Switch Fingers Adjustment................................................................................................6-18
Minor Replacements and Adjustments, Page 6—1
Page 56
Page 6—2, Section 6: Minor Replacements and Adjustments
Page 57

Section 6: Minor Replacements and Adjustments

This section is provided as an aid for performing minor replacements and adjustments on the Hematek Slide Stainer. Fully review and understand the procedures before attempting them, and they must be performed with care. For any adjustments or replacements other than those given in this section, or if any procedure appears to be too complex, refer to Supplies and Replacement Parts on page 8-6, for instructions on service for your instrument.

Fuse Replacement

Use the following procedure to replace the instrument line fuse for all voltage instruments. The fuse is 1.0 amp, 5mm x 20mm, Type T (Part No. 40151108).

1. Turn the instrument off and remove the line cord from the AC electrical outlet.

2. Using a small, blade-type screwdriver, pry the fuse cover open.

.
Figure 6-1. Opening the Fuse Cover
Section 6: Minor Replacements and Adjustments, Page 6—3
Page 58

3. Pull the fuse holder out of the instrument.

4. Remove the fuse from the holder and discard it into an appropriate waste container.

Figure 6-2. Replacing the Fuse
CAUTION
Use only the specified fuse to avoid damage to the instrument.

5. Replace the defective fuse with an identical fuse, snapping it into place. Fuse specifications are in Section 1: Introduction, while ordering information is in Section 8: Service, Supplies, and Replacement Parts.

A spare fuse, located in the small, enclosed compartment in the fuse holder, is shipped with the instrument. Push the fuse out of the compartment using a small screwdriver.

6. With the flat side up, return the fuse holder to its position in the instrument.

7. Firmly press the fuse cover until it snaps into place and is flush with the power module plate.

Replacement of Light Assemblies

Use the following procedure to replace the Power Light Assembly (PN 94000787) or Low Stain Light Assembly (PN 94001073).

1. Turn the instrument off and remove the line cord from the AC electrical outlet.

2. Raise the lid of the instrument.

Page 6—4, Section 6: Minor Replacements and Adjustments
Page 59

3. Loosen the two thumbscrews that are inserted through the notches in the back of the circuit board cover.

Figure 6-3. Location of Thumbscrews
4. Lift the panel from the front wall, exposing the printed circuit board and the connectors for the
LOW STAIN and POWER lights.
Figure 6-4. Raising the Circuit Board Cover
Section 6: Minor Replacements and Adjustments, Page 6—5
Page 60

5. Disconnect the wires from the circuit board for the light that is to be replaced.

Using a small, blade-type screwdriver, loosen the two small screws located on the top of the connector (the screws will not come out completely), then pull on the connector wires to remove them from the connector.
Figure 6-5. Removing the Connector Wires

6. Unsnap the burned-out light assembly from the circuit board cover by pinching together the two plastic retaining tabs on the light assembly.

If the tabs are too stiff, use a screwdriver to press against one side, then push that edge partially through the opening. Repeat with the other side.
Figure 6-6. Unsnapping the Light Assembly
Page 6—6, Section 6: Minor Replacements and Adjustments
Page 61

7. Discard the burned-out light assembly into an appropriate waste container.

8. Insert the new light assembly into the circuit board cover, threading the wires through the opening from the front of the cover and pressing firmly on the light until it snaps into place.

9. Insert the connector wires into the holes on the connector on the printed circuit board.

Each wire can go into either hole.

10. Tighten the small screws to just past finger tight.

11. Return the circuit board cover to its normal position, between the circuit board panel and the back edge of the screws, and tighten the two thumbscrews.

Be sure the panel is not resting on top of the screw heads before tightening.

12. Replace the line cord into the AC electrical outlet and turn the instrument on to check the light operation.

Replacement of the Pump Tubing

If any one of the three pumps fails to deliver the proper amount of solution at the adjusted volume setting, it is recommended that all pump tubes be replaced. Regular flushing of the stain tubing with methanol prolongs the life of the tubing. All pump tubes should be replaced after three Stain Paks have been used. If regular cleaning is not performed, or if usage is very heavy, the tubing should be changed more frequently. Use the following procedure to replace the pump tubes with new ones in the Hematek Pump Tube Set (PN
4482).

1. Remove the three cannulas from the Stain Pak.

2. Raise the operating lever to the the tubes.
Figure 6-7. Priming the Tubing
PRIME position until the reagents are pumped out of
Section 6: Minor Replacements and Adjustments, Page 6—7
Page 62

3. Push the operating lever down to the UNLOCK position.

4. Disconnect each pump tube from its cannula and from the recessed nipple located in the wall in front of the pumps.

Figure 6-8. Removing Tubing from Cannula
Figure 6-9. Removing Tubing from Nipple
Page 6—8, Section 6: Minor Replacements and Adjustments
Page 63

5. Remove each tube from the pump assembly.

a. Push the thumb tab on the pump arm to the left as far as possible.
b. While holding the thumb tab in this position, pull the plastic cuff on the tube until
the tube is completely removed from the pump arm.
c. Discard the tube in an appropriate waste container.
Figure 6-10. Removing Tubing from Pump Assembly

6. Remove the new tubes from the Hematek Pump Tube Set.

Each tube is numerically coded to correspond with the numbers shown on the Stain Pak.
Tube Code
stain tube
buffer tube
rinse tube
1
2
3

7. Attach the coded end of the new tube to its respective cannula.

a. Replace the cannula if it appears damaged or bent using the Hematek Cannula Set
(PN 4483).
b. If you are re-using any cannulas, make sure the cannula previously used for stain
is used only with the stain tubing.
Section 6: Minor Replacements and Adjustments, Page 6—9
Page 64

8. Insert tubing into the pump assembly.

a. Extend the new tube to its respective pump and thread the end into the hole in the
pump arm.
b. Push the thumb tab to the extreme left, as before, and push the tube through until
the plastic cuff is flush against the pump arm.
c. Release the pump arm.
NOTE:
If you encounter difficulty in threading the tubing through the pump housing, lift the operating lever to
PRIME for just a few seconds. This will cause the rollers inside the
pump housing to rotate slightly and relieve the interference.
Figure 6-11. Inserting Tubing into Pump Assembly

9. Connect the tube to its proper recessed nipple.

NOTE:
If difficulty is encountered in connecting a tube to its nipple, use forceps or a hemostat to grasp and attach the tube.
CAUTION
Take care to not damage the tubing when using forceps or hemostats.

10. Replace the cannulas into their respective reagent bottles.

Page 6—10, Section 6: Minor Replacements and Adjustments
Page 65

11. Prime the pumps until the solution in each tube is clear of all air bubbles.

NOTE:
New tubing is sometimes difficult to prime the first time and may need assistance. While holding the operating lever in the
PRIME position, push the pump arm inwards (toward the
pump) until the reagent fills the tubing, then release the pump arm. It may also be helpful for you to pinch the tubing several times with your fingers, pinching in the area between the cannula and the pump arm.

12. Verify the pump timings and reset the volume controls if necessary. See Pump Volume Adjustment on page 6-14.

Replacement of the Underplaten Tubing

The underplaten tubing is the tubing between the spout under the platen and the nipple behind the circuit board. The buffer and rinse tubing should be replaced after approximately ten Stain Paks have been used. The stain tubing needs to be changed more frequently, especially if it is not regularly flushed with methanol. Depending on usage and cleaning patterns, the stain tubing may need to be changed as often as after every four Stain Paks. Use the following procedure to replace the underplaten tubing with new tubing in the Hematek Underplaten Tubing pack (Product No. 4484).

1. Remove the three cannulas from the Stain Pak.

2. Raise the operating lever to the
PRIME position until the reagents are pumped out of
the tubes.
Figure 6-12. Priming the Tubing

3. Turn the instrument off and remove the line cord from the AC electrical outlet.

4. Raise the lid of the instrument.

Section 6: Minor Replacements and Adjustments, Page 6—11
Page 66

5. Loosen the two thumbscrews that are inserted through the notches in the back of the circuit board cover.

Figure 6-13. Location of Thumbscrews

6. Carefully rotate the panel up and back.

The circuit board cover is connected to the instrument by the two connectors of the
LOW STAIN and POWER lights. Rotate the panel carefully so the connectors are not
pulled loose.
7. Disconnect the
LOW STAIN and POWER lights from the circuit board, as directed in
step 5 of Check and Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15 in this section.
Page 6—12, Section 6: Minor Replacements and Adjustments
Page 67

8. Disconnect the stain tubing from the nipple that is located behind the circuit board.

Figure 6-14. Disconnecting Underplaten Tubing from Nipple

9. Remove the waste tank and reach underneath the platen through the waste tank area.

10. Disconnect the stain tubing from the platen by pulling the tubing free from behind the circuit board, and then disconnect it from the spout under the platen.

Figure 6-15. Removing the Underplaten Tubing

11. Select one of the new sections of underplaten tubing and connect it to the stain-tubing nipple behind the circuit board.

Make sure at least 7 mm (¼ inch) of tubing is connected onto the nipple.

12. Thread the tubing behind the circuit board through the channels provided until it extends under the platen.

Section 6: Minor Replacements and Adjustments, Page 6—13
Page 68

13. Connect the tubing to the stain spout under the platen, and then check all connections to make sure the tubing is securely in place.

Figure 6-16. Connecting Underplaten Tubing to Platen Spout

14. Repeat steps 8 through 13 for the buffer and rinse tubing.

Connect each tubing section to the appropriate nipple behind the circuit board and then connect the tubing to the appropriate spout under the platen.

15. Place the waste tank back in the instrument.

16. Connect the section.

17. Return the circuit board cover to its normal position, between the circuit board panel and the back edge of the screws, and tighten the two thumbscrews.

Be sure the panel is not resting on top of the screw heads before tightening.
LOW STAIN and POWER lights to the circuit board, as directed in this

Pump Volume Adjustment

The Hematek Slide Stainer is designed to produce stained slides of consistent quality in a continuous process. This is accomplished by having a fixed length of time in each of the three phases of stain, buffer, and rinse, as well as a predetermined ratio of stain-to-buffer volumes in the buffer phase.
The stain-to-buffer ratio is adjustable to provide ratios of at least 1:2 to 1:3. This feature allows the user flexibility in obtaining the desired stain results. A small change in the stain-to-buffer ratio can result in either lighter or darker staining. The amount of reagent pumped in each of the three phases is adjustable through the use of the volume control knobs located in the front control panel.
Page 6—14, Section 6: Minor Replacements and Adjustments
Page 69

Check and Correct Instrument Operation Prior to Pump Volume Adjustment

Follow these steps to check and correct instrument operation when slides are not satisfactory or pump volumes do not appear to be optimum. If any of these steps correct the situation, you do not need to adjust the pump volumes.
1. Replace both sets of tubing.
Refer to Replacement of the Pump Tubing on page 6-7 and Replacement of the Underplaten Tubing on page 6-11.
2. Clean the platen, including the front and back drain troughs, and make sure that there is no residue in any mixing groove.
Refer to Cleaning the Platen on page 5-3 for instructions to clean the platen and the drain troughs.
3. Observe the circular bubble level to make sure the instrument is level to provide optimal buffer mixing.
If leveling is required, follow the Level the Instrument on page 2-4.
4. Make sure the Stain Pak has been vented.
Refer to Install the Stain Pak on page 2-9.
5. Check the adjustment of the sensing switch fingers.
6. Observe several slides as they move across the platen and visually ensure all of the following events occur and they are activating the pumps at the proper time.
• The stain pump should be activated after approximately ½ of the slide width has
passed over the stain orifice.
• The buffer pump should be activated after about 1/8 inch of the slide has crossed
over the buffer orifice.
• The rinse pump should be activated just as the leading edge of the slide crosses over
the rinse orifice.
7. If the starting times for each phase are not correct, follow the procedure in this section.

Adjust the Stain, Buffer, and Rinse Volumes

Optimal staining results are achieved when the Hematek Slide Stainer is properly adjusted. Normally, proper adjustment of the instrument can be determined by observation of the amount of stain and buffer required to fill the capillary gap between the slide and the surface of the platen. If the proper amounts of stain and buffer are being pumped, proper mixing of the reagents will occur, resulting in uniform staining of blood smears. Use the following procedure to adjust the pump volumes.
1. Locate the pump volume control knobs, which are located in the control panel on the front-right corner of the instrument.
Each knob contains a centering catch on, which provides a central reference point.
The instrument is manufactured to give a stain-to-buffer ratio of approximately 1:2.5 and a rinse volume of approximately 1.0 mL/slide when each knob is at the centering catch.
Section 6: Minor Replacements and Adjustments, Page 6—15
Page 70
Volume is lowest when a knob is turned fully counterclockwise and highest when turned fully clockwise.
Volume increments are marked on each knob:
• Stain and buffer knobs: Each evenly labeled increment changes the volume by
approximately 0.1 mL per 10 slides (0.01 mL/slide).
• Rinse knob: Each evenly labeled increment changes the volume by approximately
0.5 mL per 10 slides (0.05 mL/slide).
2. Prime the instrument by placing the operating lever in the
PRIME position and holding
it there until solution is being pumped at all three stations (stain, buffer, and rinse).
3. Wipe the platen after priming, wiping from right to left only.
4. Prime the platen by loading five priming slides and allowing them to be transported across the platen.
This procedure primes and wets the platen to achieve proper surface tension characteristics.
5. Adjust the stain volume as follows:
a. Run five blood smear slides across the platen immediately following the priming
slides.
b. Adjust the stain control knob so the stain just fills, but does not overfill, the
capillary space between the platen and the slide.
c. If the slide is receiving an inadequate amount of stain, adjust the control knob
clockwise.
d. If overflow of solution is visible on either side of the slide or an excess of stain is
visible hanging over the front beveled edge of the platen, adjust the stain control knob counterclockwise.
The adjustment is correct when proper fill has been achieved on two consecutive slides.
6. Adjust the buffer volume as follows:
a. If there has been a gap of more than two consecutive slides across the platen since
the stain volume adjustment, re-prime the platen with three priming slides.
b. Run five blood smear slides across the platen immediately following the priming
slides.
When the slide passes the buffer section of the platen, mixing of the stain and buffer occurs because of the mixing grooves. This is where the proper stain-to- buffer ratio is most important. The slide looks under filled as it travels over the mixing grooves because the capillary gap is slightly greater.
7. Adjust the buffer control knob so the stain-buffer mixture just fills the capillary space under the slide when the slide is past the mixing grooves and onto the smooth portion of the platen, just before the rinse area.
Ensure that the stain-buffer mix does not extend out from either side of the slide.
Page 6—16, Section 6: Minor Replacements and Adjustments
Page 71
NOTE:
As the slide moves along, the buffer should pulse from the buffer orifice and move around the two mixing grooves for complete mixing of the stain and buffer. Too much stain and/or buffer will reduce the pulsing action, causing inconsistent staining across the slide. If partial flooding occurs, it should quickly re-drain as the slide continues to move across the grooves. Complete flooding of the second groove may indicate an excessive volume of either the stain or buffer.
8. Adjust the rinse volume as follows:
a. Examine the slides under a microscope.
b. If adjustment is necessary, use the rinse control knob to make the correction.
c. If excess stain is left on the slide appearing as precipitation under the microscope,
increase the rinse volume.
d. If the staining is pale due to over washing, reduce the rinse volume.
NOTE:
Any change in the rinse setting may cause a slight color change due to the small amount of methanol in the rinse.
9. Recheck the adjusted pump volumes:
a. Prime the platen with three priming slides.
b. Run at least three blood smears to ensure that the capillary spaces between the
platen and slide fill completely with stain at the stain area and with stain-buffer mixture just before the rinse area.
NOTE:
Occasional small voids may occur on the slide after proper filling. These voids are acceptable and will not adversely affect stain quality. Do not try to compensate for small voids by increasing the stain volume. Doing so may upset the optimum stain-to-buffer ratio and may affect the stain intensity.

Volume and Ratio Determination

Once adjustments have been made that appear visually to be giving appropriate slide results, the pump volumes and stain-to-buffer ratio can be measured as follows:

1. Remove the waste tank and disconnect the underplaten tubing leading to the stain, buffer, and rinse spouts on the platen.

2. Place the tubing ends into a beaker or other small container, and prime by using the operating lever.

3. As soon as all three-pump tubes have been primed and are free of air, place the free end of the stain and buffer tubes into separate 10 mL graduated cylinders. Place the free end of the rinse tube into a 25 mL graduated cylinder.

4. Process 10 blank slides and record the volume of stain, buffer, and rinse solution pumped into each cylinder. The following should be noted:

Section 6: Minor Replacements and Adjustments, Page 6—17
Page 72
The stain-to-buffer ratio should be approximately 1:2 to 1:3, (i.e., the volume of buffer should be about 2 to 3 times that of the stain).
Example:
1.6 mL of stain per 10 slides = 0.16 mL/slide
4.1 mL of buffer per 10 slides = 0.41 mL/slide
0.41 ÷ 0.160 = a ratio of 1:2.56 between stain and buffer
The rinse volume should be approximately 10 mL for the 10 slides.
A properly stained slide is the result of an interactive process involving the pump volumes, the mixing process, and the stain-to-buffer ratio. A minimum ratio of 1:2 is recommended; however, if the platen is filling properly, good mixing is occurring, and the stained slides are acceptable under the microscope, the stain-to-buffer ratio should not be adjusted.

5. When the determinations are complete, reconnect the three tubes to their appropriate spouts and replace the waste tank.

Sensing Switch Fingers Adjustment

The sensing switch fingers must be properly positioned in order to accurately activate the reagent pumps. If the reagents are not being dispensed, check the location of the fingers. Not following recommended cleaning practices may force the sensing switch fingers out of adjustment. Use the following procedure to adjust the sensing switch fingers.

1. Turn the instrument off and remove the line cord from the AC electrical outlet.

2. Raise the clear plastic lid on the instrument.

3. Loosen the two thumbscrews that are inserted through the notches in the back of the circuit board cover.

Figure 6-17. Location of Thumbscrews
Page 6—18, Section 6: Minor Replacements and Adjustments
Page 73

4. Carefully rotate the panel up and back.

The circuit board cover is connected to the instrument by the two connectors of the
LOW STAIN and POWER lights. Rotate the panel carefully so the connectors are not
pulled loose.

5. Check the location of each of the sensing switch fingers for the following two criteria:

Each finger should point straight down into the center of the back trough of the platen, without touching the bottom.
If it is not, bend the finger slightly forward or backward as necessary, as directed below.
The sensing switch should be activated before the finger is lifted over the top surface of the slide.

6. Lay a slide lengthwise along the back edge of the platen and slowly push the slide from right to left past the sensing switch finger, listening for a slight click as the sensing switch is activated.

If the click is not occurring before the finger has been lifted onto the top of the slide, bend the finger slightly to the right, as directed below.
Figure 6-18. Checking for Adjustment of Sensing Switch Fingers
Section 6: Minor Replacements and Adjustments, Page 6—19
Page 74

7. If adjustment to a sensing switch finger is necessary, support the wire near the center and carefully bend the lower part of the wire.

If bending to the right, bend the wire no more than 7 mm (1/4 in.) past perpendicular.
Figure 6-19. Adjusting the Sensing Switch Finger

8. Verify the final adjustment of each of the sensing switch fingers by processing several slides across the platen as in Step 5 of Check and Correct Instrument Operation Prior to Pump Volume Adjustment on page 6-15.

If these adjustments do not result in correct timing, contact your local technical support provider or distributor.
Page 6—20, Section 6: Minor Replacements and Adjustments
Page 75
Section 7: Troubleshooting ..................................................................................... 7-3
Checklist for Quality of Stained Blood Smears..............................................................................7-3
Troubleshooting Chart.......................................................................................................................7-4
Troubleshooting, Page 7—1
Page 76
Page 7—2, Section 7: Troubleshooting
Page 77

Section 7: Troubleshooting

The Hematek Slide Stainer is designed to give trouble-free operation when the directions for operating and cleaning the instrument are followed. If a problem occurs, refer to troubleshooting pointers in this section for help in solving the problem.

Checklist for Quality of Stained Blood Smears

When stained blood smears do not have the desired quality, the first tendency is to assume the staining reagents or the stainer is at fault. Frequently, however, other factors are the cause of the poorly stained blood smears. The following checklist can help isolate the problem when you question the quality of stained blood smears.
Are the pump volumes properly adjusted? Improper settings may lead to poorly
stained blood smears.
Is the blood old or does it contain an incorrect anticoagulant? Use fresh blood
(less than 8 hours old), EDTA is the anticoagulant of choice.
Was the blood thoroughly mixed before making the smear?Is the blood smear too thick, too thin, or spread unevenly?Does the feathered edge of the smear end at least 1/8 inch from the end of the
slide?
Was the blood smear thoroughly dry before staining?Are the slides clean? Even new slides are not necessarily clean.Are the slides flat? Variable thickness of a given slide may cause uneven staining.Have the Hematek Stain Pak containers been vented? Venting is necessary for
even delivery of the stain, buffer, and rinse reagents.
Has the reagent tubing on the Hematek Slide Stainer been primed before staining
the slides? Air bubbles in the tubing will cause poorly stained smears.
Has the platen been primed and wetted before running blood smears? Process two
to four blank slides across the platen to wet the platen surface before each run of blood smears.
Is the stain tubing, cannula, nipples, or orifice plugged? Daily cleaning of the stain
tubing is recommended to prevent plugging.
Are the platen and grooves clean and free of residue? A dirty platen or grooves
will result in uneven, low-quality staining.
Is the microscope clean and adequately illuminated? A dirty microscope or
inadequate lighting may give the impression of a poorly stained blood smear.
Section 7: Troubleshooting, Page 7—3
Page 78

Troubleshooting Chart

The following instrument troubleshooting chart lists possible problems relating to electrical and mechanical operations that could occur during instrument operation. Probable causes and recommended corrective actions are also included, so that many isolated problems can be quickly corrected. When dealing with any problem with the Hematek Slide Stainer, it is essential to determine which portion of the system is the source of the trouble. A systematic approach should be employed to isolate the problem.
If additional assistance is required concerning an instrument problem, copy and complete the Hematek Preservice Checklist in Appendix A, and contact your local technical support provider or distributor.
Only Hematek Stain Paks should be used with the Hematek Slide Stainer. Do not use any other stain solution. For problems or questions concerning the Stain Paks, contact your local technical support provider or distributor.
Symptom Possible Cause Corrective Action
Unit fails to turn on when the main instrument power switch is turned
POWER light
Green
ON
fails to illuminate but drive, pump, and fan motors operate
Pumps run but do not deliver solution during or after priming
Instrument not plugged into wall outlet or
Ensure line cord is plugged into outlet and instrument
instrument
Line fuse is blown Unplug the unit and replace the
fuse. Refer to Fuse Replacement on page 6-3.
Circuit breaker tripped
Reset the circuit breaker in electrical circuit of building
Instrument electrical failure
Contact your local technical
support provider or distributor
Light is burned out Replace light assembly. Refer
to Replacement of Light
Assemblies on page 6-4.
Open circuit in associated wiring
Stain Pak is empty or near empty
Contact your local technical
support provider or distributor
Replace with a new Hematek
Stain Pak. Refer to Install the
Stain Pak on page 2-9.
Page 7—4, Section 7: Troubleshooting
Tubing has accidentally been pulled off platen nipple
Place tubing back on the nipple
Page 79
Cannula openings are clogged
Remove cannulas from solution
bottles and clean away debris
with cloth and alcohol
Yellow
LOW STAIN
light is not functioning properly
Pump tubing is collapsed from use, or is perforated
Underplaten tubing (from the circuit board area to the platen) is clogged with foreign
Replace pump tubing. Refer to
Replacement of the Pump
Tubing on page 6-7.
Replace with new tubing. Refer
to Replacement of the
Underplaten Tubing on page 6-
11.
matter
Stain Pak is not properly positioned in the well
Reposition the Stain Pak so it
moves freely up and down on
the right end
Light is burned out Replace light assembly. Refer
to Replacement of Light
Assemblies on page 6-4.
Light switch is not operating correctly
Contact your local technical
support provider or distributor
for information on adjusting the
switch
Switch or associated circuit is defective
Contact your local technical
support provider or distributor
Improper mixing action between stain and buffer on platen
Uneven staining Mixing grooves on the
Section 7: Troubleshooting, Page 7—5
Instrument is not properly leveled
Inadequate pump volume setting
platen are dirty
Too much stain or buffer solution is being pumped causing poor mixing of stain and buffer
Adjust the front feet of the
instrument so the bubble is
centered in the circular bubble
level. Refer to Level the
Instrument on page 2-4.
Refer to Pump Volume
Adjustment on page 6-14.
Contact your local technical
support provider or distributor
if unable to adjust properly.
Clean mixing grooves. Refer to
Cleaning the Platen on page 5-
3.
Refer to Pump Volume
Adjustment on page 6-14
Page 80
Slides are of variable thickness
Replace slides
Pale staining Proper stain-to-buffer
ratio has not been achieved
Stain volume is adjusted too high. Increasing stain volume may decrease stain intensity
Rinse volume is too high Refer to Volume and Ratio
Deteriorated or outdated Stain Pak
Refer to Volume and Ratio
Determination on page 6-17
Refer to Pump Volume
Adjustment on page 6-14
Determination on page 6-17
Replace Stain Pak
Page 7—6, Section 7: Troubleshooting
Page 81
Section 8: Service, Supplies, and Replacement Parts .......................................... 8-3
When to Call for Service...................................................................................................................8-3
For Service .........................................................................................................................................8-3
Siemens Authorized Representative .........................................................................................8-3
Siemens Offices Worldwide ....................................................................................................8-4
Returning the Instrument for Repair, Exchange, Replacement, or Loaner................................8-7
Supplies and Replacement Parts ....................................................................................................8-8
Hematek Stain Pak – Modified Wright’s Stain .............................................................................8-8
Hematek Stain Pak – Modified Wright-Giemsa Stain ...................................................................8-8
Hematek Cannula Set ..............................................................................................................8-8
Hematek Pump Tube Set .........................................................................................................8-9
Hematek Underplaten Tubing ...................................................................................................8-9
Ordering Information for Supplies and Replacement Parts ...........................................................8-10
Service, Supplies and Replacement Parts, Page 8—1
Page 82
Page 8—2, Section 8: Service, Supplies and Replacement Parts
Page 83

Section 8: Service, Supplies, and Replacement Parts

When problems arise during operation of the Hematek Slide Stainer, refer first to Section 7: Troubleshooting. Avoid problems by carefully following proper operating and cleaning procedures.

When to Call for Service

A call for assistance is appropriate under the following circumstances:
• If the problem cannot be solved by performing the steps described in the Troubleshooting Chart on page 7-4.
• If additional assistance is required concerning an instrument or reagent problem.
• If the problem is beyond the scope of this manual.
Before calling for service, copy and complete the Hematek Preservice Checklist in Appendix A.

For Service

To contact the legal representative for Siemens within the European community, contact the Siemens Authorized Representative. For service, contact your local technical support provider or distributor.
Siemens Authorized Representative
Section 8: Service, Supplies, and Replacement Parts, Page 8—3
Page 84
Siemens Offices Worldwide
www.siemens.com/diagnostics
Page 8—4, Section 8: Service, Supplies, and Replacement Parts
Page 85

Returning the Instrument for Repair, Exchange, Replacement, or Loaner

When instructed to return the instrument to Siemens for repair, exchange, replacement or the return of a loaner, follow these steps:
1. Move the

2. Follow all steps in Decontamination and Removal from Operation on page 5-9.

3. Pack the instrument in the original bag and container, or as instructed by your local technical support provider or distributor.

4. Seal the container securely.

5. Return the container to your local technical support provider or distributor.

NOTE:
If Siemens does not receive your defective instrument within 15 days, you will be billed for the replacement.
ON / OFF switch to the OFF position.
Section 8: Service, Supplies, and Replacement Parts, Page 8—5
Page 86

Supplies and Replacement Parts

To obtain the best results with the Hematek Slide Stainer, use only Hematek brand supplies.
Hematek Stain Pak – Modified Wright’s Stain
The Stain Pak contains a polychrome methylene blue-eosin stain and specially prepared buffer and rinse solutions. The Stain Pak is contained in a single carton that fits easily into the well inside the Slide Stainer. Also supplied are buffer and rinse solutions that have been optimally designed for use on the Hematek system. Each Stain Pak contains sufficient solutions to stain approximately 1000 slides.
WARNING
Highly Flammable! Toxic! Contains Methanol. Toxic by inhalation, in contact with skin and if swallowed. Danger of very serious irreversible effects through inhalation, in contact with skin and if swallowed. Keep container tightly closed. Keep away from sources of ignition. Avoid contact with skin. Wear suitable protective clothing and gloves. In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible).
Hematek Stain Pak – Modified Wright-Giemsa Stain
This is a modified Wright-Giemsa stain for laboratories that prefer a Giemsa stain component for preparation of their blood smears. This Stain Pak contains a modified polychrome methylene blue-eosin stain based on the original stain proposed by Romanowsky. Also supplied are buffer and rinse solutions that have been optimally designed for use on the Hematek system. Each Stain Pak contains sufficient solutions to stain approximately 1000 slides.
WARNING
Highly Flammable! Toxic! Contains Methanol. Toxic by inhalation, in contact with skin and if swallowed. Danger of very serious irreversible effects through inhalation, in contact with skin and if swallowed. Keep container tightly closed. Keep away from sources of ignition. Avoid contact with skin. Wear suitable protective clothing and gloves. In case of accident or if you feel unwell, seek medical advice immediately (show the label where possible).

Hematek Cannula Set

The Cannula Set is available as a replacement item. The cannulas are positioned in the Stain Pak bottles and connected to the pump tubing.
Page 8—6, Section 8: Service, Supplies, and Replacement Parts
Page 87

Hematek Pump Tube Set

The Pump Tube Set consists of three tubes, each labeled with an identifying number specific to the reagent to be carried in the tubing. Each tube is a specific diameter and length to ensure precise measurement of the reagents. The tubing is made of a special type of rubber that is resistant to leaching and brittleness.

Hematek Underplaten Tubing

The Underplaten Tubing is available for periodic replacement. The tubing set consists of three tubes of equal diameter and length and, therefore, are not specific to a reagent.

Ordering Information for Supplies and Replacement Parts

To order the following Hematek brand products, contact your local technical support provider or distributor.
Part Number Description
4481 Hematek Stain Pak – Wright’s Stain
4405 Hematek Stain Pak – Wright-Giemsa Stain
4482
A Hematek Pump Tube Set
4
A Hematek Cannula Set
483
484
4
A Hematek Underplaten Tubing
40151108 Line fuse – 1.0 amp, 5mm x 20mm, Type T
94000787
94001073
POWER Light Assembly (Green)
LOW STAIN Light Assembly (Yellow)
Section 8: Service, Supplies, and Replacement Parts, Page 8—7
Page 88
Page 8—8, Section 8: Service, Supplies, and Replacement Parts
Page 89

Appendix A: List of Symbols

Symbols Used With This System

The following table explains the symbols used on the Hematek Slide Stainer, on the Stain Pak, and in this document.
Symbol Description
This symbol alerts you to a potential biohazard.
All products or objects that come in contact with human or animal body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of infectious agents in health-care settings as recommended for potentially infectious specimens in Protection of
Laboratory Workers from Infectious Disease Transmitted by Blood, Body Fluids, and Tissue, 2d edition; Approved Guideline (1997) Document
M29-A, National Committee for Clinical Laboratory Standards (NCCLS). This document contains complete information on user protection and it can be used as reference material for instructions on laboratory safety.
These symbols are used for both Warning and Cautions. A warning indicates the risk of personal injury or loss of life. A caution indicates the possibility of loss of data or damage to or destruction of equipment.
Appendix A: List of Symbols, Page A—1
This symbol indicates that the content of the package is toxic.
This symbol indicates that the content of the package is flammable.
This is the symbol for the On/off position indicator for the power switch
This is the symbol for the prime position of the stainer operating lever.
Page 90
This is the symbol for the unlock position of the stainer operating lever
This is the symbol for the lock position of the stainer operating lever
This symbol indicates that the product has a temperature limitation. In this example, you need to store the product between 15 and 30 °C.
Identifies Hematek stain.
Identifies Hematek buffer.
Identifies Hematek rinse.
Indicates that there is power to the instrument.
Indicates low stain.
This symbol indicates that you should consult the instructions for use.
This is the fuse symbol.
This symbol indicates that the product is UL approved for safety (United States and Canada).
This symbol indicates that the product is CSA approved for safety (Canada).
This symbol indicates that the product complies with the applicable directives of the European Union.
Page A—2, Appendix A: List of Symbols
This symbol indicates the serial number of a part or product.
Page 91
This symbol indicates the revision letter of a part or product.
This symbol indicates the number used for ordering a part or product.
This symbol indicates the expiration date of the product.
This symbol indicates the lot number of the product.
This symbol indicates the name and location of the product manufacturer.
This symbol indicates the date of the manufacture of the product.
This symbol indicates the manufacturer's authorized representative within the European community.
This symbol indicates an in vitro diagnostic device.
This symbol informs you not to use the product if the package is damaged.
This symbol is intended to encourage recycling.
This symbol is intended to encourage recycling.
This symbol indicates that the materials are recycled.
Appendix A: List of Symbols, Page A—3
Page 92
This symbol indicates that the product or container should be oriented in the direction of the arrows.
Page A—4, Appendix A: List of Symbols
Page 93

Appendix B: Hematek Preservice Checklist

For reference, record the following information.
Instrument Serial Number
Installation Date
Has Section 7, Troubleshooting, been reviewed?
In the unit plugged into a live AC electrical outlet?
Ye s N o
Does the
Has the line fuse been checked and replaced if defective?
Is the slide drying system operating properly?
Are the conveyor spirals operating?
Is the
Do all three stations pump reagent properly?
Do all stations pump reagent at the proper time?
Is the instrument level?
Has the platen been cleaned each day?
Has the stain tubing been cleaned each day?
Are the sensing fingers properly adjusted?
When was the pump tubing last changed?
POWER light illuminate when the power switch is ON?
LOW STAIN light operating properly?
When was the underplaten tubing last changed?
Stain Pak Information (Check one) Wright’s Wright-Giemsa
Give a brief description of the problem:
Appendix B: Hematek Preservice Checklist, Page B—1
Stain:
Buffer/Rinse:
Lot #: Exp. Date:
Page 94
Page B—2, Appendix B: Hematek Preservice Checklist
Page 95

Index

Numerics
12 volts DC, line current converted 1- 12
A
adjust reagent 1- 8 adjustments
buffer volume 6- 15 pump volume 6- 14 rinse volumen 6- 15 sensing switch fingers 6- 18
stain volume 6- 15 Assemblies, pump 1- 9 authorized service representative 8- 3
D
daily cleaning
platen 5- 3 stain tubing and cannula 5- 4
waste tank 5- 6 decontamination 5- 9 drain trough
cleaning 5- 6 drawer, slide
description 1- 7
E
environmental factors
humidity range 2- 3
temperature range 2- 3
B
blood smears
loading the blood smear slides 3- 5 positioning smears 1- 6 procedure 4- 3 quality checklist 7- 3 staining the smears 3- 6
blood stains
making quality stains 4- 3
bone marrow smears
making quality smears 4- 5 procedure 4- 5
C
calling for service 8- 3 cannulas
cleaning 5- 4 cannulas 1- 9 characteristics
physical 1- 5 checklist for quality of stained blood smears 7­3 circular level 1- 5 cleaning
emptying the waste tank 5- 6
platen 5- 3
stain tubing and cannula 5- 4
tubing after use 3- 7 conveyor spirals
loading slides 1- 6
F
feet, levelers 1- 5 fuse
replacement procedure 6- 3
fuse holder, power module 1- 13
G
general cleaning 5- 3 grooves, opposing (conveyor spiral) 1- 6 guidelines for optimal staining 3- 3
H
Hematek
Pump Tube Set, description 1- 9
underplaten tubing 8- 9 Hematek cannula set 8- 8 Hematek pump tube set 8- 9 Hematek Stain Pak
Modified Wright’s Stain 8- 8
Modified Wright-Giemsa stain 8- 8 humidity range 2- 3
I
instrument
location 2- 3
return procedure 8- 7 intended use 1- 3
K
knobs
volume control 1- 8
Index-1
Page 96
L
levelers
circular 1- 5 feet 1- 5
lever
operating, in electrical system 1- 11 light 1- 11 light assemblies
replacement procedure 6- 4 line cord receptacle, power module 1- 12 line voltage 1- 8 loading the blood smear slides 3- 5 location of instrument 2- 3 Low Stain light 3- 7
M
maintenance
cleaning drain troughs/rear guide rail 5- 6
cleaning the tubing after use 3- 7
daily cleaning
cleaning the platen 5- 3 cleaning the stain tubing/cannula 5- 4
emptying the waste tank 5- 6 general cleaning 5- 3 Stain Pak replacing 3- 7
Modified Wright’s Stain 8- 8 Modified Wright-Giemsa stain 8- 8
N
numbering
on pump tubes 1- 9
O
offices, Siemens 8- 4 operating lever
electrical system 1- 11
operating procedures
clean the tubing after user 3- 7 load the blood smear slides 3- 5 prime the tubing 3- 4 stain the smears 3- 6 start the instrument 3- 3 turn the instrument off 3- 7
operation
theory of 1- 4
opposing grooves 1- 6
ordering supplies and replacement parts 8- 10
P
parallel spirals, conveyor 1- 6 peripheral blood smears 4- 3 physical characteristics 1- 5 platen
cleaning procedure 5- 3 description 1- 6
main functions 1- 6 position of blood smear 1- 6 power light
electrical system 1- 12 power module
description 1- 12 priming the tubing
first time priming 6- 11
procedure 3- 4
when to 3- 4 Pump
Assemblies 1- 9 pump assemblies
description 1- 8
four rollers 1- 8
pump arm -1 8
pump cap 1- 8
pump motor 1- 8 Pump Tube Set 1- 9 pump tubing
numbering 1- 9
replacement procedure 6- 7 pump volume
adjustment 6- 14 pump volume adjustment
adjust stain, buffer, and rinse volumes 6- 15
correct instrument operation 6- 15
R
ration determination 6- 17 reagents
adjusting 1- 8 rear guide rail
cleaning 5- 6 removal from operation 5- 9 replacement parts
ordering information 8- 10
Index-2
Page 97
replacements and adjustments
fuse replacement 6- 3 light assemblies 6- 4 pump tubing 6- 7 underplaten tubing 6- 11
volume and ration determination 6- 17 replacing the Stain Pak 3- 7 returning the instrument to Siemens 8- 7
S
sensing switch fingers adjustment 6- 18 sensing switches
description 1- 10
illustration 10 service
when to call 8- 3 service calls 8- 3 service representative 8- 3 Siemens offices
authorized representatives 8- 3 worldwide offices listing 8- 4 slide drawer
description 1- 7 slide drying system, description 1- 7 slide transport 1- 6 slide transport system 1- 6 slides
loading 1- 6 Specifications
dimensions 1- 13
fuse ratings 1- 13
line leakage current 1- 13
power requirements 1- 13
throughput 1- 13
weight 1- 13 specimens
overview 4- 3 spirals, conveyor 1- 6 Stain Pak
replacement 3- 7 stain tubing
cleaning 5- 4
staining the smears 3- 6 stain-to-buffer ratio 6- 18 starting the instrument 3- 3 supplies
Hematek cannula set 8- 8 Hematek pump tube set 8- 9 Hematek Stain Pak
Modified Wright’s Stain 8- 8
Modified Wright-Giemsa stain 8- 8 Hematek underplaten tubing 8- 9 ordering information 8- 10
symbols
used with system 1- 1
system waste 1- 14
T
temperature range 2- 3 theory of operation 1- 4 transport system 1- 6 troubleshooting
chart 7- 4 checklist for blood smear quality 7- 3
troubleshooting chart 7- 4 turning the instrument off 3- 7
U
underplaten tubing
replacement procedure 6- 11
V
volume control knobs 1- 8 volume control panel 1- 8 volume determination 6- 17
W
waste disposal 1- 14 waste tank
description 1- 7 emptying procedure 5- 6 when to empty and rinse 1- 7
weekly cleaning
drain troughs 5- 6 rear guide rail 5- 6
Index-3
Page 98
Index-4
Loading...