The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.07
2005
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
All tools, measurement devices, and accessories with the exception of those marked with a "*" are listed along with their specifications in the STC (Service Tools Catalog).
•Torque wrench 20 Nm to 100 Nme.g. part no. 80 86 159 RE999
•Optimol Viscogen KL 300, 50 ml Part no. 72 79 107
•Sealing compound*Part no. 20 49 716 SRN 6002
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 6 of 98
Medical Solutions
Prerequisites7
Text emphasis1.1
DANGER
WARNING
CAUTION
NOTICE
DANGER indicates an immediate danger of death or serious physical injury.
¹ n.a.
WARNING indicates a danger that may lead to death or serious
physical injury.
¹ n.a.
CAUTION used with the safety alert symbol indicates a risk of
minor or moderate physical injury and/or damage to property.
¹ n.a.
NOTICE used without the safety alert symbol indicates a risk that
if disregarded will lead or may lead to a situation which may result
in an undesirable result or state other than death, physical injury
or damage to property.
NOTE
¹ n.a.
NOTE contains information provided with special emphasis to
facilitate proper use of the equipment or proper execution of a
procedure, i.e. hints, tips.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 7 of 98
8Prerequisites
Symbols1.2
XChecks and adjustments that must be performed with radiation ON are identified by the
radiation warning symbol.
VThis symbol means "Dangerous voltage".
CThis symbol means "Attention, refer to the documentation".
EThis symbol indicates components sensitive to electrostatic discharge (ESD).
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 8 of 98
Medical Solutions
Prerequisites9
Service, shutdown, hibernation, handover to the customer1.3
WARNING
NOTE
Switch off via hibernation (version VB13C and later) after service
or before handover to the customer is not sufficient.
Various error messages can appear after the next system boot
and configuration changes are not adopted.
¹ For correct saving of the changed configurations, a shut-
down must be performed. Shut down the system via the
upper monitor menu bar <Options>-<End Session>-<Shut Down System> and then press the off key on
the monitor trolley.
Before the system is handed over to the customer, the system
must be shut down via the menu bar <Options>-<End Session>-<Shut Down System>.
Switch off via hibernation (on/off button on the monitor trolley) is
not sufficient.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 9 of 98
10Prerequisites
Safety information and protective measures1.4
General safety information (in existing documents)0
WARNING
Danger of injuries, death or material damage.
Non-compliance can lead to death, injury or material damage.
Please note:
¹ The product-specific safety notes in these instructions,
¹ The general safety information in TD00-000.860.01... and
¹ The safety information in accordance with ARTD Part 2.
General electrical safety information0
WARNING
Electrical safety!
Non-compliance can lead to severe injury or even death, as well
as material damage.
¹ Parts under electrical voltage are accessible when the
covers are open. To avoid danger, disconnect the system
from the power supply before opening the covers. Disconnect the power plug.
CAUTION
¹ If an uninterruptible power supply (UPS) is installed in
the system, the voltage output of the UPS must also be
deenergized or the voltage output plug must be disconnected.
¹ If work steps must be performed using electrical power,
the general safety information according to
TD00-000.860.01 must be observed.
Electrical voltage!
Non-compliance can result in material damage.
¹ When working on the system, ESD regulations must be
observed.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 10 of 98
Medical Solutions
Prerequisites11
Radiation safety information0
WARNING
X¹ Checks requiring the release of radiation are identified by
X-ray radiation!
Non-compliance can lead to illness, irreversible damage to body
cells and the genotype, severe injury and even death.
During work on the system in which radiation must be released,
the radiation protection directives and the rules for radiation protection according to ARTD-002.731.02.. must be complied with.
¹ Stay as far away as possible from the radiation source.
¹ Release radiation only if necessary.
¹ Set the radiation activity as low as possible. (Low kV and
mA values, short radiation time)
¹ Release radiation for as short a time as possible.
the radiation warning symbol shown on the left.
Safety information, mechanical0
CAUTION
Danger of burns from hot parts or components!
Non-observance can lead to mild to moderately severe burns, particularly on the hands.
¹ When the covers are opened, parts and components (e.g.
power supply components, heat sinks, electromagnetic
brakes) are accessible which during operation can reach
temperatures of > 50 degrees Celsius. To avoid burns,
switch off the system and let it cool down for at least 5
minutes before touching any parts or components.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 11 of 98
12Prerequisites
CAUTION
Danger of injuries from mechanical parts!
Non-observance can lead to mild to moderately severe injuries,
particularly to the hands.
¹ When the cover panels are opened, parts such as plugs,
threaded bolts, cut-off cable ties, and sharp component
edges are exposed. Inattention can lead to skin injuries
(crushing, scrapes, cuts), particularly to the hands.
¹ Work steps carried out on such parts and components
must be performed with particular attention and caution.
¹ Wear work gloves if necessary.
Safety information - risk of infection0
WARNING
Risk of infection due to pathogens!
Non-compliance can lead to severe injury and even death.
¹ This product can be contaminated with infected blood or
other bodily fluids.
¹ Avoid all contact with blood or other bodily fluids!
¹ Strictly observe the safety information in
ARTD-002.731.37.. regarding prevention of infectious diseases during customer service calls.
Laser light localizer option0
CAUTION
NOTE
Laser emissions!
This product contains class 2 lasers. (USA: Laser class 2)
Non-observance can lead to injury, particularly to the retina of the
eye, and can thus lead to irreversible visual impairment.
¹ Observe the safety information in ARTD-002.731.03...
When using the laser light localizer, do not look directly
into the laser beam.
Laser emissions!
There is no immediate danger to the eye (blinking reflex). Nevertheless, do not look directly into the laser beam.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 12 of 98
Medical Solutions
Prerequisites13
Information on the protective conductor resistance test0
Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
The protective conductor resistance must be measured after every intervention in the system.
However, documentation of the measured values is required only during periodic safety
checks.
If parts/components that can significantly influence the protective conductor resistance
(e.g., replacement of the power cable, replacement of the power-up module, replacement
of multi-pole connection cables which also create the protective conductor connection
between system parts (e.g., monitor cable or C-arm cable)) are replaced or if protective
conductor connections have been repaired, the protective conductor resistance must be
measured. The values must be documented and evaluated in the protective conductor
resistance protocol.
NOTE
For evaluation purposes, the first measured value and the values
documented during maintenance or safety checks must be compared to the measured values. A sudden or unexpected increase
in the measured values may indicate a defect in the protective
conductor connections - even if the limit value of 0.2 ohms is not
exceeded. (Protective conductor or contacts).
The measurement must be performed according to DIN VDE 0751, Part 1 (see ARTD Part
2). The protective conductor resistance for all touchable conductive parts must be measured during the normal operating state of the system.
Make sure that control cables or data cables between the components of the system are
not mistaken for protective conductor connections.
During the measurement, the power cable and additional connection cables which also
create the protective conductor connection between system parts (e.g. monitor cable
between the basic unit and monitor trolley) must be moved section by section to detect
cable breaks.
The protective conductor resistance must not exceed 0.2 Ohms.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 13 of 98
14Prerequisites
Fig. 1:Measuring circuit for measuring the protective conductor resistance for units that are
disconnected from power, in compliance with DIN VDE 0751-1/2001-10, Fig. C2.
Pos. 1= Syste m
Pos. 2= Application part type B (if available)
Pos. 3= Measurement setup (integrated into measuring device)
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 14 of 98
Medical Solutions
Prerequisites15
System leakage current measurement information0
NOTE
If parts in the primary circuit (e.g. power cable, line filter, power
transformers, or complete ON/OFF assemblies) are replaced during service work, the system leakage current measurement must
be subsequently performed and recorded as a repeat measurement.
However, the first measured value must be newly determined and
a new protocol be must created under the following conditions:
- Lack of system leakage current measurement documentation.
- When local line voltage or line frequency deviates from the line
voltage and line frequency values documented in the protocol
(e.g., in the event of a site/operator change)
- When a different procedure for measuring the system leakage
current than the one documented in the protocol is used.
For the purpose of traceability, reference to the new protocol must
be written in the old protocol. The reason for newly determining
the first measured value must be documented and confirmed with
a name and signature.
Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 15 of 98
16Prerequisites
WARNING
Electrical voltage!
Non-compliance can lead to severe injury and even death.
¹ The leakage current measurement may be performed on
systems of protection class I only after the protective
conductor test has been passed.
First measured value
The first measured value was already determined and documented in the system leakage
current protocol. The measuring procedure was also recorded.
The measurement was performed with the recorded line voltage, line frequency and with
the recorded measuring equipment.
Measurement
Perform the measurement according to DIN VDE 0751, Part 1 (see ARTD-002.731.17....),
and record the determined value.
The measuring procedure indicated in the protocol must be used.
If the first measured value has to be newly determined (see previous information), a measuring procedure can be selected (direct measurement or differential measurement).
Measurement of the system leakage current according to the differential current method
(measurement setup according to (Fig.2/p.16)) must be given preference, since this
method is not dangerous to the person performing the measurement and other persons.
However, please note the minimum resolution of the leakage current measuring instrument and any additional manufacturer's data restricting the use of the measuring device.
Fig. 2:Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 16 of 98
Medical Solutions
Prerequisites17
class I.
Pos. 1= Syste m
Pos. 2= Application part type B (if available)
Pos. 3= Measurement setup (integrated into measuring device)
If the direct measurement of the system leakage current is used (measurement setup
according to (Fig.3/p.17)), the system must be insulated during the measurement and
must not be touched.
Fig. 3:Measuring circuit for direct measurement of the system leakage current in compliance
with DIN VDE 0751-1/2001-10, Fig. C5 for protection class I.
Pos. 1= Syste m
Pos. 2= Application part type B (if available)
Pos. 3= Measurement setup (integrated into measuring device)
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 17 of 98
18Prerequisites
WARNING
Electrical voltage!
Non-compliance can lead to severe injury and even death.
¹ No housing parts of the system may be touched during
direct measurement of the leakage current (measurement setup according to (Fig.3/p.17)).
¹ Third-person access to the system must be prevented.
The system must be switched on during measurement. Measuring devices with automated measuring sequences must therefore be set to manual measurement.
Enter the highest value in the system leakage current protocol.
This value must not exceed the permissible system leakage current values according to
DIN VDE 0751-1/2001-10, Table F.1, line "system leakage current for units according to
remarks 1 and 3" of 2.5 mA.
Do not exceed 2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a measurement at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
In the case of repeat measurements, the measured value must also be evaluated.
NOTE
For evaluation purposes, the first measured value and the values
documented during maintenance or safety checks must be compared to the measured values. A sudden or unexpected increase
in the measured values may indicate that a fault has occurred in
the primary power supply circuit (insulation damage, damage
from moisture, defective interference suppressor, etc.) - even if
the limit value of 2.5 mA is not exceeded.
.
The evaluation is not necessary in the case of a new determination.
File the protocol sheet in the system binder or log book.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 18 of 98
Medical Solutions
Prerequisites19
Information on navigation systems installed as options0
This ARCADIS system can be used in combination with navigation systems supplied by
various navigation system manufacturers.
The use of the navigation system lies within the responsibility of the customer.
Some service tasks can affect the accuracy of the navigation system.
WARNING
Service work steps that can affect the accuracy of the installed navigation system:
Some service work steps performed on the ARCADIS system will
lead to inaccuracy in a navigation system installed optionally.
Failure to take the necessary steps to address this situation can
lead to severe injuries in the patient.
¹ If service work steps are performed on the ARCADIS sys-
tem that can affect the accuracy of optionally installed
navigation systems (see list below), the customer must
be notified verbally and/or in writing that the accuracy of
the installed navigation system is no longer guaranteed
after such service work steps have been performed, and
that the accuracy of the navigation system must be
checked and certified before it is used again.
•All work steps that affect the geometry of the C-arm and its components, such as:
¹ Removing or installing the I.I.
¹ Removing or installing the I.I. housing
¹ Removing or installing the I.I. mounting ring on the I.I.
¹ Removing or installing the I.I. grid
¹ Removing or installing the I.I. optics
¹ Removing or installing the CCD camera
¹ Removing or installing any mechanical components of the C-arm
•Any adjustments that alter the geometry of the imaging components, such as:
¹ Adjustment of camera optics (optical sharpness, image size)
¹ Adjustment of camera rotation (positioning with respect to I.I. optics)
¹ Adjustment of I.I. geometry and sharpness (I.I. mini voltage supply)
•Any subsequent installation of released options that affect the geometry of the C-arm
¹ 2D Navigation Option (Attachment of altered mounting ring to I.I.)
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 19 of 98
20Prerequisites
Cover panels1.5
NOTE
Observe the safety information in this chapter!
Main system, rear cover0
Removing the rear cover
1. Engage the foot brake.
2. Remove the screws from the rear cover.
3. Pull the cover back approximately 25 cm.
4. Unscrew the ground wire from the cover panel.
5. Pull the cover completely off and tilt it down.
6. Lift up the cover and raise both lateral metal brackets from the guide rails.
Installing the back cover
1. Fit the lateral metal brackets back into the guide rail.
2. Reattach the protective conductor to the rear cover.
3. Lift the cover and push it forward. Be careful with the EMC spring contact.
4. Reattach the cover and tighten the screws.
Main system, SIREPHOS cover panel0
NOTE
1. Loosen the cover screws of the SIREPHOS cover.
2. Use a sharp knife to cut open the sealing compound all around the cover.
3. Lift off the SIREPHOS cover.
4. Remove the sealing compound residue from the cover and from the SIREPHOS.
5. On completing service, place the SIREPHOS cover on the SIREPHOS and completely
seal it with sealing compound.
6. Use a paper towel to wipe off excess sealing compound.
7. Once again, use sealing compound 2049716 to seal the SIREPHOS cover completely.
8. Refasten the SIREPHOS cover with the cover screws. If necessary, apply side pressure to the SIREPHOS cover until the sealing compound hardens.
The SIREPHOS cover panel is sealed all around against dirt contamination using a sealing compound (part no. 20 49 716).
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 20 of 98
Medical Solutions
Prerequisites21
Monitor trolley0
Removing the cover panels
1. Remove the 4 screws from the middle back cover to access the imaging system (rear).
2. Remove the upper back cover plate of the log book compartment.
3. Remove the lower back cover to access the power supply.
4. Remove the front cover to access the imaging system (front).
Installing the cover panels
1. Install the covers in reverse order.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 21 of 98
22Prerequisites
Concluding steps1.6
NOTE
Observe the guidelines on the protective conductor test and the
leakage current measurement in this chapter.
•If necessary, perform the leakage current measurement.
•After completing all work steps and closing all the cover panels, perform the protective
conductor test in accordance with ARTD-002.731.17...
¹ The protective conductor resistance must not exceed 0.2 Ohms.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 22 of 98
Medical Solutions
Main System - Replacing Components23
D1 control board0
2-2Main System - Replacin g Components
WARNING
Replacement and additional work steps0
1. Make sure that a current backup of the main system is available.
2. Read out the load counter of the Sirephos (Path: <Service>-<Main System>-<Adjustment>-<Load Counter>).
3. Document this value because it is necessary for a Sirephos replacement. If the D1 is
defective, the load count is available in the offline report (path:<Service>-<Reports>-<Main System>-<Configuration Offline").
4. Disconnect and replace the D1 board.
Electrical voltage!
See chapter 1, Safety Information.
¹ Make sure that LED V400 on the D2 board is off prior to
performing any work on boards D1 and D2. This should
occur approximately 3 minutes after switching OFF the
ARCADIS system.
5. Install the new D1 board and reconnect all cable connections. Be sure that the shielding and ground wire connections are connected correctly.
6. Set the jumpers and switches according to the wiring diagram.
7. Perform the download procedure for the D1 control board (path: <Service>-<Main System> -<Download>-<C-Arm>).
NOTE
If the EE-PROM of the replacement D1 is not completely deleted, it
will be automatically cleared after the host software has been
downloaded. The delete procedure is indicated as "d" on the
7-segment display and requires approximately 5 minutes. During
the delete procedure, do not switch off the system or perform additional programming. After successful deletion, the 7-segment display returns to the normal status display.
8. After download, wait for the system to reboot.
NOTE
With software version VB11A, the system serial number (1xxxx)
must be entered and saved manually under <Main System>-<Configuration>-<Main System> before a restore of the main system is
carried out.
9. Restore the main system parameters (path: <Service>-<Backup & Restore> "Packages" - "Main System").
10. Perform a generator adjustment (path: <Service>-<Main System>-<Adjustment>-<Generator Adjust.>).
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 23 of 98
24Main System - Replacing Components
11. Check the dose rate control and adjust it as necessary (path: <Service>-<Main System>-<Adjustment>-<Dose rate>).
12. Ensure that camera rotation is functioning properly. Adjust, if necessary.
13. Ensure that the collimator is functioning properly. Adjust, if necessary.
14. Check the display functions and the setting of the blades on the monitor. Adjust, if necessary.
15. Ensure that the area dose measurement device (if present) is functioning correctly.
16. Test the FL/PFL/DR and direct exposure functions of the system.
WARNING
When these service tasks have been performed, the accuracy of
an optionally installed navigation system is no longer guaranteed.
See the chapter "Prerequisites," section "Safety information and
protective measures."
¹ If an optionally installed navigation system is present,
the customer must be notified that the accuracy of the
installed navigation system is no longer guaranteed, and
that the navigation system must be checked and certified
before it is used again.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 24 of 98
Medical Solutions
Main System - Replacing Components25
D2 power board2.1
WARNING
1. Disconnect D1 and D2 boards completely.
2. Disconnect the cables between boards D1 and D2.
3. Remove boards D1 and D2.
4. Apply heat conducting paste to the heat sink for the power semiconductor on the new
D2 board.
5. Reinstall boards D2 and D1 and reattach all connections.
6. Ensure that the shielding and ground connections are positioned correctly.
7. Perform the kV offset adjustment and the generator adjustment (filament learning).
8. Test the FL/PFL/DR and direct exposure functions, if there is a cassette holder.
Electrical voltage!
See chapter 1, Safety Information.
¹ Make sure that LED V400 is off prior to performing any
work on boards D1 and D2. This should occur approximately 3 minutes after switching OFF the ARCADIS system.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 25 of 98
26Main System - Replacing Components
D3 interface board2.2
WARNING
1. Replace the D3 interface board.
2. Check the +26.75 V voltage for the I.I. mini-voltage supply. Adjust it, if necessary. Refer
to chapter 3 of these instructions, Checking the Operating Voltages.
3. Check the radiation release/format switchover/vertical column movement functions.
Electrical voltage!
See chapter 1, Safety Information.
¹ Switch system power supply off and disconnect the
power plug.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 26 of 98
Medical Solutions
Main System - Replacing Components27
D40 board for downward movement of the lifting column2.3
WARNING
1. Replace the D40 board.
2. Using the operating instructions for the basic system, check both the movements of the
C-arm and the signal messages.
Electrical voltage!
See chapter 1, Safety Information.
¹ Switch system power supply off and disconnect the
power plug.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 27 of 98
28Main System - Replacing Components
M14 power supply, +5 V/+15 V/-15 V2.4
WARNING
1. Replace the power supply.
2. Check the power supply voltage and adjust it if necessary. Refer to chapter
(Voltages / p. 89).
Electrical voltage!
See chapter 1, Safety Information.
¹ Switch the system power supply off and disconnect the
power plug.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 28 of 98
Medical Solutions
Main System - Replacing Components29
M13 power supply 230V/13V2.5
WARNING
1. Replace the power supply.
2. Check the power supply voltage according to chapter (Voltages / p. 89).
Electrical voltage!
See chapter 1, Safety Information.
¹ Switch the system power supply off and disconnect the
power plug.
Siemens AGSPR2-310.841.01.07.02ARCADIS Varic
Medical Solutions
08.07CS PS SP
Page 29 of 98
30Main System - Replacing Components
I.I. mini-voltage supply2.6
WARNING
WARNING
Electrical voltage!
See chapter 1, Safety Information.
¹ Switch the system power supply off and disconnect the
power plug.
Electrical voltage!
See chapter 1, Safety Information.
¹ Prior to removing the mini-voltage supply, the system
must be switched off for at least 3 minutes, until the high
voltage in the system and in the I.I. mini-voltage supply
dissipates.
Roederstein I.I. mini voltage supply0
Fig. 4:I. I. mini-voltage supply
1. Remove the cylindric cover on the I.I..
2. Replace the I.I. mini-voltage supply (Fig.4/p.30).
3. Refer to the I.I. test certificate 1 for the E1/E2/E3 and A voltages. Check and adjust
them as necessary according to the "Voltages" chapter.
4. Check the functions and adjustments of the collimator. Readjust them, if necessary.
5. Check the display functions and the setting of the blades on the monitor. Adjust them, if
necessary.
ARCADIS VaricSPR2-310.841.01.07.02Siemens AG
08.07CS PS SP
Page 30 of 98
Medical Solutions
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.