The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPR2-310.891.01.01.02 Doc. Gen. Date:04.04
Replaces: n.a.
Page 2
0 - 2Revision
ChapterPageRevision
allall01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
ARCADIS VaricSPR2-310.891.01Page 2 of 4Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Siemens AGSPR2-310.891.01Page 3 of 4ARCADIS Varic
Medical SolutionsRev. 0104.04CS SD 21
Page 4
0 - 4Contents
ARCADIS VaricSPR2-310.891.01Page 4 of 4Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Page 5
General Notes1
1 - 1
General Notes 1
- With distribution of these revision level, all preceding planning guides, Speed Infos (PGs) and drafts
lose their validity.
- All layouts issued by the Planning Departments must bear a note referring to the installation and
delivery conditions of Siemens Medical Engineering. The installation and delivery conditions must
be submitted with the layouts.
- Unless otherwise specified, all dimensions are indicated in "mm".
- The symbol indicates a change (see revision status).
- Orientation points
Points specific to system components to which reference is made when positioning system
components to each other or in the room.
The isocenter of a radiographic system is always illustrated as the orientation point.
-Fixpoints
Clearly marked points on system components, installation ceiling, walls or floor on which cable
outlets are located.
Illustration in the drawings: octagon with letter/number-combination.
The cable lengths specify the maximum fixpoint distances and thus the maximum distances
between the individual system components.
- Room height
The room height is the distance measured from the top surface of the floor to the bottom surface of
the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
- Room lighting
According to DIN 68 68-57 (international standard in preparation), the lighting in rooms in which
image playback devices (monitors) are used for diagnosis, the following requirements must be
met:
adjustable, no anti-glare screen, reproducible adjustment of the lighting (e.g. dimmer with
scale),
no glare or reflection from windows, lights and light boxes in the standard working position of
the monitors.
Hotline+ 49 (9191) 18 - 8080
Siemens AGSPR2-310.891.01Page 1 of 2ARCADIS Varic
Medical SolutionsRev. 0104.04CS SD 21
Page 6
1 - 2General Notes
Safety1
- The provisions of the relevant fire safety regulations must be observed for the premises.
- The system has been developed according to EN 60601 - 1.
- Minimum dimensions (e.g. room heights, safety distances) indicated in the planning guides are
marked "min."
- Basic strength against electromagnetic sources of interference.
Result of lightning discharges.
The protection targets of the different lightning protection areas up to the unit connection are also
specified in the IEC 1024, DIN 48810, VDE 0675 and in the DEMVT recommendations.
ARCADIS VaricSPR2-310.891.01Page 2 of 2Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Page 7
Room Planning2
2 - 1
ARCADIS Varic 2
1 : 20
Horizontal travel:200 mm
Vertical travel:450 mm
Angulation:190° (
Orbital pivot range:+40° / -90
Horizontal pivot range:± 12.5°
Siemens AGSPR2-310.891.01Page 1 of 2ARCADIS Varic
Medical SolutionsRev. 0104.04CS SD 21
± 95°)
°
Page 8
2 - 2Room Planning
Monitor Trolley Dimensions2
*1
*1 Option black/white Display
1 : 20
ARCADIS VaricSPR2-310.891.01Page 2 of 2Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Page 9
Preparation for Installation3
3 - 1
On-site Electrical Installation3
Recommendation for On-site Power Distributor per DIN VDE 0100-710 or National Regulations
*1
CEE socket
Must be provided on-site
*1FI switch IdN = 30 mA leakage current from AC voltage or DC/pulsed DC voltage
On-site Requirements for SRS3
If networking of the ARCADIS Varic is not planned, a remote connection, e.g. in the storage room or
equipment room, should be provided.
Siemens AGSPR2-310.891.01Page 1 of 4ARCADIS Varic
Medical SolutionsRev. 0104.04CS SD 21
Page 10
3 - 2Preparation for Installation
SIEMENS Remote Services
Siemens Remote Server
Public telephone system
ISDN router
Switched or customer’s
own network
TM
(SRS)3
USC
Analog modem. If
an ISDN connection is not available
New modality
Any existing modality
Any existing modality
Remote diagnostics and "Pro-active Event Management" require remote LAN access to our systems.
A router is required for this, which we provide to our customers at no charge during the warranty period
and beyond that period if a service contract is concluded.
The router remains the property of Siemens AG. The required telephone connection (preferably ISDN)
and the power supply must be provided on site. Even if several Siemens systems are connected to the
same data net, only a single central router is required.
Of course, we observe the confidentiality of patient data in the framework of remote diagnostics and
have provided security against access to it. Further information on the topic of data security can be
found in the system-specific publications.
TM
NOTE
For detailed information about SIEMENS Remote Services
(SRS)
see PG TDIT-000.891.01... (http://www-td.med.siemens.de/rootcollection/
Med_TD/Product_Information/PG/general)
NOTE
For the on-site requirements for SRS, see Page 3-1.
ARCADIS VaricSPR2-310.891.01Page 2 of 4Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Page 11
Preparation for Installation3 - 3
Network3
The image system is equipped with a network card for Fast Ethernet (100 Base T).
DICOM3
For detailed information about DICOM, see PG TDIT-000.891.02...
(http://www-td.med.siemens.de/rootcollection/Med_TD/Product_Information/PG/general).
Siemens AGSPR2-310.891.01Page 3 of 4ARCADIS Varic
Medical SolutionsRev. 0104.04CS SD 21
Page 12
3 - 4Preparation for Installation
This page intentionally left blank.
ARCADIS VaricSPR2-310.891.01Page 4 of 4Siemens AG
Rev. 0104.04CS SD 21Medical Solutions
Page 13
Technical Data4
4 - 1
Electrical Data 4
Power supplyInternal Line ResistancePower
100/110/120/127/200/230/
240 V,
50/60 Hz,
± 10%
± 1 Hz
<
0.30 Ohm 100/110/120/127 V
0.80 Ohm 200/230/240 V
<
Long-term: 1.65 kVA
Consumption
Short-term: 2.9 kVA
Fuses,
Internal
15 A slow-blow
from 200 V
20 A slow-blow
up to 127 V
Weight and Heat Dissipation 4
Weight [kg]Heat Dissipation [W]
ARCADIS Varicapprox. 236approx. 500
Monitor trolleyapprox. 180approx. 150
Environmental Conditions 4
OperationTransportStorage
Permissible ambient
temperature
Permissible relative air
humidity
+ 15
° C ... + 37° C- 20° C ... + 37° C- 20° C ... + 37° C