Siemens Arcadis Varic Planning Guide

Page 1
ARCADIS Varic
Planning Guide
System
SP
© Siemens AG 2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English Print No.: SPR2-310.891.01.01.02 Doc. Gen. Date: 04.04
Replaces: n.a.
Page 2
0 - 2 Revision
Chapter Page Revision
all all 01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery.
Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
ARCADIS Varic SPR2-310.891.01 Page 2 of 4 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 3
Contents 0 - 3
Page
1 _______General Notes _________________________________________________ 1 - 1
General Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
2 _______Room Planning ________________________________________________ 2 - 1
ARCADIS Varic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Monitor Trolley Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
3 _______Preparation for Installation_______________________________________ 3 - 1
On-site Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
On-site Requirements for SRS . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
SIEMENS Remote ServicesTM (SRS) . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 3
DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 3
4 _______Technical Data _________________________________________________ 4 - 1
Electrical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Weight and Heat Dissipation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Packing and Transport Routes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Paint Colors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 1
Siemens AG SPR2-310.891.01 Page 3 of 4 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
Page 4
0 - 4 Contents
ARCADIS Varic SPR2-310.891.01 Page 4 of 4 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 5

General Notes 1

1 - 1

General Notes 1

- With distribution of these revision level, all preceding planning guides, Speed Infos (PGs) and drafts lose their validity.
- All layouts issued by the Planning Departments must bear a note referring to the installation and delivery conditions of Siemens Medical Engineering. The installation and delivery conditions must be submitted with the layouts.
- Unless otherwise specified, all dimensions are indicated in "mm".
- The symbol indicates a change (see revision status).
- Orientation points Points specific to system components to which reference is made when positioning system components to each other or in the room. The isocenter of a radiographic system is always illustrated as the orientation point.
-Fixpoints Clearly marked points on system components, installation ceiling, walls or floor on which cable outlets are located. Illustration in the drawings: octagon with letter/number-combination. The cable lengths specify the maximum fixpoint distances and thus the maximum distances between the individual system components.
- Room height The room height is the distance measured from the top surface of the floor to the bottom surface of the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
- Room lighting According to DIN 68 68-57 (international standard in preparation), the lighting in rooms in which image playback devices (monitors) are used for diagnosis, the following requirements must be met:
adjustable, no anti-glare screen, reproducible adjustment of the lighting (e.g. dimmer with scale),
no glare or reflection from windows, lights and light boxes in the standard working position of the monitors.
Hotline + 49 (9191) 18 - 8080
Siemens AG SPR2-310.891.01 Page 1 of 2 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
Page 6
1 - 2 General Notes

Safety 1

- The provisions of the relevant fire safety regulations must be observed for the premises.
- The system has been developed according to EN 60601 - 1.
- Minimum dimensions (e.g. room heights, safety distances) indicated in the planning guides are marked "min."
- Basic strength against electromagnetic sources of interference. Result of lightning discharges. The protection targets of the different lightning protection areas up to the unit connection are also specified in the IEC 1024, DIN 48810, VDE 0675 and in the DEMVT recommendations.
ARCADIS Varic SPR2-310.891.01 Page 2 of 2 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 7

Room Planning 2

2 - 1

ARCADIS Varic 2

1 : 20
Horizontal travel: 200 mm Vertical travel: 450 mm Angulation: 190° ( Orbital pivot range: +40° / -90 Horizontal pivot range: ± 12.5°
Siemens AG SPR2-310.891.01 Page 1 of 2 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
± 95°)
°
Page 8
2 - 2 Room Planning

Monitor Trolley Dimensions 2

*1
*1 Option black/white Display
1 : 20
ARCADIS Varic SPR2-310.891.01 Page 2 of 2 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 9

Preparation for Installation 3

3 - 1

On-site Electrical Installation 3

Recommendation for On-site Power Distributor per DIN VDE 0100-710 or National Regulations
*1
CEE socket
Must be provided on-site
*1 FI switch IdN = 30 mA leakage current from AC voltage or DC/pulsed DC voltage

On-site Requirements for SRS 3

If networking of the ARCADIS Varic is not planned, a remote connection, e.g. in the storage room or equipment room, should be provided.
Siemens AG SPR2-310.891.01 Page 1 of 4 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
Page 10
3 - 2 Preparation for Installation
SIEMENS Remote Services
Siemens Remote Server
Public tele­phone system
ISDN router
Switched or customer’s own network
TM
(SRS) 3
USC
Analog modem. If an ISDN connec­tion is not avail­able
New modality
Any existing modality
Any existing modality
Remote diagnostics and "Pro-active Event Management" require remote LAN access to our systems. A router is required for this, which we provide to our customers at no charge during the warranty period and beyond that period if a service contract is concluded.
The router remains the property of Siemens AG. The required telephone connection (preferably ISDN) and the power supply must be provided on site. Even if several Siemens systems are connected to the same data net, only a single central router is required.
Of course, we observe the confidentiality of patient data in the framework of remote diagnostics and have provided security against access to it. Further information on the topic of data security can be found in the system-specific publications.
TM
NOTE
For detailed information about SIEMENS Remote Services
(SRS) see PG TDIT-000.891.01... (http://www-td.med.siemens.de/rootcollection/ Med_TD/Product_Information/PG/general)
NOTE
For the on-site requirements for SRS, see Page 3-1.
ARCADIS Varic SPR2-310.891.01 Page 2 of 4 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 11
Preparation for Installation 3 - 3

Network 3

The image system is equipped with a network card for Fast Ethernet (100 Base T).

DICOM 3

For detailed information about DICOM, see PG TDIT-000.891.02... (http://www-td.med.siemens.de/rootcollection/Med_TD/Product_Information/PG/general).
Siemens AG SPR2-310.891.01 Page 3 of 4 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
Page 12
3 - 4 Preparation for Installation
This page intentionally left blank.
ARCADIS Varic SPR2-310.891.01 Page 4 of 4 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions
Page 13

Technical Data 4

4 - 1

Electrical Data 4

Power supply Internal Line Resistance Power
100/110/120/127/200/230/
240 V,
50/60 Hz,
± 10%
± 1 Hz
<
0.30 Ohm 100/110/120/127 V
0.80 Ohm 200/230/240 V
<
Long-term: 1.65 kVA
Consumption
Short-term: 2.9 kVA
Fuses,
Internal
15 A slow-blow
from 200 V
20 A slow-blow
up to 127 V

Weight and Heat Dissipation 4

Weight [kg] Heat Dissipation [W]
ARCADIS Varic approx. 236 approx. 500
Monitor trolley approx. 180 approx. 150

Environmental Conditions 4

Operation Transport Storage
Permissible ambient temperature
Permissible relative air humidity
+ 15
° C ... + 37° C- 20° C ... + 37° C- 20° C ... + 37° C
15% ... 75% 15% ... 75% 15% ... 75%
Permissible air pressure 700 hPa ... 1060 hPa 700 hPa ... 1060 hPa 700 hPa ... 1060 hPa

Packing and Transport Routes 4

Largest crate L 2430 x W 960 x H 1550 mm
Heaviest single part approx.
approx.
Min. door opening for transport (without packaging)
min. 940 mm door width
500 kg, overseas packaging 380 kg (with pallet)

Paint Colors 4

Main color Medical White C610
Combination color Medium Basic C701; Medical Blue C750
For data regarding spray cans and paint sticks, see the Project Manager Handbook (PMH), chapter 1
Siemens AG SPR2-310.891.01 Page 1 of 2 ARCADIS Varic Medical Solutions Rev. 01 04.04 CS SD 21
Page 14
4 - 2 Technical Data
This page intentionally left blank.
ARCADIS Varic SPR2-310.891.01 Page 2 of 2 Siemens AG Rev. 01 04.04 CS SD 21 Medical Solutions