Siemens ARCADIS Orbic Start-up Instructions

Würthwein
ARCADIS Orbic
Start-up
System
SP
Start-up Instructions
08080009
Print No.: Replaces: SPR2-320.815.02.01.02
SPR2-320.815.02.02.02
© Siemens AG
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
English Doc. Gen. Date: 02.05
2004

2 Revision / Disclaimer

1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 2 of 46
Medical Solutions
Table of Contents 3
1- 0Table of Contents
1 _______ General Information______________________________________________ 5
Additionally Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Tools, aids (partially for IQ test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Note on the log book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notes on adapting the power plug to local conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Note on the power connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Notes on equivalent leakage current measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Applicability and regulations for regional companies . . . . . . . . . . . . . . . . . . . . . . . . 10
Initial measured value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Repeat measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Switching the system to be free of voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Parts conducting voltage after system OFF and/or with power plug pulled out . . . . . . . 13
Main system control board D1 and power board D2. . . . . . . . . . . . . . . . . . . . . . . . . 13
Monitor cart, monitors and image system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 _______ General Startup ________________________________________________ 14
Power line connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Measuring the Power Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
The line voltage can be adapted at the transformer in the monitor cart. . . . . . . . . . . . . 15
Adapting the line voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the system leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring arrangements for system leakage current with tester VDE 0751 October
2001 (preferred measuring method) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring arrangements for system leakage current with tester VDE 0751 October
1990 (e.g. Bender tester) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Function check of the emergency stop switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Checking the temperature indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Functional checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Checking the ground wires. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3 _______ Specific system settings_________________________________________ 22
System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Service login: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Customer address configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Main system configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Country-specific adaptation of the maximum tube current . . . . . . . . . . . . . . . . . . . . 29
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4 _______ Final work steps________________________________________________ 36
Paper printer (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Sony printer LPR-1000MD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Codonics printer 1660L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Sony printer UP-D72 X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 3 of 46
4 Table of Contents
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
IQ test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Configuring the network (optional) and remote service. . . . . . . . . . . . . . . . . . . . . . . 38
Completing protocols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Completing the "ARCADIS Orbic Installation" protocol . . . . . . . . . . . . . . . . . . . . . . . 38
5 _______ Reports _______________________________________________________ 39
C-arm movement and emergency stop. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Leakage current measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Network checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6 _______ Changes to previous version _____________________________________ 46
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 4 of 46
Medical Solutions

General Information 5

1General Information
2-

Additionally Required Documents 0

ARCADIS Orbic Operating Instructions
ARCADIS Orbic System Wiring Diagram
IQ test (quality assurance)
Image intensifier laser light localizer (optional) assembly and setting instructions
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 5 of 46
6 General Information

Tools, aids (partially for IQ test) 0

Tool kit 97 02 457 Y1971
Dynamic test kit 37 90 156 X1963
Copper filter set 44 06 120 RV090
Densitometer 97 02 416 Y1996
Radiation filter set (incl. 25 mm Al) 97 98 596 G5321
Resolution test 28 71 820 RE999
Safety tester (leakage current and ground wire testing device)
e.g. Unimed 1100
(Leakage current measurement only required in the case of a deviation of the local line voltage from the system delivery sta­tus).
51 38 727 Y0766
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 6 of 46
Medical Solutions
General Information 7

Note on the log book 0

The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 7 of 46
8 General Information

Notes on adapting the power plug to local conditions. 0

The customer can have the power plug on the supplied power cable replaced with an appropriate local plug by an electrician, provided that:
The power plug used can conduct the power required for the operation of the ARCADIS
Orbic (refer to current and voltage values stated on the line voltage label, back of moni­tor cart).
The required line internal resistance is reached (refer to ARCADIS Orbic Planning
Guide).
The selectivity of the fuse protection is assured - according to the relevant national
standards.
It must also be ensured by the customer that connection to an unsuitable power supply is prevented (e.g. by identification or design measures).
As concluding measures, perform and document protective ground wire measurements (in Germany, e.g., in the medical device book).
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 8 of 46
Medical Solutions
General Information 9

Note on the power connection 0

The system delivery state is provided by the power connection values on the the line volt­age label (back of the monitor cart). Necessary adjustments to local conditions are to be performed as necessary according to the adjustment instructions of the ARCADIS Orbic system.
WARNING
Danger high voltage!
Disregarding safety precautions can result in death or serious bodily injury.
¹ Before the line voltage is adapted, the power plug on the
UPS in the monitor cart must be pulled out, since the monitor cart is not free of voltage after the power plug is pulled out.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 9 of 46
10 General Information

Notes on equivalent leakage current measurement 0

Applicability and regulations for regional companies 0

The equivalent leakage current must be measured where applicable under the require­ments of DIN VDE 0751 Part 1. Outside of the scope of application of DIN VDE 0751, the following rules are to be adhered to for the regional companies: (also see ARTD part 2, Security Rules for Assembly and Maintenance).
The national regulations apply primarily for the regional companies. In the event that there are no existing regulations, the following provisions should be adhered to in the interest of the safety of customers, patients, employees and third parties as well as the company.

Initial measured value 0

The equivalent leakage current measurement was performed at the factory and the mea­sured value was entered in test protocol 1b. The measurement was made at the line volt­age and line frequency recorded in test protocol 1b. The test protocol 1b is filed in the log book, register 3. When the line voltage and line frequency match, the value recorded in test protocol 1b is to be transferred as the initial measured value to the equivalent leakage current/protocol.
If the local line voltage or line frequency deviates from the delivery state of the ARCADIS Orbic, the values listed in test protocol 1b are invalid. The values are to be labeled invalid.
(Check the comment "values invalid" and confirm with name, date, and signature). The equivalent leakage current measurement must be repeated. Perform the measurement according to DIN VDE 0751, Part 1 (see ARTD, Part 2), and record the determined value as the initial measured value. This value must not exceed the maximum value for equiva­lent leakage currents for devices according to IEC 601 Part 1/VDE 0750 Part 1 of 2 mA. The initial measured value is to be entered in the equivalent leakage current protocol
(General Startup / p. 14). Comply with the measuring arrangement according to (Fig.2/p.16). The system must be switched on during the measurement. When the
Bender safety tester is used, this must therefore be set to manual measurement. Sepa­rate the page with the protocol from these instructions and file it in the log book, system folder.

Repeat measurement 0

When service or repair work is performed on the primary power supply circuit (e.g. repairs to the power-on circuit or replacement of the line filter), the equivalent leakage current measurement must be repeated. Comply with the measuring arrangement according to
(Fig.2/p.16). The system must be switched on during the measurement. When the
Bender safety tester is used, this must therefore be set to manual measurement. The val­ues measured in the repeat measurement may not exceed the threshold value of 2 mA as specified in VDE 0751, Part 1 (see also ARTD, Part 2). In addition, they may not exceed the initial measured value by more than 50%. The system must be repaired if the limit is exceeded. Document the measured value in the equivalent leakage current protocol.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 10 of 46
Medical Solutions
General Information 11

Safety instructions 0

When performing the work steps and checks, the general safety information for medi-
cal products must be observed.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 11 of 46
12 General Information

Switching the system to be free of voltage 0

To switch the system to be free of voltage in all its parts, bring the system switch
into the OFF position.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 12 of 46
Medical Solutions
General Information 13

Parts conducting voltage after system OFF and/or with power plug pulled out

Main system control board D1 and power board D2 0

WARNING

Monitor cart, monitors and image system 0

WARNING
Danger high voltage!
When the system is switched off, there is still electrical voltage on boards D1 and D2 of the main system.
Disregarding safety precautions can result in death or serious bodily injury.
¹ Therefore before working on one of these boards, make
sure that the LED V400 has gone out. This is the case 3 minutes after the ARCADIS Orbic system is switched off.
Danger high voltage!
When the system is switched off and even when the power plug is pulled out, the line voltage is still applied to the power assembly of the monitor cart, to the monitors and to the image system.
0
Disregarding safety precautions can result in death or serious bodily injury.
¹ Before work is done on the monitor cart, the power plugs
must be pulled out of the UPS.
WARNING
X ¹ These checks and adjustments are explicitly designated
Dangerous radiation during checks and adjustments!
Risk of death or serious bodily injury!
¹ For checks and adjustments that must be performed with
the radiation switched on, the prescribed radiation safety measures must be observed; if necessary, wear radiation protective clothing (see also ARTD-002.731.02.xx.xx and ARTD-002.731.38.xx.xx “General Guidelines for Techni­cal Service”).
on the following pages with the radiation warning sym­bol.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 13 of 46

14 General Startup

2General Startup
3-

Power line connection 0

Measuring the Power Line Voltage 0

Measure the voltage and frequency at the intended outlet for the ARCADIS Orbic sys-
tem.
Compare the measured line voltage with the line voltage label on the back of the moni-
tor cart.
¹ The measured voltage must agree with the voltage read off from the label.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 14 of 46
Medical Solutions
General Startup 15

The line voltage can be adapted at the transformer in the monitor cart.

WARNING

Adapting the line voltage 0

Remove the back, bottom covers (Fig.1/p.15) from the monitor cart.
Pull the power plug out of the UPS in the monitor cart.
Adapt the voltage by changing the connections to transformers T1 and T2. Refer to the
ARCADIS Orbic wiring diagram.
Danger high voltage!
Disregarding safety precautions can result in death or serious bodily injury.
¹ Before the line voltage is adapted, the power plug on the
UPS in the monitor cart must be pulled out, since the monitor cart is not free of voltage after the power plug is pulled out.
0
Fig. 1: Monitor cart line voltage
Pos. 1 Power plug
Pos. 2 T1 and T2
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 15 of 46
16 General Startup

Measuring the system leakage current 0

NOTE
Perform the system leakage current measurement only if the line voltage on the ARCADIS Orbic system must be adapted to the lo­cal line voltage.
See also (Notes on equivalent leakage current
measurement / p. 10)
Measuring arrangements for system leakage current with tester VDE 0751 Octo­ber 2001 (preferred measuring method)
0
Fig. 2: Measuring circuit for measuring the system leakage current according to the differential
current method
Perform measurement configuration according to (Fig.2/p.16). Application part
(1/Fig.2/p.16) is not available in ARCADIS Orbic.
Perform measurement and repeat measurements.
Document the measured values in the "Leakage current test protocol", (Leakage cur-
rent measurement / p. 40).
Measuring arrangements for system leakage current with tester VDE 0751 Octo­ber 1990 (e.g. Bender tester)
0
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 16 of 46
Medical Solutions
General Startup 17
Fig. 3: Measuring circuit for the direct measurement of the system leakage current
Measurement configuration according to (Fig.2/p.16) application part
(1/Fig.3/p.17) is not available in ARCADIS Orbic.
For safety reasons, place the system on an insulated base.
Perform measurement and repeat measurements.
Document the measured values in the "Leakage current test protocol" (Leakage cur-
rent measurement / p. 40).
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 17 of 46
18 General Startup

Function check of the emergency stop switches 0

Connect the system to the voltage and boot it.
Click "Emergency" for the patient registration.
Activate the EMERGENCY STOP switch on the C-arm.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
Disengage the EMERGENCY STOP switch.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop (C-arm movement and emergency stop / p. 39).
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 18 of 46
Medical Solutions
General Startup 19

Checking the temperature indicator 0

Check the temperature indicator on the outside of the image intensifier housing.
¹ If the inner, square field of the indicator is white, the temperature has not been
exceeded. Remove the temperature indicator.
¹ If the indicator is discolored (inner field black), proceed according to IQ docu-
ment RXD0-000-038.01.xx.xx.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 19 of 46
20 General Startup

Functional checks 0

Operation of the ARCADIS Orbic, see operating instructions
Stand and C-arm system movements
X X
Semi-transparent slot and iris diaphragm in fluoroscopy procedure.
Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce-
dure.
Mechanical functions of the monitor cart
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 20 of 46
Medical Solutions
General Startup 21

Checking the ground wires 0

The system must be free of voltage.
Measuring instru-
ment:
Measuring instru-
ment:
Ground wire testing device; e.g. safety tester Unimed 1100
Max. 0.2 Ohm (observe country-specific regulations!) (current: 10 A)
Test procedure: Measure between all touchable, conductive parts of the system
and the ground wire at the ARCADIS Orbic power plug.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 21 of 46

22 Specific system settings

3Specific system settings
4-

System configuration 0

Please clarify in advance with the customer whether the following system configura-
tions are required.

Service login: 0

Fig. 4: Service login
Select <Options> <Customer service> <Local service>.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 22 of 46
Medical Solutions
Specific system settings 23
Fig. 5: Authentication_
Enter the 6-character password and click OK (see system folder for password).
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 23 of 46
24 Specific system settings

Customer address configuration 0

Fig. 6: Configuration_
Select <Configuration>.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 24 of 46
Medical Solutions
Specific system settings 25
Fig. 7: Configuration
Click <Next>.
Fig. 8: Configuration_Local Host
Click <Site Info>.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 25 of 46
26 Specific system settings
Fig. 9: Configuration_Local Host_Site Info
The marked customer-specific data may be adapted. Under no circumstances can you
change the "serial no." or "station name".
Click <Save>.
Click <Home>.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 26 of 46
Medical Solutions
Specific system settings 27

Main system configuration 0

Select “Main system” in the service software (see Fig. ...).
Fig. 10: Main_system
Click <Main system>.
Fig. 11: Main system_
Select ARCADIS Orbic and click <Next>.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 27 of 46
28 Specific system settings
Fig. 12: Main System_Configuration_Buzzer
Under Configuration click <Buzzer>.
Make changes according to the following description:
Block time settings
- Warning time limit: The buzzer sound starts when the entered radiation time has elapsed.
- Block time limit: After the entered radiation time has elapsed - without reset of the “warning time limit" - the radiation is blocked the next time it is selected. Every reset of the “warning time limit” also resets the “block time limit” to “0”.
Buzzer mode (direct, cassette)
- After exposure:Buzzer sound after the exposure.
- During exposure:Buzzer sound during the exposure.
Buzzer mode (indirect, I.I.)
- Buzzer off:Buzzer sound off. Exceptions: Push mode and “warning time limit” has ended.
- Buzzer on: Buzzer sound always during radiation.
The country-specific regulations must be observed for the max. fluoroscopy time and radi­ation blockage.
Factory setting:
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 28 of 46
Medical Solutions
Specific system settings 29
Fluoroscopy alarm Fluoroscopy blocking
USA
Service mask input in 1/10 min.
Europe
Service mask input in 1/10 min.
4.5 min.
45
5 min.
50
5 min.
5
10 min.
50
Make country-specific changes and click <Save>.
Click <Home>.
If a change is made to the fluoro time, check this function.

Country-specific adaptation of the maximum tube current 0

NOTE
Start service program and log in as described under "Service Login".
A maximum tube current limit is required - as far as known - only in Denmark.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 29 of 46
30 Specific system settings
Fig. 13: Main_system
Click <Main system>.
Fig. 14: Main system_
Select ARCADIS Orbic and click <Next>.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 30 of 46
Medical Solutions
Specific system settings 31
Fig. 15: Main System_Configuration_Current Limitation
Click <Current limitation>.
Make country-specific changes (only Denmark) and click <Save>.
Click <Home>.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 31 of 46
32 Specific system settings

Backup 0

NOTE
A backup must be created following software installation, every software update, and every system adjustment.
Select "Backup & Restore" in the “Service Software Home Menu”
Place the backup CD (located in the monitor cart service compartment) in the CD ROM
drive.
Fig. 16: Backup & Restore_Command
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 32 of 46
Medical Solutions
Specific system settings 33
Select the <Backup> command.
Fig. 17: Backup & Restore_Command_Backup
Select the respective drive under “Drives”.
- Save to CD ROM (standard save).
Fig. 18: Backup & Restore_Command_Backup_Packages
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 33 of 46
34 Specific system settings
SW - Settings02
Select “SW-Settings02” under Packages.
Click <Go> and wait until the message “Ready” appears in the footer.
Fig. 19: Backup & Restore_Command_Backup_SW-Settings02
ASPIA Settings
Select “ASPIA settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
ExamSet
Select “ExamSet” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Security-Settings
For systems with HIPAA configuration
Select “Security-Settings” under Packages.
Click <Go>
- Wait until the message “Ready” is displayed in the footer
MainSystem
Select “MainSystem” under Packages.
3D-Settings
Under packages, select “3D settings”
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 34 of 46
Medical Solutions
Specific system settings 35
Click <Go>
- Wait until the message “Ready” is displayed in the footer
Exiting Backup
Close the window via <Home>.
Remove the backup CD ROM from the CD ROM drive.
Store the CD ROM with the other system documentation.
NOTE
It is not possible to perform a backup of the database (patient im­ages)!
Store the patient images in an archive.
Saving to CD-ROM as long-term archive is prohibited.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 35 of 46

36 Final work steps

4Final work steps
5-

Paper printer (optional) 0

NOTE
The software (driver) for the supplied printer was preinstalled at the factory.

Sony printer LPR-1000MD 0

Perform the following steps according to the "Startup Manual" of the printer:
Unpacking (see the chapter "Preparation").
Installing the head cassette (see the chapter "Preparation").
Please note: Remove retaining mechanisms and safety mechanism from the head cassette.
Installing the ink cartridges (see the chapter "Preparation").
Switching on the printer (see the chapter "Preparation").
Please note:Place the printer in advance on the monitor cart, connect the power and USB cables (ready for use in the printer slot of the cart) to the printer, and set the power switch to "1".
Installing the head cleaning unit (see the chapter "Preparation").
Installing the automatic paper feed tray (see the chapter "Preparation").
NOTE
Insert printer paper with the white side facing “down”.
Remove the protective film from the ink cartridge.

Codonics printer 1660L 0

Perform the following steps according to the "User Manual" of the printer:
Transport instructions (see preface).
Please note:Unpack the printer and remove the two transport pins. Place the printer in the cart, connect the power cable and the 25-pin plug to the printer (cables are ready for use in the printer slot of the cart), and set the power switch to "1".
Install the print media>(General Startup / p. 14)
NOTE
Keep the transport safety materials. They are required when re­placing or shipping the printer.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 36 of 46
Medical Solutions
Final work steps 37

Sony printer UP-D72 X 0

Perform the following steps according to the user manual for the printer (CD-ROM):
Unpack the printer.
Fill the paper tray with paper (follow instructions on paper package).
Plug in the USB cable.
Print a test page after the automatic test.
Follow the instructions on the CD-ROM from the manufacturer in the event of problems.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 37 of 46

38 Final work steps

Final work steps 0

IQ test 0

Perform IQ test according to instructions SPR2-310.820.02.xx.xx.
¹ Remove the supplied IQ test protocols from the monitor cart service compart-
ment.

Configuring the network (optional) and remote service 0

Configure the network according to the Configuration Guide, SPR2-310.843.02.01.xx.xx and the remote connection according to Installation, Siemens Remote Service, SP00-000.816.02.01.xx.xx.

Completing protocols 0

After completing all adjustment work steps and check measurements, the responsible technician must sign and date the test protocols and country-specific protocols to confirm that all values have been correctly determined and recorded.

Completing the "ARCADIS Orbic Installation" protocol 0

The manufacturer of this product requires information and the legislator demands proofs that a product delivered free of defects from the factory continues to possess the required and certified quality properties on installation and start-up.
It is therefore absolutely necessary that the installation report with the installation and start-up data is sent without delay after completion of the work to the address stated on the installation report.
You will find the report in the system folder.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 38 of 46
Medical Solutions

Reports 39

5Reports
6-

C-arm movement and emergency stop 0

System: ARCADIS Orbic
Material no.: .........................................................................
Serial number: .........................................................................
Tab. 1 C-arm movement and emergency stop
Movement OK
Yes/No
Releasing/locking the brakes
Downward movement, position 1
Downward movement, position 2
The warning signals sound as described.
Function, EMERGENCY_STOP_button
Name Date: Initials
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 39 of 46
40 Reports

Leakage current measurement 0

System: ARCADIS Orbic
Material no.: .........................................................................
Serial number: .........................................................................
Measuring arrangement: see reverse:
Tab. 2 Leakage current measurement
Initial measured value
Repeat measurement 1
Repeat measurement 2
Repeat measurement 3
Repeat measurement 4
Repeat measurement 5
Repeat measurement 6
Repeat measurement 7
Equivalent
Unit leakage current
Measuring instrument used, serial no.
Name Date Initials
Repeat measurement 8
Repeat measurement 9
Repeat measurement 10
Repeat measurement 11
Repeat measurement 12
Repeat measurement 13
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 40 of 46
Medical Solutions
Reports 41
Repeat measurement 14
Repeat measurement 15
Repeat measurement 16
Repeat measurement 17
Repeat measurement 18
Repeat measurement 19
Repeat measurement 20
Equivalent
Unit leakage current
Measuring instrument used, serial no.
Name Date Initials
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 41 of 46
42 Reports

Network checklist 0

Network checklist (1+) For SAP order no.:_______________________
Will the system have a network connection (HIS/RIS, Camera, PACS)?
Mandatory for image system with network connection!
To ensure that the system is preconfigured at the factory and thus optimize installation at the customer site, we
request that you complete and return this list if a network connection is available (and used).
Person responsible for network configuration at customer site
Name:
Phone/Fax:
E-mail:
System network data
Tab. 3 System network data
Not networked
Identification:
Computer name *
Yes No
Configuration/ Local host/ TCP IP address
IP address:
TCP/ IP address
Subnet mask * Configuration/ Local host/ TCP IP address
Gateways * Configuration/ Local host/ TCP IP address
General: Local AE title for:
Study transfer * Configuration/ DICOM/ General
Remarks
Network data for HIS/RIS network partner (only with HIS/RIS option)
* Configuration/ Local host/ TCP IP address
HIS/RIS *
Configuration/ DICOM/ General
Print * Configuration/ DICOM/ General
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 42 of 46
Medical Solutions
Reports 43
Tab. 4 Network data for HIS/RIS
not available
Product: Version: Manufacturer:
Host properties:
Host name
TCP/ IP address * Configuration/ DICOM/ HIS RIS nodes
General Node Properties
Logical name
Application entity
AE title
Port number * Configuration/ DICOM/ HIS RIS nodes
Remark:
Network checklist (2+)
Network data for reporting console / archive system (only with DICOM Basic option) (1)
Tab. 5 Network data for reporting console/archiving system
Information on each: Reporting console/
archive system
* Configuration/ DICOM/ HIS RIS nodes
* Configuration/ DICOM/ HIS RIS nodes
* Configuration/ DICOM/ HIS RIS nodes
not available
Product: Version: Manufacturer:
Host properties:
Host name
* Configuration/ DICOM/ Network nodes
TCP/ IP address * Configuration/ DICOM/ Network nodes
General node properties:
Logical name
* Configuration/ DICOM/ Network nodes
Application entity:
AE title
* Configuration/ DICOM/ Network nodes
Port number * Configuration/ DICOM/ Network nodes
Supported DICOM services:
Storage yes no
* Configuration/ DICOM/ Network nodes
Storage Commitment yes no * Configuration/ DICOM/ Network nodes
Node name: * Configuration/ DICOM/ Network nodes
Query yes no * Configuration/ DICOM/ Network nodes
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 43 of 46
44 Reports
Retrieve yes no * Configuration/ DICOM/ Network nodes
Remarks:
Network data for reporting console/archive system (only with DICOM Basic option) (2)
Tab. 6 Network data for reporting console/archiving system
Information on each: Reporting console/
not available
archive system
Product: Version: Manufacturer:
Host properties:
Host name
* Configuration/ DICOM/ Network nodes
TCP/ IP address * Configuration/ DICOM/ Network nodes
General node properties:
Logical name
* Configuration/ DICOM/ Network nodes
Application entity:
AE title
* Configuration/ DICOM/ Network nodes
Port number * Configuration/ DICOM/ Network nodes
Supported DICOM services:
Storage yes no
* Configuration/ DICOM/ Network nodes
Storage Commitment yes no * Configuration/ DICOM/ Network nodes
Node name: * Configuration/ DICOM/ Network nodes
Query yes no * Configuration/ DICOM/ Network nodes
Retrieve yes no * Configuration/ DICOM/ Network nodes
Remarks:
Network checklist (3-)
Network data for a DICOM camera in the network (only with the DICOM Print option) (1)
Tab. 7 Network data for DICOM camera in the network
Information for Every camera No camera available
Product: Version: Manufacturer:
Default camera yes no
Host properties:
Host name
* Configuration/ DICOM/ Print devices
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 44 of 46
Medical Solutions
Reports 45
TCP/ IP address * Configuration/ DICOM/ Print devices
General node properties:
Logical name
* Configuration/ DICOM/ Print devices
Application entity:
AE title
* Configuration/ DICOM/ Print devices
Port number * Configuration/ DICOM/ Print devices
Remarks:
Network data for a DICOM camera in the network (only with the DICOM Print option) (2)
Tab. 8 Network data for DICOM camera in the network
Information for Every camera No camera available
Product: Version: Manufacturer:
Default camera yes no
Host properties:
Host name
* Configuration/ DICOM/ Print devices
TCP/ IP address * Configuration/ DICOM/ Print devices
General node properties:
Logical name
Application entity:
Remarks:
* Configuration/ DICOM/ Print devices
AE title
* Configuration/ DICOM/ Print devices
Port number * Configuration/ DICOM/ Print devices
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 45 of 46

46 Changes to previous version

6Changes to previous version
7-
The revised Completing the "ARCADIS Orbic Installation" protocol section was added to the "Final Work Steps" chapter.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 46 of 46
Medical Solutions