Siemens ARCADIS Orbic Start-up Instructions

Würthwein
ARCADIS Orbic
Start-up
System
SP
Start-up Instructions
08080009
Print No.: Replaces: SPR2-320.815.02.01.02
SPR2-320.815.02.02.02
© Siemens AG
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
English Doc. Gen. Date: 02.05
2004

2 Revision / Disclaimer

1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 2 of 46
Medical Solutions
Table of Contents 3
1- 0Table of Contents
1 _______ General Information______________________________________________ 5
Additionally Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Tools, aids (partially for IQ test) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Note on the log book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notes on adapting the power plug to local conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Note on the power connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Notes on equivalent leakage current measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Applicability and regulations for regional companies . . . . . . . . . . . . . . . . . . . . . . . . 10
Initial measured value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Repeat measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Switching the system to be free of voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Parts conducting voltage after system OFF and/or with power plug pulled out . . . . . . . 13
Main system control board D1 and power board D2. . . . . . . . . . . . . . . . . . . . . . . . . 13
Monitor cart, monitors and image system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 _______ General Startup ________________________________________________ 14
Power line connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Measuring the Power Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
The line voltage can be adapted at the transformer in the monitor cart. . . . . . . . . . . . . 15
Adapting the line voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Measuring the system leakage current. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring arrangements for system leakage current with tester VDE 0751 October
2001 (preferred measuring method) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring arrangements for system leakage current with tester VDE 0751 October
1990 (e.g. Bender tester) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Function check of the emergency stop switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Checking the temperature indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Functional checks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Checking the ground wires. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3 _______ Specific system settings_________________________________________ 22
System configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Service login: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Customer address configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Main system configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Country-specific adaptation of the maximum tube current . . . . . . . . . . . . . . . . . . . . 29
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4 _______ Final work steps________________________________________________ 36
Paper printer (optional). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Sony printer LPR-1000MD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Codonics printer 1660L. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Sony printer UP-D72 X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 3 of 46
4 Table of Contents
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
IQ test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Configuring the network (optional) and remote service. . . . . . . . . . . . . . . . . . . . . . . 38
Completing protocols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Completing the "ARCADIS Orbic Installation" protocol . . . . . . . . . . . . . . . . . . . . . . . 38
5 _______ Reports _______________________________________________________ 39
C-arm movement and emergency stop. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Leakage current measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Network checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6 _______ Changes to previous version _____________________________________ 46
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 4 of 46
Medical Solutions

General Information 5

1General Information
2-

Additionally Required Documents 0

ARCADIS Orbic Operating Instructions
ARCADIS Orbic System Wiring Diagram
IQ test (quality assurance)
Image intensifier laser light localizer (optional) assembly and setting instructions
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 5 of 46
6 General Information

Tools, aids (partially for IQ test) 0

Tool kit 97 02 457 Y1971
Dynamic test kit 37 90 156 X1963
Copper filter set 44 06 120 RV090
Densitometer 97 02 416 Y1996
Radiation filter set (incl. 25 mm Al) 97 98 596 G5321
Resolution test 28 71 820 RE999
Safety tester (leakage current and ground wire testing device)
e.g. Unimed 1100
(Leakage current measurement only required in the case of a deviation of the local line voltage from the system delivery sta­tus).
51 38 727 Y0766
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 6 of 46
Medical Solutions
General Information 7

Note on the log book 0

The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 7 of 46
8 General Information

Notes on adapting the power plug to local conditions. 0

The customer can have the power plug on the supplied power cable replaced with an appropriate local plug by an electrician, provided that:
The power plug used can conduct the power required for the operation of the ARCADIS
Orbic (refer to current and voltage values stated on the line voltage label, back of moni­tor cart).
The required line internal resistance is reached (refer to ARCADIS Orbic Planning
Guide).
The selectivity of the fuse protection is assured - according to the relevant national
standards.
It must also be ensured by the customer that connection to an unsuitable power supply is prevented (e.g. by identification or design measures).
As concluding measures, perform and document protective ground wire measurements (in Germany, e.g., in the medical device book).
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 8 of 46
Medical Solutions
General Information 9

Note on the power connection 0

The system delivery state is provided by the power connection values on the the line volt­age label (back of the monitor cart). Necessary adjustments to local conditions are to be performed as necessary according to the adjustment instructions of the ARCADIS Orbic system.
WARNING
Danger high voltage!
Disregarding safety precautions can result in death or serious bodily injury.
¹ Before the line voltage is adapted, the power plug on the
UPS in the monitor cart must be pulled out, since the monitor cart is not free of voltage after the power plug is pulled out.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 9 of 46
10 General Information

Notes on equivalent leakage current measurement 0

Applicability and regulations for regional companies 0

The equivalent leakage current must be measured where applicable under the require­ments of DIN VDE 0751 Part 1. Outside of the scope of application of DIN VDE 0751, the following rules are to be adhered to for the regional companies: (also see ARTD part 2, Security Rules for Assembly and Maintenance).
The national regulations apply primarily for the regional companies. In the event that there are no existing regulations, the following provisions should be adhered to in the interest of the safety of customers, patients, employees and third parties as well as the company.

Initial measured value 0

The equivalent leakage current measurement was performed at the factory and the mea­sured value was entered in test protocol 1b. The measurement was made at the line volt­age and line frequency recorded in test protocol 1b. The test protocol 1b is filed in the log book, register 3. When the line voltage and line frequency match, the value recorded in test protocol 1b is to be transferred as the initial measured value to the equivalent leakage current/protocol.
If the local line voltage or line frequency deviates from the delivery state of the ARCADIS Orbic, the values listed in test protocol 1b are invalid. The values are to be labeled invalid.
(Check the comment "values invalid" and confirm with name, date, and signature). The equivalent leakage current measurement must be repeated. Perform the measurement according to DIN VDE 0751, Part 1 (see ARTD, Part 2), and record the determined value as the initial measured value. This value must not exceed the maximum value for equiva­lent leakage currents for devices according to IEC 601 Part 1/VDE 0750 Part 1 of 2 mA. The initial measured value is to be entered in the equivalent leakage current protocol
(General Startup / p. 14). Comply with the measuring arrangement according to (Fig.2/p.16). The system must be switched on during the measurement. When the
Bender safety tester is used, this must therefore be set to manual measurement. Sepa­rate the page with the protocol from these instructions and file it in the log book, system folder.

Repeat measurement 0

When service or repair work is performed on the primary power supply circuit (e.g. repairs to the power-on circuit or replacement of the line filter), the equivalent leakage current measurement must be repeated. Comply with the measuring arrangement according to
(Fig.2/p.16). The system must be switched on during the measurement. When the
Bender safety tester is used, this must therefore be set to manual measurement. The val­ues measured in the repeat measurement may not exceed the threshold value of 2 mA as specified in VDE 0751, Part 1 (see also ARTD, Part 2). In addition, they may not exceed the initial measured value by more than 50%. The system must be repaired if the limit is exceeded. Document the measured value in the equivalent leakage current protocol.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 10 of 46
Medical Solutions
General Information 11

Safety instructions 0

When performing the work steps and checks, the general safety information for medi-
cal products must be observed.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 11 of 46
12 General Information

Switching the system to be free of voltage 0

To switch the system to be free of voltage in all its parts, bring the system switch
into the OFF position.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 12 of 46
Medical Solutions
General Information 13

Parts conducting voltage after system OFF and/or with power plug pulled out

Main system control board D1 and power board D2 0

WARNING

Monitor cart, monitors and image system 0

WARNING
Danger high voltage!
When the system is switched off, there is still electrical voltage on boards D1 and D2 of the main system.
Disregarding safety precautions can result in death or serious bodily injury.
¹ Therefore before working on one of these boards, make
sure that the LED V400 has gone out. This is the case 3 minutes after the ARCADIS Orbic system is switched off.
Danger high voltage!
When the system is switched off and even when the power plug is pulled out, the line voltage is still applied to the power assembly of the monitor cart, to the monitors and to the image system.
0
Disregarding safety precautions can result in death or serious bodily injury.
¹ Before work is done on the monitor cart, the power plugs
must be pulled out of the UPS.
WARNING
X ¹ These checks and adjustments are explicitly designated
Dangerous radiation during checks and adjustments!
Risk of death or serious bodily injury!
¹ For checks and adjustments that must be performed with
the radiation switched on, the prescribed radiation safety measures must be observed; if necessary, wear radiation protective clothing (see also ARTD-002.731.02.xx.xx and ARTD-002.731.38.xx.xx “General Guidelines for Techni­cal Service”).
on the following pages with the radiation warning sym­bol.
Siemens AG SPR2-320.815.02.02.02 ARCADIS Orbic Medical Solutions
02.05 CS SD 24
Page 13 of 46

14 General Startup

2General Startup
3-

Power line connection 0

Measuring the Power Line Voltage 0

Measure the voltage and frequency at the intended outlet for the ARCADIS Orbic sys-
tem.
Compare the measured line voltage with the line voltage label on the back of the moni-
tor cart.
¹ The measured voltage must agree with the voltage read off from the label.
ARCADIS Orbic SPR2-320.815.02.02.02 Siemens AG
02.05 CS SD 24
Page 14 of 46
Medical Solutions