Siemens ARCADIS Orbic Maintenance Instructions

Würthwein
ARCADIS Orbic
Maintenance Instructions
System
SP
Maintenance Instructions
The protocol SPR2-320.832.02.01.02 is required for these instructions
Print No.: Replaces: n.a.
SPR2-320.831.02.01.02
08080009
© Siemens AG
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.
English Doc. Gen. Date: 08.04
2004

2 Revision / Disclaimer

1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 2 of 30
Medical Solutions
Table of Contents 3
0 Table of Contents
1 _______ General information______________________________________________ 4
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Required documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Systems equipped with a laser light localizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Systems equipped with an I.I. laser light localizer . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required tools, measurement and auxiliary devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Spare parts which may be needed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Emphasized texts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety information and protective measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Explanation of abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
System maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance interval for 3D reconstruction in combination with a navigation system .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Technical Safety Checks (TSC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 _______ Inspection of exterior and surroundings____________________________ 18
Inspection of exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Inspection of surroundings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Power outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3 _______ Safety inspection _______________________________________________ 20
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4 _______ Maintenance, operating value/functional inspection __________________ 25
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Operating value inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5 _______ Final result/quality inspection and general maintenance ______________ 29
Final work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 3 of 30
4 General information
1-

1 General information

1.1 Requirements

NOTE
If the system is equipped with 3D reconstruction, in combination with a navigation system, the required maintenance has to be per­formed together with a navigation system engineer.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 4 of 30
Medical Solutions
General information 5

1.2 Required documents

Safety information according to ARTD, part 2
Maintenance protocol SPR2-320.832.02..
Main system adjustment instructions SPR2-310.842.02..
Spare parts list
Replacement of Parts: SPR2-320.841.02.01.01
Technical safety checks - protocol* SP00-000.834.01...

1.2.1 Systems equipped with a laser light localizer

Setting instructions SPR2-230.812.03..

1.2.2 Systems equipped with an I.I. laser light localizer

Installation and setting instructions RXR2-120.814.01..
Replacement instructions of the available
TFT display
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 5 of 30
6 General information

1.3 Required tools, measurement and auxiliary devices

NOTE
All indicated tools, measurement and auxiliary devices, with the exception of those marked with a " * ", are described in the ARTD (Part 3).
Standard tool kit*
Set of Allen keys*
Spring balance rated for up to 200 N e.g. 44 15 113 RH090
Equivalent leakage current tester e.g. Safety Tester
Unimet 1100
Ground wire tester e.g. Safety Tester
Unimet 1100
51 38 727Y0766
51 38 727Y0766
Internal line resistance tester 84 28 104 Y4337
DVM e. g. Fluke 8060A 97 02 101 Y4290
Luminous density measuring device e.g. SPOTMETER for
SMfit ACT 77 52 848
Densitometer (for previous MULTI-
SPOT
e.g. 97 02 416Y1996
2000 option)
Dosimeter, e.g. PTW Diados 97 17 612 Y0388
Dynamic test kit 37 90 156 X1963
Precision X-ray filter 99 00 598 XE999
Set of resolution tests 28 71 820 RE999
Set of radiation filters (10 x 0.3 mm) e.g. 44 06 120 RV090
Set of radiation filters e.g. 97 98 596 G5321
25 mm AL measuring stand, type
26765 acc. to DIN 6868 Part 50*
or
e.g.
1.2 mm Cu (e.g. of a set of radiation fil-
ters 44 06 120) plus
17 micrometer Cu strips 11 67 662
Centering cross (only with Diamentor) e.g. 96 60 051 RE999
Calibration phantom
77 51 620 G5486
(for previous 3D reconstruction option)
Cable for connecting service PC to host 99 00 440 RE999
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 6 of 30
Medical Solutions
General information 7
Service software* (see logbook)
Sealing compound 34 43 009
Special oil (Optimol Optipit) 55 07 525
Special oil (Slic Pac PTFE) 55 07 517
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 7 of 30
8 General information

1.4 Spare parts which may be needed

Cable deflector steering castor large/ARCADIS Orbic stand
Cable deflector wheels small/ARCADIS Orbic stand
Touch-up paint
Spray paint
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 8 of 30
Medical Solutions
General information 9

1.5 Emphasized texts

!
DANGER
!
WARNING
!
CAUTION
NOTICE
Fig. 1: Safety Notes
DANGER indicates when there is an immediate danger that l e a d s to death or serious physical injury.
WARNING indicates a risk of danger that m a y l e a d to death or serious physical injury.
CAUTION used with the safety alert symbol indicates a risk of danger that leads to slight or moderate physical injury and/ or damage to property.
NOTICE used without the safety alert symbol indicates a risk of danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state other than death, physical injury or property damage.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 9 of 30
10 General information

1.6 Safety information and protective measures

NOTICE
The following instructions must be observed!
¹ When performing the work steps and tests, comply with
the following:
¹ - The product-specific safety information in the docu-
mentation, safety information TD00-000.860.01.03.01,
¹ - As well as the general safety information contained in
ARTD, part 2.
¹ Pull the power plug before starting to work on the ARCA-
DIS Orbic.
¹ Ensure compliance with the general safety requirements
when working with the system connected to the mains.
¹ Observe ESD guidelines!
¹ Before beginning maintenance, perform the ground wire
test according TD00-000.860.01.... The ground wire resis-
tance must not exceed 0.2 Ohms.
¹ Switch off the ARCADIS Orbic before replacing modules
or boards.
¹ Perform the ground wire test according to
TD00-000.860.01... upon completion of the work and after all covers have been reattached. Take down and evaluate the determined values. The ground wire resistance must not exceed 0.2 Ohms.
¹ After all work has been completed and all cover panels
have been installed, perform the ground wire measure­ment according to ARTD-002.731.17.
¹ The ground wire resistance must not exceed 0.2 Ohms.
¹ When performing service work on the power-on module,
(replacing the power-on module or replacing the power cable), the equivalent leakage current must be measured and recorded.
¹ Checks and settings that need to be performed under
X-ray radiation are identified by the radiation warning
symbol x . While performing adjustments labeled in
this manner, radiation protective clothing must be worn.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 10 of 30
Medical Solutions
General information 11

1.7 Explanation of abbreviations

Abbrev. Explanation
SI Safety inspection
SIE Electrical safety
SIM Mechanical safety
PM Preventive maintenance
PMP Periodic preventive maintenance
PMA Preventive maintenance adjustments
PMF Preventive check of operating values/functions
Q System quality, image quality
QIQ Image quality
QSQ System quality
SW Software maintenance
The steps identified by these abbreviations are listed as checkpoints in the Maintenance Protocol and must be checked off there accordingly.
NOTE
The sequence for the complete inspection and maintenance is lo­cated in the following pages.
Each step must be performed annually, unless otherwise speci­fied.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 11 of 30
12 General information

1.8 Maintenance interval

1.8.1 System maintenance interval

NOTE
The maintenance interval applies to all maintenance work, except for maintenance of the 3D reconstruction in combination with a navigation system.
12 months
1.8.2 Maintenance interval for 3D reconstruction in combination with a navi­gation system
NOTE
NOTE
If the system is equipped with the 3D reconstruction option in combination with a navigation system, perform the maintenance work listed below in the specified maintenance interval.
Take into account that the maintenance work listed below has to be performed together with an engineer of the navigation system manufacturer.
- SIM navigation system (if present)
I.I. ring damage
Correction of the seating of the I.I. cap
- SIE 3D reconstruction (if present)
Calibration of 3D reconstruction
Check of the limiting resolution
- SIE 3D reconstruction with navigation system (if present)
Check the overall navigation accuracy
6 months or after 150 patients at the latest (approx. 300x 3D scans)
UPS (Uninterrupted Power Supply)
The lead-acid batteries of the UPS in the monitor cart are to be replaced every 36
months.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 12 of 30
Medical Solutions
General information 13

1.9 Technical Safety Checks (TSC)

Abbreviation: TSC = Technical Safety Checks
NOTE
Technical safety checks,
Test protocol check
Within the purview of DIN VDE 0751-1, the operator of medical en­gineering products has to perform technical safety checks at reg­ular intervals.
The checks listed in these maintenance instructions contain all technical safety checks according to DIN VDE 0751-1 except for the check of the completeness, presence and legibility of the nec­essary operator documents.
A separate protocol with the print number SP00.000.834.01... is available for each technical safety check.
Within the purview of DIN VDE 0751-1, we recommend document­ing the maintenance results both in the maintenance protocol and in the TSC protocol. The protocols should be filled out completely and handed over to the client after maintenance is completed.
In the table below each technical safety check listed in the docu­ment SP00.000.834.01... is assigned to the corresponding mainte­nance work.
Maintenance instructions check/
comments
1 Identification data Please fill in the fields.
2 General checks n.a.
2.1 Visual inspection: Is there any damage affecting safety?
2.2 Are all cables and cable guides secure and without any visible damage?
3.1 Mechanical safety,
Checks:
SIM cover panels
SIM cable deflectors
SIM foot brake
SIM brakes
SIM emergency stop switch
SIM monitors/displays
3.2 Electrical safety, check: SIE cable and plug
SIE voltage discharge rubber
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 13 of 30
14 General information
Technical safety checks,
Test protocol check
2.3 Are undamaged accessories being used?
2.4 Are all unit radiation protection devices present and undamaged?
(No on-site radiation shield)
Maintenance instructions check/
comments
3.1 Mechanical safety,
checks:
SIM I.I. laser light localizer mechanics
SIM I.I. laser light localizer function
SIM laser light localizer mechanics
SIM laser light localizer function
SIM navigation system
SIM cassette holder
3.2 Electrical safety,
checks:
SIE fluoroscopy timer
SIE audible warning signal
SIE compulsory radiation switch-off
SIE check of the radiation release switches
SIE dose rate
SIE radiation indicators
SIE iris collimator
SIE setting of cassette images
SIE Area dose product measuring unit
2.5 Are the required operator documents complete, present and legible?
Operating instructions must be available.
Ask the customer for additional required operator documents and check them for completeness, availability, and legibility.
2.6 Are all warning labels in place and rec­ognizable?
3.1 Mechanical safety,
checks:
SIM warning signs
SIM ID labels
2.7 Are the operating symbols, light indica­tors and button labels OK?
3.2 Electrical safety,
check:
SIE check of the operating functions
3. Electrical checks n.a.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 14 of 30
Medical Solutions
General information 15
Technical safety checks,
Test protocol check
3.1 Measurement of ground wire resis­tance
3.2 Measurement of leakage current or equivalent leakage current
3.2.1 Leakage current measurement (for units with fixed mains connection)
3.2.2 Equivalent leakage current measure­ment (for units without fixed mains con­nection)
Maintenance instructions check/
comments
3.2 Electrical safety,
check:
SIE ground wire test
5.1 Final work steps,
check:
SIE ground wire test
n.a.
n.a.
3.2 Electrical safety,
check:
Equivalent leakage current
5.1 Final work steps,
check:
SIE equivalent leakage current
3.3 Measurement of patient leakage cur-
n.a.
rent or equivalent patient leakage current
3.3.1 Patient leakage current measure-
n.a.
ment
3.3.2 Equivalent patient leakage current n.a.
4. Mechanical checks n.a.
4.1. Are all wall, ceiling and floor mount­ings secure and without damage?
3.1 Mechanical safety,
checks:
SIM monitor(s) / displays
4.2 Are all mechanically moved system parts clean and smoothly running (lubri­cated, if required)?
3.1 Mechanical safety,
checks:
SIM cover panels
SIM C-arm
SIM lifting column
4.3 Are the wire cables, belts and spindles free of wear and tear?
3.1 Mechanical safety,
checks:
SIM lifting column
SIM 3D reconstruction
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 15 of 30
16 General information
Technical safety checks,
Test protocol check
4.4 Mobile equipment: Are the wheels, castors and brakes OK?
Maintenance instructions check/
comments
3.1 Mechanical safety,
checks:
SIM foot brake
SIM brakes
SIM wheels and castors
4.5 Is there any unusual noise during operation (e.g. gearing)?
3.1. Mechanical safety,
checks:
SIM C-arm
SIM lifting column
5. Functional checks n.a.
5.1. Are the emergency stop switches functional?
3.1. Mechanical safety,
check:
SIM emergency stop switch
3.2 Electrical safety,
5.2 Are the warning devices functioning properly?
5.3. Do all system movements stop prop­erly in their end positions?
5.4 Does the collision protection stop sys­tem movements properly?
5.5 Are the safety distances (wall, floor, ceiling) met or ensured by other measures (e.g. light barriers)?
check:
SIE 3D reconstruction
3.2 Electrical safety,
checks:
SIE audible warning signal
SIE radiation indicators
3.1 Mechanical safety,
check:
SIM lifting column
3.2 Electrical safety,
check:
SIE 3D reconstruction
Check the collision protection, provided that it is installed.
Otherwise: n.a.
Check the measures provided that they are installed.
Otherwise: n.a.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 16 of 30
Medical Solutions
General information 17
Technical safety checks,
Test protocol check
5.6 Are other safety cutoffs (e.g. safety step plate) functioning properly?
Maintenance instructions check/
comments
Check the measures provided that they are installed.
Otherwise: n.a.
5.7 Are the table and system positions
n.a. reproducible (e.g. zero positions, slice height, etc.)?
6. Product-specific checks 3.2 Electrical safety,
checks
SIE 3D reconstruction
SIE 3D reconstruction in combination with
a navigation system
SIE IQ quick test
7. Test result/evaluation: Evaluate the long-term trends of the
ground wire resistance and the equivalent
leakage current by comparing the current
measuring values to those of preceding
technical safety checks.
A sudden or unexpected increase in the
measured values may indicate a
safety-related defect - even if the limit val-
ues are not exceeded.
Follow the instructions for checks in the
section on final work steps.
Enter the check result.
Have the operator, or a person authorized
by the operator, sign the TSC protocol.
Hand the TSC protocol over to the opera-
tor, or a person authorized by the opera-
tor.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 17 of 30
18 Inspection of exterior and surroundings
2-

2 Inspection of exterior and surroundings

2.1 Inspection of exterior

PMP Damage
Inspect the entire system for damage, such as damage to the housing or paint.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 18 of 30
Medical Solutions
Inspection of exterior and surroundings 19

2.2 Inspection of surroundings

2.2.1 Power outlets

SIE Damage
SIE Line voltage
SIE Internal line impedance
Inspect the power outlets used for system operation for damage.
Measure the line voltage and compare it to the line voltage label on the monitor trolley.
Measure the internal line impedance.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 19 of 30
20 Safety inspection
3-

3 Safety inspection

3.1 Mechanical safety

SIM Cover panels
Remove the covers from the ARCADIS Orbic stand and monitor trolley.
Check the cover panels for mechanical damage.
SIM Cable deflectors
Inspect the cable deflectors on the stand and on the monitor trolley.
SIM I.I. laser light localizer mechanics (if present)
Inspect the I.I. laser light localizer for mechanical damage.
Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating.
When doing this, pay special attention to the tension band and its closure.
Check the I.I. ring for damage.
SIM I.I. laser light localizer function (if present)
Maintain the I.I. laser light localizer according to the installation and setting instructions
RXR2-120.814.01..
SIM Laser light localizer mechanics (if present)
Inspect the I.I. cap of the laser light localizer for mechanical damage.
Install the I.I. cap of the laser light localizer on the I.I. and ensure proper locking and
seating.
Check the I.I. ring for damage.
SIM Laser light localizer function (if present)
Maintain the laser light localizer according to the setting instructions
SPR2-230.812.03...
SIM Navigation system (if present)
Check the I.I. ring for damage.
Install the I.I. cap of the navigation system and ensure proper locking and seating.
SIM Foot brake
Check the braking effect of the foot brakes of the ARCADIS Orbic stand and monitor
cart on a flat surface.
SIM Brakes
Use the spring balance to check whether the orbital brake, angulation brake, swivel
brake, horizontal lift brake, and vertical lift brake reach the defined braking values in the braked state (Replacement of Parts Instructions for ARCADIS Orbic).
SIM C-arm
Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
SIM Wheels and castors
Move the ARCADIS Orbic in a straight line on a level floor.
Evaluate the straight and quiet movement of the ARCADIS Orbic.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 20 of 30
Medical Solutions
Safety inspection 21
Replace any defective wheels or castors.
SIM Lifting column
NOTE
There may not be any additional weight on the C-Arm, e.g. lead aprons or other cover panels, during these checks and adjust­ments.
Switch on the system.
Electrically move the lifting column over its entire lift range.
- while listening for noises and checking for play in the bearings.
- the lifting column movement must switch off automatically when the end positions are reached.
Operate the emergency stop button.
- The motorized vertical lift is blocked.
Unlock the emergency stop button by pressing it lightly and turning to the left.
Switch off the system.
Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIM Emergency stop switch
Press the emergency stop switch
The lifting column may not be raised or lowered.
Release the emergency stop switch
The lifting can be moved again.
SIM Warning signs
Ensure that all required warning labels are attached and in good condition.
- Replace any labels that are difficult to read.
SIM ID labels
Ensure that all required ID labels are attached and in good condition.
- Replace any labels that are difficult to read.
SIM TFT monitor(s)
Check monitor(s) for damage.
Check monitor(s) for proper attachment to the monitor cart.
Check additional monitors for damage and proper attachment to the wall support/moni-
tor support system (if applicable).
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 21 of 30
22 Safety inspection

3.2 Electrical safety

SIE Cables and plugs
Check visible system cables and plugs for damage.
SIE Fluoroscopy timer
Check: See section on "Compulsory radiation switch off".
SIE Acoustic warning signal
Check: See section on "Compulsory radiation switch off".
SIE Compulsory radiation switch off
Switch the system on
X
SIE Check the radiation release switch
Test the functioning of the audible warning signal and compulsory radiation switch off (if
required) according to the country-specific regulations. Also check the functioning of the fluoroscopic timer in this process.
NOTE
The acoustic warning signal must sound every 4, 5.5, or 9.5 min­utes of the fluoroscopic time depending on the programming. It sounds again every 5 or 10 minutes. It is turned off by pressing the
-0- button once. Pressing this button again resets the fluoroscopic timer to "0". Depending on the country-specific programming, ei­ther the required radiation shut off does not occur or it occurs ev­ery 5 or 10 minutes and then every 5 or 10 minutes after that.
Check the functioning of the hand and foot switches for radiation release.
X
Check the cables of the radiation release switches for mechanical damage.
Move the cables and inspect them for breakage.
SIE Dose rate
Switch the system on
X
SIE Radiation indicator
Check the dose rate (see the adjustment instructions for ARCADIS Orbic)
Activate fluoroscopy briefly.
X ¹ The radiation indicator on the operating part of the ARCADIS Orbic stand and
the radiation indicator on the monitor cart must light.
Switch the system off.
SIE Iris collimator
Check the iris collimator and correct it if necessary.
X - Select I.I. full format and activate fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIE Voltage discharge rubber
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 22 of 30
Medical Solutions
Safety inspection 23
Check the voltage discharge rubber on the ARCADIS Varic stand and the wheels and
castors of the monitor cart for damage or contamination, and replace or clean them as necessary.
SIE Ground wire test
Test the ground wire with the ground wire tester according to ARTD-002.731.17.
NOTE
SIE Equivalent leakage current
The equivalent leakage current is to be measured and documented using the equivalent leakage current tester according to ARTD-002.731.17 in the scope of validity of DIN VDE
0751.
NOTE
SIE 3D reconstruction (if present)
NOTE
The equivalent leakage current is to be measured and documented using the equivalent leakage current tester according to ARTD-002.731.17 in the scope of validity of DIN VDE 0751. Refer to the section "Final work steps".
The instructions according to ARTD-002.731.17 are to be adhered to outside the scope of validity of DIN VDE 0751.
The calibration of the 3D reconstruction is part of the safety in­spection. It can, however, also be performed together with the IQ quick test.
See chapter 5, "IQ quick test".
Check the belt tension of the toothed belt for motor-controlled orbital movement.
Press the emergency stop switch and perform a 3D scan.
The motor-controlled orbital movement must not start up.
Deactivate (disengage) the emergency stop switch.
X
Start a further 3D scan. The 3D scan must be executed properly. The motor-controlled
orbital movement must stop at the end position of the C-arm.
Calibrate the 3D reconstruction. See "IQ quick test".
Check the 3D limiting resolution. See "IQ quick test".
If the system is equipped with a navigation system, the required calibration of the 3D re-
construction has to be performed together with a navigation system engineer. Before starting calibration, make sure that the I.I. cap of the navigation system is in place. See "IQ quick test".
SIE 3D reconstruction with navigation system (if present)
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 23 of 30
24 Safety inspection
NOTE
X Check the overall accuracy of the navigation system together with a engineer of the
navigation system manufacturer. See "IQ quick test".
SIE Voltage discharge rubber
Checking the overall accuracy of the navigation system in combi­nation of the 3D reconstruction is part of the safety inspection. It can, however, also be performed together with the IQ quick test.
See chapter 5, "IQ quick test".
Inspect the voltage discharge rubber on the SIREMOBIL stand and monitor trolley for
damage or dirt. If necessary, clean or replace it.
SIE Ground wire test
Measure the ground wire with the ground wire tester according to TD00-000.860.01... .
NOTE
The ground wire test is part of the safety inspection. However, it cannot be performed until all maintenance work has been complet­ed. Refer to the "Final work steps" section.
SIE Equivalent leakage current
Perform the equivalent leakage current test, with a closed system, according to
TD00-000.860.01... . Take down and evaluate the determined values. Document the
measuring procedure.
The equivalent leakage current must not exceed 2 mA.
NOTE
The equivalent leakage current test is part of the safety inspection. However, it is recommended to perform it only at the end of main­tenance, with a closed system. Refer to the "Final work steps" sec­tion.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 24 of 30
Medical Solutions

Maintenance, operating value/functional inspection 25

4-
4 Maintenance, operating value/functional inspec­tion

4.1 Maintenance

PMP Cleaning the system
Clean the entire system:
- Visible cables
- Exterior surfaces
- Contact surface of wheels
- Interior space
PMP System ventilation
Clean the ventilation slots of the ARCADIS Orbic stand and monitor cart.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 25 of 30
26 Maintenance, operating value/functional inspection

4.2 Operating value inspection

SIE Dose rate
X
Switch he system on.
Check the dose rate (see the ARCADIS Orbic service instructions for this purpose).
PMF Event log
Read out and evaluate the system event log.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 26 of 30
Medical Solutions
Maintenance, operating value/functional inspection 27

4.3 Functional inspection

SIE TFT monitor
Check the TFT monitor with the image quality assurance test.
SIE Area dose product measuring unit (if present)
Check the area dose product measuring unit and the calibration. (See the Main System
Adjustment Instructions SPR2-310.842.01...).
PMF Laser camera connection (if present)
Check the functioning of the laser camera connection.
PMF Check the operating function.
Check all system operating functions.
PMF Monitor display of the iris collimator aperture
Select I.I. full format.
Close the iris collimator (X-iris) completely.
Mark the diameter of the collimated iris collimator aperture on the monitor.
X
Release fluoroscopy briefly. The actual diameter of the iris collimator is visible. The
previously indicated monitor display must match the position and diameter of the actual iris collimator aperture.
Open the iris collimator completely.
Mark the diameter of the collimated iris collimator opening on the monitor.
X
Release fluoroscopy briefly. The actual diameter of the iris collimator is visible. The
previously indicated monitor display must match the position and diameter of the actual iris collimator aperture.
Select I.I. zoom format.
Close the iris collimator (X-iris) completely.
Mark the diameter of the collimated iris collimator aperture on the monitor.
X
Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible. The previously indicated monitor display must match the position and diameter of the actual iris collimator aperture.
Open the iris collimator completely.
Mark the diameter of the collimated iris collimator opening on the monitor.
X
PMF PMF Monitor display of the slot diaphragm positions
Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible. The previously indicated monitor display must match the position and diameter of the actual iris collimator aperture.
Select I.I. full format.
Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 27 of 30
28 Maintenance, operating value/functional inspection
X Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is visible. The previously displayed monitor display must match the position and angle of rotation of the actual slot diaphragm position.
Open the slot diaphragm completely and rotate it again.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
X
X
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is visible. The previously displayed monitor display must match the position and angle of rotation of the actual slot diaphragm position.
Select I.I. zoom format.
Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is visible.
The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
Open the slot diaphragm completely and rotate it again.
Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is visible. The previously displayed monitor display must match the position and angle of rotation of the actual slot diaphragm position.
NOTE
PMF Battery replacement in the UPS
The lead-acid battery installed in the UPS must be replaced preventively every 36 months.
The replacement procedure is described in the supplied installation instructions from the manufacturer as well as in Replacement of Parts Instructions SPR2-310.841.01.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
Perform this check with the lowest possible kV values so that the front edges of the slot diaphragm plate are effectively visualized.
08.04 CS SD 24
Page 28 of 30
Medical Solutions
Final result/quality inspection and general maintenance29
5-
5 Final result/quality inspection and general mainte­nance
SIE Image quality (IQ) quick test
Perform the image quality test according to the image quality assurance test instruc-
tions
for ARCADIS Orbic. Include additional monitors (if applicable).
PMP General maintenance
Entire system: Touch up any paint damage as needed.
Siemens AG SPR2-320.831.02.01.02 ARCADIS Orbic Medical Solutions
08.04 CS SD 24
Page 29 of 30
30Final result/quality inspection and general maintenance

5.1 Final work steps

SIE Ground wire test
SIE Ground wire resistance
Perform the ground wire test with the system closed (ARTD- 002.731.17.).
Document and evaluate the determined values.
The ground wire resistance must not exceed 0.2 Ohms.
NOTE
SIE Leakage current
Measure the leakage current according to ARTD-002.731.17....
Document and evaluate the determined values.
The limit value of 2.5 mA may not be exceeded.
NOTE
Evaluate the results by comparing the first measured value to the corresponding values documented during preceding maintenance procedures or safety checks.
A sudden or unexpected increase in the measured values may in­dicate a defect in the ground wire connections (ground wire or contacts) - even if the limit value of 0.2 Ohms is not exceeded.
Evaluate the results by comparing the first measured value to the corresponding values documented during preceding maintenance procedures or safety checks.
A sudden or unexpected increase in the measured values may in­dicate that an error occurred in the primary circuit of the power supply (damaged insulation, damage caused by water ingress or humidity, defective interference suppressor, etc.) - even if the limit value of 2.5 mA is not exceeded.
ARCADIS Orbic SPR2-320.831.02.01.02 Siemens AG
08.04 CS SD 24
Page 30 of 30
Medical Solutions