The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.04
2004
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 2 of 30
Medical Solutions
Table of Contents3
0Table of Contents
1 _______ General information______________________________________________ 4
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 3 of 30
4General information
1-
1General information
1.1Requirements
NOTE
If the system is equipped with 3D reconstruction, in combination
with a navigation system, the required maintenance has to be performed together with a navigation system engineer.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 4 of 30
Medical Solutions
General information5
1.2 Required documents
•Safety information according to ARTD, part 2
•Maintenance protocolSPR2-320.832.02..
•Main system adjustment instructionsSPR2-310.842.02..
1.2.1 Systems equipped with a laser light localizer
•Setting instructionsSPR2-230.812.03..
1.2.2 Systems equipped with an I.I. laser light localizer
•Installation and setting instructionsRXR2-120.814.01..
•Replacement instructions of the available
TFT display
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 5 of 30
6General information
1.3 Required tools, measurement and auxiliary devices
NOTE
All indicated tools, measurement and auxiliary devices, with the
exception of those marked with a " * ", are described in the ARTD
(Part 3).
•Standard tool kit*
•Set of Allen keys*
•Spring balance rated for up to 200 Ne.g.44 15 113 RH090
•Equivalent leakage current testere.g. Safety Tester
Unimet 1100
•Ground wire testere.g. Safety Tester
Unimet 1100
51 38 727Y0766
51 38 727Y0766
•Internal line resistance tester84 28 104 Y4337
•DVMe. g. Fluke 8060A97 02 101 Y4290
•Luminous density measuring devicee.g. SPOTMETER for
SMfit ACT 77 52 848
•Densitometer (for previous MULTI-
SPOT
e.g.97 02 416Y1996
2000 option)
•Dosimeter,e.g. PTW Diados97 17 612 Y0388
•Dynamic test kit37 90 156 X1963
•Precision X-ray filter99 00 598 XE999
•Set of resolution tests28 71 820 RE999
•Set of radiation filters (10 x 0.3 mm)e.g.44 06 120 RV090
•Set of radiation filterse.g.97 98 596 G5321
•25 mm AL measuring stand, type
26765 acc. to DIN 6868 Part 50*
or
e.g.
•1.2 mm Cu (e.g. of a set of radiation fil-
ters 44 06 120) plus
•17 micrometer Cu strips11 67 662
•Centering cross (only with Diamentor)e.g.96 60 051 RE999
•Calibration phantom
77 51 620 G5486
(for previous 3D reconstruction option)
•Cable for connecting service PC to host99 00 440 RE999
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 6 of 30
Medical Solutions
General information7
•Service software*(see logbook)
•Sealing compound34 43 009
•Special oil (Optimol Optipit)55 07 525
•Special oil (Slic Pac PTFE)55 07 517
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 7 of 30
8General information
1.4 Spare parts which may be needed
•Cable deflector steering castor large/ARCADIS Orbic stand
•Cable deflector wheels small/ARCADIS Orbic stand
•Touch-up paint
•Spray paint
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 8 of 30
Medical Solutions
General information9
1.5 Emphasized texts
!
DANGER
!
WARNING
!
CAUTION
NOTICE
Fig. 1:Safety Notes
DANGER indicates when there is an immediate danger that
l e a d s to death or serious physical injury.
WARNING indicates a risk of danger that m a y l e a d to death
or serious physical injury.
CAUTION used with the safety alert symbol indicates a risk of
danger that leads to slight or moderate physical injury and/ or
damage to property.
NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads or may lead to a potential
situation which may result in an undesirable result or state other
than death, physical injury or property damage.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 9 of 30
10General information
1.6 Safety information and protective measures
NOTICE
The following instructions must be observed!
¹ When performing the work steps and tests, comply with
the following:
¹ - The product-specific safety information in the docu-
mentation, safety information TD00-000.860.01.03.01,
¹ - As well as the general safety information contained in
ARTD, part 2.
¹ Pull the power plug before starting to work on the ARCA-
DIS Orbic.
¹ Ensure compliance with the general safety requirements
when working with the system connected to the mains.
¹ Observe ESD guidelines!
¹ Before beginning maintenance, perform the ground wire
test according TD00-000.860.01.... The ground wire resis-
tance must not exceed 0.2 Ohms.
¹ Switch off the ARCADIS Orbic before replacing modules
or boards.
¹ Perform the ground wire test according to
TD00-000.860.01... upon completion of the work and after
all covers have been reattached. Take down and evaluate
the determined values. The ground wire resistance must
not exceed 0.2 Ohms.
¹ After all work has been completed and all cover panels
have been installed, perform the ground wire measurement according to ARTD-002.731.17.
¹ The ground wire resistance must not exceed 0.2 Ohms.
¹ When performing service work on the power-on module,
(replacing the power-on module or replacing the power
cable), the equivalent leakage current must be measured
and recorded.
¹ Checks and settings that need to be performed under
X-ray radiation are identified by the radiation warning
symbol x . While performing adjustments labeled in
this manner, radiation protective clothing must be worn.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 10 of 30
Medical Solutions
General information11
1.7 Explanation of abbreviations
Abbrev.Explanation
SISafety inspection
SIEElectrical safety
SIMMechanical safety
PMPreventive maintenance
PMPPeriodic preventive maintenance
PMAPreventive maintenance adjustments
PMFPreventive check of operating values/functions
QSystem quality, image quality
QIQImage quality
QSQSystem quality
SWSoftware maintenance
The steps identified by these abbreviations are listed as checkpoints in the Maintenance
Protocol and must be checked off there accordingly.
NOTE
The sequence for the complete inspection and maintenance is located in the following pages.
Each step must be performed annually, unless otherwise specified.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 11 of 30
12General information
1.8 Maintenance interval
1.8.1 System maintenance interval
NOTE
The maintenance interval applies to all maintenance work, except
for maintenance of the 3D reconstruction in combination with a
navigation system.
•12 months
1.8.2 Maintenance interval for 3D reconstruction in combination with a navigation system
NOTE
NOTE
If the system is equipped with the 3D reconstruction option in
combination with a navigation system, perform the maintenance
work listed below in the specified maintenance interval.
Take into account that the maintenance work listed below has to
be performed together with an engineer of the navigation system
manufacturer.
-SIM navigation system (if present)
I.I. ring damage
Correction of the seating of the I.I. cap
-SIE 3D reconstruction (if present)
Calibration of 3D reconstruction
Check of the limiting resolution
-SIE 3D reconstruction with navigation system (if present)
Check the overall navigation accuracy
•6 months or after 150 patients at the latest (approx. 300x 3D scans)
UPS (Uninterrupted Power Supply)
•The lead-acid batteries of the UPS in the monitor cart are to be replaced every 36
months.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 12 of 30
Medical Solutions
General information13
1.9 Technical Safety Checks (TSC)
Abbreviation: TSC = Technical Safety Checks
NOTE
Technical safety checks,
Test protocol check
Within the purview of DIN VDE 0751-1, the operator of medical engineering products has to perform technical safety checks at regular intervals.
The checks listed in these maintenance instructions contain all
technical safety checks according to DIN VDE 0751-1 except for
the check of the completeness, presence and legibility of the necessary operator documents.
A separate protocol with the print number SP00.000.834.01... is
available for each technical safety check.
Within the purview of DIN VDE 0751-1, we recommend documenting the maintenance results both in the maintenance protocol and
in the TSC protocol. The protocols should be filled out completely
and handed over to the client after maintenance is completed.
In the table below each technical safety check listed in the document SP00.000.834.01... is assigned to the corresponding maintenance work.
Maintenance instructions check/
comments
1 Identification dataPlease fill in the fields.
2 General checksn.a.
2.1 Visual inspection: Is there any damage
affecting safety?
2.2 Are all cables and cable guides secure
and without any visible damage?
3.1 Mechanical safety,
Checks:
SIM cover panels
SIM cable deflectors
SIM foot brake
SIM brakes
SIM emergency stop switch
SIM monitors/displays
3.2 Electrical safety, check: SIE cable and
plug
SIE voltage discharge rubber
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 13 of 30
14General information
Technical safety checks,
Test protocol check
2.3 Are undamaged accessories being
used?
2.4 Are all unit radiation protection devices
present and undamaged?
(No on-site radiation shield)
Maintenance instructions check/
comments
3.1 Mechanical safety,
checks:
SIM I.I. laser light localizer mechanics
SIM I.I. laser light localizer function
SIM laser light localizer mechanics
SIM laser light localizer function
SIM navigation system
SIM cassette holder
3.2 Electrical safety,
checks:
SIE fluoroscopy timer
SIE audible warning signal
SIE compulsory radiation switch-off
SIE check of the radiation release
switches
SIE dose rate
SIE radiation indicators
SIE iris collimator
SIE setting of cassette images
SIE Area dose product measuring unit
2.5 Are the required operator documents
complete, present and legible?
Operating instructions must be available.
Ask the customer for additional required
operator documents and check them for
completeness, availability, and legibility.
2.6 Are all warning labels in place and recognizable?
3.1 Mechanical safety,
checks:
SIM warning signs
SIM ID labels
2.7 Are the operating symbols, light indicators and button labels OK?
3.2 Electrical safety,
check:
SIE check of the operating functions
3. Electrical checksn.a.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 14 of 30
Medical Solutions
General information15
Technical safety checks,
Test protocol check
3.1 Measurement of ground wire resistance
3.2 Measurement of leakage current or
equivalent leakage current
3.2.1 Leakage current measurement (for
units with fixed mains connection)
3.2.2 Equivalent leakage current measurement (for units without fixed mains connection)
Maintenance instructions check/
comments
3.2 Electrical safety,
check:
SIE ground wire test
5.1 Final work steps,
check:
SIE ground wire test
n.a.
n.a.
3.2 Electrical safety,
check:
Equivalent leakage current
5.1 Final work steps,
check:
SIE equivalent leakage current
3.3 Measurement of patient leakage cur-
n.a.
rent or equivalent patient leakage current
3.3.1 Patient leakage current measure-
n.a.
ment
3.3.2 Equivalent patient leakage currentn.a.
4. Mechanical checksn.a.
4.1. Are all wall, ceiling and floor mountings secure and without damage?
3.1 Mechanical safety,
checks:
SIM monitor(s) / displays
4.2 Are all mechanically moved system
parts clean and smoothly running (lubricated, if required)?
3.1 Mechanical safety,
checks:
SIM cover panels
SIM C-arm
SIM lifting column
4.3 Are the wire cables, belts and spindles
free of wear and tear?
3.1 Mechanical safety,
checks:
SIM lifting column
SIM 3D reconstruction
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 15 of 30
16General information
Technical safety checks,
Test protocol check
4.4 Mobile equipment: Are the wheels,
castors and brakes OK?
Maintenance instructions check/
comments
3.1 Mechanical safety,
checks:
SIM foot brake
SIM brakes
SIM wheels and castors
4.5 Is there any unusual noise during
operation (e.g. gearing)?
3.1. Mechanical safety,
checks:
SIM C-arm
SIM lifting column
5. Functional checksn.a.
5.1. Are the emergency stop switches
functional?
3.1. Mechanical safety,
check:
SIM emergency stop switch
3.2 Electrical safety,
5.2 Are the warning devices functioning
properly?
5.3. Do all system movements stop properly in their end positions?
5.4 Does the collision protection stop system movements properly?
5.5 Are the safety distances (wall, floor,
ceiling) met or ensured by other measures
(e.g. light barriers)?
check:
SIE 3D reconstruction
3.2 Electrical safety,
checks:
SIE audible warning signal
SIE radiation indicators
3.1 Mechanical safety,
check:
SIM lifting column
3.2 Electrical safety,
check:
SIE 3D reconstruction
Check the collision protection, provided
that it is installed.
Otherwise: n.a.
Check the measures provided that they
are installed.
Otherwise: n.a.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 16 of 30
Medical Solutions
General information17
Technical safety checks,
Test protocol check
5.6 Are other safety cutoffs (e.g. safety
step plate) functioning properly?
Maintenance instructions check/
comments
Check the measures provided that they
are installed.
Otherwise: n.a.
5.7 Are the table and system positions
n.a.
reproducible (e.g. zero positions, slice
height, etc.)?
6. Product-specific checks3.2 Electrical safety,
checks
SIE 3D reconstruction
SIE 3D reconstruction in combination with
a navigation system
SIE IQ quick test
7. Test result/evaluation:Evaluate the long-term trends of the
ground wire resistance and the equivalent
leakage current by comparing the current
measuring values to those of preceding
technical safety checks.
A sudden or unexpected increase in the
measured values may indicate a
safety-related defect - even if the limit val-
ues are not exceeded.
Follow the instructions for checks in the
section on final work steps.
Enter the check result.
Have the operator, or a person authorized
by the operator, sign the TSC protocol.
Hand the TSC protocol over to the opera-
tor, or a person authorized by the opera-
tor.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 17 of 30
18Inspection of exterior and surroundings
2-
2Inspection of exterior and surroundings
2.1Inspection of exterior
PMPDamage
Inspect the entire system for damage, such as damage to the housing or paint.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 18 of 30
Medical Solutions
Inspection of exterior and surroundings19
2.2 Inspection of surroundings
2.2.1 Power outlets
SIEDamage
SIELine voltage
SIEInternal line impedance
•Inspect the power outlets used for system operation for damage.
•Measure the line voltage and compare it to the line voltage label on the monitor trolley.
•Measure the internal line impedance.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 19 of 30
20Safety inspection
3-
3Safety inspection
3.1Mechanical safety
SIMCover panels
•Remove the covers from the ARCADIS Orbic stand and monitor trolley.
•Check the cover panels for mechanical damage.
SIMCable deflectors
•Inspect the cable deflectors on the stand and on the monitor trolley.
•Inspect the I.I. laser light localizer for mechanical damage.
•Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating.
When doing this, pay special attention to the tension band and its closure.
•Check the I.I. ring for damage.
SIMI.I. laser light localizer function (if present)
•Maintain the I.I. laser light localizer according to the installation and setting instructions
RXR2-120.814.01..
SIMLaser light localizer mechanics (if present)
•Inspect the I.I. cap of the laser light localizer for mechanical damage.
•Install the I.I. cap of the laser light localizer on the I.I. and ensure proper locking and
seating.
•Check the I.I. ring for damage.
SIMLaser light localizer function (if present)
•Maintain the laser light localizer according to the setting instructions
SPR2-230.812.03...
SIMNavigation system (if present)
•Check the I.I. ring for damage.
•Install the I.I. cap of the navigation system and ensure proper locking and seating.
SIMFoot brake
•Check the braking effect of the foot brakes of the ARCADIS Orbic stand and monitor
cart on a flat surface.
SIMBrakes
•Use the spring balance to check whether the orbital brake, angulation brake, swivel
brake, horizontal lift brake, and vertical lift brake reach the defined braking values in the
braked state (Replacement of Parts Instructions for ARCADIS Orbic).
SIMC-arm
•Perform all C-arm movements, checking for slackness of the bearings and bearing
noises.
SIMWheels and castors
•Move the ARCADIS Orbic in a straight line on a level floor.
•Evaluate the straight and quiet movement of the ARCADIS Orbic.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 20 of 30
Medical Solutions
Safety inspection21
•Replace any defective wheels or castors.
SIMLifting column
NOTE
There may not be any additional weight on the C-Arm, e.g. lead
aprons or other cover panels, during these checks and adjustments.
•Switch on the system.
•Electrically move the lifting column over its entire lift range.
- while listening for noises and checking for play in the bearings.
- the lifting column movement must switch off automatically when the end positions are
reached.
•Operate the emergency stop button.
- The motorized vertical lift is blocked.
•Unlock the emergency stop button by pressing it lightly and turning to the left.
•Switch off the system.
•Lubricate the lifting column.
- Lubricate the bearing surfaces of the lifting column guide and the spindle.
SIMEmergency stop switch
•Press the emergency stop switch
The lifting column may not be raised or lowered.
•Release the emergency stop switch
The lifting can be moved again.
SIMWarning signs
•Ensure that all required warning labels are attached and in good condition.
- Replace any labels that are difficult to read.
SIMID labels
•Ensure that all required ID labels are attached and in good condition.
- Replace any labels that are difficult to read.
SIMTFT monitor(s)
•Check monitor(s) for damage.
•Check monitor(s) for proper attachment to the monitor cart.
•Check additional monitors for damage and proper attachment to the wall support/moni-
tor support system (if applicable).
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 21 of 30
22Safety inspection
3.2 Electrical safety
SIECables and plugs
•Check visible system cables and plugs for damage.
SIEFluoroscopy timer
•Check: See section on "Compulsory radiation switch off".
SIEAcoustic warning signal
•Check: See section on "Compulsory radiation switch off".
SIECompulsory radiation switch off
•Switch the system on
X
SIECheck the radiation release switch
•Test the functioning of the audible warning signal and compulsory radiation switch off (if
required) according to the country-specific regulations. Also check the functioning of
the fluoroscopic timer in this process.
NOTE
The acoustic warning signal must sound every 4, 5.5, or 9.5 minutes of the fluoroscopic time depending on the programming. It
sounds again every 5 or 10 minutes. It is turned off by pressing the
-0- button once. Pressing this button again resets the fluoroscopic
timer to "0". Depending on the country-specific programming, either the required radiation shut off does not occur or it occurs every 5 or 10 minutes and then every 5 or 10 minutes after that.
•Check the functioning of the hand and foot switches for radiation release.
X
•Check the cables of the radiation release switches for mechanical damage.
•Move the cables and inspect them for breakage.
SIEDose rate
•Switch the system on
X
SIERadiation indicator
•Check the dose rate (see the adjustment instructions for ARCADIS Orbic)
•Activate fluoroscopy briefly.
X¹ The radiation indicator on the operating part of the ARCADIS Orbic stand and
the radiation indicator on the monitor cart must light.
•Switch the system off.
SIEIris collimator
•Check the iris collimator and correct it if necessary.
X- Select I.I. full format and activate fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
- Select a zoom format and release fluoroscopy briefly.
- The collimator blades must be clearly visible at the edges of the image.
SIEVoltage discharge rubber
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 22 of 30
Medical Solutions
Safety inspection23
•Check the voltage discharge rubber on the ARCADIS Varic stand and the wheels and
castors of the monitor cart for damage or contamination, and replace or clean them as
necessary.
SIEGround wire test
•Test the ground wire with the ground wire tester according to ARTD-002.731.17.
NOTE
SIEEquivalent leakage current
The equivalent leakage current is to be measured and documented using the equivalent
leakage current tester according to ARTD-002.731.17 in the scope of validity of DIN VDE
0751.
NOTE
SIE3D reconstruction (if present)
NOTE
The equivalent leakage current is to be measured and documented
using the equivalent leakage current tester according to
ARTD-002.731.17 in the scope of validity of DIN VDE 0751. Refer to
the section "Final work steps".
The instructions according to ARTD-002.731.17 are to be adhered
to outside the scope of validity of DIN VDE 0751.
The calibration of the 3D reconstruction is part of the safety inspection. It can, however, also be performed together with the IQ
quick test.
See chapter 5, "IQ quick test".
•Check the belt tension of the toothed belt for motor-controlled orbital movement.
•Press the emergency stop switch and perform a 3D scan.
The motor-controlled orbital movement must not start up.
•Deactivate (disengage) the emergency stop switch.
X
•Start a further 3D scan. The 3D scan must be executed properly. The motor-controlled
orbital movement must stop at the end position of the C-arm.
•Calibrate the 3D reconstruction. See "IQ quick test".
•Check the 3D limiting resolution. See "IQ quick test".
•If the system is equipped with a navigation system, the required calibration of the 3D re-
construction has to be performed together with a navigation system engineer. Before
starting calibration, make sure that the I.I. cap of the navigation system is in place. See
"IQ quick test".
SIE3D reconstruction with navigation system (if present)
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 23 of 30
24Safety inspection
NOTE
X•Check the overall accuracy of the navigation system together with a engineer of the
navigation system manufacturer. See "IQ quick test".
SIEVoltage discharge rubber
Checking the overall accuracy of the navigation system in combination of the 3D reconstruction is part of the safety inspection. It
can, however, also be performed together with the IQ quick test.
See chapter 5, "IQ quick test".
•Inspect the voltage discharge rubber on the SIREMOBIL stand and monitor trolley for
damage or dirt. If necessary, clean or replace it.
SIEGround wire test
•Measure the ground wire with the ground wire tester according to TD00-000.860.01... .
NOTE
The ground wire test is part of the safety inspection. However, it
cannot be performed until all maintenance work has been completed. Refer to the "Final work steps" section.
SIEEquivalent leakage current
•Perform the equivalent leakage current test, with a closed system, according to
TD00-000.860.01... . Take down and evaluate the determined values. Document the
measuring procedure.
The equivalent leakage current must not exceed 2 mA.
NOTE
The equivalent leakage current test is part of the safety inspection.
However, it is recommended to perform it only at the end of maintenance, with a closed system. Refer to the "Final work steps" section.
•Check the TFT monitor with the image quality assurance test.
SIEArea dose product measuring unit (if present)
•Check the area dose product measuring unit and the calibration. (See the Main System
Adjustment Instructions SPR2-310.842.01...).
PMFLaser camera connection (if present)
•Check the functioning of the laser camera connection.
PMFCheck the operating function.
•Check all system operating functions.
PMFMonitor display of the iris collimator aperture
•Select I.I. full format.
•Close the iris collimator (X-iris) completely.
•Mark the diameter of the collimated iris collimator aperture on the monitor.
X
•Release fluoroscopy briefly. The actual diameter of the iris collimator is visible. The
previously indicated monitor display must match the position and diameter of the actual
iris collimator aperture.
•Open the iris collimator completely.
•Mark the diameter of the collimated iris collimator opening on the monitor.
X
•Release fluoroscopy briefly. The actual diameter of the iris collimator is visible. The
previously indicated monitor display must match the position and diameter of the actual
iris collimator aperture.
•Select I.I. zoom format.
•Close the iris collimator (X-iris) completely.
•Mark the diameter of the collimated iris collimator aperture on the monitor.
X
•Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible. The previously indicated monitor
display must match the position and diameter of the actual iris collimator aperture.
•Open the iris collimator completely.
•Mark the diameter of the collimated iris collimator opening on the monitor.
X
PMFPMF Monitor display of the slot diaphragm positions
•Release fluoroscopy briefly.
The actual diameter of the iris collimator is visible. The previously indicated monitor
display must match the position and diameter of the actual iris collimator aperture.
•Select I.I. full format.
•Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
•Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible. The previously displayed monitor display must match the position and angle of
rotation of the actual slot diaphragm position.
•Open the slot diaphragm completely and rotate it again.
•Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
X
X
X
•Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible. The previously displayed monitor display must match the position and angle of
rotation of the actual slot diaphragm position.
•Select I.I. zoom format.
•Completely close the slot diaphragm and rotate the slot diaphragm from its basic posi-
tion.
•Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
•Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible.
•The previously displayed monitor display must match the position and angle of rotation
of the actual slot diaphragm position.
•Open the slot diaphragm completely and rotate it again.
•Mark the distance and angle of rotation of the displayed slot diaphragm position on the
monitor.
•Release fluoroscopy briefly.
The actual position (distance of the plates and angle of rotation) of the slot diaphragm is
visible. The previously displayed monitor display must match the position and angle of
rotation of the actual slot diaphragm position.
NOTE
PMFBattery replacement in the UPS
The lead-acid battery installed in the UPS must be replaced preventively every 36
months.
The replacement procedure is described in the supplied installation instructions from the
manufacturer as well as in Replacement of Parts Instructions SPR2-310.841.01.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
Perform this check with the lowest possible kV values so that the
front edges of the slot diaphragm plate are effectively visualized.
08.04CS SD 24
Page 28 of 30
Medical Solutions
Final result/quality inspection and general maintenance29
5-
5Final result/quality inspection and general maintenance
SIEImage quality (IQ) quick test
•Perform the image quality test according to the image quality assurance test instruc-
tions
for ARCADIS Orbic. Include additional monitors (if applicable).
PMPGeneral maintenance
•Entire system: Touch up any paint damage as needed.
Siemens AGSPR2-320.831.02.01.02ARCADIS Orbic
Medical Solutions
08.04CS SD 24
Page 29 of 30
30Final result/quality inspection and general maintenance
5.1 Final work steps
SIEGround wire test
SIEGround wire resistance
•Perform the ground wire test with the system closed (ARTD- 002.731.17.).
Document and evaluate the determined values.
The ground wire resistance must not exceed 0.2 Ohms.
NOTE
SIELeakage current
•Measure the leakage current according to ARTD-002.731.17....
Document and evaluate the determined values.
The limit value of 2.5 mA may not be exceeded.
NOTE
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding maintenance
procedures or safety checks.
A sudden or unexpected increase in the measured values may indicate a defect in the ground wire connections (ground wire or
contacts) - even if the limit value of 0.2 Ohms is not exceeded.
Evaluate the results by comparing the first measured value to the
corresponding values documented during preceding maintenance
procedures or safety checks.
A sudden or unexpected increase in the measured values may indicate that an error occurred in the primary circuit of the power
supply (damaged insulation, damage caused by water ingress or
humidity, defective interference suppressor, etc.) - even if the limit
value of 2.5 mA is not exceeded.
ARCADIS OrbicSPR2-320.831.02.01.02Siemens AG
08.04CS SD 24
Page 30 of 30