222
HARTREY AVE r EVANSTON ILL. CABLE SHUREMICRO
EVANSTON PHONE OA
89000 . CHICAGO PHONE SH
3
1600
DATA SHEET
I
MODEL M7D STEREO
CUSTOM DYNETIC
CARTRIDGE
(Stylus Model N7D)
1
SPECIFICATIONS
Response Frequency Characteristics:
15,000 c.p.s.
OU+~U+ Voltage:
Channel Separation:
~ecommended Load Impedance:
nel) for Stereo. 22,000 ohms for Monaural (see connections).
Higher values will produce a
sponse.
Stylus:
Number N7D.
Radius .0007" (0.018mm) Diamond.
Compliance:
Tracking Force:
Inductance:
D. C. Resistance:
Terminals:
each channel. (See Fig.
Mounting:
or 4 machine screws. .437 (I l.lmm) mounting centers using #2
machine screws.
Weight:
4 terminals.
Standard
7.9 grams 1.28 ounces).
Packaged Weight:
General:
ridge has been developed for use in all high fidelity applications.
It
has been designed to connect into magnetic and constant ve-
locity inputs.
the Model M7D will
Discs Monaurally, and Stereo Discs Monaurally without excessive
wear and without distortion.
employed in both the Studio and Professional
the same important features, such as:
The Model M7D Stereo Custom Dynetic Phono Cart-
The Model M7D is completely compatible.
The Model M7D utilizes the same Moving Magnet principle
High needle compliance.
Low needle talk.
Low tracking forces.
Wide range frequency response.
Magnetic induction from external fields is reduced to
a minimum.
No magnetic attraction to steel turntables.
5
millivolts per channel.
More than 20 db at 1000 c.p.s.
47,000 ohms (per chan-
slight increase in high frequency re-
1
VERTICAL
HORIZONTAL
4 to 7 grams (5 grams optimum).
420 millihenries.
280 ohms.
2).
1/2'1
2 ounces (56.7 g).
play Stereo Discs Stereophonically, Monaural
3.5 x 10-4 centimeters
1
per dyne
An independent ground terminal for
(12.7mm) mounting centers using
From 20
In other words,
Dynetics and offers
to
#3
Mounting:
mounting centers. Two #3-48 machine screws are supplied with each
cartridge for mounting purposes.
cartridge can be mounted on
is important during installation, that the line on the front of the
cartridge be perpendicular to the turnable or record surface.
no adjustment is available to do this, a paper or metal shim can
be used on one mounting post to adjust angle of cartridge.
5
grams. When replacing magnetic cartridges, adjustment is
is
generally unnecessary. Should
usually be made
of arm. Forces up to
proper
The M7D Cartridge has standard
By
using #2 machine screws, the
.437 (I l.lmm) mounting centers.
DIMENSIONAL DRAWING
FIGURE
The recommended needle point force for optimum results
by adjusting the counter balance or spring at rear
7 grams may be used
cycling of the record changer.
I
it
be necessary, adjustment may
if
needed to provide
y2"
(12.7mm)
It
If
Copyright
27A142
1958
(10-63)
Shure Brothers.
Inc.
Printed
in
U.S.A.
Connections:
ment for connections having a separate ground terminal for
channel. (See Figure 2).
LEFT
CHANNEL
TERMINAL
GROUND
TERMINAL
For Stereo reproduction terminal R and its ground terminal
represent the right channel (outside groove wall). Terminal
its ground terminal represents the left channel (inside groove wall).
To use a 4-lead arrangement connect each section of the
cartridge to separate amplifiers. To prevent "ground loops" and
hum, no common connection should exist between the two amplifier
systems.
When used with a 3-lead arrangement, the
should be connected to both of the ground terminals at the cartridge. No other common ground connection should exist.
For single channel reproduction of Monaural or Stereo recordings, connect the "hot" lead to both R and
connect the ground or shield lead to both of the ground terminals
marked
6.
For )-lead arrangement and for single channel reproduction, a "double loop" pin
connection between the two ground terminals-simply slide the
loop ends onto the
made before connecting the regular terminal jacks to the cart-
ridge. Connect ground pin jack to either ground terminal.
Caution:
nals. Make all solder connections to terminals jacks provided.
Phasing:
loudspeakers. Although the phasing of the M7D cartridge is carefully controlled, there are many possibilities where a particular
High Fidelity system can cause opposite phasing of the sound from
the speakers. The phasing can be easily checked by
Play a monaural record using the M7D. Reverse the connections of
the wires to one speaker. This can be done at the amplifier or the
speaker, which ever is more convenient. When the sound appears
to come from
phase. Also, bass reproduction will be much fuller when speakers
are correctly phased.
Associated Equipment:
you require the same basic equipment that you need for playing
monaural records. However, you will need two amplifiers, two preamplifiers and two speakers. If you already have a monaural system
then you need to add the extra preamplifier, amplifier and speaker.
You will also have to add another set of leads to your tone arm
the second channel. Figure 3 shows typical stereo systems.
If your system is represented by Figure
the connection instructions that accompanied your stereo preamplifiers.
If
your stereo
your system is represented by Figure 3B or 3C, consider
The M7D cartridge utilizes a 4 terminal arrange-
each
RIGHT
I
1
1
\--
1
CHANNEL
TERMINAL
RIGHT
CHANNEL
GROUND
TERMINAL DIAGRAMS
FIGURE 2
iack is provided to make a convenient
ground terminals. This connection should be
Do not make solder connections to cartridge termi-
It
is important to maintain correct phasing of both
a
point halfway between the speakers, they are In
For reproducing stereo records
Tqiiir-1
POWER AMPLIFIER
COMBINATION COMBINATION
TYPICAL STEREO SYSTEMS
FIGURE 3
dynetic as two separate cartridges and connect one set
TERMINAL
L
and
shield lead
L
terminals and
a
listening test.
foi
POWER
AMPLIFIER
3A, merely follow
of leads (see connection) to each
leads should be connected to the "low level" phono input. Again,
the instructions that come with each of your
your best guide in making these connections.
After making the cartridge connection,
balance the system by adjusting the volume from each channel un-
til
they are equal. To do this, play a monaural record with the
stereo cartridge. Adjust the volume control on one preamplifier
until you have the sound as loud as you want
volume on the second preamplifier to match the first.
It
tion that you will have to
every time you wish to raise or lower the volume. There are several
rel'atively inexpensive devices available that provide a solution to
this problem by means of a master volume control that controls the
sound level from both channels with a single control. The dealer
from whom you purchased your stereo
you information on a device of this type.
will become apparent to you when performing the opera-
adiust the balance between the channels
Speaker Placement:
greatly from room to room,
rule of speaker location for optimum results. In general, the
speakers should be placed along the wall of the room opposite the
most comfortable listening area. The speakers should be separated
by a distance equal to
which they are placed.
farther apart the speakers are the farther away you will be able
to sit and still achieve good results. Also, the closer together the
speakers are, the closer you will have to sit and the fewer the
number of people that will be able to listen at one time. By ex-
perimenting with the placement of the speakers, you can determine
the location that will give the best results in your particular room.
1/2
to
It
is well to keep in mind the fact that the
preamplifier. In general, these
~ream~lifiers will be
it
is necessary to
it.
Then, adjust the
dynetic will be able to give
Because acoustic factors vary
it
is impossible to establish a definite
3/4
of the width of the wall against
SO
STYLUS REPLACEMENT
Special Note:
such an extremely important part in the overall performance of the Stereo Custom
certain that any replacement stylus is a genuine Shure
"Stereo Custom
The stylus assembly, when installed in the cartridge is practically immune to damage during normal usage. However, care
should be taken to avoid bending or distorting the stylus assembly
when
it
is installed or removed.
Stylus replacement is simple and fast. (See Figure 4). To
replace (Step I)-Insert fingernail file (or a similarly pointed
tool) behind the lip at the front of the stylus spade. Care must
be taken not to deform needle shank (portion holding needle tip)
or spring wire in the stylus assembly. (Step 2)-Withdraw stylus
by pulling forward out of cartridge-lift completely out with thumb
and forefinger. (Step 3)-grasp new stylus between thumb and
forefinger and insert into stylus socket cartridge. (Step 4)-Press
stylus into socket until
front portion of the spade. Care must be taken not to allow the
finger to slip off of the front of the stylus spade resulting in damage to the stylus tip.
Guarantee:
guaranteed to be free from electrical and mechanical defects for
one year from date of shipment from the factory, provided all
instructions are complied with fully. The Guarantee does not cover
needle wear nor does
abuse or mishandling.
Patent Notice:
3,077,521, 3,077,522. and
Because the stylus-magnet assembly plays
Dynetic Phono Reproducer, be
Dynetic" stylus.
NEEDLE REPLACEMENT
FIGURE
it
is firmly seated. Apply pressure at the
The M7D Stereo Custom Dynetic Cartridges are
it
cover damage to the needle assembly from
Manufactured under
D187.230. Other patents pending.
4
U.S.
patents 3,055,988,