This manual contains information on correct handling and operational procedures as well as safety
consideration pertinent to ACCUREF-K 9001.
Before carrying out measurement and/or adjustment, read the instructions thoroughly so that effective
operation is ensured. As this constitutes an important reference and user guide, keep it on hand at all times.
NOTE
The information contained in this manual is subject to change without notice.
While reasonable efforts have been made in the preparation of this document to ensure its accuracy,
you should contact your local distributor immediately, if any quarries arise due to editorial errors or
omissions etc.
If you find any imperfect collating or missing pages, contact your local distributor for replacement.
SAFETY CONSIDERATION
ACCUREF-K 9001 is a Class I, Type B medical instrument as well as LED Class 2 product.
This instrument complies with Medical Device Directive 93/42/EEC.
A great deal of consideration has gone into the design and manufacturing of this instrument with regard to
its operational ease, the patient's safety and well-being as well as to the reliability of the product.
For safer and more effective use, however, follow the points described in this manual.
This instrument is designed for professional use.
~ General Definitions of Safety Symbols ~
Personal injury or physical damage may occur when this warning
CAUTION
LED Class 2 Product
NOTE
is ignored.
General warning. Caution. Risk of danger.
Denotes general ban or prohibition.
General mandatory action.
Caution of invisible LED radiation.
Avoid exposure to beam.
Indicated inside of the device.
This device is an LED class 2 product.
Additional information which is important to the text or is
useful/convenient to know.
Symbol for “MANUFACTURER”.
EC
REP
Symbol for “AUTHORISED REPRESENTATIVE IN THE
EUROPEAN COMMUNITY”.
RB-400-B02Q1
CAUTION
Always take great care when operating ACCUREF-K. Malfunction or damage to the instrument could
occur.
Cut the power immediately if malfunction occurs during operation. Damage to the equipment or
personal injury will result. Consult your dealer, if repair work needs to be carried out.
At no time attempt to remodel or disassemble ACCUREF-K. Damage to the instrument or personal
injury will result.
As ACCUREF-K is a precision optical instrument, operations must be carried out at all times by
experienced, authorized personnel. Damage to the equipment or personal injury will result.
Electromagnetic wave generated by TV, radio, mobile phone, radio transceiver, etc. may cause
malfunction of this instrument. This instrument may also generate noise in the TV, radio, mobile
phone, radio transceiver, etc.
Avoid introducing or installing instruments which may have adverse influence to the circumstance.
Avoid use this instrument adjacent to other instruments or stacked in plies. Failure or malfunction of
the instrument may occur.
Avoid installation near TV or radio. The reception can be disturbed by electrical noise. Follow the
manual for the proper installation.
Never remove the plug from the outlet if your hands are wet. Electric shock or personal injury could
result.
Make sure the power cord is not damaged. Fire or electric shock may occur.
Do not touch the optical parts. Measurement accuracy will be adversely affected.
The power cord must be firmly connected to an electrical ground (safety ground) at the power outlet.
Personal injury may result from electric shock, etc.
If the instrument fails to work properly, you should not try to repair the fault. Consult your dealer
immediately.
The instruction in this manual ensures correct operations.
Observe the following environmental conditions for used and storage. Avoid dew condensation at all
time.
TemperatureRelative humidity
Use+10°C to +40°C30% to 85%
Storage–10°C to +60°Cbelow 70%
Avoid the following conditions for storage and use of the instrument.
Where noxious gases or air pollutants are present.
Where dust and grit may occur.
Where oil fumes or greasy substances are emitted.
Where there are atmospheric concentrations of salt.
Near gas generation areas and places where dust accumulates.
Keep in a secure, stable situation. Do not expose to strong vibrations (areas of seismic activity) and
sudden shocks (this includes transportation) etc.
Where there is an inclination of more than 10 degrees.
Where voltage from the power sources rises or falls sharply during loading.
Where fluctuations in the voltage of the power source occurs.
Direct contact with sunlight.
If the instructions above are not followed, damage to the equipment or personal injury will ensue.
RB-400-B02Q2
Guidance and manufacturer’s declaration – electromagnetic emissions
ACCUREF-K9001 is intended for use in the electromagnetic environment specified below. The customer or the
user of ACCUREF-K9001 should assure that it is used in such an environment.
Group 1ACCUREF-K9001 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class A
Class A
Complies
ACCUREF-K9001 is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Guidance and manufacturer’s declaration – electromagnetic immunity
ACCUREF-K9001 Image Intensifier is intended for use in the electromagnetic environment specified below.
The customer or the user of ACCUREF-K9001 Image Intensifier should assure that it is used in such an
environment.
Immunity testIEC 60601
test level
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for
input/output lines
1 kV differential
mode
2 kV common mode
<5 %
U
U
U
T
T
T
(>95 % dip in
for 0.5 cycle
40 %
(60 % dip in
for 5 cycles
70 %
(30 % dip in
for 25 cycles
U )
U )
T
U )
T
T
Compliance levelElectromagnetic environment –
guidance
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for
input/output lines
1 kV differential
mode
2 kV common mode
<5 %
U
U
U
T
U )
T
U )
T
U )
(>95 % dip in
for 0.5 cycle
40 %
(60 % dip in
for 5 cycles
70 %
(30 % dip in
for 25 cycles
T
T
T
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.Iftheuserof
ACCUREF-K9001 Image Intensifier
requires continued operation during
power mains interruptions, it is
recommended that ACCUREF-K9001
Image Intensifier be powered from an
uninterruptible power supply or a
battery.
<5 %
U
(>95 % dip in
for 5 s
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
Uis the a.c. mains voltage prior to application of the test level.
T
3 A/m3 A/mPower frequency magnetic fields
T
U )
T
<5 %
U
(>95 % dip in
for 5 s
T
U )
T
should be at levels characteristic of a
typicallocationinatypical
commercial or hospital environment.
RB-400-B02Q3
Guidance and manufacturer’s declaration – electromagnetic immunity
P
ACCUREF-K9001 is intended for use in the electromagnetic environment specified below. The customer or the
user of ACCUREF-K9001 should assure that it is used in such an environment.
Immunity testIEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any partof
ACCUREF-K9001, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Pd2.1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Pd2.180 MHz to 800 MHz
Pd3.2800 MHz to 2.5 GHz
where
is the maximum output power rating of
the transmitter in watts (W) according to the
transmittermanufactureranddisthe
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
a
b
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
ACCUREF-K9001 is used exceeds the applicable RF compliance level above, ACCUREF-K9001 should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating ACCUREF-K9001.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RB-400-B02Q4
Recommended distance between portable/mobile RF communication equipment and ACCUREF-K9001
P
ACCUREF-K9001 intends to be used in the radiated RF contolled environment. Customers or users of
ACCUREF-K9001 shall maintain the minimum safe distance between portable/mobile RF communication
equipment(transmitter) and ACCUREF-K9001 to prevent electromagnetic interference. The minimum distance
shall be accordance with the maximum output of the communication equipment as recommended below.
Separation distance according to frequency of transmitter
mRated maximum output
power of transmitter
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
W
Pd2.1
Pd2.1
Pd3.2
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
■ Use accessories specified by us to avoid any malfunction or failure.
■ Use of accessory (power cord) other than specified below may adversely affect
Extra care should be taken for storage of a model eye. Avoid where the lens of the
model eye may be damaged as well as any dusty or humid/steamy environments.
Avoid direct sunlight, humidity and high temperature when storing printer paper
which is a thermal paper.
NameModel No.Length
other instruments and/or cause malfunction of the instrument. Always use the
accessory specified by us.
RB-400-B02Q7
1. Parts Identification
ACCUREF
K
90
Fuse Holder
LCD Monitor
Power Indicator
IOL Switch
Measurement Mode
Selection Switch
Menu Switch
Print Switch
Anti-sliding lock
AUTO REFKERATOMETER
‐
SHIN-NIPPON
Printer
Body
Base
Head Rest
Eye Mark
Measurement
Start Switch
Power Switch
Power Plug Connector
View Window
Chin Rest
Joystick
Nameplate
External Interface Connector
Output Terminal for External Monitor
RB-400-B02Q8
2. Conveyance and Handling Procedure
be carried out.
When transporting the instrument, make sure that the body has been securely locked.
Center the body onto the base so that their edges are aligned. Push up the lock while turning right until
both body and base are firmly fixed to each other.
3. Installation Environment
1)Do not expose the instrument's view window directly
to the sunlight or bright lighting from other sources.
Great care should be taken and avoid that the
examinee is exposed to strong light or glare. The
pupil will contract too much for measurement to
2)Do not operate in places where either dust or rubbish
accumulates. Environment with extremes in heat and
humidity should also be avoided.
Always follow the environmental requirements below
for installation.
3)Keep away from inflammable or explosive gases as
well as storage areas housing medical supplies and chemicals.
4)Avoid installing where dew condensation may accumulate.
Also, avoid where the radical temperature changes may
occur.
5)Keep away from sites that may experience strong vibrations
or sudden shocks.
6)Malfunction is likely to occur if the instrument is improperly stabilized or accidentally overturns.
To prevent internal/external damage caused by sudden impact, set the instrument on a solid and
secure surface. Do not store in high, 'out of reach' places.
Du s t
Gases
RB-400-B02Q9
4. Safeguard Summary
1.ACCUREF-K is a precision optical instrument. Always handle with care and avoid dropping it
accidentally.
2.Ensure that the instrument is properly grounded when connected to the power source.
3.Do not touch the optical parts with fingers and be sure to avoid dust, as their measuring accuracy
could be adversely affected and incorrect values may result.
When dust or fingerprints appear on the optical part, use a soft cloth to wipe off
the build-up. In case that the build-up is hard to remove, absolute alcohol is
recommended. Take great care when cleaning these parts as they are particularly
sensitive and fragile.
4.If the surfaces of the measuring unit and main unit including the
control panel are dirty, gently wipe with a dry cloth. For hard to
Avoid using organic solvents that will
damage the water based paint finish of
the instrument.
Benzene
5.Clean the chin rest and head rest with the neutral cleanser. For disinfecting them, especially where the
examinee may contact, hydrogen peroxide (Oxydol) is recommended.
6.If the instrument is not used for any length of time, remove the
power cord from the outlet.
7.When not in use, protect the instrument with a supplied
dustproof cover.
8.When the instrument fails to function properly, neverattempt to perform internal service or adjustment. Contact
your nearest registered agent, distributor or retail outlet.
RB-400-B02Q10
5. Preparation
5.1 Setting
(1) Set a roll of printer paper in the printer. Refer to '11.1 Reloading Printer Paper' for the procedure.
(2) Set and fix the chin rest liners with the chin rest
liner pins on the chin rest. Refer to the figure on
the right.
For sanitary consideration, disposing
NOTE
a sheet of the used chin rest liners after
every measurement is recommended.
Always use the chin rest liners following above.
For sanitary consideration, disinfecting the chin rest with Oxydol is
recommended.
Chin Rest Liners
Chin Rest Liner Pins
Chin Rest
RB-400-B02Q11
5.2 Applying Power
(1) Confirm that the power is 'OFF' (○).
O
POWER
I
T 2A 250V
(2) Insert the power cord into the instrument's power plug
connector. Then insert the plug into a general purpose
outlet.
Always make sure that the cable is
grounded.
(3)Turn the power switch 'ON'(|).
O
POWER
I
T 2A 250V
RB-400-B02Q12