Shellab SMI39-2 User Manual

SHELDON MICROBIOLOGICAL INCUBATORS
MODELS: SMI31 SMI31-2
SMI39 SMI39-2
Previously designated as
RI28, RI28-2 RI40, RI40-2
11/2013
4861573
INSTALLATION AND OPERATIONAL MANUAL
Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113
EMAIL: tech@Shellab.com INTERNET: http://www.Shellab.com/~Shellab
1-800-322-4897 (503) 640-3000 FAX (503) 640-1366
2
These units are TUV CUE listed as General Incubators for professional, industrial or educational use where the preparation or testing of materials is done at approximately atmospheric pressure and no flammable, volatile or combustible materials are being heated. These units have been tested to the following requirements:
CAN/CSA C22.2 No. 61010-1:2012 CAN/CSA C22.2 No. 61010-2-010 + R:2009 UL 61010-1:2004 + R:2005-07 + R:2008-10 UL 61010A-2-010:2002 UL 61010-1:2012 EN 61010-1:2010 EN 61010-2-010:2003
IEC 61010-1:2010 IEC 61010-2-010:2003
3
TABLE OF CONTENTS
SECTION 1.0 RECEIVING AND INSPECTION SECTION 2.0 INSTALLATION SECTION 3.0 GRAPHIC SYMBOLS SECTION 4.0 CONTROL OVERVIEW SECTION 5.0 OPERATION SECTION 6.0 CHART RECORDER INSTALLATION SECTION 7.0 MAINTENANCE SECTION 8.0 TROUBLESHOOTING SECTION 9.0 PARTS LIST
UNIT SPECIFICATIONS SCHEMATICS
4
RECEIVING AND INSPECTION
Section
Your satisfaction and safety require a complete understanding of this unit. Read the instructions thoroughly and be sure all operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its intended application; any alterations or modifications will VOID your warranty.
1.1 Inspection: The carrier, when accepting shipment, also accepts responsibility
for safe delivery and is liable for loss or damage. On delivery, inspect for visible exterior damage, note and describe on the freight bill any damage found, and enter your claim on the form supplied by the carrier.
1.2 Inspect for concealed loss or damage on the unit itself, both interior and exterior. If necessary, the carrier will arrange for official inspection to substantiate your claim.
1.3 Return Shipment: Save the shipping crate until you are sure all is well. If for any reason you must return the unit, first contact your customer representative for authorization. Supply nameplate data, including model number and serial number. Please see the manual cover for information on where to contact customer service.
1.4 Accessories: Verify that all of the equipment indicated on the packing slip is included with the unit. Carefully check all packaging before discarding. These units are equipped with 6 shelves, 24 shelf clips and 4 leveling feet.
5
INSTALLATION
Section
Local city, county or other ordinances may govern the use of this equipment. If you have any questions about local requirements, please contact the appropriate local agency. Installation may be performed by the end user.
Under normal circumstances this unit is intended for use indoors, at room temperatures between 5 and 40C, at no greater than 80% Relative Humidity (at 25C) and with a supply voltage that does not vary by more than 10%. Customer service should be contacted for operating conditions outside of these limits.
2.1 Power Source: The electrical supply circuit to the incubator must conform to all national and
local electrical codes. Consult the incubator’s serial data plate for the voltage and ampere
requirements before making connection. VOLTAGE SHOULD NOT VARY MORE THAN 10% FROM THE SERIAL PLATE RATING. This unit is intended for 50/60 Hz application. A separate circuit is recommended to prevent possible loss of product due to overloading or failure of other equipment on the same circuit.
2.2 Location: When selecting a site for the incubator, consider all conditions which may affect performance, such as extreme heat from steam radiators, stoves, ovens autoclaves, etc. Avoid direct sun, fast-moving air currents, heating/cooling ducts, and high traffic areas. To ensure air circulation around the unit allow a minimum of 10cm between the unit and any walls or partitions which might obstruct free airflow.
2.3 Lifting / Handling: These units are heavy and care should be taken to use appropriate lifting devices that are sufficiently rated for these loads. Units should only be lifted from their bottom surfaces. Doors, handles and knobs are not adequate for lifting or stabilization. The unit should be completely restrained from tipping during lifting or transport. All moving parts, such as shelves and trays should be removed and doors need to be positively locked in the closed position during transfer to prevent shifting and damage.
2.4 Leveling: The unit must sit level and solidly. Leveling feet (supplied) are to be installed in the
2.5 Cleaning: The incubator was cleaned at the factory, but not sterilized. Remove all interior parts,
holes at the base of the incubator. Turn the leveling feet counterclockwise to raise level. If the unit must be moved, turn the leveling feet in all the way to prevent bending of the feet.
including shelves and shelf clips and clean all parts and the chamber with a disinfectant that is appropriate for your application. A thorough periodic cleaning is strongly recommended.
WARNING: Never clean the unit with alcohol or flammable cleaners with the unit connected to the electrical supply. Always disconnect the unit from the electrical service when cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply.
6
2.6 Place shelves into chamber at desired position. See Figure 1.
Figure 1
7
GRAPHIC SYMBOLS
Section
Your incubator is provided with a display of graphic symbols on the control panel which are designed to help identify the use and function of the adjustable components.
1. Indicates that you should consult your manual for further
description and discussion of a control or user item.
2. Indicates “Temperature”
3. Indicates “Overtemperature”
4. C Indicates “Degrees Centigrade”
5. Indicates “AC Power
6. Indicates “Manual Adjustment”
7. Indicates “Potential Shock Hazard” behind partition
8. Indicates “Earth Ground”
9. Indicates “Unit should be recycled” (Not disposed of in
land-fill)
8
CONTROL PANEL OVERVIEW
Section
4.1 Power Switch: The main power I/O (on/off) switch controls all power to the unit and must be in the I/ON position before any systems are operational.
4.2 Main Temperature Control: The Main Temperature Control consists of the digital display and UP/DOWN arrow pads for inputting set point temperatures and calibration.
4.3 HEATING Lamp: This green pilot lamp in ON when the unit is heating up to set point and is blinking when controlling temperature at set point.
4.4 Overtemperature Thermostat: Marked SET OVERTEMPERATURE, the Thermostat is a completely independent control that acts as an override in the event that the Main control fails in the ON position. The Thermostat will regulate chamber temperature at approximately 1C above the set point of the Main controller.
4.5 OVER TEMP Lamp: This red pilot lamp is On when the Overtemperature
4.6 Circuit Breaker: (Non-CE units) Located adjacent to the power cord, the circuit
4.7 Fuse: (CE units) Located adjacent to the power cord in place of the circuit
Thermostat has taken control of the unit. Under normal operating conditions this lamp should never be on.
breaker is an added measure of protection against power source variations that, if tripped, can be reset by pushing in the button once the reason for the interruption has been cleared.
breaker. The fuse is an added measure of protection against power source variations, and if blown must be replace once the source of the interruption has been cleared.
9
OPERATION
Section
5.1 Check power supply against unit data plate, they must match. Plug service cord into the grounded electrical outlet, and turn the unit on.
5.2 Turn the Overtemperature Thermostat to its maximum position, clockwise, using a coin or a flat edged tool.
5.3 There are four (4) electrical outlets inside the chamber for use with electrical equipment not exceeding 1 amp.
5.4 Operating Electrical Apparatus Inside Chamber: Place the shaker, roller, spinner, etc. inside the incubator chamber. Plug apparatus into 1 amp outlet at rear of chamber. Be sure apparatus draws 1 amp or less (the power switch turns power on & off to outlets inside chamber).
5.5 Set Main Temperature Controller: Enter desired set point temperature. To enter set point mode on the controller, press either the Up or Down arrow pad one time. The digital display will start to blink, going from bright to dim. While blinking, the digital display is showing the set point. To change the set point, use the Up and Down arrow pads. If the arrow pads are not pressed for five (5) seconds, the display will stop blinking and will read the chamber temperature. Note that the Overtemperature Thermostat should be turned to its maximum position, (clockwise) until the unit has stabilized at desired set point temperature. Allow the incubator at least 24 hours to stabilize.
5.6 Calibration: It is recommended that calibration is done once the unit is installed in its
5.7 Set Overtemperature Thermostat: As mentioned in step 5.2, the Overtemperature
working environment and has been stabile at set point for several hours. Place a certified reference thermometer in the chamber where it can be easily viewed through the window. Be certain the thermometer is not touching any shelving. Allow the temperature to stabilize again until the thermometer reads a constant value for one hour. Compare the digital display with the reference thermometer. If there is an unacceptable difference, put the display into calibration mode by pressing both the Up and Down arrow pads at the same time for five (5) seconds until the two outside decimal points begin to flash. While the decimal points are flashing the display can be calibrated by pressing the Up or Down arrow pads until the display reads the correct value. Allow the incubator temperature to stabilize again, and recalibrate if necessary.
Thermostat should be initially set to its maximum position, to allow the unit to stabilize. Once the incubator is stabile at the desired set point, turn the Thermostat counterclockwise until the OVER TEMP light turns on. Next, turn the Thermostat clockwise just until the light turns off. Then turn the Thermostat clockwise two of the smallest divisions on its scale past the point where the light went out. This will set the Overtemperature Thermostat at approximately 1C above Main temperature set point.
10
CHART RECORDER INSTALLATION
Section
Please note that the following information is a general guide for installation. Chart recorders are available from your dealer. Before attempting installation please read the instructions provided with your chart recorder thoroughly for specific installation instructions. Note: Unplug unit from the power supply before installing the Chart Recorder.
6.1 Remove cover for Chart Recorder, located on the right side of control panel.
6.2 Remove steel plate from the back of the Chart Recorder.
6.3 Open the front glass door on the Chart Recorder and loosen the two screws on
the face of the recorder. (Screws are on the top and bottom right of the face.)
6.4 Inside the unit, behind the cover plate, you will see two wires that say “Recorder Power”. Cut off the butt connectors and strip the wires about 3/8 inch.
6.5 Put the Recorder Power wires through the hole in back of your Chart Recorder and connect the Recorder power wires to the terminal inside.
6.6 Run the Chart Recorder probe through the probe hole and slide into probe clips provided inside the chamber of the unit.
6.7 Inside the Chart Recorder you will see a piece of masking tape and a metal clip; remove these; screw face bolts back in; screw in Chart Recorder; shut the glass door on the Chart Recorder; plug unit into power supply; then turn on the unit.
6.8 In moving or mounting a chart, the pen arm and stylus must be raised by
6.9 To remove and replace a chart, unscrew the hub knob and swing it 90
6.10 Now the pen arm may be gently released and the chart knob rotated
pressing on the tab at the uppermost end of the arm. Even if the recorder is not to be used, a fresh chart should be mounted or some sort of padding placed under the stylus to protect the sapphire tip.
clockwise, or to the 3 o’clock position. Press the upper tab on the pen arm; lift the chart forward off the hub and slide down and away. DO NOT release the pen arm, but install a new chart.
counterclockwise back to the hub. Align the proper chart time with the chart plate index.
11
MAINTENANCE
Section
NOTE: Disconnect the power cord from the power source before performing any service or maintenance on this unit.
7.1 Cleaning: Clean the incubator interior and remove and clean shelves on a
regular basis. Use a disinfectant that is suitable for your application. A thorough periodic cleaning is strongly recommended. Use care when cleaning the door gasket to prevent damage which could impair the positive door seal.
WARNING: Never clean the unit with alcohol or flammable cleaners with the unit connected to the electrical supply. Always disconnect the unit from the electrical service when cleaning and assure all volatile or flammable cleaners are evaporated and dry before reattaching the unit to the power supply.
7.2 Storage: To prepare the unit for storage, remove all shelves and shelf clips and disconnect the power supply. Be certain that the chamber is clean and dry and the door is positively locked in the closed position. See Section 2.3, Lifting/Handling, for proper transport procedures.
7.3 No maintenance is required on the electrical components. If electrical or temperature problems occur, please see the Troubleshooting guide, Section 8.0, before calling for service.
12
TROUBLESHOOTING
TEMPERATURE
Temperature too high-display and reference thermometer don’t match
1/ Controller set too high-see section 5.5. 2/ Controller failed on – call Customer Service. 3/ Wiring error – call Customer Service.
Display reads "HI" or "400"+
Probe is unplugged, is broken or wire to sensor is broken – trace wire from display to probe; move wire and watch display to see intermittent problems.
Chamber temperature spikes over set point and then settles to set point
Recalibrate – see section 5.6.
Temperature too low-display and reference thermometer don’t match
1/ High limit set too low - see section 5.7 2/ Controller set too low – see section 5.5. 3/ Unit not recovered from door opening – wait for display to stop changing. 4/ Unit not recovered from power failure or being turned off – incubators will need 24 hours to warm up and stabilize. 5/ Element failure – see if HEATING light is on; compare current draw to data plate. 6/ Controller failure – confirm with front panel lights that controller is calling for heat. 7/ Overtemperature failure – confirm with front panel lights that Thermostat is operating correctly. 8/ Wiring problem – check all functions and compare wiring to owner’s manual – especially around any areas recently worked on. 9/ Loose connection – call Customer Service.
Display reads "LO"
1/ If ambient temperature is lower than range of unit – compare set points and ambient temperature to rated specifications in section 9.0. 2/ Sensor is plugged in backwards – call Customer Service.
Unit will not heat over a temperature that is below set point
1/ Confirm that fan moving and that amperage and voltage match data plate – check fan motor motion by removing top panel of the unit. 2/ Confirm that set point is set high enough –turn Thermostat all the way clockwise and see if HEATING light or OVERTEMP light comes on.
Section
FOR PERSONAL SAFETY, ALWAYS DISCONNECT THE POWER BEFORE SERVICING.
Always make a visual inspection of the incubator and control panel when troubleshooting. Look for loose or disconnected wires that may be the source of trouble.
13
3/ Check connections to sensor. 4/ Check calibration – using independent thermometer, follow instructions in section 5.6.
Unit will not heat up at all
1/ Verify that controller is asking for heat by looking for HEATING light – if pilot light is not on continuously during initial start up, there is a problem with the controller. 2/ Check amperage – amperage should be virtually at maximum rated (data plate) amperage. 3/ Do all controller functions work? 4/ Is the Overtemperature set high enough? – for diagnostics, should be fully clockwise with the pilot light never on. 5/ Has the fuse/circuit breaker blown?
Indicated chamber temperature unstable
1/ ±0.1 may be normal. 2/ Is fan working? – remove top panel and verify movement of cooling fan. 3/ Is ambient radically changing – either door opening or room airflow from heaters or air conditioning? – stabilize ambient conditions. 4/ Sensor miss-located, damaged or wires may be damaged - check mounts for control and Overtemperature sensors, then trace wires or tubing between sensors and controls. 5/ Calibration sensitivity – call Customer Service. 6/ Overtemperature set too low – be sure that its setting is more than 5 degrees over desired set point; check if pilot light is on continuously; turn controller knob completely clockwise to see if problem solved then follow instructions in section 5.7 for correct setting. 7/ Electrical noise – remove nearby sources of RFI including motors, arcing relays or radio transmitters. 8/ Bad connection on temperature sensor or faulty sensor – check connectors for continuity and mechanical soundness while watching display for erratic behavior; check sensor and wiring for mechanical damage. 9/ Bad connections– check connectors for mechanical soundness and look for corrosion around terminals or signs of arcing or other visible deterioration.
Will not maintain set point
1/ Assure that set point is at least 5 degrees over ambient. 2/ See if ambient is fluctuating – check for adjacent open doors or HVAC duct openings, stabilize ambient conditions.
Display and reference thermometer don’t
match
1/ Calibration error – see section 5.6. 2/ Temperature sensor failure – evaluate if pilot light is operating correctly. 3/ Controller failure – evaluate if pilot light is operating correctly. 4/ Allow at least 24 hours to stabilize at set point temperature. 5/ Verify that reference thermometer is certified.
Can't adjust set points or calibration
1/ Turn entire unit off and on to reset. 2/ If repeatedly happens, call Customer Service.
Calibrated at one temperature, but not at another
This can be a normal condition when operating temperature varies widely. For maximum accuracy, calibration should be done at or as close to the set point temperature as possible.
MECHANICAL
Door not sealing
1/ Check physical condition of gasket.
14
2/ Assure that gasket is in original location. 3/ Adjust hinge blocks or twist the door. 4/ Confirm that unit has not been damaged and body is square.
Motor doesn't move
1/ If shaft spins freely: check connections to motor and check voltage to motor. 2/ If shaft rubs or is frozen, relieve binding and retest.
Motor makes noise
1) Make sure that the fan or blower wheel is not contacting its housing. Adjust the motor mounting bracket position to re-center the fan or blower wheel, if necessary.
2) Check the fan or blower wheel for damage or out of balance condition. Replace the fan or blower wheel if it is damaged or out of balance.
3) Turn the motor shaft to make sure that it spins freely. If it binds or the bearings make a rubbing or scrapping sound then replace the motor.
OTHER
Controller on at all times - "locked-up"
1/ Turn unit off and on to reset. 2/ If cannot change any condition on the front panel, call Customer Service.
Front panel displays are all off
Check for wire damage.
Unit or wall fuse/circuit breaker is blown
1/ Check wall power source. 2/ Compare current draw and compare to specs on data plate. 3/ See what other loads are on the wall circuit.
Unit will not turn on
1/ Check wall power source. 2/ Check fuse/circuit breaker on unit or in wall. 3/ See if unit is on, e.g., fan or heater, and just controller is off. 4/ Check all wiring connections, especially around the on/off switch.
Unit is smoking – Out of box
This is not an uncommon occurrence when first operating new units. Put unit under vent and run at full power for one hour. – smoking is normal during first cycle to temperature.
Contamination in chamber
1/ See cleaning procedure in section 2.5. 2/ Develop and follow standard operating procedure for specific application; include definition of cleaning technique and maintenance schedule.
If following these troubleshooting suggestions does not solve the problem call your customer service representative for assistance. Please see the manual cover for information on where to contact customer service.
15
PARTS LIST
Description
115V
220V
Blower Motor
4880548
4880548
Blower Wheel
2600535
2600535
Cooling Fan, Control Panel
4880564
4880563
Cord Set
1800510
1800500
EMI Filter
2800503
2800503
Fuse 10 Amp
N/A
3300516
Fuse 16 Amp
3300513
N/A
Heating Element, SMI31 SMI31-2 (RI28 RI28-2)
2350563
2350554
Heating Element, SMI39 SMI39-2 (RI40 RI40-2)
2350557
2350566
Overtemperature Thermostat
1750982
1750983
I/O (On/Off) Power Switch
7850579
7850579
Moisture Proof Plug
1650530
1650530
Main Control w/Sensing Probe
1750582
1750583
Outlet, Interior
1650531
1650531
Pilot Light, Green
4650554
4650554
Pilot Light, Red
4650553
4650553
Shelf, SMI31 SMI31-2 (RI28 RI28-2)
5400500
5400500
Shelf, SMI39 SMI39-2 (RI40 RI40-2)
5401146
5401146
Section
16
UNIT SPECIFICATIONS
Weight
Shipping
Net
SMI31 SMI31-2
(RI28 RI28-2)
610 lbs.
500 lbs.
SMI39 SMI39-2
(RI40 RI40-2)
730 lbs.
850 lbs.
Dimensions
Exterior WxDxH
Interior WxDxH
SMI31 SMI31-2
(RI28 RI28-2)
38.5x34.0x75.3
30.5x26x62
SMI39 SMI39-2
(RI40 RI40-2)
41x34x87
35x26x76.5
Capacity
SMI31 SMI31-2
(RI28 RI28-2)
28 Cubic Feet
SMI39 SMI39-2
(RI40 RI40-2)
40 Cubic Feet
Temperature
Range
Uniformity
Sensitivity
SMI31 SMI31-2
(RI28 RI28-2)
Amb. +8 to 70C
+ .5C @ 37
+ .05C
SMI39 SMI39-2
(RI40 RI40-2)
Amb. +8 to 70C
+ .5C @ 37
+ .05C
17
MAIN POWER SWITCH CHERRY ROUND
7850532
POWER LIGHT ~ 4650554 (G)
OTP LIGHT 4650553
(R)
HEAT LIGHT 4650554(G)
CHAMBER FAN ~ 4880548
½ TB1
L1
½ TB1
L2
1 2
4
OTP
1750862
CONVENIENCE
OUTLETS
6100525 x4
DESIGN = 750W / 120V / 19.2
BLK BLK
BLK WHT
WHT
BLK
BLK HT WHT HT
BLK
BLK
WHT
WHT
BLK
BLK BLK
BLK
BLK BLK
BLK
WHT
WHT
WHT
WHT
BLK WHT
CHART RECORDER
CONNECTIONS
RED BLUWHT
BLK
WHT
BLK
BLK
WHT
BLK BLK
0900503
0.47 F / 250 VAC
HOT
NEUTRAL
GROUND
LOAD
TEMPERATURE
CONTROL
1750982
2800503
IEC FUSED INLET
4200505
BLK
WHT
BLK WHT
WIRE DIAGRAM
SMI31 (RI28) 120V
(9851242)
18
2800503
MAIN POWER SWITCH CHERRY ROUND
7850532
POWER LIGHT ~ 4650554 (G)
OTP LIGHT 6540553
(R)
HEAT LIGHT 4650554 (G)
CHAMBER FAN ~ 4880548
½ TB1
L1
½ TB1
L2
IEC FUSED INLET
4200505
1 2
4
12
4
½ OTP
1750813
½ OTP
1750813
WHTRED
BLU & BLK
EURO OUTLETS
6100531 x4
DESIGN = 750W 83
0900503
0.47 F / 250 VAC
BLK
WHT
WHT
BLK
WHT
BLK BLK
BLK WHT
WHT
BLK
BLK HT WHT HT
WHT
BLK
BLK
BLK
WHT
BLK
BLK BLK
BLK
BLK BLK
BLK
WHT
WHT
WHT
WHT
BLK
WHT
CHART RECORDER
CONNECTIONS
HOT
NEUTRAL
GROUND
LOAD
TEMPERATURE
CONTROL
1750983
WIRE DIAGRAM
SMI31-2 (RI28-2) 230V
(9851243)
19
MAIN POWER SWITCH CHERRY ROUND
7850532
POWER LIGHT ~ 4650554 (G)
OTP LIGHT 4650553
(R)
HEAT LIGHT 4650554(G)
CHAMBER FAN ~ 4880548
½ TB1
L1
½ TB1
L2
1 2
4
OTP
1750862
CONVENIENCE
OUTLETS
6100525 x4
DESIGN = 1000W 12.4
BLK BLK
BLK WHT
WHT
BLK
BLK HT WHT HT
BLK
BLK
WHT
WHT
BLK
BLK BLK
BLK
BLK BLK
BLK
WHT
WHT
WHT
WHT
BLK WHT
CHART RECORDER
CONNECTIONS
RED BLUWHT
BLK
WHT
BLK
BLK
WHT
BLK BLK
0900503
0.47 F / 250 VAC
HOT
NEUTRAL
GROUND
LOAD
TEMPERATURE
CONTROL
1750982
2800503
IEC FUSED INLET
4200505
BLK
WHT
BLK
WHT
WIRE DIAGRAM
SMI39 (RI40) 120V
(9851244)
20
2800503
MAIN POWER SWITCH CHERRY ROUND
7850532
POWER LIGHT ~ 4650554 (G)
OTP LIGHT 6540553
(R)
HEAT LIGHT 4650554 (G)
CHAMBER FAN ~ 4880548
½ TB1
L1
½ TB1
L2
IEC FUSED INLET
4200505
1 2
4
12
4
½ OTP
1750813
½ OTP
1750813
WHTRED
BLU & BLK
EURO OUTLETS
6100532 x4
DESIGN = 1000W 58.7
0900503
0.47 F / 250 VAC
BLK
WHT
WHT
BLK
WHT
BLK BLK
BLK WHT
WHT
BLK
BLK HT WHT HT
WHT
BLK
BLK
BLK
WHT
BLK
BLK BLK
BLK
BLK BLK
BLK
WHT
WHT
WHT
WHT
BLK
WHT
CHART RECORDER
CONNECTIONS
HOT
NEUTRAL
GROUND
LOAD
TEMPERATURE
CONTROL
1750983
WIRE DIAGRAM
SMI39-2 (RI40-2) 230V
(9851245)
Loading...