Shellab Bactron OPERATION MANUAL

BACTRON ANAEROBIC
CHAMBERS
OPERATION MANUAL
MODELS: I, II & IV
Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113
1-800-322-4897 (503) 640-3000 FAX (503) 640-1366
INTERNET: http://www.Shellab.com/~Shellab
______________________________________________________________________________
Orientation Page 3 Component Drawings Page 4 - 6
1.0 Receiving and Inspection Page 6
2.0 Installation Page 6 - 7
3.0 Controls Overview Page 8 - 9
4.0 Hose and Electrical Connections Page 9 -10
5.0 Manometer Page 11
6.0 Catalyst Cartridge Page 11 - 12
7.0 Thermoelectric Condensate Controller Page 12
8.0 Incubator Temperature Controller/Incubator Safety Page 13
9.0 Incubator Calibration Page 13 - 14
10.0 Vacuum Gauge Adjustment Page 14
11.0 Pressure Adjustment Page 15
12.0 Testing The Chamber For Gas Leaks Page 15 - 16
13.0 Making The Chamber Anaerobic/Desired Environment Page 16 - 18
14.0 Chamber Operation Page 18 - 19
15.0 Entering The Chamber Page 19 - 20
16.0 Exiting The Chamber Page 20
17.0 Maintenance Page 21
18.0 Chamber QC Sheet Page 22
19.0 Accessories Page 23
20.0 Chamber System Drawings: A) Keeping The Chamber Anaerobic Page 24 B) The System Controller and Controls: View From Behind Panel Page 25 C) Solenoid Page 26 D) The System Controller Indicator Guide Page 27
21.0 Troubleshooting Section Page 28 - 36
22.0 Bactron Operation Tips Page 37 - 41
23.0 Wiring Diagrams Page 42 – 46
24.0 Parts List Page 47
25.0 Loop Sterilizer Page 48
TABLE OF CONTENTS
Rev 4/02
4861118M
Bactron Operations Manual 2
______________________________________________________________________________
IMPORTANT: READ THIS INSTRUCTION MANUAL IMMEDIATELY.
Your satisfaction and safety require a complete understanding of this unit, including its proper function and operational characteristics. Be sure operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its
intended application; any alterations or modifications will void your warranty.
Orientation
Figure 1a. Bactron II Chamber
Note: Bactron I Chamber is similar; it has three incubator compartments.
Figure 1. Bactron IV Chamber
Bactron Operations Manual 3
______________________________________________________________________________
Component Drawings
Figure 2. Sleeve/Cuff Assembly
Consists of a neoprene sleeve, a plastic cuff ring, O-Ring, and a soft rubber cuff. The assembly allows gloveless chamber operation.
Figure 3. Armport Doors
CAUTION: The Bactron chambers have an efficient method of sealing the chamber by utilizing the
Armport doors as shown in Figure 3. The tightening knob on the armport doors should be turned just until the knob begins to “grab”. DO NOT OVER TIGHTEN. Damage can occur if too much force is placed on the Armport door assembly. See Section 13.5
Doors are used to seal the armports while the chamber is not in use. Doors can be stored using the holders provided while the chamber is in use. Assembly consists of an armport door, tightening knob and locking bar. Doors are secured by tuning the locking bars horizontally to pass through the cutouts in chamber, then turning locking bars verticals and tightening knobs.
Bactron Operations Manual 4
______________________________________________________________________________
Figure 4. Foot Pedals
Utilize to exchange the sleeve atmosphere (room) with the desired chamber environment (gas mixture). See pages 16 and 17 for further details.
Figure 5. Catalyst Cartridge
Two catalyst containing Palladium catalyst pellets are supplied to remove trace amounts of oxygen. See page 11 and 12 for additional data.
Bactron Operations Manual 5
______________________________________________________________________________
Figure 6. Air Lock
Provides ability to take material into and out of the chamber. Air Lock can be operated either automatically or manually. The Air lock shelf (Models II and IV) allows for materials to be conveniently rolled from the Air Lock into the chamber.
1.0 RECEIVING AND INSPECTION
1.1 The carrier, when accepting shipment, also accepts responsibility for safe delivery
and is liable for loss or damage claims. On delivery, you must inspect for visible exterior damage. Note and describe on the freight bill any damage found and enter your claim on the form the carrier supplies.
1.2 Inspect for concealed loss or damage on the unit itself, both interior and exterior. If any, the carrier will arrange for official inspection to substantiate your claim. Save the shipping crate until you are sure the unit has been delivered in good condition.
1.3 If for any reason you must return the unit, contact your sales representative for authorization and supply nameplate data.
2.0 INSTALLATION
2.1 Local city, county, or other ordinances may govern the use of this equipment. If you
have any questions about local requirements, please contact the appropriate local agency.
2.2 Under normal circumstances these units are intended for use indoors, at room temperatures between 5° and 40°C, at no greater than 80% relative Humidity (at
Bactron Operations Manual 6
______________________________________________________________________________
25°C) and with a supply voltage that does not vary by more than 10%. These ovens should not be operated at an altitude exceeding 2000 meters. Installation category is CAT-II.
2.3 Pollution Degree 2. Customer service should be contacted for operating conditions outside of these limits. Installation may be performed by the end user. It is unnecessary for this unit to be installed by a technician.
2.4 Location: When selecting a site for the unit, consider conditions which may affect performance, such as heat from steam radiators, ovens, autoclaves, etc. Avoid direct sun, fast-moving air currents, heating/cooling ducts, and high traffic areas. To ensure air circulation around the unit, allow a minimum of 2” between chamber rear and sides and any walls, partitions, or obstructions to free airflow.
2.5 Power Source: The power source must match the voltage, cycle, phase and amperage requirements listed on the data plate. Plug the cord into a grounded outlet. VOLTAGE OF THE OUTLET SHOULD NOT VARY MORE THAN 10% FROM THE DATA PLATE RATING. A separate circuit is recommended to precluded loss of product due to overloading or circuit failure. Note that the electrical supply to the unit must conform to all national and local electrical codes.
2.6 Gas Source: Install the gas regulator(s) on the tank(s) of gas. Chain the gas tanks(s) to a secure position on the wall. Set the regulator(s) to 10 psi.
2.7 If you are using an anaerobic gas mixture (AMG), we recommend 5% (H2), 5% Carbon Dioxide (CO2) and 90% Nitrogen(N2).
2.8 If you are using a two gas anaerobic application, we recommend one tank of AMG gas and the second tank of 100% Nitrogen (N2).
Figure 7. Warning Label
2.9 Cleaning: The Bactron unit was cleaned at the factory, however, a general cleaning
is recommended. Use BENZALKONIUM CHLORIDE to clean your chamber. See page 23 for a list of cleaning products.
Bactron Operations Manual 7
______________________________________________________________________________
Figure 8. Control Panel
Bactron Operations Manual 8
______________________________________________________________________________
3.0 CONTROLS OVERVIEW
3.1 Power: The lighted power “I/O” switch controls all power to the Bactron Chamber. It
must be in the “I” position before any systems are operational, including control circuitry.
3.2 Digital Display: The actual incubator temperature will be digitally displayed to + .1°C.
3.3 Main Temperature Control: This control consists of the Up/Down buttons and the digital display.
3.4 Vacuum Gauge: Analog dial shows the vacuum level of the Air Lock in inches of mercury.
3.5 Incubator Safety: The Safety is an independent thermostat that monitors incubator temperature. It guards against any failure of the incubator controller which would allow temperature to rise past set point. If temperature rises to the Safety set point, the Safety will take over control of the heating element and an audible alarm will be activated. The Safety controller allows for continued use of the chamber until service can be arranged.
3.6 Air Lock Controls. Manual Cycle: The manual cycle switch controls the vacuum and the gassing process of the Air Lock. Push VAC to evacuate the Air Lock then GAS to fill the Air Lock with the desired gas mixture. This process should be done three times.
3.7 Air Lock Controls. Auto Cycle: The automatic cycle switch will evacuate the Air Lock and then fill the chamber with the desired gas mixture. The process will cycle three times and upon completion the ANAEROBIC light will activate. The cycle can be aborted by pressing the switch to the START/RESET and holding it down for five seconds.
3.8 Incubator Heating: Light indicates that the heating element is activated and that the incubator is heating.
3.9 Safety Activated: Light indicates that the Incubator Safety audible alarm is activated and controlling the incubator temperature.
3.10 Chamber Gassing: Light indicates that the chamber is demanding gas. The light activates when the operating the Air Lock and sleeves, in addition to maintaining positive pressure in the working chamber.
3.11 Air Lock Doors Sealed: Light indicates that both the Air Lock doors are closed.
Bactron Operations Manual 9
______________________________________________________________________________
3.12 Gas Activated: In the manual cycle the GAS light indicates that the Gas switch is activated. In the automatic cycle, third cycle, light indicates that the GAS is activated in the Air Lock.
3.13 Vacuum Activated: Light indicates that the VAC switch of the manual cycle Air Lock is activated. Light is also on during automatic vacuum cycles.
3.14 Auto Cycle Aerobic: Light indicates that the Auto cycle has not been activated. When the automatic cycle has been activated, the Auto Cycle Anaerobic lights will alternate.
3.15 Auto Cycle Anaerobic: Light indicates that the Auto Cycle has been activated and is complete. When the automatic cycle has been activated, the Auto Cycle Anaerobic lights will alternate.
Indicates “Earth Ground”
4.0 HOSE AND ELECTRICAL CONNECTIONS
4.1 Hose ports for the gas and vacuum supplies and foot pedal controls are provided on
the top back left side of the control module section. The Bactron models have five access ports. (Older models have four access ports.) Figure 9 shows a typical hose connection for a five port model.
Figure 9. Hose Connections
4.2 Install the hose from the gas regulator of the working gas, for example AMG, (GAS
1) tank to the port marked “GAS 1” on the chamber module.
Bactron Operations Manual 10
______________________________________________________________________________
4.3 Install the hose from the gas regulator of the purging gas, for example Nitrogen, (GAS 2) tank to the port marked “GAS 2” on the chamber module.
NOTE: If using only one gas supply, install the “T” tube/fitting assembly (provided in the accessory pack) between the “GAS 1 IN” port and the “GAS 2 IN” port on the control module. Then attach the hose from GAS 1 tank regulator, as in 4.2, to the unused brass “T” tube end. Failure to follow this procedure will render the purging cycles of a single gas system inoperative.
4.4 Connect the 3/8” hose from the vacuum pump or a 3/8” hose from a suitable in-
house vacuum system, to the port marked VAC IN. In-house vacuum systems must be capable of evacuating to a minimum of 25 inches of mercury.
4.5 Plug the vacuum pump power cord into the outlet receptacle at the top left of the control module. This outlet is marked “5 AMPS” (installations using in-house vacuum supplies will not follow this step).
NOTE: CONNECTING HOSES. the hose into the appropriate connector until the hose stops. Pull back slightly to ensure hose will stay in place. Reinsert if hose pulls out.
4.6 Connect the center “T” section from the foot pedal assembly to the tube fittings
located on the inner sides of the armport access openings on the front of the chamber.
4.7 Connect the tube from the GAS side of the foot pedal assembly to port marked GAS OUT on the control module section.
4.8 Connect the tube from the VAC side of foot pedal assembly to the port marked VAC OUT on the control module section.
4.9 Plug power cord from Chamber into a 115 VAC, 60 Hz (220 volt units are also
available) electrical outlet. Turn the power switch on to verify power to the system.
4.10 When the main power switch is turned on, the vacuum pump should come on for a few seconds and then go off. If the pump does not come on, or comes on and will not go off, consult the Troubleshooting Guide for adjustment.
4.11 When the power switch to the Chamber is turned on, you should see Power ON, CHAMBER GASSING, AIR LOCK DOORS SEALED and AUTO CYCLE AEROBIC
lights illuminate, along with a readout of the incubator temperature. If any of these appear not to work, consult the Troubleshooting Guide.
Hose connections are intended to be simple. Insert
Bactron Operations Manual 11
______________________________________________________________________________
5.0 MANOMETER
5.1 Located below the right end of the control panel and inside the chamber, the Manometer provides a visual way to measure pressure in the chamber. It also functions as a pressure release system to prevent damage to the chamber in the event over-pressurization occurs. By measuring the difference between the inner and outer water levels one can obtain the chamber pressure in inches of water. If pressure becomes too great, excess gas will push out through the inner tube and bubble through the water thus releasing pressure from the chamber.
5.2 Purpose: The function of the manometer is two-fold:
a. Provides visual way to measure the pressure in the chamber. As the
pressure inside the chamber increases, the water level in the inner tubes moves down a distance proportional to the pressure increase. By measuring the difference of the inner and outer water levels, the user can obtain the chamber pressure in inches of water. A water level indicator notifies user that water is high or low.
b. It acts as a pressure release system to prevent damage to the chamber in
case of inadvertent over-pressurization. If the pressure becomes too great, the excess gas simply bubbles through the water and out the back of the chamber.
5.3 Filling Procedure: Use a pipette or plastic wash bottle to fill the manometer with distilled water to a level of two or three inches. Lines on the manometer provide a guide. Fill water to the top line. This may be done from the top of the manometer inside the chamber.
6.0 CATALYST CARTRIDGE
6.1 The catalyst cartridge removes ant trace amounts of oxygen from the chamber. Two catalyst cartridges containing Palladium catalyst pellets are provided. See drawing pg. 4
6.2 One cartridge should always be in the chamber. The second spare cartridge is provided so that reactivation of the catalyst can be done after each chamber usage.
6.3 The catalyst cartridge is installed on the left wall inside the chamber. To install slide the cartridge into the metal tracks and secure the handle into the bracket.
6.4 Reactivate spare catalyst daily by heating for a minimum of 1 –1/2 to 2 hours at 160°C (320°F). You may keep the unused cartridge in the heating oven at 160° prior
Bactron Operations Manual 12
______________________________________________________________________________
to its use. (NOTE: Catalyst cartridge should be reactivated for the first time before installing into the unit.)
6.5 Quality control catalyst monthly by flowing anaerobic gas over the reactivated pellets. If the pellets become warm to hot, they are working properly and are ready for use in the chamber. If they do not heat up when the gas is applied, reheat to 180° to 200°C and flow gas over catalyst again to remove any buildup of sulfides that might hinder its effectiveness.
6.6 The entire catalyst cartridge can be placed in the heating oven for reactivating. However, if necessary, the handle on the cartridge can be removed.
6.7 A catalyst cartridge can also be placed inside the Air Lock to remove any trace amounts of oxygen.
7.0 THERMOELECTRIC CONDENSATE CONTROLER
7.1 The thermoelectric condensate controller collects excess humidity and eliminates
chamber condensation. Moisture is funneled into a tube for collection inside your chamber.
7.2 The thermoelectric condensate controller is located on the left side of the chamber interior behind the catalyst cartridge. A plastic tube drains excess moisture into the working chamber, on the back left side.
7.3 Place a container, e.g., glass flask or beaker, under the tube for collection. Empty the container DAILY.
7.4 Desiccants are not recommended in the chamber. Desiccants are drying agents that can regrettably pull moisture from your samples.
Bactron Operations Manual 13
______________________________________________________________________________
8.0 INCUBATOR TEMPERATURE CONTROLLER/INCUBATOR SAFETY
8.1 Setting Incubator Controller: The INCUBATOR TEMPERATURE controller
regulates the incubator temperature by use of Up/Down buttons. To enter set point mode on the control, press either the Up or Down button one time. The digital display will start to blink, going from bright to dim. While blinking, the Digital Display is showing the set point. To change the set point, use the Up and Down buttons. If the buttons are not pressed for five (5) seconds, the display will stop blinking and will read the temperature of the unit. Note that the INCUBATOR SAFETY should be turned to its maximum position, (clockwise) until the unit has stabilized at desired set point temperature. Allow the incubator at least 24 hours to stabilize. Then re­calibrate the digital display to your reference thermometer, follow the calibration instructions given in Section 9. This incubator was calibrated at the factory at 37°C.
8.2 Setting Incubator Safety: Set the control to the desires incubator temperature. Then set the INCUBATOR SAFETY to its maximum position. Allow 24 hours for stabilization before proceeding.
If, after 24 hours the temperature is not at desired level, adjust the INCUBATOR
TEMPERATURE up or down until the precise desired temperature is achieved. (See Section 9 for calibration procedures)
When stabilization at the desired temperature has been achieved, turn the
INCUBATOR SAFETY control knob counter-clockwise until the INCUBATOR SAFETY ACTIVATED light comes on and audible alarm is activated. Next, carefully turn the INCUBATOR SAFETY knob clockwise until the light is just off. The Safety is now set; the INCUBATOR SAFETY ACTIVATED light should remain OFF during normal operation.
NOTE: An accurate thermometer should be used inside the incubator
as a reference when setting and calibrating the incubator temperature.
8.3 It is a good idea to mark the INCUBATOR SAFETY label with the desired
setting or position as a backup in case the knob is moved accidentally. If the SAFETY ACTIVATED light is on and audio alarm is activated at any time, check the INCUBATOR TEMPERATURE controller setting to be sure that it is not set above the INCUBATOR SAFETY setting.
9.0 INCUBATOR CALIBRATION
9.1 Compare the reading of the digital temperature display to an accurate reference
thermometer. If there is a difference, put the display into calibrate mode by pressing both the Up and the Down buttons at the same time and holding them in for about five (5) seconds or until the two outside decimal points start to flash on and off.
Bactron Operations Manual 14
______________________________________________________________________________
When the decimal points are flashing, the display can be calibrated to match the
reference thermometer by pressing the Up or Down button until the display reads correctly.
10.0 VACUUM ADJUSTMENT
10.1 The vacuum gauge for the Automatic Air Lock Cycle is factory calibrated. It is
recommended that customers note the setting of the vacuum gauge during Auto Air Lock Cycle to ensure that the setting is correct. Evacuation and the vacuum pump capabilities can effect the calibration.
10.2 The vacuum setting for the Auto Air Lock Cycle are factory set at the High Limit 18 and the Low at 4.
10.3 Vacuum calibration procedure: High Limit
A) Activate the Auto Air Lock Cycle and note the high level reached by the
vacuum gauge. If the vacuum gauge is not at the desired High setting of 18, adjust the high vacuum switch located on the left outside if the control housing box.
B) This switch is illustrated in the “CONTROLS: View From Behind Control
Panel” Page 25, #10.
C) Adjust the High setting by turning the large knob until the desired value of
18 is reached. (Black notch on knob toward 0 to decrease or toward 30 to increase). See page 25.
D) Activate the Auto Air Lock Cycle and observe the vacuum gauge on the
high level reached.
E) Repeat steps C and D.
Vacuum calibration procedure: Low Limit
A) Activate the Auto Air Lock Cycle and note the low level reached by the
vacuum gauge. If the vacuum is not at the desired Low setting of 4, adjust the Low vacuum switch located in the same area as the high vacuum switch noted above.
B) Repeat the procedure noted above under High vacuum to adjust the low
Vacuum setting except adjust for the lower value of 4.
Bactron Operations Manual 15
Loading...
+ 34 hidden pages