Shellab Bactron OPERATION MANUAL

BACTRON ANAEROBIC
CHAMBERS
OPERATION MANUAL
MODELS: I, II & IV
Sheldon Manufacturing Inc. P.O. Box 627 Cornelius, Oregon 97113
1-800-322-4897 (503) 640-3000 FAX (503) 640-1366
INTERNET: http://www.Shellab.com/~Shellab
______________________________________________________________________________
Orientation Page 3 Component Drawings Page 4 - 6
1.0 Receiving and Inspection Page 6
2.0 Installation Page 6 - 7
3.0 Controls Overview Page 8 - 9
4.0 Hose and Electrical Connections Page 9 -10
5.0 Manometer Page 11
6.0 Catalyst Cartridge Page 11 - 12
7.0 Thermoelectric Condensate Controller Page 12
8.0 Incubator Temperature Controller/Incubator Safety Page 13
9.0 Incubator Calibration Page 13 - 14
10.0 Vacuum Gauge Adjustment Page 14
11.0 Pressure Adjustment Page 15
12.0 Testing The Chamber For Gas Leaks Page 15 - 16
13.0 Making The Chamber Anaerobic/Desired Environment Page 16 - 18
14.0 Chamber Operation Page 18 - 19
15.0 Entering The Chamber Page 19 - 20
16.0 Exiting The Chamber Page 20
17.0 Maintenance Page 21
18.0 Chamber QC Sheet Page 22
19.0 Accessories Page 23
20.0 Chamber System Drawings: A) Keeping The Chamber Anaerobic Page 24 B) The System Controller and Controls: View From Behind Panel Page 25 C) Solenoid Page 26 D) The System Controller Indicator Guide Page 27
21.0 Troubleshooting Section Page 28 - 36
22.0 Bactron Operation Tips Page 37 - 41
23.0 Wiring Diagrams Page 42 – 46
24.0 Parts List Page 47
25.0 Loop Sterilizer Page 48
TABLE OF CONTENTS
Rev 4/02
4861118M
Bactron Operations Manual 2
______________________________________________________________________________
IMPORTANT: READ THIS INSTRUCTION MANUAL IMMEDIATELY.
Your satisfaction and safety require a complete understanding of this unit, including its proper function and operational characteristics. Be sure operators are given adequate training before attempting to put the unit in service. NOTE: This equipment must be used only for its
intended application; any alterations or modifications will void your warranty.
Orientation
Figure 1a. Bactron II Chamber
Note: Bactron I Chamber is similar; it has three incubator compartments.
Figure 1. Bactron IV Chamber
Bactron Operations Manual 3
______________________________________________________________________________
Component Drawings
Figure 2. Sleeve/Cuff Assembly
Consists of a neoprene sleeve, a plastic cuff ring, O-Ring, and a soft rubber cuff. The assembly allows gloveless chamber operation.
Figure 3. Armport Doors
CAUTION: The Bactron chambers have an efficient method of sealing the chamber by utilizing the
Armport doors as shown in Figure 3. The tightening knob on the armport doors should be turned just until the knob begins to “grab”. DO NOT OVER TIGHTEN. Damage can occur if too much force is placed on the Armport door assembly. See Section 13.5
Doors are used to seal the armports while the chamber is not in use. Doors can be stored using the holders provided while the chamber is in use. Assembly consists of an armport door, tightening knob and locking bar. Doors are secured by tuning the locking bars horizontally to pass through the cutouts in chamber, then turning locking bars verticals and tightening knobs.
Bactron Operations Manual 4
______________________________________________________________________________
Figure 4. Foot Pedals
Utilize to exchange the sleeve atmosphere (room) with the desired chamber environment (gas mixture). See pages 16 and 17 for further details.
Figure 5. Catalyst Cartridge
Two catalyst containing Palladium catalyst pellets are supplied to remove trace amounts of oxygen. See page 11 and 12 for additional data.
Bactron Operations Manual 5
______________________________________________________________________________
Figure 6. Air Lock
Provides ability to take material into and out of the chamber. Air Lock can be operated either automatically or manually. The Air lock shelf (Models II and IV) allows for materials to be conveniently rolled from the Air Lock into the chamber.
1.0 RECEIVING AND INSPECTION
1.1 The carrier, when accepting shipment, also accepts responsibility for safe delivery
and is liable for loss or damage claims. On delivery, you must inspect for visible exterior damage. Note and describe on the freight bill any damage found and enter your claim on the form the carrier supplies.
1.2 Inspect for concealed loss or damage on the unit itself, both interior and exterior. If any, the carrier will arrange for official inspection to substantiate your claim. Save the shipping crate until you are sure the unit has been delivered in good condition.
1.3 If for any reason you must return the unit, contact your sales representative for authorization and supply nameplate data.
2.0 INSTALLATION
2.1 Local city, county, or other ordinances may govern the use of this equipment. If you
have any questions about local requirements, please contact the appropriate local agency.
2.2 Under normal circumstances these units are intended for use indoors, at room temperatures between 5° and 40°C, at no greater than 80% relative Humidity (at
Bactron Operations Manual 6
______________________________________________________________________________
25°C) and with a supply voltage that does not vary by more than 10%. These ovens should not be operated at an altitude exceeding 2000 meters. Installation category is CAT-II.
2.3 Pollution Degree 2. Customer service should be contacted for operating conditions outside of these limits. Installation may be performed by the end user. It is unnecessary for this unit to be installed by a technician.
2.4 Location: When selecting a site for the unit, consider conditions which may affect performance, such as heat from steam radiators, ovens, autoclaves, etc. Avoid direct sun, fast-moving air currents, heating/cooling ducts, and high traffic areas. To ensure air circulation around the unit, allow a minimum of 2” between chamber rear and sides and any walls, partitions, or obstructions to free airflow.
2.5 Power Source: The power source must match the voltage, cycle, phase and amperage requirements listed on the data plate. Plug the cord into a grounded outlet. VOLTAGE OF THE OUTLET SHOULD NOT VARY MORE THAN 10% FROM THE DATA PLATE RATING. A separate circuit is recommended to precluded loss of product due to overloading or circuit failure. Note that the electrical supply to the unit must conform to all national and local electrical codes.
2.6 Gas Source: Install the gas regulator(s) on the tank(s) of gas. Chain the gas tanks(s) to a secure position on the wall. Set the regulator(s) to 10 psi.
2.7 If you are using an anaerobic gas mixture (AMG), we recommend 5% (H2), 5% Carbon Dioxide (CO2) and 90% Nitrogen(N2).
2.8 If you are using a two gas anaerobic application, we recommend one tank of AMG gas and the second tank of 100% Nitrogen (N2).
Figure 7. Warning Label
2.9 Cleaning: The Bactron unit was cleaned at the factory, however, a general cleaning
is recommended. Use BENZALKONIUM CHLORIDE to clean your chamber. See page 23 for a list of cleaning products.
Bactron Operations Manual 7
______________________________________________________________________________
Figure 8. Control Panel
Bactron Operations Manual 8
______________________________________________________________________________
3.0 CONTROLS OVERVIEW
3.1 Power: The lighted power “I/O” switch controls all power to the Bactron Chamber. It
must be in the “I” position before any systems are operational, including control circuitry.
3.2 Digital Display: The actual incubator temperature will be digitally displayed to + .1°C.
3.3 Main Temperature Control: This control consists of the Up/Down buttons and the digital display.
3.4 Vacuum Gauge: Analog dial shows the vacuum level of the Air Lock in inches of mercury.
3.5 Incubator Safety: The Safety is an independent thermostat that monitors incubator temperature. It guards against any failure of the incubator controller which would allow temperature to rise past set point. If temperature rises to the Safety set point, the Safety will take over control of the heating element and an audible alarm will be activated. The Safety controller allows for continued use of the chamber until service can be arranged.
3.6 Air Lock Controls. Manual Cycle: The manual cycle switch controls the vacuum and the gassing process of the Air Lock. Push VAC to evacuate the Air Lock then GAS to fill the Air Lock with the desired gas mixture. This process should be done three times.
3.7 Air Lock Controls. Auto Cycle: The automatic cycle switch will evacuate the Air Lock and then fill the chamber with the desired gas mixture. The process will cycle three times and upon completion the ANAEROBIC light will activate. The cycle can be aborted by pressing the switch to the START/RESET and holding it down for five seconds.
3.8 Incubator Heating: Light indicates that the heating element is activated and that the incubator is heating.
3.9 Safety Activated: Light indicates that the Incubator Safety audible alarm is activated and controlling the incubator temperature.
3.10 Chamber Gassing: Light indicates that the chamber is demanding gas. The light activates when the operating the Air Lock and sleeves, in addition to maintaining positive pressure in the working chamber.
3.11 Air Lock Doors Sealed: Light indicates that both the Air Lock doors are closed.
Bactron Operations Manual 9
______________________________________________________________________________
3.12 Gas Activated: In the manual cycle the GAS light indicates that the Gas switch is activated. In the automatic cycle, third cycle, light indicates that the GAS is activated in the Air Lock.
3.13 Vacuum Activated: Light indicates that the VAC switch of the manual cycle Air Lock is activated. Light is also on during automatic vacuum cycles.
3.14 Auto Cycle Aerobic: Light indicates that the Auto cycle has not been activated. When the automatic cycle has been activated, the Auto Cycle Anaerobic lights will alternate.
3.15 Auto Cycle Anaerobic: Light indicates that the Auto Cycle has been activated and is complete. When the automatic cycle has been activated, the Auto Cycle Anaerobic lights will alternate.
Indicates “Earth Ground”
4.0 HOSE AND ELECTRICAL CONNECTIONS
4.1 Hose ports for the gas and vacuum supplies and foot pedal controls are provided on
the top back left side of the control module section. The Bactron models have five access ports. (Older models have four access ports.) Figure 9 shows a typical hose connection for a five port model.
Figure 9. Hose Connections
4.2 Install the hose from the gas regulator of the working gas, for example AMG, (GAS
1) tank to the port marked “GAS 1” on the chamber module.
Bactron Operations Manual 10
______________________________________________________________________________
4.3 Install the hose from the gas regulator of the purging gas, for example Nitrogen, (GAS 2) tank to the port marked “GAS 2” on the chamber module.
NOTE: If using only one gas supply, install the “T” tube/fitting assembly (provided in the accessory pack) between the “GAS 1 IN” port and the “GAS 2 IN” port on the control module. Then attach the hose from GAS 1 tank regulator, as in 4.2, to the unused brass “T” tube end. Failure to follow this procedure will render the purging cycles of a single gas system inoperative.
4.4 Connect the 3/8” hose from the vacuum pump or a 3/8” hose from a suitable in-
house vacuum system, to the port marked VAC IN. In-house vacuum systems must be capable of evacuating to a minimum of 25 inches of mercury.
4.5 Plug the vacuum pump power cord into the outlet receptacle at the top left of the control module. This outlet is marked “5 AMPS” (installations using in-house vacuum supplies will not follow this step).
NOTE: CONNECTING HOSES. the hose into the appropriate connector until the hose stops. Pull back slightly to ensure hose will stay in place. Reinsert if hose pulls out.
4.6 Connect the center “T” section from the foot pedal assembly to the tube fittings
located on the inner sides of the armport access openings on the front of the chamber.
4.7 Connect the tube from the GAS side of the foot pedal assembly to port marked GAS OUT on the control module section.
4.8 Connect the tube from the VAC side of foot pedal assembly to the port marked VAC OUT on the control module section.
4.9 Plug power cord from Chamber into a 115 VAC, 60 Hz (220 volt units are also
available) electrical outlet. Turn the power switch on to verify power to the system.
4.10 When the main power switch is turned on, the vacuum pump should come on for a few seconds and then go off. If the pump does not come on, or comes on and will not go off, consult the Troubleshooting Guide for adjustment.
4.11 When the power switch to the Chamber is turned on, you should see Power ON, CHAMBER GASSING, AIR LOCK DOORS SEALED and AUTO CYCLE AEROBIC
lights illuminate, along with a readout of the incubator temperature. If any of these appear not to work, consult the Troubleshooting Guide.
Hose connections are intended to be simple. Insert
Bactron Operations Manual 11
______________________________________________________________________________
5.0 MANOMETER
5.1 Located below the right end of the control panel and inside the chamber, the Manometer provides a visual way to measure pressure in the chamber. It also functions as a pressure release system to prevent damage to the chamber in the event over-pressurization occurs. By measuring the difference between the inner and outer water levels one can obtain the chamber pressure in inches of water. If pressure becomes too great, excess gas will push out through the inner tube and bubble through the water thus releasing pressure from the chamber.
5.2 Purpose: The function of the manometer is two-fold:
a. Provides visual way to measure the pressure in the chamber. As the
pressure inside the chamber increases, the water level in the inner tubes moves down a distance proportional to the pressure increase. By measuring the difference of the inner and outer water levels, the user can obtain the chamber pressure in inches of water. A water level indicator notifies user that water is high or low.
b. It acts as a pressure release system to prevent damage to the chamber in
case of inadvertent over-pressurization. If the pressure becomes too great, the excess gas simply bubbles through the water and out the back of the chamber.
5.3 Filling Procedure: Use a pipette or plastic wash bottle to fill the manometer with distilled water to a level of two or three inches. Lines on the manometer provide a guide. Fill water to the top line. This may be done from the top of the manometer inside the chamber.
6.0 CATALYST CARTRIDGE
6.1 The catalyst cartridge removes ant trace amounts of oxygen from the chamber. Two catalyst cartridges containing Palladium catalyst pellets are provided. See drawing pg. 4
6.2 One cartridge should always be in the chamber. The second spare cartridge is provided so that reactivation of the catalyst can be done after each chamber usage.
6.3 The catalyst cartridge is installed on the left wall inside the chamber. To install slide the cartridge into the metal tracks and secure the handle into the bracket.
6.4 Reactivate spare catalyst daily by heating for a minimum of 1 –1/2 to 2 hours at 160°C (320°F). You may keep the unused cartridge in the heating oven at 160° prior
Bactron Operations Manual 12
______________________________________________________________________________
to its use. (NOTE: Catalyst cartridge should be reactivated for the first time before installing into the unit.)
6.5 Quality control catalyst monthly by flowing anaerobic gas over the reactivated pellets. If the pellets become warm to hot, they are working properly and are ready for use in the chamber. If they do not heat up when the gas is applied, reheat to 180° to 200°C and flow gas over catalyst again to remove any buildup of sulfides that might hinder its effectiveness.
6.6 The entire catalyst cartridge can be placed in the heating oven for reactivating. However, if necessary, the handle on the cartridge can be removed.
6.7 A catalyst cartridge can also be placed inside the Air Lock to remove any trace amounts of oxygen.
7.0 THERMOELECTRIC CONDENSATE CONTROLER
7.1 The thermoelectric condensate controller collects excess humidity and eliminates
chamber condensation. Moisture is funneled into a tube for collection inside your chamber.
7.2 The thermoelectric condensate controller is located on the left side of the chamber interior behind the catalyst cartridge. A plastic tube drains excess moisture into the working chamber, on the back left side.
7.3 Place a container, e.g., glass flask or beaker, under the tube for collection. Empty the container DAILY.
7.4 Desiccants are not recommended in the chamber. Desiccants are drying agents that can regrettably pull moisture from your samples.
Bactron Operations Manual 13
______________________________________________________________________________
8.0 INCUBATOR TEMPERATURE CONTROLLER/INCUBATOR SAFETY
8.1 Setting Incubator Controller: The INCUBATOR TEMPERATURE controller
regulates the incubator temperature by use of Up/Down buttons. To enter set point mode on the control, press either the Up or Down button one time. The digital display will start to blink, going from bright to dim. While blinking, the Digital Display is showing the set point. To change the set point, use the Up and Down buttons. If the buttons are not pressed for five (5) seconds, the display will stop blinking and will read the temperature of the unit. Note that the INCUBATOR SAFETY should be turned to its maximum position, (clockwise) until the unit has stabilized at desired set point temperature. Allow the incubator at least 24 hours to stabilize. Then re­calibrate the digital display to your reference thermometer, follow the calibration instructions given in Section 9. This incubator was calibrated at the factory at 37°C.
8.2 Setting Incubator Safety: Set the control to the desires incubator temperature. Then set the INCUBATOR SAFETY to its maximum position. Allow 24 hours for stabilization before proceeding.
If, after 24 hours the temperature is not at desired level, adjust the INCUBATOR
TEMPERATURE up or down until the precise desired temperature is achieved. (See Section 9 for calibration procedures)
When stabilization at the desired temperature has been achieved, turn the
INCUBATOR SAFETY control knob counter-clockwise until the INCUBATOR SAFETY ACTIVATED light comes on and audible alarm is activated. Next, carefully turn the INCUBATOR SAFETY knob clockwise until the light is just off. The Safety is now set; the INCUBATOR SAFETY ACTIVATED light should remain OFF during normal operation.
NOTE: An accurate thermometer should be used inside the incubator
as a reference when setting and calibrating the incubator temperature.
8.3 It is a good idea to mark the INCUBATOR SAFETY label with the desired
setting or position as a backup in case the knob is moved accidentally. If the SAFETY ACTIVATED light is on and audio alarm is activated at any time, check the INCUBATOR TEMPERATURE controller setting to be sure that it is not set above the INCUBATOR SAFETY setting.
9.0 INCUBATOR CALIBRATION
9.1 Compare the reading of the digital temperature display to an accurate reference
thermometer. If there is a difference, put the display into calibrate mode by pressing both the Up and the Down buttons at the same time and holding them in for about five (5) seconds or until the two outside decimal points start to flash on and off.
Bactron Operations Manual 14
______________________________________________________________________________
When the decimal points are flashing, the display can be calibrated to match the
reference thermometer by pressing the Up or Down button until the display reads correctly.
10.0 VACUUM ADJUSTMENT
10.1 The vacuum gauge for the Automatic Air Lock Cycle is factory calibrated. It is
recommended that customers note the setting of the vacuum gauge during Auto Air Lock Cycle to ensure that the setting is correct. Evacuation and the vacuum pump capabilities can effect the calibration.
10.2 The vacuum setting for the Auto Air Lock Cycle are factory set at the High Limit 18 and the Low at 4.
10.3 Vacuum calibration procedure: High Limit
A) Activate the Auto Air Lock Cycle and note the high level reached by the
vacuum gauge. If the vacuum gauge is not at the desired High setting of 18, adjust the high vacuum switch located on the left outside if the control housing box.
B) This switch is illustrated in the “CONTROLS: View From Behind Control
Panel” Page 25, #10.
C) Adjust the High setting by turning the large knob until the desired value of
18 is reached. (Black notch on knob toward 0 to decrease or toward 30 to increase). See page 25.
D) Activate the Auto Air Lock Cycle and observe the vacuum gauge on the
high level reached.
E) Repeat steps C and D.
Vacuum calibration procedure: Low Limit
A) Activate the Auto Air Lock Cycle and note the low level reached by the
vacuum gauge. If the vacuum is not at the desired Low setting of 4, adjust the Low vacuum switch located in the same area as the high vacuum switch noted above.
B) Repeat the procedure noted above under High vacuum to adjust the low
Vacuum setting except adjust for the lower value of 4.
Bactron Operations Manual 15
______________________________________________________________________________
11.0 PRESSURE ADJUSTMENT
11.1 The chamber pressure of your unit is factory calibrated but local elevations may
make adjustment of this value necessary.
11.2 If the manometer is bubbling when no one is working inside the chamber or the water level of the manometer is less than ½ inches, the pressure should be adjusted. See manometer diagrams on page 11 for a reference. Following is the adjustment procedure:
A) The chamber pressure switch is located on the outer left side of the
control housing box. It is adjusted by means of an Allen wrench adjustment screw (clockwise to increase/counterclockwise to decrease) see page 25.
B) Adjust the setting of this valve only if the manometer level is above the
½ inch mark or is bubbling as noted above.
C) Repeat the adjustment, if necessary, observing carefully the result,
until the proper performance is obtained.
12.0 TESTING THE CHAMBER FOR GAS LEAKS
12.1 Each chamber is individually tested at the factory before shipping. The following
procedure, however, should still be followed to detect any possible damage due to shipping.
12.2 Make sure the catalyst is removed from the chamber during initial leak testing.
12.3 Check to ensure Manometer is filled.
12.4 Close the outer and inner air Lock doors.
12.5 Secure the Armport doors. Turn the locking bars horizontally to pass through the cutouts in the chamber, then turn bars so they are vertical, tighten knobs. Do not over tighten. See Note at the end of Section 13.0
12.6 Open the gas tank valve all the way on and set the gas regulator to 10 psi.
12.7 Turn the power switch on. As before, with the power switch on, you should see the POWER SWITCH, INCUBATOR TEMPERATURE display, CHAMBER GASSING, AIR LOCK DOORS SEALED, and AUTO CYCLE AEROBIC light illuminate.
12.8 The CHAMBER GASSING light will turn OFF when the water level in the manometer
reaches a level of approximately ½” of water. The AIR LOCK DOORS SEALED light will turn ON when the inner and outer Transfer Module doors are both closed. If
Bactron Operations Manual 16
______________________________________________________________________________
this light is not ON, either the doors are not closed properly or one or both of the door sensors are improperly adjusted (See Troubleshooting Guide).
12.9 If there are no leaks in the system, the CHAMBER GASSING light will remain OFF, and the water level in the manometer will remain stable at approximately ½” water. Normally, a “leak-tight” chamber will go from 15 to 30 minutes without the
CHAMBER GASSING light coming on again.
12.10 If the system has a gas leak, the CHAMBER GASSING light will flash ON and
OFF. Flashing every 10-30 seconds indicates a very large leak, whereas flashing every 10 minutes or so indicates a small leak. This is accompanied by a “clicking” sound (originating from the chamber gas solenoid). In this case, check the following:
A) Armport doors for adequate seal of O-rings. B) Air Lock outer/inner door gaskets for adequate contact with doors.
12.11 If the system still has a leak, refer to:
A) Chamber Maintenance: Leak Detection B) Trouble Shooting: Excessive Gas Consumption
13.0 MAKING THE CHAMBER ANAEROBIC/DESIRED ENVIROMENT
Prior To Purging The Chamber
13.1 Turn Power switch to the “O” (Off) position.
13.2 Place about six oxygen indicators inside the system. Place one unopened indicator
in the incubator, and place one unopened indicator in the work area. The others can be stores in the work are for future use.
13.3 Place catalyst in the chamber. For more information about the correct placement of the catalyst, see Section 6.0. Catalyst Cartridge page 11 and 12.
13.4 Open the incubator doors before leaving the chamber. This allows the incubator to become anaerobic also.
13.5 Put armport doors in place.
NOTE: The Bactron chambers have an efficient method for sealing the chamber by
utilizing the Armport doors as shown in Figure 3 on page 4 . The tightening knob on the Armport doors should be snug, DO NOT OVER TIGHTEN too much force is placed on the Armport door assembly. When the knob just begins to “grab”, the tightening bolt has reached its maximum point. Over tightening will cause the bolt to slip inside the handle and cause leakage.
. Damage can occur if
Bactron Operations Manual 17
______________________________________________________________________________
Purging the Chamber
13.6 Turn Power Switch to the “I” (On) position.
13.7 Note the reading on the gas regulator.
13.8 Loosen the right armport door to create a small leak. (Remove sleeves from the
armports if they have been placed there. This will make it easier to adjust the armport doors to create the proper leak.) Adjust the doors so that the CHAMBER GASSING light flashes ON and OFF for equal periods. It is important that the gas flows into the chamber in “pulses” rather than as a “stream”. If the gas “streams”, it tends to go to the open door only.
13.9 Bleed the gas mixture through the chamber until the reading on the tank regulator gauge is 300-400 psi less than the original reading. Normally, it takes 1-1/2 to 2 hours to flush out all atmosphere from the chamber.
13.10 After the gas purge, tighten the armport doors to prevent leaking. The CHAMBER
GASSING light will turn off, flow of gas into the chamber will stop, and the water
level difference in the manometer should again be approximately ½” and should remain stable.
13.11 At this point it is important to check if the chamber is completely anaerobic before
using it. Check this by first replacing the sleeves on the armports, and the entering the chamber. For more information about entering the chamber see Section 15.0, Entering the Chamber page 19.
13.12 A) Once inside the chamber, open one indicator strip in the incubator, and one
strip in the work area.
NOTE: Please note and follow the directions for use for the type of indicator
strip you are using.
B) If the strips indicate that the atmosphere is anaerobic, the chamber is ready
for use.
C) If the strips indicate an aerobic or slightly aerobic atmosphere, then exit the
chamber. Create a small leak as detailed in Step 13.8. Flush an additional 100-200 psi through the chamber. After this step, proceed to step 13.11, until indicator strips indicate an anaerobic atmosphere.
D) The gas flowing through the chamber may cause previous indicator strips to
dry out. Use fresh indicator strips stored in step 13.2 each time you test the atmosphere.
Bactron Operations Manual 18
______________________________________________________________________________
NOTE: During the initial purge you may notice heavy condensation within the chamber. This is due to catalytic formation of water from hydrogen gas and oxygen initially in the chamber. Most of this condensation will dissipate by the end of the gas purge cycle.
CAUTION: The catalyst cartridge will be HOT at this point. If you must touch it for any reason, use extreme care.
Summary
The chamber is ready when:
A) The CHAMBER GASSING light remains off indicating no gas leaks. B) The incubator doors are OPEN. C) Oxygen indicators are in the chamber. D) The catalyst cartridge is in place.
14.0 CHAMBER OPERATION
Air Lock Operation
14.1 Load the Air Lock with samples, materials, supplies, etc.
14.2 Close the outer and inner Air Lock Doors.
14.3 Activate the Air Lock by one of two methods:
A) Manual Cycle (Manual Gas/Vacuum Switch):
i. Locate the 3 position Manual Gas/Vacuum Cycling Switch which should be in
the middle.
ii. Turn the Manual GAS/VAC switch to VAC until a vacuum of 17 to 20-in. Hg is
reached, as indicated by the VACUUM GAUGE.
iii. Flip the GAS/VAC switch to GAS to refill the Air Lock with the gas mixture
until the AIR LOCK VACUUM GAUGE reads approximately 4-in. Hg on the first two cycles. On the third and last cycle leave the gas on until the gauge reaches zero to ensure that there is no vacuum still present in the Air Lock, then return the switch to the middle or off position.
iv. It is essential to perform the cycle at least three times. Once this is done, you
may enter the chamber through the armports, using the procedure specified in the next section, and take your materials into the chamber through the inner Air Lock door.
Bactron Operations Manual 19
______________________________________________________________________________
NOTE: When cycling the Air Lock manually, the AUTO CYCLE ANAEROBIC light on the
control panel will not illuminate when the cycle is completed, as it does at the end of the automatic cycle.
B) Automatic Cycle:
i. Make sure both Air Lock doors are closed. The automatic cycle will not start
otherwise.
ii. Press the “AUTO CYCLE START/RESET” button to initiate the cycle. The
AUTO CYCLE AEROBIC/AUTO CYCLE ANAEROBIC lights will alternately
flash on and off, indicating the cycle is in progress. When the cycle is complete, the system will stop automatically with the AUTO CYCLE
ANAEROBIC light on.
iii. If the cycle does not follow the sequences in the previous section on manual
operation (eg: cycle won’t start, cycle does not go through three times, cycle cot within specified limits, door won’t open upon completion, etc.), consult the Troubleshooting Guide for adjustments.
NOTE: Do not hold Auto Cycle Start Switch down. Simply Press the button and let go to activate.
iv. If you WANT TO abort the cycle, press and hold the “AUTO CYCLE
START/RESET” switch for 5 seconds. The vacuum gauge will go to 0 and
stop. Then the AUTO CYCLE AEROBIC/AUTO CYCLE ANAEROBIC light will stop flashing.
15.0 ENTERING THE CHAMBER
15.1 The rubber cuffs of the sleeve should be secured around your arm as opposed to
clothing. Remove watches, bracelets, etc., as they may damage the cuffs.
15.2 Insert your hands and forearms into the sleeves. The cuffs must be firmly secured around your bare forearms. Keep your hands four to six inches away from the armport doors.
15.3 Depress the VAC foot pedal (this procedure evacuated both sleeves), and continue depressing until the sleeves have collapsed completely and firmly around your forearms.
15.4 Flush the sleeves with the gas mixture by depressing the GAS foot pedal. Do not overfill the sleeves. Stop when you feel a small separation of space between your hands and the sleeves.
Bactron Operations Manual 20
______________________________________________________________________________
15.5 Repeat steps 15.3 – 15.4 three more times to ensure that the sleeves are completely anaerobic/desired atmosphere before entering the chamber.
NOTE: ALWAYS enter the chamber with both arms simultaneously.
15.6 Loosen the armport door knobs a few turns and rotate the locking bars to a horizontal position. The armport doors can now be pushed forward into the chamber interior and can be secured on the armport door holders.
NOTE: Arm movements that are to forceful may cause the sleeves to dislodge from the armports.
16.0 EXITING THE CHAMBER
16.1 Before leaving the chamber the incubator doors and the inner Air Lock door must be
closed.
16.2 Depress the GAS foot pedal to allow the sleeves to fill with gas. This procedure keeps the sleeves from collapsing as you move outward, making it easier to reposition the doors.
16.3 Hold the locking bars in a horizontal position, and close the doors. Rotate the locking bars to a vertical position, and gently tighten both armport door knobs, just until they “grab”.
16.4 Before removing arms from the sleeves:
A) Check to be sure that armport doors are securely sealed by slowly pushing
both arms forward (this action generates a pressure within the sleeves).
B) If the Manometer bubbles when the arms are pushed forward, this indicates
that the doors are improperly sealed. Loosen the armport door knobs again, remove the doors. Check the O-rings for any particulate matter on the sealing areas, and repeat Step “A”. If Manometer remains stable, the doors are sealed properly.
16.5 When the armport doors are properly sealed, slowly and carefully withdraw both arms from the sleeves.
Bactron Operations Manual 21
______________________________________________________________________________
17.0 MAINTENANCE
DAILY/ROUTINELY:
1. Before using any cleaning or decontamination method except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment.
2. Exchange chamber catalyst cartridge with reactivated one.
3. Remove & empty condensate collection container.
4. Change oxygen indicators.
5. Disinfect the chamber interior.
6. Check the cuffs on the sleeve system. If the cuffs have holes or any tears, replace.
7. Check incubator temperature. See Example Checklist page 22.
8. Record gas tank reading.
9. Change/Rejuvenate anatox.
Anatox absorbs volatile fatty acids and hydrogen sulfide in the chamber. Recommended use is 250 grams placed in a beaker in the chamber.
Anatox-1000 grams (packaged in 250 gram packages)
Instructions:
A. Empty two 250gram packages into two separate 500 ml Pyrex beakers.
B. Day one – place one beaker in the chamber (new anatox can be placed directly into
the chamber).
C. Day two – remove beaker in the chamber and replace with the second beaker of
anatox (also new).
D. Day three – reactivate the first beaker of anatox by heating at 160°C for 2 hours.
Then replace the beaker in the chamber with this reactivated beaker.
E. Day four – six months – continue switching the reactivated beakers daily until six
months is complete, then discard the old anatox and repeat the procedure again.
Bactron Operations Manual 22
______________________________________________________________________________
18.0 CHECKLIST
Month
Record Change Change Drain Record Temperature Catalyst Indicator Condensate Tank Pressure
Gauge Reading
Date
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
CHAMBER QC SHEET
Bactron Operations Manual 23
______________________________________________________________________________
19.0 ACCESSORIES
Tank Switch Chamber Stand
Switches and monitors dual gas tank supply. Provides work bench to support Bactron Provides automatic switch-over to second gas chambers. Chamber Stand height 30”. tank alarm sounds when both tanks are empty. Stands feature locking casters and storage Tanks can be switched manually at any time. Space.
Model # 2002-B Model: Bactron I, II, and IV
Nitrogen Gas Regulator Microscope Recommendation
Optional accessory used in a two gas chamber The LEICA brand of microscope is a
commonly used unit with Bactron System.
Model # 7150500 Microscope: # 9990516
Illuminator: # 4650502 Extended Service Contracts
A LUKAS illumination system is also The contract offers a twelve month warranty on available, which includes a 250 watt parts and labor. Ensures customer of product fiber optic light, light ring, 250 watt satisfaction and optimal performance. Lamp house, and 0.3x lens.
Cleaner Illumination system: #4650503 Spare Lamp: # 4650504
Benzalkonium Chloride
#AS-972V recommended to use in Manometer. Parts From Anaerobe Systems Co. #408/782-7557 Catalog No.
Anatox 9990502 Catalyst Holder Filled
#B90010 From Anaerobe Systems Co. 5110730 Petri-Plate Rack 2x13 #408/782-7557 9490513 Light Fixture
9990514 Sleeves, Complete Plexiglass Cleaner 9490507 UV Light Source
9490512 Start Up Kit-MDL I
8oz size; # 1060503 9490511 Start Up Kit-MDL II,IV,X 3600500 Rubber Cuffs, Sm Size 7
3600501 Rubber Cuffs, Med Size 8
Plexiglass Scratch Remover 3600502 Rubber Cuffs, Lrg Size 9
6000504 O-Ring For Cuff
6000501 O-Ring For Door
8oz size: # 1060504 3450506 Gasket-Airlock, Bactron I
3450507 Gasket-Airlock, Bactron
II,IV,X
Product
Bactron Operations Manual 24
______________________________________________________________________________
20.0 CHAMBER SYSTEM DRAWINGS
Gas
Anaerobic mixture is 5% hydrogen, 5% carbon Dioxide, and 90% nitrogen. Do not exceed 5% hydrogen for safety reasons. Test for the presence of hydrogen in the gas mixture by flowing the gas over the active catalyst. The catalyst will get hot if hydrogen is present.
Catalyst
Reactive by heating for 2 hours at 160° C. Test by flowing the anaerobic gas mixture over
the catalyst. If the catalyst is active, it will get hot. The catalyst can be “cleaned” by heating to 200°C, then flowing the anaerobic gas mixture over the catalyst.
Positive Pressure
The Manometer will indicate if positive pressure is present in the chamber.
Bactron Operations Manual 25
______________________________________________________________________________
System Controller And Controls: View From Behind Panel
Bactron Operations Manual 26
______________________________________________________________________________
Solenoid
J:\0100000\100016EX.PLT
Bactron Operations Manual 27
______________________________________________________________________________
System Controller Indicator Guide
Function At Rest Start Vacuuming High Gassing Low Completed Vacuum Vacuum No action. Start Vacuum on. Switch Gas on. Switch Airlock Both doors button Vacuum Point Vacuum Point anaerobic closed. pressed.* Going up. Reached going down Reached Airlock Lights Lights Lights aerobic. flashing Flashing flashing
Inputs
x0 Start Off On* Off Off Off Off Off x1 Outer Door On On On On On On On x2 Inner Door On On On On On On On x3 High Vacuum On On On Off** On On On x4 Low Vacuum On On to Off Off Off Off On On
Output
y0 Vacuum Solenoid Off On On On to Off Off Off to On Off y1 Gas (AMG) Solenoid Off Off Off Off to On
(1) y2 Anaerobic Light Off Flashing Flashing Flashing Flashing Flashing On y3 Aerobic Light On Flashing Flashing Flashing Flashing Flashing Off
y4 Door Light On On On On On On On y5 Nitrogen** Off Off Off Off to On
(2)
*Input 0 is only when the start button is pushed ** Input 3 comes on only for a short moment (1) Gas comes on for the last cycle only. (2) Nitrogen comes on for the first two cycles.
On (1)
On (2)
On to Off
(1)
On to Off
(2)
Off
Off
Bactron Operations Manual 28
______________________________________________________________________________
CAUTION: Extreme caution must be exercised any time access is made into areas housing electrical components. Repair, replacement or adjustment of components in these areas must only be done by qualified technicians familiar with electrical circuitry and the operation of the anaerobic chamber.
ATTENTION: Des Précautions extrèmes sont requises à chaque fois que vous accédez à des endroits qui abritent les composantes électriques. La réparation, le remplacement oi ìajustement de composanted dans ces endroits ne doivent être effectués que par un technicien qualifié familier avec les circuits électriques et ì opération des étuves anaérobiques.
21.0 Troubleshooting
General Troubleshooting Guide
I. GAS CONSUMPTION
Problem Possible Cause Solution
CHAMBER GASSING A. Armport door not sealing.
light comes on every
five (5) minutes or less
B. Leak in seams. panel with leak detector.
C. Manometer leak, indicated by very large leak out manometer hole. ( Confirm with soap solution)
D. Oxygen present-hydrogen being consumed by
combing with oxygen.
CHAMBER GASSING A. Inner door gasket not light comes on more sealing-check with leak often with the outer door detector. open than with it closed
A. Tighten knob until it "grabs", Check for worn O-rings.
B. Check front plexiglass
Call for service.
C. Seal inner tube to block, and block to mounting plate with solvent cement.
D. No action necessary.
A. Check for worn gasket.
Bactron Operations Manual 29
______________________________________________________________________________
General Troubleshooting Guide Continued
I. Gas Consumption
Problem Possible Cause Solution
Chamber pressure A. Chamber pressure A. Tap switch to jar Increase, Manometer sensor switch stuck in contacts loose, then Bubbles, CHAMBER "on" position. replace switch. GASSING light on.
B. Pressure set too B. Turn adjustment high. Adjust pressure screw counter­ sensor. See diagram clockwise to page 25. decrease pressure.
Excessive gas usage - A. Foot pedal leaking. A. Disassemble and clean no chamber leaks GAS valve or replace.
B. Gas regulator/gas B. Check for and repair input connections any leaks. Check leaking. regulator for 10 psi
C. Leak in supply line. C. Check all gas lines
Chamber pressure increases, A. Chamber gassing A. Disassemble and clean Manometer bubbles. Solenoid leaking, inlet or replace solenoid valve CHAMBER GASSING gas pressure normal. if necessary. See light off. Pinch hose to chamber Diagram pg. 26.
to see that bubbling
stops.
B. Chamber gassing B. Set inlet gas pressure solenoid leaking, inlet to 10 psi. gas pressure too high.
C. Chamber gassing C. See Figure 9, pg. 9. solenoid hosed incorrectly.
D. Air Lock gas solenoid D. Disassemble and clean leaks. Pinch hose to or replace solenoid module, see if it stops. valve. See pg. 26.
E. Incubator temperature E. No action necessary. increase on startup.
maximum output.
for leaks.
Bactron Operations Manual 30
______________________________________________________________________________
General Troubleshooting Guide Continued
Problem Possible Cause Solution
Indicator strips are A. Low chamber pressure. A. Adjust chamber pressure Activated , I.e. - turn color. sensor by turning
clockwise to increase. See diagram pg. 25.
B. Catalyst inactive. B1. Use freshly heated catalyst, flush chamber with approx. 50 to 100 psig of gas depending on how long chamber has been aerobic and size of chamber.
B2. Check to see that gas makes catalyst hot. If not, heat to 200 C and blow gas mix through to remove sulfide layer.
C. Circulating blower not C. Replace fan. functioning.
D. Air lock not evacuating D. Turn High side switch to high enough vacuum clockwise to increase. ( 17 - 18 inches Hg ). See diagram on page 25.
E. Outer Air Lock door not E. Check for damaged sealing properly. gasket - replace if necessary.
F. User did not perform F. Check technique of all sleeve technique chamber users. properly upon entering chamber.
G. Gas tank mixture G1. Check for very low contains an insufficient or empty tank. concentration of Hydrogen. G2. Check label for at least 5% Hydrogen.
II. AEROBIC CHAMBER CONDITIONS
Bactron Operations Manual 31
______________________________________________________________________________
General Troubleshooting Guide Continued
Problem Possible Cause Solution
Air Lock does not hold A. Split gasket inner & A. Check gasket. If split, Vacuum. outer door. replace.
B. Inlet fitting to module B. Tighten/reseal with not sealed. teflon tape if necessary.
C. Air Lock solenoids hosed C. See diagram pg. 26. incorrectly.
D. Vacuum solenoid leaking D. Disassemble and clean Pinch hose to module solenoid - replace if and see if leak stops. necessary. WD40 can be used to clean solenoid.
E. Defective vacuum gauge. E. Replace gauge.
Automatic cycle will not start. A. Input/output terminal A. Press down on the
buss is not seated terminal buss so that it correctly. snaps into place. NOTE: Power should be turned off.
B. User pushed start switch B. Push start switch once & held inadvertently held then release. See pg. 19. down. Thus aborting Auto Cycle.
C. Make sure both doors C. Adjust location of the are closed. Door sensors door sensor so that not making contact-check DOORS light is on that DOORS light is ON. when both doors are If not, check programmable closed. Controller input display IN 0001 (outer door) and IN 0002 (inner door) should be on.
III. AIR LOCK
Bactron Operations Manual 32
______________________________________________________________________________
General Troubleshooting Guide Continued
Problem Possible Cause
Automatic cycle will not D. Start button inoperative - D. Replace start button. start. (Continued) check input display - IN
0000 should be on when start button is pushed, OFF when not.
E. No power to controller - E1. Check for 24 volt DC check power light on out of power supply. Controller. E2. Check for continuity between power supply and controller.
Chamber pressure goes A. Inner door gasket not A. Replace door gasket down ( Manometer level rises) Making proper seal. if necessary when cycle is started.
Cycle only goes through A. System Controller A. Replace System two cycles. Malfunction. Controller.
Switch will not go from A. Insufficient vacuum A. Adjust switch on VAC Vacuum to Gas. supply. pump to allow higher
vacuum before shut off, around 20 inches Hg.
B. High vacuum set point B. Turn High side switch too high, it should be counterclockwise to around 17 - 18 inches Hg. decrease. See diagram pg. 25.
Doors difficult to open A. Low vacuum set point A. Turn Low side switch upon completion of cycle. too high. Counter clockwise to
Decrease. See diagram pg. 25.
III. AIR LOCK
Solution
Bactron Operations Manual 33
______________________________________________________________________________
General Troubleshooting Guide Continued
IV. INCUBATOR
Problem Possible Cause Solution
Steady digital readout A. Temperature Controller A. Turn to pg. 13 to follow Matches set point, but out of calibration. Complete Temperature actual temperature does not. Controller instruction.
B. Incubator door open. B. Close incubator door and allow incubator to stabilize.
Digital readout and actual A. Incubator fan not A. Replace fan, Model Temperature continue to functioning. Model Bactron IV. Increase past set point. Bactron IV, See drawing
pg. 44.
B. Output relay stuck - B. Replace Temperature Check to see if Controller. " Incubator Heating" Indicator light stays on while temperature is Increasing.
C. Incubator door open. C. Close incubator door and allow incubator to stabilize.
Audible alarm sounds. A. Temperature has risen A. Determine cause of
Beyond the set range temperature increase for the High Limit Safety. from prior section and take appropriate action. Call for service.
Bactron Operations Manual 34
______________________________________________________________________________
General Troubleshooting Guide Continued
V. VACUUM PUMP
Problem Possible Cause Solution
Pump continually turns on A. Leaking check valve - A. Disassemble check valve off in rapid succession. Normally due to debris clean off any debris from Ideally, pump should remain Between poppet and O - poppet, O - ring, off for 30 minutes or more ring seal. reassemble. Before coming on again.
B. Vacuum foot pedal valve B. Remove hose barbs may be leaking. Check from valve - there may to see if pinching its hose be some debris holding Stops problem. valve open. Replace valve if necessary.
C. Leak in pump plumbing - C. Tighten any loose pipe fitting switch, hose, fittings, check hoses etc. for cuts, holes, leaks around hose barbs.
D. Air Lock vacuum D1. Disassemble solenoid, Solenoid leaking. clean plunger, O - ring spring, etc. and reassemble.
D2. Make sure vacuum hose from pump is not pinched.
Vacuum motor does A. Broken vanes in A. Remove the end of turn on. Pumping section ( the the pump and replace
end of pump with filter any broken vanes. jars. )
B. Electrical problem with B. Replace pump. Motor.
C. Check pump power C. Check circuit breaker Supply. on back of chamber.
Pump runs continuously A. Set point on pump A. See pg. 25, " Setting when there is no demand Vacuum switch set too vacuum High switch". for vacuum. high.
Bactron Operations Manual 35
______________________________________________________________________________
General Troubleshooting Guide Continued
VI. MISCELLANEOUS
Problem Possible Cause Solution
Excessive moisture A. Thermoelectric A. Replace fan. buildup in chamber. condensate controller
fan is not operating.
B. Ambient room B1. Maintain room temperature too low. temperature a bit higher.
B2. Cover chamber with a blanket overnight as means of
installation.
C. Air conditioner unit C. Relocate chamber, directly above
chamber.
conditioning duct.
D. Thermoelectric D. Empty thermoelectric condensate controller condesate controller not emptied regularly. container daily. as means of
Loop sterilizer does A. Loop covered with A. Use a new loop. not work. oxidized layer which
insulates from power supply.
redirect the air
installation.
Bactron Operations Manual 36
______________________________________________________________________________
General Troubleshooting Guide Continued
VI. MISCELLANEOUS
Problem Possible Cause Solution
Loop sterilizer does B. Sterilizer unit not getting B1. See that the unit is not work. ( Continued) power. Plugged into internal
Outlet.
B2. See that the internal Outlet is getting power.
B3. Replace.
Bactron Operations Manual 37
______________________________________________________________________________
22.0 BACTRON OPERATION TIPS
Keeping the Bactron Chamber Anaerobic: The Bactron chamber will be anaerobic if it has the following three
things: Hydrogen, Positive Pressure, and an Active Catalyst. See Instruction Manual drawing on page 24.
User should place oxygen indicators inside the chamber or an oxygen meter to ensure that they have an anaerobic environment. One indicator should be placed inside the working chamber and another in the incubator. The indicators turn color in the presence of oxygen, and are available through distributors and supply companies.
If the indicators appear “very blue” purge the chamber with 100 PSI and wait for 30 minutes. The open a new indicator. If the new indicator is “white” in color your chamber in anaerobic. The greater the color of the indicator, the more gas to purge through the chamber and the longer users should wait before re-checking system.
Gas Source: Sheldon Manufacturing, Inc. recommends an AMG, Anaerobic Mixed
Gas, mixture of 5% CO2, 5% Hydrogen, and 90% Nitrogen for anaerobic applications.
This mixture is available marked “NON-FLAMMABLE”.
Two national sources of AMG gas are AIRCO and Air Products.
How much gas will a Bactron chamber use?: Gas usage is dependent upon two things; the number of times a user
enters their chamber daily and operating technique.
Our product literature states the Air Lock uses approximately 96 liters
of gas and the sleeve system uses approximately 3 liters of gas. A standard tank contains 2,200 PSI of gas. During chamber installation, 300 PSI of gas is used. There is about 70-80 Air Lock cycle uses per tank of gas.
Methods to Conserve Gas:
Users can utilize the Air Lock as a holding station. As samples are
received they can be stored in an anaerobic environment by placing them in the Air Lock and activating either the automatic or manual Air Lock cycle.
Bactron Operations Manual 38
______________________________________________________________________________
1. Users should evaluate and try to limit the number of times the Air Lock is opened. For example, placing samples in the Air Lock once in the morning and once in the afternoon as opposed to four times a day is desirable.
2. Gas is also conserved by bringing samples in through the sleeve system. Easy samples to transport through the sleeves are transport tubes as well as plates that are sealed.
3. Placing an object in the Air Lock will reduce the amount of cubic area therefore reducing the amount of atmosphere exchanged and gas utilized.
4. Good sleeve system technique. During the first vacuum cycle, remove all the atmosphere from the sleeve. Users should be able to see the outline of their fingers. Next, bring enough gas in to the sleeve (by pressing the GAS foot pedal) as to eliminate the definition of the fingers or when the sleeve is nor clinging to the arm.
Users should not “balloon” the sleeve out when flushing with gas. This action will waste gas.
5. The bubbling of the manometer and the activation of the “chamber gassing” indicator light on the control panel notifies the user of usage will working in the chamber.
When a user is in the “working Area” they should be aware of the effects their movements have on gas consumption.
Gas consumption can be reduced by balancing arm movements. For example, when extending the right arm into the chamber, bring the left arm close to the front of the chamber, thus balancing the pressure and reducing gas displaced by area.
Aborting Auto-Cycle:
How much does AMG gas cost?:
Sheldon recommends that customers get competitive bids on their gas
The automatic Air Lock can be aborted by pressing and holding the
“Auto Start/Reset” button for five (5) seconds. If a user forgets to place a sample in the Air Lock the cycle can be aborted in order to eliminate the unnecessary use of gas.
The price that a gas supplier charges is often determined by the
amount of the gas the customer uses. The price per tank will decrease as the quantity ordered decreases.
to ensure that they are receiving the “best” market price.
Bactron Operations Manual 39
______________________________________________________________________________
The average cost of AMG gas is $30.00 per tank, and $10.00 for
nitrogen in the United States. The cost range of AMG gas is between $8.00 and $58.00 per tank.
Dual Gas Systems:
Refer to the Instruction Manual “Gas Connections” for installation
A dual gas system, the nitrogen gas is utilized in the first two flushes of
Therefore, in a dual gas system, less AMG gas is used. AMG gas is
Customers using a dual gas system need to have two regulators. An
Catalyst:
Customers need to place a reactivated catalyst in their chamber each
The catalyst can be tested to ensure it is activated by flowing
The catalyst can be “cleaned” by heating to 200° Celsius, then flowing
Positive Pressure:
The manometer provides users with a visual indication that they have
Two catalyst cartridges containing palladium pellets are provided with
The Bactron product line comes standard with the capability to operate
a chamber with a single gas or a dual gas system. This is an exclusive feature that provides customer flexibility.
details.
the automatic cycle of the Air Lock. The third and final flush is AMG gas. The sleeve system always utilizes AMG gas.
more expensive than nitrogen, and the users overall gas expense is reduced.
AMG regulator is supplied with the Bactron Chamber. A nitrogen regulator is listed in the Instruction Manual accessory parts list.
the Bactron Chamber. The catalyst must be active in order to ensure an anaerobic environment.
day the chamber is used. The catalyst can be reactivated by heating for a minimum of 1-1/2 to 2 hours at 160° Celsius.
anaerobic gas over the pellets. If the pellets become warm or hot, they are working properly and are ready for use in the chamber.
the anaerobic gas mixture over the catalyst.
If the chamber is installed properly and is connected to a gas supply
the system will have positive pressure.
positive pressure. When the manometer bubbles, for example when a user enters the chamber, it is releasing chamber atmosphere/pressure.
Bactron Operations Manual 40
______________________________________________________________________________
Role of Condensate Controller: The Condensate Controller is an exclusive Bactron feature of Sheldon
Manufacturing, Inc. Our competitors use desiccants. Desiccants are a drying agent and often dry out samples.
The Condensate Controller ensures the following:
1. Eliminates condensation in the chamber.
2. Samples will not dry out.
3. Chamber will not fog.
The Condensate Controller system is located on the left side of the chamber interior behind the catalyst cartridge. Excess moisture is drained into a tube for easy removal.
Place a container i.e. a flask or beaker under the tube for collection. Customers should empty the container routinely.
Chamber Checklist: It is recommended that customers review a checklist before they enter
the Bactron Chamber. The checklist serves as a reminder of supplies and samples they need to bring into the unit.
Common items listed on a checklist are: re-activated catalyst, samples
and loops. The checklist can be placed on the front Plexiglass panel.
How to Determine if A Chamber is Leaking: Users should note the level of their gas tank(s) on a routine basis. This
provides a way to monitor chamber gas consumption.
Users can perform the following check before they exit the chamber
through the sleeve system to ensure the chamber is closed properly. Upon closing the arm port doors and before removing arms form the sleeves, user should push on the sleeves. If the arm port doors are not closed properly, the manometer will bubble.
Hand Held Leak Detector: If a chamber leak is suspected use the hand held leak detector
provided to identify where the leak is located. The detector senses hydra-carbons, in our case hydrogen.
Be aware that when using the leak detector, the manometer vent hole
on the back of the chamber will activate the detector. This is because hydrogen is being release from the chamber through the manometer.
Bactron Operations Manual 41
______________________________________________________________________________
To check the manometer for leaks, use a soap and water solution over the manometer and observe for bubbles.
Amount of Water Collected in the Dehumidifier: The amount of water collected in the dehumidifier depends on the
Size of Sleeve Cuffs: It is critical that the sleeve cuffs seal around the user’s arm during
The location that the cuffs seal on the user’s arm is determined by
The size of the cuffs that come standard with the Bactron Chamber is
Replacement of Cuffs: When and how often the cuffs on the sleeve need to be replaced
Customers should watch for small holes that can appear where the
Electrical
Model Volts Amps VA Cycles BacI 120 8 1100 50/60 hz 240 4 1100 50/60 hz BacII 120 8 1100 50/60 hz 240 4 1100 50/60 hz BacIV 120 11 1350 50/60 hz 240 6 1350 50/60 hz
number and types of samples in the chamber. The typical amount collected is around 50cc each day of chamber use.
chamber operation to ensure the desired anaerobic environment.
individual preference.
medium, size 8 glove. Other sizes are available; large size 9, and small size 6 ½.
depends on customer care and technique. Typically the cuffs are replaced about every three months.
cuffs are placed on the “ring cuff”. If holes are present the cuffs should be replaced to avoid gas leak. Watches and jewelry can damage the cuffs and are not recommended to be worn during chamber use.
Bactron Operations Manual 42
______________________________________________________________________________
23.0 Wiring Diagram
Bactron Operations Manual 43
______________________________________________________________________________
Bactron Operations Manual 44
______________________________________________________________________________
Bactron Operations Manual 45
______________________________________________________________________________
Bactron Operations Manual 46
______________________________________________________________________________
Bactron Operations Manual 47
______________________________________________________________________________
24.0 BACTRON PARTS LIST
DESCRIPTION PART NUMBER PART NUMBER QUANTITY
115 VOLT 220 VOLT
Gasket 9x9 Soft Lt. Blue (Bac I) Gasket 12x12 Soft Lt. Blue (Bac II,
IV)
3450506 3450506 1 3450507 3450507 1
Power Cord 1800516 101990 1 Main Circuit Breaker 1100505 1100505 1 Main Switch 103351 103351 1 Incubator Safety Alarm 250501 250502 1 Incubator Safety Thermostat 1750500 1750500 1 Incubator Safety Indicator 200020 200020 1 Heating Element
Bactron I, II 2350502 2350502 2
Bactron IV 890081 1
Bacrton IV 890081 2 Vacuum Pump Ckt Bkr 1100505 1100505 1 Incubator Fan (Bac IV) 210002 210001 1 Vacuum Pump Outlet 100020 101483 1 Door Sensors 103208 103208 1 Start Switch 101970 101970 1 Vacuum Control Switch 7850502 7850502 3 System Control Switch 103207 103207 1 Gas/Vacuum Switch X1000411 X1000411 1 Gas 2 Solenoid Valve 8600528 8600529 1 Door Indicator 200020 200020 1 Aerobic Indicator 200020 200020 1 Anaerobic Indicator 200020 200020 1 Gas Indicator 200020 200020 1 Vacuum Indicator 200020 200020 1 Gas 1 Solenoid 8600528 8600529 1 Chamber Gas Solenoid 8600528 8600529 1 Chamber Gas Indicator 200020 200020 1 Vacuum Solenoid 8600509 8600508 1 Thermoelectric
Condesensate Controller 1070502 1070502 1 Cooling Fan X1000300 X1000300 1 Chamber Pressure Switch 7850508 7850508 1 Bacloop / Light Box Switch X1000124 X1000124 1EA Bacloop / Light Box Cord 100014 101990 1EA Bacloop / Light Box Breaker 1100505 1100505 1EA Light Fixture X1000292 X1000751 1 Bacloop Pilot Light 200020 200020 1 Bacloop Transformer 103352 103352 1
Bactron Operations Manual 48
______________________________________________________________________________
25.0 Loop Sterilizer
Directions For Use
1. Touch the end of in inoculator loop or needle to the front and rear shorting posts as shown.
2. As soon as the inoculator needle or loop glows red, remove it from the contact with the shorting bars.
3. Allow the needle or loop to cool before putting into use or setting aside.
Bactron Operations Manual 49
Loading...