8.0 Incubator Temperature Controller/Incubator Safety Page 13
9.0 Incubator Calibration Page 13 - 14
10.0 Vacuum Gauge Adjustment Page 14
11.0 Pressure Adjustment Page 15
12.0 Testing The Chamber For Gas Leaks Page 15 - 16
13.0 Making The Chamber Anaerobic/Desired Environment Page 16 - 18
14.0 Chamber Operation Page 18 - 19
15.0 Entering The Chamber Page 19 - 20
16.0 Exiting The Chamber Page 20
17.0 Maintenance Page 21
18.0 Chamber QC Sheet Page 22
19.0 Accessories Page 23
20.0 Chamber System Drawings:
A) Keeping The Chamber Anaerobic Page 24
B) The System Controller and Controls:
View From Behind Panel Page 25
C) Solenoid Page 26
D) The System Controller Indicator Guide Page 27
IMPORTANT: READ THIS INSTRUCTION MANUAL IMMEDIATELY.
Your satisfaction and safety require a complete understanding of this unit, including its proper
function and operational characteristics. Be sure operators are given adequate training before
attempting to put the unit in service. NOTE: This equipment must be used only for its
intended application; any alterations or modifications will void your warranty.
Orientation
Figure 1a. Bactron II Chamber
Note: Bactron I Chamber is similar; it has three incubator compartments.
Consists of a neoprene sleeve, a plastic cuff ring, O-Ring, and a soft rubber cuff. The assembly
allows gloveless chamber operation.
Figure 3. Armport Doors
CAUTION: The Bactron chambers have an efficient method of sealing the chamber by utilizing the
Armport doors as shown in Figure 3. The tightening knob on the armport doors should be turned
just until the knob begins to “grab”. DO NOT OVER TIGHTEN. Damage can occur if too much
force is placed on the Armport door assembly. See Section 13.5
Doors are used to seal the armports while the chamber is not in use. Doors can be stored using
the holders provided while the chamber is in use. Assembly consists of an armport door,
tightening knob and locking bar. Doors are secured by tuning the locking bars horizontally to pass
through the cutouts in chamber, then turning locking bars verticals and tightening knobs.
Provides ability to take material into and out of the chamber. Air Lock can be operated either
automatically or manually. The Air lock shelf (Models II and IV) allows for materials to be
conveniently rolled from the Air Lock into the chamber.
1.0 RECEIVING AND INSPECTION
1.1 The carrier, when accepting shipment, also accepts responsibility for safe delivery
and is liable for loss or damage claims. On delivery, you must inspect for visible
exterior damage. Note and describe on the freight bill any damage found and enter
your claim on the form the carrier supplies.
1.2 Inspect for concealed loss or damage on the unit itself, both interior and exterior. If
any, the carrier will arrange for official inspection to substantiate your claim. Save
the shipping crate until you are sure the unit has been delivered in good condition.
1.3 If for any reason you must return the unit, contact your sales representative for
authorization and supply nameplate data.
2.0 INSTALLATION
2.1 Local city, county, or other ordinances may govern the use of this equipment. If you
have any questions about local requirements, please contact the appropriate local
agency.
2.2 Under normal circumstances these units are intended for use indoors, at room
temperatures between 5° and 40°C, at no greater than 80% relative Humidity (at
25°C) and with a supply voltage that does not vary by more than 10%. These ovens
should not be operated at an altitude exceeding 2000 meters. Installation category is
CAT-II.
2.3 Pollution Degree 2. Customer service should be contacted for operating conditions
outside of these limits. Installation may be performed by the end user. It is
unnecessary for this unit to be installed by a technician.
2.4 Location: When selecting a site for the unit, consider conditions which may affect
performance, such as heat from steam radiators, ovens, autoclaves, etc. Avoid
direct sun, fast-moving air currents, heating/cooling ducts, and high traffic areas. To
ensure air circulation around the unit, allow a minimum of 2” between chamber rear
and sides and any walls, partitions, or obstructions to free airflow.
2.5 Power Source: The power source must match the voltage, cycle, phase and
amperage requirements listed on the data plate. Plug the cord into a grounded
outlet. VOLTAGE OF THE OUTLET SHOULD NOT VARY MORE THAN 10% FROM THE DATA PLATE RATING. A separate circuit is recommended to
precluded loss of product due to overloading or circuit failure. Note that the electrical
supply to the unit must conform to all national and local electrical codes.
2.6 Gas Source: Install the gas regulator(s) on the tank(s) of gas. Chain the gas
tanks(s) to a secure position on the wall. Set the regulator(s) to 10 psi.
2.7 If you are using an anaerobic gas mixture (AMG), we recommend 5% (H2), 5%
Carbon Dioxide (CO2) and 90% Nitrogen(N2).
2.8 If you are using a two gas anaerobic application, we recommend one tank of AMG
gas and the second tank of 100% Nitrogen (N2).
Figure 7. Warning Label
2.9 Cleaning: The Bactron unit was cleaned at the factory, however, a general cleaning
is recommended. Use BENZALKONIUM CHLORIDE to clean your chamber. See
page 23 for a list of cleaning products.
3.1 Power: The lighted power “I/O” switch controls all power to the Bactron Chamber. It
must be in the “I” position before any systems are operational, including control
circuitry.
3.2 Digital Display: The actual incubator temperature will be digitally displayed to +
.1°C.
3.3 Main Temperature Control: This control consists of the Up/Down buttons and the
digital display.
3.4 Vacuum Gauge: Analog dial shows the vacuum level of the Air Lock in inches of
mercury.
3.5 Incubator Safety: The Safety is an independent thermostat that monitors incubator
temperature. It guards against any failure of the incubator controller which would
allow temperature to rise past set point. If temperature rises to the Safety set point,
the Safety will take over control of the heating element and an audible alarm will be
activated. The Safety controller allows for continued use of the chamber until service
can be arranged.
3.6 Air Lock Controls. Manual Cycle: The manual cycle switch controls the vacuum
and the gassing process of the Air Lock. Push VAC to evacuate the Air Lock then
GAS to fill the Air Lock with the desired gas mixture. This process should be done
three times.
3.7 Air Lock Controls. Auto Cycle: The automatic cycle switch will evacuate the Air
Lock and then fill the chamber with the desired gas mixture. The process will cycle
three times and upon completion the ANAEROBIC light will activate. The cycle can
be aborted by pressing the switch to the START/RESET and holding it down for five
seconds.
3.8 Incubator Heating: Light indicates that the heating element is activated and that the
incubator is heating.
3.9 Safety Activated: Light indicates that the Incubator Safety audible alarm is activated
and controlling the incubator temperature.
3.10 Chamber Gassing: Light indicates that the chamber is demanding gas. The light
activates when the operating the Air Lock and sleeves, in addition to maintaining
positive pressure in the working chamber.
3.11 Air Lock Doors Sealed: Light indicates that both the Air Lock doors are closed.
3.12 Gas Activated: In the manual cycle the GAS light indicates that the Gas switch is
activated. In the automatic cycle, third cycle, light indicates that the GAS is activated
in the Air Lock.
3.13 Vacuum Activated: Light indicates that the VAC switch of the manual cycle Air
Lock is activated. Light is also on during automatic vacuum cycles.
3.14 Auto Cycle Aerobic: Light indicates that the Auto cycle has not been activated.
When the automatic cycle has been activated, the Auto Cycle Anaerobic lights will
alternate.
3.15 Auto Cycle Anaerobic: Light indicates that the Auto Cycle has been activated and
is complete. When the automatic cycle has been activated, the Auto Cycle
Anaerobic lights will alternate.
Indicates “Earth Ground”
4.0 HOSE AND ELECTRICAL CONNECTIONS
4.1 Hose ports for the gas and vacuum supplies and foot pedal controls are provided on
the top back left side of the control module section. The Bactron models have five access
ports. (Older models have four access ports.) Figure 9 shows a typical hose connection for
a five port model.
Figure 9. Hose Connections
4.2 Install the hose from the gas regulator of the working gas, for example AMG, (GAS
1) tank to the port marked “GAS 1” on the chamber module.
4.3 Install the hose from the gas regulator of the purging gas, for example Nitrogen,
(GAS 2) tank to the port marked “GAS 2” on the chamber module.
NOTE: If using only one gas supply, install the “T” tube/fitting assembly (provided in
the accessory pack) between the “GAS 1 IN” port and the “GAS 2 IN” port on the
control module. Then attach the hose from GAS 1 tank regulator, as in 4.2, to the
unused brass “T” tube end. Failure to follow this procedure will render the purging
cycles of a single gas system inoperative.
4.4 Connect the 3/8” hose from the vacuum pump or a 3/8” hose from a suitable in-
house vacuum system, to the port marked VAC IN. In-house vacuum systems must
be capable of evacuating to a minimum of 25 inches of mercury.
4.5 Plug the vacuum pump power cord into the outlet receptacle at the top left of the
control module. This outlet is marked “5 AMPS” (installations using in-house
vacuum supplies will not follow this step).
NOTE: CONNECTING HOSES.
the hose into the appropriate connector until the hose stops. Pull back slightly to
ensure hose will stay in place. Reinsert if hose pulls out.
4.6 Connect the center “T” section from the foot pedal assembly to the tube fittings
located on the inner sides of the armport access openings on the front of the
chamber.
4.7 Connect the tube from the GAS side of the foot pedal assembly to port marked GAS
OUT on the control module section.
4.8 Connect the tube from the VAC side of foot pedal assembly to the port marked VAC
OUT on the control module section.
4.9 Plug power cord from Chamber into a 115 VAC, 60 Hz (220 volt units are also
available) electrical outlet. Turn the power switch on to verify power to the system.
4.10 When the main power switch is turned on, the vacuum pump should come on for a
few seconds and then go off. If the pump does not come on, or comes on and will
not go off, consult the Troubleshooting Guide for adjustment.
4.11 When the power switch to the Chamber is turned on, you should see Power ON,
CHAMBER GASSING, AIR LOCK DOORS SEALED and AUTO CYCLE AEROBIC
lights illuminate, along with a readout of the incubator temperature. If any of these
appear not to work, consult the Troubleshooting Guide.
Hose connections are intended to be simple. Insert
5.1 Located below the right end of the control panel and inside the chamber, the
Manometer provides a visual way to measure pressure in the chamber. It also
functions as a pressure release system to prevent damage to the chamber in
the event over-pressurization occurs. By measuring the difference between
the inner and outer water levels one can obtain the chamber pressure in
inches of water. If pressure becomes too great, excess gas will push out
through the inner tube and bubble through the water thus releasing pressure
from the chamber.
5.2 Purpose: The function of the manometer is two-fold:
a. Provides visual way to measure the pressure in the chamber. As the
pressure inside the chamber increases, the water level in the inner tubes
moves down a distance proportional to the pressure increase. By measuring
the difference of the inner and outer water levels, the user can obtain the
chamber pressure in inches of water. A water level indicator notifies user that
water is high or low.
b. It acts as a pressure release system to prevent damage to the chamber in
case of inadvertent over-pressurization. If the pressure becomes too great,
the excess gas simply bubbles through the water and out the back of the
chamber.
5.3 Filling Procedure: Use a pipette or plastic wash bottle to fill the manometer with
distilled water to a level of two or three inches. Lines on the manometer provide a
guide. Fill water to the top line. This may be done from the top of the manometer
inside the chamber.
6.0 CATALYST CARTRIDGE
6.1 The catalyst cartridge removes ant trace amounts of oxygen from the chamber. Two
catalyst cartridges containing Palladium catalyst pellets are provided. See drawing
pg. 4
6.2 One cartridge should always be in the chamber. The second spare cartridge is
provided so that reactivation of the catalyst can be done after each chamber usage.
6.3 The catalyst cartridge is installed on the left wall inside the chamber. To install slide
the cartridge into the metal tracks and secure the handle into the bracket.
6.4 Reactivate spare catalyst daily by heating for a minimum of 1 –1/2 to 2 hours at
160°C (320°F). You may keep the unused cartridge in the heating oven at 160° prior
to its use. (NOTE: Catalyst cartridge should be reactivated for the first time before
installing into the unit.)
6.5 Quality control catalyst monthly by flowing anaerobic gas over the reactivated pellets.
If the pellets become warm to hot, they are working properly and are ready for use in
the chamber. If they do not heat up when the gas is applied, reheat to 180° to 200°C
and flow gas over catalyst again to remove any buildup of sulfides that might hinder
its effectiveness.
6.6 The entire catalyst cartridge can be placed in the heating oven for reactivating.
However, if necessary, the handle on the cartridge can be removed.
6.7 A catalyst cartridge can also be placed inside the Air Lock to remove any trace
amounts of oxygen.
7.0 THERMOELECTRIC CONDENSATE CONTROLER
7.1 The thermoelectric condensate controller collects excess humidity and eliminates
chamber condensation. Moisture is funneled into a tube for collection inside your
chamber.
7.2 The thermoelectric condensate controller is located on the left side of the chamber
interior behind the catalyst cartridge. A plastic tube drains excess moisture into the
working chamber, on the back left side.
7.3 Place a container, e.g., glass flask or beaker, under the tube for collection. Empty
the container DAILY.
7.4 Desiccants are not recommended in the chamber. Desiccants are drying agents that
can regrettably pull moisture from your samples.
8.0 INCUBATOR TEMPERATURE CONTROLLER/INCUBATOR SAFETY
8.1 Setting Incubator Controller: The INCUBATOR TEMPERATURE controller
regulates the incubator temperature by use of Up/Down buttons. To enter set point
mode on the control, press either the Up or Down button one time. The digital
display will start to blink, going from bright to dim. While blinking, the Digital Display
is showing the set point. To change the set point, use the Up and Down buttons. If
the buttons are not pressed for five (5) seconds, the display will stop blinking and will
read the temperature of the unit. Note that the INCUBATOR SAFETY should be
turned to its maximum position, (clockwise) until the unit has stabilized at desired set
point temperature. Allow the incubator at least 24 hours to stabilize. Then recalibrate the digital display to your reference thermometer, follow the calibration
instructions given in Section 9. This incubator was calibrated at the factory at 37°C.
8.2 Setting Incubator Safety: Set the control to the desires incubator temperature.
Then set the INCUBATOR SAFETY to its maximum position. Allow 24 hours for
stabilization before proceeding.
If, after 24 hours the temperature is not at desired level, adjust the INCUBATOR
TEMPERATURE up or down until the precise desired temperature is achieved. (See
Section 9 for calibration procedures)
When stabilization at the desired temperature has been achieved, turn the
INCUBATOR SAFETY control knob counter-clockwise until the INCUBATOR
SAFETY ACTIVATED light comes on and audible alarm is activated. Next, carefully
turn the INCUBATOR SAFETY knob clockwise until the light is just off. The Safety is
now set; the INCUBATOR SAFETY ACTIVATED light should remain OFF during
normal operation.
NOTE: An accurate thermometer should be used inside the incubator
as a reference when setting and calibrating the incubator temperature.
8.3 It is a good idea to mark the INCUBATOR SAFETY label with the desired
setting or position as a backup in case the knob is moved accidentally. If the
SAFETY ACTIVATED light is on and audio alarm is activated at any time,
check the INCUBATOR TEMPERATURE controller setting to be sure that it is
not set above the INCUBATOR SAFETY setting.
9.0 INCUBATOR CALIBRATION
9.1 Compare the reading of the digital temperature display to an accurate reference
thermometer. If there is a difference, put the display into calibrate mode by pressing
both the Up and the Down buttons at the same time and holding them in for about
five (5) seconds or until the two outside decimal points start to flash on and off.
When the decimal points are flashing, the display can be calibrated to match the
reference thermometer by pressing the Up or Down button until the display reads
correctly.
10.0 VACUUM ADJUSTMENT
10.1 The vacuum gauge for the Automatic Air Lock Cycle is factory calibrated. It is
recommended that customers note the setting of the vacuum gauge during Auto Air
Lock Cycle to ensure that the setting is correct. Evacuation and the vacuum pump
capabilities can effect the calibration.
10.2 The vacuum setting for the Auto Air Lock Cycle are factory set at the High Limit 18
and the Low at 4.
10.3 Vacuum calibration procedure:
High Limit
A) Activate the Auto Air Lock Cycle and note the high level reached by the
vacuum gauge. If the vacuum gauge is not at the desired High setting of
18, adjust the high vacuum switch located on the left outside if the control
housing box.
B) This switch is illustrated in the “CONTROLS: View From Behind Control
Panel” Page 25, #10.
C) Adjust the High setting by turning the large knob until the desired value of
18 is reached. (Black notch on knob toward 0 to decrease or toward 30 to
increase). See page 25.
D) Activate the Auto Air Lock Cycle and observe the vacuum gauge on the
high level reached.
E) Repeat steps C and D.
Vacuum calibration procedure:
Low Limit
A) Activate the Auto Air Lock Cycle and note the low level reached by the
vacuum gauge. If the vacuum is not at the desired Low setting of 4,
adjust the Low vacuum switch located in the same area as the high
vacuum switch noted above.
B) Repeat the procedure noted above under High vacuum to adjust the low
Vacuum setting except adjust for the lower value of 4.
11.1 The chamber pressure of your unit is factory calibrated but local elevations may
make adjustment of this value necessary.
11.2 If the manometer is bubbling when no one is working inside the chamber or the water
level of the manometer is less than ½ inches, the pressure should be adjusted. See
manometer diagrams on page 11 for a reference. Following is the adjustment
procedure:
A)The chamber pressure switch is located on the outer left side of the
control housing box. It is adjusted by means of an Allen wrench
adjustment screw (clockwise to increase/counterclockwise to
decrease) see page 25.
B) Adjust the setting of this valve only if the manometer level is above the
½ inch mark or is bubbling as noted above.
C) Repeat the adjustment, if necessary, observing carefully the result,
until the proper performance is obtained.
12.0 TESTING THE CHAMBER FOR GAS LEAKS
12.1 Each chamber is individually tested at the factory before shipping. The following
procedure, however, should still be followed to detect any possible damage due to
shipping.
12.2 Make sure the catalyst is removed from the chamber during initial leak testing.
12.3 Check to ensure Manometer is filled.
12.4 Close the outer and inner air Lock doors.
12.5 Secure the Armport doors. Turn the locking bars horizontally to pass through the cutouts in the chamber, then turn bars so they are vertical, tighten knobs. Do not
over tighten. See Note at the end of Section 13.0
12.6 Open the gas tank valve all the way on and set the gas regulator to 10 psi.
12.7 Turn the power switch on. As before, with the power switch on, you should see the
POWER SWITCH, INCUBATOR TEMPERATURE display, CHAMBER GASSING,
AIR LOCK DOORS SEALED, and AUTO CYCLE AEROBIC light illuminate.
12.8 The CHAMBER GASSING light will turn OFF when the water level in the manometer
reaches a level of approximately ½” of water. The AIR LOCK DOORS SEALED
light will turn ON when the inner and outer Transfer Module doors are both closed. If
this light is not ON, either the doors are not closed properly or one or both of the door
sensors are improperly adjusted (See Troubleshooting Guide).
12.9 If there are no leaks in the system, the CHAMBER GASSING light will remain OFF,
and the water level in the manometer will remain stable at approximately ½” water.
Normally, a “leak-tight” chamber will go from 15 to 30 minutes without the
CHAMBER GASSING light coming on again.
12.10 If the system has a gas leak, the CHAMBER GASSING light will flash ON and
OFF. Flashing every 10-30 seconds indicates a very large leak, whereas flashing
every 10 minutes or so indicates a small leak. This is accompanied by a “clicking”
sound (originating from the chamber gas solenoid). In this case, check the
following:
A) Armport doors for adequate seal of O-rings.
B) Air Lock outer/inner door gaskets for adequate contact with doors.
12.11 If the system still has a leak, refer to:
A) Chamber Maintenance: Leak Detection
B) Trouble Shooting: Excessive Gas Consumption
13.0 MAKING THE CHAMBER ANAEROBIC/DESIRED ENVIROMENT
Prior To Purging The Chamber
13.1 Turn Power switch to the “O” (Off) position.
13.2 Place about six oxygen indicators inside the system. Place one unopened indicator
in the incubator, and place one unopened indicator in the work area. The others can
be stores in the work are for future use.
13.3 Place catalyst in the chamber. For more information about the correct placement of
the catalyst, see Section 6.0. Catalyst Cartridge page 11 and 12.
13.4 Open the incubator doors before leaving the chamber. This allows the incubator to
become anaerobic also.
13.5 Put armport doors in place.
NOTE: The Bactron chambers have an efficient method for sealing the chamber by
utilizing the Armport doors as shown in Figure 3 on page 4 . The tightening knob on
the Armport doors should be snug, DO NOT OVER TIGHTEN
too much force is placed on the Armport door assembly. When the knob just begins
to “grab”, the tightening bolt has reached its maximum point. Over tightening will
cause the bolt to slip inside the handle and cause leakage.
13.8 Loosen the right armport door to create a small leak. (Remove sleeves from the
armports if they have been placed there. This will make it easier to adjust the
armport doors to create the proper leak.) Adjust the doors so that the CHAMBER GASSING light flashes ON and OFF for equal periods. It is important that the gas
flows into the chamber in “pulses” rather than as a “stream”. If the gas “streams”, it
tends to go to the open door only.
13.9 Bleed the gas mixture through the chamber until the reading on the tank regulator
gauge is 300-400 psi less than the original reading. Normally, it takes 1-1/2 to 2
hours to flush out all atmosphere from the chamber.
13.10 After the gas purge, tighten the armport doors to prevent leaking. The CHAMBER
GASSING light will turn off, flow of gas into the chamber will stop, and the water
level difference in the manometer should again be approximately ½” and should
remain stable.
13.11 At this point it is important to check if the chamber is completely anaerobic before
using it. Check this by first replacing the sleeves on the armports, and the entering
the chamber. For more information about entering the chamber see Section 15.0,
Entering the Chamber page 19.
13.12 A) Once inside the chamber, open one indicator strip in the incubator, and one
strip in the work area.
NOTE: Please note and follow the directions for use for the type of indicator
strip you are using.
B) If the strips indicate that the atmosphere is anaerobic, the chamber is ready
for use.
C) If the strips indicate an aerobic or slightly aerobic atmosphere, then exit the
chamber. Create a small leak as detailed in Step 13.8. Flush an additional
100-200 psi through the chamber. After this step, proceed to step 13.11, until
indicator strips indicate an anaerobic atmosphere.
D) The gas flowing through the chamber may cause previous indicator strips to
dry out. Use fresh indicator strips stored in step 13.2 each time you test the
atmosphere.
NOTE: During the initial purge you may notice heavy condensation within the chamber.
This is due to catalytic formation of water from hydrogen gas and oxygen initially in the
chamber. Most of this condensation will dissipate by the end of the gas purge cycle.
CAUTION: The catalyst cartridge will be HOT at this point. If you must touch it for any
reason, use extreme care.
Summary
The chamber is ready when:
A) The CHAMBER GASSING light remains off indicating no gas leaks.
B) The incubator doors are OPEN.
C) Oxygen indicators are in the chamber.
D) The catalyst cartridge is in place.
14.0 CHAMBER OPERATION
Air Lock Operation
14.1 Load the Air Lock with samples, materials, supplies, etc.
14.2 Close the outer and inner Air Lock Doors.
14.3 Activate the Air Lock by one of two methods:
A) Manual Cycle (Manual Gas/Vacuum Switch):
i. Locate the 3 position Manual Gas/Vacuum Cycling Switch which should be in
the middle.
ii. Turn the Manual GAS/VAC switch to VAC until a vacuum of 17 to 20-in. Hg is
reached, as indicated by the VACUUM GAUGE.
iii. Flip the GAS/VAC switch to GAS to refill the Air Lock with the gas mixture
until the AIR LOCK VACUUM GAUGE reads approximately 4-in. Hg on the
first two cycles. On the third and last cycle leave the gas on until the gauge
reaches zero to ensure that there is no vacuum still present in the Air Lock,
then return the switch to the middle or off position.
iv. It is essential to perform the cycle at least three times. Once this is done, you
may enter the chamber through the armports, using the procedure specified in
the next section, and take your materials into the chamber through the inner
Air Lock door.
NOTE: When cycling the Air Lock manually, the AUTO CYCLE ANAEROBIC light on the
control panel will not illuminate when the cycle is completed, as it does at the end
of the automatic cycle.
B) Automatic Cycle:
i. Make sure both Air Lock doors are closed. The automatic cycle will not start
otherwise.
ii. Press the “AUTO CYCLE START/RESET” button to initiate the cycle. The
AUTO CYCLE AEROBIC/AUTO CYCLE ANAEROBIC lights will alternately
flash on and off, indicating the cycle is in progress. When the cycle is
complete, the system will stop automatically with the AUTO CYCLE
ANAEROBIC light on.
iii. If the cycle does not follow the sequences in the previous section on manual
operation (eg: cycle won’t start, cycle does not go through three times, cycle
cot within specified limits, door won’t open upon completion, etc.), consult the
Troubleshooting Guide for adjustments.
NOTE: Do not hold Auto Cycle Start Switch down. Simply Press the
button and let go to activate.
iv. If you WANT TO abort the cycle, press and hold the “AUTO CYCLE
START/RESET” switch for 5 seconds. The vacuum gauge will go to 0 and
stop. Then the AUTO CYCLE AEROBIC/AUTO CYCLE ANAEROBIC light
will stop flashing.
15.0 ENTERING THE CHAMBER
15.1 The rubber cuffs of the sleeve should be secured around your arm as opposed to
clothing. Remove watches, bracelets, etc., as they may damage the cuffs.
15.2 Insert your hands and forearms into the sleeves. The cuffs must be firmly secured
around your bare forearms. Keep your hands four to six inches away from the
armport doors.
15.3 Depress the VAC foot pedal (this procedure evacuated both sleeves), and continue
depressing until the sleeves have collapsed completely and firmly around your
forearms.
15.4 Flush the sleeves with the gas mixture by depressing the GAS foot pedal. Do not
overfill the sleeves. Stop when you feel a small separation of space between your
hands and the sleeves.
15.5 Repeat steps 15.3 – 15.4 three more times to ensure that the sleeves are completely
anaerobic/desired atmosphere before entering the chamber.
NOTE: ALWAYS enter the chamber with both arms simultaneously.
15.6 Loosen the armport door knobs a few turns and rotate the locking bars to a
horizontal position. The armport doors can now be pushed forward into the chamber
interior and can be secured on the armport door holders.
NOTE: Arm movements that are to forceful may cause the sleeves to dislodge from
the armports.
16.0 EXITING THE CHAMBER
16.1 Before leaving the chamber the incubator doors and the inner Air Lock door must be
closed.
16.2 Depress the GAS foot pedal to allow the sleeves to fill with gas. This procedure
keeps the sleeves from collapsing as you move outward, making it easier to
reposition the doors.
16.3 Hold the locking bars in a horizontal position, and close the doors. Rotate the locking
bars to a vertical position, and gently tighten both armport door knobs, just until they
“grab”.
16.4 Before removing arms from the sleeves:
A) Check to be sure that armport doors are securely sealed by slowly pushing
both arms forward (this action generates a pressure within the sleeves).
B) If the Manometer bubbles when the arms are pushed forward, this indicates
that the doors are improperly sealed. Loosen the armport door knobs again,
remove the doors. Check the O-rings for any particulate matter on the sealing
areas, and repeat Step “A”. If Manometer remains stable, the doors are
sealed properly.
16.5 When the armport doors are properly sealed, slowly and carefully withdraw both
arms from the sleeves.
1. Before using any cleaning or decontamination method except those recommended by the
manufacturer, users should check with the manufacturer that the proposed method will not
damage the equipment.
2. Exchange chamber catalyst cartridge with reactivated one.
Switches and monitors dual gas tank supply. Provides work bench to support Bactron
Provides automatic switch-over to second gas chambers. Chamber Stand height 30”.
tank alarm sounds when both tanks are empty. Stands feature locking casters and storage
Tanks can be switched manually at any time. Space.
Model # 2002-B Model: Bactron I, II, and IV
Nitrogen Gas Regulator Microscope Recommendation
Optional accessory used in a two gas chamber The LEICA brand of microscope is a
commonly used unit with Bactron System.
Model # 7150500 Microscope: # 9990516
Illuminator: # 4650502
Extended Service Contracts
A LUKAS illumination system is also
The contract offers a twelve month warranty on available, which includes a 250 watt
parts and labor. Ensures customer of product fiber optic light, light ring, 250 watt
satisfaction and optimal performance. Lamp house, and 0.3x lens.
Anaerobic mixture is 5% hydrogen, 5% carbon Dioxide, and 90% nitrogen. Do not exceed
5% hydrogen for safety reasons. Test for the presence of hydrogen in the gas mixture by
flowing the gas over the active catalyst. The catalyst will get hot if hydrogen is present.
Catalyst
Reactive by heating for 2 hours at 160° C. Test by flowing the anaerobic gas mixture over
the catalyst. If the catalyst is active, it will get hot. The catalyst can be “cleaned” by heating
to 200°C, then flowing the anaerobic gas mixture over the catalyst.
Positive Pressure
The Manometer will indicate if positive pressure is present in the chamber.
Function At Rest Start Vacuuming High Gassing Low Completed
Vacuum Vacuum
No action. Start Vacuum on. Switch Gas on. Switch Airlock
Both doors button Vacuum Point Vacuum Point anaerobic
closed. pressed.* Going up. Reached going down Reached
Airlock Lights Lights Lights
aerobic. flashing Flashing flashing
Inputs
x0 Start Off On* Off Off Off Off Off
x1 Outer Door On On On On On On On
x2 Inner Door On On On On On On On
x3 High Vacuum On On On Off** On On On
x4 Low Vacuum On On to OffOff Off Off On On
Output
y0 Vacuum Solenoid Off On On On to OffOff Off to OnOff
y1 Gas (AMG) Solenoid Off Off Off Off to On
(1)
y2 Anaerobic Light Off Flashing Flashing FlashingFlashing Flashing On
y3 Aerobic Light On Flashing Flashing FlashingFlashing Flashing Off
y4 Door Light On On On On On On On
y5 Nitrogen** Off Off Off Off to On
(2)
*Input 0 is only when the start button is pushed
** Input 3 comes on only for a short moment
(1) Gas comes on for the last cycle only.
(2) Nitrogen comes on for the first two cycles.
CAUTION: Extreme caution must be exercised any time access is made into areas housing
electrical components. Repair, replacement or adjustment of components in these areas
must only be done by qualified technicians familiar with electrical circuitry and the
operation of the anaerobic chamber.
ATTENTION: Des Précautions extrèmes sont requises à chaque fois que vous accédez à
des endroits qui abritent les composantes électriques. La réparation, le remplacement oi
ìajustement de composanted dans ces endroits ne doivent être effectués que par un
technicien qualifié familier avec les circuits électriques et ì opération des étuves
anaérobiques.
21.0 Troubleshooting
General Troubleshooting Guide
I. GAS CONSUMPTION
Problem Possible Cause Solution
CHAMBER GASSING A. Armport door not sealing.
light comes on every
five (5) minutes or less
B. Leak in seams.
panel with leak detector.
C. Manometer leak, indicated
by very large leak out
manometer hole. ( Confirm
with soap solution)
D. Oxygen present-hydrogen
being consumed by
combing with oxygen.
CHAMBER GASSING A. Inner door gasket not
light comes on more sealing-check with leak
often with the outer door detector.
open than with it
closed
A. Tighten knob until it
"grabs", Check for worn
O-rings.
B. Check front plexiglass
Call for service.
C. Seal inner tube to
block, and block to
mounting plate with
solvent cement.
Chamber pressure A. Chamber pressure A. Tap switch to jar
Increase, Manometer sensor switch stuck in contacts loose, then
Bubbles, CHAMBER "on" position. replace switch.
GASSING light on.
B. Pressure set too B. Turn adjustment
high. Adjust pressure screw counter sensor. See diagram clockwise to
page 25. decrease pressure.
Excessive gas usage - A. Foot pedal leaking. A. Disassemble and clean
no chamber leaks GAS valve or replace.
B. Gas regulator/gas B. Check for and repair
input connections any leaks. Check
leaking. regulator for 10 psi
C. Leak in supply line. C. Check all gas lines
Chamber pressure increases, A. Chamber gassing A. Disassemble and clean
Manometer bubbles. Solenoid leaking, inlet or replace solenoid valve
CHAMBER GASSING gas pressure normal. if necessary. See
light off. Pinch hose to chamber Diagram pg. 26.
to see that bubbling
stops.
B. Chamber gassing B. Set inlet gas pressure
solenoid leaking, inlet to 10 psi.
gas pressure too high.
C. Chamber gassing C. See Figure 9, pg. 9.
solenoid hosed
incorrectly.
D. Air Lock gas solenoid D. Disassemble and clean
leaks. Pinch hose to or replace solenoid
module, see if it stops. valve. See pg. 26.
E. Incubator temperature E. No action necessary.
increase on startup.
Indicator strips are A. Low chamber pressure. A. Adjust chamber pressure
Activated , I.e. - turn color. sensor by turning
clockwise to increase.
See diagram pg. 25.
B. Catalyst inactive. B1. Use freshly heated
catalyst, flush chamber
with approx. 50 to 100
psig of gas depending
on how long chamber
has been aerobic and
size of chamber.
B2. Check to see that gas
makes catalyst hot. If
not, heat to 200 C and
blow gas mix through
to remove sulfide layer.
C. Circulating blower not C. Replace fan.
functioning.
D. Air lock not evacuating D. Turn High side switch
to high enough vacuum clockwise to increase.
( 17 - 18 inches Hg ). See diagram on page 25.
E. Outer Air Lock door not E. Check for damaged
sealing properly. gasket - replace if
necessary.
F. User did not perform F. Check technique of all
sleeve technique chamber users.
properly upon entering
chamber.
G. Gas tank mixture G1. Check for very low
contains an insufficient or empty tank.
concentration of
Hydrogen. G2. Check label for at
least 5% Hydrogen.
Air Lock does not hold A. Split gasket inner & A. Check gasket. If split,
Vacuum. outer door. replace.
B. Inlet fitting to module B. Tighten/reseal with
not sealed. teflon tape if necessary.
C. Air Lock solenoids hosed C. See diagram pg. 26.
incorrectly.
D. Vacuum solenoid leaking D. Disassemble and clean
Pinch hose to module solenoid - replace if
and see if leak stops. necessary. WD40 can
be used to clean
solenoid.
E. Defective vacuum gauge. E. Replace gauge.
Automatic cycle will not start. A. Input/output terminal A. Press down on the
buss is not seated terminal buss so that it
correctly. snaps into place.
NOTE: Power should be
turned off.
B. User pushed start switch B. Push start switch once
& held inadvertently held then release. See pg. 19.
down. Thus aborting
Auto Cycle.
C. Make sure both doors C. Adjust location of the
are closed. Door sensors door sensor so that
not making contact-check DOORS light is on
that DOORS light is ON. when both doors are
If not, check programmable closed.
Controller input display IN
0001 (outer door) and IN
0002 (inner door) should
be on.
Automatic cycle will not D. Start button inoperative - D. Replace start button.
start. (Continued) check input display - IN
0000 should be on when
start button is pushed,
OFF when not.
E. No power to controller - E1. Check for 24 volt DC
check power light on out of power supply.
Controller.
E2. Check for continuity
between power supply
and controller.
Chamber pressure goes A. Inner door gasket not A. Replace door gasket
down ( Manometer level rises) Making proper seal. if necessary
when cycle is started.
Cycle only goes through A. System Controller A. Replace System
two cycles. Malfunction. Controller.
Switch will not go from A. Insufficient vacuum A. Adjust switch on VAC
Vacuum to Gas. supply. pump to allow higher
vacuum before shut off,
around 20 inches Hg.
B. High vacuum set point B. Turn High side switch
too high, it should be counterclockwise to
around 17 - 18 inches Hg. decrease. See diagram
pg. 25.
Doors difficult to open A. Low vacuum set point A. Turn Low side switch
upon completion of cycle. too high. Counter clockwise to
Steady digital readout A. Temperature Controller A. Turn to pg. 13 to follow
Matches set point, but out of calibration. Complete Temperature
actual temperature does not. Controller instruction.
B. Incubator door open. B. Close incubator door
and allow incubator to
stabilize.
Digital readout and actual A. Incubator fan not A. Replace fan, Model
Temperature continue to functioning. Model Bactron IV.
Increase past set point. Bactron IV, See drawing
pg. 44.
B. Output relay stuck - B. Replace Temperature
Check to see if Controller.
" Incubator Heating"
Indicator light stays on
while temperature is
Increasing.
C. Incubator door open. C. Close incubator door
and allow incubator to
stabilize.
Audible alarm sounds. A. Temperature has risen A. Determine cause of
Beyond the set range temperature increase
for the High Limit Safety. from prior section and
take appropriate action.
Call for service.
Pump continually turns on A. Leaking check valve - A. Disassemble check valve
off in rapid succession. Normally due to debris clean off any debris from
Ideally, pump should remain Between poppet and O - poppet, O - ring,
off for 30 minutes or more ring seal. reassemble.
Before coming on again.
B. Vacuum foot pedal valve B. Remove hose barbs
may be leaking. Check from valve - there may
to see if pinching its hose be some debris holding
Stops problem. valve open. Replace
valve if necessary.
C. Leak in pump plumbing - C. Tighten any loose
pipe fitting switch, hose, fittings, check hoses
etc. for cuts, holes, leaks
around hose barbs.
D. Air Lock vacuum D1. Disassemble solenoid,
Solenoid leaking. clean plunger, O - ring
spring, etc. and
reassemble.
D2. Make sure vacuum
hose from pump is
not pinched.
Vacuum motor does A. Broken vanes in A. Remove the end of
turn on. Pumping section ( the the pump and replace
end of pump with filter any broken vanes.
jars. )
B. Electrical problem with B. Replace pump.
Motor.
C. Check pump power C. Check circuit breaker
Supply. on back of chamber.
Pump runs continuously A. Set point on pump A. See pg. 25, " Setting
when there is no demand Vacuum switch set too vacuum High switch".
for vacuum. high.
Keeping the Bactron
Chamber Anaerobic: The Bactron chamber will be anaerobic if it has the following three
things: Hydrogen, Positive Pressure, and an Active Catalyst. See
Instruction Manual drawing on page 24.
User should place oxygen indicators inside the chamber or an oxygen
meter to ensure that they have an anaerobic environment. One
indicator should be placed inside the working chamber and another in
the incubator. The indicators turn color in the presence of oxygen,
and are available through distributors and supply companies.
If the indicators appear “very blue” purge the chamber with 100 PSI
and wait for 30 minutes. The open a new indicator. If the new
indicator is “white” in color your chamber in anaerobic. The greater
the color of the indicator, the more gas to purge through the chamber
and the longer users should wait before re-checking system.
Gas Source: Sheldon Manufacturing, Inc. recommends an AMG, Anaerobic Mixed
Gas, mixture of 5% CO2, 5% Hydrogen, and 90% Nitrogen for
anaerobic applications.
This mixture is available marked “NON-FLAMMABLE”.
Two national sources of AMG gas are AIRCO and Air Products.
How much gas will a
Bactron chamber use?: Gas usage is dependent upon two things; the number of times a user
enters their chamber daily and operating technique.
Our product literature states the Air Lock uses approximately 96 liters
of gas and the sleeve system uses approximately 3 liters of gas. A
standard tank contains 2,200 PSI of gas. During chamber installation,
300 PSI of gas is used. There is about 70-80 Air Lock cycle uses per
tank of gas.
Methods to
Conserve Gas:
Users can utilize the Air Lock as a holding station. As samples are
received they can be stored in an anaerobic environment by placing
them in the Air Lock and activating either the automatic or manual Air
Lock cycle.
1. Users should evaluate and try to limit the number of times the Air Lock
is opened. For example, placing samples in the Air Lock once in the
morning and once in the afternoon as opposed to four times a day is
desirable.
2. Gas is also conserved by bringing samples in through the sleeve
system. Easy samples to transport through the sleeves are transport
tubes as well as plates that are sealed.
3. Placing an object in the Air Lock will reduce the amount of cubic area
therefore reducing the amount of atmosphere exchanged and gas
utilized.
4. Good sleeve system technique. During the first vacuum cycle, remove
all the atmosphere from the sleeve. Users should be able to see the
outline of their fingers. Next, bring enough gas in to the sleeve (by
pressing the GAS foot pedal) as to eliminate the definition of the
fingers or when the sleeve is nor clinging to the arm.
Users should not “balloon” the sleeve out when flushing with gas. This
action will waste gas.
5. The bubbling of the manometer and the activation of the “chamber
gassing” indicator light on the control panel notifies the user of usage
will working in the chamber.
When a user is in the “working Area” they should be aware of the
effects their movements have on gas consumption.
Gas consumption can be reduced by balancing arm movements. For
example, when extending the right arm into the chamber, bring the left
arm close to the front of the chamber, thus balancing the pressure and
reducing gas displaced by area.
Aborting Auto-Cycle:
How much does
AMG gas cost?:
Sheldon recommends that customers get competitive bids on their gas
The automatic Air Lock can be aborted by pressing and holding the
“Auto Start/Reset” button for five (5) seconds. If a user forgets to place
a sample in the Air Lock the cycle can be aborted in order to eliminate
the unnecessary use of gas.
The price that a gas supplier charges is often determined by the
amount of the gas the customer uses. The price per tank will decrease
as the quantity ordered decreases.
to ensure that they are receiving the “best” market price.
Role of Condensate
Controller: The Condensate Controller is an exclusive Bactron feature of Sheldon
Manufacturing, Inc. Our competitors use desiccants. Desiccants are a
drying agent and often dry out samples.
The Condensate Controller ensures the following:
1. Eliminates condensation in the chamber.
2. Samples will not dry out.
3. Chamber will not fog.
The Condensate Controller system is located on the left side of the
chamber interior behind the catalyst cartridge. Excess moisture is
drained into a tube for easy removal.
Place a container i.e. a flask or beaker under the tube for collection.
Customers should empty the container routinely.
Chamber Checklist:It is recommended that customers review a checklist before they enter
the Bactron Chamber. The checklist serves as a reminder of supplies
and samples they need to bring into the unit.
Common items listed on a checklist are: re-activated catalyst, samples
and loops. The checklist can be placed on the front Plexiglass panel.
How to Determine if
A Chamber is Leaking: Users should note the level of their gas tank(s) on a routine basis. This
provides a way to monitor chamber gas consumption.
Users can perform the following check before they exit the chamber
through the sleeve system to ensure the chamber is closed properly.
Upon closing the arm port doors and before removing arms form the
sleeves, user should push on the sleeves. If the arm port doors are
not closed properly, the manometer will bubble.
Hand Held Leak
Detector: If a chamber leak is suspected use the hand held leak detector
provided to identify where the leak is located. The detector senses
hydra-carbons, in our case hydrogen.
Be aware that when using the leak detector, the manometer vent hole
on the back of the chamber will activate the detector. This is because
hydrogen is being release from the chamber through the manometer.
number and types of samples in the chamber. The typical amount
collected is around 50cc each day of chamber use.
chamber operation to ensure the desired anaerobic environment.
individual preference.
medium, size 8 glove. Other sizes are available; large size 9, and
small size 6 ½.
depends on customer care and technique. Typically the cuffs are
replaced about every three months.
cuffs are placed on the “ring cuff”. If holes are present the cuffs should
be replaced to avoid gas leak. Watches and jewelry can damage the
cuffs and are not recommended to be worn during chamber use.