1 General
1.1 Warranty
The following instructions must be complied with in order for any
warranty to be applicable regarding functionality and safe operation of this equipment. This product must only be commissioned
by qualied professionals who are familiar with the legal requirements (Germany: DVGW).
● Read these operating instructions prior to operating the product.
● Use the product only as intended.
● Repairs and maintenance must only be carried out by special-
ist technicians or other suitably trained personnel. Only spare
parts approved by Hermann Sewerin GmbH may be used when
performing repairs.
● Use only suitable battery types, otherwise the device will not
be explosion-proof.
● Changes or modications to this product may only be carried
out with the approval of Hermann Sewerin GmbH.
● Use only Hermann Sewerin GmbH accessories for the product.
Hermann Sewerin GmbH shall not be liable for damages resulting
from the non-observance of this information. The warranty con-
ditions of the General Terms and Conditions (AGB) of Hermann
Sewerin GmbH are not aected by this information.
In addition to the warnings and other information in these Operating Instructions, always observe the generally applicable safety
and accident prevention regulations.
The manufacturer reserves the right to make technical changes.
1 General
1
Page 10
1 General
1.2 Purpose
The VARIOTEC 480 EX and the models 460 EX, 450 EX and
400 EX are hand-held measuring devices which can be used for
all gas pipeline testing applications.
The devices are designed for professional industrial use and
require the necessary specialist knowledge for working in gas
pipelines. Example applications are described in /3/.
Models 480 EX and 460 EX can be optionally tted with a detector
for ethane analysis to help you safely distinguish between natural
gas and swamp gas.
All devices can also be individually tted with electrochemical
sensors.
Catalytic combustion sensors operate on the catalytic combustion
principle. Thermal conductivity sensors measure the specic conductivity of gases. Electrochemical sensors operate on the electrochemical cell principle. The gas-sensitive semiconductor reacts
to changes in conductivity brought about by reducible gases.
Note:
These operating instructions describe the VARIOTEC 480 EX with
all additional equipment (rmware version 1.XXX). All descriptions
refer to the device as delivered (factory settings) and apply to all
models. The manufacturer reserves the right to make changes.
2
Page 11
1.3 Intended use
This device is intended for professional residential and commercial
use, in small rms and commercial operations and in industry. The
appropriate specialist knowledge is required to operate the device.
The device may only be used to measure the following gases
(depending on the device model and additional equipment):
● Methane CH
● Oxygen O
● Carbon monoxide CO
The device must not be used for:
● Gas analysis of technical processes
● Monitoring liquids
The device can be used up to a temperature of 40 ºC. However,
high temperatures reduce the lifetime of the sensors and rechargeable batteries.
If a device with an electrochemical sensor is exposed to gas
concentrations above the measuring range limit, this can reduce
the lifetime of the sensor.
/ propane C3H8 / butane C4H
4
2
1 General
10
3
Page 12
1 General
1.4 General safety information
● The device has been tested to ensure that it is explosion-proof
in accordance with European standards (CENELEC).
● The device must only be switched on with fresh air.
● Do not use this device in oxygen-enriched atmospheres, oth-
erwise it will not be explosion-proof.
● Only probe hoses with a hydrophobic lter may be used.
Exception:
If the probe has a built-in hydrophobic lter, the hose does not
require any other lters.
● The device must only be tested and adjusted with test gases
in well ventilated rooms or in the open air. Test gases must be
handled in a professional manner.
● Always carry out a device inspection after the device has suf-
fered an impact (for example, if dropped accidentally) (see
Section 5.1).
● Always carry out a device inspection (see Section 5.1) after
the device has been exposed to gas concentrations above the
measuring range limit of the LEL range.
● The device complies with the limits of the EMC directive. Always
observe the information in the manuals of (mobile) radio equipment when using the device close to (mobile) radio equipment.
Note:
Follow the advice regarding explosion protection (see Sec-
tion 2.3).
4
Page 13
1.5 Allocation of tasks to applications
Refer to the table below to help you select the appropriate appli-
cation for a given activity (according to /3/).
1 General
Gas measuring
● Measuring the gas concentration
Inspection above
ground
– Ground
– Gas line
commissioning or decommissioning gas systems
– Purge to verify gas purity or absence of gas, for example, when
– Possible leakage points
● Distinguishing between natural gas and swamp gasEthane analysis
Warning ExTox
Warning %LEL
area monitoring
● Warning against explosive gas concentrations through work
Measuring in bar
holes
– Determine gas dispersion (detection limit)
– Classify leaks
– Locate a probable gas escape (repair point)
monitoring
● Warning against explosive and toxic gases through work area
– Prevent possible dangers
● Measuring the gas concentrationEnclosed spaces
House
● Locating the source of gas
● Finding leaks at internal connections
LocationActivityApplication
Gas lines, gas systems,
…
Buried gas lines● Measuring minimal gas concentrations above:
Gas lines, gas systems,
… in houses, enclosed
spaces and shafts
In the ground● Measuring the gas concentration to:
In enclosed spaces
and shafts with an in-
creased potential of gas
dispersal
In the house● Measuring very low gas concentrations
5
Page 14
2 Features
2 Features
The device comes in four models:
VARIOTEC 480 EX
VARIOTEC 460 EX
VARIOTEC 450 EX
VARIOTEC 400 EX
The models are suitable for the following applications:
Application480 EX 460 EX 450 EX 400 EX
Inspection above ground×××
Measuring in bar holes
O
2
Enclosed spaces××
House××
Gas measuring×××
Warning %LEL××
Warning ExTox
CO
O
2
Ethane analysis○○
2.1 Visual and audible signals
The device features two alarms:
● Signal light on top of device (visual signal)
● Buzzer on side of device (audible signal)
The signals indicate alarms and faults. The device also emits
signals when it is switched on and o.
×
○
×
○
○
× standard○ Optional
×
○
×
○
×
○
○
6
Page 15
2 Features
If this symbol appears on the display, the audible signal
can be switched o.
When an audible signal has been switched o it cannot
be switched back on while the concentration level remains above the alarm threshold.
This symbol appears at the top left of the display as soon
as the audible signal has been switched o. It disappears
automatically if the level falls below the alarm threshold.
Operating signal
When using the Warning %LEL and Warning ExTox applications,
the device emits a visual and audible signal at regular intervals.
This indicates that the device is working properly.
Alarm
The device can monitor several gases at the same time. If the
measured gas concentration of one or more gases exceeds spec-
ied limit values (alarm thresholds) the device gives a warning. It
emits both audible and visual signals, which are distinctly dierent
from the operating signal.
WARNING! Danger of death due to hazardous gas
A
There is detailed information on alarms in Section 7.2.
concentrations
The alarms AL1, AL2 and AL3 always indicate danger.
Alarm AL4 indicates the possibility of danger.
● Take all necessary measures for your own safety and
the safety of others immediately.
7
Page 16
2 Features
2.2 Sensors
The device features four types of sensor:
● Gas-sensitive semiconductor (SC)
● Catalytic combustion sensor (CC)
● Thermal conductivity sensor (TC)
● Electrochemical sensor (EC)
ApplicationGasMeasuring rangeSensors
Inspection above
ground
Measuring in bar
holes
CH
CH
1 ppm – 10 % vol.SC, TC
4
0.0 – 100 % vol.TC
4
Enclosed spacesCH
HouseCH
CO0 – 500 ppmEC
Gas measuringCH
Warning %LEL and
Warning ExTox
CH
O
CO0 – 500 ppmEC
Ethane analysisCH
C
C3H
1 ppm – 100 % vol. SC, TC
4
1 ppm – 100 % vol. SC, TC
4
0.0 – 100 % vol.TC
4
0 – 100 % LELCC
4
0 – 25 % vol.EC
2
---Gas chroma-
4
tograph, SC,
2H6
8
TC
8
Page 17
2.3 Explosion protection
2.3.1 Passive explosion protection
The device is assigned to the following explosion-proof groups:
2 Features
Explosion-proof
group
II2G Ex d e ib IIB T4 Gb – Methane CH
II2G Ex d e ib IIC T4 Gb – Methane CH
For the following atmospheres
– Propane C3H
– Butane C4H
4
8
10
– Carbon monoxide CO
– Propane C3H
– Butane C4H
4
8
10
When using
Device
without
carrying
bag TG8
Device with
carrying
bag TG8
– Carbon monoxide CO
– Hydrogen H
2
EC type-examination certicate: TÜV 07 ATEX 553353 X
DANGER! Risk of explosion due to sparks
A
● Only open the battery compartment outside of explo-
sive areas.
● Only charge the device outside of explosive areas.
● Only use the USB port outside of explosive areas.
● Use only suitable battery types.
● When working with hydrogen, always use the carrying
bag TG8 for the device.
9
Page 18
2 Features
2.3.2 Active explosion protection
The functional safety test applies to:
Applications:Warning %LEL
Gas types:Measuring range:
– Methane CH
– Propane C
4
3H8
Gases:Measuring range:As per:
– Oxygen O
2
– Carbon monoxide CO0 – 500 ppm CO/5/
Tested accessories: – Test set SPE VOL
Type examination
Warning ExTox
0 – 100 % LEL
0 – 100 % LEL
0 – 25 % O
2
– Flexible hand probe, 1 m
– Floating probe 2 m, 6 m
/7/
10
Testing institute:DEKRA EXAM GmbH
Certicates:PFG 08 G 002 X N1
BVS 09 ATEX G 001 X N1
The following points were not part of the type examination:
● Saving measurement data (see Section 3.2.12)
● Saving protocols from the integrated device inspection (see
Section 5.1.1.4)
● Disposable alkaline batteries for the power supply (see Sec-
tion 4.1)
Page 19
3 Operation
3.1 General information on operation
3.1.1 Keys and jog dial
The ON/OFF key is the only control on the device that does not
change its function.
When switched on, the device is operated using the jog dial and
function keys to navigate the display.
ControlActionFunction
ON/OFF keyPress● Switches the device on
● Switches the device o
Function keys
F1, F2, F3
Jog dialTurn● Selects functions, settings,
Press● Variable
● As indicated on the display at
the bottom of the screen
● Function keys may also have
no function assigned in some
cases
measurement data, etc.
● Modies values
Press● Opens the next program lev-
el (e.g. menu item, function,
measurement data, selectable
values)
● Accepts values
3 Operation
11
Page 20
3 Operation
3.1.2 Selecting/exiting menus and menu items
Functions, applications and settings etc. are selected via the
main menu (for short: Menu). This menu has submenus and
menu items. Refer to Section 3.2.1 for information on accessing
the menu.
Selecting submenus/menu items
Submenus and menu items are selected and opened using the
jog dial and/or function keys.
In measuring mode the name of the selected application is in-
dicated by the symbol at the top left of the display. You can nd
detailed information on how to select or switch applications in
Section 3.1.4.
Exiting menus/menu items
There are generally two ways to exit open menus/menu items
and go back up a level:
● Press Esc
● Select Exit menu item
3.1.3 Switching the device on
Note:
Always switch the device on with fresh air.
1. Press the ON/OFF key. The device switches on.
The switching on process involves an internal check.
ProcessTest purpose
Buzzer emits audible
signal.
Signal light gives visual
signal.
Display is inverted.Are there pixels missing from the
12
Is the audible signal working?
Is the visual signal working?
display?
Page 21
3 Operation
The start screen appears on the display.
Display:
– Device type: VARIOTEC 480 EX
®
VARIOTEC 480 EX
– User:
Frank Smith
Frank Smith
City Council
Leakage Delivery
V1.200
Fig. 3: Start screen
22.04.2013
13:02
City Council
Leakage Delivery
– Firmware version: V1.200
– Date and time
– Capacity disposable battery/
rechargeable battery
The start screen for the application set under Autostart then
appears (see Section 3.3.4).
Display:
– Gas type: CH
4
– Application as symbol (top left)
and text:
Inspection above ground
– Measuring range:
0 ppm ... 10 % vol.
Fig. 4: Start screen for inspec-
tion above ground application
– Alarm threshold:
AL4, 3 ppm
– Symbol for next application that
can be selected via function
key F3: Measuring in bar holes
The device switches to measuring mode. The device warms
up. The reading ashes.
While the device is warming up, the prompt Add fresh air! is
displayed as a reminder.
2. Make sure the device is actually drawing in fresh air. Change
its location if necessary.
3. Wait until the reading stops ashing.
The device is ready for use.
13
Page 22
3 Operation
0
1010010
Fig. 5: Inspection above ground measuring mode
A
3.1.4 Selecting/switching application
A
0,1
1,0
PPM
CH4
0
WARNING! Danger of death due to incorrectly adjusted or faulty devices
Gas warning instruments must be inspected before use
at regular intervals.
● Carry out a device inspection every day before starting
work.
WARNING! Danger of death due to incorrectly adjusted or faulty devices
If the device is used as a gas warning instrument (Warning %LEL and Warning ExTox applications), it must be
regularly inspected before use.
● Carry out a device inspection every day before starting
work.
Display:
– Current reading: zero when
device is switched on with fresh
air
14
Note:
You may only switch applications when the device is drawing in
fresh air.
The current application is indicated by the symbol at the top left
of the display. The symbol at the bottom right shows the next application that can be selected via function key F3. The Settings
Page 23
3 Operation
under System (see Section 3.3.4) are used to specify which ap-
plication is rst activated when you switch the device on.
● Press Menu. Select the menu item for the application you
want to use.
OR
a) Press function key F3. The device switches to the next ap-
plication.
b) Repeat until the symbol for the application you want to use
appears at the top left.
3.1.5 Dierences between measuring mode and settings mode
The device is operated in two modes:
● Measuring mode (see Section 3.2)
Measurements are taken in measuring mode. All functions
needed to take readings can be accessed from one menu.
● Settings (see Section 3.3)
The device settings can be changed in settings mode. Information about the device can also be retrieved. Measurements
cannot be taken in settings mode.
Settings are accessed via the menu in measuring mode. The
settings are access-protected by a PIN code.
A
WARNING!
Danger of death due to lack of alarm signal
The device only issues alarms in measuring mode. As
soon you access the menu, alarms are no longer triggered.
● Only change the settings outside of explosive areas and
away from toxic or low-oxygen atmospheres.
15
Page 24
3 Operation
3.2 Measuring mode
When switched on, the device is in measuring mode. The current
measurement values are always shown in measurement mode
(Fig. 5). Depending on the application, measurements will have
to be saved or started and then stopped (see Section 3.2.12).
WARNING!
A
Danger of death due to operating signal failure
When the device is used as a gas warning device (Warning %LEL and Warning ExTox applications), an operat-
ing signal must always sound in measuring mode. If the
operating signal fails, the device is not safe to use.
● Stop using the device immediately.
● Move away from explosive areas or toxic or low-oxygen
atmospheres immediately.
16
Page 25
3 Operation
Zero point
Inspection above ground
Measuring in bar holes
Ethane analysis
Enclosed spaces
House
Gas measuring
Warning %LEL
Warning ExTox
Settings
Start measurement
Protocol
Purge
Device inspection
Gas type CxHy
Device information
Exit
3.2.1 Accessing the menu (measuring mode menu structure)
In measuring mode F1 can be used to access the menu.
Fig. 6: Menu with submenus (menu items)
Once you have started a measurement, Start measurement in
the menu becomes Stop measurement. In some applications
this menu item is called Save measurement.
You can nd detailed information on starting, stopping and saving
measurements in Section 3.2.12.
Protocol does not appear in the menu until you save a protocol
for the rst time.
Purging only appears in the menu in the ethane analysis application.
Gas type CxHy only appears in the menu if the device is equipped
for at least one further gas type apart from methane CH4.
17
Page 26
3 Operation
3.2.2 Zero point
The zero point can be set manually in the Zero point menu item.
This is only necessary if the displayed fresh air measurement is
not zero after the end of the warm-up period.
GasContent in fresh air Correct zero point on device
0 % vol.0.0 % vol.
CH
4
CO0 ppm0 ppm
O
20.9 % vol.20.9 % vol.
2
The manual zero point setting is not saved. The zero point can be
corrected by adjustment as often as zero point deviations occur
(see Section 5.2).
Requirements for correct setting of the zero point
● Device has been switched on with fresh air.
● Device continues to draw in fresh air.
Setting zero point (manual zero point setting)
1. Press Menu.
2. Select Zero point menu item. The values are automatically
adjusted. The device returns to measuring mode.
18
The Setting zero point function can also be accessed using
the relevant symbol in the Inspection above ground, Enclosed
spaces and House applications.
Page 27
3.2.3 Inspection above ground
This menu item allows you to change the measuring mode to
Inspection above ground .
Range of use
– Measuring minimal gas concentrations above the ground, the
gas pipe or possible leakage points
Symbol
Unit
– ppm (parts per million)
– % vol.
Measuring range
Gas-sensitive semiconductor 0 to 10,000 ppm
Thermal conductivity sensor 1 to 10 % vol.
3 Operation
Measurement data display
0
1010010
0,1
1,0
– Digit, e.g.
0.90 % vol. CH
– Bar graph with quasi-logarith-
mic scale
AL4
0.90
Fig. 7: Inspection above ground measuring mode
VOL%
CH4
4
19
Page 28
3 Operation
3.2.4 Measuring in bar holes
This menu item allows you to change the measuring mode to
Measuring in bar holes.
Range of use
– Measuring gas dispersal in the ground
– Locating the possible leakage point and classifying the leak
Symbol
Unit
– % vol.
Measuring range
Thermal conductivity sensor 0.0 to 100 % vol.
20
Electrochemical sensor
(oxygen O
)
2
0.0 to 25.0 % vol.
Measurement data display
– Digit, e.g.
20.9 % vol. O
0.60 % vol. CH
– Bar display with quasilogarith-
mic scale (for CH4)
Fig. 8: Measuring in bar holes measuring mode
2
4
Page 29
3.2.5 Ethane analysis
Selecting Ethane analysis from the menu opens an overview
of the detectable gases. The device automatically returns to
measuring mode.
Note:
The ethane analysis cannot be started with the Ethane analysis
menu item.
Information about carrying out the ethane analysis can be found
in Section 3.2.5.3.
Range of use
– Distinguishing between natural gas and swamp gas
Symbol
Unit
3 Operation
– ppm (parts per million)
– % vol.
Measuring range
Thermal conductivity sensor 0.0 to 100 % vol.
Electrochemical sensor
(oxygen O
)
2
Gas chromatograph/
0.0 to 25.0 % vol.
1 ppm to 1.2 % vol.
gas-sensitive semiconductor
21
Page 30
3 Operation
Measurement data display
– Digit, e.g.
20.9 % vol. O
9.0 % vol. CXH
– Bar display with quasilogarith-
mic scale (for CxHy)
Fig. 9: Ethane analysis measuring mode
3.2.5.1 General information on ethane analysis
Ethane analysis is used to demonstrate the presence of natural
gas and to distinguish between natural gas and swamp gas.
This analysis utilises the fact that natural gas contains ethane,
but swamp gas does not. Thus, if ethane is detected, it can be
concluded that natural gas is present.
Note:
Ask the network operator about the composition of natural gas.
Request an analysis certicate.
2
Y
22
Ethane can denitely be detected in natural gas at a minimum
concentration of 1 % vol. If the concentration of ethane in natural
gas is less than 0.5 % vol., the gas cannot be detected. If the
concentration falls between these two values, the analysis can
be carried out, but the result must be assessed subsequently.
During the ethane analysis, the gas sample is always analysed
for the presence of the following three gases:
● Methane CH
● Ethane C2H
● Propane C3H
4
6
8
Page 31
3 Operation
Requirement
The ethane analysis only works if the gas sample exhibits a spe-
cic concentration.
The device checks the concentration at the start of the analysis
and prevents the analysis being carried out if the concentration
is too low.
Gas sample
concentration
> 1 % vol.Denitely
Analysis
is …
Device response
Analysis can be carried out
possible
1 % vol. – 0.5 % vol.Technically
possible
Analysis can be carried out,
but Ethane analysis criti-cal message appears
< 0.5 % vol.Not
possible
Analysis cannot be carried
out because Take sample
symbol is not displayed
Displayed measurement values
In measuring mode, the device displays the actual concentra-
tions (Fig. 9).
During the analysis, the device automatically dilutes the gas sample to 1 % vol. to ensure optimal utilisation of the semiconductor
sensor’s measuring range. This means:
● The concentration ratios of all components of the gas sample
are retained.
● The maximum concentration of a gas component is 1 % vol.,
even if the actual concentration in the gas sample is higher.
In the graph, the maximum peak may be cut o as necessary
(Fig. 12 right).
Number of measurements
Always carry out 2 – 3 measurements in the same location to
obtain certain results. Note the information on evaluation in Section 3.2.5.3
23
Page 32
3 Operation
3.2.5.2 Purging the detector
The detector for ethane analysis must be kept clean at all times
to prevent distortion of measurement results. When the device
is in use, however, higher hydrocarbons (e.g. propane, butane)
can accumulate in the detector and contaminate it.
The detector is automatically purged after every ethane analysis.
An additional purge of the detector is required in these situations:
● Cancelled ethane analysis
If an ethane analysis is cancelled, the message Purging re-quired and the Take sample symbol appear automatically.
● Detector contamination suspected
Note:
The purge always takes 4 minutes. A new ethane analysis can
only be started when the purge cycle has been completed.
The device is in measuring mode.
1. Only when detector contamination is suspected:
a) Press Menu.
b) Select Purging from the menu. The message Purging re-
quired and the Take sample symbol appear.
24
NOTICE! Risk of damaging the sensor
A
2. Make sure the device is drawing in fresh air.
3. Press Take sample.
After the purge cycle is complete, the device will revert to measuring mode.
The ethane detector will be damaged if the air supply is
contaminated or moist.
● Always add clean, dry fresh air.
The purge cycle starts automatically as soon as the gas concentration drops below 50 ppm. The time remaining until the
end of the purge cycle is displayed.
Page 33
3.2.5.3 Carrying out an ethane analysis
The device is in measuring mode.
1. Make sure the device is drawing in fresh air.
2. Change to the Ethane analysis application.
3. The ethane analysis is generally performed on a bar hole. Use
the localisation probe and a probe hose.
a) Insert the probe into the bar hole.
b) Connect the probe hose to the device.
As soon as the measurement value satises specic conditions
(gas concentration > 1% vol., stable measurement value), the
Take sample symbol appears.
4. Press Take sample. A message appears.
Note:
An ethane analysis takes approximately four minutes. After Esc
is used to cancel the analysis, the detector must be purged (see
Section 3.2.5.2). This purge cycle also takes 4 minutes.
3 Operation
5. Conrm the Take sample? prompt by pressing OK. The gas
sample is taken.
The Addfresh air note appears.
6. Add fresh air.
a) To do so, remove the probe hose from the device.
b) Move away from the bar hole. Note the direction of the wind.
As soon as the gas concentration drops below 50 ppm, the
analysis of the gas sample taken starts automatically. The gas
sample analysis is plotted on the display.
25
Page 34
3 Operation
150 200050 100
C3H8
Analysis
Yes
CH4
Yes
C2H6
Yes
C3H8
250 s
1,0
CH4
C2H6
0,1
100
10
088 Seconds
Fig. 10: Graph of an analysis in progress
When the analysis is complete, the Save symbol appears.
7. Press Save.
8. If necessary enter a Comment on the analysis.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored comments. A list of the stored com-
ments will appear.
− Select the desired comment. Open the comment with OK.
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter a comment.
9. Conrm your entry with OK. The comment is saved together
with the protocol name (date, time).
3.2.5.4 Evaluating an ethane analysis
Fully completed ethane analyses are saved as protocols. These
can be accessed and cleared at any time.
26
Page 35
3 Operation
CH4
1,0
0,1
100
10
15.09.2008
11:21 12/47
Analysis
Yes
CH4
C2H6 ????
C3H8 ----
150 200050 100
250 s
Fig. 11: Ethane analysis protocol
A protocol contains the following
information about the gas sample:
– Analysis of the gas compo-
nents:
CH4 Yes
C2H6 ????
C3H8 – – – –
– Curve; peaks of the gas com-
ponents denitely present in
the sample are labelled: CH
4
– Date and time the protocol was
saved
– Protocol number: 12/47, i.e.
12th protocol of 47 protocols in
total
The symbols next to the analysed gas components are dened
as follows:
SymbolGas component is …
YesDenitely present
????Possibly present
– – – –Not present
Ideally, the result of the ethane analysis is a curve with at least
one clear, steep peak (Fig. 12 left). However, occasionally, the
analysis does not provide a clear determination of the gas com-
ponents present (Fig. 12 right). In such cases, you must decide
whether the analysis quality is sucient.
CH4
1,0
C2H6
0,1
100
10
15.09.2008
11:35 13/47
C3H8
150 200050 100
Analysis
Yes
CH4
Yes
C2H6
Yes
C3H8
250 s
Fig. 12: Quality assessment of an ethane analysis;
left: good quality, right: poor quality
CH4
1,0
0,1
100
10
15.09.2008
11:57 14/47
Analysis
Yes
CH4
C2H6 ????
C3H8 ----
150 200050 100
250 s
27
Page 36
3 Operation
Criteria for determining good analysis quality
Criteria for determining the presence of natural gas
Ethane analyses of insucient quality
Note:
Never use poor-quality ethane analyses to demonstrate the presence of natural gas.
Poor-quality ethane analyses can exhibit the following features,
for example:
● Methane is denitely present.
● Methane concentration is 1 % vol.
● Peaks of the gas components present are clearly recognisable.
● Good analysis quality (see above)
● Ethane is denitely present.
● Curve does not contain any clear peaks (smooth concentra-
tion plot)
28
Reason:Residual gases from previous analyses have
accumulated.
Corrective
action:
Purge with test gas, e.g.:
– 1 % vol. CH
in synthetic air
4
OR
– 100 ppm C2H6 and 1 % vol. CH4 in synthetic air
● Curve and symbols (Analysis) provide contradictory informa-
tion about gas components
Reason:Extremely high or low temperatures
(optimal working temperature: 20 °C)
Page 37
3.2.6 Enclosed spaces
This menu item allows you to change the measuring mode to
Enclosed spaces.
Range of use
– Measuring gas concentrations in enclosed spaces where
there is increased potential of gas dispersal
Symbol
Unit
– ppm (parts per million)
– % vol.
Measuring range
Gas-sensitive semiconductor 0 to 10,000 ppm
Thermal conductivity sensor 0.1 to 100 % vol.
3 Operation
Measurement data display
0,1
100
0
10
10
1,0
100
– Digit, e.g.
8 ppm CH
– Bar graph with quasi-logarith-
mic scale
PPM
8
CH4
Fig. 13: Enclosed spaces measuring mode
4
29
Page 38
3 Operation
3.2.7 House
This menu item allows you to change the measuring mode to
House.
Range of use
– Measuring minimal gas concentrations in buildings
– Locating the source of gas
Symbol
Unit
– ppm (parts per million)
– % vol.
Measuring range
Gas-sensitive semiconductor0 to 10,000 ppm
Thermal conductivity sensor0.1 to 100 % vol.
30
Electrochemical sensor
(carbon monoxide CO)
Measurement data display
Fig. 14: House measuring mode
0 to 500 ppm
– Digit, e.g.
0 ppm CO
4 ppm CH
4
– Bar display with quasilogarith-
mic scale (for CH4)
Page 39
3.2.8 Gas measuring
This menu item allows you to change the measuring mode to
Gas measuring.
WARNING!
A
Range of use
Symbol
Danger of death due to lack of alarm signal
In gas measuring, the device does not issue alarms.
● Before switching to this application, you must check that
the there are no ignition sources nearby.
● Do not smoke.
● Exercise extreme care.
– Demonstrating gas purity / absence of gas in gas pipes
3 Operation
Unit
– % vol.
Measuring range
Thermal conductivity sensor 0.0 to 100 % vol.
Measurement data display
– Digit, e.g.
30 % vol. CH
– Bar graph with quasi-logarith-
mic scale
VOL%
CH4
100
1,0
0
10
30
Fig. 15: Gas measuring mode
4
31
Page 40
3 Operation
3.2.9 Warning %LEL
This menu item allows you to change the measuring mode to
Warning %LEL.
WARNING!
A
Danger of death due to operating signal failure
If the operating signal fails, the device is not safe to use.
● Stop using the device immediately.
● Move away from explosive areas or toxic or low-oxygen
atmospheres immediately.
Range of use
– Testing work environments where explosion is possible, e.g.
working on gas pipes or gas systems
Symbol
32
Unit
– % LEL or % vol.
Measuring range (methane CH
)
4
Catalytic combustion sensor 0 to 100 % LEL or
0.00 to 4.40 % vol.
Measurement data display
010
50
30
40
70
80
20
60
90 100
– Digit, e.g.
14 % LEL CH
4
– Bar display with linear scale,
alarm thresholds marked AL1
AL1
%LEL
14
Fig. 16: Warning %LEL measuring mode
CH4
and AL2
Page 41
3.2.10 Warning ExTox
This menu item allows you to change the measuring mode to
Warning ExTox.
WARNING!
A
Range of use
Symbol
Danger of death due to operating signal failure
If the operating signal fails, the device is not safe to use.
● Stop using the device immediately.
● Move away from explosive areas or toxic or low-oxygen
atmospheres immediately.
– Warning of explosive and toxic gas concentrations as well as
If you switch from Warning ExTox to another application,
data which may have triggered an STEL or LTEL alarm
is reset.
● Only switch from Warning ExTox to another application
once you have left the work area being monitored.
34
Page 43
3.2.11 Settings
Settings in the menu allows you to change the device settings
and view information about the device (see Section 3.3).
3.2.12 Start/stop/save a measurement
Depending on the application, measurements will have to be
saved or started and then stopped. The ethane analysis is an
exception (see Section 3.2.5.3).
3 Operation
Application
start/stopsave
Measurement
Inspection above ground×
Measuring in bar holes×
Enclosed spaces×
House×
Gas measuring×
Warning %LEL×
Warning ExTox×
Dierence between starting/stopping and saving
Selecting Start measurement followed by Stop measurement
saves a measurement plot.
Selecting Save measurement saves an individual measurement,
the current one.
Note:
Measurements cannot be cancelled. The only way to cancel a
measurement is to stop it.
Up to 80 measurements can be saved.
The measured values can be saved with or without a comment.
Comment entries are saved automatically (ring memory with max.
10 entries).
35
Page 44
3 Operation
The stored measurements can be displayed on a computer using
a readout program. The program is available at www.sewerin.
com.
Starting a measurement
1. Press Menu.
2. Select Start measurement from the menu. This starts meas-
Measurement plot recording must always be concluded with Stop
measurement.
Stopping a measurement
1. Press Stop measurement.
2. Answer Yes to the warning prompt.
3. Enter a comment for the measurement.
Once the rst comment has been entered, the Open
stored comments function will become available.
urement plot recording.
OR
a) Press Menu.
b) Select Stop measurement from the menu.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored comments. A list of the stored com-
ments will appear.
− Select the desired comment. Open the comment with OK.
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter a comment for the
measurement.
The measurement is saved as a protocol. The protocol name
is formed from the date, time and comment.
36
Page 45
Saving a measurement
1. Press Menu.
2. Select Save measurement from the menu.
3. Enter a comment for the measurement.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored comments. A list of the stored com-
ments will appear.
− Select the desired comment. Open the comment with OK.
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter a comment for the
measurement.
The measurement is saved as a protocol. The protocol name
is formed from the date, time and comment.
3.2.13 Protocols
You can retrieve or clear protocols of saved data under Protocol
in the menu. When saved, the protocols are assigned to dierent
protocol types.
The following protocol types are available:
● Ethane analysis
● Device inspection
● Measurements
Protocols can only be cleared individually.
You can nd information on how to clear all protocols in one pro-
tocol type in Section 3.3.7.
3 Operation
37
Page 46
3 Operation
3.2.14 Device inspection
The device inspection can be used to check the general status
and the indication accuracies. Device inspection only appears
in the menu when the integrated device inspection is switched on.
Note:
The integrated device inspection is switched o in the factory
settings. More detailed information about the device inspection
can be found in Section 5.1.
The frequency of the device inspection depends on the applica-
tion (see Section 5.1.1.2).
If the integrated device inspection is switched on, the device will
remind you to perform a device inspection.
The Carry out device inspection symbol will appear
when the inspection is due. It is visible in the display
until the complete integrated device inspection has been
carried out successfully.
3.2.15 Gas type CxHy
You can temporarily change the gas type under Gas type CxHy
in the menu if the device is equipped for other gas types. Gas
types available for selection:
● Methane CH
● Propane C3H
● Butane C4H
10
The temporary gas type change is not saved. If you need to set
a dierent default gas type, you can do this in the Settings menu
under System.
38
4
8
Page 47
3.2.16 Device information
The following device information is shown under Device information in the menu:
● Service: date of the last service, date of the next service
3.3 Settings
The following menus and menu items are included under Settings:
● Adjustment
● System
● Alarms
● Date/time
● Memory
You can nd information on selecting and exiting menus and menu
items in Section 3.1.2.
3 Operation
3.3.1 Opening Settings
1. Press Menu.
WARNING!
A
2. Select Settings from the menu.
Danger of death due to lack of alarm signal
The device only issues alarms in measuring mode. As
soon you access the menu, alarms are no longer triggered.
● Only change the settings outside of explosive areas and
away from toxic or low-oxygen atmospheres.
Access is protected by a PIN code. The default setting is
always PIN code 0001.
39
Page 48
3 Operation
Note:
You can change the PIN code at any time.
SEWERIN recommends setting a dierent PIN code after initial
start-up, so only authorised personnel have access to the settings.
3. Enter the PIN code from left to right. The active digit is always
displayed with a black background.
DigitTo changeTo conrm
1st digit
Press the jog dial
2nd digitPress the jog dial
3rd digitPress the jog dial
Turn the jog dial
4th digit
If the PIN code has been entered correctly, the Settings menu
will appear once the last digit has been conrmed (Fig. 18).
Otherwise the device will revert to measuring mode.
40
Adjustment
System
Alarms
Date/time
Memory
Exit
Fig. 18: Settings menu
Page 49
3.3.2 Settings menu structure
PIN Code
SettingsAdjustmentAdjustment CxHy PPM
Adjustment CxHy
Adjustment O2
Adjustment CO
Adjustment gas mixture
Test gas
Inspection OK
Exit
SystemPIN Code
Service interval
Display
Battery
Autostart
Gas type CxHy
Unit %LEL
PPM multiplicator
Device inspection
Reset
Language
Exit
Alarms
Date/time
MemoryClear
Interval
Memory mode
Exit
Exit
Measuring mode
3 Operation
Fig. 19: Settings menu structure for VARIOTEC 480 EX
(gas type: methane)
Note:
The number and names of available menu items depend on the
device model and optional additional equipment.
41
Page 50
3 Operation
3.3.3 Adjustment
The Adjustment menu is used to set the sensors.
WARNING! Danger of death due to incorrect adjust-
A
A detailed description of adjustment along with important information is provided in Section 5.2.
Adjustment CxHy ppm
Used to adjust the gas-sensitive semiconductor for gas type C
(methane CH4, propane C3H8, butane C4H10) in the ppm range.
Applications: – Inspection above ground
ment
Incorrect adjustment can lead to incorrect measurement
results. This means that the user may not be warned about
dangerous gas concentrations in time.
● Only specialist technicians may perform adjustments
● Adjustments must be made in well ventilated rooms or
in the open air.
– Ethane analysis
– Enclosed spaces
– House
xHy
42
Adjustment CxHy
Used to adjust the catalytic combustion and thermal conductivity
sensors for gas type C
C4H10) in the % vol. range / LEL range.
Applications: – Inspection above ground
(methane CH4, propane C3H8, butane
xHy
– Measuring in bar holes
– Ethane analysis
– Enclosed spaces
– House
– Gas measuring
– Warning %LEL
– Warning ExTox
Page 51
3 Operation
Adjustment O2
Used to adjust the electrochemical sensor for oxygen O
in the
2
% vol. range.
Applications: – Measuring in bar holes
– Ethane analysis
– Warning ExTox
Adjustment CO
Used to adjust the electrochemical sensor for carbon monoxide CO in the ppm range.
Application: – Warning ExTox
Adjustment gas mixture
Used to adjust the catalytic combustion and electrochemical sen-
sors for all components of the test gas (gas mixture).
Applications: – Warning %LEL
– Warning ExTox
Test gas
Used to adjust the concentration of the test gases used.
Inspection OK
Conrms the device is in proper working order. This extends the
service interval.
43
Page 52
3 Operation
3.3.4 System
General information and specications for operation are set in
the System menu.
PIN code
Used to change or reset the PIN code.
Note:
If you lose the PIN code, you must contact SEWERIN Service.
If the PIN code is set to 0000, you will not be asked to enter it.
The settings can then be accessed by anyone.
Service interval
Species the regular inspections/maintenance required for the
device. You can also activate the automatic switch-o function
once the set interval has passed.
Display
Used to set how long the display remains illuminated after any
key is pressed as well as the display contrast.
44
Battery
Used to set the type of disposable/rechargeable battery used.
CAUTION! Damage possible due to device overheating
If the battery type is not correctly set, the device can overheat.
● Always enter the correct battery type.
Page 53
3 Operation
Autostart
Sets the application that is automatically activated when the device is switched on.
Gas type CxHy
Sets the gas type (methane CH
, propane C3H8, butane C4H10)
4
which is automatically used when the device is switched on.
Unit %LEL
Used to set the unit of measurement. Only applies to Warning
%LEL and Warning ExTox applications.
PPM multiplicator
Sets the amplication factor for the lower ppm range.
Measurement x factor = displayed measurement
Application: – Inspection above ground
Device inspection
Used to switch the integrated device inspection on or o.
Reset
Used to reset the device settings to the factory settings.
Language
Sets the language.
45
Page 54
3 Operation
3.3.5 Alarms
Sets the alarm thresholds for the gas types and gases.
Alarm AL3 cannot be set. It always occurs at the end of the
measuring range.
There is detailed information on alarms in Section 7.2.
AL1 alarm
Used to set the pre-alarm.
Applications: – Enclosed spaces
AL2 alarm
Used to set the main alarm.
Applications: – Enclosed spaces
– House
– Warning %LEL
– Warning ExTox
– House
– Warning %LEL
– Warning ExTox
AL4 alarm
Used to set the alarm threshold in excess of signicant gas concentrations which indicate a gas leak.
Application: – Inspection above ground
3.3.6 Date/time
Used to set the time, day, month and year. There are two formats
available for the date.
3.3.7 Memory
The Memory menu is used to specify how measurement data
and protocols are handled.
46
Page 55
3 Operation
Clear
Used to clear protocols.
The dierent protocol types must each be cleared separately. All
protocols in one protocol type are cleared at once.
You can nd information on clearing individual protocols in Sec-
tion 3.2.13.
Interval
Sets the interval at which measurement data is automatically
saved.
Leaking electrolytes can shorten the creepage distance
and air gap between the poles. As a result, the requirements for the batteries may no longer be met.
● Replace leaking batteries immediately.
● Clean the battery compartment (and, if necessary, the device) before inserting the new disposable/rechargeable batteries.
48
A
WARNING! Risk of explosion due to unsuitable batteries
To ensure that the device remains explosion-proof as per
/16/, only certain disposable/rechargeable batteries may
be used.
● Only use batteries supplied by SEWERIN. Other dis-
posable/rechargeable batteries, which have not been
supplied by SEWERIN, may only be used if they meet
the specications in /8/.
● In each battery compartment use only batteries that are
identical with respect to type (disposable or recharge-
able), capacity and manufacturer.
Page 57
4 Power supply
Disposable battery requirements
● Alkaline disposable batteries
● Size: AA, type: LR6 as per /11/
● The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with/8/.
Rechargeable battery requirements
● NiMH rechargeable batteries
● Size: AA, type: HR6 as per /13/
● The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with /8/.
● The rechargeable batteries must be fast charging (I > 1.25 A)
and remain within the temperature range.
Note:
A device operated with disposable alkaline batteries cannot be
charged. A note to this eect is shown on the display.
4.2 Operation with rechargeable batteries
The operating time of the device depends on the battery capacity.
If the device is not used or not kept in the docking station, the
batteries will lose their charge due to self-discharge. The self-discharge intensity depends on the battery type.
4.2.1 Charging
The device can be charged via:
● Connection for power supply
● Docking station TG8
49
Page 58
4 Power supply
DANGER! Risk of explosion due to sparks
A
For charging you will need either:
Please note the following points:
4.2.2 Rechargeable battery maintenance
If the device is not used for a long period of time, it is advisable
to fully discharge the battery before recharging it again.
A full discharging and recharging process takes approx. 11 hours
(8 hours to discharge + 3 hours to recharge). The duration depends on the capacity of the rechargeable batteries used.
High charging current occurs when batteries are being
charged.
The power supply is not explosion-proof.
● Only charge the device outside of explosive areas.
● AC/DC adapter M4
● Vehicle cable M4
● The device or docking station must not be directly connected
to a 24-V on-board power supply in the vehicle. The voltage is
too high for the charging process.
● The battery should be charged at approximately room tem-
perature.
50
DANGER! Risk of explosion due to sparks
A
High charging current occurs when batteries are being
charged.
The power supply is not explosion-proof.
● Only charge the device outside of explosive areas.
● Connect the device (switched on) to the power supply via the
side connection
OR
Place the device (switched on) into the docking station.
Page 59
The rechargeable batteries will be fully discharged. Once the
device has been discharged, it will automatically switch to
charging mode.
4.3 Battery alarm
As soon as the remaining capacity of the batteries gets low, a
battery alarm will go o:
Level 1:Battery almost empty
Level 2:Battery empty
4 Power supply
– Capacity disposable battery/rechargeable
battery symbol ashes
– Audible signal (one-o)
– Operating signal doubles
– Remaining operating time: approx. 15 min
A 2.5 mm Allen key (supplied) is required to open the battery
compartment on the back of the device.
1. Loosen the two screws securing the battery compartment. Re-
2. Lift out the battery compartment.
3. Remove the disposable/rechargeable batteries and insert new
When the housing is open, the device is not explosion-proof.
● Only open the battery compartment outside of explo-
sive areas.
move the screws by repeatedly turning them alternately a short
way; this ensures that the battery compartment does not twist.
ones. Ensure that the batteries are inserted with the correct
polarity.
51
Page 60
4 Power supply
4. Replace the battery compartment so it ts neatly into place
and secure rmly with the screws.
5. When you switch the device back on again, you will be asked
which battery type is in use. Enter the correct battery type.
If it takes longer than 120 seconds to replace the batteries, the
date and time will have to be reset the next time you switch the
device on. All the other data will be saved.
52
Page 61
5 Maintenance
In accordance with the legal regulations, device maintenance
comprises the following elements:
● Device inspection including test of indication accuracy
● Adjustment
● Maintenance
All inspections must be documented. The documentation must
be retained for at least one year.
WARNING! Danger of death due to incorrectly ad-
A
justed or faulty devices
If the device is used as a gas warning instrument (Warning %LEL and Warning ExTox applications), it must be
regularly inspected before use.
● Carry out a device inspection every day before starting
work (as per /1/, /2/, /6/, /10/).
5 Maintenance
5.1 Device inspection
5.1.1 General information on the device inspection
5.1.1.1 Scope
The device inspection includes the following tests:
● Analysis of the general status
● Test of the indication accuracy with supply of fresh air
● Test of the indication accuracy with supply of test gas
53
Page 62
5 Maintenance
5.1.1.2 Frequency
The frequency of the device inspection depends on the application.
ApplicationWhen to testLegal basis
Inspection above
ground
Measuring in bar
holes
Enclosed spaces
House
Gas measuring
Warning %LEL Every day before starting
Warning ExTox/1/, /2/, /6/,
Ethane analysisQuarterly
If the integrated device inspection is switched on, the device will
remind you to perform a device inspection.
The applications are grouped together for the device inspection.
The device inspection must be carried out separately for each
group.
Before starting work and if
/3/
out of use for long periods
Weekly to every six months /3/
/2/, /10/
work
/10/
54
Devices with multiple gas types
For devices that are designed for multiple gas types, the device
inspection must be carried out for the gas type currently being
used. The device inspection can also be carried out for other
available gas types.
Note:
If the device inspection symbol appears after changing the gas
type, the device inspection must always be carried out immediately.
Page 63
5.1.1.3 Documentation
The device inspection procedure must be documented. There
are two ways of doing this:
● On paper
● Saved electronically supported by the device (integrated de-
vice inspection)
Only the integrated device inspection is described in these operating instructions.
Note:
If the integrated device inspection is switched o, the device inspection must be documented on paper.
5.1.1.4 Integrated device inspection
The integrated device inspection is accessed via the menu
(Fig. 6).
The results of the device inspection are stored in the device as
a protocol.
The device inspection protocols can be opened in the device at
any time (see Section 3.2.13). They can also be displayed on a
computer using a readout program. The program is available
at www.sewerin.com.
5 Maintenance
The Carry out device inspection symbol appears when
a device inspection is due. It is visible in the display until
the complete integrated device inspection has been carried out successfully for the selected application. If the
device inspection was completed but the device failed
on some points, the symbol will remain visible.
The integrated device inspection is switched o in the factory
settings. The integrated device inspection has to be switched on
(once only) before it can be performed.
55
Page 64
5 Maintenance
Switching on the integrated device inspection
1. Press Menu.
2. Select Settings.
3. Enter your PIN code.
4. Select System.
5. Select Device inspection.
6. Select Yes.
7. Accept the setting with OK.
8. Exit the settings with Exit.
5.1.1.5 Order
You can carry out the device inspections and the associated
tests for the applications (groups) that are due to be inspected
in any order you wish. You can repeat the tests as often as you
wish provided you have not yet concluded the device inspection
for a group.
5.1.1.6 Test gases for the device inspection
The following variables depend on the application:
● Number of tests with test gas
● Test gases to be used
56
Page 65
5 Maintenance
The following test gases can be used to check the indication accuracy when supplying test gas:
Application
(group)
Inspection above
ground
Bar holes/
measuring
House/SpacesTest gas CH41 % vol. CH
TestRecommended
test gases
Test gas 10 ppm
10 ppm CH
*)
4
in 10 seconds
Test gas CH4100 % vol. CH
Test gas C3H8100 % vol. C3H
Test gas CO40 ppm CO
Test gas C3H81 % vol. C3H
4
8
*)
4
*)
*)
8
WarningGas mixtureExTox IR
Test gas C3H81 % vol. C
Test gas C4H101 % vol. C4H
Ethane analysisTest gas
50/100 ppm C2H6
100 ppm C2H6 in
1 % vol. CH4
3H8
*)
*)
10
(Rest: synthetic air)
*)
in synthetic air
If a test can be carried out either with a gas mixture or with an
individual gas, SEWERIN recommends you use the gas mixture.
If a gas mixture is used, but the device is equipped for other gas
types and gases which are not contained in the gas mixture,
these must be additionally tested with the relevant individual gas.
Note:
Use of test gases not provided by SEWERIN can cause interference.
The concentration of the test gas used must match the specied
test gas concentration.
57
Page 66
5 Maintenance
Inspection above ground
Bar holes/measuring
Houses/Spaces
Warning
Ethane analysis
Changing the test gas concentration
If there is no test gas with the specied concentrations available
for the inspection, the values can be adjusted to the test gas
used in the adjustment menu under Test gas (see Section 3.3.3).
5.1.2 Carrying out the device inspection
5.1.2.1 Accessing the device inspection
The device is in measuring mode.
1. Press Device inspection.
OR
a) Press Menu.
b) Select Device inspection from the menu.
The Device inspection menu appears.
Fig. 20: Device inspection menu
2. Select an application (group).
The Dev. Test ... menu appears.
3. Select a test (under General status, Fresh air, Test gas …).
Note:
The number and names of the available menu items depend on
the device features.
58
Page 67
4. Carry out the test.
For detailed information, refer to the following sections:
− General status Section 5.1.3
− Fresh air Section 5.1.4
− Test gas … Section 5.1.5
5.1.2.2 Concluding the device inspection
After all the tests have been carried out as described in Section 5.1.3 to 5.1.5, the Save symbol will appear in the display.
An integrated device inspection is concluded by saving it. Up to
40 device inspections can be saved. The following information
can be stored along with the device inspection:
● Inspector (e.g. inspector's name or initials)
● Password to protect the protocol from being accessed by un-
authorised people
Inspector entries are saved automatically (ring memory with max.
10 entries).
Once the rst inspector has been entered, the Open
stored inspectors function will become available.
5 Maintenance
1. Press Save.
2. Enter the name of the inspector.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored inspectors. A list of the stored in-
spectors will appear.
− Select the desired inspector. Open the inspector with OK.
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter an inspector for the
device inspection.
59
Page 68
5 Maintenance
3. Enter a password.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
b) Then conrm your entry with OK.
OR
Press Esc if you do not wish to enter a password for the
device inspection.
The device inspection is saved as a protocol. An overview with
the device inspection results is displayed.
This overview includes a list of all gas types for which the de-
vice is congured. Gas types for which the indication accuracy
has been successfully tested as part of the device inspection
are marked with OK. Gas types that are available but have
not been tested are agged with ----.
4. Conrm the overview by pressing OK. The device returns to
measuring mode.
5.1.3 Testing the general status
The general status test is part of the device inspection (see Sec-
tion 5.1.1.1). It is based on estimations by the user. The following
must be tested:
● Housing
● Signals
● Probe
● Filter
● Pump
The battery charge status and the working condition of the controls
are automatically tested during the integrated device inspection.
60
Page 69
The device inspection has been opened.
1. Select General status from the Dev. Test ... menu.
2. Test all associated subitems as described in Section 5.1.3.1
to 5.1.3.5.
3. Conrm the prompt General status OK? by pressing Yes if
all subitems show no faults during testing. General status
OK appears on the display.
This concludes the General status test.
5.1.3.1 Housing
● Is the housing free from external damage?
5.1.3.2 Signals
During the integrated device inspection the signals are emitted
at short intervals.
● Can the audible signal be heard?
● Is the visual signal visible?
5 Maintenance
5.1.3.3 Probe
Probes are accessories. They only need to be tested if they are
likely to be used in the course of the working day.
● Are the probes free from external damage?
Probe hoses are tested with a simple leak check.
1. Connect the probe hose to the gas input.
2. Seal the free end of the probe hose.
An error message should appear after approx. 10 seconds.
This indicates that the probe hose is in good condition.
61
Page 70
5 Maintenance
5.1.3.4 Filter
The ne dust lter is located behind the gas input. It is tested by
means of a visual inspection.
1. Unscrew the gas input.
2. Remove the ne dust lter.
3. Check that there is no dirt in the ne dust lter.
As soon as there are any signs of deposits, the lter must be
replaced. If you do not replace the lter, you must reinsert it
exactly as you found it.
5.1.3.5 Pump
The pump function is tested with a simple leak check.
1. Seal the gas input.
After a maximum of 10 seconds an error message should appear. This indicates that the pump is working correctly.
If the error message does not appear, the pump may be faulty.
The device must be tested by SEWERIN Service.
2. Release the gas input again.
After approximately 5 seconds, the error message should
disappear again. Otherwise there is a fault (see Section 6).
5.1.4 Testing indication accuracy with supply of fresh air
The indication accuracy with supply of fresh air test is part of the
device inspection (see Section 5.1.1.1).
The device inspection has been opened.
1. Make sure that only fresh air is being drawn in.
2. Select Fresh air from the Dev. Test ... menu.
3. Wait until the displayed readings are stable. A Status: OK
message will appear.
4. Press OK to conrm. Fresh air OK will appear on the display.
This concludes the Fresh air test.
62
Page 71
5 Maintenance
If the Status: OK message does not appear within a reasonable
amount of time, the air inow does not correspond to the limit
values stored in the device (see Section 7.3). Move the device
to another location and repeat the test.
If the Status: OK message still does not appear when the test
is repeated, the device must be re-adjusted (see Section 5.2).
5.1.5 Testing indication accuracy with supply of test gas
The indication accuracy with supply of test gas test is part of the
device inspection.
All test gases specied in the device must be tested. The number
of specied test gases depends on the model and the optional
additional equipment.
The following resources are needed for the test:
● Test gas (e.g. gas mixture)
Information on test gases for the device inspection can be found
in Section 5.1.1.6.
● Test set for the supply of test gas (e.g. SPE VOL)
Note:
Details of how to use the test set can be found in the accompanying operating instructions.
The procedure for testing with a gas mixture and individual gas
is the same.
The device inspection has been opened.
1. On the Dev. Test … menu, select the menu item to be tested
(e.g. Gas mixture, Test gas C3H8, etc.).
2. Check whether the test gas concentration specied by the
device matches the test gas you intend to use. To do this
press Information.
63
Page 72
5 Maintenance
Note:
When testing the indication accuracy for Inspection above
ground using test gas 10 ppm in 10 seconds it is not possible
to check the test gas. Instead, a message will appear explaining
what steps need to be carried out.
3. Add the test gas
4. Wait until the displayed readings are stable. A Status: OK
5. Press OK to conrm.
6. Stop the test gas supply.
7. Repeat for further test gases if applicable.
Once all tests have been successfully completed, the complete
test of the indication accuracy with supply of test gas is also
complete.
message will appear.
The device switches back to the Dev. test... menu. A Test gas OK message will appear.
64
Test gas test unsuccessful
If a Test gas … test was not carried out successfully, the message
Test gas not OK appears.
A test may be unsuccessful for the following reasons:
CauseCorrective action
Connections leakingRepeat check, checking the
seal on the connections
Measurement values outside
the specied limit values
Adjustment required
(see Section 5.2)
(see Section 7.3)
Page 73
5.2 Adjustment
WARNING! Danger of death due to incorrect adjust-
A
5.2.1 Scope
Adjustments must be made separately for each measuring range.
Note:
For each measuring range always adjust the zero point rst, followed by the sensitivity.
ment
Incorrect adjustment can lead to incorrect measurement
results. This means that the user may not be warned about
dangerous gas concentrations in time.
● Only specialist technicians may perform adjustments
● Adjustments must be made in well ventilated rooms or
in the open air.
● Zero point
● Sensitivity
5 Maintenance
When carrying out an adjustment, only adjust the gases whose
measurement values are outside the specied limits (see SSection 7.3).
65
Page 74
5 Maintenance
5.2.2 Test gases for the adjustment
The following test gases can be used for adjustment:
If more than one test gas can be used to adjust a gas, it does not
have to be adjusted with all test gases. However, adjusting with
more than one test gas increases the measurement quality.
Every gas must be adjusted separately, except for adjustment
with a gas mixture (see Section 5.2.3).
Replacement test gases are suitable for adjustment even though
they do not contain the gas to be tested, although the adjustment
error can be up to -15 % or +30 %.
Fresh air contains 20.9 % vol. O
and is therefore used with ox-
2
ygen to adjust the sensitivity.
Propane C
and butane C4H10 can only be adjusted with indi-
3H8
vidual gas.
Page 75
5 Maintenance
Note:
Use of test gases not provided by SEWERIN can cause interference.
The concentration of the test gas used must match the specied
test gas concentration.
5.2.3 Special features of adjustment with gas mixture
If you are using a SEWERIN gas mixture as the test gas, the following gases can be adjusted in a single step via Adjustment gas mixture:
● Methane CH
● Oxygen O
4
2
● Carbon monoxide CO
SEWERIN recommends a test gas with 2.2 % vol. CH
CO2, 40 ppm CO and 17.5 % vol. O2 as the gas mixture.
Zero point
The zero point is adjusted under Fresh air in the menu. The
following parameters are adjusted in a single step:
● Zero points of CH
● Sensitivity of O
and CO
4
2
, 2.0 % vol.
4
Sensitivity
The sensitivity is adjusted under Gas mixture in the menu. The
following parameters are adjusted in a single step:
● Sensitivities of CH
, CO2 and CO
4
67
Page 76
5 Maintenance
5.2.4 Preparation
An adjustment always requires time. Leave yourself plenty of time
to prepare the necessary steps of the procedure.
● Have all necessary tools available.
● Let the device run for several minutes to guarantee that the
temperature is correct, for example.
Note:
When adjusting the gas-sensitive semiconductor (Inspection
above ground and House/Spaces groups) a conditioner must
be applied.
● SEWERIN recommends applying separate conditioners for
methane CH
on the one hand and propane C3H8 / butane C4H10
4
on the other.
● The conditioner must already be connected during the run-
ning-in period.
5.2.5 Performing the adjustment
The zero point and sensitivity are adjusted following the same
procedure for all gases. The adjustment of oxygen is an excep-
tion. For this reason it is described separately (see Section 5.2.6).
You can nd detailed information on the adjustment
of various gases (for example, test gas concentration,
installation date of the sensor, date of last adjustment)
under Information.
The symbol appears after the corresponding Adjust-
ment… menu item has been selected.
5.2.5.1 Adjusting the zero point
For all gases except oxygen O
ing the same procedure.
1. Make sure that only fresh air is being drawn in.
2. Open Settings.
3. Select Adjustment menu item.
68
, the zero point is adjusted follow-
2
Page 77
4. Select the desired adjustment (e.g. Adjustment CH4, Adjust-
ment gas mixture).
5. Wait at least 1 minute. The displayed reading must be stable.
Note:
For Adjustment gas mixture, all values must be stable. The time
required for this can vary depending on the specic gas.
6. From the menu select the method you wish to use to adjust
the zero point.
− e.g. for Adjustment CH4: Zero point
− e.g. for Adjustment gas mixture: Fresh air
7. Press OK to conrm.
This adjusts the zero point. The reading shows zero (0.00 %
vol. / 0 ppm). (Except for Adjustment gas mixture menu item:
the value for oxygen O
5.2.5.2 Adjusting the sensitivity
For all gases except oxygen O
ing the same procedure.
The following resources are needed for adjusting the sensitivity:
● Test gas
Information on test gases for adjustment can be found in Section 5.2.2.
● Test set for the supply of test gas (e.g. SPE VOL)
5 Maintenance
is 20.9 % vol.)
2
, the sensitivity is adjusted follow-
2
Note:
Details of how to use the test set can be found in the accompanying operating instructions.
1. Connect the device to the test set.
2. Open Settings.
3. Select Adjustment menu item.
69
Page 78
5 Maintenance
4. Select the desired adjustment (e.g. Adjustment CH4, Adjust-
5. Select the menu item that species the sensitivity to be tested.
6. Press and hold the release button on the test set. The test gas
7. Wait at least 1 minute. The displayed reading must be stable.
Note:
For Adjustment gas mixture, all values must be stable. The time
required for this can vary depending on the specic gas.
8. Press OK to conrm. The device is adjusted. The reading
9. Let go of the release button on the test set.
ment gas mixture).
− e.g. for Adjustment CH4: 2.20 % VOL. CH4
− e.g. for Adjustment gas mixture: Gas mixture
Do not conrm with OK yet.
is added. Do not let go of the release button.
shows the specied value (e.g. 2.20 % vol. CH
).
4
5.2.6 Carrying out an oxygen adjustment
As oxygen is a component of fresh air, the procedure for adjusting
oxygen is dierent from the procedure for all other gases.
5.2.6.1 Adjusting the zero point for oxygen
The zero point of oxygen must be adjusted using a gas that does
not contain any oxygen and which will not damage the sensor.
70
Page 79
5 Maintenance
The following resources are needed for adjusting the zero point
of oxygen:
● Test gas
Information on test gases for adjustment can be found in Section 5.2.2.
● Test set for the supply of test gas (e.g. SPE VOL)
Note:
Details of how to use the test set can be found in the accompanying operating instructions.
1. Connect the device to the test set.
2. Open Settings.
3. Select Adjustment from the menu.
4. Select Adjustment O2 menu item.
5. Select Zero point menu item. Do not conrm with OK yet.
6. Press and hold the release button on the test set. The test gas
is added. Do not let go of the release button.
7. Wait at least 1 minute. The displayed reading must be stable.
8. Press OK to conrm. The device is adjusted. The reading
shows zero (0.0 % vol.).
9. Let go of the release button on the test set.
71
Page 80
5 Maintenance
5.2.6.2 Adjusting the sensitivity for oxygen
The sensitivity for oxygen is adjusted with fresh air.
1. Make sure that only fresh air is being drawn in.
2. Open Settings.
3. Select Adjustment from the menu.
4. Select Adjustment O2 from the menu.
5. Wait until the displayed reading is stable. (The reading may
still ash.)
6. Select 20.9 % VOL. (fresh air) from the menu (select and
conrm with OK ). This adjusts the sensitivity. The reading
shows 20.9 % vol.
5.3 Servicing
The device must only be serviced and repaired by SEWERIN
Service.
● Send the device to SEWERIN for repairs and for annual main-
tenance.
72
Note:
If there is a service agreement in place, the device can be serviced by the mobile maintenance service.
The inspection plate on the device shows con-
rmation of the last maintenance and the next
scheduled maintenance.
Fig. 21: Inspection plate
Page 81
6 Faults
If a fault occurs during operation, an error message will appear
on the screen.
Error messages are displayed in the order in which they occur.
Up to ve errors can be displayed.
Error messages continue to be displayed until the error is corrected.
Overview of possible error messages
6 Faults
Error
code
8No calibration
9No calibration
10Adjustment failed
11Zero point
52XFLASH
59Error unknown
60PX sensorError can only be corrected by
61PX sensorError can only be corrected by
100Pump error
200I2C HOST – IR
201I2C HOST – EC
202I2C HOST – EX
Error message on the
display
PPM sensor adjustment
IR/PX sensor adjustment
Test gas
Adjustment required
SEWERIN Service
SEWERIN Service
Probe/lter
SEWERIN Service
SEWERIN Service
SEWERIN Service
Error correction
Adjustment CxHy ppm required
Adjustment required (Adjustment CxHy or Adjustment gas mixture)
Check test gas
Adjustment required
Error can only be corrected by
SEWERIN Service
Error can only be corrected by
SEWERIN Service
SEWERIN Service
SEWERIN Service
Check all lters, probes and hose
connections for porosity and dirt
Error can only be corrected by
SEWERIN Service
Error can only be corrected by
SEWERIN Service
Error can only be corrected by
SEWERIN Service
73
Page 82
7 Appendix
7 Appendix
7.1 Specications and permitted operating conditions
Dimensions (W×D×H):Approx. 148 × 57 × 205 mm
Approx. 148 × 57 × 253 mm with supporting
bracket
Weight:Approx. 1000 g, depending on equipment
Operating position:Any
Protection rating:IP54
Power supply:4 cells, either:
– Rechargeable batteries: NiMH
– Disposable batteries: Alkaline
Operating time:Minimum 8 h
Charging time for re-
chargeable batteries:
Charging voltage:12 V DC (max. 1 A)
Operating temperature: -20 ºC – +40 ºC
Storage temperature:-25 ºC – +60 ºC
Pressure:800 – 1100 hPa
Permissible relative
Zero point deviation:7 ppm
Response time:t
Decay time:t10 < 24 s
Drift:< 10 % within 6 months
Temperature range:-20 ºC – +40 ºC
Interference at 20 ºC.
– 3000 ppm H
– 100 ppm NO
Lifetime:
– Warranted
– Expected
Test gases:
– Zero point
– Sensitivity
2
< 15 s
90
2 years
3 years
100 % vol. CH
or 100 % vol. N
4
2
Fresh air (20.9 % vol.)
0 – 500 ppm
4 ppm
±5 ppm (long-term stability) as per /4/
< 30 s
90
Approx. 1000 ppm CO
Approx. 25 ppm CO
24 months
36 months
Fresh air
40 ppm CO
83
Page 92
7 Appendix
7.6 Technical information
7.6.1 Sensitivity of the catalytic combustion sensor
Oxygen-decient atmospheres can reduce the sensitivity of the
catalytic combustion sensor (sensor suocation).
Gaseous constituents of silicones, oils and phosphate esters, for
example, have a damaging eect on the sensor. They permanently
reduce the sensitivity.
Contamination of the measurement environment with halogens,
burnt neoprene, PVC or trichloroethene, for example, also lowers the sensitivity of the sensors, however, this can be restored.
7.6.2 Identication sticker (back of device)
The symbols on the sticker mean the following:
Only ever open the battery compartment outside of
explosive areas.
Read the operating instructions.
7.6.3 Cleaning
The device must only be cleaned with a damp cloth.
CAUTION! Damage possible due to unsuitable
A
84
cleaning agents
Unsuitable cleaning agents can cause chemical corrosion
on the housing surface. Vapours from solvents and substances containing silicone can penetrate the device and
damage the sensors.
● Never clean the device with solvents, petrol or cockpit
spray containing silicone or similar substances.
Page 93
7.6.4 Electrostatic charge
Avoid electrostatically charging the device. Electrostatically unearthed objects (e.g. including metallic housing without an earth
connection) are not protected against applied charges (e.g.
through dust or dispersed ows).
DANGER! Risk of explosion due to sparks
A
When working with hydrogen, electrostatic charging can
occur.
● When working with hydrogen, always use the TG8 car-
rying bag.
7 Appendix
85
Page 94
7 Appendix
7.7 Accessories and consumables
Accessories
PartOrder number
Docking station TG8LP11-10001
AC/DC adapter M4LD10-10001
Vehicle cable M4, 12 V= portableZL07-10100
Vehicle cable M4, 12 V= installedZL07-10000
Vehicle cable M4, 24 V= portableZL09-10000
"Vario" carrying system3209-0012
Carrying bag TG83204-0040
Case TG8-RÜZD29-10000
Compact case TG8ZD31-10000
Carpet probe PROZS01-12000
Bell probe D80ZS05-10500
Localisation probe 345 mmZS03-10300
Flexible hand probeZS32-10000
Floating probeZS21-10100
Probe hoseZS25-10000 (e.g.)
Test set SPE VOLPP01-90101
Test set SPE ppmPP01-40101
Test set SPE DUOPP01-60001
Test case kit PPMZP03-12001
Test plateZP06-10000
86
Consumables
PartOrder number
Fine dust lter2499-0020
Hydrophobic lter2491-0050
Special lter element2499-0005
Rechargeable NiMH battery1354-0009
Disposable alkaline battery1353-0001
Test gas (2.2 % vol. CH
CO and 17.5 % vol. O2),
test gas can 1 l, pressure approx. 12 bar
Test gas 10 ppm CH4 in synthetic air,
test gas can 1 l, pressure approx. 12 bar
, 2.0 % vol. CO2, 40 ppm
4
ZT32-10000
ZT24-10000
Other accessories and consumables are available for the product. Please contact our SEWERIN sales department for further
information.
Page 95
7.8 EU declaration of conformity
Hermann Sewerin GmbH hereby declares that the VARIOTEC
480/460/450/400 EX fulls the requirements of the following
guidelines:
● 2014/34/EU
● 2014/30/EU
Gütersloh, 2016-04-20
Dr. S. Sewerin (CEO)
The complete declaration of conformity can be found online.
Sample inspection protocol for inspection with gas mixture.
88
Page 97
7.0 Measuring in bar holes / Gas measuring
7.1
Zero point CH
4
– Display -1.0 – +1.0 % vol.
7.2
Test gas 100 % vol. CH
4
– Display 98 – 102 % vol.
8.0 Ethane analysis
8.1
Test gas 100 ppm C
2H6
, 1 % vol. CH
4
–
CH4: Yes, C2H6: Yes, C3H8: No
9.0 Comments
– Housing damaged
– Adjustment, repair
– Inspection at factory
– Or similar
10.0 Inspection
– Day
– Month
– Year
– Signature
- 2 -
7 Appendix
89
Page 98
7 Appendix
7.10 Advice on disposal
The European Waste Catalogue (EWC) governs the disposal of
appliances and accessories.
Description of wasteAllocated EWC waste code
Device16 02 13
Test gas can16 05 05
Disposable battery, rechargeable battery
16 06 05
End-of-life equipment
Used equipment can be returned to Hermann Sewerin GmbH.
We will arrange for the equipment to be disposed of appropriately
by certied specialist contractors free of charge.
90
Page 99
7.11 Terminology and abbreviations
% vol.● Percent concentration of a gas in a gas
mixture with respect to the volume
AL● Alarm
CC● Catalytic combustion sensor
CENELEC● European Committee for Electrotechnical
Standardization
EC● Electrochemical sensor
Gas type● Hydrocarbon C
ured with a SC, TC or CC
● One of the available gas types must al-
ways be set at a time, as it is not possible to measure more than one gas type
at the same time.
LEL● Lower Explosion Limit
LTEL● Long-time exposure limit
NiMH● Nickel metal hydride
OEL● Occupational exposure limit
ppm● Parts per million
Ring memory● Type of data storage in the device
● If the available storage space is full, the
oldest le is automatically overwritten by
the current le.
SC● Gas-sensitive semiconductor
Stack memory● Type of data storage in the device
● If the available storage space is full, you
are prompted to conrm whether the
oldest le should be overwritten by the
current le.
ance Association); Code of Practice T 023: Gas Warning Devices for
Explosion Protection – Use and Operation
Available for download at: www.bgchemie.de
/3/ DVGW G 465-4
Deutsche Vereinigung des Gas- und Wasserfaches e. V. (German
Association of Gas and Water Specialists); Regulation G 465-4:
Gasspür- und Gaskonzentrationsmessgeräte für die Überprüfung
von Gasanlagen (Gas-Detection and Gas-Concentration Measurement Devices for Inspection of Gas Systems)
Available for download at: www.dvgw.de
/4/ EN 45544
/5/ EN 45544-2
/6/ EN 45544-4
/7/ EN 50104:2010
/8/ EN 60079-7:2007
/9/ EN 60079-29-1
/10/ EN 60079-29-2
/11/ EN 60086-1
/12/ EN 61779-1
/13/ EN 61951-2
/14/ IEC 60079-20
/15/ TRGS 900: 2006
Technical regulations for hazardous substances: Occupational expo-
sure limits
Available for download at: www.baua.de
/16/ 94/9/EC (ATEX 100a)
92