sewerin Multitec 560 Operating Instructions Manual

Page 1
Multitec® 560
20.04.2016 a – 106810 – en
Operating instructions
Page 2
Multitec® 560
Connector
Supporting bracket
Buzzer
USB port
ON/OFF key
Connection for power supply
Alarm
Gas
Signal light
Gas input
Display
Function keys
Jog dial
Connector
Measured value Unit
Warning ExTox symbol
Capacity disposable battery/ rechargeable battery
Fig. 2: Multitec 560 display
Current assignment of function keys F1 – F3
Page 3

Display symbols

Menu Fault
OK Carry out device inspection
Esc
Cancel
Buzzer off Clear
Save Information
Stop measurement Warning ExTox
Capacity disposable battery/ rechargeable capacity
Tab
(jump to next input eld)
Gas measuring
Open stored comment Open stored inspector
Page 4

Information about this document

The warnings and notes in the document mean the following:
DANGER!
A
A
A
Risk of personal injury. Results include serious injury or death.
WARNING!
Risk of personal injury. Can result in serious injury or death.
CAUTION!
Risk of personal injury. Can result in injury or a risk to health.
CAUTION!
Risk of damage to property.
Note:
Tips and important information.
Enumerated lists (numbers, letters) are used for:
● Instructions that must be followed in a specic sequence
Bulleted lists (bullet points, dashes) are used for:
● Lists
● Instructions comprising only one action
Numbers enclosed by forward slashes /.../ refer to referenced doc­uments.
Page 5
Contents Page
1 General .....................................................................................1
1.1 Warranty ....................................................................................1
1.2 Purpose .....................................................................................2
1.3 Intended use .............................................................................3
1.4 General safety information ........................................................4
2 Features ...................................................................................5
2.1 Visual and audible signals .........................................................5
2.2 Sensors .....................................................................................6
2.3 Explosion protection .................................................................. 7
2.3.1 Passive explosion protection ..................................................7
2.3.2 Active explosion protection ....................................................8
3 Operation .................................................................................9
3.1 General information on operation .............................................. 9
3.1.1 Keys and jog dial .................................................................... 9
3.1.2 Selecting/exiting menus and menu items .............................10
3.1.3 Switching the device on ....................................................... 11
3.1.4 Selecting/switching application ............................................13
3.1.5 Differences between measuring mode and settings mode ..13
3.2 Measuring mode .....................................................................14
3.2.1 Accessing the menu (measuring mode menu structure) ......15
3.2.2 Zero point ............................................................................. 15
3.2.3 Gas measuring ..................................................................... 16
3.2.4 Warning ExTox .....................................................................16
3.2.5 Settings ................................................................................16
3.2.6 Starting/stopping a measurement ........................................17
3.2.7 Protocols ..............................................................................18
3.2.8 Device inspection ................................................................. 19
3.2.9 Device information ...............................................................19
3.3 Settings ...................................................................................20
3.3.1 Opening Settings ..................................................................20
3.3.2 Settings menu structure .......................................................22
3.3.3 Adjustment ...........................................................................23
3.3.4 System .................................................................................24
3.3.5 Alarms ..................................................................................26
3.3.6 Date/time .............................................................................. 26
3.3.7 Memory ................................................................................26
I
Page 6
Contents Page
4 Power supply .........................................................................27
4.1 Suitable disposable/rechargeable battery types .....................27
4.2 Operation with rechargeable batteries ....................................28
4.2.1 Charging ............................................................................... 28
4.2.2 Rechargeable battery maintenance .....................................29
4.3 Battery alarm ........................................................................... 30
4.4 Replacing disposable/rechargeable batteries .........................30
5 Maintenance ..........................................................................32
5.1 Device inspection ....................................................................32
5.1.1 General information on the device inspection ...................... 32
5.1.1.1 Scope ................................................................................ 32
5.1.1.2 Frequency .........................................................................33
5.1.1.3 Documentation .................................................................. 33
5.1.1.4 Integrated device inspection .............................................33
5.1.1.5 Order .................................................................................34
5.1.1.6 Test gases for the device inspection .................................34
5.1.2 Carrying out the device inspection ....................................... 35
5.1.2.1 Accessing the device inspection ....................................... 35
5.1.2.2 Concluding the device inspection......................................36
5.1.3 Testing the general status .................................................... 37
5.1.3.1 Housing ............................................................................. 38
5.1.3.2 Signals ..............................................................................38
5.1.3.3 Probe.................................................................................38
5.1.3.4 Filter ..................................................................................39
5.1.3.5 Pump .................................................................................39
5.1.4 Testing indication accuracy with supply of fresh air ............. 39
5.1.5 Testing indication accuracy with supply of test gas .............. 40
5.2 Adjustment ..............................................................................41
5.2.1 Scope ...................................................................................42
5.2.2 Test gases for the adjustment .............................................. 42
5.2.3 Special features of adjustment with gas mixture .................. 43
5.2.4 Preparation ........................................................................... 44
5.2.5 Performing the adjustment ................................................... 44
5.2.5.1 Adjusting the zero point.....................................................44
5.2.5.2 Adjusting the sensitivity ..................................................... 45
5.2.6 Carrying out an oxygen adjustment .....................................46
5.2.6.1 Adjusting the zero point for oxygen ................................... 46
5.2.6.2 Adjusting the sensitivity for oxygen ................................... 47
5.3 Servicing .................................................................................48
II
Page 7
Contents Page
6 Faults ......................................................................................49
7 Appendix ................................................................................50
7.1 Specications and permitted operating conditions .................. 50
7.2 Alarms .....................................................................................51
7.2.1 Features ...............................................................................51
7.2.2 Occupational exposure limits (OELs) and
excess factors (STEL and LTEL) ..........................................54
7.2.3 Alarm thresholds (factory settings) .......................................54
7.2.4 Setting ranges for gas types ................................................55
7.3 Limit values for the device inspection .....................................56
7.4 Memory capacity .....................................................................57
7.5 Sensors ...................................................................................58
7.5.1 Infrared sensors (IR) ............................................................58
7.5.1.1 Methane CH4 for warning ExTox ....................................... 58
7.5.1.2 Methane CH4 for gas measuring ....................................... 58
7.5.1.3 Carbon dioxide CO2 for warning ExTox ............................. 59
7.5.1.4 Carbon dioxide CO2 for gas measuring .............................59
7.5.2 Electrochemical sensors (EC) .............................................. 60
7.5.2.1 Oxygen O2 ........................................................................ 60
7.5.2.2 Carbon monoxide CO .......................................................60
7.5.2.3 Hydrogen sulphide H2S for warning ExTox .......................61
7.5.2.4 Hydrogen sulphide H2S for gas measuring .......................62
7.6 Technical information ..............................................................63
7.6.1 Identication sticker (back of device) ...................................63
7.6.2 Cleaning ............................................................................... 63
7.6.3 Electrostatic charge ..............................................................63
7.7 Accessories and consumables ............................................... 64
7.8 EU declaration of conformity ...................................................65
7.9 Inspection protocol ..................................................................66
7.10 Advice on disposal ..................................................................68
7.11 Terminology and abbreviations ...............................................69
7.12 Referenced documents ...........................................................70
8 Index .......................................................................................71
III
Page 8
1 General

1.1 Warranty

The following instructions must be complied with in order for any warranty to be applicable regarding functionality and safe opera­tion of this equipment. This product must only be commissioned
by qualied professionals who are familiar with the legal require­ments (Germany: DVGW).
● Read these operating instructions prior to operating the product.
● Use the product only as intended.
● Repairs and maintenance must only be carried out by special-
ist technicians or other suitably trained personnel. Only spare parts approved by Hermann Sewerin GmbH may be used when performing repairs.
● Use only suitable battery types, otherwise the device will not
be explosion-proof.
● Changes or modications to this product may only be carried
out with the approval of Hermann Sewerin GmbH.
● Use only Hermann Sewerin GmbH accessories for the product.
Hermann Sewerin GmbH shall not be liable for damages resulting from the non-observance of this information. The warranty con-
ditions of the General Terms and Conditions (AGB) of Hermann
Sewerin GmbH are not affected by this information.
In addition to the warnings and other information in these Oper­ating Instructions, always observe the generally applicable safety and accident prevention regulations.
The manufacturer reserves the right to make technical changes.

1 General

1
Page 9
1 General

1.2 Purpose

The Multitec 560 can be used as a gas measuring and gas warning device.
The device is tted with infrared sensors for measuring hydro­carbons CXHY and carbon dioxide CO2 as standard. It can also be equipped with electrochemical sensors for measuring O2, H2S and CO.
The infrared sensors operate on the principle of absorption via infrared-active gases, and the electrochemical sensors operate on the electrochemical cell principle.
Multitec 560 as a gas warning device
As a warning device it is used to monitor the ambient air at the workplace. The device can measure up to ve gases simultane­ously, thus offering comprehensive protection against dangerous gas concentrations. It provides warnings including:
● explosive gas concentrations
● lack of oxygen/excessive oxygen
● toxic gases
Multitec 560 as a gas measuring device
As a gas measuring device it determines the composition of gas
mixtures that are formed in biological processes (biogas, landll gas). It measures the concentration of several gases in the gas
mixture simultaneously. The device is ideal for use in waste dis­posal sites, sewage plants and biogas plants.
Note: These operating instructions describe the Multitec 560 with all
additional equipment (rmware version 1.XXX). All descriptions refer to the device as delivered (factory settings). The manufac-
turer reserves the right to make changes.
2
Page 10

1.3 Intended use

This device is intended for professional residential and commer-
cial use including small rms and commercial operations. The
appropriate specialist knowledge is required to operate the device.
The device may only be used to measure the following gases
(depending on additional equipment):
● Methane CH
● Carbon dioxide CO
● Oxygen O
● Hydrogen sulphide H2S
● Carbon monoxide CO
The device must not be used for:
● Gas analysis of technical processes
● Monitoring liquids
The device can be used up to a temperature of 40 ºC. However, high temperatures reduce the lifetime of the sensors and re­chargeable batteries.
If a device with an electrochemical sensor is exposed to gas concentrations above the measuring range limit, this can reduce
the lifetime of the sensor.
1 General
4
2
2
3
Page 11
1 General

1.4 General safety information

The device has been tested to ensure that it is explosion-proof in accordance with European standards (CENELEC).
● The work area must be well ventilated.
● The device must only be switched on with fresh air.
● In measuring mode, only change applications with fresh air.
● When switching from the Gas measuring application (using
the device as a gas measuring device) to the Warning ExTox application (using as a gas warning device), the device should
only be used as a gas warning device once the zero points have appeared.
● Do not use this device in oxygen-enriched atmospheres, oth- erwise it will not be explosion-proof.
● Observe the relevant safety regulations when working at agri- cultural biogas plants.
● Only probe hoses with a hydrophobic lter may be used.
Exception:
If the probe has a built-in hydrophobic lter, the hose does not require any other lters.
● The device must only be tested and adjusted with test gases in well ventilated rooms or in the open air. Test gases must be handled in a professional manner.
● Always carry out a device inspection after the device has suf-
fered an impact (for example, if dropped accidentally).
● The device complies with the limits of the EMC directive. Always observe the information in the manuals of (mobile) radio equip­ment when using the device close to (mobile) radio equipment.
Note:
Follow the advice regarding explosion protection (see Section. 2.3 on page 7).
4
Page 12
2 Features

2.1 Visual and audible signals

The device features two alarms:
● Signal light on top of device (visual signal)
● Buzzer on side of device (audible signal)
The signals indicate alarms and faults (see Section. 6 on page 49). The device also emits signals when it is switched on and off.
If this symbol appears on the display, the audible signal can be switched off.
When an audible signal has been switched off it cannot be switched back on.
This symbol appears at the top left of the display as soon as the audible signal has been switched off. It disappears automatically if the level falls below the alarm threshold.

2 Features

Operating signal
Only applies to Warning ExTox application.
The device emits a visual signal and an audible signal at regular intervals. This indicates that the device is working properly.
Alarm
Only applies to Warning ExTox application.
The device can monitor several gases at the same time. If the measured gas concentration of one or more gases exceeds spec-
ied limit values (alarm thresholds) the device gives a warning. It
emits both audible and visual signals, which are distinctly different from the operating signal.
5
Page 13
2 Features
WARNING! Danger of death due to hazardous gas
A
concentrations
An alarm always indicates danger.
There is detailed information on alarms in Section. 7.2 on page 51.

2.2 Sensors

The device features two types of sensor:
● Infrared sensor (IR)
● Electrochemical sensor (EC)
● Take all necessary measures for your own safety and
the safety of others immediately.
Application Gas Measuring
range
Warning ExTox CH
0 – 100 % LEL IR ×
4
CO 0 – 500 ppm EC
CO
0 – 5 % vol. IR ×
2
H2S 0 – 100 ppm EC
O
Gas measuring CH
0 – 25 % vol. EC
2
0.0 – 100 % vol. IR ×
4
CO 0 – 500 ppm EC
CO
0 – 100 % vol. IR ×
2
H2S 0 – 2000 ppm EC
O
0 – 25.0 % vol. EC
2
Sensor Fea-
tures
× standard
○ optional
6
Page 14

2.3 Explosion protection

2.3.1 Passive explosion protection

The device is assigned to the following explosion-proof groups:
2 Features
Explosion-proof group
II2G Ex d e ib IIB T4 Gb – Methane CH
For the following at­mospheres
– Propane C3H – Butane C4H – Nonane C9H
4
8
10
20
When using
Device with­out carrying
bag TG8
– Hydrogen sulphide H2S – Carbon monoxide CO
II2G Ex d e ib IIC T4 Gb – Methane CH
– Propane C3H – Butane C4H – Nonane C9H
4
8
10
20
Device with carrying bag TG8
– Hydrogen sulphide H2S – Carbon monoxide CO – Hydrogen H
2
EC type-examination certicate: TÜV 07 ATEX 553353 X
DANGER! Risk of explosion due to sparks
A
● Only open the battery compartment outside of explo-
sive areas.
● Only charge the device outside of explosive areas.
● Only use the USB port outside of explosive areas.
● Use only suitable battery types.
● When working with hydrogen, always use the carrying
bag TG8 for the device.
7
Page 15
2 Features

2.3.2 Active explosion protection

The functional safety test applies to:
Application: Warning ExTox
Gas types: Measuring range: As per:
– Methane CH
4
Gases: Measuring range: As per:
– Oxygen O
– Carbon dioxide CO
2
2
– Carbon monoxide CO 0 – 500 ppm CO /4/, /5/
– Hydrogen sulphide H2S 0 – 100 ppm /4/, /5/
Tested accessories: – Test set SPE VOL
Type examination
Testing institute: DEKRA EXAM GmbH
0 – 100 % LEL /9/
0 – 25 % O
0 – 5 % CO
2
2
/7/
/4/, /5/
– Flexible hand probe, 1 m – Floating probe 2 m, 6 m
Certicates: PFG 08 G 002 X
BVS 09 ATEX G 001 X
The following points were not part of the type examination:
● Saving measurement data (see Section. 3.2.6 on page 17)
● Saving protocols from the integrated device inspection (see Section. 5.1.1.4 on page 33)
● Disposable alkaline batteries for the power supply (see Sec- tion. 4.1 on page 27)
8
Page 16
3 Operation

3.1 General information on operation

3.1.1 Keys and jog dial

The ON/OFF key is the only control on the device that does not change its function.
When switched on, the device is operated using the jog dial and function keys to navigate the display.
Control Action Function ON/OFF key Press Switches the device on
● Switches the device off
Function keys F1, F2, F3
Jog dial Turn Selects functions, settings,
Press ● Variable
● As indicated on the display at
the bottom of the screen
● Function keys may also have
no function assigned in some cases
measurement data, etc.
● Modies values
Press ● Opens the next program lev-
el (e.g. menu item, function, measurement data, selectable
values)
Accepts values

3 Operation

9
Page 17
3 Operation

3.1.2 Selecting/exiting menus and menu items

Functions, applications and settings etc. are selected via the main menu (for short: Menu). This menu has submenus and menu items. Refer to Section. 3.2.1 on page 15 for information on accessing the menu.
Selecting submenus/menu items
Submenus and menu items are selected and opened using the jog dial and/or function keys.
In measuring mode the selected application is indicated by the symbol at the top left of the display.
Exiting menus/menu items
There are generally two ways to exit open menus/menu items and go back up a level:
Press Esc
● Select Exit menu item
10
Page 18

3.1.3 Switching the device on

Note:
Always switch the device on with fresh air.
● Press the ON/OFF key. The device switches on.
The switching on process involves an internal check.
Process Test purpose
Buzzer emits audible sig­nal.
Signal light gives visual signal.
Display is inverted. Are there pixels missing from the
The start screen appears on the display.
®
Multitec 560
Frank Smith City Council Leakage Delivery
V1.000
Fig. 3: Start screen
22.04.2013
13:02
3 Operation
Is the audible signal working?
Is the visual signal working?
display?
Display:
– Device type: Multitec 560 – User:
Frank Smith City Council
Leakage Delivery – Firmware version: V1.000 – Date and time – Capacity disposable battery/
rechargeable battery
The start screen for the application set under Autostart then appears (see Section. 3.3.4 on page 24).
11
Page 19
3 Operation
Fig. 4: Start screen for gas measuring
Fig. 5: Measuring mode – display of current readings
Display:
– Symbol: Warning ExTox
Warning ExTox Range CH4 4.40 VOL% CO2 5.00 VOL% O2 25.0 VOL% H2S 100 PPM CO 500 PPM
AL1 AL2 STEL LTEL
0.45 2.20 - -
0.50 1.00 1.00 0.50
18.0 23.0 - ­10 20 10 10 30 60 30 30
– Detectable gases – Measuring range – Alarm thresholds – Capacity disposable battery/
rechargeable battery
The device switches to measuring mode.
Display:
– Current readings: zero when
device is switched on with fresh air
The device is ready for use.
12
A
WARNING! Danger of death due to incorrectly ad­justed or faulty devices
Gas warning instruments must be inspected before use at regular intervals.
● Carry out a device inspection every day before starting
work.
Page 20

3.1.4 Selecting/switching application

WARNING! Danger of death due to incorrectly ad-
A
Note:
You may only switch applications when the device is drawing in fresh air.
The current application is indicated by the symbol at the top left of the display. The symbol at the bottom right shows the next ap­plication that can be selected via function key F3. The Settings under System (see Section 3.3.4) are used to specify which ap-
plication is rst activated when you switch the device on.
justed or faulty devices If the device is used as a gas warning instrument (Warn­ing ExTox application), it must be regularly inspected
before use.
● Carry out a device inspection every day before starting
work.
● Press Menu. Select the menu item for the application you
want to use.
OR a) Press function key F3. The device switches to the next ap-
plication.
b) Repeat until the symbol for the application you want to use
appears at the top left.
3 Operation

3.1.5 Differences between measuring mode and settings mode

The device is operated in two modes:
Measuring mode (see Section. 3.2 on page 14)
Measurements are taken in measuring mode. All functions needed to take readings can be accessed from one menu.
13
Page 21
3 Operation
Settings (see Section. 3.1 on page 9)
The device settings can be changed in settings mode. Infor­mation about the device can also be retrieved. Measurements cannot be taken in settings mode.
Settings are accessed via the menu in measuring mode. The settings are access-protected by a PIN code.
WARNING!
A

3.2 Measuring mode

When switched on, the device is in measuring mode. In meas­uring mode, the current measurements are always displayed (Fig. 5). However, to save the data from a measurement, you must always start the measurement manually (see Section. 3.2.6 on page 17).
Danger of death due to lack of alarm signal
The device only issues alarms in measuring mode. As soon you access the menu, alarms are no longer triggered.
● Only change the settings outside of explosive areas and
away from toxic or low-oxygen atmospheres.
14
A
WARNING! Danger of death due to operating signal failure
If the operating signal fails in the Warning ExTox appli- cation, the device is not safe to use.
● Stop using the device immediately.
● Move away from explosive areas or toxic or low-oxygen
atmospheres immediately.
Page 22
3 Operation

3.2.1 Accessing the menu (measuring mode menu structure)

In measuring mode F1 can be used to access the menu.
Zero point
Gas measuring
Warning ExTox
Settings
Start measurement
Protocol
Device inspection
Device information
Exit
Fig. 6: Menu with submenus/menu items
Once you have started a measurement, Start measurement in the menu becomes Stop measurement. You can nd detailed information on starting and stopping measurements in Sec­tion. 3.2.6 on page 17.
Protocol does not appear in the menu until you save a protocol
for the rst time.

3.2.2 Zero point

The zero point can be set manually in the Zero point menu item. This is only necessary if the displayed fresh air measurement is not zero after the end of the warm-up period.
Gas Content in fresh air Correct zero point on device
CH40 % vol. 0.0 % vol.
CO 0 ppm 0 ppm
CO20.04 % vol. 0.04 % vol.
H2S 0 ppm 0 ppm
O
20.9 % vol. 20.9 % vol.
2
The manual zero point setting is not saved. The zero point can be corrected by adjustment as often as zero point deviations occur.
15
Page 23
3 Operation
Requirements for correct setting of the zero point
● Device has been switched on with fresh air.
● Device continues to draw in fresh air.
Setting zero point (manual zero point setting)
1. Press Menu.
2. Select Zero point menu item. The values are automatically
adjusted. The device returns to measuring mode.

3.2.3 Gas measuring

You can nd an overview of detectable gases under Gas meas­uring in the menu. The device automatically returns to measuring
mode.

3.2.4 Warning ExTox

You can access the overview of detectable gases and corre­sponding alarm thresholds under Warning ExTox in the menu. The device automatically returns to measuring mode.
CAUTION! Health risk due to late alarm
A

3.2.5 Settings

You can change the device settings and access information about the device under Settings in the menu (see Section. 3.3 on page 20).
16
If you switch from Warning ExTox to Gas measuring, data that could initiate a STEL or LTEL alarm is reset to zero.
● Only switch from Warning ExTox to Gas measuring
once you have left the work area being monitored.
Page 24

3.2.6 Starting/stopping a measurement

Measurements must always be started and then stopped. When the measurement is stopped, the measurement data is stored
in a le.
Note:
Measurements cannot be cancelled. The only way to cancel a measurement is to stop it.
Up to 80 measurements can be saved.
The measured values can be saved with or without a comment. Comment entries are saved automatically (ring memory with
max. 10 entries).
Once the rst comment has been entered, the Open
stored comments function will become available.
The stored measurement les can be displayed on a computer
using a readout program. The program is available at www. sewerin.com.
3 Operation
Starting a measurement
1. Press Menu.
2. Select Start measurement from the menu. This starts meas-
urement plot recording.
Measurement plot recording must always be concluded with Stop measurement.
17
Page 25
3 Operation
Stopping a measurement
1. Press Stop measurement.
2. Answer Yes to the warning prompt.
3. Enter a comment for the measurement.
OR
a) Press Menu. b) Select Stop measurement from the menu.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored comments. A list of the stored com-
ments will appear.
− Select the desired comment. Open the comment with OK.
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter a comment for the measurement.
The measurement is saved as a protocol. The protocol name is formed from the date, time and comment.

3.2.7 Protocols

You can retrieve or clear protocols of saved data under Protocol in the menu. When saved, the protocols are assigned to different protocol types.
The following protocol types are available:
● Device inspection
● Measurements
Protocols can only be cleared individually. You can nd information on how to clear all protocols of one pro-
tocol type in Section. 3.3.7 on page 26.
18
Page 26

3.2.8 Device inspection

The device inspection can be used to check the general status and the indication accuracies. Device inspection only appears in the menu when the integrated device inspection is switched on.
Note:
The integrated device inspection is switched off in the factory settings. More detailed information about the device inspection can be found in Section. 5.1 on page 32.
If the integrated device inspection is switched on, the device will remind you to carry out a device inspection.
The Carry out device inspection symbol will appear when the inspection is due. It is visible in the display until the complete integrated device inspection has been carried out successfully.

3.2.9 Device information

The following device information is shown under Device infor­mation in the menu:
● Installed electrochemical sensors: gas, installation date, war-
ranted/expected lifetime
● Firmware: version, date
● Service: date of the last service, date of the next service
3 Operation
19
Page 27
3 Operation

3.3 Settings

The following menus and menu items are included under Set­tings :
● Adjustment
● System
● Alarms
● Date/time
● Memory
You can nd information on selecting and exiting menus and menu
items in Section. 3.1.2 on page 10.

3.3.1 Opening Settings

1. Press Menu.
WARNING!
A
Danger of death due to lack of alarm signal
The device only issues alarms in measuring mode. As soon you access the menu, alarms are no longer triggered.
● Only change the settings outside of explosive areas and
away from toxic or low-oxygen atmospheres.
20
2. Select the Settings menu item.
Access is protected by a PIN code. The default setting is always PIN code 0001.
Note:
You can change the PIN code at any time. SEWERIN recommends setting a different PIN code after initial start-up, so only authorised personnel have access to the settings.
3. Enter the PIN code from left to right. The active digit is always displayed with a black background.
Page 28
3 Operation
Exit
Digit To change To conrm
1st digit
Press the jog dial
2nd digit Press the jog dial
3rd digit Press the jog dial
Turn the jog dial
4th digit
If the PIN code has been entered correctly, the Settings menu
will appear once the last digit has been conrmed (Fig. 7). Otherwise the device will revert to measuring mode.
Adjustment System Alarms Date/time Memory
Fig. 7: Settings menu
21
Page 29
3 Operation

3.3.2 Settings menu structure

Settings Adjustment Adjustment gas mixture warning
Measuring mode
PIN Code
Adjustment gas mixture measuring Test gas Adjustment CH4 Adjustment CO2 Adjustment O2 Adjustment CO Adjustment H2S Inspection OK Exit
System PIN Code
Alarms
Date/time
Memory Clear
Exit
Service interval Display Battery Autostart Unit %LEL Device inspection Language Exit
Interval Memory mode Exit
Fig. 8: Settings menu structure for Multitec 560 (gas type: methane)
22
Note:
The number of available menu items depends on the optional additional equipment.
Page 30

3.3.3 Adjustment

The Adjustment menu is used to set the sensors.
WARNING! Danger of death due to incorrect adjust-
A
A detailed description of adjustment along with important infor­mation is provided in Section. 5.2 on page 41.
Adjustment gas mixture warning
Used to adjust the infrared and electrochemical sensors for all
components of the test gas (gas mixture).
ment
Incorrect adjustment can lead to incorrect measurement results. This means that the user may not be warned about dangerous gas concentrations in time.
● Only specialist technicians may perform adjustments
● Adjustments must be made in well ventilated rooms or
in the open air.
3 Operation
Application: – Warning ExTox
Adjustment gas mixture measuring
Used to adjust the infrared sensors for methane CH4 and carbon dioxide CO2.
Application: – Gas measuring
Test gas
Used to adjust the concentration of the test gases used.
Adjustment CO2
Used to adjust the infrared sensor for carbon dioxide CO2 in the % vol. range.
23
Page 31
3 Operation
Adjustment O2
Used to adjust the electrochemical sensor for oxygen O2 in the % vol. range.
Adjustment CO
Used to adjust the electrochemical sensor for carbon monox­ide CO in the ppm range.
Adjustment H2S
Used to adjust the electrochemical sensor for hydrogen sulphide H2S in the ppm range.
Inspection OK
Conrms the device is in proper working order. This extends the
service interval.

3.3.4 System

General information and specications for operation are set in
the System menu.
24
PIN code
Used to change or reset the PIN code.
Note:
If you lose the PIN code, you must contact SEWERIN Service. If the PIN code is set to 0000, you will not be asked to enter it. The settings can then be accessed by anyone.
Service interval
Species the regular inspections/maintenance required for the
device. You can also activate the automatic switch-off function once the set interval has passed.
Page 32
3 Operation
Display
Used to set how long the display remains illuminated after any key is pressed as well as the display contrast.
Battery
Used to set the type of disposable/rechargeable battery used.
CAUTION! Damage possible due to device overheating
If the battery type is not correctly set, the device can overheat.
● Always enter the correct battery type.
Autostart
Sets the application that is automatically activated when the de­vice is switched on.
Unit %LEL
Used to set the unit of measurement.
Device inspection
Used to switch the integrated device inspection on or off.
Language
Sets the language.
25
Page 33
3 Operation

3.3.5 Alarms

Sets the alarm thresholds for methane CH
There is detailed information on alarms in Section. 7.2 on page 51.
AL1 alarm
Used to set the pre-alarm.
AL2 alarm
Used to set the main alarm.

3.3.6 Date/time

Used to set the time, day, month and year. There are two formats available for the date.

3.3.7 Memory

The Memory menu is used to specify how measurement data and protocols are handled.
4.
26
Clear
Used to clear protocols.
The different protocol types must each be cleared separately. All protocols in one protocol type are cleared at once.
You can nd information on clearing individual protocols in Sec­tion. 3.2.7 on page 18.
Interval
Sets the interval at which measurement data is automatically saved.
Memory mode
Switches between ring memory and stack memory.
Page 34

4 Power supply

4 Power supply
This device can be operated using:
● Disposable (non-rechargeable) alkaline batteries
● Rechargeable NiMH batteries
The device comes with nickel metal hydride rechargeable batter­ies. The corresponding settings are stored.
WARNING! Risk of explosion due to leaking batteries
A

4.1 Suitable disposable/rechargeable battery types

Leaking electrolytes can shorten the creepage distance and air gap between the poles. As a result, the require­ments for the batteries may no longer be met.
● Replace leaking batteries immediately.
● Clean the battery compartment (and, if necessary, the device) before inserting the new disposable/recharge­able batteries.
A
WARNING! Risk of explosion due to unsuitable bat­teries
To ensure that the device remains explosion-proof as per /16/, only certain disposable/rechargeable batteries may be used:
● Only use batteries supplied by SEWERIN. Other dis- posable/rechargeable batteries, which have not been supplied by SEWERIN, may only be used if they meet the specications in accordance with /8/.
● In each battery compartment use only batteries that are identical with respect to type (disposable or recharge-
able), capacity and manufacturer.
27
Page 35
4 Power supply
Disposable battery requirements
● Alkaline disposable batteries
● Size: AA, type: LR6 as per /11/
● The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with /8/.
Rechargeable battery requirements
● NiMH rechargeable batteries
● Size: AA, type: HR6 as per /13/
● The creepage distance and air gap between the poles must
not be less than 0.5 mm in accordance with /8/.
● The rechargeable batteries must be fast charging (I > 1.25 A)
and remain within the temperature range.
Note:
A device operated with disposable alkaline batteries cannot be charged. A note to this effect is shown on the display.

4.2 Operation with rechargeable batteries

The operating time of the device depends on the battery capacity.
If the device is not used or not kept in the docking station, the batteries will lose their charge due to self-discharge. The self-dis­charge intensity depends on the battery type.

4.2.1 Charging

The device can be charged via:
● Connection for power supply
● Docking station TG8
28
Page 36
DANGER! Risk of explosion due to sparks
A
For charging you will need either:
Please note the following points:

4.2.2 Rechargeable battery maintenance

If the device is not used for a long period of time, it is advisable to fully discharge the battery before recharging it again.
A full discharging and recharging process takes approx. 11 hours (8 hours to discharge + 3 hours to recharge). The duration de­pends on the capacity of the rechargeable batteries used.
High charging current occurs when batteries are being charged. The power supply is not explosion-proof.
● Only charge the device outside of explosive areas.
● AC/DC adapter M4
● Vehicle cable M4
● The device or docking station must not be directly connected
to a 24-V on-board power supply in the vehicle. The voltage is too high for the charging process.
● The battery should be charged at approximately room tem-
perature.
4 Power supply
DANGER! Risk of explosion due to sparks
A
High charging current occurs when batteries are being charged. The power supply is not explosion-proof.
● Only charge the device outside of explosive areas.
● Connect the device (switched on) to the power supply via the
side connection.
OR
Place the device (switched on) into the docking station.
The rechargeable batteries will be fully discharged. Once the device has been discharged, it will automatically switch to charging mode.
29
Page 37
4 Power supply

4.3 Battery alarm

As soon as the remaining capacity of the batteries gets low, a battery alarm will go off:
Level 1: Battery almost empty
Capacity disposable battery/rechargeable
battery symbol ashes
Audible signal (one-off) – Operating signal doubles – Remaining operating time: approx. 15 min
Level 2: Battery empty
– Blank display apart from Capacity disposable
battery/rechargeable battery symbol
– Continuous audible signal – Measuring mode unavailable – Device shuts off

4.4 Replacing disposable/rechargeable batteries

DANGER! Risk of explosion due to sparks
A
When the housing is open, the device is not explo­sion-proof.
● Only open the battery compartment outside of explo-
sive areas.
30
A 2.5 mm Allen key (supplied) is required to open the battery
compartment on the back of the device.
1. Loosen the two screws securing the battery compartment. Re­move the screws by repeatedly turning them alternately a short way; this ensures that the battery compartment does not twist.
2. Lift out the battery compartment.
3. Remove the disposable/rechargeable batteries and insert new ones. Ensure that the batteries are inserted with the correct polarity.
4. Replace the battery compartment so it ts neatly into place
and secure rmly with the screws.
5. When you switch the device back on again, you will be asked which battery type is in use. Enter the correct battery type.
Page 38
4 Power supply
If it takes longer than 120 seconds to replace the batteries, the date and time will have to be reset the next time you switch the device on. All the other data will be saved.
31
Page 39

5 Maintenance

5 Maintenance
In accordance with the legal regulations, device maintenance comprises the following elements:
● Device inspection including test of indication accuracy
● Adjustment
● Maintenance
All inspections must be documented. The documentation must be retained for at least one year.
WARNING! Danger of death due to incorrectly ad-
A

5.1 Device inspection

justed or faulty devices
Gas warning instruments must be inspected before use at regular intervals.
● Carry out a device inspection every day before starting work (as per /1/, /2/, /6//10/).

5.1.1 General information on the device inspection

5.1.1.1 Scope
The device inspection includes the following tests:
● Analysis of the general status
● Test of the indication accuracy with supply of fresh air
● Test of the indication accuracy with supply of test gas
32
Page 40
5.1.1.2 Frequency
The frequency of the device inspection depends on the appli­cation.
Application When to test Legal basis
Warning ExTox Each work day /1/, /2/, /4/, /6/, /10/ Gas measuring Weekly to every six
If the integrated device inspection is switched on, the device will remind you to carry out a device inspection.
5.1.1.3 Documentation
The device inspection procedure must be documented. There are two ways of doing this:
● On paper
● Saved electronically supported by the device (integrated de-
vice inspection)
Only the integrated device inspection is described in these op­erating instructions.
5 Maintenance
/3/
months
Note:
If the integrated device inspection is switched off, the device in­spection must be documented on paper.
5.1.1.4 Integrated device inspection
The integrated device inspection is accessed via the menu (Fig. 6).
The results of the device inspection are stored in the device as a protocol.
The device inspection protocols can be opened at any time. They can also be displayed on a computer using a readout program. The program is available at www.sewerin.com.
33
Page 41
5 Maintenance
The integrated device inspection is switched off in the factory settings. The integrated device inspection has to be switched on
(once only) before it can be performed.
Switching on the integrated device inspection
1. Press Menu.
2. Select Settings.
3. Enter your PIN code.
4. Select System.
5. Select Device inspection.
6. Select Yes.
7. Accept the setting with OK.
8. Exit the settings with Exit.
The Carry out device inspection symbol appears when a device inspection is due. It is visible in the display un­til the complete integrated device inspection has been carried out successfully.
If the device inspection was completed but the device failed on some points, the symbol will remain visible.
5.1.1.5 Order
You can carry out the tests that make up the device inspection in any order you wish. You can repeat the tests as often as you wish provided you have not yet concluded the device inspection.
5.1.1.6 Test gases for the device inspection
The following test gases can be used to check the indication ac­curacy when supplying test gas:
Test Recommended test gases
Gas measuring Gas mixture Bio IR
Warning Gas mixture ExTox IR
34
Page 42
Note:
Gas measuring
Warning
General status
Fresh air
Gas mixture Gas measuring
General status
Fresh air
Gas mixture Warning
Use of test gases not provided by SEWERIN can cause inter­ference.
The concentration of the test gas used must match the specied
test gas concentration.
Changing the test gas concentration
If there is no test gas with the specied concentrations available
for the inspection, the values can be adjusted to the test gas used in the adjustment menu under Test gas (see Section 3.3.3).

5.1.2 Carrying out the device inspection

5.1.2.1 Accessing the device inspection
The device is in measuring mode.
1. Press Device inspection.
OR
a) Press Menu. b) Select Device inspection menu item.
The Device inspection menu appears.
5 Maintenance
Fig. 9: Device inspection menu
2. Select the application for which you wish to perform the de-
vice inspection (Gas measuring or Warning). A submenu will appear.
Fig. 10: Submenus for the device inspection:
Dev. test Gas measuring (left) / Dev. Test Warning (right)
35
Page 43
5 Maintenance
3. Select a test from the menu (Device inspection, Fresh air,
Gas mixture…).
4. Carry out the test.
For detailed information, refer to the following sections:
− General status Section. 5.1.3 on page 37
− Fresh air Section. 5.1.4 on page 39
− Test gas … Section. 5.1.5 on page 40
5.1.2.2 Concluding the device inspection
After all the tests have been carried out as described in Sec­tion 5.1.3 to Section 5.1.5, the Save symbol will appear in the display.
An integrated device inspection is concluded by saving it. Up to 40 device inspections can be saved. The following information can be stored along with the device inspection:
● Inspector (e.g. inspector's name or initials)
● Password to protect the protocol from being accessed by un-
authorised people
Inspector entries are saved automatically (ring memory with max.
10 entries).
Once the rst inspector has been entered, the Open
stored inspectors function will become available.
36
1. Press Save.
2. If necessary enter the name of the inspector.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
OR
− Press Open stored inspectors. A list of the stored in-
spectors will appear.
− Select the desired inspector. Open the inspector with OK.
Page 44
b) Then conrm your entry/selection with OK.
OR
Press Esc if you do not wish to enter an inspector for the device inspection.
3. Enter a password.
a) Select the characters required using the jog dial. Conrm
each character using the jog dial.
b) Then conrm your entry with OK.
OR
Press Esc if you do not wish to enter a password for the device inspection.
The device inspection is saved as a protocol. An overview with the device inspection results is displayed.
4. Conrm the overview by pressing OK. The device returns to
measuring mode.

5.1.3 Testing the general status

The general status test is part of the device inspection. The gen­eral status test is based on estimations by the user. The following must be tested:
● Housing
● Signals
● Probe
● Filter
● Pump
The battery charge status and the working condition of the controls are automatically tested during the integrated device inspection.
5 Maintenance
37
Page 45
5 Maintenance
The device inspection has been opened.
1. Select the General status menu item from the Dev. Test …
submenu.
2. Test all associated subitems as described in Section 5.1.3.1 to Section 5.1.3.5.
3. Conrm the prompt General status OK? by pressing Yes if
all subitems show no faults during testing. General status OK appears on the display.
This concludes the General status test.
5.1.3.1 Housing
● Is the housing free from external damage?
5.1.3.2 Signals
During the integrated device inspection the signals are emitted at short intervals.
● Can the audible signal be heard?
● Is the visual signal visible?
5.1.3.3 Probe
Probes are accessories. They only need to be tested if they are likely to be used in the course of the working day.
● Are the probes free from external damage?
Probe hoses are tested with a simple leak check.
1. Connect the probe hose to the gas input.
2. Seal the free end of the probe hose.
An error message should appear after approx. 10 seconds. This indicates that the probe hose is in good condition.
38
Page 46
5.1.3.4 Filter
The ne dust lter is located behind the gas input. It is tested by
means of a visual inspection.
1. Unscrew the gas input.
2. Remove the ne dust lter.
3. Check that there is no dirt in the ne dust lter.
As soon as there are any signs of deposits, the lter must be replaced. If you do not replace the lter, you must reinsert it
exactly as you found it.
5.1.3.5 Pump
The pump function is tested with a simple leak check.
1. Seal the gas input.
After a maximum of 10 seconds an error message should ap­pear. This indicates that the pump is working correctly.
If the error message does not appear, the pump may be faulty. The device must be tested by SEWERIN Service.
2. Release the gas input again.
After approximately 5 seconds, the error message should disappear again. Otherwise there is a fault.
5 Maintenance

5.1.4 Testing indication accuracy with supply of fresh air

The indication accuracy with supply of fresh air test is part of the device inspection.
The device inspection has been opened.
1. Make sure that only fresh air is being drawn in.
2. Select the Fresh air menu item from the Dev. Test … sub-
menu.
3. Wait until the displayed readings are stable. A Status: OK
message will appear.
4. Press OK to conrm. Fresh air OK will appear on the display.
This concludes the Fresh air test.
If the Status: OK message does not appear within a reasonable amount of time, the air inow does not correspond to the limit val-
39
Page 47
5 Maintenance
ues stored in the device (see Section. 7.3 on page 56). Move the device to another location and repeat the test.
If the Status: OK message still does not appear when the test is repeated, the device must be re-adjusted (see Section. 5.2 on page 41).

5.1.5 Testing indication accuracy with supply of test gas

The indication accuracy with supply of test gas test is part of the device inspection.
All test gases specied in the device must be tested. The number of specied test gases is dependent on the optional additional
equipment.
The following resources are needed for the test:
● Test gas (e.g. gas mixture ExTox IR)
Information on test gases for the device inspection can be found in Section. 5.1.1.6 on page 34.
● Test set for the supply of test gas (e.g. SPE VOL)
Note:
Details of how to use the test set can be found in the accompa­nying operating instructions.
40
The device inspection has been opened.
1. Select the menu item Gas mixture measuring/ Gas mixture warning menu item from the Dev. Test… submenu.
2. Check whether the test gas concentration specied by the device matches the test gas you intend to use. To do this press Information.
3. Add the test gas
4. Wait until the displayed readings are stable. A Status: OK message will appear.
5. Press OK to conrm.
6. Stop the test gas supply.
The device switches back to the Dev. test... menu. A Test gas OK message will appear.
Page 48
Once all tests have been successfully completed, the full test of the indication accuracy with supply of test gas is also complete.
Test gas test unsuccessful
If a Gas mixture… test was not carried out successfully, the message Test gas not OK appears.
A test may be unsuccessful for the following reasons:
Cause Corrective action
Connections leaking Repeat check, checking the
Measurement values outside
the specied limit values (see Section. 7.3 on page 56)

5.2 Adjustment

WARNING! Danger of death due to incorrect adjust-
A
ment
Incorrect adjustment can lead to incorrect measurement results. This means that the user may not be warned about dangerous gas concentrations in time.
● Only specialist technicians may perform adjustments
● Adjustments must be made in well ventilated rooms or
in the open air.
5 Maintenance
seal on the connections
Adjustment required
41
Page 49
5 Maintenance

5.2.1 Scope

Adjustments must be made separately for each measuring range.
● Zero point
● Sensitivity
Note:
For each measuring range always adjust the zero point rst, fol­lowed by the sensitivity.
Only gases for which the readings are outside the predened
limits need to be adjusted during the adjustment process (see Section. 7.3 on page 56).

5.2.2 Test gases for the adjustment

The following test gases can be used for adjustment:
Gas Suitable test gases for
Zero point Sensitivity
CO ● Fresh air ● Gas mixture ExTox IR
CO
CH
2
4
● Fresh air ● Gas mixture ExTox IR
● Fresh air ● Gas mixture ExTox IR
H2S ● Fresh air ● Gas mixture ExTox IR
O
2
● Gas mixture ExTox IR
● Gas mixture Bio IR
and gas mixture Bio IR
and gas mixture Bio IR
and gas mixture Bio IR
● Fresh air
42
If more than one test gas can be used to adjust a gas, it does not have to be adjusted with all test gases. However, adjusting with more than one test gas increases the measurement quality.
Fresh air contains 20.9 % vol. O2 and is therefore used with ox­ygen to adjust the sensitivity.
Page 50
Note:
Use of test gases not provided by SEWERIN can cause inter­ference.
The concentration of the test gas used must match the specied
test gas concentration.

5.2.3 Special features of adjustment with gas mixture

If a SEWERIN gas mixture is used as a test gas, several gases can be adjusted in a single step:
Gases which can be adjusted simultaneously with a SEWERIN gas mixture
5 Maintenance
Test gas: ExTox IR Menu item:
Adjustment gas mixture warning
● Methane CH
4
● Carbon dioxide CO
● Hydrogen sulphide H2S
● Oxygen O
2
2
Test gas: Bio IR Menu item:
Adjustment gas mixture measuring
● Methane CH
● Carbon dioxide CO
4
2
● Hydrogen sulphide H2S
● Oxygen O
2
● Carbon monoxide CO
Zero point
The zero point is adjusted under Fresh air in the menu. The following parameters are adjusted in a single step:
● Zero points of CH4, CO2, H2S and CO
● Sensitivity of O2
Sensitivity
The sensitivity is adjusted under Gas mixture in the menu. The following parameters are adjusted in a single step:
● Sensitivities of CH4, CO2, H2S and CO
● Zero point of O2
43
Page 51
5 Maintenance

5.2.4 Preparation

An adjustment always requires time. Leave yourself plenty of time to prepare the necessary steps of the procedure.
● Have all necessary tools available.
● Let the device run for several minutes to guarantee that the
temperature is correct, for example.

5.2.5 Performing the adjustment

The zero point and sensitivity are adjusted following the same procedure for all gases. The adjustment of oxygen is an excep- tion. It is therefore described separately.
You can nd detailed information on the adjustment
of various gases (for example, test gas concentration,
installation date of the sensor, date of last adjustment)
under Information.
The symbol appears after the corresponding Adjust-
ment… menu item has been selected.
5.2.5.1 Adjusting the zero point
For all gases except oxygen O2, the zero point is adjusted follow­ing the same procedure.
44
Note:
When adjusting the zero point of carbon dioxide CO2, a carbon dioxide lter must be used. This applies to both the zero point adjustment for Adjustment CO2 in the menu and Adjustment gas mixture warning.
1. Make sure that only fresh air is being drawn in.
2. Open Settings.
3. Select the Adjustment menu item.
4. Select the desired adjustment (e.g. Adjustment CH4, Adjust-
ment gas mixture warning).
5. Wait at least 1 minute. The displayed reading must be stable.
Page 52
Note: For Adjustment gas mixture warning and Adjustment gas mixture measuring, all values must be stable. The time required
for this can vary depending on the specic gas.
6. From the menu select the method you wish to use to adjust
the zero point.
− e.g. for Adjustment CH4: Zero point
− for Adjustment gas mixture warning: Fresh air
7. Press OK to conrm.
This adjusts the zero point. The reading shows zero (0.00 % vol. / 0 ppm). (Except for Adjustment gas mixture: the value for oxygen O2 is 20.9 % vol.)
5.2.5.2 Adjusting the sensitivity
For all gases except oxygen O2, the sensitivity is adjusted follow­ing the same procedure.
The following resources are needed for adjusting the sensitivity:
● Test gas
Information on test gases for adjustment can be found in Sec­tion. 5.2.2 on page 42.
● Test set for the supply of test gas (e.g. SPE VOL)
5 Maintenance
Note:
Details of how to use the test set can be found in the accompa­nying operating instructions.
Note:
When adjusting the sensitivity, never use a carbon dioxide lter.
45
Page 53
5 Maintenance
1. Connect the device to the test set.
2. Open Settings.
3. Select the Adjustment menu item.
4. Select the desired adjustment (e.g. Adjustment CH4, Adjust-
5. Select the menu item that species the sensitivity to be tested.
6. Press and hold the release button on the test set. The test gas
7. Wait at least 1 minute. The displayed reading must be stable.
Note: For Adjustment gas mixture warning and Adjustment gas mixture measuring, all values must be stable. The time required
for this can vary depending on the specic gas.
ment gas mixture warning).
− e.g. for Adjustment CH4: 2.20 %VOL. CH4
− for Adjustment gas mixture warning: Gas mixture
Do not conrm with OK yet.
is added. Do not let go of the release button.
8. Press OK to conrm. The device is adjusted. The reading
shows the specied value (e.g. 2.20 % vol. CH4).
9. Let go of the release button on the test set.

5.2.6 Carrying out an oxygen adjustment

As oxygen is a component of fresh air, the procedure for adjusting oxygen is different from the procedure for all other gases.
5.2.6.1 Adjusting the zero point for oxygen
The zero point of oxygen must be adjusted using a gas that does not contain any oxygen and which will not damage the sensor.
The following resources are needed for adjusting the zero point of oxygen:
46
Page 54
5 Maintenance
● Test gas
Information on test gases for adjustment can be found in Sec­tion. 5.2.2 on page 42.
● Test set for the supply of test gas (e.g. SPE VOL)
Note:
Details of how to use the test set can be found in the accompa­nying operating instructions.
1. Connect the device to the test set.
2. Open Settings.
3. Select the Adjustment menu item.
4. Select the Adjustment O2 menu item.
5. Select Zero point menu item. Do not conrm with OK yet.
6. Press and hold the release button on the test set. The test gas
is added. Do not let go of the release button.
7. Wait at least 1 minute. The displayed reading must be stable.
8. Press OK to conrm. The device is adjusted. The reading
shows zero (0.0 % vol.).
9. Let go of the release button on the test set.
5.2.6.2 Adjusting the sensitivity for oxygen
The sensitivity for oxygen is adjusted with fresh air.
1. Make sure that only fresh air is being drawn in.
2. Open Settings.
3. Select the Adjustment menu item.
4. Select the Adjustment O2 menu item.
5. Wait until the displayed reading is stable. (The reading may
still ash.)
6. Select 20.9 % VOL. (fresh air) from the menu (select and
conrm with OK ).
This adjusts the sensitivity. The reading shows 20.9 % vol.
47
Page 55
5 Maintenance

5.3 Servicing

The device must only be serviced and repaired by SEWERIN Service.
● Send the device to SEWERIN for repairs and for annual main-
tenance.
Note:
If there is a service agreement in place, the device can be ser­viced by the mobile maintenance service.
Fig. 11: Inspection plate
The inspection plate on the device shows con-
rmation of the last maintenance and the next
scheduled maintenance.
48
Page 56
6 Faults
If a fault occurs during operation, an error message will appear on the screen.
Error messages are displayed in the order in which they occur.
Up to ve errors can be displayed.
Error messages continue to be displayed until the error is cor­rected.
Overview of possible error messages

6 Faults

Error code
9 No calibration
10 Adjustment failed
11 Zero point
52 XFLASH
59 Error unknown
62 IR sensor Error can only be corrected by
80 Error unknown
100 Pump error
105 Pump error
200 I2C HOST – IR
201 I2C HOST – EC
202 I2C HOST – EX
Error message on the display
IR sensor adjustment
Test gas
Adjustment required
SEWERIN Service
SEWERIN Service
SEWERIN Service
Probe/lter
Gas outlet
SEWERIN Service
SEWERIN Service
SEWERIN Service
Error correction
Adjustment required (Adjustment CH4, Adjustment CO2 or Adjust­ment gas mixture ...)
Check test gas
Adjustment required
Error can only be corrected by SEWERIN Service
Error can only be corrected by SEWERIN Service
SEWERIN Service
Error can only be corrected by SEWERIN Service
Check all lters, probes and hose
connections for porosity and dirt
Check pressure at gas input/gas outlet
Error can only be corrected by SEWERIN Service
Error can only be corrected by SEWERIN Service
Error can only be corrected by SEWERIN Service
49
Page 57

7 Appendix

7 Appendix
7.1 Specications and permitted operating conditions
Dimensions (W×D×H): Approx. 148 × 57 × 205 mm
Approx. 148 × 57 × 253 mm with supporting
bracket Weight: Approx. 1000 g, depending on equipment Operating position: Any Protection rating: IP54 Power supply: 4 cells, either:
– Rechargeable batteries: NiMH
– Disposable batteries: Alkaline Operating time: Minimum 8 h Charging time for re-
chargeable batteries: Charging voltage: 12 V DC (max. 1 A) Operating temperature: -20 ºC – +40 ºC Storage temperature: -25 ºC – +60 ºC Pressure: 800 – 1100 hPa Permissible relative
humidity: Sensors: – IR for CO2, CH
Warm-up time: – < 30 s
PC connection: USB Memory: 8 MB Display: 320 × 240 pixels Buzzer: Frequency: 2.4 kHz
Signal light: Red Pump: Vacuum: > 250 mbar
Pressure at gas input: max. 100 mbar Operation: – ON/OFF key
Approx. 3 h (full charge),
depending on capacity
5 – 90 % r.h., non-condensing
Optional:
– EC for O2, H2S, CO
– Up to 90 s for EC
Volume: 80 dB (A) / 1 m
Volume ow: Typically 50 l/h ±20 l/h Pump error (F100) dependent on volume ow:
≤ 20 l/h F100 certain
> 20 l/h – ≤ 35 l/h F100 possible
– Jog dial
– 3 function keys
4
50
Page 58

7.2 Alarms

Only applies to Warning ExTox application.
WARNING! Danger of death due to hazardous gas
A
concentrations
An alarm always indicates danger.
The type of action depends on the situation. In the event of a
pre-alarm, it may be sufcient to ventilate thoroughly. In the event
of a main alarm, it may be necessary to leave the danger zone immediately.

7.2.1 Features

AL1
Type: Pre-alarm Adjustable: Yes Latching: No Trigger: Alarm threshold AL1 exceeded Indicator: – Audible signal
Acknowledge­ment:
Reset: – Automatic when level falls below alarm threshold
7 Appendix
● Take all necessary measures for your own safety and
the safety of others immediately.
– Visual signal – AL1 message on display
– Possible for audible signal when alarm threshold
AL1 is exceeded
AL1
Exception for oxygen: alarm AL1 is triggered when the level falls below alarm threshold AL1.
51
Page 59
7 Appendix
AL2
Exception for oxygen: alarm AL2 is not latching.
AL3
Type: Main alarm Adjustable: Yes Latching: Yes Trigger: Alarm threshold AL2 exceeded Indicator: – Audible signal
– Visual signal – AL2 message on display
Acknowledge­ment:
Reset: – By acknowledgement after level falls below alarm
Type: End of measuring range Adjustable: No Latching: Yes Trigger: Alarm threshold AL3 exceeded Indicator: – Audible signal
Acknowledge­ment:
Reset: – By acknowledgement after level falls below alarm
– Possible for audible signal when alarm threshold
AL2 is exceeded
– Possible overall after level falls below alarm thresh-
old AL2
threshold AL2
– By switching device off
– Visual signal – AL3 message on display – Reading ashes
– Possible after level falls below alarm threshold AL2
threshold AL2
– By switching device off
52
Page 60
7 Appendix
STEL
Type: Main alarm (short-time exposure limit) Adjustable: No Latching: Yes Trigger: Sum of the concentrations of a gas is greater than
the product of the OEL and the excess factor over the averaging time
Indicator: – Audible signal
– Visual signal – STEL message on display
Acknowledge­ment:
Reset: – By switching device off
– Not possible
LTEL
Type: Main alarm (long-time exposure limit) Adjustable: No Latching: Yes Trigger: Sum of the concentrations of a gas is greater than the
OEL over the averaging time
Indicator: – Audible signal
– Visual signal – LTEL message on display
Acknowledge­ment:
Reset: – By switching device off
– Not possible
53
Page 61
7 Appendix

7.2.2 Occupational exposure limits (OELs) and excess factors (STEL and LTEL)

The short-time exposure limit (STEL) is calculated by multiplying
the OEL value by the excess factor over an averaging time of 15 minutes, as per /15/.
The long-time exposure limit (LTEL) is obtained from the OEL
value over an averaging period of 8 hours, as per /15/.
Gas OEL Excess factor
CO 30 ppm 1
CO
2
H2S 5 ppm 1
0.50 % vol. 2

7.2.3 Alarm thresholds (factory settings)

Gas AL1 AL2 AL3 STEL LTEL
CH410 % LEL 50 % LEL 100 % LEL
CO 30 ppm 60 ppm 500 ppm 30 ppm 30 ppm
CO20.50 % vol. 1.00 % vol. 5 % vol. 1.00 % vol. 0.50 % vol.
O
18.0 % vol. 23.0 % vol. 25 % vol.
2
H2S 5 ppm 20 ppm 100 ppm 5 ppm 5 ppm
54
Page 62

7.2.4 Setting ranges for gas types

LEL values are specied as per /14/.
The setting for AL1 must not exceed the setting for AL2.
Gas type AL1 AL2
All CxHy in % LEL
CH
CO
CO
H2S
O
2
Threshold 10 % LEL 50 % LEL
Threshold 0.45 % vol. 2.20 % vol.
Setting range 0.20 – 2.60 % vol. 0.25 – 2.65 % vol.
4
Increment 0.05 % vol. 0.05 % vol.
Threshold 30 ppm 60 ppm
Setting range 5 – 499 ppm 6 – 500 ppm
Increment 1 ppm 1 ppm
Threshold 0.5 % vol. 1.0 % vol.
Setting range 0.1 – 4.99 % vol. 0.11 – 5.0 % vol.
2
Increment 0.01 % vol. 0.01 % vol.
Threshold 5 ppm 20 ppm
Setting range 5 – 99 ppm 6 – 100 ppm
Increment 1 ppm 1 ppm
Threshold 18.0 % vol. 23.0 % vol.
Setting range 0.5 – 20.5 % vol. 21.2 – 25.0 % vol.
Increment 0.1 % vol. 0.1 % vol.
7 Appendix
Factory settings in bold
55
Page 63
7 Appendix

7.3 Limit values for the device inspection

Warning ExTox
Gas Zero point Sensitivity
Specication Deviation Specication Deviation
CH40.00 % vol. ±0.15 % vol. 2.20 % vol. ±0.20 % vol. CO 0 ppm ±3 ppm 40 ppm ±3 ppm CO20.04 % vol. ±0.06 % vol. 2.00 % vol. ±0.20 % vol. O
0 % vol. ±0.5 % vol. 20.9 % vol. ±0.5 % vol.
2
H2S 0 ppm ±3 ppm 40 ppm ±3 ppm
Gas measuring
Gas Zero point Sensitivity
Specication Deviation Specication Deviation
CH40.00 % vol. ±2 % vol. 100 % vol. ±5 % vol. CO20 % vol. ±2 % vol. 100 % vol. ±5 % vol.
56
Page 64

7.4 Memory capacity

The total memory capacity of the device is divided up as follows:
Protocol type Maximum number of storable protocols
Device inspection 40
Measurement 80
There is a choice of two memory modes (see Section. 3.3.7 on page 26). The selected memory mode applies for all protocol types.
Measurements
Note: A le is saved after each Start measurement – Stop meas- urement cycle, regardless of whether the memory capacity is
exhausted.
Each le has a maximum memory capacity of 1800 data records. This means that a le can record data for 30 mins (0.5 h) at a
save interval of 1 second. After this, data recording continues
automatically in the next le.
7 Appendix
Save interval
1 s 0.5 h 40 h
2 s 1 h 80 h
5 s 2.5 h 200 h
10 s 5 h 400 h
20 s 10 h 800 h
Factory settings in bold
Save time for 1 le (1800 data records)
Save time for 80 les (max. memory capacity)
57
Page 65
7 Appendix

7.5 Sensors

Note:
Probes increase the stated response times.

7.5.1 Infrared sensors (IR)

7.5.1.1 Methane CH4 for warning ExTox
Type: Infrared sensor Measuring range: 0 – 100 % LEL Measuring error:
– Linearity – Short-term stability
– Long-term stability Response times: t50 < 8 s t90 < 14 s Temperature range: -20 ºC – +40 ºC Interference: All hydrocarbons CxH Lifetime, expected: 5 years Test gases:
– Zero point
– Sensitivity
As per /9/
≤ 3 % LEL ≤ 2 % LEL ≤ 4 % LEL (2 months)
Fresh air
2.20 % vol. CH
y
4
7.5.1.2 Methane CH4 for gas measuring
Type: Infrared sensor Measuring range: 0 – 100 % vol. Measuring error CH4: ±2 % vol. Response times: t50 < 9 s t90 < 17 s Temperature range: -20 ºC – +40 ºC Interference: All hydrocarbons CxH Lifetime, expected: 5 years Test gases:
– Zero point
– Sensitivity
58
Fresh air 100 % vol.
y
Page 66
7.5.1.3 Carbon dioxide CO2 for warning ExTox
Type: Infrared sensor Measuring range:
– Lower limit
Measuring error:
– Linearity – Long-term stability
0 – 5 % vol.
0.02 % vol. As per /4/
≤ 4 % from measured value / ≤ 0.02 % vol. ≤ 8 % from measured value / ≤ 0.02 % vol. (3 months)
Zero point deviation: ≤ 0.04 % vol. Response time: t90 < 20 s Decay time: t10 < 14 s Temperature range: -20 ºC – +40 ºC Interference: None Lifetime, expected: 5 years Test gases:
– Zero point – Sensitivity
Fresh air
2.00 % vol. CO
7.5.1.4 Carbon dioxide CO2 for gas measuring
Type: Infrared sensor Measuring range: 0 – 100 % vol. Measuring error: ±1.5 % vol. Response time: t90 < 20 s Temperature range: -20 ºC – +40 ºC Interference: None Lifetime, expected: 5 years Test gases:
– Zero point – Sensitivity
Fresh air 100 % vol. CO
7 Appendix
2
2
59
Page 67
7 Appendix

7.5.2 Electrochemical sensors (EC)

7.5.2.1 Oxygen O2
Type: Electrochemical sensor Measuring range: 0 – 25 % vol. Resolution: 0.1 % vol. Measuring error:
– Linearity
As per /7/
≤ 1.5 % from measured value / ≤ 0.3 % vol.
– Long-term stability
≤ 0.2 % vol. (3 months)
Response time: t90 < 30 s Temperature range: -20 ºC – +40 ºC Interference: None Lifetime, expected: 24 months Test gases:
– Zero point
– Sensitivity
100 % vol. CH4 or 100 % vol. N
Fresh air (20.9 % vol.)
7.5.2.2 Carbon monoxide CO
Type: Electrochemical sensor Measuring range:
– Lower limit Resolution: 1 ppm Measuring error:
– Linearity
– Long-term stability
Zero point deviation: 7 ppm Response time: t90 < 30 s Decay time: t10 < 24 s Temperature range: -20 ºC – +40 ºC Interference at 20 ºC.
– 3000 ppm H
– 100 ppm NO
2
Lifetime, expected: 36 months Test gases:
– Zero point
– Sensitivity
0 – 500 ppm 4 ppm
As per /4/
≤ 2 % of measured value / ≤ 1 ppm ≤ 10 % of measured value / ≤ 1 ppm (3 months)
Approx. 1000 ppm CO Approx. 25 ppm CO
Fresh air 40 ppm CO
2
60
Page 68
7.5.2.3 Hydrogen sulphide H2S for warning ExTox
Type: Electrochemical sensor Measuring range:
– Lower limit Resolution: 1 ppm Measuring error:
– Linearity
– Long-term stability
Zero point deviation: 1 ppm Response time: t90 < 60 s Decay time: t10 < 90 s Temperature range: -20 ºC – +40 ºC Alarm thresholds
(as delivered):
Interference at 20 ºC.
– 100 ppm CO
– 1 % vol. H
– 100 ppm NO
2
2
Lifetime, expected: 24 months Test gases:
– Zero point
– Sensitivity
0 – 100 ppm 1 ppm
As per /4/
≤ 2 % from measured value / ≤ 1 ppm ≤ 10 % from measured value / ≤ 1 ppm (3 months)
AL1: 5 ppm AL2: 20 ppm AL3: 100 ppm
Approx. 1 ppm H2S Approx. 10 ppm H2S Approx. 3 ppm H2S
Fresh air 40 ppm H2S
7 Appendix
61
Page 69
7 Appendix
7.5.2.4 Hydrogen sulphide H2S for gas measuring
Type: Electrochemical sensor Measuring range:
– Lower limit
Resolution: 1 ppm (to 100 ppm)
Measuring error:
– Linearity – Long-term stability
Zero point deviation: 2 ppm Response time: t90 < 60 s Decay time: t10 < 90 s Temperature range: -20 ºC – +40 ºC Interference at 20 ºC.
– 100 ppm CO – 1 % vol. H – 100 ppm NO
2
2
Lifetime, expected: 24 months Test gases:
– Zero point – Sensitivity
0 – 2000 ppm 1 ppm
2 ppm (from 100 ppm)
≤ 3 % of measurement value ≤ 10 % of measurement value / ≤ 5 ppm (3 months)
Approx. 2 ppm H2S Approx. 10 ppm H2S Approx. 4 ppm H2S
Fresh air 180 ppm H2S
62
Page 70

7.6 Technical information

7.6.1 Identication sticker (back of device)
The symbols on the sticker mean the following:
Only ever open the battery compartment outside of explosive areas.
Read the operating instructions.

7.6.2 Cleaning

The device must only be cleaned with a damp cloth.
CAUTION! Damage possible due to unsuitable
A
cleaning agents
Unsuitable cleaning agents can cause chemical corrosion on the housing surface. Vapours from solvents and sub­stances containing silicone can penetrate the device and damage the sensors.
● Never clean the device with solvents, petrol or cockpit
spray containing silicone or similar substances.
7 Appendix

7.6.3 Electrostatic charge

Avoid electrostatically charging the device. Electrostatically un­earthed objects (e.g. including metallic housing without an earth
connection) are not protected against applied charges (e.g. through dust or dispersed ows).
DANGER! Risk of explosion due to sparks
A
When working with hydrogen, electrostatic charging can occur.
● When working with hydrogen, always use the carrying
bag TG8 for the device.
63
Page 71
7 Appendix

7.7 Accessories and consumables

Accessories
Part Order number
Docking station TG8 LP11-10001 AC/DC adapter M4 LD10-10001 Vehicle cable M4, 12 V= portable ZL07-10100 Vehicle cable M4, 12 V= installed ZL07-10000 Vehicle cable M4, 24 V= portable ZL09-10000 "Vario" carrying system 3209-0012 Carrying bag TG8 3204-0040 Case TG8-RÜ ZD29-10000 Compact case TG8 ZD31-10000 Flexible hand probe ZS32-10000 Floating probe ZS21-10100 Probe hose ZS25-10000 (e.g.) Test set SPE VOL PP01-90101 Universal test head PP01-B1700 Gas outlet adapter MG04-Z2000
64
Consumables
Part Order number
Fine dust lter 2499-0020 Hydrophobic lter 2491-0050 Carbon dioxide lter ZG02-11000
Rechargeable NiMH battery 1354-0009 Disposable alkaline battery 1353-0001 Test gas ExTox IR,
test gas can 1 l, pressure approx. 12 bar Test gas Bio IR,
test gas can 1 l, pressure approx. 12 bar
ZT47-10000
ZT49-10000
Other accessories and consumables are available for the prod­uct. Please contact our SEWERIN sales department for further information.
Page 72

7.8 EU declaration of conformity

Hermann Sewerin GmbH hereby declares that the Multitec® 560
fulls the requirements of the following directives:
● 2014/34/EU
● 2014/30/EU
Gütersloh, 2016-04-20
Dr. S. Sewerin
(General Manager)
The complete declaration of conformity can be found online.
7 Appendix
65
Page 73
7 Appendix
– Capacity disposable battery/
rechargeable battery capacity (e.g.: ¼)
Warning ExTox (Test gas Fresh air)
Zero point CH
Zero point CO2 (0.04 % vol.)
Fresh air O2 (20.9 % vol.)
Zero point H2S
CH4(2.20 % vol.)
CO2 (2.00 % vol.)
O2 (0 % vol.)
H2S 40 ppm

7.9 Inspection protocol

Sample inspection protocol for inspection with gas mixture.
TEST PROTOCOL Gas mixture
Serial no. (e.g.: 066 01 0501)
1.0 General status
1.1 – Perfect condition (e.g.: Y / N)
1.2 – Fine dust filter correct (e.g.: Y / N)
1.3
2.0 Pump check
2.1 – Pump error F100 in seal
3.0
3.1
– Display -3 – +3 % LEL or – Display -0.15 – +0.15 % vol.
3.2
– Display -0.02 – +0.10 % vol.
3.3
– Display 20.4 – 21.4 % vol.
3.4 Zero point CO – Display -3 – +3 ppm
3.5
– Display -3 – +3 ppm
4.0 Warning ExTox (Test gas ExTox IR)
4.1
– Display 45 – 55 % LEL or – Display 2.00 – 2.40 % vol.
4.2
– Display 1.80 – 2.20 % vol.
4.3
– Display -0.5 – +0.5 % vol.
4.4 CO (40 ppm) – Display 37 – 43 ppm
4.5
– Display 37 – 43 ppm
4.6 Visual alarm (e.g.: Y / N)
4.7 Audible alarm (e.g.: Y / N)
4
Multitec
®
560
30.03.2011
66
- 1 -
Page 74
5.0 Gas measuring (Test gas fresh air)
Zero point CH
Zero point CO
Zero point H2S
Fresh air O2 (20.9 % vol.)
CH4 (60 % vol.)
CO2 (40 % vol.)
H2S (180 ppm)
O2 (0 % vol.)
5.1
– Display -0.3 – +0.3 % vol.
5.2
– Display -0.3 – +0.3 % vol.
5.3
– Display -3 – +3 ppm
5.4
4
2
– Display 20.4 – 21.4 % vol.
6.0 Gas measuring (Test gas Bio IR)
6.1
– Display 58 – 62 % vol.
6.2
– Display 38 – 42 % vol.
6.3
– Display 175 – 185 ppm
6.4
– Display -0.5 – +0.5 % vol.
7.0 Comments
– Housing damaged – Adjustment, repair – Inspection at factory – Or similar
8.0 Inspection
– Day – Month – Year – Signature
7 Appendix
- 2 -
67
Page 75
7 Appendix

7.10 Advice on disposal

The European Waste Catalogue (EWC) governs the disposal of
appliances and accessories.
Description of waste Allocated EWC waste code
Device 16 02 13
Test gas can 16 05 05
Disposable battery, rechargeable battery
End-of-life equipment
Used equipment can be returned to Hermann Sewerin GmbH. We will arrange for the equipment to be disposed of appropriately
by certied specialist contractors free of charge.
16 06 05
68
Page 76

7.11 Terminology and abbreviations

% vol. Percent concentration of a gas in a gas
mixture with respect to the volume
AL ● Alarm CENELEC ● European Committee for Electrotechnical
Standardization
EC ● Electrochemical sensor Gas type ● Hydrocarbon CxHy, which can be meas-
ured with the IR
● One of the available gas types must al-
ways be set at a time, as it is not possi­ble to measure more than one gas type at the same time.
IR ● Infrared sensor LEL ● Lower Explosion Limit LTEL ● Long-time exposure limit NiMH ● Nickel metal hydride OEL ● Occupational exposure limit ppm ● Parts per million Ring memory ● Type of data storage in the device
If the available storage space is full, the oldest le is automatically overwritten by the current le.
Stack memory ● Type of data storage in the device
● If the available storage space is full, you
are prompted to conrm whether the oldest le should be overwritten by the current le.
STEL Short-time exposure limit
7 Appendix
69
Page 77
7 Appendix

7.12 Referenced documents

The following standards, guidelines and regulations are referred to in these operating instructions:
/1/ BGI T 021
Berufsgenossenschaft Chemie (Chemical Employers' Liability Insur-
ance Association); Code of Practice T 021: Gaswarneinrichtungen
für toxische Gase/Dämpfe und Sauerstoff – Einsatz und Betrieb (Gas Warning Devices for Toxic Gases/Vapours and Oxygen – Use and Operation) Available for download at: www.bgchemie.de
/2/ BGI T 023
Berufsgenossenschaft Chemie (Chemical Employers' Liability Insur-
ance Association); Code of Practice T 023: Gas Warning Devices for
Explosion Protection – Use and Operation Available for download at: www.bgchemie.de
/3/ DVGW G 465-4
Deutsche Vereinigung des Gas- und Wasserfaches e. V. (German
Association of Gas and Water Specialists); Regulation G 465-4: Gasspür- und Gaskonzentrationsmessgeräte für die Überprüfung
von Gasanlagen (Gas-Detection and Gas-Concentration Measure­ment Devices for Inspection of Gas Systems)
Available for download at: www.dvgw.de /4/ EN 45544 /5/ EN 45544-2 /6/ EN 45544-4 /7/ EN 50104:2010 /8/ EN 60079-7:2007 /9/ EN 60079-29-1 /10/ EN 60079-29-2 /11/ EN 60086-1 /12/ EN 61779-1 /13/ EN 61951-2 /14/ IEC 60079-20 /15/ TRGS 900: 2006
Technical regulations for hazardous substances: Occupational expo-
sure limits
Available for download at: www.baua.de /16/ 94/9/EC ATEX 100a
70
Page 78
8 Index

8 Index

A
Accessories 64 Adjustment 23, 41
Carrying out 44 CO 24 CO2 23 Gas mixture 43 Gas mixture measuring 23 Gas mixture warning 23 H2S 24 O2 24, 46 Preparation 44 Scope 42 Sensitivity 45
Zero point 44 Adjustment menu 23 Alarm 5, 26, 51
Setting ranges for gas types 55 Alarm thresholds 54 Application
Selecting 13
Switching 13 Autostart 25
B
Battery alarm 30
C
Carbon dioxide lter 44 Cleaning 63 Clear 26 Comment 18 Consumables 64
D
Date 26 Device
Switching on 11 Device information 19 Device inspection 19, 25, 32
Carrying out 35
Concluding 36
Documentation 33
Frequency 33
Integrated 33 Limits 56 Opening 35 Order 34 Scope 32
Switching on 34 Display 25 Display contrast 25 Display illumination 25 Disposable battery 25
Replacing 30
Requirements 28
Setting the type 25 Disposal 68
E
Electrostatic charge 63 Error message 49 Excess factor 54 Explosion protection
active 8
Passive 7
F
Faults 49 Filter 39 Fine dust lter 39 Function key 9
G
Gas measuring 16 General status 37
H
Housing 38
I
Identication plate 63 Indication accuracy
With fresh air 39
With test gas 40 Inspection OK 24 Interval 26
71
Page 79
8 Index
J
Jog dial 9
K
Keys 9
L
Language 25 LTEL 54
M
Maintenance 32, 48 Measurement
Protocol 18 Starting 17 Stopping 18
Measuring mode 13, 14
Menu structure 15 Memory 26, 57 Memory mode 26 Menu 10, 15
Exiting 10
Opening 15
Selecting 10 Menu item
Exiting 10
Selecting 10 Menu structure 15, 22
O
OEL 54 Operating signal 5 Operation 9 Oxygen 46
P
PIN code 20, 24 Power supply 27 Probe 38 Protocols 18 Pump 39
R
Rechargeable battery 28
Charging 28
Maintenance 29 Replacing 30 Requirements 28 Self-discharge 28 Setting the type 25
Ring memory 26
S
Sensitivity
Adjusting 45, 47
Sensor 6, 58
Electrochemical 6, 60 Infrared 6, 58
Installation date 19 Service interval 24 Settings 14, 16, 20
Menu structure 22
Opening 20 Signal 38
Audible 5
Visual 5 Stack memory 26 STEL 54 System 24
T
Test gas 23
Changing 35
For adjustment 42
For the device inspection 34 Time 26
U
Unit of measurement
Setting 25 Use
Intended 3
W
Warning ExTox 16
Z
Zero point 15
Adjusting 44, 46
72
Page 80
Hermann Sewerin GmbH
Robert-Bosch-Straße 3 33334 Gütersloh, Germany Tel.: +49 5241 934-0 Fax: +49 5241 934-444 www.sewerin.com info@sewerin.com
SEWERIN SARL
17, rue Ampère – BP 211 67727 Hoerdt Cedex, France Tél. : +33 3 88 68 15 15 Fax : +33 3 88 68 11 77 www.sewerin.fr sewerin@sewerin.fr
SEWERIN IBERIA S.L.
Centro de Negocios “Eisenhower” Avenida Sur del Aeropuerto de Barajas 28, Of. 2.1 y 2.2 28042 Madrid, España Tel.: +34 91 74807-57 Fax: +34 91 74807-58 www.sewerin.es info@sewerin.es
Sewerin Ltd
Hertfordshire UK Phone: +44 1462-634363 www.sewerin.co.uk info@sewerin.co.uk
Sewerin Sp.z o.o.
ul. Twórcza 79L/1 03-289 Warszawa, Polska Tel.: +48 22 675 09 69 Faks: +48 22 486 93 44 Tel. kom.: +48 501 879 444 www.sewerin.pl info@sewerin.pl
Loading...