Setra SRIM1 Specifications

Model SRIM1
Room Isolation Monitor
The SRIM1 is Setra’s standard non-BACnet product mea­suring low dierential pressure in critical applications. The SRIM1 is an ideal solution for anyone who requires cost-ef­fective local monitoring and alarming of a single pressure re­lationship, but does not require BACnet protocol. The SRIM1 has a two-line LCD display with easy menu navigation and congurable visual/audible alarm setup. The SRIM1 has eld selectable output and uses a unique removable faceplate de­sign, allowing the user to fully calibrate the unit without the hassle of removing plumbing or wiring.
Why Use Anything But the Best?
In the most critical of environments where contamination can result in nancial loss and even the loss of life, it is no secret why Setra dierential pressure sensors are the most trusted in the industry. Setra’s capacitive stainless steel de­sign oers the most stable very low dierential pressure sen­sor for building automation systems.
n Maximize Patient Safety n Save on Calibration & Installation n Low-Cost Reliable Solution
Premium Performance at Affordable Price
The SRIM1 is designed for the facility that needs local alarm­ing on pressure, without paying a premium for the bells and whistles of higher priced devices. The SRIM1 has a congu­rable audible/visual alarm for pressure, which is easily cong­ured through the 2-line LCD display.
Save Time and Money on Calibration
With requirements to calibrate pressure sensors anywhere from 1-3 times annually, the Setra SRIM1 oers a solution to help you save on calibration time. The SRIM1 allows the end user to remove the sensor without detaching any wiring or plumbing, saving time and money by completing the cali­bration in minutes.
Model SRIM1 Features:
• Dead- Ended On-board Sensor - Industry Best Accuracy
• 2-Line LCD Display for Easy Setup and Room Display
• Congurable Audible & Visual Alarm
• Flush Mount (43mm wall depth) and Surface Mount
Available
• Field Congurable Output (0-5VDC, 0-10VDC, 4-20mA)
• Removable Faceplate for No Hassle Calibration
• True Selectable 2-Wire 4-20mA Output
Where We’re Installed:
• Battelle Laboratories
• Bella Vista Hospital
• Fort Lauderdale Hospital
• Harvard Medical School
• Russell Medical Center
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Model SRIM1
74
2.9
144
5.67
[ mm ] in.
74
2.9
144
5.67
[ mm ] in.
74
2.9
144
5.67
112.8
4.44
157.2
6.19
6.9
0.27
PG9
KNOCKOUT
4.7
0.19
43.4
1.71
[ mm ] in.
144
5.67
[ mm ] in.
[ mm ] in.
Room Isolation Monitor
ORDERING INFORMATION
S R I M
Model Pressure Ranges Type** Output** Mounting/Logo Accuracy Pressure Snubber
SRIM1 = SRIM1
Example: Part No. SRIM2R05WD11WLC1 = Model SRIM2, 0 to 0.05 in. W.C. Pressure Range, Unidirectional, 4 to 20 mA Output, Wall Mount with Logo, ±1.0% FS Accuracy, 1 Snubber
1
(in W.C.) (Pascals)
R05W* 0 to 0.05 ±0.05 Z02L 0 to 12.5 ±12.5
0R1W 0 to 0.1 ±0.1 025L 0 to 25 ±25
R25W 0 to 0.25 ±0.25 050L 0 to 50 ±50
OR5W 0 to 0.5 ±0.5 100L 0 to 100 ±100
001W 0 to 1.0 ±1.0 250L 0 to 250 ±250
2R5W 0 to 2.5 ±2.5 500L 0 to 500 ±500
005W 0 to 5.0 ±5.0 10CL 10CL 10CL
25CL 0 to 2500 ±2500
Dierential
11 4 to 20mA WL Wall Mount w/ Logo
C ±1.0% FS N None
D Unidirectional 2B 0 to 5 VDC DL Duct Mount w/ Logo F ±0.25% FS w/ cal. cer t 1 Quantity 1
B Bidirectional 2C 0 to 10 VDC WN Wall Mount, No Logo H ±.05% FS w/ cal. cert 2 Quantity 2
*available in unidirectional only
**Field Congurable but factory congured for cal certs.
DN Duct Mount, No Logo G ±1.0% FS w/ cal. cert
DIMENSIONS GENERAL SPECIFICATIONS
157.2
112.8
4.44
43.4
1.71
1/2" CONDUIT
KNOCKOUT
6.19
PRESSURE FITTING FOR 3/16 I.D. / (4.4mm) TUBING
4.7
0.19
Performance Data Environmental Data
Code F Code H Code C/G Operating Temp. °F (°C) 22 to +140 (-6 to +60)
Accuracy RSS* ±0.25% FS ±0.5% FS ±1.0% FS Operating Humidity 5 to 95% RH (Non-Condensing)
Non-Linearity (BFSL) ±0.22% FS ±0.49% FS ±0.98% FS
Hysteresis ±0.1% FS ±0.1% FS ±0.1% FS
Non-Repeatability ±0.05% FS ±0.05% FS ±0.05% FS Output (eld selectable) 0 to 5 VDC, 0 to 10 VDC, 4 to 20 mA
Zero/Span Setting Tol. ±0.5% FS ±0.5% FS ±1.0% FS Excitation 18 to 32 VDC
Thermal Eects Current Consumption 5 mA (voltage output mode)
Compensated Range
40 to 120 (4.5 to 50)
°F (°C)
Zero/Span Shift
±0.02% FS/deg
C
Typ. Electrical Connection Screw Terminal
%FS**
Overpressure Up to 10 PSI (Range Dependent) Weight 8.9 oz (Duct) 9.8 oz (Wall)
Pressure Media
Air or non-conductive, non-explosive gases. Pressure Fittings Barbed Fittings for 1/4” Tubing
Certications
CE
RoHS
EN61326-1 & EN61326-2-3 BASIC Immunity & Class B Emission
Electrical Data
Circuit 3-Wire (Exc, Out, Com)
2-Wire (+Exc, Com)
Physical Description
Display Custom 2-Line Character LCD
Case Fire Retardant Plastic UL94V-0
*RSS of Non-Linearity, Non-Repeatability & Hysteresis at constant temp. **Units calibrated at nominal 21ºC. Max thermal error computed from this datum.
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
43.4
1.71
SSP-SRCM REV. D 12/13