Model SRCM
Room Pressure and Condition Monitor
GA UGE PRESSURE
ROOM PRESSURE MONITORS
FEATURES
■ True Pressure Measurement
• High accuracy Setra low dierential technology
• Dead ended solution —no contamination or clogging
• Standard on-board sensor and optional remote sensor
■ Display 4 Ambient Parameters
• Pressure, Temperature, Humidity, User-Dened (ex., CO2, LUX)
■ Flush Mount Design
• No visible mounting fasteners
• Snap-in ush bezel
• Face is sealed for cleaning or wipe-down
NOTE: Setra quality standards are based on ANSI-Z540-1.
The calibration of this product is NIST traceable.
U.S. Patent nos. 6019002; 6014800
DESCRIPTION
In a hospital, laboratory or animal research facility, the
integrity of the ventilation control system is at the heart
of a contaminant free environment. Whether a room is
to be maintained at a negative pressure to prevent contaminants from escaping into adjacent areas or positive
pressure to protect patients with compromised immune
systems, the proper pressurization of the room is essential. To assure proper pressurization is maintained in
these critical environments, a room pressure monitor is
employed to measure and alert sta and personnel of
any change in pressure—no matter how small. A fail-safe
solution to monitoring these very low pressure changes is
Setra’s Model SRCM room pressure monitor, which utilizes
highly accurate capacitance sensing technology to measure and display true low pressure dierential.
"BACnet" is a registered
trademark of ASHRAE"
Setra’s “2 in 1” Solution
■ Full Banner Feature
• Utilize same monitor for room condition
• Clearly display condition with facility specic nomenclature
■ Clone Feature
• Display rotates up for access to USB port
• On-board USB port—cloning of congurations
for multiple unit installation
■ BACnet® Communications
• Installed or eld upgrade for in-situ installation
• BACnet• MSTP/ASC
• All setups congurable through touch screen display
■ Alarm Capabilities
• Local Audible and Visual alarming
• Remote annunciator alarming capability
• Alarm delay feature—prevent nuisance alarms
■ Ease of Installation
• Mounts in o-the-shelf electrical gang box
• 4-screw self leveling mount
1
The Environment is Critical, the Control is Easy
Ambient Parameters
+
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
=
SRCM
ROOM PRESSURE MONITORS
Attractive, Intuitive, and Congurable Design
Simultaneously Monitor and Display 4 Parameters Per Room (up to 2 rooms)
Selectable Room Condition Banners
1/3 Screen Banners
Full Sceen Banners
Room Pressure and Condition Monitor
Display 4 Ambient Parameters (slide bar o)
Pressure, Temperature, Humidity, and User-Dened
Parameter
Setup Menu
Password Protected Entry to
Menu Structure
Room I.D.
Customer Congurable Feature - Customize Room Name
Pressure Slide Bar On
Graphic Display of Pressure
Reading
Relative to Alarm Setpoints
Model SRCM
Room Pressure Mode
Display Positive, Negative, or Neutral Mode of
Room
Room Pressure Status
Visual Indication of Normal,
Warning, and Alarm Con-
dition
Displayed Pressure
Displays Room Pressure
in User Selectable Units
GA UGE PRESSURE
Setup Display Banner
• User Dened Text: Free form data
entry for room name
• Room Status: Change room from
Isolation to No Isolation
Fingertip Access for Easy Setup
Setup Display Advanced
• Display Contrast: Change brightness of display
• Display Averaging: Improve display resolution
in unstable
ambient pressure environments
• Display Ambient Parameters: Display primary
and
secondary rooms or toggle between 2 rooms
• Enable Password: Administrator and Supervisor
Setup Unit Operation
• Setup primary and secondary room
• Change analog output
• Free form data entry for room name
©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
2
Model SRCM
Room Pressure and Condition Monitor
SPECIFICATIONS
Performance Data Environmental Data Electrical Data (Voltage)
3
Operating
Code F Code H
1
(at constant temp) ±0.25% FS ±0.5%
Accuracy RSS
Non-Linearity (BFSL based) ±0.24% ±0.49%
GA UGE PRESSURE
Hysteresis ±0.05% ±0.05%
Non-Repeatability ±0.05% ±0.05%
ROOM PRESSURE MONITORS
Zero/Span Tolerance ±0.05% FS ±0.05% FS
Pressure Media
Air or Non-Conductive, Non-Explosive Gases.
Certications
Electro-Magnetic Compatibility
CE
CSA
1
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2
Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
3
Operating Temperature limits of the electronics only.
4
Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater.
Specications subject to change without notice.
Directive 2004/108/EC
Low-Voltage Directive 2006/95/EC
RoHS Directive 2011/65/EC
CAN/CSA - C22.2 No. 61010-1-04
ANSI/UL 61010-1, 3rd Edition
Temperature
°F (°C)
Storage Temperature °F (°C) -20 to +160 (-30 to +70) Output
Operating Humidity 5 to 95% RH (non-condensing) Excitation 18-32 VAC, 50-60 HZ
Physical Description
Case Fire Retardant Plastic UL94 V-0 Alarm Output SPDT Relay: 0.6A @ 120 VAC, 2A @ 30 VDC
Dimensions 5.84”H x 7.45”W x 0.38”D (14.84 x 18.92 x 0.95 cm)
Electrical Connection Removable Terminal Block Circuit 2-Wire
Pressure Fittings Barbed Fittings for 1/4” O.D. Tubing Output 4 to 20 mA
Weight (approx.) 1 lb. 3.2 ounces (554 grams) External Load 0 to 510 ohms
Mounting Mounts to a triple gang double-deep electrical box Excitation 18-32 VAC
Communications Option
BACnet® MS/TP ASC Compensated Range °F (°C) ±0.03% FSI (±0.05% FS)
Display Overpressure ±1 PSI (15” W.C. for ≤ 0.10” W.C. F.S.)
LCD 4.3” TFT, 480 x 272, Dimmable
32 to + 120 (0 to +50) Circuit 3-Wire (Exc, Out, Com),
4
Power Consumption 10 W max., 3 W typ.
Electrical Data (Current)
Thermal Eects
0 to 5 VDC, 0 to 10 VDC
2
DIMENSIONS
14.84
5.84
WIRING
18.92
7.45
POWER
Earth Ground
HIGH or room pressure tubing
contacts here
L1 24 Vac
L2 24 Vac
RELAY
COM COM
+15 +15 Vdc
NO Normally Open
RC Relay COM
NC Normally Closed
COM
B (+)
A (-)
GND
0.95
0.38
5.08
2.00
ANALOG (ANL) OUTPUT
COM COM
mA Current Output (4-20 mA)
Vdc Voltage Output (0-5 Vdc, 0-10 Vdc)
INPUTS (wired to door N.O. contact)
COM COM Door
DI Digital Input
COM COM Analog Input2
IN2 Analog Input2 (0-5, 0-10Vdc)
+15 15 Vdc
COM COM Analog Input1
IN1 Analog Input1 (0-5, 0-10Vdc)
+15 15 Vdc
LOW or reference pressure tubing
connects here
1.49
0.59
7.03
2.77
13.15
5.18
cm
in.
3
REAR VIEW
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
ROOM PRESSURE MONITORS
Model SRCM
Room Pressure and Condition Monitor
ORDERING INFORMATION
S R C M
Model Range Code Excitation/Output Accuracy Pressure Snubber* Faceplate
SRCM = SRCM See Table 1 Below A1 24 VAC/4-20 mA or 0-5 and 0-10
VDC
A2 24 VAC w/ BACnet® F ±0.25% FS
Table 1. Range Specication
RANGE
CODE
R05WB
0R1WB
R25WB
0R5WB
001WB
2R5WB
005WB
INCHES
W.C. RANGE CODE PASCALS
±0.05
±0.10
±0.25
±0.50
±1.00
±2.50
±5.00
Z02LB
025LB
050LB
100LB
250LB
500LB
10CLB
±12.5
±25
±50
±100
±250
±500
±1000
H ±0.5% FS N
0
1 1
2 2
"BACnet" is a registered
trademark of ASHRAE"
S Setra
B Blank
GA UGE PRESSURE
Logo
Ordering Example: Part No. SRCMR05WBA1HNS for A SRCM, ±0.05”WC Range, 24VAC/4-20 mA, 0.5% Full Scale Accuracy,
NO Pressure Snubber
* For other pressure tting congurations, please contact factory.
ACCESSORIES
Model SRAN
Remote Annunciator
Green LED, Normal Indication
Red LED, Alarm Indication
Buzzer, Audio Alarm, ADJ. from SRPM Acknowledge Switch
Order Part Number:
S R A N
Pressure Snubber
Room Pressure Snubber (Wall Mount Pressure Taps)
The RPS is a stainless steel room static pressure sensor that has the same footprint (2.75”
W X 4.5” H) as your standard electrical wall
plate.
©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
SSP-SRCM Rev. D 12/18/13
4