Model SRCM
Room Pressure and Condition Monitor
GAUGE PRESSURE
ROOM PRESSURE MONITORS
FEATURES
True Pressure Measurement
• High accuracy Setra low dierential technology
• Dead ended solution —no contamination or clogging
• Standard on-board sensor and optional remote sensor
Display 4 Ambient Parameters
Flush Mount Design
• No visible mounting fasteners
• Snap-in ush bezel
• Face is sealed for cleaning or wipe-down
NOTE: Setra quality standards are based on ANSI-Z540-1. The calibration of this product is NIST traceable.
U.S. Patent nos. 6019002; 6014800
DESCRIPTION
In a hospital, laboratory or animal research facility, the integrity of the ventilation control system is at the heart of a contaminant free environment. Whether a room is to be maintained at a negative pressure to prevent con­taminants from escaping into adjacent areas or positive pressure to protect patients with compromised immune systems, the proper pressurization of the room is es­sential. To assure proper pressurization is maintained in these critical environments, a room pressure monitor is employed to measure and alert sta and personnel of any change in pressure—no matter how small. A fail-safe solution to monitoring these very low pressure changes is Setra’s Model SRCM room pressure monitor, which utilizes highly accurate capacitance sensing technology to mea­sure and display true low pressure dierential.
"BACnet" is a registered
trademark of ASHRAE"
Setra’s “2 in 1” Solution
Full Banner Feature
• Utilize same monitor for room condition
• Clearly display condition with facility specic nomenclature
Clone Feature
Display rotates up for access to USB port
• On-board USB port—cloning of congurations
for multiple unit installation
BACnet® Communications
• Installed or eld upgrade for in-situ installation
• BACnet• MSTP/ASC
• All setups congurable through touch screen display
Alarm Capabilities
• Local Audible and Visual alarming
• Remote annunciator alarming capability
• Alarm delay feature—prevent nuisance alarms
Ease of Installation
• Mounts in o-the-shelf electrical gang box
• 4-screw self leveling mount
1
The Environment is Critical, the Control is Easy
Ambient Parameters
+
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
=
SRCM
Page 2
ROOM PRESSURE MONITORS
Attractive, Intuitive, and Congurable Design
Simultaneously Monitor and Display 4 Parameters Per Room (up to 2 rooms)
Selectable Room Condition Banners
1/3 Screen Banners
Full Sceen Banners
Room Pressure and Condition Monitor
Display 4 Ambient Parameters (slide bar o)
Pressure, Temperature, Humidity, and User-Dened
Parameter
Setup Menu
Password Protected Entry to Menu Structure
Room I.D.
Customer Congurable Feature - Cus­tomize Room Name
Pressure Slide Bar On
Graphic Display of Pressure Reading Relative to Alarm Setpoints
Model SRCM
Room Pressure Mode
Display Positive, Nega­tive, or Neutral Mode of
Room
Room Pressure Status
Visual Indication of Normal, Warning, and Alarm Con-
dition
Displayed Pressure
Displays Room Pressure in User Selectable Units
GAUGE PRESSURE
Setup Display Banner
• User Dened Text: Free form data entry for room name
• Room Status: Change room from Isolation to No Isolation
Fingertip Access for Easy Setup
Setup Display Advanced
• Display Contrast: Change brightness of display
• Display Averaging: Improve display resolution in unstable ambient pressure environments
• Display Ambient Parameters: Display primary
and
secondary rooms or toggle between 2 rooms
• Enable Password: Administrator and Supervisor
Setup Unit Operation
• Setup primary and secondary room
• Change analog output
• Free form data entry for room name
©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
2
Page 3
Model SRCM
Room Pressure and Condition Monitor
SPECIFICATIONS
Performance Data Environmental Data Electrical Data (Voltage)
3
Operating
Code F Code H
1
(at constant temp) ±0.25% FS ±0.5%
Accuracy RSS
Non-Linearity (BFSL based) ±0.24% ±0.49%
GAUGE PRESSURE
Hysteresis ±0.05% ±0.05%
Non-Repeatability ±0.05% ±0.05%
ROOM PRESSURE MONITORS
Zero/Span Tolerance ±0.05% FS ±0.05% FS
Pressure Media
Air or Non-Conductive, Non-Explosive Gases.
Certications
Electro-Magnetic Compatibility
CE
CSA
1
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2
Units calibrated at nominal 70˚F. Maximum thermal error computed from this datum.
3
Operating Temperature limits of the electronics only.
4
Calibrated into a 50K ohm load, operable into a 5000 ohm load or greater.
Specications subject to change without notice.
Directive 2004/108/EC Low-Voltage Directive 2006/95/EC RoHS Directive 2011/65/EC
CAN/CSA - C22.2 No. 61010-1-04 ANSI/UL 61010-1, 3rd Edition
Temperature
°F (°C)
Storage Temperature °F (°C) -20 to +160 (-30 to +70) Output
Operating Humidity 5 to 95% RH (non-condensing) Excitation 18-32 VAC, 50-60 HZ
Physical Description
Case Fire Retardant Plastic UL94 V-0 Alarm Output SPDT Relay: 0.6A @ 120 VAC, 2A @ 30 VDC
Dimensions 5.84”H x 7.45”W x 0.38”D (14.84 x 18.92 x 0.95 cm)
Electrical Connection Removable Terminal Block Circuit 2-Wire
Pressure Fittings Barbed Fittings for 1/4” O.D. Tubing Output 4 to 20 mA
Weight (approx.) 1 lb. 3.2 ounces (554 grams) External Load 0 to 510 ohms
Mounting Mounts to a triple gang double-deep electrical box Excitation 18-32 VAC
Communications Option
BACnet® MS/TP ASC Compensated Range °F (°C) ±0.03% FSI (±0.05% FS)
Display Overpressure ±1 PSI (15” W.C. for ≤ 0.10” W.C. F.S.)
LCD 4.3” TFT, 480 x 272, Dimmable
32 to + 120 (0 to +50) Circuit 3-Wire (Exc, Out, Com),
4
Power Consumption 10 W max., 3 W typ.
Electrical Data (Current)
Thermal Eects
0 to 5 VDC, 0 to 10 VDC
2
DIMENSIONS
14.84
5.84
WIRING
18.92
7.45
POWER
Earth Ground
HIGH or room pressure tubing contacts here
L1 24 Vac L2 24 Vac
RELAY
COM COM +15 +15 Vdc NO Normally Open RC Relay COM NC Normally Closed
COM
B (+) A (-) GND
0.95
0.38
5.08
2.00
ANALOG (ANL) OUTPUT
COM COM mA Current Output (4-20 mA) Vdc Voltage Output (0-5 Vdc, 0-10 Vdc)
INPUTS (wired to door N.O. contact)
COM COM Door DI Digital Input
COM COM Analog Input2 IN2 Analog Input2 (0-5, 0-10Vdc) +15 15 Vdc
COM COM Analog Input1 IN1 Analog Input1 (0-5, 0-10Vdc) +15 15 Vdc
LOW or reference pressure tubing connects here
1.49
0.59
7.03
2.77
13.15
5.18
cm
in.
3
REAR VIEW
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Page 4
ROOM PRESSURE MONITORS
Model SRCM
Room Pressure and Condition Monitor
ORDERING INFORMATION
S R C M
Model Range Code Excitation/Output Accuracy Pressure Snubber* Faceplate
SRCM = SRCM See Table 1 Below A1 24 VAC/4-20 mA or 0-5 and 0-10
VDC
A2 24 VAC w/ BACnet® F ±0.25% FS
Table 1. Range Specication
RANGE
CODE
R05WB 0R1WB R25WB 0R5WB 001WB 2R5WB 005WB
INCHES
W.C. RANGE CODE PASCALS
±0.05 ±0.10 ±0.25 ±0.50 ±1.00 ±2.50 ±5.00
Z02LB 025LB 050LB 100LB 250LB 500LB 10CLB
±12.5 ±25 ±50 ±100 ±250 ±500 ±1000
H ±0.5% FS N
0
1 1
2 2
"BACnet" is a registered
trademark of ASHRAE"
S Setra
B Blank
GAUGE PRESSURE
Logo
Ordering Example: Part No. SRCMR05WBA1HNS for A SRCM, ±0.05”WC Range, 24VAC/4-20 mA, 0.5% Full Scale Accuracy, NO Pressure Snubber
* For other pressure tting congurations, please contact factory.
ACCESSORIES
Model SRAN
Remote Annunciator
Green LED, Normal Indication Red LED, Alarm Indication Buzzer, Audio Alarm, ADJ. from SRPM Acknowl­edge Switch
Order Part Number:
S R A N
Pressure Snubber
Room Pressure Snubber (Wall Mount Pressure Taps)
The RPS is a stainless steel room static pres­sure sensor that has the same footprint (2.75” W X 4.5” H) as your standard electrical wall plate.
©2013 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
SSP-SRCM Rev. D 12/18/13
4
Loading...