Setra SPC5000 Manual

Page 1
SPC5000 Series
Operating Instructions
Setra Systems, Inc.
159 Swanson Road, Boxborough, MA 01719
800.257.3872 • www.setra.com
Page 2
Copyright 2018 by Setra Systems, Inc., All Rights Reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of Setra Systems, Inc. The information contained in this document constitutes proprietary trade secrets of Setra Systems. You are not allowed to disclose or allow to be disclosed such information except as allowed by Setra Systems in writing. No patent liability is assumed with respect to the use of the information contained herein. While every precaution has been taken in the preparation of this manual, Setra Systems, Inc. assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of the information contained herein.
Neither Setra Systems, Inc. nor its affiliates shall be liable to the purchaser of this product or third parties for damages, losses, costs, or expenses incurred by purchaser or third parties as a result of: accident, misuse, or abuse of this product or unauthorized modifications, repairs, or alterations to this product, or failure to strictly comply with Setra Systems operating and maintenance instructions.
Setra Systems, Inc. shall not be liable against any damages or problems arising from the use of any options or any products or accessories other than those designated as Original Setra Systems Products or Setra Systems approved.
Setra Systems, is a registered trademark of Setra Systems, Inc.
MODBUS is a registered trademark of Schneider Automation Inc.
Microsoft™, Windows™ and Excel™ are trademarks of Microsoft Corporation
NOTICE: The contents of this manual are subject to change without notice.
Product Name: Setra Systems, Inc. 5000 Series Remote Particle Counters
Model Numbers: 5301 & 5501
The following standards are applied only to the particle counters that are so labeled. EMC is tested using Setra Systems power supplies.
North America: EMI: FCC/ICES-003 Class A
FCC Compliance Statement for American Users
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their own expense.
2
Page 3
Declaration of Conformity
In accordance with EN ISO/IEC 17050-1:2010
Manufacturer’s Name: Setra Systems, Inc.
Manufacturer’s Address: 159 Swanson Road
Boxborough, MA 01719 USA
Application of Council Directives
EMC: 2014/30/EU
RoHS 2: 2011/65/EU
Low Voltage Directive: 2006/95/EC
Application of Council Directives
EMC: EN 61326-1:2013
CISPR 11:2009+A1:2010
RoHS 2 Technical Documentation: EN 50581:2012
Safety Requirements: EN 61010-1:2010
Product Name: Handheld and Remote Particle Counters
Product Model Numbers:
Handheld followed by: 8306, 8303, 8506, or Remote followed by: 7301-AQM, 7302-AQM, 7301, 7501, 5301, 5501, 5301-AQM, 5302-AQM
We, the undersigned, hereby declare that the equipment specified above conforms to the above Directives and Standards.
Boxborough, Massachusetts
3
Page 4
Table of Contents
Contents
1-1 Important Safety Information .......................................................................................................... 6
1-2 Ergonomic Recommendations........................................................................................................ 7
1-3 Warnings for Use of Wireless Devices ........................................................................................... 7
1-4 Overview .......................................................................................................................................... 8
1-5 Specifications ....................................................................................... Error! Bookmark not defined.
1-6 Included Accessories .................................................................................................................... 11
1-7 Optional Accessories .................................................................................................................... 12
1-8 Product Views ................................................................................................................................ 14
2-1 Unpacking and Inspecting the Instrument ................................................................................... 15
2-2 Registering Your Product .............................................................................................................. 15
2-3 Contacting Setra Systems ............................................................................................................. 15
2-4 Storing and Shipping the Instrument............................................................................................ 15
2-5 Power Considerations and Connecting to AC Power .................................................................. 16
2-6 Turning the Unit On ....................................................................................................................... 16
3-1 Control and Menu Icons (ALL MODELS) ...................................................................................... 17
4-1 Operational Flow Chart – Menu Map ............................................................................................. 21
5-1 Operation – Initial Power Up – First Time Use .............................................................................. 22
5-2 Display ........................................................................................................................................... 22
5-3 Taking a Sample ............................................................................................................................ 23
5-4 Real-Time Meter, Graphing, and Environmental Screens ............................................................ 23
5-5 Recorded Data ............................................................................................................................... 26
5-6 Reports ........................................................................................................................................... 27
6-1 Settings .......................................................................................................................................... 30
6-2 Sampling Setup ............................................................................................................................. 30
6-3 Channel Management .................................................................................................................... 32
6-4 Locations & Recipes...................................................................................................................... 33
6-5 Configuration ................................................................................................................................. 37
6-6 Printer Setup .................................................................................................................................. 38
6-7 Communication ............................................................................................................................. 39
6-8 Environment................................................................................................................................... 42
6-9 Passwords ..................................................................................................................................... 42
6-10 Screens ........................................................................................................................................ 43
7-1 Power Management ....................................................................................................................... 44
8-1 Volume Controls ............................................................................................................................ 45
4
Page 5
9-1 Instrument Management Software (IMS)....................................................................................... 45
10-1 Remote Operation ........................................................................................................................ 46
Appendix – A .......................................................................................................................................... 46
Appendix – B .......................................................................................................................................... 46
!
5
Page 6
1-1 Important Safety Information
This section presents important information intended to ensure safe and effective use of this product. Please read this section carefully and store it in an accessible location.
Do not use near explosive, flammable, or reactive gases
Do not attach directly to pressurized gases or liquids
Do not improperly discard electronic instruments, only dispose of in accordance with local
regulatory requirements or contact Setra Systems for trade-in option
Defective or non-working Lithium-Ion batteries must be recycled, do not throw in trash
This device contains a Class I laser product that is not accessible during normal operation, do not
take this device apart, exposure to harmful laser radiation can occur
Taking the device apart will void all warranties
Do not use this device for any purpose other than measuring of particles in ambient environments
Do not operate the instrument with the inlet capped or plugged, as this can cause damage to the
vacuum pump
Do not allow water or any other liquid to enter the inlet of the particle counter; this will damage the unit
Any changes or modifications to Setra Systems equipment not expressly approved by Setra Systems could void the user’s authorization to operate the equipment, can risk serious injury, and will void all warranties
Key to Symbols
The symbols in this manual are identified by their level of importance, as defined below.
Read the following carefully before handling the product.
WARNING:
Warnings must be observed carefully to avoid serious bodily injury.
CAUTION:
Cautions must be observed to avoid minor injury to yourself or damage to your equipment.
NOTE: The laser in this product is completely enclosed within a sensor with no user serviceable parts. In addition, the emission level does not exceed the AEL (Accessible Emission Limit) of Class 1 under all conditions of operation, maintenance, service and failure.
6
Page 7
1-2 Ergonomic Recommendations
CAUTION: In order to prevent or reduce the potential risks of ergonomic injury, follow
the recommendations below. Consult with your local Health & Safety Manager to ensure that you are adhering to your company’s safety programs to prevent employee injury.
Reduce or eliminate repetitive motion
Maintain a natural position while holding the instrument
Reduce or eliminate excessive force
Keep objects that are used frequently within easy reach
Perform tasks at correct heights
Utilize a tripod or the built-in stand with the instrument in a freestanding mode
1-3 Warnings for Use of Wireless Devices
Please observe all warning notices with regard to the usage of Setra Systems' particle counters with optional Wi-Fi communications module installed.
Safety in Hospitals
Wireless devices transmit radio frequency energy and may affect medical electrical equipment. Wireless devices should be switched off wherever you are requested to do so in hospitals, clinics, or health care facilities. These requests are designed to prevent possible interference with sensitive medical equipment.
Pacemakers
Pacemaker manufacturers recommend that a minimum of 15cm (6 inches) be maintained between a wireless device and a pacemaker to avoid potential interference with the pacemaker. These recommendations are consistent with independent research and recommendations by Wireless Technology Research.
Persons with Pacemakers:
Should ALWAYS keep the device more than 6 inches (15cm) from the pacemaker if turned ON
Should not carry the device on your chest
Should use the arm furthest from the pacemaker to minimize the potential for
interference
If you have any reason to suspect that interference is taking place, turn OFF your device
Other Medical Devices
Please consult your physician or the manufacturer of the medical device to determine if the operation of your wireless product may interfere with the medical device.
7
Page 8
1-4 Overview
Thank you for purchasing a Setra Systems particle counter, the most advanced battery powered, remote instrument available.
This user manual will provide the detailed explanation and instructions for the proper use and operation of this feature-rich particle counter.
The Setra Systems particle counters have a large dynamic range measuring from 0.3 µm to 25.0 µm, with true variable binning for channel size adjustment settings to 0.01 µm. This instrument utilizes 7 or more processors to maintain and manage the various functions of operation. The advanced processing also allows for many operations to take place simultaneously, even while the unit is sampling. This includes adding annotations to the current sample in progress, or adding annotations to previously recorded data while sampling is in progress.
The Real-Time Meter function is unique in its ability to fine-tune the instrument's sensitivity in order to locate particulate sources with visual and audible indications. This versatile particle counter's ability to count higher than typical particle concentrations allows the Real-Time Meter to find point source contamination in cleanrooms as well as locating higher particle concentrations being generated in many industrial environments.
This instrument also has a mass concentration mode, providing for particle mass monitoring of an environment for industrial health and safety regulatory purposes. The unit can measure all (6) adjustable particle size channels and capture PM levels indicated in µg/m³ with values corrected for particle density and refractive index correction.
The technology designed into these particle counters includes advanced power management functions, which includes the industry’s first sleep mode. This permits the instrument to take intermittent samples over the course of a few months, minimizing power consumption and increasing system life.
The Setra Systems particle counters have versatile communication methods and protocols including: Modbus TCP, USB Host, USB Client and (optional) Wi-Fi, (optional) MODBUS RTU and ASCII.
The remote web server hosting feature allows for monitoring and control of the particle counter from any PC, smartphone or tablet simply by inputting the IP address of the particle counter on the local area network and entering it into any browser. The main processors allow for multiple connections, all with simultaneous access to review, monitor and control the operation of the instrument.
Thank you,
8
Page 9
1-5 Specifications
Size Channels (Model 5301)
Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning
Size Range (Model 5301)
0.3 to 25 μm
Counting Efficiency (Model 5301)
50% @ 0.3 μm; 100% for particles >0.45 μm per JIS
Size Channels (Model 5501)
Factory calibrated at 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm variable binning
Size Range (Model 5501)
0.5 to 25 μm
Counting Efficiency (Model 5501)
50% @ 0.5 μm; 100% for particles >0.75 μm per JIS
Flow rates
0.1 CFM (2.83 LPM) (optional long-life internal brushless pump available)
Light Source
Long life laser diode
Zero Count
<1 count / 5 minutes (<2 particles / ft³) (per ISO 21501-4 & JIS)
Count Modes
Automatic, manual, cumulative/differential, mass concentration, count or concentration
Count Alarms
1 to 9,999,999 counts
Calibration
NIST traceable
Display
4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Printer (Optional)
External thermal printer
Vacuum Requirements
External vacuum >15” (38.1 cm) of Hg
Vacuum Source (Optional)
Internal 0.1 CFM pump assembly (Part Number: UM-27100A)
Filtered Exhaust (Optional)
Internal HEPA filter (included when ordering UM-27100A)
Number of Channels
6
Custom Size Channels
Calibration for custom size channels available
Audible Alarm
Adjustable built-in alarm
Reports
ISO 14644-1, EU GMP Annex 1, FS 209E
Recipes
50 user-configurable recipes
Communication Modes
Ethernet and USB
Optional Communication Modes
Wireless 802.11 b/g, RS485 or RS232
Environmental Sensor
Optional temperature and relative humidity probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC), 15-90% ±2% relative humidity (optional)
Alarm
Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow
Standards
ISO 21501-4 and JIS B9921
9
Page 10
Calibration
Recommended minimum once per year
External Surface
Stainless Steel
Dimensions (L x W x H)
5.2” x 2.3” x 8.9” (13.3 cm x 5.8 cm x 22.6 cm) includes barb fittings
Weight
2.6 lb (1.18 kg)
Accessories
Operating manual & quick start guide on USB flash drive, barbed fitting, IMS software, & power supply
Optional Accessories
Printed manual, carrying case, external printer, wall mount bracket, purge filter assembly, USB cable and isokinetic probes
Buffer Memory
45,000 sample records (rotating buffer) including particle count data, environmental data, locations and times. Scrollable on screen or printout.
Sample Locations
Up to 1,000 locations 20 characters long
Sample Time
1 second to 99 hours
Power
110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions
41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing
Storage Conditions
32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing
Warranty
2 Years. Extended warranties available.
Please note that specifications are subject to change without notice.
!
10
Page 11
1-6 Included Accessories
Description
Part Number
Image
Isoprobe Threaded 0.1 CFM (2.83 LPM) Nickle Plated Aluminum
PS-12041
Power Supply 15V ~2amp 100-240VAC (Select adapter -US, -EU, -UK or -CN)
EE-80081-XX
5000 Series User Manual and Data Download Software (USB Key)
MN-24004
11
Page 12
1-7 Optional Accessories
Description
Part Number
Image
1/8" Inlet Barbed Fitting Nickle Plated Aluminum
PS-12004
Wireless 802.11 b/g Output with Internal Antenna
EE-80092
Temperature / RH Probe 32-122°F (0-60°C) ±1°F (0.5°C), 15 - 90% ±2%
EE-80014A
External Thermal Printer with 2 Rolls of Paper
AS-99011
External Thermal Printer Paper - 1 Roll
AS-99012
External Thermal Printer Cleanroom Paper - 10 Pack
AS-99013
External Thermal Printer Spare Battery
AS-99014
Purge Filter Assembly 0.1 CFM (2.83 LPM)
AS-99002A
USB Cable 6' (1.8m)
AS-99010
12
Page 13
Description
Part Number
Image
Isoprobe Threaded 0.1 CFM (2.83 LPM) Stainless Steel
PS-12070
Isoprobe Barbed 0.1 CFM (2.83 LPM) Stainless Steel
PS-12022
1/8" Inlet Barbed Fitting Stainless Steel
PS-12005
Sample Tubing 1/4" OD (1/8" ID) Per Foot
AS-99018
5000 Series User Manual (Printed)
MN-24004P
5000 Series Validation Manual
MN-24006
Certificate of Origin
MN-24000
BACnet Gateway
229998
!
13
Page 14
1-8 Product Views
Connecting Ethernet Cable
Connecting USB - B Client
Connecting USB - A Host
Front View
Optional Temp & RH Probe Universal Sensor Port Optional Isokinetic Probe
4.3” Color Touch Screen Display
External Vacuum Connection Port (not present when optional internal pump is installed)
1/8” Barbed Fitting
Side View
External Power Connection
USB-B Client Connection
UBS-A Host Connection
Ethernet RJ45 Connection
!
!
! !
14
Page 15
2-1 Unpacking and Inspecting the Instrument
Careful consideration was given to our packing material to ensure that the Setra Systems
instrument will reach you in perfect condition. If the instrument has been subject to excessive handling during shipping, there may be visible damage to the shipping carton. In the event of damage, keep the shipping container and packing material for the carrier’s inspection. Carefully unpack the instrument from its shipping container and inspect the contents for damaged or missing items. If the instrument appears damaged or something is missing, contact the carrier and Setra Systems immediately. Please save the container and packing material in case you have to return the instrument.
2-2 Registering Your Product
All Setra Systems instruments are automatically registered upon sale for the warranty period, and tracked by Serial number.
2-3 Contacting Setra Systems
To order accessories, receive technical assistance, report damaged or missing items from your shipment, or get contact information for your nearest Setra Systems authorized reseller, call +1-800-257-3872 or +1-978-264-0292
2-4 Storing and Shipping the Instrument
This instrument utilizes a high quality advanced Lithium Ion Power Cell. This must be removed from the device prior to shipping the unit. If the instrument needs to be packed and shipped for annual calibration or service, it is recommended to use the original packing materials. If they are not available, please insure that the instrument is packaged in a box that is sturdy and that the instrument is well protected with proper packing materials to cushion and protect it from harm during transit.
To store the instrument, place it in its optional case or in a box, under cover, in an environment as stated in the Specifications in Section 1-5.
! !
15
Page 16
2-5 Power Considerations and Connecting to AC Power
The Setra Systems instrument comes with a power adapter line cord for AC powered operation and battery charging. The power adapter is designed to operate with line voltage from around the world. The correct plug adapter must be used to match your local AC power adapter standard. If the instrument power adapter does not have the proper plug configuration, please contact Setra Systems or an authorized reseller for service.
To install the country specific plug adapter, simply slide the adapter into the power supply as shown.
2-6 Turning the Unit On
The 5000 Series instruments are designed to operate as fixed installed instruments, and automatically power on when power is applied to the external power connector. The instrument can be powered down from the Power Management Screen (see section 7.1) but will only power back up if the power connection is removed and reconnected.
16
Page 17
3-1 Control and Menu Icons (ALL MODELS)
Function Name
Location/Screen
Description of Function
Setra Logo Icon
Home Screen
Press the logo and the product information screen will appear displaying your model number, serial number, manufacture date, last calibration date, next calibration due date, and number of channels activated on the unit. Press the Back Arrow icon on bottom left corner of the display to return to Home screen.
Speaker Icon
Home Screen
Press the speaker icon to adjust volume using a pop-up slider bar. Press the cross bar on slider and move up and down for volume control. Icon will have red prohibited symbol when sound is turned off or muted.
Temperature & RH Indicator
Home Screen
Press the temperature and RH value indicator, to open a larger screen that displays Temperature, Relative Humidity, and the current Barometric Pressure. Press the Back Arrow icon on bottom left corner of the display to return to Home screen.
Time and Date Indicator
Home Screen
Press the time and date to change the current time and date.
Power Management Icon
Home Screen
Press the power adapter indicator to display the power management screen.
USB Icon
Home Screen
If a USB drive is connected, the USB Icon will appear. Press the USB Icon to save the current record to the USB drive.
Printer Indicator
Home Screen
If the Setra Systems Printer is connected to USB Port, press the Printer Icon to print the current record.
System Warning Indicator
Home Screen
When the System Warning Indication Icon appears, press to display additional information. Please contact Setra Systems technical service for assistance should the Icon remain visible.
17
Page 18
Function Name
Location/Screen
Description of Function
Alarm Indicator
Home Screen
Visual indication that the instrument has exceeded the user-defined thresholds. Press icon to silence the alarm.
Run Icon
Home Screen
Press the Run icon to start sampling. Once pressed the Run icon will be replaced by the Stop icon.
Stop Icon
Home Screen
Press the Stop icon to stop the instrument from sampling.
Display Mode Icon
Home Screen
Press the Display Mode icon to switch between the Main Screen, Real-Time Meter, Graphing, and PM screen (must enable Mass Mode by going to Settings, then Channel Management for PM Screen).
Recorded Data Icon
Home Screen
Press the Recorded Data icon to display the saved data records page. All saved records can be accessed from this screen.
Reports Icon
Home Screen
Press the Reports icon to display ISO 14644-1, EU-GMP Annex 1, or Federal Standard 209E.
Particle Icon
Home Screen
Press the Particle icon to change the indicated values from particle count to count per cubic meter, count per cubic foot, or particle mass concentration (must enable Mass Mode by going to Settings, then Channel Management to display particle mass concentration).
Differential Mode Icon
Home Screen
Press the Differential Mode icon to toggle the differential data values on and off.
Cumulative Mode Icon
Home Screen
Press the Cumulative Mode icon to toggle the differential data values on and off.
Location Menu Icon
Home Screen
Press the Location Menu icon to display the Location and Recipe set-up page. This feature allows for the input of up to 1,000 locations and up to 50 unique user-defined recipes.
18
Page 19
Function Name
Location/Screen
Description of Function
Mode Indicator
Home Screen
The Mode Indicator displays the current mode of operation for the instrument. Modes include automatic, manual, and continuous.
Sample Indicator
Home Screen
The Sample Time indicator displays the current sample time duration (Hours:Minutes:Seconds). This value will count down from the set value for the sample time, displaying the amount of time left in the current sample (go to Settings, then Sampling Setup to change Sample Timing).
Hold Indicator
Home Screen
The Hold Time indicator displays the current hold time as an interval between samples. The maximum hold time is 99 hours, 59 minutes and 59 seconds (go to Settings, then Sampling Setup to change Hold Time).
Cycle Indicator
Home Screen
The Cycle indicator displays the number of count samples that will be taken at a location in automatic mode. The maximum number of possible cycles is 9,999. The value is displayed as the sample number vs. the total number of samples to be completed in this cycle (go to Settings, then Sampling Setup to change Cycles).
Record Indicator
Home Screen
The Record indicator displays the total number of sampling records saved in the instrument. The instrument is capable of storing 45,000 records in a rotating buffer.
Settings Menu Icon
Home Screen
Press the Settings Menu icon to open the Settings Screen. All aspects of the instrument’s set-up can be managed from the icon driven sub­menus.
Annotation Icon
Home Screen
Press the Annotation icon to notate up to 32 characters for each record. This action can be performed during sampling or after a sample has been taken. The green pencil indicates an annotation exists for any record (go to Settings, then Configuration to enable Annotations).
19
Page 20
Function Name
Location/Screen
Description of Function
Plus and Minus Buttons
Home Screen
Press the Plus or Minus icons to scroll through 1,000 possible locations that can be saved and uniquely identified in the Locations setup screen. Locations can have set recipes assigned to them in advance for ease-of-use during sampling.
Flow & No Flow Indicator
Home Screen
The three horizontal arrows indicate that the pump is working and that the internal flow sensor is detecting the correct flow rate through the instrument. If a red line appears diagonally through the arrows, it is an indication of a flow error.
Back Arrow Icon
Various Screens
Press Back Arrow icon to return to the previous screen.
20
Page 21
4-1 Operational Flow Chart – Menu Map
Environment
Date / Ti me
Set
Management
Audio Volume
Set
Recipes
Advance
Data
Record
Information
Print Current
Record
Setup
Management
Locations &
Recipes
Configuration
Setup
Communications
Setup
Environment
Setup
Setup
Samples
Channel Alarm
Setup
Meter
A
A
ISO 14644-1
Definition
Generate
Annex 1
Definition
Generate
Report
Definition
Generate
Setup
Configuration
Load/Save
Records
Restore
Admin Password Not Required
Admin Password Required to Edit
Admin Password Required to View
Print
Shut Down
Information
Power
Zoom
Main
Real-Time
Location &
Location
Charts
Recorded
Reports
Settings
Annotate
Print
Room
Printer
Sampling
Modbus
Report
Channel
Room
EU-GMP
Print
Report
Password
Room
Clear
Factory
FS 209E
Report
21
Page 22
5-1 Operation – Initial Power Up – First Time Use
Date Format Select Button
Choose format for Month/Day/Year, Day/Month/Year, or Year/Month/Day by selecting the corresponding button on the screen.
Numeric Keypad
Press Date within the window and a numeric keypad will appear to change the date. To change the values, use the < or > to move cursor. When complete press OK button.
Time Format Select Button
Choose 12 hour or 24 hour clock indication by selecting the corresponding button on the touchscreen.
Numeric Keypad
Press Time within the window and a numeric keypad will appear to change time. Use 24 hour clock format for time entry to properly indicate AM or PM. To change the values, use the < or > to move cursor. When complete press OK button. Time will display AM or PM or 24 hour format based on the selection.
Back Arrow Icon
Press Back Arrow icon to return to the previous screen.
Standard View
Zoomed View
After the Particle Counter turns on for the first time, a window will appear stating “Time of Day Clock Not Set”. Press OK to Set Clock.
5-2 Display
Press any blank space on the screen to zoom in or out on the Home page.
22
Page 23
5-3 Taking a Sample
Taking a Sample
Using the Run icon on the display begins the sample
To take a sample, press the Run icon on the screen. This will begin the sample according to the sampling setup parameters displayed on the right side of the Home screen (go to Settings, then Sampling Setup to change Sample Timing).
Stopping the Sample
Using the Stop icon on the display stops the sample
To stop a sample, press the Stop icon on the screen.
Data Unit of Measure Selection
Press the Particle icon to change the indicated values from particle count to count per cubic meter, count per cubic foot, or particle mass concentration (must enable Mass Mode by going to Settings, then Channel Management to display particle mass concentration).
Main Sampling Home Screen
5-4 Real-Time Meter, Graphing, and Environmental Screens
The instrument will display a bar graph visualization that rises and falls with the increase of pulses counted per second, per channel. This can be used to pinpoint the source of particles within an area. The closer the instrument is to the source, the higher the indication appears on the bar graph.
23
Page 24
Start/Stop Sampling while on Real-Time Meter or Graphing Screen
The Sampling can be started or stopped from these screens using the Run or Stop icons or by using the one touch button.
Channel Select – Radio Button
Select the channel size that is the focus of the particles being investigated by clicking on the radio select button below each channel. The channel selected also represents the data being displayed in pulses per second on the Graphing screen.
Range – Sensitivity Adjustment Slider
Press and slide the Range slider bar up towards the top of the slider to increase sensitivity and down towards the bottom of the slider to decrease sensitivity. As the instrument gets closer to the particle source the visual signal can increase and hit 100% of the indicated scale long before the actual source of the contamination is found. By moving the Range slider bar down, the sensitivity is reduced, and the indication is scaled down. This action can be repeated until the source is identified.
Switching between Main Screen, Real-Time Meter, Graphing, and Environmental Screens
Press the Real-Time Meter icon to switch between the Main screen, Real-Time Meter, Graphing, and Environmental screens (must enable Mass Mode by going to Settings, then Channel Management to display Environmental screen). NOTE: Go to Settings, then to Screens to select which screens will be available.
Range
Channel 3 selected and signal at maximum value
Range slider lowered to reduce Channel 3 value
Real-Time Meter Operation
24
Page 25
Environmental Screen Size Select
Press Back Arrow to
screen
Select the PM size
displayed
Real-Time Graph
Pressing the PM value
(shown below)
Real-Time Graph Operation
If the Graph function is chosen, this feature graphically displays the pulses per second, making the graph’s historical information useful in point source detection.
Environmental Screen
This screen displays specific PM size, temperature and humidity.
accesses the Select PM To Display screen
channel to be
return to the previous
25
Page 26
Home Screen Mass Concentration
Print or Save the Current Record being displayed
If a Printer or USB thumb drive is attached to the instrument, the current record on the Recorded Data Screen can be downloaded or printed by pressing the corresponding Printer or USB icon.
Turn off Cumulative or Differential columns
All displayed channel values can be turned off or back on from the recorded sample record by toggling the mode icons.
Data Units
Data Units can be changed to the corresponding calculated values by pressing the Particle icon. Press the icon to change between particle count, count per cubic meter, count per cubic foot, or µg/m³ (if enabled).
µg/m³ is displayed in the PM is the sum of all
size value
first column for that particle size up to the next size as a value
previous channels not including that channel
5-5 Recorded Data
The instrument stores up to 45,000 records that can be accessed by selecting the recorded data icon. The Recorded data page uses a horizontal slider bar for scrolling left and right through all records. Press white arrows for fine control in locating a specific record. Press the slider button with the stylus and drag left or right to navigate through large amounts of records quickly. NOTE: After the 45,000th sample is recorded, the software deletes one block of 250
records. After the block is deleted the next record number will become 44,751. The instrument will then store data to record 45,000 before repeating this process. Each block removed is from the oldest record first, following a first in / first out method.
Recorded Data Screen
26
Page 27
Adding Annotations to Recorded Data
Press the Annotation icon to notate up to 32 characters for each record. This action can be performed during sampling or after a sample has been taken. The green pencil indicates an annotation exists for any record (go to Settings, then Configuration to enable Annotations).
Annotations Keyboard
Saved Annotation Icon
The green pencil over a document icon denotes that an annotation exists on that record. The annotation can be accessed, edited or deleted.
Caps Lock, ?123, and Delete Keys
Statistics
Pressing the Shift Key will activate or deactivate the Caps Lock function. Pressing the ?123 Key will toggle the use of numbers and symbols. Pressing the Backspace Key will delete all text entered on the text line.
Pressing the Statistics icon will provide an onscreen display of the minimum, maximum, and average values for the records selected. This function is disabled if a printer or USB stick is connected to the instrument.
Report Functions
5-6 Reports
27
Page 28
ISO 14644-1
Press the ISO 14644-1 icon to generate a report based on the parameters and guidelines of the ISO standard.
EU-GMP Annex-1
Press the EU-GMP ANNEX 1 icon to generate a report based on the parameters and guidelines of the EU-GMP standard.
FS 209E
Press the FS 209E icon to generate a report based on the parameters and guidelines of the US Federal standard 209E.
Room parameters can be defined by pressing the Room Definition button. To generate a report, select
Generate
Create Report
Press the Generate button to display the Report Generation screen
the Generate button. This creates a report document that can be saved to an external thumb drive, or printed to a connected printer.
Press the Create Report button for 5 seconds to display a sample report with the current test records. These records will be added to the recorded data so the values and information can be reviewed prior to completing the report and saving to an external source.
Create Report
Press the Start and End Windows to enter the Start and End dates and times. Press the Create Report button to generate the report. If the values of the count or sampling set-up are incorrect, the report will indicate the errors prior to the report being generated.
To exclude an outlier measurement from the report, go to Recorded Data, find the specific recorded data to be excluded and annotate the record by writing “EXCLUDED” or “EX”. A brief description must be added after the “EXCLUDED” or “EX” explaining why the data is being excluded. Example Annotations: “EXCLUDED Sample Stopped Early” or “EX Equipment Failure”.
Room Definition
28
Page 29
Each Report has a Room Definition page to input data that defines the output of the
Printing or Saving a Report
With a thumb drive or printer connected, press the Printer icon to print a report, or press the USB icon to save a report. NOTE: A thumb drive or printer must be connected to the
instrument to display these icons.
report. The inputs include Room Size, Class, Allowed (Channel) Sizes, Room Status and Air Flow.
Sample Report Screen
Sample Report (Printed)
29
Page 30
6-1 Settings
Delay
Press the Delay window to open the Enter Time screen. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Delay time is 99 hours, 59 minutes and 59 seconds. When complete press OK button. NOTE: This allows the operator to leave an area before sampling begins.
Sample
Press the Sample window to open the Enter Time screen. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Sample time is 99 hours, 59 minutes and 59 seconds. When complete press OK button.
Settings Screen Press the Settings Menu icon to open the Settings Screen. All
6-2 Sampling Setup
aspects of the instrument’s set-up can be managed from the icon driven sub-menus. When the administrator password is in use, this screen is not available and may only be accessed and settings modified by the administrator.
Sampling Setup Screen Press the Sampling icon to display the Sample Setup screen.
30
Page 31
Volume
Press the Volume window to open the Enter Sample Volume screen. To select the value to change, use the < or > to move cursor. The volume value entered will control the length of time per sample to achieve the desired sample volume.
Sample or Volume Selection Buttons
Selecting the Sample button will cause the sample to be time based. Selecting the Volume button will cause the sample to be based on the volume of air to be measured.
Hold
Press the Hold window to open the Enter Time screen. This controls the amount of time the instrument will hold between samples in automatic mode. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Hold time is 99 hours, 59 minutes and 59 seconds. When complete press OK button.
Cycles
Press the Cycles window to open the Enter Cycles screen. This controls the number of sampling cycles to be taken at a specific location when the unit is in automatic mode. To select the value to change, use the < or > to move cursor. When complete press OK button.
NOTE: Enter (0000) to activate Continuous Sampling Mode.
Mode
Press the Mode window to open the Select Sample Mode screen. Select Automatic or Manual. When complete press OK button.
Sample Volume Units
Select Cubic Feet (ft^3), Cubic Meters (m^3) or Liters.
Use Recipes Button
The Use Recipes button activates Recipes on the Select Location & Recipe screen. Recipes associated with locations will be utilized in place of the general settings from the Sample Setup screen.
Clear All Samples
Press Clear Samples icon to open the Clear Samples window. All records on the instrument can be cleared with this function. NOTE: This action is not reversible and all data will be cleared by pressing the Yes button.
Back Arrow Icon
Press the Back Arrow icon to return to the previous screen.
31
Page 32
6-3 Channel Management
Enable Channel
Each channel can be turned on or off by pressing the corresponding radio button. When the channel is turned off all values related to that channel are ignored and will not be displayed or recorded.
Size µm
Press the Size µm window to open the Enter Channel Size screen. This controls the particle size measured and recorded for the selected channel. This feature is also known as variable binning, and is ideal for focusing on specific particle sizes. To select the value to change, use the < or > to move cursor. When complete press OK button.
Enable Alarms
Each channel can have an alarm enabled or disabled by pressing if the User Recipes button is not selected. If enabled, Press the Alarm Threshold window to open the Enter Alarm Threshold screen. To select the value to change, use the < or > to move cursor. When complete press OK button.
Alarm On
Press the Alarm On button to open the Channel Alarm Setup screen. Select Δ (Differential Count), Δft³ (Differential Cubic Feet), Δm³ (Differential Cubic Meter), Σ (Cumulative Count), Σft³ (Cumulative Cubic Feet Count), Σm³ (Cumulative Cubic Meter Count), PM (µg/m³).
Channel Management Screen
This screen provides control of many rich features such as enabling or disabling channels, setting custom channel sizes, enabling or disabling alarms by channel, and the setting of the alarm thresholds. The mass concentration mode may also be activated to allow entry of particle density and refractive index values by channel.
Mass Mode Settings
32
Page 33
Mass Mode Enable
Press the Mass Mode button to enable Mass Mode. The instrument will now display particle count data as calculated particle mass concentration in weight/volume units. The international SI unit for mass is (kg/m³), which can be translated to micro grams per milliliter (µg/ml).
Density g/ml
Press the Density g/ml window to open the Enter Density Factor screen. To select the value to change, use the < or > to move cursor. When complete press OK button.
Refractive Index
Press the Refractive Index window to open the Enter Refractive Index screen. To select the value to change, use the < or > to move cursor. When complete press OK button.
Calculation of displayed Values on Main Screen for Mass Concentration Mode
µg/m3 Indicated Values
When Mass Mode is selected and µg/m³ is selected on the Main screen, the instrument is measuring in Mass Concentration Mode. The mass value for a channel size is the particle count between that channel and the next larger channel, calculated using the average particle size of the two channels. For example, the value in the 0.50 µm data field represents the mass of all particles counted between this channel and the next highest channel, calculated as the mass of a particle that is 0.75 µm.
Particle Mass Indicated Values
When Mass Concentration Mode is selected, µg/m³ is the measured value in the first column. The column labeled PM shows the total particle mass of particles that are less than the displayed channel size. For example, the value displayed in the PM column for the 2.5 µm channel is the particle mass (µg/m³ column) of all particles with a size less than 2.5 µm, generally referred to as PM2.5.
6-4 Locations & Recipes
Press the Locations & Recipes icon to open the Select Location & Recipe screen. Up to 1,000 unique location names can be created. Pressing the User Recipes button allows the creation of location specific recipes which utilize preset sample Delay, Sample or Volume, Hold, Cycles, Mode and Channel Alarm.50 unique recipes can be created and assigned to any number of locations. NOTE: This feature is only useful if your device is portable, and not fixed­mounted.
Select Location & Recipe Screen
33
Page 34
Location Auto Advance
Select the Location Auto Advance button to auto advance to the next location.
Slider
The Select Location & Recipe screen uses a vertical slider bar for scrolling through all locations. Press white arrows for fine control. Press the slider button with the stylus and drag to navigate through large amounts of locations quickly.
Change Sequence of Locations
Press the silver and green up and down arrow buttons on the left side of the locations list to change the sequence of locations.
New Location Entry
Press the New button to open the Enter Name For New Location screen. Use the keyboard to enter the name of the new location. Press OK to save the new location name or ESC to return to the previous screen.
Edit Location
Press the Edit button to open the Enter Name For New Location screen. Use the keyboard to edit the name of the location. Press OK to save the new location name or ESC to return to the previous screen.
Remove Location
Press the Remove button to delete the currently highlighted location.
Use Recipes
Press the Use Recipes button to display the recipes settings. Select the Location in the Location box and the recipe in the Recipe box to assign the recipe to that location. The location can have one of 50 possible recipes assigned to it.
34
Page 35
New Recipe Entry
Press the New button to create a new recipe. The Enter Name for New Recipe screen will appear. Use the touchscreen keyboard to enter the name of the new recipe. Press OK to return to the previous screen.
Edit Recipe
Press the Edit button to modify the existing recipe highlighted on the Location list. Pressing this button will display the Enter New Name for Recipe screen. Use the touchscreen keyboard to modify the name of the existing recipe. Press OK to return to the previous screen.
Remove Recipe
Press the Remove button to delete the recipe that is highlighted on the Location list.
Used By Tab
The Used By window displays all locations currently set for the current recipe.
Edit 1 Tab
Press the Edit 1 tab to adjust the settings for Delay, Sample, Volume, Hold, Cycles, and Mode for the selected recipe.
Edit 1 Delay
Press the Delay window to open the Enter Time screen. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Delay time is 99 hours, 59 minutes and 59 seconds. When complete press OK button. NOTE: This allows the operator to leave an area before sampling begins.
Edit 1 Sample
Press the Sample window to open the Enter Time screen. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Sample time is 99 hours, 59 minutes and 59 seconds. When complete press OK button.
Edit Recipe for Selected Location
Press the Edit button on lower righthand side of the screen to open the Edit Recipes screen.
35
Page 36
Edit 1 Volume
Press the Volume window to open the Enter Sample Volume screen. To select the value to change, use the < or > to move cursor. The volume value entered will control the length of time per sample to achieve the desired sample volume.
Sample or Volume Selection Buttons
Selecting the Sample button will cause the sample to be time based. Selecting the Volume button will cause the sample to be based on the volume of air to be measured.
Hold Edit
Press the Hold window to open the Enter Time screen. This controls the amount of time the instrument will hold between samples in automatic mode. To select the value to change, use the < or > to move cursor. The entry is made in Hours:Minutes:Seconds (HH:MM:SS). The maximum Hold time is 99 hours, 59 minutes and 59 seconds. When complete press OK button.
Edit 1 Cycles
Press the Cycles window to open the Enter Cycles screen. This controls the number of sampling cycles to be taken at a specific location when the unit is in automatic mode. To select the value to change, use the < or > to move cursor. When complete press OK button.
NOTE: Enter (0000) to activate Continuous Sampling Mode.
Edit 1 Mode
Press the Mode window to open the Select Sample Mode screen. Select Automatic or Manual. When complete press OK button.
Edit 2 Tab
Press the Edit 2 tab to enable alarms for each channel size.
Back Arrow Icon
Press the Back Arrow icon to return to the previous screen.
36
Page 37
6-5 Configuration
Language
Select and highlight the language to be used with your device.
Store Partial Samples
Select the Store Partial Samples button to save values from a prematurely ended sampling. If this option is left unchecked, the instrument will ignore partial sampling events.
Alarm Acknowledge
Select the Alarm Acknowledge button and the visual alarm bell and the audible sounder will continue to sound until the alarm bell icon is pressed on the Main screen.
Enable Annotations
Select the Enable Annotations button to allow annotations to be entered for each sample record. These annotations will be included in downloaded record data.
Number Format
Select between 1,000.0, 1.000,0 or 1 000,0
Factory Restore
Press the Factory Restore button to open the Confirm Factory Restore screen. Select Settings to Restore and press the Confirm button.
USB Icon
When a USB thumb drive is plugged into the USB host port, the USB icon will appear on the Configuration screen. Press the USB icon to open the Configuration Load/Save screen. Select Settings to Load/Save and press Load Settings or Save Settings. NOTE: This feature can be used to load the instrument’s configuration to any other Setra Systems Particle Counter.
Press the Configuration icon to select Language, Store Partial Samples, Alarm Acknowledgement, Enable Annotations, Number Format, Save Configuration to a USB thumb drive, and Factory Restore.
Configuration Screen
37
Page 38
Configuration
Settings to Load/Save
Select the check boxes to choose Locations and Recipes and/or User Settings to load or save to a USB thumb drive.
Load Settings
Press the Load Settings icon to initiate loading a previously saved settings configuration into the instrument.
Save Settings
Press the Save Settings icon to initiate saving the current configuration to a USB thumb drive.
Back Arrow Icon
Press the Back Arrow icon to return to the previous screen.
Include in Printout
Select each item to be included in printed values and reports.
Automatic Printing
Select Print Options Automatic button to enable automatic printing On Sample or On Alarm.
Load/Save Screen
6-6 Printer Setup
Press the Printer Setup icon to open the Printer Setup screen.
38
Page 39
6-7 Communication
Wi-Fi Communication
The instrument can support an optional Wi-Fi communication module. Select the Wi-Fi radio button for Wi-Fi connection.
Use DHCP
When the DHCP button is selected the instrument will obtain an IP address, subnet mask and gateway information from the router automatically.
Wi-Fi IP Address, Subnet Mask, & Gateway
If the DHCP button is not selected Press the IP Address, Subnet Mask and Gateway windows to enter the desired values for your router. A numeric keypad will be displayed
Ethernet Communication
Select the Ethernet radio button for Ethernet connection.
Use DHCP
When the DHCP button is selected the instrument will obtain an IP address, subnet mask and gateway information from the router automatically.
IP Address, Subnet Mask, & Gateway
If the DHCP button is not selected Press the IP Address, Subnet Mask and Gateway windows to enter the desired values for your router. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
The instrument has multiple modes of communication for uploading or downloading data or configurations for operation. The modes of communication include Ethernet, RS485, RS232, USB Host or Client, and Optional Wi-Fi or Optional BACnet. Press the Communication icon to open the Communications screen.
Communications Screen
39
Page 40
for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
Wi-Fi SSID
Press the Wi-Fi SSID window to enter the Wi-Fi SSID and press OK.
NOTE: Case Sensitive
Wi-Fi Password
Press the Wi-Fi Password window to enter the Wi-Fi Password and press OK.
NOTE: Case Sensitive
Modbus Setup
Press the Modbus Setup icon to open the Modbus Setup screen.
Modbus Address
Press the Modbus Address window to open the Enter Modbus Address (0-247) screen. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
TCP Port
Press the TCP Port window to open the Enter Modbus TCP Port screen. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
TCP Link Timeout
The TCP Link Timeout is useful when a connection to a remote computer is absent for longer than the expected time (perhaps due to a router connection drop). This could allow the connection to be automatically reestablished.
Press the TCP Link Timeout window to open the Enter Modbus TCP Link Timeout screen. A numeric keypad will be displayed for entering the number of timeout seconds. To select the value to change, use the < or > to move cursor. When complete press OK button.
NOTE: Setting the TCP Link Timeout to 0 disables the timeout function.
Mode
Select ASCII, RTU, or TCP.
Baud
Select 9600, 19200, 38400, 57,600, or115200.
Parity
Select Parity, Odd, Even, or None.
40
Page 41
Register Set
Select Register Set.
BACnet via Modbus Gateway
If using the Modbus to BACnet gateway, set up Modbus and then proceed to the BACnet Gateway Operating Instructions to continue the setup of BACnet connectivity. Once installed, the Modbus to BACnet gateway is a Modbus register.
RS-232 or RS-485
Select RS-232 or RS-485 radio button.
NOTE: The Baud, Parity, and RS-232 or RS-485 are not used if TCP is selected
Apply Settings
Press the Apply Settings icon to apply Modbus Setup settings.
Back Arrow Icon
Press the Back Arrow icon to return to the previous screen.
MODBUS Register Map
The MODBUS register map can be found in the Appendix at the back of this manual.
Internet of Things (IOT)
The instrument can communicate to remote servers over a network or the Internet. The setup of this JSON protocol for IOT can be found in the IMS Manual.
!
41
Page 42
6-8 Environment
Units
Select Barometric Pressure (BP) Units in Inches of Mercury (inHg) or Millibar (mbar).
Select Temperature (Temp) Units in Fahrenheit (°F) or Celsius (°C).
Show on Home
Select Show on Home Button to display the selected value on the Main screen.
Enable Alarms
Select the Enable Alarm button to Enable Alarms.
Low Alarm
Press the Low Alarm window to open the Enter Low Threshold screen. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
High Alarm
Press the High Alarm window to open the Enter High Threshold screen. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
Press the Environment icon to open the Environment screen to set Barometric Pressure (BP) Units, Temperature (Temp) Units, Enable Low and High Alarm, Setting Low and High Alarm and Show on Home.
Environment Setting Screen
6-9 Passwords
Press the Password Setup icon to open the Passwords screen.
Password Setup Screen
42
Page 43
User Password
Select the User Password button to enable User Password, preventing any unauthorized use of the instrument. The User Password allows the use of the instrument in its current configuration and saving samples.
New Password Confirm New Password
Press the New Password window to open the Enter New User Password (4-15 Chars) screen. Press the Confirm New Password window to open the Confirm New User Password (4-15 Chars) screen. Press Confirm Change Button to save changes.
NOTE: The default User Password is 1234.
Admin Password
Select the Admin Password button to enable Admin Password, preventing access to Settings screens, Time/Date, and configurations by Users.
New Password Confirm New Password
Press the New Password window to open the Enter New Admin Password (4-15 Chars) screen. Press the Confirm Admin Password window to open the Confirm New Admin Password (4-15 Chars) screen.
NOTE: The default Admin Password is 4321.
Password Timeout
Press the Password Timeout window to enter the number of minutes (1-99) the instrument will timeout due to inactivity. A numeric keypad will be displayed for entering these values. To select the value to change, use the < or > to move cursor. When complete press OK button.
NOTE: The default is five (5) minutes.
Lock Now
Press the Lock Now icon to immediately lock the instrument.
LOST PASSWORD: Please call or email Setra Systems for a temporary password that will be valid for the day the password is requested. NOTE: This unique password will automatically expire at the end of the day it is issued.
6-10 Screens
Press the Screens icon to open the Config. Screens window. Select Particles, Real­time meter, Graph, and/or PM environmentals to activate the screens shown when repeatedly pressing the Screens icon on the Main page.
43
Page 44
7-1 Power Management
Sleep Between Samples
Selecting the Sleep Between Samples button activates the feature for powering down the instrument between samples to conserve battery life. Press the Sleep Between Samples Wait window to open the Enter Wait Time screen. A numeric keypad will be displayed for entering the Wait Time in seconds from 0 to 65,535. To select the value to change, use the < or > to move cursor. When complete press OK button. NOTE: A Hold Time has to be set for a minimum of 38 seconds in order to activate Sleep Between Samples.
Sleep When Idle
Selecting the Sleep When Idle button activates the feature for powering down the instrument when idle. Press the Sleep When Idle Wait window to open the Enter Wait Time screen. A numeric keypad will be displayed for entering the Wait Time in seconds from 10 to 65,535. To select the value to change, use the < or > to move cursor. When complete press OK button.
Dim Screen When Idle
Selecting the Dim Screen When Idle button activates the feature for dimming the instrument’s screen when idle. Press the Dim Screen When Idle Wait window to open the Enter Wait Time screen. A numeric keypad will be displayed for entering the Wait Time in seconds. To select the value to change, use the < or > to move cursor. When complete press OK button. NOTE: The minimum Dim Screen When Idle Wait Time is 10 seconds.
Brightness Slider
Press and slide the Brightness Slider to increase and decrease screen brightness.
Shutdown Icon
Press the Shutdown Icon to immediately Shut Down the instrument.
Back Arrow Icon
Press the Back Arrow icon to return to the previous screen.
Press the AC Adapter icon to open the Power Management screen.
Power Management Screen
44
Page 45
8-1 Volume Controls
Software Installation
Plug the provided thumb drive into your Windows PC and select the IMS Install web installer application file. Follow the installation prompts.
Start Software
Use the IMS Icon in the program folder or on the desktop to start the program.
Connection Indication
Connect the instrument to a PC with the provided USB cable to establish a connection between IMS and the instrument. When a connection has been established the Green USB Connection indication will appear.
Volume Control Icon
Press the Volume Control icon to enable the Volume Control Slider. Slide the Volume Control Slider Bar to adjust the volume. A numeric value appears at the bottom of the slider to indicate relative volume. Moving the Volume Control Slider Bar to 0 mutes the instrument.
9-1 Instrument Management Software (IMS)
The instrument includes a PC based software utility for real-time graphing, downloading data from the instrument, field calibration of sensors, firmware updates, remote diagnostics direct to a service technician, and more. NOTE: Detailed instructions for the use of the IMS software are contained in the IMS Software Manual on the included USB thumb drive.
45
Page 46
10-1 Remote Operation
The instrument may be accessed with a web browser via Ethernet (wired or Wi-Fi), using the instruments IP address found in the Communications screen. The following functions can be performed remotely: Start and Stop sampling, change Data Units, select Locations and turn on/off displayed channel values.
The Instrument must be placed in Ethernet or optional Wi-Fi mode and the unit must also be connected to a working router on the same network as the PC.
In the Communications screen, select the Use DHCP button and the unit will automatically display the router issued settings. Enter the displayed IP address into any device browser to access the remote screen.
NOTE: The URL address is the IP address that the DHCP Router assigned to the instrument. A static IP address can be provided for the device through the Communications screen by deselecting the Use DHCP button and entering the IP address.
Appendix – A Modbus Register Map
The MODBUS Register Map can be found on the USB thumb drive.
Appendix – B Environmental Sensor Data Logging
Environmental sensor data can be logged without running the pump or laser by deselecting all particle channels on the Channel Management screen. An All Channels Are Disabled screen will appear when the Back Arrow button is pressed. Press Yes to confirm and No to cancel.
46
Page 47
Setra Systems, Inc.
159 Swanson Road, Boxborough, MA 01719
SSP-SPCMN-24004P Rev. A 7/2020
800.257.3872 • www.setra.com
Loading...