Model MRC
Multi-Range Critical
Pressure Transducer
Designed for Critical Environments, The Setra Model MRC is
Setra’s newest dierential pressure transducer. This is the rst
multi-range transducer designed for stringent requirements
of dicult applications. The MRC oers class leading ±0.5%
FS accuracy in selectable ranges down to 0.1”W.C., which is
required for critical applications. Setra’s MRC comes with 3
dierent housing congurations; duct probe, DIN Rail mount
and a universal model to cover any installation changes on
the job site.
0.5% FS Accuracy
The Setra MRC is the rst multi-range transducer designed
for use in Critical Environments. No other multi-range transducer product oers better than 1% FS accuracy; a requirement in most critical applications. The Setra MRC is available
down to 0.1”W.C. with 0.5% FS accuracy.
n Ideal For Critical Environments
n 0.5% FS Accuracy
Field Selectable Universal Design
The Setra MRC has 4 selectable ranges and 3 selectable outputs, giving the exibility to make changes on the job site.
The MRC is eld congurable for range, mounting (DIN Rail,
wall mount and duct mount), output (mA or voltage) and engineering units (W.C. or Pascals). This exibility means that
the contractor can use the MRC for all of their critical needs.
IP67 Rated Housing
The MRC housing is a robust IP67 rated design and is sealed
with a gasket to make it wash-down capable for dicult applications. The MRC also has a conduit tting, making installation and wiring easier.
Capacitive Sensing Technology
Only Setra can claim ownership to the stainless steel capacitive design used in all of our HVAC/R sensors. Our advanced
capacitive element provides excellent stability and linearity,
while standing above the competitors in our ability to measure low pressure (<0.0001”W.C.) at high accuracy. Our technology has been used in over 8 million installations and has
the highest eld acceptance rate in the industry.
n Universal Design
Model MRC Features:
• Field Congurable Duct Probe
• 4 Digit LCD
• Field Selectable Range
• Field Selectable Output
• Simple 5-Step Setup
• Field Accessible Push-Button Zero & Span
• External Mounting Tabs & Optional DIN Rail
• Unregulated AC/DC Operation
• IP67 Rated Housing
Target Uses:
• Service/Retrot Friendly
• Sub-Contractors- Quick Installation
• Flexible for Building Specication Changes
• Service Technicians- Quick & Accurate
Reconguration
• Hospitals
• Isolation Rooms
• Vivariums
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
Model MRC
Multi-Range Critical Pressure Transducer
ORDERING INFORMATION
M R
C
Model Field Selectable Ranges Conguration Options
MRC Unidirectional
Pressure Ranges
0.1”W.C. 25 Pa ±0.1”W.C ±25 Pa U Universal
Bidirectional
Pressure Ranges
S Standard (Base
Mount)
1
C 1/2” Conduit w/ Cal Certication
D PG9 w/ Cal Certication
2
2
0.25”W.C. 50 Pa ±0.25”W.C. ±50 Pa D DIN Rail
P Duct Probe
1
Universal unit includes Duct Probe and DIN Rail options.
2
Calibration certicate is standard and is provided for highest range ±0.25”W.C. on ±50 Pa.
DIMENSIONS GENERAL SPECIFICATIONS
Performance Data Environmental Data
Standard
1
(at constant temp) ±0.5% FS
Accuracy RSS
Compensated Range ºF (ºC) 32 to 122°F (0 to 50°C)
Thermal Eects2 %FS/ºF(ºC) 0.03 (0.054)
Maximum Line Pressure 10 PSI
Overpressure
Long Term Stability (max.) 1.0% FS/YR
1 PSI Field Selectable Output
Position Eect
Zero Oset %FS/G 0.5% Loop Resistance
(Unit is factory calibrated at 0g eect in the vertical position) Approval CE & RoHS Compliant
Operating Temperature
Electrical Data
Excitation Range 13 to 30 VDC/18 to 24 VAC
Current Consumption 30mA (max)
Mis-Wiring Reverse Excitation Protection
Output Resistance
(Voltage Output)
Load Resistance
(Voltage Output)
(4-20mA)
3
32 to 122°F (0 to 50°C)
(Voltage Output) 13 to 30 VDC (4
to 20mA output at terminals)
4
0 to 5 V, 0 to 10V (3-wire), 4 to
20mA (2-wire)
10 Ohms (max)
10 K-Ohms (min)
0 to 800 Ohms
Physical Description Pressure Media
Case Fire-Retardant
Polycarbonate (UL 94 V-0
Approved), Hinged Lid
Mounting Two Screw Holes Vertical
Position
Electrical Connection Block Removable Screw
Terminal
Pressure Fitting 3/16” O.D. Barbed Brass
Zero Push Button
Span Push Button
Weight (approx.) 8 Ounces
Typically air or similar non-conducting gases.
1
RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2
Units calibrated at nominal 70˚F. Maximum thermal error
computed from this datum.
3
Operating temperature limits of the electronics only. Pres-
sure media temperatures may be considerably higher.
4
Calibrated into a 50K ohm load, operable into a 10K ohm
load or greater.
5
Span (Full Scale) output factory set to within 1%.
6
Calibrated at factory with a 24 VDC loop supply voltage and
a 250 ohm load.
7
Span (Full Scale) output factory set to within ±0.16mA.
Specications subject to change without notice.
138.5
5.45
80.4
3.17
94.1
3.71
49.7
1.96
31.8
1.25
191.4
7.5
109.1
4.30
1/2" CONDUIT
OR
PG9
STATIC PRESSURE
PROBE
DIN RAIL
BACKPLATE
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com © Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.