Accurate measuring & display
Setra Lite provides a simple, cost-effective, and accurate way to measure
and display room differential pressure in areas today that require manual
verication. For local staff managing an area, Setra Lite has a light ring
that is green or red, showing a clear indication of room conditions. For
the facility manager, using Setra Lite means the frequency of preventative
maintenance walk-by inspections can be reduced because Setra Lite is
monitoring the space.
Setra Lite features
Setra Lite has many features found only in larger, more complex room
pressure monitors. These include a bright light ring that can be seen at
an angle or down a hallway, a digital display of the pressure value that is
always-on or momentary, an alarm delay for door entry, and POS or NEG
room mode. The Setra Lite sensor has Setra’s accuracy and reliability you
know you can trust.
™
Visual Pressure Indicator
Features
• Simple single-room pressure indication for all
pressurized healthcare spaces
• Saves preventive maintenance from manually
checking each room
• Sleek and attractive single wall plate
• Simple to install in minutes
• Visual alarm via red/green light ring
• Silent operation, no audible alarm
• Offset light pipe that can be seen at an angle or
down a hallway
• Digital display can be always-on or momentary
• Full conguration is easy, using just 3 buttons
• Analog output provides pressure reading to BMS
system
• Works with Setra FLEX to serve rooms with multiple
doors
Simple installation
Setra Lite is about the size of a standard light switch, and just as a simple
to install. It ts in a single-gang US electrical box, international 86-box or
international round back-box, and is powered by 24VAC or 24VDC. The
reference side pressure pickup is integral with the Lite faceplate. An analog
output provides the building management system with the pressure signal
for any additional trending or alarming needed.
The Setra Lite also integrates seamlessly with the Setra FLEX
environmental monitor. Mount Setra Lite at additional doors that enter the
space, and use an on-screen badge to display its value.
Applications
• Sterile processing
• Sterile storage
• USP 800
• Laundry
• Endoscopy
• Bronchoscopy
The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
© Setra Systems, Inc. All rights reserved.
Pressure indication Simple green status Red indicates alarm
800.257.3872 | www.setra.com
Setra Lite™ | Visual Pressure Indicator
Specications
Physical description
Dimensions
Mounting
Case Fire-retardant plastic UL94 V-0
Weight 8 oz (0.5 lbs)
Display
Alarm indication
Single-gang US electrical box, China 86 box,
European back box, or equivalent interior
Resolution: 128x160, backlight: LED
Red illuminated LED light ring , 11mm standoff
from faceplate, viewable at 90° angle from side
3.42” W x 4.88”H x 1.83”D
(67 mm x 124 mm x 46.5 mm)
dimensions
Color: RGB, 65K
Performance
1
Accuracy
Non-linearity (BFSL) ±0.44% FS
Hysteresis ±0.10% FS
Non-repeatability 0.05% FS
Span Tolerance ±1% Rdg
Thermal effects
Zero shift %FS/°F (°C) ±0.033 (±0.06)
Span shift %FS/°F (°C) ±0.033 (±0.06)
Long-term stability ±0.5% FS per year
Overpressure ±1 PSI (15” WC for <0.10” WC FS)
Pressure media Air or non-conductive, non-explosive gases
Pressure ttings 3/16” barbed tting
Position
2
Can be mounted vertical or horizontal. Re-zero
±0.5% FS
after installation
Environmental data
Operating temp. °F (°C) 32 to 120 (0 to 50)
Storage temp. °F (°C) -20 to 160 (-30 to 70)
Operating humidity 5 to 95% RH (non-condensing)
Ingress protection (IP) rating IP54
Display
Light ring Green, red, or off
Numeric value 4 digits
Features
Re-zero Single press-and-hold to re-zero sensor
Alarm thresholds Set numeric threshold in “WC or Pa
Door delay Set door alarm delay for 0-999 seconds
POS or NEG indicator Set display to show POS, NEG, or off
Display on/off
Signicant digits 1,2, or 3 after decimal point, POS or NEG
Light ring
Units Set units on the display to show “WC or Pa
Set display to always-on, or momentary press to
show pressure value
Set LED ring to 1). steady green, steady red if
alarm,
2). Green OFF, and steady red if alarm
3). always of
Electrical
Power
Power draw 2W max, 1W typical
Analog output
Wire Stranded unshielded twisted-pair, 18-24 AWG
Terminals Fixed, non-removable
3
Set for 0-5V, 0-10V, or 4-20mA signal
24 VAC (+/-10%)
24VDC (18-32)
50-60 HZ
Regulatory compliance
CE, RoHS
1
RSS of non-linearity, non-repeatability, and hysteresis
2
Units calibrated at nominal 70°F maximum thermal error computed from this datum
3
Zero tolerance negated by zero push button
© Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. | www.setra.com | 800.257.3872