Setra AccuSense AXD Operating Manual

Model AXD
Operating Instructions
1.0 Introduction
Every sensor in the AXD product family has been tested and calibrated before shipment. Setra Systems AXD product family senses gauge, compound, vacuum gauge and converts this pressure difference to a proportional high level analog output. Three standard excitation and output versions are offered:
Excitation Output Output codes
9 to 28 VDC 4-20 mA reverse excitation protection 11 9 to 30 VDC e.g. 0.5 to 5.5 VDC - (Reverse excitation protection) 24, 2E, et.al
4.9 to 8.1 VDC Reverse excitation protection 0.5 to 4.5 VDC 45
2.0 Mechanical installation
2.1 Media compatibility
The transducers in the AXD product family are offered with two different types of wetted materials. The AXD1 is made with 17-4PH stainless steel and the AXDH is made with 316 stainless steel. Prior
to use in your application conrm that your media is compatible with the wetted materials of the
ordered sensor.
2.2 Environment
The operating temperature limits of the AXD are -40° to +257°F (-40 to +125°C) on most electrical connection options (except where noted). The compensated temperature range is -4 to +185°F (-20 to +85°C).
2.3 Pressure ttings
Typically, standard pipe ttings and procedures should be used. However, for pressure ranges in
excess of 500 psig, we suggest the use of a sealant such as Locite Hydraulic Sealant. Excessive
torquing of metal ttings may cause a slight zero shift. The use of plastic ttings typically results in no noticeable zero shift. Torquing does not appreciably affect linearity or sensitivity.
2.4 Venting
Because the reference pressure in a sealed gauge transducer will vary due to changes in temperature and will affect overall accuracy (especially in units of less than 200 psig range), all transducers in the AXD product family are available in both vented or sealed pressure types.
Vented units are ordered as gauge pressure type (e.g. PSIG) units. Sealed units are ordered as sealed gauge type (e.g. PSIS) range units. The AXD gauge transducers are vented through electrical termination as shown in table below.
Electrical termination Venting
Cable (XX) Venting through cable 3-pin packard (P1) Venting through porous lter protected by loose o-ring M12, 4-pin (M4) Venting through connector Terminal strip/conduit ("A1") Venting through porous lter on terminal strip
3.0 Electrical installation
The AXD product family is available with four electrical terminations:
• (2ft/0.5m, 6ft./2m, 12ft/4m, 25ft/8m) cable
• M12-4 PIN connector
• 3-pin packard connector
• Terminal connection w/ conduit adapter, 1/2 inch
3.1 Voltage and current output units
Red + Excitation
Black - Excitation
+
Monitor or
Control Device
_
Green + Out
White
- Out
Shield
MATERIAL
SCALE
ENGR
DATE
DATE
Boxborough, MA, USA
MATERIAL
1:1
SAINT HILAIRE
1-24-19
DRAWN BY
© SETRA SYSTEMS, INC. THIS DRAWING CONTAINS THE
PROPRIETARY INFORMATION OF SETRA SYSTEMS, INC., ALL
RIGHTS RESERVED. ALL UNAUTHORIZED REPRODUCTION,
USE AND TRANSMISSION OF THIS DRAWING AND
INFORMATION IT CONTAINS ARE STRICTLY PROHIBITED.
THIS DRAWING IS SUBJECT TO RETURN TO SETRA SYSTEMS, INC.,
UPON ITS REQUEST.
The Model AXD (voltage output) transducer is a 3-wire voltage device. It is operable in a 5000Ω load or greater. The wires for the individual conductors is as follows:
Current output units
The Model AXD (current output) transducer is a true 2-wire, 4-20 mA current output device and delivers rated current into any external load of 0-800 ohms. The 4-20 mA units are designed to have current flow in one direction only - Please observe polarity. We suggest that the electrical cable shield be connected to the system’s loop circuit ground to improve electrical noise rejection. The electrical connection is as follows:
3.2 Output wiring schematics
Voltage
Model AXD Setra transducer (Voltage Output)
CAUTION: Unit is reverse excitation protected. However, do not apply power to output lead as this could cause permanent damage.
+Out
-Out
Current output
Model AXD
Setra transducer
(4-20 mA)
+
+ Power supply
_ _
The AXD has a 2-wire cable, where red is positive and black is negative.
CAUTION: Reverse excitation will not cause damage to the unit unless voltage applied is above 50 VDC. However, the unit will not function if reverse wired.
_
+
Load
(monitor)
Model AXD: In case of electrical noise pickup we suggest connecting a 22 μF non-polar electrolytic
capacitor rated 50V between (+Exc and -Exc) terminals of power supply to improve noise rejection.
3.3 Electrical termination wiring instructions
Cable version ("XX")
Red +Exc +Exc/supply
Green +Out N/C
Black -Exc/common -Exc/return White N/C /no connect N/C
Shield Earth ground/system ground Earth ground/system ground
3-pin packard connector ("P1")
Type: P2S series 150
Voltage Current
B
A
Vent ring
(Hydrophobic lter)
C
Voltage Current
A –Exc/common –Exc/return B +Exc +Exc/supply C +Out N/C
2
Loading...
+ 2 hidden pages