Setra 595 Operating Manual

Adding a Site Offset to SDI12 Output
In some applications it is necessary to measure a site value including some reference data point, for example sea level. Assuming the sensor is installed 10m above sea level, then it is possible to have all reported readings add 10m, thus acquiring the data without the need for further modification. To add a site offset then:
1. Determine the offset to be applied.
2. Add the offset to the pressure units (e.g. aXU1,10!)
3. Save new calibration to nonvolatile memory (e.g. aXS0!)
4. Verify sensor now measures 10m for 0 applied pressure.
10. WARRANTY
The company warrants its products to be free from defects in material and workmanship in normal use and service for a period of two years from date of shipment. The Company reserves the right and option to repair or replaceupon evaluation of the returned original part. Modification, misuse, attempted repair by others, improper installation or operation shall render this guarantee null and void. The Company makes no warranty of merchantability or fitness for a part or purpose.
11. Servicing
The transducer cannot be repaired locally and if damaged should be returned to ourselves at the address shown below when a replacement/repair is required.
Setra Systems, Inc. 159 Swanson Road Boxborough, MA 01719 Attn: Repair Department
Model 595 Slimline
Submersible Pressure Transducer
12. Return to Factory
PLEASE NOTE: All returns must be accompanied with Setra’s “Calibration and Service Order” return form. The form can be downloaded at:
http://www.setra.com/tra/repairs/pdf/webrepair.pdf
Any instrument returned without Setra’s authorized form completely filled out will be quarantined and no action will occur. It may ultimately be returned to you and subject to a transportation charge.
159 Swanson Road, Boxborough, MA 01719/
Toll Free: 800-257-3872/ Phone: 978-263-1400/
Fax: 978-264-0292/Email: sales@setra.com/Web: www.setra.com
ISO
9001: 2000
Certified
Setra Systems, Inc.
Contents
1.0 Introduction ........................................................................................................................... 3
2.0 Safety Instructions ................................................................................................................. 3
3.0 Hazardous Products ............................................................................................................... 3
4.0 GENERAL ................................................................................................................................ 3
5.0 Action on Receipt ................................................................................................................... 4
To re-range the 595 sensor proceed as follows:
1. Determine the 595 sensor full scale and the user application full scale.
2. Find the required calibration gain. Divide the 595 sensor full scale by the user application full scale (e.g. 10/4)
3. Program the calibration gain into the sensor (via SDI12 interface). (e.g. aXG2,2.5!)
4. Check the residual error with zero pressure applied. Note as the gain is in­creased any offset errors ban be magnified. So for example instead of 4mA expected sensor reads 3.959mA.
5. Program offset in calibration of sensor to remove offset errors (e.g. aXA2,0.041!)
6. Save new calibration to nonvolatile memory (e.g. aXS0!) 7 Verify sensor now measures 4 to 20mA for required application range.
6.0 Installation ............................................................................................................................ 4
6.1 Lowering the Pressure Probe ......................................................................................... 4
6.2 Electrical Installation .....................................................................................................5
7.0 Operation ............................................................................................................................... 6
7.1 Analogue Output Probes ............................................................................................... 6
7.2 SDI12 Operation ............................................................................................................ 6
8. Maintenance ............................................................................................................................ 9
8.1 Self diagnostics ............................................................................................................. 9
8.2 Calibration Adjustment ............................................................................................... 10
8.3 Cleaning ...................................................................................................................... 14
9. Advanced Operation ............................................................................................................... 14
10. WARRANTY ......................................................................................................................... 16
11. Servicing ............................................................................................................................ 16
12. Return to Factory ................................................................................................................ 16
Configure a Reverse Output 20mA to 4mA. If your application demands a 20mA output with zero applied pressure and 4mA output for full scale applied pressure then the sensor output can be reversed by sending the following SDI12 commands:
1. Set calibration gain of mA output to -1 (e.g. aXG2,-1!)
2. Set calibration offset of mA output to +24 (e.g. aXZ2,24!)
3. Save new calibration to nonvolatile memory (e.g. aXS0!)
4. Verify sensor now measures 20mA to 4mA for 0 to full scale applied pressure.
Applying a Filter to the mA Output In applications that are inherently difficult to measure, for example waves on reservoir, then the 595 can be configured to apply a mean or median filter via the SDI12 interface before updating the analogue mA output. The filter command is described below:
SDI12 Command: 0XFt,nn,xxxx! Neptune Response: 0t,nn,xxxx<CR><LF>
Where:
T = type of filter 0 = MEAN 1 = MEDIAN nn = number of values to filter, maximum 12
xxxx = 220ms intervals between measurements, maximum period 1 hour To apply a filter:
1. Determine the type of filter and period suitable for the application.
2. Send the filter command to the 595 sensor (e.g. 12 reading mean, sampling period 15 minutes, aXFt,12,4091A)
3. Save the filter to nonvolatile memory (e.g. aXS0!)
4. Verify the filter performance in applications.
8.3 Cleaning
Inspection and cleaning of the product is recommended as part of a preventative maintenance procedure. The required inspection period will depend on the nature of the application environment. To inspect the product unscrew the nose cover. Figure 2, and view the sensor diaphragm. The housing should be free from deposits. Cleaning with a cotton bud (Q-TIP) is possible, but extreme
care must be taken not to deform the sensor diaphragm during cleaning.
Figure 2
Inspecting and Cleaning the Sensor
9. Advanced Operation
The 595 allows for a number of advanced operations not normally available.
Re-Range 4 to 20mA Output Often users wish to measure a specific water level that is not the same as the full range output of the transducer. For example, if the sensor output was 10MH2O and the user wishes to measure up to 4mH2O then without a down range feature the output for the 10mH2O unit in the users applications would be 0 to 40% of the range or 4 to 10.4mA. In this situation it is possible to use the SDI12 interface to program calibration coefficients into the sensor to provide 4 to 20 mA for the 0 to 4mH2O application, with the 10mH2O sensor.

1.0 Introduction

Model 595 is a level measuring transducer for precise measurements of ground and surface water levels. The transducer measures pressure and temperature signals and, by compensating these measured values for temperature effects, the relative density of water and the specific gravity in the users location, provide a highly accurate and repeatable “true” level measurement. The resultant level is available either as an analogue output or as a digital output

2.0 Safety Instructions

Important Note: Setra’s Model 595 pressure transducer is designed and manufac­tured in accordance with Sound Engineering Practice as defined by the Pressure Equipment Directive 97/23/EC. Pressure transducer products designed to meet the highest risk category “IV” of the Pressure Equipment directive are clearly marked on the label by “CE0086”. Compliance is achieved through modules “B&D”. No other products should be used as “Safety Accessories” as defined by the PED, Article 1, Paragraph 2.1.3.
EMC: Model 595 conforms with the essential protection requirements of the EMC Directive 89/336/EEC amended by certified type testing to EN 61000-6-3 and EN61000-6-2. Conformity with the requirements of the CE mark only applies when the installation conditions described in these instructions have been met.

3.0 Hazardous Products

The 595 may be classified as Electrical, Electro-Mechanical and Electronic equip­ment.
These products are tested and supplied in accordance with our published specifications or individual special requirements that are agreed in writing at time of order. They are constructed so as not to affect adversely the safety of persons and property when properly installed, maintained and used by qualified person­nel, in the applications for which they were designed and manufactured. Refer to installation and maintenance instructions for further details.

4.0 GENERAL

Transducer should not be subjected to greater than maximum allowable pressure range as defined on the transducer label.
Transducer should not be subjected to mechanical impact.
The effects of decomposition of unstable fluids should be considered by the
user when placing this device in service.
Pressure range must be compatible with the maximum level being mea­sured.
Pressure media must be compatible with the transducer wetted parts which are:
UNS31803 Stainless Steel, BS En 10088-3:1995 No. 1.4462. Polyurethane Acetal
Nitrile
Loading...
+ 5 hidden pages