Setra 270 Operating Manual

Operating Instructions
Model 270 Pressure Transducer
General Information
Your Setra transducer has been carefully calibrated before shipment to you, and it should be handled with the same care given any precision instrument. Pressure range and dimensions are reported on the specifications bulletin for the transducer.
Ambient Conditions
Do not use in ambient conditions corrosive to anodized aluminum, or submerge in liquids, subject to spray or vibration environment.
Standard
“Belden” Cable Lead Optional Function (#8723 Grey Cable) “Tensolite” Cable Lead Positive Excitation Red White Positive Output Green Yellow Negative Output White Brown Negative Excitation Black Black Case Shield Shield
Electrical
The electrical circuit is equivalent to a 4-terminal network which can be grounded at only one point, either at the negative excitation or the negative signal output lead, but must not be commoned or grounded at more than one point.
The pressure transducer must be operated with the case connected either to the negative excitation terminal or to the negative output terminal. Failure to do this may result in damage to, or unsatisfactory operation of the unit. This connection may be made by connecting shield and black (negative excitation) leads or alternatively by connecting the shield and negative output signal leads together. Best shielding against noise will be obtained by connecting the shield and negative excitation leads. Circuit is reversed voltage protected for at least 5 minutes. Internal transient suppression network is provided for short duration transients to 150 volts.
In some instances, use of long cables (several hundred feet long) , may introduce enough cable capaci­tance into the output circuit to cause output oscillation. If encountered, this oscillation may be eliminated by connecting a 100 ohm resistor (1/8 watt or larger) in series in each of the output leads at the end of the 2 foot transducer cable. These series resistors of course add to the output resistance.
Adjustments (with cover removed)
CAUTIONS: 1. Before making any adjustments, determine which circuit board diagram on
Page 2 matches the circuit board in your transducer.
2. Turn off excitation power during cover removal or replacement.
3. Touching any adjustments other than zero or span may necessitate recalibration and voids the warranty.
ZERO OUTPUT
Output can be adjusted to zero by potentiometer as shown in diagram. Unit factory adjusted to
0.00 VDC output (±5mV).
SPAN
Can be adjusted by potentiometer as shown in diagram. Unit factory adjusted to 5.00 VDC full scale output (±5mV).
LINEARITY - DO NOT TOUCH
Factory adjusted for best linearity.
Circuit Board 1 Circuit Board 2
NOTE: When circuit cover is removed, a zero shift can be noticed. The zero will return by the same amount when the cover is replaced. This displacement of the output should be taken into account when making adjustments or during recalibration
*Adjustments: Due to the high accuracy of this transducer, adjustments to zero and span should only be made with the use of primary pressure standards of known accuracy higher than that of the Model 270.
Loading...
+ 3 hidden pages